        Operating Instructions Laboratory Mortar
         Contents
1.  lift the pestle by pushing up the weight on its rod     4  After having closed the lid locking system slowly release the weight on its rod on the  sample to be ground     5  By sliding the weight on its rod  the grinding pressure on the sample can be se   lected and thus the load of the pestle can be adapted to the hardness and the  grindability of the sample      Front position  ca  3 kg load   rear position  ca  9 kg load    6  Then close the cover     Laboratory Mortar   Grinder  pulverisette 2     Page 12    FRITSCH    Advice   For preliminary crushing of larger pieces you may slowly release the pestle on the  sample in order to break them before the pestle reaches its normal operating po   sition and can be arrested     8 2 Collecting the ground sample    1  After completion of the grinding process turn the timer in its    OFF    position or the  timer stops the unit by itself     2  Lift the cover and lift up the rod with the grinding weight before the lid is opened   Then lift the pestle out of the bowl by opening the lid until the rod comes to a stand  at the rear side     3  Remove the bowl out of its recess by hand and collect the sample     Advice   You may collect your sample more complete by lifting the lid only a little and clean  up the pestle from sticking sample f e  using a brush  The sample will drop directly  into the bowl     9 Grinding during forced cooling    Whenever material which is very heat sensitive is to be ground  we recommend a for   ced cooli
2.  on itself by gentle pressure  against the radius of the bowl  If the pestle does not rotate  than the setting must be  corrected     5  Re tighten the 3 mm hexagon socket screw firmly     Laboratory Mortar   Grinder  pulverisette 2     Page 11    FRITSCH    7 3 Scraper    The Vulkollan scraper can be adjusted horizontally with the spherical button and verti   cally by loosening and re tightening the side knob     A new scraper has to be ground before use     e The part in contact with the bottom of the bowl must be ground so that its rear  edge  in eye direction  touches the bottom     e The part of the scraper in contact with the wall must be ground so that its front  edge  in eye direction  touches the bottom and the scraper wipes off the sample  from the bowl     8 Grinding    Instructions  Before closing the lid  make certain that the grinding set is firmly seated   e Close lid securely   e Pay attention to heating of material to be ground     in case of longer run time  possibly stop to cool down   In order to go easy with the grinding parts     Laboratory Mortar Grinder with lid in closed position should not be operated  the lid being closed without sample to be ground     8 1 Charging the sample    1  Open the cover and the lid of the Laboratory Mortar Grinder and fill in the sample  into the mortar bowl  The initial feed size should not exceed 3 mm and the quantity  should not be more than 150 ml     2  Switch on the instrument   3  Then close the lid  By doing this 
3.  overload   f e  after blockage of the mortar   the drive motor is protected  by an overload switch on the front panel     To continue the grinding process   after release of the blockage   press the button in  the middle of the switch     4 3 Fuses    There is a fuse in the connection plug at the rear side of the grinder  above the inlet  for the power cable     Replacement  5 x 20 TA4    Laboratory Mortar   Grinder  pulverisette 2     Page 8    WFRITSCH       5 Front Panel       WFRITSCH mortar grinder  LL   pulverisette 2    5 1 Power on   The green lamp  Power On  will light up after you turn on the master switch  black  switch  on the rear side    5 2 Activating the Laboratory Mortar Grinder    Timer    1  Setting the run time  Set knob operated switch of timer to run time    2  Setting constant operation  Set knob operated switch of timer to continuous  operation  ON     3  Maximum grinding period  1 hour    4  Before restarting  examine the grinding conditions    5  When operating with external timer  examine the temperature of the sample     Shutting down  Wait until time set on the timer has elapsed  or turn knob operated  switch on timer to  OFF      5 3 Overload switch    The motor is protected against overload   f e  after blockage of the mortar   by an  overload switch     To continue the grinding process   after release of the blockage   press the button in  the middle of the switch     Laboratory Mortar   Grinder  pulverisette 2     Page 9    FRITSCH    6 Selec
