1.               Model UT51 55  OPERATING MANUAL    A Input sensitivity   lt 100mVrms   Overload protection  250Vrms     B 8 Temperature                Range Comment Measuring Condition  Display diode forward   Forward DC current abt  voltage near value   1mABackward DC voltage  Unit    mV    abt 2 8V    Beeper sounds if Continuity Voltage at open circuit  Resistance lt  709 Display  abt 2 8V  near value  Unit   Q       A Overload protection  250V DC or AC RMS                   B 10 Transistor hFE test    Measuring Condition    Can measure NPN or  Basic polarity current abt       hFE  PNP transistor hFE     Range  0 10008 10uA Vce abt 2 8V          Model UT51 55  OPERATING MANUAL    C  Making Measurements    Caution     1  If there is no display or    Ea    is shown on the  LCD when the Meter is switched on  replace  the battery ASAP     2  Never exceed the maximum input voltage or  current limits shown besides the input jacks     A  otherwise the Meter will be damaged and  this is dangerous to life     3  Turn the rotary switch to proper range before    operating        On Off Switch   Capacitance jack    Liquid Crystal Display   Temperature jack    Rotary switch    Transistor jack    Input jack             10          Model UT51 55  OPERATING MANUAL    C 1 Measuring DC Voltage       1  Connect the black test lead to COM jack and  red test lead to    V    jack    2  Set the rotary switch to  V     3  Connect the test leads across with the object  to be measured  LCD appears the
2.    Model UT51 55  OPERATING MANUAL    Contents    A Your Meters Feature  B Specifications   C Making Measurements  D Maintenance   E Accessories   F Using Holster   G Using Strap    Introduction    One kind of UNI T brand new UT50 series  Multimeters is a 3 1 2 digit with steady function  and highly reliable hand held measuring instrument   The Meter uses large scale of integrated circuit  with double integrated A D converter as its core  and has full range overload protection  The Meter  can measure DC current  AC current  DC Voltage   AC Voltage  Resistance  Capacitance  Diode   Temperature  Frequency and Continuity  which is  an ideal tool for users        Model UT51 55  OPERATING MANUAL    Safety Rules    The UNI T UT50 series comply with   IEC 61010  in pollution degree 2   overvoltage category  CAT   1000V and  CAT Il 600V  and double insulation    Use the Meter only as specified in this  manual  otherwise the protection provided  by the Meter may be impaired    CAT I For signal level  telecommunication   electronic with small transient over voltage   CAT II  For local level  appliances  main  wall outlets  portable equipment    The meter is designed to withstand the  stated Max  voltages  If it s not possible   to exclude without doubts that impulses   transients  disturbance or for other reasons   these voltages are exceeded a suitable  prescale  10 1  must be used    Do not operate the Meter before the cabinet  has been closed and screwed safely as  terminal can carr
3.   Fax   852  2950 9303   Email  infoQuni trend com   http   www uni trend com    20       
4.  measuring  value and also the polarity of the red test lead     A Caution    1  If magnitude of the voltage is unknown  always  start with the highest range and reduce until  satisfactory reading is obtained    2  If 1    is shown on the LCD  which means the  Meter is overloaded  then set the measuring  range to higher    3     AY means never exceed the maximum input  limits 1000V  otherwise internal circuit of the  Meter will be damaged    4  Take extra care of voltage leakage when  measuring high voltage     C 2 Measuring AC Voltage   1  Connect the black test lead to COM jack and  red test lead to  V  jack     2  Set the rotary switch to    V        3  Connect the test leads across with the object  to be measured     A caution  1  Refer to   DC Voltage Caution    1  2  4   2      A    means never exceed the maximum input    limit 750V  otherwise internal circuit of the Meter  will be damaged     11       Model UT51 55  OPERATING MANUAL    C 3 Measuring DC Current    di     Connect the black test lead to    COM    jack   When measuring 200mA  UT51 is 2A  or below   connect the red test lead to mA jack  When  measuring 20A  10A  or below  connect the  red test lead to    A  jack    Set the rotary switch to A        Connect the test leads in series with the  object to be measured  the LCD display the  measuring value and polarity of red test lead        Caution    1     2     3     If magnitude of the current is unknown  always  start with the highest range and reduce until 
