Monarch MINI HUMIDITY TEMPERATURE METER 315 User's Manual
Contents
1. the second display 5 Insert the probe into a temperature controlled chamber at 40 C 104 F and again allow the system to stabilize for 10 minutes Press the MAX MIN button to initiate calibration A single long beep will indicate calibration is complete and the instrument will revert to normal operating mode
2. EDA 1604 or JIS 006P or IEC6F22 Battery Life Approx 85 hours of continuous operation with alkaline battery Low battery indication Operating Condition e Operating Temperature Humidity 0 C 40 C 32 F 104 F lt 80 RH e Storage Temperature Humidity 10 C 60 C 14 F 140 F lt 70 RH e Altitude Up to 2000 meters 6500 feet Dimension 240x54x34 mm 9 5 2 1 1 3 in Weight Approx 180g 6 4 oz Accessories Battery Instruction Manual Options 33 RH Humidity standard 75 RH Humidity standard MODEL 315 HUMIDITY TEMPERATURE METER 5 4 Notes on Temperature and Humidity Calibrations 1 During either calibration pressing the button at any time will return the instrument to the normal operating mode and no calibration data will be changed 2 While performing the humidity calibration the environment should be maintained at stable conditions i e constant temperature and constant humidity to optimize the accuracy 3 Always begin the temperature calibration process with the 0 C 32 F temperature controlled chamber and the humidity calibration process with the 32 8 RH 25 C standard cavity For maximum efficiency insert the probe into the conditioned environment 10 minutes for temperature or 30 minutes for humidity before applying power to the instrument and starting the calibration process 4 Whenever the probe is inserted into a standard humidity cavity or temperature controlled chamber alwa
3. MONARCH INSTRUMENT Instruction Manual C Monarch 315 Mini Humidity Temperature Meter 15 Columbia Drive Amherst NH 03031 2334 USA Phone 603 883 3390 Fax 603 886 3300 E mail support monarchinstrument com 1071 8041 310 Website www monarchinstrument com Monarch Instrument 2002 all rights reserved Safeguards and Precautions Read and follow all instructions in this manual carefully and retain this manual for future reference Do not use this instrument in any manner inconsistent with these operating instructions or under any conditions that exceed the environmental specifications stated Do not touch or manipulate the sensor Do not expose the sensor to direct light This may cause false readings Do not expose the sensor to static electricity Making measurements of high or low temperatures can be dangerous Keep the hand holding the temperature probe well away from the object being measured This instrument is not user serviceable For technical assistance contact the sales organization from which you purchased the product or Monarch Instrument directly LIMITED WARRANTY SELLER warrants hardware products to be free from any defect in materials or workmanship for a period of one 1 year from date of shipment to BUYER SELLER s entire liability and BUYER s sole and exclusive remedy resulting from any defect in workmanship or material in the hardware product covered by this limit
4. easing these buttons again press the HOLD button to restore factory calibration values A single long beep will indicate restoration is complete MODEL 315 HUMIDITY TEMPERATURE METER 3 0 Symbol Definition and Button Locations Data Hold indication Minmunm indication Maximum indication Low battery indication Relative indication Humidity display Unit indication Auto power off indication Temperature display Feature Locations 1 Dust mask 2 LCD display 3 ON OFF button 4 MAX MIN button 5 HOLD button 6 RELATIVE button 7 C F control button 8 Battery cover e Bgg amp d t MODEL 315 HUMIDITY TEMPERATURE METER 4 0 Operating Instructions 41 4 2 Preparation for Measurement To measure ambient humidity and temperature hold the instrument in the environment of interest and allow sufficient time for the sensors to respond to ambient conditions See Section 2 0 Response Time Power Press the button to turn the instrument ON or OFF 4 3 Selecting the Temperature Scale 4 4 4 5 4 6 When first turned on the instrument defaults to reading in Celsius C The instrument toggles between Celsius C and Fahrenheit F each time the C F button is pressed The instrument remembers the scale setting when last turned off and powers on in that setting the next time Data Hold Function The present reading is held on the display by pr
