Home
        OAKTON hygrothermograph manual
         Contents
1.   the groove     See figure O    8  Install the new chart paper on the  Chart Roll     e Load the new chart paper onto the  Chart Roll  making sure that the slot  on the chart paper roll fits into the  ridge on the lower disk     e Make sure that the chart paper is  loaded correctly  The side of the  paper with the oval shaped sprocket  holes should be at the upper disk   top   while the side with the round  sprocket holes should be at the lower  disk  bottom      e Replace the upper plastic disk of the  Chart Roll by turning it clockwise     See figure Q    Take up roll        upper disk  J  stopper  slot         cardboard  tube          15    9  Install the Chart Roll with the new  chart paper back into the chart  assembly     e Place the pointed spindle tip  pro   truding through the upper disk  into  the receptacle at the left  on the  upper part of the assembly   Push to  the left  and reinsert the right end of  the starting roll into the groove  on  the lower part of the assembly   Make  sure that the roll is securely installed  and can pivot easily     See figure     e Make sure paper is positioned so  it will run in a counterclockwise  direction around the Chart Roll     See figure  1     10  Pull out the chart paper about 6  inches  Draw the paper around the  Timing Roll  Make sure the sprock   et holes in the paper are placed  correctly over the sprockets on the  Timing Roll     See figure 0    11  Tape the end of the chart paper to  the Take Up Roll  Remove any oth
2.  08368 20   2  or 3 speed 14 to 122  F 7 day 2 hour 100  WD  08368 30   2 or3 speed    10 to 50  C 1 day 15 minute 100  WD  08368 40   2 or3 speed    10 to 50  C 7 day 2 hour 100  WD 08368 22   3 speed 14 to 122  F 32 day 6 hour 25  WD 08368 42   3 speed  10 to 50  C 32 day 6 hour 25  WD 37250 60   Long cycle  20 to 50  C 1 month 2 hour 1 roll  WD 37250 62   Long cycle  20 to 50  C 3 month 2 hour 1 roll  Extra Pens    For use with all OAKTON hygrothermographs   WD 08368 75 Blue pens  6 pack  WD 08368 70 Red pens  6 pack    Axial Fan    WD 37250 50 Optional axial fan  Use with 2 speed  3 speed and long cycle models  for more uniform sample measurements in fluctuating environments  110 VAC    WD 01578 02 Transformer  Use to operate axial fan at 220 VAC  100 W    26    9  Appendix 1  Readings in areas with high humidity       The specially processed human hair bundle  such as the one used in this instrument   is the best sensor for ambient humidity measurements in terms of linearity   repeatability  and temperature compensation  Human hair will elongate linearly as  humidity increases or decreases in the range of 5 to 90  RH     However  one of the characteristics of human hair is that exposure to humidity  levels above 90 RH causes it to undergo a change of state  This condition is called  the    saturated state     and occurs with even short exposure to high humidity levels   When the hair bundle reaches the saturated state  it contracts and causes a 7 to 9   downward shift in
3.  L  Pen tip for temperature 6  Protective cover drum is completely lowered over the   2  Cylinder drum with quartz clock 7  Pen arms center shaft    3  Pen tip for humidity 8  Pen guide See figure Q   4  Paper holder 3  Replace the hold down nut on the top of    the center shaft   5  Pen lift lever                4  Hygrothermograph operation    ES Powering your hygrothermograph    Minidrum models  08369 50 and  70         These models require one AA battery  The battery  compartment is located on top of the cylinder  drum  To open the battery compartment  slide the  cover of the battery compartment to the side  Make  sure that the   end of the battery is facing up     Spring wound  2 speed model  37250 10         This model is spring wound  requires no batteries    The spring winding shaft is located on top of the  cylinder drum  To wind up the unit  turn the shaft  counterclockwise until it is too stiff to move   any further  This hygrothermograph will make a     clicking    sound while running  The drum will  maintain an even rotation speed even when the  spring is winding down     3 speed model  37250 00         This model requires two C batteries  The battery  compartment is located on top of the cylinder  drum  To open the battery compartment  pull the  silver knob on top of the battery compartment  away from the center of the cylinder drum  Note  polarity on the lid of the battery compartment     This model features a battery indicator  To check  the battery  press the
