Operators Manual SMon5010
Contents
1. 69 RIELE Operating Instructions Serial Monitor 5010 ROBERT RIELE GmbH amp Co KG Software Version 3 14 Documentation Version 01 2003 SMonDocu doc 20 01 03 The software Serial Monitor 5010 was developed for the RIELE product Photometer 5010 upwards serial number 3500 The operator controls the photometer via a PC which employs menu driven soft ware Several tasks can be performed e Like the method editor of Photometer 5010 the software Serial Monitor 5010 supports the adapting of methods to the photometer With special functions the user can upload and download pre programmed methods Additionally methods can be created and edited on the PC Customer defined methods can be stored in an external file e The operating of Photometer 5010 by the remote control can be demon strated The user can easily recognize the meaning and performance of the defined instruction set e General operating software contained in FLASH MEMORY of Photometer 5070 can be updated Data file can be downloaded from Internet floppy disk or CDROM e The EDP can be demonstrated in a special terminal window The contents of the window can easily transferred to other applications ROBERT RIELE GmbH amp Co KG Page 2 Introduction to Serial Monitor 5010 SMonDocu doc 20 01 03 This chapter describes requirements and procedures for starting Serial Monitor 5010 2 1 System requirements The minimum and recommended system requirements needed t
2. as follows Baud rate 9600 bit s Data format 7 data bits 1 stop bit Parity even For further information see Service manual of Photometer 5070 chapter 6 1 Communication protocol 4 4 Set device selects different models of Photometer 5070 ROBERT RIELE GmbH amp Co KG Page 6 Menu CONFIGURATION SMonDocu doc 20 01 03 4 5 Download operating system updates general operating software contained in FLASH MEMORY of Photometer 5010 The general operating software is supplied as a file The actual file has the name P_100 hex It can be downloaded by internet see http www riele de Products Photometer 5010 Operating system Besides specific files with different names are existing and can be used Before proceding with uploading the specific file of general operating software should be copied to the directory DI_5010 see chapter 2 3 Organisation of software Download operating system to photometer X Password 100 Cancel Before starting the procedure of download the Photometer 5010 has to be set to DOWNLOAD by MODE SERIAL 5 and ENTER The Photometer 5010 displays DOWNLOAD Attention During the following procedure the FLASH MEMORY of Photometer 5070 will be erased and programmed electrically The complete proce dure needs up to seven minutes During this operation be careful that there is no power fai
3. system initialization by applying the tool 1 With this function the default parameters will be set e Make a free methods initialization by applying the tool S2 This operation is optional Attention With this function all methods edited by the user are deleted without call back In the case of doubt these methods should be checked before Scroll through all methods with the method editor and note them down for an optional later new programming e Make a parameter initialization by applying the tool S3 This operation is optional Attention The extended memory for the parameters of all nonlinear methods is deleted by this function without call back and the corresponding user defined methods become useless e Make a setting of attribut of PCB layout by applying the tool 13 Software applications may be dependent on this attribut e Make a patients data initialization by applying the tool 14 Attention The memory of patients data mangement is cleared completely In case of a migration to operating system version 3 5 the operation has to be done e Define the laboratory name by applying the utility U18 The name of the laboratory can be stored permanently In case of a migration to operating system version 3 5 the operation has to be done e Define the user names by applying the utility U19 The names of maximum five users can be stored permanently In case of a migration to operating system version 3 5 the op
4. OUT TO SCREEN all instructions are embedded in a STX ETX BCC protocol Each information represented by an ASCll string starts with a start character start of text STX and stops with an end character end of text ETX After each ETX a block code character called BCC is send The BCC is the result of a XOR function including the information and the ETX Protocols are accepted by the receiver with an acknowledgement ACK or a nega tive acknowledgement NAK character e Select the menu PRINTOUT TO SCREEN e Activate on Photometer 5070 the utility program EDP by MODE SERIAL 2 and ENTER The Photometer 5070 displays Die Oleg 1 ON 0 OFF e Select the option 1 Photometer 5070 displays EDP ON 0 6 B 5 1 CR LE e Select the option 0 and confirm with ENTER Photometer 5010 displays METHOD NO TPNEUT 12350 e The program shows the following window after selecting e g method 1 C F PHOTOMETER 5010 3 2a 31 05 2002 DATE 06 09 02 TIHE 16 43 12 HETHOD 1 ir FACTOR 36 8 HANELENGTH S46nn TEMPERATURE OFF DELAY s HEASURE BLANK ABS SULT ROBERT RIELE GmbH amp Co KG Page 12 Menu TERMINAL SMonDocu doc 20 01 03 The menu METHODS enables the user to upload and download pre programmed methods Additionally methods can be created and edited on the PC All methods are stored in a database called Methods dbf situated in t
