Home

Hasselblad Flextight 848 User's Manual

image

Contents

1. Figure 7 Replacing the Light Tubes User s Guide Flextight 848 Hardware Reference Flextight 848 Cleaning the External Surface of the Scanner 27 continued 8 Remove the defective tube e Press the flaps backwards to release the tube e Pull the tube gently out of the holders and then sideways out of the scanner 9 Insert the replacement tube and mount it into the holders making sure that it is properly held in place by the flaps 10 Connect the plug F 11 IMPORTANT The T on the lamp board cover G indicates the socket for the transparencies tube 12 Lower the feed table E 13 Mount the side panels C and D the screws B and the front panel A following the reverse of the procedure used when removing them NOTE When mounting the front panel A make sure to give it a firm push at the bottom to click it properly into position 14 Turn on the scanner and check function of the new tube Cleaning the External Surface of the Scanner If the external surface of the scanner becomes dirty then clean it with a damp cloth Be careful not to get any moisture inside the scanner or on any of the connectors Do not use alcohol or other solvents Disposal If you need to dispose of the scanner it must be delivered to an authorized waste plant for electronics equipment User s Guide Flextight 848 Hardware Reference Flextight 848 Technical Specifications 28 Technical Spec
2. User s Guide Flextight 848 2004 Hasselblad A S All rights reserved Flextight 848 User s Guide Part No 70030049 revision A The information in this manual is furnished for informational use only is subject to change without notice and should not be construed as a commitment by Hasselblad A S Hasselblad A S assumes no responsibility or liability for any errors or inaccuracies that may appear in this manual Hasselblad A S assumes no responsibility or liability for loss or damage incurred during or as a result of using Hasselblad software or products Hasselblad Hasselblad FlexColor and Flextight are trademarks of Hasselblad A S Adobe and Adobe Photoshop are trademarks of Adobe Systems Inc Macintosh Mac OS FireWire and ColorSync are registered trademarks of Apple Computer Inc Printed in Denmark User s Guide Flextight 848 Table of Contents 3 Table of Contents Hardware Reference Flextight 848 5 Warnings and Restrictions 6 System Requirements Front Panel 9 Back Panel amp Connectors 10 Cut Away View 11 Setting up the Scanner 12 Environmental Requirements 12 Electrical Requirements 12 Installation Procedure FireWire Interface 13 Operating Instructions 15 Resolutions 19 True Resolutions 19 Calibrating the Scanner 20 Focus Calibration 21 White Calibration for Reflectives 22 The Light Tubes 26 General 25 Replacing the Light Tubes 25 Cleaning the External Surface of the Scanner 27 Disp
3. at full scanning wdd 35 mm Portrait 35 mm Landscape 45 x 60 mm Portrait 60 x 60 60 x 70 mm 4 x 5 Portrait 13 x 18 Portrait A4 Portrait Table 1 True Resolutions User s Guide Flextight 848 Hardware Reference Flextight 848 Calibrating the Scanner 20 Calibrating the Scanner The scanner requires two types of calibration e Focus calibration adjusts the positioning of the scanners optics for each zoom level If your scans are coming out unfocused or at slightly the wrong size then you may need to make a new focus calibration to adjust the zoom mechanism The focus calibration is stored in the scanners flash PROM so it is preserved when you move the scanner to a new computer See Focus Calibration on page 21 for more information e White calibration is only required for reflective scans it is done automatically for transparencies During calibration the scanner scans a white target to establish the white point along the full length of the scanners CCD The results of the white calibration scan are saved on your computers hard disk so you must make a new white calibration if you move the scanner to a new computer See White Calibration for Reflectives on page 22 for more information Both of the calibration procedures are simple and are initiated from the FlexColor program User s Guide Flextight 848 Hardware Reference Flextight 848 Calibrating the Scanner 21 Focus Ca
4. Do not place your fingers or any other object into the scanner while it is connected to power e Before servicing or opening the scanner the power supply must be disconnected from the mains unplugged It is not sufficient simply to press the on off button e Install the scanner in a location where children can not get to it It contains small openings and moving parts that can cause injury User s Guide Flextight 848 Hardware Reference Flextight 848 7 System Requirements System Requirements Below are the basic hardware requirements for the PC or Macintosh system to which the scanner is to be connected For information about the processor operating system RAM and harddisk requirements please refer to the Software Reference manual that comes with the FlexColor software e Screen resolution of 800 x 600 pixels with true colors 24 bits e Mouse or other pointing device e FireWire interface User s Guide Flextight 848 8 Hardware Reference Flextight 848 User s Guide Flextight 848 Front Panel Hardware Reference Flextight 848 Front Panel LOOKS SS S82 lt lt LLP ROLLS LOR KLEE LICKEL I oS BOC LK OL SOR KSEE CREEL tet ses Figure 1 Front panel features of Flextight 848 A Feed Table must be in lower position for normal scanning When scanner is not in use you can turn it to upright position flushing with the scanner front to protect it from dust etc Adjust to
