1. Model code Suffix code Description Part Name Quantity 710105 DLM2022 Digital Oscilloscope 2ch 20OMHz Power cord 1 710110 DLM2024 Mixed Signal Oscilloscope 4ch 200MHz Passive probe model 701938 200 MHz Per number of channels 710115 DLM2032 Digital Oscilloscope 2ch 350MHz oe MOSER AUTOA AIT 710120 DLM2034 Mixed Signal Oscilloscope 4ch 350MHz E ee o rer number OF channa 710125 DLM2052 Digital Oscilloscope 2ch 500MHz 710125 7101307 710130 DLM2054 Mixed Signal Oscilloscope 4ch 500MHz Protective front cover 1 Power cord D UL CSA standard Soft carrying case for probes 1 F VDE standard Printer roll paper for B5 option 1 roll Q BS standard User s manuals 1 set R AS standard 1 The 701938 probes are not included when EX22 or EX24 is selected H GB standard The 701989 probes are not included when EX52 or EX54 is selected 3 Operation guide as the printed material and User s manual as CD ROM are included N NBR standard Language HE English Menu and Panel Additional Option License for DLM2000 HC Chinese Menu and Panel Model Suffix code Description HK Korean Menu and Panel 709810 G2 User defined math 4 ch model only HG German Menu and Panel G3 Power supply analysis function 4 ch model only HF French Menu and Panel G4 Power supply analysis function includes G2 4 ch model only HL Italian Menu and Panel F1 UART trigger and analysis 4 ch model only HS
2. Sampling modes Real time interpolation repetitive sampling Accumulation Select OFF Intensity waveform frequency by brightness or Color waveform frequency by color Accumulation 100 ms to 100 s Infinite time Roll mode Enabled at 100 ms div to 500 s div depending on the record length setting 1 V to 10 V div 1 of setting 20 mV Frequency characteristics 3 dB attenuation wh en inputting a sinewave of amplitude 3div T Zoom function Two zooming windows can be set independently Zoom1 Zoom2 Zoom factor x2 to 2 5 points 1Odiv in zoom area DLM202x DLM203x DLM205x 1 MQ when using passive probe 100 mV to 100 V div DC to 200 MHz DC to 850 MHz DC to 500 MHz Scroll Auto Scroll Search Edge Edge Qualified State Pulse Width State Width functions C option SPI option UART option CAN option LIN option Flexray option History memory Max data 10 000 record length 1 25 kPoints with M1 or M1S option 20 000 record length 1 25 kPoints with M2 option 50 000 record length 1 25 kPoints with M3 option History search Select Rect WAVE Polygon or Parameter mode Replay Automatically displays the history waveforms function sequentially Display Specified or average waveforms Cursor Types AT AV AT amp AV Marker Degree Snapshot Currently displayed waveform can be retained on screen Computation
3. on an edge of any channel an edge or state when conditions on other channels are met or when the width of a pulse is either more or less than a specified time Dedicated triggers are also available for serial bus options Edge trigger Edge Enhanced triggers Edge OR Edge qualified State Pulse width State width Serial optional CAN LIN UART I C SPI standard user defined TV NTSC PAL SDTC HDTV user defined B triggers E A Delay B A to B n Dual bus combination trigger of 2 serial busses YOKOGAWA Trigger on arbitrary serial bus patterns In order to support any type of serial bus and thus the ability to trigger on any combination of ID and data etc a user define trigger is provided On one input channel a pattern of up to 128 bits can be set and other channels can also be used for clock chip select and latch signals HHHH LHLH LLLL Trigger Saeed lala ale mi tual ead toca ton Arbitrary pattern xe Sam ew eee Hee ey eer is y tE Cerf GE gri ee se ER Mit E A SE a E E Combination triggers Via the B trigger menu two serial bus triggers can be combined which means that the DLM2000 for example will trigger when signal conditions on either a CAN bus or a LIN bus are met LIN CAN Input signal A CAN Input signal B LIN The A to Bin trigger provides the ability to trigger when a specific number of edges has occurred on input B This enables measurements on signals
4. DLM2000 s internal storage can be recognized by a PC as an external USB storage device Transferring files is easy even when a USB thumb drive can t be used ce Gl om Supports USB yp l Sends waveform screen and settings data Remote control Comprehensive software tools DLM2000 Series A comprehensive suite of software tools 12 to support and complement complex measurement tasks Comprehensive software tools XviewerLlTE Basic display and measurement Provides zooming vertical cursors and data conversion Xviewer Advanced analysis Xviewer can display acquired waveforms Off line to CSV format transfer files and control instruments waveform remotely In addition to simply displaying display and the waveform data Xviewer features many analysis of the same functions that the DLM2000 offers zoom display cursor measurements calculation of waveform parameters complex waveform math and FFT Binary Wetec Xwirepuller waveform data call ae a to monitoring S CSV Excel or Floating Point Decimal format onaPC e DLM2000 can be simply controlled using a PC and mouse via an Ethernet USB or GP IB interface Data transfer When the software program starts a simulation of toaPC the oscilloscope appears on the PC display LabVIEW drivers MATLAB toolkit By using the LabVIEW driver written for the The MATLAB tool kit can be used to control DLM2000 a developer can dramatically reduce the the DLM2000 and to transfer data via GP IB
