Home
        CON 400 and CON 410 - Oakton Instruments
         Contents
1.    23        Instruction Manual CON 400 410  7 1 Advanced SETUP Mode Overview    Press the SETUP key to enter Set up mode    Press the MI A or MR    keys to scroll through sub groups    Press ENTER key to enter a particular parameter    See Addendum 4 on page 48 for a table of meter factory default settings     P1 0  Memory clear  P1 0 Clear all stored readings    P2 0  Viewing previous calibration data  P2 1 First range calibration point   P2 2 Second range calibration point  P2 3 Third range calibration point   P2 4 Fourth range calibration point  P2 5 Fifth range calibration point       SETUP P3 0  Viewing electrode data  P3 1 Effective cell constant for first range  E E E P3 2 Effective cell constant for second range  P3 3 Effective cell constant for third range    D 3 g P3 4 Effective cell constant for fourth range  g P3 5 Effective cell constant for fifth range    Kess      SETUP P4 0  Unit configuration  P4 1 READY indicator and auto endpoint function      TUE select on or off  SS P4 2 Select   C or F  P4 3 Select Automatic or Manual Temperature  H UU Compensation  P4 4 Setting TDS conversion factor  CON 410  meter only      25     Instruction Manual       SETUP    LEL    Psg    SETUP    LI  P10    SETUP MEAS    r St    P80       CON 400 410    P5 0  Temperature  P5 1 Adjusting temperature coefficient  P5 2 Adjusting normalization temperature    P6 0  Selecting cell constant  P6 1 Selecting cell constant   K   1 0  10  or 0 1    P7 0  Setting Clock  e Setting Year    e
2.    5  Calculate the TDS factor of your solution  See A    Addendum 2 on page 46 for information on how to  calculate the TDS factor    6  Press the MI A or MR V keys to select your A J 5 ppt  calculated TDS conversion factor  i SR   7  Press the ENTER key to confirm selection and to E dS  return to the subgroup menu  Press the CAL MEAS             key to return to measurement mode     Figure 29   Change of TDS factor    2332    Instruction Manual CON 400 410    7 6 P5 0 Temperature  7 6 1 P5 1 Adjusting the temperature coefficient    The temperature coefficient is the amount of change in conductivity per degree of temperature   it is expressed in percent per   C  Entering the exact temperature coefficient of your solution  lets you accurately compensate temperature for almost any solution  You can adjust 0 0 to  10 0   per   C  Meter default is 2 1  per   C  Adjustment to 0 0  will not correct for  temperature effects so the displayed value will be the actual value at the displayed  temperature     From measurement mode   1  Press SETUP key to enter Set Up mode    2  Press the MI A or MR V keys to scroll through  subgroups until you view parameter P5 0    3  Press the ENTER key to select parameter 5 1  The    display shows  t CO    on the upper display  I   4  Press the ENTER key again  The upper display  shows the temperature coefficient and the lower  display shows    t CO       5  Press the MI A or MR V keys to select the  temperature coefficient of your solution    6  Pre
3.   This will retain the  meter   s old calibration data in the measuring  range of the calibration     You can offset the conductivity reading up to   40  from factory default value  If your  measured value differs by more than  40    clean or replace probe if needed    The minimum offset reading allowable is  limited to 10  of full scale reading of the    CON 400 410        Immerse   probe beyond  upper steel  band    Figure 7   Proper Immersion of  the conductivity probe    range you are working in  The maximum offset reading allowed is the range full scale reading    Eutech Instruments  Oakton Instruments offers a wide selection of high quality calibration    standards  See page 44 for more information      11     Instruction Manual CON 400 410    4 5 TDS Calibration  4 5 1 Calibrating for TDS directly    MEAS  For CON 410 meter only d 5 5  The factory default setting for the TDS conversion factor is ppm  0 50  If your solution has a different TDS factor  you can 258      ATC    improve calibration accuracy by setting the TDS factor prior to  calibration  See page 33 for directions     1  If necessary  press the MODE key to select TDS mode    2  Rinse the probe thoroughly with DI water or a rinse  solution  then rinse with a small amount of calibration  standard    3  With the yellow probe guard attached  dip the probe into  the calibration standard  Immerse the probe tip beyond  the upper steel band  Stir the probe gently to create a  homogeneous sample  Allow time for the
4.   minute    digits will flash    14  Press the MI A or MR V keys to toggle to the correct  minutes    15  Press the ENTER key to confirm the minutes and  move to    second    digits will flash  See Figure 34    16  Press the MI A or MR V keys to toggle to the correct  seconds    17  Press the ENTER key to confirm the seconds and  return to    century    selection    18  Press the CAL MEAS key to return to the subgroup  menu  Press the CAL MEAS key again to return to  measurement mode        NOTES  Press the CAL MEAS key at any point while setting the  time to return to the subgroup menu        Figure 34   Selecting hour   minutes and seconds      38      Instruction Manual CON 400 410    7 9 P8 0  Resetting to factory default settings    Program 8 0 lets you reset all parameters to factory default settings  This clears all calibration  data  memory  and any other setup functions you might have changed  It does not clear clock  settings     From measurement mode SETUP MEAS  1  Press SETUP key to enter Set Up mode  St  2  Press the MI A or MR V keys to scroll through r    subgroups until you view parameter P8 0    3  Press the ENTER key to enter parameter P8 0    4  Press the MI A or MR V keys to toggle between NO  and YES  NO retains current settings  YES resets to  factory default settings     5  Press the ENTER key to confirm selection and to SETUD MEAS  return to the measurement mode  Otherwise press P St  CAL MEAS key to return to measurement mode  without resetting to fac
5.  1 00  available in TDS  400 and CON 410 meters only    P5 1 Temperature coefficient SER Adjustable from 0 0 to 10    P5 2 Normalisation temperature   25 C Adjustable from 15 to 30  C   P6 1 Cell constant 1 0 Select from k 1 0  0 1 or 10   P8 0 Factory default No Retains your current settings   NOTES                 The P7 0 clock function retains the year and time that you set  even after reverting to factory  default settings       48      Instruction Manual CON 400 410          17 WARRANTY       This meter is supplied with a three  year warranty  six month warranty for probe against  significant deviations in material and workmanship    If repair or adjustment is necessary and has not been the result of abuse or misuse within the  designated period  please return     freight pre paid     and correction will be made without  charge  Eutech Instruments  Oakton Instruments will determine if the product problem is due  to deviations or customer misuse     Out of warranty products will be repaired on a charged basis     Exclusions  The warranty on your instrument shall not apply to defects resulting from     Improper or inadequate maintenance by customer    Unauthorised modification or misuse    Operation outside of the environment specifications of the products     49     Instruction Manual CON 400 410          18 RETURN OF ITEMS       Authorisation must be obtained from our Customer Service Department or authorised  distributor before returning items for any reason  A    Retu
