Home
        EndoStim User Manual
         Contents
1.   systems  and implantable drug infusion pumps  Physical  proximity may cause sensing problems and inappropriate device  responses  Clinicians involved with both devices should evaluate  any potential interference problems before surgery     Transcutaneous Electrical Nerve Stimulator    Transcutaneous Electrical Nerve Stimulator  TENS  therapy is  generally contraindicated for patients who have active  implantable medical devices  The high voltage pulses delivered  by TENS units to the body can interfere with the operation of the  IPG     If a TENS device must be used  place the TENS electrodes as far  from the IPG and lead as possible  The TENS electrodes should  also be placed as close as possible to each other to reduce current  spread  The operation of the IPG should be monitored closely  during TENS use  As a precaution  programming the IPG to the  OFF mode reduces the possibility of adverse effects     Contraindication  Patients and physicians should be advised  that exposure to TENS in the abdominal region is  contraindicated for patients implanted with the EndoStim  System     Home Appliances    Home and commercial microwave ovens in good condition  and  used as intended  will not affect the IPG  Even a defective oven  that exposes the IPG to direct microwave energy may not  damage the unit itself  Ovens using electromagnetic induction  can cause the device to go into magnet mode  disable stimulation  therapy output      10 EndoStim Lower Esophageal Sphincter Stimulat
2.  System during the MRI examination     Notes     High specific absorption rate  SAR  RF power levels can  increase the risk of heating     Excessive heating of the lead may occur even if the IPG 1s  explanted     Magnetic Field Interactions    MRI uses a powerful static magnetic field which may exert force  and torque on the metallic objects that could result in movement  or displacement  Tests performed at 1 5 Tesla and 3 Tesla  indicated that this is not a concern for the EndoStim System     Warnings and Precautions  EndoStim Programmer System    The EndoStim Programmer contains ferromagnetic materials   These materials could make the Programmer susceptible to the  powerful static magnetic field of the MR system  Therefore  the  EndoStim Programmer System is considered MR Unsafe   meaning that it cannot be present and or used in the same room  as the MR system  The pre  and post exposure implanted pulse  generator programming requirements  detailed below  must be  performed outside of the MR system room     Other Implanted Systems    Prior to MRI examination  the patient should be evaluated for  other implanted medical devices  especially electronically   activated devices  An x ray may be used to determine if the  patient has other implanted medical devices     22 EndoStim Lower Esophageal Sphincter Stimulation System    The decision to perform an MRI examination must be made by  the MRI trained radiologist  and the patient   s physician  in  consideration of the risks and
3.  The EndoStim System may fail to properly operate for any  number of reasons  including but not limited to  random  component failure  including the battery   lead failure  including  electrical shorts  opens  and insulation faults   and software  errors  The frequency of these events cannot be predicted     12 EndoStim Lower Esophageal Sphincter Stimulation System    System Overview    This section provides descriptions of all the components for the  EndoStim   Lower Esophageal Sphincter Stimulation System   Throughout the instructions in this manual  you may see the  terms stimulator and IPG  implantable pulse generator   These  terms refer to the same type of device     The system consists of an IPG  bipolar stimulating lead  and a  Programmer        ndoStym    LES PROGRAMMER     A  NG       Figure 1 EndoStim LES System  Implantable Pulse Generator    General Description    The EndoStim II implantable pulse generator  IPG   Figure 2   is  an internally powered  lithium battery  device that delivers  electrical stimulation pulses to the lower esophageal sphincter   LES   The IPG is hermetically sealed in a titanium case  It  delivers electrical pulses to the LES via a bipolar IS 1 BI header  connector port  Figure 3  that connects the IPG and lead  The  IPG is sterilized with ethylene oxide     EndoStim Lower Esophageal Sphincter Stimulation System 13    EndoSttm II    LES Stimulator    qr  IS 1 BI    Figure 3 IPG Header    Figure 2 EndoStim II LES Comedor    Stimulator    
4.  W  according to the transmitter manufacturer  and d  is the recommended separation distance in meters  m      e Field strengths from fixed radio frequency  RF   transmitter  as determined by an electromagnetic site  survey     should be less than the compliance level in each  frequency range        e Interference may occur in the vicinity of equipment  marked with the ionizing radiation symbol     A    At 80 MHz and 800 MHz  the higher frequency range applies     These guidelines may not apply in all situations  Electromagnetic  propagation is affected by absorption and reflection from  structures  objects and people        Field strengths from fixed transmitters  such as base stations for  radio  cellular cordless  telephones and land mobile radios   amateur radio  AM and FM radio broadcast and TV broadcast  cannot be predicted theoretically with accuracy  To assess the  electromagnetic environment due to fixed RF transmitters  an  electromagnetic site survey should be considered  If the  measured field strength in the location in which the EndoStim  Programmer is used exceeds the applicable RF compliance level  above  the EndoStim Programmer should be observed to verify  normal operation  If abnormal performance is observed   additional measures may be necessary  such as relocating the  EndoStim Programmer     Over the frequency range 150 kHz to 80 MHz  field strengths  should be less than 3 V m     34 EndoStim Lower Esophageal Sphincter Stimulation System    Recommended S
5.  Warranty           ccccccscscscscscscscscscscscecscececceccceees 37    EndoStim Lower Esophageal Sphincter Stimulation System V    List of Acronyms and Abbreviations    Radio Interference    Magnetic Resonance Imaging  Radio Frequency  T2DM Type 2 Diabetes Mellitus    TENS Transcutaneous Electrical Nerve  Stimulation       vi EndoStim Lower Esophageal Sphincter Stimulation System    Device Description  Contents of all Packages    Model 1011 EndoStim LES Stimulator and Lead  Package    e 1 Model 1006 EndoStim II Implantable Pulse Generator    e   Model 1003 EndoStim Implantable Bipolar Stimulating  Lead    e   Torque wrench  Allen type   e 1 Set peel off labels for IPG  e 1 Set of peel off labels for Lead  Model 1006 EndoStim LES Stimulator Package  e   Model 1006 EndoStim II Implantable Pulse Generator  e   Torque wrench  Allen type   e 1 Set peel off labels for IPG  Model 1012 EndoStim LES Programmer System  e   Model 1504 EndoStim LES Programmer USB Wand    e   Model 1522 Tablet PC with EndoStim LES Programmer  Software    Accessories    e Standard pulse generator and lead accessories     magnet   lead cap  lead extenders  torque wrench    EndoStim Lower Esophageal Sphincter Stimulation System 1    Indications and Contraindications    Indications    The EndoStim LES Stimulation System is intended for the  treatment of patients with chronic Gastroesophageal Reflux  Disease  GERD  with symptom duration of 6 months or  longer     Contraindication    The EndoStim Lower Esophag
