Home
        Fluke PM3370B Hunting Equipment User Manual
         Contents
1.                                                                                                                                                                                                                         576728    DELAYED TIMEBASE TRIGGER    Using the Delayed Timebase  Del   d  to select  capture  and display a small portion  of the Main Timebase display results in a magnification of the intensified part   Depending on the ratio between the MainTB and Del   dTB settings  this  magnification can be very significant     If the input signal contains jitter or any other form of timing instability  this jitter will  be magnified in the same ratio  This can be so much  that the display becomes  unusable  To address such a problem  the Delayed Timebase  Del dTB  can be    made to trigger on the input signal after the delay time has passed   Start with the STANDARD SETUP       Connectthe Probe Adjust signal to the                                                                                                                               Channel 1 input  pcm    Simultaneously press the STATUS and                                TEXT OFF keys       Switch channel 1 to AC coupling      mr    Shiftthe trace to the upper half of the screen  Wo    Press the Delayed Timebase  DTB  key to    Nm  enter the DELAYED TIMEBASE menu               Turn on the Delayed Timebase     4 66 HOW TO USE THE INSTRUMENT       The third softkey in the menu is labeled    starts trig   d        Star
2.                                                                                                               ST6481      Turn the TRACK control counterclockwise so that the Average factor is  8 again     HOW TO USE THE INSTRUMENT 4 33       ACQUIRE ENVELOPE  The ENVELOPE mode records the minimum and the maximum of the signal over  a number of acquisitions       Press the ACQUIRE key     Press the ENVELOPE    softkey to turn it on                       jme ald ha a ENVELOPE pamm paraa           on                                                      hy                mei                                                                                             56480           AVERAGE mode and the ENVELOPE modes are mutually exclusive   Observe that AVERAGE is automatically switched off  The text           appears       the bottom of the screen to indicate that the function is active    The minimum and maximum of the input signal is stored and becomes visible on  the screen  This process continues until it is turned off by the user       Turn the POS control of Ch1     Observe that every time you change the settings of the scope  like trace POS  the  ENVELOPE process automatically starts again       Press the ENVELOPE    softkey to turn off the ENVELOPE mode     ACQUIRE PEAK DETECTION  PEAK DETection automatically catches peak values  up to glitches of 10 ns width   of the input signal during a single acquisition      Press the    PEAK DET    softkey to turn it on     The tex
3.                                                                                                    POSITION A CURSOR  ON NEGATIVE SLOPE  OF PULSE          i    O  us    FUNCTION REFERENCE       CURSORS READOUT    Three readouts can be selected     AV  Gives the voltage difference  between the cursors    V1 V2  Gives the absolute voltage  with respect to ground for each  cursor  V1 and V2 have to be  selected separately    AV ratio  The readout is a  percentage that can be reset  using the  A2100      softkey  This  can be used for amplitude  comparisons  The figure shows an  example  the percentage of  overshoot compared with 100    pulse amplitude is determined     Key sequence     AT 1       AT ratio  CURSORS        ph T trg  BOTH  AM  V1 amp V2  AV ratio    AV 100   AT 100   AT 360      cursor  track  yes no           ie       576846  9312    Remote commands     VOLT       ch 1   AV 100           TRACK             O               0         POSITION CURSORS ON TO    PRESS AV 100                             OF SIGNAL             ch J    4    5               O                                                            POSITION A CURSOR ON TOP OF OVERSHOOT    CPL  QM  command to query measured values     Refer to Chapter 6 for full details     MAT4211    Softkey pair to switch between time readout AT   1 AT  AT ratio  phase  and T trg     Toggle function softkey to switch between  voltage readout AV  V1  V2 or AV ratio     Softkey to reset AV ratio to 100   AT ratio to  100  o
4.                                                                 cea  4     boo 2 9 omg     S M MM                                                                           ST6680  9312  Figure 3 4                         switching setup  Step 1 Press AUTOSET  The scope performs an AUTOSET in analog mode   Step 2 Press the ANALOG key to change over to the digital mode  Check that    the picture is identical to the one in the analog mode  The text DIGITAL  MODE    is displayed briefly at the bottom of the screen        3 8 GETTING STARTED   Step 3 Press AUTOSET again  This time the scope performs the autoset in  digital mode    Step 4 Press the RUN STOP key and observe that the trace is frozen and  stays on screen even after removing the probe    Step 5 Press the RUN STOP key to display the actual input signal again   Reconnect the probe to display the Probe Adjust signal again    Step 6 Press the ANALOG button once again to return to the analog mode  In    the bottom of the screen  the text    ANALOG MODE is briefly displayed     3 6 VERTICAL SETUP    This section deals with setting of the input circuits of the four channels  The main  adjustments are AMPLitude  POSition  and the channel input coupling selection  for GND  DC  and AC              VERI MENU AVERAGE                                                                           8 RANGECHI  CH2 RANGE  as   4L J     a Se os                                                                   GND ON                     Figur
5.                                                     576704    GETTING STARTED 3 5       Step 4 To prevent measurement errors  check the pulse response before any  measurement  If the pulse shows overshoot or undershoot  you can  correct this by using the trimmer in the probe   s body  Chapter 4  describes how to adjust the pulse response                                                                                                575962                                                                                                       In most cases  using AUTO SETUP is sufficient for a good initial display of the  signal s   After the initial AUTOSET  and to optimize the signal for a more detailed  view  continue with the paragraphs below     NOTE  Ifyou get lost when adjusting your instrument  just press AUTOSET     3 6    GETTING STARTED       3 5 MODE SWITCHING BETWEEN ANALOG AND  DIGITAL OPERATING MODES    You can use the yellow ANALOG key to switch from the analog mode to the digital  mode and back at any time  The signal acquisition and display functions of both  operating modes are very similar  However  the nature of the signals you are  using may determine which operating mode you prefer to use  For more  information  refer to the following table     SIGNAL CRITERIA    Repetitive signals of  30 Hz and higher    Repetitive signals  below 30 Hz    Single events    Repetitive signals that are    amplitude modulated    Repetitive signals that  are modulated in frequen
6.                                  pds stop bits    pds handshake Y CR            baud     75  110  150  300  600  1200  2400  4800  9600  19200   38400     parity     O EorN    lt data bits gt    7or8    lt stopbits gt           lt handshake gt    XONXOFF _ to enable Xon Xoff handshake  HWL to enable hardwareline handshake    If left out  both handshake methods are disabled     THE CPL PROTOCOL                            Response    acknowledge CR  Note   Example   100 PARS  2400 N 8 1   110 CTL    XONXOFF   120 PRINT  130 GOSUB 1000  140 CLOSE  1  150 WT TIMER  160 WHILE  TIMER WT   170 COMMS    180 OPEN COMMS AS  1       approx  0 5 sec after an  lt acknowledge gt    0 is received  the    communication parameters are changed to the new values                                       lt     1  PC  PAR  CTL        5               END      comm parameters      XONXOFF handshake    Send command     Sync on acknowledge    Close the COM Port       Wait 0 5 second     COM1   PARS   CS0 DS0 CDO       Reprogram COM Port    6 18 THE CPL PROTOCOL       PROGRAM SETUP PS    Purpose     To configure the oscilloscope using compact setup strings    This Program Setup  PS  command and the Query Setup  QS  command can be  used together to restore and retrieve a complete setup or partial of the  oscilloscope     The format of the programming strings must be the same as the format of the  received setup strings  Each setup string describes a  node  in the oscilloscope  setup    It is possible to s
7.                                 5 5  5 PRINTING AND PLOTTING             5 55        ALT CHOP         ehe 5 7  4 8 ANALOG AND DIGITAL MODES          4 9  5                                            5 8  5 INPUT ATTENUATOR                  5 39  5   CALIBRATION AUTOCAL               5 16  5                                              5 9  5 INPUT ATTENUATOR                  5 39  5   AUTORANGE                         5 9  5                                               5 46  5 AUTORANGE                         5 9  4 16 AUTOSET AND SETUP UTILITIES       4 70  5  VAUTOSET za 284 ed eG kde ee fice      5 10  5   AUTOSET SEQUENCE                5 11  5   AUTOSET USERPROG                5 12  5  AVERAGE    or             5 14  5  BANDWIDTHLIMITER                 5 15  bo  CURSORS seed ick  weed ELO 5 19  5   CALIBRATION AUTOCAL               5 16  5   CHANNEL TRACE SELECTION          5 17  Reference manual   5   CONFIDENCE CHECK                 5 18  5  ALT CHOP oui                        5 7       1 4 FUNCTION INDEX  Entry Refer to chapter function Page  Clock 5   PRINTING AND PLOTTING             5 55  Command switch 5   TOUCH  HOLD  amp  MEASURE     MODE    5 78  Common mode 5 ADD INVERT SUBTRACT               5 4  Confidence 5 CONFIDENCE                           5 18  Cursor limited  measurements 5 MEASURE MENU                     5 47  5  TIMEMEASUREMENS                 5 76  5 VOLT MEASUREMENTS                5 93  Connections OM emnes uendere TE DK e XI  Cursors 5  CURSO
8.                              MEASURE SELECT       MEAS n  MEAS 1 3 volt J  pkpk  gt  time  ch2 E delay    on off de A  1 rms J  MEAS 2 5        1  rise  gt   max        2    kpk E  on off high v  CURSOR   chi J  LIMIT amp  ch2     STATIST              RETURN 4  CURSOR  LIMIT amp   STATIST  CURSOR 1  LIMIT  yes noz  STATIST  on off J  RETURN 4                            RETURN 4       SELECT  MEAS n  volt  pall time  delay 5  f                 chi        h2            J  FNY  A chi J      cha     ext J  RETURN 4                   5  7673         1         DEL   D TB  menu structure       Appendix F          DEL D TB  menu structure          TB MODE   EVENTS  ge DELAY  cx  DTB          Sen     TRIGGER   DELAY TB FF 1y  MAIN TB  on off  starts  trig d TRACK       ST6564  9303    SAVE RECALL MENU STRUCTURE       Appendix G  SAVE RECALL menu structure    SAVE    RECALL                                                 1  OPTIONAL          SAVE ACQ   ees CLEAR CLEAR          TO PROTECT                 MEMORY  MEMORY MEMORY CONFIRM CONFIRM  o mi   TRACK omi TRACK      em  D       m2 oO  e m3 5 em3       dva PROTECT    J on off J J  clear   yes yes c    copr   clear   ARE YOU   OVERRULE    5 clear SURE  3j PROTECT  3  J all J J  CLEAR amp         PROTECT      1          RETURN 4 no no        RECALL RECALL COPY  REGISTER REGISTER MEMORY  MEMORY  MEMORY   TRACK TRACK         2            emit O O emt       mi2 3 o m2 TRACK  DISPLAY DISPLAY            off              CLEAR CLEAR    DISPL
9.                           ST7664E A                          KEYS FOR AUTO RANGING VERTICAL AUTO RANGING EXTERNAL TRIGGER  MENU OPERATION ATTENUATOR CONTROLS  TIME BASE INPUT CONTROLS    Figure 3 1 Front panel layout    Note that the front panel shown is that with the most functions  Differences are  explained in Section 4 1  For this getting started procedure  only CH1 and CH2  are used  These are identical for all models     3 2    GETTING STARTED       Typical operation of your instrument will be     3 2    Switching on the instrument  Initial standard setup   Screen controls   Auto setup   Analog Digital mode switching  Vertical setup   Timebase setup   Magnify  Expand    Direct trigger setup   Pretrigger view   More advanced features  Cursor operation   More advanced trigger functions  More signal detail with the DTB  Trace storage     see Section 3 2    see Section 3 2    see Section 3 3    see Section 3 4    see Section 3 5    see Section 3 6    see Section 3 7    see Section 3 8    see Section 3 9    see Section 3 10   see Section 3 11   see Section 3 12   see Section 3 13   see Section 3 14   see Section 3 15             SWITCHING ON THE INSTRUMENT    Connect the power cord and set the front panel power switch to ON  For any line  source between 100V to 240V nominal  50 400 Hz  the instrument automatically  turns on  After performing the built in power up routine  the instrument is  immediately ready for use  The instrument   s settings will be identical to those  wh
10.                 im      2  500       CHP MTB 200    hir        576970      Press the ANALOG key to switch to the Digital mode     Observe that switching to the digital mode turns off the third trace  The  1 2   indication disappears       Press the CH1 CH2 key once again     Observe that in the digital mode the  CH1  CH2  trace can be used as well   Simultaneous display of both channels together with the sum of both channels is  not possible in the digital mode     CH1 CH2            i m      IcHt  sodmv   14   CH2  50dmV M  B 00   hir                                                                                                                             ST6720      Press the ANALOG key to return to the analog mode     HOW TO USE THE INSTRUMENT 4 19       VERTICAL INVERT    The INVERT function in Channel 2 can be used to make it easier to do out of   phase signal comparisons  The most common use of the INVERT function is to  obtain the display or make the acquisition of the voltage difference between two  channels  This is done by displaying or capturing the sum of Ch1 and Ch2 as  follows   This is referred to as the differential mode        Using two probes  connect the Probe Adjust signal to Ch1 and Ch2      Press AUTOSET      Both Ch1 and Ch2 are now displayed      Adjust POS and AMPL to get a display as illustrated on the left below     Press the CH1 CH2 key      Press the INV key of channel 2     Since channel 2 is inverted before it is added to channel 1  the result 
11.              4 71  4 17 OTHER FEATURES   lt  o peri ei Rex eb        4 75   5 FUNCTION REFERENCE                               5 1   6 THE CPL PROTOCOL                                   6 1  6 1 INTRODUCTION                                         6 1  6 2 EXAMPLE PROGRAM FRAME                             6 3  6 3 COMMANDS IN FUNCTIONAL ORDER                       6 4  6 4 COMMANDS IN ALPHABETICAL ORDER                    6 5  6 5 COMMAND REFERENCE                                  6 6  6 6 ACKNOWLEDGE                                       6 47  6 7 STATUS re i          UN                   6 48  6 8 SETUP taittaa Rom Meee aes hate Eg a bode aeta 6 50       CONTENTS XI  Appendix A ACQUIRE menu structure                        A 1  Appendix B CURSORS menu structure                       B 1  Appendix C DISPLAY menu structured                      C 11  Appendix D MATHEMATICS menu structure                   D 1  Appendix E MEASURE menu structure                       E 1  Appendix F         DEL D TB  menu structure                  F 1  Appendix G SAVE RECALL menu structure                   G 1  Appendix H SETUPS menu structure                         H 1  Appendix J TB MODE menu structure                         J 1  Appendix K TRIGGER menu structure                        K 1  Appendix L UTILITY menu structure                          L 1  Appendix M VERTICAL menu structure                       M 1  Appendix N RS 232 Cable configurations                     N 1             
12.             on acknowledge    Read text characters    Print setup text    THE CPL PROTOCOL       QUERY WAVEFORM    Purpose     To obtain a complete waveform from the oscilloscope     You can use the PW command to send the data received with the QW query back    to the oscilloscope     The option   A  can be used to recall the date and time in the customized format   When compatibility with initial versions of these oscilloscopes is required this          option should be omitted     Note         number of data bits of the RS 232 interface must be 8  not 7      otherwise an execution error will be given                                                                                               Command    QW  pds wave nr      CR  Response 1   acknowledge CR  Response 2   admin     count d sample checksum                                  Note  A trace transfer can be canceled by sending the ESC character     27 decimal         6 36    THE CPL PROTOCOL       wave_nr    admin    count    sample    The oscilloscope waveform source    001   004 for CH1   CH4 00e for EXT TRIG   011   084 for m1 1   m8 4 01   08   for m1 e   m8 e   091   504 for m9 1   m50 4  Extended Memory only    09e   50e for m9 e   m50 e  Extended Memory only     xx e for 2 ch  versions only  xx3 and xx4 for 4 ch  versions only     16 parameters  separated by a        PARAMETER          EXAMPLE NOTES       trace name string m4 1   Y unit string V   X unit string S   Y zero number 3 Y position  reverse   X zero 
13.            emt                                     P 1  FUNCTION INDEX  see Chapter 5                         1 1  INDEX                              E l 3    XII       FRONT VIEW                                                                                                                                                                                                                                                                                                                                                                                                                                  ue AUTOSET CAL SETUPS UTILITY ANALOG ACQUIRE SAVE RECALL MEASURE MATH DISPLAY HARD COPY    JOO COQOOOOOOO  STATUS  Rack CURSORS 4 HOLD OFF TRIGGER MAGNIFY 4                wo Sv al O  0 2  2  0  CO caneomessse            HER LO     C  egg  eee           sTIME DIV ns TB MODE AR  ROTATION            O  1 CB    VERT MENU GIHOR      Pos ePos          POS  eat                                        AUTI     AUTO     UT AUTO        mu pes AMPL PANGE INV AMPL RANGECH3 CH4 AMEL PANGE INV     TEXTOFF Sa B na mi a AE ans     j      Ed D Et  POWER UO  i     9          no        1     25     Oz   Oz  IMO 25pF       5  7721    IEEE 488 2  SOCKET RS 232 MAINS FUSE BATTERY LOCATION OF   OPTIONAL  SOCKET INPUT HOLDER COMPARTMENT TYPE PLATE                                    E gt                           L3                 E gt  co              es Ce                 SaS         E E   E E              E E E
14.          programming the AUTOSET key to be used as the    recall next setup    key   This selection can be done in the UTILITY  gt  gt  AUTOSET menu       Byusing a probe with a COMMAND switch  Each instrument is delivered with  two such probes  as standard accessory  The appropriate mode for this probe  command can be selected in the UTILITY      PROBE menu     To return to the    normal    oscilloscope mode  recall the STANDARD SETUP by  simultaneously pressing the STATUS and TEXT OFF key     FUNCTION REFERENCE 5   69       STANDARD SETUP   FRONT PANEL RESET    Description     A factory programmed set of default settings is available to put the instrument in  a defined state  The default settings  std  are reached in the menu under the  SETUPS menu key  Another method to perform a front panel reset is by pressing  the STATUS and TEXT OFF keys simultaneously     The TRACK control is used to Select  std  and the memories s1     s10  refer to  SETUPS   The settings saved in  std  become the actual settings when the  recall   softkey is pressed     The most important  std  settings are      Analog mode      AUTOSET USERPROG is set to  off       Setting readout is switched on      Ground indicator is set to    on         Usertext is set to    off         Cursors are    off         Input is dc coupled and input impedance is 1 MQ      Input sensitivity is set to 50mV and VAR off      CH1 is on  CH2  CH3         and EXT TRIG are off      Vertical POS is set to center of screen     
15.         displayed signal        be expanded horizontally so that more  DID signal detail becomes visible  In the analog mode  a  10  E magnification is possible and the entire sweep length can be made  visible by turning the X POS control     Pressthe right MAGNIFY key   The text MGN  appears in the bottom of the screen to indicate that the function is    active  Observe that the timebase indication is changed from 1 00 ms div to  100 us div     EME   o5    LEE           200mV MTB 1004  hi   CH1  200mV         _ 1004s ich1f                                                                                                                                                    576504      Press the ANALOG key to set the scope      the digital mode     Press the right MAGNIFY key     Observe that the signal expands in  1   2   4     steps to a maximum of 32 times   This magnification factor is displayed briefly on the screen       Setthe MAGNIFY to  4     HOW TO USE THE INSTRUMENT 4 25       TIMEBASE X POS    X POS    With X POS the displayed signal is shifted horizontally across the  display       Turn the X POS control clockwise               5     N          4199                                                                                                                                                             Observe that a bar graph              is displayed  The block on the bar graph  shows which part of the digital trace is displayed as expanded  This block on the  bar gr
16.       zug              ch1  ch2 pair of softkeys  When       more than one channel is on  the 3  desired channel for voltage TRACK             readout must be selected with the      ch1  ch2 pair of softkeys  In digital      mode stored traces          also be used for voltage measurements  The figure shows the required settings   The cursor positioning with the TRACK and A controls is also shown  The  example shows how peak peak voltage measurement is done                                                        FUNCTION REFERENCE 5 21                                  CURSORS BOTH   In this mode  both voltage and CURSORS   time cursors are active  The o     C3   TRACK and A controls operate as  ua  c3   in VOLT or TIME mode  as            selected with the CONTROL key        ch2       Qr   i   Cj  TRAGK     READOUT  C    O   rr       Key sequence     on off    CURSORS J           Remote commands                                              TRACK       570554  E        Toggle function softkey to switch between cursors on and off     Toggle function softkey to switch between volt  time or both  cursors     Softkey pair to select CH1 or CH2 for measurements  Selects in  digital mode also stored traces     Toggle function softkey to switch between volt and time cursors  for positioning  This selection is usable only in the    both    cursor  mode     Control to shift both cursors simultaneously  or to shift the  track cursor only  Selection in the CURSORS READOUT  menu determines 
17.      CURSORS READOUT AT RATIO PHASE    First  the reference distance between the two cursors is set  This then is set to  100    360    by pressing the AT 100    360    softkey  Changing the distance  between the cursors now results in a        reading of the reference       Press the softkey next to AT until AT ratio  phase  is intensified     Set the time cursors to be exactly one signal period apart     Press the softkey next to AT 100   360        In the cursor readout line of the screen  the value for AT is now    AT 100       AT 360      Observe that the 100   360    reference does not change when the TRACK  control is turned           50                 50                                                                                                                            86733    TRACK A TRACK A        5   0                            Adjust the cursors to a distance of half a signal period     Now the AT in the readout area will read        50      AT 180     This is the way to use the cursors for duty cycle measurements     HOW TO USE THE INSTRUMENT 4 49       4 11 MEASUREMENT FUNCTIONS    When operating in the analog mode  you can use the cursors to measure  amplitude and time data     When operating in the digital mode  the scope has an extensive set of fully  automated amplitude and time measurement functions  You can select two  measurements to be performed simultaneously  These measurements are  updated automatically  so that when signals change  the measu
18.      Response 2        count    node CR                                               4  more nodes to follow     lt count gt  number of strings to follow   lt node gt  string of decimal characters representing a setup node in  the oscilloscope  e g  16   10 hex      6 32 THE CPL PROTOCOL                Example    Complete setup query   90 DIM              25    Reserve space  100 PRINT  1  QS      Send command  110 GOSUB 1000   Sync on acknowledge  120 INPUT  1 N   Read number of node strings  130 FOR    1         140 INPUT   41 SETUPS I    Read all node strings  150 PRINT  N    I SETUPS I   160 NEXT I       example response   PM3394   11 0106hlhl      OPAT AN  0306  1  1       CR         These SETUP  array elements can be stored and sent back later to the  oscilloscope with the  PS  command      Optional setup query to read only one setup node        100 PRINT  1  0S  4   Query setup node 4   110 GOSUB 1000   Sync on acknowledge   120 INPUT  1 N   Read number of setup nodes  130 INPUT  1 SETUPS   Read the setup node   140 PRINT SETUPS   Print the setup string       THE CPL PROTOCOL 6 33       QUERY TEXT QT    Purpose   To query text from an oscilloscope     If the S parameter is specified  setup text is returned  The text will be queried from  one of the setup registers of the oscilloscope  The parameter n specifies the setup  register  A text field of 22 characters is returned     If the S parameter is not specified  user text is returned  The text will be queried  from th
19.     to select                 Observe that the menu displays both traces of register 1  m1 1 and m1 2   Both are  filled  m1 1 with the signal of channel 1 and m1 2 with the signal from channel 2       Turn the TRACK control to select m1 1     Press the DISPLAY on off    softkey to off     Observe that the stored trace of channel 1 is not displayed any more       Press the    CLEAR DISPLAY    softkey to clear all stored traces before  continuing to the next section     4 44 HOW TO USE THE INSTRUMENT       4 10 CURSORS FUNCTIONS            CURSORS Cursors are provided to make fast and accurate amplitude    O   and time measurements  These can be done in digital   0 mode as well as in analog mode   The analog mode of the oscilloscope is often used to accurately display complex  waveforms  such as in AM  FM  and Video   The digital mode will often be the preferred operating mode for single accuracies  which have to be studied or analyzed        When operating in digital mode  the cursors are always set to the optimum place  for that particular measurement and the readout is directly displayed on the  Screen    This is the fastest and easiest way for most of the measurements     But for measurements on very complex signals with different waveforms in a  trace  the analog cursor measurements gives you a very helpful tool to find your  measurement     STANDARD SETTING    Start with the standard setting  This ensures you have the correct start condition       Press the STATUS and TE
20.    50e for m9 e   m50 e  Extended Memory only     01e   08e for m1 e   m8 e     xx e for 2 ch  models only  xx 3 and xx 4 for 4 ch  models only        admin 16 parameters  separated by a  pds    ACTION BY  PARAMETER TYPE EXAMPLE OSCILLOSCOPE NOTES  trace name string m4 1 ignored  Y_unit string V interpreted  X unit string S interpreted  Y zero number  3 interpreted  X_zero number  8 625E 6 ignored  Y_resolution number 78 13E 3 interpreted  X_resolution number 1E 6 interpreted  Y_range number 65535 ignored  date string 25 01 1994 interpreted 3   time string 17 10 25 40 interpreted 4   dT corr number 375E 3 interpreted 1   min max number 0 interpreted 2   mult_shot_tot number 32 interpreted  mult_shot_nr number 5 interpreted  reserved string 0 ignored  reserved string 0 ignored  Notes  1  dT   dT corr   X resolution  2  1   min max trace envelopes 0   waveform trace  3  date format  European   E25 01 1994  American     01 25 1994  Japanees   J1994 01 25  for compatibility reasons  initial versions   European format without leading character can  be used   4  for compatibility reasons  initial versions  hundreds of  seconds can be omited  count The number of samples in the waveform     512  1024  2048  4096  8192  16384 or 32768   Extended Memory  2048 and higher     6 24 THE CPL PROTOCOL       sample 2 bytes  Most Significant Byte   Least Significant Byte   representing  the 16 bit sample value  bit 16    32768  bit 15   16384       bit 1   1    Range   32k  down  to  32k  up  
21.    CPL  QW  Command to query a waveform   Refer to Chapter 6 for full details     GLITCH TRIGGER    Glitch triggering examines a single channel  on one or more time conditions     Any one of the channel inputs can be used as trigger source and is selected with  the keys    TRIG1  TRIG2    or EXT TRIG  The same keys are used to toggle  between positive and negative glitch triggering     Time qualification is selected with softkeys  These time conditions are     n  t1 triggers when the glitch is longer than the selected time t1  The  TRACK control is used to adjust t1     n  t2 triggers when the glitch is shorter than the time t2  The TRACK  control is used to adjust t2     range triggers when the glitch duration is between two time limits as  specified by t1 and t2  The TRACK control is used to adjust t1 and  the control is used to adjust t2     FUNCTION REFERENCE 5   37       Key sequence     edge tv  glitch    TRIGGER J     Toggle key to select positive or negative  glitch detection     Toggle softkey to select positive or  negative glitch detection     Softkey pair to select the additional trigger  condition     Control to adjust the conditional time or the  beginning of the range     Control to adjust the end of the range        HOLD OFF    Description     The HOLD OFF control determines the hold off time    In the analog mode  the hold off time is an additional    dead    time after each  timebase sweep during which the MAIN TB trigger is inhibited  Variable HOLD  OFF ena
22.    Example  1000000011000111   value  32569    byte 1    byte 2      byte 1   128  value  32768   byte 2   128 64 4 2 1   199  value  199        checksum 1 byte checksum of all sample bytes   add one by one  result modulus 256     Example     In this example  a waveform of 512 points is generated in the computer and sent to  the oscilloscope  The display of the waveform on the oscilloscope screen consists  of two sine waves  amplitude     3 vertical divisions and period  2x    5 horizontal  divisions   The Y scale will be set at 500 Volts per division  and the X scale at  50 0 us per division  The following values can be calculated from these settings              number of waveform amplitude points      divisions   6400 points div    19200   used in the example program       The number of waveform time points   trace length  512    2   2      40 744   used in the example program     500 V div         Y resolution                               78 13E 3V  6400 dots div       X resolution      PUN    1E 6s  50 dots div     After running the example program  perform the following     m Select the default setup by pressing the front panel keys STATUS and TEXT  OFF at the same time        Select the digital mode by pressing the ANALOG key     m Select the recall menu by pressing the RECALL key   Use the TRACK button and the software keys to turn the m4 register display  on and the acq display off     m Press the STATUS key     THE CPL PROTOCOL       Example program     100  110  1
23.    ST7415  9312    You can turn off the text by pressing the TEXT OFF key  This can be done to free  up the display area       Press the TEXT OFF key three times     Observe that the text mode follows the following sequence  menu off  settings off   both on     4 8 HOW TO USE THE INSTRUMENT       CAL SIGNAL AND PROBE ADJUSTMENT    Each measuring probe has been checked and adjusted before delivery  However   to match the probe to your oscilloscope  you must perform the following  procedure to optimize the pulse response of the combination of oscilloscope input  and probes       Connect the probe body to channel 1     Connect the probe tip to the Probe Adjust output of the oscilloscope     Press the AUTOSET key     If the display looks like one of the two displays shown on the left  you must adjust  your probe to get the display shown on the right    The probe output impedance can be adjusted through a hole in the compensation  box of the probe to obtain optimum pulse response  Refer to the following figure                                                                                                                                                                                     Adjust the probe until the screen shows the correct compensation     The probe is now adjusted for optimum pulse response with this oscilloscope  If  you connect the probe to another channel or oscilloscope  it must be adjusted  again to that oscilloscope input       Repeat this adjustment for the 
24.   4K 2 registers of 2 traces each 6  6    2K 2 registers of 4  3  traces each 12  9   512 8 registers of 4  3  traces each 40  30                 memory plus acquisition  2 ch  models between brackets       FUNCTION REFERENCE 5 3       MEMORY EXPANSION    With the Extended Memory Option installed  the acquisition memory can be made  as long as 32K  When shorter acquisitions are selected  more traces can be  stored in memory with a maximum of 208  156  traces              Acquisition Reference memory Traces     1 x 32K 2 registers of 1 traces each 3  3    2x 16K 2 registers of 2 traces each 6  6    3 x 8K 2 registers of 4  3  traces each 12  9    3   512 50 registers of 4  3  traces each 208  156                    memory plus acquisition  2 ch  models between brackets    Key sequence     TB MODE LENGHT    4ch   512 pts        Softkey to select 512 points with confirm menu          4ch   2k pts        Softkey to select 2K points with confirm menu           2ch   4k pts         Softkey to select 4K points with confirm menu           1ch   8k pts       Softkey to select 8K points with confirm menu           576751  9303    5 4 FUNCTION REFERENCE       ADD INVERT SUBTRACT    Description     The CH1   CH2  CH3   CH4  key in the control section for CH1  CH3  can be  used to display additional traces of the sums of these channels  CH3   CH4 is  only present in the 4 channels models     The invert key  INV  in the control section for CH2  CH4  can be used for signal  inversion     A 
25.   55 50        5   Read measured frequency   Measurement    TYPES MEASS  SUFFIXS    Print measured frequency                            Example   of multiple measurements     100    110  120    130    140    150                PRINT  1  QM  10  L    Query for delta   voltage measurement   GOSUB 1000   Sync on acknowledge   INPUT  1 MEASS     Read measured  delta voltage   PRINT  Measurement   USMEASS  Print measured  delta voltage   IF INKEYS    THEN GOTO 120   Do next measurement  until a key is pressed   PRINT  1 CHRS  27      Send ESC character          THE CPL PROTOCOL 6   29       QUERY PRINT         Purpose    To obtain the graphical data of the display from the oscilloscope    The Query Print  QP  command can be used to retrieve a complete display  picture  which can be stored as a HPGL file in the PC  This file can then be used  in for example FLUKEVIEW  Word Perfect or Word files     Command            QP  1 CR                         Response 1                       acknowledge CR       Response 2                       print data null       print data HPGL data bytes   In the example below the print data is read byte after byte by the PC  and appended to the HPGL HGL file    null When input buffer stays empty data transfer has ended                                                                                              6 30 THE CPL PROTOCOL  Example  Written in Quick Basic   CLS  CLOSE  OPEN    COM2 19200 N 8 1  CSO  DSO  CDO    FOR RANDOM AS  1  OPEN  HPGL HGL
26.   Entry Refer to chapter function Page  G  Glitch 5 PEAK DETECTION                    5 51  Glitch trigger 5 GLITCHTRIGGER                    5 36  GND 5 INPUT                                   5 40  H  Hard copy 5 PRINTING AND PLOTTING             5 53  High level 5 VOLT MEASUREMENT                 5 92  Hold off        HOLD ORF                 ry ee 5 37  Horizontal deflection 4 5 HORIZONTAL DEFLECTION            4 21  Horizontal functions 4 8 ADVANCED HORIZONTAL FUNCTIONS   4 34              488 2 5 PRINTING AND PLOTTING             5 54  5 REMOTE CONTROL IEEE 488 2         5 58  Input attenuator 5 INPUT ATTENUATOR                  5 39  Input coupling 5 INPUT                                   5 40  Input impedance 5                                                5 41  Interpolation 5 DISPLAY                               5 30  Installation 2 INSTALLATION INSTRUCTIONS           2 1  5                                            N 1  Invert 5 ADD INVERT SUBTRACT                5 4  L  Logic trigger 5 LOGIC                                         5 44  Low level 5 VOLT MEASUREMENTS               5 92  M  Magnify 5 MAGNIFY HORIZONTAL               5 44  5 MAGNIFY                                   5 45  Maintenance 5 UTIL                                         5 88  Main Timebase 5 HOLD  ORF         ade        aes 5 37  5 MAIN                                       5 46  Mathematics 4 12 PROCESSING FUNCTIONS             4 52  5 MATHEMATICS                       5 47  Maximum vo
27.   FOR OUTPUT AS  2  Open output file  PRINT   1              1    Query print  Start    TIMER  WHILE   TIMER  lt  Start    1  AND  LOC 1    0    WEND  IF LOC 1     0 THEN    Syne on data  PRINT  Response       DO  Byte    INPUTS S LOC 1   51    Read each data byte  PRINT Byte     Display databyte  PRINT  2  ByteS    Append databyte to file  Start    TIMER  WHILE   TIMER  lt  Start    1  AND  LOC 1    0    WEND  LOOP WHILE LOC 1   gt  0  ELSE    Error message  PRINT    No Response     END IF  CLOSE             THE CPL PROTOCOL 6 31       QUERY SETUP Qs    Purpose   To query the oscilloscope for its current setup     This Query Setup  QS  command and the Program Setup  PS  command can be  used together in order to retrieve and restore a complete or partical setup of the  oscilloscope  Optionally a parameter can be added to the command to query a  particular part of the setup  When this parameter is omitted  the complete setup  is returned     The response can be stored as an array of strings in the computer to be sent back  later using the PS command  Each string that is received describes a setup node  in the oscilloscope     Refer to Program Setup  PS  and section 6 8 SETUP     Command   bypass to ask all nodes                    Qs  pds node number CR                                lt node number gt  optional parameter  in decimal ASCII  to query only one  setup node  Refer to section 6 8 SETUP for the setup node  numbers     Response 1           acknowledge CR              
28.   handshake configuration     The LQ1500 printer can have a 6 pole DIN connector with the  following connections    TxD on pin 1 DTR on pin 2   RxD on pin 3 S GND on pin 5       2 RS 232 CABLE CONFIGURATIONS       B  Cable to printer plotter with hardware handshake    parameters   3 wire 7 wire    7 wire   XON XOFF on off    on or off                OSCILLOSCOPE PRINTER PLOTTER  DTE  DTE   SIGNAL 9 PIN 9 PIN 25 PIN    female   male   male   TxD 3           3 2  RxD 2        2 3         220 Qr   S7  4  CTS 8          8 5  DTR 4 o    O 4 20  DSR 6 E    6 6  DCD 1 o        8  S GND 5 o o 5 7  F GND CASE       CASE 1       C  Cable to controller with hardware handshake    parameters   3 wire 7 wire    7 wire   XON XOFF on off    on or off    The communication is confirm the        protocol  This means that signal RfR is  active when the oscilloscope can receive data                             OSCILLOSCOPE CONTROLLER  SONAL              female  ferris   TxD 3              3 2         2            o 2 3  RIR 7           7 4  CTS 8 o   X      o 8 5  DTR 4 Q    lt     4 20  DSR 6      X     Oo 6 6  DCD 1 o  ee 1 8  S GND 5 o        5  7  F GND CASE       CASE 1       NOTE  When during data transfer the signal DSR on the oscilloscope becomes  inactive  the serial communication buffers of the instrument are cleared   The hard copy action or waveform transfer stops then     APPENDIX P    1       Appendix       CHANNELS SIMULTANEOUSLY SAMPLED AT 100 MS s    The scope has two 8 bit flash
29.   the mostly used  default  parameters are   9600  8      3wire                                set up the layout of the printer hard copy       Press the LAYOUT softkey     Set the layout parameters  grid  paper length and trace info      Press the RETURN softkey     Now you are ready to make the hard copy     Press the HARD COPY key     The printer starts printing   Observe that the screen displays   HARDCOPY BUSY       DONE    to indicate  the progress of the printing     4 70 HOW TO USE THE INSTRUMENT       The preceding section describes the setup of a printer using the RS 232 interface   If you wish to use an IEEE 488 equipped printer  all steps are the same  except  for the RS 232 setup     chi                                                          y Div  Timebase  TRACE  200mV 200us     1    576836  9303    UTILITY PLOT SETUP    For a hard copy with a plotter  please refer to the previous section on how to set  up the oscilloscope to be used with a printer  Instead of selecting a printer  you  can select a plotter from the following models or emulations   HP7440  HP7550   HP7475A  HP7470A  HPGL    standard HPGL   PM8277 and PM8278     HOW TO USE THE INSTRUMENT 4 71       4 16 AUTOSET AND SETUP UTILITIES    This oscilloscope has a number of utilities that assist you to quickly get to the  setup you need    One utility is the SET STANDARD utility to set the oscilloscope to a factory   defined  known state    The most frequently used utility will be AUTOSET  AUTOSET is 
30.   year month day      The adjusted time becomes active after pressing  the softkey ENTER  amp  RETURN     print  plot cik Toggle softkey to select printer  plotter or  clock menu   TRACK 2  Control to select printer  plotter  type   LAYOUT Access to LAYOUT menu for printer or  e plotter   IEEE  RS232 Toggle softkey to select IEEE or RS 232    interface  only visible if IEEE interface is  present      ST7410  9312    FUNCTION REFERENCE       PRINT amp   UTILITY   PLOT amp CLK       PRINTER  LAYOUT yes no    TRACE  info  yes no     C    576763  9303       GRID  PLOTTER  LAYOUT yes no    N   zl    TRACE  INFO  yes no    COLORS  yes no       576764  9303    Toggle softkey to select printer  plot  or clock menu     Toggle softkey to select the clock  time stamp  to be  adjusted with the up down softkey pair     Up down softkey pair to adjust the time     Toggle softkey to select European  USA  or Japanese  format of the date     Toggle softkey to print the grid     Toggle softkey to select paper length     Toggle softkey to print trace information     Toggle softkey to plot the grid     Toggle softkey to select paper size     Toggle softkey to select plot format     Toggle softkey to plot trace information     Toggle softkey to plot traces in colors     FUNCTION REFERENCE 5 57       PROBE UTILITIES    Description     The PROBE SWITCH    setting of the  UTILITY  gt  gt  PROBE    menu determines the  instrument s reaction when you press the command button on the probe  You can  sele
31.  65 Mathematics 1 settings   42 66 Mathematics 2 settings   Mathematics bits  mathematics type  source 1 low high  source 2  low high  scale low high  scale minimum low high  offset low high   window width low high    50 80 Display settings   settings display on off  ground level indicator on off  trigger level  indicator on off  dots join on off  x versus Y on off  status view  on off  backup status view  window on off     51 81 Trace intensity settings   analog trace intensity low high  mtb dtb intensity ratio    52 82 Display trace position setting  X position  Y position    60 96 Setup label text  max 22 characters    70 112 Real time clock setting  format    FO 240 Special node for service purposes    It is NOT necessary to send all strings to the oscilloscope  when a setting must  be changed     6 52 THE CPL PROTOCOL       Applications         setup nodes for different timebase settings can be stored  separately  They can be used afterwards as fixed    templates    to  change only the oscilloscope timebase setup     The layout of each setup node string is   DnllxxxXxxXx   dias         where all characters are in the hexadecimal range  0  9 A  F    Each pair forms the hexa decimal representation of a byte  in high low order   eg  hex 2A   bin 00101010   dec 42     nn the setup node number  ll the number of bytes in the setup node    each byte is represented by two characters   xx the actual setup node    ACQUIRE MENU STRUCTURE       Appendix A  ACQUIRE menu structure    
32.  C  Key to start the fine adjustment procedure  Press this key    for at least 2 seconds to start the procedure     Remote commands   CPL  CL  Command for Calibrate   Refer to Chapter 6 for full details    FUNCTION REFERENCE 5 17       CHANNEL TRACE SELECTION    Description     In this family of instruments  the distinction is made between    channel    and    trace      A channel is referred to as an input channel  complete with AMPL and POS  settings  A trace represents a waveform which has been stored in one of the  register memories  Once such a waveform is recalled from memory  it is displayed  on the screen as a trace     Each channel can be turned ON and OFF by the ON key located near the channel  keys on the front panel  In the digital mode  this key switches the acquisition and  the display on off     In the digital mode  a waveform which has been previously stored in a register  can  be recalled and displayed via the RECALL menu  One of the registers  MO  is the  acquisition register  where the newly acquired data from the input channels is stored     The RECALL function only switches on and off the display of the trace and does  not influence the acquisition  It allows the running acquisition to be hidden to give  a better view of the processed results calculated from an acquisition running in  the    background        In addition to the acquisition register  MO   the scope provides eight different  register memory locations  50 for the extended memory version   Ea
