Drive Medical Design AS8800 User's Manual
Contents
1. drive Drive Medical Design amp Manufacturing 99 Seaview Boulevard Port Washington NY 11050 Toll Free 877 224 0946 Local 516 998 4600 www drivemedical com REV2 12 17 13 526001 5150 V1 2 Low air loss Alternating pressure mattress Instruction Manual PLEASE READ ALL INSTRUCTIONS BEFORE USE Contents Important Safeguards rrr trent nennnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnn 2 Pere eee eee eee een eee een eee ee 3 Product Description vtttttttmnnmnnnnnnnnnnnniiinnininnnnnnnnnnnnnnnnnnnniiain 4 E E Ce eae cee 7 Operation rr rinnnnnnnnsnnnnnnaininnnnnuinininnnnaiininnnniininnnnniiinnnnniiisn 9 Cleaning crn 1 Storage corr innnnnnnnnnnunnnnnnnainnnnnnnniinininnnnnniinnnnnnnaiiinnnnnn 1 SAPIENS tee eee 40 Technical Description nnee 13 SAVE THESE INSTRUCTIONS GROUNDING INSTRUCTIONS This product should be grounded In the event of an electrical short circuit grounding reduces the risk of electric shock by providing an escape wire for the electric current This product is equipped with a cord having a grounding wire with a grounding plug The plug must be plugged into an outlet that is properly installed and grounded DANGER Improper use of the grounding plug can result in a risk of electric shock If repair or replacement of the cord or plug is necessary do not connect the grounding wire to either flat blade terminal The wire with insulation having an outer surface that is green with or without yellow stripes is t
2. it on a soft surface such as a bed or couch where their openings may be blocked Keep the air opening free of lint hair and other similar particles 7 Never drop or insert any object into any opening or hose 8 Connect this product to a properly grounded outlet only See Grounding Instructions NOTE CAUTION AND WARNING STATEMENTS NOTE Indicate some tips CAUTION Indicate correct operating or maintenance procedures in order to prevent damage to or destruction of the equipment or other property WARNING Calls attention to a potential danger that requires correct procedures or practices in order to prevent personal injury YMBOLS BF symbol indicate this product is according to the degree of protecting against electric shock for type BF equipment oO Oo Grounding terminal for 230V system Functional earth for UL only Attention should read the instructions Disposal of Electrical amp Electronic Equipment WEEE This product should be handed over to an applicable collection point for the recycling of electrical and electronic equipment For more detailed information about the recycling of this gues Product please contact your local city office household waste disposal service or the retail store where you purchased this product m B Class II for UL only 1 INTRODUCTION This manual should be used for initial set up of the system and for reference purposes 1 1 GENERAL INFORMATION
3. soak pump unit By wiping the mattress unit with a damp cloth pre soaked with warm water not to exceed 65 C containing a mild detergent and keep it away from dust The cover may also be cleaned using sodium hypochlorite diluted in water All parts should be air dried thoroughly before use CAUTION Do not use phenolic based product for cleaning CAUTION After cleaning dry the mattress without direct exposure of sunlight The carrying bag if available should be turned inside out and completely wiped down using the disinfectant solutions Allow it to air dry thoroughly Once the inside is dry turn it back and wipe down the outside of the bag with disinfectant solutions 6 STORAGE 1 To store lay the mattress out flat and upsides down 2 Roll from the head end towards the foot end 3 Foot end strap can then be stretched around the rolled mattress to prevent unrolling Note Do not fold crease or stack the mattresses 1 CPR 2 Pump unit 3 Mattress System 4 Quick Connectors 2 2 PUMP UNIT 4 3 PRESSURE SET UP 2 Users can adjust pressure of air mattress to a desired softness by adjusting the comfort keys by weight Please consult your physician for a suitable setting Front 1 Quick Connectors Patient can lie on the mattress after the pump is running under the alternating mode this indicates r 2 Front Panel the pressure is reached to the desired softness NNOTE Check to see if the suitable pressure is selected b
