Home
        RF Generator User Manual - STOCKERT medical solutions
         Contents
1.        The ablation created a  steam pop       The catheter was  pushed suddenly into  the tissue       The spike cut off value  is too low     Indicators   V   Visual  S   Sound     Alert    V  A and Alarm indicator turns  on orange    S   none  Ablation Stop Reason    V   Ablation Summary and  catheter receptacle flashes red for  5 seconds    S   Ablation stop tone    Ablation Stop Reason    V  A and Ablation Summary  and catheter receptacle flashes red  for 5 seconds    S   Ablation stop tone    7 Troubleshooting   Page 76    What To Do    Impedance below minimum    If the catheter is connected  through an additional system   such as a navigation system  try  replacing the system parts that are  between the generator and the  catheter  or connect the catheter  directly to the generator     If the problem persists  contact  Customer Support     It is not recommended  but if  you know the reason for the  low impedance and still want  to perform an ablation  you can  disable this cut off by setting its  value to Off   See Section 6 3 3   Editing a Preset      Impedance spike cut off    Consider possible clinical reasons     If you suspect that there is no  reason for the impedance to  spike  try increasing the cut off to  avoid getting this message due to  measurement noise     If the problem persists  contact  Customer Support     If it is clinically appropriate  you  can disable this cut off by setting  its value to Off   See Section 6 3 3     Editing a Preset      RF Gener
2.       Connection status      Type of catheter  selected on PRESET  screen      Connection status    Action  N A    Foot pedal  test  for the  foot pedal  connected to  the generator     N A    N A    N A    Action Description    N A    Press the foot pedal connected to the  generator to verify that it is working   During the foot pedal test  a tone is  emitted and the image changes toa  pressed foot pedal  RF energy is not  delivered during the foot pedal test   while the Foot Pedal device screen is  displayed      N A    N A    See Section 6 3 2  Selecting a Preset     6 Operation  Connections   Page 37    RF Generator User Manual       Table 4   System Settings Screen Options        In    Oo    ma tio n Dis o  aye fo Action       Date Press a gray date button and use the Data Entry Knob O to change the  date    Time Press a gray time button and use the Data Entry Knob O to change the  time    Language Select a language  if supported by the software version     Calibrate Touch Screen Press Calibrate Touch Screen to improve the performance of the touch   button screen  see Section 7 5  Calibrating the Touch Screen     Service button This button is for use by service technicians    Export to USB button Press Export to USB to export ablation and event log data to a USB flash  drive  see Section 6 5 1  Exporting Ablation and Event Log Data to a USB  Flash Drive      P    6 Operation  Connections   Page 38    RF Generator User Manual       6 3 Presets    PRESET Screen    ThermoCo
3.      When catheters with cooled ablation electrodes are used  the temperature measurement  reflects the temperature of the cooled electrode  not the temperature of the tissue    The temperature of the tissue may be distinctly higher and the risk of pop  explosion of  steam bubbles  may increase  Therefore  use only moderate RF power output  Follow the  recommendations in the instructions for use provided with the therapeutic catheters that  are used with the system     3 Warnings and Precautions   Page 10    RF Generator User Manual SMART       4  Keepin mind when setting the temperature that only the temperature of the electrode and  not the temperature of the heart tissue is measured  Because of the cooling effect of blood  flow  the temperature of the heart tissue may be higher than the temperature measured at  the ablation electrode     5  Avoid sudden increases in impedance to minimize charring at the ablation electrode   Charring at the ablation electrode can result in reduced RF energy delivery and or an  embolism     6  Make sure that the active RF electrode of the ablation catheter is not in contact with another  catheter or with another metallic conductor  such as an implanted pacemaker lead  This  could lead to uncontrolled conduction of the RF energy to other parts of the body  or to an  uncontrolled increase in the effective size of the active RF electrode     7  Do not set any extreme  unrealistic limit values  The limit values serve to trigger an alarm  when a 
4.     Products Law     UL 60601 1 Medical electrical equipment   Part 1  General requirements for safety    8 11 EMC Information and Technical Description    The use of accessories  transducers and cables other than those specified  with the exception  of transducers and cables approved by the manufacturer of the SMART  BLATE  RF Generator  as replacement parts for internal components  may result in increased emissions or decreased  immunity of the SMaRTABLATE    RF Generator     Electromagnetic Emissions    The SMARTABLATE    RF Generator is intended for use in the electromagnetic environment specified  below  The customer or user of the SMARTABLATE    RF Generator should ensure that it is used in such an  environment     Emissions Test Compliance Electromagnetic Environment   Guidance  RF emissions Group 1 The SMARTABLATE    RF Generator uses  CISPR 11 RF energy only for its internal function     Therefore  its RF emissions are very low and  are not likely to cause any interference in  nearby electronic equipment     RF emissions Class A  CISPR 11 The SmarTABLaTe    RF Generator is suitable      for use in all establishments other than   Harmonic emissions Class A   domestic and those directly connected to the  IEC 61000 3 2   public low voltage power supply network  Voltage fluctuations flicker Complies that supplies buildings used for domestic  emissions purposes   IEC 61000 3 3    8 Technical Data   Page 92    RF Generator User Manual       Electromagnetic Immunity    The 
5.    The SMARTABLATE    RF Generator meets the requirements of the following standards     Standard  ANSI   AAMI HF 18    CAN CSA C22 2  No  601 1  B 90    CMDR   DIN EN 980   DIN EN 1041   DIN EN ISO 14971  EU Dir  2002 96 EC    EU Dir  2006 66 EC    IEC 60812    IEC EN 60601 1    IEC EN 60601 1 1    IEC EN 60601 1 2    IEC EN 60601 1 4    IEC EN 60601 1 6    IEC EN 60601 1 8    IEC EN 60601 1 9    IEC EN 60601 2 2    IEC EN 62304    Title  Electrosurgical devices    Medical electrical equipment   Part 1  General requirements for safety    Canadian Medical Devices Regulations   Symbols for use in the labeling of medical devices   Information supplied by the manufacturer of medical devices  Medical devices   Application of risk management to medical devices    EU Directive 2002 96 EC of 27 January 2003 on waste electrical and electronic  equipment  WEEE     Directive 2006 66 EC of the European Parliament and of the Council of 6  September 2006 on batteries and accumulators and waste batteries and  accumulators and repealing Directive 91 157 EEC    Analysis techniques for system reliability   Procedure for failure mode and  effects analysis  FMEA     Medical electrical equipment   Part 1  General requirements for basic safety and  essential performance    Medical electrical equipment   Part 1 1  General requirements for  safety   Collateral standard  Safety requirements for medical electrical systems    Medical electrical equipment   Part 1 2  General requirements for basic safe
6.    V  A and Ablation Summary    S none    Foot pedal released   Remote      The pressed foot pedal  that is connected to  the remote control was  released     Ablation Stop Reason    V  A and Ablation Summary    S none      Disconnect the foot pedal    from the remote control  then  reconnect it  If the problem  persists  either use a different foot  pedal or operate the generator    using the Ablation Start    and    Stop buttons on the generator  or remote control       To resume ablation  press and    hold the foot pedal       To resume ablation  press and    hold the foot pedal     7 Troubleshooting   Page 74    RF Generator User Manual       Table 8   Warnings  Alerts  and Ablation Stop Reasons    Message Indicators  What To Do  Reasons  V   Visual  S   Sound   Impedance above maximum    Theimpedance exceeded Alert   Verify that the catheter is properly  the maximum cut off V  A and Alarm indicator turns inserted in the body and is out of  value set in the preset  on orange the sheath   Possible reasons are      Disconnect the catheter and the    The catheter is out of zur indifferent electrode from the  the patient s body oris Ablation Stop Reason generator  then reconnect them   covered by a sheath    Improve the contact of the    V   Ablation Summary and    catheter receptacle flashes red for indifferent electrode with the    patient s skin or use a different      The indifferent  electrode has become    5 d a  disconnected from the m indifferent electrode   See  gene
7.    When using a split  indifferent electrode   the contact quality with  the patient   s skin has  deteriorated since the  indifferent electrode was  applied to the patient     Bubble      The pump sensor  identified a bubble in the  irrigation tubing       The temperature of   the catheter ablation  electrode exceeded the  cut off value set in the  preset  Possible reasons  are       The ablation electrode  is too hot       The cut off value is too  low     Indicators   V   Visual  S   Sound     Bad contact indifferent electrode      Alert    V  A and Alarm indicator turns  on orange    S   none  Ablation Stop Reason    V   Ablation Summary and both  indifferent electrode receptacles  flash blue for 5 seconds    S   Ablation stop tone    Alert    V  A and Alarm indicator turns  on orange    S   none  Ablation Stop Reason  V   Ablation Summary    S   Ablation stop tone    Catheter temperature above maximum    Alert    V  A and Alarm indicator turns  on orange    S   none  Ablation Stop Reason    V   Ablation Summary and  catheter receptacle flashes red for  5 seconds    S   Ablation stop tone    7 Troubleshooting   Page 69    What To Do      When using a split indifferent    electrode  improve the contact of  the indifferent electrode with the  patient   s skin   See 3 4  Handling  Indifferent Electrodes        If the problem persists  use a    different indifferent electrode ora  different cable       Patient Danger         Press the Stop button on the    pump       Flush th
8.   Data Entry Knob    4 System Overview and Setup   Page 15    RF Generator User Manual       4 3 Controls on the Back of the Generator    DEVICE LABEL       Item Description  1 Remote Control   Receptacle for SMARTABLATE    Remote Control cable  2 USB port  3 Serial data port 1   for electrophysiology recording system  4 Mains rating label  5 Fiber optic ports  not implemented   6 Mains socket  7 Fuse cover  8 Potential equalization port  9 Serial data port 2   for Carto   System  10 Ethernet port  not implemented   11 Pump   Receptacle for SMARTABLATE    Irrigation Pump cable  12 Foot Pedal   Receptacle for SMARTABLATE    System foot pedal cable  13 Device label    4 System Overview and Setup   Page 16    RF Generator User Manual       4 4 Functional Test and On Site Training    The SmarTABLATE    RF Generator is a medical device  In some countries  in order to comply with  local regulations  the generator may not be put into service until the distributor has subjected  the generator to a functional test on site and has trained the person responsible for operating the  device in the handling of the generator per this user manual  All applicable requirements of local  regulations must be followed     A functional test of the generator can be performed using a STOCKERT test box for the  SMARTABLATE    RF Generator  which is available for purchase  see Section 5 2  Accessory Lists   See  the instructions for use that are provided with the test box     4 5 Setup  4 5 1 Setting U
9.   Edit the Preset      Follow the instructions in Section 6 3 3  Editing a Preset     3  Save the Preset Temporarily      Press the ABLATION workflow button or another workflow button     or      Press Save O then press Save temporarily     Notes   1     When you save temporarily     EDITED PRESET     is displayed above the list of  presets and none of the presets in the list is selected  To understand which  preset has been temporarily edited  look at the name of the preset on the  PRESET workflow button   the name will be followed by an asterisk         Temporary changes are in effect only until a different preset is selected or the  system is turned off  When a different preset is selected  the temporary changes  to the previous preset are discarded     The most common time to perform a temporary save is when you are preparing  to start ablation or during a pause in ablation when you determine that a  permanently saved preset does not contain the exact settings that you want     To save the temporary preset as a new preset  follow the instructions in Section  6 3 5  Creating a New Preset     All changes made at the ABLATION screen are temporary  see Section 6 4 3   Making Temporary Changes to Ablation Settings      6 Operation  Presets   Page 44    RF Generator User Manual       6 3 5 Creating a New Preset    Keyboard Screen    PHYSICIAN jum  lt       Backspace        erases one  character     OK    closes keyboard  and saves physician  name and preset  name     Escape   c
10.   The Ethernet port on the back of the generator  see Item 10 in Section 4 3  Controls on  the Back of the Generator  is provided for future use        DODDDDDD    4 5 9 USB Connection    Only USB mass storage devices  such as USB flash drives  may be connected to the  USB port on the back of the generator  see Item 2 in Section 4 3  Controls on the Back  of the Generator   Do not connect any other USB devices  such as a printer  mouse  or  scanner   these other devices are not supported by the generator        Note  The USB ports on the back of the tilting front panel of the generator are for use  by service technicians only     4 5 10 Serial Data Connections    lOlOll Devices that exchange data with the generator  such as the Carto   3 System  the  CARTO   XP System  compatible electrophysiology recording systems  and other external  compatible monitoring systems can be connected to the serial data ports on the   back of the generator  see Items 3 and 9 in Section 4 3  Controls on the Back of the  Generator         4 5 11 Fiber Optic Connections    The fiber optic ports on the back of the generator  see Item 5 in Section 4 3  Controls  on the Back of the Generator  are provided for future use        4 5 12 Setting Up Irrigated Catheters with the Irrigation Pump and Tubing    Connect the SmarTABLATE    RF Generator to the SMARTABLATE    Irrigation Pump by plugging one  end of the connection cable into the Pump receptacle on the back of the generator  see Item  11 in Section 
11.   the increase will be determined from a baseline  measurement taken just before ablation     Impedance Settings Pane    Max  Cut Off The maximum impedance measurement between the ablation  electrode and the indifferent electrode above which RF energy  will automatically shut off     Spike Cut Off If the impedance spikes higher than the number set here in any  half second interval  the RF energy will automatically shut off     Min  Cut Off The lowest impedance measurement between the ablation  electrode and indifferent electrode below which the RF energy  will automatically shut off     Irrigation Control Settings Pane    This pane appears only when an irrigated catheter type is selected  see Section 6 3 10   Irrigation Control      Other  Audio Settings 2  See Section 6 3 11  Changing Audio Settings     Display Settings See Section 6 3 12  Selecting Ablation Display Settings     a  6 Operation  Presets   Page 43    RF Generator User Manual       6 3 4 Editing a Preset   Temporary Save    EDITED  PRESET   indicates   temporary save    Asterisk     after  preset name  indicates  temporary save  of that preset    PRESET Screen After Temporary Save    4mm Standard  D  a Edit  Set Default    4mm    8mm    ThermotCool      ThermoCool   SF       Preset 4mm Standard     Quick Steps   1  Display the Edit Preset Screen    Press the PRESET workflow button     Select the preset to edit     Press Edit     or      When the ABLATION screen is displayed  press Edit on the Navigation Bar   2
12.   the safety of the device and system is not guaranteed     5 2 1 Devices       SMARTABLATE    Irrigation Pum  p       Worldwide  except USA  Japan  China M4900103  USA M490003   Japan M4900203  China M4900403       5 Accessories   Page 23    RF Generator User Manual       SMART  BLATE   Remote Control       Worldwide  except USA  Japan  China M4900104  USA M490004  Japan M4900204  China M4900404  STOCKERT Testbox III 39F45X                            5 2 2 Foot Pedal       Foot Pedal  length of connection cable  3 m       Worldwide  except China M490005  China M4900405          ES   Foot Pedal Extension Cable  length  7 m M490026     Redel 6  gt  Redel 6  black     5 2 3 SMARTABLATE    Cables    Connection Cable from SMARTABLATE    RF    Generator to SMARTABLATE    Remote Control   Redel 10  gt  Redel 10  yellow     length  10 m M490010  length  15 m M490029  length  25 m M490012  length  30 m M490013    Connection Cable from SMARTABLATE    RF    Generator to SMARTABLATE    Irrigation Pump   Redel 10  gt  Redel 10  blue     length  1 m M490027  length  3 m M490016  length  5 m M490028  length  10 m M490030       5 Accessories   Page 24    RF Generator User Manual    Item  Mains Cable  length  3 m  Australia  Brazil  China  EU  India   South Africa  Israel  Japan  Russia  Switzerland  Taiwan  UK   Singapore   Hong Kong  USA  Grounding Cable  length  3 m  Worldwide  except Japan    Japan    ECG Connection Cable  length  3 m   Redel 16  gt  4x 2 mm tip pins  green     Serial Da
13.  Bar    CONNECTIONS screen    Navigation Bar    Manual Section  6 3 11  Changing Audio Settings    The icon is gray if this feature is not  supported by the software version     Press the icon to close the screen     6 3 7  Deleting a Preset    6 3 12  Selecting Ablation Display  Settings    Press the icon to close the screen  without saving changes     6 2  Connections    6 3 10  Irrigation Control    The icon is gray until the ABLATION  workflow button is pressed     6 4 1  Starting a New Procedure    The icon is gray until the ABLATION  workflow button is pressed     Press the icon to close the screen and  save changes     6 3 3  Editing a Preset    6 4 3  Making Temporary Changes to  Ablation Settings    6 3 5  Creating a New Preset    6 7  System Control    6 8  Volume  Brightness  and Foot  Pedal Test    6 2  Connections    7 2  Warnings  Alerts  and Ablation  Stop Reasons       6 Operation  Icons on the Screens   Page 62    RF Generator User Manual    Name Icon Where Manual Section                System Status  Navigation Bar 7 2  Warnings  Alerts  and Ablation  Warning or Alert Stop Reasons  System Status  Navigation Bar 7 3  System Errors  System Error  USB  To USB PRESET screen 6 3  Presets  Also see 6 5 1  Exporting Ablation and  Event Log Data to a USB Flash Drive   The icon is gray if this feature is not  supported by the software version   USB  From USB PRESET screen 6 3  Presets  The icon is gray if this feature is not  supported by the software version   Wait 
14.  Do not modify accessories  Visually inspect all  accessories on a regular basis  Make sure that the connected cables are not damaged  When  using sterile accessories  be sure to maintain the sterility of those accessories     3 Warnings and Precautions   Page 6    RF Generator User Manual SMART       10  Place the generator on a secure  non slip surface such as the CarTo   System cart  If the  generator is placed on a mounting plate  make sure it is securely fastened  Do not place the  generator directly above another device and do not place any other device directly on the  generator  Make sure that there is enough free space on all sides of the generator to allow  the heat created by the generator to escape     11  The entire surface of the indifferent electrode must be as close as possible to the operating  field and must have fully reliable contact with the patient s body  The skin surface must be  free of excessive oil and body hair   See Section 3 3  Warnings and Precautions  During an  Ablation Procedure      12  The generator is equipped with a contact quality monitoring feature for split indifferent  electrodes  If the contact quality falls below a defined value  the generator displays an alert  message  However  if two split indifferent electrodes are used at the same time and one  loses contact quality but the other maintains contact quality  no alert is triggered  If a solid  indifferent electrode detaches from the patient s skin  no alert is triggered  When usin
15.  Generator User Manual SMART       Preface    The SMARTABLATE    Radiofrequency  RF  Generator is a highly specialized device intended for use  during RF ablation therapy of the human heart  This device is used in conjunction with specialized  therapeutic catheters and a dispersive pad  indifferent electrode  to create a closed electrical circuit  capable of delivering specified doses of RF energy  The RF energy is delivered to cardiac tissue that  forms unwanted electrical pathways that either drive or maintain arrhythmias  The RF energy heats  the tissue such that it becomes denatured and no longer functional  This interrupts and or destroys  the unwanted electrical pathways  thereby restoring normal heart function     The SmarTABLATE    RF Generator actively generates specified doses of RF energy and continuously  monitors  displays  and coordinates the amount of RF energy  the temperature of the catheter   s  ablation electrode  and the tissue impedance during ablation therapy  The temperature of the  catheter   s ablation electrode is measured by a sensor in the electrode and is transmitted to the  generator  Simultaneously  high resolution impedance measurement circuits built into the generator  measure the heart tissue impedance allowing detection of small tissue changes prior to  during   and after treatment  Before and during treatment  electrical signals generated by the heart are  also detected and transmitted by the catheter through the generator to connected moni
16.  M4900615  M4900616    A E LE    Indifferent Electrodes and Cables available from STOCKERT GmbH    Other indifferent electrodes that meet the requirements specified in Section 3 4  Handling  Indifferent Electrodes  are also compatible with the SMARTABLATE    RF Generator     Item    Erbe Indifferent Electrode for Children   reusable  17x 11cm    Erbe Indifferent Electrode for Adults  reusable   17 5 x 29 5 cm    Connection Cable for 3M Indifferent Electrode  with Clip  length  3 m    Connection Cable for Erbe Indifferent  Electrode with 4 mm Phone Jack  length  3 m    Catheters    Part    39D 15X  a  39D 19X  M490034     Eim      For recommended catheters  contact Customer Support     Note  The UL mark on this generator is only valid with accessories tested by Underwriters Laboratories        See http   www stockert de hleadmin downloads Tested_accessories_by_UL pdf     5 Accessories   Page 26    RF Generator User Manual       5 3 Caring for Accessories    ann    Do not modify any accessories    5 3 1 Connection Cables  Visually inspect all connection cables before each use  Do not use damaged cables   To avoid damage to the connection cables  do not wrap the cables around the generator or  other apparatus  Coiling the connection cables during normal operation of the generator creates  inductive components which can lead to measurement errors  The values indicated in error can    lead to misinterpretations     When plugging in a cable or unplugging a cable  hold the cable by 