6.  benefits to the patient     MR System Type    1 5 Tesla 64 MHz and 3 Tesla 128 MHz MR systems are the  only types of equipment permitted for use for a patient with the  EndoStim System  There is no restriction on the MRI  manufacturer     Other MR systems  e g   0 6 Tesla  etc   are not permitted for  use and may cause serious patient injury and or damage to the  EndoStim System     MR System RF Coil   Use of Transmit Receive RF  Local Coil  ONLY    A transmit receive RF local coil is the only type of RF coil  permitted for use in a patient with the EndoStim System     The transmit receive RF body coil must not be used     No portion of the RF transmit receive local coil can cover any  part of the implantable system  The diagram below illustrates  the region that is considered unsafe for MRI exposure       J LEAD       PULSE    EXCLUSION ZONE    gt  j  L IA MWA GENERATOR    Dm um     L AIN os 1    EndoStim Lower Esophageal Sphincter Stimulation System 23    Figure 7 Unsafe MR Imaging Region  highlighted in grey     This exclusion zone is dependent on the typical placement of the  EndoStim II IPG and Lead  Implanted system components may  be located as close as Ocm to the lower  caudad  edge of the  transmit receive RF local coil  but no part may be inside of it     Brain head imaging should be performed using a  transmit receive RF head coil only  The diagram below  illustrates the required placement of the transmit receive RF head  coil  as shown in grey  Under no circumsta
7.  cobalt nickel  conductors  and platinum iridium stitch electrodes  A curved  stainless steel suture needle is affixed to the end of each stitch  electrode  The lead is 45 cm in length from the IS 1 BI  connector tip to the end of the platinum iridium electrodes  The  lead bifurcates after 35 cm into individual  unipolar leads  with  the length of each unipolar lead being 10 cm     Handling and Storage    Do not implant the lead if the package is damaged or if the lead  has been dropped from a height of 30 cm or more  Return  damaged packages to EndoStim  B V     Store the lead at a temperature range of 20  C to 25  C and ina  dry location  Do not allow the packaging to become damp or wet  because sterilization of the contents may be compromised     Re Sterilization and Re Use    The lead is a single use device  Do not re sterilize the lead or  silicone stoppers  Do not re use an explanted lead  the device  must not be used for implantation in another patient     Programmer  General Description    Use the EndoStim LES System Programmer to interrogate and  program the IPG  The Programmer software runs on an  IEC60950 certified tablet personal computer  PC   which runs on  battery power  Communication between the Programmer and the  IPG is accomplished with the Programmer placed directly over  the implant site  The Programmer communicates via magnetic  induction telemetry with the IPG implanted in the patient     Note  Avoid skin contact between the patient   s skin and the  Pr
8.  have an  additional Programmer present at each implant and clinic visit     Implant the EndoStim System    Laparoscopic surgery is generally used for the implantation of  the EndoStim System  An open  bariatric surgical approach may  also be appropriate for some patients  Minimal use of narcotics is  recommended  Follow standard sterile technique and adhere to  standard operating room procedures     Warning  The implanting physician is required to undergo  formal training by an EndoStim expert representative prior to  implanting the system in humans     MRI Information for the EndoStim System    It is important to read the section in its entirety before  conducting any magnetic resonance imaging  MRI  examination  on a patient implanted with the EndoStim System     Contact an EndoStim representative with any questions  regarding the following MRI information     If all the instructions below are followed  MRI examinations of  the regions using a transmit receive RF local coil  e g   head coil   only may be safely performed at 1 5 Tesla 64 MHz or 3   Tesla 128 MHz  Failure to carefully follow the specific  information presented in this document may result in serious  patient injury and or damage to the EndoStim System     EndoStim Lower Esophageal Sphincter Stimulation System 21    Potential Interactions with the EndoStim System    MRI Related Heating    The MRI uses radiofrequency  RF  energy which may cause  excessive heating of the pulse generator and lead of the  EndoStim
9.  radiopaque marker placed within the IPG allows the model  number and year of manufacture to be identified by normal X   ray techniques  The manufacturer   s identification code for  EndoStim is E  the code for the EndoStim II IPG model is B  and  the code for the year of manufacture is XX  where XX is  replaced by the last two digits of the calendar year in which the  device was manufactured     EndoStim Lower Esophageal Sphincter Stimulation System 15    Implantable Bipolar Stimulating Lead    General Description    The EndoStim lead  called the Implantable Bipolar Stimulating  Lead  is used in conjunction with the IPG  The lead has an IS 1   BI connector    at the proximal end for attachment to the IPG   Figure 4   The lead is sterilized with ethylene oxide     The lead delivers stimulation pulses to the tissue through stitch  electrodes at the distal end  During implantation  the stitch  electrodes of the leads are sutured into the LES and secured into  place  The electrodes of the bipolar lead are implanted in the  sero muscular layer of the LES  Figure 5      Figure 4 EndoStim Lead and  IPG Figure 5 Implanted EndoStim    Lead and IPG       Implants for Surgery     Cardiac Pacemakers     Part 3  Low profile connectors   IS 1  for implantable pacemakers  ISO 5841 3 2000 E     16 EndoStim Lower Esophageal Sphincter Stimulation System    Physical Characteristics    The implantable lead is constructed of biocompatible materials   an inner and outer silicone rubber sheathing 
10.  than 3 times the battery voltage  When battery voltage  falls below 2 3V  the device will turn off all stimulation outputs   When the battery voltage falls below 2 1V  communication with  the Programmer will no longer be possible     The longevity of the battery can be estimated based on the  different stimulating parameters applied  as described in Table 3     28 EndoStim Lower Esophageal Sphincter Stimulation System    Table 3 Estimated IPG Battery Longevity    Parameter    Pulse  Width     Pulse  Frequency     Pulse  Amplitude    Hours per  Day    Estimated  Life       This is the recommended value    Condition 1    215 usec    20 Hz    5 0 mA    3 0    8 6 years    Condition 2 Condition3 Condition 4    215 usec 215 usec 215 usec    20 Hz 20 Hz 20Hz    5 0 mA 3 5 mA 7 5 mA    4 0 4 0    6 6 years 7 8 years 5 8 years    P This assumes a 12 month shelf life prior to implant    Lead Characteristics and Specifications    Table 4 Lead Physical Description and Materials    Connector  Length  Mass  Electrodes    Bifurcation    Lead    Specification    IS 1 Bl   45 cm   4g   Platinum iridium   At 35 cm  lead has two 10cm  unipolar segments    EndoStim Lower Esophageal Sphincter Stimulation System 29    Programmer Specifications    Table 5 Programmer Specifications    Description Value  Height 140 mm  Width 62 7 mm  Thickness 30 5 mm  Mass 251 9  Storage temperature 5  C to 37  C    Safety Shutdown with a Magnet    Therapy signal delivery of the IPG can be shut down for 24  hours 