6.  1413 uS in the 0 to 1999 uS  range and you recalibrate at 1500 uS  also in the O to 1999 uS range   the meter will replace  the old calibration data  1413 uS  in that range  The meter will retain all calibration data in  other ranges     To completely recalibrate your meter  or when you use a replacement probe  it is best to clear  all calibration data in memory  To erase all the old conductivity and TDS calibration data  completely from memory  see SETUP section Program 8 0 on page 39     For information on how to calibrate your meter   e See section 4 3 for Temperature Calibration   e See section 4 4 for Conductivity  e See section 4 5 for TDS Calibration  CON 410 meter only    e See Addendum 1 for more calibration tips     Instruction Manual CON 400 410    4 2 Preparing the Meter for Calibration    Before starting calibration  make sure you are in the correct measurement mode  When you  switch on the meter  the meter begins with the units you shut it off in     For best results  select a standard value close to the sample value you are measuring   Alternatively use a calibration solution value that is approximately 2 3 the full scale value of  the measurement range you plan to use  For example  in the 0 to 1999 uS conductivity range   use a 1413 uS solution for calibration     Do not reuse calibration solutions after calibration  Contaminants in the solution can affect the  calibration  and eventually the accuracy of the measurements  Use fresh calibration solution  each 
7.  Setting Date  month day   e Setting Time  hour   minute   second     P8 0  Reset to factory defaults  P8 0 Reset meter to factory defaults     26      Instruction Manual CON 400 410    7 2    P 1 0  Memory Clear  CLr     Use this parameter to clear all stored memory values when you need to store a new series of  values  This lets you avoid confusing the old values with the new ones   NO is the default setting     NOTE  Selecting YES will wipe out all memory     From measurement mode   1   2     Press the SETUP key to enter Set Up mode    Press the MI A or MR V keys to scroll through  subgroups until you view the parameter P1 0    Press the ENTER key to enter parameter P1 0   Press the MI A or MR V keys to toggle between  NO and YES  NO retains current memory  YES  clears all memory  See Figure 23    Press the ENTER key to confirm selection and  return to the subgroup menu  Press CAL MEAS key  to return to measurement mode        Figure 23   P1 0  Memory Clear     27     Instruction Manual    7 3 P2 0  Viewing calibration data    This mode lets you recall previous calibration data  which  helps you know when to re calibrate your meter  This is  a    view only    mode  See Figure 24     From measurement mode    1  Press the SETUP key to enter Set up mode    2  Press the MI A or MR V keys to scroll through  subgroups until you view parameter P2 0    3  Press ENTER key repeatedly to view previous  calibration data  The meter will first display the  calibration point  and then di
8.  You  can use this feature to alert you to the need for a new  probe prior to total failure      29     CON 400 410          Figure 25   View probe data for  each measurement range    Instruction Manual CON 400 410    7 5 P4 0  Unit configuration    7 5 1 P4 1  READY indicator and auto  endpoint function   Program P4 1 lets you select    READY indicator on    to   indicate when your measurement is stable  or select    READY indicator off    for faster meter response     Program P4 1 also lets you switch the Auto endpoint  function on or off  Select auto endpoint on to  hold    your  measurement when it is stable for more than 5 seconds   The display automatically freezes  and HOLD indicator  appears on the left side of the display  Press the HOLD  key to release the display and access other functions   Select auto endpoint off to deactivate this feature     SETUP  READY    From measurement mode   1  Press SETUP key to enter Set Up mode    2  Press the MI A or MR V keys to scroll through  subgroups until you view parameter P4 0  READY   3  Press ENTER key to select parameter 4 1    4  Press the MI A or MR V keys to select the  configuration you require    e OFF switches the READY indicator off  ON  switches the READY indicator on  ON and HOLD  together switches the auto endpoint feature on    5  Press ENTER key to confirm selection and to  proceed to step 4 of P4 2  You can also press the READY   CAL MEAS key to return back to measurement Ser   mode        NOTE  Meter default 
9.  an auto endpoint feature  When this feature is switched on  and  when a reading is stable for more than 5 seconds  the display will automatically     HOLD    the reading  The    HOLD    indicator appears  Press the HOLD key to release  the reading  To switch on or off the auto endpoint feature  see SETUP program P4 1  on page 30       20      Instruction Manual CON 400 410       6 MEMORY AND DATA INPUT FUNCTIONS       6 1 Memory Input MEAS    READY  uS  Your meter stores up to 50 sets of data  Data sets 8 d 3    include conductivity  temperature  date and time  d 7 3   c  D ATC    To store a reading   1    During any measurement function  MEAS   press    MI A key to input any data into the memory    2  MEM     Sto    and memory number will flash  The  meter then returns to measurement mode    3  If necessary  measure the next sample solution and  press MI A key to input the next data into the  memory    See figure 14     NOTE  If the memory is full  the first value stored will be  erased to create space for the new value     MEAS    READY 5 6 o us  Ce H ie       MEM  MEM d  Figure 14   Memory input     21     Instruction Manual    6 2 Memory Recall    This function recalls the previous readings stored in the  memory  You can access MR from the measurement  mode only  Memory recall is in  Last In First Out  order     To recall readings   1  Press the MR V key once to retrieve the last    reading stored  The memory location screen     MEM      Loc    and the memory number     
10.  helps you to meet the Good Laboratory Practice  GLP   standards     From measurement mode   1  Press SETUP key to enter Set Up mode    2  Press the MI A or MR V keys to scroll through  subgroups until you view parameter P7 0    3  Press the ENTER key to enter parameter P7 0  The  meter lets you select the century     19       or    20        The  century digits will flash    4  Press the MI A or MR V keys to toggle to the correct  century    5  Press ENTER key to confirm the century and move to     year    selection  The    year    digits will flash    6  Press the MI A or MR V keys to toggle to the correct  year  See Figure 33    7  Press ENTER key to confirm the year and move to     month    selection  The    month    digits will flash    8  Press the MI A or MR V keys to toggle to the correct  month    9  Press the ENTER key to confirm the month and move  to    date    selection  The    date    digits will flash  See  Figure 33    10  Press the MI A or MR V keys to toggle to the correct  date    11  Press the ENTER key to confirm the date and move to     hour    selection  The    hour    digits will flash  See Figure  34  next page                  Figure 33   P5 0  Setting century   year  month and date     37     Instruction Manual CON 400 410    12  Press the MI A or MR V keys to toggle to the correct  hour  Note the    AM    and    PM    indicator on the lower  portion of the display    13  Press the ENTER key to confirm the hour and move  to  minute    selection  The
11.  measurement throughout the entire range of the meter  If a  range was not calibrated  the meter automatically detects the closest range calibrated and  uses that calibration information  However  only the ranges that were calibrated have  maximum accuracy     e If you are measuring in ranges near to or greater than 20 mS  10 ppt   or  near to or lower than 100 uS  50 ppm   calibrate the meter at least once a  week to get specified  1  F S  accuracy   e If you are measuring in the mid ranges and you washed the probe in  deionized water and stored it dry  calibrate the meter at least once a month   e  Wet the probe for 10 minutes before calibrating or taking readings to saturate the probe  surface and minimize drift  If you make measurements at extreme temperatures  calibrate the  meter at least once a week     You should only use the conductivity   TDS   temperature probe specified for these meters   These probes have a built in temperature sensor  If you use a different probe without a  temperature sensor  you must measure the solution temperature separately and manually  enter the solution temperature  see manual temperature compensation section on pages 16   17       45     Instruction Manual CON 400 410       14 ADDENDUM 2  CALCULATING TDS CONVERSION  FACTORS       You can calibrate your meter using TDS calibration standard solutions  The calibration  standard only needs to give the TDS value at a standard temperature such as 25   C  To  determine the conductivity to TDS 