4.  best results  we recommend you use our chart paper  See the    Accessories     section on page 26 to order extra chart paper rolls     When the chart paper is nearing the end of its cycle  a red stripe will appear on the  lower edge of the chart paper  The stripe means there are only three days left  and  that the paper should be changed by that time     Timing roll knob    To change the chart paper     1  Remove the protective cover  Pull the  pen lift lever towards you so the pen  tips do not touch the paper  Write the  time and date on the chart paper     2  Wind the old chart paper onto the  Take Up Roll by turning the Timing  Roll knob clockwise  in the direction  of the arrow      See figure Q             3  Release the right and left chart  assembly locks located on the base  of the instrument near the chart  drive mechanism     See figure O    4  Pull the chart assembly back using  the handle located behind the chart  assembly  To prevent the unit from  tipping over  hold the base of the  instrument with your other hand   Use the handle to keep the chart  assembly in a horizontal position     See figure O       5  Remove the empty cardboard tube  from the Chart Roll  see sticker on  top of chart assembly for position      See figure  D    e Push the Chart Roll towards the top  of the chart assembly  which removes  it from the groove at the base of the  assembly     e Pull the Chart Roll out from the  bottom     e Remove the upper plastic disk of  the Chart Roll by tu
5.  it from the post     See figure  B     3  Using two screwdrivers  lift the gear  currently installed on the bottom of  the cylinder drum     See figure  c   4  Insert the other gear onto the oppo   site shaft on the bottom of the drum     e The one day gear slides on the shaft  marked    24 H   Z   22     This gear has  22 teeth and a larger diameter than the  7 day gear     e The seven day gear slides onto the  shaft marked 168H   Z   18  This gear  has 18 teeth     5  Place the gear not in use over the  post at the corner of the thermohy   grometer   s base for storage     6  Install the cylinder drum onto  the cylinder shaft  See page 7 for  directions     20       Three speed hygrothermograph 37250 00       This model lets you select between a  1 7  and 32 day drum rotation  When  you receive your hygrothermograph  from the factory  it is set for a 7 day  rotation           o0  See figure    _    IMPORTANT  make sure you have the  appropriate chart paper for the rotation  speed you select  See page 26 to order  more chart paper                             1  To switch from 7 day operation to  1 day operation  flip the switch on  top of the cylinder drum to 1D     See figure Q   2  To switch from 7 day operation to  32 day operation  flip the switch on  top of the cylinder drum to 32D     See figure Q             21    5  Installing optional axial fan    NOTE  The axial fan is not for use with minidrum models 08369 50 and 08369 70     The external draft fan  optional  att
6.  red check button next to the  battery compartment  Note the color of the LED  indicator     e A green LED means the batteries are in  good condition    e A red LED means the batteries need to  be replaced          Top of Drum             Close up of battery indicator                I    BATTERY                   Long cycle model  37250 20         This model requires one C battery  The battery  compartment is located on top of the cylinder  drum  To open the battery compartment  flip  open the lid on top of the cylinder drum  Note  polarity inside the battery compartment     This model features a battery indicator  To check  the battery  press the red check button next to the  battery compartment  The power indicator has  green  white and red zones     e the green zone means the battery is in  good condition    e The white zone means the battery is low  and will have to be replaced soon    e The red zone means the battery is very  low and should be replaced immediately           Close up of battery indicator    AN    BATTERY    E Cartridge pens    To operate the pens        1  Pull the pen lift lever towards you to move the  temperature and humidity pens to their resting  position     See figure Q  2  Remove the pen caps by rotating them counter  clockwise     3  Push the pen lift lever away from you until the  temperature and humidity pens are touching the  paper on the cylinder drum     NOTE  Do not touch the pen tips  because skin oils  will prevent the ink from flowing sm