11.  to these  electromagnetic fields  When appropriate  give specific  warnings     EndoStim Lower Esophageal Sphincter Stimulation System 11    Cellular Phones    Cellular and other portable telephones can interfere with the  operation of the IPG  Potential effects may result from either the  radio frequency emitted by these telephones or the magnet  within the phone   s speaker  These effects may include inhibition  or inappropriate triggering of electrical stimulation delivery  when the phone is in close proximity  within 25 cm  to the IPG  and the lead     Advise patients to hold the phone to the ear opposite the side of  the implanted IPG  Patients should not carry the phone in a  breast pocket or on a belt over or within 25 cm of the implanted  IPG because some phones emit signals when they are turned on  though not in use     Airport Screening Systems    Globally  passenger screening systems encountered in airports  may interfere with the IPG  Most commonly  interference may  result in electrical stimulation inhibition  Instruct patients to  inform security personnel about the implanted medical device  to  show their identification card  and to walk at a normal pace when  passing through the portal of these systems  No damage to the  implanted system should occur     Magnets    Application of a magnet directly over the IPG for at least 2  consecutive seconds  followed by 2 consecutive seconds without  a magnet  will terminate therapy for 24 hours     General Precaution   
12. EndoSt m    Lower Esophageal  sphincter Stimulation  system    Clinician Manual     With EndoStim II LES Stimulator Model 1006     C q 0344 1588    Year of Authorization  2013    EndoStim   is a registered trademark of EndoStim B V     Read all accompanying documentation before using  the device     reen    EndoStim B V   Prinsessegracht 20  2514 AP The Hague  The Netherlands    The information contained in this document may be changed  without notice     No part of this manual may be reproduced or transmitted in any  way or by any means  either electronic or mechanical  for any  purpose without the express written consent of EndoStim B V     Explanation of Symbols on Labels    DESCRIPTION    Manufacturer    Date of Manufacture    Affixed in compliance with  European directives   0344   Notified Body Number  for Directive 90 385 EEC    Affixed in compliance with  European directives   0344   Notified Body Number  for Directive 90 385 EEC  1588   Notified Body Number  for Directive 1999 5 EC    Consult instructions for use     Do Not Use if Package is  Damaged or Opened    Transport Temperature Limits    STERILE EO Sterilized with Ethylene Oxide     gt  lt  Use By    EndoStim Lower Esophageal Sphincter Stimulation System 1       SYMBOL DESCRIPTION       LOT  ane          BR cme     E Torque Wrench        Oo    Intentional Radiator    Signal Strength Indicator    USB Connector    MR Conditional       ul EndoStim Lower Esophageal Sphincter Stimulation System    SYMBOL DESCRIPTIO
13. N    D MR Unsafe       EndoStim Lower Esophageal Sphincter Stimulation System ili    TABLE OF CONTENTS    Explanation of Symbols on Labels               scsecscscecsecscscees    TABLE OF CONTENTS  ciiaicicasisieticasesasieatiiecsevienasaacteeteasess IV  List of Acronyms and Abbreviations              cccscsesscsceeeees vi  Device DESCIIDUION a isa ais ias sm an cota lis tacstesudcasiievecsees 1  Contents of all Packases aus isa cesso To seis nes rn aaa 1  Indications and Contraindications             ceccccscscscscscscces 2  ITIGU GATIONS ssa ssa teria ai ai bodas ossada ditas 2  Contraindication seas sines naso nos pisis aaa Rede ai Calda 2  PrecaAUTIONS  sacras aa losi eai ra Rio iiired nica donde miau bssastasdiindas 2  Potential Adverse Effects Events                                     3  Implantation of the System             scsscscscsccscscsccccccscececcecsces 3  Use of the System assada as adia SN oia 4  Warnings and Precautions           cccscscsscscscsccscscsceccecscscees 5  Cremation and INCINEratiON           cceccccccccccececscscecscscscscsceces 5  ROUS Css cvsccsnacasenenasensbaceavamsnestcendenecwasd eeawaaetasasses senses encetecwens 5  Environmental Hazards           csccscsssccscsceccccscscsccccccscsceccccsces 5  BIC CUMOCAULCIY coisirean asirio a 6  RFADIAUO Macnee aaa demos 7  Medical Diathermy            ccccsccscscscsccscscsccccscsceccccccscescecccsces 7  DERI FIA ON qe NPR RREO RR DO e NR 8  Therapeutic Radiation             cccccccscsssccscscsccccccsc
14. RF power should be limited to a specific absorption rate of  1 W kg or less  in the head and whole body  for each pulse  sequence     EndoStim Lower Esophageal Sphincter Stimulation System 25    The spatial gradient should be limited to 250 Gauss cm or less     The isocenter location landmark should be the region of interest  ONLY     Preparing the Patient prior to MRI Procedure    The patient should be informed to bring the Patient ID card to  the MRI procedure appointment and to inform the MR system  operator that he she has an active implantable medical device   prior to entering the MR system room     The patient should be informed that the therapy will be turned  off until after the MRI procedure is completed     The IPG should be programmed to OFF  reference section  Program Stimulation Amplitude and Polarity      Important Note  If the IPG is in End of Life  reference section  IPG Battery Characteristics and Specifications   then the IPG  mode cannot be programmed or confirmed via telemetry and the  MRI procedure cannot be performed     The patient should be informed of all risks associated with  undergoing an MRI procedure with a transmit receive RF local  coil only  reference the Warning and Precautions section in this  chapter      The patient should be informed to immediately inform the MR  system operator of any discomfort  stimulation  shocking  or  unusual sensation that may occur during the MRI procedure     The patient should be conscious throughout the MRI  e