12.  reading to  stabilise    4  Press CAL MEAS to enter TDS calibration mode  The  CAL indicator will appear in the upper right corner of the READY  display    5  Press the MI A or MR V to change the value on the  primary display to match the value of the calibration  standard referenced to your normalisation temperature   usually 25  C     6  Press ENTER to confirm the calibration value  The meter  returns to the MEAS  measurement  mode  MEAS    7  Repeat steps 1 to 6 for other measuring ranges  nen J Og  ppm    NOTES 258 i    To exit from TDS Calibration mode without confirming  calibration  DO NOT press the ENTER key in step 6  Press  CAL MEAS instead  This will retain the meter   s old calibration  data in the measuring range of the calibration    You can offset the TDS reading up to  40  from the default  setting  If your measured value differs by more than  40    clean or replace probe as needed    The minimum offset reading allowable is limited to 10  of full scale reading of the range you  are working in  The maximum offset reading allowed is the range full scale reading   Eutech Instruments  Oakton Instruments offer a wide selection of high quality calibration  standards  See page 44 for more information        Figure 8   TDS Calibration     12     Instruction Manual CON 400 410    4 5 2 Calibration with Conductivity Standard and TDS factor  For CON 410 meter only    The concentration of salts dissolved in solution increases the conductivity of that solution  This  rela
13.  when  turned ON    Unstable readings    Slow response    Batteries not in place or  battery housing is not  attached to meter correctly    Batteries not in correct  polarity    and     position      Weak batteries  Air bubbles in probe   Dirty probe     Probe not deep enough in  sample     External noise pickup or  induction caused by nearby  electric motor     Broken probe   probe guard not in place    Dirty   Oily probe       41      CON 400 410    Check that batteries are in  place and making good  contact     Re insert batteries with  correct polarity     Replace batteries     Tap probe to remove  bubbles     Clean the probe and re   calibrate     Make sure sample entirely  covers the probe sensors     Move or switch off  interfering motor     Replace probe   Attach probe guard    Clean probe  See    Probe  Care  amp  Maintenance     page  40        Instruction Manual          10 ERROR MESSAGES          CON 400 410       LCD Display    Indicates    Cause    Solution       Err annunciator    Unrecognised input  from keypad    Wrong input in selected    mode     Release key  Select valid  operations depending on  mode        CAL  amp  Err  annunciators on    Buffer and  electrode  indicators blink     Calibration error     Wrong value input at  calibration     Dirty probe     Check your input value   clean probe     See Calibration sections or  Probe Maintenance  section        Battery indicator  blinks    Low battery level     Need new batteries or    battery connection i
14.  with  k 1 0  EC CONSEN91W      EC CONWP410 03    35608 10       Waterproof CON 410 Hand held conductivity TDS  meter complete with conductivity temp probe with  k 1 0  EC CONSEN91W   packed in hard plastic  carry case    EC CONWP410 03K       Ultem body conductivity temp probe for CON 400  series meters  k  1 0  16 X 110 mm  1 m cable    EC CONSEN91W    35608 50       Ultem body conductivity temp probe for CON 400  series meters  k  1 0  16 X 110 mm  5 m cable    35608 57       Epoxy body conductivity temp probe for CON 400  series meters  k  0 1  12 X 110 mm  1 m cable    Epoxy body conductivity temp probe for CON 400  series meters  k  10 0  12 X 110 mm  1 m cable    EC CONSEN72W    EC CONSEN73W    35608 55    35608 51       Hard carrying case    WPDRYKIT    35632 98       Calibration Solutions       Calibration standard solutions referenced to 25  C    Eutech Instruments    Oakton Instruments                               Ordering Code Ordering Code  10 uS single use pouches 20 mL x 20 pcs EC CON 10BS 35653 09  447 uS single use pouches 20 mL x 20 pcs EC CON 447BS 35653 10  1413 uS 480 mL bottle  1 pint  EC CON 1413BT 00653 18  1413 uS single use pouches 20 mL x 20 pcs EC CON 1413BS 35653 11  2764 uS 480 mL bottle  1 pint  EC CON 2764BT 00653 20  12 88 mS 480 mL bottle  1 pint  EC CON 1288BT 00606 10  15 00 mS 480 mL bottle  1 pint  7 35653 13         44         Instruction Manual CON 400 410          13 ADDENDUM 1  CALIBRATION TIPS       You only need one calibration for
15. Instruction Manual    CON 400 and CON 410    Waterproof Hand held Conductivity TDS Meter          68X248904    nology Made Easy     ISO 9001 Rev3 01 04    109001    Preface    This manual serves to explain the use of the Waterproof CON 400 410 hand held meters  It  functions in two ways  firstly as a step by step guide to help you to operate the meter   Secondly  it serves as a handy reference guide  It is written to cover as many anticipated  applications of the Waterproof CON 400 410 meters as possible  If there are doubts in the  use of the meter  please do not hesitate to contact the nearest Authorised Distributor     Eutech Instruments  Oakton Instruments cannot accept any responsibility for damage or  malfunction to the meter caused by improper use of the instrument     The information presented in this manual is subject to change without notice as improvements  are made  and does not represent a commitment on the part of Eutech Instruments Pte Ltd   Oakton Instruments     Copyright    1999   Eutech Instruments Pte Ltd  Oakton Instruments  All rights reserved    Rev 3  01 04    TABLE OF CONTENTS                  INTRODUG THON BE 1   DISPLAY AND KEYPAD FUNCTIONS 2  2 1 DISPLAY NTE SETTE 2  2 2 ENEE EE 3   PREPARATION Haut kaiene EE 4  3 1 INSERTING THE BATTERIES EE 4  3 2 CONDUCTIVITY ELECTRODE INFORMATION     s ssssssskssisitsitintnitttstttintintatantntntatnnenna ena 5  3 3 CONNECTING THE PROBE TO THE METER  6   GALIBRATIONES SAS 7  4 1 IMPORTANT INFORMATION ON METER CA
16. MI A or MR V keys to select Automatic or   Manual Temperature Compensation    e YES   Automatic Temperature Compensation  activated    e NO   Manual Temperature Compensation  activated     Press the ENTER key to confirm selection and to KS  return to the subgroup menu  Press the CAL MEAS    key to return to measurement mode  SETUP  HEL  no            Figure 28   Selecting Automatic or  Manual Temperature Compensation      30     Instruction Manual CON 400 410    7 5 4 P4 4 Setting the TDS factor   For CON 410 meter only   The concentration of salts dissolved in solution increases the conductivity of that solution  This  relationship varies from salt to salt and is roughly linear over a given range for a given salt   The TDS conversion factor is the number used by the meter to convert from conductivity to  TDS     To determine the conductivity to TDS conversion el    factor for your solution    Addendum 2 on page 46 explains the calculation in  determining the conductivity to TDS conversion factor   You can set the TDS conversion factor between 0 40 and  1 00  meter default is 0 50     From measurement mode   1  Press SETUP key to enter Set Up mode    2  Press the MI A or MR V keys to scroll through  subgroups until you view parameter P4 0    3  Press the ENTER key five times to select parameter  4 4  The upper display shows    tdS    and the lower  display shows    P4 4       4  Press the ENTER key again  The upper display  shows a value and the lower display shows    tdS    