5.  satisfactory reading is obtained    11    is shown on the LCD  which means the  Meter is overloaded  then set the measuring  range to higher       A    means never exceed the maximum input  limit 200mV   UT51 is 2A   otherwise will cause  the burn of fuse  20A range does not have  fuse protection while UT51 at 10A range has     C 4 Measuring AC Current    1     Connect the black test lead to    COM    jack   When measuring 200mA  UT51 is 2A  or below   connect the red test lead to mA jack  When  measuring 20A  10A   connect the red test  lead to    A    jack     2  Set the rotary switch to A       3  Connect the test leads in series with the object  to be measured    A caution    1     Please refer to DC Current Caution 1  2  3     12          Model UT51 55  OPERATING MANUAL    C 5 Measuring Resistance    1  Connect the black test lead to COM   jack and  red test lead    Q    jack    2  Set rotary switch to 0      3  Connect the test leads across with the object  to be measured     A Caution    1  111   is shown on the LCD  which means the  Meter is overloaded  then set a higher measuring  range  If resistance is above 1MQ  the reading  will only be steady after few seconds which is  normal for measuring high value of resistance    2     1 is displayed when open circuit or no input    3  Make sure all objects  circuit and components  to be measured are without voltage    4  200M9 short circuit has 10 digits which need  to be deducted after the reading when making  measuremen
6.  screw and rubber feet  Replacement of fuses is seldom required   Burning of a fuse always results from the  improper operation       Accessories     A book of users manual     Apair of test lead     A pair of WRN 01B bread temperature probe     for UT53 and UT55 ONLY       A piece of holster  if selected     Using Holster    Three different ways to use holster     1     2     3     Set holster parallel on the table  do not open  the tilt stand  see diagram 1     Set holster in a small angle on the table  tilt  it up by the first part of tilt stand  see diagram 2   Set holster in a large angle on the table  tilt it  up by all two parts of tilt stand  see diagram 3          RA    figure1 figure2 figure3          Model UT51 55  OPERATING MANUAL    G  Using Strap    1  Put the front end of the strap through the round  metal of the Meter  see part 1 of the below  diagram    2  Put the bottom end of the strap through the  front part and tide it up  see part 2 of the below  diagram          END      The manual is subject to changes without separate notice       18       Model UT51 55  OPERATING MANUAL       Model UT51 55  OPERATING MANUAL    Manufacturer  UNI TREND TECHNOLOGY DONG GUAN LIMITED   Address  Dong Fang Da Dao  Bei Shan Dong Fang Industrial  Development District  Hu Men Town  Dong Guan City   Guang Dong Province  China   Headquarters  Uni Trend International Limited   Address  Rm901  9 F  Nanyang Plaza 57 Hung To Road  Kwun Tong Kowloon  Hong Kong   Tel   852  2950 9168 
7. 2A  range    10A  250V fast acting fuse 5x20mm at 10A range     For UT52 53 54 55    315mA  250V fast acting fuse o5x20mm  No fuse  at 20A range     Max current input     For UT51  10A  The measurement time for high  current should be less than 10 seconds maximum   and the interval time between two measurements  should be greater than 15 minutes    For UT52 53 54 55  20A  The measurement time  for high current should be less than 15 seconds  maximum  and the interval time between two  measurements should be greater than 15 minutes   Measuring voltage drop  Full range is 200mV   Display  Average Value  RMS of Sine Wave           Model UT51 55  OPERATING MANUAL    B 5 Resistance                         Range   Resolution Accuracy  UT51  UT 52  UT53 UT54 UT55  2002 0 19   0 8   3   2KQ 19  20KQ 10Q  200KQ   1002   0 8   1   2MQ 1KQ  20MQ 10KQ H1   2  200MQ 100KQ  H5   10   10              voltage at open circuit  790mV  200M2 range   open circuit voltage around 3V     Overload protection  All ranges 250VDC or AC    RMS      Caution  At 200MQ range  as test lead is short  circuit the LCD display 10 digits is normal  deduct  the 10 digits from the measured reading during    measuring     B  6 Capacitance       Range   Resolution    Accuracy       UT51  UT 52 UT53 UT54  UT55                   A Testing signal  around 400Hz 40mVrms    B 7 Frequency                        Accuracy  Range   Resolution 711075310753  UT54 UTES  2kHz 1Hz E   2  5        20kHz 10Hz   1 5  5               
