Home

ScopeMeter® 120 and 190 Series

image

Contents

1. CQnnectan ICW ScopeMeter 120 and 190 Series incl 225C and 215C ScopeMeter 215C 225C and 190C Series Speed performance and analysis power For demanding applications the ScopeMeter 215C 225C and 190C Series high performance oscillo scopes offer specifications usually found on top end bench instru ments With up to 200 MHz band width 2 5 GS s real time sampling and a deep memory of 27 500 points per input they re ideal for engineers who need the full capabilities of a high performance scope in a handheld battery pow ered instrument e Dual input 200 100 or 60 MHz bandwidth e Up to 2 5 GS s real time sampling per input Bus Health Test capability for industrial buses 225C and 215C High waveform resolution of 3000 datapoints per channel Frequency Spectrum using FFT analysis Connect and View automatic triggering a full range of manual trigger modes plus external trig gering Digital Persistence for analyzing complex dynamic signals like on an analog oscilloscope Fast display update rate for seeing dynamic behavior instantaneously Automatic capture and replay of 100 screens 21 500 points and more per input record length using ScopeRecord mode Technical Data 1 000 V CAT II and 600 V CAT II safety certified Up to 1 000 V independently floating isolated inputs 5000 count DMM and paperless recorder built in e Four hours rechargeable NiMH battery pack Scop
2. 10 pF 2 pF Advanced meter functions Auto manual ranging relative measurements zero reference TrendPlot recording General specifications Input voltage ratings Maximum probe voltage CAT II 1 000 V CAT M 600 V Maximum voltage between 10 1 probe tip VPS210 and reference lead Floating voltage CAT II 1 000V CAT III 600 V Maximum voltage between earth ground and any terminal signal input or shielding Independently isolated inputs CAT II 1 000 V CAT III 600 V Maximum voltage between any terminal of one input or probe VPS210 and any other terminal of another input or probe VPS210 Maximum voltage on BNC input directly input A or B CAT III 300 V Maximum voltage on meter input CAT II 1 000 V CAT III 600 V ScopeMeter 215C 225C and 190C Series 120 Series Fluke Corporation 3 Memory save and recall Scope memories 15 memory locations that each can contain two waveforms plus corresponding setup With each storage action a user specified name 20 ASCII characters long can be assigned to the stored data for easier reference Recorder memories 2 memory locations that each can contain 100 captured dual input scope screens or a dual input ScopeRecord 27 500 Min Max pairs per input or a dual input Trendplot 18 000 min max pairs average values Real time clock Time and date stamp for ScopeRecord 100 captured screens and TrendPlots Mechanical data Size 256 mm x 169 mm x 64 mm 10 1 i
3. VPS200 and any other terminal of another input or probe VPS200 Maximum voltage on BNC input directly input A or B CAT Il 300 V Maximum voltage on meter input CAT II 1 000 V CAT III 600 V Environmental Operating temperature O C to 50 C Storage temperature 20 C to 60 C Humidity 10 C to 30 C 95 RH non condensing 30 C to 40 C 75 RH non condensing 40 C to 50 C 45 RH non condensing Maximum operating altitude 3 000 m 10 000 feet Maximum storage altitude 12 km 40 000 feet Electro Magnetic Compatibility EMC EN 61326 1 for emission and immunity Optically isolated PC printer interface To printer Supports HP Laserjet DeskJet Epson FX LQ Seiko DPU 414 and Postscript printers via optional PAC 91 To PC Transfer instrument settings screen images and waveform data compatible with FlukeView software for Windows via optional OC4USB or PM9080 Warranty Three years parts and labor on mainframe instrument One year on accessories Accessories BP 190 BC 190 Rechargeable battery pack installed Line voltage adapter battery charger Voltage probes and accessories 10 1 voltage probe VPS200 1 red 1 grey including hook clip ground lead with mini alligator clip ground spring for probe tip Multimeter test leads TL75 Hard Point test lead set 1 red 1 black Multi lingual CD ROM Three breakout adapt BHT190 ers for bus systems using RJ45 DBY and
