Omron Healthcare HEM-711AC User's Manual

image

Contents

1. Modelo HEM 711AC Pantalla Pantalla LCD digital Espectro de medici n Presi n 0 a 299 mmHg Pulso 40 180 min Memoria 60 mediciones con fecha y hora Precisi n calibrado Presi n 3 mmHg o 2 de la medici n Pulso 5 de la medici n Inflado Autom tico mediante bomba el ctrica Desinflado V lvula de liberaci n autom tica de presi n Liberaci n r pida de aire V lvula de escape autom tico Detecci n de presi n Sensor capacitivo de presi n M todo de medici n M todo oscilom trico Detecci n de la onda de pulso Sensor capacitivo de presi n Fuente de alimentaci n Adaptador de CA o 4 pilas AA Duraci n de las pilas 300 usos aproximadamente cuando se usa una vez al d a durante 2 minutos con 4 pilas nuevas de manganeso est ndar Temperatura de 10 C a 40 C 50 F a 104 F 30 a 85 humedad operaci n Humedad relativa Temperatura de 4 F a 140 F 20 C a 60 C 10 a 95 de almacenamiento Humedad humedad relativa Peso de la unidad principal Aproximadamente 12 1 2 oz 355 g Dimensiones de la Aprox 6 3 32 pulg largo x 5 5 32 pulg unidad principal ancho x 3 5 16 pulg alto 155 mm x 131 mm x 84 mm Tama o del brazalete Aproximadamente 5 fi pulg ancho x 19 pulg largo 140 mm x 483 mm Tubo del brazalete 23 5 8 pulg largo 600 mm Circunferencia del brazalete Se ajusta a circunferencias de brazos de 9 pulg a 13 pulg 229 mm
2. CUIDADO Y MANTENIMIENTO Para mantener su monitor digital de presi n arterial en buen estado y proteger la unidad contra da o siga las instrucciones siguientes Guarde el monitor en su estuche de almacenamiento cuando no lo use Aseg rese de que el adaptador de CA est colocado debajo de la unidad principal para no da ar la pantalla No doble a la fuerza el brazalete ni el tubo de aire No los doble ajustadamente Limpie el monitor con un pa o suave seco No use limpiadores abrasivos ni vol tiles No trate de limpiar el brazalete Nunca sumerja el monitor ni ninguno de los componentes en agua Guarde el monitor en un lugar seguro y seco No exponga el monitor a temperaturas extremas ya sean calientes o fr as a la humedad ni a la luz solar directa Evite exponer el monitor a impactos fuertes como dejar caer la unidad al suelo Quite las pilas si la unidad no se va a usar por tres o m s meses Siempre reemplace todas las pilas simult neamente con pilas nuevas Use la unidad de acuerdo con las instrucciones descritas en este manual Use s lo piezas y accesorios autorizados S19 INDICADORES DE ERROR S MBOLO CAUSA CORRECCI N Qu tese el brazalete Consulte la secci n Colocaci n del brazalete Espere 2 a 3 minutos Realice otra medici n El monitor no detect la frecuencia de pulsos Brazalete inflado Qu tese el brazalete insuficientemente Lea la secci n C mo realizar una medici n
3. 5 Envuelva firmemente el brazalete alrededor de su brazo usando la tira de cierre 13 COLOCACI N DEL BRAZALETE COLOCACI N DEL BRAZALETE EN EL BRAZO DERECHO Cuando realice una medici n usando el brazo derecho emplee la instrucci n del Paso 5 Coloque el brazalete en la parte superior de su brazo derecho de manera que la flecha azul y el tubo de aire queden al interior de su codo La parte inferior del brazalete debe estar aproximadamente 1 2 pulgada arriba de su codo NOTA Tenga cuidado de no apoyar su brazo sobre el tubo de aire Esto restringir el flujo de aire al brazalete S14 C MO REALIZAR UNA MEDICI N El monitor determina autom ticamente el nivel de inflado ideal En raras circunstancias cuando puede ser necesario un mayor nivel de inflado el monitor autom ticamente vuelve a inflar el brazalete a un valor 30 mmHg m s alto que el nivel de inflado inicial y reinicia la medici n 1 Presione el bot n START STOP inicio paro Aparecen todos los s mbolos en la pantalla El brazalete comienza a inflarse autom ticamente A medida que se infla el brazalete el monitor autom ticamente determina su nivel de inflado ideal Puesto que este monitor detecta el pulso incluso durante el inflado no mueva su brazo y permanezca quieto hasta que todo el proceso de medici n haya concluido 388 dBBe Dn pee 588 eo 88 8 08 8 00 29 NOTA Para detener el inflado o la medici n p
4. en la p gina 16 Brazalete inflado Espere 2 a 3 minutos excesivamente Realice otra medici n Reemplace las cuatro pilas Consulte la secci n Instalaci n de las pilas Las pilas est n gastadas Qu tese el brazalete Espere 2 a 3 minutos Realice otra medici n Permanezca quieto durante la medici n Movimiento durante la medici n 20 SUGERENCIAS PARA LA RESOLUCI N DE PROBLEMAS No hay alimentaci n Reemplace las pilas No aparece ninguna pantalla Verifique la instalaci n de las pilas en la unidad para asegurar la correcta ubicaci n de las polaridades de las pilas La presi n arterial var a constantemente Muchos factores entre ellos el estr s la hora del d a c mo se coloca el brazalete pueden afectar la presi n arterial Repase las secciones Antes de medir la presi n arterial y C mo realizara una medici n Los valores de medici n aparecen demasiado altos o demasiado bajos S21 GU A DE REFERENCIA R PIDA Para asegurar una lectura confiable evite comer fumar y hacer ejercicios los 30 minutos anteriores a la medici n Debe tratar de medir la presi n arterial a la misma hora cada d a Evite realizar mediciones durante momentos de estr s 1 Qu tese la ropa apretada de la parte superior del brazo 2 Si ntese en una silla con los pies sobre el piso y coloque su brazo sobre la mesa de manera que el brazalete quede al mismo nivel que su coraz n