8. ATING MANUAL    5   6     7     2  Turn the Meter power OFF  3   4  Remove rubber feet and screws from the case    Remove the holster from the Meter     bottom  and separate the case bottom from   the case top    Remove the battery from the battery compartment   Replace the battery with a new 9V battery    NEDA 1604 or 6F22 or 006P     Rejoin the case bottom and case top  and install  the screws and rubber feet     lll Replacing the Fuses    Caution    To avoid electrical shock or arc blast  or personal  injury or damage to the Meter  use specified  fuses ONLY in accordance with the following  procedure    To replace the Meter   s fuse     i     Disconnect the connection between the testing  leads and the circuit under test  and remove  the testing leads away from the input terminals  of the Meter       Turn the Meter power OFF     Remove the rubber feet and screws from the    case bottom  and separate the case bottom  from the case top       Remove the fuse by gently prying one end loose     and then take out the fuse from its bracket       Install ONLY replacement fuses with the    identical type and specification as follows and  make sure the fuse is fixed firmly in the bracket   UT51  2A  250V fast acting fuse o5x20mm   below 2A range    10A  250V fast acting fuse o65x20mm  at 10A  range     16          VN  m    F     Model UT51 55  OPERATING MANUAL    UT52 53 54 55  315mA  250V fast acting fuse       5x20mm      Rejoin the case bottom and the case top  and    install the
9. FE   1  Set rotary switch to hFE    2  Identify NPN or PNP  connect objects to the  correspondent transistor jack    3  LCD displays measuring value    4  Measuring condition   Ib   10pA  Vce  2 8V    14       Model UT51 55  OPERATING MANUAL    C 11Auto Power Off function Only for 53 54 55    1  The Meter equips with auto power off function   Itwill be in a sleep condition once it has worked  abt 15 minutes which only consume 7uA current  during that time    2  Press the on off switch two times to power up  again    D  Maintenance   I  General Service   The Meter is a highly precise electrical testing   instrument  do not attempt to change the circuit   of your Meter on your own  Take a note of the   following few points    1  Do not input to DC Voltage above 1000V or  AC above 750V RMS    2  Do not input Voltage when the rotary switch  is in    Current Range        OH and   3  Do not operate the Meter if battery is not inside  the Meter or bottom cabinet is not securely  screwed    4  Disconnect the test leads and power off the    Meter before replacing the Battery and Fuses    ll  Replacing the Battery   A caution   To avoid false readings  which could lead to   possible electric shock or personal injury    replace the battery as soon as the battery  indicator appears    To replace battery    1  Disconnect the connection between the testing  leads and the circuit under test  and remove  the testing leads away from the input terminals  of the Meter        Model UT51 55  OPER
10. load protection  200mV is 250VDC or AC  RMS  All other ranges is 750Vrms or 1000Vp p     B 2 Alternating Current Voltage  AC Voltage     F Accuracy  UT51 UT 52JUT53JUT54  200mV   100uV   1 2   3                   200V_   100mV    750V 1V  1 2   3          AA Input impedance  All ranges are 10Mo   Frequency  40Hz 400Hz     Overload protection  200mV is 250VDC or AC  RMS  All other ranges is 750Vrms or 1000Vp p   Display  Average Value  RMS of Sine Wave         Model UT51 55  OPERATING MANUAL    B 3 Direct Current Current  DC Current                               0 8  1   2       goa    Oma pees  12755   A Overload protection   For UT51   2A  250V fast acting fuse  5x20mm below  2A range   10A  250V fast acting fuse p5x20mm at 10A  range     For UT52 53 54 55     315mA  250V fast acting fuse    5x20mm  No  fuse at 20A range     Max current input     For UT51  10A  The measurement time   for high current should be less than 10  seconds maximum  and the interval time  between two measurements should be greater  than 15 minutes    For UT52 53 54 55  20A  The measurement  time for high current should be less than 15  seconds maximum  and the interval time  between two measurements should be greater  than 15 minutes           Model UT51 55  OPERATING MANUAL    B 4 Alternating Current Current  AC Current        Accuracy       Range  Resolution    200uA 0 1UA  2mA 1UA             200mA   100 pA  AA          A overload protection    For UT51    2A  250V fast acting fuse  65x20mm below 