5. a histogram of acquired waveforms Min input voltage 701988 500 mVp p 701989 300 mVp p Input range Model 701988 40 V Model 701989 threshold 6V Max nondestructive input voltage 40 V DC ACpeak or 28 Vrms when using 701989 User defined math G2 option The following operators can be arbitrarily combined in equations X SIN COS TAN ASIN ACOS ATAN INTEG DIFF ABS SQRT LOG EXP LN BIN DELAY P2 power of 2 PH DA MEAN HLBT PWHH PWLL PWHL PWLH PWXX FV DUTYH DUTYL The maximum record length that can be computed is the same as the standard math functions Threshold level setting range Model 701988 40 V setting resolution of 0 05 V Model 701989 6 V setting resolution of 0 05 V Power supply analysis G3 G4 option Input impedance 701988 Approx 1 MQ approx 10 pF 701989 Approx 100 kQ approx 3 pF Power analysis For Pwr1 and Pwr2 selectable from 4 analysis types De skewing between the voltage and current waveforms can be executed automatically Maximum sampling rate 1 25 GS s Maximum record length M1 M1S option Repeat 6 25 MPoints Single 25 MPoints Switching loss Total loss switching loss power waveform display Automatic measurement and statistical analysis of power analysis items Wp Wp Wp Abs Wp P P P Abs P Z Safety operation area SOA analysis by X Y display
6. amp Analysis Functions Parameter measurement Max Min P P High Low Rms Mean Sdeyv IntegTY Integ TY OVER OVER Pulse Count Edge Count V1 V2 AT Freq Period Avg Freq Avg Period Burst Rise Fall Width Width Duty Delay Statistical computation of parameters Min Max Mean standard deviation Count Statistics modes Continuous Cycle History Trend Histogram display of wave parameters Up to 2 trend or histgram display of specied wave parameters Computations MATH X Filter Delay Moving Avg IIR Lowpass IIR Highpass Integ Count Edge Rotary user defined math optional Computable no of traces 2 Math1 Math2 1 trace for 2ch model Max computable memory length M1 M18S option 25 MPoints M2 option 62 5 MPoints M3 option 125 MPoints 20 mV to DC to 150 MHz DC to 300 MHz DC to 400 MHz 50 mV div 50 Q 10 mV to DC to 200 MHz DC to 350 MHz DC to 500 MHz 500mV div 2 mV to DC to 150 MHz DC to 300 MHz DOC to 400 MHz 5 mV div Isolation between channels 34 dB analog bandwidth typical value Residual noise level The larger of 0 4 mV rms or 0 05 div rms typical value A D resolution 8bit 25LSB div Max 12 bit in High Resolution mode Bandwidth limit FULL 200 MHz 100MHz 20 MHz 10 MHz 5 MHz 2 MHz 1 MHz 500 kHz 250 kHz 125 kHz 62 5 kHz 32 kHz 16 kHz 8 kHz can be set for each channel Maximum sample rate Real t
7. the user s specific requirements Serial bus triggering and analysis Serial bus triggering and analysis Up to 4 buses simultaneously Dedicated trigger and analysis options are available for FlexRay CAN CAN FD LIN SENT UART 12C and SPI serial buses From most buses a wide variety of trigger combinations can be set including ID and Data combinations which can also be combined with conventional edge triggers A serial bus auto setup enables the MSO to be quickly configured The user therefore does not need detailed knowledge of the bus frame format Analysis can be performed at high speed simultaneously on up to four different buses operating at different speeds This is enhanced by the extensive search facilities allowing the user to look for specific data in the very long memory The dual zoom facility means that different buses can be viewed and debugged alongside each other Four bus decode and list display Simultaneous analyses of I C and SPI DLM2000 Series 10 Power supply analysis options The G3 and G4 options enable switching loss joule integral i2t SOA safe operating area harmonics based on EN61000 3 2 and other power parameters to be measured and analyzed Switching loss analysis Using the long memory the switching loss of the voltage and current input waveforms can be computed V t X i t over long time periods The turn on off loss the loss including the continuity loss and the
8. time from trigger position Time ms Data Bin Hex display ASCII display and Information Auxiliary analysis functions Data search Analysis result save function Analysis list data can be saved to CSV format files CAN Bus Signal Analysis Functions F4 F6 F7 and F8 Options Applicable bus CAN version 2 0A B Hi Speed CAN ISO11898 Low Speed CAN ISO11519 2 Analyzable signals CH1 to CH4 M1 to M2 Bit rate 1 Mbps 500 kbps 250 kbps 125 kbps 83 3 kbps 33 3 kbps User defined an arbitrary bit rate from 10 0 kbps to 1 000 Mbps with resolution of 100 bps CAN bus Trigger modes SOF ID DATA ID OR Error enabled when loading physical values symbol definitions Auto setup function Auto setting of bit rate threshold value time axis scale voltage axis scale and display of analysis results Internal Storage Standard model C9 Option Standard model 300 MB C9 option 7 2 GB Capacity Built in Printer B5 Option Built in printer 112 mm wide monochrome thermal USB Peripheral Connection Terminal Connector USB type A connector x 2 front panel x 1 rear panel x 1 USB 2 0 compliant Low Speed Full Soeed High Speed USB Printer Class Ver 1 0 compliant ERSON HP PCL ink jet printers USB Mass Storage Class Ver 1 1 compliant mass storage devices Please contact your local Yokogawa sales office for model names of verified devices Electromech