5. las limitaciones establecidas para dispositivos digitales Clase B de conformidad con la secci n 15 de las Normas de la FCC El prop sito de estas limitaciones es brindar protecci n razonable contra interferencias perjudiciales en el hogar Este producto genera usa y puede irradiar energ a de frecuencia radioel ctrica y si no se lo instala y emplea seg n las instrucciones correspondientes puede provocar interferencias en las comunicaciones por radio No obstante no existe garant a alguna de que no vayan a producirse interferencias en alguna instalaci n en particular Si el producto produjera alg n tipo de interferencia perjudicial para la recepci n de radio o televisi n lo cual puede determinarse apagando y volviendo a encender el producto se recomienda que el usuario realice alguna de las siguientes operaciones para intentar corregir la interferencia Cambie la orientaci n o ubicaci n de la antena de recepci n Aumente la distancia que separa el producto del receptor Conecte el producto a un tomacorriente que integre un circuito diferente de aqu l donde est conectado el receptor Solicite asistencia al vendedor o a un t cnico de radio TV experimentado POSIBILIDAD DE INTERFERENCIA EN RADIO TELEVISI N para Canad nicamente Este aparato digital no excede las limitaciones de Clase B para emisiones de ruido de radio provenientes de aparatos digitales de conformidad con la norma para equipos que causan i
6. tiempo de espera seg n sus caracter sticas fisiol gicas individuales S4 CONOZCA SU UNIDAD Unidad principal Pantalla Compartimiento Ny de las pilas Bot n START STOP l inicio paro Bot n MEMORY memoria Conector Bot n de ajuste de fecha hora de aire Conector del adaptador de CA Componentes Brazalete Tubo de aire Enchufe de aire se conecta al conector de aire Estuche de almacenamiento Adaptador de CA S5 PANTALLA DE LA UNIDAD Presi n arterial sist lica Presi n arterial dist lica i E S mbolo de latidos del coraz n Indicador del pulso S mbolo de desinflado S mbolo de la memoria S mbolo de valor promedio S mbolo de error por movimiento Indicador de fecha hora S mbolo de pilas bajas S mbolo de latidos irregulares del coraz n S MBOLO DE LATIDOS IRREGULARES DEL CORAZ N Cuando un monitor detecta un ritmo irregular dos o m s veces durante la medici n el s mbolo de latidos irregulares del coraz n CQ aparecera en la pantalla con los valores de medicion Se denomina ritmo irregular de los latidos UE del corazon a un ritmo que varia por mas Pulso A A A A A A 5 del 25 con respecto al ritmo promedio Presi n NN NN JN detectado mientras el monitor mide arterial la presi n arterial sist lica y diast lica Latido irregular ael coraz n Corto Largo Si el s mbolo de latidos irregulares del did coraz n KO
7. aparece con sus resultados de Presi n PIA IAN arterial medici n recomendamos que se comunique con su m dico Siga las instrucciones de su m dico S6 INSTALACI N DE LAS PILAS 1 Presione el indicador W situado en la cubierta de las pilas y deslice la cubierta en direcci n de la flecha para extraerla 2 Instale 4 pilas AA de manera que las polaridades positiva y negativa coincidan con las polaridades indicadas en el compartimiento de las pilas 3 Vuelva a colocar la cubierta de las pilas REEMPLAZO DE LAS PILAS tX Indicador de pilas bajas Cuando aparezca en la pantalla el indicador de pilas baja quite todas las pilas Reempl celas con cuatro pilas nuevas Se recomienda el uso de pilas alcalinas de larga duraci n NOTA Los valores de medici n almacenados en la memoria no se eliminar n durante el reemplazo de las pilas Despu s de reemplazar las pilas reajuste la fecha y la hora Consulte la p gina 10 Ajuste de la fecha y hora S7 USO DEL ADAPTADOR DE CA Use s lo el adaptador de CA Omron autorizado n mero de modelo HEM ADPT1 con este monitor El uso de cualquier otro adaptador puede causar da o al monitor y anular la garant a del usuario Para conectar el adaptador de CA O Inserte el enchufe del adaptador de CA en el conector del adaptador de CA situado en la parte trasera del monitor tal como se ilustra 2 Enchufe el adaptador de CA en una toma de corriente de 120 VC
8. informaci n sobre servicio cubierto por la garant a est disponible en nuestro sitio web www omronhealthcare com Incluya el comprobante de compra Incluya una carta con su nombre direcci n n mero de tel fono y descripci n del problema espec fico Embale cuidadosamente el producto para evitar que se da e durante el transporte Debido a la posibilidad de p rdida durante el transporte recomendamos asegurar el producto y pedir acuse de recibo TODAS LAS GARANT AS IMPL CITAS ENTRE LAS QUE FIGURAN LAS GARANT AS IMPL CITAS DE COMERCIABILIDAD E IDONEIDAD PARA PROP SITOS PARTICULARES QUEDAN LIMITADAS A LA DURACI N DE LA GARANT A ESCRITA CORRESPONDIENTE DESCRITA Algunas provincias no permiten limitaciones respecto al plazo de las garant as impl citas por lo tanto la limitaci n anterior puede no aplicarse en su caso OMRON NO SER RESPONSABLE POR DANOS OCASIONADOS DEBIDO A LA P RDIDA DE USO NI POR NING N OTRO DA O GASTO O COSTO INCIDENTAL INDIRECTO O RESULTANTE Algunos estados no permiten la exclusi n o restricci n de da os resultantes o incidentales por lo tanto las exclusiones anteriores pueden no aplicarse en su caso Esta garant a le otorga derechos legales espec ficos y adem s usted puede tener otros derechos que pueden variar seg n el estado de residencia PARA SERVICIO AL CLIENTE Visite nuestro sitio web www omronhealthcare com Llame sin cargo 1 800 634 4350 25 Especificaciones