17.  Mapping Physician  System Presets  System       Er eS a Te S a    oo    Ablation           Summary       Ablation Setting      Serr     Pane ee een   bot Warnings and     eee    Alerts Window   Remote     eee eee eee   Control Temperature    Di m ee Se en      Settings Pane         Foot Pedal   oo   wer   ee    Esophageal      Temperature   Indifferent   Settings Pane    Electrode HERREN    Impedance 7  Esophageal   Settings Pane      probe   N rain    Gr u DZ ee ze      Irrigation Control  Catheter   Settings Pane    System Irrigation  Settings Control  High Flow  Audio  Settings  Display  Settings     The Esophageal Probe and the Esophageal Temperature Settings Pane  shown above are not applicable for the US and JP versions of the generator      N  S        3 N  O  za  Wr       6 Operation  Flowchart of the Screens   Page 61    RF Generator User Manual       6 10    Name    Audio Settings    Close    Delete    Display Settings    Escape    Front Back    Irrigation Control    New Procedure    OK    Save    Save As    System Control    System Settings    System Status     Ready for Ablation    Icons on the Screens    Icon       pu       ESG             Where    Edit Preset screen    Top right corner of  many screens    PRESET screen    Edit Preset screen    Many screens    CONNECTIONS screen    Navigation Bar    Also appears on the  ABLATION screen  when fluid is low during  ablation     Navigation Bar    Many screens    Edit Preset screen    Edit Preset screen    Navigation
18.  Save     Notes   1  The ablation values that are displayed when you first access the ABLATION  screen are determined by the selected preset for the catheter tyoe   either  the default preset or a preset that you selected  see Sections 6 3 2  Selecting a  Preset  and 6 3 3  Editing a Preset      2  When Power Control Mode is selected in the preset     Power Control    is  displayed at the top of the screen   See the sample screen in Section 6 4   Ablation      3  Onthe ABLATION screen  before or during ablation  you can change the  time  s  and power  W  values  In Temperature Control Mode  you can also  change the temperature set point    C   For irrigated catheters  you can also  change the flow rate  mL min   These changes take effect immediately  without  confirmation     4  When you change the flow rate on the ABLATION screen     AUTO    at the top  of the screen changes to    MAN  It reverts to    AUTO    at the end of the current  ablation     5  All changes made as described above are temporary  An asterisk     appears  after the preset name on the PRESET workflow button to indicate that  temporary changes have been made but have not been saved  The changes  stay in effect until a different preset is selected or until the system is turned off   If you want to save these settings in a new preset  follow the steps in Section  6 3 5  Creating a New Preset     6 Operation  Ablation   Page 53    RF Generator User Manual       6  The catheter type selected for the preset im
19.  ablation catheter or to the  indifferent electrode that are not  approved by the manufacturer     7 Troubleshooting   Page 79    RF Generator User Manual       Table 8   Warnings  Alerts  and Ablation Stop Reasons    Message    Reasons    No low  standby  flow      An irrigated catheter  preset is selected  the  catheter temperature and  the impedance indicate  that the catheter is in the  patient  but the pump is  stopped     Power output too high      The generator was  trying to deliver a power  output that was beyond  its intended capability   Possible reasons are       There is a compatibility  problem between the  generator and external  devices       There is an internal  problem in the  generator     Pump door open      The pump door is open     Indicators   V   Visual  S   Sound     Warning    V  A and orange border around  Irrigation Control icon flashes for  20 seconds    S   Warning tone    Ablation Stop Reason    V  A and Ablation Summary  and catheter receptacle flashes red  for 5 seconds    S   Ablation stop tone    Alert    V  A and Alarm indicator turns  on orange    S   none  Ablation Stop Reason  V   Ablation Summary    S   Ablation stop tone    7 Troubleshooting   Page 80    What To Do      Press START Low Flow on the    generator s Irrigation Control  screen or press Low  Standby     Flow SA on the pump to  start the pump flowing at the low   standby  flow rate   See Section  6 3 10  Irrigation Control        Disconnect any devices connected    to the ab
20.  by other equipment  even if  that other equipment complies with CISPR EMISSION requirements     The SmarTABLaTE    RF Generator internally generates RF electromagnetic energy for the purpose  of its operation  The following frequency bands are used     e 45 kHz 1 kHz is used for impedance measurement between the catheter and the indifferent  electrode     e 480 kHz   1 kHz is used for impedance measurement between the catheter and the  indifferent electrode     e 100 kHz   1 kHz is used for impedance measurement of the split indifferent electrode     ECG signals have very small amplitudes  0 1 mV   For this reason  the SMARTABLATE    RF Generator  uses a sensitive measurement unit for the ECG signals  This measurement unit is also sensitive   to externally radiated RF energy and can be disturbed by other systems placed close to the  SMARTABLATE    RF Generator  Possible sources of interference are  for example  neon lamps and  power transformers     The following instructions must be followed to avoid disturbance of the measurement quality   e The SmarTABLATE    RF Generator may be operated only with cables provided by or  recommended by the manufacturer     e The SmartABLATE    RF Generator must be connected with the grounding cable to a separate  grounding point     e All devices in the patient area that are connected to the SMaARTABLATE    RF Generator must be  on the same mains     e All devices in the patient area that are connected to the SMARTABLATE    RF Generator mu
21.  connect a pump with a flushed tubing set  see the  instructions for use for the SMARTABLATE    Irrigation Pump  the instructions for use    for the SMARTABLATE    Irrigation Tubing Set  and Section 6 6  Flushing to Eliminate  Air in the Tubing and Catheter      6 1 8 Typical Procedure Process   Full    Quick Steps     Turn on the generator and remote control   6 1 1  System Startup       Atthe CONNECTIONS screen  review the device  6 2  Connections   connections     Physically connect all applicable devices to the  generator     Physically connect all applicable devices to the  patient       Atthe PRESET screen  select or create a preset   6 3  Presets       Go to the ABLATION screen  then position the  catheter at the first ablation site     Press Ablation Start    or press and hold the foot  6 4  Ablation   pedal to start ablation     Notes   1  Ata minimum  connect a catheter and an indifferent electrode to the  generator  For an irrigated catheter  also connect a pump and a flushed tubing  set  see the instructions for use for the SMARTABLATE    Irrigation Pump  the  instructions for use for the SMARTABLATE     Irrigation Tubing Set  and Section 6 6   Flushing to Eliminate Air in the Tubing and Catheter      2  Optionally  connect other devices shown on the CONNECTIONS screen   3  You can also start ablation by pressing Ablation Start    on the front of the    remote control or by pressing the foot pedal connected to the remote control  when the remote control is the sys
22.  in Radiofrequency Catheter Ablation  Circulation  86  No  5  1648 1650    Wittkampf FHM   Hauer RNW   Robles de Medina EO   1989   Control of Radiofrequency Lesions by  Power Regulation  Circulation 80  962 968    Wittkampf FHM   Nakagawa H   Yamanashi W   Imai S   Jackman WM   1996   Thermal Latency in  Radiofrequency Ablation  Circulation Vol 93  No  6  1083 1086    Wittkampf FHM   Simmers TA   Hauer RNW   Robles de Medina EOR   1995   Myocardial  Temperature Response during Radiofrequency Catheter Ablation  Pace Vol  18  307 317    SSS  11 References   Page 102    RF Generator User Manual    Manufactured By      STOCKERT GmbH  B  tzinger Stra  e 72  79111 Freiburg  Germany  Fax   49 761 20716 20  Email  info stockert de  http   www stockert de    Customer Support   In USA    Biosense Webster  Inc        Tel   1 866 4SERVCE   1 866 473 7823   toll free in USA     Email  complaint bwius jnj com    Customer Support   Outside USA    Biosense Webster  Tel   32 2 7463 463  Email  techsupport its jnj com    Distributed By    Outside USA   Biosense Webster   A Division of Johnson  amp  Johnson Medical NV SA  Leonardo da Vincilaan 15   1831 Diegem  Belgium   Tel   32 2 7463 401   Fax   32 2 7463 403    In USA   Biosense Webster  Inc    3333 Diamond Canyon Road  Diamond Bar  California 91765 USA  Tel   1 866 473 7823   Fax   1 909 468 2786  http   www biosensewebster com    Carto    CeLsius    SMARTABLATE     and THERMOCOOL   are trademarks of Biosense Webster  Inc  The  third part
23.  mapping system of the connected mapping system   The mapping system image on the    CarTo   3 CONNECTIONS screen will change   system  when this screen closes    CARTO   XP   System  or   Other     Pi    6 Operation  Connections   Page 35    RF Generator User Manual       Device    Recording  System    Pump    Table 3   Device Screens    Information Displayed Action    After selecting the TRANSMIT  mapping system   images and values  appear for verifying  transmission of  impedance data  STOP    Image of generatorand TRANSMIT  recording system    Values for verifying  transmission of data  STOP    Active flow rate  or Flush    dashes if the pump is  active when    not connected  the pump is    Serial number  or connected   dashes if the pump is  not connected     Message to connect   the pump and select an   irrigated catheter preset    if the pump is not   connected or if a non    irrigated catheter preset   is selected    START Low  Flow   visible when  the pump is  connected and  is stopped   STOP   visible when  the pump is    connected and    is running     Action Description    Press TRANSMIT to test the  communication on the mapping   system connection port  see Item 9 in  Section 4 3  Controls on the Back of the  Generator   The color in the arrow moves  from left to right to indicate that the test  is running     The number on the screen is the  impedance value being sent to the  mapping system     Press STOP to stop the transmission test     Press TRANSMIT to test 
24.  open the System Control  screen     Volume    Press the volume bar  then turn the Data Entry Knob   to change the volume  of the audio output     Brightness  Select Automatic to have the system determine the screen brightness     Select Manual  then press the brightness bar and turn the Data Entry Knob O  to adjust the screen brightness manually     Foot Pedal    Press the foot pedal connected to the generator to verify that it is working   During the foot pedal test  a tone is emitted and the image changes to a pressed  foot pedal  RF energy is not delivered during the foot pedal test  while the  System Control screen is displayed      Notes 1  This screen is displayed only on the local unit  generator or remote control  and  the settings affect only the local unit  The local unit can be the system Master  or the System monitor when you access this screen   See Section 6 7  System  Control for information about the system Master      2  The volume setting affects the RF tone and all alarm and informational tones   Make sure that you select the appropriate volume so that you can hear the  tones     3  When the generator is the system Master  you can also perform the foot  pedal test on the Foot Pedal device screen  which is accessible from the  CONNECTIONS screen  see Table 3   Device Screens  in Section 6 2   Connections      6 Operation  Volume  Brightness  and Foot Pedal Test   Page 60    RF Generator User Manual       6 9 Flowchart of the Screens    PRESET ABLATION POST 
25.  remove the Warnings  and Alerts window from the screen     A   There is a System Error or a Fatal System Error  The Event window appears   See Sections  7 3 1  System Error and 7 3 2  Fatal System Error      Ablation Summary with Ablation Stop Reason    ABLATION SUMMARY  RF time total 60s  Power max  10W  Power avg  9  9W  Energy total   Temp  max    Temp  avg     Imp  max     Imp  drop Ablation Stop    Imp  min  1000    Irrigation 19mL Reason    Ablation Stop Reason     Timer completed       Ablation a on Kr      System Status  icon    When ablation stops because the preset ablation time has been reached or because ablation    has been intentionally stopped  the A System Status icon is displayed  The Ablation Summary  appears displaying an Ablation Stop Reason such as    Timer completed        When ablation stops because an Alert has occurred  the A System Status icon is displayed and  the Ablation Summary screen appears displaying an Ablation Stop Reason that explains the Alert   You also have the option of pressing the icon to display the Warnings and Alerts window  Resolve  the Alert  see Table 8   Warnings  Alerts  and Ablation Stop Reasons below  and then press Close  to remove the Warnings and Alerts window from the screen        7 Troubleshooting   Page 68    RF Generator User Manual       Table 8 describes the Warnings  Alerts  and Ablation Stop Reasons that may appear on the screen     Table 8   Warnings  Alerts  and Ablation Stop Reasons    Message      Reasons   
26.  the  Generator      Verify that you are using the correct catheter  connection cables for the selected catheter  type and for the selected mapping system    See Section 5 2 3  SMARTABLATE    Cables      Replace the catheter and or connection  cable s  with new ones     Warnings and Alerts    Ablation       Active WARNINGS    No active warnings    Active ALERTS Warnings  Pump not connected an d A   e rts  window    System  Status  icon    The System Status icon is visible on the Navigation Bar whenever ablation is not being    performed  There are three different System Status icons  A A A    A   Ablation can be performed and there are no active Warnings or Alerts        7 Troubleshooting   Page 67    RF Generator User Manual       A   There is aWarning or an Alert     A Warning indicates a condition that is not critical  Ablation can be performed even when  there is a Warning  For example     Low fluid        An Alert indicates a condition that is critical and must be fixed before ablation can be  performed  An Alert may occur when ablation is not being performed or it may stop  ablation  For example     Catheter temperature above maximum        The System Status icon changes from A to A when a Warning or an Alert occurs  Press  the icon to display the Warnings and Alerts window  Resolve the Warning or Alert  see  Table 8   Warnings  Alerts  and Ablation Stop Reasons in this section   As each Warning or  Alert is resolved  it disappears from the screen  Press Close        to
27.  to drop before  the cut off value set on orange restarting ablation   in the preset  Possible   See the instructions in this table  I aves a  One for Esoph  temperature above    The ablation is heating Ablation Stop Reason maximum  Warning    the esophagus     V   Ablation Summary and  esophageal probe receptacle  flashes red for 5 seconds      The cut off value is too  low     S   Ablation stop tone    Esoph  temperature above maximum  not applicable for the US and JP versions of the generator       The esophageal probe Warning   Reduce the power setting used for  temperature exceeded A the ablation or stop the ablation   the warning value set   u punpa an re   Ifitis clinically appropriate   in the preset  Possible An reposition the catheter farther  reasons are  orange border flashes    away from the esophagus     The ablation is heating S  Warning tone      Ifthe problem persists  contact  the esophagus     Customer Support     The warning value is      You can disable this warning  too low     by setting its value to Off   See  Section 6 3 3  Editing a Preset      7 Troubleshooting   Page 72    RF Generator User Manual SMART       Table 8   Warnings  Alerts  and Ablation Stop Reasons    Message Indicators    What To D    Reasons  V   Visual  S   Sound  Se    Esophageal temperature invalid  not applicable for the US and JP versions of the generator       The generator did Alert   Verify that the esophageal probe  not recognize a valid V  A and Alarm indicator turns is co
28. 4 3  Controls on the Back of the Generator  and the other end of the cable into the  appropriate receptacle on the back of the pump     Connect a SMARTABLATE    Irrigation Tubing Set with the pump and with an irrigated catheter as  described in the SmarTABLATE    Irrigation Pump user manual  Also see Section 6 6  Flushing to  Eliminate Air in the Tubing and Catheter     4 5 13 Tilting the Generator   s Front Panel  Generators with a Tilt Panel button     After the generator has been turned on  see Section 6 1 1  System Startup   the front panel can  be tilted for easier viewing  Press the Tilt Panel button  see Item 7 in Section 4 2  Controls on the  Front of the Generator  to release the lock and gently pull the bottom of the front panel forward     Note  The generator   s front panel cannot be adjusted when the generator is off  Before moving  the generator  make sure that the front panel is closed     4 System Overview and Setup   Page 19    RF Generator User Manual SMART       4 5 14 Tilting the Generator   s Front Panel  Generators without a Tilt Panel button     The front panel can be tilted for easier viewing  the generator does not have to be turned on    Gently pull the bottom of the front panel forward until it is at the desired angle  then release it   The front panel will lock into place  To lower the front panel  gently pull the front panel forward  slightly and then let it glide backward     4 6 Cleaning and Disinfecting the Generator    Pens    e The SmarTABLAT
29. 80 MHZ to calculated from the equation applicable to the  25 GHz frequency of the transmitter     Recommended separation distance   d 1 17VP 150 kHz to 80 MHz  d 1 17VP 80 MHz to 800 MHz  d 2 33 P 800 MHz to 2 5 GHz    where P is the maximum output power rating   of the transmitter in watts  W  according to   the transmitter manufacturer and d is the  recommended separation distance in meters  m      Field strengths from fixed RF transmitters  as  determined by an electromagnetic site survey      should be less than the compliance level in each  frequency range        Interference may occur in the vicinity of  equipment marked with the following symbol      4     Note 1  At 80 MHz and 800 MHZ  the higher frequency range applies     Note 2  These guidelines may not apply in all situations  Electromagnetic propagation is affected by  absorption and reflection from structures  objects  and people     a Field strengths from fixed transmitters  such as base stations for radio  cellular cordless  telephones  and land mobile radios  amateur radio  AM and FM radio broadcast  and TV broadcast cannot be  predicted theoretically with accuracy  To assess the electromagnetic environment due to fixed RF  transmitters  an electromagnetic site survey should be considered  If the measured field strength in  the location in which the SMARTABLATE    RF Generator is used exceeds the applicable RF compliance  level above  the SMARTABLATE    RF Generator should be observed to verify normal operati
30. Aa SS on N OE AIOE E E AONT 17   4 5 4 Potential Equalization        eesesesesesesesesssenenenenenennnnnnnnnnenennnnnenennnennnnnnnnnnnnnnnnnnnnnnnnnenenenenenennenne 18   4 5 5 Girc  lar Plug CONNEENONS    nennen 18   4 5 6 Connection Cables and Catheters            s ssesssessesseeseessessesseessesseessesseoseessesseoseeseessessesseeseess 18   4 5 7 Indifferent Electrode Connections  neun 18   458 Ethernet COMME ELON een rahenlseurn 19   A39 NSB OM CIO Aes ee ee 19   4 5 10 Serial Data CONNECTIONS      usesesesesesesenenenenenenenenenenenenenenennenenenenenenenenenenennenenennnnnenenensnnenenenene 19   ASIT FIDEROBUCCONDECHOm Lo een ee ee 19   4 5 12 Setting Up Irrigated Catheters with the Irrigation Pump and Tubing                    19   4 5 13 Tilting the Generator   s Front Panel  Generators with a Tilt Panel button             19   4 5 14 Tilting the Generator   s Front Panel  Generators without a Tilt Panel button     20   4 6 Cleaning and Disinfecting the Generator             cscsscssessssccscssessessesseseesssssssessessessesessessesseassesseaeens 20   A7 Protecting Against Damage tate rosea scncaseaseuschtaceats eaeshsnchsacvstesteancschsansasesccheagestuniesecschtace lt tisaeseacentartenegeds 20   Ao PEDO 7  cape sz ceeccceccctec A E eierarsiaversceieneisedeostneint 21    RF Generator User Manual       5 ACCESSORIES 22  5 1 Diagram of the Generator and ACCeSSOrieS          ssesseessseesssesssessseesseesssessseesseesserosseessseenserssersseeossee 22  FCS OL een ee 
31. D Many screens This icon appears when the system is  busy     6 Operation  Icons on the Screens   Page 63    RF Generator User Manual       7 Troubleshooting    7 1 Device Connections    The CONNECTIONS screen  see Section 6 2  Connections  displays 9 to indicate that a device    is properly connected and bd to indicate a problem with the connection  Table 7 describes the  specific meaning of these icons for each device     Table 7   Device Connections    Device Meaning of Icons  Pump Communication between the generator and the SMARTABLATE    Irrigation Pump was    established and is functional  The pump is now controlled by the generator     1 Possible Reasons What To Do    The pump is not connected to the Connect the pump to the generator     generator    The pump is powered off  Turn on the power to the pump    There are communication Replace the communication cable between   problems between the generator the generator and the pump    and the pump    The pump has a software version Check the software version on the pump   that is not compatible with the and on the generator   See Table 3   Device   generator   s software version  Screens in Section 6 2   Then consult Customer  Support     There may bea problem with the See the SmartABLaTe    Irrigation Pump User  pump  for example  the pump Manual    door may be open or there may   be a bubble in the irrigation   tubing     7 Troubleshooting   Page 64    RF Generator User Manual       Table 7   Device Connections    Devic
32. E    RF Generator must be turned off and the mains cable must be  disconnected from the generator before cleaning and disinfecting the generator     e The SMART  BLATE   RF Generator  mains cable  grounding cable  serial connection cable  and  the indifferent electrode cable must not be sterilized  Note  Some accessories may need  sterilization  follow the instructions for use that are provided with the accessories     Use a dry microfiber cloth to clean the generator   s touch screen  If damp cleaning is necessary   moisten a soft cotton or linen cloth with commercial glass cleaner that does not contain alcohol  and wipe the screen  Do not spray the cleaner directly onto the screen     Use a soft  damp  lint free cloth to clean and disinfect the housing of the generator  Use only non   flammable and non explosive substances  Water or soapy water is recommended for cleaning   Make sure that no liquid penetrates the inside of the generator     The following disinfectants may be used for disinfecting the generator and its cables     e Lysoformin   spezial   e Kodan     e Meliseptol     e Commercial disinfectants that do not contain ethyl alcohol    Caution  The following substances must not be used  acetone  benzene  acids of any kind   including acetic acid and citric acid   scouring agents  nitro dilutions and other organic solvents   Do not use any agent containing ethyl alcohol to clean the generator  lodine or disinfectants  containing dyes cause discoloration of the housi
