1. 24 Routine Maintenance eese e teten tatnen teta tontnin 24 SECTION 8 TROUBLESHOOTING GUIDE 27 SECTION 9 OPTIONAL ACCESSORIES 29 RECYCLING INFORMATION 30 LIMITED WARRANTY cccccccccccccccccccccccccccccces 3 I Part No 1150713 3 Perfecto Series SPECIAL NOTES SPECIAL NOTES Signal words are used in this manual and apply to hazards or unsafe practices which could result in personal injury or property damage Refer to the following table for definitions of the signal words SIGNAL WORD MEANING Danger indicates an imminently hazardous situation which if not avoided will result in death or serious DANGER injury Warning indicates a potentially hazardous situation which if not avoided could result in death or WARNING serious injury Caution indicates a potentially hazardous situation which if not avoided may result in property CAUTION damage or minor injury or both NOTICE The information contained in this document is subject to change without notice A DANGER DO NOT SMOKE while using this device Keep all matches lit cigarettes or other sources of ignition out of the room in which this product is located and away from where oxygen is being delivered NO SMOKING signs should be prominently displayed Textiles and other materials that normally would not burn are easily ignited and burn with great intensity in
2. O ee Attention Consider Accompanying Documents Class II Double Insulated 230 VAC x 10 253 207 VAC 50 Hz Altitude Up to 1828 meters 6 000 ft above sea level without degradation of concentration levels Not recommended for use above 1828 meters 6 000 ft Oxygen Output All IRCSPO2AW IRCBPAW models Concentration Levels 87 to 95 6 at 0 5 to 5 L min NOTE Stated concentration levels achieved after initial warm up period approximately 30 minutes Maximum Outlet Pressure 34 5 kPa 3 45 kPa 5 psi 0 5 psi Part No 1150713 13 Perfecto Series SECTION 5 TYPICAL PRODUCT PARAMETERS Flow Range 0 5 to 5 L min maximum For flowrates less than L min we recommend the use of the Invacare Pediatric Flowmeter Accessory IRCPFI6AW Low Flow Alarm 0 L min to 0 5 L min Rapid Audible Alarm Beeping No Accessories Connected Average Power 300 W Consumption 280 W lt 3 L min Pressure Relief Mechanism 241 kPa 24 1 kPa 35 psi 3 5 psi Operational at Change in maximum 0 7 L min recommended flow when back pressure of 7kPa is applied Cabinet Outlet HEPA and Compressor Inlet Safety System Current overload or line surge shutdown High temperature compressor shutdown High Pressure Alarm w compressor shutdown Low Pressure Alarm w compressor shutdown Battery Free Power Loss Alarm SensO Oxygen System SensO Model Low Flow Alarm 38 1 cm 1 cm 15 3 8 i
3. Humidifier Humidifier Bottle w o Cap Bottle with Cap FIGURE 6 1 Filling the Humidifier Perfecto Series 18 Part No 1150713 SECTION 6 OPERATING INSTRUCTIONS 3 Insert a flathead screwdriver in the plate groove on the top edge of the filter access door and gently pry the filter access door off FIGURE 6 2 4 Pull up and remove the humidifier bottle adapter FIGURE 6 2 5 Replace the filter access door and install the air scoop onto the filter access door by inserting the six tabs on the air scoop into the six slots on the sides of the filter access door Humidifier Bottle Insert Flat b Adapter Screwdriver Filter Access Here Door Cabinet Filter Air Scoop Tabs FIGURE 6 2 Humidifer Bottle Adapter 6 Attach the humidifier bottle adapter to the humidifier bottle by turning the wing nut on the humidifier bottle counterclockwise until it is securely fastened See Detail A in FIGURE 6 3 7 Place the humidifier bottle adapter assembly in the humidifier compartment on the concentrator See Detail B in FIGURE 6 3 8 Attach oxygen tubing from the humidifier bottle adapter assembly to oxygen outlet connector on the concentrator See Detail B in FIGURE 6 3 9 Attach the cannula patient supply tubing to the humidifier bottle outlet See Detail B in FIGURE 6 3 Part No 1150713 19 Perfecto Series SECTION 6 OPERATING INSTRUCTIONS 10 After assembly en