9. middle finger The air tube runs down the inside of your arm The bottom of the cuff should be approximately 1 2 above your elbow 1 2 inch 1 2 cm 5 Wrap the cuff firmly in place around your arm using the closure strip 13 APPLYING THE ARM CUFF APPLYING THE CUFF ON THE RIGHT ARM When taking a measurement using the right arm use this instruction for Step 5 Apply the cuff to your right upper arm so the blue arrow and the air tube runs along the inside of your elbow The bottom of the cuff should be approximately 1 2 above your elbow NOTE Be careful not to rest your arm on the air tube This will restrict the flow of air to the cuff 14 TAKING A MEASUREMENT The monitor automatically determines your ideal inflation level In rare circumstances when a higher inflation may be necessary the monitor automatically re inflates the cuff up to 30 mmHg higher than the initial inflation and restarts the measurement 1 Press the START STOP button All display symbols appear on the screen The cuff starts to inflate automatically As the cuff begins to inflate the monitor automatically determines your ideal inflation level Because this monitor detects the pulse even during inflation do not move your arm and remain still until the entire measurement process has been completed 388 8B Ch 18 1 908 cu 38788 8 00 8 00 NOTE To stop the inflation or measurement push the STA
10. 3 Coloque el brazalete en la parte superior de su brazo izquierdo de manera que la flecha azul est hacia el interior de su brazo y alineada con su dedo medio El tubo de aire pasa por el lado interior de su brazo La parte inferior del brazalete debe estar aproximadamente 1 2 pulgada arriba de su codo 4 Envuelva firmemente el brazalete alrededor de su brazo usando la tira de cierre 1 2 pulgada 1 2 cm Marcador Tubo de aire 22 GU A DE REFERENCIA R PIDA 5 Presione el bot n START STOP inicio paro Aparecen s mbolos indicadores El brazalete comienza a inflarse autom ticamente Permanezca quieto y no hable durante la medici n NOTA Para detener el inflado o la medici n presione el bot n START STOP El monitor detendr el inflado iniciar el desinflado y se apagar 6 Cuando termina la medici n el brazalete del brazo se desinfla completamente Aparece en la pantalla su presi n arterial y frecuencia de pulsos NOTA Espere 2 a 3 minutos entre mediciones El tiempo de espera permite que las arterias vuelvan al estado en que estaban antes de la medici n de tensi n arterial Es posible que necesite aumentar el tiempo de espera seg n sus caracter sticas fisiol gicas individuales 823 DECLARACI N DE LA FCC Nota POSIBILIDAD DE INTERFERENCIA EN RADIO TELEVISI N para los EE UU nicamente Las pruebas a las que se someti este producto comprobaron que cumple con
11. A 60 ciclos Para desconectar el adaptador de CA primero extraiga el adaptador de CA de la toma de corriente y luego quite del monitor el enchufe del adaptador de CA NOTA El monitor est dise ado para no consumir alimentaci n de las pilas cuando se utiliza el adaptador de CA S8 AJUSTE DE LA FECHA Y LA HORA El monitor de la presi n arterial autom ticamente almacena hasta 60 valores de medici n con la fecha y hora El monitor tambi n calcula una lectura promedia basada en los ltimos tres valores de medici n Ajuste el monitor con la fecha y hora actual antes de realizar la primera medici n Si no se ajusta la fecha y la hora aparecer en la pantalla una fecha y hora predeterminadas 1 PARA COMENZAR EL AJUSTE Presione el bot n de Ajuste de fecha hora para ajustar la fecha y la hora El a o parpadea en la pantalla 2 AJUSTE DEL A O El a o puede fijarse entre 2004 y 2030 Cuando la pantalla llegue al a o 2030 regresar al a o 2004 Presione el bot n de MEMORIA para hacer avanzar los d gitos uno a la vez Presione el bot n de Ajuste de fecha hora para fijar el a o actual El mes parpadea en la pantalla S9 AJUSTE DE LA FECHA Y LA HORA 3 AJUSTE DEL MES Presione el bot n de MEMORIA para avanzar en incrementos de un mes Presione el bot n de Ajuste de fecha hora para fijar el mes actual el d DAMEN LAN E donus A El d a parpadea en la pantall
12. Notas de seguridad importantes 000 0 002000 3 Antes de medir la presi n arterial 0 000 d Instrucciones de operaci n Conozca su Unidad ex mind a oes ERES 5 Pantalla de la unidad sls esses 6 S mbolo de latidos irregulares del coraz n 6 Instalaci n de las pilas 0 eee oo 7 Uso del adaptador de CA oooooooooooooommommmoooo 8 Ajuste de la fecha y la hora 2 00 02 eee eee eee 9 Colocaci n del brazalete 0 0 0c eee eee 12 C mo medir la presi n arterial 2 2 2 0 0 0 0 e eee eee 15 Uso de la funci n de memoria 0 000000 eee 17 Cuidado y mantenimiento Cuidado y mantenimiento 0 0 0 c eee eee eee 19 Indicadores de erfor spas eee rbd dnt ne 20 Sugerencias para la resoluci n de problemas 21 Gu a de referencia r pida 0 0 cee eee eee 22 Declaraci n de la FCC seners minnein nE E E e 24 Informaci n sobre la garantia 0 0 00 eee eee eee 25 Especificaciones ssa ssa sca bes ee iss RR ER dea 26 S1 INTRODUCCI N Gracias por comprar el monitor autom tico de la presi n arterial HEM 711AC IntelliSense de Omron Llenar para futura referencia FECHA DE COMPRA NUMERO DE SERIE Engrape su recibo de compra aqu Su nuevo monitor digital de la presi n arterial usa el m todo oscilom trico de medici n de la presi n arterial Esto sign
