Home

A&D PV-200 User's Manual

image

Contents

1. Instruction Manual AD A amp D WEIGHING Clearly a Better Value 1555 McCandless Drive Milpitas CA 95035 Tel 800 726 3364 408 263 5333 Fax 408 263 0119 Fax on Demand 800 726 7099 Press 2 Website www andweighing com Email scales andweighing com ELECTRONIC PRECISION SCALE USER MANUAL FOR MODEL PV 60 PV 100 PV 200 and PV 500 PLEASE READ BEFORE USING SCALE A amp D WEIGHING Clearly a Better Value GENERAL DESCRIPTION This is a portable top loading digital electronic scale The sensing mechanism is a load cell INSTALLATION A Set the scale on a flat rigid surface located away from dust vibration extreme heat sources electromagnetic sources such as electric motors or generators and fluorescent lamp starters within 3 feet typically Any slope may lead to inaccurate readings For greater accuracy recalibrate the balance see below whenever the balance is relocated In the event of temperature change allow 1 2 minutes for the scale to stabilize PANEL LAYOUT AND DISPLAY FORMAT A Panel Layout a ON key used to turn on the scale b TARE key used to zero the scale or for taring a container to eliminate its weight c MODE key a short press selects grams oz troy oz or pennyweight d OFF key used to turn off the scale B Display Format a This is a 4 digit liquid crystal display Weighing units are shown on the right hand side of th
2. ITNESS FOR A PARTICULAR PURPOSE ALL SUCH WARRANTIES ARE EXPRESSLY DISCLAIMED IN NO EVENT WILL A amp D S LIABILITY UNDER THE FOREGOING WARRANTY EXCEED THE PURCHASE PRICE PAID FOR THE PRODUCT GIVING RISE TO THE LIABILITY IN NO EVENT SHALL A amp D BE LIABLE FOR ANY LOSS INCONVENIENCE OR DAMAGE INCLUDING WITHOUT LIMITATION DIRECT INDIRECT SPECIAL INCIDENTAL OR CONSEQUENTIAL DAMAGES RESULTING FROM OR RELATED TO THE USER OR INABILITY TO USE A ORDUCT WHERHER RESULTING FROM BREACH OF WARRANTY OR ANY OTHER LEGAL THORY EVEN IF A amp D HAS BEEN ADVISED OF THE POSSIBILITY OF SUCH INCONVENIENCE DAMAGE OR LOSS A amp D WEIGHING 1555 MCCANDLESS DRIVE MIPITAS CA 95035 8001
3. e display b When first turned on the display shows all possible display segments confirming that power is on and then the scale begins its search for zero c EE or EE indicates overload or underload respectively When the E E signal blinks it indicates that the tare process is prohibited die to the overload or underload conditions Cb indicates the power from the battery is too low for normal operation e Battery Saver feature scale will automatically shut off after an average of 5 7 minutes to save battery life CALIBRATION PROCEDURE A Make sure scale is in the gram mode B Press TARE key scale will read zero C Place correct calibration weight on platform PV 500 200g calibration weight PV 200 100g calibration weight PV 100 50g calibration weight PV 60 50g calibration weight Any electronic balance may be calibrated periodically to assure it accuracy D Press the OFF and MODE keys simultaneously until display shows CAB then release The scale is now calibrated The display should read the same as the calibration weight being used 5 CAUTIONS a Never leave a load on the scale for a prolonged period of time as this may damage the loadcell b Always place a load gently onto the platform Dropping the weight on may cause PERMANENT DAMAGE to loadcell c Never try to disassemble the scale which may PERMANENTLY DAMAGE the mechanism and electronic component
4. ork the first time we suggest you keep trying PV 60 50g calibration weight PV 100 50g calibration weight PV 200 100g calibration weight PV 500 200g calibration weight A amp D Engineering Inc A amp D warrants this product to be free from defects in materials and workmanship and to conform to its published product specifications under normal use and service for the period of one year commencing from the date of purchase The foregoing warranty is extended only to the first purchaser and applies only to product failures due to defective materials and or workmanship Without limiting the generality of the foregoing the foregoing warranty will not apply if upon inspection it is found that the product was misused or abused used for a purpose for which it was not designed mishandled placed in an improper environment repaired by unauthorized personnel or improperly installed or adjusted The sole and exclusive remedy in the event of any breach of the foregoing warranty shall be for the original purchaser to return the product to A amp D freight and insurance prepaid for repair replacement or credit at A amp D s option Any product repaired or replaced under the foregoing warranty shall be shipped back to customer freight and insurance prepaid EXCEPT AS PROVIDED ABOVE A amp D MAKES NO WARRANTY CONCERNING THE PRODUCTS EXPRESSED OR IMPLIED INCLUDING BUT NOT LIMITED TO THE IMPLIED WARRANTIES OR MERCHANTABILITY AND F
5. s There are no user serviceable parts inside the scale Maintenance should be done by properly trained technicians For maximum accuracy place loads as near to the center of the platform as possible e Do not expose the scale to direct sunlight or store in high humidity Serious jarring of the scale should be avoided If your scale reads E or EE Sometimes an electronic scale can be shifted out of its working range possibly by a jolt in shipping or transportation When this occurs your scale tells you by displaying EE or EE This can be corrected 1 Press OFF and TARE simultaneously The display will show In meaning input You will be able to input the info needed to get the scale back to its normal function 2 Press OFF and TARE simultaneously a second time a number will appear that will count down pause for a few seconds and eventually stop at a final number Press TARE AND the scale will display zero 4 Place the calibration weight onto the scale and press OFF and MODE simultaneously until CAb appears on the display A number will then appear but not the actual weight of the calibration weight 5 Remove the calibration weight from the scale and turn off the scale to get out of the input function The scale now can be turned back on and should be calibrated normally At this point you should be back in business 9 gt If this procedure does not w

Download Pdf Manuals

image

Related Search

Related Contents

DR5 適E-2(クサラブレーク)  User manual Saey NEEMA  VL6000 - Velleman  430 515 BA AgroLyt lambda  LG 22LB4510 Energy Guide : Free Download, Borrow, and Streaming : Internet Archive  Terrano 3-8 FX - Horsch Maschinen GmbH  SERVICE MANUAL - Appliance Factory Parts  水中蛍光灯 取扱説明書 pdf  USER MANUAL - Rohde & Schwarz  konzeption eines web-frameworks zur gestenerkennung auf basis  