33.  CONTROL softkey so that          is intensified      Turn the TRACK conirol and observe that only the time cursors move     HOW TO USE THE INSTRUMENT 4 47       CURSORS READOUT      Press the softkey next to READOUT        The  CURSOR READOUT     selection menu for horizontal and  vertical measurements is now  displayed  You can make the  following selections from this menu   AT  1 AT  AT ratio  T trg and phase  for time measurements    AV  V1  V2  AV ratio for voltage  measurements                                                                       AT is the time difference between the time cursors  The readout is in seconds  s   or fractions of seconds    1 AT results in the inverted value of the time difference  and is displayed in Hz   When the time cursors are set exactly one period apart this represents the  frequency of the signal    AT ratio allows for a relative measurement of two cursor readings  It is displayed  as a percentage    Phase is used to measure a phase difference between two signals  or to  determine the phase of a detail within one signal    T trg gives the timedifference between the cursors and the trigger point     AV is the voltage difference between the voltage cursors  readout  AV    V     V1  V2 displays the absolute voltages for each voltage cursor in relation to ground   readout  V1 Vreference  V2 Vdelta     AV ratio allows for a relative measurement of two cursor readings  It is displayed  as a percentage     4 48 HOW TO USE THE INSTRUMENT  
34.  Chapter 6     Note 1  Finding your way through the menus     Some functions are attained via successive steps through a tree structure of  menus  An example is how to activate the key CLICK function      Press the UTILITY menu key      Press the SCREEN  amp  SOUND softkey      Press the SOUND softkey      Press the CLICK softkey to activate    on        The shortform annotation for these steps is  UTILITY  gt  gt  SCREEN  amp  SOUND   gt  gt  SOUND  gt  gt  CLICK on        Note 2  A complete function index is part of this Operating Manual  This    function index contains all function names and reference words in  alphabetical order including the relevant chapter and page number  where more detailed information can be found  It can be found at the  end of this manual     5 2 FUNCTION REFERENCE       ACQUISITION LENGTH    Description     The oscilloscope allows the user to select the acquisition length  or record length  that best suits the needs of the application  The default acquisition length is 512  data points for each trace  It is possible to increase the length of a trace from 512  points up to a maximum 8K points  or 32K points if the memory expansion option  is installed   This results in a trace length of 16 screens  or 160 divisions     The TB MODE menu offers four choices to select the acquisition length  This  menu is accessed via the key sequence       MODE  gt  gt  ACQ LENGTH  gt  gt     Selection of the desired acquisition length is then made with the soft key
35.  Command                  G L  CR                   Response                       acknowledge CR       Example   100 PRINT  1  GL    Send command  110 GOSUB 1000   Sync on acknowledge    Local  Remote  Remote Local Lockout States          Go to Remote      d    Local State Remote State   4          Go to Local      Reset Instrument                                            lt STATUS LOCAL button gt    Local Lockout   Remote with   Go to Local  Local Lockout   Reset Instrument  State             6 12 THE CPL PROTOCOL       GO to REMOTE GR    Purpose     Puts the oscilloscope in the Remote State  In the Remote State none of the  oscilloscope functions are accessible via the front panel buttons and knobs     Going back to the Local State is achieved by sending the Go to Local  GL   command or by pressing the STATUS LOCAL key on the frontpanel  Refer also  to  Local Lockout  and  Go to Local       Command            GR  CR                         Response           acknowledge CR                   Example   100 PRINT  1  GR       Send command  110 INPUT  1 ACK   Sync on acknowledge    Local  Remote  Remote Local Lockout States    d   Go to Remote        d     Local State Remote State                  Go to Local      Reset Instrument                                         lt STATUS LOCAL button gt    Local Lockout   Remote with   Go to Local              4 Local Lockout   Reset Instrument  State             THE CPL PROTOCOL 6 13       IDENTIFICATION ID    Purpose     Returns t
36.  Delayed Timebase     Probe correction     Beep and click signals     Noisy signal triggering     Confidence check    Most of these functions can be operated via the UTILITY menu    For further information on all of these features  refer to the cross  reference index  and Chapter 5  This chapter describes all oscilloscope functions in alphabetical  order    This oscilloscope has been designed to give you many years of dependable  service  We are sure that you will feel confident with your Fluke oscilloscope   Fluke is in the process of constantly improving products and documentation  For  any problems or suggestions  please contact the Fluke Service Center nearest you   A complete listing of addresses for the Fluke Service Centers can be found in the  Reference Manual     FUNCTION REFERENCE 5 1       5 FUNCTION REFERENCE    This chapter contains an alphabetized description of each oscilloscope function   For easy reference  the functions are organized in the following order     1     The Function description  Explanation and detailed information about the function   Key sequence    Tells the operator which keys controls to use to select the desired function   The first key control mentioned is always one of the front panel menu selection  keys  and the other keys are the softkeys     Remote commands    Gives information about the command to be programmed for the operation of  the relevant function via a remote control interface  For more detailed  programming information  see
37.  E          YOUT ZMOD MTB EXT DTB  BAIE TRIG GATE    STORAGE SPACE       OPTIONAL       L  ONLY IN    FOR MAINS CABLE    4 CH MODELS    ST7665E  970117    XIII       FRONT PANEL CONNECTIONS    to OOOOSO    Probe Adjust  Squarewave output signal for e g  probe calibration   Amplitude is calibrated     CH1  BNC input socket for vertical channel 1 with probe  indication contact     CH2  BNC input socket for vertical channel 2 with probe  indication contact     CH3  BNC input socket for vertical channel 1 with probe  indication contact   only in 4 channel models     CH4  BNC input socket for vertical channel 1 with probe  indication contact   only in 4 channel models     EXT TRIG   BNC input socket used as an extra external trigger  input with probe indication contact  only in 2 channel  models     Ground socket  banana   same potential as safety  ground     The measuring ground socket and the external  conductor of the BNC sockets are internally  connected to the protective earth conductor of the  three core mains cable  The measuring ground  socket or the external conductor of the BNC sockets  must not be used as a protective conductor terminal     XIV       REAR PANEL CONNECTIONS    Z MOD  BNC input socket for external intensity modulation  of the CRT trace     RS 232 BUS  EIA 232 D   Input output socket to connect the oscilloscope to an  RS 232 Interface        NC NOT CONNECTED    576065    LINE IN  Line input socket  Fuse holder is built in                    XV       OPTION
38.  MAIN TB on and X DEFL    off         MAIN TB in 1 ms and auto mode  VAR and 10x MAGNIFY are off      MAINTB triggering in edge  level pp on  positive slope  ac coupled and trigger  source CH1     Note  Error messages appear in the CRT viewing area when incorrect  commands are given    Key sequence    setups  TRACK   std memory selection    recall       softkey to recall  std  settings     ST6767  9303    Remote commands   CPL  DS  Command for a front panel reset    Refer to Chapter 6 for full details    5 70 FUNCTION REFERENCE       STATUS SCREEN    Description     Normally a maximum of four lines of setting information are given in the lower  screen area  More extensive setting information can also be displayed  The  STATUS toggle key switches between normal and extensive settings information     The status screen gives the following additional setting readouts      Channel input coupling is indicated by AC  DC  or GND instead of symbols      Settings of inactive channels are given in addition to the active channels      Channel off or on is displayed      The probe type is given for each channel      Acontinuous indication of HOLD OFF  LEVEL MAIN TB and LEVEL DEL D TB  is given      MAIN TB and DELD TB trigger coupling are indicated      The trigger position is displayed      TV mode settings are displayed     When the SAVE or RECALL menu is active  the STATUS SCREEN will show the  X and Y settings of the displayed traces     Note         STATUS key has a third function wh
39.  Pattern  State and Glitch triggering  2 ns   2 channel models  4ns Glitch  triggering only    Event delay and pretriggering and posttriggering    TV triggering including HDTV and TV line selection    Serial interface for printing and plotting    Averaging to reduce signal noise and to increase the vertical resolution from  8 to 16 bits    Advanced mathematics  including digital low pass filtering  A Math  option  adds integration  differentiation  histogramming  and FFT    Sine interpolation and magnification which enables true to life four channel  single shot acquisitions with a timebase up to 625 ns div  32x magnified    A delayed timebase with full trigger features         RS 232  EIA 232 D  interface  standard  and an GPIB IEEE 488 interface   optional     Autocal for automatic fine tuning of all circuitry to achieve maximum accuracy  under all user conditions    Closed case calibration for efficient maintenance of traceable calibration at  minimum cost     The following options are available        MATH  option with more automated measurement functions including  envelope and measurement pass fail testing  Also included in this option are  Integration  Differentiation  Histogramming  and FFT    Memory extension offering 32K acquisition length and the ability to store 156  traces  of 512 samples each  in memory for 2 channel versions    IEEE 488 2 interface using the new SCPI  Standard Commands for  Programmable Instruments  industry standard for remote control of test 
40.  Right side switches x10 MAGNIFY on       LD Left side switched x10 MAGNIFY off   i Digital mode  Right side increases the MAGNIFY factor   Left side decreases the MAGNIFY factor     MAGNIFY    MAGNIFY VERTICAL    Description     In the digital mode the displayed signal s  can be expanded vertically up to 32  times for detailed signal examination  Vertical magnification is activated in the  DISPLAY menu    The TRACK control is then used to select the magnification factor in a x1  x2  x4      X92 sequence  The selected factor is displayed in the DISPLAY menu          actual vertical deflection coefficient in V div is automatically updated and  displayed in the text area at the bottom of the screen     You can use the POS control of the displayed channel s  to move the trace and  display the part of interest     If you selected a trace stored in one of the memory locations m1     mn  using the  RECALL menu  to be displayed and magnified  you can change the position of  such magnified traces using the Y POS control as indicated      the RECALL menu                                    i  i                   VVVM       E                      Function of vertical magnifier and Y POS control  Key sequence     Display TRACK    Control to select the vertical magnification     576759  9303    5 46 FUNCTION REFERENCE       MAIN TIMEBASE MANUAL   AUTOMATIC    Description     The Main Timebase  MAIN TB  has a wide range of time div settings  This  enables the display of signals of various 
41.  TEXT function      Softkey to select sine   linear   or no interpolation   in digital mode only      FUNCTION REFERENCE 5 33       ENVELOPE    Description     If a waveform is changing over time  because of drift  jitter  or intermittent faults    the    history    of the changing waveform can be collected using the envelope mode   Inthe envelope mode the minimum and maximum signal values are stored  taking  the values of a large number of successive waveform acquisitions  If an  interpolation mode is switched on  the area between the maximum and minimum  values is shaded     Each time the settings of the scope are changed  like trace POS   the envelope  process is restarted by clearing the acquisition register  A new envelope process  starts and continues until stopped or until a change of settings is made                                                                                                                       0  Display of signal with Display of signal with frequency  amplitude variation  AM   variation  FM or jitter      ENVELOPE and AVERAGE functions are mutually exclusive  When the AVERAGE  function is selected  the ENVELOPE mode is automatically switched off     Key sequence     ENVELOPE  ACQUIRE onoff Toggle softkey to switch Envelope on or off     576790  9303    5 34 FUNCTION REFERENCE       EXTERNAL TRIGGER  2 CHANNELS MODELS ONLY     Description     The External Trigger input provides an extra input that can be used as the trigger  source for the Main 
42.  THEN GOTO 1100   1040 PRINT  ERROR    ACK    L050 ON ACK GOTO 1060 1070 1080 1090       L060 PRINT  SYNTAX ERROR    END  1070 PRINT  EXECUTION ERROR    END  1080 PRINT  SYNCHRONISATION ERROR    END  1090 PRINT  COMMUNICATION ERROR    END  1100 PRINT  UNKNOWN ACKNOWLEDGE    END  1110 RETURN    First the serial port of the PC is opened  line 10  with the settings of the  oscilloscope communication parameters       Following that  the example program lines from the COMMAND REFERENCE  SECTION can be executed     The subroutine to synchronize on the   acknowledge   returned from the  oscilloscope starts at line 1000        The oscilloscope communication parameters are stored in battery backup  memory when the oscilloscope is turned off  On power up the parameters are  restored     6 4    THE CPL PROTOCOL       6 3 COMMANDS IN FUNCTIONAL ORDER       group name command  Communication Program Communication PC  Setup Auto Setup AS  Default Setup DS  Program Setup PS  Program text PT  Query Setup Qs  Query text QT  Recall Setup RS  Save Setup 55  Calibrate CL  States Go to Local GL  Go to Remote GR  Local Lockout LL  Measurement Arm Trigger AT  Program Wavefrom PW  Query measurement QM  Query Wavefrom QW  Trigger Acquisition TA  Query Print QP  Miscellaneous IDentification ID  Reset Instrument RI  STatus query ST    THE CPL PROTOCOL 6 5       6 4 COMMANDS IN ALPHABETICAL ORDER       command name   AS Auto Setup   AT Arm Trigger   CL Calibrate   DS Default Setup   GL Go to Local   GR G
43.  The male 9 pin connector and its  connections are shown in the figure below     For correct functioning Communication parameters must be adjusted  This is  done in the menu UTILITY RS 232 SETUP  This menu is reached via the key  sequence  UTILITY  gt  gt  RS 232 SETUP  gt  gt            Possible Selections are     Baud rate  To be selected with the TRACK control     Number of DATAbits and PARITY  Combinations are  dataBITS  7 7 8 8 8  PARITY  Odd Even Odd Even No  There is always one stopbit     Hardware handshake is selected with  3 wire 7 wire   In the 7 wire position the  hardware handshake signals DSR DTR and CTS are active     Software handshake is selected with  XON XOFF on off        NC        ob    Operation of the front panel key STATUS LOCAL  passes the control of the oscilloscope from the  interface  remote  to the front panel keys  local         NC NOT CONNECTED    ST6065    5 60 FUNCTION REFERENCE                   Key sequence   WITH IEEE   REMOTE RS232  UTILITY SETUP SETUP  RS232 TRACK  SETUP    NO IEEE     a Control to adjust baud rate   BITS  7 8  Toggle softkey to select number of  databits   PARITY  no odd  even  Toggle softkey to select parity   3 wire  7 wire  Toggle softkey to select hardware  handshake   XON XOFF  on off  Toggle softkey to select software  handshake   STATUS  LJ Key to switch from remote to local   LOCAL    Remote commands     CPL  PC  Command to program communication parameters   GL  Has same result as operation of STATUS LOCAL key   
44.  The number of samples at each  amplitude level is counted  A peak in the number of samples at one level  indicates the high level and is referred to the 100  level  If no such point is  present  e g   sine wave   the high level is equal to the maximum voltage     5 94 FUNCTION REFERENCE         Oversh  OVERSHOOT   Measures the overshoot  in     related to the amplitude of the signal  There are  two types of overshoot  rising overshoot and falling overshoot  undershoot      max  of 1st rising slope   high     100   high   low    rising oversh         low   min  of 1st falling slope   high   low    falling oversh    100          presh  PRESHOOT   Measures the preshoot  in     related to the amplitude of the signal  There are  two types of preshoot  rising preshoot and falling preshoot     low   min  of 1st rising slope      100   high   low    rising presh         max  of 1st falling slope   high  high   low     100        falling presh      It is possible to perform measurements on the part of the waveform between the  two cursors  This function is called cursor limited measurements and is turned on  via the key sequence MEASURE  gt  gt  CURSOR LIMIT  amp  STATIST  gt  gt  CURSOR  LIMITED yes  Cursor operation is done via the TRACK and A controls and via the  CURSOR menu     Results of the measurements MEAS1 and MEAS2 are displayed in the top left  corner of the screen  When you press the key sequence MEASURE  gt  gt  CURSOR  LIMIT  amp  STATIST  gt  gt  STATIST on  the scree
45.  The upper trace is the main timebase trace  This first trace shows an intensified  part  Adjust the TRACE INTENSITY with the control as necessary  The lower  trace is the delayed timebase trace and is an expanded representation of the  intensified part in the upper trace     Step 3    Step 4    Step 5    Turn the DELAY knob to shift the intensified part and to select which  part of the main timebase you want to magnify     The delayed timebase TIME DIV keys are used to select the   magnification factor     Notice the changing delayed timebase  TIME   DIV    readout at the bottom of the screen     The     symbol at the fourth blue softkey indicates that the cursor  TRACK control can be used to make adjustments  In this menu the  cursor TRACK control is used to change the TRACE SEParation  which  is the distance between the main timebase and the delayed timebase     The delayed timebase can be used in the triggered mode  The triggered mode is  selected with the STARTS TRIG   D softkey  The function of the triggered mode will  be explained in Chapter 4  For this part of  Getting Started   remain in the  STARTS mode     Step 6    Switch the menu off with the TEXT OFF key  Notice that the delayed  timebase is still active and that the most important functions  DELAY  control and TIME DIV key pair  still allow you to operate the delayed  timebase     3 22 GETTING STARTED       3 15 TRACE STORAGE    In the digital mode you not only have the ability to store traces on the screen   u
46.  USE THE INSTRUMENT       TRIGGER LINE      Connect a sine wave signal of 4 V 300 Hz to channel 1      Simultaneously press the STATUS and TEXT OFF keys      Press AUTOSET      Select a timebase speed of 1 ms div      Press the TRIGGER key      Press the softkey next to the function    ch1 line    to select    line        When  line  is selected  this function is intensified  The line frequency is used as  the trigger source       Slowly vary the frequency of the sine wave input signal     Observe that when the input frequency is close to an integer multiple of the line  frequency  the running sine wave slows down or even stands still  Line triggering  can be used to display signals or signal components that are related to the line  frequency  e g   hum or power supply ripple      HOW TO USE THE INSTRUMENT 4 39       4 9 MEMORY FUNCTIONS    The next section deals with storing and recalling traces in memory for later use   Functions related to trace storage and recall are easily accessible via the menus  selection keys labeled  SAVE  and RECALL        STANDARD SETTING    Before continuing with the memory functions  you must first set the instrument to  the default setting to ensure a correct start situation       Simultaneously press the STATUS and TEXT OFF keys     Connect the Probe Adjust signal to channel 1     Press AUTOSET     The Probe Adjust signal  now supplied to the input  is a square wave with a lower  level of OV and a top level of 600 mV     TRACE STORAGE CLEAR  a
47.  USERPROG allows the user to customize the AUTOSET function for specific  applications     Example  To program the AUTOSET function for dc coupling of the inputs  do  the following       Press the UTILITY key      Press the softkey labeled  AUTOSET       Select  userprog      Press the  VERT  gt     softkey  select  dc   and press RETURN     Press the AUTOSET key     The Probe Adjust signal is now displayed as dc coupled  instead of ac coupled as  would be the case after a STANDARD AUTOSET     SETUPS           All settings of all controls can be stored in any of 10 memory locations     This is a useful feature when performing routine tasks involving complex  setup procedures  Rather than changing all parameters to go from one  measurement to another  the oscilloscope can recall even the most complex  settings combinations with the touch of one button   The SETUPS key is used to store and recall previously defined instrument  settings     First clear all memory locations     Press the SETUPS key    Press the  CLEAR amp PROTECT  softkey   Press the    clear all    softkey    Press the  yes  confirm softkey     Example  Set the attenuator to 500 mV div  and set the timebase to 500 us div       Position the trace in the bottom of the screen     HOW TO USE THE INSTRUMENT 4 73       This setup can be saved in memory as follows       Turn the TRACK control until memory location    s7    is selected     Press the softkey    save                          FRONTS  Os6    Ss  5       The
48.  UTILITY  gt  gt  AUTOSET  gt  gt   userprog     This gives access to VERT  TRIG and PROBE submenus where the  selections are made  The following table shows possible selections  Unaffect  means that existing selections are not overruled by AUTOSET                       Description Standard Userprog Menu  AUTOSET alternatives UTILITY  gt  gt AUTOSET   gt  gt userprog gt  gt    Channel selection volt div  on off unaffect VERT  Input coupling ac dc  unaffect VERT  Input Impedance 1MQ 500  unaffect 2  VERT  Bandwidth limiter off on  unaffect VERT  Triggering edge  ac  unaffect TRIG  level pp on   auto  LEVEL  MTB center  Probe manual probe unaffect PROBE  selection set manual probe  to 1 1 selections          1  200 MHz models only    FUNCTION REFERENCE    5 13       Key sequence              AUTOSET  off  default  userprog  UTILITY AUTOSET SS  CHANNELS  scan  VERT unaffect  ac       unaffect  1MQ 500   unaffect  BWL  on off  unaffect  UNAFFECT  TRIG    s  PROBE  1 1  PROBE unaffect    576066  9303    Remote commands   CPL  AS  Command for an Auto Setup   Refer to Chapter 6 for full details     Toggle softkey to activate userprog    Toggle softkey to preset channels  on off and input attenuator after  AUTOSET     Toggle softkey to preset input  coupling after AUTOSET    Toggle softkey to preset input  impedance after AUTOSET    Toggle softkey to preset reaction of  bandwidth limiter after AUTOSET     Toggle softkey to preset trigger  settings after AUTOSET     Toggle softkey to 
49.  actual setting of the front is             now stored in memory location undo        s7     The indication in front of eave C3  memory location number  s7  7     TEXT p  changes from an open circle to    J    LEAR  closed circle                                                                              1_500      ITB  50015 1       To recall a previously stored setting      Press the AUTOSET key     The signal is displayed again with an amplitude of three divisions and with four  periods       Press the SETUPS key      Use the TRACK control to select the same front number    s7    as in the  previous example      Press the softkey next to recall        The stored settings are recalled and the trace is displayed in the same way as  when the setup was stored    A STANDARD FRONT setup can be recalled at any time by pressing the STATUS  and TEXT OFF keys simultaneously     4 74 HOW TO USE THE INSTRUMENT       SETUP TEXT LABEL    In the setup menu each stored setups can be given a label of user defined text   This is done in the submenu TEXT of the SETUP menu  The cursor controls are  used for editing text  The TRACK control is used as  cursor   The A control is used  for selecting the character     SETUP RECALL A SEQUENCE    You can create a sequence of setups by storing setups to successive locations    e g  51  52  53  and clearing the next location  e g  54   The AUTOSET key then  can be programmed to step through this sequence  This is very useful for semi   automated manu
50.  align the trace with the    graticule    Control for focusing of trace  text and cursors     O S    GRATICULE  ILLUMINATION    Control for illumination intensity of measuring graticule          SCREEN MESSAGES    Description     User messages show up in the center of the CRT viewing area  Messages warn  of incorrect settings and error conditions  The following table shows the important  messages     Message Meaning  Refer to function 7          ALWAYS PARITY IF 7 BITS Always parity in case of 7 bits   Function   REMOTE CONTROL RS 232      AUTOCAL    approx4 min  Indicates that autocalibration has started   Function CALIBRATION AUTOCAL   AUTOCAL NECESSARY Indication that scope is out of its specified    temperature range and that an  autocalibration is necessary   Function CALIBRATION AUTOCAL     AUTO SETTING    Indicates that instrument performs an  autoset   Function AUTOSET        5 64    FUNCTION REFERENCE       AUTO SETTING  USERPROGRAM     CALIBRATION COMPLETED    CALIBRATION ERROR        CH  500 OVERLOAD    DESTINATION PROTECTED    FUNCTION ONLY IN ANALOG    HARDCOPY ABORTED    HARDCOPY BUSY       DONE    HARDCOPY DONE    INVALID REGISTER SELECTION    INVALID SELECTION    KEY INACTIVE WHEN STOPPED    NO AVERAGE IN ROLL MODE    NO BATTERY BACKUP    Indicates that instrument performs an  userprogrammed autoset   Function AUTOSET USERPROG        Autocalibration is completed   Function CALIBRATION AUTOCAL     Autocal not successfully completed   Function CALIBRATION AUTOCA
51.  analog to digital converters  ADCs   each with a  maximum sample rate of 100 MS s  This allows two channels to be captured  simultaneously in single shot mode  with sample rates of up to 100 MS s for each  channel  The horizontal resolution is then 10 ns  i e   1 100 MS s      DOUBLE SAMPLING MODE FOR 200 MS s SINGLE CHANNEL ACQUISITIONS    Input stage switching expands the capability of both ADCs so they can be  interleaved  allowing a maximum sample rate of 200 MS s on a single channel   The entire acquisition memory can be used so that the record length for a single  channel acquisition is can be equal to the full memory of the instrument  i e   8K  for the standard instrument  and 32 K for an instrument equipped with Memory  Expansion   The interleaving is done by feeding the input signal of a single  channel to both ADCs that are sampling at a phase difference of 5 ns     GLITCH DETECTION    The sampling rate of a DSO decreases when you lower the timebase setting  As  long as the glitch capture feature is activated  the ADCs continue to sample at  their maximum speed  A fast digital circuit analyzes all of these samples at real   time speed  100 MHz signal processing      Only the maximum and minimum values  i e   an envelope of min max pairs  are  stored in memory  Such a glitch detection circuit allows very narrow glitches to be  captured even at low timebase settings and even if they occur only once     The glitch capture circuit can also be used to avoid aliasing  T
52.  buttons in each of the  vertical channel sections    When the PATTERN condition is true for a specified amount of time  the  oscilloscope triggers     FUNCTION REFERENCE 5 43       The time condition        be set in the softkey menu     enter  exit    if gt t1    if lt t2    range    Glitch    triggers when the pattern becomes true   triggers when the pattern changes from true to false     triggers when the pattern is true and its duration exceeds a  specified time  The TRACK control is used to adjust time limit t1     triggers when the pattern condition exists for a time which is shorter  than t2  Triggering actually occurs when the pattern changes from  true to false in a time shorter than t2  The TRACK control is used  to adjust t2     triggers when pattern remains true for a time longer than t1 and  shorter than t2  Triggering actually occurs when the pattern  changes from true to false within the specified time limits  The  TRACK control is used to adjust t1  and the control is used to  adjust t2     Refer to GLITCH TRIGGER function     Key sequence     TRIGGER    state  edge tv pattern TRIG  logic glitch Toggle key to select the logic state for    CH1  CH2  CH3 and CH4     Softkey pair to select the trigger source  CH1     CH4 as clock input       sar     p  RRA    enter          Vx       if lt t2  Softkey pair to select the additional trigger  condition     99    4  2   gt       A    Control to adjust the conditional time or the  beginning of the range        Control
53.  buttons simultaneously to select the VARiable  mode  This mode is used to make fine adjustments of the input  amplitude settings between the 1 2 5 steps    When the VARiable mode is turned off  the oscilloscope selects the  nearest  1 2 5  value            4 VAR AJZ                     Enter the VAR mode by simultaneously pressing both AMPL keys     Adjust amplitude with either AMPL key     Observe that the sensitivity steps that can be selected are much finer than before  and that the displayed amplitude is continuously adjustable     Note         VAR values are calibrated amplitude settings as well  This enables  you to make accurate measurements and readouts even when  intermediate settings are used       Adjust          to 220          Press both AMPL keys to turn the VARiable mode off     4 18 HOW TO USE THE INSTRUMENT       VERTICAL CH1 CH2      Using two probes  connect the Probe Adjust signal to Ch1 and Ch2      Press AUTOSET      Both Ch1 and Ch2 are now displayed      Adjust POS and AMPL to get a display as illustrated on the left below     Press the CH1 CH2 key     In the analog mode  a third trace will appear on the screen  This trace has twice  the amplitude of the Probe Adjust signal  The position of the third trace is affected  by the position controls of both Ch1 and Ch2    The display mode is indicated in the screen as 1 2        RR                      CH1 CH2     gt                                                                                            
54.  corner of the screen reads 100 mV div    The amplitude of the signal is then  100 mV div times 6 div   600 mV                                                        CH1 100m V         50gmv4                                                                         AMPLidiv AMPLidiv         1  5    Try other sensitivity settings as follows      Use the AMPL keys to step through the attenuator range     Observe that the sensitivity readout changes in steps following a 1  2  5  sequence  The AMPL keys allow you to step up and down through the  sensitivities from 5 V div to 2 mV div and vice versa  This sequence enables a  quick selection between the oscilloscope   s sensitivity positions and are such that  almost every input signal can be made visible with sufficient amplitude       Adjust          to 100 mV     A waveform with an amplitude of six divisions is displayed     HOW TO USE THE INSTRUMENT 4 17       VERTICAL AUTO RANGE    ANG        AUTO RANGE function results in an amplitude display of 2 to 6       divisions     Press the AUTO RANGE key     Observe that the amplitude of the signal changes from 6 divisions to 3 divisions   In the upper right corner of the display ATT 1 is displayed  This indicates that the  AUTO RANGE function is active on input channel 1       Press the upper AMPL key once     The AUTO RANGE function is switched off and the ATT 1 indication disappears  The attenuator is now back in manual control     VERTICAL VARIABLE AMPLITUDE    AMPE Press both AMPL
55.  factor is displayed in the  DISPLAY menu  At the same time the vertical deflection  volts div  in the bottom  text area is adjusted     X versus Y is an XY display capability in the digital mode  It is similar to X deflection  in the analog operating mode  The TRACK control is used to select the source for  vertical direction  Y   The source can be any newly acquired trace but it can also be  a saved trace in a memory location  e g   m3   For horizontal direction  X  you can  choose from any active channel or a signal saved in memory  e g   m3 1  as well     FUNCTION REFERENCE 5 31       Horizontal source selection is made with softkeys               sources must always be  traces from the same register memory     This is done to avoid errors because the traces have to be sampled simultaneously  to give a useful and correct X vs Y display     The TEXT submenu allows you to display    user text    as additional information in  the viewing area  This can be very useful when making photographs or hard  copies on printers or plotters  For a detailed description  refer to the USER TEXT  function  The submenu is also used for on off switching of the trigger level  indicators for MAIN TB   DEL D  TB  and EVENT  triggering  and the ground level  indicators for each channel     INTERPOLATION determines if and how the spaces between sample points are   displayed  The choices can be dots  samples only   interpolation tuned off   linear   or sine interpolation activated  Linear interpol
56.  has a dedicated submenu for usertext  See the USERTEXT function for  more information    Presettings for the remote control interfaces  Refer to REMOTE CONTROL  RS 232 and REMOTE CONTROL IEEE 488 2   Presetting for suitable printer or plotter and real time clock adjustment   Settings for maintenance and repair such as tests and calibration data  Data  affecting the instrument   s accuracy can only be changed by entering a  password  This is explained in detail in the Service Manual     The UTILITY menu is reached via the UTILITY menu key  The following figure  shows the structure of the UTILITY menu related to instrument operation     Key sequence     UTILITY        5        Access to AUTOSET programming     Access to PROBE functions        ot        Ld Access to UTILITY SCREEN  amp  SOUND functions   RS232      TEES   SETUP  REMOTE CONTROL when IEEE 488 option is present              Access to interface settings  REMOTE functions      PRINT amp   Polat Access to PRINT  amp  PLOT  amp  CLOCK settings menu        y    MAINTE  Access to UTILITY MAINTENANCE functions  for service  technicians only               y    576779  9312    FUNCTION REFERENCE 5 91       UTILITY SCREEN  amp  SOUND    Description     The UTILITY SCREEN  amp  SOUND menu is used to select on screen text  trigger  and ground level indicators  and user text  Settings for acoustic feedback  beep  and click  are set in this menu  The menu can be reached with the key sequence   UTILITY  gt  gt  SCREEN  amp  SOUN
57.  key  the AUTOSET key  or the main timebase  AUTORANGE key     In VARiable timebase mode the AUTO RANGE function tries to keep the  displayed trace unchanged     Key sequence     AUTO RANGE    Toggle key to switch the AUTO RANGE function of main  timebase and vertical channels     5 10 FUNCTION REFERENCE       AUTOSET    Description     The AUTOSET function sets the oscilloscope so that an optimum display of the   input signals is obtained within the same mode analog or digital  Operating the   AUTOSET key results in      Channels with an input signal are switched on  others are switched off      Input coupling is set to ac  because of this autoset does not function at very  low signal frequencies      Input impedance is set to 1 MQ     Input attenuator settings are optimized  VAR function off      Bandwidth limiter and INVert are switched off      ALT or CHOP mode selected to most optimal display      Vertical POS selection in center screen      Edge triggering on positive slope is activated      Trigger source is the channel with the lowest frequency  at equal frequency  the lower channel number is selected  Trigger source is EXT TRIG input when  a signal is present  and Trigger View becomes active      Trigger coupling is set to ac and level pp on      Horizontal mode is switched to MAIN TB only      Cursors and usertext are switched off  settings display is switched on     These settings are suitable for most signal conditions  You can also customize  AUTOSET to your own 
58.  mode  the trigger system determines which part of the trace is  acquired and placed in memory  As in every other Digital Storage Oscilloscope   signal acquisition through the ADC system is a continuous process  until it is  stopped by the trigger signal  If nothing else would be done  the signal placed in  memory would be the signal prior to the trigger moment and the trigger moment  would be displayed at the end of the screen  For most oscilloscope users this  would be confusing  since most users were taught that the trigger point  determines the beginning of a sweep  as is the case for an analog scope      Therefore  in actuality  a delay time equivalent to the time of one sweep length is  added  so that it appears as if the trigger moment is displayed at the beginning of  the screen     The TRIGGER POSITION control allows you to change the time delay between  the trigger point and the digital acquisition  If set to zero  the acquisition is stopped  after a time equivalent to one sweep length has been added so that the trigger  point appears at the beginning of the trace     Any of the inputs can be used as the trigger source  The source is selected with  the keys    TRIG1  TRIG   or EXT TRIG    in the respective control sections  The  same keys are used to toggle between the positive       and negative  Y  slope   In the digital mode  dual slope      is available  The oscilloscope must be set to   REAL TIME ONLY    yes in the TB MODE menu  Dual slope triggering can be  se
59.  no full period        3rd mesial crossing  last sample if no full period   S n    nth sample  j lt  n lt  k     FUNCTION REFERENCE 5   93         rms  ROOT MEAN SQUARE VOLTAGE   Measures the rms value within one period  related to the ground level of the  signal  If no full period is present  all input samples are included in the  calculation        rms   x S n      k j 1    where  j  st mesial crossing  first sample if no full period        3rd mesial crossing  last sample if no full period   S n    nth sample  j lt  n lt  k       min  MINIMUM VOLTAGE   Measures the minimum voltage level of the signal  including undershoot  It is  calculated over all samples of the displayed signal     min   min  S n    where         nth sample              MAXIMUM VOLTAGE   Measures the maximum voltage level of the signal  including overshoot  It is  calculated over all samples of the displayed signal     max   max  S n    where         nth sample      pkpk  PEAK TO PEAK VOLTAGE   Measures the peak to peak voltage  the difference between the absolute  minimum voltage and absolute maximum voltage of the signal      pkpk   max   min      low  LOW LEVEL   Measures the low level of the signal  The number of samples for each  amplitude level is counted  A peak in the number of samples at one level  indicates the low level and is referred to 0  level  If no such point is present   e g   sine wave   it is equal to the minimum voltage       high  HIGH LEVEL   Measures the high level of the signal 
60.  number  Yrng Y range 65535 number  dT corr dT correction 4688E 04 number  N sample count 512 number  Y 1  sample 1 9603 binary  Y 2  sample 2 9612 binary  YIN  sample N  9599 binary    The following definitions and formula   s that enable you to interpret the data above   The same definitions  symbols and formula   s will be used to describe the discrete  mathematical functions of the MATH  processing functions     THE CPL PROTOCOL 6   39       Definitions    A waveform can be described as a set of two one dimensional arrays  S 1   N   and T 1   N   The index n corresponds with the number of the samples and is an  integer value in the range 1   N  The waveform of a sampled signal describes for  a range of sample moments  T 1    T N   usually expressed in seconds  the value  of the signal at these moments  S 1    S N   usually expressed in volts     A waveform is sometimes expressed in other units  An FFT waveform for example  gives for a range  of frequencies  T 1    T N   expressed in Hz  the signal  amplitude of these frequency components  S 1    S N   expressed in dB   Another  example is the possibility to multiply an input voltage with a floating point value of  another dimension     Formula s    S n  S n  is the value of sample n  expressed in Y units   Y units are usually expressed in volts     S n     Yz    Y n    Yr     Yu    Sensitivity Usually the sensitivity is expressed in volt div  In the oscilloscope   the range of the Y n  values  Yrng  always equals 2 16   6
61.  peak  peak to peak  low level  high level    The following time measurements are available   frequency  period  pulse width   rise time  fall time  duty cycle    The    delay    measurements include the time delay between leading or trailing  edges between any two channels or traces   Channel or trace must be on the  display     MEASURE MEAS1 RMS      Select  volt measurements     Turn the TRACK control to select    rms        Observe that the screen displays the result of the rms measurement   ch1 rms      V        HOW TO USE THE INSTRUMENT 4 51       MEASURE MEAS 2 FREQ    Press the RETURN    softkey to return to the  MEASURE  menu     Turn on MEAS 2     MEAS 2 performs a frequency measurement on the same signal  The screen  displays this result as    ch1 freq      KHz        Remove the Probe Adjust signal from channel 1    Because of the absence of an input signal  no frequency can be measured  This  results in the display  ch1 freq      Hz        MEASURE DELAY      Connect the Probe Adjust signal to channels 1 and 2      Turn      channel 2      Shiftthe signal on channel 1 to the upper half of the screen and the  signal on channel 2 to the lower half of the screen      Press the softkey next to MEAS1     The SELECT MEAS1 submenu is displayed       Press the first softkey to select  delay     SELECT                                  The delay measurement is MEAS 1   performed automatically  vot     Observe that the first delay    measurement is linked to the   f j CJ  cha
62.  proper ly   trained personnel only  Adjustment  maintenance and repair of the exposed  equipment shall be carried out only by qualified personnel     1 2 SAFETY PRECAUTIONS    For the correct and safe use of this instrument it is essential that both operating  and service personnel follow generally accepted safety procedures in addition to  the safety precautions specified in this manual  Specific warning and caution  statements  where they apply  will be found throughout the manual  Where  necessary  the warning and caution statements and or symbols are marked on  the apparatus     1 3 CAUTION AND WARNING STATEMENTS    CAUTION  Is used to indicate correct operating or maintenance  procedures in order to prevent damage to or destruction of the  equipment or other property     WARNING  Calls attention to a potential danger that requires correct  procedures or practices in order to prevent personal injury     1 2 OPERATORS SAFETY       1 4 SYMBOLS    Read the safety information in the manual       Earth     C     Conformit   Europ  enne     oy Recycling information   aw    1 5 IMPAIRED SAFETY PROTECTION    The use of the instrument in a manner not specified may impair the protection  provided by the equipment  Before use  inspect the instrument and accessories  for mechanical damage    Whenever it is likely that safety protection has been impaired  the instrument  must be made inoperative and be secured against any unintended operation  The  matter should then be referred to q
63.  result  trace is stored as a new trace in a separate register  This implies that the original  waveform or trace is never disturbed by the process  allowing you to  experiment   with different filter factors     A typical use of this digital low pass filter is to suppress noise even after a signal  acquisition  Since the FILTER function is a post acquisition process  it can also be  used on single event waveforms     Any newly acquired trace or previously stored trace can be used as the source for  the filter process     The result trace is automatically written in memory location m1 for MATH1 and m2  for MATH2 and instantly displayed on the screen     For each sample point of the trace      1   cos X  weighted sum is calculated over  a window of N samples  convolution   N is adjusted with the A control  The window  value is displayed in the  MATH FILTER PARAM  menu  The resulting cut off  frequency is the result of the sample rate set by the time base and N  The  3 dB  point is displayed in the bottom area of the screen     5 36 FUNCTION REFERENCE       Key sequence     MATH     MATH1 2  TRACK      Control to select the filter function from the MATH 1    or MATH 2 menu     A   O Control to select the source signal     Toggle key to switch selected function on or off           TRACK  PARAM Control to select the number of samples for the filter  window   DISPLAY  SOURCE  yes no    Toggle key to switch the display of the source trace  on or off     576767       Remote commands