13. OMRON Q0 usi Automatic Blood Pressure Monitor with Arm Cuff Model HEM 711AC ESPA OL ENGLISH CONTENTS Before Using the Monitor tods raio la also cda 2 Important Safety Notes 2 eee eee cee ee eee eee 3 Before Taking a Measurement 00 00 esses 4 Operating Instructions Know Your Unit assa S rap asin peed ade SIN qe ERES 5 Unit Display ias un ese 6 Irregular Heartbeat Symbol 0 00 00 eee eee 6 Battery Installation 0 0 cece eee eee 7 Using the AC Adaptor 00 0 cece eee eee 8 Setting the Date and Time 00 000 eot n eae 9 Applying the Arm Cuff san isses ses AR serene ade 12 Taking a Measurement 0 00 0 esee 15 Using the Memory Function ssslsse esses 17 Care and Maintenance Care and Maintenance 0 00 ce ee 19 Error Indicators due dae latido ia re wee oe Eee 20 Troubleshooting Tips 0 0 ce eee eee ee eee 21 Quick Reference Guide 0 cece cece eee 22 FCC Statement cnn cde dee mne aaa lees 24 Warranty Information sses essnee ui eee eee eee e 25 Specifications icic ae e ss eraot den sae ess qug aka Ea s 26 INTRODUCTION Thank you for purchasing the Omron HEM 711AC IntelliSense Automatic Blood Pressure Monitor Fill in for future reference DATE PURCHASED SERIAL NUMBER Staple your purchase receipt here Your new digital blood pressure monitor uses the oscillome
14. RT STOP button The monitor will stop inflating start deflating and will turn off 2 Inflation stops automatically and the measurement is started As the cuff deflates decreasing numbers appear on the display The Heart Symbol ep flashes iac 46 125 at every heartbeat En 00 Ln n E 00 TS TAKING A MEASUREMENT 3 When the measurement is complete the arm cuff completely deflates Your blood pressure and pulse rate are displayed 4 Press the START STOP button to turn the monitor off NOTE The monitor will automatically turn off after five minutes INSTRUCTIONS FOR SPECIAL CONDITIONS If your systolic pressure is known to be more than 220 mmHg press and hold the START STOP button until the monitor inflates 30 to 40 mmHg higher than your suspected systolic pressure NOTES Do not apply more pressure than necessary The monitor will not inflate above 299 mmHg 16 USING THE MEMORY FUNCTION Every time you complete a measurement the monitor stores the blood pressure and pulse rate in the memory The blood pressure monitor automatically stores up to 60 measurement values When 60 sets of measurement values are stored in the memory the oldest record is deleted to save the most recent measurement values The monitor also calculates an average reading based on the values of the three most recent measurements taken If the date and time is not set the measurement values will not be stored in t
15. Wait 2 3 minutes between measurements The wait time allows the arteries to return to the condition prior to taking the blood pressure measurement You may need to increase the wait time depending on your individual physiological characteristics 23 FCC STATEMENT Note POTENTIAL FOR RADIO TELEVISION INTERFERENCE for U S A only This product has been tested and found to comply with the limits for a Class B digital device pursuant to part 15 of the FCC rules These limits are designed to provide reasonable protection against harmful interference in a residential installation The product generates uses and can radiate radio frequency energy and if not installed and used in accordance with the instructions may cause harmful interference to radio communications However there is no guarantee that interference will not occur in a particular installation If the product does cause harmful interference to radio or television reception which can be determined by turning the product on and off the user is encouraged to try to correct the interference by one or more of the following measures Reorient or relocate the receiving antenna Increase the separation between the product and the receiver Connect the product into an outlet on a circuit different from that to which the receiver is connected Consult the dealer or an experienced radio TV technician for help POTENTIAL FOR RADIO TELEVISION INTERFERENCE for Canada onl
16. a 4 AJUSTE DEL D A Presione el bot n de MEMORIA para avanzar en incrementos de un d a Presione el bot n de Ajuste de fecha hora para fijar el d a actual La hora parpadea en la pantalla 5 AJUSTE DE LA HORA La hora se ajusta usando AM o PM Presione el bot n de MEMORIA para avanzar en incrementos de una hora Presione el bot n de Ajuste de fecha hora para fijar la hora actual Los minutos parpadean en la pantalla S10 AJUSTE DE LA FECHA Y LA HORA 6 AJUSTE DE LOS MINUTOS Presione el bot n de MEMORIA para avanzar en incrementos de un minuto Presione el bot n de Ajuste de fecha hora para fijar los minutos El afio parpadea en la pantalla 7 Presione el bot n START STOP inicio paro para apagar la pantalla S11 COLOCACI N DEL BRAZALETE 1 Aseg rese de que el enchufe de aire est insertado firmemente en la unidad principal 3 Si ntese en una silla con los pies en el suelo Coloque su brazo sobre una mesa de manera que el brazalete quede a nivel de su coraz n 12 COLOCACI N DEL BRAZALETE 4 Coloque el brazalete en la parte superior de su brazo izquierdo de manera que la flecha azul est centrada en el interior de su brazo y alineada con su dedo medio El tubo de aire pasa por el lado interior de su brazo La parte inferior del brazalete debe estar aproximadamente 1 2 pulgada arriba de su codo 1 2 pulgada 1 2 cm Marcador Tubo de aire
17. a 330 mm Incluye Unidad principal brazalete adaptador de CA Manual de instrucciones y estuche de almacenamiento C digo UPC 0 73796 07115 8 NOTA Estas especificaciones est n sujetas a cambio sin previo aviso 26 Fabricado para OMRON HEALTHCARE INC 1200 Lakeside Drive Bannockburn Illinois 60015 www omronhealthcare com Hecho en China O 2005 Omron Healthcare Inc 711ACNINST Rev