17. ON 400 410          5 MEASUREMENT       This meter is capable of taking measurements with automatic or manual temperature  compensation  Factory default is set with automatic temperature compensation  ATC  on     5 1 Automatic Temperature Compensation    For automatic temperature compensation  ATC  simply   plug the conductivity TDS probe into the meter  see page READY   6 for directions   The ATC indicator will light on the LCD    NOTE  If the ATC indicator does not light  manual    temperature compensation may be selected in 245 E  the meter s SETUP mode  See Program P4 3 on  page 32 for directions on selecting Automatic  Temperature Compensation     MEAS    1402          Figure 11   ATC annuftciator will  light up when connected to  temperature probe     15     Instruction Manual CON 400 410    5 2 Manual Temperature Compensation    IMPORTANT  For manual compensation  you must deactivate the automatic temperature  compensation feature     5 2 1 Selecting Manual Temperature SETUP  Compensation g A E   To select between Automatic Temperature Compensation    ATC  and Manual Temperature Compensation in the D u A    SETUP program P4 3  Meter default is ATC on     P4 0 Configuration Setup    From measurement mode  Se l 1  Press SETUP key to enter Set Up mode   2  Press the MI A or MR V keys to scroll through  subgroups until you view parameter P4 0     HE E 3  Press the ENTER key three times to select parameter  l 4 3  The upper display shows    ATC    and the lower  H q J   c d
18. Press CAL MEAS key  to return to measurement mode        Figure 31   Adjusting the  normalization temperature     35     Instruction Manual CON 400 410    7 7 P6 0 Selecting the cell constant    Your meter lets you select a cell constant of K   1 0  10  or 0 1    Use a cell of K   1 0 for midrange measurements   Use a cell of K   10 for high range measurements  above 20 mS or 10 ppt    Use a cell of K   0 1 for low range measurements  below 20 uS or 10 ppm    The cell included with your meter has a cell constant of K   1 0     From measurement mode   1  Press SETUP key to enter Set Up mode    2  Press the MI A or MR V keys to scroll through  subgroups until you view parameter P6 0    3  Press the ENTER key to select parameter 6 1    4  Press the MI A or MR V keys to select the cell  constant between K   1 0  0 1  or 10    5  Press the ENTER key to confirm selection and to  return to the subgroup menu  Press the CAL MEAS  key to return to measurement mode     NOTES   When using a cell of K   0 1  the lowest measuring range  will be  0 to 1 999 uS  0 to 0 999 ppm   The 0 to 199 9 ms   10 to 200 ppt  range will not be accessible    When using a cell of K   10  the highest measuring range  will be  0 to 1999 mS  100 to 2000 ppt   The 0 to 19 99 uS   0 00 to 9 99 ppm  range will not be accessible        Figure 32   Selecting cell constant      36      Instruction Manual CON 400 410    7 8 P7 0  Setting the real time clock    The meter features a real time calendar and clock  This 
19. Setup mode    Press the MR    key until the lower display reads P5 0    Press the ENTER key twice  The lower display reads tCO and the upper display  shows the temperature coefficient value  should be 0  as per step 1 above     Press the MI A key until the upper display shows your calculated temperature  coefficient    Press ENTER key to confirm the value    Press CAL MEAS key twice to return to measurement mode     The calculated temperature coefficient will not be applied to all the meter readings      47     Instruction Manual    CON 400 410          16 ADDENDUM 4  METER FACTORY DEFAULT                                  SETTINGS   Type Parameter Default Remarks   P1 0 Memory clear No Retains current memory   P2 1 Viewing previous    No calibration data for 1  range   P2 2 calibration data     No calibration data for 2   range   P2 3     No calibration data for 2 range   P2 4     No calibration data for 4  range   P2 5     No calibration data for 5  range   P3 1 Viewing probe data 1 0 No offset for effective cell constant  1  range    P3 2 1 0 No offset for effective cell constant  279 range    P3 3 1 0 No offset for effective cell constant  3   range    P3 4 1 0 No offset for effective cell constant IEN range    P3 5 1 0 No offset for effective cell constant  5  range    P4 1 Ready indicator   auto READY ON Ready indicator on  auto endpoint off   endpoint   P4 2 Select   C   F   C Degrees Celsius   P4 3 ATC on or off ATC on       P4 4 TDS factor 0 50 Adjustable from 0 40 to
20. UBRATION   7  4 2 PREPARING THE METER FOR CALIBRATION pr  4 3 TEMPERATURE CALIBRATION  Gustad Ce Mea   4 4 CONDUCTIVITY CALIBRATION   arnnnrenvnnnvenvrnnvenvrrnnnnvernnnnrrsvrnnnennnrnnesvrrnnrsnrsnnnnresnernnrnnennn  4 5 TDS CALIBRATION        ccsecseseeeeesees   4 5 1 Calibrating for TDS directly   4 5 2 Calibration with Conductivity Standard and TDS factor       svrrnrnnvrnnrnnvrrnrnnrnr 13   MEASUREMENN vs ER  15  5 1 AUTOMATIC TEMPERATURE COMPENSATION   sresrrs ven renrnnnrnevenrennrnerrerrerreneenernrrrerrennene 15  5 2 MANUAL TEMPERATURE COMPENSATION         cesccesseseeseeseeesecseeeeecseseeeaeeaeeeseeseeseeeaeens 16   5 2 1 Selecting Manual Temperature Compensaton  eane 16   5 2 2 Setting a manual temperature compensation value 17  5 3 TAKING MEASUREMENTS EE 18  5 4 USING MANUAL RANGING FUNGCTION   19  5 5 HOLD  FUNGTIOSN   knes er 20   MEMORY AND DATA INPUT FUNCTIONS    rrrrrrnrrnnnnrrrrrrnnrnnnrrrrnnennnnnnnnnrvnnnn 21  6 1 MEMORYINPUT cruna ne n R O A a A  6 2 MEMORY RECALL   ADVANCED SETUP FUNCTIONS   mmnnnnnnnrvvnennnnnnnnrrnvernnnnnnnnrrnnnnssnnrnnnrrnnnnssnnnnn 23  7 1 ADVANCED SETUP MODE OMERVIEwW  25  7 2 P 1 0  MEMORY CLEAR  Clm 27  7 3 P2 0  VIEWING CALIBRATION DATA    28  7 4 P3 0  VIEWING PROBE DATA  29  7 5 P4 0  UNIT CONFIGURATION  esrrnnvrnvrrnvrnvrrnvrnvernrnnvennrrnnennrrnnnsvrrnnns nes nennnesvennnnsnssnnnnnenne 30   7 5 1 P4 1  READY indicator and auto endpoint function       aeearennvrnnvrrnnvrnvvennvre 30        7 5 2 P4 2 Selecting  Got F oae a
21. ature coefficient of your sample solution use this formula     Where     DEES CH    FEIN 5    tc   Temperature coefficient 25   25   C  Cr    Conductivity at Temp 1 Cr2   Conductivity at Temp 2  Ti   Temp 1 T2   Temp 2    NOTE  A controlled temperature water bath is ideal for this procedure     1  Immerse the probe into a sample of your solution and adjust the temperature coefficient  to 0   that is  no compensation  by performing the following     Err  A    EN    From measurement mode  press the SETUP key to enter Setup mode    Press the MI A or MR V key until the lower display reads P5 0    Press the ENTER key twice  The lower display reads tCO and the upper display  shows the temperature coefficient value    Press the MR V key until the upper display shows 0 0    Press ENTER key to confirm the value    Press CA MEAS key twice to return to measurement mode     Wait for 5 minutes  Note Ti and Cy   conductivity at T4    Condition the sample solution and probe to a temperature  T2  that is about 5   C to 10   C    different from T   and note the conductivity reading Cro     NOTE  Record your results for future reference  Ideally Ti and T   should bracket your    measurement temperature  and should not different by more than 5   C     4  Calculate the temperature coefficient of your solution according to the formula shown  above   5  Enter the temperature coefficient you calculated into the meter     A   B   C   D    E   F    From measurement mode  press the SETUP key to enter 
22. conversion factor use the following formula     Factor   Actual TDS   Actual Conductivity   25   C    Definitions   e Actual TDS  Value from the solution bottle label or as a standard you make  using high purity water and precisely weighed salts   e Actual Conductivity  Value measured using a properly calibrated  Conductivity Temperature meter     Both the Actual TDS and the Actual Conductivity values must be in the same magnitude of  units  For example  if the TDS value is in ppm the conductivity value must be in uS  if the TDS  value is in ppt the conductivity value must be in mS     Check your factor by multiplying the conductivity reading by the factor in the above formula   The result should be in TDS value     14 1    Preparing your own TDS calibration standards        If you are trying to measure a concentration of a fertilizer  sea salt  or other sample that uses  multiple salts that may not be relevant to commercially available KCI based standards  it may  be desirable to prepare your own TDS standards     Since ppm is expressed as milligrams per Liter  the preparation is relatively simple with an  analytical balance  One gram of water   1 mL   1 cubic centimeter     To prepare a TDS calibration standard of 750 ppm  weigh out 750 milligrams  0 75 grams  of  your salt or fertilizer and dissolve into 1000 grams  1 liter  of purified water      46      Instruction Manual CON 400 410       15 ADDENDUM 3  CALCULATING TEMPERATURE  COEFFICIENTS       To determine the temper
