Home
        California State University, Fresno Department of Electrical and
         Contents
1.  be used by an Electrical or Computer Engineer  In the  process of learning about the functionality of an oscilloscope  you will also learn how to measure the voltage and  period frequency of various signals     The Tektronix TDS 210 Oscilloscope    Oscilloscopes are used by engineers and technicians to display voltage waveforms and signals as a function of time   Older displays consist of a glass cathode ray tube  CRT   an electron gun  x  and y deflection plates  and a phosphor  screen that emits light when struck with a focused electron beam  The spot of light can be moved by horizontally and  vertically deflecting the beam of electrons     if moved rapidly enough  the moving spot will have the appearance of  displaying an image on the screen  Horizontal deflection typically occurs from left to right at a constant speed called  the sweep rate  while vertical deflection occurs in response to the strength of the applied voltage  resulting in a plot  of the waveform  The horizontal sweep is generated by applying a triggered  sawtooth voltage waveform  created  internally  across the x deflection plates  Although newer LCD screens on digital oscilloscopes do not use a CRT   understanding the principles of older oscilloscopes is benefitial toward learning how the current generation of digital  scopes operate     Figure 1 illustrates the    sawtooth    voltage waveform applied to the horizontal deflection points  At point A of the  cycle  the beam starts to sweep the screen  uni
2.  the settings discussed so far  except   perhaps when using XY mode  to center an oscilloscope trace on the display  and to adjust the INTENSITY  if nec   essary   The position knobs  temporarily grounding the inputs  and locating the traces for CH1 and CH2 allow the  reference level  0 V  of the display to be arbitrarily set  so become familiar with them  Once you are finished  set the  SEC DIV knob to 500 us  and turn OFF CH 2     Sensitivity Controls   Vertical deflection of the electron beam is produced by applying an external  direct  or alternating current  DC or AC   signal to the input s of the scope  The degree of vertical deflection is affected by both the magnitude of the input signal  as well as the sensitivity of the scope  which can be adjusted by using the VOLTS DIV knob  Pressing the CH 1  MENU button  you can choose what kind of Coupling you wish to allow into the oscilloscope  Ground  DC  or AC   When the Coupling button is pressed so that Ground is chosen  the input signal grounded   preventing any vertical  deflection of the beam  This position is typically useful for establishing a reference position somewhere on the display  and or finding the beam  While the input is grounded  wherever the beam is moved to on the screen vertically  which  can be done arbitrarily using the VERTICAL POSITION knob  will determine the vertical location that corresponds  to 0 V  or ground  You should always know where ground is referenced to on the display for both CH1 and C
3. 2 as necessary  but this time center the trace  toward the top of the screen  Once CH 2 has been adjusted  press the MATH MENU button and select the     Oper   ation  Select CH1 CH2 and you will see a third trace displayed on the screen  which is the algebraic subtraction of  the two signals  This corresponds  physically  to the vector subtraction  or interference  of two parallel signals  You  can also select CH2 CH1 if you wish  or choose to add the two signals together by selecting   under Operation  By  adjusting the VERTICAL POSITION of either beam  you will see the vector addition or subtraction move as well   Once finished  press the MATH MENU soft key button once more to turn off the Math Operation     Next  press the CH 1 MENU button and be sure that the Coupling is set to Ground  Repeat for CH 2  Then adjust both  VERTICAL POSITION knobs so that the beams are exactly centered at 0 00 div  0 00 V   Then press the DISPLAY  button and change the Format option from YT to XY  which will turn your line into a dot  you will not be able to see  it unless you adjust one of the VERTICAL POSITION knobs  In this mode  the oscilloscope is capable of plotting  two dimensional graphs representing a function y   f x   The plot here is achieved by eliminating the variable time t  between the two functions f x  and y t   where x t  is the CH 1 input signal and y r  is the CH 2 input signal  The  physical meaning of this operation is the vector summation  or interference  of the two p
