Contents

1. Scientific f Instruments Model HSPC 1 INSTRUCTION MANUAL V 1 3 January 2004 ALA Scientific Instruments Inc 1100 Shames Drive Westbury NY 11590 1746 Tel 516 997 5780 Fax 516 997 0528 E mail support alascience com www alascience com Page 1 of 15 Table of Contents Page SAFETY INFORMATION AND GENERAL WARNINGS cccccccceeeeeeeeeeeeeeeeeeeeeeeeeeeees 3 ACKNOWLEDGEMENTS wixeucoi ce cee eee ee ee ae eG 4 SHIPPING CONTENTS i shes ae a aeaea aeara aren aaae a Danaa aiaa enaar 5 DESCRIPTION OF COMPONENTS FRONT PANEL cceccccceeeeeeeeeeeeeeeeeeeeeeeeeeeeeees 6 DESCRIPTION OF COMPONENTS REAR PANEL amp HEADSTAGE 00000 7 NTRODUC TIO N sacs r a r eats tect e aar Sara ia Ekas a Se coer EEEN Aas Ease 8 INSTALLATION OF THE HSPC 1 ccccssseeeeeeeeeeeeeeeseeeeeeeeeeeeeasessseeeeeeeeeeessenessneeeeees 9 Positioning of COMPONENIS dntcsenezcbiras ttle disedirdti ctl sterieeba sad ces sedldduateclednsuatdaeataatts 9 Connecting the tubing and electrical cords cccceeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeenes 9 INSTRUCTIONS FOR USE DURING PATCH RECORDING csssseeeeeeeeeeeeeeeeeeees 11 Initial zeroing of the InStrument nnnenuccnneesecenenenenenenenenenenemae 11 Formation of cell attached and whole cell configuration cceeeeeeeeeeeeeeeeees 11 Execution of pressure clamp protocols in the patch clamp configuration 12 Aspirati
2. controller settings after an idle is initiated see Instructions for use during patch recording below 7 Power on off switch 8 Display select toggle switch Selects display of setpoint or measured pressure 1 Page 6 of 15 Description of components Rear Panel amp Headstage oA ty oe E mea OS E 2 wast nase ue a p y Ke 1 C3 fF wee y deene ron 000 HEAD MOTOR O amp f STAGE COC sie C OGOO6O oTe 2 1 Controller rear panel 9 Power cord input This instrument is preset to 110 120 V AC input North American or 220 240 V AC European at the factory 10 DB9 cable input Input for custom cable that connects headstage and controller Substitution of standard cable with DB9 connector may damage instrument 11 Monitor output 20 mV mmHg analog output from pressure sensor Use to display or record measured pressure Standard BNC 12 Command input 20 mV mmHg analog input Use to control pressure from function generator or acquisition software Standard BNC Headstage connections F1 Output connector blue for silicone tubing going to electrode holder R1 DB9 connector for cable from headstage to controller R2 Input source pressure connector blue Luer connector R3 Input source vacuum connector yellow Luer connector Page 7 of 15 Introduction The development of the voltage clamp enabled stepping of excitable membrane preparations to pro
3. input air pressure is from a cylinder of purified air or from a house air supply filtered with at least a 5 u filter 3 When not using the valve i e when pipettes are being changed toggle the reset idle switch 6 to the idle position This causes a small volume of air to flow out of the valve and prevents dust or moisture from entering To resume recording toggle the switch to the reset position Page 12 of 15 Maintenance and Troubleshooting With normal usage no routine maintenance procedures are required other than immediately wiping up liquids spilled on the instrument and cleaning or replacing the contaminated components following moisture triggered alarms The HSPC 1 has no user serviceable parts If you have any problems with the HSPC 1 please contact ALA Scientific Instruments if bought within North America or your local distributor Contact ALA Scientific Instruments by phone or email at the number or address below Email support alascience com Phone 516 997 5780 Fax 516 997 0528 Page 13 of 15 References Bertil Hille 1992 Classical biophysics of the squid giant axon In lonic Channels of Excitable Membranes Sunderland MA Sinauer Associates Inc pp 23 58 Besch S R Suchyna T and Sachs F 2002 High speed pressure clamp Pflugers Arch 445 161 166 Hamill O P and Martinac B 2001 Molecular basis of mechanotransduction in living cells Physiol Rev 81 685 740 McBride D W Jr and
4. Hamill O P 1992 Pressure clamp a method for rapid step perturbation of mechanosensitive channels Pflugers Arch 421 606 612 Peter Jonas 1995 Fast Application of Agonists to Isolated Membrane Patches In Single Channel Recording Bert Sakmann and Erwin Neher eds New York Plenum Press pp 231 243 Reinhold Penner 1995 Introduction to patch clamping In Single Channel Recording Bert Sakmann and Erwin Neher eds New York Plenum Press pp 3 30 Page 14 of 15 Warranty ALA Scientific Instruments agrees to warranty this product for a period of one year from the date of delivery against any and all manufacturer s defects in material and or workmanship other than as noted below Remedy will consist of repair or replacement at ALA s discretion Please report any problems promptly so as not to jeopardize the warranty coverage ALA Scientific Instruments does not assume any liability based on the use of this product whether correct or incorrect except as specified under law Warranty rights may vary from state to state ALA Scientific Instruments will not warranty any of the plastic connectors or tubing Use of the HSPC 1 in any way that contradicts the instructions of this manual may void the warranty If the product does need repair it must be returned to the factory freight prepaid freight sent collect will be refused and in clean condition If returned parts have been in contact with any liquid substance documentat