15. The operators of the Programmer include physicians in charge of  either implanting or monitoring an IPG  and trained medical  personnel who assist physicians  Operators will be familiar with  operation of electronic medical equipment  particularly IPGs   and programmers and   or the operators will have been trained on  the operation of the EndoStim Programmer     The Programmer is used in an operating room where the IPG is  being implanted  When in the operating room  the Programmer  should be brought into the sterile field through a sterile sleeve  while the tablet PC remains outside the sterile field  The surgeon  should place the Programmer over the IPG while an appropriate  person outside the sterile field operates the Programmer     The Programmer is also used in a clinical room where patients  with an implanted IPG are monitored  When in the clinic  the  clinician should place the Programmer over the IPG and have  either an assistant or the patient hold the Programmer in place   The patient can sit or lie down  The tablet PC should be outside  the patient environment  at least 1 5 meters away  and operated  by the clinician     20 EndoStim Lower Esophageal Sphincter Stimulation System    To be compliant with electromagnetic compatibility  requirements  interference characteristics   the Programmer  should not be used adjacent to or stacked with other equipment   If adjacent or stacked use is necessary  verify that the  Programmer functions properly  It is recommended to
16. The programmable IPG communicates with the external  Programmer via telemetry  Medical and technical personnel use  the Programmer to program parameters that control the IPG  function  The IPG may be programmed specifically for  individual patients by changing the settings parameters within  the implanted device  In addition  the Programmer can access  performance data that the IPG collects during its normal  operation     14 EndoStim Lower Esophageal Sphincter Stimulation System    Handling and Storage    Do not implant the IPG if the package is damaged or if the IPG  has been dropped from a height of 30 cm or more  Return  damaged packages to EndoStim B V     Store the IPG at a temperature range of 20  C to 25  C and in a  dry location  Do not allow the sterile packaging to become damp  or wet because sterilization of the contents may be compromised   Exposure to temperatures below 0  C may cause a change in the  parameter values of the IPG to the DOWN mode  no output    Once the IPG is removed from the extreme environment  it will  either return to the programmed settings  or it will remain in the  DOWN mode  In the latter case  follow the Reset procedure  If  unsuccessful  the unit should be returned to EndoStim B V     Re sterilization and Re Use    The IPG and torque wrench are single use devices  Do not re   sterilize the IPG or torque wrench  Do not re use an explanted  IPG  the device must not be used for implantation in another  patient     X Ray Identification    A
17. Voltage Rsplicabi   to the public low voltage  Fluctuations Flicker PP supply network that supply  Emissions buildings used for domestic  IEC 61000 3 2 purposes     RF emissions  CISPR 11       EndoStim Lower Esophageal Sphincter Stimulation System 31    ESD and Power Fluctuations    Guidance and manufacturer   s declaration      electromagnetic immunity    The EndoStim Programmer is intended for use in the  electromagnetic environment specified in Table 7  The customer  or user of the EndoStim Programmer should assure that it is used  in such an environment     Table 7 ESD and Power Fluctuations  Electromagnetic  Immunity    Immunity   IEC 60601   Compliance   Electromagnetic  test Test level environment    guidance    Electrostatic  6 kV  6 kV contact   Floors should be wood   discharge contact concrete or ceramic tile    ESD   8 kV air If floors are covered with   8 kV air E synthetic material  the  IEC 61000 4 2 E relative humidity should  be at least 30    Electrical fast Not Not Mains power should only  transient burst Applicable Applicable be used to recharge the  tablet PC battery while    IEC 61000 4 4 the unit is powered off     Surge Not Not Mains power should only  Applicable   Applicable be used to recharge the  tablet PC battery while  PEO the unit is powered off   Voltage dips  Not Not Mains power should only  short Applicable Applicable be used to recharge the  interruptions tablet PC battery while  and voltage the unit is powered off   variations on  power su
18. ccscsceccccccscsceececs 20  Implant the EndoStim System             sccccscscsscscscececcecsces 21  MRI Information for the EndoStim System                       21  Potential Interactions with the EndoStim System               22  Warnings and PrecautiOns           ccccscscscscscscscscscecscececccceces 22  Preparing the Patient prior to MRI Procedure                     26  Beginning of the MRI Procedure             ccscscsccscscscessscscscees 26  During the MRI Procedure           cscscscsccscscessccscsceccccscscecees 27  After the MRI Procedure              ccscsecscsecsceccsceccsceccsceccececs 27  System Specifications            ceceecsceccccccscececcecscececcecscecess 28  IPG SDECITICATIONS asas sas  sisias sous riisist ienai aa 28  IPG Battery Characteristics and Specifications                    28  Lead Characteristics and Specifications             cscscsssscscseees 29  Programmer Specifications            ccsccccccscscecsccscsceccccscecscees 30  Safety Shutdown with a Magnet           ccccsccccscsscccscscsceceees 30  Electromagnetic Interference            ccsccscssessccscscsceccecscscees 30  Electromagnetic EMISSIONS            ccccscsccscscecsccscsceccccecscscees 31  ESD and Power Fluctuations              ccccscsccecsccccsceccsceccnceees 32  Radiated Electromagnetic Fields             cscscsccscscsssssccscsces 33  Recommended Separation Distances          cececccsccecececscecs 35  Range and Tolerance of Displayed Values                            36  Service and
19. coil    ONLY  Other conditions must be followed to ensure patient  safety  Failure to carefully follow all labeling conditions may  result in serious patient injury and or damage to the EndoStim  System     Lithotripsy    Direct exposure of an IPG to lithotripsy shock waves can cause  damage to the IPG  If the implant site is outside of the shock   wave path  no clear contraindication to the use of lithotripsy can  be established     AS a precaution  programming the IPG to the OFF mode reduces  the possibility of adverse effects  Check the IPG for proper  operation immediately following the procedure  If the unit is  found in the DOWN mode  follow the Reset procedure     Therapeutic and Diagnostic Ultrasound    Direct exposure of an IPG to diagnostic ultrasound can cause  damage to the IPG  The IPG may inadvertently concentrate the  ultrasonic field and cause harm to the patient     Therapeutic ultrasound may be used if the implant site 1s distant  and clearly outside of the ultrasonic field  Programming the IPG  to OFF mode reduces the possibility of adverse effects  Check  the IPG to assure operation immediately following the  procedure  If the unit is found in the DOWN mode  follow the  Reset procedure     EndoStim Lower Esophageal Sphincter Stimulation System 9    Effects on Other Implanted Devices    The EndoStim system may affect the operation of other  implanted devices  such as cardiac pacemakers  implantable  cardioverter defibrillators  ICDs   other neurostimulation