7.  the humidity indications  If the humidity then stays below 80    the hair bundle will elongate linearly for approximately two weeks  After this time  period  the hair bundle is in its dry state     The hair bundle has two curves  depending on the state of the hair bundle  These  characteristics can be taken into account when using the instrument  The instrument  is calibrated at the factory with the hair bundle in the dry state  Generally  the  instrument is used indoors with humidity levels below 90      When used in air conditioned areas or other areas of moderate humidity  you do not  need to adjust the instrument     How to use the instrument in high humidity areas     For use above 85  RH  turn the humidity fine adjustment screw counterclockwise  to raise the humidity reading by 7 to 9   Readings will be within specifications over  the entire range when the instrument is exposed to high humidity conditions  85 to  100   at least once every two weeks     Note any calibration change either on the chart paper or on the instrument itself   This way you can change the instrument back to its original calibration if required     27    
8. 10  Warranty    OAKTON warrants this instrument to be free from significant deviations in material  and workmanship for a period of one year from date of purchase  If repair or  adjustment is necessary and has not been the result of abuse or misuse within   the warrantied time period  please return   freight prepaid   and correction will   be made without charge  OAKTON alone will determine if the product problem   is due to deviations or customer misuse     Out of warranty products will be repaired on a charge basis     11  Return of items    Authorization must be obtained from our Customer Service Department before  returning items for any reason  When applying for authorization  please include  data regarding the reason the items are to be returned  For your protection  items  must be carefully packed to prevent damage in shipment and insured against possi   ble damage or loss  We will not be responsible for damage resulting from careless or  insufficient packing  A restocking charge will be made on all unauthorized returns     NOTE  We reserve the right to make improvements in design  construction  and  appearance of products without notice     00703 07    OAKION       OPERATING INSTRUCTIONS    OAKTON    Models 08369 50   70  37250 00   10   20    grothermographs                                                                                                    Printed in the U S A  1 00    Write down the name and information of your OAKTON distributor here     Table of Co
9. F  41  C    1 8  F  bimetallic strip  1  C   2  F  GE al WLL    1 day  26 hr  or  7 day  172 hr     Spring wound  13 3 W x 11 5 H x 5 1 D  9 Ibs                      Three speed  hygrothermograph    37250 00  5 to 90  RH     3  from 10 to 90  RH    5  from 5 to 10  RH    0 to 100  RH  human hair bundle  1  RH   10 to 50  C  14 to 122  F   41  C   41 8  F  bimetallic strip  T  C   2  F  Gye ial xe US IL     1 day  26 hr   7 day  172 hr   or 32 day  810 hr     Two C batteries  included   13 3 W x 11 5 H x 5 1 D  9 Ibs       Long cycle  hygrothermograph    37250 20  5 to 90  RH     3  from 10 to 90  RH    5  from 5 to 10  RH    0 to 100  RH  human hair bundle  1  RH   20 to 50  C   1  C  bimetallic strip  1  C    6 6 H x 114  L  1 month   6 6 H x 300  L  3 month     1 month  38 day max  or  3 month  100 day max     One C battery  included   13 3 W x 11 5 H x 5 1 D  9 Ibs    25    8  Ordering information    Additional Hygrothermographs       WD 08369 70 Minidrum hygrothermograph   WD 08369 70 Domed minidrum hygrothermograph  WD 37250 10 Spring wound  two speed hygrothermograph  WD 37250 00 Three speed hygrothermograph  WD 37250 20 Long cycle hygrothermograph    Extra Chart Paper                                        Model Use with Temperature   Rotation Increments   Quantity   number model type range pack  WD 08369 60   Mini drum 22 to 104  F 7 day 2 hour 100  WD 08369 55   Mini drum  6 to 40  C 7 day 2 hour 100  WD  08368 10   2  or 3 speed 14 to 122  F 1 day 15 minute 100  WD 
10. aches to the protective cover  It improves instru   ment response and accuracy by directly pulling surrounding air into the instrument   Use the fan in areas where ventilation is poor     If you want to order an axial fan  see the    Accessories    section on page 26     To attach the fan     1  Install the fan on the single  column of five vents located on  the protective cover     See figure Q   2  Position fan so that the spring clips  on the fan go into the upper vent  slots on the hygrothermograph  cover     See figure      3  Let the fan slide down until it stops   Make sure both spring clips are fully  engaged in the cover     See figure O   4  Position the instrument so the  flow of air from the fan will not  be blocked              5  Plug the power cord into an 110 to  120 volt  50 60 Hz outlet  For 220 V  operation  order optional 220 V  transformer  sold separately in     Accessories    section on page 26                     22    1  Do not use your hygrothermograph     4     e in direct sunlight   e in temperatures below  20  C or above 50  C   e near ovens  stoves  or other heating equipment   e near harsh chemicals such as paint thinner or ammonia   e in dusty or wet environments   e within magnetic fields   e in areas with strong vibrations  such as loudspeakers or motors    For best results  use only the pens and chart paper listed under the    Accessories     section of this manual  page 26   We cannot guarantee results with other pens  and other paper     D