4. The system is a high quality and affordable mattress system suitable for stage Il Ill bed sore treatment and prevention of pressure ulcers The system has been tested and successfully approved with the following standards EN 60601 1 EN 60601 1 2 C EN 55011 Class B IEC61000 3 2 IEC 61000 3 3 N FOR US AND CANADA ONLY Medical Equipment Air Pump with respect to electrical shock fire and mechanical hazards only in accordance with UL60601 1 AND CAN CSA C22 2 NO 601 1 E245395 53DG Le produit t test avec des quipements m dicaux et respecte les normes UL 60601 1 amp CAN CSA C22 2 No 601 1 pr venant les choc lectrique le feu et les risques de blessures physiques 1 2 INTENDED USE This product is intended to help and reduce the incidence of pressure ulcers while optimizing patient comfort It also provide following purposes e Individual home care setting and long term care of whom suffering from pressure ulcer e Pain management as prescribed by physician A NOTE Equipment not suitable for use in the presence of a flammable anesthetic mixture with air or with oxygen or nitrous oxide A NOTE L quipement ne peut tre utilis s il y a risque de m lange d un anesth sique inflammable avec l air ou l oxyg ne ou oxyde nitreux 7 MAINTENANCE 7 1 GENERAL 1 Check main power cord and plug if there are abrasions or excessive wears 2 Check mattress cover for signs of wear or damage Ensure m
5. attress cover and tubes are stubbed together correctly 3 Check airflow from the air hose connector The airflow should alternate between each connector every half cycle time if it s in alternating mode 4 Check the air hoses for any kink or break For replacement please contact our local dealers 7 2 FUSE REPLACEMENT 7 3 AIR FILTER REPLACEMENT Milh 2 ESA O 12 Disconnect the plug form mains power when a blown fuse is suspected Remove the cover of the fuse holder by means of a small screwdriver Insert a new fuse of the correct rating in and replace the cover of the fuse holder back The fuse should be rated as T1A 250V type and VDE approved Loosen the screws from the air filter plate located at the back of pump The filter is reusable and can be washed gently with a mild detergent and water Dry the filter before use Check and replace air filter regularly if environment is dirty 5 CLEANING 2 PRODUCT DESCRIPTION poeni l l 2 1 PUMP AND MATTRESS SYSTEM It is important to follow the cleaning procedures before using the machine on human bodies otherwise patients and or doctors may have the possibility of getting infection By wiping the pump unit with a damp cloth pre soaked with a mild detergent and keep it away from dust If other detergent is used choose one that will have no chemical effects on the surface of the plastic case of the pump unit CAUTION Do not immerse or
6. ct for any damage which may have occurred during shipment If there are any damages please contact your dealer immediately 3 1 PUMP amp MATTRESS INSTALLATION Please note for the foot end It is recommended to use this mattress system with a bed frame with adequate side rails to prevent falling 2 Hang the pump onto bed rail foot end and adjust hangers to best upright position of the pump or place the pump on a flat surface 3 Connect air hose connectors from air mattress to the pump unit When a click sound is felt or heard the connection is completed and secured NOTE Check and ensure the air hoses are not kinked or tucked under mattress 1 Place the mattress or pad on top of the bed frame 4 Plug the power cord into electrical outlet NOTE 1 Make sure the pump unit is suitable for the local power voltage 2 The plug is also served to disconnect the device A NOTE 1 S assure que la pompe est compatible au voltage local ou disponible 2 Lappareil est galement muni d une fiehe de connexion a l lectricit CAUTION The pump can only be applied to the mattress recommended by the manufacturer Do not use it for any other purpose applied part air mattress AVERTISSEMENT La pompe ne peut tre utilis e qu avec le nee ee ee A matelas design ou r 4 ar eet o J recommand par le f pol manufactureier Ne pas server d ysage autre que i celui auq