9. Spanish Menu and Panel F2 C SPI trigger and analysis 4 ch model only Option LN No switchable logic input 4 ch model only F3 UART I C SPI trigger and analysis 4 ch model only B5 Built in printer F4 CAN LIN trigger and analysis 4 ch model only Standard continuous measurement 8 25 Mohs Shale made 25 ee ee Mpoints when interleave mode ON 62 5 Mpoints F6 CAN LIN FlexRay trigger and analysis 4 ch model only as ecto pester er P M oe node 62 5 F7 CAN CAN FD LIN trigger and analysis 4 ch model only ee H E a a F8 FlexRay CAN CAN FD LIN trigger and analysis 4 ch model only continuous measurement 25 Mpoints Single mode 125 F9 SENT analysis 4 ch model only Mpoints when interleave mode ON 250 Mpoints M1S Memory expansion option 2 ch model only During X1 F4 gt F7 or F6 gt F8 add CAN FD Standard Po eal ML eds risa Separately sold license product customer installable P2 Probe power for 2 ch models Accessory Models P4 3 Probe power for 4 ch models Name Model Specification IC1 4 GP IB Interface Logic probe PBL100 701988 1 MQ input resistance toggle frequency of 100 MHz C10 Fthernet Interface Logic probe PBL250 701989 100 kQ input resistance toggle frequency of 250 MHz C11 4 GP IB Ethernet Interface Passive probe 701938 10 MQ 10 1 200 MHz 1 5 m CQ Internal storage 7 2 GB Passive probe 701939 10 MQ 10 1 500 MHz 1 3 m 1G2 5 User defined math 4 ch model only Miniature pa
10. Test Measurement YOKOGAWA oil WOE Ct oO r Precise control and flexibility DLM2000 Series Mixed Signal Oscilloscopes Bulletin DLM2000 E E The DLM2000 embodies everything a user would expect in an oscilloscope It s a family of products that goes beyond the demands and needs of users Typical of a company focused on quality the DLM2000 has been built to last decades crafted by engineers to meet current demands and future proofed to keep track with the ever rapid changes in technology A product designed for the future but at today s prices The DLM2000 is a series of bench top oscilloscopes made for electronic design and debug lts ergonomic easy to use and complete with all the features and more you would expect in today s oscilloscope With bandwidths from 200 to 500 MHz and memory from 1 to 250 MPoints there is a DLM2000 to meet your application and budget Why choose the DLM2000 Quality As a bench top instrument an oscilloscope is the most used piece of test and measurement equipment To meet the rigors of everyday use it needs to be reliable Yokogawa s reputation for high quality products ensures the user is never let down and can depend on the DLM2000 Innovation he ever increasing demands of today s test needs means oscilloscopes must be versatile and adapt to all sorts of different applications Yokogawa s DLM2000 series is equipped with all the fea
11. amount of work required to enable a PC to control USB or Ethernet from within MATLAB the instrument from within the LabVIEW environment TMCTL library This DLL Dynamic Link Library enables Microsoft Visual studio programs such as Visual C Command and Visual Basic to be quickly developed to control communicate between the PC and the DLM2000 It supports GPIB USB and Ethernet interfaces Custom software DLTerm development The command line tool can be used with the TMCTL library to develop communication programs Prototype code can be rapidly created to automate sequences of capture measurement and analysis tasks before writing a fully custom software routine Symbol editor Physical value symbol definition files for CAN serial bus analysis can be created and edited CANdb files can also be imported 13 Recommended Probes and accessories Recommended Probes and accessories The extensive range of Yokogawa probes includes models which are designed and optimized for specific applications For power electronics testing PBC050 PBC100 Current probes DC to 50 MHz 100 MHz 30 Arms L PBDH0150 differential probe a SS N DC to 150 MHz E H 1400 V DC ACpeak differential and common mode voltage N N Ay 701926 differential probe DC to 50 MHz 5000 Vrms 7000 Vpeak KI 701936 deskew correction signal source Higher current and other differential probes are availab
12. anical specifications Supported transfer standards Supported devices USB PC Connection Terminal Connector USB type B connector x 1 USB 2 0 compliant Electromechanical specifications Analyzable no of frames 100 000 frames max Supported transfer standards High Speed Full Speed Analysis results displays Analysis no time from trigger position Time ms Frame type ID DLC Data CRC presence absence of Ack information Auxiliary analysis functions Data search and field jump functions Analysis result save function Analysis list data can be saved to CSV format files CAN FD Bus Signal Analysis Functions F7 and F8 Options Applicable bus CAN FD Version 1 0 Analyzable signals CH1 to CH4 M1 to M2 Bit rate Arbitration 1 Mbps 500 kbps 250 kbps User Define an arbitrary bit rate from 20 kbps to 1 Mbps with resolution of 100 bps Data 8 Mbps 5 Mbps 4 Mbps 2 Mbps 1 Mbps 500 kbps User Define an arbitrary bit rate from 250kbps to1OMbps with resolution of 100 bps CAN FD bus trigger modes SOF ID ID OR Error Frame Message enabled when loading physical values symbol definitions Auto setup function Auto setting of bit rate recessive Level threshold value time axis scale voltage axis scale and display of analysis results Analyzable no of frames 50 000 frames max Analysis results displays Analysis