11. mbols                                EX   LowBattery     Earth Ground  A   Safety Rules    E    Double Insulated  rw   AC    Diode   vee   DC  1     Buzzer   E   Fuse       Dangerous Voltages                Model UT51 55  OPERATING MANUAL    Your Meters Feature    32 ranges    Liquid Crystal Display  digits height is 27mm   Overload display I    Maximum display 1999  3 1 2 digit     Full range overload protection   Auto Power Off  For UT53  UT54 and UT55  ONLY     Temperature    Operating 0  C to 40  C  32   F to 104   F    Storing  10  C to 50  C  14   F to 122   F     Altitude    Operating  2000m   Storing  10000m    Relative Humidity Max  relative humidity  80  for temperature up to 31  C decreasing  linearly to 50  relative humidity at 40 C   Low Battery display  4      Battery Type  9V NEDA 1604 or 6F22 or 006P  Strap for easy carry    Tilt stand design  three observation angles   is in favor of reading display    Dimension  190mm x 88mm x 34mm   Weight  Meter only excluding test leads   about 270g  Meter   holster   tilt stand  about 550g           Model UT51 55  OPERATING MANUAL    B  Specifications   Accuracy is specified for one year after calibration   at operating temperatures 23 C     5 C  with relative  humidity at  lt  75  Accuracy specifications take  the form of      a  readings  digits      B 1Direct Current Voltage  DC Voltage        Accuracy  UT51 UT 52  UT53 UT54  UT55          0 5   1           0 8   2        A Input impedance  All ranges are 10M2    Over
12. t  For example  when measuring  100M9   it displays 101 0  10 digits need to be  deducted     C 6 Measuring Capacitance   Before measuring capacitance  remember it takes  time for zeroing when changing ranges  Floating  reading does not effect accuracy     1  To avoid damage of the Meter or the equipment  under testing  disconnect circuit powers and  discharge capacitors before measuring  capacitance    2  Connect capacitor to the capacitance jack    3  Stabilizing reading takes some time when  measuring high capacitance    4  Unit  1pF 10  uF  1nF   10  F    13       Model UT51 55  OPERATING MANUAL    C 7 Measuring Frequency    1  Connect red test lead to HZ jack and black  test lead to COM jack   2  Set the rotary switch to kHZ       3  Connect the test leads across with the obje  being measured  LCD appears the measurin  value     C 8 Measuring Temperature   Connect one end of the bread temperature probe  to the Meter and the other end to top or inside of  the object being measured  LCD displays the  measuring value with unit as   C      C 9 Measuring Diode and Continuity beeper   1  Connect the black test lead to COM   jack and  red test lead to    V  jack    2  Set the rotary switch to     n   3  Connect the test lead across with the object  being measured  LCD appears the measuring  value    4  Connect the test lead to two ends of the object  being measured  the beeper sounds if the  resistant value between the two ends is less  than 700      C 10 Measuring Transistor h
13. y voltage    Make sure before each measurement the  Meter is set to the suitable range    Before using the Meter  please inspect the  cabinet and test leads for damaged insulation  or exposed metal    Connect the red and black test lead to the  correct measuring input jack properly    Do not input values over the maximum  range of each measurement to avoid damages  of the Meter    Do not turn the rotary function switch  during Voltage and Current measurement   otherwise the Meter could be destroyed           Model UT51 55  OPERATING MANUAL      Make sure to use new fuses with proper  rating in stead of bad fuses       To avoid electric shock or damages  do  not apply more than 1000V between the   COM  terminals and   earth ground      Use caution when working with Voltages  above 60V  DC  or 30Vrms  AC   These  Voltages pose shock hazard      Replace the battery as soon as the battery  indicator    4  appears  With a low battery   the Meter might produce false readings  that can lead to electric shock and personal  injury      Turn off the Meter once finished measuring   fetch out the battery  when the meter will  not be used for long period      Do not operate the Meter under adverse  environmental condition especially humid  area      To avoid damages and dangerous  do not  change the circuit      Periodically wipe the case with a damp  cloth and mild detergent  Do not use  abrasives or solvents      The Meter is suitable for indoor use only      International Electrical Sy
    