4. Lo Fluke 225C_ Fluke 215C_ Fluke 199C _ Fluke 196C_ Fluke 192C Fluke 125 Fluke 124 Fluke 123 Bandwidth 200 MHz 100 MHz 200 MHz 100 MHz 60 MHz 40 MHz 40 MHz 20 MHz Max real time sample rate 2 5 GS s 1 GS s 2 5 GS s 1 GS s 500 MS s 25 MS s 25 MS s 25 MS s sample rate 10 2 om monochrome LCD Y Visual Reference and Automatic Pass Fail testing Max record length in Scope mode in ScopeRecord mode Number of inputs 2 plus external DMM input all isolated from each other and from ground 2 opt Isolated External Trig thru ITP120 Number of digitizers Independently floating isolated inputs Up to 1000 V between inputs references and ground 2 mV div 100 V div 5 mV div 500 V div Glitch capture Up to 3 ns using Pulse Width triggering 50 ns peak detect at 5 us div to 1 min div 40 ns Timebase range in scope 10 ns div 20 ns div Maser tes Connect and View Free Run Single Shot Edge Delay Video Frame Video Line Connect and ViewTM Free Run Single Shot gger typ selectable pulse width and External Dual slope trigger and Event trigger n cycle Edge Video iq S 3000 points per input channel allowing for high time resolution signal analysis using 512 min max points per input Zoom 27 500 points per input or more 5 ms div 2 min div As 124 Power Cursor 26 56 autowatic 7 cursor measurements 30 automatic measurements Automatic Vrms and watts measuremen
5. free RS232 port or USB port e PM9080 optically isolated RS232 adapter cable or e OC4USB optically isolated USB interface adapter cable available separately or included in SCC120 SCC190 kit and in ScopeMeter S versions Supported instruments Full support for Fluke 199C 199B 199 196C 196B 196 192B 192 125 124 and 123 starting release V4 5 the Fluke 225C 215C and 192C are supported ScopeMeter 215C 225C and 190C Series 120 Series Fluke Corporation 7 Rechargeable battery BP190 BP120MH pack installed Line voltage adapter BC190 PM8907 battery charger Voltage probes and 10 1 voltage probe VPS210 including hook STL120 Shielded Test lead set VPS40 high accessories clip ground lead with mini alligator clip impedance 10 1 probe 40 MHz 1 black included ground spring for probe tip with Fluke 125 124 ground leads with mini alligator clips AC120 alligator clips BB120 BNC to Shielded banana adapter Multimeter test leads TL75 Hard Point test lead set TL75 Hard Point test lead set 1 black 1 red 1 black Current clamp 1400s Current Clamp included with Fluke 125 only Multi lingual CD ROM Multi lingual CD ROM Bus test BHT190 included with Fluke 225C and 215C BHT190 optional for use with Fluke 125 only connection support acts as break out adaptor for DB 9 RJ 45 and M12 industrial bus connection systems Next to the above standard accessories Fluke offers a wide range of optional acce
6. shot roll Trigger and delay Source Input A input B external via optional ITP120 Modes Automatic Connect and View Free Run Edge single Shot Video Video Line Connect and View Advanced automatic triggering that recognizes signal patterns and automatically sets up and continuously adjusts triggering time base and amplitude Automatically displays stable pictures of complex and dynamic signals like motor drive and control signals Video triggering NTSC PAL PAL SECAM Includes line select Time delay Up to 10 divisions pre trigger view Measurements Vdc Vac Vac dc Vpeak max Vpeak min Vpeak to peak frequency Hz positive pulse width negative pulse width positive duty cycle negative duty cycle Amp ac Amp dc Amp ac dc Phase Temperature C Temperature F dBV dBm into 50 Q and 600 Q Amps C or F with optional probes Cursor measurements 125 124 Sources Input A Input B Modes Single or dual vertical cursor dual horizontal cursor risetime or falltime Measurements single vertical line Average min value max value time from start of recording in roll mode Dual vertical lines AV at markers time between cursors 1 T between cursors in Hz Dual horizontal lines High low or AV readout risetime and falltime transition time O level 100 level with markers at 10 and 90 Accuracy As oscilloscope Bus health tester Fluke 125 only Bus health automatically anal