64.  selected  Once selected  the results are  automatically updated with each new signal acquisition and the result is displayed  in the top left corner of the screen    Measurements can be performed on live signals or signals stored in any of the  registers  It is possible to perform measurements on the part of the waveform  between the two cursors  This function is called cursor limited measurements and  is turned on via the key sequence MEASURE  gt  gt  CURSOR LIMIT  amp  STATIST  gt  gt   CURSOR LIMITED yes  Cursor operation is done via the TRACK and A controls  and via the CURSOR menu     There are three measurement dimensions  amplitude  volt   time  and delay  The  lower part of the menu reflects the type of measurement selected     5 48 FUNCTION REFERENCE       The following measurements are available     volt  dc  rms  min peak  max peak  pk pk  low level  high level  overshoot   positive and negative   preshoot  positive and negative    time  frequency  period  pulse width  rise time  fall time  duty cycle   delay  channel to channel on leading or trailing edges    Refer to the functions TIME MEASUREMENT  VOLT MEASUREMENT  and  DELAY MEASUREMENT for more information     Execution of these measurements can also be controlled by pressing the  command switch on the measuring probe in the TOUCH  HOLD  amp  MEASURE     mode   Refer to the appropriate function for more information      Appendix E gives the complete menu structure of the MEASURE menu     All calculated measur
65.  the screen shows the message    HARDCOPY BUSY      DONE     and  during a print or plot action  all front panel keys except HARDCOPY are inactive   Pressing the HARDCOPY key again stops the hard copy action  Before  continuing  wait for the message    HARDCOPY DONE        5 54 FUNCTION REFERENCE       Interface    The instrument is equipped with an RS 232 Interface as standard  This interface  can be used with an RS 232 printer or plotter     The IEEE 488 2 Interface is available as factory installable option  This interface  can be used with IEEE 488 compatible printers or plotters     For correct functioning  correct interface parameters must be set  This is done in  the menu UTILITY  gt  gt  PRINT amp PLOT amp CLOCK     Press the IEEE   RS232 softkey to select the IEEE or the RS232 interface     Printer or plotter  This selects the menu to set up a printer or plotter being used     Printer type Supported printers  HP2225  FX80  LQ1500   HPLASER  150 dpi   HP540 DeskJet or  compatibles  The TRACK control is used to make  the selection     Plotter type Supported plotters  HP7440  HP7550  HP7475A   HP7470A  HPGL  Can be used to import  screenplots in a suitable word processing files    PM8277  PM8278 and dump m1  This selects a  trace dump to arbitrary waveform generator  PM5150  The dump action is started with the front  panel key HARD COPY or PLOT  Data transfer is  possible via RS232 or IEEE option  The TRACK  control is used to make the selection     Layout    A submenu 
66.  to adjust the end of the range     o  gt    5            89   8a   88    5 44 FUNCTION REFERENCE       MAGNIFY HORIZONTAL    Description     In the analog mode  the MAGNIFY key pair switches between the normal trace  and horizontal expansion of the trace by a factor of 10  The maximum timebase  speed is then increased from 20 ns div to 2 ns div     In the digital mode  the same MAGNIFY key pair gives horizontal expansion of  x1  x2  x4 up to 32 times for detailed viewing of captured signals  This results in  an expansion of the timebase range from 2 ns div to 62 5 ps div  The MAGNIFY  keys also allow horizontal compression of up to 32 times so that full records of 8K   32K optional  can be displayed on a single screen without loss of information   The current magnification factor is temporarily displayed on the screen when you  are making a change  When the magnify function is active  the magnification  factor is permanently displayed on the screen     In both analog and digital modes  the X POS control positions the trace with respect  to the screen to display the part of interest  When you adjust X POS  a bargraph is  temporarily displayed to show which part of the trace is currently being displayed             B               X  Lok   2  2N                             4      5   576672                    Function of the timebase magnifier        X POS control    FUNCTION REFERENCE 5   45       Key sequence     Key pair to adjust the horizontal magnification   Analog mode 
67.  to do this      Press the TRIGGER key once again      Press the TEXT OFF key    The TEXT OFF key operates in a 1 2 3 cycle  and allows you to blank  the bottom text line as well              RIGGER    MAIN TB    cht  i                   TEXT OFF  levellpp                   EN                                                            YY      CHI 200           ITB  2006   CH1  CHf 200mV Y        200us   CH T                                                                                                                         TEXT OFF                T6679    S    Step 2 Use both methods to familiarize yourself with turning the menus and  the bottom text line on and off     GETTING STARTED 3 17       3 12 CURSOR OPERATION    Cursors are used for accurate amplitude or time measurements of the signal                gt          00000000        oooe  O omm          CURSORS  TRACK                                                 E     ST6431  9303    Figure 3 9   Cursor setup    Step 1 Before you continue  reset the instrument with the STANDARD  SETUP  To do this  press the STATUS key and TEXT OFF key  simultaneously  Now the instrument is set in the default condition and  operates in analog mode     Step2 Press the AUTOSET key     Step 3 Press the CURSORS key to enter the cursors menu      The menu is now displayed on the screen and the  cursors are turned on  on off    Step 4 Use the second blue softkey from the top to select za    one of the three cursor modes       Amplitude cur
68.  to turn the TRIGGER  POSITION control clockwise  a positive delay between trigger point  and acquisition is set  The delay is no longer read out in divisions of  pretrigger information  but in seconds  or fractions of seconds to  indicate how much delay is used     3 16 GETTING STARTED       3 11 MORE ADVANCED FEATURES    All basic functions are accessed by dedicated keys for fast and easy operation   Some of the more advanced features are menu operated  Menus are called up by  pressing one of the keys identified with blue text on the front panel  After you  press one of these keys  a menu is displayed on the right side of the screen  This  menu gives you access to the more advanced functions of the oscilloscope  Use  the blue softkeys to the right of the screen to select the desired functions  the  selected function is indicated by the highlighted text     TRIGGER WITH  MENUKEY    TRIGGER    C                    0000                            oJ o           a                 nus      o3   9505045095                                                          9 9 Qe                                              000                               gloosoo                ST6433    SOFTKEYS 9303    Figure 3 8          keys and softkeys    Step 1 Press the key marked TRIGGER   Check that the TRIGGER MAINTB    menu is displayed at the right side  of the screen     After changing the setting  you can deactivate the menu again to use  the full screen for the signal    There are two ways
69.  trace on the display  even with  no signal applied to the oscilloscope inputs  Refer for a list of the standard setting  to the section  STANDARD SETUP FRONT PANEL RESET  of Chapter 5     DISPLAY ADJUSTMENT    Before going through the examples that introduce you to   features of this oscilloscope  adjust the display as follows      Turnthe TRACE INTENSITY control for optimum trace  brightness          Turnthe TEXT INTENSITY control for optimum     display brightness of the text    m   Adjust trace alignment with the graticule by using the      screwdriver control TRACE ROTATION      Turn the FOCUS control until a sharp trace is  obtained      You can turn the GRATICULE ILLUMINATION control  to illuminate the graticule lines as desired           TRACE  INTENSITY    TEXT  INTENSITY    TRACE  ROTATION    GRATICULE                   ST6156A    HOW TO USE THE INSTRUMENT       DISPLAY LAYOUT    The following illustration shows the layout of the display with a maximum amount    of text     Most text is active only when specific functions are activated     MENU HEADER                                                                                                                                                             CHANNEL IDENTIFIER  AND GROUND LEVEL CURSOR READ OUT l  INDICATION  CH1  1     UA  CH2  2        EXT  E  C3  REGISTERS  ere  __                    TRIGGER LEVEL      INDICATION pr  MTB  T                      ER           MAX  2 LINE  EVENT  E  ES Ux B     s SETTING
70.  values are measured over the total number of acquisitions for that  particular measurement and are updated instantly  The statistic measurement    values are as follows       absolute minimum value    absolute maximum value    mean value    Key sequence        volt  time  MEASURE MEAS  2  425   TRACK       A  on off    CURSOR CURSOR          LIMIT amp  LIMIT   STATIST yes no  STATIST   on off       576770  9312    Remote commands     Toggle softkey to select time measurements in menu  MEAS1 or MEAS 2     Control to choose the time measurement     Control to select the signal on which the measurement  must be made     Toggle softkey to switch the measurement on and off     Toggle softkey to switch cursor limited measurements  on and off     Toggle softkey to switch the statist function on and off     CPL  QM  Command to query a measurement     Refer to Chapter 6 for full details        5 78 FUNCTION REFERENCE       TOUCH  HOLD  amp  MEASURE     MODE    Description     The TOUCH  HOLD  amp  MEASURE     mode is a quick way to freeze the trace and  to display four main measurements instantly  This is done by pressing the  COMMAND switch on the measuring probe of the required channel     The four measurements are  dc voltage level  peak to peak value  rms value  and  frequency  The four values are displayed in the upper left corner of the screen     This function is a fast and easy way of troubleshooting  The COMMAND switch is  always where it should be  near the place to be meas
71. 20  130  140  150  160  170  180  190  200       210  220  230  240  250  260  270  280  290  300  310  320  330  340  350  360    370  380  390  400    10  20  30  40  50  60       Sod BoB oa as    COUNT 512   PRINT  Generating trace samples  DIM WAVE  2 COUNT    FOR I 1 TO COUNT                     Number of samples    Display operator message    Declare waveform array    For all samples    SAMPLE   INT 19200 SIN I 40 744     Generate sine value  LSB   SAMPLE AND  amp HFF   Calculate Least Sign  Byte  MSB     SAMPLE LSB    256  AND  amp HFF   Calculate Most Sign  Byte  WAVE 2 I    LSB   Put LSB in array  WAVE 2 I 1    MSB   Put MSB in array  NEXT  PRINT  Transmitting trace administration         Display operator message  PRINT E PW  4l    Program waveform  m4 1   PRINT   m4 l     Trace name  ignored   PRINT           Y_unit  Volts   PRINT         X unit  seconds   PRINT WEST   Y zero  3 V offset   PRINT   8 625E 6    X zero  ignored   PRINT  18 13   3    Y resolution  PRINT  1   6    X resolution  PRINT 765535    Y range  ignored   PRINT   E25 01 1994     Date  PRINT 14321956800     Time  PRINT pol SE 33   Delta T correction  PRINT 505   Min Max trace off  PRINT poe   Mult shot tot  PRINT 525   Mult  shot nr  PRINT prom  mo     Reserved field 1 2   ignored   PRINT  1 COUNT   Number of 16 bit samples  GOSUB 1000   Sync on acknowledge  SUM 0  PRINT  Transmitting trace samples      Display operator message  FOR I 1 TO 2 COUNT  72 bytes per sample  PRINT  1 CHRS  WAVE  1      S
72. 5535   which corresponds with 10 24 divisions on the screen  This gives a  maximum vertical resolution of 6400  levels  per division   Sensitivity is expressed in Y units division     Sensitivity   6400   Yr   Yu div    Offset Offset is the vertical screen position of the signal ground level   indicated by  1    excluding Y pos shift   The offset can be adjusted  with the front panel Y POS controls  For calculated traces the offset  can be adjusted with the delta controls in the MATHematics menu   Offset is expressed in Y units  volts      Offset    Yz   Yu    Note  Traces can also be positioned with the Y pos control in the  RECALL menu  but this does not affect the offset  The  reason is that Y pos is a pure display function  which  therefore does not affect the signal sampling  nor the  calculation of a trace using a mathematics function     T n  T n  is the sample moment  expressed in X units   X units are usually expressed in seconds                Xz    n 1    Xr     dT corr   Xr     Xu    6 40    THE CPL PROTOCOL       Time base    Delay    Notes     In case of a sampled signal the horizontal scale is called the time  base  The time base setting is usually expressed in  seconds division  Without magnification or compression the  oscilloscope always displays 50 samples per horizontal division     Note         horizontal scale as well as the FFT scale can differ  In  that case it is expressed in X units per division     Time base   50   Xr   Xu div    Delay is the time d
73. AC DC GND coupling  capabilities    In VERTMENU each channel is set to 500 or 1 MQ input impedance     4 2 HOW TO USE THE INSTRUMENT       PM3390B 200 MHz 2 Channel Oscilloscope    The PM3390B has the same capabilities as the PM3394B on the channels 1 and  2  The channels 3 and 4 are replaced by an external trigger channel  This channel  can only be used as an additional trigger input channel  Signal manipulation as in  the full channels 1 and 2 is not possible  The external trigger signal can be  displayed by using the function TRIG VIEW     PM3384B 100 MHz Full Four Channel Oscilloscope    Bandwidth is 100 MHz   The PM3384B has the same capabilities as the PM3394B  VERT MENU only  offers BW LIMIT selection     PM3380B 100 MHz 2 Channel Oscilloscope    The PM3380B has the same capabilities as the PM3384B on the channels 1 and  2  The channels 3 and 4 are replaced by an external trigger channel as in the  PM3390B  VERT Menu offers only BWLIMIT selection     PM3370B 60 MHz 2 Channel Oscilloscope    The PM3370B has a bandwidth of 60 MHz  All other features are equal to those  of the PM3380B     HOW TO USE THE INSTRUMENT       FRONT PANEL LAYOUT                                                                                                                                                                AUTO SET DELAYED CONTROLS HORIZONTAL  SCREEN DISPLAY AND OTHER CURSORS TIME BASE FOR DIGITAL AND TRIGGER  CONTROLS AREA UTILITIES   CONTROLS CONTROLS MODE CONTROLS    7        FRUT
74. AL REAR PANEL CONNECTIONS    SHIELD SRQ NDAC DAV DIO4 DIO2    ATN            NR  FD    E01 plo3 DI01    CH1 Y OUT  BNC output socket with a signal derived from the  Channel 1 input signal     MAIN TB GATE   BNC output socket with a signal that is  high  when  the Main Timebase is running and  low  for the other  conditions     DTB GATE   BNC output socket with a signal that is  high  when  the Delayed Timebase is running and  low  for the  other conditions     EXT TRIG  only in 4 channel models   BNC input socket used as an extra external trigger  input for the Main Timebase    IEEE 488 2 BUS OPTION                ot                           LOGIC GND Gi  GND 10                            IGND GND GND REN DIO7 DIOS  9 7       IND GND DIO8 DIO6  8 6    If installed you will find here the input output  socket to connect the oscilloscope to an  IEEE 488 interface                 The external conductor of the BNC sockets  and the screening of the interface bus  connectors are internally connected to the  protective earth conductor of the three core  mains cable  The external conductor of the  BNC sockets and the screening of the  interface bus connectors must not be used as  a protective conductor terminal     OPERATORS SAFETY 1 1       1 OPERATORS SAFETY    ATTENTION         instrument is designed for indoor use only   Read this page carefully before installation and use of the  instrument     1 1 INTRODUCTION    The instrument described in this manual is designed to be used by
75. ALLATION INSTRUCTIONS                       2 1   2 1 SAFETY INSTRUCTIONS                                  2 1   2 1 1 Protective earthing                                 2 1   2 1 2 Mains voltage cord  mains voltage range and fuses       2 1   2 2 MEMORY BACK UP BATTERIES                           2 3   2 2 1 General information                                2 3   2 2 2 Installation of batteries                              2 3   2 3 THE FRONT COVER                                      2 3  2 4 HANDLE ADJUSTMENT AND OPERATING   POSITIONS OF THE INSTRUMENT                          2 4   2 5 IEEE 488 2 IEC 625 BUS INTERFACE OPTION                2 4   2 6 RS 232 C SERIAL INTERFACE                             2 5   2 7 RACK MOUNTING                                       2 5   2 8                                                               diu eui E EXE 2 5    CONTENTS IX       3 GETTING                                                   3 1  3 4 FRONT PANEL LAYOUT                                  3 1  3 2 SWITCHING ON THE INSTRUMENT                         3 2  3 8 SCREEN CONTROLS                                     3 3  3 4   AUTO SETUP  oor ental Aiea 3 4  3 5 MODE SWITCHING BETWEEN ANALOG AND   DIGITAL OPERATING                                         3 6  3 6 VERTICAL                                                  3 8  3 7                  SETUP                                  3 11  3 8 MAGNIFY                                                   3 12  3 9 DIRECT TRIGGER 
76. AY J DISPLAY A A  Y pos    A Y pos       x xxD     x xxD           X pos       xx xxD 4  trace 7 trace 5  register                     RETURN 4 NEUE             1 SETUPS MENU STRUCTURE       Appendix H  SETUPS menu structure                            SETUPS    CLEAR CLEAR  Se SETUPS SETUPS         TRACK  SETUPS  TRACK CONFIRM CONFIRM  o 1  e S1    e s2    recall    undo     yes yes I   save      ARE YOU  RUNE  SURE  4 PROTECT   TEXT      44    CLEAR amp  1  PROTECT    no no       E    TRACK          ST6746  9303    TB MODE MENU STRUCTURE J 1       Appendix J  TB MODE menu structure                                Temone ANALOG   TB MODE EVENT     DELAY   1 7       auto    on off    o   trig J J TRACK     single COUNT            10220         1 CHANNEL         1 2         a  ANALOG   1                    N 1 499 8mV     1       P      alt chop RETURN 4   1     L     1   i   1  mul DIGITAL         TB MODE ACQ  LENGTH  ch     512 pts  ch    2k pts  2ch    4k pts    ich    8k p               n  4      79      2  1    REALTIME  NLY               yes no    EVENT  DELAY    ACQ  LENGTH                         1  OPTIONAL 976563  9303       K 1    TRIGGER menu structure       Appendix K    TRIGGER menu structure    TRIGGER    TRIGGER             TB    ANALOG           TRIGGER    1  X only in SINGLE mode  when  REAL TIME ONLY     in TB MODE menu is yes         1  1  1     1  1  1  1     1  1  1  4  1  1  1  1  1  1  1  1  1  1  1  1  1       ANALOG MODE                     TRIGGER TRIG
77. Autoranging  CombiScope   Instrument    PM3370B PM3380B PM3390B    PM3384B PM3394B    Users Manual    2 1  Nov 1998                                                                                                                                                                                                                                                                   mc   AUTOSET CAL SETUPS UTILITY    ANALOG ACQUIRE SAVE RECALL MEASURE MATH DISPLAY HARD COPY  Ce OOCQO CGQQ9OOOODOOCOO  STATUS  qpacx CURSORS     HOLD OFF TRIGGER MAGNIFY     4XPOSP  x    6 9936  INTENSITY LOCAL RUNJSTOP FANGE SINGLE ARMD  TRIGGER    TRIGGER  DELAYED TIME BASE DELAY POISITION      LEVEL         sTIME DIV ns      MODE w VAR w  TRACE LJ  ROTATION    VERT MENU ALT CHOP  Pos     Pos Pos Pos  TRIG  TRiG2 TRGS TRIG                  0   0        AUTO 0 AUTO AUTO AUTO  AMPL RANGECHIAGH2 AMPL RANGE INV AMPL RANGECHSACHY AMPL RANGE INV  mi  GRATICULE   nl 4     ILLUMINATION    T ACDC   ACDC            E ACDC                  D           GND S             5                4  g E    l  v v v     POWER Probe Adjust  Bl    a     AMA 25pF        TL  a  ST7721A       FLUKE     IMPORTANT    In correspondence concerning this instrument please give the model number and  serial number as located on the type plate on the rear of the instrument     NOTE  The design of this instrument is subject to continuous development and  improvement  Consequently  this instrument may incorporate minor  changes in detail from t
78. D        The following selections are possible      Trigger level indicators  TRIG IND  for MAIN TB and DEL D TB  Level pp and  TV mode must be off  The horizontal marker is the level position      Ground level indicators  GND IND  for each channel are visible only when  channel is on  The horizontal line is the ground level position      MAIN TB int contrast between MAIN TB trace and intensified part is adjusted  with the TRACK control      USERTEXT leads to a submenu for usertext  Refer to the function  USERTEXT for more information     The following audio signals are activated with the key sequence    UTILITY  gt  gt   SCREEN  amp  SOUND  gt  gt  SOUND         BEEP on off  the signal sounds to indicate messages or AUTOSET      CLICK on of  the signal indicates operation of keys and rotary controls     Key sequence        TRIG IND  SCREEN amp  ff  UTILITY SOUND         Toggle softkey to display trigger level  indication   GND IND  on off    Toggle softkey to display ground level  E indication   BEEP  SOUND on off          2  Toggle softkey to select warning signal   CLICK  on off Toggle softkey to select rotary control key  Ly  operation signal   TRACK  Control to adjust the intensity ratio in    mtbi     mode   USERTEXT               oe Access to editing menu for USERTEXT     576087  9303    5 92 FUNCTION REFERENCE       VOLT MEASUREMENTS    Description     Voltage measurements can be made using the cursors or using automated   calculated measurement routines  Cursors oper
79. DIGITAL MODE   ACQUIRE     7 ACQUIRE  TRACK  AVERAGE      256    PEAK         on off    ENVELOPE  ono          1 CURSORS menu structure       Appendix B  CURSORS menu structure    ANALOG  EE ANALOG___ Cy SS DIGITAL                          CURSORS                           V1 V2        AV ratio   AV 100    cursor cursor  track track  yes no yes no    RETURN 4       i    ST6567 1  9312    CURSORS menu structure B 2           CURSORS               cursor  track    cursor  track        ST6567 2  9312    DISPLAY MENU STRUCTURE       Appendix C    DISPLAY menu structured    DISPLAY    ANALOG    DISPLAY              1      1      1  1  1  1           4   5  1  1         1  1         1  1  1  1    ANALOG MODE                       X DEFL   on off 1  A SE           2     ch3 7  ch4     line d  RETURN   4             DISPLAY edi           l  text  gt        DIGITAL MODE           TEXT          RETURN 4          TRIG IND   on off      GND IND 1  on off    USER          TEXT  gt         delete    insert          ENTER 4    ST6560  9303    D 1 MATHEMATICS MENU STRUCTURE       Appendix D  MATHEMATICS menu structure    MATH       MATH n                     MATH 2  m2  d    filter  acq    on off  PARAM      i    DISPLAY  SOURCE                  Ex                                          1 Div    WINDOW   21 3mU    31  D       samples          OFFSET     26 8mU                      RETURN 4          576748  9303    MEASURE MENU STRUCTURE       Appendix E  MEASURE menu structure    MEASURE  
80. ES                   5 73                                               N 1  Readout 5 CURSOR READOUT                  5 22  Real time clock 5   PRINTING AND PLOTTING             5 55  Recall 5   CHANNEL TRACE SELECTION          5 17  5 SETUPS  oo  ices REV des 5 67  Rise time 5  TIMEMEASUREMENTS                5 76  RMS voltage 5          MEASUREMENTS                5 92  RS 232 5   PRINTING AND PLOTTING             5 54  5 REMOTE CONTROL RS 232            5 59  Run bu                                            5 61  RS 232 cables                                            N 1       FUNCTION INDEX 1 7             Refer to chapter function Page  5  Safety 1 OPERATORS                               1 1  Sampling 5  TIMEBASE MODES                   5 73                                             N 1  Save 5    SETUPS                                5 67  Screen 5   SCREEN CONTROLS AND GRATICULE   5 62  5 SCREEN MESSAGES                 5 63  5 UTILITY SCREEN  amp  SOUND            5 90  Setting the clock 5 PRINTING AND PLOTTING             5 55  Setups 4 16 AUTOSET AND SETUP UTILITIES       4 70  b  SOBEDUBSLEZI eI T A E 5 66  5 SETUPS SEQUENCE                  5 68  Single shot 5  TIMEBASE MODES                   5 73  Sound 5 UTILITY SCREEN  amp  SOUND            5 90  Statistics 5 TIME MEASUREMENTS                5 77  5 VOLT MEASUREMENTS               5 93  Status 5 STATUS SCREEN                     5 70  Subtract 5   ADDINVERT SUBTRACT               5 4  5 SUBTRACT  MATHEMA
81. GER  MAIN TB MAIN TB              edge iva edge tv   edge tv    field 1 field 1   m   field 2 Fo field 2    lines lines TRACK   level pp LINE NBR   on off 32        noise   on off     pos neg       ac ds VIDEO          SYSTEM P    hf rej hdtv               DIGITAL MODE     TRIGGER  MAIN TB       TRIGGER TRIGGER  MAIN TB MAIN TB    ST6566 1  9303    level pp is then replaced by     trigger gap    0 50 div          TRACK                 TRIGGER menu structure           pattern  glitch            glitch  LHxH    glitch  LHxH                                enter enter   exit exit 4   if  gt t1 if  gt t    if   t2 S if   t2  i  gt    range range      TRACK TRACK  RANGE   t   X XXXMS   X XXXMS xx xxms A  A       ST6566 2  9303    L 1    UTILITY menu structure       Appendix L  UTILITY menu structure    UTILITY                                                                                                                                                             TRACK  UTILITY UTILITY EDIT    UTILITY    SCREEN  amp     USER A    SOUND  SOUND TEXT        AUTOSET  gt   TRIG ND on off peu    GND IND   1 CLICK    PROBE  gt   on off Space   on off    EO souno    1       TRACK      REMOTE j  MTBint o  delete  PRINT  amp  USER i      PLOT amp CLK  gt   TEXT    inset   1  MAINTE  J REFER      J 5 J  NANCE          SERVICE RETURN 4 RETURN 4 RETURN  gt   MANUAL  UTILITY UTILITY UTILITY  L           amp  PRINT  amp  PRINT  amp   PLOT PLOT PLOT  printer   printer printer  plot      plot clk p
82. IG1  and  INV keys     GETTING STARTED 3 11       3 7 TIMEBASE SETUP    The next step is the adjustment of the main timebase controls  TIME DIV  X  POSition  and MAGNIFY keys                                                                                                                                                                                                                                                             soea MAGNIFY 4            8  oTo l KID     5 span        ninm RANGE     Ha o o      29                                   TIME DIV  fi       Qe Gm   2  ST6435  9312   Figure 3 6                  setup   Step 1 Press the AUTOSET key    Step 2 Use the TIME DIV keys on the right hand side of the instrument to  decrease or increase the number of periods of the signal on the  screen                            50015        100      Step 3 Select a timebase of 1 ms div    Step 4 Press AUTO RANGE and see the signal display change to a more  suitable timebase  The AUTO RANGE function automatically selects a  timebase that displays 2 to 6 signal periods    Step 5 Press the ANALOG key to switch the scope to the analog mode    Step6   Turn the X POS control to shift the signal horizontally  left or right     across the screen     3 12    GETTING STARTED       3 8 MAGNIFY  EXPAND     Step 1    Step 2    Step 3    NOTE     You can use the MAGNIFY keys to expand the signal on the screen   The  MGN  indication and the corrected timebase setting are displayed  in the text li
83. ILITY SOUND TEXT f  Toggle softkey to activate user text        text has to be edited     TRACK   O Control to determine the position where  A    Control to determine the character to be  added     Softkey to make space     Softkey to delete a character     insert          Softkey to insert a character     0    576055    Remote commands     CPL  PT  command to program user text   QT  command to query user text     Refer to Chapter 6 for full details     FUNCTION REFERENCE 5   89       UTILITY MAINTENANCE    Description     The UTILITY MAINTENANCE menu is used to calibrate the oscilloscope and for  repair and testing  Calibration data is protected by a password and by operation  of a pinhole switch that can be sealed  Calibration is of vital importance for the  instrument s high accuracy  The menu is meant for calibration and for use by  service technicians  and is therefore not explained in this Users Manual    An extensive description can be found in the Service Manual     5 90 FUNCTION REFERENCE       UTIL MENU    Description     The UTIL menu is used to make presettings for instrument settings that do not  need to be changed frequently     The operation of the AUTOSET key  For a description  refer to the AUTOSET  and AUTOSET USERPROG functions    Selections in relation to probes  Refer to PROBE UTILITIES function for details   Amount of information in CRT viewing area and audio signals  Refer to the  UTILITY SCREEN  amp  SOUND function  The UTILITY SCREEN  amp  SOUND  menu
84. If your oscilloscope is equipped with the IEEE 488 2 interface  it can be used in a  bus system configuration  The protocol used is SCPI  Standard Commands for  Programmable Instruments   For setup information  refer to the function  REMOTE CONTROL IEEE 488 2 in Chapter 5     The IEEE 488 2 interface is a factory installed option     INSTALLATION INSTRUCTIONS 2 5       2 6 RS 232 C SERIAL INTERFACE    Your oscilloscope is equipped with an RS 232 C interface as standard  The  interface can be used in a system for serial communication  The protocol used is  CPL  Compact Programming Language   CPL is a small set of very powerful  commands that can be used for full remote control  Detailed information about this  interface and the CPL protocol is given in Chapter 6 in this manual  For setup  information  refer to the REMOTE CONTROL RS 232 function in Chapter 5  Function Reference        2 7 RACK MOUNTING    The rackmount kit  PM 8960 04  allows you to install the oscilloscope in a  standard 19 inch rack     It is not necessary to open the oscilloscope itself to mount the rackmount kit   Installation can be done easily by the user     2 8 VERSIONS    The model number of your oscilloscope  e g  PM33     is indicated on the text strip  above the CRT  This model number is also represented by the digits 6  7  8 and 9  of the 12  digit code on the type plate on the rear panel  The  A or  B  series is  indicated by a 1 or 2 on the 5th digit     The instrument s serial number is also giv
85. L     Input voltage at 50Q input impedance is  too high   Function    INPUT IMPEDANCE       MATH 1  2  is activated while register M1   2  is protected    Functions  ADD    FILTER     MATHEMATICS    MULTIPLY  and    SUBTRACT      Function activated in digital mode  but is  only present in analog mode   Function  ANALOG MODE        Print or plot action has been aborted   Function PRINTING AND PLOTTING        Instrument is busy with a print plot action   Function PRINTING AND PLOTTING        Print plot or abort action has been done   Function PRINTING AND PLOTTING        Copying to the same register is not  possible     Selected source is not valid   FunSELECTIONctions    ADD     FILTER      MATHEMATICS     MULTIPLY    and   SUBTRACT      Pressed key is inactive when trace is  frozen   Function  RUN STOP      Average and Roll modes cannot be  combined   Function  AVERAGE  and TB MODE        Battery backup not possible because of  absence of batteries     FUNCTION REFERENCE    5 65                    ACQ  TOO LONG    NO DTB IN ROLL MODE    NO ENVELOPE IN ROLL MODE    NO EVENT DELAY  DUAL SLOPE    NO EVENT DELAY IN TV MODE    NO IND WITH THIS COUPLING    NOT PART OF X VS Y REGISTER    PLEASE FIRST SWITCH TO DSO    PLOT KEY   ABORT    PRESS 2 SEC FOR AUTOCAL    PROBE DETECTED  NO CHANGE    Del d TB only possible at acquisition  length of 512 sample points    Functions  ACQUISITION LENGTH    and   DELD TB       Del d TB and Roll modes cannot be  combined    Function  DEL D TB    an
86. LEVEL control to adjust the level   Turn the control  The precise position in relation to the maximum signal  amplitude  between  100   and  100    is displayed on the screen     SUMMARY   The previous steps covered the basic adjustments  Now you are ready to look  at the special features of the oscilloscope  This includes the use of the cursors    advanced trigger functions and using the second  delayed  timebase for signal   details     GETTING STARTED 3 15       3 10 PRE TRIGGER VIEW    One of the powerful features in the digital mode is the ability to capture and view  signal contents prior to the actual trigger  The amount of pretrigger information  can be as long as one full acquisition record  The trigger position is adjusted with  the TRIGGER POSITION control     Step 1 Turn the TRIGGER POSITION control counter clockwise  Now the  triggering edge shifts to the center of the screen  A trigger point marker   s  indicates the trigger point  The part to the left of the marker is called  pretrigger view  The pretrigger view is indicated in the bottom of the  screen  in divisions     TRIGGER  POSITION                                     5 000                                                                                                                                       ST6686  PRE TRIGGER  VIEW    Step2        TRIGGER POSITION control can also be used to adjust time  delay  Rotate the TRIGGER POSITION control clockwise until the  readout displays  0   When you continue
87. LL  Inhibits front key STATUS LOCAL     Refer to Chapter 6 for full details     FUNCTION REFERENCE 5 61       RUN STOP    Description     The RUN STOP button operates in the digital mode only  When the STOP  function is active  any new signal acquisition is stopped and the trace is    frozen      The status of the STOP function is displayed in the bottom right side of the screen     With the acquisitions STOPped  the following actions are still possible      Plot actions     Display changes  also Y POS      Cursor measurements     Calculated measurements     Mathematics and signal analysis     Settings of the acquisition system  actions that influence the contents of the  register are not possible       Data        be saved in memory     Front settings actions  also recalls     Pressing the RUN STOP key returns the oscilloscope to the acquisition mode that  was in effect before the key was pressed     Key sequence     RUN STOP       Key to stop the acquisition and to freeze the trace and to  start the acquisition again     5 62 FUNCTION REFERENCE       SCREEN CONTROLS AND GRATICULE    Description   The screen controls are located to the left of the CRT viewing area     Brightness of trace s  and text can be adjusted separately with the TRACE  INTENSITY and TEXT INTENSITY controls  Intensity of the trace s  can also be  determined by a voltage applied to the rear panel socket Z MOD  Refer to  Characteristics    Chapter 1 of the Reference Manual for input amplitudes     The FOC
88. MATHEMATICS        Recalling settings from an empty memory  location is not possible  Protection of an  empty front is impossible    Function SETUPS        Settings cannot be saved in protected  memory location   Function SETUPS        Save settings in the    std    front is not  possible   Function SETUPS        Set timebase to fastest Roll mode  position   Function       MODE        Maximum allowed traces is eight     Ten complete front panel settings can be saved into a battery backed up memory   This feature is useful for routine measurements  Setups are stored in memory  location labeled s1       10  These memories are accessible under the menu key  SETUPS  The menu structure is shown in Appendix H     Selection of the memories s1     s10 is done with the TRACK control  Memory  location  std  is a factory stored set of standard settings that can be used to put  the instrument in a defined state  For a detailed specification  refer to the    STANDARD FRONT function     FUNCTION REFERENCE 5 67       save  The actual set of settings is saved      the selected memory location     recall  The settings saved in the selected memory location become the actual  settings     undo  The settings previous to the last recall action become active again     CLEAR  amp  PROTECT  in the CLEAR  amp  PROTECT SETUPS submenu    PROTECT  on off can be used to protect a memory location by prohibiting save actions to this  location     clear all       the CLEAR  amp  PROTECT SETUPS submenu    cl
89. MATICS  MEASURE MENU  MULTIPLY    PEAK DETECTION  POSITION   POWER SUPPLY  PRINTING AND PLOTTING  PROBE UTILITIES    REMOTE CONTROL  RUN STOP    SCREEN CONTROLS AND GRATICULE  SCREEN MESSAGES   SETUPS   SETUPS RECALL SEQUENCE   STANDARD FRONT FRONT PANEL RESET  STATUS SCREEN   SUBTRACT  MATHEMATICS     TB MODE   TEXT OFF   TIME MEASUREMENTS  TOUCH HOLD  amp  MEASURE MODE  TRIGGER COUPLING   TRIGGER DELD TB   TRIGGER LEVEL   TRIGGER MAIN TB   TV TRIGGER    USERTEXT   UTILITY MENU   UTILITY SCREEN  amp  SOUND  UTILITY MAINTENANCE    VOLT MEASUREMENTS  X DEFLECTION    FUNCTION INDEX       INDEX    The overall index contains all function names and reference words in alphabetical  order  It refers to the relevant section and page number  mainly of Chapter 5   Function Reference   In this chapter more detailed information can be found     Entry    500  1 MQ    A            Acquisition  Add    Adjusting the clock  Alternate  Analog    Attenuator  Autocal  Automatic attenuators    Automatic timebase    Auto range  Autoset    Average    B  Bandwidth  Bar graph    C    Calibration  Channel  Characteristics  Checking  Chop    Refer to chapter function Page  5                                                  5 41  5 INPUT IMPEDANCE                    5 41  5 INPUT                                     5 40  5 TRIGGER COUPLING                 5 79  5   ACQUISITION LENGTH                 5 2  5   CHANNEL TRACE SELECTION          5 17  5 ADD INVERT SUBTRACT               5 4  5   ADD         
90. MEBASE MODES    Description     The Main Timebase  MAIN TB  can function in three different modes   AUTO   TRIG  or SINGLE  The choice is mainly determined by the frequency of the trigger  signal  The modes are explained below     AUTO In    auto    mode  the MAIN TB always gives a display  even when  no signal is present  It checks for the occurrence of the trigger  signal for some 100 ms  but if none is found  it automatically  starts a new sweep or acquisition  On repetitive signals this mode  is most common except for signal frequencies lower than 10 Hz  where unstable displays will occur     TRIG In the triggered  trig  mode  traces are displayed only when the  MAIN TB is actually triggered by a signal  This mode is  recommended for frequencies lower than 10 Hz because it  assures a stable display  The ARM   D indicator indicates if no  trigger occurs     SINGLE In single shot mode  the MAIN TB runs only once after a trigger   Pressing the SINGLE key resets the scope for single shot  operation  the trigger level is automatically set to 0 5 division  above ground  indicated by T    The ARM   D indicator lights when  the MAIN TB is waiting for a trigger  MAIN TB armed   The single  mode is used to capture and observe  photograph  events that  happen only once     In the digital mode the single mode can be used in combination with pre triggering  to observe and store single events  along with signal information occurring before  the trigger point     Auto  trig  and single 
91. ND    The input coupling after AUTOSET is ac  Since the Probe Adjust signal is a pulse  type signal with a 50  duty cycle  its mean value is at the signal   s 50  amplitude  level  When the input is ac coupled  the mean value will be displayed at the  ground level of the oscilloscope  As a result of this  the displayed waveform is  centered on the screen    AC coupling can be used to examine small ac components that are superimposed  on large dc voltages       Press the AC DC GND key once for DC input coupling     Since the Probe Adjust signal is a pulse type signal with a lower level of OV  and  a higher level of 600 mV  and since the oscilloscope adjusts the display position  of the ground level to coincide with the screen center  switching from ac coupling  to dc coupling results      an upward shift of the display position of the signal     The coupling sign in the lower left hand corner of the screen changes from    ac   to    dc      The ground level for each of the channels is indicated by a dash after the channel  identifier  i e    1   for the ground level of channel 1     0 5ms        600mV         ov    Input Signal                                           T          AUTO SET ag oc   2    A E    i HE  BHH    AC DC                                                                                                                         m   Les 2bon v        _                   200                     4 t    AC Coupled DC Coupled GND Coupled                          Pr
92. OMPARTMENT TYPE PLATE                        Q                                                                         CIC        0                0 i  AA    a         2  s    s        cee               AC 2            THEMES                                                                                      5   0000 0                                                                    YOUT ZMOD MTB EXT DTB  SE TRIG         STORAGE SPACE  OPTIONAL FOR MAINS CABLE 577665  L    ONLY IN  4 CH MODELS       Figure 2 1 Rear view of the instrument showing the mains input fuse holder  and back up battery compartment     When the apparatus is connected to its supply  terminals may be live  and the  opening of covers or removal of parts  except those to which access can be  gained by hand  is likely to expose live parts     The apparatus shall be disconnected from all voltage sources before it is opened  for any replacement  maintenance or repair     Capacitors inside the apparatus may still be charged even when the apparatus  has been disconnected from all voltage sources     Any maintenance and repair of the opened apparatus under voltage shall be  avoided as far as possible and  when inevitable  shall be carried out only by a  skilled person who is aware of the hazard involved     INSTALLATION INSTRUCTIONS 2 3       2 2 MEMORY BACK UP BATTERIES    2 2 1 General information    Memory backup is provided to store the oscilloscope   s settings when switched off  so that the instrument r
93. OS   QURE SAVE RECALL MEASURE MATT DISPLAY AARD           TATU TRACK A HOLD OFF TRIGGER MAGNIFY 4      5    wir c O Of 0 2820     cun        ong             sTIME DIVns                               _      m  Pos      5  Focus             Berma  Te  re ee GE    mu  ILLUMINATION       zpomy                  cht    T  er 8 0           aN AUG        gU BO ad  Ch omKO                        Stasi   SOFTKEYS AND PROBE ADJUST VERTICAL CONTROL TRIGGER SOURCE EXTERNAL TRIGGER  TEXT ON OFF OUTPUT AND CHANNEL AND SLOPE CONTROL AND  CONTROLS SIGNAL INPUTS CONTROLS CHANNEL INPUT    The controls on the front panel are grouped by function  In this chapter  a  description for each group of controls is given in the following sequence     Display and probe adjustment  Analog and digital modes   Vertical deflection   Horizontal deflection and triggering  Digital acquisition and storage  Advanced vertical functions  Advanced horizontal and trigger functions  Memory functions   Cursor functions   Measurement functions  Processing functions   Display functions   Delayed timebase   Hard copy facilities   AUTOSET and other utilities    see section 4 2  see section 4 3  See section 4 4  see section 4 5  see section 4 6  see section 4 7  see section 4 8                          see section 4 9                         See section 4 10   See section 4 11   see section 4 12   see section 4 13   see section 4 14   see section 4 15         see section 4 16    4 4    HOW TO USE THE INSTRUMENT       Study t
94. RE MENU                     5 47  5 VOLT MEASUREMENTS               5 91          Waveform parameters 6                                           6 36   Windows 5 DISPLAY                                5 30   X   X Deflection 5 DISPLAY                                5 30  5 X DEFLECTION                       5 94   X vs Y 5 DISPLAY                               5 30  5                                            5 94    
95. REFERENCE 5   83       Leveling in DEL D TB is adjusted with the A control  It is activated in the triggered  DEL D TB mode  The range is from  8      8 divisions     Trigger levels for MAIN TB and DEL D TB can be displayed  Refer to the  description of the UTILITY SCREEN  amp  SOUND function                                M                                                                              m  TRIGGER TRIGGER   LEVEL LEVEL  CE   TIME BASE  RANGE TRIGGERED  PEAK PEAK  SIGNAL  VALUE  ie Re me NOR           M         lt   ARMD  level                   gt  ACE  Pace  level        off    576791    TRIGGER LEVEL control ranges      level pp      and off  Key sequence     TRIGGER  LEVEL    Control for MAIN TB trigger level     level pp  TRIGGER on off    Toggle softkey to switch MAIN TB level pp on and off     starts  DTB trid d A  Control for DEL D TB trigger level     576775  9303    Remote commands   CPL  TA  Command for trigger Acquisition   Refer to chapter 6 for more details     5 84 FUNCTION REFERENCE       TRIGGER MAIN TB    Description     This section deals only with    edge    triggering of the MAIN TB  For TV triggering   Logic triggering or DEL D TB triggering  refer to the appropriate sections     In the analog mode triggering determines the start point of the MAIN TB sweep   The sweep starts at the moment the signal crosses the trigger level in positive or  negative direction  The slope is selectable     X   as is shown in the figure below     In the digital
96. REFERENCE 5 15       BANDWIDTH LIMITER    Description     The bandwidth limiter cuts the bandwidth of all vertical channels to 20 MHz and  makes noisy input signals look smoother  The bandwidth limiter does not affect  triggering  The following figure shows the effect of the bandwidth limiter     0dB    Effect of bandwidth limiter  30B                  BANDWITH        FREQ   Key sequence   BW LIMIT  VERT MENU      off Toggle softkey to switch the vertical bandwidth limiter    on off    576556  9312    5 16 FUNCTION REFERENCE       CALIBRATION AUTOCAL    Description     The CAL key is used to make a fine adjustment of the oscilloscope   s input  trigger   and timebase circuitry to achieve high accuracy even under extreme  environmental conditions such as very high or very low temperatures  Ina  workshop or laboratory environment  a fine adjustment once a week or even  every month is sufficient     It is recommended that you do the fine adjustment after the instrument has  warmed up  For a complete calibration  advised once a year or every 2000  service hours   a special submenu is available in the maintenance menu   Calibration data are protected by a keyword and a seal  Calibration should be  done by qualified personnel only  For details  refer to the chapter  Calibration  Adjustment Procedure  in the service manual     Attention  Calibration autocal data disappears after having removed back up  batteries while the oscilloscope is not powered by line     Key sequence     CAL 