23. cting the CON 400 410 waterproof portable meter  This meter is a  microprocessor based instrument that is designed to be user friendly and allow one hand  operation  It has a built in real time clock  expanded memory  and many other user friendly  features  all of which are accessible through the membrane keypad  It is completely  WATERPROOF     and it FLOATS  You have one of the following models     e CON 400 meter  reads conductivity and temperature   e CON 410 meter  reads conductivity  TDS and temperature     Your meter includes 4    AAA    batteries and a conductivity electrode  cell constant K   1 0  with  built in temperature sensor  Please read this manual thoroughly before operating your meter        Instruction Manual CON 400 410          2 DISPLAY AND KEYPAD FUNCTIONS       2 1 Display    The LCD has a primary and secondary display   e The primary display shows the measured conductivity or TDS reading   e The secondary display shows the measured temperature   The display also shows error messages  keypad functions and program functions     See Figure 1     Primary Display       20  19  18  17 ooqatt  16 SL ATC  15 1413 11 Secondary Display 10  Figure 1  Full LCD Screen  1  SETup mode indicator 8  parts per million indicator  CON 15  Probe indicator  410 meter only   2  MEASurement mode indicator 9  Temperature indicator 16  Calibration solution indicator  3  CALibration indicator 10  Automatic Temperature 17  Cell constant indicator  Compensation indicator  4  MEMo
24. ductivity or TDS calibration mode  press  the MODE key to enter temperature calibration mode   The primary display shows the current temperature  setting and the secondary display shows the default  value 25   C  77   F  or its last set value    5  Check the temperature of your sample using an  accurate thermometer    6  Press the MI A or MR V keys to offset the  temperature to the measured value from step 5    7  Press ENTER to confirm the selected temperature and Figure 9   Manual Temperature   SC Compensation  to return to the conductivity or TDS measurement  mode        The meter will now compensate conductivity or TDS readings for manually set temperature   NOTES     To exit this program without confirming the manual temperature compensation value  DO NOT  press ENTER in step 7  Press CAL MEAS instead      17     Instruction Manual CON 400 410    5 3 Taking Measurements    To take readings  MEAS   1  Rinse the probe with DI or distilled water before use READY 4 LU 5     to remove any impurities adhering to the probe     j  body  Shake or air dry  To avoid contamination or  dilution of your sample  rinse probe with a small  volume of your sample liquid    2  Press ON to switch on meter and the MEAS  annunciator appears on the top center of the LCD    3  Dip the probe into the sample     213        Figure 10   During measurement    When dipping the probe into the sample  take care to ensure that the liquid  level is above its upper steel band  Stir the probe gently in the 
25. e a i eaea eenaa aaa ea O teates tile  avd 31  7 5 3 P4 3 Selecting Automatic or Manual Temperature Compensation                32  7 5 4 P4 4 Setting the TDS factor AAA 33  7 6 P5 0 TEMPERATURE  geet as ata da en d y dat akebakke 34       7 6 1 P5 1 Adjusting the temperature coefficient         rrrrrnrvnornrrrnrnnrrennnrrrrnn renere 34  7 6 2 P5 2 Adjusting the normalization temperature        rrrvnnnnvrenonnrrennnnrrnnnn renn ner 35  7 7 P6 0 SELECTING THE CELL CONSTANT        c  cceceeeeeeeeeeeeeeeeeeeeaeeseeeeeeeseaeeeaeeeeaeeseeeaes 36  7 8 P7 0  SETTING THE REAL TIME CLOCK        ccccceseceeeeeeseeeeeeeeeeeeaeeeeeeeeeeeseaeeeeeeeaeeneeeaaes 37        7 9 P8 0  RESETTING TO FACTORY DEFAULT SETTINGS  39    8 PROBE CARE AND MAINTENANCE    envrrnvonnrennrernrervrnrvnnvvnnvvennrernrerverrennrennen 40    9 TROUBLE SHOOTING GUIDE ee eee cee teee tees teeter eter seaeeeeeeeeeeas 41  10 ERROR  MESSAGES  innredn 42  11 SPECIFICATIONS 2 2 ea eege ccs tie ES ei ths E ee eel 43  12 AGGESS  RIES unne 44  13 ADDENDUM 1  CALIBRATION TIPS    renvnnnvnnnvennvennrernrenvrnnvonnrnnnvennrenneennnn 45  14 ADDENDUM 2  CALCULATING TDS CONVERSION FACTORS                46   14 1     PREPARING YOUR OWN TDS CALIBRATION STANDARDE  46  15 ADDENDUM 3  CALCULATING TEMPERATURE COEFFICIENTS           47  16 ADDENDUM 4  METER FACTORY DEFAULT SETTINGS ee 48  17 WARRANTY  pusser eee 49    18 RETURN OF ITEMS urent nd d nisse 50    Instruction Manual CON 400 410       1 INTRODUCTION       Thank you for sele
26. hile in Measurement mode  The first N JH   3 us  range will appear on the display and the    MEAS       indicator blinks    2  Press the RANGE key again  if needed  until desired  range is selected    3  To re select the Auto ranging function  repeatedly  press the RANGE key until the    MEAS    indicator  appears without blinking     ped     mem  MEAS    sed READY 1413     If the value of the solution you are measuring is higher than A    the range selected    Or    will appear on the primary display  SE arc  Press RANGE until the correct range is selected        Figure 11   Manual ranging     MEAS  READY A us The meter resets to the Auto ranging function once it is  r turned off  You will have reset the manual ranging function  each time you turn the meter off     ee    Figure 12   Out of range        19     Instruction Manual CON 400 410    5 5 HOLD Function    This feature lets you freeze the display for a  delayed observation  HOLD can be used any time  in MEAS mode     1  To hold a measurement  press the HOLD  key while in measurement mode     HOLD    will  appear on the display    2  To release the held value  press the HOLD  again  Continue to take measurements        Figure 13   HOLD function    NOTE    e     This meter shuts off automatically after 20 minutes of nonuse  without key press     e If the meter is shut off either automatically or manually  the HOLD value will be lost   For longer storage  use the memory functions  see pages 21 22     e     Your meter has
27. is set for Ready Indicator on  and auto    endpoint function off   Figure 26   Configuring the  READY feature      30      Instruction Manual CON 400 410    7 5 2 P4 2 Selecting   C or F    You can select between   C and   F units for temperature readings  Meter default is   C     From measurement mode    1   2     See Figure 27     Press SETUP key to enter Set Up mode  l  Press the MI A or MR V keys to scroll through ec  subgroups until you view parameter P4 0    Press the ENTER key two times to select parameter  4 2    Press the MI A or MR V keys to toggle between    C and   F    Press the ENTER key to confirm selection and to  proceed to step 3 of P4 3  Press the CAL MEAS  key to return to measurement mode        Figure 27   Change Temperature  measurement unit     31     Instruction Manual CON 400 410    7 5 3 P4 3 Selecting Automatic or Manual Temperature Compensation       This feature lets you select between Automatic Temperature Compensation  ATC  and  Manual Temperature Compensation  Meter default is ATC activated     1    From measurement mode Leeft    Press SETUP key to enter Set Up mode  GW    2     Press the MI A or MR V keys to scroll through   subgroups until you view parameter P4 0      Press the ENTER key three times to select parameter HE L   4 3  The upper display shows    ATC    and the lower S   display shows    P4 3     P H 3 ATC   Press the ENTER key again  The upper display   shows    ATC    and the lower display shows    YES    or      NO       Press 
28. isplay shows    P4 3        ATC    4  Press the ENTER key again  The upper display shows     ATC    and the lower display shows    YES    or    NO      5  Press the MI A or MR V to select between Automatic  or Manual Temperature Compensation feature   e YES   Automatic Temperature Compensation  activated   e NO   Manual Temperature Compensation  activated   6  Press the ENTER key to confirm selection and to    return to the subgroup menu  Press the CAL MEAS  KS key to return to measurement mode           SETUP  NOTE   GEL Manual Temperature Compensation should only be  selected if the ATC fails  For non compensated  ng BG measurements  change the temperature coefficient to    0 0   see section 7 6 1     Selecting Automatic or Manual  Temperature Compensation      16     Instruction Manual CON 400 410    5 2 2 Setting a manual temperature compensation value       To use manual temperature compensation  you need to  manually enter the temperature value of your process into  the meter  You can select any temperature between 0 and  100   C  32 to 212   F   Default value is 25   C     To select a manual temperature compensation value   1  Switch the meter on  Press the MODE key to select  measurement mode    2  If necessary  select ATC off as described in section  5 2 1 on page 16  The ATC indicator will not appear  on the display    3  Press the CAL MEAS key to enter conductivity or TDS  calibration mode  The CAL indicator will appear above  the primary display    4  While in con