4. California State University  Fresno  Department of Electrical and Computer Engineering    ECE 90L Principles of Electronic Circuits Laboratory    Experiment No  1  Introduction to the Digital Storage Oscilloscope    Introduction  Welcome to the Principles of Electronic Circuits Laboratory  You will find the following equipment on your lab bench       Tektronix TDS 210 or 1002 Two Channel Digital Real Time Oscilloscope    GW Instek GDM 8245 Digital Multi Meter  DMM      GW Instek SFG 1013 DDS Function Generator     Mastech DC Power Supply HY3005D 3 Function Generator     Dual   12 V power supply with breadboard       Within the cabinets are  BNC cables  banana plug adapters  variable resistors  capacitors  and inductors  additional  power supplies  and other equipment that will be made available to you over the course of the semester as you perform  your experiments  Identify all of the equipment  listed above  on the lab bench before proceeding     Objective    The first experiment is intended to be an introduction to some of the hardware available to you in the laboratory     specifically the Tektronix TDS 210 or 1002 oscilloscope  the GW Instek GDM 8245 Digital Multi Meter  DMM   and  the GW Instek SFG 1013 DDS Function Generator  An oscilloscope  scope  is an electronic instrument that displays  a voltage input as a function of time  and is typically used to display and measure repetitive waveforms or transient  signals  It is one of the important diagnostic tools that can
5. H2 while  examining signals on the oscilloscope     Unlike Ground  when the Coupling button is in the AC position  only the DC  constant  component of the signal will  be filtered from the input  which means that the AC  changing  component of the signal will still be displayed  And  finally  it is only when the button is in the DC position that the oscilloscope will display the actual waveform input into  the device   which includes both both the DC and AC components of the signal  The advantage of using AC coupling  over DC coupling occurs when you are trying to measure small variations of a signal that contain a large DC bias  As  an example  consider the relatively small    ripple     5 mV  on the output of a typical DC power supply  say 5 V   With  DC coupling  increasing the input sensitivity in an attempt to see the small variation will send the trace off the top of  the screen  However  with AC coupling  the 5 V level will be suppressed and the sensitivity can be greatly increased  while retaining the trace on the center of the display     Use the 5 V DC power supply to confirm this behavior  turn ON the power to the Mastech DC Power Supply  HY3005D 3   Use a BNC banana adapter in conjunction with a BNC cable to hook up the  5 V 3A voltage supply  on the Mastech Power Supply to the oscilloscope  Make sure the tab marked GND on the banana plug adapter is  plugged into the ground plug of the voltage supply  However  before you attach the output of the power supply to th
6. e  scope  first verify that you have a 5 V signal by connecting the output to the leads of the GW Instek Digital Multimeter   DMM   Note that you must plug the leads for the DMM into the appropriate plugs  and have selected the DC voltage  button  to measure the voltage of the power supply  Once this has been done  using DC coupling on the oscilloscope   re verify that the output voltage of the power supply is correct  If you then try to examine the ripple more closely by  increasing the sensitivity  turning the VOLTS DIV knob to the right   the trace will soon driven off the edge of the  display  With AC coupling  the DC level is suppressed and the ripple can be observed in detail  measure its amplitude  and compare it to the DC level  What happens if you gently wiggle the cable  Why     Trigger Controls   The triggering circuitry is the means by which the oscilloscope is able to provide repeated    snapshots    of a repetitive  signal  By controlling the threshold voltage of triggering and slope polarity  the user is provided with great flexibility  in the appearance of the display and the utility of the scope for capturing elusive signals  Some further details of scope  operation are required to understand the concept of triggering     As mentioned previously  the horizontal sweep of the electron beam is controlled by the timebase circuitry  the  SEC DIV knob   The beam sweep is driven by the application of a ramp voltage to the horizontal deflection plates   which serves 