13. ation MATLAB tool kit 701991 MATLAB plug in Xviewer 701992 SP01 For DL DLM Series standard version LAX NOTE Before operating the product read the user s manual thoroughly for proper and safe operation 701992 GP01 For DL DLM Series with MATH functions YOKOGAWA EUROPE B V Phone 31 88 4641000 Fax 31 88 4641111 Subject to change Euroweg 2 3825 HD Amersfoort The Netherlands MINOU TONG E mail tmi nl yokogawa com tmi yokogawa com YOKOGAWA METERS amp INSTRUMENTS CORPORATION Global Sales Department Phone 81 42 534 1413 Fax 81 42 534 1426 SUNSU DUVZ000 Ee E mail tm cs jp yokogawa com EALEN YOKOGAWA CORPORATION OF AMERICA Phone 1 770 253 7000 Fax 1 770 254 0928 Copyright Yokogawa 2014 YOKOGAWA ENGINEERING ASIA PTE LTD Phone 65 62419933 Fax 65 62412606 Printed in The Netherlands 2015
14. can then be easily seen The variation of parameters can also be displayed as histograms thus providing a visual method of assessing them statistically Automatically measure time and voltage differences Line or waveform marker cursors can be placed on different displayed waveforms and the absolute values of voltage and time and their differences can be simply displayed A degree cursor can also be used by converting the time axis into a position degree axis FFT frequency domain analysis 2 FFT analyses can be performed simultaneously The source data can be either from input channels or the results of mathematical computations As well as standard Power Spectrum calculations a full suite of FFT functions are available using the G2 user define math option User define math Up to 2 math channels are available The standard DLM2000 provides arithmetic and filtering functions on computations of up to 125 MPoints By installing the G2 option the oscilloscope offers comprehensive user defined mathematics Equations can be arbitrarily created using a suite of operators such as trigonometric and logarithmic operators integration and differentiation pulse width operators phase measurement and digital to analogue conversion in the image When used in combination with cursor and automatic waveform measurements on the computed waveforms the DLM2000 Is able to provide meaningful results according to
15. et 1 MQ 1V 2 mV div to 50 mV div setting range 10V 100 mV div to 500 mV div 100V 1 V div to 10 V div 50 Q 1V 2 mV div to 50 mV div 5V 100 mV div to 500 mV div DC accuracy 1 5 of 8 div offset voltage accuracy Offset voltage accuracy 2 mV to 50mV div 1 of setting 0 2 mV 100 mV to 500 mV div 1 of setting 2 mV Trigger type trigger Source A triggers Serial Bus C optional CH1 to CH4 Logic SPI optional CH1 to CH4 Logic UART optional CH1 to CH4 Logic FlexRay optional CH1 to CH4 CAN optional CH1 to CH4 CAN FD optional CH1 to CH4 LIN optional CH1 to CH4 SENT optional CH1 to CH4 Logic User defined CH1 to CH4 AB triggers A Delay B 10 ns to 10 s Edge Edge Qualified State Serial Bus A to B N 1 to 10 Edge Edge Qualified State Serial Bus Dual Bus Serial bus only Trigger level setting range CH1 to CH4 4 div from center of screen Trigger level setting CH1 to CH4 0 01 div TV trigger 0 1 div resolution Trigger level accuracy CH1 to CH4 0 2 div 10 of trigger level Window Comparator Center Width can be set on individual Channels from 1 4 CH1 to CH4 Display Display 8 4 inch TFT color liquid crystal display 1024 x 768 XGA Functions Waveform acquisition modes Normal Envelope Average High Resolution mode Max 12 bit the resolution of the A D converter can be improved equivalently by placing a bandwidth limit on the input signal
16. ime Interleave OFF 1 25 GS s sampling Interleave ON 2 5 GS s mode Repetitive 125 GS s sampling mode Maximum record length 2 ch model Repeat Single Single Interleave M1S 6 25 M 25 M 62 5 MPoints 4 ch model Repeat Single Single Interleave M18S 6 25 M 25 M 62 5 MPoints 4 ch model Repeat Single Single Interleave M2 12 5 M 62 5 M 125 MPoints 4 ch model Repeat Single Single Interleave M3 25 M 125 M 250 MPoints Ch to Ch deskew 100 ns Time axis setting range 1 ns div to 500 s div steps of 1 2 5 Time base accuracy 0 002 Reference function Up to 2 traces REF1 REF2 of saved waveform data can be displayed and analyzed Max acquisition rate Approx 20 000 waveform sec ch Accumulation mode Dead time in N Single mode Approx 2 2us approx 450 000 waveforms sec ch Action ON trigger Modes Rect Wave Polygon Parameter Logic Signal Input 4 ch model only Number of inputs 8 bit excl 4 ch input and logic input Maximum toggle frequency Logic probe 701988 100 MHz Logic probe 701989 250 MHz Compatible probes 701988 701989 8 bit input 701980 701981 are available GO NO GO Actions Buzzer Print Save Mail XY Displays XY1 to XY2 and T Y simultaneously FFT Number of points 1 25k 12 5k 25k 125k 250k Window functions Rectangular Hanning Flat Top FFT Types PS LS RS PSD CS TF CH are available with G2 or G4 option Histogram Displays