29. rn Goods Authorisation     RGA  form  is available through our authorised distributor  Please include data regarding the reason the  items are to be returned  For your protection  items must be carefully packed to prevent  damage in shipment and insured against possible damage or loss  Eutech Instruments   Oakton Instruments will not be responsible for damage resulting from careless or insufficient  packing  A restocking charge will be made on all unauthorised returns    NOTE  Eutech Instruments Pte Ltd  Oakton Instruments reserves the right to make  improvements in design  construction  and appearance of products without notice       50      NOTES    For more information on Eutech Instruments  Oakton Instruments    products  contact    your nearest distributor or visit our website listed below     Oakton Instruments   P O Box 5136    Vernon Hills  IL 60061  USA  Tel   1  888 462 5866   Fax   1  847 247 2984  E mail  info 4oakton com  Web sites   www 40akton com       www oaktoninstruments com       Eutech Instruments Pte Ltd   Bik 55  Ayer Rajah Crescent     04 16 24 Singapore 139949  Tel   65  6778 6876   Fax   65  6773 0836   E mail     marketing Geutechinst com    Web site  www eutechinst com       Distributed by           
30. robe with 6 pin  connector    den ae                _ locking ring    top view of  6 pin connector  of probe       Figure 5   Connection for Conductivity probe  6 pin connector     Instruction Manual CON 400 410          4 CALIBRATION          4 1 Important Information on Meter Calibration    Your meter has five measuring ranges  You can calibrate one point in each of the measuring  ranges  up to five points   If you are measuring values in more than one range  make sure to  calibrate each of the ranges you are measuring     To view current calibration points  see SETUP section Program 2 0 on page 28   The following table lists the corresponding conductivity and TDS ranges  You should calibrate    each range using a solution that falls between the values in the    recommended calibration  solution range    column                                Conductivity Recommended TDS Measuring Recommended  Measuring Calibration Solution Range Calibration Solution  Range Range Range   0 00  gt  19 99 uS 6 00 to 17 00 uS 0 00  gt  9 99 ppm 3 00 to 8 50 ppm   0 0  gt  199 9 uS 60 0 to 170 0 uS 10 0  gt  99 9 ppm 30 0 to 85 0 ppm  0  gt  1999 uS 600 to 1700 uS 100  gt  999 ppm 300 to 850 ppm  0 00  gt  19 99 mS 6 00 to 17 00 mS 1 00  gt  9 99 ppt 3 00 to 8 50 ppt  0 0  gt  199 9 mS 60 0 to 170 0 mS 10 0  gt  200 ppt 30 0 to 170 ppt    When you recalibrate your meter  old calibrations are replaced on a range by range basis  For  example  if you previously calibrated your conductivity meter at
31. ry recall mode indicator 11  Clock indicator 18  ON   OFF indicator  5  millisiemens indicator 12  ERRor indicator 19  HOLD indicator  6  microsiemens indicator 13  MEMory location indicator 20  READY indicator  7  parts per thousand indicator 14  Low battery indicator   CON 410 meter only     Instruction Manual CON 400 410    2 2 Keypad    The large membrane keypad makes the instrument easy to use  Each button  when pressed   has a corresponding graphic indicator on the LCD  See Figure 2  Some buttons have several  functions depending on its mode of operation        Key Function    ON OFF Powers on and shuts off the meter  When you switch on the meter  the meter starts up  in the mode that you last switched off from  For example  if you shut the meter off in  TDS measurement mode  only in CON 410 meter   the meter will be in TDS  measurement mode when you switch the meter on           HOLD Freezes the measured reading  To activate  press HOLD while in measurement mode   To release  press HOLD again        MODE Selects the measurement parameter      CON 400 meter  Toggles between conductivity and time      CON 410 meter  Toggles between conductivity  TDS and time   CAL MEAS   Toggles between Calibration and Measurement mode     NOTE  Temperature calibration is available from conductivity calibration mode  see  page 9 for directions        ENTER   ENTER function  Press to confirm values in Calibration mode and to confirm selections  RANGE in SETUP mode     RANGE function  Pres
32. s  bad     Clean battery contacts   Replace batteries with fresh  ones  noting polarity        Err 1  in primary  display     Memory write error     Hardware failure     Turn meter on and off  again  If message persists   return unit         Err  2  in primary  display     Memory checksum  error     Hardware failure     Turn meter on and off  again  If message persists   return unit         Err  3  in primary  display     A D converter error     Hardware error     Turn meter on and off  again  If message persists   return unit         Err  4  in primary  display           Keypad error        One or more keys on    the keypad are stuck or    fault in keypad        Turn meter on and off  again  If message persists   return unit           See    Warranty    and    Return of Items    on pages 49 and 50   If an error message appears in the primary display  the  upper row of larger digits   switching off the meter and  switching it on again may eliminate the error message   Refer to diagram on right     Err I    If error persists  or the meter shows incorrect values  return the meter   For a complete diagram of the display see page 2      42        Instruction Manual CON 400 410          11 SPECIFICATIONS    SPECIFICATIONS DESCRIPTIONS CON 400   CON 410    0 to 19 99  199 9  1999 uS cm   EE 0 to 19 99  199 9 mS cm ARA  0 00 to 9 99  0 0 to 99 9  0 to 999 ppm  mn  TDSR  alt 0 00 to 9 99ppt  0 0 to 199 9 ppt  0 05   Full Scale Wem Leen   1  F S   1 digit    0 0 to 100 0   C  Temperat
33. s to enter manual ranging function   The MEAS indicator blinks while in manual ranging function        In Measurement mode     MI A and Press MI A  memory input  to store values with its corresponding temperature values in  MR V the memory  Press MR V  memory recall  to retrieve data from memory     In Calibration mode    Press to scroll through calibration values    In SETUP mode    Press to scroll through the setup subgroup programmes        SETUP Takes you into the SETUP mode  This mode lets you customise meter preference and  defaults  and view calibration  electrode offset data and select cell constant                      amp     FOGH  OOO             Figure 2   Keypad    Instruction Manual CON 400 410       3 PREPARATION       3 1 Inserting the Batteries    Four AAA batteries are included with your meter   1  Use a Philips screwdriver to remove the two screws holding the battery cover  See Figure  3 below     2  Remove battery cover to expose batteries   3  Insert batteries  Follow the diagram inside the cover for correct polarity   4  Replace the battery cover into its original position using the two screws removed earlier        Remove these  two screws to  access battery  compartment    Battery  compartment             Figure 3   Back panel of meter showing meter compartment    Instruction Manual CON 400 410    3 2 Conductivity Electrode Information    The CON400 410 meter uses a conductivity   TDS cell with a  sturdy 6 pin connector     Your meter includes a conduc
