1. 8 Vdc 10 6 Vdc 5 15 Vdc stabilisized 1 3 mA with 10 6 Vdc Umweltanalytische Mess Systeme GmbH GmunderstraBe 37 D 81379 Miinchen Tel 49 0 89 12 6652 0 Fax 49 0 89 12 6652 20 eMail info ums muc de www ums muc de All rights reserved 19
2. are dissolved at atmospheric presseure will become gasiform with rising water tension If the soil dries out the tensiometer water vaporises at a pressure of about 900 hPa expands and runs out of the cap If the soil dries out further the water film in the pores of the ceramic disintegrates allowing air to enter When this happens a refill of the tensiometer is necessary At the upper end of the shaft two diametrically opposed holes are drilled A silicone rubber ring is stretched over them This allows excess pressure built up through too rapid placement of the sensor cap on the shaft to be released through these holes excess water can escape The pressure sensor is thus protected from excess pressure 3 General Instructions 3 1 How to Connect Your Tensiometer to a Display Unit or a Data Logging System The pressure transducer works as an asymmetric Wheatstone full bridge Tensiometers with plug new version since 2001 that have a black cable and the 4 pin plug M12 IP67 have the following configuration brown PIN 1 Supply white PIN2 Signal blue PIN3 Supply GND screen black PIN4 Signal GND Male plugon Female plug on sensor cable connecting cable Tensiometers without plug and the grey cable have this configuration brown wire is power supply minus white wire is power supply plus yellow wire is signal minus green wire is signal plus Note Do not connect the signal minus to the power supply minus
3. cap again in order to refill the shaft to the top If the sensor cap is now replaced the water tension should increase rapidly if a dry cloth is wrapped around the ceramic cap Wrap the ceramic cap in a damp cloth or place it in a water filled vessel so that at least the ceramic cap is covered with water Your T3 is now ready for use 10 3 6 Calibration The most important thing is to check the zero offset since this value drifts very much faster than the gain The stability of the zero point over one year is typically 0 5 of full scale To maintain an accuracy of 5 hPa calibration has to be done annually To recalibrate the slope you need the UMS Calibration Kit 3 6 1 Checking the zero offset For checking the zero offset it is not necessary to use the UMS Calibration Kit This check can be done much more easily gate N Place the tensiometer in a beaker een until the ceramic end is 10 mm below water level KIT mm The tensiometer must have previously been filled in the correct way After about 5 minutes or less time depends on the filling level you should get a constant value on your display unit or data logging system This value should not vary more than 5 hPa from the 0 hPa value in the calibration certificate If it exceeds this the new value should be written into your calibration certificate with the calibration date How to change the zero offset of your display recording unit should be shown in the cor
4. the referece port of the pressure transducer inside the body through the white membrane on the cable This membrane must always have contact to the atmospheric pressure After the tensiometer is filled the sensor cap is placed on the shaft using the silicon cone The protective cap also contains a pull protection for the cable However force should be avoided Secure the cable to the tensiometer shaft so that no one can trip over it Although the protective cap reflects the greater part of the sunlight striking it direct exposure to the sun should be avoided Temperature drift The sensor surface is electrical isolated to prevent influence of galvanic soil potentials 2 1 2 Tensiometer Shaft The tensiometer shaft consists of the ceramic cup also tensiometer cup the acryl glass shaft and the excess pressure protector The tenisometer ceramic acts like a porous filter Firstly it is water permeable so that the water tension of the soil can be transmitted to the tensiometer water and via this to the pressure transducer Secondly when wet it is impermeable to air below a certain pressure so that a water tension under pressure can build up in the tensiometer shaft Every ceramic is selected by UMS for a defined pore size and homogeneous pore structure The tensiometer water must be degased and deionized so it will transmit differing pressures hydraulically to the pressure transducer largely without changed volume Gases which
5. User manual Version 7 Pressure Transducer Tensiometer 2002 UMS GmbH Munich Umweltanalytische Mess Systeme GmbH Gmunderstr 37 D 81379 Muenchen Tel 49 0 89 12 66 52 0 Fax 49 0 89 12 66 52 20 eMail info ums muc de www ums muc de UMS measure to know Index 1 Sketch of the T3 Tensiometer 3 2 General Description 4 2 1 Tensiometer Construction 5 2 1 1 Sensor Cap 5 2 1 2 Tensiometer Shaft 6 3 General Instructions 7 3 1 How to Connect your Tensiometer to a Display Unit or a Data Logging System 7 3 2 Installation 8 3 3 How to Remove 9 3 4 How to Check the Filling Level refill on site 10 3 5 Refilling the Tensiometer 10 3 6 Calibration 11 3 6 1 Checking the zero offset 11 3 7 Maintenance and Storage 12 4 Tensiometric Principle and Limits 13 4 1 Theory of Water as Measuring Variable 13 4 2 Theory of Water as Pressure Transmitting Medium 14 4 3 Measuring Range and Standing Time 15 4 3 1 Pressure Transducer 15 4 3 2 Pore Size of the Ceramic Material Used 15 4 3 3 State of the Tensiometer Water 16 4 3 4 Ambient Pressure 18 4 3 5 Ambient Temperature 18 5 Technical Data 19 UMS measure to know 1 Sketch of the T3 Tensiometer screwed cable gland with strain relief pressure transducer silicon stopper protection ring transparent shaft water deionized cup ceramic cup 2 General Description The UMS T3 tensiometer is a probe that can be used i
