Transmitter Calibration
Contents
1. 4 177 mA 100 0 C 1T6 90 3 1 390m Int Ref 24 050 Pigure 7 8 Select Auto Test and watch the DPC take O 50 and 100 measurements of cur rent 4 12 and 20 mA nomi nal while sourcing 100 C 200 C and 300 C 9 Once the DPC has completed this evaluation it will display the post test summary screen with the SOURCE MEASURE and ERROR columns Failures in the error column will be in reverse video Figure 8 Depress Done 4 176 mA 12 277 mA 20 563 A HL 100 0 8 200 09 300 0 C Figure 8 10 Now you are presented with the Tag entry screen Figure 9 AH ire LOWER G ami Pres ENTER to Change Tag TT 644 S N 1235456789 10 GS Figure 9 Press the ENTER key to edit the entries Depress Done when complete until the split screen reappears with the soft key options of As Left Adjust Save and More Choices on the display of the DPC 11 At this point select Adjust and the DPC will source O and measure for 4 mA to 2 Fluke Corporation 750 Series Documenting Process Calibrators 0 25 As long as the error exceeds 25 the error win dow in the measure half of the screen will be in reverse video Adjust the zero on the transmitter until the error is less than 0 25 Next select Go to 100 and adjust the transmitter span in the same way You may also Go to 50 to check linearity at mid
2. calibration management software Process Track All trademarks are the property of their respective holders Additional information is available for download from www fluke com on these products Pressure Calibration Application Note Custom Units Application Note Switch Test Application Note HART Calibration Application Note Process Tools Accessories ISO 9001 8 Fluke Corporation 750 Series Documenting Process Calibrators Fluke Keeping your world up and running Fluke Corporation PO Box 9090 Everett WA 98206 U S A Fluke Europe B V PO Box 1186 5602 BD Eindhoven The Netherlands For more information call In the U S A 800 443 5853 or Fax 425 446 5116 In Europe M East Africa 31 0 40 2675 200 or Fax 31 0 40 2675 222 In Canada 800 36 FLUKE or Fax 905 890 6866 From other countries 1 425 446 5500 or Fax 1 425 446 5116 Web access http www fluke com 2010 2011 Fluke Corporation Specifications subject to change without notice Printed in U S A 8 2011 3792201B A EN N Modification of this document is not permitted without written permission from Fluke Corporation
3. Transmitter Calibration with the Fluke 750 Series Documenting Process Calibrator The precision sourcing and measurement capabilities of the rugged Fluke Documenting Process Calibrator DPC make it an ideal choice for the calibration of process instrumentation The following examples show how to address trans mitter calibrations quickly and effectively using the Fluke 753 or 754 Documenting Process Calibrators Rosemount 444 Figure 1 Connect a type K mini connector into the TC input of the DPC and connect to the thermocouple input of the transmitter Figure 1 It is necessary that you use the correct type of mini connec tor for this connection to the DPC Connect the center two ter minals of the DPC to the con nections on the transmitter where loop power is applied Figure 1 Note If you prefer to use house loop power rather than supplying loop power from the DPC connect these two leads in series with the loop supply to the transmitter Fluke DPC Application Note Once your connections are made starting from the DPC power up state or measure screen select the mA but ton to measure milliamps Figure 2 Source OFF 0 000 mA Scale Hange More Figure 2 4 Next depress the MEAS SOURCE button now you are in the source screen Since you want to source a thermocouple press TC RTD once Press ENTER to select TC Use the down
4. arrow to move the cursor to K and then press ENTER Select Linear T and press ENTER The DPC will respond with ture to source Depress 100 and ENTER Figure 3 Figure 3 100 0 C IT5 90 3 11 Om Scale More Choices 5 Next press the MEAS SOURCE button once and you will be in the split screen mode Figure 4 100 0 C IT5 90 3 119m 5 More Choices Int Hef 24 4 C AS Found Pigure 4 Note Skip to step 7 if you are measuring Current in series using the house loop power source mentioned in step 2 6 Next depress the SETUP button on the DPC The cursor should begin on loop power Depress ENTER down arrow until the DPC cursor is on Enabled and depress ENTER Next depress the Done softkey Figure 5 LOUP 4 103 mA 100 0 C ITS 90 3 148mV _ More Choices Int Ref 23 7 C AS Found Figure 5 Z At this point depress As Found and select Instrument and press ENTER then fill in the template use the default Delay O seconds 4 000 mA 20000 mA 0 25 Ve LOO Valle Tolerance Delay os Loos Value Test Strategy User Custom Abort yame tints Doe Figure 6 Depress DONE You should now see the split screen and are presented with soft key choices of Abort Auto Test and Manual Test Figure 7 MEASUHE