5. al Methods l COM1 online Method no B5 ERRRD _ ERRTD _ _ Llose ERR TOut i ROBERT RIELE GmbH amp Co KG Page 14 Menu METHODS SMonDocu doc 20 01 03 6 3 Edit The browse view write protected Monitor Photometer 5010 iol x Eile Edit Help Editor for methods TORT MECE EN 405 nm Browse view Form view Ji dee teen ne ne ann ui B00 aos m Delete record pie pen foou ann 90 joe nere p few pem en Pre Append record ee en ou Go bottom Next record Previous record TAMYLeosll 1 4 per p e m Pe s ass un ooo Som 705 are 46 ASATifec yk l 800 ul 340 nm 29 10 37 C a ke The form view for fields to edit Monitor Photometer 5010 lolx Eile Edit Help a aale doa Editor for methods Method no Ei Name JAcPHOSna Dimension General UA FF Uninear Filter 2 405 nm Protected Factor Standad 740 E C F SbRb a G S C S Rb No of Standards gt C S Sb Temperature gt C S SbRb Volume fiz feco ul FTK F Rb C OFF open FTK S Ab TO s 30 C 25 C KIN S Ab a PIs TRANSM T1 Deltas Time s 13 E 24 C C Minimum 2 fo Maximum fo ROBERT RIELE GmbH amp Co KG Page 15 Menu METHODS SMonDocu doc 20 01 03 6 4 clear deletes and packs all records 6 5 pack packs the database Methods dbf 6 6 Organization of the worki
6. eration has to be done 4 6 Exit quits the application ROBERT RIELE GmbH amp Co KG Page 9 Menu CONFIGURATION SMonDocu doc 20 01 03 The items of menu TERMINAL demonstrate the operating of Photometer 5010 by the remote control and EDP 5 1 Remote control The menu REMOTE CONTROL gives the user the possibility to remote the photo meter All instructions are embedded in a STX ETX BCC protocol Each informa tion represented by an ASCll string starts with a start character start of text STX and stops with an end character end of text ETX After each ETX a block code character called BCC is send The BCC is the result of a XOR function including the information and the ETX Protocols are accepted by the receiver with an acknowledgement ACK or a nega tive acknowledgement NAK character e Select the menu REMOTE CONTROL e Activate on Photometer 5070 the utility program REMOTE by MODE SERIAL 3 and ENTER The Photometer 5070 displays REMOTE ON e The program shows the following window Monitor Photometer 5010 O x Configuration Terminal Methods l COM1 online Send button for instruction Close button for interface Single line edit for instruction History of data dialog Actual COM port AC ComOpen 1 RC ComSetNotifyMessageAnyChar 1 ROBERT RIELE GmbH amp Co KG Page 10 Menu TERMINAL SMonDocu doc 20 01 03 Example for using remote instr
7. he subdirec tory Files The data base has the format of a dBase Ill file The file may be re placed with a pre programmed file 6 1 read from photometer fills the database Methods dbf with all user defined methods The function also reads the status of the photometer and defines the range of the free programma ble methods If the database is not empty the methods read will be merged with the existing ones e Select the menu READ FROM PHOTOMETER e Activate on Photometer 5070 the utility program REMOTE by MODE SERIAL 3 and ENTER The Photometer 5070 displays REMOTE ON e The program shows the following window during reading from Photometer 5010 Monitor Photometer 5010 Configuration Terminal Methods l COM1 online ERRAD ERRTxD _ __ Shove ERR TOut LI ROBERT RIELE GmbH amp Co KG Page 13 Menu METHODS SMonDocu doc 20 01 03 6 2 write to photometer stores all user defined methods of the database Methods dbf to the photometer The database methods will be merged or overwritten with the existing ones of Photometer 5010 e Select the menu WRITE TO PHOTOMETER e Activate on Photometer 5070 the utility program REMOTE by MODE SERIAL 3 and ENTER The Photometer 5010 displays REMOTE ON e The program shows the following window during writing to Photometer 5010 z Monitor Photometer 5010 Configuration Termin
8. l out Do not interrupt the connection between PC and Photometer 5010 The password is 5010 ROBERT RIELE GmbH amp Co KG Page 7 Menu CONFIGURATION SMonDocu doc 20 01 03 DownLoad operating system to photometer Select the first option Download to photometer and confirm with ENTER Attention Do not select the second option Resume after malfunction without having a malfunction In case of a malfunction the Photometer 5010 displays more than10 minutes after start DOWNLOAD Vx x Contact your dealer for getting help PLEASE WAIT Offnen ae 21x Som A A fea An open dialog is shown Open the file e g P_100 hex Never open the file D_load hex The download begins The window is closed and the progressbar shows the proceding of operation After three up to seven minutes the Photometer 5010 stops the transferring of data with a reset operation The new operating system is working ROBERT RIELE GmbH amp Co KG Page 8 Menu CONFIGURATION SMonDocu doc 20 01 03 After updating the general operating software of Photometer 5010 several initialization procedures have to be done by using service tools and utilities A detailed description of the service tools is contained in the service manual of Photometer 5010 see chapter 3 4 Service tools A detailed description of the utilities is contained in the operating manual of Photometer 5010 see chapter 8 Utilities e Make a