33. Not valid for  versions JP and US          D  z      Q          Q         O  0 50 100 150 200 250 300 350 400 450 500  Load Impedance  ohms   Maximum Output Voltage vs  Output Control Setting  Load Impedance      ohm  300    250  200   gt      150  5  2  100      Limit of  versions  JP and US v  50 rms         100 110    sel Control Me  W        8 Technical Data   Page 88    RF Generator User Manual       Output Power vs  Output Control Setting  Load Impedance 100   200 ohm    W     Output Power                 T of          gt    Tolerance max   versions    Nominal    JP and US  e Tolerance min     I HE HE m    Control Be  W           8 7 Safety Inspections    To ensure the safety of the SMART  BLATE   RF Generator  inspect the generator before each use   Perform the following verifications     e Verify that all of the visual and audible indicators available on the generator are functional   see Section 6 1 1  System Startup      e Verify that preventive maintenance for the SmarRTABLATE    RF Generator is not overdue  see the  description of the generator screen in Table 3   Device Screens in Section 6 2  Connections      e Verify that the Data Entry Knob O is not loose and that the values on the screen change as  expected when the knob is turned     e Verify that the foot pedal is not damaged and that it functions correctly  see the description  of the foot pedal screen in Table 3   Device Screens in Section 6 2  Connections      e Verify that the connections to all accesso
34. OLINA T CTRL 40DEG    ZHANG 20W    ThermoCool      ThermoCool   SF       Provides access to selecting  editing  creating  and deleting  presets     Provides access to the Irrigation Control screen  the Audio  Settings screen  and the Display Settings screen     ABLATION 6 4  Ablation         When the ABLATION screen is displayed  you can press  Ablation Start    and Stop or the foot pedal to start or  stop ablation       The screen displays ablation measurement information  and provides the ability to change common settings while  viewing the ABLATION screen     When the Navigation Bar is displayed with the ABLATION  screen  before or after ablation   you can press Edit on the  Navigation Bar to change additional settings  You can press  Irrigation Control to modify irrigation settings for an  irrigated catheter  You can press System Control to  change the volume or the brightness of the screen and to  indicate whether the generator or remote control has control  of the system     When ablation stops  the Ablation Summary appears     6 Operation  Quick Overview   Page 30    RF Generator User Manual       Table 1   Screens  Screen Description  POST    Section    6 5  Post  Procedure  1 Procedure  Jate  06 September 2010    Total Ablation Time  00 03 13        The Max  Esophageal temperature shown above is not applicable  for the US and JP versions of the generator      Displays the POST  Post Procedure  screen     6 1 4 Screen Features    Sample Screen  PRESET Screen     Butto
35. SMART    RF Generator  User Manual    SAG_UM_04 00_EN    Manufactured by Distributed by  Stockert GmbH gd  Biosense Webster   a fohwonGohwren company       RF Generator User Manual       Table of Contents   PREFACE IV  1 FUNCTIONAL PRINCIPLES 1  1 1 Functional Principles of RF Ablation reed 1   1 2 Power Controlled Ablation Method  nennen 2   1 3  Temperature Controlled Ablation Method an    cc essesesesesesesesesesssssssesssssesssssesssssessssssssssesseseeees 2   1 4 Unipolar Application Method ersehen ee 3   1 5 Measuring Impedance ernennen 3   2 INDICATIONS 5  3 WARNINGS AND PRECAUTIONS 6  31 Warnings  Generator Usage arena innen ne ehe statuses ae aaa 6   3 2 Warnings  System Safety and CONNECTIONGS          ccccscssssssssessssssecsssssscsssssecssssscsssesscassscassesscssseesessess 9   3 3 Warnings and Precautions  During an Ablation Procedure     uusssssssssssesseesesenenenenenenenenenenenene 10   3 4 Handling Indifferent Electrodes 0    esessssssessssssscsssscscsesessescsesscsescsesesesseseseseseeseseseseescacseesesesesessess 11   4 SYSTEM OVERVIEW AND SETUP 14  4 1 Structure and Features ou    esessssssssssesssscscscscscsessssssssesssescsesescscseseseseseseacacsesesssesseseseseseacacecsseseseseseacacs 14   42    Controls on the Front of the Genese 15   AS Controls on the Back of the Generator ansehe 16   4 4 Functional Test and On Site Training euere 17   Ia OU ea A A A EA A EE EE T 17  4 5 1 Setting Up the Generator sssss EHE 17   452 Malhs COIS TOM esc een cent 17   
36. SMARTABLATE    RF Generator is intended for use in the electromagnetic environment specified  below  The customer or user of the SMARTABLATE    RF Generator should ensure that it is used in such an    environment     Immunity Test    Electrostatic  discharge  ESD   IEC 61000 4 2    Electrical fast  transient burst  IEC 61000 4 4    Surge  IEC 61000 4 5    Voltage   dips  short  interruptions  and voltage  variations on  power supply  input lines  IEC 61000 4 11    Power frequency   50 60 Hz   magnetic field  IEC 61000 4 8    IEC 60601  Test Level      6 kV contact     8 kV air      2 kV for power  supply lines      1 kV differential  mode     2 kV common  mode     lt  5  U     gt  95  dip in U   for 0 5 cycle    40  U    60  dip in U    for 5 cycles    70  U    30  dip in U    for 25 cycles     lt  5  U      gt  95  dip in U   for 5 sec    3 A m    Compliance  Level      6 kV contact      8 kV air      2 kV for power  supply lines      1 kV differential  mode      2 kV common  mode     lt  5  U     gt  95  dip in U   for 0 5 cycle    40  U    60  dip in U    for 5 cycles    70  U    30  dip in U    for 25 cycles     lt  5  U     gt  95  dip in U   for 5 sec    3 A m    Electromagnetic Environment   Guidance    Floors should be wood  concrete  or ceramic  tile  If floors are covered with synthetic  material  the relative humidity should be at  least 30      Mains power quality should be that of a  typical commercial or hospital environment     Mains power quality should be tha
37. Visual  S   Sound  aie      See the Troubleshooting section    in the user manual for the pump       Ifthe problem persists  contact    Customer Support     S   Ablation stop tone      If an irrigated catheter is being    used  connect the pump to the  generator       Ifa non irrigated catheter is being    used  select an appropriate preset       Disconnect the remote control    and operate the generator  directly from the generator   s user  interface       To resume ablation  press    Ablation Start or press and  hold the foot pedal     S none       RF Generator User Manual       Table 8   Warnings  Alerts  and Ablation Stop Reasons    Message    Reasons    Stop button   Remote      The Stop    button on  the remote control was  pressed     System overheating      The internal hardware  temperature of the  system is too high     Timer completed      The ablation time has  reached the set value on  the ABLATION screen     Voltage output too high      The setup is causing  the generator to deliver  higher voltage than it  was designed for     Indicators   V   Visual  S   Sound     Ablation Stop Reason    V  A and Ablation Summary    S none    Alert    Vo A and Alarm indicator turns  on orange    S   none  Ablation Stop Reason  V   Ablation Summary    S   Ablation stop tone    Ablation Stop Reason    V  A and Ablation Summary    S none    Ablation Stop Reason    V  A and Ablation Summary  and catheter receptacle flashes red  for 5 seconds    S   Ablation stop tone    What To D
38. a service center authorized by the manufacturer  The warranty is voided   if improper modifications or repairs are performed or if the device is not maintained according  to the instructions in this user manual  The warranty is not extended or renewed by services  performed under warranty     10 3 Product Liability  Product liability is assumed by STOCKERT GmbH under the following conditions     e The generator has been operated only with the original accessories provided by or  recommended by the manufacturer  see Chapter 5  Accessories      e Repairs to the generator and to the accessories have been performed only by the  manufacturer  a service engineer authorized by the manufacturer  or by a service center  authorized by the manufacturer     10 Warranty   Page 100    RF Generator User Manual SMART       11 References    Dorwarth U   Mattke S   Muller D   Hoffmann E   Steinbeck G   1993   Temperaturkontrollierte  Raidofrequenz Katheterablation  Vergleich Mono  und Bipolarer Nekrosebildung  Z Kardiol 82   Suppli   12    Hoffmann E   Bach P   Haberl R   Mattke S   Steinbeck G   1992   Antiarrhythmische und  Pro arrhythmische Effekte einer Oralen Cibenzolintherapie bei Anhaltenden Ventrikul  ren  Tachykardien  Z Kardiol 81  389 393    Hoffmann E   Dainat C   Mattke S   Dorwarth U   Steinbeck G   1989   Optimized Radiofrequency  Catheter Ablation by Temperature Controlled Bipolar Asymmetrical Coagulation  Eur Heart J 10   Suppl   221    Hoffmann E   Dorwarth U   Pulter R   Gok
39. ances  temperature  measurement may not prevent overheating  The temperature indicator is a guide  not an  absolute measurement     1 Functional Principles   Page 2    RF Generator User Manual SMART       The difference between actual tissue temperature and the temperature measured by the  integrated temperature sensor  moreover  depends greatly on the relevant catheter technology   When standard catheters that are not irrigated are used  the deviation in temperature is  proportional to the electrode size  due to passive cooling by blood flow   The larger the ablation  electrode  the larger the temperature difference  Regardless of the electrode used  the actual  tissue temperature is the same as or higher than the indicated temperature  When an actively  cooled ablation electrode is used  for example  an irrigated electrode cooled with saline solution    the electrode no longer provides a basis for accurate tissue temperature  In this case  the  temperature measurement is an indicator of the irrigation function and is completely dissociated  from the tissue temperature     1 4 Unipolar Application Method    To perform ablation with RF current  a closed circuit is required  The RF current flows from the  generator via the connection cable to the catheter   s ablation electrode  tip electrode   The  current then passes through the biological tissue and is passed back to the generator through  a second electrode  indifferent electrode   In order for the tissue heating to occur 
40. at only the  interface between the ablation electrode and the tissue  the second electrode must have a  considerably larger surface area than the ablation electrode  This arrangement is known as a  unipolar application     Because the current density is highest at the smaller ablation electrode  most of the RF energy is  converted into heat in the area of this electrode  The second  considerably larger RF electrode  is  preferably a skin adhesive electrode  This second electrode is known as an indifferent electrode  or a neutral electrode  For safety reasons  care must be taken to ensure that the area of the  indifferent electrode is considerably larger than the area of the ablation electrode  If the ratio  between the electrodes is too small  severe skin burns can occur at the site of the indifferent  electrode  With this type of application  the information provided by the manufacturer of the  indifferent electrode and the information in Chapter 3  Warnings and Precautions  must be  heeded     1 5 Measuring Impedance    Measurement of tissue impedance  the electrical resistance in the tissue  is used for two  purposes     1  To differentiate between the various types of tissue so that the ablation electrode can be  positioned correctly     2  To confirm that a lesion is being formed during RF current application     The SMARTABLATE    RF Generator measures impedance between the surface of the ablation  electrode and the indifferent electrode     1 Functional Principles   Pa
41. ated with that name  see Section 6 3 9  Adding a Name to the  Physician Presets List      6 Operation  Presets   Page 46    RF Generator User Manual       6 3 9 Adding a Name to the Physician Presets List    Quick Step   Create a new preset containing the physician   s name  see Section 6 3 5  Creating a    New Preset      Note   When you create a new preset  you are prompted to select an existing physician   s  name or type a new physician s name  If there are fewer than 10 physician names  in the Physician Presets list when you type a new physician   s name  that name  will appear automatically in the list the next time that you press the button in the  Physician Presets pane  see Section 6 3 8  Displaying One Physician   s Presets      6 3 10 Irrigation Control    Irrigation Control Screen    Monitor Low Flow N    High Flow   gt  gt   17  30 mL min  Low Flow   gt   2 mL min  Pre RF Time    Post RF Time       Monitor Fluid f    Low Fluid Warning mL    1000  mL    Volume Remaining 8 5mL    Quick Steps   1  Display the Irrigation Control Screen  Press the PRESET workflow button     Select an irrigated catheter type     Select a preset to edit   Press Edit     Press the gray rectangle in the Irrigation Control Settings pane     or    Press the ABLATION workflow button     Press the Irrigation Control  gt  icon on the Navigation Bar     2  Change Settings    Use the Data Entry Knob O to change values  see Section 6 1 5  Data Entry    Knob      Press OK ey     See Table 6   Irri
42. ator User Manual       Table 6   Irrigation Control Settings  Setting Description  Flush Press Flush  When a message appears  press and hold the Data Entry Knob O    to flush air bubbles from the irrigation tubing and the catheter  Use this function  carefully  For details  see Section 6 6  Flushing to Eliminate Air in the Tubing and    Catheter   START Low Flow Press START Low Flow to start the pump rotation at the Low Flow  standby flow    visible only when rate   the pump is stopped   STOP Press STOP to stop the pump      visible only when  the pump is running     6 3 11 Changing Audio Settings    Quick Steps     Press the PRESET workflow button     Press Edit       Press Audio Settings D      Select the options that you want       Press OK 07     Note   This feature is not available in some software versions     6 3 12 Selecting Ablation Display Settings    Display Settings Screen  Impedance  0  v i     Voltage  v  M  999 999 999 999 99  f 99 MA 99 mL 99  E  C    Eoo  BE z    Irrigation  delivered  mL  M    1 ee  ae see ee i    Grid Lines M All Values White M  Setpoint Lines M Impedance 4  i Separate    or   Ex        The Esophageal    C  selection shown above  is not applicable for the US and JP versions of the generator      6 Operation  Presets   Page 49    RF Generator User Manual       Quick Steps     Press the PRESET workflow button     Press Edit     Ir      Select the options that you want       Press OK ey     Notes 1  The options that you select determine what i
43. ator User Manual       Table 8   Warnings  Alerts  and Ablation Stop Reasons       Incompatible catheter cable      The catheter connection Alert   If using an incompatible catheter  cables have an electrical V  A and Alarm indicator turns connection cable  replace it  identification component on orange with a cable that was provided  that the system uses to by or recommended by the  verify that an approved S   none manufacturer     type of cable is being  used  This alert appears  when the system   does not detect the  identification data   Possible reasons are     If using a compatible catheter  connection cable  replace it with a  new compatible connection cable     If the problem persists  contact  Customer Support       The cable was not  provided by or  recommended by the  manufacturer       The cable was  damaged during  re sterilization       The cable is faulty       There is a system  problem     7 Troubleshooting   Page 77    RF Generator User Manual       Table 8   Warnings  Alerts  and Ablation Stop Reasons    Message Indicators    Reasons  V   Visual  S   Sound     What To Do    Incompatible esophageal probe cable  not applicable for the US and JP versions ofthe generator       The esophageal probe Alert  connection cables have V  A and Alarm indicator turns  an electrical identification on orange  component that the  S   none    system uses to verify  that an approved type of  cable is being used  This  alert appears when the  system does not detect  the identi
44. can endanger the patient   s  life  Make sure that the leakage current of the connected system  in any combination  never  exceeds the maximum permissible value  patient leakage current  lt  0 05 mA      Take into account that the system   s RF leakage current can be negatively influenced by other  system components  The maximum allowable values are specified in IEC EN 60601 2 2   Displacements of the ECG baseline surface signals is a sign of uncontrolled RF current  leakage through the body surface ECG electrodes  Heating of these electrodes gives rise to   a shift in direct current voltage which  in turn  leads to a displacement of the baseline  In this  case  check the whole system combination to avoid uncontrolled current leakage  It may  also be the case  however  that the electrocardiograph is unsuitable for this use or that the  body surface ECG electrodes have too high a contact impedance  Make sure that the ECG  device is appropriate for this application  Skin burns at the body surface ECG electrodes can  be a sign of uncontrolled current leakage     7  There are many electrocardiographs to which a stimulator can be directly connected  The  stimulator must be galvanically isolated  or physically disconnected  before the RF current is  turned on  The reason for this is the parallel connection between the output of the generator  and the stimulator  If the connection is not isolated  the stimulator   s patient interface and  safety insulation may be damaged  The transf
45. ching it  even when wearing surgical gloves  or by tapping it with a pen     The generator features redundant protection circuits  if one safety mechanism fails  a second  safety mechanism indicates the failure to ensure safety   A central computer processor operates  the device  Several other microprocessors control the various functions  such as RF power  output and calculation of impedance  The generator is equipped with a memory unit that stores  the settings even after the generator has been turned off  These settings are active when the  generator is turned on again  Operating errors and defects in the attached devices are detected  and error messages are issued  For a list of possible error messages  see Sections 7 2  Warnings   Alerts  and Ablation Stop Reasons and 7 3  System Errors     ET  4 System Overview and Setup   Page 14    RF Generator User Manual       4 2 Controls on the Front of the Generator          Item    1    WM  P  WIN    O     10    12  13    Description   RF On indicator   Ablation Start button   Stop button   Catheter 1   receptacle for ablation catheter cable    Catheter 2   receptacle for esophageal probe cable  not functional in the US  and JP versions of the generator     Receptacles for indifferent electrode cables  Tilt Panel button  on side of front panel   not available on some models     Receptacle for ECG cable  external stimulator   electrophysiology recording system     Touch screen  Alarm indicator  Stand by button  Stand by indicator
46. communication  through the recording system port    see Item 3 in Section 4 3  Controls on  the Back of the Generator   The color   in the arrow moves from left to right   to indicate that the test is running  The  numbers on the screen are the values  being sent to the recording system     Press STOP to stop the transmission test     Press Flush  When a message appears   press and hold the Data Entry Knob    O to flush air bubbles from the  irrigation tubing and the catheter  Use  this function carefully  For details  see  Section 6 1 5  Data Entry Knob     Press START Low Flow to start the  pump flowing at the low  standby   flow rate  see Section 6 3 10  Irrigation  Control      Press STOP to stop the pump     6 Operation  Connections   Page 36    RF Generator User Manual       Device    Remote    Foot Pedal    Indifferent  Electrode    Esophageal  Probe  not  applicable for  the US and JP  versions ofthe  generator     Catheter    Table 3   Device Screens    Information Displayed      Device name      Serial number  or dashes  if the remote control is  not connected       Image of foot pedal      Green bar and  number s  indicating  impedance if indifferent  electrode is connected  and is functioning  properly     Red bar and dashes if  indifferent electrode is  not connected      Red bar and number s   if split indifferent  electrode is connected  but contact quality is  below the threshold      Temperature of  esophageal probe  or  dashes if there is no  probe attached 
47. d all  new data is logged as part of the new procedure  The POST screen summarizes  the activities  see Section 6 5  Post Procedure      2  Ifthe generator is off for more than 6 hours  a new procedure starts  automatically when you turn on the generator and start ablation     6 4 2 Starting and Stopping Ablation    Quick Steps     Press the ABLATION workflow button       To start ablation  press Ablation Start    on the front of the generator or  remote control  or press and hold the foot pedal        To stop ablation  press Stop    on the front of the generator or remote control   or release the foot pedal      Notes   1  Either the generator or the remote control can be the system Master  see  Section 6 7  System Control   When the remote control is the system Master   ablation can still be started from the generator     2  When you start ablation  the RF On indicator on the generator and on the  remote control light up and a continuous tone is emitted  These turn off when  you stop ablation     3  Ifa Pre RF time has been set for fluid control  RF energy will not start until after  the fluid has flowed for the Pre RF amount of time  see Section 6 3 10  Irrigation  Control      4  lf ablation does not stop when you press Stop     try any of these alternatives     If you pressed Stop    on the generator  press Stop    on the remote  control     If you pressed Stop on the remote control  press Stop on the  generator    Press and release the foot pedal on the generator    Pre
48. d external  devices     S   Ablation stop tone      There is an internal  problem in the  generator     For a dual sensor    DS  catheter  the  difference between the  measurements of the  two sensors is too high   This may be caused   by a problem with the  catheter or the cables     Catheter temperature warning      The temperature    measured by the catheter    may have exceeded the  warning value set in the  preset     Current output too high      The generator was trying    to deliver a current  output that was beyond  its intended capability     Warning    V  A appears near temperature  reading and its orange border  flashes    S   Warning tone    Ablation Stop Reason    V  A and Ablation Summary    and catheter receptacle flashes red    for 5 seconds    S   Ablation stop tone    7 Troubleshooting   Page 71    Reduce the power setting used for  the ablation or stop the ablation     If it is clinically appropriate   modify the temperature warning  value for the preset that was  used   See Section 6 3 3  Editing a  Preset        See the instructions in this table    for Power output too high        RF Generator User Manual       Table 8   Warnings  Alerts  and Ablation Stop Reasons    Message Indicators    What To D    Reasons  V   Visual  S   Sound  Se    Esoph  temperature above maximum  not applicable for the US and JP versions of the generator       The esophageal probe Alert   Wait a few seconds for the  temperature exceeded V  A and Alarm indicator turns temperature