5. e longevity These innovations include supporting the piezo bimorph at both ends minimizing valve dead volume placing the pressure sensor on the valve body and incorporating a piezo element that does not require a static bias voltage A final innovation the moisture sensitive optical sensor protects the piezoelectric valve from contact with moisture by triggering an automated application of positive outward pressure when moisture is sensed Besch et al 2002 Page 8 of 15 Installation of the HSPC 1 ALA strongly recommends the use of a Faraday cage to minimize radiative pickup during patch clamp recording and the use of an anti vibration table to minimize damage to seals during recording The headstage of the HSPC 1 should be positioned close enough to the electrode holder of the patch pipette to permit connection to the pipette holder with the 10 cm length of tubing provided Use of exceptionally long gt 10 cm or distensible tubing not supplied by ALA Scientific Instruments will delay and distort the pressure waveform between the headstage and pipette holder It is not necessary or desirable to bolt the HSPC 1 headstage to the microscope or table prior to use All other electrical components such as the HSPC 1 controller and electrically driven pressure or vacuum sources should be positioned outside of the Faraday cage Connecting the tubing and electrical cords Connect the output of the HSPC 1 headstage F1 to the electrode h
6. grammed voltages under closed loop control This enormously advanced understanding of the function of voltage gated channels Bertil Hille 1992 Similarly the concentration clamp technique enabled rapid application of agonists to ligand sensitive excised patch preparations and advanced understanding of the function of ligand gated channels Peter Jonas 1995 Mechanosensitive ion channels are also important physiologically Hamill and Martinac 2001 but the lack of instruments capable of establishing a pressure clamp to enable stepping of mechanosensitive preparations to programmed pressures under closed loop control has impeded progress in understanding their function ALA Scientific Instruments is proud to present the HSPC 1 an instrument that enables any investigator familiar with the patch clamp technique to perform sophisticated pressure clamp experiments on mechanosensitive preparations in the excised patch or whole cell patch clamp configuration Drs Frederick Sachs Thomas Suchyna and Stephen Besch built the earliest version of the HSPC 1 at the University of Buffalo Besch et al 2002 The HSPC 1 is similar in some respects to an earlier device by McBride and Hamill that employed piezoelectric valves to partition pressure and vacuum sources under closed loop control McBride Jr and Hamill 1992 The current HSPC 1 features a miniaturized piezoelectric with several innovations that improve response time stability and valv
7. ion of substances used must be provided to ALA This product is intended for use in biophysical experiments with membrane or cellular preparations only THIS EQUIPMENT IS NEITHER INTENDED NOR APPROVED FOR CLINICAL USE Page 15 of 15
8. le can be connected from the monitor output 11 to an oscilloscope or to an A D converter to observe or record the pressure measured at the headstage sensor Finally the power cable should be plugged from the back of the instrument 9 into a high quality surge protector At this point all electrical and pressure vacuum connections are complete and the system can be powered on for initial zeroing and use Page 10 of 15 Instructions for use during patch recording This manual presumes a basic understanding of patch clamp technique by the reader Many excellent reviews are available e g Reinhold Penner 1995 Initial zeroing of the instrument After the instrument has been set up as described above attach an appropriately filled electrode to the electrode holder The HSPC 1 may be turned on at this point and the instrument zeroed according to the following procedure 1 Toggle the display select switch 8 to the setpoint position Then adjust the setpoint knob until the display reads 00 0 2 Toggle the display select switch 8 to the sensor position Then adjust the zero knob until the display 1 reads 00 0 3 Toggle the display select switch back to the setpoint position The display 1 should read 00 0 If it does not turn the setpoint knob 2 to make the slight adjustment required 4 At this point the instrument is zeroed properly and ready to receive an input command voltage to control press