4. 5 TYPICAL PRODUCT PARAMETERS IRCSPO2AW INDICATORS LABEL O PURITY INDICATOR LIGHTS SYMBOL LED SYSTEM OKAY GREEN Indicator Light O2 O over 85 2 O Between 73 3 YELLOW Indicator light to 85 2 A YELLOW Solid B YELLOW Flashing Sensor Failure Call a qualified technician SYSTEM FAILURE RED Indicator Light O Below 73 3 Continuous Audible Alarm Sieve GARD Compressor Shutdown IRCSPAW INDICATORS LABEL STATUS INDICATOR LIGHTS SYMBOL LED SYSTEM OKAY GREEN Indicator Light SYSTEM FAILURE RED Indicator light G Continuous Audible Alarm Sieve GARD Compressor Shutdown Call a qualified technician Perfecto Series 16 Part 1150713 SECTION 6 OPERATING INSTRUCTIONS SECTION 6 OPERATING INSTRUCTIONS Introduction Your oxygen concentrator is intended for individual use It is an electronically operated device that separates oxygen from room air It provides high concentration of oxygen directly to you through a nasal cannula Clinical studies have documented that oxygen concentrators are therapeutically equivalent to other types of oxygen delivery systems Your provider will show you how to use your oxygen concentrator He She should be contacted with any questions or problems regarding your oxygen concentrator This owner s manual will tell you about your concentrator and will serve as a reference as you use your concentrator Select
5. outlet Concentrator not operating Power Switch On Beeeeeeeep Beep Beep 3 Tripped circuit 3 Push reset circuit breaker breaker If problem recurs call Service Provider Alarm System Failure la Remove and clean cabinet Continuous l Unit overheating due filters to blocked air intake 2 Insufficient power at outlet 3 Internal repairs required Ib Move oxygen concentrator at least three inches away from walls draperies or furniture 2 DO NOT use extension cords Move to another electrical outlet or circuit 3 Call Service Provider Part No 1150713 27 Perfecto Series SECTION 8 TROUBLESHOOTING GUIDE SYMPTOM YELLOW or RED light Illuminated Alarm Continuous On RED light only Only applies to IRCSPO2AW models PROBABLE CAUSE Low oxygen purity SOLUTION Clean or Replace filters 2 Kinked or blocked tubing cannula or humidifier 2 Inspect for kinks or blockages Correct clean or replace item Once corrected turn power Off for 60 seconds and then turn power back On 3 Flowmeter set at 0 5 L min 3 Check flowmeter is set to 1 0 L min or more Refer to Typical Product Parameters on page 13 4 Unit overheating due to blocked air intake 4a Remove and clean cabinet filters 4b Move concentrator at least three inches from walls draperies and furniture 5 Insufficie
6. oxygen enriched air Failure to observe this warning can result in severe fire property damage and cause physical injury or death Perfecto Series 4 Part No 1150713 SPECIAL NOTES CAUTION Caution Statutory law can restrict this device to sale by oron order of a physician or any other practitioner licensed by the law of the governmental agency in which he she practices to use or order the use of this device Invacare recommends an alternate source of supplemental oxygen in the event of a power outage alarm condition or mechanical failure Consult your physician or equipment provider for the type of reserve system required This equipment is to be used as an oxygen supplement and is not considered life supporting or life sustaining Part No 1150713 5 Perfecto Series LABEL LOCATION LABEL LOCATION Perfecto FOR OPERATING INSTRUCTIONS or ALARMS See Operator s Manual A DANGER RISK OF FIRE NO SMOKING Keep ALL sources of ignition out of the room in which this product is located and away from areas where oxygen is being delivered Textiles oil and other combustiles are easily ignited and burn with great intensity in oxygen enriched air DANGER ELECTRIC SHOCK DO NOT remove cover Refer servicing to qualified service personnel A Flowrates below 0 5 L min will trigger Low Flow Alarm rapid audible alarm A Set Flowmeter at 2 5 L min or less when filling HomeFill cylinders HomeFill Co