7. 1010 1 04 including CSAys approval ANSI ISA 82 02 01 ScopeMeter 215C 225C and 190C Series 120 Series Input voltage ratings Maximum input voltage CAT IIT 600 V Maximum voltage between input and reference lead Maximum input voltage using VPS40 Probe CAT III 600 V CAT II 1000 V Maximum voltage between probe tip input and reference lead Floating voltage CAT III 600 V Maximum voltage between earth ground and any terminal signal input or reference lead Maximum voltage between reference leads Instrument has common grounds connected via self recovering fault protection For different ground potential measurements between inputs use DP120 differential voltage probe or a Fluke 190C Series instrument Environmental According to MIL PRF 28800F Class 2 Operating temperature O C to 50 C Storage temperature 20 C to 60 C Humidity 10 C to 30 C 95 RH non condensing 30 C to 40 C 75 RH non condensing 40 C to 50 C 45 RH non condensing Maximum operating altitude 2 000 m 6 500 ft 4 500 m 15 000 ft voltages lt 300 V Maximum storage altitude 12 km 40 000 ft Electro Magnetic Compatibility EMC EN61326 1 for emissions and immunity Optically isolated PC printer interface To printer Supports HP Laserjet Deskjet Epson FX LQ and postscript printers via optional PAC91 To PC Transfer instrument settings screen images and data compatible with FlukeView software for Wind
8. Continuous roll for the duration of the full recordable timespan Measurement speed 5 measurements per second Horizontal scale Time from start time of day Zoom Up to 64x zoom Memory Up to 2 TrendPlot recordings can be saved for later recall and analysis Cursor measurements all recorder modes Source Input A B or DMM input Dual vertical lines Min Max or Average voltage Time between cursors Single vertical line Min Max or Average voltage Absolute date and time or time from start Meter mode Via 4 mm banana inputs Fully isolated from scope inputs and scope ground The specified accuracy is valid over the temperature range 18 C to 28 C 65 F to 82 F Add 10 of specified accuracy for each degree C below 18 C or above 28 C Maximum resolution 5 000 counts Voltmeter ranges 500 mV 5 V 50 V 500 V 1 000 V Accuracy V de 0 5 5 counts V ac true rms 15 Hz to 60 Hz 1 10 counts 60 Hz to 1 kHz 2 5 15 counts V ac dc true rms dc to 60 Hz 1 10 counts 60 Hz to 1 kHz 2 5 15 counts Ohms Ranges 500 Q 5 kQ 50 kQ 500 kQ 5 MQ 30 MQ Accuracy 0 6 5 counts Other meter functions Continuity Beeper on lt 50 Q 30 Q Diode test Up to 2 8 V Amps A dc A ac A ac dc using an optional current clamp or shunt Scaling factors 0 1 mV A 100 V A Temperature C F With optional accessories Scale factors 1 mV C or 1 mV F Input impedance 1 MQ 1
9. M12 connectors User manual ScopeMeter 215C 225C and 190C Series 120 Series Oscilloscope mode Vertical deflection Bandwidth risetime e with VPS40 probes e input A and B directly 40 MHz 40 MHz 12 5 MHz 20 MHz 20 MHz 12 5 MHz e with STL120 Shielded Test Leads Instrument risetime 8 75 ns 17 5 ns input directly Number of inputs two Input coupling AC dc with ground level indicator Input sensitivity 5 mV to 500 V div with the included VPS40 Fluke 125 124 and STL120 shielded test leads measure up to 600 Vrms CAT MI Input voltage 600 V CAT III See general specifica tions for more detailed information Vertical resolution 8 bit Accuracy 1 of reading 0 05 x range div Input impedance 1 MQ 1 225 pF with STL120 shielded test leads 1 MQ 1 20 pF 3 pF with BB120 5 MQ 1 15 5 pF with VPS40 10 1 voltage probe Horizontal Max sample rate both channels simultaneously Fluke 125 124 2 5 GS s for repetitive signals 25 MS s for single shot Fluke 123 1 25 GS s for repetitive signals 25 MS s for single shot Number of digitizers two Time base range 10 ns div to 1 min div Fluke 125 124 Maximum record length 512 Min Max points per input Accuracy 0 1 of reading 1 pixel Glitch detect 40 ns Display and acquisition Display modes Input A input A and B envelope smooth Acquisition modes Normal including glitch capture single