4. 00  92 607 up    Labor M  rsermiihle P2 100 110 200 220 240V    laboratory mortar grinder p2    Bezeichnung   Haube  Acrylglas   Lagerkopf   Pistillwelle mit Lagerung  Sichtscheibe  Acrylglas  V Ring  V 170A  VerdeckverschluB  Kondensator  16uF  240V MP  Getriebemotor  220 380V DS  Flachrundschraube  M6x35  Schaber lasche   Kugelknopf M5 D 16  Randelmutter M6  Gewindestange M10  Randelmutter M10  Schaberhalterung  Innensechskantschl  ssel Gr 3  T  rscharnier   Gewicht   Ringkerntrafo 250 VA  Kaltger  testecker   Schalter 1 polig 5 Raster  Signalleuchte  gr  n  Glimmlampe 220V    berstromschutzschalter 1 2A  Knebelknopf D 40mm kpl   Zeitschalter  Glasrohrsicherung 10A T    bersteckkappe D 10    O b   Le La eeu 4  dy 4    Un lee Moo     94 406 vO Uu len Fur Lape    18 12 91    Description   acrylic hood   bearing head   pestle shaft with storage  acrylic observation glass  v ring V 170 A   covering seal   capacitor  l6uF  240V MP  motor  220 380V 3 ph   saucer head screw  M6x35  scraper mounting  spherical knob   knurled knob M6   threaded rod M10   knurled knob M10   lid mounting   wrench  size 3   hinge   slider weight  transformer 250 VA   mains input socket  switch 1 pole 5 latches  signal lamp  green   glow lamp 220V  over current relay 1 2A  knob for timer   timer   fuse 10A slow   cap D 10    34481090   j27 aSl48Ajnd    NIJLSH38D 3V0I 0869 a  HOW HISLIAA    HdWD HOSLIHA  Ad  LH9IUAIOI    YA 0S2  HaWJOjSUDJ  JOD auo    03047  40ds   usarnBuly    VS 36042 0A 3ndu    434  0
5. WFRITSCH    Operating Instructions    Laboratory Mortar Grinder   gt  pulverisette 2           Fritsch GmbH    Manufacturers of  Laboratory Instruments    FRITSCH    Operating Instructions  Version of December 1  1995   Valid from Ser  No  5000     Copyright 1995     Diese Betriebsanleitung ist auch in Deutsch erhaltlich     FRITSCH GmbH Telephone 06784700  Manufacturers of Laboratory Instruments Fax 067847011  Industriestra  e 8 Telex  426203fritsd    D   55743 Idar Oberstein    Fritsch GmbH    On June 24  1995  Fritsch GmbH was certified by the TUV Zertifizierungsgemeinschaft  e V      gt     DIN ISO 9001    An audit was conducted  Report No  Q94 107  to verify that Fritsch meets the  requirements of    DIN ISO 9001   EN 29001     Certificate No  71 100 4 008        Laboratory Mortar Grinder    pulverisette 2   Page 1    Table of Contents    10 3  10 4    11  12    Safety Signs  Application and Use  Safety Instructions   Method of Operation   Operating Safety   Setting up   Electrical Connections   Checking for accordance of voltage and current  Motor   Fuses   Front Panel   Power on   Activating the Laboratory Mortar Grinder  Overload switch   Selection of mortar bowl and pestle  Installation of the grinding set  Mortar bowl   Pestle   Scraper   Grinding   Charging the sample to be ground  Collecting the ground sample  Grinding during forced cooling  Maintenance   Cleaning   Re grinding the grinding set   Drive motor   Instrument   Warranty   Technical Data   List of Spare
6. Y4DSuN     SsBunuuvodssBuvGBuj3    Maul       unyjoy gt sy1az       
7. at the hinges        Laboratory Mortar Grinder  pulverisette 2   page 5    FRITSCH    Operating Instructions  Laboratory Mortar   Grinder   gt  pulverisette 2        The    pulverisette 2    is an automatic mortar grinder for the universal dry  or wet grind   ing down to analytical size     When preparing samples for the synthesis  the Laboratory Mortar Grinder is used for  mixing or homogenising dry samples  emulsions or pastes     The Laboratory Mortar Grinder grinds inorganic as well as organic samples for the  analysis  quality control and material testing as for example     Plants leafs coniferous needles spices nuts  almonds food gelatine sugar grain pellets  feed drugs dragees tablets soil coal  ashes chemicals fertilisers chalk clay porcelain  gypsum lime clinker cement silicate cinder    The initial feed size should not exceed 3 mm and the quantity of a charge should not  be more than 150 ml   Larger pieces shall be reduced to small pieces before charging  into the grinder      The final fineness will depend upon the grinding time   1 um as a mean particle size  can be reached     Laboratory Mortar   Grinder  pulverisette 2   Page 6  951201VB    FRITSCH    1 Method of Operating    The size reduction from dried foods with hand mortar and pestle is the oldest known  grinding procedure  Archaeologists in Jordan have found a mortar with stage whose  age could be determined at approx  10 000 years     Similar to hand mortar and pestle by using the Laboratory Mortar Grinde