7. these devices will be best for you They should also give you advice on the proper usage maintenance and cleaning NOTE Updated versions of this manual can be found at www invacare com Perfecto Series 2 Part No 1150713 TABLE OF CONTENTS ACCESSORIES e eee ee eee ees soeeececcccccceccccccoeoe 2 SPECIAL NOTES A LABEL LOCATION e e eee eecoooeeeecccccoocceccccs SECTION I GENERAL GUIDELINES 7 Radio Frequency Interference eerte 9 SECTION 2 FEATURES 66 SECTION 3 HANDLING 6611 Unpacking Rte pe Hear nh eee 11 INSPEC OR E Storage A Ea RE 11 SECTION 4 TECHNICAL DESCRIPTION 12 SECTION 5 TYPICAL PRODUCT PARAMETERS 13 SECTION 6 OPERATING INSTRUCTIONS 17 ENTRY 17 Selecta Location osito easement DAN 17 ad 18 mentor er cech dea oe 21 SensO Oxygen Purity Indicator Model IRCSPO2AW 22 Initial Startup of the Concentrator sss 22 Explanation of Indicator Lights Model IRC5PAW 23 Elapsed Time Meter Hour 23 SECTION 7 MAINTENANTCE
8. Operator s Manual Perfecto Series Oxygen Concentrators without SensO Oxygen Concentrators with SensO HomeFill Compatible Model IRC5PO2AW Model IRC5PAW Dealer This manual MUST be given to the end 0120 user User BEFORE using this product read this 4 manual and save for future reference For more information regarding Invacare products parts and services please visit www invacare com Yes you can WARNING A WARNING DO NOT use this product or any available optional equipment without first completely reading and understanding these instructions and any additional instructional material such as owner s manuals service manuals or instruction sheets supplied with this product or optional equipment If you are unable to understand the warnings cautions or instructions contact a healthcare professional dealer or technical personnel before attempting to use this equipment otherwise injury or damage may occur A ACCESSORIES WARNING Invacare products are specifically designed and manufactured for use in conjunction with Invacare accessories Accessories designed by other manufacturers have not been tested by Invacare and are not recommended for use with Invacare products ACCESSORIES There are many different types of humidifiers oxygen tubing cannulas and masks that can be used with this device You should contact your local home care provider for recommendations on which of
9. a Location A WARNING NEVER block the air openings of the product or place it on a soft surface such as a bed or couch where the air opening may be blocked Keep the openings free from lint hair and the like Keep unit at least 7 5 cm 3 in away from walls draperies furniture and the like You may select a room in your house where using your oxygen concentrator would be most convenient Your concentrator can be easily rolled from room to room on its casters Part No 1150713 17 Perfecto Series SECTION 6 OPERATING INSTRUCTIONS Your oxygen concentrator will perform best when operated under the conditions outlined in the Typical Product Parameters on page 13 Usage in environments other than those described may result in the need for increased equipment maintenance The air intake of the unit should be located in a well ventilated area to avoid airborne pollutants and or fumes Set Up 1 Plug in power cord to an electrical outlet 2 Connect Humidifier if so prescribed A WARNING DO NOT overfill humidifier DO NOT reverse the oxygen input and output connections Water from the humidifier bottle will travel through the cannula back to the patient NOTE For this procedure refer to FIGURE 6 1 FIGURE 6 2 on page 19 FIGURE 6 3 on page 20 1 Remove cap from bottle 2 Fill humidifier with distilled water to the level indicated by the manufacturer Replace the humidifier cap and securely tighten
10. be used only for filling oxygen cylinders with the REAR VIEW HomeFill home oxygen compressor The outlet fitting does not affect concentrator performance Cabinet Refer to the HomeFill Fiter owner s manual part number 1116444 for connection and operating instructions When not in use the plug provided with the concentrator should be inserted into the outlet fitting For more information about the HomeFill contact your Invacare dealer Power HF Outlet Cord Fitting Perfecto Series 10 Part 1150713 SECTION 3 HANDLING SECTION 3 HANDLING The concentrator should ALWAYS be Kept in the upright position to prevent cabinet damage while being transported If the concentrator is to be reshipped by common carrier additional cartons are available from Invacare Unpacking NOTE For this procedure refer to FIGURE 3 1 1 Check for any obvious _ damage to the carton or inn its contents If damage is XA evident notify the 1 Lo carrier or your local dealer 2 Remove allloose packing lt V from the carton DX d 3 Carefully remove all the components from the carton FIGURE 3 1 Unpacking NOTE Unless the oxygen concentrator is to be used IMMEDIATELY leave concentrator in its packaging for storage until use of the concentrator is required Inspection 1 Inspect examine exterior of the oxygen concentrator for nicks dents scratches or other damages Insp
