        Pyle PLMT21 lightmeter
         Contents
1. A Figure B       
2. PYLE    www pyleaudio com    MODEL PLMT21    Digital Lux Meter    Operation Manual    A  Features    Precise and instant readout   Wide range of measurements  Reading hold function   Low battery indicator   Level range selection   Compact  light weight  with excellent  operation    B  Operation instruction   1  Install 9V battery into battery com    partment properly   2  Push slide switch to turn on and select  your desired measurement range   2000 Lux LCD display  X1LUX  20000 Lux LCD display X10 LUX  100000 Lux LCD display  X100 LUX  3  Before removing the light sensor cap   The LCD display should read   000  LUX   Remove the sensor cap and take measu   rement digital reading shows out immedi   ately   If the LCD display    1    that means  over load  please select higher range  for measurement   4  Push the slide switch to HOLD to hold  the reading  LCD display icon   5  When the battery   s power drop into  6 5 7 0V  LCD display icon    indicates low battery  please replace  the battery    6  After measurement  cover the  sensor cap and turn off the unit    7  Clean the surface of the sensor  with a moist cotton cloth     Specification   Range  1 10 0000 LUX  Accuracy    lt  10000 LUX   4  10dgt     10000 LUX   5  10dgt   Sampling rate  2 times sec  Temperature characteristic  0 1  F  Repeatability     2   Power supply  9V battery  Operation temperature   327 to 104F  Operation humidity   10RH  to 9ORH           Install  9V battery                                  Figure
    