20. e the defective or non conforming  components free of charge  The Warranty Period for a replaced  component shall be the longer of the time remaining on the  original Warranty Period or nine months from delivery of the  replaced item     If a system component completely fails to function within the  first 72 hours of operation  EndoStim will replace the failed item  with a new one     EndoStim shall not be liable under this warranty if testing  and examination discloses that the alleged defect or non   conformity in the system component does not exist or was  caused by the end user s misuse  neglect  improper  implantation or testing  unauthorized attempts to repair  or  by accident  fire  lightning or other hazard     EndoStim Lower Esophageal Sphincter Stimulation System 37    EndoSt m    EndoStim B V   Prinsessegracht 20  2514 AP The Hague  The Netherlands   1 866 510 1003    Internet  www  endostim com       EndoStim B V  2015 Part Number CM 08 Rev B  All Rights Reserved    
21. eal Sphincter Stimulation  System is contraindicated for individuals with the following  conditions or needs     e Significant cardiac arrhythmia  or ectopy  or significant  cardiovascular disease   e Pregnant or nursing   The following treatments are contraindicated for patients  implanted with the EndoStim device   e Medical diathermy  e Transcutaneous Electrical Nerve Stimulation  TENS   exposure in the abdominal region    Precautions    The EndoStim Lower Esophageal Sphincter Stimulation  System has not been evaluated in the following  populations  Patients with the following conditions should  be considered prior to implant  e Large  greater than 3 cm  hiatal hernia that has not been  repaired  e Severe Grade D esophagitis  e Long segment Barrett   s esophagus or Barrett   s esophagus  with dysplasia  e Significant uncontrolled autoimmune disorder such as  Scleroderma  Dematomyositis  CREST syndrome     Sjogren   s Syndrome  or Sharp   s Syndrome  that affects  esophageal motility    2 EndoStim Lower Esophageal Sphincter Stimulation System    e Severe obesity that in the opinion of the surgeon increases  the risk of a laparoscopic implant procedure   e Type 1 diabetes mellitus   e Uncontrolled type 2 diabetes mellitus  T2DM  defined as  HbA Ic  gt  9 5 in the previous 6 months  or has T2DM for  more than 10 years   e Suspected or confirmed esophageal or gastric cancer   e Esophageal or gastric varices   e Dysphagia due to severe esophageal peptic structure   excluding Sc
22. eccccccscscecsecs 8  Magnetic Resonance Imaging  MRI             ccscscsscsssscseccsceees 9  LINOIS sos ssnea saidas asas Siena soda SS Sn Disse 9  Therapeutic and Diagnostic Ultrasound             cccecsscscsceeeees 9  Effects on Other Implanted Devices              csccscsssssecscscees 10  Transcutaneous Electrical Nerve Stimulator                       10  Home Appliances sis essas cuinvbotsenvcnvactsnsbicsa nese Oda o RU ana 10  Antitheft Systems scg samsasi sos vvedusencendccduecadesabe vet Sof anida Daio dit 11  Industrial Machinery            scccsccccscscsccscscsceccccscsceccccecncssees 11  Radio TransmitterS          seesessescessessesoesoscescescoesoessesseseeseesee 11  Cellular PRONE Siueni naai nei aei 12  Airport Screening SystemMS          sesssessescescoessesceecoescesseecceeo 12    iv EndoStim Lower Esophageal Sphincter Stimulation System    Ma NUS viccsctust atin ser ectscu sonia a tes sunee eeeeener  12    General Precaution         sesesesesesessececesesesoecececesesescececeseseeee 12  System OVELVICW       cccccceccccccccccccscscscscscscscccscscscescscscecs 13  Implantable Pulse GeneratoOr           csceccscscecsccscscececcscscscees 13  Implantable Bipolar Stimulating Lead                 cccecescscseees 16  PROS ram sasosaaisarenats se casswisuuassroneoessisaceensvecesssereiseeseusanes 17  IVIGINCEN ANCE assista eira E 19  Handling and Storage           cccccccccscscsccscscsccccccscsceccccscscssees 19  Service Life and Disposal             cccecsccscscecc
23. eparation Distances    Recommended separation distances between portable  and mobile RF communications equipment and the  EndoStim Programmer    The EndoStim Programmer is intended for use in an  electromagnetic environment in which radiated RF disturbances  are controlled  Table 9   The customer or user of the EndoStim  Programmer can help prevent electromagnetic interference by  maintaining a minimum distance between portable and mobile  RF communications equipment  transmitters  and the EndoStim  Programmer as recommended  according to the maximum  output power of the communications equipment      Table 9 Recommended Separation Distances Between  RF Equipment and EndoStim Programmer  Rated Separation distance according to  maximum frequency of transmitter  output 150 kHz to 80 MHz to 800 MHz to  power of 80 MHz    transmitter   Z 117NP   d 0 35VP   d 0 70VP       Notes     For transmitters rated at a maximum output power not listed  above  the recommended separation distance  d  in meters  m   can be estimated by using the equation applicable to the  frequency of the transmitter  where P is the maximum output  power rating of the transmitter in watts  W  according to the  transmitter manufacturer     e At 80 MHz and 800 MHz  the higher frequency range  applies     EndoStim Lower Esophageal Sphincter Stimulation System 35    These guidelines may not apply in all situations  Electromagnetic  propagation is affected by absorption and reflection from  structures  objects  and 
24. hatzki   s ring   e History of any advanced malignancy   e History of previous surgery such that directly involves the  lower esophageal sphincter or the EndoStim electrode  implant site  e g   fundoplication or an esophageal  myotomy   e Other implanted electro medical devices    e Younger than 21 years old    Potential Adverse Effects Events  Implantation of the System    Potential adverse effects events associated with the implantation  of the implantable pulse generator  IPG  and lead include  but  are not limited to  the following  death  pulmonary embolism   partial or complete ileus  peritonitis  esophageal perforation by  the electrodes  infection  inflammation  injury to organs within  the abdominal cavity  intravenous site complications   pneumonia  bleeding  incisional hernia  allergic or abnormal  reaction to anesthetic agents  pain  and fever     EndoStim Lower Esophageal Sphincter Stimulation System 3    Use of the System    Additional adverse effects that could be associated with the  EndoStim System include  but are not limited to  the following   lead electrode dislodgement  lead erosion or perforation into the  esophagus or stomach  IPG migration in the subcutaneous space   IPG erosion through the skin  diaphragmatic stimulation   stimulation of abdominal muscle  irritation and or inflammatory  response to the IPG and or the lead  allergic reaction to  materials  hematoma  infection  dysphagia  odynophagia  cardiac  arrhythmia  nausea  and discomfort  LES st