97. RETURN 4  UTILITY 1  UTILITY                 REMOTE RS232  SETUP SETUP               TRACK  J 1200 O    BITS    7 8  IEEE TRACK  parity 3  ADDRESS      no odd  8 D  even 4  3 wire  J 7 wire J  RS232     XON XOFF  SETUP J      of J  RETURN 4 RETURN 4       M 1 VERTICAL MENU STRUCTURE       Appendix M  VERTICAL menu structure    VERT MENU       ST7412  9312    RS 232 CABLE CONFIGURATIONS N 1       Appendix N  RS 232 Cable configurations    This appendix supplies additional information about the RS 232 cable  configurations between the oscilloscope and a connected device  The  oscilloscope and most of the devices are Data Terminal Equipment  DTE     configurated     Communication lines     TxD   Transmitted Data   RxD   Received Data   RTS   Request To Send   CTS   Clear To Send   DTR   Data Terminal Ready   DSR   Data Set Ready    Modem Ready   DCD   Data Carrier Detect   S GND   Signal ground    F GND   Safety ground    A  Cable to printer plotter controller with software handshake    parameters     ATTENTION     NOTE      9 wire 7 wire    3 wire    XON XOFF on off    on   OSCILLOSCOPE DEVICE   DTE   DTE     SIGNAL 9 PIN 9 PIN 25 PIN   female     TxD          RxD    RfR    DTR  DSR    3   2   7   CTS 8  4   6   DCD 1  5    3  2  7  8  4  6  1  5    S GND                                             Oo 0 00 00 0 0       F GND   CASE CASE       5  6          The maximum Baud rate for the PM8277            8278 plotters is  4800 baud  For a higher Baud rate you must use a hardware
98. RS uo Lys        5 19  5 CURSORS READOUT                 5 22  D  d c  5 INPUT COUPLING                    5 40  DC voltage 5 VOLT MEASUREMENTS                5 91  Delay 5  DELAY Sr eL                  eth 5 24  Delay measurement 5 DELAY MEASUREMENT               5 26  Delayed timebase 4 14 DELAYED TIMEBASE                  4 62  5   DELAYED TIMEBASE                 5 27  5 TRIGGER DELD TB                   5 81  Differential mode 5   ADDINVERT SUBTRACT               5 4  Digital mode 4 3 ANALOG AND DIGITAL MODES          4 9  5 DIGITAL                                5 29  Display 4 2 DISPLAY AND PROBE ADJUSTMENT    4 06  4 13 DISPLAY FUNCTIONS                  4 56  5 DISPLAY                               5 30  Dual slope trigger 5 TRIGGER                                5 84  dump m1 5   PRINTING AND PLOTTING             5 54  Duty cycle 5 TIME MEASUREMENTS               5 76  E  Envelope 5 ENVELOPE 4205                                5 33  Event delay 5 DELAY                       yee eh ene 5 24  Expansion 5 ACQUISITION LENGTH                 5 2  5                            RR EL RES N 1  External trigger 5 EXTERNAL TRIGGER                 5 34  F  Fall time 5 TIME MEASUREMENTS               5 76  Filter 5                                                                        T PES 5 35  Frequency 5  TIMEMEASUREMENTS                5 75  Front Aale                              ss Lor foco elas    4 2  5 STANDARD                              5 69       FUNCTION INDEX 1 5
99. Reserved field   250 INPUT      5265   Reserved field   260 INPUT ARESSS   Reserved field   270 INPUT  RES4S   Reserved field   280 INPUT   COUNT   Sample count   290 DIM WAVE  2 COUNT    Declare waveform array  300 SUM 0   310 PRINT  Reading trace samples        Display operator message  320 FOR I 1 TO 2 COUNT  72 bytes per sample   330 WAVE  1   ASC  INPUTS  1   1      Read each sample byte  340 SUM   SUM WAVE  1    MOD  256    Calculate checksum   350 NEXT I   360 CHECKSUM ASC  INPUTS  1   1      Read checksum   370 IF SUM  lt  gt  CHECKSUM THEN GOTO 380   Test checksum    380  390    END    PRINT      Checksum error           Display error message    6 38 THE CPL PROTOCOL       Interpretation of waveform data    As aresponse to the QW command the oscilloscope sends a set of data describing  the waveform  The following example shows the interpretation of that data  The  example assumes the Probe Adjust signal to be applied to input channel 1     For information about MATHPLUS  refer to the supplement for the operation guide     EXAMPLE                                                                    Settings Attenuator 200 mV div  Coupling DC  Offset  1 5 div    300 mV offset   Time base 200 us div  Delay  2 50 div   500 us pre trigger view  Display  d   CH1  200                200   2 50       Data SYMBOL PARAMETER VALUE TYPE  Yu Y unit V string  Xu X unit S string  Yz Y zero 0 3 number  Xz X zero  0 0005 number  Yr Y resolution 3 125e 005 number  Xr X resolution 4e 006
100. S READ OUT  SYMBOLS  1   INVERT ON 1 2  CHANNEL 1   CHANNEL 2      DC COUPLED INPUT 1 2  CHANNEL 1   CHANNEL 2     AC COUPLED INPUT L          4 BWL  BANDWIDTH LIMITER       L   GROUNDED INPUT ALT  ALTERNATED DISPLAY OF TRACES      PROBE CORRECTION ON CHOP   CHOPPED DISPLAY OF TRACES  12   500 INPUT IMPEDANCE  ONLY VISIBLE WHEN CHANGING   SPECIAL FEATURES  TEMPORARELY DISPLAYED  Y ji Y v t MAGNIFY  CHi 102 VAL  1 2  TRIGGER SOURCE                          10 2 V   Lz BWL MTB  1 15u5 MG         f   POS TRIGGER SLOPE  EXT  10 1 VA       DTB Ons  1 99          1   NEG TRIGGER SLOPE  UHH                 5  7679    CHANNEL SETTING TIME BASE DELAY    TIME    HOW TO USE THE INSTRUMENT 4 7       MENUS TEXT OFF    A menu appears when a key with blue text is pressed   The menu functions can be selected by pressing the blue softkeys to the right side  of the screen       Press the ANALOG key to select the digital mode     Press the DISPLAY key to activate the menu     DISPLAY    Each menu starts with a menu name  which  corresponds with the key that was used to select the  menu  This is sometimes followed by a second name of  the softkey that initialized a submenu        The windows function can be switched on and off by  pressing the corresponding blue softkey        indicates that a selection must be made with the  TRACK control     The arrow      behind TEXT indicates that there is a  submenu     linear      The bottom softkey switches  toggles  between the  sine    three functions  
101. SETUP                                3 13  3 10 PRE TRIGGER VIEW                                    3 15  3 11 MORE ADVANCED FEATURES                            3 16  3 12 CURSOR OPERATION                                   3 17  3 13 MORE ADVANCED TRIGGER FUNCTIONS                  3 19    3 14 MORE SIGNAL DETAIL WITH THE DELAYED TIMEBASE     3 20  3 15 TRACE STORAGE                                      3 22    4 HOW TO USE MORE ADVANCED FUNCTIONS    OF THE INSTRUMENT                                  4 1  41 INTRODUCTION    obli b I DD EDS SA 4 1  4 2 DISPLAY AND PROBE ADJUSTMENTS                      4 5  4 3 ANALOG AND DIGITAL MODES                            4 9  44 VERTICAL DEFLECTION                                 4 13    4 5 HORIZONTAL DEFLECTION AND TRIGGERING              4 22       X CONTENTS  4 6 DIGITAL ACQUISITION AND STORAGE                     4 30  4 7 ADVANCED VERTICAL FUNCTIONS                       4 31  4 8 ADVANCED HORIZONTAL AND TRIGGER FUNCTIONS       4 34  4 9                                                                     4 39  4 10 CURSORS FUNCTIONS                                  4 44  4 11 MEASUREMENT FUNCTIONS                             4 49  4 12 PROCESSING FUNCTIONS                               4 54  4 13 DISPLAY FUNCTIONS                                   4 57  4 14 DELAYED                                                     4 63  4 15 HARD COPY FACILITIES                                 4 68  4 16 AUTOSET AND SETUP UTILITIES            
102. TANDARD SETTING    Simultaneously press the STATUS and TEXT OFF keys     Connect the Probe Adjust signal to channel 1     The Probe Adjust signal  now supplied to the input  is a square wave with a lower  level of OV and a top level of 600 mV  The oscilloscope always operates in analog  mode after a recall of the standard setting       Press the AUTOSET key for optimum signal display     ANALOG TO DIGITAL MODE SWITCHING    Switching from analog to digital mode and vice versa can be done by pressing the  yellow ANALOG key  If no Digital mode only features were activated  you can  switch between the modes with no changes in the display       Press the ANALOG key again     The oscilloscope is automatically set in the digital mode  This is indicated by a  message DIGITAL           that appears briefly on the screen     4 12 HOW TO USE THE INSTRUMENT       TRACE STORAGE RUN STOP    The digital mode offers a set of powerful features  One of the most important  advantages is that you can store one or more traces in memory or on screen       Press the RUN STOP key to stop the signal acquisition and freeze the  display on the screen     The trace is frozen and stays on the screen  Observe that the signal stays stored  even when you remove the probe    Now the instrument has been STOPped  and most keys have been disabled  The  only keys that continue to function are those directly related to display functions   This includes trace shift using the POS control  All measurement functions 
103. TB and DEL D TB     Key sequence     TB MODE       auto  trig       single Softkey pair to select between auto  trig  single and  multi multi  optional    ROLL  on off  ROLL Toggle softkey to switch ROLL on and off                    TRIGGER  yes ne Toggle softkey to stop acquisition in ROLL mode on  a trigger   REALTIME a  ONLY Er   yes no  Toggle softkey to inhibit random sampling mode   ARM D  e        Key to reset MAIN           single shot mode  with             ARMPD indicator        Remote commands   CPL  AT  has same result as the operation of the SINGLE key   Refer to Chapter 6 for full details     FUNCTION REFERENCE 5 75       TIME MEASUREMENTS    Description     Time measurements can be made using the cursors or using automated   calculated measurement routines  Cursors operate in the analog mode as well as  in the digital mode    Calculated time measurements are available in the digital mode only    Using the cursors  three time interval readouts can be selected      AT gives the time interval  seconds  between the cursors      1AT gives an inverted readout of the time interval  and is read in Hz      AT ratio gives a percentage  96  that can be reset by using the AT 100  softkey   For more details  refer to CURSORS READOUT     In the digital mode  the MEASURE menu accesses automatic time  measurements  This menu is reached via the key sequence    MEASURE  gt  gt   MEAS 1 2   gt  gt  time     The TRACK control us used to select the type of  measurement and the A c
104. THE CPL PROTOCOL 6 7       AUTOSET AS    Purpose     To start the AUTOSET function  With this command the oscilloscope will select  the optimum settings  volts  time base  trigger mode  etc   for the connected  signal s      The AutoSet  AS  command performs the same function as pressing the front  panel AUTOSET button     Command               AS  CR                      Response                       acknowledge CR       Note  The  lt acknowledge gt  will be sent after the AUTOSET has been fully  completed     Example     100 PRINT  1  AS       Send command  110 GOSUB 1000   Sync on acknowledge    6 8 THE CPL PROTOCOL       ARM TRIGGER AT    Purpose     Will reset the timebase and rearm the triggering for a new timebase trigger   Issuing this command during a time base sweep will immediately stop the sweep   reset the timebase and rearm the triggering     The Arm Trigger  AT  command performs the same function as pressing the  frontpanel SINGLE ARM D button     Command            AT  CR                         Response           acknowledge CR                   Example     100 PRINT  1  AT       Send command  110 GOSUB 1000   Sync on acknowledge    THE CPL PROTOCOL 6 9       CALIBRATE CL    Purpose     To start the internal Auto Calibration procedure  This procedure optimizes the  input  trigger and time base circuitry of the oscilloscope  This calibration takes  approximately one minute and completion is signalled by the acknowledge     The Calibrate  CL  command performs 
105. TICS             5 71  T  Text 5 DISPLAY                               5 31  b  TEXT OFF             ER ESI 5 74  5 USERTEXT   uit REX See    5 88  Timebase 5                                               5 75  5 DELAYED TIMEBASE                  5 27  Time cursors 5 CURSORS ein              5 19  Time measurements 5 CURSORS READOUT                 5 22  5 MEASURE MENU                     5 47  5 TIME MEASUREMENTS                 5 77  Time to trigger 5 CURSORS READOUT                 5 22  Touch  Hold  amp  Measure 4 11 TOUCH  HOLD  amp  MEASURE     MODE    4 52  5   TOUCH  HOLD  amp  MEASURE     MODE    5 78  Trace 5   CHANNEL TRACE SELECTION          5 17  Trace separation 5 DELAYED TIMEBASE                  5 28  Trigger 5 TRIGGER COUPLING                  5 79  5 TRIGGER DELDTB                   5 81  5 TRIGGERLEVEL                     5 82  5 TRIGGER                                5 84  Trigger position 5 DELAY ekle 5 24  Trigger View 5 EXTERNAL TRIGGER                 5 34  T trg 5 CURSORS READOUT                 5 22  TV 5                                             5 86       1 8 FUNCTION INDEX   U   User text 5 USER TEXT        wie              5 87   V   Vertical deflection 4 4 VERTICAL DEFLECTION               4 13   Vertical functions 47 ADVANCED VERTICAL FUNCTIONS      4 29   Vertical magnify 5 DISPLAY                                 5 30   Volt cursors 5 CURSORS e             diu cue ue ae 5 19   Volt measurements 5 CURSORS READOUT                 5 22  5 MEASU
106. TRACE STORAGE CLEAR    You can also clear a memory location       Select location    m2        Press the    clear    softkey     This clears memory location 2  This is indicated by the open circle     HOW TO USE THE INSTRUMENT 4 41       TRACE STORAGE COPY    A trace can be copied from one memory location to another     Press the  COPY  softkey     The  COPY MEMORY is now displayed  The source   FROM   and destination            memory locations can be selected in this submenu       Turn the TRACK control to select  m3  as source from     Turn the A control to select  m2  as destination to     Press the  copy  softkey     The contents of  m3  have now been copied into  m2  without destroying the data  in   m3        4 42 HOW TO USE THE INSTRUMENT       TRACE STORAGE RECALL        If you completed all of the previous steps  three memory locations           occupied by trace information  You can display these stored traces at  any time and in any combination       Press the RECALL key     Turn the TRACK control to select    m1        Press the DISPLAY on off    softkey to display m1     Observe that the stored trace on memory location  m1  is displayed and that the  circle before m1 is solid    An open circle indicates an empty memory location    A circle with a dot indicates that there is a trace stored  but it is not displayed    A solid circle indicates trace data in memory are displayed on the screen   Empty memory locations cannot be displayed     You can shift traces sepa
107. The DELAY control adjusts the delay between the start of the MAIN TB and DEL D  TB sweep  Refer to the TRIGGER DEL D TB function for more details     The DEL D TB time scale is adjusted in steps with the delayed timebase  TIME DIV key pair     5 28 FUNCTION REFERENCE       In the DELAYED TIMEBASE menu  which is selected with the DTB key  the  delayed timebase can be switched on  This is done with the first softkey called     DELD TB on off     Once activated  the delayed timebase trace is displayed  The  main timebase trace can be switched off using the    MAIN TB on off    softkey  If both  main and delayed time bases are displayed  you can separate them by using the  TRACE SEP function     DELAY DTB START DIRECTLY AFTER DELAY TIME  STARTS        DELAY TIME       Function of  mtb  dtb   DELAY  and TRACE SEP     TRACK             576750    Key sequence     DEL    DTB       ff        Toggle softkey to switch the delayed timebase on or off   MAIN TB  on off               Toggle softkey to switch the main timebase      or off   TRACK  Control to adjust the vertical distance  TRACE SEP   between MAIN TB and DEL    TB traces   S TIME DIV ns  K  CD  Key pair to adjust the DEL D TB time scale   577356  9303  DELAY    Control to adjust delay time between start of MAIN TB and DEL D  TB sweeps     FUNCTION REFERENCE 5   29       DIGITAL MODE    Description     The yellow ANALOG key is used to switch from the analog mode to the digital  mode and back at any time  The signal acquisition 
108. Time Base  MTB     The External Trigger input chargacteristics are simular to those of the input  channels 1 and 2  The input has probe range indication detection and two  attenuator positions  0 1 V div  and 1 V div    If External is selected as trigger  source the same trigger modes  Edge  TV and Glitch  are availlable as for the  trigger sources CH1 and CH2     When the External Trigger input is selected as trigger source   the signal on this  input can be made visible by pressing the function key TRIG VIEW  When another  Trigger source is selected  Trigger View is automatically switched off     Trigger View can be used in Single  and Multiple shot in timebases up to 10ps div   In the other timebase modes Trigger View automatically selects Alternating mode  for timebases faster than 10us div  This result in the simultaneous acquisition of  the External Trigger input     Key sequence     EXT TRIG  Key to select the external trigger input as trigger source   Selects trigger slope when External trigger is selected   TRIG VIEW    Key to turn on and off the display of the external trigger  signal     Toggle key to switch between the attunuator settings  1 V div and 0 1 V div                FUNCTION REFERENCE 5 35       FILTER    Description     The FILTER function is a waveform MATH function  It is a post  acquisition  algorithm which can be used to simulate the effect of a low pass filter process on  a trace  The cut off frequency of the low pass filter can be adjusted and the
109. US control is used to optimize display sharpness     A graticule is a provided with 1cm x 1cm divisions  Each vertical centimeter  equals the indicated channel sensitivity  A horizontal centimeter equals the  indicated timebase setting  The graticule has 0  10  90  and 100   lines that can  be used for risetime measurements  For these  the signal peaks are exactly  positioned on the 0 and 100   lines  The risetime of the pulse is readout between  the 10 and 90   lines as shown in the following figure  Intensity of graticule  illumination is controlled with the GRATICULE ILLUMINATION control     Trace alignment is done with the screwdriver operated TRACE ROTATION control                                                                                                                             eRILLIANCE    1       xum mE     BRILLIANCE 71                                         Screen controls and function of MS n A                    A    T17    TRACE ROTATION        gt  lt   DISPLAY              SHARPNESS    A Y ATESA rm SER                                 A  MESES  i               we       EXT  GRATICULE i IE Ipzmi  ILLUMINATION               1                         Rise time measurement using the graticule                                                              4217    FUNCTION REFERENCE 5   63       Key sequence     TRACE INTENSITY  Control for trace intensity     O    TEXT INTENSITY  Control for text intensity     O    TRACE ROTATION Srewdriver operated control to
110. XT OFF keys simultaneously     Connect the Probe Adjust signal to channel 1     The Probe Adjust signal now supplied to the input is a square wave with a lower  level of OV and a top level of 600 mV       Press the AUTOSET key     HOW TO USE THE INSTRUMENT 4 45       CURSORS ON OFF    There are two sets of cursors   amplitude cursors and time cursors  Amplitude  cursors are two horizontal lines  and Time cursors are two vertical lines  The  dashed lines are referred to as reference cursors and the dotted lines are referred  to as delta  A  cursors    Cursor measurements can be made in both operating modes of the instrument   Amplitude and time cursors can be displayed at the same time       Press the ANALOG key to set the scope in mode     Press the CURSORS key to enter the cursors menu     The menu is now displayed  Time cursors       are selected by default  In the  digital mode  two    x   s mark the actual measuring point where the cursors intersect  with the signal  In the analog mode these  x indicators are not available     CURSORS TIME    Using the TRACK and A controls  you can adjust the position of the cursors  The  time and amplitude difference between the cursors is read in the top of the screen   Cursor readout parameters can be   changed and are selected via the  CURSORS    READ OUT  softkey        on off      Turnthe TRACK control and            observe that both cursors cht       move  chi  ATE 35ltus    584mV     Turn the Delta  A  control  and observe that o
111. ace with amperes over time     The resulting trace is expressed in watts and gives the power over time     After integration this gives the Pulse Power in joules     THE CPL PROTOCOL 6 41       RESET INSTRUMENT         Purpose     Resets all of the software of the oscilloscope  including the CPL protocol handler  and all of the input and output buffers    Oscilloscope settings remain the same  Interface parameters are not changed in  order to keep the communication alive     When the Reset has been completed the oscilloscope responds with an   lt acknowledge gt      There is no frontpanel equivalent for this command     Command                  RI  CR                   Response                       acknowledge CR       Note      acknowledge   will be sent after the Reset has been completed  Example     100 PRINT  1  RI       Send command  110 GOSUB 1000   Sync on acknowledge    6 42 THE CPL PROTOCOL       RECALL SETUP RS    Purpose     To recall an internally stored setup from one of 10 setup registers  This setup must  have been stored in the oscilloscope manually or with the Save Setup  SS   command     The command performs the same as the frontpanel key SETUPS together with  the softkey RECALL                                            Command    RS  pds setup reg CR   lt setup reg gt  represents the setup register number ranging from 1   10  Response           acknowledge CR                   Note      acknowledge   will be sent after the internal setup has been complete
112. and  measurement equipment     VI       INITIAL INSPECTION    Check the contents of the shipment for completeness and note whether any  damage has occurred during transport  When the contents are incomplete or  there is damage  file a claim with the carrier immediately  Then notify the FLUKE  Sales or Service organization to arrange for the repair or replacement of the  instrument or other parts  FLUKE addresses are listed in the back of the  REFERENCE MANUAL     The following parts should be included in the shipment     Service ordering number  or model number    1 Oscilloscope PM3370B  PM3380B or  PM3390B  PM3384B or  PM3394B   1 Front cover 5322 447 70121    1 Users Manual  1 Reference Manual  1 Line cord  European type  or 5322 321 21616    1 Line cord  North American type  or 5322 321 10446    1 Line cord  British type  or 5322 321 21617  1 Line cord  Swiss type  or 5322 321 21618  1 Line cord  Australian type  5322 321 21781    2 Probes 10 1  2 Batteries AA  LR6     1 Spare fuse 3 15 AT 4822 070 33152   located inside fuse holder     The performance of the instrument can be tested by using the PERFORMANCE  TESTS in the REFERENCE MANUAL     VII       INSIDE THIS MANUAL    This operating guide contains information on all of the oscilloscope s features  It  starts with a general introduction  a summary of main capabilities  initial  inspection note and a front and rear view     Operators safety    Installation instructions    Getting started    How to use more advanced  fu
113. and display functions of both  operating modes are very similar  However  the nature of the signal used or the  choice of the measurement may determine when it is best to use the digital mode     Digital storage is used for the following      recording of single events  or very low frequencies   e pretrigger and posttrigger view     glitch triggering     Saving  recalling and comparing multiple traces     automatic measurements     Signal mathematics     Signal analysis     making hard copies  plot print      digital mode offers the convenience of the auto range function for the vertical  channels and the main time base     The same waveform is usually displayed in the same manner  regardless of the  operating mode selected  However  some functions are not implemented in both  modes  In the analog mode  for instance  no  pretrigger  information can be made  visible    If switching from one mode to another results in an unsatisfactory display  press  the yellow key once more to return to the previous situation     Key sequence     ANALOG    Cy Toggle switch to switch between analog mode and digital  mode     5 30 FUNCTION REFERENCE       DISPLAY MENU    Description     The DISPLAY menu offers a set of powerful display functions for the analog mode  and the digital mode    In the analog mode  the scope can be set to the X DEFLECTION mode  In this  mode XY displays can be generated from a combination of any of the input  channels for X and Y  while the analog timebase generator 
114. aph moves as you use the XPOS control       Press the RUN STOP key     Observe that the MAGNIFY key and the X POS control can still be used after the  acquisition has STOPped  This allows you to look at signal details even after the  signal has been acquired       Press the RUN STOP key     TRIGGERING      Simultaneously press the STATUS and TEXT OFF keys     Press AUTOSET     4 26 HOW TO USE THE INSTRUMENT       TRIGGER SOURCE    Press the TRIG 2 key in the CH2 section of the front panel to select  channel 2 as trigger source     The indication in the lower right hand readout area of the screen now displays    ch2      Observe that the signal is not triggered  The ARM D LED is on       Press the TRIG 1 key in the CH1 section of the front panel to select  channel 1 as trigger source     Observe that the indication has now changed into    ch1    and the signal is triggered  again                                                                                                                          TRIG2 TRIG  C     2 PER       CH1 hn   ICH1 in1T   TRIGGER SLOPE    TRGI        trigger source selection key is also used to select the trigger slope  between the positive and negative going edge of the triggering signal       Pressthe TRIG 1 key in the CH1 section of the front panel a second time     Observe that the signal now starts with the trailing edge instead of a leading edge   Also the trigger slope symbol in the lower right hand corner of the readout area  indicates a trai
115. application or preference  For information about  AUTOSET programming  refer to AUTOSET USERPROG     Key sequence     AUTOSET    a Key to start the autoset    Remote commands   CPL  AS  Command for Auto Setup   Refer to Chapter 6 for full details     FUNCTION REFERENCE 5 11       AUTOSET SEQUENCE    Description     If front panel settings are stored in memory locations    s1 to s5    and  s6  is cleared   then the range of  s1 to s5    becomes a sequence of front panel settings  Such a  sequence can be used as  part of  a step by step testing procedure  The two  following methods can be used to quickly step through such a sequence  This  saves time by not without having to recall each setting from a menu       Programming the AUTOSET key to be used as the  recall next setup  key  This  selection is done in the UTILITY  gt  gt  AUTOSET menu       Using aspecial probe with a COMMAND switch  Two probes with COMMAND  Switch are delivered with each oscilloscope as standard accessory  The  appropriate mode for this probe command can be selected in the UTILITY  gt  gt   PROBE menu     To return to the    normal    oscilloscope mode  press the STATUS and TEXT OFF  key simultaneously to recall the STANDARD SETUP     5 12    FUNCTION REFERENCE       AUTOSET USERPROG    Description     A number of instrument settings after AUTOSET can be customized to your  special needs when the standard default settings do not fit on your application     The selections are reached via the key sequence
116. ar character Ey   37    39 ASCII characters    amp      40    47 ASCII characters               48    57 ASCII characters 0123456789   58    64 ASCII characters    lt   gt      65    77 ASCII characters ABCDEFGHIJKLM  78    90 ASCII characters NOPQRSTUVWXYZ  91  95 ASCII characters     _   96 Menu selection indication   97    109 ASCII characters abcdefghijklm   110    122 ASCII characters nopqrstuvwxyz   123 Menu return indication   124   character   125 and 126   Menu rocker key up and down characters   127 Control up down character   amp     128    255 Control up down character  conform char  127                 Table 6 1 Character code table for oscilloscopes     THE CPL PROTOCOL 6 21       Examples   of user text     1  Program the following user text to be displayed on the screen of the  oscilloscope  Measurement 15       100 PRINT  1  PT    Program user Text command  110 GOSUB 1000   Sync on acknowledge   120 PRINT  1  Measurement 15   Send user text   130 GOSUB 1000   Sync on acknowledge    2  Inthe next example user text  containing non keyboard characters  0 25  decimal and s 2125 decimal   is programmed to be displayed on the screen  of the oscilloscope  e g   Ohm Q  Up s         100 PRINT  1  PT    Program user Text command  110 GOSUB 1000   Sync on acknowledge  120 PRINT  1  Ohm   CHRS 25     Up   CHRS 125            Send user text  130 GOSUB 1000   Sync on acknowledge    Example   of setup text     Program the following text to set up register 7 of the oscillosc
117. are still  operable    A frozen trace can be used for comparisons with other traces  mathematics   cursor measurements  automatic measurements and more    The RUN STOP key is used to end the STOP mode and start the aquisition again       Press the RUN STOP key     This starts the acquisition again so that the actual input signal is displayed  You  can use the RUN STOP key at any time                       RUN STOP                                                                                                                RUNNING FROZEN Sese    NOTE  The following section explains the basic functions regardless of  operating mode  The oscilloscope will react almost identically in either  mode  Where necessary  different behavior will be explained     HOW TO USE THE INSTRUMENT 4 13       4 4 VERTICAL DEFLECTION             MENU AVERAGE        section shown on the left contains all    direct vertical deflection controlls for the input  channels 1 and 2        Refer to Section 4 1 for the differences  between model numbers     To start this section with the settings ina  predefined state  you must recall the  standard setting                 STANDARD SETTING      Simultaneously press the STATUS and TEXT OFF keys     Connect the Probe Adjust signal to channel 1     The Probe Adjust signal  now supplied to the input  is a square wave with a lower  level of OV and a top level of 600 mV       Press the AUTOSET key     4 14 HOW TO USE THE INSTRUMENT       VERTICAL COUPLING AC  DC  G
118. ate in the analog mode as well as  in the digital mode    Calculated volt measurements are available in the digital mode only    Using the cursors  three voltage readouts can be selected      AV gives the voltage difference between the cursors      V1 amp V2 gives the absolute voltage with respect to ground for each cursor      AV ratio gives a percentage that can be reset by using the AV 100  softkey   For more details refer to the section CURSOR READOUT     In the digital mode  the MEASURE menu provides automatic voltage  measurements    This menu is reached via the key sequence MEASURE      MEAS 1 2       volt    The TRACK control is used to select the type of measurement  and the A control  is used to select the waveform    The result of the measurement is displayed in the top left corner of the screen                                     nie PRESHOOT    OVERSHOOT     12                    Hian The following   RMS measurements are possible                Refer to section MEASURE  MENU for details of signal  LOW parameters                                                                               4           4      UNDERSHOOT       T6670            DC VOLTAGE   Measures the absolute dc level  related to the ground level of the signal  This  is the average voltage of all samples in one period  If no full period is present   all input samples in the signal are included in the calculation     dc         1                        where  j  1st mesial crossing  first sample if
119. ation is the default  The following   description applies    dots  The space between the sample points is blanked  Only the real  samples are shown    linear  The space between the sample points is interpolated linearly  Straight  lines are drawn between sample points    sine  The space between the sample points is interpolated using a sine  wave interpolation algorithm  This mode is used to obtain a more  accurate display of mainly sinusoidal signals which were  undersampled     Key sequence     DISPLAY     X DEFL Softkey to select the X DEFLection mode  refer    tothe X  DEFLECTION function  in analog mode  only    WINDOWS  DISPLAY on off Separates the displayed signals and sets them in    a window  in digital mode only      VERT  DISPLAY MAGNIFY Softkey to turn on the vertical magnification       in digital mode only    TRACK  Control to determine the magnification factor        576784  9303    FUNCTION REFERENCE       DISPLAY X vs Y    DISPLAY TEXT    dots  linear  sine    DISPLAY       576755  9303         off    0    TRACK          X SOURCE       i    576785  9303    TRIG IND  on off    GND IND  on off         USERTEXT               576786    Softkey to select the X vs Y mode  see function  X DEFLECTION  in digital mode only      Control to select the Y source     Softkey to select the X source     Toggle softkey to display trigger level indication     Toggle softkey to display or blank ground level  indicators     Access to editing menu for USERTEXT  refer to  the USER
120. bles the user to synchronize the display rate to the signal being  examined  This eliminates the problem of    double    triggering on pulse train  information as shown in the figure    For most signals the hold off must be minimal  0    so that the sweep repetition  rate can be the highest permitted by the trigger signal  Turning the HOLD OFF  control clockwise increases the hold off time  When the hold off time is set longer  than is necessary to synchronize the signal  loss of light output will result    In the digital mode  the variable hold off time serves a similar function  and the  extra time is used for processing previously captured data    In digital mode  triggering on certain pulses can be achieved via the logic trigger  mode    pattern     For details refer to the LOGIC TRIGGER function     5 38 FUNCTION REFERENCE                   2 1 2 1 2  SIGNAL  TRIG  TRIG TRIG  SWEEP        HOLD OFF HOLD OFF HOLD OFF   no triggers  no triggers  no triggers  accepted  accepted  accepted   WAVEFORM  1  ON SCREEN  1 2 1 2 1 2  SIGNAL             TRIG     TRIG     TRIG            WAVEFORM     2  ON SCREEN   DOUBLE  TRIGGERING                                Using HOLD OFF to suppress double triggering     Key sequence     HOLD OFF  O Control to adjust MAIN TB hold off time     FUNCTION REFERENCE 5 39       INPUT ATTENUATOR MANUAL   AUTOMATIC  Description     The oscilloscope   s input has a wide range of sensitivities  This enables signals of  different amplitudes to be display
121. cal direction  Y  is selected by the TRACK control  It can be  the acquired trace or a saved track in a memory location  e g   m3     For horizontal direction  X  you can choose from an active channel signal saved  in memory  e g   m3 1        Press the ANALOG key to select the digital mode     Press the DISPLAY key      Press the X vs Y softkey      Press the    on off    softkey to turn it on      Press the X SOURCE softkey to select    ch1                                 Now two horizontal lines  Ch 1 against Ch 2  are  displayed  Horizontal deflection is determined by the Sab aban   Se che od  amplitude of the sine wave                                                 The X vs Y submenu displays the selection menu for the X and Y sources   Observe that you can select the Y source  REGISTER  using the TRACK control   Use the softkeys to choose the X source       Remove the signal from channel 2     4 60 HOW TO USE THE INSTRUMENT       DIGITAL DISPLAY VERT MAGNIFY    In the digital mode  the displayed signal can be expanded vertically after it has  been captured  This allows you to look to signal details  Please note that the  magnified representation of the signal has the same resolution as the originally  acquired signal       Press the ANALOG key to select the digital mode      Press AUTOSET      Remove the probe tip from the Probe Adjust output      Turn the TRIGGER POSITION control to mid screen  counterclockwise      Press the SINGLE key to prepare the scope for a sing
122. can be chosen in the menu under the TB MODE menu key     In the digital mode the TB MODE menu is expanded to include MULTI  only with  the optional           functions installed   ROLL and REAL TIME ONLY mode     FUNCTION REFERENCE       ROLL    REAL TIME    ONLY    Part of MATH  option  Refer also to MATH  Users Manual  The  MULTIple shot mode enables a number of single shot acquisitions  to be made in rapid succession  It is basically the same as SINGLE  but with automatic rearming following each acquisition  Each  acquired waveform is stored in a different memory location     In the ROLL mode  the MAIN TB can be used with sweep speeds  from 200 ms div down to 200 s div  Initially the trace fills the  screen from the right to the left  Once the screen is full  the trace  flows  through  the screen from left to right  When ROLL is  selected  the TB MODE menu provides a  STOP ON TRIGGER   choice  If this is set to    off     the trace flows through the screen  continuously until you press the STOP key  If    STOP ON  TRIGGER    is set to    on     the trace stops when the scope is  triggered and the trigger point reaches the left of the screen     In the digital mode the scope operates in random sampling mode  for sweep speeds 200 ns and faster  Setting    REAL TIME ONLY     to    on    means that these MAIN TB positions are inhibited  This  measuring errors on signals that are not truly repetitive or  identical  REAL TIME ONLY also inhibits alternating acquisitions  of MAIN 
123. ch memory  location can hold a maximum of one of the following   4  3  waveforms of 512 samples     3  waveforms of 2K samples     4  3   2  2  waveforms of 4K samples   1  1  TRACE                                                                             1  waveform of 8K samples  MEMORY  The number of stored waveforms depends on PES ee       the selected acquisition length and the number acq i ch2  of channels that were active at the moment of mi    ext  the acquisition       E  To switch single traces on and off instead of a ms      total register  the bottom key in the RECALL      i     menu can be used to select TRACE instead of           REGISTER         RECALL menu is then trace me 2244    oriented instead of register oriented  m7           mea  ms m8 2  Example       changes in m1 1 and   1 2  Uc   m8 ext  representing the labels for the ch1 sT7700    trace in m1 and the ch2 trace in m1   respectively  Label m1 e is for the  External Trigger input signal in m1     5 18 FUNCTION REFERENCE       Key sequence     ON         Key to switch a channel ON or OFF   RecaLL TRACK  Control to select a register or a trace to be recalled  from memory   DISPLAY  on off     Toggle softkey to turn the display of the selected  trace or register ON or OFF   trace  register  Toggle softkey to deselect traces or deselect registers        576557  9303    CONFIDENCE CHECK    Description     After turning the oscilloscope on  a confidence check starts automatically  The  following is teste
124. ct between the start of AUTOSET  performing a TOUCH  HOLD  amp  MEASURE      selecting the next setup or switching between analog and digital mode     For non Fluke probes or probes without an indication ring  the attenuation factor  can be programmed  As a result  the combined input sensitivity of the probe and  oscilloscope is given in the readout area  The selections in the UTIL PROBE  CORR menu are reached via the key sequence  UTILITY  gt  gt  PROBE  gt  gt  PROBE  CORR     Attenuation factors 1 1  10 1  20 1  50 1 and 100 1 can be selected  This  can be done for each channel individually     Key sequence     autoset  setups       q meas              UTILITY   PROBE CA  Softkey pair to select oscilloscope s reaction  on pressing pushbutton on probe  7                   ext  CoRR A Softkey pair to select for which channel the   7   probe attenuation is valid  ch1  ch2  ch3  ch4       or Ext Trig        Softkey pair to select probe attenuation factor    Pag  TZ   4 4  10 1  20 1                   5 58 FUNCTION REFERENCE       REMOTE CONTROL IEEE 488 2    Description     An IEEE 488 2 Interface is available as an option  This interface can be used to  control oscilloscope functions by an external computer  All of the oscilloscope   s  current settings can be read by the computer     The programming language is called SCPI  Standard Commands for  Programmable Instruments   SCPI is an IEEE standardized language designed  for remote control of programmable test and measuring e
125. cted by using the A control in the EVENT DELAY sub  menu     Event delay and trigger position can be used in combination to offer powerful  delay capabilities     ed      o             i ACQUISITION   CH1  SLOPE X       CH2  EVENT             8     TRIGGER POSITION       ms    JL LIU             576768    Key sequence     TRIGGER  POSITION    Control to adjust of the trigger position     EVENT    TB MODE DELAY on      Toggle softkey to switch the Event mode on and off     When the Event mode is turned on  further menu  selections are displayed       Control to change the number of events  Readout in  the menu is called    COUNT        Toggle softkey to select the channel on which the  events must be counted     Control to set the events level  Readout in the menu is  called    LEVEL        Toggle softkey to select the positive or negative slope     5 26 FUNCTION REFERENCE       DELAY MEASUREMENT    Description     In the digital mode  the MEASURE menu provides a    delay    measurement  This  is an automatic measurement of the time between two 50  levels   mesials   of  the first leading or trailing edge of two signals  The menu is reached with the key  sequence    MEASURE  gt  gt  MEAS 1  2   gt  gt  delay    The lower part of the menu is then  automatically changed to the delay measurement     Newly acquired signals or signals stored in memory can be used as sources for  this measurement  The A and TRACK controls are used to select the two sources   Sources have to be trace
126. ction on off   AMPL  E  Toggle key to switch between two vertical input sensitivities   Cy of the External Trigger input channel     INPUT COUPLING    Description     The characteristics of the oscilloscope s inputs can be selected for each channel   A channel can be switched on off with the toggle key ON  Display of the External  Trigger input signal is switched on off with the toggle key TRIG VIEW  CH1 is  always switched on when all other channels are switched off     In the dc coupled mode the complete signal including dc components and  extremely low frequencies    10 Hz  are displayed  In the ac coupled mode  dc  components are suppressed  This results in a display of the ac components   210 Hz      Ground coupling  GND  interrupts the input signal  The position of the trace is at  the 0 volt level  A continuous ground level indication for each channel can also be  activated  Refer to the description of the UTILITY gt  gt SCREEN amp SOUND MENU  function     The selections are made with the toggle keys AC DC GND for the channel inputs   or AC DC for the External Trigger input     The type of coupling is given in the readout area using the symbols  v  ac      dc   and 1  GND               10Hz FULL BANDWITH DC FULL BANDWITH      FREQ       FREO    Effect of ac dc input coupling    FUNCTION REFERENCE 5 41       Key sequence     ON  Cy Toggle key to switch a channel on off     TRIG VIEW  Toggle key to switch the display of the External Trigger  signal on off    AC DC GND  T
127. cy    Long serial data streams    Video signals    OTHER CRITERIA    Need to see pretrigger    information    ANALOG MODE    Usable   Causes display  flickering  Displayed for the  duration of    the event    Preferred    Preferred    Preferred when  Delayed sweep  is not used     Preferred when  Delayed sweep  is not used     Not possible    DIGITAL MODE    Usable    Preferred    Can capture and  display for long  term display    May cause alaising  Use Peak detect or  Envelope mode    May cause aliasing   Use Envelope mode     When using delayed  sweep to observe  details  Digital mode  provides better   light output     When using delayed  sweep to observe  details  Digital mode  provides better   light output     Up to full acquisition  length       GETTING STARTED 3 7  SIGNAL CRITERIA ANALOG MODE DIGITAL MODE  You need to make adjustments Fastest Slower  to the circuitry and watch display display  the signal change update update  Automatic measurements Can t use Fully implemented  Signal Math Add  Subtract All functions  Add  Subtract  Multiply  Signal Analysis Not available Full analysis  Integration   optional   Differentiation  FFT  Automatic Pass Fail test Not available Fully implemented   optional   Autorange attenuator Not available Results in a displayed  signal with an ampli   tude of 2 to 6 4 divi   sions  Autorange timebase Not available Results in a signal    display of 2 to 6  waveform periods    ANALOG    C                                                    