7. e from 6 Hz all the way to 6 MHz by powers of 10 and  adjust the time base setting as necessary to continue to provide a    clean    display  Comment on your results     Returning to a timebase around 500 us div and a frequency of 600 Hz  confirm with the oscilloscope that the period of  the 5 V AC signal is 600 Hz  Adjust the trigger level if necessary to achieve a stable display  Measure the horizontal  distance between the two closest points where the waveform crosses the baseline  0 V level  with the same slope    Note that there are two zero crossings per cycle  each with different slope   Multiply this number of divisions by the  timebase setting  time div   the resulting time T is known as the period     How short a time can be read by this oscilloscope  This can be estimated by the smallest timebase setting  In practice   this setting may result in a trace too weak to be seen   thus the so called writing speed of the oscilloscope may also  limit the shortest time that can be accurately determined  as will the bandwidth of the input amplifiers     Dual Trace Operation   The availability of two independent input channels greatly facilitates comparison of two signals  To do this  turn on  both Channels  While displaying the 5 V  600 Hz sinusoidal signal on CH 1  using it as a trigger source  observe  the signal from a loose BNC cable on CH 2  You will probably observe    spikes    and other types of electrical noise   depending on the sensitivity level  You should attemp
8. erpendicular signals  When  both signals are periodic  the resulting graph is known as a    Lissajoux    figure  Press the Format button once again to  return to the previous mode of operation YT        Figure 2  The front panel of a Tektronix Oscilloscope Model 210  In this figure  the input signal is applied to Channel  1  whose sensitivity is set to 10 0 V div  The time base is set to 250 us div  meaning that the beam will travel each  division in 250 us  or the entire screen in 2 5 ms  The trace on the screen shows a continuous change of the input  voltage from    2 0 divx 10 V div     20 0 V to  2 0 divx 10 V div  20 V with a period over 4 div x250 us div    1 0 ms time interval     Time Base Controls   The time base  SEC DIV  is an important feature of these oscilloscopes  as it allows for the measurement of different  frequencies and time scales within an electrical circuit  Adjusting the time base adjusts how slowly quickly the signal  sweeps across the screen  Since the sweep rate is constant  the distance  which is measured in screen divisions   and  more specifically  centimeters  traveled by the spot is a measure of the corresponding travel time  Experiment a bit  with the SEC DIV knob  noting the different rates which the display can be swept  You should notice that for very  slow sweep rates  you can see the voltage versus time signal slowly sweep from left to right across the screen     Whenever you turn on the scope  you should be able to quickly adjust all of
9. formly  from left to right  At point B  the beam passes through the center  of the screen and  at C  moves to the end of its trace on the right  The beam then quickly    jumps    back to the starting  point at A    within a few nanoseconds  where a new cycle begins  Both the sweep rate and the vertical sensitivity can  be adjusted by using controls on the front panel of the oscilloscope  an example of which is shown in Figure 2        Figure 1  The    sawtooth    shape of the sweeping voltage V t   At point A of the cycle  the beam starts to sweep the  screen  uniformly  from left to right  At point B  the beam passes through the center of the screen and  at C  is at the  right end of the trace  The beam then quickly    jumps    back to the starting point at A    within a few nanoseconds  where  a new cycle begins     Display Controls   To begin  turn on the oscilloscope by pressing the POWER button on the top of the device  Because the Tektronix  210 1002 is a dual input oscilloscope  you can choose to display a voltage waveform from either 1 of 2 inputs  or both  simultaneously   which are marked CH1 and CH2  respectively  on the front panel  Repeatedly pressing the MENU  button for CH 1 will turn the scope trace for Channel 1 ON and OFF  Likewise  pressing the MENU button for CH  2 will turn the scope trace for Channel 2 ON and OFF  You have the option of displaying one of the scope traces at  a time  or both simultaneously if multiple signals need to be measured  Initiall