49. d on the back of the generator  see Item 2 in Section 4 3   Controls on the Back of the Generator      2  Ifyou press Today  all data from the current day will be exported  If you press  Last 7 Days  all data from the last 7 days will be exported  If you press All  all  data on the generator will be exported  These buttons change from blue to  gray during the export process     3  A light green dot appears on the export status line and continually gets darker  as the export process progresses  It changes to when the export status is    100  complete  It changes to if you press Cancel export or if the export  fails  Any files that have already been exported to the USB flash drive  will  remain on the flash drive  Any file that was only partially exported  will be  completely exported the next time the export process is started     4  The export process may take minutes or more than an hour depending on the  amount of data     6 Operation  Post Procedure   Page 57    RF Generator User Manual       6 6 Flushing to Eliminate Air in the Tubing and Catheter    mn    e Irrigated catheters must be flushed before they are inserted in the patient     e When flushing a catheter  the catheter must not be inside the patient  When flushing the  SMARTABLATE     Irrigation Tubing Set  either the catheter must be outside the patient or the  4 way stopcock on the tubing set must be turned so that the fluid is not routed to the  catheter     Quick Steps     Ensure that the catheter is not in th
50. de or a  different cable     RF Generator User Manual       Table 8   Warnings  Alerts  and Ablation Stop Reasons    Message    Reasons    Insufficient irrigation      High  ablation  flow was  stopped  Possible reasons  are       The Stop    button on  the pump was pressed       The Low  Standby     Flow button on    the pump was pressed       The foot pedal on the  pump was released     Low fluid      The generator has  calculated that the  amount of fluid in the  irrigation bag is low     Measurement integrity error      The generator has  two independent  measurement channels  for critical parameters   this message will appear  if the two channels  do not agree  Possible  reasons are       The cable is faulty       There is a compatibility  problem between the  generator and external  devices       There is an internal  problem in the  generator     Indicators   V   Visual  S   Sound     Ablation Stop Reason    V  A and Ablation Summary    S   Ablation stop tone    Warning    V  A and orange border around  Irrigation Control icon flashes for  20 seconds    S   Warning tone    Alert    V  A and Alarm indicator turns  on orange    S   none  Ablation Stop Reason  V   Ablation Summary    S   Ablation stop tone    What To Do      Restart the flow       Check the fluid level  If it is low   change the bag and after the  ablation  update the fluid bag size  on the Irrigation Control screen    See Section 6 3 10  Irrigation  Control        Disconnect any devices connected  to the
51. des  and electrodes of low quality can impair contact quality  and increase the risk of a skin burn     e Heat produced by thermal blankets or other sources of heat adds to the heat arising under  the indifferent electrode  This increases the risk of skin burns  Do not use such sources of heat  in immediate proximity to the indifferent electrode     Feen    An unsuitable indifferent electrode or an incorrectly applied indifferent electrode can lead to  burns on the skin surface  Check the indifferent electrode and the connection cable before use   Do not use these if they are damaged or modified     3 Warnings and Precautions   Page 13    RF Generator User Manual SMART        YSTEM       4 System Overview and Setup    The SmarTABLaTE    RF Generator is a radiofrequency generator designed for use in cardiac  ablation procedures  The generator can be configured to work with a variety of other devices and  meets the highest standards for safety  precision  and ease of operation     u recording    System    LAR TOR 3    Be Remote  System    en     FF    Catheter    Foot Fedal    Esophageal Frobe    Connections  gt              The Esophageal Probe shown above is not applicable for the US and JP versions of the generator      The generator is compatible with a large number of catheters and devices  For a listing of devices  and accessories  see Section 5 2  Accessory Lists     The generator is operated via an easy to use user interface  Selections can be made on the screen  by tou
52. e   Do not use any contact gel with single use electrodes  Use the single use electrode for  only one procedure  If the indifferent electrode becomes loose or needs to be moved  use a  new indifferent electrode  Repeated use of an already applied electrode can mean a loss of  adhesiveness and thus lead to distinctly poorer contact quality     e Forareusable electrode  Make sure that the electrode contact surface is not dry  If it is dry   use a small  evenly distributed quantity of conductive gel     Kr                        3 Warnings and Precautions   Page 12    RF Generator User Manual       e Carefully select the contact area  Choose a muscular area on the back of the patient that is  as near the heart as possible and that has sufficient blood flow  Do not place the indifferent  electrode near wounds     e Carefully prepare the contact area  Prepare the contact area on the patient and on the  indifferent electrode according to the instructions for use provided with the indifferent  electrode  Place the indifferent electrode closer to the ablation site than any ECG electrode or  other products that could represent an alternative lead  Shave and degrease the skin     e Carefully apply the indifferent electrode to the contact area  Make sure that the whole  surface of the indifferent electrode forms a closed contact with the patient   s back  There  should be no pockets of air between the skin and the indifferent electrode  Oil  hair  dirt   repeatedly used adhesive electro
53. e  Brightness  and FOOT Pedal TeSticsscscssscovesssessasscsconessnansncovssceseisescisssssessdessavcanensosveseecsesesssss 60   89  FOwcChart othe SCLC SING nen ee eh 61  6 10  Tons On LS CS een are 62   7 TROUBLESHOOTING 64  Al DOCE OM SCIONS senn r ER E E EE E E 64   7 2 Warnings  Alerts  and Ablation Stop Reasons           cscesssesesssscssscsesessssssssesesescscscsesesssssessescacscsesees 67   13 Sy ENON porra R 83  Pek VOM EN OT gen a E errs rea E ere rer 83   732 FP E aa AA E seeanaen 83   74A Other Troubleshooting sera nenne 84   7 93 Calibrating the TOuUcHSereen een 84   8 TECHNICAL DATA 85  3 1 SIDS CEM CUO 1  Seien een ee ee O E es 85   82 Storage and Shipping    ONG One eareensenane 86   8 gt  Operating Condi nennen ernennen 86   8 4 THERMOCOOL   Catheter Preset RANGES       cscsssessessssessessessssessessssesssssssesssssssessessssessessssssaesecsesseeseeees 86   85 Alarm and Touch Specifications        e ssesseseeessessesseessessesseeseesseessensesseessesseoseeseesseoseoseesseeseesesseeseeseesesss 87   39 OUUU PEE OE nee een een 88   er Salat  LASSE OIG ee E ENSE S AE AES 89   88 MONONA O E E E r E ORE 90   o R O r A A E E E AS 90  8 10 Conformance to Standards    na 91   8 11 EMC Information and Technical Description        eseseseseseseesesesesenenenenenenenenennnnnnnnnnenenennnnnnnnnnnnenenene 92   9 SYMBOLS 97  10 WARRANTY 100  101 Transit Damage era 100   1O N enian EE een ee ns 100   103 Produc OaD co seer arroen a SEE ET A AAAA 100   11 REFERENCES 101    RF
54. e Meaning of Icons    Remote    Communication between the generator and the SMART  BLATE   Remote Control was    established and is functional   Possible Reasons    The remote control is not  connected to the generator     The remote control is powered  off     There are communication  problems between the generator  and the remote control     The communication cable  between the generator and the  remote control is damaged     The remote control has a software  version that is not compatible  with the generator s software  version     What To Do    Connect the remote control to the generator     Turn on the power to the remote control     Replace the communication cable between  the generator and the remote control     Replace the communication cable between  the generator and the remote control     The software version on the remote control  needs to be updated  Contact Customer  Support     Foot Pedal The foot pedal is connected to the generator and has passed the foot pedal test     1 Possible Reasons    The foot pedal is not connected  to the generator  or has become  disconnected      The foot pedal test has not been  performed     The foot pedal has failed the foot  pedal test     The foot pedal automatic ID  check failed     What To Do    Connect the foot pedal to the generator   Perform the foot pedal test   See Table 3    Device Screens in Section 6 2 or Section 6 8   Volume  Brightness  and Foot Pedal Test      Press the image of the foot pedal and perform  the foo
55. e patient  or that the 4 way stopcock on the  tubing is turned so fluid is not routed to the catheter        Open the Pump device screen or the Irrigation Control screen      Press Flush on the screen or Flush on the pump      When a message appears  press and hold the Data Entry Knob      Release the Data Entry Knob       Notes   1  The Flush button is available on the generator   s Pump device screen  see Table  3   Device Screens in Section 6 2  and on the Irrigation Control screen  see  Section 6 3 10  Irrigation Control      2  Toensure that all bubbles have been flushed from the catheter or tubing set   flush for at least 5 additional seconds after an image appears on the pump   s  display indicating that no bubbles were detected and after there are no visible  bubbles in the tubing  see the instructions for use provided with the pump      3  Opening the door ofthe pump can move the tubing slightly  causing the  bubbles to move out of the bubble sensor area  Inspect the tubing carefully     6 Operation  Flushing to Eliminate Air in the Tubing and Catheter   Page 58    RF Generator User Manual       6 7 System Control    When you first turn on the generator  the generator is the system Master and the remote control  is the Monitor  The Master controls all system functions  except the volume  brightness  and foot  pedal testing of the Monitor   At the remote control  you can request that it become the Master   the generator then becomes the Monitor   When you want the gen
56. e stimulator are galvanically  connected with the ablation catheter  Only cables and plugs approved for a CF type of  device  CF safety classification  may be used  see Section 5 2  Accessory Lists      2  All devices with these cables and plugs must fulfill the requirements stated in MDD MPG for  class 1 devices  Fulfillment of the requirements stated in IEC EN 60601 1 1 and MDD 93 42   EEC must be monitored and documented in a suitable way     3  The person who connects the generator and accessories to each other or who uses the  generator and accessories is responsible and liable for installation and operation that  complies with IEC EN 60601 1 1     4  All system components must comply with all applicable requirements and standards and be  labeled pursuant to these standards     5  If there are any concerns regarding the use of a component  contact the distributor of the  component to obtain further information     6  If several devices are connected to the generator and to each other  they should be as safe   both individually and together  as specified in IEC EN 60601 1 and its sub standards and  IEC EN 60601 1 2  All devices and accessories  when located within the patient area  must  comply with IEC EN 60601 1 and its sub standards     3 Warnings and Precautions   Page 9    RF Generator User Manual SMART       Keep in mind that the ECG connection cable for the generator creates a direct electrical  connection to the patient   s heart  Incorrect use of this connection 
57. e tubing   See Section 6 6     Flushing to Eliminate Air in the  Tubing and Catheter        Resume low  standby  flow       Wait a few seconds for the    temperature to drop before  restarting ablation       If the problem persists  modify the    temperature cut off value in the  preset that was used   See Section  6 3 3  Editing a Preset        Consider possible clinical reasons    for the ablation electrode to  overheat       If the problem persists  contact    Customer Support     RF Generator User Manual       Table 8   Warnings  Alerts  and Ablation Stop Reasons    Message  Reasons    Indicators   V   Visual  S   Sound     Catheter temperature change      The catheter temperature  changed at a rate that  was greater than 20  degrees within 200  ms  This may occur  when there is electrical  interference between the  ablation electrode and  the temperature sensor     Ablation Stop Reason    V  A and Ablation Summary    for 5 seconds    S   Ablation stop tone    Catheter temperature invalid      The generator does Alert  not recognize a valid V  A and Alarm indicator turns  temperature reading  on orange  Possible reasons are   S none      A catheter is not  connected or is  connected to the  wrong receptacle     Ablation Stop Reason    V   Ablation Summary and  catheter receptacle flashes red for      The cable or catheter is  5 seconds    faulty       There isa compatibility      Ablation stop tone    problem between the  generator and external  devices       There is an 
58. ed from the generator  by performing the steps in Section    6 5 1  Exporting Ablation and Event Log Data to a USB Flash Drive     Alarm Tones and Alarm Indicator  Al  Light     See Sections 7 2  Warnings  Alerts  and Ablation Stop Reasons and 7 3  System Errors      Warning  Fatal   Ablation Stop Reason System Error System Error  Comment Information tone Information tone High priority alarm  Pulse Pattern ti ti ti ti ti ti ti  ti ti          ti ti ti  ti ti  Frequency 600 Hz 900 Hz 1000 Hz  Duration 600 ms 400 ms 4 25  Repetition None 20s 10s  Al Off On  orange Flashes  red  f   2Hz  duty   cycle 50   Note  All alarms are technical alarms   Signal Lights    See Section 7 2  Warnings  Alerts  and Ablation Stop Reasons    Indicator Color Frequency Duration  Indifferent electrode receptacles Blue 5 Hz 55  Catheter receptacle Red 5Hz 55  Esophageal probe receptacle  not functional Red 5 Hz 55    in the US or JP versions of the generator     Duration of warning    Border of System Control icon Orange 0 5 Hz    Border of Irrigation Control icon Orange 0 5 Hz    Duration of warning    Border of System Status icon Orange 0 5 Hz    Touch Screen Tones    Duration of warning    Duration    Event Frequency    Touch is recognized and button is valid 4 kHz    40 ms  20 ms    Touch is recognized but button or area is not valid 2 kHz    8 Technical Data   Page 87    RF Generator User Manual       8 6 Output Specifications    Output Power vs  Load Impedance    pf ff of ff ff  IN ee ee HE    
59. ee 23   ZI    eer ee ee ae esse en 23  522 TOPCO graces ce se cacy contacter E E A 24  523 SWARTABLEIE  Cable ee rer 24  5 2 4 Indifferent Electrodes and Cables available from STOCKERT GmbH                      26  323     CNS SMe een ae ee ee Teen 26  53 MIO CS 5 SS een 27  5 3 1 Connection Cables         ss sesesseesseesseesseessesseesseesseosseoseesseesseosseesseoseeoseroseenseoseesseesseoserosereseeoseeosee 27  3 32  naderen ISCO CGS cueir eana EEN sneunteswaarts 27  33 OO POGI eaa een 27  5 4 Cleaning  Disinfecting  and Sterilizing Accessories     esenssesesesenesensenenenenennenenenennnnenenenennenenenennene 27  6 GENERATOR OPERATION 28  Bl EOV IO aan ar ee ee 28  5 1   SYE U e E AE E E E 28  612 E E E a E E ee 29  SM ie   6           ee E 29  0 14     Scieen l  9      u a 31  6 15 Boll BAY KN0D  een 32  5 10 Foot Pedal een ee 32  6 1 7 Typical Procedure Process     Quick          csssssssssscsssecessssssssecscsssssssecscsssssesesscssseccsssesesecseens 33  6 1 8 Typical Procedure Process     Full aaa ana 33  5 2  ONEC DONS een ren 34  OS ass crear EBEN EEE esc case cme vacate ccs ELDER EERIEESNOEENEENEE OO ENCEULIERERREERRERIE DIENEN EIERN SPREGEEIER 39  081  PIESELOVEV Wesen ae er 39  632 SOUS CBAC a PIE ee ee 39  63 EAN a PrE O aE E anne 40  6 3 4 Editing a Preset     Temporary Save    44  035 CAEN eS OI ernennen een 45  6 3 6 Designating a Default Preset    ceesesscscsscesscsescsssssssscssssssesesssesessesssesseescseseesescseseseeseees 46  6 3 7 Deleting a Preset    
60. efore the generator is turned on  Beware of flammable endogenous gases  Materials such  as cotton and gauze  when saturated with oxygen  can be ignited by sparks that arise even  during normal use of the generator   The foot pedal is suitable for use in operating rooms      20  Be aware during ablation procedures that in patients with cardiac pacemakers or pacemaker  leads  there is a risk of interference with the pacemaker function or damage to the  pacemaker  In case of doubt  consult the manufacturer of the device     21  The electromagnetic radiation emitted by the generator can interfere with the function of  other electrical devices  Conversely  other electrical devices can influence the function of the  generator if they are operated at the same time in the immediate vicinity of the generator     22  The plugs on the catheter and the catheter connection cable must not be brought into  contact with parts that have high voltage  such as mains outlets  or metallic objects  This can  lead to the patient s death by electrocution     23  To avoid damage to the connection cables  do not wrap the cables around the generator or  other apparatus  Coiling the connection cables during normal operation of the generator  creates inductive components  which can lead to measurement errors  The values indicated  in error can lead to misinterpretations     24  Special safety measures with regard to electromagnetic compatibility  EMC  must be  taken with electrical medical devices  This 
61. el M   Steinbeck G   1992   Bedeutung Physikalischer  Parameter fur die Effektivitat der Radiofrequenz Katheterablation  Biomed Technik 37  62 68    Hoffmann E   Gerth A   Remp T   Muller D   Steinbeck G   1992   Linksventrikul  re Katheterablation  des AV Uberleitungssystems mit Radiofrequenzstrom  Z Kardiol 81  378 384    Hoffmann E   Haberl R   Pulter R   Dainat C   1987   Phase Displacement between Voltage und  Current during Radiofrequency Catheter Ablation  Circulation 76  Suppl IV   278  Abstract     Hoffmann E   Haberl R   Pulter R   Gokel M   Steinbeck G   1992   Biophysical Parameters of  Radiofrequency Catheter Ablation  Inter J Cardiol 37  213 222    Hoffmann E   Mattke S   Dorwarth U   et al   1989   Temperature Controlled Impedance Guided  Radiofrequency Catheter Ablation in Swine  Circulation 80  Suppl II   4  Abstract     Hoffmann E   Mattke S   Dorwarth U   Muller D   Haberl R   Steinbeck G   1993   Temperature   Controlled Radiofrequency Ablation with a New Electrode Catheter  First Clinical Experience  Eur  Heart J 14  57 64    Hoffmann E   Mattke S   Knecht M   et al   1990   Predictors of Successful Temperature Controlled  Radiofrequency AV Ablation  Eur Heart J 11  155  Abstract     Hoffmann E   M  ller D   Gerth A   Remp T   Steinbeck G   1992   Bedeutung des Impedanz   Monitoring wahrend Radiofrequenz Katheterablation  Z Kardiol 81  Suppl 1   142  Abstract     Hoffmann E   Muller D   Remp T   Gerth A   Steinbeck G   1992   Radiofrequenzablation der  La
62. equipment generates  uses  and can radiate  radiofrequency energy and  if not installed and used in accordance with this user manual   may cause harmful interference to radio communications  Similarly  portable and mobile  communication devices may cause harmful interference to the functioning of the generator     25  If error messages repeatedly appear  stop using the generator and contact Customer  Support     26  To avoid damage to the generator and its accessories  use only appropriate cleaning agents   see Sections 4 6  Cleaning and Disinfecting the Generator  and 5 4  Cleaning  Disinfecting   and Sterilizing Accessories      27  To avoid the risk of electric shock  connect the mains cable from the generator mains socket   see Item 6 in Section 4 3  Controls on the Back of the Generator  to a 3 prong outlet that  meets the specifications in Section 8 1  Specifications     28  The generator may be opened only by persons authorized by the manufacturer  When the  generator is open  parts that have high voltage or are very hot are accessible and can cause  injury  If the generator is opened by an unauthorized person  any claims on the warranty are  void  No modification of the generator is permitted     29  If fluid penetrates into the generator  stop using the generator and contact Customer  Support     3 Warnings and Precautions   Page 8    RF Generator User Manual SMART       30  Inthe unlikely event that there is a fatal system error  the system will stop and the Alarm  
63. er of RF energy into the electrodes connected  to the stimulator can also cause injury to the patient     8  Only classified electrical medical devices may be connected to the generator  If a PC system  does not fulfill the requirements of IEC EN 60601 1   UL 60601 1 and their sub standards   the distance between the PC system and the patient must be at least 1 83 m  6 ft  and the  PC system must fulfill the requirements of IEC 950   UL 60950 1  All medical devices that are  connected electrically to the generator must fulfill the requirements of IEC EN 60601 1 1   the standard for medical systems     3 3 Warnings and Precautions  During an Ablation Procedure    1  Blood vessel perforation is a risk inherent in the placement of an electrophysiology catheter   The catheter needs to be moved carefully to avoid damage or perforation of blood vessels     2  When performing ablation of the posterior wall of the left atrium  beware of the risk of  forming a lesion in the esophagus     3  Avoid high ablation temperatures  High ablation temperatures can lead to clot formation   charring of the heart tissue or blood  and or evaporation of interstitial intracellular fluid     Note  The temperature indicated on the generator is not the tissue temperature  The  indicated temperature is the temperature of the catheter   s ablation electrode  which does  not necessarily represent the tissue temperature  This applies especially when catheters with  saline cooled ablation electrodes are used
64. erator to take control again   make it the Master as described in this section     Note  Screen calibration also affects only the Master  see Section 7 5  Calibrating the Touch  Screen      System Control Screen  When Generator Is Monitor     Device status  Become  Master    MONITOR    Volume Brightness Foot Pedal Gen     Press to test    f COOUNEEOEE 4  Automatic      no RF energy will be    delivered    Manual        Net  y pressed    3 TEP    l j    Preset  THERMOCOOL   Standard       6 7 1 Making the Generator the System Master    Quick Steps     Press System Control WES on the Navigation Bar     Press Become Master     Notes   1  The Ablation Start    button on the generator and the foot pedal on the  generator can be used to start ablation regardless of whether the generator    is the Master or the Monitor  Similarly  the Stop button and the foot pedal  on the generator can be used to stop ablation  regardless of whether the  generator is the Master or the Monitor     2  When the remote control is the Monitor  it can only be used to stop ablation     6 Operation  System Control   Page 59    RF Generator User Manual       6 8 Volume  Brightness  and Foot Pedal Test    System Control Screen    Device status  Become  Master    MASTER    Volume Brightness Foot Pedal Gen     Press to test    ef COOUREEEEE 4  Automatic       no RF energy will be  delivered    Manual           Not      pressed    3 TT x al j       u   Quick Steps   Press System Control on the Navigation Bar to
65. es     At least once a year  verify that the SMARTABLATE    RF Generator and its accessories are installed  per Section 4 5  Setup     At least once a year perform the touch screen calibration per Section 7 5  Calibrating the Touch  Screen     The SmarTABLaTe    RF Generator contains measurement circuitry  To ensure that the generator  stays in compliance with its performance specifications  the generator   s calibration must be  maintained  Every other year  and any time that loss of accuracy is suspected   contact Customer  Support and schedule preventive maintenance that includes calibration and a full functional test   The calibration and the full functional test may be performed only by the manufacturer  a service  engineer certified by the manufacturer  or a service center certified by the manufacturer     At any time  if any problems are found  report them to Customer Support and send the generator  or broken accessory for repair     If desired  a service contract can be arranged  Contact Customer Support    89 Repairs  There are no parts in the SMARTABLATE    RF Generator that may be serviced by the user   Exception   The fuses may be replaced per Section 4 5 3  Fuses   Repairs or any changes to the generator and    the accessories must be performed only by the manufacturer  a service engineer certified by the  manufacturer  or a service center certified by the manufacturer     8 Technical Data   Page 90    RF Generator User Manual       8 10 Conformance to Standards 