11. blocked or kinked tubing or a defective humidifier bottle After rated flow is restored to more than 0 5 L min the LOW FLOW audible alarm will go off NOTE The use of some accessories such as the pediatric flowstand and the HomeFill compressor will deactivate the Low Flow Alarm 7 M 3 Flow Knob Flowmeter 0 5 L min Ball DT FIGURE 6 5 Flowrate Part No 1150713 21 Perfecto Series SECTION 6 OPERATING INSTRUCTIONS SensO Oxygen Purity Indicator Model IRCSPO2AW This feature monitors the purity of oxygen generated by the oxygen concentrator If purity falls below factory preset standards indicator lights on the control panel will illuminate Initial Startup of the Concentrator NOTE Concentrator may be used during the initial start warm up time approximately 30 min while waiting for the O purity to reach maximum When the unit is turned on the GREEN light will come on SYSTEM OK O greater than 85 After five minutes the oxygen sensor will be operating normally and will control the indicator lights depending on oxygen concentration values Explanation of Oxygen Purity Indicator Lights Model IRC5PO2AW NOTE For this procedure refer to FIGURE 6 6 on page 23 GREEN light O Normal Operation YELLOW light A Immediately call supplier You may continue to use the concentrator unless instructed otherwise by your supplier Be certain that backup oxygen is nearby RED light Q To
12. ect all components Storage 1 Store the repackaged oxygen concentrator in a dry area 2 DO NOT place objects on top of repackaged concentrator Part No 1150713 lI Perfecto Series SECTION 4 TECHNICAL DESCRIPTION SECTION 4 TECHNICAL DESCRIPTION The Invacare Perfecto concentrator is used by patients with respiratory disorders who require supplemental oxygen The device is not intended to sustain or support life The oxygen concentration level of the output gas ranges from 87 to 95 6 The oxygen is delivered to the patient through the use of a nasal cannula The Invacare Perfecto concentrator uses a molecular sieve and pressure swing adsorption methodology to produce the oxygen gas output Ambient air enters the device is filtered and then compressed This compressed air is then directed toward one of two nitrogen adsorbing sieve beds Concentrated oxygen exits the opposite end of the active sieve bed and is directed into an oxygen reservoir where it is delivered to the patient The Invacare Perfecto concentrator is capable of operation by the patient in a home environment or in an institutional environment Device operates at a nominal 230 VAC 50 Hertz supply Service information will be available upon request to qualified technical personnel only Perfecto Series 12 Part No 1150713 SECTION 5 TYPICAL PRODUCT PARAMETERS SECTION 5 TYPICAL PRODUCT PARAMETERS Alternating C t Type BF equipment
13. humidifier every day 2 Wash it in soapy water and rinse it with a solution of ten parts water and one part vinegar 3 Rinse thoroughly with hot water and refill with distilled water to the level shown on the humidifier Part No 1150713 25 Perfecto Series SECTION 7 MAINTENANCE PREVENTIVE MAINTENANCE RECORD Model No IRC Serial No On Each Inspection Record Date Of Service Record Elapsed Hours on Hour Meter Clean Cabinet Filters Check Prescribed L min Flow Rate Annually During Preventive Maintenance Schedule Or Between Patients Clean Replace Cabinet Filters Check outlet HEPA Filter Check Compressor Inlet Concentration SensO models Check Power Loss Alarm As Required Clean Heat Exchanger Replace Exhaust Muffler Every 180 Days for IRCSPAW Check oxygen Concentration NOTE Refer to Preventive Maintenance section of Service Manual NOTE Model not available within EU countries Perfecto Series 26 Part No 1150713 SECTION 8 TROUBLESHOOTING GUIDE SECTION 8 TROUBLESHOOTING GUIDE SYMPTOM Alarm PROBABLE CAUSE Main Power Loss SOLUTION Short beeps long pause 1 Power cord not plugged in Insert plug into outlet Concentrator not operating power switch On 2 No power at outlet 2 Inspect house circuit breakers or fuses If problem recurs use a different