34. sample to  create a homogenous sample     4  Allow time for the reading to stabilise  Note the reading on the display   5  CON 410 meter only  Press the MODE key to toggle between conductivity and TDS  readings     Taking measurements with READY indicator selected on  If the READY indicator has been activated  the READY annunciator lights when the reading is    stable  Switch the READY indicator on or off in SETUP program P 4 1  See page 30 for  directions     Taking measurements with the auto endpoint feature selected on  When a reading is stable for more than 5 seconds  the auto endpoint feature will automatically       hold    the reading  The    hold    indicator appears on the left side of the display  Press the HOLD  key to release the reading  Switch the Auto endpoint feature on or off in SETUP program P 4 1   see page 30 for instructions      18     Instruction Manual CON 400 410    5 4 Using Manual Ranging Function    Your meter automatically selects the range in which your readings appear  The manual  ranging function lets you select the specific range  and corresponding resolution  that you  want to work in     CON 400 and CON 410 meters  CON 410 meter   he 0 00     19 99 uS 1  0 00     9 99 ppm  2 0 0     199 9 uS 2 0 0     99 9 ppm  3 0     1999 uS 3   0    999 ppm  4  0 00     19 99 mS 4  0 00     9 99 ppt  5 0 0     199 9 mS 5 0     100 ppt   up to 199 9 ppt depending on TDS factor  setting   1  To select the desired measuring range  press the MEAS  RANGE key w
35. splay the date and time  of calibration to help you comply with the Good  Laboratory Practice  GLP  standards    4  When you have scrolled through all calibration data   you will automatically return to the subgroup menu   Press CAL MEAS key to return to measurement  mode     NOTES   If there is no previous calibration data at a particular point   the primary display will show                 CON 410 meter only  If you entered Set Up mode from  Conductivity measurement mode  calibration data will be  in uS or mS  If you entered Set Up mode from TDS  measurement mode  calibration data will be in ppm or ppt      28        CON 400 410    Figure 24   View calibration data  and its relevant time and date    Instruction Manual    7 4 P3 0  Viewing probe data    Program 3 has five  view only    options that let you check  the probe s parameters for diagnostic purposes  These    options show you the effective cell constant for each range     The cell constant is adjusted according to your calibration     From measurement mode   1  Press the SETUP key to enter Set Up mode    2  Press the MI A or MR V keys to scroll through  subgroups until you view parameter P3 0    3  Press the ENTER key repeatedly to view the effective  cell constant for each range    4  When you have scrolled through all probe data  you  will automatically return to the subgroup menu  Press  the CAL MEAS key to return to measurement mode    See Figure 25     NOTES     Cell constants may degrade with time and usage 
36. ss the ENTER key to confirm selection and to    proceed to step 3 of P5 2  Press the CAL MEAS key      twice to return to measurement mode           SETUP  NOTES  d l  If you do not know the temperature coefficient of your D  solution you can determine the correct value using the L ro    formula in Addendum 3    Calculating Temperature  Coefficients    on page 47     ls    5e  E CO    e     Figure 30   Changing the  temperature coefficient     34      Instruction Manual CON 400 410    7 6 2 P5 2 Adjusting the normalization temperature   Your meter will normalize its conductivity measurements to a standard temperature that you  can select  You can adjust the normalization temperature from 15 to 30   C  59 to 86   F   Meter  default is 25   C  77   F      From measurement mode  1   2     NOTE   Most calibration standards are referenced to 25  C  If a  normalization temperature other than 25  C is entered  the  calibration standard value at this temperature must be  known     Press SETUP key to enter Set Up mode    Press the MI A or MR    keys to scroll through  subgroups until you view parameter P5 0    Press the ENTER key three times to select  parameter 5 2  The display shows    t nr    on the  upper display    Press the ENTER key again  The upper display  shows the normalisation temperature and the lower  display shows  nr     Press the MI A or MR V keys to select the  normalisation temperature    Press the ENTER key to confirm selection and to  return to the subgroup menu  
37. time you calibrate your meter     NOTE  Your meter is factory set to a temperature coefficient of 2 1  per  C  For most  applications this will provide good results  See Program P5 1 on page 34 to set the  temperature coefficient to different value  See Addendum 3     Calculating Temperature  Coefficients    to determine the appropriate temperature coefficient for your solution    NOTE  The factory default value for normalization temperature is 25   C  If you need to  normalize to a value other than 25   C  see Program P5 2 on page 35     Instruction Manual CON 400 410    4 3 Temperature Calibration    Your probe features a built in temperature sensor  The  temperature sensor is factory calibrated  Calibrate your  sensor only if you suspect temperature errors that may  have occurred over a long period of time or if you have a  replacement probe     Temperature Calibration  1  Make sure the cell is attached to the 6 pin connector     The ATC annunciator will appear at the right hand  side of the LCD     2  Switch the meter on  Press the MODE key to select  conductivity or TDS mode    3  Press the CAL MEAS key to enter conductivity or  TDS calibration mode  The CAL indicator appears  above the primary display    4  While in conductivity or TDS calibration mode  press  the MODE key to enter temperature calibration  mode  The primary display shows the current  temperature reading and the secondary display  shows the factory default temperature value    5  Dip the cell into a solu
38. tion of known temperature  i e   A temperature bath   Allow time for the built in  temperature sensor to stabilise    6  Scroll with the MI A or MR V keys to set the correct  temperature value  i e  the temperature of the  temperature bath   You can adjust the reading in  increments of 0 1   C    7  Once you have selected the correct temperature   press the ENTER key       If the ATC indicator does not light up  see Program P4 3  to switch it on     NOTES  fey odie    H You can offset the temperature reading up to  5  C from  the original reading  default reading     H To exit this program without confirming the temperature  calibration value  DO NOT press ENTER  Press  CAL MEAS instead    H If the ATC indicator does not light  see Program P4 3   on page 32 to switch it on    H Since temperature readings affect the accuracy of the  conductivity TDS measurements  it is strongly       recommended to carry out   a _ conductivity TDS    calibration after a temperature calibration is done  Figure 10   Temperature  Calibration   9     Instruction Manual CON 400 410    4 4 Conductivity Calibration    The CON 400 and CON 410 meters are capable of up  to 5 point conductivity calibration at one point per  conductivity range  0 00   19 99 uS  0 0   199 9 uS  0    1999 uS  0 00   19 99 mS  0 0   199 9 mS      All new calibration data will over ride existing stored  calibration data for each measuring range calibrated     Calibrating for Conductivity   1  If necessary  press the MODE key 
39. tionship varies from salt to salt and is roughly linear over a given range for a given salt   The TDS conversion factor is the number used by the meter to convert from conductivity to  TDS     Instead of calibrating for TDS directly  described in Section 4 5 1   you can calibrate the CON  410 meter by    1  calibrating to conductivity standards  as described on pages 10 to 11  and then   2  entering the appropriate TDS conversion factor into the meter     To determine the conductivity to TDS conversion factor for your solution   e Addendum 2 explains the formula to calculate the conductivity to TDS    conversion factor     Enter the TDS conversion factor into your meter as described under Section 7 5 4  in P4 4  Setting the TDS Factor on page 33     zie    Instruction Manual    From measurement mode    1     Press Setup key to enter Set Up mode     2  Press the A and YW keys to scroll through subgroups  until you view parameter P4 0    See Figure 9    3  Press the ENTER key again  The upper display shows  a value and the lower display shows    tdS       4  Calculate the TDS factor of your solution  See  Addendum 2 on page 46 for information on how to  calculate the TDS factor    5  Press the A and    keys to select your calculated TDS  conversion factor    6  Press the ENTER key to confirm selection and to    return to the subgroup menu  Press the CAL MEAS  key to return to measurement mode      14     CON 400 410       Figure 9   TDS Conversion Factor    Instruction Manual C
