Sencor SBD 1470 blood pressure unit

image

Contents

1. 44mm Memory capacity 60 records Measuring range Pressure 0 300 mmHg 0 40 kPa Pulse 40 199 pulses minute Measurement accuracy Pressure 3 mmHg 0 4 kPa at an ambient temperature of 5 40 C 5 mmHg 0 7 kPa at an ambient temperature of 0 4 C and 41 45 C Pulse 5 Adjustable size of the 13 5 19 5 cm Power source 2 x LRO3 AAA Electric shock protection class Medical electrical device with an internal power supply Applied part type B XR Degree of protection against the intrusion of water IPXO Safety of use in the presence of anaesthetic combustible mixtures The device is not suitable for use in the presence of combustible anaesthetic and air mixtures or combustible anaesthetic and oxygen mixtures or mixtures containing oxides of nitrogen Operating mode Continuous operation with short term loading Operating conditions Ambient temperature 0 C to 45 C humidity lt 80 Storage conditions Ambient temperature 20 C to 60 C humidity 10 93 Dimensions of the device 68 x 75 31mm Weight of the device 120 g without batteries Accessories 2 x LRO3 AAA type batteries user s manual We reserve the right to change text and technical specifications EN 12 Copyright 2012 Fast CR a s Revision 11 2012 Digital Blood Pressure Monitor with a Wrist Cuff SBD 1470 INSTRUCTIONS AND INFORMATION REGAR
2. a s Revision 11 2012 Digital Blood Pressure Monitor with a Wrist Cuff SBD 1470 TROUBLESHOOTING In this chapter you will find solutions to problems that you may encounter when using this device If you were unable to remedy the problem according to the following instructions contact an authorised service centre Problem error Possible cause Possible solution message After pressing the On Batteries are flat Replace the batteries Off button the display The batteries are Insert the batteries with the correct polarity does not turn on inserted incorrectly direction as shown in the battery compartment The symbol is shown Batteries are almost Replace the old batteries with new ones on the display flat E1 The cuff is not attached Turn off the device using the On Off button to the wrist or is Attach the cuff to the wrist according to the attached to the wrist instructions in chapter Attaching the cuff and too loosely correct posture of the body and hand during measurement and repeat the measurement E2 The cuff is too tight Turn off the device using the On Off button Attach the cuff to the wrist according to the instructions in chapter Attaching the cuff and correct posture of the body and hand during measurement and repeat the measurement E3 The pressure the cuff Rest for 4 to 5 minutes and then repeat the was exceeded measurement E10 or E11 The device
3. of the heart the measurement results may not be accurate During measurement sit calmly relaxed and don t talk Do not move the hand to which the cuff is attached 3 3 Measuring blood pressure 3 3 1 Turn on the device by pressing the On Off button In a short time all the elements will light up on the LCD display 3 3 2 If residual air remains in the cuff the symbol will appear on the display for a short time and the air will be released The value 0 mmHg or kPa and the time of measurement will appear on the display 3 3 3 The device will automatically pressurise the cuff While the cuff is being pressurised the pulse rate is being detected This is indicated by the flashing y symbol on the LCD display 3 3 4 Then the pressure in the cuff is continuously released and the values of the systolic SYS and diastolic DIA pressure pulse rate and the blood pressure are automatically determined The blood pressure categories are defined in the following table Blood pressure category Measured blood pressure value in G1 G2 G3 mmHg Optimal Normal Normal high Mild Medium Heavy hypertension hypertension hypertension SYS systolic value lt 120 120 129 130 139 140 159 160 179 2180 DIA diastolic value lt 80 80 84 85 89 90 99 100 109 gt 110 If the symbol D appears on the display the device has detected cardiac arrhythmia 3 3 5 Turn off the device by pressing the On Off button If you do not t
4. DING THE DISPOSAL OF USED PACKAGING MATERIALS Dispose of packaging material at a public waste disposal site DISPOSAL OF USED BATTERIES Batteries contain environmentally damaging compounds and therefore do not belong in standard household waste Take the batteries to an appropriate collection point which will provide for their ecological disposal You can obtain the contact for the nearest collection point from you town council or from your retailer DISPOSAL OF USED ELECTRICAL AND ELECTRONIC EQUIPMENT The meaning of the symbol on the product its accessory or packaging indicates that this product shall not be treated as household waste Please dispose of this product at your applicable collection point for the recycling of electrical amp electronic equipment waste Alternatively in some states of the European Union or other European states you may return your products to your local retailer when buying an equivalent new product The correct disposal of this product will help save valuable natural resources and help in preventing the potential negative impact on the environment and human health which could be caused as a result of improper liquidation of waste Please ask your local authorities or the nearest waste collection centre for further details The improper disposal of this type of waste may fall subject to national regulations for fines For business entities in the European Union If you wish to dispose of an electrical or elect