5. bel 12 13 Press the ENTER key to edit the entries press Done when complete until the split screen reappears with the soft key options of As Left Adjust Save and More Choices At this point select Adjust and the DPC will source O instruct you to Go to 3 PSI and measure for 4 mA to 2 As long as the error exceeds 2 the Error win dow in the measure half of the screen will be in reverse video Adjust the zero on the P to I until the error is less than 2 Next select Go to 100 and adjust span in the same way After adjustment is complete select As Left Done and Manual Test and repeat step 9 If the adjustment was successful all the errors in the post run summary should be in normal video Figure 28 3 003 psi g r psi 15 004 pai 12 008 mA 20 000 A Figure 28 14 Select Done Done and Done again and you are done changes to More Choices Fluke Corporation 750 Series Documenting Process Calibrators 7 Note on Uploading Results to Your PC If you are using a 743 744 or 753 754 you may choose an instrumentation management software package from this list Fluke DPC TRACK2 D A m7 Plant Resource Manager AMS from Emerson Process PRM Plant Resource Management formerly Manager from Yokogawa Fisher Rosemount Electric Corporation Weme ROG a echn ologies
6. n Figure 13 4 Next depress the MEAS Figure 13 SOURCE button now you are Oa se save tore in the source screen Figure 14 MEASUHE FLUKE rQoPDOS 3 008 psi Figure 15 Select Function Scale More Choices Figure 14 4 Fluke Corporation 750 Series Documenting Process Calibrators At this point depress As Found select Instrument and press ENTER then fill in the template as shown in figure 16 FLUKE QOOPDS 3 000 psi 15 000 psi 045 Value 100 Value Tolerance Delay Os 40000 ray 100 Value 20 0000 ma Test Strategy User Custom Abort yae units Poe Figure 16 Press DONE You should now see the split screen with soft key choices of Abort Auto Test and Manual Test Select Manual Test and watch the DPC start at O Once the reading has settled depress accept point Per form the 50 and 100 measurements the same way Once the 753 754 has com pleted this evaluation it will display the post test summary screen with the SOURCE MEASURE and ERROR columns Failures appear in reverse video in the Error column Figure 17 Depress Done UHL 4 000 mA 12 000 mA 20 01000 mA 0 47 pai 15 490 psi Figure 17 9 Now you see the Tag entry screen Figure 18 Press ENTER to Change Tag PTLOO S N 123456789 10 Figure 18 10 11 Press the ENTER key to edi
7. nable the loop 24 volt power from the setup screen source the pertinent RTD for the specific transmit ter being evaluated and set the measure mode to mA Fluke DPC At this point enter the split screen mode of the DPC and perform the As Found Adjust and As Left tests as you did in the previous thermocouple transmitter example Fluke Corporation 750 Series Documenting Process Calibrators 3 Pressure Module Connect the pressure module to the DPC Connect the out Figure 12 put of the I to P to the input of the pressure module If you are using a differential pres sure module make sure for this example that it is a 15 PSI module and connect to since you want to source a current depress mA Depress ENTER which chooses Source poe mA when given the choice of the input labeled H Pressure Button Source mA or Simulate Trans Figure 12 mitter When the DPC responds 2 Connect from the center two with asking for a current to banana type connections of EE e p e source depress 4 and ENTER the DPC to the input of the I o oeg to P where control current is x 5 Next depress the MEAS SOURCE button once and you should be in the split screen mode Figure 15 supplied Figure 12 3 Once your connections are made assuming you are in the power up state or mea sure screen select the Pres sure button directly above the Range butto
8. orporation 750 Series Documenting Process Calibrators 6 Since we wish to source a pressure using a hand pump depress the pressure button directly above the HART or range button Figure 22 Pressure Button Figure 23 Figure 22 8 At this point depress As Found select Instrument and press ENTER then fill in the template per Figure 24 4 0000 mA 100 Value 20 000 m Tolerance 2 00 3h Delay Os FLUKE 7 O0PDS 4 000 psi 100 Value 5 00000 psi Test Strategy User Custom bot yae units Poe Figure 24 Depress DONE You should now see the split screen and are presented with soft key choices of Abort Auto Test and Manual Test 9 Select Manual Test and watch the DPC for the prompt Go to 3 PSI Figure 25 Error 0 12 4 015 mA FLUKE 700PDS 2 997 psi Go To 3 psi Figure 25 At that point using the hand pump pump until the DPC display is reasonably close to 3 PSI and depress Accept Point Now the DPC should display Go to 9 PSI Again using the hand pump pump up reasonably close to 9 PSI and depress Accept Point When the DPC responds with Go to 15 PSI pump to that value and Accept Point It is important to understand a key feature of the DPC at this point When the error is being calculated on the measured mA by the DPC it will make the calculation based on the deviation of