128. d       MODE        Envelope and Roll modes cannot be  combined   Function ENVELOPE    and  TB MODE        Event delay and dual slope cannot be  combined   Function DELAY        Event delay and TV trigger mode cannot  be combined   Function DELAY    and       TRIGGER        Trigger level indication not possible at dc  input coupling and ac or If rej trigger  coupling    Function  UTILITY SCREEN  amp  SOUND     X source only possible out of chosen  register    Functions  DISPLAY MENU  and     X DEFLECTION      Activated function only possible in digital  mode   Function ANALOG DIGITAL MODE        Indicates that the print plot action can  only be aborted with PLOT key   Function PRINTING AND PLOTTING        CAL key must be pressed for more then  two sec to start autocalibration   Function CALIBRATION AUTOCAL     The automatic probe detection overrules  manual selection when a probe with  indication ring is used    Function PROBE UTILITIES        5 66    FUNCTION REFERENCE       REGISTER EMPTY    REGISTER PROTECTED    REGISTER USED FOR MATH    SETUP EMPTY    SETUP PROTECTED    STD SETUP   RECALL ONLY    TIME DIV ADJUSTED    FOR ROLL  TOO MANY TRACES    SETUPS    Description     Recalling traces from an empty register is  not possible  Protection of an empty  register is impossible    Function  RECALL     Register cannot be saved in protected  memory location   Function SAVE        Trace cannot be saved in register M1 or  M2  because it is in use for a MATH  function    Function 
129. d   Example     100 PRINT  1  RS  5   Recall setting 5  110 GOSUB 1000   Sync on acknowledge    THE CPL PROTOCOL 6   43       SAVE SETUP ss    Purpose     To save the current oscilloscope setup in one of 10 internal setup registers  This  setup can be recalled manually or by sending the Recall Setup  RS  command     The command performs the same as the frontpanel key SETUPS together with  the softkey SAVE  An execution error will be returned if the setup register is write  protected                                               Command    SS  pds setup reg CR   lt setup reg gt  represents the setup register number  ranging from 1     10  Response           acknowledge CR                   Example     100 PRINT  1  SS  3   Save setup in reg 3  110 INPUT 1 ACK   Sync on acknowledge    6 44 THE CPL PROTOCOL       STATUS ST    Purpose     To obtain a more detailed status report  The response gives more information  about the conditions  causing a wrong acknowledge     The status is returned as one or more signed decimal integers    Each bit of the equivalent 16 bit binary status word represents a status condition   If a bit is set  the corresponding status event has occurred    The 16 th  sign  bit of a status word indicates that another status word follows   The last status word is the error status     After the reply  the value of the status is reset to zero  The complete description  of the status word can be found in section 6 7  STATUS      The function of the front panel ke
130. d DEL D TB  the triggering becomes less sensitive to noise     The MAIN TB trigger coupling can be selected in the menu under the TRIGGER  menu key  For DEL D TB  the menu under the DTB menu key is used  the  selection is possible in  triggered  mode  ch1  ch2 and ext trig  only     ac coupled It reject hf reject             30kHz    10Hz    FULL FULL FULL  BANDWITH BANDWITH BANDWITH  FREQ      FREQ          FREQ                        Effect of trigger coupling modes    Presence of trigger level indicators  T   D   as a function of trigger coupling and  other oscilloscope settings                    Trigger Vertical Trigger level indicator  coupling input coupling   ac      off   dc dc on   If rej dc off   hf rej dc on   all settings ac on   if level pp is on off                5 80    FUNCTION REFERENCE       Key Sequence     ac       If rej    TRIGGER       noise  on off       516772  9303    ac dc             DTB hf rej     1673  9303    Toggle softkey to select MAIN TB trigger coupling modes       dc   If reject  and hf reject     Toggle softkey to switch  noise  mode for MAIN TB DELD TB     Toggle softkey to select DEL D TB trigger coupling modes ac  dc   If reject and hf reject  Not selectable when    starts    is active in the  menu under the DTB key     FUNCTION REFERENCE 5 81       TRIGGER DEL   D         Description     The Delayed Timebase has two operating modes  starts and triggered  In both  modes the main timebase must be triggered first  and the delay time mu
131. d back from the oscilloscope  The following list  gives the possibilities for data separators     program data separator  space SP  ASCII32       tab HT  ASCII9       comma    ASCII 44        more than one space or more than one tab can be used as a  separator    response data separator  comma    ASCII 44     Message Terminators   At the end of a command or response a terminator must be sent  For both  programming and response messages the terminator has been defined as     program message terminator  carriage return CR  ASCII 13     response message terminator  carriage return CR  ASCII 13     THE CPL PROTOCOL 6 3       6 2 EXAMPLE PROGRAM FRAME    In the COMMAND REFERENCE SECTION a very short programming example is  given for each command  All examples are written in GW Basic and able to run  on an IBM compatible PC  The example program expects the oscilloscope to be  connected via COM1 port  RS 232  with a RS 232 null modem cable and to be   setup at 9600 baud  8 databits  no parity  3 wire  xon xoff   off  Menu UTILITY      The following program lines are an example frame work  The frame work lets you  embed any of the example programs shown in the COMMAND REFERENCE  SECTION     Program frame   10 OPEN  COM1 9600 N 8 1  CS0  DSO  CDO  AS  1    open serial port              Insert Example Programs Here                      999 END   1000      synchronize on acknowledge  amp  handle error   1010 INPUT 1 ACK   1020 IF  ACK 0  THEN GOTO 1110   1030 IF  ACK lt 0  OR  ACK gt 4 
132. d bitwise and only when both bits are 1  the  result bit is 1     Example   Status   38 and must be checked for  data out of range     4     38   0000 0000 0010 0110  4   0000 0000 0000 0100  logical AND 0000 0000 0000 0100    First status word  CPL status          Bit Mask Value Meaning   Position hex dec   0  amp HO0001 1 Unknown header   1  amp H0002 2 Data format of body is wrong   2  amp H0004 4 Data out of range   3  amp H0008 8 Invalid instruction   4  amp H0010 16   reserved  normally zero   5  amp H0020 32 Invalid number of parameters   6  amp H0040 64 Wrong number of databits   7  amp H0080 128 Flash ROM not present   8  amp H0100 256 Invalid Flash software   9  amp H0200 512 Conflicting oscilloscope settings  10  amp H0400 1024 User request  front panel key pressed  11  amp H0800 2048 Flash write error   12  amp H1000 4096   reserved  normally zero   13  amp H2000 8192 Invalid password entered   14  amp H4000 16384   reserved  normally zero   15  amp H8000 32768 Next status available             Second status word  Auto calibration status  Third status word  Acquisition status    THE CPL PROTOCOL 6 49       Example program to investigate status        2000 PRINT  STATUS IS    STATUS  2010 PRINT  MEANING      2020 IF  STATUS AND  amp H0001     0 T  2030 IF  STATUS AND  amp H0002     0 THEN PRINT  WRONG DATA FORMAT   2040 IF  STATUS AND  amp   0004   gt  0 THEN PRINT  DATA OUT OF RANGE   2050 IF  STATUS AND  amp H0020   gt  0 THEN PRINT  INVALID   PARAMS     gt     g
133. d in sequence      The instruments internal control bus      The communication between front panel and internal microprocessor      The settings in the memory  with backup batteries installed only       Interface to digitizer circuitry     The selftest takes less than a second  A message appears on the screen when  errors are found     The settings stored in memory become active with backup batteries installed   These are the same settings present when the instrument was last switched off   With no batteries present  the standard default setting  std  is activated  Refer to  STANDARD SETUPS for details     Key sequence     POWER ON OFF    Toggle key to switch the oscilloscope on off  Starts  confidence check    FUNCTION REFERENCE 5 19       CURSORS    Description     Cursors are on screen measuring lines  They can be moved using the TRACK  and A controls  Cursors can be positioned on signal details of interest and can be  used for accurate measurements     Basically there are two types of cursors  vertical lines      called time cursors and  horizontal lines     called volt cursors  A simultaneous display of both cursor  types     is possible     If the acquisition length in digital mode is more than 10 horizontal screen  divisions  it is possible to position the cursors horizontally along the complete  acquisition  When a cursor that is visible on the screen is moved outside the  screen  it will drag the trace with it  This means that the X POSition is changed  A  bar graph 
134. differential mode is provided by adding CH1  CH3  and    Inv    CH2          together     In analog mode the two channel and the added trace are displayed at the same  time  In digital mode however channel and added trace are displayed separately   Switching between both is done with the CH1 CH2  CH3   CH4  toggle key   Activating the function will turn off AUTO RANGE AMPL     CH1          Suppression of common mode  components in two signals using the  differential mode     ONLY THE SQUARE      WAVE WILL BE  DISPLAYED        Key sequence  inus  CH      CH     Toggle key to switch the addition of CH1 and CH2 or CH3  and CH4 on off    INV    Cy Toggle key to switch the inverted display of CH2  CH4   on off    FUNCTION REFERENCE 5 5       ADD  MATHEMATICS     Description     The ADD function performs a point to point addition of two traces  related to the  two ground levels  indicated as           The result of the ADD function is a new trace  in a different register  This trace can be scaled and positioned     Scaling is the correction of the resultant trace to fit in the screen  The TRACK  control adjusts scaling    Vertical positioning is called offset  It offsets each sample in the resultant trace  with a value so that the trace can be  moved  vertically  The A control adjusts the  vertical positioning  The scale factor and the offset factor are displayed in the     MATH SCALE    menu  Pressing the    autoscale    softkey automatically selects the  settings so the trace fit
135. e  marked with a circle in front of the memory location  number  e g       m3      Press the second blue softkey     save      You have  now saved the acquisition signal into memory     register    m3  A single register can contain a set of up  to three traces  e g   CH1  CH2  and EXT  trigger  view    In this case only one input channel was  turned on  so that only one was stored     Remove the probe from CH1  Now recall the stored  trace        Press the RECALL key  576705    Select the previously filled memory register m3 with the TRACK  control  A memory register with trace information is indicated with       Press the second blue softkey to turn on the display of this register   Indicated by      Turn the A control to separate the acquisition  live signal  and the trace  recalled from memory     You are now able to operate nearly all the oscilloscope   s functions in  most routine applications  Please continue with Chapter 4 for a more  detailed description of the oscilloscope   s many advanced features     Memory indications        Empty register     Filled register    Displayed register    HOW TO USE THE INSTRUMENT 4 1       4 HOW TO USE MORE ADVANCED  FUNCTIONS OF THE INSTRUMENT    This chapter allows more experienced oscilloscope users to learn more about the  advanced features of this instrument and how to use them  For a complete  description of each function  refer to the next chapter in this manual   Function  Reference     This chapter explains the basics of eac
136. e 3 5    Step 1    Step 2       5  7670    Vertical setup    Adjust the absolute ground level by disconnecting the signal and using  the POS control to position the trace in the middle of the screen  A  marker with the channel number     1      at the left of the screen indicates  the ground reference                                                                                                                          MAT4191    Reconnect the probe to the Probe Adjust signal for display     GETTING STARTED 3 9       Step 3    Step 4    Step 5    You can change the amplitude of the signal in a 1  2  5 sequence by  pressing one of the AMPL keys  Note that the bottom of the screen  shows the AMPL DIV setting of CH1                                   3 FEE  x   gt                                                                                                100mV  CH1   500MV N          ST6681    Press the ON button of CH2 and observe that a second trace is now  visible  The position and amplitude of this channel can be adjusted  similar to the adjustment of CH1  The channel settings are also  displayed in the bottom of the screen     Press the ON key of CH2 once again to turn this channel off     Press the AC DC GND key of CH1 so that a 1 sign is displayed in the  bottom text line  This interrupts the input signal and connects the input  to the ground  In this case  only the  base  line is visible     Press the AC DC GND key once again for ac input coupling  the  bottom text li
137. e has a bandwidth of 60 MHz   Beside the 2 channel models with EXT TRIG input  there is a choice of two models  with four fully featured channels  all shown in the following table     Type Number  Bandwidth Sample rate Number of Input  Channels Impedance  PM3370B 60 MHz 200 MS s 2 1 MQ  PM3380B 100 MHz 200 MS s 2 1 MQ  PM3384B 100 MHz 200 MS s 4 1 MQ  PM3390B 200 MHz 200 MS s 2 1      500   PM3394B 200 MHz 200 MS s 4 1      500                 In the same instrument family  there are two 200     2        two 100 MHz analog  oscilloscopes that have specifications similar to the above mentioned analog   digital combination oscilloscopes operating in analog mode     All analog digital combination oscilloscopes listed above have the following features      Autoranging attenuators      Realtime clock      32K sample acquisition memory in 4 channel versions                        acquisition memory  expandable to 32K      2 channel versions      Up to 40 waveforms stored in memory or 204 waveforms with optional  memory extension      Autoset function for an instant optimized signal display at the touch of a button      Autoranging timebase      Cursor measurements with 1  accuracies      Extensive set of fully automated voltmeter and time measurement functions      Probe operated    Touch Hold and Measure    function freezes the display and  instantly displays the signal frequency  amplitude and dc voltage level        Peak detection for the capture of glitches as narrow as 5 ns   
138. e parameter n         not be used     The programmed text can be read with the Query Text  QT  command     Command 1               PT  pds      pds n   CR                                                    Response 1                       acknowledge CR       Command 2     X char CR    Response 2                                            acknowledge CR                   6 20 THE CPL PROTOCOL        S  setup text will be specified for register n  if  S  and n are left out   user text is specified   n one of the setup registers  ranging from 0 to 10      0 selects the current setup   char a character byte  range   0 to 12  14 to 255   refer to character code table 6 1     The following table contains the decimal codes of the character set for the screen  of an oscilloscope           CHAR CHARACTER DESCRIPTION   0    12 Control up down character  conform char  127   13 Command terminator ASCII CR  may not be used   14 Control up down character  conform char  127   15 and 16 Positive and negative slope characters   17 and 18 Positive and negative pulse characters   19 Dual slope character   20 Special marker  X    21  22  23 Delta  degrees  micro characters   24 and 25 Low impedance  low  2  and omega  Ohm  characters  26 and 27 Arrow up and down  char  27 is also ESC character   28 and 29 AC and ground  channel coupling  characters  30 Register off  but filled  character   31 Filled o character   32 ASCII space character   33    35 ASCII characters        36 External Trigger ch
139. e screen of the oscilloscope  A text field of 64 characters is returned  The  parameter n may not be used     The queried text can be programmed again with the Program Text  PT  command                                                                                                                 Command    OT  pds      pds n T CR   6  setup text of register    will be returned  if  S  and n are left out   user text is returned  n one of the setup registers  ranging from 0 to 10      0 selects the current setup  Response 1   acknowledge CR  Response 2              CR          char acharacter byte  range   15 to 127   refer to PROGRAM TEXT character code table 6 1     6 34    THE CPL PROTOCOL       Example   of user text     Read the user text from the screen of an oscilloscope     90 DIM USERTXTS  64    100 PRINT  1  QT    110 GOSUB 1000   120 LINE INPUT   1 USERTXT   130 PRINT USERTXTS                           Query user Text command    Sync on acknowledge    Read text characters    Print user text    The text on the screen of the oscilloscope will be printed  e g      Measurement 15    Note  The ASCII presentation of the oscilloscope character set is    printed  e g         dec 25   oscilloscope character Q  e dec 25   ASCII character      Example   of setup text     Read the text from setup register 7 of an oscilloscope     90 DIM SETUPS  22    100 PRINT  1  QT S 7   110 GOSUB 1000   120 LINE INPUT  1 SETUPS   130 PRINT SETUPS                     Query setup Text command 
140. e to 200 us     The signal is displayed with four periods on the screen     HOW TO USE THE INSTRUMENT 4 23       TIMEBASE AUTO RANGE    ANG        AUTO RANGE function continuously adjusts the timebase to a     display of 2 to 6 waveform periods       Press the AUTO RANGE key     Observe that the timebase of the signal display changes to display 2 to 6 signal  periods  AUTO TB in the upper right corner of the display indicates that the AUTO  RANGE timebase function is active       Press one of the TIME DIV keys     The AUTO RANGE function is switched off and the AUTO TB indication  disappears     TIMEBASE MAIN TB VAR    siemens Fine adjustment of timebase speeds between the 1 2 5 steps    can be made by simultaneously pressing the MainTB TIME DIV  keys   WS    ST6710                  Activate the VAR function by simultaneously pressing the TIME DIV keys     Press one key of the MainTB VAR keys     Observe that the timebase indication is now changing continuously instead of  following the 1 2 5 step sequence    The VAR timebase readout values are calibrated values  This enables accurate  timing measurements using VAR timebase settings       Adjust MainTB VAR to 850 us div     Turn off the VAR function by pressing the MainTB TIME DIV key pair  simultaneously     Observe that the timebase is set to the nearest step value  1 ms div      In the Digital mode  the timebase speeds are determined by an XTal oscillator     4 24 HOW TO USE THE INSTRUMENT       TIMEBASE MAGNIFY    MAGNIFY
141. e wave with a lower  level of OV and a top level of 600 mV     HOW TO USE THE INSTRUMENT 4 55       MATHEMATICS FILTER      Press the ANALOG key to select the digital mode     Press the MATH menu key     The MATH menu is displayed on the screen  Observe that MATH 1 is the default     Press the softkey next to MATH 1 to enter the MATH 1 submenu     This menu is used to select one of the four mathematic functions     add  sub  mul   filter        Observe that the default process for MATH 1 is the filter    function  Typical use of  the digital low pass filter is to suppress noise even after a  single shot  acquisition  has to be captured    You can select between the acquired signal or a stored signal to be the source   The trace information in the source trace is mathematically filtered by a low pass  filter    The resulting trace is automatically stored in memory location  m1 and displayed  on the screen       Turn the TRACK control to select  acq      Press the ENTER    softkey     Press the key next to    on off    to turn MATH 1 on     BuU     SdB 4 25kHi 3dB 1 112kH           200mV        20045 hir  CH1  200     MTB 200us h1z                                                                                                                             576745      Turn the TRACK control to change the cut off frequency     Switch off MATH 1 again before continuing to the next section     4 56 HOW TO USE THE INSTRUMENT       MATHEMATICS MULTIPLY    You can activate a second proc
142. ear all    leads to a  confirm menu  The memory locations s1 to s10 are emptied altogether when  yes   is selected  Selecting  no  stops the clear action     Note    Screen messages are given in the CRT viewing area when incorrect  commands are given and the action is prohibited  Example  When one  attempts to save settings to a protected memory location     Key sequence     SETUPS  Settings memory selection        2   gt            save Softkey to recall settings in selected memory    location          redi Softkey to recall settings from before the last recall    action          undo          Softkey to save the actual front settings     CLEAR amp     Softkey that leads to a dedicated menu to switch  PROTECT        the memory protection of the settings and to clear  memory contents              576766  8303    Remote commands     CPL  SS  RS  PT  QT           To save a front in a desired memory    To recall a front from a desired memory   To program text into a setup register    To query text from a setup register                   Refer to Chapter 6 for details    5 68 FUNCTION REFERENCE       SETUPS SEQUENCE    Description     If front panel settings are stored in memory locations    s1 to s5    and    s6    is cleared   then the range of  s1 to s5  is referred to as a sequence  Such a sequence can be  used as  part of  a step by step testing procedure  There are two possible ways  to step through a sequence without have to recall each individual setup using the  menus   
143. easurements  You need only one single push button  This  push button is located nearest to the place of measurement on the probe   therefore  this button is referred to as COMMAND switch                                                                                                                 oooo                   0 0 0 o o 1m0 o        000       gt         O  O oom     a  2 03  D 4  nm     r3 o o       o n o oO   800800800800                                     Before using the COMMAND  fi e            switch on the probe  you must  m       first adapt the instrument  reaction to this function  This is  done in the  UTILITY  gt  gt     COMMAND PROBE  menu        2 576837    9303    The procedure to set up for TOUCH  HOLD and MEASURE      is       Press the UTILITY key     Press the softkey labeled  PROBE      Select    q meas          Press the COMMAND push button on the probe     Observe that the screen displays the signal frequency and amplitude dc voltage  level  dc  pkpk  rms  and freq  measurements     The probe COMMAND switch gives you the fastest and easiest method of  trouble shooting  You can keep your eyes on the circuit to be measured without  having to move your eyes to press a push button on the scope     NOTE  The probe COMMAND switch can also be used for other functions   Programming of other functions is done in the UTILITY menu   see section 5     4 54 HOW TO USE THE INSTRUMENT       4 12 PROCESSING FUNCTIONS    MATH     Most oscilloscopes  including m
144. ed on the available screen area  Sensitivity  adjustment is done with key pair AMPL VAR or a single AMPL toggle key  for the  External Trigger input      Input sensitivity increases when the top key  mV  is pressed  it decreases when  the lower key  V  is pressed  The sensitivity adjustment can be done in steps or  continuously  switching between these modes is done by pressing both keys  simultaneously  The CRT readout resolution correspondingly changes  The  amplitude of a signal can be determined as described under the SCREEN  CONTROLS AND GRATICULE  function  The AUTO RANGE key activates a  continuous automatic adjustment of input sensitivity  The result is a signal  amplitude on the screen between 2 and 6 4 divisions  This functions in digital  mode for input sensitivities up to 50 mV div  For more information  refer to the  AUTO RANGE function     For the External Trigger input  the input sensitivity can be adjusted to the two most  commonly used settings  1 V div and 100 mV div   Selection is done with a single  toggle key AMPL                                                                                                                          RANGE           y         CH1 100m   CH1 500        4  AMPL div AMPLidiv          Function of key pair AMPL VAR  Key sequence   AMPL y  e Key pair to adjust the vertical input sensitivity         coarse or fine steps      gt       V    ST6182    5 40 FUNCTION REFERENCE       AUTO   RANGE Toggle key to switch the AUTO RANGE fun
145. ed with the same TRIG 1  TRIG    and EXT TRIG  keys that are used for the main timebase trigger source and slope selection  The  delayed timebase source and slope have their own readout  This readout can be  found at the bottom right corner of the screen  below the readout for the main  timebase trigger source and slope     If the channel and trigger coupling are the same  e g   both ac or both dc   the  trigger level is indicated on the screen by D      For more details refer to the DEL   D TB section     Key sequence        DEL D TB  DTB on off        C Toggle softkey to switch DEL D TB  on    starts  trig d  Toggle softkey to select the DEL D TB trigger mode            Control to adjust the DEL D TB trigger level   ac       re            Toggle softkey to select the DEL D TB trigger coupling modes ac         If  reject and hf  reject     576774  9303    TRIGGER LEVEL    Description     LEVEL selects the signal level at which the timebase will trigger  For triggering   the level must be within the peaks of the signal     TRIGGER LEVEL sets the trigger level for the main timebase  In level pp    on     mode  the level range is clamped within the peak peak values of the signal  Then  the MTB will always be triggered as shown in the following figure  When level pp  is    off     the level range is from  8      8 divisions     Leveling in DEL D TB is adjusted with the A control  It is activated in the triggered  DEL D TB mode  The range is from  8      8 divisions     FUNCTION 
146. edge or TV triggering  In digital mode  also logic triggering     F XX   Toggle softkey to select positive or negative trigger edge  slope    In digital mode also dual slope          ch    ext  trig  line    Toggle softkey to select MAIN TB trigger source  CH1 to CH4 are  to be selected with TRIG1 to TRIG4        ST7691      TRIG   Toggle key to select CH1  CH   or EXTTRIG as MAIN TB         trigger source  Subsequent key presses cause the trigger    Cy slope to change between positive and negative trigger  edge     5 86 FUNCTION REFERENCE       TV TRIGGER    Description     In addition to edge and glitch triggering  explained under TRIGGER MAIN TB    there are extensive video triggering possibilities  These enable stable triggering  on video frames and lines from various TV standards  There is no need to adjust  the trigger level     Triggering is possible on video signals with positive  pos  or negative  neg  signal  polarity  Supported video standards include NTSC  PAL  SECAM  and HDTV  For  HDTV systems a three  level sync trigger is used  and the line count of the system  may be selected     The MAIN TB can be triggered on a TV line  all lines are superimposed   or field  1 or field 2 of any of the four channels  In the field mode  the TRACK control can  be used select triggering on any specific TV line for individual line display     The TV trigger settings are selected from the menu under the TRIGGER menu  key  TV trigger is not available when the optional External Tri
147. eference only  Its position can differ slightly from  the actual trigger level     4 36 HOW TO USE THE INSTRUMENT       TRIGGER MODE TRIGGERED      Press the TB MODE key      EM  In the screen  the timebase mode  TB MODE  menu appears     auto    is     1 intensified  With the  auto  mode turned on  the timebase will run free as  soon as no triggering signal is detected  in order to provide a base line   The    Auto    mode works well with most signal that have frequencies of  30 Hz or higher       Press the softkey next to the  auto trig single    function to select  trig    In the triggered mode  a signal is visible only when proper triggering occurs       Move the indicator  T   in and out of the signal range     TRIGGER                 O     In the Analog mode  the result is a stable and triggered signal when the indicator  is inside the signal range  and no trace appears on the screen when the indicator  is outside the signal range    Disabling the  auto  function by selecting    trig    can be useful for displaying very low  frequency signals  The scope waits for a real trigger before the sweep starts even  if the signal frequency is very low    In the Digital mode  the same criteria exist for the use of the    auto    function and  the    trig    function  but there is a difference  When the scope is in the    trig    mode  and triggering is lost  the acquisition will stop  but the display will reflect the last  information before the triggering stopped  This may lead 
148. elay between the trigger moment and the  moment that the first sample is taken  If the delay is negative  the  first sample is taken before the trigger moment  This is called pre   trigger view  In case of pre trigger view the trigger point  horizontal  position of which T n  0  is indicated on the screen  The delay or  the position of the trigger point can be adjusted by means of the  TRIGGER POSITION control    Delay is usually expressed in X units  Seconds      Note  Traces can also be positioned in horizontal direction with  the X  POS control  but this does not affect the delay  The  reason is that X POS is a pure display function and  therefore does not affect the signal sampling    Delay     Xz    dT corr   Xr     Xu   Trigger pos    Xz    dT corr   Xr     Xu   These values can also be expressed in divisions     Delay       Xz    dT corr   Xr             1 50   div   Trigger pos      Xz   dT corr   Xr             1 50   div      The mathematic functions only give useful results  if there is no conflict  between the units  For example  it makes no sense to add amperes to volts       The trigger delay and the time base setting of the two traces must be equal   This is always guaranteed  if the traces were acquired at the same time   because the trace elements are from the same register       Consider the following useful example of a mathematics function     Perform the multiplication of two traces with different units  i e  a trace  representing volts over time and a tr
149. els  These functions are not available in the analog mode     FUNCTION REFERENCE 5 9       If switching from one mode to another results in an unsatisfactory display  press  the yellow key a second time to return to the original situation     Key sequence     ANALOG  Toggle key to switch between analog mode and digital  mode   AUTO RANGE  Description     The AUTO RANGE function automatically selects the input sensitivity of the  vertical channels and time div of the main timebase to obtain an optimum display  of the input signal s   AUTO RANGE is a continuous function  vertical channel or  time div settings are adapted if the input signal changes     The AUTO RANGE function of the vertical channels automatically selects the  input sensitivity  The result is that the input signal is displayed with 2 to 6 4  divisions of amplitude  The amplitude is between 1 and 3 2 divisions with  DISPLAY WINDOWS on  AUTO RANGE can be turned on separately for each  vertical channel  AUTO RANGE functions at input sensitivities between 5V and  50 mV div  Input coupling is put to ac if AUTO RANGE is switched on  dc or  ground coupling can be selected manually  AUTO RANGE is switched off after  operation of the AMPL key  the AUTOSET key  the channel   s AUTORANGE key   channel ADD key or changing Acquisition length     The AUTO RANGE function of the main timebase  MAIN TB  adjusts the  timebase automatically so that 2 to 6 waveform periods are displayed  To turn off  AUTO RANGE  press the TIME DIV
150. ements are made using the HISTOGRAM method  The  HISTOGRAM method is used to assure that measurements on signals with  distortion  overshoot  ringing  or noise  establish the most probable signal levels  as references for the 0   and 100   amplitude levels     The next figure shows all parameters used in this method     MAX  HIGH    ee       a a a     2nd DISTAL 2                calx lc                                              icc eot         tte 50   151 MESIAL 2nd MESIAL 3rd MESIAL       2nd PROXIMAL 3rd PROXIMAL                                                   a      dtu M  AGRI             10                      LOW  MIN    ST6743          All calculated measurements are made following a fixed sequence     1  Using the HISTOGRAM algorithm  determine the HIGH LOW levels and the  MIN MAX levels  These parameters are used to define the 096 and 10096  voltage levels of the signal     2  Calculate the MESIAL  DISTAL  and PROXIMAL levels  These voltage levels  are derived from the HIGH and LOW     FUNCTION REFERENCE 5 49       3  Calculate all other signal parameters  The formulas for all time and volt  measurements are given in the sections for DELAY MEASUREMENT  TIME  MEASUREMENT  and VOLT MEASUREMENT     The HISTOGRAM method determines the voltage levels  This method is as follows       The input data    trace  is used to create a histogram  This histogram  evaluates all signal amplitude values that occur in a given signal  plus the  number of occurrences for each ampl
151. en on the type plate  This number  consists of a six digit code preceeded by the characters  DM      The instrument version can also be displayed on the CRT after having pressed  menu key UTILITY and then softkey MAINTENANCE     GETTING STARTED 3 1       3 GETTING STARTED    This chapter provides a 10 minute tutorial intended for those who are not familiar  with Fluke oscilloscopes  Those who are already familiar can skip this chapter and  continue to Chapter 4     3 1 FRONT PANEL LAYOUT    This oscilloscope is a combination of an analog oscilloscope and a digital storage  oscilloscope in the same instrument  The basic signal acquisition and display  functions are identical in both operating modes  Differences will be explained in  the text  Switching between the two operating modes is done with the yellow  ANALOG key     The front panel of the oscilloscope is organized into functional areas  The areas  are discussed in order of typical operation   DELAYED CONTROLS MAIN TIME BASE                                                                                                                                                                   SCREEN CURSOR TIME BASE FOR DIGITAL AND TRIGGER  CONTROLS AUTO SET CONTROLS CONTROLS MODE CONTROLS          m      5 Ur Track CURSORS    HOLD        TRIGGER LU Cb 4X Po   TEXT  lt      Sy a         Ji                 CB Q                              Focus p po EE              m                           p 7 58         aa                    
152. en the oscilloscope is under  remote control  This second function is called  go to LOCAL  and is  indicated in small text below the STATUS key  For more information   refer to the  REMOTE CONTROL RS 232  and  HEMOTE CONTROL  IEEE 488 2  functions     Key sequence     STATUS    Toggle key to select amount of settings information on  screen    LOCAL    FUNCTION REFERENCE 5 71       SUBTRACT  MATHEMATICS     The subtract mode is available in the digital mode of operation   Description     The SUBTRACT function performs a point to point subtraction of two traces   Each point related to the respective ground levels  indicated as           The result of  the SUBTRACT function is a new trace in a different register  This trace can be  scaled and positioned     Scaling is the correction of the resultant trace to fit in the screen  Scaling can be  adjusted by using the TRACK control    Vertical positioning is called offset  It offsets each sample in the resultant trace  with a certain value so that the trace can be  moved  vertically  It is adjusted with  the A control    The scale factor and the offset factor are displayed in the    MATH SCALE    menu   Pressing the    autoscale    softkey automatically selects the settings for the trace to  fit the screen     When subtracting signals with different amplitude settings  the size of the smaller  signal is automatically adjusted to match the size of the larger signal     Newly acquired traces or previously stored traces can be used as 
153. en the oscilloscope was switched off  with the batteries installed     To ensure that you will get the same setup in all cases  press the STATUS key  and TEXT OFF key simultaneously  This will set the instrument in a predefined  default condition  STANDARD SETUP  and a trace will appear on the screen   Text is also displayed at the bottom of the screen     GETTING STARTED 3 3       3 3 SCREEN CONTROLS    The screen controls can be adjusted for optimum trace  text and spot quality by  the controls to the left of the screen                                  Le                   INTENSITY            Olobs          0  2 g                  o E              OJ amp XO OJEXO               0      olgo olgo olg oo INTENSITY    o            9 04                                                                                   988                                                                TRACE  ROTATION    FOCUS    GRATICULE  ILLUMINATION          Figure 3 2 Screen control area EOC reagan    The brightness on the screen is adjusted by two controls  one for the trace and  one for the text       Turn the TRACE INTENSITY control clockwise and verify that only the  brightness of the trace increases       Turn the TEXT INTENSITY control clockwise and verify that only the  brightness of the text increases     The sharpness of the trace and text is optimized by the FOCUS control        When you are making photographs   or are in a dark environment  you   can use the ILLUMINATION control   t
154. ence  51 T1 trg 5 Time between cursor 1 and trigger  52 T2 trg   Time between cursor 2 and trigger  60 FFT_freq Hz Frequency at cursor    61 FFT_ampl dB dBm    dBuV Vrms X Amplitude at cursor      Notes  1  The meas type and suffix unit of the measurement 1 and 2 values  depend on the selections  made with the MEASURE MENU via the  front panel key MEASURE  Refer to the MEASURE MENU of  chapter 5  FUNCTION REFERENCE     2  Delta T can only be obtained  if X Deflection is off    3  Delta X can only be obtained  if X Deflection is on    4  FFT only when MATH  option is availlable     Response 1              acknowledge CR                Response 2                                meas value                               meas type                   e meas value   suffix_unit CR                                  Response 2       Condition        type value unit    value    value  value     No V or L parameter  V parameter specified  L parameter specified    6 28    THE CPL PROTOCOL       meas_type    meas_value    suffix_unit    string of characters  specifying the type of the measured value   e g   Tr     string of characters  specifying the measured value in floating  point notation  e g   98934E 09     string of characters  specifying the unit of the measured value    V        or  Hz     Example   of a single measurement     100    110  120  130    PRINT    GOSUB     1  QM  21   Query for frequency  measurement  1000   Sync on acknowledge       INPUT  PRINT        1         5     
155. end back partial setups because the setup nodes can be send  individually     Refer to  Query Setup  and section 6 8 SETUP     Command      PS  d      pds count           node CR                                                                send more nodes        lt count gt  number of node strings to follow    lt node gt  string of hexadecimal characters  0  9 A  F  representing a  setup node in the oscilloscope   Response   acknowledge CR                      Note           acknowledge   is sent after the setup has been completed     Example   100 PRINT  1  PS      Setup command  110 PRINT  1 N      of strings    120 FOR    1          130 PRINT  1     SETUPS I       Send strings sep   d by SP      140 NEXT I   150 PRINT  1      Send CR to end the setup  160 GOSUB 1000   Sync on acknowledge              These are the strings that are received after a Query Setup  QS  command    THE CPL PROTOCOL 6 19       PROGRAM TEXT PT    Purpose   To program text to an oscilloscope     If the S parameter is specified  setup text is programmed  The text will be set into  one of the setup registers of the oscilloscope  The parameter n specifies the setup  register  A maximum number of 22 characters is allowed  The remainder of the  text field is set to spaces        If the S parameter is not specified  user text is programmed  The text will be  displayed on the screen of the oscilloscope  A maximum number of 64 characters  is allowed  The remainder of the text field is set to spaces     Th
156. end each sample byte  SUM   SUM WAVE  I  MOD  256    Calculate checksum  NEXT I  PRINT  1 CHRS  SUM     Send checksum  GOSUB 1000   Sync on acknowledge    6 26 THE CPL PROTOCOL       QUERY MEASUREMENT QM    Purpose   To obtain measurement data from an oscilloscope   If the V  Value  parameter is specified  one measured value will be returned     If the L  Logging  parameter is specified  an infinite number of measured values  will be returned  Cancelling is possible by sending the ASCII character ESC     If no V or L parameter is specified  one measured value is returned  preceded by  its type  and concluded by its suffix     Note  If averaging is on  the values will be averaged over the number of  average counts     Command                                  OM  n my CR    A                                                     V  only the numerical value of the measurement result is returned         the numerical values of an infinite number of measurement  results is returned  cancelling is possible by sending ASCII  character ESC    27 decimal     n decimal number  specifying the type of measurement to perform   see table on next page     THE CPL PROTOCOL          n meas type 1  suffix unit 7  Note    1 variable variable measurement 1   2 variable variable measurement 2  10 dV V delta Voltage between cursors  11 V1 V Volt cursor 1  12 V2 V Volt cursor 2  13 Vdc V Volt dc  20 dT s delta Time between cursors 2  21 F Hz Frequency  1 dT   30 dx V delta X 9  40 dPh    Phase differ
157. ess MATH 2  This way  two processes can run at  the same time       Connect the Probe Adjust signal to channel 1 and channel 2 and turn on  both channels     Press the    MATH 2    softkey      the MATH menu     The menu displays the MATH 2 functions  MATH 2 offers the same functions and  parameters as MATH 1    For MATH 2 the default parameter is  mult     Using MULT    means that two selected traces are multiplied together and that the  result is stored in    m2          Press the softkey next to MATH 2 to enter the MATH 2 submenu     Observe that the submenu is now optimized for the multiplying process  You can  select between all active traces or memory locations to be the source traces     Turn the TRACK control to select ch1    Turn the A control to select ch2    Press the ENTER softkey to return to the main menu   Press the softkey next to    on off    to turn MATH 2 on     The multiplied trace of channel 1 and channel 2 is instantly displayed on the  screen  At this time the resulting trace is stored in memory location  m2    One example of this process is to multiply voltage across a component and  current through that component to find dissipated instantaneous power       Switch MATH 2 off again     HOW TO USE THE INSTRUMENT 4 57       4 13 DISPLAY FUNCTIONS    When operating in the analog mode  you can set the scope to display XY displays  by selecting     DEFLECTION    mode  In this mode the horizontal deflection is  obtained from one of the input signals  while t
158. ess the      DC GND key to obtain ground coupling     A straight line is now displayed  This is the OV  ground  level of the input  This  level serves as the 0 volt reference for amplitude measurements   The coupling sign 1    now indicates ground coupling     HOW TO USE THE INSTRUMENT 4 15       VERTICAL POSITION    Pos Use the POS control to adjust the ground level to any desired vertical  O position on the screen                                                                                                                          MATA171      Usethe position control to position the line in the middle of the screen   Observe that the channel identifier  1   shifts with the trace     Press the AC DC GND key again to obtain ac input coupling     Since the Probe Adjust signal is a pulse type signal with a 5096 duty cycle  its  mean value is at the 50  amplitude level of the input signal  When the input is ac  coupled  the mean value will be displayed at the ground level of the oscilloscope   As a result of this  the displayed waveform is centered on the screen     4 16 HOW TO USE THE INSTRUMENT       VERTICAL AMPLITUDE      Press the upper key of the AMPL keys    AMPL  Pressing the upper key increases the amplitude of the displayed   Als waveform  As the amplitude of the displayed waveform changes  the  screen readout of the input sensitivity changes as well    m If you adjust the displayed amplitude to 6 divisions  you will notice that   y the readout in the lower left hand
159. ess the STATUS and TEXT  OFF keys      Connect the Probe Adjust signal to channel 1      Shiftthe trace to the upper half of the screen                                                                                                                                                                                                                                                                               as indicated   CH1 500mVi        1ms  nts  DELAYED TIMEBASE MainTB and Del dTB    Pressthe Delayed Timebase DTB key to enter Um  the DELAYED TIMEBASE menu                 Press the  DEL D TB on off    softkey to turn on            the Delayed Timebase  THES  The main timebase trace and delayed timebase  trace are displayed simultaneously  cul sobmve   bTB      Bi                                  The upper trace is the Main Timebase trace  The Main Timebase trace shows an  intensified portion  It may be necessary to adjust the trace intensity with the  TRACE INTENSITY control to the left of the screen    The lower trace is the Delayed Timebase trace  This represents the intensified  part of the upper trace     Since all primary functions have dedicated keys and controls on the front panel   the menu can now be turned off       Press the TEXT OFF button to turn off the menu     4 64 HOW TO USE THE INSTRUMENT       DELAYED TIMEBASE DELAY    Use the DELAY control to select the start of intensified part of the signal in the  upper trace  The intensified part acts like a window over 