25. ibly lose statistical data  Electrocautery may  damage the IPG and lead  Application of electrocautery close to  an IPG can also cause damage to the LES tissue  possibly  producing burns     If electrocautery is required  apply it in short bursts  position the  ground plate on the patient to minimize current flow through the  IPG and lead  To reduce the possibility of adverse effects   program the IPG to the OFF mode  Check the IPG for proper  operation immediately following the procedure  If the unit is in  the DOWN mode  follow the Reset procedure     6 EndoStim Lower Esophageal Sphincter Stimulation System    RF Ablation    RF Ablation can cause the IPG to revert to its DOWN mode   with possible loss of statistics data  If sufficient energy 1s  coupled into the system  the unit may be damaged  Application  of RF ablation in close proximity to the electrodes of an  implanted IPG can also cause a direct coupling of radio   frequency energy through the leads and electrodes to the LES  tissue  possibly producing burns     If RF ablation is necessary  position the ground plate as far from  the IPG and lead as possible  Avoid direct contact with the  ablation catheter and the IPG and lead  Program the IPG to the  OFF mode to reduce the possibility of adverse effects  Check the  IPG for proper operation immediately following the procedure   If the unit is found in the DOWN mode  follow the Reset  procedure     Medical Diathermy    Medical diathermy  short wave thermal induction  
26. if it is exposed to a magnetic field  A standard pacemaker  safety magnet can be used     Electromagnetic Interference    The center frequency of the EndoStim Programmer transmitter   to the EndoStim II IPG  is 20 kHz with a bandwidth of 18 kHz  to 22 kHz     The center frequency of the EndoStim Programmer receiver   from the EndoStim II IPG  is 10 kHz to 28 kHz     30 EndoStim Lower Esophageal Sphincter Stimulation System    Electromagnetic Emissions    Guidance and manufacturer   s declaration      electromagnetic emissions    The EndoStim II IPG and Programmer are intended for use in  the electromagnetic environment specified in Table 6  The  customer or user of the EndoStim II IPG and Programmer should  assure that it is used in such an environment     Table 6 Electromagnetic Emissions    The EndoStim II IPG and Programmer are intended for use in the  electromagnetic environment specified below  The customer or user  of EndoStim II IPG and Programmer should assure that it is used in   such an environment     Emissions test Compliance   Electromagnetic environment    guidance    The EndoStim II IPG and  Programmer use RF energy  only for its internal function   Therefore  their RF emissions  are very low and are not likely  to cause any interference in  nearby electronic equipment   The EndoStim II IPG and  CISPR 11 Programmer are suitable for  Harmonic use in all establishments other  Emissions than domestic establishments  IEC 61000 3 2 Not and those directly connected  
27. imulation ceases  when the battery in the IPG 1s completely discharged     There is a potential that any system component could  malfunction  e g   software bug   become damaged  e g   lead  fracture   or the patient   s incision could become infected  System  component malfunction or other clinical circumstances  e g    sepsis  may require noninvasive corrective actions or possibly  even a surgical revision  repositioning  replacement  or removal   of the malfunctioning component s      It is recommended to shut the system down if the patient  experiences severe sensation or muscle or diaphragm  stimulation     Notes     e If necessary  use the Programmer to adjust the stimulation  amplitude to eliminate any pain  discomfort  or sensation  that may be related to stimulation  Reducing stimulation  amplitude is the only recommended adjustment  After  verifying the IPG is functioning as previously  programmed  it is recommend reducing stimulation  amplitude  incrementally in voltage steps of 0 5mA  until  the adverse effect resolves  If such attempts are not  successful  it may be necessary to use the Programmer or  a magnet to shut down the system     e A dramatic increase of the lead impedance may indicate a  partial or complete lead dislodgement  Patient symptoms  should be monitored and  if deteriorating  x ray or  fluoroscopy is recommended to assess the lead and  electrode location     4 EndoStim Lower Esophageal Sphincter Stimulation System    Warnings and Precautions  C
28. ion System    Inform patients about the possibility of interference from some  electric razors  electric power tools and electrical ignition  systems  including those used on gasoline powered devices  In  general  patients who have an IPG may operate gasoline   powered devices if protective hoods  shrouds  and other  shielding remain in place     Antitheft Systems    Certain types of antitheft devices  such as those used at  entrances exits of retail stores  libraries and other establishments   can interfere with the IPG  Most commonly  interference may  result in electrical stimulation delivery inhibition  Instruct  patients to walk at a normal pace and avoid lingering when  passing through the entrances and exits of these establishments     Industrial Machinery    High voltage power lines  electric arc welders  electric smelting  furnaces  and power generating equipment can interfere with the  operation of the IPG  For this reason  the intensities and  modulation characteristics of the electromagnetic fields  encountered by patients as a result of their occupation and  lifestyle should be considered  When appropriate  give specific  warnings     Radio Transmitters    Communications equipment such as radio and TV transmitters   including amateur    ham    transmitters  microwave transmitters  and CB transmitters with high power linear amplifiers  and radar  transmitters can interfere with the operation of the IPG  Discuss  this with patients whose occupations may expose them
29. is generally  contraindicated for patients implanted with active medical de   vices  It is unknown how the IPG will operate if subjected to the  intense fields of energy  Although damage to either the circuitry  of the IPG or LES tissue is improbable  it cannot be ruled out     If medical diathermy is required  apply it away from the  immediate vicinity of the IPG and lead  Programming the IPG to  the OFF mode reduces the possibility of adverse effects  Check  the IPG for proper operation immediately following the  procedure  If the unit is in the DOWN mode  follow the Reset  procedure     Contraindication  Patients and physicians should be advised  that exposure to medical diathermy is contraindicated in patients  implanted with the EndoStim System     EndoStim Lower Esophageal Sphincter Stimulation System d    Defibrillation    Any implanted active medical device can be damaged by cardiac  defibrillation procedures  In addition  the defibrillation current  can cause damage to LES tissue adjacent to the electrodes and or  to tissue surrounding the IPG  The defibrillation current may also  cause the IPG to revert to its DOWN mode  with possible loss of  statistics data  If sufficient energy is coupled into the system  the  unit may be damaged  If defibrillation is necessary  position the  paddles as far away from the implanted system as possible  avoid  placing the IPG in the defibrillation current path between the  paddles     Following defibrillation  closely monitor the pe