18. ect your blood pressure Review the sections Before Taking a Measurement and Taking a Measurement Measurement values appear too high or too low 21 QUICK REFERENCE GUIDE To ensure a reliable reading avoid eating smoking and exercising for 30 minutes before taking a measurement You should try to measure your blood pressure at the same time each day Avoid taking measurements during stressful times 1 Remove any tight fitting clothing from your upper arm 2 Sit in a chair with your feet flat on the floor and place your arm on a table so the cuff is at the same level as your heart 3 Apply the cuff to your left upper arm so the blue arrow is on the inside of your arm and aligned with your middle finger The air tube runs down the inside of your arm The bottom of the cuff should be approximately 1 2 above your elbow 4 Wrap the cuff firmly in place around your arm using the closure strip 1 2 inch 1 2 cm Marker Air Tube 22 QUICK REFERENCE GUIDE 5 Press the START STOP button Display symbols appear The cuff starts to inflate automatically Remain still and do not talk during the measurement NOTE To stop the inflation or measurement push the START STOP button The monitor will stop inflating start deflating and will turn off 6 When the measurement is complete the arm cuff completely deflates Your blood pressure and pulse rate are displayed NOTE
19. guo se elimina para guardar el registro m s reciente El monitor tambi n calcula una lectura promedio basada en los valores de las tres mediciones m s recientes Si no se han ajustado la fecha y la hora los valores de medici n no se guardar n correctamente en la memoria PARA MOSTRAR EL VALOR PROMEDIO of lll S mbolo de valor promedio 1 Presione el bot n de MEMORY memoria El s mbolo de valor promedio aparecer en la pantalla con la lectura promedia S17 USO DE LA FUNCI N DE MEMORIA PARA MOSTRAR LOS VALORES DE MEDICI N La fecha y hora se muestran alternativamente con los valores de medici n 1 Presione el bot n de MEMORY memoria para mostrar en la pantalla los valores de medici n m s recientes Presione el bot n de MEMORY memoria repetidamente para mostrar los valores de medici n m s antiguos ieg Pantalla alternada NOTAS Presione el bot n repetidamente para mostrar los siguientes valores Presione y mantenga el bot n presionado para mostrar los valores m s r pidamente 2 Presione el bot n START STOP inicio paro para desconectar la alimentaci n PARA ELIMINAR LOS VALORES ALMACENADOS EN LA MEMORIA Usted no puede eliminar parcialmente los registros almacenados en la memoria Se eliminar n todos los registros 1 Presione y mantenga presionado el bot n de MEMORY memoria y el bot n START STOP inicio paro simult neamente durante m s de 2 segundos S18
20. he memory correctly TO DISPLAY THE AVERAGE VALUE on Moll Average Value Symbol 1 Press the MEMORY button The average value symbol will display on the screen with the average reading 17 USING THE MEMORY FUNCTION TO DISPLAY THE MEASUREMENT VALUES The date and time is alternately displayed with the measurement values 1 Press the MEMORY button to display the most recent measurement values on the screen Press the MEMORY button repeatedly to display the oldest measurement values i Given 311 003 Alternating display NOTES Press the button repeatedly to display the next values Press and hold the button to display the values faster 2 Press the START STOP button to turn the power off TO DELETE ALL VALUES STORED IN THE MEMORY You cannot partially delete records stored in the memory All records will be deleted 1 Press and hold the MEMORY button and START STOP button simultaneously for more than 2 seconds CARE AND MAINTENANCE To keep your digital blood pressure monitor in the best condition and protect the unit from damage follow the directions listed below Keep the monitor in the storage case when not in use Make sure the AC Adapter is placed under the main unit so that is does not damage the display Do not forcefully bend the arm cuff or air tube Do not fold tightly Clean the monitor with a soft dry cloth Do not use any abrasive or volatile cleaners Do not attempt
21. ifica que el monitor detecta el movimiento de la sangre a trav s de la arteria braquial y convierte el movimiento a una lectura digital Un monitor oscilom trico no necesita estetoscopio por lo tanto el monitor es f cil de usar La unidad HEM 711AC viene con los siguientes componentes Monitor Brazalete est ndar para adulto Adaptador de CA Manual de instrucciones impreso en ingl s y espa ol Estuche de almacenamiento El monitor de la presi n arterial de inflado autom tico HEM 711AC IntelliSense est dise ado para uso domiciliario GUARDE ESTAS INSTRUCCIONES S2 NOTAS DE SEGURIDAD IMPORTANTES Para asegurar el uso correcto del producto deben tomarse medidas de seguridad b sicas entre ellas las precauciones listadas a continuaci n M Antes de usar la unidad lea toda la informaci n descrita en el manual de instrucciones y cualquier otra documentaci n incluida en la caja M Comun quese con su m dico para obtener informaci n espec fica acerca de su presi n arterial Siga las instrucciones de su proveedor de atenci n m dica MI Use la unidad s lo seg n lo indicado No la use para ning n otro prop sito V La unidad est dise ada para medir la presi n arterial y frecuencia del pulso en pacientes adultos No use la unidad en beb s ni en personas que no pueden expresar sus intenciones M No utilice un tel fono celular cerca de la unidad Esto puede causar una falla de funcionamiento V N