14. ith other oxygen concentrators or oxygen therapy devices Perfecto Series 8 Part No 1150713 SECTION GENERAL GUIDELINES Radio Frequency Interference This equipment has been tested and found to comply with EMC limits specified by IEC EN 60601 1 2 These limits are designed to provide a reasonable protection against electromagnetic interference in a typical medical installation Other devices may experience interference from even the low levels of electromagnetic emissions permitted by the above standards To determine if the emissions from the Perfecto is causing the interference turn the Perfecto Off If the interference with the other device s stops then the Perfecto is causing the interference In such rare cases interference may be reduced or corrected by one of the following measures e Reposition relocate or increase the separation between the equipment e Connect the equipment into an outlet on a circuit different from that to which the other device s is connected Avoid creation of any spark near medical oxygen equipment This includes sparks from static electricity created by any type of friction Part No 1150713 9 Perfecto Series SECTION 2 FEATURES SECTION 2 FEATURES Oxygen Outlet Lights Fault and Power P Indicator Lights Flowmeter Circuit Breaker Elapsed Time Power Switch Meter i 8 px y NOTE This outlet fitting is to
15. mpatible Perfecto Series Serial Number Label is located on the resonator intake assembly Specification Label is located on the back near the bottom PA A Part No 1150713 SECTION GENERAL GUIDELINES SECTION GENERAL GUIDELINES In order to ensure the safe installation assembly and operation of the Perfecto concentrator these instructions MUST be followed A WARNING This section contains important information for the safe operation and use of this product A DANGER Risk of electric shock DO NOT disassemble Refer servicing to qualified service personnel No user serviceable parts TO REDUCE THE RISK OF BURNS ELECTROCUTION FIRE OR INJURY TO PERSONS Avoid using while bathing If continuous usage is required by the physician s prescription the concentrator MUST be located in another room at least 2 5 m 7 ft from the bath DO NOT come in contact with the concentrator while wet DO NOT place or store product where it can drop into water or other liquid DO NOT reach for product that has fallen into water Unplug IMMEDIATELY If the concentrator has a damaged cord or plug if it is not working properly if it has been dropped or damaged or dropped into water call qualified technician for examination and repair A spontaneous and violent ignition may occur if oil grease or greasy substances come in contact with oxygen under pressure These substances MUST be kept a
16. n 58 4 cm cm 23 in 3 8 in 30 5 cm cm I2 in 3 8 in 20 5 kg kg 45 165 2 Ibs Shipping Weight 22 7 kg kg 50 Ibs 2 Ibs Operating Ambient 10 C 35 C 50 F 95 F at 20 60 relative Temperature humidity Exhaust Cooling Air Less than Ambient 19 C 35 F Temperature Oxygen Output Less than Ambient 3 C 6 F Temperature Cabinet Impact Resistant flame retardant plastic cabinet that conforms to UL 94 V0 Perfecto Series 14 Part No 1150713 SECTION 5 TYPICAL PRODUCT PARAMETERS No AP APG Not suitable for use in the presence of a flammable anaesthetic mixture Standards and Regulatory IRCBPO2AW IRCSPAW Listing IEC EN61000 3 2 IEC EN61000 3 3 IEC EN 60601 1 Al A2 IEC EN 60601 1 2 IRC5PO2AW Models Only 1508359 MDD 93 42 EEC Annex and IX CE marked model 5 2 Placement No closer than 7 5 cm 3 in from any wall furniture draperies or furniture to assure sufficient air flow Avoid deep pile carpets and heaters radiators or hot air registers Floor location only Tubing 2 m 7 ft cannula with a maximum 15 m 50 ft of Crush Proof Tubing DO NOT pinch Time of Operation Up to 24 hours per day Recommended Storage 29 C to 65 C 20 F to 150 F at 15 95 and Shipping Temperature relative humidity Environment Smoke and soot free No confined spaces Example No closets Part No 1150713 15 Perfecto Series SECTION