17. l Analysis Functions F2 amp F3 Options Applicable standard J2716 JAN2010 and older Trigger types 3 wire 4 wire After assertion of CS compares data after arbitrary byte count and triggers Analyzable signals CH1 to CH4 Logic input M1 to M2 Byte order MSB LSB Analyzable signals CH1 to CH4 logic input or M1 to M2 Clock period 1 us to 100 us with resolution of 0 01 us Data type Fast channel Nibbles User Defined Slow channel Short Enhanced SENT trigger modes Start of fast channel Auto setup function Auto setting of threshold value time axis scale voltage axis scale and display of analysis results Auto setup function Auto setting of clock period nibble number pause pulse threshold value time axis scale voltage axis scale and display of analysis results Analyzable no of data 300 000 bytes max Analyzable no of frames 100 000 frames max Decode bit length Specify data interval 1 to 32 bits decode start point and data length Analysis results displays Analysis no time from trigger position Time ms Data 1 Data 2 Auxiliary analysis functions Data search function Analysis results displays Analysis no time from trigger position Time ms Fast channel Sync Cal period Tick Status amp Comm Data CRC frame length information Slow channel Analysis no time from trigger position Time ms ID Data CRC informa
18. le See page 16 YOKOGAWA For serial bus testing PBDH1000 1GHz differential probe 1 M ohm 1 1pF input 25 V differential voltage input Compatible with the FlexRay standard 701920 701922 differential probes 500 MHz 200 MHz 12 V 20V differential voltage input Logic probes PBL100 PBL250 8 bit logic probes 100 MHz 250 MHz toggle frequency 1 M ohm 100 K ohm input PBA1000 1 GHz Active probe 100 k ohm 0 9 pF input Specifications DLM2000 Series Models Model name Frequency Input terminal Max sample rate bandwidth DLM2022 710105 200MHz DLM2032 710115 350MHz 2 analog channels 1 25GS s DLM2052 710125 500MHz interleave mode off DLM2024 710110 200MHz 4 analog channels 2 5GS s DLM2034 710120 3850MHz 3 analog channels interleave mode on DLM2054 710130 500MHz 8bit logic Basic Specifications Analog Signal input Input channels Analog input DLM20x2 CH1 CH2 DLM20x4 CH1 to CH4 CH1 to CH3 when using logic input Input coupling setting AC DC DC50 Q GND Input impedance Analog input 1 MQ 1 0 approximately 20 pF 50 Q 1 0 VSWR 1 4 or less DC to SOOMHz Voltage axis sensitivity 1MQ 2 mV div to 10 V div steps of 1 2 5 setting range 50 Q 2 mV div to 500 mV div steps of 1 2 5 Max input voltage 1 MQ 150 Vrms 50 Q Must not exceed 5 Vrms or 10 Vpeak Max DC offs
19. loss over many cycles of the 50 Hz 60 Hz power line can be calculated and analyzed Power measurement The MSO can also be used as a power meter by providing automated measurement of power parameters for up to two pairs of voltage and current waveforms such as the active power apparent power and power factor These values can then be statistically processed and calculated Switching loss analysis Power parameter measurement 11 Connectivity Connectivity Ehernet optional Supports 1000BASE T 100BASE TX 10BASE T Go No Go output terminal TTL level output of the result of the GO No GO function RGB video signal output terminal Check the image of the waveform on a external monitor USB PC connection terminal Enables control from a PC USB perpheral connection terminal Supports USB storage USB keyboards USB printers Probe power terminal optional Power supply for current and differential probes GP IB connection terminal optional Enables control from a PC External trigger input An input for a trigger signal separate from the channel signals Trigger output Output a CMOS 3 8V level trigger signal YOKOGAWA 1000BASE T 100BASE TX 10BASE T Sends waveform screen and settings data Remote control Mail transmission GO NO GO action Y USB standard on rear panel Ethernet C10 C11 options PaaS
20. mplies with System Management Bus Analyzable signals CH1 to CH4 M1 to M2 Analyzable signals CH1 to CH4 Logic input or M1 to M2 Bit rate 10 Mbps 5 Mbps 2 5Mbps 12C Trigger modes Every Start Address amp Data Non Ack General Call Start Byte HS Mode FlexRay bus Trigger modes Frame Start Error ID Data ID OR Analysis results displays Analysis no time from trigger position Time ms 1st byte address 2nd byte address R W Data Presence absence of ACK information Auto setup function Auto setting of bit rate threshold value time axis scale voltage axis scale and display of analysis results Analyzable no of frames 5 000 Auto setup function Auto setting of threshold value time axis scale voltage axis scale and display of analysis results Analysis results displays Analysis no time from trigger positions Time ms Segment Static or Dynamic Indicator FramID PayLoad length Cycle count Data Information Analyzable no of data 300 000 bytes max Auxiliary analysis functions Data search Search function Searches data that matches specified address pattern data pattern and acknowledge bit condition Analysis result save function Analysis list data can be saved to CSV format files Analysis results save function Analysis list data can be saved to CSV format files SENT Signal Analysis Functions F9 Option SPI Bus Signa
