P25T Power Supply User Manual

image

Contents

1. 18 6 4 Operating Status Display The Power Pack P25 T is in one of the following three operating modes voltage limitation current limitation and power limitation depending on the selected voltage and current settings and the ohmic resistance of the electrophoresis chamber The operating mode is indicated by the two LEDs located below the display The following rules apply LED indication Display indication CURRENT LED illuminated Current limitation VOLTAGE LED illuminated Voltage limitation CURRENT LED and Power limitation VOLTAGE LED illuminated In the voltage control mode the output voltage will remain equal to the selected value see also chapter 5 8 Working with Constant Voltage Should the output current exceed the selected default value the current limitation function of the Power Pack P25 T is activated The output voltage is now adjusted to the ohmic resistance of the electrophoresis chamber The output current remains limited to the selected default value see also chapter 5 9 Working with Constant Current Should the user s selected voltage and current defaults exceed the maximum allowable output power of 200W the device will reduce the maximum allowable output current according to figure 1 in chapter 5 10 Both LED indicators are illuminated when the current limitation is accessed to indicate that the outp
2. 5 9 Working with Constant Current Adjust the current controller s set value to the desired output current see chapter 5 5 Current Setting when working with constant current The voltage controller s set value will then have been selected so high that the device will switch to current limit mode see chapter 5 4 Voltage Setting When the current output has reached the desired set value the Power Pack P25 T will limit the current The voltage output will then adjust to the ohmic resistance of the electrophoresis chamber and the current output will remain constant The LED indicator CURRENT see chapter 6 4 Operating Status Display is illuminated throughout the current limitation mode Should the set value for the voltage controller be selected too low the output voltage will therefore not suffice to reach the desired current The output power of the Power Pack P25 T is limited to 200W The maximum allowable current output of the device is limited to 200V as shown in figure 1 of chapter 5 10 Application example A semi dry blot should be conducted with a constant current of 800 mA For safety reasons the output voltage should not exceed SOV The following set values will be applied Output voltage 50V Output current 800mA 14 5 10 Power Limitation The output power of the Power Pack P25 T is limited to 200W Should the user s selected voltage and current defaults exceed the maximum allowable output power
3. biometra a part of Analytik Jena Standard Power Pack P25T Order No 040 850 Manual May 2009 Attention VAN Please read this manual carefully N prior to operating the device Biometra GmbH Rudolf Wissell Stra e 30 D 37079 G ttingen P O Box 1544 D 37005 G ttingen Phone 49 0 5 51 50 68 6 0 Fax 49 0 5 51 50 68 6 66 E mail Info biometra com Internet http www biometra com Service Division Rudolf Wissell Stra e 14 16 D 37079 G ttingen Phone 49 0 5 51 50 68 6 10 or 12 Fax 49 0 5 51 50 68 6 11 E mail Service biometra com 1 0 Intended Uses and Specification The Biometra Power Pack P25 T is suitable for PAGE SDS PAGE agarose gel electrophoresis blotting and electroelution of proteins and DNA RNA from gels The Power Pack P25 T is intended for use with any electrophoresis applications designed to operate below 400 volts and 500 milliamps DC electrophoresis or 200 volts and 1000 milliamps DC blotting Four sets of outputs operate in parallel to provide 0 to 400 volts up to 1000 milliamps and up to 200 watts in constant voltage or constant current mode Voltage and current setting is adjustable in steps of 1V and 1mA In addition the Power Pack P25 T offers a timer function which shuts down the output voltage after a defined time The timer can be set between 1 minute and 1999 minutes A single LCD display provid
4. P O Box 1544 D 37005 Gottingen Phone 49 0 5 51 50 68 6 0 Fax 49 0 5 51 50 68 6 66 E mail Info biometra com Internet http www biometra com 27