5. Digital Blood Pressure Monitor with a Wrist Cuff User s manual Time Date SENCOR SBD 1470 Prior to using this product please read the user s manual thoroughly even in cases when has already familiarised themselves with previous use of similar types of products Use the product only as described in this manual Keep this manual for later use Caution Not following the instructions contained in this user s manual may lead faulty operation of the device or its damage We recommend saving the original package packaging material receipt and warranty card for the duration of warranty In the case of transportation pack the product using the original packaging materials only EN 1 Copyright 2012 Fast CR a s Revision 11 2012 Digital Blood Pressure Monitor with a Wrist Cuff SBD 1470 CONTENTS WHAT YOU SHOULD KNOW ABOUT BLOOD PRESSURE 3 IMPORTANT INSTRUCTIONS qa Seow etek EEE eae ee Caw woe ae es 4 BASIC FUNCTIONS AND FEATURES OF THE BLOOD PRESSURE MONITOR 5 DESCRIPTION OF THE BLOOD PRESSURE MONITOR 0 2 0 002 cece ee 5 DESCRIPTION OF THE DISPLAY cc tae manto Sea oe ho eae hae baw bt Pee 6 USING THE BLOOD PRESSURE MONITOR 0 0 0 0c cee eee 6 TROUBLESHOOTING 2 dees weg hee EE Soe ee he oe oe 10 MAINTENANGE AND CLEANING cue eae ge iw ce ey Chae he ea
6. by pressing the SET button The device will automatically switch to the month setting mode 2 4 Use the MEM button to set the current month Confirm the setting by pressing the SET V button The device will automatically switch to the day setting mode 2 5 Set the current day of the month using the MEM button Confirm the setting by pressing the SET button The device will automatically switch to the hour format setting mode 2 6 Press the MEM button to set 12 hour or 24 hour time format Confirm the setting by pressing the SET button The device will automatically switch to the hour setting mode 2 7 Use the MEM button to set the current hour Confirm the setting by pressing the SET button The device will automatically switch to the minute setting mode 2 8 Use the MEM button to set the current minutes Confirm the setting by pressing the SET button The device will automatically switch to the measuring unit selection mode 2 9 Select the measuring unit mmHg or kPa using the MEM button Confirm the setting by pressing the SET button Note The standard measuring unit for the measurement of blood pressure is mmHg millimetres of a mercury column 2 10 Setup of the date time and measuring units is complete The device will gradually show the set values and then turn itself off automatically Note Setting range year 2000 2050 time format 12 hour 24 hour In the 12 hour time format the morn
7. detected Movement may affect the measurement result movement during the Rest for 4 to 5 minutes and then repeat the measurement measurement E20 or E21 Measurement error Rest for 4 to 5 minutes and then repeat the measurement If a different error message in the format E number code or Ee number code appears on the screen that is not included in the table turn off the device take the batteries out of it Wait a while and then reinsert the batteries After a few minutes repeat the measurement If the problem persists contact your vendor or an authorised service centre MAINTENANCE AND CLEANING Keep the device clean Wipe off dust using a lightly damp cloth Do not wash the device or the blood pressure cuff under running water or submerge it in water Do not use abrasive cleaning products or petrol for cleaning Otherwise the device may be damaged EN 10 Copyright 2012 Fast CR a s Revision 11 2012 Digital Blood Pressure Monitor with a Wrist Cuff SBD 1470 STORAGE If you will not using the device for an extended period of time remove the batteries Protect the device against impacts and falls Store the device in clean dry place out of reach of children Do not expose the device to direct sunlight or extreme temperature changes CALIBRATION Recommendation To ensure accurate measurement results we recommend the device is calibrated after two years of operation All costs associated