66. etting Descriptions in this section        Use the Data Entry Knob O to change the value  see Section 6 1 5  Data Entry  Knob        Press Save O then press In selected preset   or  Press Escape Mis  then press No to exit the screen without changing the preset     Notes   1  If you don t see the preset that you want to edit  select a different catheter  type  see Section 6 3 2  Selecting a Preset  or display the presets for a different  physician  see Section 6 3 8  Displaying One Physician s Presets      2  The choices in each pane vary depending on the catheter type you select     3  Before making changes  select Temperature or Power in the Ablation Settings  pane to select Temperature Control Mode or Power Control Mode  The  selections in the panes vary depending on the mode      Temperature Control Mode is not available for all catheter types and is not  applicable for the US and JP versions of the generator      4  The lrrigation Control Settings pane appears only when you select an irrigated  catheter     5  When you press Save    a message appears asking whether to save In  selected preset or Save temporarily  Select In selected preset   See Section  6 3 4  Editing a Preset   Temporary Save  for information on Save temporarily      6  When you press Save O and then press In selected preset  the selected  preset is saved with the revised values  If you want to save the revisions as a    different preset  press Save As and follow the instructions in Section 6 3 5   Creat
67. fication data   Possible reasons are       The cable was not  provided by or  recommended by the  manufacturer       The cable was  damaged during  re sterilization       The cable is faulty       There is a system  problem     Indifferent electrode not detected    No indifferent electrode Alert  was detected  V  A and Alarm indicator turns    A split indifferent on orange  electrode is connected to  S   none    the generator but is not    applied to the patient  Ablation Stop Reason    V   Ablation Summary and both  indifferent electrode receptacles  flash blue for 5 seconds    S   Ablation stop tone    7 Troubleshooting   Page 78    If using an incompatible  esophageal probe connection  cable  replace it with a cable that  was provided by or recommended  by the manufacturer     If using a compatible esophageal  probe connection cable  replace  it with a new compatible  connection cable     If the problem persists  contact  Customer Support     If a split indifferent electrode is  connected to the generator  apply  it to the patient     Disconnect the indifferent  electrode from the generator   then reconnect it  Check all other  connections to the indifferent  electrode such as the connection  between the cable and the  electrode itself     Plug the indifferent electrode into  the other indifferent electrode  receptacle on the front of the  generator  Contact Customer  Support even if the problem is  resolved     If the problem persists  use a  different indifferent electro
68. g a  solid indifferent electrode  the hospital staff is responsible for ensuring proper contact of the  electrode with the patient s skin     13  The patient must not be in contact with grounded metal components or with metal  components that have a large grounded area  for example  the operating table supports    For this purpose  the use of sufficiently insulating antistatic covers on the operating table  is recommended  Electrostatic discharge  ESD  can give rise to extremely high current  densities at the catheter tip  which can injure the patient  Therefore  do not touch the pins  in the plug at the end of the catheter or the pins in the plug at the end of the cable after the  catheter has been placed in the patient s body     14  Skin contact between parts of the patient   s body  for example  between the arms and the  body  should not occur  Such contact can be avoided by using dry gauze  for example     15  Ifthe generator and physiological monitoring devices are used on a patient at the same  time  all monitoring electrodes without protective resistances or RF filters should be applied  to the patient s body as far away as possible from the ablation electrodes  Needle electrodes  are not recommended for monitoring purposes  In all cases  it is appropriate to use  monitoring electrodes and other monitoring devices that limit the RF current     16  Position the connection cables of the ablation electrodes in such a way that they do not  touch either the patient or othe
69. gation Control Settings in this section        6 Operation  Presets   Page 47    RF Generator User Manual    SMART       Setting    Monitor Low Flow    Use automatic flow  adjustment    High Flow    Low Flow    Pre RF Time    Post RF Time    Monitor Fluid    Low Fluid Warning    New Bag    Volume Remaining    Table 6   Irrigation Control Settings  Description    Select Monitor Low Flow if you want the system to warn you when either the  catheter is in the patient and the pump hasn t started working yet or the flow is  stopped     Select Use automatic flow adjustment if you want the rate at which fluid flows  during ablation to be automatically adjusted by the system     When Use automatic flow adjustment is selected  you will see    AUTO      automatic  next to the set flow rate on the ABLATION screen  When   Use automatic flow adjustment is not selected  you will see    MAN     manual   next to the set flow rate on the ABLATION screen   See Section 6 4  Ablation  for  a sample screen      Note  If the flow rate is changed manually during ablation as described in Section  6 4 3  Making Temporary Changes to Ablation Settings  all system controls over  the flow rate are disabled  The user is responsible for ensuring that the flow rate is  appropriate for the power setting     The rate at which fluid flows during Pre RF  Post RF  and ablation  If   Use automatic flow adjustment is selected  press High Flow to access the High  Flow screen  where you can set the flow for each powe
70. ge 3    RF Generator User Manual       RF heating of the tissue changes the impedance because of the heating and dehydration of the  tissue  During the increase in tissue temperature  the tissue impedance first falls to a minimum  value and remains at this level for some time  Then the impedance rises based on the supplied  power  see Figure 4   Impedance should be monitored during ablation  A rapid increase in  impedance is an indicator that there is a problem and ablation should be stopped  One possible  cause for the rapid increase is char formation  another possible cause is that the electrode has  become buried in the tissue     Figure 4       a    1 Functional Principles   Page 4    RF Generator User Manual       2 Indications    The use of the SMARTABLATE    RF Generator and all accessories is indicated in combination with  compatible therapeutic catheters for use in conventional intracardiac RF ablation procedures   Read and follow the instructions for use that are provided in this user manual before using  the system in a clinical application  Also refer to instructions for use that are supplied with the  accessories and with the compatible therapeutic catheters     USE   2 Indications   Page 5    RF Generator User Manual SMART       3 Warnings and Precautions  3 1 Warnings  Generator Usage    1  Read this user manual carefully before using the SMARTABLATE    RF Generator for the first  time  Note especially the instructions in Section 4 5  Setup     2  Cardiac ablat
71. heter type  that you selected when you created the preset     8  See Section 6 3 4  Editing a Preset   Temporary Save  for an explanation of a  temporary save     Designating a Default Preset    Quick Steps     Select the preset that you want as the default     Press Set Default     Notes   1  The default preset is the preset that is automatically selected after you select a  catheter type     2      D     appears after the name of the default preset     Deleting a Preset    Quick Steps      Select the preset to delete     Press Delete Ea      Press Yes     Notes   1  The standard presets that come with the generator cannot be deleted     2  If you delete all of the presets for one physician  the physician s name will  be deleted from the Physician Presets list  see Section 6 3 8  Displaying One  Physician s Presets      Displaying One Physician s Presets    Quick Steps     Press the PRESET workflow button     Select the catheter type     In the Physician Presets pane  press the button       Select the physician whose presets you want to see     Notes   1  When the button in the Physician Presets pane says All  all of the presets for  the catheter type that you selected appear in the Preset pane  When the button  says the name of a physician  only the presets for that physician for the selected  catheter type  and the standard preset for the selected catheter type  appear     2  Ifa physician   s name is not listed on the Physician Presets screen  there are no  presets associ
72. ific  location in the tissue  This current density is a factor not only of the generator output  but also  of the ablation electrode surface area and the contact between the ablation electrode and the  cardiac tissue  In order to assess the effect of the RF power  the user must watch for intracardiac  ECG signal reduction in the area of the ablation     1 3 Temperature Controlled Ablation Method    The temperature of the ablation electrode is measured by a temperature sensor  In the case of  non irrigated ablation catheters  the temperature of the ablation electrode is an indication of the  temperature of the tissue in contact with the electrode  During temperature controlled ablation   the RF power output is regulated automatically by the generator so that the temperature  selected by the user is reached and maintained at the desired level  see Figure 3   Thus the risk of  charring  tissue adhesion  or pop is greatly reduced     Figure 3       The size of a lesion is determined primarily by the ablation time and temperature  With the  temperature controlled ablation method  it is not necessary to set the initial RF power level  manually     Note  Because the ablation electrode measures only the temperature of the electrode with  its integrated temperature sensor and not the actual tissue temperature deep inside the  myocardium  tissue overheating can still occur  Therefore  temperature measurement is  suitable only for limiting overheating  However  in some physical circumst
73. indicator will flash red   immediately disconnect the mains cable from the generator     31  Electrodes and probes for monitoring and stimulation devices can be electrical conductors  of RF current  Reduce the risk of burns by placing the electrodes and probes as far away as  possible from the site of ablation and from the indifferent electrode     32  To avoid possible injury to patients and medical personnel  do not start RF power  application before the catheter is positioned in the intended ablation area     33  To minimize electrical noise on the ECG recordings  position catheter connection cables so  that they do not touch either the patient or other cables  For an optimal ECG trace  keep  unused active surface electrodes at a distance from the patient     34  To prevent fluid from impairing system performance  ensure that sterile catheters and cable  plugs are completely dry     35  To prevent a system malfunction  inspect all reusable accessories regularly  Do not use  damaged cables     36  Ablation data stored electronically must not be used for diagnosis or therapy  The intended  use is for archiving or research only     3 2 Warnings  System Safety and Connections    The SmarTABLaTe    RF Generator can be connected directly or indirectly to many other devices  and accessories  The generator and the connected group of devices and accessories are referred  to in the following warnings as the    system        1  The generator connections for the ECG signals and for th
74. ing a New Preset    7  You cannot save changes to the standard presets that come with the generator   If you make changes to a standard preset and want to save the changes  press  Save As and follow the instructions in Section 6 3 5  Creating a New Preset        6 Operation  Presets   Page 41    RF Generator User Manual       Table 5   Preset Setting Descriptions    Setting  Ablation Settings Pane    Time    Max  Power    Power    Power Ramp Duration    Temperature Settings Pane    Target    Warning    Cut Off    Control Algorithm    Description    The time limit for a single ablation  Ablation will stop  automatically after this time      only in Temperature Control Mode  Upper power limit for  the temperature control algorithm  The temperature control  algorithm will automatically set the power without exceeding  this value  Consult the instructions for use for the catheter to  determine the appropriate power setting for the catheter      only in Power Control Mode  The output power to be maintained  by the generator during the ablation  Consult the instructions for  use for the catheter to determine the appropriate power setting  for the catheter      only in Power Control Mode  The time it will take the generator  to increase power from OW to the value set in the Power field      only in Temperature Control Mode  The temperature that you  want the ablation electrode to reach  The generator will generate  power up to the Max  Power setting in an attempt to reach the  Targe
75. ings pane    20  Off  Tem peratu re    Temperature       Power    Irrigation Control    Settings pane IRRIGATION CONTROL SETTINGS  43 High Flow G gt    Settings pane       Low Flow  Normal and Fast re Re    appears only  i   Post RF Time 5 for irri ated  are only available ESOPHAGEAL TEMPERATURE SETTINGS Erland wenn r    for 4mm catheters Max        warning OCut Off catheters  Esophageal Rise    Warning    Cut Off  Temperature orecer THERMOCOOL        Standard    Settings pane   The Esophageal Temperature Settings pane shown above  is not applicable for the US and JP versions of the generator      Edit Preset Screen in Temperature Control Mode  Non Irrigated Catheter      OD Temperature EO    O Normal    ESOPHAGEAL TEMPERATURE SETTINGS  Max     Warning OCut Off  Rise    Warning OCut Off    Preset 4mm Standard         Temperature Control Mode is not available for some irrigated catheters      Temperature Control Mode for irrigated catheters and  the Esophageal Temperature Settings pane are not  applicable for the US and JP versions of the generator      6 Operation  Presets   Page 40    RF Generator User Manual       EDITING A PRESET  continued     Quick Steps   1  Display the Edit Preset Screen    Press the PRESET workflow button     Select the preset to edit     Press Edit     2  Change the Control Mode  not available for all catheter types      Inthe Ablation Settings pane  select Temperature or Power    3  Change Settings     Press a displayed value  see Table 5   Preset S
76. internal  problem in the  generator     and catheter receptacle flashes red    What To Do      Try a different cable or catheter   If the problem persists  contact  Customer Support       Verify that the catheter is  connected to the appropriate  cable and that the cable is  connected to the red Catheter  1 receptacle on the front of the  generator   See Item 4 in Section  4 2  Controls on the Front of the  Generator        Verify that you are using a  catheter and a cable or cables that  are compatible with the generator       Try a different cable or catheter   If the problem persists  contact  Customer Support     7 Troubleshooting   Page 70    RF Generator User Manual       Table 8   Warnings  Alerts  and Ablation Stop Reasons    Message    Reasons    Indicators    What To D   V   Visual  S   Sound  ann    Catheter temperature measurementerror      The generator recognizes Alert   Verify that you are using a  a valid temperature V  A and Alarm indicator turns catheter and a cable or cables that  reading but the on orange are compatible with the generator   redundant eaS ANENA   Try a different cable  or if you  system reflects a reading S  none    are using a dual sensor  DS   catheter  try a different catheter   If the problem persists  contact  Customer Support         Possible reasons Ablation Stop Reason  V   Ablation Summary and  catheter receptacle flashes red for    5 seconds      The cable is faulty       There is a compatibility  problem between the  generator an
77. ion procedures should be performed only by physicians who have been  thoroughly trained in RF catheter ablation techniques  Catheter ablations should be  performed only in medical rooms in compliance with IEC EC 60601 1  Annex A 4  Clause 16     3  In catheter ablation procedures  fluoroscopy is commonly used  When using fluoroscopy   there is the risk of exposure to considerable radiation  The intensity of the X rays and the  duration of the radiation can lead to acute radiation damage or to an increased risk of a  physical or genetic defects for both patients and medical personnel  Therefore  perform  catheter ablations only after carefully considering the potential radiation dose  The  advantages and disadvantages of X ray exposure should be considered carefully before  catheter ablation is performed in pregnant women  Also  the long term risk of prolonged  X ray imaging has not been researched  Therefore  the advantages and disadvantages  should be considered carefully before this procedure is performed in prepubertal children   Consider using a navigation system without fluoroscopy to reduce exposure to X rays     4  Do not apply RF energy while a stimulator is connected to the generator unless specifically  indicated by the stimulator   s instructions for use  Doing so might induce ventricular  fibrillation     5  Always verify that the generator   s visual and acoustic alarms are working before using the  generator  see Section 6 1 1  System Startup      6  When the ge
78. king according to Chinese Administration on the Control  of Pollution Caused by Electronic Information Product   ACPEIP   standard SJ  11364 2006     The device is protected against solid foreign objects of 12 5 mm diameter and  greater and is protected against vertically falling water drops     Certificate of origin in Germany  Hardware revision number    Mode of operation  Indication of duty cycle giving the maximum activation  ON   time and the minimum deactivation  OFF  time     9 Symbols   Page 99    RF Generator User Manual       10 Warranty  10 1 Transit Damage    The SMARTABLATE    RF Generator and accessories must be examined for transit damage  immediately after receipt  If the generator and accessories are damaged  notify Customer  Support  Damage claims can be validated only if the seller or the shipping agent is informed  immediately  For your convenience  an    Error Report    form is shipped with the generator  you can  use this form to record any damage     10 2 Warranty    The warranty for the SmarTABLATE    RF Generator warrants that the generator will conform to   its product specifications for a period of one year from the date of shipment or for the period   of time defined in the purchase contract  The warranty does not cover normal wear and tear   damage due to improper handling  accessories  connection cables  and disposable parts  Repairs  or any changes may be performed only by the manufacturer  a service engineer authorized by  the manufacturer  or 
79. lates the amount of fluid delivered and estimates the  remaining fluid amount based on the bag size you select  When you press  New Bag  the remaining fluid level is reset to be the same as the bag size you  select  Because the volume remaining calculation is affected by the preset you  select  the value of the remaining volume may not correctly reflect the actual  volume remaining if you select a different preset during an ablation     When you turn off the generator and turn it on again within 6 hours  the  volume remaining value is retained  If you turn the generator on again after 6  hours  the volume remaining is reset to the selected bag size     Before you start ablation  the Irrigation Control Edo  icon is available on  the Navigation Bar  After you start ablation  the Navigation Bar disappears   During ablation  if the fluid level reaches the warning limit set on the Irrigation    Control screen  the Irrigation Control icon appears at the bottom of the  ABLATION screen with a flashing orange border     If you press Irrigation Control Ego when the ABLATION screen is displayed  and you then make changes on the Irrigation Control screen  an asterisk      appears after the name of the preset on the PRESET workflow button to  indicate that temporary changes have been made  see Section 6 4 3  Making  Temporary Changes to Ablation Settings      The Irrigation Control screen indicates the volume remaining in the selected  bag size  When you press New Bag  this number resets 
80. lation catheter or to the  indifferent electrode that are not  approved by the manufacturer       If the problem persists  try    disconnecting any other devices  connected to the patient that are  not mandatory       Ifthe problem still persists  turn    the generator off  Let it cool down  for 5 seconds  then turn it on  again       Contact Customer Support even if    the problem is resolved       Close the pump door     RF Generator User Manual       Table 8   Warnings  Alerts  and Ablation Stop Reasons    Message  Reasons    Pump error      There is an internal error    in the pump     Pump not connected      An irrigated catheter  preset is being used and  no pump connection was  detected     Remote disconnected      The remote control    that had control   of the generator is  disconnected or there  isa communication  error between it and the  generator     Stop button   Generator      The Stop button  on the generator was  pressed     Indicators    Alert    V  A and Alarm indicator turns  on orange    S   none  Ablation Stop Reason    V   Ablation Summary    Alert    V  A and Alarm indicator turns  on orange    S   none  Ablation Stop Reason  V   Ablation Summary    S   Ablation stop tone    Warning    V  A and orange border around  System Control icon flashes for 20  seconds    S   Warning tone  Ablation Stop Reason  V   Ablation Summary    S   Ablation stop tone    Ablation Stop Reason    V  A and Ablation Summary    7 Troubleshooting   Page 81    What To Do     V   
81. limit value is exceeded  If unrealistic values are set  important alarm functions will be  triggered too late or not at all     8  To minimize the risk to the patient  keep the ablation duration to one minute or less     9  When using the usual operating settings  a low output RF power or a problem with the  ablation device can be a sign that the indifferent electrode is not correctly positioned or has  poor contact with its connection cable     10  Continuously monitor the generator   s impedance measurement during RF energy  application  If a sudden increase in impedance is observed  stop the RF energy delivery   Remove the catheter from the patient   s body and clean the ablation electrode of the  catheter with a sterile cloth to remove any adherent materials     11  If there is any doubt about an unintended increase in RF energy or the proper functioning  of the touch screen  Data Entry Knob    foot pedal  or remote control during RF delivery     immediately stop the RF energy delivery by pressing the Stop button on the generator   releasing the foot pedal  or pressing the Stand by button  If none of these measures  stops the RF energy delivery  disconnect the mains cable from the generator     3 4 Handling Indifferent Electrodes    An indifferent electrode is used for unipolar ablations  see Section 1 4  Unipolar Application  Method      The indifferent electrode is monitored at all times by the generator     e To ensure patient protection  the generator prevents unipola