11. en the time printed on the  chart paper and pen tips line up about  two hours past the present time     4  Turn the cylinder counterclockwise to  the correct present time     5  To take up slack in the chart paper   turn the chart paper roll counterclock   wise until the paper is taut  Be sure not  to offset the alignment of present time  and pen tips     See figure  E        EZ Chart paper    NOTE  See pages 13 17 for directions on  changing the chart paper on long cycle  model 37250 20     Your cylinder drum comes with one sheet  of chart paper already attached  See the     Accessories    section on page 26 to order  extra chart paper     To replace chart paper  for models 08369 50   08369 50  37250 10  37250 00        1  Wind the new chart onto the cylinder so  that the two ends of the paper overlap     See figure Q  2  Secure this overlapping section with the  paper holder     See figure O  3  Make sure the markings on the paper  coincide where the overlap occurs     4  Follow directions on bottom of page 11  to set the correct time     12                To replace chart paper for long cycle model 37250 20        The long cycle hygrothermograph comes with one roll of paper installed  As the  unit records  paper moves from the Chart Roll to the Take Up Roll  When the paper is  completely used  you need to replace the chart paper     NOTE  do not discard cardboard tube from the original chart paper roll  You will  need this cardboard tube for use on the Take Up Roll     For
12. er  paper that might still be attached  to the Take Up Roll  Be sure the  paper is straight and even so that it  tracks properly     See figure Q    16                   Tipe    i     J    5    Tl  DEEDS  g  R     ry  A       l  iy       12  Advance the chart with the timing  knob     13  Return the chart assembly upright  to its original position     See figure       14  Tighten the right and left assembly  locks to hold chart assembly in  place     See figure Q  15  To set time  see bottom of page 11     16  Make sure to check that pen car   tridges and batteries are still fresh     Notes       Be sure to check at least once a month  to guarantee accuracy  since the instru   ment operates over a long period of  time  30 or 90 days      Also make sure to periodically check  that the pens are operating correctly     Vibrations could loosen the chart paper  in about 60 days  The paper should be  tightened for accurate time readings               XJ Calibration adjustment    Your hygrothermograph is factory calibrated  However  vibration during shipment  may make a slight readjustment necessary     To adjust calibration        1  Compare the hygrothermograph reading to another temperature  humidity  instrument known to be accurate     2  Raise and lower the reading by turning the adjustment screws See drawings on  these two pages for location of the adjustment screws     Calibration adjustment screw locations       Minidrum hygrothermograph   model 08369 70           On the min
13. eturn Of IteMS       sssssssssnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnn nnmnnn nnmnnn nnmnnn nnana 28  3    1  Introduction    Thank you for selecting an OAKTON thermohygrometer  These precision instru   ments are useful for monitoring temperature and relative humidity conditions over  time  They create a permanent chart recording of these measurements for your  records  Their quartz controlled drive maintains an even drum speed even when  batteries are weak  or when the spring winding runs down      Typical applications include general weather measurements  agriculture   horticulture  monitoring the condition of test laboratories  humidity chambers   computer rooms  precision machinery rooms  food storage facilities  warehouses   museums  bank vaults  film and book storage facilities     You have one of five available models     Minidrum hygrothermograph  Model number  08369 70  Special features    e Compact size   e Records on 7 day rotation    e Protective acrylic cover locks  into place                      Domed minidrum  hygrothermograph    Model number  08369 50    Special features        e Our most compact model    e Records on 7 day rotation             e Protective glass dome                                                                                                                                        Spring wound  two speed  hygrothermograph    Model number  37250 10  Special features     e Spring wound     no batteries needed     e Records on 1  or 7 day 