9. ndows is updated The software product is named with Serial Monitor 5010 Also a special EXE file can be used for deinstallation The name of the file is Caindreg exe Its location is the directory of the application Serial Monitor 5010 In any case the file Method dbf containing all application data will not be cleared The location of this file is in the directory Files ROBERT RIELE GmbH amp Co KG Page 4 INSTALLING SMonDocu doc 20 01 03 Before starting the software connect PC and photometer in following manner e Switch off PC and Photometer 5010 e Connect PC and Photometer 5010 with a data interface cable For further information see Service manual of Photometer 5070 chapter 6 1 Communication protocol e Switch on PC and Photometer 5010 e Start the program Serial Monitor 5010 EXE e The program shows the following window e At the first time control and correct the language and COM port with the menu CONFIGURATION ROBERT RIELE GmbH amp Co KG Page 5 STARTING SMonDocu doc 20 01 03 The menu CONFIGURATION makes the setting of different parameters possible The settings are saved by the program and therefore correctly adjusted at the next program call 4 1 Deutsche Sprache selects German language 4 2 English language selects English language 4 3 Configure Interface selects COM port baud rate parity data bits stop bits The default settings of the serial interface are
10. ng memory of methods The working memory of Photometer 5010 is divided in two areas e The area of write protected methods including basic methods with all available calculation procedures The area starts with method 1 End point with factor With the function Methods write to photometer it is not possible to change the contents of this area e The area of not write protected and write protected methods With the function Methods write to photometer it is possible to change the con tents of this area 6 7 Write protection of methods A method of Photometer 5010 can be protected against clearing und editing The form view of the edit mode allows setting a checkbox with the name protected Setting and resetting of the checkbox is only possible with password After writing to photometer a protected method can be cleared in the database But the protected method of the Photometer 5010 can not be cleared ROBERT RIELE GmbH amp Co KG Page 16 Menu METHODS SMonDocu doc 20 01 03 The menu employs the item e Info shows information about application name and version Information ROBERT RIELE GmbH amp Co KG Page 17 Menu
11. o install and run the program are as follows Photometer Version 3 0 or higher Windows 95 Version 4 00 950 or higher Windows NT Version 4 0 or higher 2 2 Installation of software e Insert the RIELE CD into the CD ROM drive e Inthe run dialog box enter e g D Riele_CD Ph5010 Smon5010 setup exe Ausfuhren 21x Geben Sie den Namen eines Programms Ordners Dokuments oder einer Internetressource an ffnen pr iele_CD Ph50104SMon501 O setup exe v Abbrechen Durchsuchen e Select the button lt OK gt to invoke the Serial Monitor 5010 Installer e The Serial Monitor 5010 installation program starts Serial Monitor 5010 To install Serial Monitor 5010 enter a new path below or click Continue to install into the default directory C Cavo25AR SMon5010 e Confirm or enter new path of Serial Monitor 5010 C Cavo25RR Smon5010 e Select the button lt Continue gt to run the installer and follow the instructions ROBERT RIELE GmbH amp Co KG Page 3 INSTALLING SMonDocu doc 20 01 03 2 3 Organisation of software Serial Monitor 5010 software is completely installed in a directory e g C Cavo25RR Smon5010 DI_5010 Files Other directories like Windows System are not changed or affected The registry of Windows is updated The directory C Ca_appsw is empty and may be deleted 2 4 Deinstallation of software Use of the official deinstallation tool of Windows is recommended as the registry of Wi
12. uctions e Enter the character B to the single line edit for instruction and push the button Send The command B gives the information about the firmware of Photometer 5010 e The PC sends a protocol B to Photometer 5010 In history of data dialog the line appears TxD B TxD means Transfer of data from PC to Photometer 5010 e Photometer 5010 accepts the instruction with an acknowledgement character In history of data dialog the line appears RxD ACK RxD means receipt of data from Photometer 5010 e Photometer 5010 sends to the PC the information about the software version In history of data dialog the line appears RxD B V3 2a 31 05 2002 1 3 1 0 e The PC accepts the instruction with an acknowledgement character In history of data dialog the line appears TxD ACK e After a complete dialog Photometer 5010 sends MO Module 0 MO and C Command C and is ready to accept a new instruction ST tox Configuration Terminal Methods l COM1 online D ACK Fil TxD ACK RxD M0 TxD ACK RxD C TxD ACK RC ComOpen 1 AC ComSetNotityMessageAnpChar 1 For further details see Service Manual chapter 6 1 communication protocol ROBERT RIELE GmbH amp Co KG Page 11 Menu TERMINAL SMonDocu doc 20 01 03 5 2 Printout to screen The menu PRINTOUT TO SCREEN shows the operating of EDP mode The EDP mode of Photometer 5010 supports STX ETX BCC protocol or CR LF protocol In menu PRINT