160. eturns to the same settings when turned on  Two AA  LR6   Alkaline batteries are used     Note         batteries are not factory installed and must be installed at the  customer s site     Note  This instrument contains batteries  Do not dispose of these batteries with          other solid waste  Used batteries should be disposed of by a qualified   es 9 recycler or hazardous materials handler  Contact your authorized Fluke   Service Center for recycling information     2 2 2 Installation of batteries    Proceed as follows      Remove all input signals and disconnect the instrument line power      Remove the plastic cover of the battery compartment so that the battery  holder becomes accessible      Install two penlight batteries  AA  in the battery holder as indicated on the  battery holder      Reinstall the cover of the battery compartment     Note  Frontsettings and autocalibration data disappear after exchange of the  batteries with the instrument disconnected from the line power  After  battery exchange  it is necessery to press the CAL key after the  recommended warming up time     CAUTION  Never leave the batteries in the oscilloscope at ambient temper   atures outside the rated range of the battery specifications be   cause of possible damage that may be caused to the  instrument  To avoid battery damage  do not leave the batteries  in the oscilloscope when it is stored longer than 30 days     2 3 THE FRONT COVER    For ease of removal and reinstallation  the front c
161. evel indicator is visible  D       If you have set up for triggered operation of the Del dTB as described in the  previous section  and the Del   dTB trace is not displayed  the trigger level needs  adjustment    Adjust the A control until the Del dTB signal is visible                                                        O                      a    516729                                                                                     D          TRIGGER LEVEL INDICATOR    DELAYED TIMEBASE COUPLING    The Del dTB permits the same trigger coupling selection as for the MainTB  ac   dc  LF rej  HF rej   The level indication  D    for Del   dTB  is present only when the  combination of the Del dTB trigger coupling and the input coupling is useful     4 68 HOW TO USE THE INSTRUMENT       4 15 HARD COPY FACILITIES    The oscilloscope offers the capabilities to make a hard copy of the screen  information on a printer or a plotter  The hard copy can include the recorded  waveform s   the relevant scope settings  trace identification  cursors   measurement results and screen graticule     Before you are able to print or plot a hard copy of the information on the CRT it is   necessary to make some preparations      Oscilloscope and printer plotter must be connected through a suitable cable      The oscilloscope and the printer plotter must be set up to the correct interface  parameters     All setup actions are done in the UTILITY menu     SET STANDARD    First set the instrument t
162. f the square wave on channel 1                                                               CH1  204       CH2  1 00       lt  lt          Press the X SOURCE softkey so that    line    is inten    Turn off channel 1     516739    Horizontal deflection is now obtained from the line voltage  On the screen channel  2  sine wave  is displayed against the line voltage  The signals on the horizontal  deflection  line  and the vertical deflection  ch2  have different frequency       Vary the frequency of the generator to get an almost stable picture     The displayed figure is called a Lissajous figure  This lissajous figure has 5  60 Hz  line  or 6  50 Hz line  tops depending on the line frequency  The number tops  multiplied by the horizontal  line  frequency results in the frequency of the vertical  signal                                                                                                                    ch   A  ae         2 ch3   5  ch4  line W C3       2                   50Hz 60Hz               Xzlhe                576006    If the generator is equal or close to the line frequency  the Lissajous figure is an             whose shape depends on the phase difference  Measuring phase difference  using the X Deflection mode is one of the typical applications for this mode     HOW TO USE THE INSTRUMENT 4 59       DIGITAL DISPLAY X versus Y    Similar to the analog mode  the digital mode allows you to display one trace as a  function of another    The source for verti
163. facturing  To program the AUTOSET to be used for STEP TO   NEXT SETUP IN THE SEQUENCE select the UTILITY  gt  gt  AUTOSET submenu  and select setups     AUTOCAL    The AUTOCAL function is used for fine adjustment of the oscilloscope   s input   trigger  and timebase circuitry  This allows you to achieve the high accuracies  specified for this instrument  even under extreme environmental conditions like  very high or very low temperatures    When the oscilloscope is always kept under the same environmental conditions   e g  in a workshop or lab    it is sufficient to perform this AUTOCAL once every  month    It is recommended that you perform an AUTOCAL after the instrument has  warmed up     To perform an AUTOCAL      Press the CAL key for at least 2 seconds     The calibration is fully automatic and takes about 4 minutes    For purposes of full traceability  an official calibration in a fully equipped and  traceable calibration laboratory should be done once a year or after every 2000  hours of use     HOW TO USE THE INSTRUMENT 4 75       OTHER FEATURES    You are now an experienced user of this Fluke oscilloscope     The following features were not covered by this step by step introduction     Trigger and ground level indicators     TV and HDTV trigger modes and line selection     Trigger filters     State pattern and glitch triggering  Glitch only in 2 ch  models     Event counter delay     Record length selection 8k 4k 2k 512 samples     User text     Intensity ratio Main
164. frequencies to be displayed with optimum  resolution  The time scale is adjusted with the key pair TIME DIV  VAR    Adjustment can be done in steps or in a continuous range  VAR   Switching  between these modes is done by pressing both keys simultaneously  For low  frequency signals  the MAIN TB speed must be slow  this is obtained by pressing  the    s    side of the key pair  For high frequency signals  the    ns    side is pressed   Time values of a signal can be determined as described under the  screen  controls and graticule  function  The AUTO RANGE key activates a continuous  automatic adjustment of the TIME DIV of the main timebase  The result is a  display of between 2 and 6 waveform periods on the screen  This functions in  digital mode  For more information  refer to the AUTO RANGE function    If AUTO RANGE is turned on when VAR is active  the timebase is changed in  such a way that the trace on display is kept unchanged     In the digital mode  the timebase speeds are determined by an XTal oscillator   VARiable control of the timebase speeds in digital mode happens in bigger steps  than in analog mode          AUTO  RANGE             TIME DIV  S            ns                                                                            Function of key pair TIME DIV  VAR      Key sequence     TIME DIV  Sw           ns    Key pair to adjust the MAIN TB time scale for fine or  K  D  coarse steps    ST6839  9303    AUTO  RANGE Toggle key to switch the AUTO RANGE functio
165. g identified as H High  L Low  or X Don   t care    Example  HHxL      There are three logic trigger modes  pattern  state  and glitch     To enter the logic trigger mode select    logic    from the TRIGGER MAIN TB menu   The menu now displays the choices  and you use a softkey to select one of the  three logic trigger modes  pattern  state  or glitch       State   State triggering  sometimes called conditional triggering  is used in applications  with clocked or synchronous systems  One of the four input signals must be  selected as the    clock    signal  Triggering occurs on this clock edge  when the other  3 inputs match the selected three bit pattern  i e   are TRUE   The clock edge may  be chosen as a rising or falling clock edge  In this menu you use a softkey pair to  select the trigger slope in combination with the keys TRIG1 through TRIG4 that  give the state  HIGH  LOW  or don   t care  for the three other trigger sources     Example of a STATE trigger word  HLL  in which ch1 is used as positive clock  edge       Pattern   The PATTERN mode is used to trigger the oscilloscope when a combination of up  to four signals is true for a specified amount of time  In other words  triggering  occurs only when two conditions are met  the combination and the time limits   The trigger pattern is set as a combination of HIGH  LOW  or don t care  of the  four input channels of the oscilloscope     Example of a PATTERN  HHxL     The PATTERN condition is set or edited with the TRIG
166. g readout and the   X  indicating where the trace and the cursors intersect  Since the trace  is digitized  the cursors can be really smart  In the digital mode you can  measure time differences  AT  and amplitude differences  AV  at the  same time              cht  AT  500us 3 600mV CURSORS    on off           1         h                      FREAD                                                  aoe  2          W    CH1 200 ROOS              Step 12 Use the first blue softkey to turn the cursors off  The cursor menu  disappears     GETTING STARTED 3 19       3 13 MORE ADVANCED TRIGGER FUNCTIONS    Most of the trigger functions  source  slope  and level  can be controlled with  direct access to the functions  see Section 3 9   A CRT menu is used for more  advanced trigger functions     TRIGGER                         nad                         o  oper o  nnns   boo O         ARM D    2   18 4  0400 0            LEVEL                         a  EogoaBgaaoasoogagaoug                                MODE       Q                        000000 00                               gloosoo                ST6432  9303  Figure 3 10 More advanced trigger setup    Press the menu key TRIGGER  This TRIGGER  turns the menu on  An extensive set of MAIN TB  functions is now displayed     AL    edge tv  glitch  ch1  ext  trig  line  level pp  on off    noise  on 7X    LILL  AJLLLLBDBDLELLLDLBILLLLILLLLI    ac Zo   If rej   hf rej  ST7688       All functions are explained in Chapter 4  For most a
167. gger is selected as  trigger source     Note     trigger edge can not be selected    FUNCTION REFERENCE       Key sequence     edge tv  glitch    TRIGGER                            Toggle softkey to switch between edge and tv triggering   In digital mode also glitch triggering     Toggle softkey to switch between TV triggering on field 1   field 2  or line sync pulses     Control to select the line number     Toggle softkey to select between pos itive  and  neg ative  signal polarity     Softkey pair for video system selection     Softkey pair for selection of max  number of HDTV lines     Selects the channel and toggles between pos itive  and  neg ative  signal polarity  no edge selection     5 88 FUNCTION REFERENCE       USERTEXT    Description     Two lines of user definable text can be displayed in the CRT viewing area  The  text may be useful as additional information when taking photographs  The  selections are reached via the key sequence  UTILITY  gt  gt  SCREEN  amp  SOUND  gt  gt   USERTEXT  gt  gt  on        The EDIT USER TEXT menu gives the following editing modes      The TRACK control determines the position where text has to be edited  The  position is indicated by a flashing cursor      The A control is used for character selection      Softkey    space    changes the character under the cursor into a space      Softkey    delete    erases text under the cursor      Softkey  insert  adds a space under the cursor     Key sequence     SCREEN amp  USER on off  UT
168. h function and gives examples are given  in a step by step sequence    Less experienced oscilloscope users should read Chapter 3 before beginning this  chapter     4 1 INTRODUCTION    All of the oscilloscope models in the PM337xB  PM338xB and PM339xB family  combine the features and operation of an analog oscilloscope with that of a full   featured Digital Storage Oscilloscope  DSO   Switching between one mode of  operation to the other is done by pressing one  yellow  push button    Most signal acquisition functions are identical for both modes of operation  even  though the digital mode allows for pretrigger acquisition and display  plus more  powerful logic triggering    Delayed sweep operation is available in both operating modes    Cursors operate in both modes    The digital mode provides access to many powerful calculated measurement and  signal analysis functions    It also has a completely new feature called AUTO RANGE    AUTO RANGE automatically adjusts the attenuator or the timebase setting when  the amplitude or the frequency of the signal has changed     This family of oscilloscopes is available with a 60 MHz  100 MHz or 200 MHz  bandwidth  There are full four channel instruments as well as 2 2 channel  models    The economy version have 2 full channels and an external trigger channel     PM3394B 200 MHz Full Four Channel Oscilloscope     The PM3394B offers a 200 MHz bandwidth  Four channels provide equal  bandwidth  and sensitivity ranges  Each channel has full 
169. hat channel 2 is selected as the trigger  source  The result is that the signal on channel 1 is no longer triggered   not stable   The ARM D LED is on  to indicate that the oscilloscope is  not triggered  Check also that the right side of the bottom text line  indicates the trigger source     ch2         Only in 2 channel models   Press EXT TRIG key to select External Trigger input as the trigger  source  Check that the rightside of the bottom textline indicates the  trigger source    EXT        Press the TRIG1 key  Channel 1 is now selected as the trigger source   The  ch1  symbol is displayed in the bottom text line  Triggering  resumes  Turn channel 2 off  by pressing ON again     Press the ns key of the MainTB TIME DIV keys until the timebase is  set to  2 us div        3 14 GETTING STARTED       Step 6 The same TRIG1 key that was used to select the trigger source is also  used to select the trigger slope  Repeatedly pressing the TRIG1 button  changes the triggering so that it occurs on the leading or trailing edge  of the input signal  Note that the slope is also displayed in the bottom  text line                                         Nee    CHI 200mV           24s   CH1  CH 00mV           246   CHIL    576685                                                                                        Step 7 For repetitive signals  you obtain a stable display when each  successive timebase sweep is triggered at the same stable level of the  input signal  You use the TRIGGER 
170. he analog timebase generator is  turned off    In the digital mode the DISPLAY    menu is extended     STANDARD SETTING    Start with the standard setting  This ensures that you have the correct start       Simultaneously press the STATUS and TEXT OFF keys     Connect the Probe Adjust signal to channel 1     Press AUTOSET     The Probe Adjust signal now supplied to the input is a square wave with a lower  level of OV and a top level of 600 mV     ANALOG DISPLAY X DEFL    In X DEFLection mode   XY  displays are used for direct comparisons of two or  more signals   one as a function of one or more others     Vertical deflection is selected in the same way as for  normal  displays using the  timebase  Horizontal deflection is obtained by selecting one of the sources as  displayed in the X DEFL menu on the screen       Connecta sine wave of approx 300 Hz to channel 2     Press AUTOSET     Observe that the scope displays the two traces on the screen in analog mode       Press the DISPLAY key to enter the analog DISPLAY MENU     Press the X DEFL softkey to enter the X DEFL sub menu     Press the  on off    softkey so that  on  is intensified     Two vertical lines appear on the screen   The menu now displays the source for the X axis intensified   X2ch1       4 58 HOW TO USE THE INSTRUMENT       The signal on channel 1 is a square wave  Consequently   there will be only two vertical levels to display    The signal on channel 2 is a sine wave  which is now  displayed as a function o
171. he front panel of your oscilloscope and observe what functions the different  controls and push buttons  keys  perform  There are three different styles of push  buttons  plus the blue softkeys adjacent to the screen  The push button functions  are as follows     C     C   s      C3        Direct function key  These keys provide direct access to specific  functions as labeled on the front panel  Examples include  AUTOSET for automatic setup of the oscilloscope and AC DC   GND  for selection of the instrument s input coupling     Menu initialization key  with blue text  Press to produce a menu on  the screen from which you can select functions that are related to  the function name of this key     Key pair  These pairs serve as up down controls  They are used to  select a value from a range     Softkey  Press to select a function from the menu that has been  initialized by pressing one of the menu initialization keys     Rotary control  Used for continuously variable control of a function     HOW TO USE THE INSTRUMENT 4 5       4 2 DISPLAY AND PROBE ADJUSTMENTS    To help you follow the step by step descriptions  each section begins by recalling  the standard setting as outlined below    If you get  lost   you can return to the beginning of each section  because all  functions are set to a predefined state to create a correct start situation     RECALL STANDARD SETTING      Simultaneously press the STATUS and TEXT OFF keys     Recalling the standard setting always results in a
172. he identification of the oscilloscope  It gives information about the model  number  the version numbers of all software modules and the installed options     This Identification  ID  command gives the same information as can be read from  the oscilloscope screen after pressing the frontpanel knob UTILITY and the  softkey MAINTENANCE     Command            ID  CR                         Response 1           acknowledge CR                   Response 2           identity string CR                   Note  The items in the identity string are separated by a      ASCII 59     Example    100 PRINT  1  ID    Send command   110 GOSUB 1000   Sync on acknowledge  120 INPUT  1 ID    Input Identity string    130 PRINT ID    Print Identity string    6 14 THE CPL PROTOCOL       example response     FLUKE  PM 3380B 0 SW3394BI V4 0 1996 10 02 UHM V1 0 UFO V2 0  IEEE                    p x 3 ML M EE RE       1   manufacturer  2   model number of the oscilloscope  3   information about the oscilloscope software  4   information about the micro controller software  5   information about the frontpanel control software  6   information about installed options  e g      IEEE  IEEE interface     EXT  External Trigger option  only in 4 channel models     EM  Extended Memory     MP  Math Plus     THE CPL PROTOCOL 6 15       LOCAL LOCKOUT LL    Purpose     This instruction will inhibit the Go to Local function of the STATUS LOCAL key on  the frontpanel     Once activated  the Local Lockout State is d
173. he information provided in this manual     Fluke Corporation Fluke Industrial B V   P O  Box 9090 P O  Box 680  Everett WA 7600 AR Almelo  98206 9090  USA The Netherlands    Copyright   1997  1998 Fluke Corporation  All rights reserved  No part of this manual may be reproduced by any means or in  any form without written permission of the copyright owner     Printed in the Netherlands                                                                                                                                                                                     TRACE AUTOSET CAL SETUPS UTILITY ANALOG ACQUIRE SAVE RECALL MEASURE MATH DISPLAY HARD COPY  INTENSITY    OOO         STATUS  trac   CURSORS HOLDOFF TRIGGER MAGNIFY     4XPOSP               INTENSITY LOCAL RUNISTOP RANGE SINGLE ARMD  1 1   TRIGGER    TRIGGER  DELAYED TIME BASE DELAY postmon LJ LEVEL  p               sTME DIV ns TB MODE AT  TRACE          ROTATION                LJ  Pos   Pos     Pos   Pos       pO BO BO                  AUTO        AUTO   AMPL RANGECHIiCH2 AMPL RANGE INV AMPL RANGECH3 CH4 AMPL RANGE INV     hw V   yw  GRATICULE 4 4L  4  ILLUMINATION                      4 ACDC    ATC     TEXT OFF SOW m 5      S      S ON  4 4 4 4              v                                                         POWER Probe Adjust                        Oz               1M 25pF IMA 25pF IMA 25pF IMA 25      ST7721A             Thank you for purchasing this FLUKE oscilloscope  It has been designed and  manufactured t
174. he screen  and  calibrated in divisions  The trigger point is indicated on the screen with an arrow   Pretrigger recording can be adjusted and is limited to one full screen  or  10  divisions       Turn the TRIGGER POSITION control clockwise     When the control is rotated clockwise  delay is added  The trigger point shifts to  the left  The delay can be adjusted to a maximum of1000 divisions  so that it is  outside the screen  Positive trigger delay is indicated in seconds or in fractions of  seconds       Set the delay to 0 with TRIGGER POSITION    HOW TO USE THE INSTRUMENT 4 29       TRIGGER SINGLE SHOT      Make sure the instrument is in the Digital Mode     Press AUTOSET     This sets up the scope with the proper amplitude and timebase settings       Remove the probe tip from the Probe Adjust output     Turn the TRIGGER POSITION to mid screen  counterclockwise      Press the SINGLE key     This automatically prepares the scope for a single acquisition    Observe that the screen has been cleared  and that the ARM   D LED is on  This  indicates that the scope is armed and waits for a trigger  The trigger level is  automatically set to  5 divisions  indicated by    T            Look at the screen and touch the Probe Adjust output with the probe tip     Observe that the oscilloscope    runs    once and that a signal appears on the screen   The ARM D LED is turned off  At the moment you touched the Probe Adjust output  with the probe  the scope was triggered  The trigger mome
175. he upper curve and  the lower curve of an AM modulated signal appear on screen as an envelope of  min max pairs  and the area in between is shaded     In the two channel mode  glitches down to 10 ns  i e   1 100 Ms  can be captured   In single channel mode  even glitches down to down to 5 ns can be captured     GLITCH CAPTURE IN FOUR CHANNEL MODE    If more than two channels are activated and the glitch detector is switched on  the  scope automatically selects the alternating mode  Since the alternate mode  allows each ADC to run at full speed  10 ns glitch detection is now possible on all  four channels     Single shot acquisition with glitch detection is possible only in one  and two   channel modes        2 APPENDIX P       CHANNEL ACQUISITIONS    For three  or four channel acquisition  the ADCs are used in conjunction with the  chopped mode or alternate mode     In the chopped mode the first ADC takes a single sample on channel 1  then one  on channel 3  then one on channel 1 again and so on  The second ADC chops  between channels 2 and 4  The chopping rate is so high that the effective sample  rate is 5 MS s on four channels in parallel  This is more than adequate for single  shot capture in most applications  power switching  audio  process control  etc    and certainly for electro mechanical applications  which are typically in the kHz  range     The advantage of a high speed chopper is that it allows single shot acquisitions  to be taken on four channels simultaneousl
176. information from the adc is then stored in a Acquisition Memory  The data  acquisition in this memory can be stopped to freeze the trace on the screen  You  can also store  save  complete registers or separate signals into a  background   memory  called Register Memory  Traces can be stored as long as you like and  can be recalled at any time    The CombiScopes in this range have a standard memory of 8K  The record length  for each trace is the maximum number of samples divided by the number of  traces  With the standard scope  the memory depth of 8K is available for two  channels  The memory is partitioned so that 1x8K  2x4K or 4x2K records can be  captured  For maximum update rate  records can be made as short as 512 points   If the memory expansion option is installed  the record length can be adjusted  between 32K and 512 points     HOW TO USE THE INSTRUMENT 4 31       4 7 ADVANCED VERTICAL FUNCTIONS    All basic functions of the oscilloscope are accessible via direct action front panel  keys    More advanced functions are easily accessible via the menus behind the menu  initialization keys  keys with their function name in blue text      STANDARD SETTING  Before continuing with the advanced vertical functions  you must set the  instrument to a predefined state to ensure a correct start situation       Simultaneously press the STATUS and TEXT OFF keys     Connect the Probe Adjust signal to channel 1     The Probe Adjust signal  now supplied to the input  is a square wave 
177. is provided to customize the output to the printer or plotter  This  submenu changes  depending on the printer or plotter that has been selected     Grid If    yes     the complete grid  screen graticule  with  divisions is printed plotted  If    no     only the square  border in printed plotted    Paper size This can be 11  or 12  for printer and A4  A size  or  A3  B size  for plotters    Plot format  plotters only  This can be 1 1 and 2 1     Trace information If    yes     the settings readout and measuring results  are printed plotted   Colors  plotters only  If yes     the traces are plotted in different colors  If       no     all traces are plotted in one color     FUNCTION REFERENCE 5 55       Real time clock    The real time clock is used to make time stamps on a hardcopy  Hardcopies will  be stamped with two timestamps  the time of operation of the HARD COPY key  and the time of trigger of the acquisition     Adjust block    Date format    Confirm    Key sequence     PRINT amp   UTILITY     PLOT amp CLK       Three softkeys are used to adjust the clock  The  softkey in the middle determines if the adjustment  is done for day  month  year  hour  minutes or  seconds  The item to be adjusted is displayed  intensified  The adjustment is done with the upper  and lower softkey  The upper softkey s gives an  increase  the lower softkey t a decrease     The date can be selected to be in the European  format  day month year   USA format   month day year  or Japanese format 
178. is turned off    In the digital mode  the display menu is extended with additional display functions  such as  WINDOWS  VERTICAL MAGNIFY  X vs Y  TEXT GENERATOR and  INTERPOLATION     For the X DEFLECTION function for the analog mode  refer to X DEFLECTION  further on in this chapter     WINDOWS automatically separates the traces on the screen when more than one  trace is displayed  Each trace is then displayed in a window    When two channels are displayed  one is displayed in the upper half of the  screen  and the other is displayed in the bottom half  When three traces from  different channels are displayed  the screen is divided into four windows of two  divisions each    Even when only two divisions are allocated to a trace  for each trace is displayed  with the full dynamic range of 8 bits and the full 256 ADC levels are still used  The  displayed amplitude of the signal and its screen readout  has however  been  adapted to the new window space  Also the range of the POS controls is limited  to the new window  The windows mode is especially useful in multichannel  applications  It eliminates trace positioning and scaling  while retaining maximum  resolution  accuracy and trigger sensitivity     VERT MAGNIFY is used to vertically expand signals without changing the  AMPL DIV setting of the input channels  It can be used to study signal details in  high resolution acquisitions such as an averaged waveform  The signals can be  expanded up to 32 times  and the magnification
179. isabled by sending the Go to Local   GL   the Reset Instrument  RI  command or by cycling power OFF and ON    Refer also to  Go to Remote  and  Go to Local       There is no front panel equivalent for this command     Command            Dp  CR                         Response           acknowledge CR                   Example   100 PRINT  1  LL       Send Command  110 GOSUB 1000   Sync on acknowledge    Local  Remote  Remote Local Lockout States                 Go to                    Local State Remote State   1            Go to Local      Reset Instrument                                            lt STATUS LOCAL button gt    Local Lockout   Remote with   Go to Local  Local Lockout   Reset Instrument  State             6 16 THE CPL PROTOCOL       PROGRAM COMMUNICATION PC    Purpose     To program baudrate  parity mode  number of data and stopbits and the  handshake method for computer communication     After the command is sent  an  lt acknowledge gt  will be returned with the old  communication parameters still active    If the  lt acknowledge gt    0  the new communication parameters will be valid  approximately 0 5 seconds later  The communication parameters are stored in  battery backup memory and restored on power up     This command performs the same function as the UTILITY   RS232 SETUP menu     Command                     PC  pds baud pds parity pds data bits                                                       to bypass handshake                                  
180. ising slope of the input signal  At least one rising slope must  be displayed on the screen to get a valid measurement     rise  time of distal of first rising slope   time of proximal of first rising slope      fall  FALL TIME    Measures the fall time of a signal  It is calculated between the proximal and  distal of the first falling slope of the input signal  At least one falling slope must  be displayed on the screen to get a valid measurement     falla time of proximal of first falling slope   time of distal of first falling slope    duty     DUTY CYCLE      Measures the percentage time of the positive part of the first period of the signal  as compared to the period time  At least one complete signal period must be  acquired to get a valid calculation     2nd mesial of pos part   1st mesial of pos part  time of 3rd mesial   time 1st mesial       duty     100     It is possible to perform measurements on the part of the waveform between the  two cursors  This function is called cursor limited measurements and is turned on  via the key sequence MEASURE      CURSOR LIMIT  amp  STATIST      CURSOR  LIMITED yes  Cursor operation is done via the TRACK and A controls and via the  CURSOR menu     FUNCTION REFERENCE    5 77       Results of the measurements MEAS1 and MEAS2 are displayed in the top left  corner of the screen  When you press the key sequence MEASURE  gt  gt  CURSOR  LIMIT  amp  STATIST  gt  gt  STATIST on  the screen displays three values per MEAS  function  These
181. itude value found in the trace  The  highest number of occurrences at the high end of amplitude range determines  the HIGH level  when the count is over 5  of the total number of samples   Under the same condition  the highest count in the low end of the amplitude  range determines the LOW level        If the 5  limit is not reached by any count  the HIGH level is set to the MAX level  and LOW is set to the MIN level  MAX and MIN are the highest and the lowest  detected amplitude levels  This occurs for example when the signal is a sinewave       The DISTAL  MESIAL  and PROXIMAL level are now derived as follows   DISTAL   0 9    HIGH LOW   MESIAL   0 5    HIGH LOW   PROXIMAL   0 1    HIGH LOW     Remote commands   CPL  QM  Command to query a measurement   Refer to Chapter 6 for full details     MULTIPLY  MATHEMATICS     Description     The MULTIPLY function performs a point to point multiplication of two traces  The  value of each data point is related to ground  The result of the MULTIPLY function  is a trace in a register     The result trace can be scaled and positioned     Scaling is the correction of the resulted trace to fit in the screen  The TRACK  control us used to adjust scaling    Vertical positioning is called offset  It offsets each sample in the resulting trace  with a certain value so that the trace can be  moved  vertically  The A control is  used to adjust vertical positioning    The scale factor and the offset factor are displayed in the    MATH SCALE    me
182. le acquisition      Touch the probe tip to the Probe Adjust output     The scope    runs    once and a signal appears on the screen       Press the DISPLAY key     Press the VERT MAGNIFY softkey to turn it on     Turn the TRACK control clockwise     The signal magnifies a maximum of 32 times  The magnification factor is  displayed in the DISPLAY menu  At the same time the vertical deflection in the  bottom text area reflects the expanded deflection factor       Use the TRACK control to adjust VERT MAGNIFY to 8     Observe that you can observe  the signal in more detail  It has  the same vertical resolution as  when captured  Notice the adc  noise being displayed                 Vertical magnify is very useful    when you want to observe a TEXT pl  signal detail with very high dots  resolution  You can use it in sine  combination with the horizontal  gt    MAGNIFY function        25 pmV v    MTB 20  us  5        cht sT6725    9303             UUUUUU                                                          Note  If you want to adjust the displayed position of the magnified trace  you  must enter the Recall menu  and use the A control to make any  adjustments to the Y position     HOW TO USE THE INSTRUMENT 4 61       DIGITAL DISPLAY INTERPOLATION    Interpolation is a mathematical way to calculate displayed dots between actually  captured signal samples  Interpolated displays help in the recognition of trace  waveforms  even when the number of samples is too low to render an acc
183. lected only by the TRIGGER menu key and is restricted to the single shot  modes  The medium level is adjusted by the TRIGGER LEVEL knob  The trigger  gap between positive and negative slope is adjusted by the TRACK knob  For line  frequency related signals  a Line trigger source is available  The Line trigger  source is selected by a softkey under the TRIGGER menu key  The External  Trigger input provides an extra trigger input  This input can be used as the trigger  source for the Main TimeBase  MTB  and Delayed Time Base  DTB   The input  characteristics are simular to the input channels  To reduce the effect on triggering  on noisy signals  trigger filters        be used  For this  If rej or hf rej        be  selected    Refer to TRIGGER COUPLING and TRIGGER LEVEL for details     FUNCTION REFERENCE 5 85       The MAIN TB trigger settings are selected in the menu under the TRIGGER menu  key  The toggle softkey    ch   line    selects the trigger source in combination with the  keys TRIG1  TRIG   and EXT TRIG that give direct front panel access to select  the trigger source  This is combined with the slope selection     SLOPE  SLOPE   POSITIVE NEGATIVE         E   0  dI Ne of kee            LEVEL    WAVEFORM WAVEFORM  ON SCREEN ON SCREEN    SLOPE   NEGATIVE    pa    TRIGGER  LEVEL            SLOPE   POSITIVE    F       WAVEFORM WAVEFORM  ON SCREEN ON SCREEN    Function of TRIGGER LEVEL and slope    Key sequence     edge tv   glitch     TRIGGER       Toggle softkey to select 
184. lection  Selection is possible only with X vs Y on     Softkey pair to select source for horizontal deflection   Selection is possible only with X vs Y on     THE CPL PROTOCOL 6 1       6 THE CPL PROTOCOL    6 1 INTRODUCTION    The oscilloscope can be controlled via the RS 232 serial interface using the  Compact Programming Language  CPL  protocol  In this protocol a small but very  powerful set of commands is defined     The main characteristics of the CPL protocol     It is kept simple and straightforward and is fully tailored to use simple  communication facilities like those of BASIC     Special emphasis is put on the ease of programming       easy input format with a    forgiving    syntax   Commands always consist of two characters that can be upper or lower  case   Parameters that sometimes follow the command may be separated from  it by one or more separation characters       Strict and consistent output format   Alpha character responses are always in UPPERCASE   Parameters are always separated by only one comma        ASCII 44    Responses always end with a carriage return  CR   ASCII 13        Synchronization between input and output  After receipt of every command  CPL returns an acknowledge character  and a carriage return  CR   ASCII 13   to indicate reception and or  execution of the command     Commands   All commands consist of a header of two alpha characters  Some of the  commands are followed by parameters to give the oscilloscope more information   The pa
185. ling edge     ARUL  8  FUE    hif hit         4    576025                                                                                                                                           HOW TO USE THE INSTRUMENT       TRIGGER    TRIGGER LEVEL    TRIGGER After each AUTOSET  the trigger level is always clamped within the    LEVEL    signal amplitude range to assure stable triggering on most signals  The     trigger level is adjustable  but it is limited between the minimum    100   and the maximum   100   amplitude levels of the signal       Turn the TRIGGER LEVEL control     As you turn the trigger level control  the  trigger level readout in the bottom of the  screen shows the trigger level relative  to the signal amplitude  For example  it  may appear as    Level   14                                                   17                          4                         TRIGGER  LEVEL     O     576547       4 28 HOW TO USE THE INSTRUMENT       TRIGGER TRIGGER POSITION                   Press the ANALOG key to select the digital mode        Turn the TRIGGER POSITION control counterclockwise     One of the outstanding capabilities of a Digital Storage Oscilloscope is the ability  to capture and display signal details before the trigger moment    When you turn the TRIGGER POSITION control counterclockwise  the trigger  point  A  is shifted to the right  This allows you to look at the signal before the  trigger point  The pretrigger view is displayed in the bottom of t
186. lot       J TRACK 5 TRACK E  HP2225 pi _ HP7475A e      IM   HP7470A L     J HPGL   E  dd mm yy  LAYOUT Pj LAYOUT P hh mm ss     IEEE IEEE    RS232 RS232  5 5 dd mm yy 3  mm dd   E 2 yy mm dd  J  RETURN 4 RETURN 4 RETURN 4  UTILITY UTILITY         PRINTER L gt  PRINTER  LAYOUT LAYOUT  GRID GRID  yes no yes no    1  7         12    M    7 1 1 7  1 2     1  When IEEE option is not present  UTILITY TRACE 7 TRACE 7  REMOTE SETUP menu s are not   0         present and REMOTE SETUP is replaced ee COLORS 1  by RS323 SETUP                    RETURN 4 RETURN 4                      ST7677 1    UTILITY me    nu structure                                                                                                                                     5  7677 2                                           UTILITY UTILITY UTILITY UTILITY  Lp  AUTOSET       AUTOSET   AUTOSET   AUTOSET  PROBE PROBE VERT  AUTOSET   PROBE UNAFFECT 1 CHANNELS 1  off 1 1 yes no scan  default 3 unaffect 5 unaffect 5  userprog ac dc  setups J J uneffect J  1 1 1 1MQ 500                   1 uneffect  TRIG       3 5 BWL 7  PROBE  gt  on off  E     unaffect        RETURN 4 RETURN 4 RETURN 4 RETURN 4  UTILITY UTILITY UTILITY  PROBE                             PROBE  CORR SCALE  PROBE     cht  SWITCH chi ch2  autoset           3 ls 8  setups   e                  5 14 5 5  analog w 10 1 A on off    20 1   SCALE J TRACK  501   2449      J 100 1 J E 06 5             1 UNITS    CORR   SCALE P               RETURN 4 RETURN 4 
187. ls for  horizontal direction  Also the line voltage or EXT TRIG channel can be used for  X deflection  Any of the input channels can be selected for vertical deflection     In the digital mode  X vs Y also uses a signal from one of the input channels    X vs Y displays can also be displayed from signals previously stored in memory   In this case the memory register  e g   m3  must first be selected with the TRACK  control  The horizontal deflection source can then be any of the active trace  registers  e g   m3 2      The selections are made in the DISPLAY menu under the X DEFL or X vs Y menu  softkey  After the function is switched    on     the source selection becomes visible   In all cases the active X deflection source is displayed at the bottom right of the  screen and the Y deflection source s  at the bottom left of the screen     To achieve correct scaling and positioning  use the channel AMPL front panel key  pair  their POS controls and the XPOS control in the analog mode  For the digital  mode  use the VERT MAGNIFY under the DISPLAY menu and the XPOS control     FUNCTION REFERENCE       Key sequence     ANALOG MODE   DISPLAY X DEFL    on off              chi ch2  ext line    DIGITAL MODE   DISPLAY X vs Y on off  TRACK  X SOURCE       Toggle softkey to switch X  Deflection on     Softkey pair to select X DEFL source  Selection is  possible only with X  DEFLection on     Toggle softkey to switch X vs Y on     Control to select the register as source for vertical  def
188. ltage 5 VOLT MEASUREMENTS               5 92  Measure 4 11 MEASUREMENT FUNCTIONS           4 49  5 DELAY MEASUREMENT               5 26  5 MEASURE MENU                     5 47  5          MEASUREMENTS                5 75  5 VOLT MEASUREMENTS               5 91       1 6 FUNCTION INDEX  Entry Refer to chapter function Page  Memory 4 9 MEMORY FUNCTIONS                4 49  Memory back up 2 2 MEMORY BACK UP BATTERIES         2 3  Memory bar 5  CURSORS     eium 5 19  Memory expansion 5   ACQUISITION LENGTH                 5 2  Minimum voltage 5           MEASUREMENTS               5 92  Multiply                                                               5 49  O  Overshoot 5 VOLT MEASUREMENTS                5 93      Peak detection 5 PEAK DETECTION                    5 51  Performance test Reference manual  Period 5 TIME MEASUREMENTS               5 76  Pkpk voltage 5 VOLT MEASUREMENTS                5 92  Plot 4 15 HARD COPY FACILITIES               4 67  5   PRINTING AND PLOTTING             5 53  Post trigger 5 DELAY teo bur be abe 5 24  Power up check 5   CONFIDENCE CHECK                 5 18  Preshoot 5        MEASUREMENTS                5 93  Pre trigger 5 DELAY    ie                  E eee R 5 24  Print 4 15 HARD COPY FACILITIES               4 67  5   PRINTING AND PLOTTING             5 53  Probe 5 PROBE UTILITIES                    5 57  5 UTILITY MENU                       5 89  Pulse width 5   TIME MEASUREMENT                 5 76  R  Random sampling 5  TIMEBASE MOD
189. mode     In the acquisition system of a Digital Storage Oscilloscope  the sample distance  is determined by the time base speed  At higher time base speeds  the distance  between the samples is short  at lower time base speeds  the sample distance is  longer     To capture high frequency signals  or signal details of short duration  a high time  base speed is usually selected  A high time base speed captures a time    window     that is usually equivalent to the length of the screen  So at higher time base  speeds  the time    window    is shorter than at lower time base speeds     Some applications require the time    window    to be long  This is then achieved by  selecting a slow time base speed  to fit the requirement  But at lower time base   speeds  the time distance between the actual samples increases  The result is    that signal details shorter than the sample distance will be missed     The peak detection mode allows the Analog to Digital Convertors  ADC  to  operate at their highest speed  even when a lower time base speed has been  selected  The result is that  even at lower time base speeds  the maximum or  minimum peaks of the signals are placed in memory and displayed  This  technique is referred to as oversampling     Using peak detection  PEAK DET  allows you to capture peak values of the input  signal which may otherwise occur between the samples  PEAK DET operates on  repetitive signals as well as on signals taken in a single acquisition     The shortest e
190. mp  PROTECT  SAVE  n    Press the ANALOG key to select the digital mode    1   Press the SAVE key to enter the  SAVE ACQ TO MEMORY    menu     To be sure that all unnecessary stored traces are removed  you first will clear all  memory locations       Press the  CLEAR  amp  PROTECT    softkey     Press the    clear all    softkey     The scope now displays the confirmation submenu to prevent you from removing  traces you did not want to remove       Press the    yes    confirm softkey     If any memory locations were protected  PROTECT ON    in CLEARS PROTECT  MEMORY submenu   a second confirmation must be given       Press the    yes    confirm softkey again     4 40 HOW TO USE THE INSTRUMENT       TRACE STORAGE SAVE      Turn the TRACK control to select the memory location in which to store  the trace data     Observe that eight memory locations  m1 to m8   or 50 memory locations when  extended memory is installed  scroll through the menu  All locations are marked  with a circle before the memory number       Select memory location    m1        Press the  save  softkey     The signal is saved in memory location m1   Observe that the circle before  m1  is solid now to indicate that this memory  location is occupied       Setthe MainTB to 500 us      Select location  m2       Press the  save  softkey      Set the MainTB to 200 us div     Select location    m3         Press the    save    softkey     Observe that memory locations  m2  and  m3  are now solid bullets as well     
191. n displays three values per MEAS  function  These values are measured over the total number of acquisitions for that  particular measurement and are updated instantly  The statistic measurement  values are as follows       absolute minimum value    absolute maximum value    mean value    FUNCTION REFERENCE 5 95       Key sequence        volt  time  MEASURE MEAS1 2  delay Toggle softkey to select time measurements in menu  C  2 7 MEAS1      MEAS2   TRACK    Control to choose the volt measurement     Control to select the signal on which the measurement  must be done        be         on off  Toggle softkey to switch the measurement on and off     CURSOR CURSOR          LIMIT amp  LIMITED  STATIST yes no         C  Toggle softkey to switch cursor limited measurements  on and off   STATIST  on off  Toggle softkey to switch the statist function on and off        ST6780  9312    Remote commands   CPL  QM  Command to query a measurement   Refer to Chapter 6 for full details     5 96 FUNCTION REFERENCE       X DEFLECTION  X DEFL  X vs Y     Description     X deflection creates X Y displays in which one input signal is displayed as  function of another  To obtain X Y displays in the analog mode  the oscilloscope  can be set to X DEFLection mode  In the digital mode a similar function is called  X vs Y mode  In both cases a display of amplitude vs amplitude instead of the  display of amplitude vs time is visible     In the analog mode  X DEFL uses a signal from one of the input channe