7. he grounding wire NOTE If the repair or replacement of the cord is necessary please contact a qualified electrician or serviceman To reduce the risk of electric shock do not modify the cord or plug in any way Check with a qualified electrician or serviceman if the grounding instructions are not completely understood or if in doubt as to whether the product is properly grounded 14 8 TECHNICAL DESCRIPTION Specifications Item Specification Power Supply Note See AC 120V 50 60 Hz 0 17A for 120V system rating label on the product AC 230V 50 60 Hz 0 07A for 230V system Fuse Rating T1A 250V Cycle time Four selectable Dimension ec x W x H 29 1 x 20 x 11 7 cm or 11 4 x 7 9 x 4 6 F Wagt o 2 2 Kg or 5 Ib Operation 10 C to 40 C 50 F to 104 F Temperature Storage 15 C to 50 C 5 F to 122 F Shipping 15 C to 70 C 5 F to 158 F Environment i i Operation 10 to 90 non condensing Classification Humidity Storage 10 to 90 non condensing Shipping 10 to 90 non condensing Class II with functional earth NOTE 5 Type BF IPXO Applied Part Air Mattress Not suitable for use in the presence of a flammable anesthetic mixture No AP or APG protection Mattress Specification 8 Mattress 200 x 90 x 20 3 cm Dimension L x W x H Weight 7 5 kg or 16 5 Ib NOTE or 78 7 x 35 4 x 8 1 Consult the distributor or EU representative for further tech
8. ible D Seat Inflation The seat inflation features additional supports to the patient during upright position without bottoming out User can select this additional feature under either static or alternate mode 4 Panel Lock Should the panel remain untouched for 3 minutes the panel lock feature will lock the panel with green LED light on to prevent accident from changing the setting during normal operation To unlock simply by pressing the Panel lock button for 3 second to unlock 5 Alarm Mute Press alarm mute button to temporary suspend the low pressure alarm LED light and buzzer When there is an emergency to perform CPR on the patient quickly pull the CPR valves to release air from mattress The CPR valves are located at the head end right hand side of the mattress The quick connector found from the pump unit can be disconnected for even faster deflation Should the situation not resolved within 5 mins the alarm shall resume to notify the patient A PFA Power Failure Alarm During power failure situation the Power Failure LED light will lit on with buzzer By pressing the mute button to disable both buzzer and LED B Low Pressure Indicator When low pressure LED lights up the pressure inside of air mattress is below normal Please refer to troubleshooting 6 Service This feature will lit during mechanical failure situation User can notify the technician for repair 3 INSTALLATION Unpacking the box to inspe
9. nical documents 2 The specification is also suitable for other areas operating with same power supply 3 Mattress dimension and weight is measured without foam cushion 4 The manufacturer reserves the right to modify the specification without notice 5 The earthing terminal is served as functional earth not protective earth for UL only 13 IMPORTANT SAFEGUARDS READ ALL INSTRUCTIONS BEFORE USING DANGER To reduce the risk of electrocution 1 Always unplug this product immediately after using 2 Do not use while bathing 3 Do not place or store product where it can fall or be pulled into a tub or sink 4 Do not place in or drop into water or other liquid 5 Do not reach for a product that has fallen into water Unplug immediately WARNING To reduce the risk of burns electrocution fire or injury to persons 1 This product should never be left unattended when plugged in 2 Close supervision is necessary when this product is used by on or near children or invalids 3 Use this product only for its intended use as described in this manual Do not use attachments not recommended by the manufacturer 4 Never operate this product if it has a damaged cord or plug if it is not working properly if it has been dropped or damaged or dropped into water Return the product to a service center for examination and repair Keep the cord away from heated surfaces Never block the air openings of this product or place
10. sa for decreasing air pressure Higher pressure output will support the heavier weight patient In order to find the correct pressure setting First inflate the patient to max pressure level have the patient lie on top of the mattress then decrease the pressure gradually until the patient feel most comfortable without bottoming out Have the caregiver insert palm facing up and place between patient s buttock s area and static cell Should the palm able to touch patients buttock increase the pressure until the skin is least 1 2 inch away from the palm Always leave at least 1 inch space between patient s buttock areas to the static cell under to prevent bottoming out 4 OPERATION A NOTE Always read the operating instruction before use 4 1 GENERAL OPERATION 4 2 CPR Switch on the main power switch found from the side of the pump The pump unit is to delivering air into mattress wait least 20 minutes for mattress to be inflated The low pressure indicator yellow LED will light up when the mattress is not fully inflated When the appropriate pressure is reached the low pressure indicator yellow LED will go off and pump will enter the alternating mode According to the weight and height of the patient adjust the pressure setting to the most comfortable level without bottoming out then the pressure in mattress will slowly increase to a certain value after the air mattress is ready to