10. function of the second  channel  based upon what you see on the screen  Next  switch the inputs  so that the sinusoid is input into CH 2 and  the DC voltage into CH 1  What changed  and why     Additional Display Options   Next press the MEASURE button and note that there are various measurements about the waveforms that can be  displayed for both Channels 1 and 2  Experiment with the various measurement options and explain as many as you  can in your laboratory notebook  Next press the CURSOR button and choose Time under the Type option  You can  then move the two VERTICAL POSITION knobs to give you a measurement of the time difference between the two  vertical lines  This can be used to help you make measurements in time  Similarly  if you choose Voltage under the  Type option  you can make similar measurements with respect to voltage  Once finished  turn the CURSOR Type to  Off     Finally  press the ACQUIRE button and note that you can continually sample the voltage versus time plot by selecting  Sample  choose to display the peak voltages by selecting Peak detect  or display an average of a number of voltages  by selecting Average and then adjusting the Averages button  Explain the difference between the waveforms for each  case  and describe what the oscilloscope is doing     Conclusion    Provide a brief conclusion based upon what you have learned today and both partners will turn in the lab book at the  beginning of the next laboratory session  No group report is nece
11. moving the trigger toward the right   the starting point  of the trace changes  in this case toward the left   Likewise  adjusting the threshold for the trigger toward a negative  voltage  moving the trigger increasingly toward the left   the starting point of the trace moves in the opposite direction   Refer to Figure 3 for a pictorial representation of the various triggering combinations  Switch the trigger SLOPE and  observe the changes to the displayed waveform  The SLOPE selector chooses whether the trigger event will occur  when moving through the threshold from a high voltage to a low voltage  falling edge negative slope  or from a low  to high voltage  rising edge positive slope   Refer again to Figure 3     Finally  setting the trigger level too low or too high  outside the bounds of the signal amplitude  results in loss of  triggering and an unstable    running    waveform  You can observe this if you decrease the input amplitude of the signal        TRIGGER         TRIGGER         TRIGGER         TRIGGER      LEVEL SLOPE  LEVEL SLOPE  LEVEL SLOPE  LEVEL SLOPE   x  vy     CF   1G  z ey   KI         2H APN ASS    threshold    threshold    threshold    threshold     slope    slope    slope    slope       Figure 3  Effects of slope and the threshold on triggering     by using the AMPLITUDE knob on the function generator  Decreasing the TRIGGER LEVEL down toward the  voltage level of waveform will restore a stable trace  Notice also what happens if you switch the trigge
12. r Source to  CH2  Explain your results for both cases     Next  examine of the effects of the external trigger selection EXT  It is with this setting that an additional  better   conditioned signal can be used to trigger the scope while observing a relatively weak signal   one that is too weak  when using internal triggering  as discussed above  For example  you may again observe the ripple on the DC power  supply or the stray signal picked up by a loose lead  using the 600 Hz  AC signal to trigger the scope externally  Ex   ternal triggering is accomplished by connecting the signal to be used for triggering to the external trigger input EXT  TRIG of the scope and selecting Ext on the trigger Source button     When you are finished  display a 600 Hz  5 V peak to peak sinewave using CH 1 on the scope  Make whatever ad   justments are necessary to display a stable waveform  Then start from 500 us div and increase the timebase to larger  settings  slower sweeps   which is simply a longer    snapshot    of the signal being displayed  but with less detail  Re   member that the 600 Hz signal has not changed  the beam sweeps more slowly  so only more transitions are displayed   Moving to much smaller timebase settings  faster sweeps   more detail can be seen at the expense of the    big picture      However  notice the decrease in intensity of the trace as the electron beam moves more swiftly over the screen at faster  time base settings  Finally  adjust the frequency of the sinewav
13. ssary for this laboratory     Acknowledgments    In addition to my own work  several people have allowed me to adapt portions of their lab manuals into this laboratory  procedure  I would like to gratefully acknowledge the following people for giving me permission to use their materials     Dr  Larry Owens  Professor  Electrical and Computer Engineering Department  California State University  Fresno   Dr  George Watson  Associate Dean  College of Arts  amp  Sciences  University of Delaware   Dr  Jeffrey Urbach  Associate Professor and Chair  Department of Physics  Georgetown University     