17. nt power at outlet 5 DO NOT use extension cords Move to another electrical outlet or circuit 6 Internal repairs required 6 Call Service Provider GREEN light with YELLOW Light Flashing Internal repairs required Call Service Provider Alarm Rapid Beep Beep Beep Beep LOW FLOW ALARM la Kinked or blocked tubing cannula or humidifier Ib Flowmeter set at 0 5 L min la Inspect for kinks or blockages Correct clean or replace item Once corrected turn power Off for 60 seconds and then turn power back ON Ib Flowrates less than L min are not recommend Perfecto Series 28 Part No 1150713 SECTION 9 OPTIONAL ACCESSORIES SECTION 9 OPTIONAL ACCESSORIES The following optional accessories are also available e Standard Adult Nasal Cannula 2 1 m 7 feet M3120 e Standard Humidifier M5120 e PreciseRX Pediatric Humidifier Flowmeter Accessory IRCPF16AW e HomeFill home oxygen compressor IOH200AW Part No 1150713 29 Perfecto Series RECYCLING INFORMATION RECYCLING INFORMATION This product has been supplied from an environmentally aware manufacturer that complies with the Waste Electrical and Electronic Equipment WEEE Directive 2002 96 CE This product may contain substances that could be harmful to the environment if disposed of in places landfills that are not appropriate according to legislation The crossed out
18. sure that oxygen is flowing through the cannula DETAIL A DETAIL Cbyygen amp Humidifier Outlet Oxygen Bottle Connector Tubing Adapter Ld Humidifier Humidifier Bottle Bottle Outlet Himin iS it Bottle ENS Humidi FIGURE 6 3 Humidifier Compartment Power Switch NOTE Forthisprocedure referto FIGURE 6 4 1 Press power switch to On position All the panel lights and the audible alarm will come on for one second indicating that the unit is functioning properly FIGURE 6 4 Power Switch Perfecto Series 20 Part No 1150713 SECTION 6 OPERATING INSTRUCTIONS Flowrate NOTE For this procedure refer to FIGURE 6 5 NOTE Take care not to set the flow above RED ring An oxygen flow greater than 5 L min will decrease the oxygen concentration 1 Turn the flowrate knob to the setting prescribed by your physician or therapist A WARNING DO NOT change the L min setting on the flowmeter unless a change has been prescribed by your physician or therapist NOTE To properly read the flowmeter locate the prescribed flowrate line on the flowmeter Next turn the flow knob until the ball rises to the line Now center the ball on the L min line prescribed 2 If the flowrate on the flowmeter ever falls below 0 5 L min for more than about one minute the LOW FLOW alarm will be triggered This is a rapid beeping of the audible alarm Check your tubing or accessories for
19. t NOTE At a minimum preventive maintenance MUST be performed according to the maintenance record guidelines In places with high dust or soot levels maintenance may need to be performed more often Refer to Preventive Maintenance Record on page 26 Routine Maintenance The following routine maintenance should be performed more frequently See recommended intervals in each section and Preventive Maintenance Record on page 26 Cleaning the Cabinet Filter CAUTION DO NOT operate the concentrator without the filter installed NOTE For this procedure refer to FIGURE 7 1 on page 25 NOTE There is one cabinet filter located on the back of the cabinet 1 Inspect the filter monthly Remove the filter and clean as needed NOTE Environmental conditions that may require more frequent inspection and cleaning of the filter include but are not limited to high dust air pollutants etc Perfecto Series 24 Part No 1150713 SECTION 7 MAINTENANCE 2 Clean the cabinet filter with a vacuum cleaner or wash in warm soapy water and rinse thoroughly 3 Dry the filter thoroughly before reinstallation FIGURE 7 1 Cleaning the Cabinet Filter Cleaning the Cabinet 1 Clean the cabinet with a mild household cleaner and non abrasive cloth or sponge Cleaning the Humidifier NOTE To clean the oxygen humidifier follow the instructions provided by the manufacturer If none are provided follow these STEPS 1 Clean the