40. tivity probe  Part No  EC   CONSEN91W   35608 50  Ultem   Stainless Steel cells with a cell  constant of K   1 0  This conductivity   TDS cell features a built in  temperature sensor for Automatic Temperature Compensation   ATC   It has a specially designed housing that provides fast  temperature response and reduces air bubble entrapment  which  makes it easy to obtain accurate  stable readings     Wetted parts include    1    Polyetherimide  Ultem       2    Polybutylterphalate  Valox       3  Stainless Steel  SS 304     NOTE  We recommend that you do not submerge the probe above  the protective yellow cap for prolonged periods  The protective  yellow cap can be removed for cleaning  but must be attached at  all times during calibration or measurment     See Section 8 on page 40 for    Probe Care and Maintenance     information        Figure 4   Conductivity probe   EC CONSEN91W  35608 50          Valox and Ultem are trademark names of General Electric Co      5     Instruction Manual CON 400 410    3 3 Connecting the probe to the meter    1  Line up the notch and 6 pins on the meter with the holes in the 6 pin connector  Push  down and turn the locking ring clockwise to lock into place  See figure below    2  To remove probe  turn the locking ring counterclockwise on the probe connector  Pull  probe away from the meter     CAUTION  Do not pull on the probe cord or the probe wires might disconnect   NOTE  Keep connectors clean  Do not touch connector with soiled hands     P
41. to select    CAL  conductivity mode   2  Rinse the probe thoroughly with de ionized  DI  l LU l 6     water or a rinse solution  then rinse with a small  amount of calibration standard  dd 3 SS  3  With the yellow probe guard attached  dip the  probe into the calibration standard  Immerse the  probe tip beyond the upper steel band  Stir the  probe gently to create a homogeneous sample   Allow time for the reading to stabilizse     i CAL  See Figure 7   ok READY uS  4  Press CAL MEAS to enter conductivity 14   4  calibration mode  The CAL indicator will appear on  in the upper right corner of the display  ee 3    ATC    5  Press the MI A or MR V key to change the  value on the primary display to match the value  of the calibration standard referenced to your  normalisation temperature  usually 25  C     6  Press ENTER to confirm calibration value  The  meter returns to the MEAS  measurement  mode    o READY  7  Repeat steps 1 to 6 for other measuring ranges  14 l 3 us    See figure 6   22 3 nc    MEAS       Figure 6   Conductivity Calibration      10     Instruction Manual    NOTES    When entering calibration mode  the meter  displays the factory default value  If the meter  was previously calibrated  the display may  seem to    jump    to the factory default value  when switching from measurement to  calibration mode  This is expected     To exit from Conductivity calibration  mode without confirming calibration  DO  NOT press the ENTER key in step 6  Press  CAL MEAS instead
42. tory default  H E 5       See Figure 35     Figure 35   P6 0  Reset to  factory default values    NOTES   See Addendum 4 on page 48 for a table of factory default  settings      39      Instruction Manual CON 400 410       8 PROBE CARE AND MAINTENANCE       Keep the conductivity probe clean  Rinse the probe twice  and  gently swirl it while you take readings  Rinse the probe with DI  or clean water before storing  Never scratch the bands with a  hard substance  Do not strike the probe against any hard  surface     Do not immerse the probe in oily solutions  Clean the  electrode thoroughly by stirring it in a mild detergent bath or  isopropyl alcohol if needed  Wipe the probe with a soft tissue  paper  Rinse thoroughly in tap water and then in DI water   Recalibrate the meter after cleaning the probe          locking notch    The conductivity probe  Order Part No  EC CONSEN91W   35608 50  which is included with your meter features a    removable probe guard to make cleaning easy  twist probe  guard clock     wise to remove  NOTE  The yellow probe guard must be attached during    measurement and calibration  Readings are unstable and not  accurate without this guard in place     To remove probe guard for cleaning   1  Grip yellow probe guard and twist clockwise  The locking  notch will release     2  Slide probe guard off end of probe  Figure 36   Conductivity probe   EC CONSEN91W  35608 50       40      Instruction Manual       9 TROUBLE SHOOTING GUIDE    Problem    No display
43. ure Range   32 0 to 212   F   0 1   C  Resolution  0 1 from 32   F to 199 9   F  and 1 0 from 200   F to 212   F   Accuracy  0 5   C    0 9   F      Cell Constat   0   1  1 0  10 0  selectable              Temperature Compensation   Automatic Manual fromotosoec             0 0 to 10 0      C poe eee    Normalization Temperature   15010300   C  adjustable      J  gt     emer  oe M   Number of calibration points 5  Maximum 1 per range         Auto ranging Yes  manual select available    HOLD Function Yes   Memory Yes   Auto Power off 20 minutes after last key operation          em       em      inputs    Gpinroundcommestor         gt       Display  O Cusombuateo S oe o oe    Power Requirements       4 AAR Batteries J o d  gt       Battery Lie S     100hous Soe So oe    i    gt  Meter  19 x 10 x 6 cm  320 g  Pre EON Boxed  24 x 23 x 7 cm  700 g pr                           43     Instruction Manual          12 ACCESSORIES          Replacement Meter and Meter accessories    CON 400 410       Description    Eutech Instruments  Ordering Code    Oakton Instruments  Ordering Code       Waterproof CON 400 Hand held conductivity meter  complete with conductivity probe of k 1 0  EC   CONSEN91W      EC CONWP400 03    35608 00       Waterproof CON 400 Hand held conductivity meter  complete with conductivity probe of k 1 0  EC   CONSEN91W  packed in hard plastic carrying case    EC CONWP410 03K       Waterproof CON 410 Hand held conductivity TDS  meter complete with conductivity temp probe
44. will flash on the  display    2  Press the ENTER key to recall the reading stored  under that memory number    3  Press ENTER key again to view the date and time  the reading was taken    4  Press ENTER key again to return to the  memory  location  screen  The display automatically moves  to the next memory location screen    5  If necessary  press the MI A or MR V keys to  select the next  memory location  screen  press the  ENTER RANGE key to select the previous  memory  location  screen    6  Repeat steps 2   5 to review additional stored data  sets    7  To exit Memory Recall  press MEAS key to return to  Measurement mode     NOTES    Readings stored in memory are retained even if the units  are turned off  To erase all readings stored in memory   use the SETUP mode P1 0 on page 27      22        CON 400 410    MEAS    116     22 3        2 24      109    Figure 20   Memory recall    Instruction Manual CON 400 410       7 ADVANCED SETUP FUNCTIONS       The advanced setup mode lets you customised your meter s preferences and defaults  The  Waterproof CON400   CON410 meters features different sub groups that organise setup  parameters    The sub groups are     1  P1 0  Memory clear  CLr    2  P2 0  Viewing calibration data  CAL   3  P3 0  Viewing electrode data  ELE   4  P4 0  Unit configuration  COF    5  P5 0  Temperature  tPr    6  P6 0  Selecting cell constant  CEL   7  P7 0  Setting clock  CLO    8  P8 0  Reset to factory defaults  rSt     See Figure 22 on next page   
    
Download Pdf Manuals
 
 
    
Related Search
    
Related Contents
Outdoor Fireplace  Soundshield 4G Userguide  5 MODE D`EMPLOI.indd  LevelOne FEU-0510 network switch  Benq MS630ST    manuel d`utilisation  SLー…ラER  「オレッセ Ent`S」カタログ(pdf)  le dossier de presse    Copyright © All rights reserved. 
   Failed to retrieve file