5. grid on the light table B to help line up the image e For reflectives make sure that no part of the original extends past the edge of the holder Use the dotted lines printed on the holder to line up the image Handle the clear plastic layer only by the edges to avoid fingerprints Curl the plastic layer back and slip the front edge into Flextight s foil holder continues User s Guide Flextight 848 Hardware Reference Flextight 848 Operating Instructions 18 continued 9 Lay the top layer of the holder down flat over the original e For transparency holders the top layer is magnetic which will hold the original fast For the reflective holder note the four metal pegs the top two are shown as A and C in Figure 5 below make sure that the clear plastic sheet is pressed all the way down over all four pegs Note also the two curved tabs at the top of the holder B in the figure below the clear sheet must slip under each of these tabs as shown Figure 5 Reflective holder top 10 Go to your computer and follow the directions given in the FlexColor software manual to take a preview make settings and take the final scan User s Guide Flextight 848 Hardware Reference Flextight 848 Resolutions 19 Resolutions True Resolutions The table below shows the max resolutions in ppi available with each of the original holders when scanning in True Resolutions mode wos Max resolution ppi
6. L NORMAL Figure 4 Operating the Flextight 848 scanner User s Guide Flextight 848 Hardware Reference Flextight 848 Operating Instructions 17 continued NOTE Each original holder has it s own unique identification code a combination of small rectangular holes at the holder s leading edge These codes combined with the scanner s ability to detect the feed table position ensure that the scanner will not operate unless the feed table position matches the selected original holder 6 Place the original holder K with the slotted tab facing into the scanner For transparencies the flexible magnetic layer must face upwards For reflectives the clear plastic sheet must face upwards Make sure that the original holder fits between the appropriate guides C or D on both sides of the light table 7 Slide the holder gently into the slot E at the top of the feed table It will slip about cm 1 4 inch into it Do not press too hard It slips in very easily and is held in place by a magnetic clasp To remove the holder simply pull it gently backwards 8 Lift the top layer of the original holder and place your original L as follows e For transparencies place your original with the emulsion side down The original must completely fill the hole with no edges showing and with a minimum overlap of 2mm along each edge Also no edges of the original must extend past the outer edges of the holder Use the
7. ardware Reference Flextight 848 Cut Away View 11 Cut Away View Figure 3 Cut away view of Flextight 848 A CCD Housing this movable platform contains the light sensor and holds the optics Positioning Lead Screws for positioning of the CCD housing and optics housing for the appropriate resolution and original size Optics Housing focuses the image of the original on the CCD Upper Light Source illuminates reflective originals Lower Light Source illuminates transparent originals Drum rotates the original into the scanner and steps it past the CCD line of focus Original Holder a transparency holder is shown mounted and with the top layer held open Mimo ow D User s Guide Flextight 848 Hardware Reference Flextight 848 12 Setting up the Scanner Setting up the Scanner Environmental Requirements Set up the scanner in a location that fulfils the following requirements e Keep the scanner away from sources of heat such as direct sunlight or a radiator Warm temperatures will degrade the quality of your scans for best results work in a cool environment e The scanner must be operated away from sources of strong electromagnetic interference Although the scanner complies with all regulations governing electromagnetic immunity and Hasselblad has taken every reasonable step to make the unit immune to electromagnetic interference it is still a precision electronic device so strong radio waves can inte