82. loses keyboard  without saving       Press the PRESET workflow button       In the Catheter pane  select the catheter type for which you want to create a    preset       In the Preset pane  select a preset with settings similar to those you want     Press Edit     Change the settings  see Section 6 3 3  Editing a Preset      ip     Press Save As vw       Select a physician s name  then type a preset name     or  Press Create New  type a physician name  press PRESET NAME  then type a  preset name       Press OK ga     NAME  PRESET  NAME  Quick Steps   Notes   1    2    3    4    5    6     The physician s name can be a maximum of 9 characters   The maximum number of physician names is 10    The preset name can be a maximum of 13 characters   The maximum number of presets is 80     You cannot use the same preset name for two catheter types  For example  you  cannot create an 8mm preset named    20W POWER LIMIT    if you already created  one with the same name for 4mm  For this reason  it is a good idea to include a  catheter description such as    4MM    in the preset name     If there are fewer than 10 physician names in the Physician Presets list  the  physician s name that you type when you create a new preset will be added to  the Physician Presets list  see Section 6 3 8  Displaying One Physician   s Presets      6 Operation  Presets   Page 45    RF Generator User Manual       6 3 6    6 3 7    6 3 8    7  When you create a new preset  it will be available only for the cat
83. n   The Alarm indicator flashes alternately red and orange for a short time to  indicate that the Alarm indicator is working  Also  the RF On indicator flashes  orange and an audible RF tone is played briefly  If the indicators listed above  do not perform as described  there is a problem with the generator and you  should contact Customer Support     3  Progress dots in the bottom left corner of the screen indicate that the system  is performing a self test  When the self test is successfully completed  a    confirmation message appears asking you to press Stop to continue  After    you press Stop O  the ABLATION screen will appear  If the progress dots do  not appear  or if the number of dots stops increasing and the logo remains on  the screen  or if the confirmation message does not appear  there could be a  problem  see Section 7 4  Other Troubleshooting      4  Always wait until the automatic self test has been successfully completed  before starting the first steps of the procedure on the patient  for example   before anesthesia or before creating percutaneous access   Also  if using a  remote control  a pump  a CARTO   System  or an electrophysiology recording  system  verify that the connections to those systems are functional prior to  starting the first steps of the procedure  Doing this helps to detect problems  with the device before the patient is in a phase of the intervention in which  interruption could lead to an increased risk to the patient s health  Also 
84. n   y ThermoCool   Standard   D    P  wa   20w   Edit  N A H    Button  owas Set Default    MOLINA T CTRL 40DEG  Icon  ZHANG 20W  Selection buttons Pane    ThermoCool   SF    N avi g atio n Ba r Dre ThermoCool        Standard    EEE  ee a  Workflow buttons Icons    System Status icon       Table 2   Screen Features  Name Description How to Use    Button Blue  raised Press the button to display another screen or perform an    action   Selection button Gray  flat    white when   selected     Press one button to make a selection from a group of buttons     ee   6 Operation  Quick Overview   Page 31    RF Generator User Manual       Table 2   Screen Features    Name Description How to Use  Workflow button Black  flat Press the button to display the screen named on the button    on Navigation Bar  The workflow button for the displayed screen has an orange   underscore    Icon Image Press the icon to perform the action represented by the  image on the icon  see Section 6 10  Icons on the Screens     System Status Image  triangle See Sections 7 2  Warnings  Alerts  and Ablation Stop Reasons    icon and 7 3  System Errors    Pane Rectangular gray Provides visual grouping of related buttons and information     section of the screen    Note  When raised buttons are gray and when icons are gray  either they are not available for use  on the screen that is displayed or their functions are not supported by the software version     6 1 5 Data Entry Knob       The Data Entry Knob O is used 
85. nerator  on  If the problem persists  contact Customer Support     See Section 7 5  Calibrating the Touch Screen     Verify that the power supply meets the specifications in Section  8 1  Specifications and then check to see whether a fuse is broken   If a fuse is broken  replace it  see Section 4 5 3  Fuses   If the fuse is  not broken or if a new fuse breaks  contact Customer Support     Calibrating the Touch Screen    Over time  the buttons on the touch screen may move a little and the location that you press on  the screen may no longer be exactly where the button is  If this happens  follow the steps below     Quick Steps      Press the CONNECTIONS workflow button       Press System Settings sa       Press Calibrate Touch Screen       Press each target after it appears  a total of 6 targets        Press Stop        Notes   1     You can only calibrate the generator   s screen when the generator is the system    Master  see Section 6 7  System Control      2  Atany time during the calibration process  you can press Stop Y to exit    calibration     3  Ifthe calibration is successful  you will see Result  f   if the calibration is    unsuccessful  you will see Result  ol If the calibration is unsuccessful  repeat  the steps until a successful calibration is achieved  If the problem persists   contact Customer Support     4  When you perform calibration  be careful to press directly on the center of the  target  If the calibration fails  try touching the center of the targe
86. nerator User Manual          rc     amp  JGRU22EBNKNAO69     O          c  UL Jus    C E0366   4     Description    Alternating current    Serial data port    Ethernet port    Output signal    Input signal  USB port    Remote control receptacle    Foot pedal receptacle    Pump receptacle    Manufacturer    Manufacturing date    Serialnumber  Lot number  Japanese version of the SMARTABLATE    RF Generator    United States version of the SMARTABLATE    RF Generator    Worldwide version of the SMARTABLATE    RF Generator  except the United States  and Japan     Medical   applied current energy equipment as to electrical shock  fire and  mechanical hazards only in accordance with UL 60601 1   CAN CSA C22 2 NO   601 1   IEC 60601 2 2  2006    ANSI AAMI ES60601 1  2005    CAN CSA C22 2 NO   60601 1  2008   IEC 60601 2 2  2009    40GF    CE marking with the number of the notified body    The device is radiating non ionizing electromagnetic radiation   See Section 8 11   EMC Information and Technical Description      Separate collection for electrical and electronic equipment       9 Symbols   Page 98    RF Generator User Manual       RoHS y        IP21    Made in Germany    Description    The device does not contain phthalates     The device is latex free     RoHS compliant  Directive 2002 95 EC of the European Parliament and ofthe  Council of 27 January 2003 on the restriction of the use of certain hazardous  substances in electrical and electronic equipment     RoHS compliant  mar
87. nerator has been turned off  power continues to be supplied to the generator  through the mains cable  To completely cut off power to the generator  disconnect the  mains cable from the generator  To allow easy disconnection in the event of severe power  failure  make sure that the generator is located where there is easy access to the mains cable  on the back of the generator     7  After turning on the generator  always wait until the automatic self test has been  successfully completed before starting the first steps of the procedure on the patient  for  example  before anesthesia or before creating percutaneous access   Also  if using a remote  control  a pump  a CarTo   System  or an electrophysiology recording system  verify that the  connections to those systems are functional prior to starting the first steps of the procedure   Doing this helps to detect problems with the device before the patient is in a phase of the  intervention in which interruption could lead to an increased risk to the patient s health   Also perform a visual inspection of the accessories  see Section 5 3  Caring for Accessories      8  Use only catheters that have been approved for use with the maximum voltage specified in  this user manual  see Section 8 1  Specifications      9  Use only accessories that have been provided by or recommended by the generator   s  manufacturer  see Chapter 5  Accessories   The use of other accessories can have a negative  effect on the technical specifications 
88. ng and should not be used  Perform disinfection  by wiping  not by spraying  Avoid condensation  If cleaning or disinfecting the generator with  flammable or explosive substances cannot be avoided  make sure these substances have  completely evaporated before the generator is turned on     Check the vents on the bottom and sides of the generator regularly for excessive deposits of dust  and foreign bodies  Carefully brush or wipe off the unwanted material     4 7 Protecting Against Damage    In addition to being operated and maintained per the instructions in this user manual  the  SMARTABLATE    RF Generator must also be protected against damage  This includes installing the  generator in a safe location and protecting the generator against moisture  contamination  and  contact with flammable or explosive substances     Make sure that all cables to devices and accessories are arranged in a way that prevents tripping     4 System Overview and Setup   Page 20    RF Generator User Manual       Make sure that the front panel of the generator is closed before moving the generator  Hold the  sides of the generator  not the front panel  when lifting the generator  Carry the generator with  two hands  To transport the generator over long distances or to send the generator for service   use the original packaging  If the original packaging is not available  contact the manufacturer or  distributor for a replacement     4 8 Disposal    Inadequately treated electrical and electronic wa
89. ngen einer Dosierten Hochfrequenzkoagulation bei Stereotaktischen  Hirnoperationen  Zentralbl fur Chir 85  1051 1062    Nakagawa H   Yamanashi WS   Pitha JV   Arruda M   Wang X   Ohtomo K   Beckmann KJ   McClelland  JH   Lazzara R   Jackman WM   1995   Comparison of In Vivo Tissue Temperature Profile and Lesion  Geometry for Radiofrequency Ablation with a Saline Irrigated Electrode versus Temperature  Control in a Canine Thigh Muscle Preparation  American Heart Association    Remp T   Hoffmann E   Gerth A   Mattke S   Ebeling F   Steinbeck G   1994   Bedeutung des  Impedanzabfalles w  hrend Radiofrequenz Katheterablation  Z Kardiol 83  Suppl   118    Remp T   Hoffmann E   Gerth A   Mattke S   Ebeling F   Steinbeck G   1994   Drop in Impedance  and Safety during Catheter Ablation  Validation of a Predictive Marker for Impedance Rise  Eur  Heart J 15  Suppl   286    Steinbeck G   1983   Differentialdiagnose der Herzrhythmusst  rungen  Tachykarde Rhythmus   St  rungen  In  Herzrhythmusst  rungen  Springer  Berlin Heidelberg New York  Handbuch der  Inneren Medizin  Bd IX 1  S  617 642     Steinbeck G   1986   Intrakardiale Ableitung und Programmiierte Stimulation  In  L  deritz B  Hrsg   Herzschrittmacher  Springer  Berlin Heidelberg New York Tokyo  S  120 161    Steinbeck G   Doliwa R   Bach P   1988   Treatment of Paroxysmal Atrial Fibrillation with Digoxin  Alone or Digoxin Plus Antiarrhythmic Drugs  Dtsch Med Wschr 113  1867 1871    Wittkampf FHM   1992   Temperature Response
90. ngsamen Bahn bei AV Reentry ohne Anderung der Normalen AV Uberleitung  Z Kardiol 81   Suppl 1   142  Abstract     Hoffmann E   Remp T   Gerth A   Mattke S   Steinbeck G   1993   Does Preablation Impedance  Measurement Improve the Safety of Radiofrequency Catheter Ablation  Eur Heart J 14  Suppl   34    Hoffmann E   Remp T   Gerth A   Mattke S   Steinbeck G   1993   Does Impedance Monitoring  during Radiofrequency Catheter Ablation Reduce the Risk of Impedance Rise  Circulation 88   I 165    Hoffmann E   Remp T   Gerth A   Mattke S   Steinbeck G   1993   Preablation 50 kHz Impedance  a  New Parameter for Assessing Myocardial Wall Contact before Radiofrequency Catheter Ablation   J Am Coll Cardiol 21  49A  Abstract     11 References   Page 101    RF Generator User Manual SMART       Hoffmann E   Steinbeck G   1993   Elektrotherapie Supraventrikul  rer Tachykardien  Vorhof   flimmern Vorhofflattern   Innere Medizin 48  446 452    Hoffmann E   Steinbeck G   1994   Katheterablation Supraventrikul  rer Tachykardien   Springer Verlag  ISBN  3 540 60301 8    Hoffmann E   Steinbeck G   1999   Interventionelle Kardiale Elektrophysiologie  Springer Verlag   ISBN  3 540 65684 7    Moringlane JR   Koch R   Schafer H   Ostertag ChB   1989   Experimental Radiofrequency  Coagulation with Computer Based Online Monitoring of Temperature and Power  Acta  Neurochirurgica 96  126 131    Mundinger F   Riechert T   Gabriel E   1960   Untersuchungen zu den Physikalischen und  Technischen Voraussetzu
91. nnected to the appropriate  temperature reading  on orange cable and that the cable is  Possible reasons are  connected to the red Catheter    An esophageal probe S   none 2 receptacle on the front of the   is not connected or Ablation Stop Reason generator   See Item 5 in Section     connected tothe 4 2  Controls on the Front of the  wrong receptacle  V   Ablation Summary and Generator      The cable is faulty  esophageal probe receptacle   Verify that you are using a probe  flashes red for 5 seconds and a cable that are compatible    S   Ablation stop tone with the generator       Try a different cable or probe   If the problem persists  contact  Customer Support       In order to continue with the  procedure  you can disable the  esophageal monitoring feature by  setting the warning and cut off  values to Off   See Section 6 3 3   Editing a Preset      Esophageal temperature rise  not applicable for the US and JP versions of the generator       The esophageal probe Ablation Stop Reason   See the instructions in this table  temperature rise for Esoph  temperature above  exceeded the cut off v     A and ablation Summary maximum  Alert and Ablation  value set in the preset  and esophageal probe receptacle Stop Reason      flashes red for 5 seconds  S   Ablation stop tone    Esophageal temperature rise  not applicable for the US and JP versions of the generator       The esophageal probe Warning   See the instructions in this table  temperature rise A for Esoph  temperature ab
92. nufacturer  Other catheters may be used only after consultation with  the distributor  For recommended accessories  see Section 5 2  Accessory Lists     To connect an ablation catheter  use a sterile cable and connect one end to the Catheter 1  receptacle on the front of the generator  see Item 4 in Section 4 2  Controls on the Front of the  Generator  and the other end to the catheter  Alternatively  if the ablation catheter is connected  to an additional system  such as the CarTo   System  connect the cable from the system to the  Catheter 1 receptacle     To connect an esophageal probe  connect the appropriate cable to the Catheter 2 receptacle  on the front of the generator  see Item 5 in Section 4 2  Controls on the Front of the Generator     An esophageal probe cannot be used with the US or JP versions of the generator  The Catheter 2  receptacle is not functional      4 5 7 Indifferent Electrode Connections      The receptacles on the front of the generator for the indifferent electrodes  see Item  6 in Section 4 2  Controls on the Front of the Generator  can be used to connect both  solid and split indifferent electrodes  Some of the indifferent electrodes recommended  for use with the SMARTABLATE    RF Generator are available with cables that have the  appropriate mating component  For other indifferent electrodes  appropriate adapter  cables are available        4 System Overview and Setup   Page 18    RF Generator User Manual SMART       4 5 8 Ethernet Connection  
93. o      To resume ablation  press    Ablation Start   or press and  hold the foot pedal       Make sure that the vents on the    bottom and sides of the generator  are not blocked       Wait for a few minutes without    turning on the RF energy and then  try to operate the system again       Contact Customer Support even if    the problem is resolved       To resume ablation  press    Ablation Start    or press and  hold the foot pedal       See the instructions in this table    for Power output too high     7 Troubleshooting   Page 82    RF Generator User Manual       7 3 System Errors    Event Window Displaying System Error Codes    EVENT  code  Active WARNINGS  l  System error  2213  10 28 No active warnings  i Active ALERTS  atheter temperature invalid    mnpedance above maximum Event WI ndow  displaying System    Error codes    System  Status  icon       7 3 1 System Error    When a System Error occurs  the A System Status icon is displayed and the Event window  appears displaying the System Error codes      NThe Alarm indicator is orange and the warning tone is emitted     Manually record the codes displayed in the Event window  If you want to save the ablation  files and event log  insert a USB flash drive containing a data transfer indicator into the back of  the generator and wait for the data to transfer  Then press Stand by SUB for approximately   3 seconds to turn the generator off and press Stand by  49 again to turn the generator on   Contact Customer Support to 
94. ol   Standard ol P   nya   20w   a Edit  DAVIS 30W N A    Set Default    Physician    Presets pane  MOLINA T CTRL 40DEG  4mm  ZHANG 20 W    smm  Preset  pane    ThermoCool   SF    Catheter  pane       6 3 1 Preset Overview    e Apreset is a named group of settings to use during ablation     e The SmART  BLATE  RF Generator comes with one standard preset for each catheter type listed  in the Catheter pane  This preset is always the first in the list and it cannot be deleted     e You can create new presets for each physician     e You can view alist of all of the presets for a selected catheter type  or you can view only the  presets for a selected physician for the selected catheter type     e When you select a catheter type  the default for that catheter type is automatically selected   You can designate one of the displayed presets to be the new default for the catheter type     6 3 2 Selecting a Preset    Quick Steps     Press the PRESET workflow button   Select the catheter type     Select a preset     Notes 1  The currently selected preset is highlighted  white background    2  The PRESET workflow button displays the name of the selected preset     3  Forinstructions on displaying the presets for only one physician  see Section  6 3 8  Displaying One Physician   s Presets     6 Operation  Presets   Page 39    RF Generator User Manual       6 3 3 Editing a Preset    Edit Preset Screen in Power Control Mode  Irrigated Catheter    Ablation Impedance  Settings pane i 60 Sett
95. on  If  abnormal performance is observed  additional measures may be necessary  such as reorienting or  relocating the SMARTABLATE    RF Generator     b Over the frequency range 150 kHz to 80 MHz  field strengths should be less than 3 V m     NSS Tee GG  8 Technical Data   Page 94    RF Generator User Manual SMART       Recommended Separation Distances between Portable and Mobile RF Communications  Equipment and the SmART  BLATE   RF Generator    The SMARTABLATE    RF Generator is intended for use in an electromagnetic environment in which radiated  RF disturbances are controlled  The customer or user of the SmarRTABLATE    RF Generator can help prevent  electromagnetic interference by maintaining a minimum distance between portable and mobile RF  communications equipment  transmitters  and the SMARTABLATE    RF Generator as recommended below   according to the maximum output power of the communications equipment     Separation Distance According to Frequency of Transmitter  m   Rated Maximum Output    Power of Transmitter  W  150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2 5 GHz  d 1 17VP d 1 17 VP d 2 33 VP  0 01 0 12 0 12 0 23  0 1 0 37 0 37 0 74  1 1 17 1 17 233  10 3 69 3 69 7 38  100 11 67 11 67 23 33    For transmitters rated at a maximum output power not listed above  the recommended separation  distance din meters  m  can be estimated using the equation applicable to the frequency of the  transmitter  where P is the maximum output power rating of the transmitter in 
96. ove  exceeded the warning V  appears near esophageal maximum  Warning      value set in the preset  temperature reading and its  orange border flashes    S   Warning tone    7 Troubleshooting   Page 73    RF Generator User Manual       Table 8   Warnings  Alerts  and Ablation Stop Reasons    Message    Reasons    Indicators   V   Visual  S   Sound     What To Do    Foot pedal disconnected   Generator  or Foot pedal disconn    Generator       The foot pedal that  was connected to  the generator when  ablation started became  disconnected during  ablation     Warning    V  A and orange border around  System Control icon flashes for 20  seconds    S   Warning tone  Ablation Stop Reason  V   Ablation Summary    S   Ablation stop tone      Disconnect the foot pedal from    the generator  then reconnect it   If the problem persists  either use  a different foot pedal or operate  the generator using the Ablation    Start    and Stop    buttons on  the generator or remote control     Foot pedal disconnected   Remote  or Foot pedal disconn    Remote       The foot pedal that  was connected to the  remote control when  ablation started became  disconnected during  ablation     Warning    V  A and orange border around  System Control icon flashes for 20  seconds    S   Warning tone  Ablation Stop Reason  V   Ablation Summary    S   Ablation stop tone    Foot pedal released   Generator      The pressed foot pedal  that is connected to the  generator was released     Ablation Stop Reason 
97. p the Generator    The SMARTABLATE    RF Generator may be used only in medical rooms in compliance with IEC EN   Annex A 4  Clause 16     Place the generator on a secure  non slip surface such as the Carto   System cart  If the generator  is placed on any other surface  make sure it is securely fastened     Do not place the generator directly above another device and do not place any other device  directly on the generator     Make sure that there is enough free space on all sides of the generator to allow the heat created  by the generator to escape     4 5 2 Mains Connection    The SMART  BLATE  RF Generator automatically adjusts for the necessary mains   620  voltage and frequency  To avoid the risk of electric shock  connect the mains cable  from the generator mains socket  see Item 6 in Section 4 3  Controls on the Back  of the Generator  to a 3 prong outlet that meets the specifications in Section 8 1   Specifications  Use only the mains cable supplied by the manufacturer or distributor for  the country where the generator is being used        4 5 3 Fuses    BEE Externally accessible fuses for the SmarTtAsLate    RF Generator are located under the  fuse cover  see Item 7 in Section 4 3  Controls on the Back of the Generator   Use only  fuses approved by the manufacturer     4 System Overview and Setup   Page 17    RF Generator User Manual SMART       4 5 4 Potential Equalization    To avoid differences in potential between the generator and other devices with power  ope