21. ngle shot using all analog channels 25 MPoints 62 5 MPoints standard memory 125 MPoints with M2 option 250 MPoints with M3 option Single shot using half the analog channels 62 5 MPoints 6 25 MPoints standard memory DLM2000 Series History memory and high speed acquisition Capture and replay up to 50 000 acquisitions With the history memory the DLM2000 can capture and replay up to 50 000 individual acquisitions These can be displayed one at a time or as an accumulation Using the search and measurement functions abnormal signals can therefore be quickly isolated analyzed and precisely categorized without needing to carefully configure triggers to capture rare events Together with a maximum continuous acquisition rate of 20 000 waveforms per second which increases to approximately 450 000 in N Single mode the history memory in the DLM2000 enables abnormal signals to be captured without needing to know what makes them different Waveform history gt ibe BY Abnormal signal a Reliable triggering Reliable triggering When just a specific event or abnormal waveform needs to be captured the flexible and reliable triggering of the DLM2000 is the solution The user can combine analog and digital inouts and select the trigger conditions appropriate to the complexity and uniqueness of the event Enhanced triggers Via the Enhanced menu the DLM2000 can be set to trigger for example
22. no time from trigger position Time ms Frame type ID DLC Data CRC presence absence of Ack information Auxiliary analysis functions Data search and field jump functions Analysis result save function Analysis list data can be saved to CSV format files LIN Bus Signal Analysis Functions F4 F6 F7 and F8 Options Applicable bus LIN Rev 1 3 2 0 2 1 Analyzable signals CH1 to CH4 M1 to M2 Bit rate 19 2 kbps 9 6 kbps 4 8 kbps 2 4 kbps 1 2 kbps User defined an arbitrary bit rate from 1000 bps to 200 kbps with resolution of 10 bps LIN bus Trigger modes Break Synch ID DATA ID OR and ERROR trigger Auto setup function Auto setting of bit rate treshold value time axis scale voltage axis scale and display of analysis results Analyzable no of frames 100 000 frames max Analysis results displays Analysis no time from trigger position Time ms ID ID Field Data CheckSum information Auxiliary analysis functions Data search and field jump functions Analysis result save function Analysis list data can be saved to CSV format files USBTMC USB488 USB Test and Measurement Class Ver 1 0 Supported class Ethernet C10 amp C11 Options Connector RJ 45 connector x 1 Ethernet 1O00BASE T 100BASE TX 10BASE T Server FTP HTTP VXI 11 Client FTP SMTP SNTP LPR DHCP DNS Transmission methods Suppo
23. rent probe PBCO50 701929 DC to 50 MHz bandwidth 30 Arms EX52 Attach two 701946 probes For 2ch 350 500 MHz models Current probe PBC 100 701928 DC to 100 MHz bandwidth 30 Arms EX54 Attach four 701946 probes For 4ch 350 500 MHz models Curent prope Ore BE TOT MEE bangin loans 1 Logic probes sold separately Please order the model 701988 701989 accessory logic probes separately Current probe 701931 DC to 2 MHz bandwidth 500 Arms 2 Only one of thes may be selected at a time l 8 Specify this option when using current probes or other differential probes that don t support probe interface Deskew correction signal source 701936 For deskew correction j w n a es ey peers a ae Printer roll paper B9988AE Lot size is 10 rolls 10 meters each a ae oi A is The 701938 probes are not included when this option is selected Carrying case 701964 Also for DL1600 DL1700E Series The 701939 probes are not included when this option is selected 1 For the accessories for 701938 701939 probe various adapters are available Please refer to DL series Accessories DLM is a pending trademark or registered trademark of Yokogawa Electric Corporation Any company s brochure as He eae ae appearing in this document are the registered trademarks or trademarks of their 2 Current probes maximum input current may be limited by the number of probes used at a time Accessory Software Name Model Specific
24. rted services General Specifications Rated supply voltage 100 to 240 VAC Rated supply frequency 50 Hz 60 Hz Maximum power consumption 170 VA External dimensions 226 W x 293 H x 193 D mm when printer cover is closed excluding protrusions Weight Operating temperature range Approx 4 2kg With no options 5 C to 40 C 1 Measured under standard operating conditions after a 30 minute warm up followed by calibration Standard operating conditions Ambient temperature 23 C 5 C Ambient humidity 55 10 RH Error in supply voltage and frequency Within 1 of rating 2 Value in the case of repetitive phenomenon The frequency bandwidth of a single shot phenomenon is the smaller of the two values DC to sampling frequency 2 5 or the frequency bandwidth of the repetitive phenomenon 3 When the input section is shorted the acquisition mode is set to Normal accumulation is OFF and the probe attenuation is set to 1 1 4 Acquisition rate does not vary with an increase or decrease in channels External Dimensions Unit mm a WEEET Model and Suffix Codes Standard Main Unit Accessories