14. idrum hygrothermograph  model  the calibration screws are locat   ed behind the two stickers on the back  of the beige case protecting the pen  mechanism     See figure Q  Remove the stickers to reach the cali   bration adjustment screws     See figure O                                                                                                    18       Domed minidrum hygrothermograph   model 08369 50     All other models   models 37250 00   10   20                 Temperature  calibration  adjustment  screw       Humidity  calibration  adjustment  screw                      Temperature calibration  adjustment    Humidity  calibration  adjustment  screw    19    EJ Changing drum rotation length  NOTE  Minidrum models 08369 50 and 08369 70 operate on a 7 day rotation only     Long cycle model 37250 20 accepts either 1 month or 3 month chart rolls   it is not  necessary to adjust drum speed     Spring wound  two speed hygrothermograph 37250 10       This model lets you select between a   1 and 7 day drum rotation  When you  receive your hygrothermograph from the  factory  it is set for a 7 day rotation     IMPORTANT  make sure you have the  appropriate chart paper for the rotation  speed you select  See page 26 to order  more chart paper     1  Look at the bottom of the cylinder  drum  You will see notches for two  gears  Only one gear will be in place     See figure Q   2  The alternate gear is located on a post  at the corner of the thermohygrome   ter   s base  Remove
15. ntents    1  Introd  ctionN Bee es eee 4 5  2  COMPONENKS     ccceeeeeeeeeeeeeeeeeeeeeeneeseeeeeeseeeeaeeenaeeeseaneeeeeeeeseeeeeseeeeeceneeeeeee 6  3  Getting Started       ccccceecceeeeeeeeneeeeeeeeeeeeaeeeeeeeeeneoeseeeeeeeeenaeeeeeeeeseaseeeeeeesenans 7   Si Tp a Cin Gs cis veel sana stewed evades ess esau esta KERE Ueatted ence ae taeertalcensoiacdeavdervarksuctuasdayeaiaieies 7       3 2 Attaching the cylinder drum    4  Hygrothermograph operation       4 1 Powering your hygrothermograp himasin EEEE 8 9  4 2 Cartridge Pens sisssrsssssissssesrsierssrsesssiisneiosetessstdstsoE rE nE E  bESE STEES ESEESE EESTE BARM SASEN ESAE SETENE 10  43 Setting Heera np iiare eiiiai EEEE ENEAS AS RELAT ENRE ARE ie 1  A A CARE papelpin eaea ia neler a VaR EARE 12 17  4 5 Calibration adjustment scis lt sczsiiessecasatantecssence hoc cadsas gesdsdeack St reias i ea hierati ieie 18 19  4 6 Changing drum rotation length xc  cciinnc ated aati aren S 20 21  5  Installing optional axial fan            ccececeseeseeeeeeeeeeaneeeeeeeeeeeeeeeeeeeeeeneneeeeee 22  6  Precautions        0  c  eee 23  7  SPCCHFICATIONS       ceeceeeeeeeeeeeeeeeeeeeeneneeeeeeeeneeeseeeeeeeeenaseeeeeaeseoanaeeeeeesnons 24 25  8  Ordering information         cccccesseeeeeeeeeeeeeeeeeeeeeeeeseeeeeeaeenaeeeeeeaesneeeeeeeenans 26  9  Appendix 1  Readings in areas with high humidity              csssseseeee 27  10  Warranty      ccccccseeeeeeeeeeenseeeeneeseeseeeeeeneueeseanegeeeneageeneagessaeeaeeeneenseneneennnas 28  11  R
16. o not repair the instrument yourself  Should repairs be necessary  please return  the instrument to place of purchase  See the back cover of this manual for infor   mation on instrument Warranty and Return of Items     If your instrument will not be used for long periods of time  make sure to replace  the pen caps and remove the battery     23    7  Specifications    Model    Model number  Humidity range    Humidity accuracy    Humidity chart range  Humidity sensor   Humidity chart graduations  Temperature range  Temperature accuracy  Temperature sensor   Temp  chart graduations    Chart size    Chart rotation    Power  Dimensions    Shipping weight    24       Minidrum  hygrothermograph    08369 70  5 to 90  RH     5  from 10 to 90  RH    7  from 5 to 10  RH    5 to 100  RH  human hair bundle  5  RH     6 to 40  C   22 to 104  F  42  C   43 6  F  bimetallic strip  2  C   2  F  SHO  lal ey INL     7 day  172 hr     One AA battery  included   6 W x 7 4 H x 3 9 D  4 lbs       Domed minidrum  hygrothermograph    W08369 50  5 to 90  RH     5  from 10 to 90  RH    7  from 5 to 10  RH    5 to 100  RH  human hair bundle  5  RH     6 to 40  C   22 to 104  F  42  C   43 6  F  bimetallic strip  2  C   2  F  3 6 H x 8 1 L    7 day  172 hr     One AA battery  included   7 5 H x 5  dia  4 lbs                Spring wound  two speed    hygrothermograph  37250 10  5 to 90  RH     3  from 10 to 90  RH    5  from 5 to 10  RH    0 to 100  RH  human hair bundle  1  RH   10 to 50  C   14 to 122  
