Home

AlcoMate Prestige (Model AL6000) USER`S MANUAL

image

Contents

1. AlcoMate Prestige Model AL6000 USER S MANUAL The AlcoMate Prestige is intended to measure alcohol in human breath Readings obtained by this device are used in the diagnosis of alcohol intoxication The accuracy of this device has been established at a Blood Alcohol Concentration of 0 008 and 0 032 Accuracy at a Blood Alcohol Concentration greater than 0 032 has not been evaluated Accuracy testing was completed according to DOT regulations The AlcoMate Prestige Model AL6000 analyzes a breath sample to detect alcohol in the lungs This reading is then converted to blood alcohol concentration BAC The AlcoMate Prestige uses an advanced oxide semiconductor sensor to detect alcohol The Prestige is the first and only breathalyzer that features Pre Calibrated Alcohol Sensor Modules to eliminate the need for recalibration Just snap out snap in and you re ready to go Digital LED Display 3 Decimal Digit ou el BAC Reading with Mouthpiece Attached WARN Indicator BATT LOW Indicator READY Indicator Device Operation If you are using the Prestige for the first time please install two 1 5V AA batteries included and attach a mouthpiece to the unit 1 Instructions a Press and hold the Power Switch to turn the unit on The cumulative test count will be displayed as you hold the Power Switch down b Release the Power Switch There will be a countdown from est to Ee as the unit warms up c Secure your lips around
2. eneral Information about DUI DWI The legal limit for intoxication DUI DWI has recently been lowered from 0 10 to 0 080 BAC across most of the USA According to the American Medical Association AMA a BAC of 0 050 is enough to impair the ability to drive You should never operate a motor vehicle when you are impaired by the consumption of alcohol drugs or lack of proper rest Warnings a Wait at least 20 minutes after eating or drinking before testing Alcohol remaining in the mouth or even excess saliva may interfere with testing This includes alcohol based mouthwash or similar products Avoid testing in high wind or enclosed spaces Poor ventilation may also lengthen the time required for warm up between tests Do not blow smoke saliva or other contaminants into the mouthpiece as the sensor may become damaged Do NOT use the AlcoMate Prestige as a tool to drink and drive Keep out of extreme temperatures and out of the reach of children To maintain accuracy we recommend replacing the alcohol sensor module inside the AlcoMate Prestige approximately every 300 tests or if damaged by improper testing or handling WARNING DISPLAY Flo If you see Flo in the display the breath sample was not strong enough for measurement Please try again and make sure the breath is strong and steady during sampling WARNING DISPLAY Cal If you see Cal in the display the unit requires recalibration or maintenance Plea
3. se use the contact information at the end of this document WARNING DISPLAY Bat If you see Bat in the display the battery power is low Please replace the batteries Remove the battery cover at the bottom of the unit and replace the old batteries with two new 1 5V AA batteries _ Make sure to properly align the batteries by looking at the marks Specifications 0 000 to 0 400 125 height X 58 width X 41 thickness 117g Ooo O Sensor Modules Sensor modules should be replaced approx 200 300 tests or if damaged by improper testing or handling Warranty The manufacturer warrants the product to be free from defects in workmanship or material excluding calibration under normal use for six 6 months from the date of purchase The manufacturer s obligation under this warranty is limited to replacing adjusting or repairing the unit if returned along with proof of purchase This warranty is void if the unit has been tampered with or maliciously damaged Recalibration amp Service For recalibration and technical support please contact us at American Screening Corporation Complete Drug Testing Solutions 6658 Youree Dr Ste 180 PMB 404 Shreveport LA 71105 Ph 316 22 4994 Paxi518 227 4996 Toll Free 1 866 526 2873 www americanscreeningcorp com
4. the mouthpiece and blow sharply and steadily into the unit for about 4 seconds You will hear two a when the device is done taking the breath sample You will see the digit moving left to right in the display as the device analyzes the sample d When analysis is complete you will see the blood alcohol concentration BAC results displayed c The results will be shown for about 15 seconds The unit will then shut itself off d If the device cannot take a successful sample you will see Gut in the display and the unit will shut itself off 2 Understanding Your Results BAC Reading Note The AlcoMate Prestige uses a 3 Decimal Digit high precision display Rather than a reading such as 0 08 the Prestige will display 080 intake warnings intake warnings 050 and over Warning range Ability to drive may Flashing WARN be impaired indicator and repeating beep Drinking alcohol in ANY amount may impair the ability to drive There is no safe level of alcohol intake before driving Don t drink and drive Note The maximum reading the device can show is 0 400 A true BAC of 0 400 or above may result in the onset of coma or even death by respiratory arrest A reading of 0 400 may indicate an acutely dangerous condition but may also indicate a sampling error such as liquid alcohol blown into the device Please carefully read the Warnings section of this manual before operating the device G

Download Pdf Manuals

image

Related Search

Related Contents

Manuel d`Installation De l`Analyseur GW475  n°003 - Marchin  アースクリーンK 取扱説明書  製品安全データシート  Da-Lite CM-CCA  User Manual: MK2430  EXRADIN A101 CT Ion Chamber  AXIS P5512 Guia de instalação  KF510 User Guide  Safeguard Pro RM User Manual  

Copyright © All rights reserved.
Failed to retrieve file