6. built up in the tensiometer assuming the above men tioned potentials are neglected minus the pressure accounted for by the vertical length of the tensiometer is equivalent to the water tension in the soil under investigation The water column in the tensiometer leads to a hydrostatic potential p acting upon the pressure receptor due to its hanging weight and should be added to the measured value according to the equation as follows p density xg xh 1 Pa Pascal 1 N m This potential however has been compensated by calibration Example Vertical difference between pressure receptor h and ceramic tip h 10 cm p 1000 kg m x 9 81 m s x 0 1 m 981 kg ms 1 N kg m s 981 N m 1 hPa Hektopascal 100 N m p 9 81 hPa 13 The following schematic view informs of the measuring principle and the hydrostatic relations cmWS pF 5 5 10 7 Ww E g N ee g water not available 240 5 for plants E 10 42 A water A 3 availa ble B 10 3 for plants A 300 2 5 v ER 60 18 10 1 2 water level figure 1 Measuring Principles and Hydrostatic Relations taken from Lehrbuch der Bodenkunde Scheffer Schachtschabel 1981 4 2 Theory of Water as Pressure Transmitting Medium Since deionised and degassed water with a modulus of compression of approx 1 can be regarded as an ideal non compressible liquid it is useful as a pressure transmitting medium in the range of atmosphe
7. ent Pressure The maximum measuring value depends directly on the ambient pressure 4 3 5 Ambient Temperature The maximum measuring value depends on the vapour point of water which itself is temperature dependent and therefore also depends on the ambient temperature see vapour point curve exemplified We wish you the best success in measuring Do not hesitate to contact us for further questions 18 5 Technical Data Dimensions Cable length Measuring principle Range Signal Impedance Temperature shift Hysteresis Stability over one year Sensor Electronic principal Common mode range Power supply Current consumption 2001 UMS GmbH M nchen Length of ceramic cap ca 50 mm Diameter of ceramic cap ca 20 mm Length of tensiometer shaft as desired max 200 cm material acrylglas Height of the sensor cap ca 40 mm Diameter of the sensor cap below ca 14 mm 5m standard as desired Measuring the soil water tension transmitted via ceramic cup onto the tensiometer water and pressure transducer giving an analog signal 0 850 hPa pF 2 9 minus shaft length 1cm 1 hPa 0 100 mVdc 3 mV calibrated exact values are recorded for each tensiometer 2 6 kQ compensated typ drift 0 5 FS above 25 C typ 0 1 FS typ 0 5 FS piezoresistive pressure transducer overpressure max 3000 hPa assym wheatstone full bridge signal distance to mass with 10 6 Vdc supply 2 8 6
8. er remain immovable pull it out by turning screw up using as little force as possible 3 4 How to Check the Filling Level refill on site In order to accertain that the tensiometer is still properly filled check that there is an air bubble in the shaft directly beneath the sensor cap This bubble should always be as small as possible or better not be present If it is longer than 2 cm pull off the sensor cap and refill the shaft to the top with deionized water Then replace the sensor cap 3 5 Refilling the Tensiometer Place the tensiometer in a vessel with deionized water The ceramic cap must be immersed to a depth of at least 20 cm The shaft itself is not filled with water since this would lead to air being trapped within the walls ofthe ceramic cap Wait about 30 minutes until the water meniscus can be seen in the tensiometer shaft above the ceramic cap Now empty the shaft and fill the tensiometer shaft to the top with deionized water Place the sensor cap on the shaft with a slight rotational motion and connect the tensiometer to your display device Wrap a damp cloth around the sensor cap simulating dry soil A water tension of 850 hPa should be attainable minus the shaft length e g for a shaft of 30 cm gt 850 30 820 hPa The shaft must be held vertically with the sensor cap pointing downwards In case bubbles appear inside the shaft wall these can be caused to rise by gently tapping sideways Then remove the sensor
9. he measuring range as can be seen from the two phase diagram for water and water vapour When the tensiometer contains dissolved gases the vapour point is raised which restricts the measuring range considerably Therefore care should be taken to degas the deionised water as completely as possible e g by boiling 16 4 UMS measure to know If there are no gases dissolved in the water the measuring range may be repeatedly traversed The diffusion of gases dissolved in water through the ceramic material may then be neglected However if the tensiometer water contains dissolved gases which get gaseous before reaching the soil water tension water will flow through the ceramic material outside Dissolved gases pass once more into the tensiometer when soil water tension gets nearer to zero In view of the resulting exponential conditions the tensiometer water has therefore to be well degassed to obtain a long lifetime time that is needed until refill is necessary If the water tension values are low the tensiometer water may be regenerated less often This also applies for water tension values which show only small changes gt m Q oO 900 5 pressure p hPa 8004 7004 600 5004 liquid phase 400 1 vapour phase vacuum a2 KA 04 T T T T j T 0 b 2 3 40 5 Mm 48 00 temperature t C figure 2 Two Phase Diagram of Water Water Vapour 17 4 3 4 Ambi