8. bes one for illuminating reflective originals from above the other for shining light through transparencies from below Replacing the Light Tubes If your scans are suddenly coming out black or all white if you are scanning negatives then one of your light tubes may be defective Make a preview using both types of scans transparent and reflective to find out which tube is affected e If transparencies are not working then the bottom tube may be burned out e If reflectives are not working then the top tube may be burned out Follow procedure below to replace a tube See Figure 7 opposite 1 Get the appropriate replacement lamp ask your Flextight dealer 2 Turn off and unplug the scanner from the primary power supply 3 IMPORTANT Let the light tubes cool down for app 10 minutes before proceeding 3 Remove the front panel A by pulling it outwards at the bottom and lift it off 4 Remove the two black allen screws B one in each side using a 3 mm allen key Do not remove any other screws 5 Remove the two side panels C and D by lifting them slightly upwards and then outwards NOTE It may require a firm push upwards to release the panels 6 Turn the feed table E to upright position 7 Locate the tube to be replaced and dismount the plug F continues User s Guide Flextight 848 Hardware Reference Flextight 848 26 The Light Tubes PULL OUT PULL OUT
9. er uses a white reference to make sure that all elements in the scanner s CCD react consistently to light brightnesses The scanner has been calibrated from the factory and the calibration data is stored in the scanner s flash PROM As the scanner ages the properties of the light tubes may change slightly Therefore you must redo the white calibration every six months or so and immediately after changing the light tubes Use procedure below to make White Calibration 1 Find the white calibration sheet that came with your scanner It is a square white piece of paper 219 x 219 mm 8 6 x 8 6 If you have lost the white calibration sheet then you can use a flat non textured perfectly white piece of paper of the same dimensions 2 Your scanner came with several clear plastic layers for the reflective original holder If you have used the reflective holder before make sure that there are no finger prints or other marks on the plastic layer Replace the layer with a new one if necessary 3 Align the top edge of the sheet with the dotted line near the top of the original holder see Figure 6 opposite Align each side of the target with the edges of the original holder 4 Select White Calibration from the Maintenance window The External White Calibration window appears External white calibration External white calibration is used to ensure high quality reflective scans The calibration must be performed after installatio
10. horizontal position when using the 35 mm slide mount holder see page 15 Light Table for viewing originals The grid helps you to align your transparent originals Transparency Holder Guide all transparency holders fit between two rails to keep them straight Original Holder Clasp all original holders slide into a slot here and are held in place by a magnetic clasp Power Switch press here to turn the unit on off Power Indicator Green Remains lit when ready to scan Flashes when first installed to indicate that firmware must be loaded it will be loaded automatically when you run FlexColor Motor Drive Indicator Yellow lights when the scanner is repositioning the optics to scan in a new format or when drum is repositioning Scan Indicator Red lights when scanning Do not touch the scanner while this indicator is lit If the light is flashing when not scanning then an error has occurred please contact your Flextight dealer for assistance User s Guide Flextight 848 10 Hardware Reference Flextight 848 Back Panel amp Connectors Back Panel amp Connectors Figure 2 Back panel features of Flextight 848 A B Power Cable Socket plug the power cable into this socket FireWire Interface Sockets plug a FireWire cable max 6 m here and connect it to your computer You can use the second plug to connect another device hard disk printer etc to the FireWire chain User s Guide Flextight 848 H
11. ht 848
12. ifications Power Connector 24 VDC 60W oe Power Consumption Max 60 W during operation Power Requirements 100 240V AC 1 3A 50 60Hz Interface to Computer FireWire IEEE 1394 Front Panel One multifunction switch on off switch Status indicator LED s Drum Force 1kg 2 2lbs Operating Environment Operating Temperature 10 35 C 50 95 F Recommended Working Temperature 10 25 C 50 77 F Humidity 20 80 RH no condensation Storage Environment Temperature 0 50 C 32 122 F Relative humidity 20 80 RH no condensation Originals Transparencies from 35mm to 4x5 inches lt 1mm thick Reflectives up to 220 x 310mm A4 oversize lt 1mm thick Standard 35 mm dias in frames optional Lamp type Color optimized fluorescent lamp Dimensions Height 645 mm 25 4 Width 385 mm 15 2 Depth 220 mm 8 7 feed table closed 440 mm 17 3 feed table open Weight 17 kg 38 lbs User s Guide Flextight 848 Hardware Reference Flextight 848 FCC Notice 29 FCC Notice This equipment has been tested and found to comply with the limits for a Class A digital device pursuant to Part 15 of the FCC Rules These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment This equipment generates uses and can radiate radio frequency energy and if not installed and used in accordance with the instructi