9. m Resonant piston pumps are recommended They are quiet and reliable The 60 Hz component of the pressure output associated with these pumps can be eliminated from the input by using filtering canisters Appropriate canister size must be determined empirically A reasonable starting size is 250 ml Page 9 of 15 Installation of the HSPC 1 Connect the headstage to the controller using the custom cable with DB9 connectors WARNING do not substitute standard cables One end should attach to the headstage socket R1 and the other to the appropriate socket in the back of the controller 10 A BNC cable is used to connect the command input 12 to an appropriate analog voltage source The analog voltage waveform can be derived from any function generator or from commercial electrophysiology experimental control data acquisition programs e g npi electronic s CellWorks Axon Instruments pClamp or HEKA Electronik Pulse PulseFit or PatchMaster that generate an output command voltage However if the desired experiment requires command waveforms to be generated simultaneously to the voltage clamp and pressure clamp the user should carefully evaluate the specifications of any experimental control data acquisition software to make certain that the selected software has the required capabilities Alternatively a custom signal generator that adheres to the required input limitations maximum input voltage 12 V may also be used A BNC cab
10. older of the patch clamp with the thick walled tubing provided Please note that with the HSPC 1 it is possible to input pressures directly from a function generator or acquisition software for gigaseal formation and membrane rupturing i e formation of the whole cell configuration Use of manual methods for application of suction e g by mouth or syringe is unnecessary and inadvisable The headstage contains ports for input of pressure and vacuum Connect the vacuum port R3 and pressure port R2 to suitable vacuum and pressure sources with the supplied tubing The pressure source must be filtered to 5 u or better to remove particulates and oil ALA Scientific Instruments manufactures a device the Pressure Vacuum system for HSPC 1 catalog number ALA PV PUMP that is designed to provide appropriate pressure and vacuum to the respective inlet ports Alternatively other commercial sources for these components are available Should you choose to configure a system from components from other vendors please bear in mind the following 1 For optimal valve performance input pressure or vacuum should not exceed 500 mmHg Higher pressures than this may not damage the valve but may prevent complete closing of the valve and thus hinder valve performance The recommended input pressure is 250 mmHg At this pressure consumption of air is approximately 3 L min N Several types of pump may be suitable for generating pressure and vacuu
11. on of moisture into the headstage eee eeeeeeeeeeeeeeeteeeeeeeeeeeeeeeeeeeeeeeeee 12 Prolonging the longevity of the piezoelectric valve 12 MAINTENANCE AND TROUBLESHOOTING cccsseeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeees 13 REP ERENG BS 5 icccncocce cecsecct cect c E EE 14 a n E Eos k sada sec ade ac cc nda vaca TT 15 Page 2 of 15 eke aA Pe ledge BEFORE BEGINNING ANY WORK WITH THE HSPC 1 PLEASE READ THE ENTIRE MANUAL FAILURE TO FOLLOW THE WARNINGS BELOW MAY RESULT IN DAMAGE TO THE INSTRUMENT OR INJURY TO PERSONNEL Safety Information and General Warnings 1 The HSPC 1 requires a connection to mains power Incorrect usage or carelessness may result in electrocution Do not operate if the HSPC 1 is immersed in liquids or if the electrical cables are damaged 2 The DB9 cable that connects the headstage to the controller is custom wired Do NOT substitute a standard cable with DB9 connectors 8 Only trained scientific personnel who are familiar with patch clamp amplifiers should use this instrument er If liquids spill on any part of the instrument shut the instrument off and wipe up the liquids before turning it on again 2 The control box and headstage are not user serviceable Unauthorized opening or modification voids the warranty 6 Do not position the HSPC 1 where someone might trip over the cords or tubing 7 x The HSPC 1 has an alarm system to prevent accidental suctioning of liq