5. check the voltage or current settings see chapter 5 4 Voltage Setting The device must always be turned off before the AN supply line to the electrophoresis chamber is to be connected or removed for any given reason power switch on 0 and LED not illuminated 5 2 Timer Mode On Off The TIMER button sets the Power Pack P25 T on timer or continuous mode When the timer is activated illuminated LED left of timer button the output voltage will switch off by itself after the given time interval has elapsed see chapter 5 3 and 5 6 to start the timer and set time intervals 5 3 Voltage On Off To turn the output voltage on or off press the OPERATION button A timer will start counting when the timer function of the device is activated The output voltage will switch off by itself after the given time interval has elapsed A green LED left of OPERATION button shows when the output voltage is turned on Operating this button does not constitute a safe A N disconnection from the power source The device must always be turned off before the supply line to the electrophoresis chamber is to be connected or removed for any given reason power switch on 0 and LED not illuminated 5 4 Voltage Setting First turn off the output voltage by using the OPERATION button Then by using the SELECT button switch to VOLT on the display Set the desired output voltage by turning the knob In n
6. alue 12 5 7 2 Monitoring Actual Values Choose the desired measurement category at any time by using the SELECT button in order to control the output voltage and current In doing so only the LCD display will be switched The supplied current or applied voltage will not be affected by this operation 5 8 Working with Constant Voltage Adjust the voltage controller s set value to the desired output voltage see chapter 5 4 Voltage Setting when working with constant voltage Then select the maximum possible or allowable value for the current controller s set value The current controller s set value will then be the maximum possible or allowable value chosen see chapter 5 5 Current Setting The Power Pack P25 T supplies a constant output voltage until reaching the selected current limit voltage control The output current then varies according to the ohmic resistance of the electrophoresis chambers The output of a constant voltage will be shown by the LED indicator VOLTAGE see chapter 6 4 Operating Status Display Should the maximum allowable output current be exceeded the Power Pack P25 T will limit the current and the output voltage will drop Application example An agarose gel electrophoresis should be conducted with a constant voltage of 150V For safety reasons the output current should not exceed 300mA The following set values will be applied Output voltage 150V Output current 300mA 13
7. d occupational medicine Do not attempt any repair yourself In need of repair consult your local qualified service representative Only qualified personnel may open the housing parts The inner device may still contain dangerous voltage several minutes after disconnection from power supply Make sure the indicated voltage and frequency range of the device corresponds with the system voltage and that the power switch is on 0 before it is connected to the power supply Connect the device to a properly grounded outlet only Pull the power cord out of the outlet to safely disconnect from the power supply ae iP fe gt gt gt gt D Use exclusively the provided power cord or one that meets the legal requirements Pay attention to the dimensioning of the fuses and replace them if required The fuse holder is located in the lower part of the power connector on the back side of the Power Pack The following fuses are required Micro fuses slow blow 2 x 5A 250 V Caution For continued protection against risk of fire replace only with same type and rating of fuse Caution Double pole neutral fusing disconnect power before changing fuse In order to ensure the safe electrical disconnection the floating outputs must never be grounded Make sure that your test setup always complies to this condition Only safe electrical cords certified according to IEC EN 61010 2 031 may be connected to the pow
8. derspannungsrichtlinie 73 23 EWG und der EC low voltage directive 73 23 EEC and the EG Richtlinie ber die elektromagnetische Vertr glichkeit 89 336 EWG entsprechen EC directive about the electromagnetic compatibility 89 336 EEC Folgende harmonisierte Normen wurden angewandt The following harmonized standards have been used EN 60601 1 2 2001 EN 55011 1998 A1 1999 A2 2002 EN 61000 3 2 2000 EN 61000 3 3 1995 A1 2001 EN 61000 6 1 2001 EN 61010 1 2001 Dr J rgen Otte Quality Manager 26 Warranty This Biometra device has been accurately produced and the quality thoroughly controlled before delivery and is therefore guaranteed to correspond to the product specifications in this manual Biometra offers a 24 month warranty on defective parts under the following conditions This warranty is valid for 24 months from date of shipment to the purchaser through Biometra or an authorized dealer and can not be assigned to a third party without Biometra s written agreement This warranty is limited to the product and if applicable to the standard original accessories It is presumed that this device will exclusively be operated in accordance with Biometra s instructions Defective parts returned within the warranty period will be repaired or replaced without any charge by Biometra This watranty does not cover damage caused by misuse neglect or normal wear Biometra GmbH Rudolf Wissell StraBe 30 D 37079 Gottingen