25. ssive probe 701946 10 MQ 10 1 500 MHz 1 3 m G3 Power supply analysis function 4 ch model only FET probe 700939 DC to 900 MHz bandwidth 2 5MQ 1 8pF IGA poe Sy oe function includes G2 Active probe PBA1000 701912 DC to 1 GHz bandwidth 100k 2 0 9pF E15 UART trigger and analysis 4 ch model only 100 1 voltage probe 701944 DC to 400 MHz 1 2 m 1000 Vrms JFZ C SPI trigger and analysis 4 ch model only 100 1 voltage probe 701945 DC to 250 MHz 3 m 1000 Vrms JFB UART C SPI trigger and analysis 4 ch model only Differential probe 701921 DC to 100 MHz bandwidth max 700 V IFAT CAN LIN trigger and analysis 4 ch model only Differential probe 701922 DC to 200 MHz bandwidth max 20 V F5 7 FlexRay trigger and analysis 4 ch model only Differential probe PBDH1000 701924 DC to 1 GHz bandwidth 1MQ max 25 V F67 FlexRay CAN LIN trigger and analysis 4 ch model only Differential probe 701926 DC to 50 MHz bandwidth 5000 Vrms 7000 Vpeak JETT CAN CAN FD LIN trigger and analysis 4 ch model only Differential probe PBDHO1 50 701927 DC to 150 MHz bandwidth max 1400V ER 7 T aA FD LIN trigger and analysis Differential probe 700924 DC to 100 MHz bandwidth max 1400 V FO SENT analvss ch model only Differential probe 700925 DC to 15 MHz bandwidth max 500 V EX22 Attach two 701946 probes For 2ch 200 MHz models Differential probe 701920 DC to 500 MHz bandwidth max 12 V 7EX24 Attach four 701946 probes For 4ch 200 MHz models Cur
26. t of an abnormality which could be on the same or different input channels They also make it possible to view and compare the details and timings of different serial buses which are running at different speeds Advanced waveform search functions Quickly find and mark abnormal signals in long and multiple waveform acquisitions Single waveform acquisitions of up to 250 MPoints can be searched using various criteria such as edges state patterns pulse widths and serial bus patterns optional The history memory can be searched to find matching criteria in up to 50 000 acquisitions Automatic parameter measurement and Statistical analysis 30 waveform parameters from a total of 29 different types can be displayed simultaneously with a high update rate These include maximum minimum peak to peak pulse width period frequency rise and fall times and the delay between channels The statistics of repetitively measured parameters can also be displayed such as the mean maximum minimum and standard deviation Additionally the Go NO GO function can be used to test the results of parameter measurements trigger conditions and other criteria and automatically save or print data send an e mail etc Features and benefits YOKOGAWA gt Parameter trend and histogram displays To observe the fluctuations of measured parameters it is possible to display them as trends Period to period changes
27. t to be nearer the eyes of the user and keeps the footprint on the bench to a minimum The intuitive controls are laid out so that a user can see at a glance what channels and features are switched on and quickly make the measurements that are needed Easy to configure 8 4 inch display Users can automatically or manually split the display to separate individual channel waveforms while maintaining their full resolution and dynamic range It is therefore easy to see the details of all signals regardless of the number of channels in use YOROGAWA Diii Triisa Soe a Me T se T Horizontal Position and Scale Knob eee Dedicated Zoom Keys Four Direction Selector Button Select key moves the cursor up down left right Jog Shuttle and Rotary Knob Logic input connector Trigger Control Keys and Level Knob Vertical Position and Scale Large screen in a compact body Snapshot key to freeze traces on screen Graphical on line help key Built in manual Fast and flexible Fast and flexible Flexible MSO input Choose to capture a mix of analog and more digital signals With a push of a button channel 4 converts into 8 digital inputs and the DLM2000 becomes a mixed signal oscilloscope This makes it possible to view 3 analog and 8 digital signals simultaneously and view more control and logic signals Digital channels can also be used to analyze I C SPI and UART serial buses which keep the analog channels available for o
28. ther signals ScopeCORE fast data processing The hardware optimized architecture and dedicated ScopeCORE IC in the DLM2000 enable measurements and signal processing to be carried out in real time This means that turning on more channels does not affect the waveform acquisition rate and measurements are always performed at high speed DLM2000 series Lineup ped no i YOKOGAWA 3 channels analog 8 bit logic The fast ScopeCORE internal processor Analog Maximum Maximum record length in single i No of channels oe Bandwidth sample rate measurement mode and with interleave on 200 MHz 4 analog or 3 analog plus 8 digital 250 MPoints with M3 option The flexibility of longer memory The flexibility of longer memory Long waveform memory Up to 250 MPoints The two advantages of a long waveform memory are the abilities to capture for long periods of time and to maintain high sample rates and hence higher effective measuring bandwidths for all time base settings With the maximum memory installed M3 option in single shot mode a 10 kHz signal lasting for more than one hour can be captured The same memory can capture a 200 millisecond signal at a sample rate of 1 25GS s sample rate Maximum memory per channel 6 25 MPoints 12 5 MPoints with M2 option 25 MPoints with M3 option Continuous 25 MPoints Standard memory 62 5 MPoints with M2 option 125 MPoints with M3 option Si