14. t to explain why this noise is    locked    to the signal on CH 1  For  a more dramatic effect  touch your finger to the CH 2 input conductor from the BNC cable  The noise probably has  much higher amplitude and is still locked to the calibration signal  Finally  while holding the bare lead in one hand   touch the metal frame of the table and observe the noise signal on CH 2  Attempt to explain your observations     Next  apply the 0   20 V input from the Mastech power supply to CH 2 on the oscilloscope using the red and black  banana plugs  making sure that the plug with the GND tab is attached to the common  black output from the power  supply  While displaying both Channels on on the oscilloscope  adjust the Coupling control to DC for both channels   and observe your results when using the MATH MENU options  As you do so  adjust the amplitude of the DC signal  using the   VOLTS knob on the dual power supply  What happens  and why  Finally  try various combinations of  AC  GND  and DC  and observe your results  When the knobs are in the AC position and you vary the amplitude of  the DC voltage from the dual power supply  what happens  Why     X Y Mode   Once you are comfortable with the dual trace operation of the oscilloscope  set the amplitude of the DC signal into CH  2 to 5 V and select XY operation of the scope using the Format option after hitting the DISPLAY button  Now  the  voltages into CH 1 and CH 2 are functions of each other  Can you justify which channel is a 
15. to move the beam across the face a distance proportional to the elapsed time from a trigger event  The  occurrence of the trigger event marks the start of the linear ramp  By slowing the timebase to about 1 s div  the actual  deflection of the beam can be slowed to the point where the resulting luminous waveform can be observed crossing  the display from left to right     To examine the effects of the trigger settings on the triggering of the display  use the output available on the SFG 1013  function generator to generate a 600 Hz triangle wave  use the WAVE button  as the input signal for CH 1  Hook up  the input to CH 1 on the scope and increase the sensitivity to 5 V div  Adjust the amplitude on the function generator  so that the amplitude of your triangle wave is exactly 5 V  peak to peak   and select a 500 us div sweep speed on the  oscilloscope  With the trigger Source set for CH1  note there are 5 options   CH1 for Channel 1  CH2 for Channel  2  and Ext for an external trigger  Ext 5 which divides the external trigger by 5  and AC Line which expects an AC  signal   the TRIGGER LEVEL knob rotated to be just above 0 V  and the Slope option set to detect on the rising  edge of a pulse  a stable trace showing the triangle wave should appear  Adjust the horizontal position of the trace until  its starting position is seen on the display  Next  adjust the TRIGGER LEVEL by adjusting the knob to the left and  right  Notice that as the threshold for the trigger is increased  
16. y  we are only interested in looking  at the voltage on Channel 1  so press the appropriate MENU buttons to turn CH 1 ON and CH 2 OFF  Under the  TRIGGER MENU  make sure that the Auto option is selected  you can toggle between options under Mode using  the 2nd soft key button from the bottom  just to the right of the LCD   Also  make sure that the Edge option is selected   you are using a Rising Slope  your Source input is CH1  and you are using DC Coupling  Next turn the SEC DIV  knob to 250 us  and make sure the VOLTS DIV knob for CH 1 is turned all the way to the left     After the scope warms up  you will see a horizontal line on the screen  If you cannot find the beam  use the VER   TICAL POSITION knob to center the line at 0 00 divs  0 00 V   Because we are only looking at CH1  only the  left hand knob for the VERTICAL POSITION will be useful  Rotating the knobs applies a constant  negative or  positive  voltage across the horizontal vertical deflection plates  causing the beam to move horizontally vertically on  the screen   depending on the applied voltage  how far you turn the knob   Experiment with the knobs for a bit to move  the trace to arbitrary locations on the display  Can the display be completely moved off the screen  What happened   and why  Explain this in your lab notebook     Once the trace has been centered  use the DISPLAY option to adjust the brightness of the trace using the Contrast  Increase and Contrast Decrease buttons  Repeat the process for CH 
    
Download Pdf Manuals
 
 
    
Related Search
    
Related Contents
MUXD 19" (B6L 19'')  Technical Pdf    DeLOCK GX53 LED    Copyright © All rights reserved. 
   Failed to retrieve file