11. uel elle a ete con u 5 Then turn the main power switch to ON position Grounding Before any connection to the output connectors is made the unit shall be connected to a protective earth conductor via the three core main cable the mains plug shall be inserted only into a socket outlet provided with a protective earth contact The protective action shall not be negated by the user of an extension cord without protective conductor Several installation tips are listed below After installation the extra length of the power cord if any should be neatly arranged to avoid any tripping accidents The EQUIPMENT should be firmly placed at position where users doctors can access easily
12. use A NOTE Every time when mattress is first setup for use it is recommended that the pressure knob to be set to Max for the quickest inflation User can then adjust air mattress to the desired softness afterward 2 Cycle Time There are four selectable cycle times found from the touch panel area By pressing icon sign user can select four different cycle times for best comfort 3 Therapy A Autofirm This mode will be forced to execute every time the power switch is turned on This insures the pump is able to reach its maximum operating pressure Once the max pressure level is reached the pump will automatically switch into alternating mode User can also use this function as full mattress inflation during ingress egress the patient for better support B Alternate There are four selectable cycle time available User can select for best comfort C Static Press STATIC button to suspend alternating function if needed The pressure inside of air tubes will be adjusted to the same softness Press the STATIC button again it will switch back to alternating mode Under the static mode cell pressure level will be lowered compare to the same pressure level from alternating mode This special feature is to protect high risk pressure sore patient from receiving additional pressure which could lead to pressure sore risks There is a 20 mins auto recovery mode back to alternating cycle to ensure patient always getting best therapy poss
13. y sliding one hand between the deflated air cells and the patient to feel patient s buttocks Users should be able to feel the minimum contact 4 4 LOW PRESSURE INDICATION FUNCTION Rear 5 1 Power Switch When an abnormal low pressure occurred the Low Pressure indicator yellow LED will light up 2 Power Cord along with the buzzer Check to ensure all connections are connected properly and that they are 3 Fuses correctly installed as per installation instructions 4 Air Filter 5 Mounting Brackets NOTE If the pressure level is consistently low check for any leakage tubes or connecting hoses If necessary replace any damaged tubes or hoses Or contact local qualified dealer for repair 2 3 FRONT PANEL The Low Pressure indicator will light up until the low pressure fault condition is resolved 4 5 STATIC MODE Press the blue STATIC button to suspend the alternating mode of mattress By pressing this button again the mattress will be back to alternating therapy mode 4 6 ALARM MUTE When low pressure situation occurred both the LED light and buzzer will sound off to warn the 1 Pressure Adjust Levels patient By pressing the button it will temporary mute the buzzer so the care giver may check for Pressure adjust levels controls the air pressure output When turning up gt the output possible air leaks Should the situation not resolved within 5 minutes the alarm will resume to beep pressure will increase Vice ver
Download Pdf Manuals
Related Search
Related Contents
Epson Perfection 1250 PHOTO Scanner Product Brochure User Manual Mode d`emploi - Nintendo of Europe Minka Lavery 4952-267B Instructions / Assembly Lenovo ThinkPad E520 TS CKSTST6711 & CKSTST6712 Especificaciones INSECTICIDE EN POUDRE DRIONE® *USAGE COMMERCIAL User Manual for 3040 Copyright © All rights reserved.