29. tion Auxiliary analysis functions Data search and trend functions Analysis result save function Analysis list data can be saved to CSV format files UART Bus Signal Analysis Functions F1 amp F3 Options Analysis list data and trend data can be saved to CSV format files Analysis result save function Bit rate 1200 bps 2400 bps 4800 bps 9600 bps 19200 bps user defined an arbitrary bit rate from 1 k to 1 Mbps with resolution of 100 bps GP IB C1 amp C11 Options Electromechanical specifications Conforms to IEEE std 488 1978 JIS C 1901 1987 Analyzable signals CH1 to CH4 logic input or M1 to M2 Protocol Conforms to IEEE std 488 2 1987 Auxiliary Input Data format Select a data format from the following 8 bit Non Parity 7 bit Data Parity 8 bit Parity UART Trigger modes Every Data Data Error Framing Parity Rear panel I O signal External trigger input DLM20x2 front panel external trigger output GO NOGO output video output Auto setup function Auto setting of bit rate threshold value time axis scale voltage axis scale and display of analysis results Probe interface terminal front panel 4 terminals DLM20x4 2 terminals DLM20x2 Probe power terminal rear panel 2 terminals P2 option 4 terminals P4 option Analyzable no of frames 300 000 frames max Analysis results displays Analysis no
30. tures and more that an engineer requires In an 01001100 0127 Foresight Users experience a short learning curve thanks to an intuitive man machine interface that is easy to use Keeping in touch with users has ensured that any Yokogawa product introduced to the market has been developed with their needs in mind Why choose Yokogawa YOKOGAWA Why choose Yokogawa Our passion for measurement Yokogawa believes that precise and effective measurement lies at the heart of successful innovation and has focused its own R amp D on providing the tools that researchers and engineers need to address their challenges both great and small Our heritage Yokogawa has been developing measurement solutions for almost 100 years consistently finding new ways to give R amp D teams the tools they need to gain the best insights from their measurement strategies Our oscilloscope design has been led by customers looking for ease of use and functionality Our commitment Yokogawa takes pride in its reputation for quality both in the products we deliver often adding new features in response to specific client requests and the level of service and advice we provide to our clients helping to devise measurement strategies for even the most challenging environments Precise control DLM2000 Series Precise control Easy to use portrait design The large display of a DLM2000 is located above the controls this enables i
31. using voltage as X axis and current as Y axis is possible M2 option Repeat 12 5 MPoints Single 62 5 MPoints M3 option Repeat 25 MPoints Single 125 MPoints Triggers Trigger modes Auto Auto Level Normal Single N Single Trigger type trigger Source A triggers Edge CH1 to CH4 Logic EXT LINE Edge OR CH1 to CH4 Edge Qualified CH1 to CH4 Logic EXT State CH1 to CH4 Logic Pulse width CH1 to CH4 Logic EXT State width CH1 to CH4 Logic TV CH1 to CH4 Harmonic analysis Basic comparison is possible with following standard Harmonic emission standard IEC61000 3 2 edition 2 2 EN61000 3 2 2000 IEC61000 4 7 edition 2 Joule integral Joule integral l2t waveform display automatic measurement and statistical analysis is possible Power Measurement Automated measurement of power parameters for up to two pairs of voltage and current waveforms Values can be statistically processed and calculated Measurement parameters Urms Unm Udc Urmn Uac U pk U pk Up p Irms Imn Idc Irmn lac l pk I pk lp p P S Q Z A Wo Wp Wp Abs Wp q q q Abs q Avg Freq voltage current YOKOGAWA 15 I2C Bus Signal Analysis Functions F2 amp F3 Options FlexRay Bus Signal Analysis Functions F5 F6 and F8 Options Applicable bus FlexRay Protocol Version2 1 Applicable bus 1C bus Bus transfer rate 3 4 Mbit s max Address mode 7 bit 10 bit SM bus Co
32. with shifted timing such as non standard video signals and motor reference position pulses and drive pulses to be easily made Input signal B T Trigger Triggering on the 7th edge of the B input signal Features and benefits Features and benefits Capable measurement and analysis DLM2000 Series The DLM2000 is packed with advanced capture and analysis features to provide quick and comprehensive answers Real time filters and post processed digital filters The DLM2000 has two types of filter Real time input filters with cut offs from 8 kHz to 200 MHz are selectable for each channel and the filtered data is stored in the internal memory Inout waveforms can also be filtered using a digital IIR filter using the mathematics MATH function This method enables the input and filtered waveforms to be simultaneously displayed and compared High and low pass filters from 0 01 Hz to 500 MHz are selectable with a high level of precision The image shows the snapshot function which freezes traces on the screen to compare old and new acquisitions Two fully independent Zoom windows Combined with the advanced search and cursor parameter measurement capabilities the two zoom windows enable users for example to see the waveform detail of two parts of the acquisition which can be separated by a long time period It is thus possible to quickly find measure and analyze the details of the cause and effec