5. ed warranty shall be limited to and fully discharged by the SELLER s option of replacement or repair of such item without charge The limited warranty provided in this clause is in lieu of all other warranties expressed or implied arising by law or otherwise ALL IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FORAPARTICULAR PURPOSE ARE EXCLUDED This limited warranty shall not be modified except by an arrangement signed by both parties specifically referencing this clause SELLER warrants that any software supplied will operate in accordance with the documentation or manual supplied therewith in all material respects when used in strict compliance with such documentation or manual Notwithstanding the foregoing BUYER acknowledges that since software is complex and therefore may have defects BUYER s sole and exclusive remedy for any such defects or breach of this warranty shall be to require SELLER within a reasonable period of time to provide all reasonable programming services to correct programming errors in the software Except as provided above SELLER MAKES AND BUYER RECEIVES FROM SELLER NO EXPRESS OR IMPLIED WARRANTIES OF ANY KIND WITH RESPECT TO ALL OR ANY PORTION OF SOFTWARE AND BUYER HEREBY AGREES AND ACKNOWLEDGES THAT IT ACCEPTS THE SOFTWARE IN AS IS CONDITION SELLER HEREBY EXPRESSLY EXCLUDES ANY IMPLIED WARRANTIES OF MERCHANTABILITY OR FITNESS FORA PARTICULAR PURPOSE WITH RESPECT TO THE SOFTWARE BUYER agrees that any spec
6. essing the HOLD button Pressing the HOLD button again releases the hold function and returns the instrument to continuous reading In the HOLD mode the AREL MAX MIN and C F buttons are disabled as indicated by two consecutive beeps when pressed Relative Operation Pressing the AREL button internally memorizes the present reading and the main display shows the difference between the new reading and the memorized value Press the AREL button again to exit the Relative operation MAXI MIN Operation Press the MAX MIN button to enter the MAX MIN mode In this mode both the maximum and minimum values are simultaneously retained in memory and updated with every new data sample The instrument first enters the MAX mode and the MAX symbol and value are both displayed Pressing MAX MIN again advances the display to the MIN symbol and value The next press of the MAX MIN button will cause both the MAX and MIN symbols to blink This indicates that the maximum and minimum values have been updated in memory and the displayed reading is the present value Each successive press of the MAX MIN button circulates the display mode among these options To exit the MAX MIN mode press and hold MAX MIN for two seconds In the MAX MIN mode the AREL and C F buttons are disabled as indicated by two consecutive beeps if pressed 4 7 4 8 4 9 MODEL 315 HUMIDITY TEMPERATURE METER Auto Power Off By default the instrument powers on in the A
7. fore conforms in accordance with 89 336 EEC EMC Directive The testing of this product was performed by GesTek EMC Lab in February and March of 2002 Ref No 0202033E Ain L Alan Woolfson VP Engineering Authorized Signature Kes 6 March 2002 Importer Amherst NH MODEL 315 HUMIDITY TEMPERATURE METER Table of Contents Section Page 1 0 ele Tok 1 errr e 1 41 4 2 4 3 4 4 4 5 4 6 4 7 4 8 4 9 Selecting the Temperature Scale Data Hold Operation Relative Operation MAX MIN Operation Auto Power Off Low Battery Condition Calibration Points 5 0 Calibration Procedure 5 1 5 2 5 3 5 4 Temperature Calibration Humidity Calibration Restoring Factory Calibration Notes on Temperature and Humidity Calibration MODEL 315 HUMIDITY TEMPERATURE METER 1 0 Introduction This instrument is a digital Humidity Temperature meter incorporating a polymer capacitive humidity sensor and a semiconductor temperature sensor in a compact self contained enclosure 2 0 Specifications Measurement Range Humidity 1 RH 99 RH Temperature 20 C 60 C 4 F 140 F Accuracy at 23 C 5 C Feature Range Accuracy Resolution Humidity 0 1 RH Temperature 0 1 C 0 1 F Response Time Humidity 180 sec in slowly moving air Temperature 10 sec in slowly moving air Display Dual four digit display Sample Rate 2 5 times per second Power 9 Volt Battery N