8. ee he ee eee E 10 STORAGE arn EE Oe ee ae ee PE eo Se ee he ee ee es 11 CALIBRATION 3 30 2 eee a ee hee ee hee EE Do ee hee 11 ELECTROMAGNETIC INTERFERENCE i 2 OA dee he ee ee he ee hehe es 11 COMPLIANCE WITHISTANDARDS 2 Oh eee de Seas ee Chee eee etd 11 TECHNICAL SPECIFICATIONS ieee arrid wae tei ME wk ot 12 INSTRUCTIONS AND INFORMATION REGARDING THE DISPOSAL OF USED PACKAGING MATERIALS 13 DISPOSAL OF USED BATTERIES priri Seow eh aretha dee ae es 13 DISPOSAL OF USED ELECTRICAL AND ELECTRONIC EQUIPMENT 13 EN 2 Copyright 2012 Fast CR a s Revision 11 2012 Digital Blood Pressure Monitor with a Wrist Cuff SBD 1470 WHAT YOU SHOULD KNOW ABOUT BLOOD PRESSURE What is blood pressure Blood pressure is defined as the pressure exerted by the blood on the walls of the arteries through which it flows Blood pressure fluctuates during the course of each heartbeat between the maximum systolic and the minimum diastolic value Blood pressure is influenced by many factors such as physical activity fear anger or by a certain time of day Blood pressure changes constantly over the course day Early in the morning it rises and before noon it falls In the afternoon it rises again and then falls in the evening hours Blood pressure may also change within an instant and so the subsequent measurement results ma
9. emove the cuff and press the On Off button to end the pressurisation of the cuff Remember that blood pressure fluctuates over the course of the day and is also affected many factors such as smoking consumption of alcohol taking medicines and physical activity The measurement results should be evaluated by a doctor or another expert who knows your long term health condition Please do not make conclusions on the basis of the results yourself By regularly measuring your blood pressure and recording the measurement results you will provide your doctor with a complete picture of your blood pressure during natural activity Blood pressure values measured using the oscillometric method when using this device are equivalent to the measurement results taken by an experienced observer using the auscultatory listening method using a blood pressure monitor with a stethoscope This device is not designed for the continuous monitoring of blood pressure during medical treatment such as for example operations etc This device is designed for domestic use and does not substitute for professional medical care Keep the device and the batteries out of reach of children EN 4 Copyright 2012 Fast CR a s Revision 11 2012 Digital Blood Pressure Monitor with a Wrist Cuff SBD 1470 BASIC FUNCTIONS AND FEATURES OF THE BLOOD PRESSURE MONITOR Measurement of the systolic and diastolic blood pressure and pulse Detection
10. ended only for adults Do not twist excessively bend the cuff Take care not to damage the cuff by sharp items such as pins needles etc Do not disassemble the device and do not make any alterations to it Do not use the device if your wrist is injured If you suffer from a circulatory system disorder such as atherosclerosis diabetes liver or kidney illness heavy hypertension external circulation disorders etc consult your doctor or an expert healthcare professional about the suitability of using a blood pressure monitor or similar devices If you are undergoing medical treatment or taking medication consult the use of this device with a doctor Rest at least 5 to 10 minutes before measuring blood pressure Wait at least 4 to 5 minutes before measuring again so that your blood circulation can return to the normal state Do not perform measurement sooner than 30 45 minutes after consuming beverages containing caffeine or after smoking cigarettes Remove all tight clothing from your wrist before taking a blood pressure measurement Use the cuff only on the wrist Do not use on another part of the body Do not start measurement until the cuff is attached to the wrist Perform the measurement in a calm and relaxed position Do not move the device during measurement The device automatically releases air when the pressure in the cuff exceeds 300 mmHg If the automatic air release does not occur r
11. he display shows the symbol the display is dim the display does not turn on Note If the polarity is reversed when the batteries are inserted the device may not only not function but may also heat up Do not combine used and new batteries or batteries of various types e g alkaline batteries and rechargeable batteries Saved records will remain stored in the device s memory even after the batteries are replaced EN 6 Copyright 2012 Fast R a s Revision 11 2012 Digital Blood Pressure Monitor with Wrist Cuff SBD 1470 2 Setting the date time and measuring units 2 1 Before taking measurements set the current date time and measuring unit The measured values will be stored in memory together with the date and time of measurement 2 2 After inserting the batteries all the elements will be momentarily lit on the display In the bottom left corner of the display the time value year will start to flash and the device is ready for setup The device will turn itself off automatically if you do not perform any settings within 1 minute In such a case it is necessary to activate the settings mode the following way Press the SET Y button The clock will appear on the screen Press the SET button again and hold it down for approx 3 seconds In the bottom left corner the value year will start to flash The device is ready to be set up 2 3 Use the MEM button to set the current year Confirm the setting