9. er outputs Make sure the Power Pack P25 T is located in a safe and dry place in proximity to the electrophoresis devices powered Keep away from liquids Should a liquid penetrate the P25 T disconnect the device immediately PPR gt D Do not operate in a damp or humid environment where condensing moisture may lead to creeping current See chapter 2 Technical Data for ambient operating temperature range When moving from a cold to a warm room allow 8 hours before reusing the device in order for it to reach the ambient temperature and for possible condensation to evaporate Use only mild non abrasive detergents Carefully clean the device to avoid liquids penetrating the inner housing When positioning the P25 T make sure the ventilation openings on the back side are not covered Should the Power Pack become extremely warm smoke or display unusual information while in operation discontinue use immediately and contact a qualified service representative 4 0 Set Up 4 1 Content e Power Pack P25 T e Manual e Power Cord 4 2 Unpack and Check Your Biometra Power Pack P25 T has been carefully packed to ensure its safety Please unpack it carefully control the integrity of its content and inspect the device Report any damage immediately to BIOMETRA Do not operate a device which is in no proper condition Should you have a complaint return the complete system in its original packing by following the return in
10. es output voltage or current readings 2 0 Technical Data Type Output Constant Voltage and Constant Current Max Voltage 400 VDC Max Current 1000mA in the range 0 200VDC lt 1000mA in the range 200 400VDC Max Power 200 W continuous Timer 1 minute to max 1999 minutes Output Display LCD display 3 5 digits selectable between display mA V and min Resolution 1 mA 1 V and 1 min respectively Accuracy 5 v M for U I gt 0 05 Umax Imax 5 of 0 05 Umax Imax for U I lt 0 05 Umax Imax Arithmetic mean over 3 s T 0 4 s Number of Output Terminals Four sets of 4 mm safety sockets Input Power 100 240V AC 50 60Hz max 300W Allowable deviation 6 10 Ambient Operating Temperature Range 0 to 40 C Humidity 0 to 70 Dimensions W x Lx H 26 5 x 25 0 x 9 5 cm Weight 3 kg The Power Pack P25 T is designed for use with AC net supply voltages between 100 Volt and 240 Volt 50 60Hz It is a switched mode power supply providing voltage and current controlled DC power respectively The DC output is free of ground and short circuit proof See also chapter 5 10 3 0 Safety Instructions The Power Pack P25 T must only be used in accordance with its designation The use of this device unconformably to the operating instructions will dissolve any responsibility and liability Observe the relevant rules for accident prevention as well as those of approved safety an
11. he Power Pack P25 T is correctly connected to the power source e the fuses in the power inlet socket are not defective Caution Disconnect the device from the power VAN supply by unplugging the power cord before any fuse replacement 17 6 2 Display Indication The actual and set values for the voltage and current outputs will alternatively be shown on the LCD display see chapter 5 7 Monitoring Voltage Current and Timer The three LEDs on the right side of the display indicate to which parameter the value shown on the display belongs The following rules apply LED indication Display indication mA LED illuminated Output current in mA VOLT LED illuminated Output voltage in volt TIMER LED illuminated Remaining time in minutes mA LED flashes Set value for the current control in mA VOLT LED flashes Set value for the voltage control in volts TIMER LED flashes Set value for the timer in minutes 6 3 TIMER and OPERATION display mode Two LEDs next to the OPERATION and TIMER buttons indicate if the timer function is activated and if the output voltage is turned on or not The following rules apply LED indication Display indication TIMER LED on Timer mode TIMER LED off Duration mode OPERATION LED on Output voltage activated OPERATION LED off Output voltage deactivated