30. mmer Wand  Do not use solvents or cleaning cloths  infused with chemical cleaning agents     Warnings     e The Programmer is not protected against the ingress of  water  IXPO   Avoid immersing the Programmer in any  fluids     e Do not use the Programmer in the presence of flammable  anesthetics     e Do not sterilize the Programmer   e Do not connect any other equipment to the Programmer     e Do not modify the Programmer in any way     Handling and Storage    Do not use the Programmer if the package 1s damaged or if the  Programmer has been dropped from a height of 1 meter or more   Return damaged packages to EndoStim B V     The Programmer environmental conditions are found in Table 1     EndoStim Lower Esophageal Sphincter Stimulation System 19    Table 1 Environmental Conditions for the Programmer    Condition Shipping Operating and  Storage  Temperature  20 70   C   4 158   F  5 37   C  41 104   F   Humidity 15 93   non condensing 15 93   non condensing  Atmospheric 54 0 101 3 kPa  7 8 14 7 54 0 101 3 kPa  7 8 14 7  Pressure psi  psi   Altitude 0 5000 m  0 16 404 ft  0 5000 m  0 16 404 ft     Environmental conditions for the tablet PC may be different than  the Programmer  Check tablet PC user manual for tablet PC  environmental ranges and assume worst case condition     Service Life and Disposal    The life of service is expected to be 5 years  The Programmer  should be returned to EndoStim B V  when disposal is required     Environment of Use and Operator Profile    
31. nces is the  transmit receive RF head coil allowed to overlap with the off   limits area as shown in Figure 7        Figure 8 Brain MR Imaging Region with Head RF Coil    Extremity imaging should be performed using an appropriate  transmit receive RF local coil only  The diagrams below  illustrate several examples of the required placement of the  transmit receive RF local coil  as shown in grey  Under no  circumstances is the transmit receive RF head coil allowed to  overlap with the off limits area as shown in Figure 7     24 EndoStim Lower Esophageal Sphincter Stimulation System     i   fi     i     i          I     t         i              I              I           i          Figure 9 MR Imaging Region Knee and Ankle with Local RF Coil    If these instructions are not followed  higher than normal heating  may occur at the lead electrodes  If the patient has a broken  lead  higher than normal heating can also occur at the break if in  the area of the transmission of the RF energy  Excessive heating  can cause tissue damage or serious patient injury     Consult the MR system manufacturer for questions regarding the  type of RF coil in use with the 1 5 Tesla 64 MHz or 3   Tesla 128 MHz MR system     MRI Parameters    The RF frequency must be 64 MHz or 128 MHz  ONLY  Other  frequencies have not been tested  These could lead to tissue  damage or serious patient injury  Furthermore  use of other  transmit frequencies could cause device damage and or  excessive heating     The 
32. ogrammer Wand  when possible  to avoid any potential cross   contamination from previous use of the Programmer     EndoStim Lower Esophageal Sphincter Stimulation System 17    The Programmer can    Read  interrogate  IPG parameters as currently  programmed    Modify IPG parameters  Retrieve statistics accumulated by the IPG as it operates  Log the activity of the IPG    Store standard programs for future use    The Programmer has 2 components  Figure 6         Programmer Wand    Programmer Tablet PC with Software    LES PROGRAMMER    Stim    Endo    Figure 6 EndoStim LES System Programmer    Electrical Characteristics    The Programmer is internally powered Type B equipment that is  suitable for continuous use     EndoStim Lower Esophageal Sphincter Stimulation System    The Programmer is considered an applied part and the  Programmer tablet PC is considered out of the patient  environment  at least 1 5 meters from the patient      The Programmer can communicate to the IPG at a maximum  distance of 5 0 cm     Maintenance    The Programmer Wand does not contain any user serviceable  parts  If any Programmer Wand parts become damaged or loose   or it does not function properly  return the Programmer Wand to  EndoStim B V     After each use  it is recommended that you disconnect the  Programmer Wand from the USB port of the tablet PC  To clean  the Programmer  use a soft cloth dampened with distilled water   methanol  or isopropyl alcohol to wipe the exterior case of the  Progra
33. people     Range and Tolerance of Displayed Values  IPG and Lead Parameters      Parameter   Range   Increment __  Tolerance  Max of  100Q and  20     Impedance 200 to 2000  Q     Stimulation Parameters    Parameter Range Increment   Tolerance    Pulse Width 30 to 975  usec    30usec Max of 5   and  15usec    Pulse 2 0 to 10 0  mA    0 5mA Max of  Amplitude 0 5mA  and 20     Pulse Rate 2 to 80  Hz  2 10Hz  1Hz Max of 1   10 40Hz  and  2Hz 100usec  40 80Hz   5Hz    Active Time 00 00 01 to  23 59 59  23 59 59    Dose Mode Parameters    Parameter Range   Increment   Tolerance    00 00 01 to  Block Time 0 5 to 4 0  hr   Dose Schedule   00 00 to 23 30    Sensing Parameters    Parameter Range   Increment   Tolerance    Supine Time 1  5  30  or 60 N A 1 min   min        10    1     Yo  36    EndoStim Lower Esophageal Sphincter Stimulation System    O          Supine Time     70  80  90  or 95   N A 1          Service and Warranty    EndoStim provides emergency device consultation on a 24 hour  basis  If you require emergency assistance  please contact the  following number  1 866 510 1003     EndoStim warrants that all IPGs and accessories  including  associated firmware and software  will be free from defects in  workmanship and materials for a period of 12 months after the  original implantation of the IPG  the    Warranty Period         If it appears that an IPG contains a defect in workmanship or  materials  or fails to conform to applicable specifications   EndoStim will replac
34. pply  input lines    IEC 61000 4 11   Power Power frequency  frequency magnetic fields should  be at levels  characteristic of a typical  location in a residential  environment      50 60 Hz        IEC 61000 4 8    32 EndoStim Lower Esophageal Sphincter Stimulation System    Radiated Electromagnetic Fields    Guidance and manufacturer   s declaration      electromagnetic immunity    The EndoStim Programmer is intended for use in the radiated  electromagnetic fields specified in Table 8  The customer or user  of the EndoStim Programmer should assure that it is used in such  an environment     Portable and mobile RF communications equipment should be  used no closer to any part of the EndoStim Programmer   including cables  than the recommended separation distance  calculated from the equation applicable to the frequency of the  transmitter     Table 8 Radiated Electromagnetic Fields   Electromagnetic Immunity    Immunity test IEC Compliance   Recommended  60601Test level Separation  level Distance  Conducted RF  IEC 3 Vrms 3 Vrms  61000 4 6  150 kHz d 1 17VP  to 80 MHz outside  ISM bands    Conducted RF  IEC 3 Vrms 3 Vrms    P  61000 4 6  150 kHz d 1 20  to 80 MHz inside   ISM bands    Radiated RF  IEC 3  ee   3 ia kaa  61000 4 3  80 MHz d 1 20  to 800 MHz    Radiated RF  IEC 3 all 3 all o  61000 4 3  800 MHz Eid   2 304 P  to 2 5 GHz       EndoStim Lower Esophageal Sphincter Stimulation System 33    Notes     e Pis the maximum output power rating of the transmitter in  watts 