192. n of main    a timebase     FUNCTION REFERENCE 5 47                           5    Description     Two mathematical functions  MATH 1 and MATH 2  are independent waveform  processes  These can be used separately  or be chained together if required   Each offers a choice of four mathematical functions  add  subtract  multiply  filter    The result of each function is always placed in a separate register     Independent operation allows two separate processes to be performed at the  same time  By placing the result of a mathematical function in a register memory   it is possible to chain functions together  An example is Ch1xCh2 with its result  being filtered to remove noise  In this example  MATH 1 is the multiply function  with the product placed in m1 and MATH 2 is the filter process  filtering the  contents of m1 and placing its results in m2     Appendix D gives the complete menu structure of the MATH menu     When the oscilloscope has been delivered with the optional    MATH      menu  the  set of functions is expanded with integration  differentiation  Fast Fourier  Transformation  FFT  analysis  and histogram analysis processes  For more  details of these  see the separate MATH PLUS manual     MEASURE MENU    Description     In the digital mode the oscilloscope can perform two calculated measurements   MEAS1 and MEAS2  simultaneously  These measurements are in addition to the  cursor measurements described under CURSORS  Each calculated  measurement can be individually
193. nal to the output amplifier in turn  This  can be done at the end of every sweep  or at a high frequency  The display modes  are referred to as Alternate or Chop modes     Alternate mode  After each sweep of the time base has been completed  the  channel switch selects the next channel to be displayed during the next sweep   The result is that each channel is displayed one after the other  At fast timebase  speeds this mode assures maximum intensity  while slow timebase speeds in  alternate mode result in a flickering display     Chopped mode  In this mode  the channel selector switch operates at a high  frequency  and is no coupled to the time base sweep  This mode is recommended  for slow timebase speeds  The display switches very fast  1 MHz  between the  channels  which results in what appears to be a simultaneous display of all  channels     ALT CHOP is not available with single trace display     In the digital mode channel 1 and channel 2 each have their own digitizer for  simultaneous sampling on both channels  In this two channel mode ALT CHOP is  not active    If a combination of channel 1 or 2  and channel 3 or 4 is selected  the ALT CHOP  is active and functions in the following way    ALT results in a complete memory acquisition of channel 1 or 2 simultaneously  sampled  followed by a complete memory acquisition of channel 3 or 4  also  simultaneously sampled    CHOP results in a single sample being made on channel 1 or 2 followed by a  single sample on channel 3 
194. nctions of the instrument    Function reference    CPL protocol    Function index    Index    IN THE APPENDICES  Menu structures    RS 232    Chapter 1 should be read before unpacking   installing  and operating the instrument     Chapter 2 describes grounding  line cord  fuses   and backup batteries     Chapter 3 provides a 10 minute tutorial intended  for those who are not familiar with Fluke  oscilloscopes     Chapter 4 provides the more experienced user  with a detailed explanation of the major functions  of the oscilloscope     Chapter 5 contains an alphabetized description of  each function  Each description includes an  explanation of local and remote control functions     Chapter 6 provides the CPL commands with an  example of each     The Function Index lists all implemented  functions in alphabetical order     The overall index contains all function names  and reference words in alphabetical order  It  includes the relevant chapter and page number  where more detailed information can be found     Cable configurations          CONTENTS   CONTENTS Page   1 OPERATORS SAFETY                                  1 1   1 1 INTRODUCTION                                         1 1   1 2 SAFETY PRECAUTIONS                                  1 1   1 3 CAUTION AND WARNING STATEMENTS                    1 1   14 SYMBOLS                     a ee a eee 1 2   1 5 IMPAIRED SAFETY PROTECTION                          1 2   1 6 MEASURING EARTH                                     1 2   2 INST
195. ndwidth limiter on off  number of traces   average  envelope     10 16 Common horizontal settings   X deflection on off  mode auto trig single multiple  X deflection  Source  magnify  acquisition length  X position    11 17 Main Timebase settings   timebase speed  triggering edge TV  MTBI on off  MAIN TB on off   trigger slope pos neg  TV trigger F1 F2 line  noise suppression  on off  continuous var  discrete  level pp on off  trigger source   trigger coupling  TV trigger normal hdtv    12 18 Delayed Timebase settings     timebase speed  trigger mode edge tv  DEL D TB on off  edge  trigger slope pos neg  edge trigger source  trigger coupling     THE CPL PROTOCOL 6 51          hex dec meaning   13 19 Events Trigger Delay settings   event counter low high  trigger level low high  trigger source  event  bits  trigger coupling    14 20 SCPI Trigger Settings  SCPI trigger source low high    20 32 Cursor settings   V cursors on off  T cursors on off  rise time on off  control V T  Vpp  on off  rise time 10 90 20 80  readout Vpp V V   cursors on off  V  readout delta absolute ratio  T readout delta 1 delta ratio  channel  FFT read out dBm dBuV Vrms  FFT reference impedance    21 33 Cursor autosearch settings   on off  flank selection  reference selection  abs rel    31 49 Measurement 1 settings   32 50 Measurement 2 settings   Measure bits  measure type  source 1 low high  source 2 low high    33 51 Pass fail test settings   on off  define  action  menu on off  hardware on off    41
196. ne    In the analog mode  magnification is limited to  10  The right key will  activate the magnification  The left key will turn off the MAGNIFY  function  On or off is indicated by  MGN  in the bottom of the screen     Press the ANALOG key to switch the scope to the digital mode     In the digital mode  pressing the right MAGNIFY key expands the  signal in  1   2   4    steps to a maximum of  32 times  Pressing the  left MAGNIFY key compresses the signal to  1 again  When you  operate the MAGNIFY buttons  or when you turn the X POSition  control  a bargraph is displayed showing which part of the digital trace  is expanded     The MAGNIFY key and X POS control can also be used after the  oscilloscope is STOPped     Pd 5 0045 p    2 10 DIV a           START OF TIME WINDOW CAN BE VARIED WITH  X POS OVER THE WHOLE SWEEP RANGE     1 00ms          40 DIV E  ST6684       GETTING STARTED 3 13       3 9 DIRECT TRIGGER SETUP    Now you are ready to set your trigger conditions  You will use one of the channel  selection keys  TRIG1  TRIG2  TRIG3  TRIG4 or EXT TRIG  and the TRIGGER  LEVEL control        ARMD  e TRIGGER  LEVEL                               Ter gue       Figure 3 7    Step 1    Step 2    Step 3    Step 4    Step 5       ST7669    Direct trigger setup    Press the AUTOSET key  The square wave signal of the Probe Adjust  output is now displayed on channel 1  Turn channel 2 on to display a  second horizontal trace  channel 2 has no input signal      Press the TRIG2 key so t
197. ne now displays             In most cases  dc input coupling is used to show ac as well as dc components of  the signal  However  in some cases where a small ac signal is superimposed on  a large dc voltage  ac input coupling must be used  Then only the ac component  is visible on the screen  The text line shows a     or            sign to indicate dc or ac  coupling  Because the calibration signal is a square wave with a low level of OV  and a high level of  600 mV  the screen shows either of the following two displays     AC INPUT COUPLING DC INPUT COUPLING    e                                          LEVEL LEVEL                                                                                        CHA 200           MTB 200   CH1  CH 200              2005                     56682    3 10    GETTING STARTED       Step 6  Step 7    Step 8    NOTE     Press the ANALOG key to enter the digital mode    Press the top one  mV  of the AMPL keys  so that the signal has  maximum amplitude     Press AUTO RANGE and see the signal change to a suitable  attenuator value  When AUTO RANGE is active  the attenuators  automatically adjust when the signal amplitude changes  to keep the  trace on screen     Press the key labeled AVERAGE  Noise in the input signal can be  reduced by using the average function  The random noise is reduced  by calculating the average over the last n scans  average factor can be  set between 2 and 4096     Refer to Chapter 4 for an explanation of the CH1 CH2  TR
198. ng correct or incorrect operation      0  Ok  normal situation     1  Syntax error  ST query may give more info    2  Execution error  ST query may give more info    3  Synchronization error   4  Communication error    Explanation and examples of the errors     Syntax Error   returned when the command is not understood by the oscilloscope for one of the  following reasons       Unknown header  wrong instructions    Data format of body is wrong  e g  alpha characters when decimal data  expected    Execution Error   returned when internal processing is not possible        Data out of  internal  range    Conflicting oscilloscope settings    Synchronization Error   returned when the oscilloscope receives a new command while it is still executing  the previous one       a new command is sent without waiting for the   acknowledge    synchronization     Communication Error   returned when any framing  parity or overrun error occurs in the received data     When an error is detected during the execution of the command       the oscilloscope sends an   acknowledge        the oscilloscope terminates further execution of the command and returns  to the idle state      the oscilloscope is then ready for a new command     6 48 THE CPL PROTOCOL       6 7 STATUS    The Status word gives more information when the acknowledge is non  zero  A  certain bit in the word can be found by performing a logical AND of the status word  with the mask defined below      Logical AND  the words are compare
199. nly the  delta cursor moves                             UUUUUU     READOUT              576730  9303                                           hizi          4 46 HOW TO USE THE INSTRUMENT       CURSORS VOLT      Use the softkeys  to select the amplitude cursors          intensified      Two amplitude cursors will appear in the display    The menu permits the channel to be selected for which the amplitude cursors  apply    When channel 1 is the only channel displayed     ch1    is the only selection                                                                                       CURSORS   on off        Turn the TRACK Sing             control to move both   qm  chi  cursors  and turn the A      Delta  A  control to oni Arr PAY Sep     move only the A         cursor                     cH1 2pomV    MTB 200us hif  CURSORS AMPLITUDE  amp  TIME    Amplitude and Time cursors can be displayed at the same time  When both  amplitude and time cursors are displayed  you must select which cursors are  affected by the TRACK and A controls  This selection is made by toggling the  softkey CONTROL    Channel selection applies only to the voltage cursors  The time cursor always  applies to all channels     To use both types of cursors at the same time  do the following       Press the softkey next to           so that     is intensified      Press the CONTROL softkey so that         is intensified      Turn the TRACK control and observe that only the amplitude cursors  move      Press the
200. nnel 1 signal  on positive m       slope  and the second is listed to           the channel 2 signal  on LICL            negative slope               The delay is displayed in the top f                      corner of the screen        ch    del      us     X               DELAY                                                      Press the RETURN softkey to return to the MEASURE menu     This measurement is very useful when you must compensate unequal cable delays     4 52 HOW TO USE THE INSTRUMENT       MEASURE CURSOR LIMIT    With cursor limited measurements it is possible to perform measurements on a  part of the waveform       Press the softkey next to CURSOR LIMIT amp  STATIST     Set CURSOR LIMITED to    yes        Turn the A control to reduce the area between the cursors     Notice that the measurement result shows dashes when the area between the  cursors gets too small       Press the RETURN softkey to return to the MEASURE menu      Turn off the measurements MEAS1 and MEAS2      Disconnect the probe from channel 2 before continuing to the next  section     HOW TO USE THE INSTRUMENT 4 53       MEASURE TOUCH  HOLD  amp  MEASURE        The probes delivered with the oscilloscope offer a unique and innovative way to  perform a number of functions directly from a push button mounted on the side of  the probe  One of these functions is called TOUCH HOLD and MEASURE           The TOUCH HOLD  amp  MEASURE     function is a very quick way to instantly  display the four main m
201. nt is displayed in the  center of the screen  The left side of the screen displays the part of the signal  before triggering  pretrigger information      SINGLE Probe Adjust   je                                                  p  mV  z JL    576505  9303                                                                                                                            In SINGLE SHOT also  Dual Slope    triggering is possible  See chapter 5  TRIGGER MAIN TB function     4 30 HOW TO USE THE INSTRUMENT       4 6 DIGITAL ACQUISITION AND STORAGE    This section gives you a short introduction to digital acquisition and storage in  order to provide the basic knowledge and terms  This information is necessary for  you to understand all digital statements in the following sections of the manual                               ANALOG    gt  INPUT       DISPLAY  DIGITAL   INPUT signal        data ACQUISITION       sme                    DISPLAY                   REGISTER  MEMORY       T6721    In the analog mode the input signals are directly displayed on the screen     In the digital mode the channels are applied to the input stage of the digital circuit   the analog to digital convertors  ADC   s   The ADC   s convert the analog signal s   into digital data  Depending on the selected timebase speed  this sa2pling and  conversion is done at a rate of up to 200 million samples per second  200 MS s    This high sampling rate allows you to observe fast signal variations     Digital 
202. nting EVENTS  Available in the digital mode only  The  use of EVENTS delay causes the acquisition to start after a user selectable  number of events has been counted     Delay  A full description is given under Delayed Timebase     Trigger position   Pretrigger is used to observe a portion of the signal that occurs before the trigger  point  This trigger point is then indicated in the screen by a marker  s   The  maximum pretrigger is the acquisition length  In this case  the trigger point marker  is positioned on the right side of the screen and the entire record contains  pretrigger information  The pretrigger view is indicated  in divisions  in the bottom  of the screen    Posttrigger is used for time delay and is expressed in seconds or fractions  thereof  The maximum delay depends on the setting of the main timebase   Maximum trigger delay is 100 divisions  the resolution is 1 50 of the timebase  TIME DIV setting     Event delay   Event delay postpones triggering until a specific number of trigger events have  occurred  The trigger event is defined in the EVENT DELAY sub menu of the TB  MODE  Its definition consists of a signal source and a trigger level  event level      FUNCTION REFERENCE 5 25       Every time the signal crosses this level the event counter is incremented by one   When the event counter reaches the selected delay value  the scope triggers and  a new signal acquisition is started  The number of events to be counted before the  acquisition starts is sele
203. nu   Pressing the    autoscale    softkey automatically selects the settings for the trace to  fit in the screen     5 50 FUNCTION REFERENCE       Newly acquired traces or previously stored traces can be used as source for this  process and can be selected with the A control  The resulting trace is  automatically written in a register memory  m1 for math1 or m2 for math2      To see the result more clearly  use the    DISPLAY SOURCE on off    softkey to turn  off the two source traces     One example of using the MULTIPLY function is the measurement of dissipated  power  by taking the voltage across a device and multiplying it by the current  through the same device     Mathematical description   Result   S1 x S2    Key sequence     add   substract   multiply  MATH         1 2  filter          Control to select the MULTIPLY process        Control to select the first source trace     Control to select the second source trace              0    on off Toggle softkey to switch the MULTIPLY function on     SCALE TRACK    Control to adjust the scale factor     Control to adjust the offset factor     00    auto  scale       Toggle softkey to select the autoscaling           DISPLAY  SOURCE    yes      Toggle softkey to switch the source traces on and off        5  7271  9303    Remote commands     CPL  QW  Command to query a waveform   Refer to Chapter 6 for full details     FUNCTION REFERENCE 5 51       PEAK DETECTION    Description   This function is available only in the digital 
204. number  8 625E 6 Trigger delay   Y resolution number 78 13E 3   X_resolution number 1E 6   Y_range number 65535   date string  F 1994 01 25 1    time string 17 11 25 40           option  17 11 25  dT corr number 375E 3        dT corr   X resolution  min max number 0 1   min max trace    mult_shot_tot number 32  mult_shot_nr number 5  reserved string 0  reserved string 0    1  When    A option is added  F   date format   E   European  A   American  J   Japanese   No option then E format     The acquisition length  512  8192  16384 or 32768   Extended Memory  8192 and higher     2 bytes  Most Significant Byte   Least Significant Byte   representing  the 16 bit sample value  bit 16    32768  bit 15   16384       bit 1   1    Range   32k  down       32k  up     Example  1000000011000111   value  32569    byte 1    byte 2      byte 1   128  value  32768   byte 2   128 64 4 2 1   199  value  199        checksum 1 byte checksum over all sample bytes                         THE CPL PROTOCOL 6 37  Example    100 PRINT 7  QW   74   Query waveform  m7 4   110 GOSUB 1000   Sync on acknowledge  120 INPUT     NM    Trace name   130 INPUT  YUNITS    Y unit   140 INPUT  XUNITS   X unit   150 INPUT   YZERO   Y zero   160 INPUT   XZERO     X_zero   170 INPUT   YRESOL   Y resolution   180 INPUT   XRESOL   X resolution   190 INPUT   YRANGE   Y range   200 INPUT        Date   210 INPUT            Time   220 INPUT  DTC   Delta T correction   230 INPUT   MINMAX   Min Max trace   240 INPUT  RES1S   
205. o illuminate the graticule of the Mee    screen          The trace is adjusted      parallel with  the horizontal graticule lines by the  screwdriver controlled TRACE   ROTATION control                                                                3 4 GETTING STARTED       3 4 AUTO SETUP    The best way to start each measurement is by using the AUTOSET key  This  automatically finds and scales all relevant parameters on all channels        AUTO SET                                               boo            o                         o  AOoso0o0s8o008oO0o0                                   9 9 9   m                                                                   000000            00900                                     576659  9303    Figure 8 8 Measuring setup    Step 1 Connect the probe as shown in figure 3 3     NOTE    AUTOSET is programmable  Because you have set the instrument in the   standard setup  before  see Section 3 2   all programmable features are  setto a predefined condition and the instrument is set in the analog mode   Programming of AUTOSET is explained in Chapters 4 and 5     Step 2 Press the AUTOSET key   The scope flashes the message  AUTO SETTING     on the screen  In  a few seconds the front panel settings are adjusted for an optimized  display of the applied signal in the analog mode     Step3        calibration signal is clearly displayed   The parameters of the channel and the timebase settings are  displayed at the bottom of the screen       
206. o the highest quality standards to give you many years of trouble  free and accurate measurements     The powerful measuring functions listed below have been combined with an easy  and logical operation to let you use the full power of this instrument each and  every day     If you have any comments on how this product could be improved  please contact  your local FLUKE organization  FLUKE addresses are listed in the back of the  REFERENCE MANUAL     The REFERENCE MANUAL also contains       CHARACTERISTICS AND SPECIFICATIONS     PRINCIPLES OF OPERATION     BRIEF CHECKING PROCEDURE     PERFORMANCE TEST PROCEDURES     PREVENTIVE MAINTENANCE PROCEDURES    IV       MAIN FEATURES    There are five models in this family of FLUKE oscilloscopes  Each of these  models is a combination of an analog real time oscilloscope and a fully featured  digital storage oscilloscope  By pressing a single key  you can switch the  instrument from the analog mode to the digital mode and back  This allows each  of the units to be used in an optimum operating mode for all kinds of signal  conditions  Complex data streams  modulated waveforms  and video signals can  often best be seen in the analog mode of operation  The digital mode of operation  is more suited for single events  signals with low repetition frequencies  and when  automatic measurements need to be performed    In this family there is a choice of five models  Two models have a bandwidth of  200 MHz  two have a bandwidth of 100 MHz and on
207. o the standard setting       Simultaneously press the STATUS and TEXT OFF keys     Connect the Probe Adjust signal to channel 1     The Probe Adjust signal  now supplied to the input  is a square wave with a lower  level of OV and a top level of 600 mV       Press AUTOSET     HOW TO USE THE INSTRUMENT 4 69       UTILITY PRINT SETUP    The oscilloscope must first be set to the correct interface parameters  Your  instrument is always equipped with an RS 232 interface as standard  The  following procedure describes how to set up the oscilloscope to use a printer  through the RS 232 interface       Press the ANALOG key to select the digital mode     Press the UTILITY key      Press the softkey    PRINT amp PLOT amp CLK         Select    print    with the first softkey     Using the TRACK control  you can choose from the following models or  emulations  HP 2225  FX 80  LQ 1500  HPLASER and HP540 Deskjet       Turn TRACK to select the printer you wish to use     Press the softkey next to  RS 232 IEEE  to select RS 232     The RS 232 settings of the oscilloscope and the printer must match  To set the  RS 232 parameters do the following       Press RETURN to go to the UTILITY main menu     Press the    RS 232 SETUP    softkey     If the IEEE option is present  the RS 232 SETUP softkey is reached via the  REMOTE SETUP softkey       Set the RS 232 parameters  baud rate  data bits  parity  connector type   and XON XOFF      Press the RETURN softkey     If you don   t know what to choose
208. o to Remote   ID IDentification   LL Local Lockout   PC Program Communication  PS Program Setup   PT Program text   PW Program Wavefrom  QS Query Setup   QM Query measurement  QP Query Print   QT Query text   QW Query Wavefrom   RI Reset Instrument   RS Recall Setup  internal   SS Save Setup  internal   ST STatus query    TA Trigger Acquisition    6 6 THE CPL PROTOCOL       6 5 COMMAND REFERENCE    In this section all commands of the CPL protocol available in the oscilloscope are  described in alphabetical order     All command descriptions have the same layout   NAME NM    Purpose   Explains the command  its parameters and limitations   Command     Shows the syntax for the programming command  The parameters are separated  by one or more PROGRAM DATA SEPARATORS  lt pds gt   Commands are  terminated by a Carriage Return  CR      Response     Shows the format of the response coming from the oscilloscope  Responses are  terminated by a Carriage Return  CR      The oscilloscope will  lt acknowledge gt  after the receipt of each programmed  command  This acknowledgment indicates the status of the oscilloscope after  command execution  For more information refer to section 6 6  ACKNOWLEDGE      To obtain a more detailed status description  the status can be fetched with the  ST command     Example     Example lines of programming code are used to demonstrate the function of the  CPL commands  The examples as shown can be embedded in the Program  Frame mentioned in section 6 2     
209. odels offer the selection of an input impedance of 50Q on all input  channels  The EXTTRIG input has no 500 possibility  The impedance selection   is a subfunction in VERT MENU    The 50Q input impedance is used to obtain a correct impedance to match signal  sources of the same impedance  For interconnection  a coaxial 50Q cable must  be used  The 50 position is indicated on the display with the Lz sign  Low Z     Low Impedance        Disconnect all signals from the inputs      Using a 50Q coaxial cable  connect a 4 Vpp  2 kHz signal from a  generator with 50   output impedance to channel 1      Press the AUTOSET key      Press VERT MENU      Press the 500 CH1 softkey                                500 CH                 0                                                                                                  575968   9312    Observe that the displayed signal amplitude changes to half of the original  amplitude  This is caused by the change of the input impedance to 50Q  which  results in a voltage division between the 500 output of the generator and the 500  input of the oscilloscope       Remove the input signal from CH1     VERTICAL PROBE RANGE INDICATOR      Connect the probe again     Since the 10 1 probe delivered with the instrument is a high  impedance probe  the 50Q termination is automatically turned off     At the same time  the sensitivity readout is adjusted automatically  detector sw  when a 10 1 or 100 1 probe is used  This way you don t have to  multi
210. oggle key for vertical input coupling of the input  channels    AC DC   Toggle key for vertical input coupling of the External  Trigger input     INPUT IMPEDANCE  200 MHZ MODELS ONLY     Description     For high frequency measurements the input impedance can be switched from  1 MO to 500  Input impedance is automatically adapted to the type of probe   Manual switching is done with the toggle key 500 in the VERT MENU menu   The 500 input is not available for the EXT TRIG     The input is protected in the 1 MQ as well as the 500 setting     1       protected for signals up to 400V     500  protected for peak voltages up to 50V or up to 5V rms     For details  refer to the Reference Manual     Key sequence     500 CH1    VERT MENU 9n      TT  Toggle key to switch 500  input impedance of CH1     500 CH    on off    Toggle key to switch 500  input impedance of CH      577917    5 42 FUNCTION REFERENCE       LOGIC TRIGGER  4 CHANNELS MODELS ONLY     Description     Logic triggering enables triggering on a combination of the four input signals   Each input is compared with a trigger level and is recognized as being either  HIGHer or LOWer than the trigger level  The four input signals together can be  regarded to be a 4 bit digital    word     If this    word    matches a certain pattern  triggering occurs  The 4 bit trigger pattern can be edited with the TRIG buttons in  each of the channel sections and is displayed in the lower right hand corner of the  screen with each input bein
211. one of the two     Control to shift the A cursor     CPL  QM  command to query measured values     Refer to Chapter 6 for full details     5 22 FUNCTION REFERENCE       CURSORS READOUT    Description     The cursors offer a wide variety of voltage and time readouts  For comparison of  signal details the ratio mode is very suitable  When in analog mode time or volt  cursors have been selected  only the relevant readouts are displayed in the  READOUT menu  In digital mode all readouts are displayed at the same time     The various readout selections for time  volt and    both    are reached via the key  sequence CURSORS  gt  gt  READOUT        CURSORS READOUT TIME    Three time interval readouts can be selected            Gives the time between the cursors      1 AT  This results is a frequency readout  The readout is correct when the  distance between the cursors equals one signal period      AT ratio  The readout is a percentage  that can be reset by using the        100      softkey  This mode can be  used for time comparisons  The figure  shows an example of a duty cycle  measurement  First the cursors are  positioned at the period and the readout           Be a      Then the pulse width is ees POSITION CURSORS     Phase  ph   The readout is a number of            PRESS AT 100   degrees that can be reset using the        360     softkey      T trg  time to trigger   The readout gives  the time between the trigger point and  each time cursors                                 
212. ontrol is used to select the waveform    The result of the measurement is displayed in the top left corner of the screen     PERIOD  PULSE             The following measurements are  possible    Tee MESIAL  Refer to section MEASURE MENU  for details of signal parameters       4      DISTAL                   m PROXIMAL                                                             stera    RISE FALL      freq  FREQUENCY    Measures the reciprocal of the PERIOD of the signal  Frequency is calculated  between the first and third mesial of the signal  At least one complete signal  period must be acquired to get a valid measurement     1  time of 3rd mesial   time 1st mesial       freq      5 76 FUNCTION REFERENCE         period  PERIOD    Measures the time between the first and third mesial of the signal  At least one  complete signal period must be acquired to get a valid measurement     period   time of 3rd mesial   time of 1st mesial      pulse  PULSE WIDTH    Measures the minimum time of the positive part and negative part of the first  period of the signal  At least one complete signal period must be acquired to  get a valid measurement     pulse   min   time 2nd mesial  time 1st mesial  pos part of signal period      time 2nd mes  time 1st mes  neg part of signal period     Note that the part with the shortest duty cycle is always defined as    pulse          rise  RISE TIME    Measures the rise time of a signal  It is calculated between the proximal and  distal of the first r
213. ope   1 25       CH1        100 PRINT  1  SS 7    Save Setup command   110 GOSUB 1000   Sync on acknowledge   120 PRINT  1  PT S 7    Program setup Text command  130 GOSUB 1000   Sync on acknowledge   140 PRINT  1  1 25 k  CHRS 25     CH1        Send setup text  150 GOSUB 1000   Sync on acknowledge    6 22 THE CPL PROTOCOL       PROGRAM WAVEFORM PW    Purpose     To send a waveform to the oscilloscope  This function is referred to as to Program  a complete Waveform in the oscilloscope     A waveform is sent  programmed  in two command sequences  The first  sequence selects the waveform register number  programs the waveform  administration data  and programs the number of samples  The second sequence  programs the samples  including a checksum  Each sequence is synchronized by  an acknowledge response     You can use the PW command to send the data received with the QW command  back to the oscilloscope     Note         number of data bits of the RS 232 interface must be 8  not 7    otherwise  an execution error will occur     Command 1                                                                     PW  pds wave nr pds admin pds count CR       Response 1           acknowledge CR                   Command 2                           sample checksum                      Response 2           acknowledge CR                   THE CPL PROTOCOL       wave      The oscilloscope waveform destination   011   084 for m1 1   m8 4  091   504 for m9 1   m50 4  Extended Memory only   09e
214. or 4  This continues until the acquisition memory for  each channel is filled  This ALT CHOP function is valid only for the real time  sampling timebase speeds     NOTE  For further details on three or four channel operation  also refer to the  ACQUISITION LENGTH and PEAK DETECTION functions     5 8 FUNCTION REFERENCE       The alternate and chopped modes are shown in the figures below     ALTERNATE MODE                                                               Key sequence                 TB MODE      chop  Toggle softkey to select ALTernate or CHOPped mode     576845  9303    ANALOG MODE    Description     You can use the yellow ANALOG key to switch from the analog mode to the digital  mode and back at any time  The signal acquisition and display functions of both  operating modes are very similar  Depending on the nature of the signal used or  the choice of the measurement  you may prefer the analog mode     This is especially true for the following types of signals     amplitude modulated signals     frequency modulated signals     fastsweeps     long serial data streams   e most video signals     observing signal changes  e g   adjustments     The same waveform is usually displayed in the same manner  whether you select  the analog or the digital mode  However  some functions are not implemented in  both modes  For instance  the digital mode permits the viewing of pre trigger  information  The digital mode also offers auto range for main time base and  vertical chann
215. ost Digital Storage Oscilloscopes  limit    their capabilities to the display of amplitude  in volts   versus time   In addition to those traditional oscilloscope functions  this range offers capabilities  to mathematically change the contents of each memory location  To do so  the  functions under MATHEMATICS    allow you to calculate new waveforms from  existing other traces serving as input data   If your oscilloscope is equipped with the MATH   option  more functions         available than the ones described in this section  Additional information on the  additional functions is described in a separate manual     The section below describes the mathematical functions implemented in the  standard oscilloscopes versions     Two mathematical functions MATH 1 and MATH 2 can be used as two  independent processes  They can be used in series or in parallel    The result of MATH 1 is always stored in memory location  m1     The result of MATH 2 is always stored in memory location    m2       When used for mathematical functions  existing memory locations are  overwritten   Use the  Copy  function to save important traces in another memory  location  before using the mathematical functions      STANDARD SETTING    Start with the standard setting to ensure that you have the correct start condition       Simultaneously press the STATUS and TEXT OFF keys     Connect the Probe Adjust signal to channel 1     Press AUTOSET     The Probe Adjust signal now supplied to the input is a squar
216. over has been designed to snap  on to the front of the instrument     The front can be removed as follows      Fold the carrying handle down so that the oscilloscope occupies a sloping  position  refer to Chapter 2 4 for how to proceed       Pull the clamping lip at the top side of the cover slightly outwards           the cover off the instrument     2 4 INSTALLATION INSTRUCTIONS       2 4 HANDLE ADJUSTMENT AND OPERATING  POSITIONS OF THE INSTRUMENT    By pulling both handle ends outwards away from the instrument  the handle can  be rotated to allow the following instrument positions       vertical position on its rear feet     horizontal position on its bottom feet     in three sloping positions on its handle     The characteristics mentioned in the REFERENCE MANUAL are guaranteed for  the specified positions or when the handle is folded down     CAUTION  To avoid overheating  ensure that the ventilation holes in the  covers are free of obstruction  Do not position the instrument  in direct sunlight or on any surface that produces or radiates  heat     In the rear panel of the instrument there is storage space for the mains cable   There is also a clamping device to fix the end of the mains cable to the rear panel   The mains plug then fits in the area where the RS232 connector is present  In this  way the instrument can also stand on its rear feet     1  Let  n     Figure 2 2 Instrument positions                               2 5 IEEE 488 2 IEC 625 BUS INTERFACE OPTION    
217. plays the two Probe Adjust signals in the    center of the screen            Press the TEXT OFF key  to turn off the bottom    text     Switch on the channels 3 and 4  when present                 screen displays two      four traces now                                                             Press the DISPLAY key     Press the key next to WINDOWS    to    on            Bt             The screen is automatically divided into two windows of  four or four of two divisions each  Each window displays                           4  one trace  Observe that the amplitude of the signal has         been adjusted to the actual window space                                           Turn the Y POS control of channel 2 and observe that you can shift the  trace within  but not outside the window     Use the window function for logic applications where amplitude accuracy is less  important than clear timing comparisons       Switch the WINDOWS off again     Disconnect the probe from channel 2 before continuing to the next  section     HOW TO USE THE INSTRUMENT 4 63       4 14 DELAYED TIMEBASE       The Delayed Timebase  DEL DTB  has two basic  functions     DELAYED TIME BASE DELAY  DTB s TIME DIV ns    OOO      To magnify and display any detail of the signal displayed with the main  timebase     To permit more accurate timing measurements              ST6719  9303    STANDARD SETTING    To start from a predefined state  you must recall the standard front setting          Simultaneously pr
218. ply the displayed amplitude by 10 or 100 when you use a Fluke probe with  range indication        4 22 HOW TO USE THE INSTRUMENT       4 5 HORIZONTAL DEFLECTION AND TRIGGERING    Before starting with the horizontal deflection functions  you must set the  instrument to a predefined state to create a correct start situation     STANDARD SETTING    Simultaneously press the STATUS and TEXT OFF keys     Connect the Probe Adjust signal to channel 1     The Probe Adjust signal  now supplied to the input  is a square wave with a lower  level of OV and a top level of 600 mV       Press AUTOSET     TIMEBASE TIME DIV    TIME DIV                     4 periods cycles of the square wave are displayed on the screen               576639                  Press the    s     left  key of the MainTB TIME DIV key pair    few times     The more you press the left     s     key  the slower the timebase will run  This results  in the number of displayed periods cycles of the input signal to increase       Press the    ns     right  key of the MainTB TIME DIV key pair a few times     The number of displayed periods cycles decreases as the timebase speed  increases   Observe that the timebase speeds are adjustable in steps following a 1 2 5    sequence   TIME DIV  Sv VAR v ns                  amp             MITB 80045 MTB 100us                                                                                                                                                     576709      Set the timebas
219. pplications  this menu is not  needed     3 20 GETTING STARTED       3 14 MORE SIGNAL DETAIL WITH THE DELAYED  TIMEBASE    When you need to study a part of a signal in more detail  a second  delayed   timebase is available  This timebase has its own timebase settings and trigger  level adjustment  Additional selections are made in the DELAYED TIMEBASE  menu     DISPLAY       MAGNIFY                         TRIGGER  DELAYED TIMEBASE pei      POSITION  DTB s TIME DIV ns TB MODE    CIO OO                                                                                              gajo osoo                               Figure 8 11 Delayed timebase setup  Step 1 Press the STATUS key and TEXT OFF key at the same time for    STANDARD SETUP   Then shift to trace to the upper half of the screen as indicated                                                                                                                                                                    CH1 50mVi MTB 1 00  5 chi    516690       GETTING STARTED 3 21       Step 2    Press the DTB key  The DELAYED TIME BASE menu is now displayed  on screen  Turn the delayed time base on with the first softkey        DELAYED  TIME BASE                  H on off  MAIN TB    on off  starts     trig d                                                                                                           T                                               TB1 00m  chi  CH1 50mV  DTB 100us 4 882ms                         516689   
220. preset probe  attenuation factor  for probes  without indication ring  after  AUTOSET     5 14 FUNCTION REFERENCE       AVERAGE    Description   Valid in digital mode only     Averaging is a process to reduce random noise without losing bandwidth   Averaging can only be used for repetitive signals     Every sample point is calculated after every subsequent acquisition as follows     new   previous                                 In this formula  previous  is the result of the average function after the previous  acquisition  on the same sample position as the measured one   C  is the  average factor  Noise is reduced by a factor which is equal to the square root of  C      Average values for      can be selected from 2 to 4096 in a binary sequence  To  do so  use the TRACK control in the ACQUIRE menu     Averaging also results in a higher vertical resolution resulting in higher  measurement accuracy     Averaging can be quickly switched on and off with the direct access key   AVERAGE  on the front panel  or via the ACQUIRE menu  Once activated   average is a continuous process until it is switched off  However  whenever the  display is changed  e g  POS  AMPL DIV  the average process will restart     AVERAGE is mutually exclusive with ENVELOPE  If the ENVELOPE mode is  selected  AVERAGE is automatically switched off     Key sequence     AVERAGE  Key to switch the AVERAGE function on or off     acquirE TRACK    Control to select the AVERAGE factor     576555  9303    FUNCTION 
221. quipment  General  information for SCPI and the instructions are located in a separate programming  manual  The 24 pole connector and its connections are shown in the figure below     For correct functioning in a IEEE 488 environment  the oscilloscope   s device  address must be selected  This is done in the UTILITY REMOTE CONTRL menu  with the TRACK control  This menu is reached via the key sequence  UTILITY  gt  gt   REMOTE CONTRL     To change the IEEE settings  first select IEEE with the   RS 232 IEEE  softkey     SHIELD SRQ NDAC DAV DIO4 DIO2    ATN  IFC   FD  EO1 DIOS DIO1                   Operation of front key STATUS LOCAL    12   passes the control of the oscilloscope from            the interface  remote  to the front keys     local                                 0107                              IGND  9 7    LOGIC GND GND GND DIO8 DIO6  GND 10 8 6              Key sequence     REMOTE TRACK    UTILITY SETUP  7   20  Control to select the IEEE device address     ST6075  9303    STATUS     Key to switch from remote to local    LOCAL    FUNCTION REFERENCE 5   59       REMOTE CONTROL RS 232    Description     The oscilloscope is equipped with an RS 232 Interface as standard  This can be  used for remote control or for setting the readout using an external controller or PC     The language used is called CPL  Compact Programming Language  and is  described in Chapter 6  CPL is a small set of very powerful commands for full  remote control of all oscilloscope functions 
222. r low repetition rate signal at  high timebase speed  This can be done by selecting the    real time sampling only     mode in the Main Time Base menu       this mode the scope never switches to  random sampling     In order to capture a longer time    window    than would be accessible with the given  time base speed  the Main Time Base mode menu allows you to select a longer  acquisition length     By means of horizontal magnification  2x  4x  8x  up to 32x  you can expand the  signal  giving you a higher effective time base speed  In the DOTS mode  single  samples become visible  and SINE and LINEAR interpolation can be used to  reconstruct a true to life waveform  In four channel single shot mode with its  maximum sample rate of 5 MS s  a timebase speed close to 1 us div is no  problem  The 8x magnified waveform is then reconstructed out of 62 samples     FUNCTION INDEX l 1       FUNCTION INDEX  see Chapter 5     ACQUISITION LENGTH   ADD INVERT SUBTRACT   ADD  MATHEMATICS   ALT CHOP   ANALOG MODE   AUTO RANGE   AUTOSET   AUTOSET RECALL SEQUENCE  AUTOSET USERPROG  AVERAGE    BANDWITH LIMITER    CALIBRATION AUTOCAL  CHANNEL TRACE SELECTION  CONFIDENCE CHECK  CURSORS   CURSORS READOUT    DELAY   DELAY MEASUREMENT  DEL   D TB   DIGITAL MODE  DISPLAY MENU    ENVELOPE  EXTERNAL TRIGGER    FILTER   GLITCH TRIGGER  HOLD OFF   INPUT ATTENUATOR  INPUT COUPLING  INPUT IMPEDANCE    LOGIC TRIGGER    1 2 FUNCTION INDEX       MAGNIFY HORIZONTAL  MAGNIFY VERTICAL  MAIN TB TIME DIV  MATHE
223. r phase to 360       Toggle function softkey to link the TRACK control  to both cursors or to the track cursor only      I Toggle function softkey to switch between time or    Ly  volt cursor for 100  reset and positioning     5 24 FUNCTION REFERENCE       DELAY    Description     Delay is the term that is used to define the time difference between the trigger  point of an acquisition and the starting point of the resulting trace  In an analog  oscilloscope  the trigger point is at the beginning of a trace  and the delay is said  to be zero  The use of a Delayed Timebase introduces a delay between the trigger  point of the Main Timebase and the start of the trace which is displayed with the  Delayed Timebase     Three types of Delay     This family of oscilloscopes has three types of adjustable delay     1  Delay controlled by the DELAY control on the front panel  This is the amount of  time delay between the start of the Main Timebase and the Delayed Timebase   This delay applies to the Analog mode  as well as to the Digital mode    2  A delay between the trigger point and a signal acquisition  This delay is used  to    position    the acquired input signal with respect to the trigger point  This type  of delay can be    positive     as well as    negative     The result is the acquisition of  posttrigger or pretrigger waveform data  respectively    This delay is available in the digital mode only and is controlled with the  TRIGGER POSITION control    3  Delay caused by cou