12. patch clamp configuration After achieving the desired patch clamp configuration as described above both pressure and voltage are controlled by independent negative feedback loops You can execute complex patch clamp protocols exactly as you would voltage clamp protocols within the constraints described in the General Warnings section above Aspiration of moisture into the headstage When using this instrument aspiration with a broken pipette or other conditions may cause moisture to enter the headstage When this happens the emergency moisture sensor circuitry senses the moisture and prevents damage to the piezoelectric valve The input command voltage is disabled the valve is immediately driven to apply positive pressure to the pipette and the alarm light 4 illuminates After a moisture alarm stop the experiment determine the source of moisture entry and rectify the problem Carefully dry the pipette holder and tubing connecting the electrode holder to the HSPC 1 headstage sometimes it may be simpler and faster to replace this tubing Toggling the reset idle switch to the reset position clears the alarm state and restores normal operation Prolonging the longevity of the piezoelectric valve The piezoelectric valve in the headstage is designed to last for years under normal working conditions Several simple steps will maximize its longevity 1 Do not defeat the moisture alarm triggering circuitry 2 Make sure that
13. t and pressure input tubing h Thick walled silicone tubing for connection of headstage output to electrode holder 1 Please open the box and the inspect contents immediately upon receipt If any components appear to be damaged or missing please contact ALA Scientific Instruments or your local distributor Page 5 of 15 Description of components Front Panel 4 Pressure mmHg Setpoint Zero Setpoint Display Select Sensor Pressure Clamp Controller front panel 1 Pressure display Digital display of either the instrument s pressure setpoint or the pressure measured by the headstage sensor Maximal display is 200 mmHg 2 Setpoint knob Manual control for setting pressure output from the controller and for zeroing of the instrument prior to recording 3 Zero knob Manual control for zero adjustment prior to recording 4 Alarm light Light will illuminate when moisture sensor detects entry of moisture or when the unit is in the idle state 5 6 Power light Illuminates when instrument is powered Reset switch Three way momentary toggle switch Middle position is default position for normal instrument functioning Reset upper position restores normal function of moisture alarm and piezoelectric valve after emergency shutoff is triggered by entry of moisture into headstage see Instructions for use during patch recording below Reset position will also restore preconfigured
14. uids into the instrument Circumventing this system voids the warranty 8 The piezoelectric valve inside the headstage is highly sensitive Follow the suggestions in the Prolonging the longevity of the piezoelectric valve p 12 section to insure long life 9 Do not apply voltages greater than 12 V to the command input 10 The HSPC 1 should be plugged into a high quality surge protector to prevent damage to sensitive components during electrical storms 11 The HSPC 1 is not approved for clinical applications ALA Scientific Instruments discourages any such use and accepts no liability for such use Page 3 of 15 Acknowledgements The HSPC 1 was developed by Drs Stephen Besch Thomas Suchyna and Frederick Sachs at SUNY Buffalo John Montalbano and Mohammad Ahmed at ALA Scientific Instruments contributed to circuit design in the manufactured product ALA Scientific Instruments is also grateful for the contributions of Drs Stephen Besch and Thomas Suchyna in editing and writing the manual This manual was written by Dr lan M Herzberg with assistance from Stuart Havel for graphics and formatting Page 4 of 15 Shipping contents The HSPC 1 shipment should include the following items a Instruction manual 1 b Power cable 1 c Instrument controller 1 Headstage 1 e Plastic bag containing various connectors 1 f Custom cable with 2 male DB9 connectors to connect headstage to controller 1 g Vacuum inpu
15. ure applied to the patch electrode Output pressure will follow input command voltages with a proportionality constant of 20 0 mmHg mV Formation of cell attached and whole cell configuration A positive outward pressure should be applied to the patch pipette when lowering the pipette through the bath medium This prevents occlusion of the pipette tip by debris If you are using a function generator to generate the command voltage the setpoint adjustment 2 can be used for this purpose If you are using a commercial software program such as pClamp Axon Instruments CellWorks npi electronic Pulse PulseFit or PatchMaster HEKA Electronic it is generally more convenient to set this positive pressure using the control panel of the respective program After the tip of the electrode is close to the cell determined by visual inspection under the microscope and or by an increase in measured resistance during application of a test pulse negative pressure is applied using the HSPC 1 as described above to achieve the cell attached configuration Negative pressure can again be applied to form the whole cell configuration Controlled application of pressure or vacuum by the HSPC 1 achieves the desired patch clamp configuration more efficiently and reproducibly than does application of pressure or vacuum by conventional methods Page 11 of 15 Instructions for use during patch clamp recording Execution of pressure clamp protocols in the