17. oothly     To replace the pens        The pen tips can be used for up to a year  but we  recommend that you replace them at least every six  months  Conditions such as high temperature or low  humidity may shorten pen life to three months     See the    Accessories    section on page 26 to order  extra pens     1  Hold the pen arm in your right hand and pull  the pen cartridge off with your left hand  If it is  difficult to pull off  use a small pliers     See figure O  2  To attach the replacement cartridge pen  insert it  over the arm until it is fully attached     10                      EE  Setting the time    Models 08369 50   08369 50  37250 10  37250 00        1  Pull the pen lift lever towards you to  move the temperature and humidity  pens to their resting position     See figure Q  2  Turn the cylinder drum one full  rotation clockwise     See figure      3  Turn the cylinder drum counterclock   wise until the pen tips align with the  correct time of day or night  Numbers  indicating time are located at the top  and bottom of the chart paper     NOTE  Make sure that the numbers on  the graph are right side up  Change the  chart paper when the cylinder has com   pleted one rotation     Long cycle model  37250 20  only        1  Pull the pen lift lever slightly towards  you to move the temperature and  humidity pens away from the chart     See figure QO    2  Turn the chart assembly knob   located on top of the chart assembly   clockwise     See figure      3  Stop wh
18. rning it  counter clockwise     e Remove the cardboard tube from the  metal spindle  spindle is inside the  cardboard tube   Save this cardboard  tube for step 7     See figure  E     6  Remove the Take Up Roll with  the old chart paper from the chart  assembly  see sticker on top of  chart assembly for position      See figure Q   e Push the Take Up Roll toward the  top of the chart assembly  which  removes it from the groove at the  base of the assembly     e Pull the Take Up Roll out from the  bottom     e Remove the upper plastic disk of  the Take Up Roll by turning it  counter clockwise     e Remove the used chart paper   complete with inner cardboard tube   from the metal spindle spindle is  inside the cardboard tube      See figure  E     14          upper disk    cardboard  tube                Take up roll    7  Replace the Take Up Roll with the  empty cardboard tube into the  chart assembly     e Place the empty cardboard tube   from step 5  back over the metal  spindle of the Take Up Roll  Make sure  that the slot on the Take Up Roll fits  into the ridge on the lower disk     e Replace the upper plastic disk of the  Take Up Roll by turning it clockwise     See figure Q   e Place the pointed spindle tip   protruding through the upper disk   into the receptacle at the left  on the  upper part of the assembly   Push to  the left  and reinsert the right end of  the take up cylinder into the bottom  groove  You may have to turn the  cylinder slightly until it slides into
19. rotation    e Protective acrylic cover locks  into place    Three speed hygrothermograph  Model number  37250 00  Special features     e Records on 1  or 7   or 32 day  rotation    e Protective acrylic cover locks  into place    Long cycle hygrothermograph  Model number  37250 20   Special features    e Records on 1 or 3 month rotation    e Protective acrylic cover locks  into place    2  Components 3  Getting started    All OAKTON hygrothermographs have these same features     EES Unpacking    1  Remove all components from the packing material     2  Carefully remove the acrylic or glass protective cover from the hygrothermo   graph  If your model has a handle  turn the carrying handle to the left until it  stops  then lift up the protective cover     3  Carefully remove the styrofoam piece located beneath the pen arm mechanism     A          Remove the white paper that secures the humidity pen     5  Lift the metal pen holding fitting up from the temperature pen until it is removed  from the pen guide        6  Move the pen lift lever towards you to pull the temperature and humidity pens to  their resting position     NOTE  Keep all original packing in case you need to transport the instrument                     EQ  Attaching the cylinder drum    NOTE  Long cycle model 37250 20 already  has an attached cylinder drum           1  Unscrew the hold down nut from the  center shaft        2  Carefully insert the cylinder drum onto  the center shaft  Make sure the cylinder      
    
Download Pdf Manuals
 
 
    
Related Search
    
Related Contents
Massive Wall light 16251/31/10  QRS (QoS Routing Simulator) Version 2.1 User's Manual  D - Commercial Floor Machine Parts Solutions are here!  MD3505 emb.indd - CONRAD Produktinfo.  Roadmaster MP3/WMA User's Manual  Table of Contents - OneStart  Prova - Idecan  MONTE-CHARGE - WEP Weisshaupt  Kondator 652-1424 mounting kit  Eureka! Tents Backcountry 1 User's Manual    Copyright © All rights reserved. 