35. remation and Incineration    The IPG contains a sealed chemical power cell  lithium battery    For this reason  never incinerate an IPG  Be sure that the IPG is  explanted before a deceased patient is cremated  Contact your  local waste management officials for other information about the  environmentally safe collection and disposal of the IPG     Dropping and Damage    Do not implant the IPG if the package is damaged or if the IPG  has been dropped from a height of 30 cm or more     Do not implant the Lead if the package is damaged or if the IPG  has been dropped from a height of 30 cm or more     Do not use the Programmer if the package is damaged or if the  Programmer has been dropped from a height of 1 meter or more    Reuse    The IPG and implanted lead are intended for single use only     Environmental Hazards    The following discussion reflects a conservative approach to the  issue of patient safety in the presence of potential environmental  hazards  Design features in the IPG minimize the potential for  such hazards  but they cannot be ruled out completely     Notes     e Do not use any other electrical equipment adjacent to the  EndoStim System  If the components cannot be kept  separate  then monitor devices to assure normal operation     e Portable and mobile RF  radio frequency  equipment can  interfere with the normal operation of the EndoStim  System  The portable and mobile RF equipment should be  considered in any situation where the EndoStim System  device
36. rformance of the  IPG  If an operational abnormality is detected  consider  repositioning or replacing the lead and or reprogramming  or  replacing  the IPG  If the IPG changes to the DOWN mode   follow the Reset procedure     Therapeutic Radiation    Therapeutic equipment that produces ionizing radiation  such as  linear accelerators and cobalt machines used in cancer treatment   can damage the type of circuitry used in most active implantable  medical devices  Since the effect is cumulative  both dose rate  and total radiation dosage determine whether  and to what extent   damage will occur  Please note that any damage to the IPG may  not be immediately detected     In addition  the electromagnetic fields generated by some  therapeutic machines as part of the energy    steering    process can  affect the operation of the IPG     The effects of radiation therapy can range from temporary  disturbance to permanent damage  Therefore  if such therapy is  used  protect the IPG with local radiation shielding  and monitor  its performance during and after treatment  If tissue near the  implant site must be irradiated  it may be advisable to relocate  the IPG     8 EndoStim Lower Esophageal Sphincter Stimulation System    Magnetic Resonance Imaging  MRI     The EndoStim System is MR conditional and high specific  conditions must be followed to ensure patient safety  This  includes using 1 5 Tesla 64 MHz or 3 Tesla 128 MHz MR  systems and a transmit receive RF local coil  e g   head 
37. s are not acting as expected  Other equipment may  interfere with these devices  even if that equipment  complies with CISPR emission limits     EndoStim Lower Esophageal Sphincter Stimulation System 5    As with any medical device system  all the components of the  EndoStim System can be affected by magnetic  electrical  and  electromagnetic signals of sufficient strength  On rare occasions   interfering signals could inhibit electrical stimulation delivery or   alternatively  trigger inappropriate delivery of electrical  stimulation signals  In addition  certain sources can couple  sufficient energy into the IPG to damage the circuitry of the IPG  and or LES tissue adjacent to the electrodes  The physician may  wish to discuss these risks with the patient     The susceptibility of a particular unit will also depend on the  location of the IPG pocket  the nature of the interference  and the  programmed operating parameters     Because of the diversity of potential causes of electromagnetic  interference  EndoStim cannot characterize and describe within  this manual the effects of all potential sources of interference     Warning  Advise patients to be cautious when in the vicinity of  equipment that generates electrical or magnetic fields and to seek  medical advice before entering an area posted with a warning for  pacemaker patients  or other medical implantable devices      Electrocautery    The surgical use of electrocautery can cause the IPG to become  inactive and poss
38. t is in the same room as the MR system     After the MRI Procedure    Verify that the patient has not experienced any adverse effects  during the MRI procedure     Once outside of the MR system room  reset the IPG using the  Programmer  Reprogram the parameter values to those prior to  the MRI procedure and enable stimulation  Any problem with  these programming steps indicates that the IPG may have been  damaged due to the MRI examination     Immediately report any device related issues to the implanting  physician and EndoStim     EndoStim Lower Esophageal Sphincter Stimulation System 21    System Specifications    This section contains specifications and characteristics for all the  devices included in the LES system  IPG  Bipolar lead  and  Programmer     IPG Specifications    Table 2 Implantable Pulse Generator Specifications    Description Value   Height 62 mm   Width 39 mm   Thickness 8 4 mm   Mass 28 5 9   Biocompatible materials in Titanium   contact with human tissue Epoxy resin  Silicone rubber set plugs   Power source Lithium carbon monoflouride  battery   Storage temperature 20  C to 25  C    IPG Battery Characteristics and Specifications    The battery voltage at beginning of life is approximately 3 3V  and the usable capacity is 2 5 Ah  When battery voltage falls  below 2 5V  the device sends information to the Programmer   upon interrogation  that the battery is near the end of life  In  addition  stimulation amplitude will be reduced if programmed at  greater
39. xamination  1 e   no sedation or anesthesia is allowed      Beginning of the MRI Procedure    Verify the patient   s weight is entered correctly into the MR  system   s console to ensure that the regional SAR 1s correctly  calculated     Verify that the MR system   s transmit receive RF local coil is  being used     Verify that the MR system transmit receive RF body coil is not  being used     26 EndoStim Lower Esophageal Sphincter Stimulation System    Verify the MRI parameters for compliance with the requirements  in this section  If the parameters are not compliant  they should  be adjusted to meet the requirements  If this is not possible  the  MRI procedure must not be performed     Verify that the Programmer is not in the same room as the MR  system     During the MRI Procedure    The EndoStim System should not activate  turn on  or conduct  any energies into a stimulation     There should be no visual artifacts from the EndoStim System as  it is not to be placed within the transmit receive RF local coil     Monitor the patient visually and audibly during  and between   the MR imaging sequences to ensure the patient is not having  any difficulty     During the MRI procedure  the patient may feel vibration of the  IPG     Discontinue the MRI procedure immediately if the patient  reports any issues  is unable to respond to any questions  or the  IPG movement or vibration is too uncomfortable     Do not  under any circumstances  attempt to program the IPG  while the patien
    