98. perform  a visual inspection of the accessories  see Section 5 3  Caring for Accessories      6 Operation  Quick Overview   Page 28    RF Generator User Manual       6 1 2 System Shutdown    Quick Steps      Stop ablation  see Section 6 4 2  Starting and Stopping Ablation      Press and hold Stand by for approximately 3 seconds to turn off the  generator       Turn offthe remote control     For irrigated catheters  turn off the pump     Note   You can turn off the pump and remote control before or after you turn off the  generator     6 1 3 Screens    The SMART  BLATE  RF Generator has four primary screens  represented by the first four buttons on  the Navigation Bar  See Section 6 9  Flowchart of the Screens  for a chart that lists other screens  that can be accessed from these primary screens     Table 1   Screens  Screen Description Section    CONNECTIONS 6 2   Connections              The Esophageal Probe shown above is not applicable for the US  and JP version ofthe generator      Shows all devices that you can connect to the SMART  BLATE    RF Generator     Indicates which devices are currently connected and  functioning properly     Provides access to information about the connected devices     Provides access to the System Settings screen     6 Operation  Quick Overview   Page 29    RF Generator User Manual SMART       Table 1   Screens  Screen Description Section    PRESET 6 3  Presets    ThermoCool   Standard  D    e   nA   20w   Edit    DAVIS 30W N A Set Default  M
99. poses restrictions on the flow rate  and the ranges of power  Consult the instructions for use for the catheter to  determine the appropriate flow rate and power     6 4 4 Ablation Summary    Ablation Summary    ABLATION SUMMARY Ablation  RF time total 60s Summa ry    Power max  10W  Power avg  9  ON  Energy total 595J    Tem p  max  44  C    Temp  avg  391C    Imp  max  1010  Imp  drop  Imp  min     Irrigation 13mL       Ablation Sto p Reason  Ablation  Timer completed Stop  Ablation    i J er Rea son    Notes   1  Afew seconds after ablation stops  the Ablation Summary appears     2  Atthe bottom of the Ablation Summary  an Ablation Stop Reason is displayed   see Section 7 2  Warnings  Alerts  and Ablation Stop Reasons      3  To remove the trend lines and the Ablation Summary  either wait for them to  disappear or press anywhere on the screen     4  To view a summary of the entire procedure  press the POST workflow button   see Section 6 5  Post Procedure      6 4 5 Low Fluid Level    Quick Steps   To Change the Fluid Level Settings for a Preset  See Section 6 3 10  Irrigation Control   To Indicate Replacement of the Fluid Bag    Wait until ablation stops  or press Stop       Press Irrigation Control Sul    If the bag size changed  select the new bag size  see Section 6 3 10  Irrigation  Control      Press New Bag     Press OK 07     Press the ABLATION workflow button        6 Operation  Ablation   Page 54    RF Generator User Manual       Notes   1     The generator calcu
100. principles of various methods of applying RF energy and the  principles of power and temperature control that are presented in this manual  Proper application  of RF energy leads to desired tissue denaturation to correct arrhythmias  incorrect application of RF  energy can lead to excessive tissue denaturation     RF Generator User Manual       The SmarTABLaTE    RF Generator may be used only by medical personnel trained and experienced  in the techniques of electrophysiology  Before using this generator for the first time in a clinical  application  the user should thoroughly read this user manual  To become familiar with operation  of the generator and to avoid damaging it  the user should practice using the generator and all  accessories by simulating ablation in a container of saline solution or by using a STOCKERT test box  for the SMarRTABLATE    RF Generator  The user should also become familiar with all accessories and  other devices connected to the generator by reading and following the instructions for use for the  accessories and devices  For information on training and a STOCKERT test box for the SMARTABLATE     RF Generator  please contact Biosense Webster     We will gladly answer any questions that you may have   Yours truly     STOCKERT GmbH    RF Generator User Manual SMART       1 Functional Principles  1 1 Functional Principles of RF Ablation    Cardiac RF ablation is the destruction of the electro mechanical function  heart beat conduction   of cardiac ti
101. r ablation in the absence of an  indifferent electrode     e Ifthe indifferent electrode is not correctly connected to the generator  the generator immediately  stops RF energy delivery and displays a message     e The generator is equipped with a contact quality monitoring feature for split indifferent  electrodes  When a split indifferent electrode is used  the generator displays an alert message  as soon as skin contact with the electrode surface decreases by an unacceptable amount  After  the indifferent electrode is applied to the patient  the generator continuously monitors the  contact quality and updates an internal value for best achieved contact quality  the minimum    3 Warnings and Precautions   Page 11    RF Generator User Manual SMART       contact impedance of the split indifferent electrode   This value is reset every time the indifferent  electrode is disconnected from the patient  The alert message appears when the contact  impedance of the split indifferent electrode increases by 40  from the best achieved contact  quality  If a solid  non split  indifferent electrode is used and the electrode partially detaches  itself  no alert is triggered  The danger of a skin burn under the indifferent electrode is thus  increased     Place the indifferent electrode within the  operating field on the patient   s back     Solid indifferent electrode       Split indifferent electrode       The generator provides the option of connecting two separate indifferent electrode
102. r cables  Keep active electrodes that are temporarily not in  use at a Safe distance from the patient     17  Set the RF power at only moderate output to avoid charring and clotting at the catheter  ablation electrode     18  When the SmART  BLATE   RF Generator is used with irrigated catheters  monitor the irrigation  flow rate to avoid hazards caused by insufficient irrigation flow  The approximate flow  rate can be estimated by observing the drip speed in the drip chamber  The hospital staff  is responsible for determining and monitoring the flow rate to avoid insufficient flow of  the irrigation solution  The hospital staff is responsible for monitoring the total amount of  solution delivered to the patient to avoid an excessive loading of the irrigation solution in  the patient  For recommended flow rates  refer to the catheter   s instructions for use     3 Warnings and Precautions   Page 7    RF Generator User Manual SMART       19  Avoid using flammable anesthetics or oxidizing gases such as nitrous oxide  NO  and  oxygen  O   if the procedure is being performed in the region of the thorax or head   unless the gases are being aspirated off or an anesthesia safe device is being used  Before  starting RF ablation  allow time for flammable substances that are used as cleaning agents   disinfectants  or solvents to evaporate  There is a risk associated with flammable liquids  under the patient or in the patient s body cavities  Wipe away the liquid in these places away  b
103. r increment displayed on  a graph  If Use automatic flow adjustment is not selected  press High Flow and    turn the Data Entry Knob to change the value  Consult the instructions for  use for the catheter to determine the appropriate flow rate for the catheter     The standby rate at which fluid flows before the Pre RF period and after the  Post RF period     The length of time  after Ablation Start    is pressed  that fluid flows at the  High Flow  ablation flow  rate without RF  After this Pre RF Time  RF turns on  A  tone is emitted during the Pre RF time    The length of time  after RF turns off  that fluid flows at the High Flow  ablation    flow  rate  After the Post RF Time  flow returns to the Low Flow  standby flow   rate  A tone is emitted during the Post RF time     Select Monitor Fluid if you want the generator to issue a warning when the fluid  level is low  Then specify the size of the bag that you are using  Press 500 mL   1000 mL  or 2000 mL  Or press User  then press the size that appears under    User  and then use the Data Entry Knob o to set a different bag size     The fluid level at which a warning will occur  Low Fluid Warning is only available  when Monitor Fluid is selected     After you attach a new bag to the pump and select the new bag size on the  screen  press New Bag     After you press New Bag  the generator   s fluid level monitor will reset and the  new bag size will appear after Volume Remaining     6 Operation  Presets   Page 48    RF Gener
104. rate effective RF power setting     A lesion  conversion of the soluble protein in the tissue into a denatured  coagulated state  arises  when the tissue is exposed to RF current densities  At low current densities  the tissue is heated  slowly and contracts as a result of fluid loss  When the tissue is exposed to excessively high RF  current densities  the tissue fluid evaporates so rapidly that the resulting steam pressure in the  cells destroys the cell membrane  If this process occurs suddenly  superheating   it is perceptible  as a steam explosion  Since this phenomenon is audible in certain cases and under some  circumstances can be felt at the catheter handle  it is also known as    pop    In extreme cases  this  can lead to undesirable tissue changes such as tears  craters  and holes  A suitable choice of RF  energy output and sufficient contact between the ablation electrode and the myocardial wall  avoids this pop effect     LOG  1 Functional Principles   Page 1    RF Generator User Manual       1 2 Power Controlled Ablation Method    As described in Section 1 1  Functional Principles of RF Ablation  a high RF current can rapidly  heat the tissue  With the power controlled ablation method  the RF power output can be  manually adjusted on the basis of values known from experience     The power adjusted by the user in this mode is the power output from the generator  The user  should be aware that the lesion formation is dependent on the RF current density in the spec
105. ration  the SmarTABLATE    RF Generator must be connected to a grounding line   as specified in IEC EN 60601 1   Connect one end of the green and yellow potential  equalization cable provided with the generator to the grounding connection on  the back of the generator  see Item 8 in Section 4 3  Controls on the Back of the  Generator   Connect the other end to the room   s central grounding line  To avoid  radiation interference  noise  in the ECG signal  always use the same grounding line for  connection of all devices involved  electrocardiograph  Carto   System  generator      4 5 5 Circular Plug Connections    Ki The circular plugs on the accessory cables are the same color as the  O  receptacles on the front and back of the generator  see Sections 4 2   Controls on the Front of the Generator  and 4 3  Controls on the Back of   CE the Generator   To connect an accessory cable to the generator  align  the arrow on the plug with the mark on the receptacle and push in  without unnecessary force until a click indicates that the connector is  locked in the receptacle  If the plug does not fit in the receptacle  verify  that the color coding matches and that the number of pins in the plug is  appropriate for the receptacle  To remove the plug  gently pull back the  sleeve on the plug and pull the plug out of the receptacle     4 5 6 Connection Cables and Catheters    Use only ablation catheters with temperature sensors and connection cables supplied by or  recommended by the ma
106. rator  S   Ablation stop tone Section 3 4  Handling Indifferent    The indifferent Been Oge gt    electrode contact with   Consider applying the indifferent  the patient is not good  electrode to a different place on    the patient s body to avoid oil   hair  and dirt that increase the  impedance     If the catheter is connected  through an additional system  such as a navigation system   make sure that this system is  not intentionally disconnecting  the generator from the tip of  the catheter  Try replacing the  system parts that are between  the generator and the catheter  or  connect the catheter directly to  the generator     If the problem persists  contact  Customer Support     It is not recommended  but if  you know the reason for the  high impedance and still want  to perform an ablation  you can  disable this cut off by setting its  value to Off   See Section 6 3 3   Editing a Preset      CE u  7 Troubleshooting   Page 75    RF Generator User Manual       Table 8   Warnings  Alerts  and Ablation Stop Reasons    Message    Reasons      The impedance was  below the minimum  cut off value set in the  preset  Possible reasons  are       There is a compatibility  problem between the  generator and external  devices       There is an internal  problem in the  generator     Rarely  such as in  pediatric cases  the  patient has low body  impedance and there is  no problem       The impedance change  exceeded the spike  cut off value set in the  preset  Possible reasons  are
107. report the codes  even if the problem is resolved     7 3 2 Fatal System Error     N The Alarm indicator flashes red and the high priority tone is emitted     Immediately turn off the generator  If pressing Stand by    49 does not turn off the generator   remove the mains cable from the generator  Contact Customer Support     CE u  7 Troubleshooting   Page 83    RF Generator User Manual       7 4 Other Troubleshooting    Problem    Bubbles are detected in the  irrigation tubing during standby     On startup  the progress dots do  not appear or the logo screen does  not disappear     On startup  after you press  Stop in response to the  message that appears  the    ABLATION screen does not appear     Nothing happens when you press a  button on the screen     The generator won t turn on     7 5    What To Do    Immediately turn the 4 way stopcock on the SMART  BLATE    Irrigation Tubing Set so that the fluid is not routed to the catheter   Then flush the tubing  see Section 6 6  Flushing to Eliminate Air in  the Tubing and Catheter   After flushing  turn the 4 way stopcock  to allow fluid to be routed to the catheter again     Press and hold Stand by for approximately 3 seconds to  turn off the generator  Then press Stand by again to turn the  generator on  If the problem persists  contact Customer Support     Unplug the foot pedal and plug it in again  Then press and hold  Stand by for approximately 3 seconds to turn off the  generator  Then press Stand by again to turn the ge
108. ries are functional     Once a year  perform the verifications listed above as well as those below     e The device label is present and readable   e The cable connection receptacles are not damaged and are not loose   e The accessory cables  especially the mains cable and the grounding cables  are not damaged     e The remote control can function as the system Master   Make the remote control the system    Master and then perform ablation using a STOCKERT test box for the SMARTABLATE    RF  Generator      8 Technical Data   Page 89    RF Generator User Manual       Because the generator is a medical device  some countries require an electrical safety test  If local  regulations require a safety test  have the generator tested by a technician certified by the local  governing agency      N Important Notice  To ensure the safety of the SmarTABLATE    RF Generator  repairs may be performed only by the  manufacturer  a service engineer certified by the manufacturer  or a service center certified by    the manufacturer  The safety inspections must be performed at least once a year  and after every  repair  according to IEC EN 60601 1 and its sub standards and IEC EN 62353     8 8 Maintenance  Perform the safety inspections specified in Section 8 7  Safety Inspections     Regularly clean and disinfect the SMARTABLATE    RF Generator per Section 4 6  Cleaning and  Disinfecting the Generator     Regularly clean the accessories per the instructions for use provided with the accessori
109. s  Both   split and solid indifferent electrodes can be connected  In this case as well  the contact quality  monitoring feature works only with split indifferent electrodes  However  if two split indifferent  electrodes are used at the same time and one loses contact quality but the other maintains  contact quality  no alert is triggered  The use of two indifferent electrodes doubles the size of  the contact surface on the patient   s skin  The use of two indifferent electrodes is recommended   especially for heavy patients with a low muscle mass  in ablations with high output   gt  50 watts    and in the case of prolonged RF energy delivery   gt  60 seconds   If two indifferent electrodes are  used  apply one electrode to the left side of the ablation site and the other electrode to the right  side of the ablation site  both at approximately the same distance from the ablation site  The two  indifferent electrodes must not overlap     Guide to the Use of Indifferent Electrodes with the SMART  BLATE  RF Generator      Use only indifferent electrodes with a surface area of  gt  124 cm    that conform with either IEC   EN 60601 2 2 or ANSI   AAMI HF 18     e Read the instructions for use for the indifferent electrodes carefully and take special note of  the warnings and precautions sections       Fora single use electrode  Make sure that the electrode contact surface is not dry  If it is  dry  replace it with a new  unused electrode before continuing with the ablation procedur
110. s displayed on the screen during  ablation  see Section 6 4  Ablation      2  The Irrigation Delivered option is displayed only for irrigated catheters  This  value is the total irrigation delivered during the procedure  excluding flushing     3  The Dual Temp option is displayed only for dual sensor  DS  8mm catheters     6 3 13 Copying Presets to a USB Flash Drive    Quick Steps     Press the PRESET workflow button     Press To USB zZ    Note   This feature is not available in some software versions     6 Operation  Presets   Page 50    6 4 Ablation    Item           COIN ISA   WM   P  WIN    ABLATION Screen in Power Control Mode  before ablation starts   1 234 5 6 7 i 9        Temperature Control Mode for irrigated catheters  is not applicable for the US and JP versions of the generator      Description   Impedance measurement   Set time   Time since start of ablation   Flow rate control indicator  AUTO   automatic  MAN   manual   Set flow rate   Set temperature  displayed only in Temperature Control Mode   Measured temperature    Set power  in Power Control Mode   Max  power  in Temperature Control Mode     Measured real power output    6 Operation  Ablation   Page 51    RF Generator User Manual       6 4 1 Starting a New Procedure    Quick Steps     Press the ABLATION workflow button     Press New Procedure L   on the Navigation Bar   Press Yes     On the PRESET screen  select a preset     Notes   1  When you start a new procedure  any accumulated parameters are reset an
111. sesssssssssssssssssssssssssescsesessssssssssesesesesescscsesssssseseseseseassssseseseseseseacaceeseeses 46  6 3 8 Displaying One Physician s Preseis   uuenee ae 46  6 3 9 Adding a Name to the Physician Presets List            sssesscsssssssecscssssssessssssecscesssseseescens 47  6 3 10 oan COMMON mern ee ana 47  63 11 Changing AUdiO Settings area ae ENA 49  6 3 12 Selecting Ablation Display SettingS          s  sseessesssesssesseesssesseesseessesseesseesseossesseesseesseesseessee 49  6 3 13 Copying Presets to a USB Flash Drive           ssessesssesssesseesseesseesseenseoseesseesseosseoseesseesseosseessee 50  CA ABITO a E 51  6 4 1 Starting a New Procedure         se ssesseessesseessessesseeseessesseoseessesseosecoseoseesseoseoseessesseoseeseessesseeseesss 52  642 Starting and Stopping Ablation werner un 52  6 4 3 Making Temporary Changes to Ablation SettingS         ss  sssesseesseessesssesseesseessesseesseessee 53  GAA PPO Mis UMMA een 54  6 4 5 Low Fluid Level           sse ssesssesssesseessesssesssesseesseesseesseesseossesseesseosseoseeoseesseoseeosseossesseesseesseoseeoseessee 54    RF Generator User Manual       03 POTRO OU ee nee eee 56  6 5 1 Exporting Ablation and Event Log Data to a USB Flash Drive    56   6 6 Flushing to Eliminate Air in the Tubing and Catheter uu    esesssscscsesssssssssssssesessssssssssssssesees 58  STI GOAT ON ee een ae 59  6 7 1 Making the Generator the System Mastel            sssssssesscsssssssecssssssesesscsssecscesesseseescees 59   6 8 Volum
112. ss and release the foot pedal on the remote control    Press Stand by on the generator     If the remote control is the system Master  press Stand by on the  remote control     Disconnect any of the following cables  catheter connection cable   indifferent electrode connection cable  mains cable        6 Operation  Ablation   Page 52    RF Generator User Manual       5  During ablation  the Navigation Bar disappears  After ablation  the Navigation  Bar reappears     6  Forirrigated catheters  if the fluid level reaches the Low Fluid Warning level     the Irrigation Control Ego  icon appears at the bottom of the ABLATION  screen as a warning  After ablation stops and the Navigation Bar appears  press    the Irrigation Control 29  icon to access the Irrigation Control screen  see  Sections 6 3 10  Irrigation Control  and 6 4 5  Low Fluid Level      7  Ablation stops if conditions are not right for ablation  see Section 7 2  Warnings   Alerts  and Ablation Stop Reasons      8  A few seconds after you stop ablation  the Ablation Summary appears  see  Section 6 4 4  Ablation Summary      9  Tominimize the risk to the patient  keep the ablation duration to one minute or  less     6 4 3 Making Temporary Changes to Ablation Settings    Quick Steps   To Change Values at the Top of the Ablation Screen      Press the displayed value          Turn the Data Entry Knob    to change the value     To Change Values at the Edit Preset Screen      See Section 6 3 4  Editing a Preset   Temporary
113. ssue through the application of RF energy     The term    radiofrequency     RF  is used to refer to an alternating current that flows through a  conductor  In the case of ablation  RF current flows through biological tissue that contains free  ions  The saline solution present in the tissue provides the electrical conductivity  The tissue  conductivity can be represented by the tissue impedance  Low impedance represents high  conductivity and high impedance represents low conductivity     The application of RF current to biological tissue causes heating of the tissue  The higher the  RF current density in the biological tissue  current per unit area   the higher the resulting  temperature  The tissue stops reacting to electrical stimulation when heated above a threshold  over a minimal period of time     The extent of heating is proportional to the RF power  current density  output  A low RF power  output causes slow heating to the tissue  See Figure 1      Figure 1 Figure 2       At high RF power output  the tissue can heat rapidly  Excessively high RF power output can  overheat the tissue  causing evaporation of the water bound in the tissue  see Figure 2   This can  result in charring of the tissue  Therefore  excessively high RF power output is not recommended   Charring reduces the energy delivery to the heart tissue  which can cause other side effects   such as thrombosis or adhesions on the ablation electrode  It is recommended that ablation be  performed using a mode