20. tal unit shutdown Immediately switch to backup oxygen supply and call supplier GREEN light with YELLOW light flashing Immediately call supplier Oxygen sensor malfunctioning you may continue to use the concentrator Perfecto Series 22 Part No 1150713 SECTION 6 OPERATING INSTRUCTIONS OXYGEN PURITY Auto Shut Down Use Backup a N Below VA TT Call Supplier o YELLOW Normal Normal Oo GREEN FIGURE 6 6 Explanation of Oxygen Purity Indicator Lights Model IRC5PO2AW Explanation of Indicator Lights Model IRCSPAW NOTE For this procedure refer to FIGURE 6 7 IRCSPAW Model Indicator Light Explanation RED light Q Total Unit Shut Down Immediately switch to a back up oxygen supply and call supplier GREEN light I O On Off System okay Auto Shut Down Use Backup a 4 RED N TT Call Supplier Normal e 4 GREEN FIGURE 6 7 Explanation of Indicator Lights Model IRCBPAW Elapsed Time Meter Hour Meter The hour meter displays the cumulative number of hours the unit has operated Part No 1150713 23 Perfecto Series SECTION 7 MAINTENANCE SECTION 7 MAINTENANCE A WARNING The Invacare concentrators are specifically designed to minimize routine preventive maintenance Only qualified personnel should perform preventive maintenance on the concentrator Unplug the concentrator when cleaning To avoid electrical shock DO NOT remove cabine
21. way from the oxygen concentrator tubing and connections and all other oxygen equipment DO NOT use any lubricants unless recommended by Invacare Part No 1150713 7 Perfecto Series SECTION GENERAL GUIDELINES Operating Information For optimum performance Invacare recommends that each concentrator be on and running for a minimum of 30 minutes ata time Shorter periods of operation may reduce maximum product life Keep the oxygen tubing cord and unit out from under such items as blankets bed coverings chair cushions clothing and away from heated or hot surfaces including space heaters stoves and similar electrical appliances DO NOT move or relocate concentrator by pulling on the power cord NEVER drop or insert any object or liquid into any opening Invacare recommends that Crush Proof oxygen tubing be used with this product and not exceed 15 2 m 50 ft in length There are no user serviceable parts This does not include normal maintenance items See maintenance section for user maintenance items A product should NEVER be left unattended when plugged in Make sure the Perfecto is Off when not in use Close supervision is necessary when this product is used near children or physically challenged individuals Additional monitoring or attention may be required for patients using this device who are unable to hear or see alarms or communicate discomfort DO NOT connect the concentrator in parallel or series w
22. wheelie bin symbol is placed on this product to encourage you to recycle wherever possible Sas Please be environmentally responsible and recycle this product through your recycling facility at its end of life Perfecto Series 30 Part No 1150713 LIMITED WARRANTY LIMITED WARRANTY United Kingdom Invacare Ltd South Road Bridgend Industrial Estate UK Bridgend CF31 3PY Tel 44 0 1656 664 321 Fax 44 0 1656 667 532 UK invacare com Ireland Invacare Ireland Ltd Unit 5 Seatown Business Campus Seatown Road Swords County Dublin Ireland Tel 353 1 810 7084 Fax 353 1 810 7085 ireland invacare com Part No 1150713 3l Perfecto Series D Yes you can Invacare Corporation www invacare com USA One Invacare Way Elyria Ohio USA 44036 2125 440 329 6000 800 333 6900 Technical Services 440 329 6593 800 832 4707 EU Representative Invacare International Sarl Route de Cit Ouest 2 1196 Gland Switzerland Tel 41 22 354 60 10 Fax 41 22 354 60 11 Part No 1150713 Invacare Corporation 2101 E Lake Mary Blvd Sanford FL 32773 800 832 4707 All rights reserved Trademarks are identified by and All trademarks are owned by or licensed to Invacare Corporation or its subsidiaries unless otherwise noted 9 2009 Invacare Corporation Rev C 4 09 0120
Download Pdf Manuals
Related Search
Related Contents
PR IMERGY ES200 取扱説明書への追記と修正 HT PV Catalog - Clean Energy Brands Manuel d`utilisation OCB Fujifilm Superia X-tra 800 135/36 Manual do Sistema SAMMED - 2014 IDEC MicroSmart User Manual Lavorwash Venti X D-Link DCS-6314BS surveillance camera Dossier De presse NoVeMBre 2011 Copyright © All rights reserved.