22. mmended NOTE Measurement values stored in the memory will not be deleted during battery replacement After replacing the batteries reset the date and time Refer to page 10 Setting the Date and Time USING THE AC ADAPTER Use only the authorized Omron AC Adapter Model Number HEM ADPTI with this monitor Use of any other adapter may cause damage to the monitor and will void the user warranty To connect the AC Adapter O Insert the AC Adapter Plug into the AC Adapter Jack on the back side of the monitor as illustrated Plug the AC Adapter into a 120V AC outlet 60 cycles To disconnect the AC Adapter remove the AC Adapter from the electrical outlet first and then remove the AC Adapter Plug from the monitor NOTE The monitor is designed not to draw power from the batteries when the AC Adapter is used SETTING THE DATE AND TIME The Blood Pressure Monitor automatically stores up to 60 measurement values with the date and time The monitor also calculates an average reading based on the last three measurement values Set the monitor to the current date and time before taking a measurement for the first time If the current date and time is not set a default date and time will display 1 TO START THE SETTING Press the Date Time Setting button to adjust the date and time The year flashes on the display 2 SETTING THE YEAR The year can be set between 2004 and 2030 When the display reaches 2030 it will
23. nterferencia titulada Digital Apparatus ICES 003 del Departamento Canadiense de Comunicaciones Los cambios o las modificaciones no aprobados por el partido responsable de conformidad pod an anular la autoridad del usuario para funcionar el equipo 24 INFORMACI N SOBRE LA GARANT A Su monitor autom tico de presi n arterial IntelliSense modelo HEM 711AC sin incluir el brazalete est garantizado contra defectos de materiales y mano de obra que surjan en un per odo de 5 a os contados a partir de la fecha de compra siempre y cuando se lo utilice seg n las instrucciones suministradas con el monitor El brazalete est garantizado contra defectos de materiales y mano de obra que surjan en el per odo de un a o a partir de la fecha de compra siempre y cuando el monitor sea utilizado seg n las instrucciones suministradas Las garant as anteriormente descritas s lo se extienden al comprador original Nos comprometemos a reparar o reemplazar sin cargo y seg n nuestro criterio cualquier monitor o brazalete cubierto por dichas garant as La reparaci n o el reemplazo es nuestra nica responsabilidad y su nico recurso bajo las garant as mencionadas Para obtener servicios cubiertos por la garant a comun quese con el Servicio al Cliente de Omron Healthcare llamando al 1 800 634 4350 para obtener la direcci n del local de reparaci n as como informaci n sobre los gastos por env o de retorno y tramitaci n La
24. o enchufe ni desenchufe el cable de alimentaci n el ctrica con las manos h medas V Los cambios o modificaciones no aprobados por Omron Healthcare anular n la garant a del usuario No desarme ni trate de reparar la unidad ni los componentes S3 ANTES DE MEDIR LA PRESI N ARTERIAL Para asegurar una lectura confiable siga estas recomendaciones 1 Evite comer fumar y hacer ejercicios los 30 minutos anteriores a la medici n Descanse por lo menos 15 minutos antes de la medici n 2 El estr s aumenta la presi n arterial Evite realizar mediciones durante momentos de estr s 3 Las mediciones deben realizarse en un lugar tranquilo 4 Qu tese la ropa apretada del brazo izquierdo 5 Si ntese en una silla con los pies en el suelo Coloque su brazo izquierdo sobre una mesa de manera que el brazalete est al mismo nivel que su coraz n 6 Permanezca quieto y no hable durante la medici n 7 Mantenga un registro de las lecturas de su presi n arterial y frecuencia del pulso para su m dico Una sola medici n no proporciona una indicaci n precisa de su verdadera presi n arterial Es necesario tomar y anotar varias lecturas durante un per odo de tiempo Trate de medir su presi n arterial a la misma hora cada d a 8 Espere 2 a 3 minutos entre mediciones El tiempo de espera permite que las arterias vuelvan al estado en que estaban antes de la medici n de tensi n arterial Es posible que necesite aumentar el
25. pid Air Release Automatic exhaust valve Pressure Detection Capacitive pressure sensor Measurement Method Oscillometric method Pulse Wave Detection Capacitance type pressure sensor Power Source AC Adapter or 4 AA batteries Battery Life Approx 300 uses when used once a day for 2 minutes with 4 new with standard manganese batteries Operating Temperature Humidity 50 F to 104 F 10 C to 40 C 30 to 85 RH Storage Temperature Humidity 4 F to 140 F 20 C to 60 C 10 to 95 RH Main Unit Weight Approximately 12 1 2 oz 355 g Main Unit Dimensions Approx 6 3 32 1 x 5 5 32 w x 3 5 16 h 155 mm x 131 mm x 84 mm Cuff Size Approximately 5 1 2 w x 19 1 140 mm x 483 mm Cuff tube 23 5 8 1 600 mm Cuff Circumference Fits arm circumferences 9 to 13 229 mm to 330 mm Contents Main Unit Arm Cuff AC Adapter Instruction Manual and Storage Case UPC Code 0 73796 07115 8 NOTE These specifications are subject to change without notice 26 Manufactured for OMRON HEALTHCARE INC 1200 Lakeside Drive Bannockburn Illinois 60015 www omronhealthcare com Made in China O 2005 Omron Healthcare Inc 711ACNINST Rev OMRON QA Intell es Monitor Autom tico de la Presi n Arterial con Brazalete de Brazo Modelo HEM 711AC ESPA OL ENGLISH NDICE Antes de usar el monitor TO UCI lid ida 2