224. rameters are separated from one another by a PROGRAM DATA  SEPARATOR  lt pds gt   At the end of the complete command  i e   header         parameters  comes the PROGRAM MESSAGE TERMINATOR CR  After the CR  is recognized by the oscilloscope  the command will be executed  Then an    acknowledge   and CR is sent to signal the end of the command processing     Notes    Literal characters are placed between double quotes  e g   AS      Literal characters may be specified in upper and lower case     6 2 THE CPL PROTOCOL       There are several IMPLICIT QUERY commands  which means that the  oscilloscope will send data back  i e   respond  to the computer after receiving  and executing the command     Acknowledge   The  lt acknowledge gt  is an automatic response from the oscilloscope to let the  computer know that the received command has been executed  The    acknowledge   also contains information about how the command was  executed     An   acknowledge   is always followed by a CR  For more information  see section  6 6 ACKNOWLEDGE     Responses   The format of the response data depends on the command which invoked the  response  When several values or strings are returned they are always separated  with a RESPONSE DATA SEPARATOR which is a comma        ASCII 44   To  signal the end of the response a RESPONSE MESSAGE TERMINATOR   CR  ASCII 13  is sent     Data Separators   Data Separators are used between parameters sent to the oscilloscope and  between values and strings receive
225. rately over the screen by means of the A control       Turn the POSition control       REGISTER                off    softkey to on   Q Iorin      Turn the A control        trace  A register             The live trace shifts  sm O  Cj    Turn the TRACK ono   CO   control to select    m2     piseuay           Press the DISPLAY      POS 950                                                                  Observe that the second stored trace shifts     HOW TO USE THE INSTRUMENT 4 43       TRACE STORAGE SAVING MULTIPLE TRACES    Each memory location can store two channels       Connect the probe to channel 1      Press AUTOSET      Switch on channel 2      Turn the channel 2 POS control counterclockwise     The upper trace displays the signal of channel 1  and the lower line represents  channel 2       Press the SAVE key     Turn the TRACK control to select memory location    m1        Press the    save    softkey to store two traces in  m1      Observe that m1 was already filled  so the confirmation submenu is displayed     Press the softkey next to    yes        This overwrites memory  m1  with the new traces     TRACE STORAGE RECALLING MULTIPLE TRACES      Press the RECALL menu key   The memory location is already preset to the last saved  m1        Press the    DISPLAY on off    softkey so that    m1    is displayed     Turn the A control counterclockwise     Observe that both traces are displayed and shifted to the lower screen     Press the softkey next to          register
226. red values  automatically changed at the same time  The read  outs are displayed in the  upper section of the screen     STANDARD SETTING    Start with the standard setting  This ensures a correct start       Simultaneously press the STATUS and TEXT OFF keys     Connect the Probe Adjust signal to channel 1     The Probe Adjust signal now supplied to the input is a square wave with a lower  level of OV and a top level of 600 mV       Press AUTOSET     4 50 HOW TO USE THE INSTRUMENT       MEASURE MEAS1 PKPK  MUST  Press the ANALOG key to select the digital mode         Press the MEASURE key     The displayed menu gives access to the two measurements MEAS 1 and  MEAS 2  Each measurement can be independently turned on and off  In this  menu  you can select the measurement in MEAS 1 and MEAS 2 function     Observe that the measurements are linked to the active channel  The two default  measurements are  pkpk  and  freq        Press the second softkey to turn MEAS1 to    on        The screen automatically displays the result of the peak to peak measurement   chi                mV        Press the softkey next to    MEAS1        The menu displays the  SELECT MEAS 1  submenu  You can choose from three  main measurements  volt  time  or delay  The lower part of the menu optimizes to  the selected main measurement    The actual type of measurement is selected by using the TRACK    control  The  following voltage measurements are available   dc  root  mean square  minimum  peak  maximum
227. return to edge        Note  In the Digital mode  an additional selection is available  This is called   logic  or  glitch   A full description of the Logic and Glitch triggering  functions can be found in Chapter 5     HOW TO USE THE INSTRUMENT 4 35       TRIGGER LEVEL      Press the  level pp  softkey to turn it off     TRIGGER The automatic level detection circuitry is turned off  The trigger level is  LEVEL      longer clamped within the peak peak range of the signal  You must     adjust the proper trigger level  While you turn the TRIGGER LEVEL  control  the actual trigger level is displayed in  m V     An indicator  T   is displayed next to the channel identifier when the combination  of trigger coupling and channel input coupling is useful  both ac coupled or both  dc coupled      When the  T   is within the signal range  a stable display of the signal is obtained   When proper triggering takes place  the ARM D LED is off     Use the TRIGGER LEVEL control to move the indicator  T   vertically     FHEEHHEE           LEVEL      When the trigger level indicator  T   is outside the signal range  triggering is lost   Because of the nature of the  Probe Adjust  input signal  it appears that only two  lines are drawn  The timebase is not properly triggered  as is indicated by the  ARM D LED being turned on                                                                                                                                576738    Note       T  is an indication for r
228. s as  indicated  Note that all registers will be cleared when the length is changed  The  user must confirm the change of the acquisition length     The X POS control is used to move the trace in horizontal direction and to display  any desired part  A bar graph shows which part of the trace is displayed on the  screen  By using the MAGNIFY key pair  you can expand the trace up to 32 times  to look at signal details  The MAGNIFY keys can also be used to compress the  signal to fit the display width and look at the total waveform  This is realized  without the loss of vertical information     The 8K acquisition memory is shared between all channels  In the  2ch O4K pts     mode the entire memory is used for the two channels  so triggerview of the  external trigger channel is not possible     In the standard mode the trace length is 512 samples and all channels can be  acquired simultaneously  In this mode  more traces can be stored in the reference  memory  and the screen update rate increases  This mode also supports delayed  timebase operation  This way  acquisition can consist of up to eight traces  four Main  TB and four Delayed TB traces   for the 2 channel models  six traces  two Main TB  traces  two Delayed Time Base traces and two Trigger View  MTB DTB  traces      The following table shows the selection possibilities for a standard oscilloscope  with 8K acquisition memory              Acquisition length Reference memory Traces     8K 2 registers of 1 trace each 3  3  
229. s in the same register memory to avoid any possible  error of making delay measurements between traces that were not part of the  same acquisition     For each source there is a softkey to select the slope of the edge  leading edge  or trailing edge      The delay is displayed in the top left corner of the screen     del         5     Delay    leading trailing edge S1   leading trailing edge S2   where   1 and S2 are source signals S1 and S2    Refer to section MEASURE MENU for details of signal parameters                                                        ST6658    DELAY      FUNCTION REFERENCE 5 27       Key sequence     MEASURE         1 2  delay Toggle softkey to select the delay measurement in  4   menu DELAY 1 or DELAY 2     J       TRACK  Control to select the reference waveform     Toggle softkey to select the slope of the reference  waveform  Referred to  zero  delay     Control to select the second waveform     Toggle softkey to select the slope of the second  waveform        JO          on off       Toggle softkey to switch the delay measurement             on and off     9303    Remote commands   CPL  QM  Command to query a measurement   Refer to Chapter 6 for full details     DELAYED TIMEBASE  DEL D TB     Description     The Delayed Timebase  DEL D TB  is used to examine a signal detail of interest   The detail to be examined is indicated as an intensified part of the MAIN TB trace  and is displayed on the full screen width using the DEL D TB time scale     
230. s the screen     When you add signals with different amplitude settings  the smaller signal is  automatically adapted to the larger signal     Either newly acquired traces or previously stored traces can be used as the  sources for this process and each can be selected with the control  The resulting  trace is automatically written in a register memory  m1 for math1 or m2 for math2      To see the result more clearly  you can use the DISPLAY SOURCE on off softkey  to turn off the two source traces     Mathematical description     Result   S1   S2    5 6    FUNCTION REFERENCE       Key sequence     add  substract  multiply    MATH _ MATH1 2  filter    EE          00 0    on off    SCALE TRACK       6 6    auto  scale             DISPLAY  SOURCE  yes no       T7271  9303    Remote commands   CPL  QW  Command to query a waveform    Refer to Chapter 6 for full details    Control to select the ADD process     Control to select the first source trace     Control to select the second source trace     Toggle softkey to switch the ADD  function on     Control to adjust the scale factor     Control to adjust the offset factor     Toggle softkey to select the autoscaling     Toggle softkey to switch the source  traces on and off     FUNCTION REFERENCE 5 7       ALT CHOP    Description     In the analog mode  when two or more channels are selected  the oscilloscope  displays multiple channels in a    time shared    mode  This implies that a fast  electronic switch connects each input sig
231. s voltage range and fuses    Before inserting the mains plug into the mains socket  make sure that the  instrument is suitable for the local mains voltage     NOTE  When the mains plug has to be adapted to the local situation  such  adaption should be done by a qualified technician only     WARNING  The instrument shall be disconnected from all voltage  sources when a fuse is to be renewed     The oscilloscope has a tapless switched mode power supply that covers most  nominal voltage ranges in use  ac voltages from 100     240 V  r m s    This  obviates the need to adapt to the local mains  line  voltage  The nominal mains   line  frequency range is 50 Hz     400 Hz     Line fuse rating  3 15 AT delayed action  250 V  for ordering code see   INITIAL INSPECTION      2 2 INSTALLATION INSTRUCTIONS       The mains  line  fuseholder is located on the rear panel in the mains  line  input  socket  When the mains  line  fuse needs replacing  proceed as follows      disconnect the oscilloscope from the mains  line       remove the cover of the fuseholder by means of a small screwdriver      fita new fuse of the correct rating and refit the cover of the fuseholder     WARNING  Make sure that only fuses with the required rated current and  of the specified type are used for replacement  The use of      makeshift fuses and the short circuiting of fuse holders                 prohibited   REAR VIEW  IEEE 488 2  SOCKET RS 232 MAINS FUSE BATTERY LOCATION OF   OPTIONAL  SOCKET INPUT HOLDER C
232. second probe     HOW TO USE THE INSTRUMENT 4 9       4 3 ANALOG AND DIGITAL MODES    ANALOG 1 245       3   2    This instrument is a combination of an analog real time oscilloscope and    a digital storage oscilloscope  which offers a variety of additional features   The combination of analog and digital modes in one instrument gives  you the advantages of both modes     In the ANALOG MODE the signal is directly written on the screen  The result is  the  traditional  real time signal representation  Because of the high update rate  and infinite resolution  this image gives signal details that are visible only on true  analog oscilloscopes     In the DIGITAL MODE the input signal is sampled  These samples are stored in  memory so that mathematics  calculated measurements  printing  and other  memory functions can be performed on the trace     You can use the yellow ANALOG push button to switch from the analog mode to  the digital mode and back at any time  The signal acquisition and display functions  of both operating modes are very similar  However  the nature of the signals you    are using may determine which operating mode you prefer to use  For more  information  refer to the following table        SIGNAL CRITERIA ANALOG MODE DIGITAL MODE  Repetitive signals of Usable Usable   30 Hz and higher   Repetitive signals Causes display Preferred    below 30 Hz    Single events    Repetitive signals that  are amplitude modulated    Repetitive signals that  are modulated in freq
233. shows the display window and cursor positions in relation to the total  acquisition length  The cursors are trace oriented rather than display oriented   The acquisition length can be selected via the TB MODE menu     The readout of the delta between the cursor lines is shown in the cursor display  area  upper part of the CRT viewing area   In analog mode the display can be in  voltage or time  and is in digital mode in voltage and time at the same time  In this  way cursors can be used for accurate on screen measurements without using the  graticule  Additional readout information can be selected under the READOUT  softkey  For details see the CURSOR READOUT function     The cursors are activated via the menus under the CURSORS menu key  The  structure of the menu is shown in Appendix B     Activating the cursors is made with the CURSORS on off    menu selection     5 20 FUNCTION REFERENCE       CURSORS TIME           CURSORS   The time cursors are used for time    measurements  The example shows the on of  required softkey settings for period  a d  measurements  The cursor positioning   with the TRACK and A controls is also BE  shown                    000000       READOUT                                                           CURSORS VOLT    The volt cursors are used for  CURSORS    voltage measurements  When       more than one channel is on  the      desired channel for voltage e     C3  readout must be selected with the Easy  NE     gt  v  C3   5     7               
234. sing the RUN STOP key   but also to store traces in memory for later use     ANALOG SAVE RECALL    Si Ss                                                                   o 00000000          ogoo RUN STOP                                     Os 00    eles  ele                ST6691  9312    Figure 3 12 Digital memory setup    Store traces on screen   Step 1 Press AUTOSET     Step 2 Make sure that the scope is in the digital mode  If not  press the  ANALOG key to enter the digital mode     Step3 With the RUN STOP key  new acquisitions are stopped and the display  is frozen  Removing the input signal or pressing a key has no effect on  the display  Stopping the acquisition is very useful to do measurements  on the signal or to make a hard copy     Step 4 Press the RUN STOP key to reactivate the acquisition   Store traces in memory     Step 5 For this step by step introduction you will first clear all memory  locations so that all unnecessary traces are removed     press the SAVE menu key     select CLEAR  amp  PROTECT     select CLEAR ALL     in the confirm menu  select YES   Sometimes there will be a second confirm menu  Select YES again to  clear protected traces as well   Press the TEXT OFF key to turn the menu off     GETTING STARTED 3 23                is how traces        stored      memory     Step 6    Step 7    Step 8    Step 9    Step 10    Step 11    Step 12    NOTE     Use the TRACK conirol to select an empty memory  location such as m1  m2  or m3  Empty locations ar
235. sor measurements  indicated by             Time cursor measurements  indicated by     for    measuring time or frequency       Amplitude and time measurements  indicated by  f  The top text line displays the result of the READ oun  measurements  AV or AT         Step 5 Press the second bluesoft key until      is highlighted   Step 6 The TRACK control moves both cursors at the same time  For    example  to measure the period time of the input signal  set the left   reference  cursor to a rising edge of the signal     Step7 The A control moves the right cursor only  Set this cursor the next rising  edge of the signal     3 18 GETTING STARTED       Step 8 The top text line now shows the pulse repetition time of the signal   e g       1  AT  500 us             11 AT  500us CURSORS             1     3  i on off   1 d                                     i 1  i OUT                                      t  CH1 200mV   MTB PoOus             516687       Step 9 Press the second blue softkey until         is highlighted  Now perform a  peak to peak measurement and check that the amplitude of the signal      AV     is 600 mV     NOTE    When you select     the fifth blue softkey is automatically activated so  that you can choose between using the controls for positioning the  vertical cursors            or the horizontal cursors         The  READOUT  submenu is explained in Chapter 4   Step 10 Select the vertical cursors again     Step 11 Now switch to the digital mode  Notice the changin
236. st have  expired  The Del d Time Base modes are selected in the DELAYED TIMEBASE  menu after switching the delayed timebase on     In the  STARTS  mode  the DEL D TB starts immediately after the DELAY time   This is explained under DEL D TB  When signal jitter is present  the effect will be  magnified by using the DEL D TB  The jitter is eliminated by changing from the     starts    to the triggered mode  The start of the DEL D TB sweep is then    postponed     to the first DEL D TB trigger after the delay time     The triggered mode is activated in the DELAYED TIMEBASE menu by selecting  the trigger source  ch1  ch2 and ext trig   For stable triggering  the DEL D TB  trigger level must be adjusted to a proper level  The trigger level for the Del d Time  Base is adjusted with the A control  The figure shows the difference between   starts  and  trig d      For more explanation of trigger functions  refer to TRIGGER MAIN TB  function   For details about trigger coupling  see the appropriate section     DELAY DTB START DIRECTLY AFTER DELAY TIME  STARTS     Q  ___   DELAY TIME          DEL D TB    starts    and             trig   d    modes  DTB START AFTER DELAY TIME UPON TRIGGER PULSE  CH1   CH4   DELAY WAITING FOR TRIGGER     Mu DELAY TIME              DTB    TRIGGER                   4214           DEL D TB trigger source and slope can be selected when     The menu is active     The    trig   d    mode is selected     5 82 FUNCTION REFERENCE       Source and slope are select
237. t              2060 IF  STATUS AND  amp H0040  0 THEN PRINT  WRONG   DATABITS                       EN PRINT  UNKNOWN HEADER           LH                E                               2070 IF  STATUS AND  amp H0200  0 THEN PRINT  CONFLICT SETTING   2080 IF  STATUS AND  amp H0400   gt  0 THEN PRINT  USER REQUEST    2090 IF ABS  STATUS AND  amp H8000     0 THEN PRINT  MORE STATUS   2100 RETURN                                              6 50    THE CPL PROTOCOL       6 8 SETUP    The Query Setup  QS  and Program Setup  PS  commands can be used together  in order to retrieve and restore a complete or partical setup of the oscilloscope     When a setup is requested from the oscilloscope  it will send its setup as a  sequence of strings  Each separate string describes a setup node  By adding a  parameter to the QS command a particular setup node can be queried     So the query program could be     PRINT  PRINT       PRINT     1   os    1   OS  2   1   OS  S    Complete setup   Only node 2  channel 2 settings    Only node S  where S must be one of the  values specified below        Node numbers  S  can have the following values        hex dec meaning   01 1 Channel 1 settings   02 2 Channel 2 settings   03 3 Not used   04 4 Ext trig settings   nodes above  attenuator  display on off  coupling DC AC  GND   Invert  509 1       continuous var  discrete  position control    OE 14 Probe Scale settings  scale value  scale unit         15 Common vertical settings   add 1 2  alt chop  ba
238. t  PKD  appears in the bottom of the screen to indicate that the function is  active  Observe that any glitch on the input signal becomes visible on the screen     Note  Regardless of the timebase speed selected  the scope is set to the  highest sample rate  so that glitches up to the sample frequency can be  caught  5 ns at 200 MS s        Press the          DET    to turn it off again     4 34 HOW TO USE THE INSTRUMENT       4 8 ADVANCED HORIZONTAL AND TRIGGER  FUNCTIONS    All basic timebase and trigger functions of the oscilloscope are accessible via  direct action front panel keys  More advanced functions are easily accessible via  the menus behind the menu initialization keys  keys with the function in blue text      STANDARD SETTING  Before continuing with the advanced functions  you must set the instrument to a  predefined state to ensure a correct start situation       Simultaneously press STATUS and TEXT OFF     Connect the Probe Adjust signal to channel 1     Press the AUTOSET key     TRIGGER MODE      Press the TRIGGER key     On the screen the TRIGGER MAIN TB    menu is displayed  The menu functions  can be selected by pressing the blue softkeys to the right of the screen  The top  softkey selects the trigger mode   edge  or  tv   The lower part of the menu is  optimized for the selected trigger modes       Press the first softkey to select  tv  trigger mode   Observe that the lower part of the menu is optimized for tv applications     Press the first softkey to 
239. t any nominal line voltage between 100 Vac and  240 Vac  with no switching and no fuse changes  After the instrument is turned on  by pressing the POWER ON OFF switch  an automatic power up test is started   For detailed information  refer to the CONFIDENCE CHECK    function     The oscilloscope starts up with its previous settings when backup batteries are  installed  In the absence of batteries  the standard default setting  std  becomes  active  Refer to STANDARD FRONT for details     Key sequence     POWER ON OFF    Cy Toggle key to switch the oscilloscope on off     PRINTING AND PLOTTING    Description     In the digital mode a hard copy of the information on the CRT can be made to a  printer or a plotter  The hard copy can consist of trace s   the trace settings  trace  identification  cursors  measurement results  and the screen graticule     The oscilloscope and printer plotter must be interconnected via a suitable cable  and must be set up to the correct interface parameters  The setup of the print or  plot action is made in the UTILITY      PRINT  amp  PLOT  amp  CLOCK menu  The setup  of the interface parameters is made in the UTILITY      REMOTE SETUP or the  UTILITY  gt  gt  RS232 SETUP menu  This is described under the REMOTE  CONTROL IEEE 488 2 and RS 232 functions  The settings of the printer or plotter  are described in the manual that comes with the printer or plotter     The front panel key HARDCOPY is used to start a print or plot  During the plot  action 
240. the Main Timebase trace                                                                                                                                                                                            o                   500            2615           500       BIB 10 mu  DELAYED TIMEBASE TIMEBASE      Press either key of the Del dTB TIME DIV keys     When you press the left key  the intensified part of this signal gets longer  More  periods of the signal are displayed on the Delayed Timebase  When you press the  right key  the intensified part of this signal gets shorter and fewer periods of the  signal are displayed on the Del   dTB  This way you can select any small portion of  a signal and capture and display it with a higher resolution                                                                                     S TIME DIV ns                                                                                                 576727  9303    HOW TO USE THE INSTRUMENT 4 65       DELAYED TIMEBASE TRACE SEP    The traces displayed by the Main Timebase and Delayed Timebase can be  separated with TRACE SEP    If the DELAYED TIMEBASE menu is activated  you will see the      symbol next  to the TRACE SEP text  The      indicates that the TRACK control can be used to  adjust the trace separation       Turn the TRACK control and observe that the DTB trace shifts                 dtb                                                                                           
241. the same function as pressing the front  panel CAL button for more than 2 seconds     Command               CL  CR                      Response                       acknowledge CR       Note           acknowledge   will be sent after the calibration has been fully  completed  A detailed error report may be queried for using the ST  command  only if acknowledge is not zero      Example     100 PRINT  1  CL    Send command  110 GOSUB 1000   Sync on acknowledge    6 10 THE CPL PROTOCOL       DEFAULT SETUP DS    Purpose   Sets the oscilloscope to the default setup conditions     The Default Setup  DS  command performs the same function as pressing the  TEXT OFF and STATUS LOCAL buttons simultaneously     The communication interface parameters will not be changed     Command               DS  CR                      Response                       acknowledge CR       Note  The  lt acknowledge gt  is sent after the completion of the change to the    default setups   Example   100 PRINT  1  DS      Send command    110 GOSUB 1000   Sync on acknowledge    THE CPL PROTOCOL 6 11       GO to LOCAL GL    Purpose     Puts the oscilloscope in the Local State  In the Local State  all oscilloscope  functions are accessible via the front panel buttons and knobs     The Go to Local  GL  command performs the same function as pressing the  STATUS LOCAL key on the front panel of the oscilloscope  when the oscilloscope  is in the Remote State  Refer also to  Go to Remote  and  Local Lockout      
242. the source for  this process and can be selected with the A control  The resulting trace is  automatically written in a register memory  m1 for math1 or m2 for math2      To see the result more clearly  you can turn off the two source traces by pressing  the  DISPLAY SOURCE on off softkey      Mathematical description   Result   81   S2    5 72 FUNCTION REFERENCE       Key sequence     add  substract  multiply    MATH  MATH1 2  filter           Control to select the SUBTRACT process        A     Control to select the first source trace   A          Control to select the second source trace   mE Toggle softkey to switch the SUBTRACT  function on   TRACK    Control to adjust the scale factor        SCALE  Oo   3          Q Control to adjust the offset factor        auto       bu  Toggle softkey to select the autoscaling   DISPLAY  SOURCE    yes no Toggle softkey to switch the source traces       on and off     ST7221  9303    Remote commands   CPL  QW  Command to query a waveform   Refer to Chapter 6 for full details     TEXT OFF    Description     Toggle key to select information in the CRT viewing area  This way  more space  is available on screen for trace display  There are three steps in the cycle      Softkey menu turned off      Instrument settings turned off      Both softkey menu and instrument settings turned on again     FUNCTION REFERENCE 5 73       Key sequence     TEXT OFF    Toggle key to cycle through three states of information  given in CRT viewing area     TI
243. ts    is highlighted as  default  This is an indication that the Delayed Timebase starts immediately after  the delay time has passed  For most signals  the    starts    mode can be used       Press the softkey to select the TRIG   D mode     When the triggered mode is selected for the Del   dTB  the start of the Del   dTB is  postponed until the first trigger occurs after the delay time  A valid trigger depends  on the proper setting of the trigger source  slope and level     The delayed timebase trigger source and slope can be selected with the same  TRIG 1  TRIG 2  TRIG 3  TRIG 4 or EXT TRIG buttons as are used for Main  Timebase triggering    When the DELAYED TIMEBASE is active  operation of the Del dTB triggering is  similar to that of the Main Timebase trigger source and slope    In the lower right hand corner of the screen  the delayed timebase trigger readout  is displayed under the readout for the Main Timebase triggering       Press the TRIG 1 key in the CH1 section a few times     Observe that the trigger slope for the Delayed Timebase changes     HOW TO USE THE INSTRUMENT 4 67       DELAYED TIMEBASE TRIGGER LEVEL    Just as is the case for Main Timebase triggering  proper triggering of the Delayed  Timebase depends on the selection of the proper trigger level     The A symbol in the Delayed Timebase trigger menu indicates the A control to be  used to adjust the trigger level     If the coupling for the vertical channel and the triggering are the same  a trigger  l
244. ualified technicians  Safety protection is likely  to be impaired when  for example  the instrument fails to perform the intended  measurements or shows visible damage     1 6 MEASURING EARTH    The measuring earth socket and the external conductor of the BNC sockets are  internally connected to the protective earth conductor of the three core mains  cable  The measuring earth socket or the external conductor of the BNC sockets  must not be used to connect a protective conductor     INSTALLATION INSTRUCTIONS 2 1       2 INSTALLATION INSTRUCTIONS    Attention  You are strongly advised to read this chapter thoroughly before  installing your oscilloscope     2 1 SAFETY INSTRUCTIONS    2 1 1 Protective earthing    Before any connection to the input connectors is made  the instrument shall be  connected to a protective earth conductor via the three core mains cable  the  mains plug shall be inserted only into a socket outlet provided with a protective  earth contact  The protective action shall not be negated by the use of an  extension cord without protective conductor     WARNING  Any interruption of the protective conductor inside or outside  the instrument is likely to make the instrument dangerous   Intentional interruption is prohibited     WARNING  When an instrument is brought from a cold into a warm  environment  condensation may cause a hazardous  condition  Therefore  make sure that the grounding  requirements are strictly adhered to     2 1 2   Mains voltage cord  main
245. uency    Long serial data streams    flickering  Display for  duration of  the event    Preferred    Preferred    Preferred when  Delayed sweep  is not used     Can capture and  display for long  term    May cause aliasing   Use Peak detect or  Envelope mode    May cause aliasing   Use Envelope mode     When using delayed  sweep to observe  details  Digital mode  provides better   light output     4 10    HOW TO USE THE INSTRUMENT       SIGNAL CRITERIA    ANALOG MODE    DIGITAL MODE       Video signals    OTHER CRITERIA    Need to see pretrigger  information    You need to make adjustments  to the circuitry and watch  the signal change    Automatic measurements    Signal Math   Add  Subtract  Multiply  Signal Analysis   Integration     Differentiation  FFT    Automatic Pass Fail test    Autorange attenuator    Autorange timebase    Preferred when  Delayed sweep  is not used     Not possible    Fastest  display update    Can   t use    Add  Subtract only    Not available    Not available    Not available    Not Available    When using delayed  sweep to observe  details  Digital mode  provides better light  output     Up to one screen    Slower display  update    Fully implemented    All functions    Full analysis  with  Math Plus option  installed    Fully implemented   with Math Plus option  installed    Results in a  displayed signal with  an amplitude of 2 to  6 4 divisions    Results in a signal  display of 2 to 6  waveform periods    HOW TO USE THE INSTRUMENT 4 11       S
246. urate  representation of the signal    Use the bottom key in the DISPLAY menu to control whether or not interpolation  is used  You can choose between dots  linear  and sine interpolation    The default is linear interpolation  Linear interpolation fills the spaces between  each sample on the screen with additional dots that are interpolated linearly  This  is used to obtain a smoothed display       Usethe MAGNIFY keys to set the horizontal magnify to  4     Turn the X POS control clockwise to display the adc noise only     The noise is displayed as continuous trace  This is because the space between  the samples is interpolated with additional dots and this filled with a continuous  line                                                                                                                                                                                   LINEAR DOTS SINE 91616      Press the bottom softkey to select    dots        Observe that only the sample dots are displayed and the space between the dots  is blanked       Press the bottom softkey to select    sine        Observe that the trace is smoothed again  but because of a different calculation   it is roundedoff as compared to the linear interpolation    Sine interpolation is recommended for signals primarily containing sine waves or  combinations     4 62 HOW TO USE THE INSTRUMENT       DIGITAL DISPLAY WINDOWS      Connect the Probe Adjust signal to channels 1 and 2     Press AUTOSET        The screen dis
247. ured  not on the  oscilloscope   This feature enables accurate signal probing and display freezing  without having to reach and touch the oscilloscope control     NOTE  The    PROBE SWITCH    setting of the    UTILITY  gt  gt  PROBE    menu must  first be set to  q meas  in order to obtain a reading on the oscilloscope  when you press the COMMAND switch on the measuring probe     Press the COMMAND switch again to go to RUN mode  TOUCH  HOLD  amp   MEASURE     results are deleted from display     Key sequence     autoset  setups  q meas    analog Toggle softkey to select the TOUCH  HOLD  amp   C mem      MEASUREMENT     mode reaction of the oscilloscope on  pressing the COMMAND switch on the measuring probe     ST7689    COMMAND             i COMMAND switch on the measuring probe to activate  A and leave the TOUCH  HOLD  amp  MEASURE     mode        FUNCTION REFERENCE 5 79       TRIGGER COUPLING    Description     Trigger coupling is used to optimize the trigger stability for different signal types     The filter modes      and dc are identical to those of the vertical inputs  Refer to  function INPUT COUPLING     Lf reject cuts off lower frequencies  triggering occurs on signals between 30 kHz  and full bandwidth     Hf reject cuts off higher frequencies  triggering occurs on signal frequencies lower  then 30 kHz  The following figure explains ac  If reject  and hf reject     Noise  on  improves trigger stability for noisy signals  By enlarging the trigger gap   of MAIN TB an
248. user  programmable as outlined below     SET STANDARD      Simultaneously press the STATUS and TEXT OFF keys     The standard setting feature resets all functions to a predefined state  At this time it  must be used to ensure that the standard setup condition applies before proceeding   SET STANDARD also resets the autoset function to the standard autoset condition     AUTOSET STANDARD    AUTO SET The AUTOSET function automatically sets all relevant functions of the  oscilloscope as they apply to the input signal  This includes the  selection of channels  input sensitivity  timebase setting  trigger   source  trigger slope  and trigger level for an optimum trace       Connect the Probe Adjust output to channel 1     Press the AUTOSET key     The result is a stable display with a number of signal periods  The amplitude is  well within the display range  In this example the Probe Adjust signal is displayed  with four periods cycles on the screen and an amplitude of three divisions  This  operating mode is referred to as AUTOSET after the SET STANDARD        STATUS                                                                                             TEXT OFF                                                                                  _5 0my  ITB 1 005                  576734          4 72 HOW TO USE THE INSTRUMENT       AUTOSET USER PROGRAMMABLE    The AUTOSET function can also be programmed so that certain functions switch  to a predefined position after an AUTOSET   
249. vents  or    glitches     that can be captured are 5 ns in the single  channel mode and 10 ns in the dual channel mode  Selection of peak detection is  made from the ACQUIRE menu     The following is an overview of peak detection possibilities                    Mode of use Description Peak detection speed  one channel any channel 5 ns  two channels Ch1  amp  Ch2 or Ch3  amp  Ch4   10ns  three channels ALT mode 10 ns  CHOP mode not applicable  four channels ALT mode 10 ns  CHOP mode not applicable          5 52 FUNCTION REFERENCE       See the function ACQUISITION LENGTH for additional information           SIGNAL  DETAIL       i                MAX SAMPLE RATE       GLITCH OFF           RD      EFFECTIVE SAMPLES    MIN    MIN GLITCH              1  1      EFFECTIVE SAMPLES  1                      1 SAMPLE PERIOD ST6660    Key sequence     ACQUIRE PEAK DET  Toggle key to switch the peak detection on off     POSITION    Description     Position controls allow the signals to be shifted across the screen to align signals  with the measuring graticule to make time and voltage measurements     Vertical positioning is done for each channel with the POS controls   Horizontal positioning of all signals is done with the X POS control   Key sequence    A POS   v Q Control to adjust vertical position of a channel      lt x POS  gt   O Control to adjust the horizontal position of all the channels    FUNCTION REFERENCE 5 53       POWER SUPPLY    Description     The instrument can be used a
250. will be CH1   CH2  This is indicated as  1 2  on the screen  The signals on both inputs are the  same  resulting in a straight line at ground level     If the line is not straight  this may be an indication that the probes are not correctly  adjusted  When one of the probes is not properly adjusted  the input signals at the  input connectors of the oscilloscope will be unequal  The difference of the two    input signals shows up in the differential mode                        CH1 CH2    C                             2                                                                       200m       CH2 20         CHP        poou                575969    4 20 HOW TO USE THE INSTRUMENT       VERTICAL MENU BANDWIDTH LIMITER    The Bandwidth Limiter reduces the bandwidth of the vertical channels to 20 MHz   This is done by activating a filter in the vertical channels  This feature can be used  both in analog mode and in digital mode to suppress high frequency noise  For  repetitive signals and when in the digital mode  averaging is the preferred  method to reduce noise without limiting the bandwidth       Press the    BW LIMIT    softkey to turn it on     Observe that the displayed line becomes  thinner  as an indication that the  amount of noise is reduced    The text    BWL appears in the bottom of the screen to indicate that the function is  active       Press the  BW LIMIT  softkey to turn it off again     HOW TO USE THE INSTRUMENT 4 21       VERTICAL MENU 500    The 200 MHz m
251. with a lower  level of OV and a top level of 600 mV       Press AUTOSET     4 32 HOW TO USE THE INSTRUMENT       ACQUIRE AVERAGE    The average function averages the input data over a number of successive  acquisitions  The average function is used to reduce the influence of random  noise in the input signal  There is no loss of bandwidth when the average function  is activated  but the signal must be repetitive  The number of samples over which  the average is calculated can be selected by the user       Use the probe to connect Ch1 to the Probe Adjust output      Press AUTOSET      Connect the second probe to channel 2  but do not connect it to a signal      Press the CH1 CH2 key to add Ch1 and Ch2 together      Press the mV key of the CH2 AMPL keys so that the noise picked up by  the probe tip of the second probe appears on the CH1 CH2 trace      Press the AVERAGE key     For maximum ease of use  the scope offers direct access to the important  average feature     The text  Average   8  appears in the bottom of the screen to indicate that the  function is active  The default value is 8  The indication changes into  AVG    Observe that the noise on the trace is reduced            change the Average constant  press the ACQUIRE key     Turnthe TRACK control clockwise     The Average factor increases when the TRACK control is turned clockwise   Observe that this reduces noise on the trace even more                       TRACK   zi   hverag    2 Averagd 286  VG    WG           
252. y  The chopper is also an advantage  for acquisitions at low timebase speeds     ALTERNATING ACQUISITION    Some applications require the use of an alternating signal acquisition mode  In  this mode each ADC can be used to its highest speed  Channels are acquired in  the following sequence      Channels 1 and 2 are acquired in parallel  with synchronized ADC acquisition   followed by channels 3 and 4  Because each ADC can be used up to 100 MS s   glitch capture is now possible for glitches down to 10 ns     RANDOM SAMPLING FOR FULL 200 MHz ACQUISITIONS AT HIGH TIME DIV  AND FOR FOUR CHANNELS IN PARALLEL     For high timebase settings  200 ns div to 2 ns div   real time signal acquisition  would require a real time sampling rate of 25 GigaSamples second  Such sample  rate is not achievable     The oscilloscope can be used with time base speeds exceeding the capabilities  of the ADCs  For such acquisitions repetitive signals are required  A sampling  method known as random sampling is used     This sampling method is automatically selected at higher time base speeds  In the  random sampling mode  the chopper is used to acquire four signals  simultaneously without restriction  The glitch detection circuit is not needed at  these high timebase speeds because the sample distance is smaller than the  glitch capture capability of 10 ns would allow  40 picoseconds at 2 ns div      APPENDIX P P 3       EXPANSION AND INTERPOLATION    Sometimes there is a need to look at a single shot o
253. y STATUS has no relation with this ST command     Command               ST  CR                      Response                       acknowledge CR       when acknowledge 0 followed by                          between  32768     32767         status CR  lt status gt    signed integer                       when status is negative       THE CPL PROTOCOL       Example     100  110  120  130  140    PRINT  1  ST    GOSUB 1000   INPUT  1 STATUS   IF STATUS  lt  0 THEN GOTO 120  GOSUB 2000       example status     6CR    0000000000000110 in binary     which means   because bit 2 and bit 1 are set       data out of range and    data format of the body is wrong      Send command     Sync on acknowledge    Read Status word    Fetch next status    Display Status          See section 6 7  STATUS  for program example     6 46 THE CPL PROTOCOL       TRIGGER ACQUISITION TA    Purpose   To perform a software trigger     This command causes an acquisition or sweep to be started  It is the software  equivalent of a normal trigger pulse  In the single shot mode the AT  Arm Trigger   command is sent to arm the triggering first     Command                  TA  CR                   Response           acknowledge CR                   Example     100 PRINT  1  TA    Send Trigger Acquisition  110 GOSUB 1000   Sync on acknowledge    THE CPL PROTOCOL 6 47       6 6 ACKNOWLEDGE            lt acknowledge gt  is a synchronization reply that is returned after each  command sent to the oscilloscope  signalli
254. you into thinking that the  Scope is properly triggered because the display appears to be stable  while in fact   you are displaying stored information only  Watch the ARM D LED                                                                                                                                 576737    HOW TO USE THE INSTRUMENT 4 37       TRIGGER MODE ROLL      Press the ANALOG key to select the digital mode     Press the TB Mode key to enter the TB mode menu    The TB MODE menu is extended with extensive timebase modes  The  differences are as follows      A Roll mode     Selection of Real time only      e Delay by events     Selection of Acquisition length      Press the softkey next to the ROLL on off    function to select    on        The    ROLL on    function is now intensified   Observe that the timebase is automatically set to 200 ms  The trace moves from  the right to the left       Press the  STOP ON TRIGGER    softkey to select    yes        The trace stops when the trigger  in this case the start of the first period  reaches  the left of the screen     Using the roll mode  you can monitor signals like temperature changes and  chemical processes at low timebase speeds  In this mode the scope operates like  a four pen plotter  Over 36 hours of events can be recorded in memory and  plotted later    Unlike a paper and pen plotter  the scope can record glitches as narrow as 5 ns   and the record can be stopped when a trigger condition is met     4 38 HOW TO
    
Download Pdf Manuals
 
 
    
Related Search
    
Related Contents
bedienungsanleitung d gb usa operating  INSTRUCTIONS MANUAL www.xenasecurity.com  Bedienungsanleitung CB-Funkgerät AE 4180  WD TV Live Streaming Media Player User Manual    Copyright © All rights reserved. 
   Failed to retrieve file