12. ore every return that the device is free of radioactive and 21 dangerous biological contamination Contaminated devices sent to Biometra will be refused The sender will be liable for the repair order in case of damage resulting from an inadequate decontamination of the device Please include a note with the following information a Sender s name and address b Contact person and phone number in case of inquiry 22 biometra a part of Analytik Jena Equipment Decontamination Certificate To enable us to comply with german law i e 71 StrlSchV 17 GefStoffV and 19 ChemG and to avoid exposure to hazardous materials during handling or repair will you please complete this form prior to the equipment leaving your laboratory COMPANY INSTITUTE ADDRESS TEL NO FAX NO E MAIL EQUIPMENT Model Serial No Tf on loan evaluation Start Date Finish Date Hazardous materials used with this equipment Has the equipment been cleaned and decontaminated YES NO delete Method of cleaning decontamination NAME POSITION HEAD OF DIV DEP INSTITUTE COMPANY SIGNED DATE PLEASE RETURN THIS FORM TO BIOMETRA GMBH OR YOUR LOCAL BIOMETRA DISTRIBUTOR TOGETHER WITH THE EQUIPMENT PLEASE ATTACH THIS CERTIFICATE OUTSIDE THE PACKAGING INSTRUMENTS WITHOUT THIS CERTIFICATE ATTACHED WILL BE RETURNED TO SENDER 23 biometra a part of Analytik Jena Note for Dispo
13. ormal mode the display shows the output voltage actual value By turning the knob the display changes from the actual value to the desired set value recognizable by the flashing VOLT LED indicator The desired value increases with every clockwise rotation and decreases with every counter clockwise rotation Depending on speed and direction of the rotation the value will augment or reduce in steps of 1V 10V or 100V A set value modified by the user will 3 seconds after the last operation of the knob precede all other values and will be set as the desired output voltage The voltage value V should be set without any device connected 10 5 5 Current Setting Turn off the output voltage by using the OPERATION button Then by using the SELECT button switch to mA on the display Set the desired output current by turning the knob In normal mode the display shows the current conduction actual value By turning the knob the display changes from the actual value to the desired set value recognizable by the blinking mA LED indicator The desired value increases with every clockwise rotation and decreases with every counter clockwise rotation Depending on speed and direction of the rotation the value will augment or reduce in steps of 1mA 10mA or 100mA A set value modified by the user will 3 seconds after the last operation of the knob precede all other values and will be set as the desired outpu
14. sal of Electric Electronic Waste This symbol the crossed out wheelie bin means that this product should be brought to the return systems and or separate systems available to end users according to yours country regulations when this product has reached the end of its lifetime For details please contact your local distributor This symbol applies only to the countries within the EEA EEA European Economics Area comprising all EU members plus Norway Iceland and Liechtenstein 24 9 0 Notes 25 biometra a part of Analytik Jena EU Konformit tserkl rung EC Declaration of Conformity G ttingen den 01 10 2004 im Sinne der EG Richtlinie ber elektrische Betriebsmittel zur Verwendung innerhalb bestimmter Spannungsgrenzen 73 23 EWG following the EC directive about electrical equipment for use within certain limits of voltage 73 23 EEC und and im Sinne der EG Richtlinie f r die elektromagnetische Vertr glichkeit 89 336 EWG following the EC directive about the electromagnetic compatibility 89 336 EEC Hiermit erkl ren wir da das folgende Power Pack mit Timerfunktion Herewith we declare that the following Power Pack with Timer Typ type Standard Power Pack P25T Rev b Best Nr Order No 040 850 den grundlegenden Anforderungen der corresponds to the basic requirements of EG Nie
15. structions see chapter 8 0 N Attention N Please send back the enclosed postcard along with the filled out Warranty Certificate to Biometra in order to claim the full warranty 4 3 Power Supply Before connecting make sure the actual voltage complies with the voltage range indicated on the type label A N Caution The Power Pack P25 T must be grounded 4 4 Location The Power Pack P25 T must be located in a dry and safe place Do not place an electrophoresis chamber on the power supply device The ventilation openings on the back side must not be covered by other devices furniture or fixtures A sufficient air flow for convection cooling must be provided 4 5 Electrophoresis Chamber Connection The electrophoresis chamber is connected to the Power Pack P25 T by safety cords Use exclusively electrophoresis devices with safety connectors Switch off the Power Pack P25 T while working on the test equipment setup 5 0 Operation 5 1 Power On Establish the connection to the electrophoresis chamber s before switching on the Power Pack P25 T Then switch on the device using the power switch A green light appears over the power switch when power is turned on When power is turned on the P25 T sets the output voltage or current to the latest value stored Should you be unsure of the latest saved settings detach all connections to the electrophoresis chamber and switch on the Power Pack P25 T in order to