9. r the results and press Tag for its associated label When you are finished reviewing results press Done until right most soft key label changes to More Choices Fluke Corporation 750 Series Documenting Process Calibrators 5 Pressure Module Figure 20 1 Connect the pressure module 5 Next depress the MEAS 7 Next depress the MEAS to the DPC Connect the input of the P to I and the output of a pressure hand pump to the input of the pressure module using a T connector Figure 20 If you are using a differential pressure module ensure for this example it is a 15 PSI module and connect to input labeled H SOURCE button now you are in the source screen Figure 21 0 00000 V 0 00 O75 1 50 2 2 3 00 Y juanma SOURCE button once and you should be in the split screen mode Figure 23 MEASUHE 0 000 mA Toe F OORDE 0 000 psi Scale Alan Hore Auto Choices Figure 21 AS More Found Choices 2 Connect from the center 2 banana type connections of the DPC to the input of the P to I where loop power is supplied Figure 20 Next depress the SETUP but ton on the DPC The cursor should begin on loop power Depress ENTER down arrow to Enabled and depress ENTER Next depress the Done softkey Once your connections are made assuming you are in the power up state or measure screen select the mA button Fluke C
10. span 12 After adjustment is complete select As Left Done and Autotest Watch the 0 50 100 evaluations If the adjustment was successful all the errors in the post run summary should be in normal video or less than 25 Figure 10 ARL T rt ie E uo I 100 0 8 40000 A 200 0 C 300 0 C 11 996 mA 19 9097 mA Figure 10 13 Select Done Done and Done again and you are done To review your results in memory select the More Choices soft key the Review Memory soft key and move your cursor to the applicable result and press enter Select As Found or As Left then ENTER and review the setup for the test select Next Page for the results and press Tag for its associated label Select Done and Done Move your cursor to the applicable As Left result and press ENTER and review the test setup select Next Page for the results and press Tag for its associated label When you are finished reviewing results press Done until right most soft key label changes to More Choices Make the RTD source connections 1 S2 S3 and S4 omit S3 and S4 for 2 wire RTDs and S4 for 3 wire RTDs S4 is only needed for 4 wire RTD sourcing Next connect the loop power and current measurement connections C1 and C2 Figure 11 Once the connections are made as per the diagram in Figure 11 e
11. t the entries Depress Done when complete until the split screen reappears with the soft key options of As Left Adjust Save and More Choices At this point select Adjust and the DPC will source O 4 mA and measure for 3 PSI to 2 As long as the error exceeds 2 the Error win dow in the measure half of the screen will be in reverse video Adjust the zero on the I to P until the error is less than 2 Next select Go to 100 and adjust span in the same way Select Go to 50 and evaluate linearity After adjustments are com plete select As Left Done and Manual Test and repeat step 8 If the adjustment was successful all the errors in the post run summary should be in normal video Figure 19 3 001 psi 0 001 psi 15 000 psi T 4 000 mA 12 000 mA 20 01000 mA Figure 19 12 Select Done Done and Done again and you are done To review your results in memory select the More Choices soft key then the Review Memory soft key Move your cursor to the applicable result and press ENTER Select As Found or As Left then ENTER Press Next Page for the results and press Tag for its associated label Select Done and Done and move your cursor to the applicable As Left result and press enter Review the setup for the test select Next Page fo
12. the amount of pressure being sourced What this means is 10 if you are sourcing 3 120 PSI 1 of span above nominal the DPC will look for a mA reading of 4 160 mA also 1 of span above nominal In order to make a proper evalu ation of the device you need only to get reasonably close to the nominal pressure val ues with the hand pump The remaining source pressure error is compensated for by the DPC during mA measure calculations Once the DPC has completed this evaluation it will display the post test summary screen with the SOURCE MEASURE and ERROR columns Figure 26 Depress Done AALE mA 12 4335 mA 20 601 m T 3 000 psi 8 999 psi 15 028 psi Figure 26 11 Now you are presented with the Tag entry screen Figure 21 Press ENTER to Change Tag PTLOl S N 123456789 10 MEN Figure 27 To review your results in memory select More Choices soft key Review Memory soft key move your cursor to the applicable and press ENTER and review the setup for the test select Next Page for the results depress Tag for its associated label Select Done and Done next move your cursor to the applicable As Left result and depress ENTER and review the setup for the test select Next Page for the results depress Tag for its associated label When you are finished review ing results press Done until right most soft key la