12. hreatening and require professional treatment Symptoms of cardiac arrhythmia Symptoms of cardiac arrhythmia strong or accelerated beating of the heart feeling of tiredness vertigo loss of consciousness lack of breath and chest pain Symptoms of bradycardia slowed down heart activity feeling of tiredness lack of breath vertigo or dizziness Symptoms of tachycardia accelerated heart activity the heartbeat may be felt in the neck or as a beat in the chest with irregular speed feeling of unease weakness lack of breath dizziness sweating and vertigo Copyright 2012 Fast CR a s Revision 11 2012 Digital Blood Pressure Monitor with a Wrist Cuff SBD 1470 Can cardiac arrhythmia be treated Cardiac arrhythmia can to a certain extent be prevented by eliminating the stimuli physical exertion stress smoking consumption of alcohol coffee or other beverages containing caffeine affecting the nervous system Many types of cardiac arrhythmias do not require treatment as they are naturally compensated by the immune system Other types of cardiac arrhythmias must be treated with medication antiarrhythmic agents defibrillator implants or pacemakers The treatment method depends on the type of cardiac arrhythmia age of the patient and their physical condition IMPORTANT INSTRUCTIONS This device is designed for non invasive blood pressure measurement The cuff with an adjustable length of 13 5 19 5 cm is int
13. ing afternoon time is indicated by the abbreviation AM PM 3 Measurement 3 1 Basic instructions for achieving the most accurate measuring results Always take measurements at the same time of day ideally in the morning at noon and in the evening under the same conditions or according to the recommendations of your doctor Do not perform measurement sooner than 30 45 minutes after consuming coffee tea or smoking a cigarette Wait at least 20 minutes after taking hot shower or bath Wait approximately 4 5 minutes before measuring again 3 2 Attaching the cuff and correct posture of the body and hand during measurement Remove any clothing watches bracelets from the left wrist etc before attaching the cuff Attach the cuff the left wrist so that the display of the blood pressure monitor is the same side as your palm The top edge of the cuff should be located 1 to 1 5 cm under the wrist joint Wind the cuff around the wrist and tighten it There should be no space between the cuff and the wrist EN 7 Copyright 2012 Fast CR a s Revision 11 2012 Digital Blood Pressure Monitor with a Wrist Cuff SBD 1470 Seat yourself comfortably on a chair and rest your left elbow on a table The palm faces upwards towards the face and the blood pressure monitor is located level with the heart Slightly open the palm and relax the fingers If the blood pressure monitor is located above or below the level
14. of cardiac arrhythmia Adjustable length cuff for wrist circumferences from 13 5 to 19 5 cm Automatic inflation and air release of the cuff Large LCD display 60 memory positions for storing measurement results including date and time Battery operation DESCRIPTION OF THE BLOOD PRESSURE MONITOR LCD display On Off button Blood pressure cuff 6 Battery compartment cover button Battery compartment SET button EN 5 Copyright 2012 Fast CR a s Revision 11 2012 Digital Blood Pressure Monitor with a Wrist Cuff SBD 1470 DESCRIPTION OF THE DISPLAY 2 D Yr BAS iS CIAL Time Date Cuff air release indicator Flat battery indicator Display of date and time Systolic pressure value Measuring unit Diastolic pressure value Pulse rate number of pulses minute Display of the memory sequence number Measured blood pressure category indicator Detection of wrist movement during measurement Detection of cardiac arrhythmia USING THE BLOOD PRESSURE MONITOR Installing and replacing batteries To power the device use two LR0O3 AAA 2 x 1 5 V type batteries Remove the battery compartment cover and insert two LRO3 AAA type batteries When inserting the batteries ensure the correct polarity as shown in the battery compartment Close the cover The batteries need to be replaced when t