8. ific right or remedy granted to BUYER hereunder with respect to any breach or default by SELLER shall be in lieu of all other rights and remedies otherwise available to BUYER at law or in equity as the result of such breach or default regardless of whether based on contract tort strict liability or other theory of liability IN NO EVENT SHALL SELLER BE LIABLE FOR ANY SPECIAL INDIRECT INCIDENTAL CONSEQUENTIAL OR PUNITIVE LOSSES OR DAMAGES INCLUDING BUT NOT LIMITED TO LOSSES OR DAMAGES FOR ANY LOST PROFITS OR LOST DATA AS THE RESULT OF ANY BREACH OR DEFAULT BY SELLER WITH RESPECT TO THE HARDWARE OR SOFTWARE EVEN IF SELLER HAS BEEN ADVISED OR MADE AWARE OF THE POSSIBILITY OF ANY SUCH LOSSES OR DAMAGES AND REGARDLESS OF WHETHER THE CLAIM IS BASED ON CONTRACT TORT STRICT LIABILITY OR OTHER THEORY OF LIABILITY This limited warranty does not extend or apply to consumables including but not limited to lamps and batteries if applicable or equipment instruments or accessories which are warranted separately by the original manufacturer of these items DECLARATION OF CONFORMITY Monarch Instrument Division of Monarch International Inc 15 Columbia Drive Amherst NH 03031 USA declares that the product Name Humidity Temperature Meter Model Monarch 315 to which this declaration relates is in conformity with the following standards EMC EN55011 1991 CISPR 11 Class A EN50082 2 1995 EN61000 4 and there
9. uto Power Off mode and will automatically shut off 30 minutes after the last key operation To disable this feature press and hold the HOLD button and then power on the meter Two successive beeps indicate that Auto Power Off is disabled Low Battery Condition When the battery voltage is at or below the minimum for proper operation the symbol will show on the display indicating that the battery must be replaced Calibration Points Room temperature 23 C 5 C Temperature calibration points 0 C and 40 C Humidity calibration points 32 8 RH 25 C and 75 3 RH 25 C 5 0 Calibration Procedure Note Conduct temperature calibration before humidity calibration 5 1 Temperature Calibration 1 Turn the unit off Press and hold the HOLD and C F buttons While these two buttons are pressed turn the power on Release all buttons and all the segments on the display will blink 2 Within 3 seconds of releasing these buttons again press the C F button to enter the calibration mode or the instrument will revert to normal operating mode 3 Upon entering the calibration mode CAL1 will appear on the second display 4 Insert the probe into a temperature controlled chamber at 0 C 32 F and allow the system to stabilize for 10 minutes Press the MAX MIN button to initiate calibration A single long beep will indicate calibration is complete and the instrument will advance into the CAL2 mode as indicated on
10. ys allow at least 30 minutes to ensure ambient conditions have stabilized MODEL 315 HUMIDITY TEMPERATURE METER 5 2 Humidity Calibration 1 Turn the unit off Press and hold the HOLD and C F buttons While these two buttons are pressed turn the power on Release all buttons and all the segments on the display will blink 2 Within 3 seconds of releasing these buttons press the AREL button to enter the calibration mode or the instrument will revert to normal operating mode 3 Upon entering the calibration mode CAL1 will appear on the second display 4 Place the probe in a stable 32 8 RH environment at 25 C and allow the system to stabilize for 30 minutes Press the MAX MIN button to initiate calibration A single long beep will indicate calibration is complete and the instrument will advance into the CAL2 mode as indicated on the second display 5 Insert the probe in a stable 75 3 RH environment at 25 C and again allow the system to stabilize for 30 minutes Press the MAX MIN button to initiate calibration A single long beep will indicate calibration is complete and the instrument will revert to normal operating mode 5 3 Restoring Factory Calibration 1 To restore factory default calibration values turn the instrument off Press the HOLD and C F buttons While these two buttons are pressed turn the power on Release all buttons and all the segments on the display will blink 2 Within 3 seconds of rel
Download Pdf Manuals
Related Search
Related Contents
MULTÍMETRO DIGITAL Philips Full-body solarium HB556 Bushnell 98-0917/04-09 Binoculars User Manual Guía de usuario del Nokia 5610 XpressMusic Nav XT User Manual - Hofmann AG, Langenthal Brochure - Luminaire Napert Artisan Technology Group is your source for quality new and JET B100-T User manual HSS-AZ Copyright © All rights reserved.