10. ic cup is incorporated for the transmission of water The ceramic when wet is impermeable to gases This is the essential requirement for the build up of a underpressure in relation to the ambient pressure in the tensiometer 15 The bubble point i e the first entrance of air into the ceramic material which also leads to a limitation of the measuring range may be calculated according to the formula of capillary tension as follows 40 D Pore diameter um 10 m D Surface tension dynes cm water 0 C 75 6 20 C 72 7 50 C 67 8 K Capillary tension hPa Air may burst in if pressure difference of ceramic inside to ceramic outside exceeds capillary tension Then the water film in the biggest pores the biggest pore channel will disintegrate Since the bubble point depends on the largest pore line from outside to inside the latter should be as homogeneous as possible To avoid the flow resistance becoming too high the pore diameter must not be too small The ceramic material used has a pore size of 1 um together with high homogeneity of pore size distribution It can block up pressures of theoretically 3000 hPa only for water satuated ceramics and therefore provides for best performance in tensiometers Because of its fine pores the shaft may become covered by microorganisms in time Rinse it in dry cleaning spirits carefully and at length 4 3 3 State of Tensiometer Water The tensiometer water limits t
11. input of your measuring unit The signal minus is not OV or Ground when the Tensiometer is supplied with 5V 5 6V because then the signal range is between 3 2 V and 6 8 V with reference to power supply As the pressure transducer is a Wheatstone full bridge it has to be connected in a certain mode Please read therefore the manual of your display unit or data logging system before connection 3 2 Installation Please note that the water column inside the shaft causes an hydrostatic pressure on the pressure sensor and has to be compensated The water column is the shaft length plus 3 cm for the cup Withdraw 1 hPa for each cm of water column from the reading Correct soil water tension hPa measured reading hPa water column cm Find the correction value depending on the resp shaft lenght on the calibration certificate Example for 30 cm shaft reading 130 hPa Correct tension 130 hPa 33 cm 97 hPa The tensiometer has an outer diameter of 20 mm An appropriately large drilling rod UMS accessory must be selected drill a hole of the required depth at the drilling location In stony soil it may be necessary to drill the hole several times Should the walls of the hole fail to hold UMS lining tubes can be used Pour a thin paste of quartz clay powder and water into the hole For loamy soil use ca 20 ml plus ca 100 ml of water for coarse clay like soil ca 30 ml for sandy soil ca 100 ml and for stony soil adj
12. n a wide variety of applications It is very compatible safe and extremely simple in handling The type of tensiometer is shown on the label not for newer versions with plug T3 30 type length of shaft em Electronic pressure transducer tensiometers are high resolution measuring devices for the continuous measurement of water tension in the soil being used in a wide range of hydrological studies soil physical studies or general studies of ecosystems The T3 is used to measure water tension in the soil in the non saturated zone The water tension is measured and converted into a defined continuous electrical signal This signal can be displayed with a voltmeter and power supply or with the UMS Infield 5 which transmits the electric signal in pressure unit hPa hecto Pascal Also the signal can be adapted to a data logging unit for logging the water tension signals automatically The water tension is a direct measure of the availability of water in the soil for plants and is therefore an important plant physiological parameter In the field of ecosystems tensiometers are used for hydrological investigations e g for input and transport studies gravitational water studies or to measure the water tension itself as a characteristic value used in mathematical models of energy or water balance Tensiometers can also be used as control sensors for UMS irrigation systems or as control sensors for UMS suction sampling systems Care sho