15. ronic device request the necessary information from your seller or supplier Disposal in other countries outside the European Union If you wish to dispose of this product request the necessary information about the correct disposal method from local government departments or from your seller EN 13 Copyright 2012 Fast CR a s Revision 11 2012 Digital Blood Pressure Monitor with a Wrist Cuff SBD 1470 EN 14 Copyright 2012 Fast CR a s Revision 11 2012
16. urn off the device it will turn itself off automatically within 1 minute of the last measurement Remove the cuff from your wrist after completing the measurement EN 8 Copyright 2012 Fast CR a s Revision 11 2012 Digital Blood Pressure Monitor with a Wrist Cuff SBD 1470 4 Recalling memory 41 To show the last measurement record press the MEM button 4 2 Toscroll through the individual measurements in memory use the MEM and SET buttons 43 For each measurement the month day and the time the measurement was taken will be shown in the bottom left hand part of the display 44 The most recent measurement stored in memory always has the sequence number of 1 The maximum memory capacity is 60 measurements As soon as the maximum memory capacity is achieved every new measurement will delete the oldest measurement 5 Deleting memory 5 1 Press the MEM button The display will show the last stored record Then hold down the MEM button for approx seconds The message del all delete all will appear on the display 5 2 Press the SET button to confirm the deletion of all measurements in memory The display will show the message del delete and done completed The device will automatically turn itself off Note If you wish to interrupt the deletion process press the On Off button 53 No values will appear on the display when the memory is subsequently retrieved Copyright 2012 Fast CR
17. with the calibration are borne by the customer ELECTROMAGNETIC INTERFERENCE To prevent measurement inaccuracies caused by electromagnetic interference do not use this device in the vicinity of mobile telephones or microwave ovens COMPLIANCE WITH STANDARDS This device complies with European standards EN 60601 1 Medical electrical devices Part 1 General basic safety and necessary functionality requirements EN 60601 1 2 Medical electrical devices Part 1 2 General basic safety and necessary functionality requirements Group norm Electromagnetic compatibility Requirements and tests EN 1060 1 Non invasive blood pressure monitors Part 1 General requirements EN 1060 3 Non invasive blood pressure monitors Part 3 Specific requirements for electromechanical systems for the measurement of blood pressure This device meets the requirements of the European directive 93 42 EEC al The manufacturing date is marked on the rating label of the device wd Manufacturer Zhongshan Transtek Electronics Co Ltd Jinan Road Jinbiao Minzhong Zhongshan Guandong China Authorised representative for the EU MDSS Medical Device Safety Service GmbH Ec Schiffgraben 41 30175 Hannover Germany EN 11 Copyright 2012 Fast R a s Revision 11 2012 Digital Blood Pressure Monitor with a Wrist Cuff SBD 1470 TECHNICAL SPECIFICATIONS Measuring method Oscillometric Display LCD display size 31 5
18. y vary Why is it important to measure your blood pressure at home Many people have increased blood pressure when they visit their doctor while at home their blood pressure is in the normal range This is the so called white coat syndrome and may affect up to 15 of the population Home blood pressure measurement eliminates the white coat syndrome and provides the doctor with a picture of the various blood pressure levels during your natural activity Blood pressure classification by the World Health Organisation The following table shows the blood pressure classification for an adult person according to the World Health Organisation WHO Blood pressure category Systolic blood pressure Diastolic blood pressure mmHg in mmHg Optimal lt 120 lt 80 Normal 120 129 80 84 High normal 130 139 85 89 Hypertension Stage 1 mild 140 159 90 99 Hypertension Stage 2 medium 160 179 100 109 Hypertension Stage 3 heavy 2180 2110 Isolated systolic hypertension 2140 lt 90 What is cardiac arrhythmia Cardiac arrhythmias are a disorder of the rhythm of the heartbeat They result from a varied creation or conduction of electrical impulses in the heart Many cardiac arrhythmias are only temporary in nature Such types of arrhythmias are considered to be harmless and include the cases where the heart misses or adds a beat This may be caused by strong emotions or exercise However there exist types of arrhythmia which may be life t

Download Pdf Manuals

image

Related Search

Related Contents

Material Safety Data Sheet  3 LIVE VIEW - iView Tech  En Lignes n° 06  ULTRA-MAX SERIES  DECCOSOL MF ME_04  CAN_LIN_vF4_rev (Page 4 - 6) - RS Components International    Cisco Systems OL-14619-01 User's Manual  Snapper MR216015T, MR216015B, MR216015BV, MRP216015B, MRP216015BV Lawn Mower User Manual  認定基準Version1との対照表  