8. et        When the lid is open  the mortar bowl can be removed and can easily be cleaned   i e   under running water  The pestle and the scraper are easily accessible when the lid is  open and grinding residues are to be removed  Use a brush too if necessary     Residual ground material on the rubber seal at the lid or  at the border of the mortar bowl may destroy the lip seal    We can also recommend alcohol as an organic cleaning media  Other solvents should  not be utilised   See paragraph 7 as to dismantling the pestle and the scraper      10 2 Re grinding the grinding set    After a long use scratches or roughness may appear on the grinding surfaces of the  grinding set  In such cases  charge about 20 gr  carborundum  silicon carbide  of 0 5  mm particle size and about 20 ml water into the bowl  close the lid and let the instru   ment operate for about 30 minutes to 1 hour  The silicon carbide paste in the mortar  bowl will re grind the grinding set to a great extend by abrasion     10 3 Drive motor  The drive motor does not require any maintenance     The switches installed on the grinder should be checked for function  smoothness of  action  in event of contamination     10 4 Laboratory Mortar Grinder    The Laboratory Mortar Grinder itself does not require any maintenance  The bearings  of the rotating components are permanently lubricated     Laboratory Mortar   Grinder  pulverisette 2     Page 14    FRITSCH    11 Warranty    The warranty card enclosed with this ship
9. ment must be returned to the supplier com   pletely filled out in order for the warranty to take effect     Fritsch GmbH of Idar Oberstein and our  Application technology laboratory  and or our  agent in your country will be glad to advise and help you     In order to answer any questions you may have  we will require the serial number  stamped on the nameplate     12 Technical Dates  Weight with mortar bowl and pestle  30 kg  Dimensions  H  W  D  450  250  450 mm    Power consumption at 22  C and 225 Volt       20 87 A  Power P   145 W  max  Power Pmax   165 W    Laboratory Mortar   Grinder  pulverisette 2     Page 15    ajyanwjasiaOWogoi  Z ayasuaa  nd    noq     01610q07  Hqw9 yosiy    Ladd    NN e KNU    as  SA AA  N  LZ    aal    444   Sew L        A    RSSSSSSSSSSSSSSSSSSSSSSSSSSSSS       4    SE QZZZZZZZZZZZZZZZZZZZZZZZZZZZZZZZZZZZZZZZZZZZZZZZZZZZZZZZZZZZZZZZZZZZZZZZ         Ersatzteilliste    spare parts list    Pos  1  2    o Oo N On ao A w    11    12  15    17  18  19  20  21  22    23  24    25  26    Art Nr   02 126   02 135   02   02   84   93   81   81   92   02   92   92   02   92   02   83   92   02   81   81     81    86   81     81  81    81     140    129   405   910    00    311    220   148   114   606   508   141   510   112   405   908   118     131    609    430   86     417    422   517    403    479   507     17  10     00    17  15  00    00  09  10  16  00  10  00  13  09  00  10     00    00  00     00    00  00  00  00  00    93 404 16    92 07 
10. ng with liquid nitrogen  The liquid nitrogen is introduced directly into the bowl  together with the material to be ground      Dry ice   solid carbon dioxide  can also be used  Keep in mind  however  that dry ice  frequently contains frozen condensation water on its surface     Caution  Wear safety glasses when working with liquid nitrogen        With some materials to be ground  grinding can be improved by using coolants   prior to grinding  For example  animal tissue or special plastics can be rendered brittle  by dipping them briefly in to liquid nitrogen or freezing them in a deep freeze before   hand  Here  too  you must protect the material to be ground against condensing water   For example  you could place the material in a PVC bag during the cooling procedure     Laboratory Mortar   Grinder  pulverisette 2     Page 13    FRITSCH    10 Maintenance    The most important process in preventive maintenance is regular cleaning     Caution     The cleaning and maintenance of the Laboratory Mortar  Grinder has to be done in accordance to the national safety  regulations   specially when the environment contains cer   tain amount oft dust so that dust explosion could happen          10 1 Cleaning of the Laboratory Mortar Grinder    The frequency with which a thorough cleaning must be performed depends on the  grinding material and its final fineness  We recommend adapting the cleaning intervals  to the use of the Laboratory Mortar Grinder by making regular checks at the outs