13. responding user manual of your display unit or data logger If the slope is also to be calibrated UMS offers this as a service 11 3 7 Maintenance and Storage 3 7 1 Cleaning the Tensiometer Clean the tensiometer with a damp cloth water or with petroleum ether Do not use acid or solvent as they could attack the adhesive joint or the material itself 3 7 2 Storage Place the tensiometer as near to the vertical as possible in a pail of water so that the ceramic cap cannot dry out To prevent the formation of algae the tensiometer should be stored in a dark place In case of storage for longer than 3 months the T3 can also be stored dry Filled tensiometers always must be protected from freezing 12 4 Tensiometric Principle and Limits 4 1 Theory of Water as Measuring Variable Water tension measurement absorptive tension measurement as a direct means for the measurement of water available in soils for plants involves the measurement of total potential of water retention in the soil under investigation excluding osmotic potential potentials of differential pressure and gravitation The saturation state of soil level of ground water resp determines the extent of water absorption with regard to the water tension in soil through the ceramic material by the tensiometer which is otherwise hermetically sealed The porous properties of the ceramic material sintered Al O are considered ideal The atmospheric underpressure
14. ric pressure p 0 hPa to atmospheric underpressure p 950 hPa in a temperature range of gt 0 C to lt 30 C without change of volume see curve of the water vapour pressure relationship This means that the tensiometer with its mechanical tightness in the said pressure range exchanges a very limited amount of water via the ceramic material during the process occuring when pressure is transfered from the ceramic tip to the pressure transducer sensor This applies to the use of relatively rigid material like acrylic glass and the pressure receptor 14 The tensiometer has been designed so that only a small amount of water is exchanged when the soil cycles from wet to dry 0 to 850 hPa This ensures that the tensiometer can last for many drying cycles provided that the water inside is free from dissolved gasses and all bubbles including microscopic ones 4 3 Measuring Range and Standing Time The extent of measuring depends on the following parameters 1 Pressure receptor 2 Pore size of ceramic cup 3 State of tensiometer water 4 Ambient pressure 5 Ambient temperature 4 3 1 Pressure Transducer The pressure transducer can tolerate pressures in the range of 3000 hPa to 3000 hPa theoretical value non destructively and measure them linearly in the range of 1000 hPa to 1000 hPa linearity error lt 0 1 4 3 2 Pore Size of the Ceramic Material Used In order to measure soil water tension a porous ceram
15. uld be taken in correlating the water tension with the volumetric water content This is only possible if the soil texture is known very accurately 2 1 Tensiometer Construction 2 1 1 Sensor Head The sensor head consists of the pressure transducer which is connected to a test supply and pressure reference lead the silicon cone a protective cap and the type plate The sensor principle is based on the piezoresistive effect of silicon semiconductors which change their specific electric resistance on deformation and are processed to a defined signal via a Wheatstone bridge The deformation is caused by the pressure or water tension on the silicon chip which is very thin and therefore extremely sensitive to pressure variations The pressure transducer is calibrated for 10 6 Vdc and so requires a 10 6 Vdc regulated power supply such as the TV BATT The signal changes in direct proportion to the power supply At 10 6 Vdc we obtain a signal of 0 hPa 0 mVdc and 1000 hPa 103 mVdc The tolerance is 3 mV depending on the pressure transducer The exact calibration values are shown in the UMS Calibration Certificate supplied with each tensiometer This is not necessary for tensiometers supplied with Infield systems Since 2001 the signals of all Tensiometers are balanced by potentiometers inside the pressure senor head Since the water tension is measured with respect to the atmospheric air pressure this pressure is transferred to
16. ust the proportions according to the stone content Now remove the plastic container from the tensiometer cup turning the container slowly clockwise when viewed from below Immediately insert the tensiometer in the hole in the soil Filling the T3 3 3 How to Remove If the tensiometer is to be removed from the soil you will need the following equipment 1 The plastic bottle protective container for the ceramic cap with ca 10 ml of deionised water 2 A five liter container of water tap water 3 Adhesive tape and a water proof marker 4 For each tensiometer a rod approximately 20 mm in diameter and of a length sufficient to allow the tensiometer to be reinserted at the same location later The rod should be rounded at the end The shape of the tensiometer cup is ideal Procedure Hold the tensiometer by its shaft and pull it vertically out of the soil avoiding sideways movements After removal place the plastic bottle half filled with water over the ceramic cup rotating it slightly clockwise Push the rod into the hole in the soil with a slight turning motion Mark the rod with the tensiometer number If the tensiometer is connected to a logger the signal channel should also be noted on the end of the cable Please note The tensiometer should not be rotated when inserted and removed If it is removed after a dry period and the tensiometer is stuck water poured along the shaft works wonders Should the tensiomet
Download Pdf Manuals
Related Search
Related Contents
Innodisk E3SS-32R1 USER'S MANUAL TTP-2410M/346M TTP DPhyDecoderHSCtl User`s Manual PLEASE READ FIRST Manual Change - Agilent Technologies Istruzioni di Uso e Manutenzione Scientech VLSI Development Platform: SR-HD1500EU SR-HD1250EU - info Use & Care Instructions MVC-CD400 Copyright © All rights reserved.