114. st be  connected to separate ground points with their own grounding cables     See Section 5 1  Diagram of the Generator and Accessories  for an illustration of the typical  configuration and the cables required for undisturbed measurement quality     8 Technical Data   Page 96    RF Generator User Manual       9 Symbols    Symbol  Distributed by  Manufactured by     gt  lm        gt    gt          gt  gt  Ba kb O     lwo i       al      U ie         l    Description  The device is distributed by    The device is manufactured by    Quantity  Catalog number    Alarm    Rotational adjustment  Clockwise rotation increases value     Follow operating instructions     Follow operating instructions     Upper and lower temperature limits  storage and shipping conditions     Upper and lower humidity limitation  storage and shipping conditions     Upper and lower atmospheric pressure limitation  storage and shipping  conditions     Keep dry    Caution    To electrophysiology recording system  ECG     Press the button to release the lock and allow the front panel of the generator to  be tilted up   not available on some models     Indifferent electrode    The indifferent electrode and catheter connections on the device have zero  voltage  Connections identified with this symbol are floating  are not grounded      All patient connections of the device are defibrillation proof type CF applied parts   Radiofrequency    Potential equalization port    Fuses       9 Symbols   Page 97    RF Ge
115. ste poses environmental and health risks  Follow  local regulations for disposal of the SMARTABLATE    RF Generator and its electrical or electronic  accessories  To ensure proper disposal  there is also the option of contacting your Biosense  Webster representative regarding return of the generator and its accessories       aaa lease SSS aaa   4 System Overview and Setup   Page 21    RF Generator User Manual       5 Accessories    5 1 Diagram of the Generator and Accessories    Patient Area                          Front View          Outside  Patient Area                Outside  Patient Area                            5 Accessories   Page 22    RF Generator User Manual       1 SMARTABLATE    RF Generator   2 Catheter   3 CARTO   3 System   4 Catheter   5 CARTO   XP System   6 Catheter   7 Esophageal probe  not applicable for the US and JP versions of the  generator    8 ECG connection   9 Split indifferent electrode   10 Split indifferent electrode   11 Erbe indifferent electrode   12 SMARTABLATE    Irrigation Pump   13 SMARTABLATE    Remote Control   14 Foot pedal   15 Grounding cable   16 Mains cable   17 Personal computer or CARTO  System   18 Electrophysiology recording system    5 2 Accessory Lists    The following is a list of accessories for the SMARTABLATE    RF Generator     emia    Use only the original accessories provided by or recommended by the manufacturer of the  SMARTABLATE    RF Generator  If you use accessories that are not authorized by the manufacturer 
116. stem    The temperature is not within a  reasonable range     The esophageal probe or  connection cable is defective     connection cable   See Section 5 2 3   SMARTABLATE    Cables      Verify that the esophageal probe connection  cable is connected to the red Catheter 2  receptacle on the front of the generator   See  Item 5 in Section 4 2  Controls on the Front of  the Generator      Verify that the esophageal probe is compatible  with the generator  Verify that the connection  cable is correct for the type of esophageal  probe     Replace the esophageal probe and or  connection cable with a new one     7 Troubleshooting   Page 66    RF Generator User Manual       Table 7   Device Connections    Device Meaning of Icons   Catheter e  The catheter is connected to the generator and the temperature is within a    and Ablation  reasonable range for the room or for the patient    Cable to     Mapping Possible Reasons What To Do   System  if The catheter is not connected to Verify that the CONNECTIONS screen reflects  applicable  the generator  the mapping system that you are using    The wrong cable type is being  used     The catheter or a cable is  defective     7 2 Warnings  Alerts  and Ablation Stop Reasons     CaRTO   3 System  CARTO   XP System  or other      Verify that the catheter is connected to   the appropriate cable and that the cable is  connected to the red Catheter 1 receptacle  on the front of the generator   See Item 4   in Section 4 2  Controls on the Front of
117. t of a  typical commercial or hospital environment     Mains power quality should be that of a  typical commercial or hospital environment   If the user of the SMARTABLATE    RF Generator  requires continued operation during power  mains interruptions  it is recommended that  the SMARTABLATE    RF Generator be powered  from an uninterruptible power supply     Note that a 5 second removal of power  as  tested  results in the device shutting down  and requires user intervention to restart   There was no interruption of operation for  other tests     Power frequency magnetic fields should  be at levels characteristic of a typical  location in a typical commercial or hospital  environment     Note  U is the AC mains voltage prior to application of the test level  The SMARTABLATE    RF Generator    was tested at 100 and 230 VAC     8 Technical Data   Page 93    RF Generator User Manual SMART       Electromagnetic Immunity    The SMARTABLATE    RF Generator is intended for use in the electromagnetic environment specified  below  The customer or user of the SMARTABLATE    RF Generator should ensure that it is used in such an  environment     Immunity Test IEC 60601 Compliance Electromagnetic Environment   Guidance  Test Level Level  Conducted RF 3 Vrms 3V Portable and mobile RF communications  IEC 61000 4 6 150 kHz to equipment should be used no closer to any part of  80 MHz the SmartABLaTe    RF Generator  including cables   Radiated RF 3V m 3V m men u    n erp rena  IEC 61000 4 3 
118. t pedal test     Replace the foot pedal with a good foot pedal  and perform the foot pedal test     In this case  although the foot pedal may have  passed the foot pedal test  it cannot be used   Replace the foot pedal with a good foot pedal  and perform the foot pedal test     7 Troubleshooting   Page 65    RF Generator User Manual       Table 7   Device Connections    Device Meaning of Icons    Indifferent Solid indifferent electrode  Indicates that the indifferent electrode is connected to    Electrode the system     Split indifferent electrode  Indicates that the indifferent electrode is connected to  the system and contact with the patient is good enough for ablation     1 Possible Reasons    Solid or split indifferent electrode   The indifferent electrode is not  connected to the generator     Split indifferent electrode  The  contact quality on the patient is  not good enough for ablation     What To Do    Connect the indifferent electrode to the  generator     Improve the contact of the indifferent  electrode on the patient   See Section 3 4   Handling Indifferent Electrodes      Esophageal The esophageal probe is connected to the generator and the temperature is within    Probe  not a reasonable range for the room or for the patient    applicable for ym     the US andJP   S Possible Reasons What To Do   versions of the The esophageal probe is not Verify that the esophageal probe is connected  generator  connected to the generator  to the appropriate SMARTABLATE    Sy
119. t temperature      only in Power Control Mode  The temperature of the ablation  electrode at which a warning will occur  RF energy will remain on     The temperature of the ablation electrode at which RF energy will  automatically shut off      only for Amm catheters in Temperature Control Mode  The  sensitivity of the temperature control algorithm to changes in  temperature  how fast it reacts how aggressively it changes the  power to reach the target temperature     Esophageal Temperature Settings Pane  not applicable for the US and JP versions ofthe generator     Max   Warning   Cut Off    Select either Warning or Cut Off     If you select Warning  then select the temperature of the  esophageal probe at which you want a warning to occur  RF  energy will remain on     If you select Cut Off  then select the temperature ofthe  esophageal probe at which RF energy will automatically shut off     6 Operation  Presets   Page 42    RF Generator User Manual       Table 5   Preset Setting Descriptions  Setting Description    Rise Select either Warning or Cut Off     Warning   Cut Off   l  If you select Warning  then select the increase of temperature    of the esophageal probe at which you want a warning to occur  during ablation   the increase will be determined from a baseline  measurement taken just before ablation  RF energy will remain  on     If you select Cut Off  then select the increase of temperature  of the esophageal probe at which RF energy will automatically  shut off 
120. t with the tip  of a pen instead of your finger        7 Troubleshooting   Page 84    RF Generator User Manual    8    8 1    Technical Data    Specifications    Impedance Unit  Measurement current  Measurement range    Indifferent electrode  monitoring current    Measurement range    Ablation Unit    Temperature    Measurement range    Preset time  RF frequency  RF output    Measurement range  RF maximum voltage    Modes    Interface    Interface    Mains    Voltage    Frequency    Power consumption    Fuses       6 5 uA  45 kHz sinusoidal  0   500 Q   max 5   3 Q   1 Q resolution    4 uA  100 kHz sinusoidal  0   390 Q   20     User configurable target  up to 90  C  1  C increments  User configurable cut off  up to 95  C    10  C   60  C   1  C  61  C   100  C   2  C  catheter   10  C   50  C   1  C  esophageal probe    esophageal probe temperature is not available for US and JP versions     Up to 999 sec  non linear increments 5 to 30 sec  480 kHz   1 kHz    100 W   1009  version WW   70W  100 Q  versions JP and US    software limits output depending on catheter type selected     0   150 W   max 8   2W     60 150      290 V FY    Temperature Control  Power Control  manually set power     RS 232  USB max  current 300 mA    115 120 V   230 V  versions US and WW   100 V  version JP     50 60 Hz  350 VA    T3 15 AH 250 V    8 Technical Data   Page 85    RF Generator User Manual       Acoustic Error Signal    Sound pressure Greater than 65 db A     Operational Limits  6 hour con
121. ta Communication Cable for CARTO      Systems or Electrophysiology Recording  Systems  Redel 9  gt  Dsub 9  gray     length  3 m  length  5 m  length  15 m    Connection Cable for Smiths Esophageal  Probe  Rectangular 2 pin  length  3 m  not  applicable for the US and JP versions of the  generator     Connection Cable for Philips Esophageal  Probe  accepts mini  1 8     phone plug   length  3 m  not applicable for the US and JP  versions of the generator     Connection Cable from STOCKERT Testbox III    to an indifferent electrode    Part      M4900601  M4900609  M4900410  M4900108  M4900604  M4900606  M4900210  M4900603  M4900605  M4900607  M4900602  M490024    M490014  M490015    M490018    M490021  M490025  M490023  M490019    M490020    M490017    5 Accessories   Page 25       0 en eal  SED     i p        E  f     mme         RF Generator User Manual       5 2 4    5 2 5    Item    Connection Cable for CELSIUS   Catheters   TC THR  Redel 26  gt  Redel 10  red   length   3m    Connection Cable for CELSIUS   Catheters   DS  Redel 26  gt  Redel 10  green   length  3 m    Connection Cables from SMARTABLATE    RF  Generator to ablation catheters  available in  Europe   length  3 m    APTMedical Triguy D  Boston Scientific   EPT  Osypka   Osypka Easy TC  Osypka Plus Flutter  Bard   St Jude Medical   IBI   St Jude Medical   DAIG  Biotronik    Medtronic    Part    D130302    F E E    D130303    A E    M4900622  M4900610  M4900611  M4900620  M4900619  M4900612  M4900613  M4900614 
122. tem Master  see Section 6 7  System Control         6 Operation  Quick Overview   Page 33    RF Generator User Manual       6 2 Connections    CONNECTIONS Screen  Displaying Carto   3 System     Recording System          Mapping Pump  System  Remote Control  Catheter  Esophageal Foot Pedal  Probe       ECG    Indifferent Electrodes     The Esophageal Probe shown above is not applicable for the US and JP versions of the generator      Quick Steps     Press the CONNECTIONS workflow button     Press the image for any device  except the ECG cable  to view information about  the device and or to test the interface between the generator and the device   see Table 3   Device Screens in this section      Press Front Back to view the connections on the back of the generator     Press System Settings Ba to open the System Settings screen  see Table 4    System Settings Screen Options in this section      Note   See Section 7 1  Device Connections  for an explanation of the meaning of and  for each device connection     Fi  6 Operation  Connections   Page 34    RF Generator User Manual       Sample Device Screen    START  Lo ur F   Di    Serial Number        Table 3   Device Screens        Information Displayed Actior    Generator   Device name N A N A      Manufacturer    Serial number      Software and hardware  versions      Last and next preventive  maintenance dates      Customer name and    number   Mapping   Image of generatorand Change Press Change  then press the image   System
123. the plug housing  not by the  cable itself     5 3 2 Indifferent Electrodes    Follow the instructions in Section 3 4  Handling Indifferent Electrodes  and the instructions for  use that are provided with the indifferent electrodes     5 3 3 Foot Pedal    Never pick up the foot pedal by its cable  Do not coil the cable closely around the foot pedal   Check the cable regularly for visible damage  Do not use a damaged foot pedal     5 4 Cleaning  Disinfecting  and Sterilizing Accessories  To ensure that the accessories are safe to use  follow the cleaning  disinfecting  and sterilizing    instructions in the instructions for use provided with the accessories  Make sure that any liquid is  removed from electrical contacts after cleaning  disinfecting  or sterilizing     5 Accessories   Page 27    RF Generator User Manual       6 Generator Operation  6 1 Quick Overview    6 1 1 System Startup    Quick Steps     Connect the generator to the mains supply   Press Stand by to turn on the generator       At the message  press Stop        Optionally  turn on the remote control     For irrigated catheters  set up and turn on the pump     Notes   1  When you press Stand by ATB  make sure that Ablation Start    and Stop  and the foot pedal are not pressed     2  When you turn on the generator  the generator   s audible and visual indicators  are tested to allow you to verify that they are functional  The Stand by  indicator lights up and the SmarTABLATE    System logo appears on the scree
124. tinuous operation with intermittent ablations  Maximum 75  of the time performing ablations at maximum of 100 W output    Maximum continuous ablation time at full power less than 30 minutes    Other   Classifications Class land defibrillation proof type CF  as defined in IEC EN 60601 1   Class Ilb  as defined in MDD 93 42 EEC   Class 3  FDA   Weight 18 5 kg   Dimensions 399 mm wide x 178 mm high x 344 mm deep   IP rating IP21    8 2 Storage and Shipping Conditions    Ambient temperature  30   to  65  C  Relative humidity 10  to 90   non condensing     Atmospheric pressure 500 to 1060 mbar    8 3 Operating Conditions    Ambient temperature 10   to 40  C  Relative humidity 10  to 85   non condensing     Atmospheric pressure 700 to 1060 mbar    8 4   THERMOCooL   Catheter Preset Ranges    THERMOCOOL   Catheter THERMOCOOL   SF Catheter  Power Pump Default Power Pump Default    W   mL min   W   mL min   1   10 17 1   10 8  11   20 17 11   20 8  21   30 17 21   30 8  31   40 30 31   40 15  41   50 30 41   50 15    8 Technical Data   Page 86    RF Generator User Manual       8 5 Alarm and Touch Specifications    After an alarm event occurs  there may be a delay of not more than 500 ms before the related    sound  light  and or message are produced     All alarm events are recorded and stored in the SMARTABLATE    RF Generator  If power is lost  alarm  event data is maintained on the generator for a minimum of 30 days  This data can be exported   copied to a USB flash drive and delet
125. to change values on the generator screens  When you press    a displayed value on the touch screen  the ring around the Data Entry Knob O turns on  The  color of the ring matches the color of the value or the color of the background surrounding the  value     When you turn the Data Entry Knob O    the value changes  Press the knob to save the new  value  The ring around the knob will turn off  You can also save the new value by pressing a  different displayed value or by waiting until the ring around the knob turns off by itself     Note  At the ABLATION screen  the changed values take effect immediately  it is not necessary  to press the knob      6 1 6 Foot Pedal    If a SMARTABLATE    System foot pedal is connected to the generator  the foot pedal can be used to  start and stop ablation  see Section 6 4 2  Starting and Stopping Ablation      Note  If a SMARTABLATE    Irrigation Pump is connected to the generator and a SMARTABLATE     System foot pedal is connected to the pump  that foot pedal controls only the pump     6 Operation  Quick Overview   Page 32    RF Generator User Manual       6 1 7 Typical Procedure Process   Quick    Quick Steps     Physically connect devices to the generator     Turn on the generator   Press the PRESET workflow button     Select a preset   Press the ABLATION workflow button     Press Ablation Start   s to start ablation     Note   Ata minimum  connect a catheter and an indifferent electrode to the generator   For an irrigated catheter  also
126. to the selected bag size     6 Operation  Ablation   Page 55    RF Generator User Manual       6 5 Post Procedure    POST Screen    Procedure  1    Date  06 September 2010    Total Ablations  4    Total Ablation Time  00 03 13  Total    Total Energy   Total Irrigation         The Max  Esophageal temperature shown above  is not applicable for the US and JP versions of the generator      Quick Step   Press the POST workflow button     Notes   1  The POST screen displays a summary of the data collected by the generator  during an entire procedure  To indicate the start of a new procedure  to restart  data collection   press New Procedure  see 6 4 1  Starting a New Procedure      2  The Total Irrigation value shown on the POST screen is the total irrigation  delivered during the procedure  excluding flushing     6 5 1 Exporting Ablation and Event Log Data to a USB Flash Drive       Export Ablation and Event Log Data  Insert USE flash drive  then select the amount of data to export     Today Last 7 Days    Export status       complete    Time remaining          Cancel Export         6 Operation  Post Procedure   Page 56    RF Generator User Manual       Quick Steps     Insert a USB flash drive into the USB port on the back of the generator     Press the CONNECTIONS workflow button       Press System Settings aa     Press Export to USB     Press Today  Last 7 Days  or All       When the export status is 100  complete  remove the USB flash drive     Notes   1  The USB port is locate
127. toring  instruments  These signals in combination with the temperature of the ablation electrode  tissue  impedance  and RF energy information are used by the operator to control the safe delivery of RF  energy to locations within the heart that require treatment     The SmarTABLaTE    RF Generator was designed to ensure simple  intuitive  and user friendly operation   This device features a touch screen for making selections and setting parameters  Among the  features of the screen is a display that shows power  temperature  and impedance in color coded   real time  continuous graphics during ablation  The generator features redundant protection circuits   if one safety mechanism fails  a second safety mechanism indicates the failure to ensure safety    Critical safety functions are controlled by buttons that are operated independently of the touch    screen  A prominent red Stop button has priority over all other functions and connected devices   thus immediately stopping RF energy delivery when pressed     The SmarTABLaTE    RF Generator is connected to the SmarTABLATE    Irrigation Pump by a  communication interface that allows both devices to be controlled from the user interface of the  generator  If an error occurs in either of these devices  the necessary notification and recovery steps  are initiated simultaneously in both devices     This user manual describes the operation of the SMARTABLATE    RF Generator  It is important that  the user understand the functional 
128. ty  and essential performance   Collateral standard  Electromagnetic compatibility    Requirements and tests    Medical electrical equipment   Part 1 4  General requirements for  safety   Collateral Standard  Programmable electrical medical systems    Medical electrical equipment   Part 1 6  General requirements for basic safety  and essential performance   Collateral standard  Usability    Medical electrical equipment   Part 1 8  General requirements for basic safety  and essential performance   Collateral Standard  General requirements  tests  and guidance for alarm systems in medical electrical equipment and medical  electrical systems    Medical electrical equipment   Part 1 9  General requirements for basic  safety and essential performance   Collateral Standard  Requirements for  environmentally conscious design    Medical electrical equipment   Part 2 2  Particular requirements for the basic  safety and essential performance of high frequency surgical equipment and  high frequency surgical accessories    Medical device software   Software life cycle processes    8 Technical Data   Page 91    RF Generator User Manual       Standard Title    IEC EN 62353 Medical electrical equipment   Recurrent test and test after repair of medical  electrical equipment    IEC EN 62366 Medical devices   Application of usability engineering to medical devices  MDD 93 42 EEC Medical Device Directive  MDD  Council Directive 93 42 EEC of 14 June 1993   MPG  Medical concerning medical devices
129. watts  W  according to the  transmitter manufacturer     Note 1  At 80 MHz and 800 MHz  the separation distance for the higher frequency range applies     Note 2  These guidelines may not apply in all situations  Electromagnetic propagation is affected by  absorption and reflection from structures  objects  and people     EMC Information and Technical Description    MEDICAL ELECTRICAL EQUIPMENT needs special precautions regarding EMC and needs to be  installed and put into service according to the EMC information provided in this user manual     Portable and mobile RF communications equipment can affect MEDICAL ELECTRICAL  EQUIPMENT      N warning  The use of ACCESSORIES  transducers  and cables other than those specified  with  the exception of transducers and cables approved by the manufacturer of the SMARTABLATE    RF  Generator as replacement parts for internal components  may result in increased EMISSIONS or  decreased IMMUNITY of the SmartABLaTE    RF Generator     AN Warning  The SMARTABLATE    RF Generator may be operated in the immediate vicinity of the  SMARTABLATE    Irrigation Pump  However  do not operate the generator in the immediate vicinity  of devices that are not approved for electromagnetic compatibility  EMC  with the generator  If  such operation is unavoidable  regularly monitor the generator to ensure proper functioning        8 Technical Data   Page 95    RF Generator User Manual        N Warning  The SMARTABLATE    RF Generator may be interfered with
130. y trademarks listed herein are trademarks of their respective owners     Caution  Federal  USA  law restricts this device to sale by or on the order of a physician     SMARTABLATE    RF Generator  Firmware 1 XX XX       Biosense Webster  Inc  2012 2013    January 2013    SAG_UM_04 00_EN    
    
Download Pdf Manuals
 
 
    
Related Search
    
Related Contents
VP-71 [仕様変更について] 自己保留/転送/パーク保留 1 2  Cisco Systems ASR 1004 User's Manual    Copyright © All rights reserved. 
   Failed to retrieve file