11. parts   Wiring Diagram    Laboratory Mortar Grinder    pulverisette 2     U  W             oOoOOOO Oo Oo OO OO ON NN R W    FRITSCH    Page 2    Safety Signs    Application and Use       Safety signs indicate dangers     Caution   Combustible ma  Materials detri  Caution Warning of dan   terials mental to health gers    A    Danger  electri  Wear eye protec   cal power tion    Use of the Laboratory Mortar Grinder    This operator   s manual describes the area of application and  the use of the Laboratory Mortar Grinder     Any use other than for grinding with the grinding parts  manufactured by Fritsch is impermissible    Modifications may be made to the Mortar Grinder only after  consultation with and written approval from FRITSCH  If  modifications are made without approval  the conformity de   clared by FRITSCH in respect of the European directives will  no longer apply     Exercise caution when working with combustible or poison   ous materials     Always comply with valid safety regulations  MAC val   ues  when handling samples or cleaning liquids which  are dangerous to your health  Where applicable  set up    the Mortar Grinder in a well ventilated safety zone        Laboratory Mortar Grinder    pulverisette 2   page 3    Safety Instructions    e Do not remove information signs    e Allow only trained personnel to perform maintenance   e Replace covers after maintenance    e Do not disable safety devices permanently    e Check safety device regularly     e    Out of se
12. r the commi   nution takes place by pressure and friction between the mortar bowl and the free ro   tating pestle     But opposite to the hand mortar the automatic mortar is rotating and the axis of the  pestle is in a stationary position  A relatively large grinding surface ensures quick and  uniform process and simple pestle adjustment  Sticky materials could be ground after  adding some liquid into the bowl     The sample is moved by the motor driven mortar bowl and taken off from the wall and  the bottom of the mortar bowl by the Vulcollan scraper  The sample is turned around  by the scraper working like a plough bringing it back in front of the pestle     Compared to the usually more plain hand mortars the grinding bowl of the Laboratory  Mortar Grinder is closed  A lip seal between grinding bowl and lid guarantees for loss   free comminution   even if suspension is to be ground     Reproducible results are obtained by the built in timer and the easy positioning of the  load of the pestle  A large acrylic safety cover permits observation of the grinding pro   cess which can be controlled by adjustment of scraper and pestle     The extremely robust construction of the driving unit  the guidance of the pestle and all  the other movable parts guarantees for a long lasting lifetime of the unit and prepared  it in everydays readiness     2 Operating safety    The operating safety of the Laboratory Mortar Grinder was proofed by the TUV   Saarland  After the successful final te
13. rvice     gt  Turn off master switch on rear side     e Allow only trained specialists to work on the mill     e Do not open Laboratory Mortar Grinder until it is com     pletely stopped   gt   e Always wear safety glasses when working     e Operate Mortar Grinder only with fixed pestle        e Do not put your hands into Mortar Grinder while it is running     e Set up the Laboratory Mortar Grinder indoors only   e Set it up on a sturdy work table    e Ambient temperature 0 to 40  C   e Dimensions  395 x 435 x 870 mm  H x W x D    e Weight  net   approx  30 kg    e Danger  Electricity    e The instrument contains electrical switches in the low voltage  range 230 Vrms    e Noise level  70 dB    e Beware of     A e Squeezing on counterweight   j e Squeezing on lid toggle catch      a e Squeezing on pestle fixing device and its spindle     Read instructions manual before use        Laboratory Mortar Grinder  pulverisette 2   page 4    Safety Instructions    Unpacking    e Attention  Weight of Laboratory Mortar Grinder   30 Kg  e Compare the content of the box with your order     1  Pull out the staples with the help of a pincers resp  cut the closing band  of the box     2  Open the cover of the box and  3  take out the accessories such as grinding set  pestle and Plexiglas lid     4 Don   nt use the rod as a hold  Take out the Mortar with both hands at the  lid     Mounting the Plexiglas Cover    1  Remove the paper  2  Fasten the cover with 4 screws   you will find the screws 