26. resione el bot n START STOP El monitor detendr el inflado iniciar el desinflado y se apagar 2 El inflado se detiene autom ticamente y comienza la medici n A medida que se desinfla el brazalete en la pantalla aparecen n meros de conteo regresivo i So n 36 B ip 5 El s mbolo del coraz n 42 EN e pA 8 00 8 00 8 0 parpadea con cada latido del coraz n S15 C MO REALIZAR UNA MEDICI N 3 Cuando termina la medici n el brazalete del brazo se desinfla completamente Aparece en la pantalla su presi n arterial y frecuencia de pulsos 4 Presione el bot n START STOP inicio paro para apagar el monitor NOTA El monitor se apagar autom ticamente despu s de cinco minutos INSTRUCCIONES PARA CONDICIONES ESPECIALES Si sabe que su presi n sist lica es m s de 220 mmHg presione y mantenga presionado el bot n START STOP inicio paro hasta que el monitor se infle a un valor 30 a 40 mmHg mayor que su presi n sist lica sospechada NOTAS No aplique m s presi n que la necesaria El monitor no se inflar a m s de 299 mmHg S16 USO DE LA FUNCI N DE MEMORIA Cada vez que usted realiza una medici n el monitor almacena los valores de presi n arterial y frecuencia de pulsos en la memoria El monitor de presi n arterial autom ticamente almacena hasta 60 valores de medici n Cuando se han almacenado 60 conjuntos de valores de medici n en la memoria el registro m s anti
27. return to 2004 Press the MEMORY button to advance the digits one at a time Press the Date Time Setting button to set the current year The month flashes on the display SETTING THE DATE AND TIME 3 SETTING THE MONTH Press the MEMORY button to advance by increments of one month Press the Date Time Setting button to set the current month The day flashes on the display 4 SETTING THE DAY Press the MEMORY button to advance by increments of one day Press the Date Time Setting button to set the current day The hour flashes on the display 5 SETTING THE HOUR The time is set using AM or PM Press the MEMORY button to advance by increments of one hour Press the Date Time Setting button to set the current hour The minutes flash on the display 10 SETTING THE DATE AND TIME 6 SETTING THE MINUTE Press the MEMORY button to advance by increments of one minute Press the Date Time Setting button to set the current minute The year flashes on the display 7 Press the START STOP button to turn the display off 11 APPLYING THE ARM CUFF 1 Make sure the air plug is securely inserted in the main unit 3 Sit in a chair with your feet flat on the floor Place your arm on a table so the cuff is level with your heart 12 APPLYING THE ARM CUFF 4 Apply the cuff to your left upper arm so the blue arrow is centered on the inside of your arm and aligned with your
28. sassemble or attempt to repair the unit or components BEFORE TAKING A MEASUREMENT To ensure a reliable reading follow these recommendations Avoid eating smoking and exercising for 30 minutes before taking a measurement Rest for at least 15 minutes before taking the measurement 2 Stress raises blood pressure Avoid taking measurements during stressful times 3 Measurements should be taken in a quiet place 4 Remove tight fitting clothing from your left arm 5 Sit in a chair with your feet flat on the floor Rest your left arm on a table so that the cuff is at the same level as your heart 6 Remain still and do not talk during the measurement 7 Keep a record of your blood pressure and pulse readings for your physician A single measurement does not provide an accurate indication of your true blood pressure You need to take and record several readings over a period of time Try to measure your blood pressure at the same time each day for consistency 8 Wait 2 3 minutes between measurements The wait time allows the arteries to return to the condition prior to taking the blood pressure measurement You may need to increase the wait time depending on your individual physiological characteristics KNOW YOUR UNIT Main Unit Display Battery Compartment MEMORY Button Date Time Setting Button AC Adapter Jack Components Arm Cutt Air Tube AC Adapter 5 UNIT DISPLAY Systolic Blood Press
29. to clean the cuff Never immerse the monitor or any components in water Store the monitor in a safe and dry location Do not subject the monitor to extreme hot or cold temperatures humidity and direct sunlight Avoid subjecting the monitor to strong shocks such as dropping the unit on the floor Remove the batteries if the unit will not be used for three months or longer Always replace all the batteries with new ones at the same time Use the unit consistent with the instruction provided in this manual Use only authorized parts and accessories 19 ERROR INDICATORS SYMBOL CAUSE CORRECTION Remove the arm cuff Refer to Monitor did not detect Applying the arm cuff pulse rate Wait 2 3 minutes Take another measurement Cuff under inflated Remove the arm cuff Read Taking a Measurement on page 16 E Wait 2 3 minutes Cuff over inflated Take another measurement to Battery Installation Remove the arm cuff Movement during Wait 2 3 minutes measurement Take another measurement Remain still during the measurment 20 Replace the four batteries Refer Batteries are worn E E E 8 TROUBLESHOOTING TIPS CAUSES AND SOLUTIONS No power Replace worn batteries No display appears on the unit Check the battery installation for proper placement of the battery polarities Blood pressure varies constantly Many factors including stress time of day how you wrap the cuff may aff