13. libration The Flextight 848 scanner uses an adjustable zoom and focus mechanism to optimize its scanning resolution for each of the original formats it can handle If you suspect your scans are not as sharp as they should be then you may need to calibrate your scanner Note that you must not calibrate the scanner every day It is intended that the scanner is calibrated not more than once every three months In most cases this procedure will never be necessary Use procedure below to make Focus Calibration 1 Locate the Focus sheet that came with your scanner It is a 8 x 9 cm 3 x 3 5 square of transparent film with a black image on it TT BASS el imacon _ 2 Load the sheet into the 6 x 6 original holder with the lines pointing towards the scanner vertically Place the sheet as straight as possible Use the scanners light table grid and the corners marked on the sheet to help align it 3 Select Focus Calibration from the Maintenance menu Mount transmission focus sheet f Cancel d Calibrate y 4 Click on the Calibrate button and the focus calibration starts 5 When the focus calibration is finalized the focus calibration window closes again Your scanner is now calibrated for all resolutions and for both transparent and reflective originals User s Guide Flextight 848 Hardware Reference Flextight 848 Calibrating the Scanner 22 White Calibration for Reflectives The Flextight scann
14. n thereafter it should be repeated every 6 months or more frequently in the case of extensive scanner usage The procedure is as follows 1 The best result is obtained by using the Imacon supplied white calibration sheet but good quality white paper can also be used It should cover the entire width of the A4 original holder the minimum lenath is 210 mm 2 Place the calibration sheet at the top of the A4 original holder aligned to the upper dotted line Make sure to use a new foil without scratches and fingerprints 3 Press Calibrate to start the calibration procedure a a User s Guide Flextight 848 Hardware Reference Flextight 848 Calibrating the Scanner 23 Align the top edge of the calibration sheet with the dotted line near the top of the original holder D Align the side edges of the calibration sheet with the side edges of the original holder Figure 6 Positioning the white calibration sheet on the reflective holder 5 Click on Calibrate The scanner will take several scans of the white target The process will take about eight minutes User s Guide Flextight 848 Hardware Reference Flextight 848 24 The Light Tubes PULL OUT PULL OUT Figure 7 Replacing the Light Tubes User s Guide Flextight 848 Hardware Reference Flextight 848 The Light Tubes 25 The Light Tubes General The Flextight 848 contains two light tu
15. on manual may cause harmful interference to radio communications Operation of this equipment in a residential area is likely to cause harmful interference in which case the user will be required to correct the interference at his own expense User s Guide Flextight 848 30 Hardware Reference Flextight 848 CE Declaration of conformity CE Declaration of conformity HASSELBLAD HASSELBLAD A S HEJREVEJ 30 DK 2400 COPENHAGEN NV TELEPHONE 45 7026 0800 TELEFAX 45 7026 0900 REG VAT NO 18635445 www hasselblad com Declaration of conformity Application of Council Directives 06 89 336 EEC amended by 92 31 EEC 93 68 EEC 91 157 EEC C73 23 EEC amended by 93 68 EEC Standards to which Conformity is declared EN61000 6 1 2001 EN61000 6 3 2001 and EN60950 2000 Manufacturer cccceccceeeceeeceeeeeeseeeeeeeees Hasselblad A S Hejrevej 30 DK 2400 Copenhagen NV Denmark Type Of EQUIPMENT ceeeeeeeeeeeeeeeeeeeeeeeees Desktop CCD scanner IMOGEI NAME citarieentecciedaiiidavauniecevealendiueagens Flextight 848 The undersigned hereby declare that the equipment specified above conforms to the above Directives and Standards PIACE ipusa raa aaaea tena cau Paa EEN Copenhagen NV Date dairies anana a Ea AGE Aa AEAN September 12th 2004 UI MANE anssen eaaa Christian Poulsen POSIIOM facechonetncedadsetcensdceutesssvadieasssecnitannats CEO Signatur User s Guide Flextig