16. t current The electric current value mA should be set without any device connected 11 5 6 Time Interval Setting By using the SELECT button switch to TIMER on the display Set the desired time interval value by turning the knob In timer mode the display shows the remaining time period actual value If time is up or the device is set on continuous operation mode the value will appear By turning the knob the display changes from the actual value to the desired set value recognizable by the flashing TIMER LED indicator The desired value increases with every clockwise rotation and decreases with every counter clockwise rotation Depending on speed and direction of the rotation the value will augment or reduce in steps of 1 minute 10 minutes or 100 minutes A default value modified by the user will 3 seconds after the last operation of the knob precede all other values 5 7 Monitoring Voltage Current and Timer 5 7 1 Monitoring Set Values Choose the desired measurement category by using the SELECT button in order to monitor the default voltage and current values Then turn the knob to the right or to the left by one detent and the LCD displays will show the desired set value of the corresponding measurement The value appearing on the display will 3 seconds after the last operation of the knob precede all other values The LCD display will then automatically be switched to the actual v
17. the device will reduce the default set value of the output current as shown in figure 1 1A Power ree Limitation 5 05A 5 gt O OA OV 200 V 400 V Output Voltage Figure 1 Automatic limitation of the output current at high output voltage values The CURRENT and VOLTAGE LED indicators are both illuminated when the current limitation is accessed see chapter 6 4 Operating Status Display to give a signal that the output current due to the power limitation is lower than the value originally set by the user 15 5 11 Switching Off The Power Pack P25 T can be turned on or off by pressing the on off switch The device must always be turned off before the N supply line to the electrophoresis chamber is to be connected or removed for any given reason power switch on 0 and LED above not illuminated 16 6 0 Status Notification 6 1 Power ON LED A green LED located above the on off switch shows along with the on off switch position if the Power Pack P25 T works properly on off switch LED Meaning position indication O dark No error Power off T green No error Power on T dark Error e No connection to power e Faulty fuse e Device failure The Power On LED remains dark although the device is switched on when there is an operation or hardware failure In such case please verify that e the power outlet supplies electricity e t
18. ut current due to the power limitation is lower than the value originally set by the user 19 7 0 Cleaning Disconnect the Power Pack Use only mild non abrasive detergents and a slightly moistened soft cloth Do not use solvents soak while cleaning or immerse in water 20 8 0 Service Should you have any problem with the device please contact our service division or your local Biometra dealer Biometra GmbH Service Division Rudolf Wissell StraBe 14 16 D 37079 Gottingen Phone 49 0 5 51 50 68 6 10 or 12 Fax 49 0 5 51 50 68 6 11 E mail Service biometra com Please follow the return instructions when returning the device Return instructions Return only faulty devices Please contact the technical support of Biometra Phone 49 0 5 51 50 68 6 10 or 12 to get an RAN number Use the original packing or a similar robust packing when returning the material Mark the outer packing with CAUTION SENSITIVE ELECTRONIC INSTRUMENT and the RAN number sticker Deliveries without RAN number can not be ac cepted Please include a precise description of the problem and preferably one that also points out by which procedure the failure occurs or where it originates Important Clean every part of the device from residues such as chemical radioactive and dangerous biological contamination Please confirm in writing Equipment Decontamination Certificate bef