30. tric method of blood pressure measurement This means the monitor detects your bloods movement through your brachial artery and converts the movements into a digital reading An oscillometric monitor does not need a stethoscope so the monitor is simple to use The HEM 711AC comes with the following components Monitor Standard Adult Arm Cuff AC Adapter nstruction Manual printed in English and Spanish Storage Case The HEM 711AC IntelliSense Automatic Inflation Blood Pressure Monitor is intended for home use SAVE THESE INSTRUCTIONS 2 IMPORTANT SAFETY NOTES To assure the correct use of the product basic safety measures should always be followed including the precautions listed below V Read all information in the instruction book and any other literature included in the box before using the unit MI Contact your physician for specific information about your blood pressure Follow the instructions of your healthcare provider V Operate the unit only as intended Do not use for any other purpose V The unit is intended for use in measuring blood pressure and pulse rate in adult patient population Do not use on infants or persons who cannot express their intentions M Do not use a cellular phone near the unit It may result in operational failure Y Do not plug or unplug the adapter power cord with wet hands Y Changes or modifications not approved by Omron Healthcare will void the user warranty Do not di
31. ure Diastolic Blood Pressure Memory Symbol Heartbeat Symbol I Pulse Display Deflation Symbol Average Value Symbol Movement Error Symbol Date Time Display Battery Low Symbol Irregular Heartbeat Symbol IRREGULAR HEARTBEAT SYMBOL When the monitor detects an irregular rhythm two or more times during the measurement the Irregular Heartbeat Symbol O will appear on the display with the measurement values The irregular heartbeat rhythm is defined Normal Heartbeat as a rhythm that varies by more than or less pulse AMAIA A than 25 from the average rhythm detected while the monitor is measuring the pressure systolic and diastolic blood pressure Irregular Heartbeat Short Long If the irregular heartbeat symbol Y Pul e displays with your measurement results we Blood SINA IAN he pressure recommend you consult your physician Follow the directions of your doctor BATTERY INSTALLATION 1 Press the W indicator on the battery cover and slide the cover off in the direction of the arrow 2 Install 4 AA size batteries so the positive and negative polarities match the polarities of the battery compartment as indicated 3 Replace the battery cover BATTERY REPLACEMENT 9X Low Battery Indicator When the Low Battery Indicator appears on the display screen remove all the batteries Replace with four new batteries at the same time Long life alkaline batteries are reco
32. ww omronhealthcare com Enclose the Proof of Purchase Include a letter with your name address phone number and description of the specific problem Pack the product carefully to prevent damage in transit Because of possible loss in transit we recommend insuring the product with return receipt requested ALL IMPLIED WARRANTIES INCLUDING BUT NOT LIMITED TO THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR PARTICULAR PURPOSE ARE LIMITED TO THE DURATION OF THE APPLICABLE WRITTEN WARRANTY ABOVE Some provinces do not allow limitations on how long an implied warranty lasts so the above limitation may not apply to you OMRON SHALL NOT BE LIABLE FOR LOSS OF USE OR ANY OTHER INCIDENTAL CONSEQUENTIAL OR INDIRECT COSTS EXPENSES OR DAMAGES Some states do not allow the exclusion or limitation of incidental or consequential damages so the above exclusions may not apply to you This warranty gives you specific legal rights and you may also have other rightswhich may vary from state to state FOR CUSTOMER SERVICE Visit our web site at www omronhealthcare com Call toll free 1 800 634 4350 25 Model HEM 711AC Display LCD Digital Display Measurement Range Pressure 0 to 299 mmHg Pulse 40 180 min Memory 60 measurements with date and time Accuracy Calibration Pressure 3mmHg or 2 of reading Pulse 596 of reading Inflation Automatic by electric pump Deflation Automatic pressure release valve Ra
33. y This digital apparatus does not exceed the Class B limits for radio noise emissions from digital apparatus as set out in the interference causing equipment standard entitled Digital Apparatus ICES 003 of the Canadian Department of Communications Changes or modifications not expressly approved by the party responsible for compliance could void the user s authority to operate the equipment 24 WARRANTY INFORMATION Your HEM 711AC IntelliSense Automatic Blood Pressure Monitor excluding the monitor cuff is warranted to be free from defects in materials and workmanship appearing within 5 years from the date of purchase when used in accordance with the instructions provided with the monitor The monitor cuff is warranted to be free from defects in materials and workmanship appearing within one year from the date of purchase when the monitor is used in accordance with the instructions provided with the monitor The above warranties extend only to the original retail purchaser We will at our option repair or replace without charge any monitor or monitor cuff covered by the above warranties Repair or replacement is our only responsibility and your only remedy under the above warranties To obtain warranty service contact Omron Healthcare s Customer Service by calling 1 800 634 4350 for the address of the repair location and the return shipping and handling fee Information for warranty service is available on our website at w