16. osal 217 Technical Specifications 28 FCC Notice 29 CE Declaration of conformity 30 User s Guide Flextight 848 User s Guide Flextight 848 Hardware Reference Flextight 848 5 Hardware Reference Flextight 848 This manual provides important information about using your Flextight 848 scanner Topics include e Important warnings and restrictions e System requirements e Diagrams of the front panel rear panel and internal structure of the Flextight 848 scanner e Installation instructions e Environmental requirements e Electrical requirements e Operating instructions e Special considerations for scanning 35mm originals e Calibration instructions e Maintenance advice e Disposal instructions e Technical specifications e Declarations of FCC and CE regulation conformity User s Guide Flextight 848 Hardware Reference Flextight 848 Warnings and Restrictions Warnings and Restrictions e Read all of the included documentation before attempting to install and use the scanner e Do not touch the originals and or the original holder while scanning e Do not start scanning or previewing until an original holder with an original has been mounted e The original holder is only to be mounted or removed when the drum is in the load position e When you turn on the scanner the drum will roll to the load position if it is not there already Do not touch the scanner while the drum is rotating e
17. rfere with your scans The surface on which the scanner is placed must be stable and free from vibrations If the scanner is shaken or moved while scanning your results may be affected If the scanner has been in a colder environment e g outside or in storage just before you set it up in a warmer room then wait about two hours before using it otherwise condensation may form which will prevent the scanner from operating correctly e Avoid using the scanner in areas where there is a high level of dust Electrical Requirements The Flextight 848 power supply requires a mains voltage of between 100 and 240 V AC at a frequency of 50 to 60 Hz This is within the normal wall socket power standards of most countries Do not attempt to use Flextight 848 with any power source outside the specified range The scanner and all devices attached to the scanner computer monitor FireWire devices etc must be grounded i e use a three point electrical connection Note that the power supply may emit a low hissing sound when plugged in but not connected to the scanner This is normal and will not damage the power supply or scanner User s Guide Flextight 848 Installation Procedure FireWire Interface 1 Hardware Reference Flextight 848 Setting up the Scanner 13 Place the scanner on a table near your computer with the front facing away from you Make sure that the location you choose meets all environmental and elect
18. rical requirements as described on page 11 When using a FireWire interface connection it is not necessary to switch of your computer and other connected units Connect a FireWire cable from one of the FireWire connectors on your scanner see Back Panel amp Connectors on page 10 to either the computer s FireWire connector or to a free FireWire connector on any other FireWire device already connected to your computer Connect the round connector from your power supply to the round power supply socket on the back of the scanner IMPORTANT Make sure that the wall socket you are planning to use meets the electrical requirements outlined earlier Plug the Flextight 848 power supply into a wall socket Note that the power supply may emit a low buzzing sound when plugged in This is normal and will not damage the power supply or scanner Turn the scanner around so that the front is facing you Your system is assembled Turn on your computer and install the FlexColor image scanning software User s Guide Flextight 848 Hardware Reference Flextight 848 Operating Instructions 14 CLOSED HORIZONTAL NORMAL Figure 4 Operating the Flextight 848 scanner User s Guide Flextight 848 Operating Instructions Hardware Reference Flextight 848 Operating Instructions 15 See Figure 4 opposite 1 Itis assumed that the scanner and FlexColor software have been properly installed 2 Ifnot alread
19. y done turn the feed table A to normal position 3 Press the red power button F on the front of the scanner to switch it on The green light G starts flashing This means that no firmware has been downloaded yet It will be downloaded automatically later when you first run the FlexColor program 4 Turn on your computer and start the FlexColor program The green light G will stop flashing indicating that firmware has been downloaded to the scanner 5 Select the original holder K that matches your original e For transparencies choose one of the holders that has a square hole in it The original must completely fill the hole with no edges showing Also no edges of the original must extend past the outer edges of the holder The feed table A must be in normal position For reflectives select the large A4 letter size holder with the clear plastic layer The feed table A must be in normal position e For 35 mm slide frames use the optional 35 mm slide mount holder The feed table A must be adjusted to horizontal position This is obtained by lifting the table slightly upwards while pushing it gently into the scanner Please note that when scanning 35 mm slide frames in the 35 mm slide mount holder it is recommended to use autofocus See the FlexColor Software Reference for details continues User s Guide Flextight 848 Hardware Reference Flextight 848 Operating Instructions 16 CLOSED HORIZONTA