14. st the TUV Saarland certified that the Laboratory  Mortar Grinder by with its construction and manufacturing meets the requirements of  DIN EN 61010  03 94  resp  IEC 1 1990  A1 1992  This was confirmed by an official  Certification of Conformity     The enclosed    Declaration of Conformity    ensures  that the Laboratory Mortar Grinder  delivered with this paper was built according to the regulations     Laboratory Mortar   Grinder  pulverisette 2     Page 7    FRITSCH    3 Setting up the Laboratory Mortar Grinder    Set up the Laboratory Mortar Grinder indoors on a flat  sturdy surface   It is not necessary to fasten it in place     The room temperature must be between 0 and 40  C   Make certain that the grinder is easy to access     4 Electrical connections    4 1 Checking for concordance of voltage and current    Before connecting the grinder  check the setting of the voltage selector on the back of  the grinder and compare it to the voltage of your local mains       gt  Only a trained specialist is permitted to change over the    supply voltage and or alter the connecting cable  WA    The slot in the axle of the switch must point to the value of your  network     100 110 200 230 240 V A C    with protective earth  fuse max  16A  Turn on the master switch  black switch  on the rear side  The green lamp  Power On   will light up     4 2 Motor    The motor  single phase brushless A C  motor with PE conductor  50 60Hz  can be  started by turning the timer     In the case of
15. tion of mortar bowl and pestle    When selecting the material of the grinding set you must consider the hardness of the  sample and the tolerated contamination arising from the abrasion of the grinding set   Even a very small contamination may qualitatively influence your analytical results     Mortar bowl and pestle can be supplied with the following  materials     N  Material Order  NR  Abrasion   Mortar and pestle   Complete set  resistance    Mineral materials   Hard Porcelain     4601100   adequate      99 9  SiO2 46 005 00 good    94  20 46 012 00    Metallic materials    Chrome Nickel Steel 46 010 00          Chrome Steel 46 009 00    Tungsten Carbide 93  WC  6  Co 46 008 00       7 Installation of the Grinding Set    Caution  The parts of the grinding set installed in the mortar grinder have been ex   actly matched to each other  Therefore mortar bowl  pestle and scraper  must always be used jointly     The setting of mortar bowl  pestle and scraper has been optimised at works     Depending of the hardness of the sample to be ground  the pestle and the scraper  must be re set from time to time  Also when a different grinding set  or a new one  is  installed  the setting must be readjusted     Laboratory Mortar   Grinder  pulverisette 2     Page 10    FRITSCH    7 1 Mortar Bowl    To replace a mortar bowl  simply lift the bowl from its support  plate and put it back by placing it on to the support  the two trac   tion screws being introduced in the two holes on the bot
16. tom of  the plastic setting of the bowl     The bowl should not    dangle        7 2 Pestle    To remount or replace a pestle you need the enclosed tools   e Fork wrench 19 mm   e Fork wrench 24 mm   e Hexagon socket screw key  Allan key  3 mm and  e Face spanner       1  Hold the pestle axle with the provided fork wrench 19 mm  d  Then loosen the counter screw with the provided fork wrench 24 mm  The pestle  can than be unscrewed from the axle by turning to the left     2  Re mounting the pestle is being done the opposite way     When the lid is closed  the pestle must be well set in the lower radius of the mortar  bowl  To enable you to do this setting the axle of the pestle is arranged eccentric   Turning this eccentric the axis will be drawn near the edge of the bowl resp  the oppo   site way     The eccentric is fixed with a hexagon socket screw   1  Fill in the bowl some sample  f e  small amount of sand      The sand will take care of the friction between bowl and pestle in order to turn around  the pestle  Only by this you will be able to do the setting correctly     2  Close the lid and fasten the locking system   3  To frees the eccentric loosen the 3 mm hexagon socket with the Allan key     4  Insert the provided face spanner into the eccentric and reset the eccentric in a way  that the pestle slightly presses against the radius of the mortar bowl     Caution  The front of the bowl should not come up     If the setting is correctly done  then the pestle will rotate
    