10. UKE or Fax 905 890 6866 From other countries 1 425 446 5500 or Fax 1 425 446 5116 Web access http www fluke com 2001 2010 Fluke Corporation Specifications subject to change without notice Trademarks are the property of their respective owners Printed in U S A 1 2010 1629083H D EN N Modification of this document is not permitted without written permission from Fluke Corporation
11. appliance outlets with short connections to service entrance CAT II Single phase receptable connected loads e Outlets and long branch circuits e All outlets at more than 10 m 30 ft from Category III source e All outlets at more than 20 m 60 ft from Category IV source CAT I Electronic e Electronic equipment e Low energy equipment with transient limiting protection To protect your instrument and more importantly high voltage energy transients To choose the right yourself choose a test tool that can withstand the tool the voltage rating alone does not determine the electrical hazards present in the environment in which safety It is the combination of voltage rating and you plan to use it installation category that determines maximum tran EN61010 establishes international safety require sient withstand capability of the tool CAT MI rated ments for electrical measurement equipment It instruments are recommended for measurement on separates the various electrical environments into industrial power distribution systems installation categories based on the danger from Ordering information Fluke 225C Color ScopeMeter 200 MHz 2 5 GS s with Bus Health Test Functions Fluke 225C S Color ScopeMeter 200 MHz 2 5 GS s with Bus Health Test SCC190 kit Fluke 215C Color ScopeMeter 100 MHz 1 GS s with Bus Health Test Functions Fluke 215C S Color ScopeMet
12. eMeter 120 Series Three in one simplicity The compact ScopeMeter 120 Series is the rugged solution for industrial troubleshooting and installation applications It s a truly integrated test tool with oscilloscope multi meter and paperless recorder in one affordable easy to use instru ment Find fast answers to problems in machinery instrumentation control and power systems e A dual input 40 MHz or 20 MHz digital oscilloscope e Two 5 000 counts true rms digital multimeters Cursor measurements Fluke 124 125 A dual input TrendPlot recorder Connect and View trigger simplicity for hands off operation Shielded test leads for oscillo scope resistance continuity and capacitance measurements Full bandwidth VPS40 10 1 40 MHz probe included standard with Fluke 124 125 Up to seven hours battery operation 600 V CAT III safety certified Optically isolated RS 232 interface Rugged compact case Bus Health test for industrial bus systems Fluke 125 Power measurements and har monics measurement Fluke 125 Oscilloscope mode Vertical deflection Bandwidth 200 MHz 100 MHz 60 MHz Bandwidth limiter User selectable 10 kHz 20 MHz or off Number of inputs Two plus external trigger All inputs references isolated from each other and ground Input coupling AC or dc with ground level indicator Input sensitivity 2 mV div to 100 V div Normal invert On both input channels switched separat
13. ely Variable attenuator Variable gain on input channel A Input voltage 1000 V CAT II 600 V CAT III rated see general specifications for further details Vertical resolution 8 bit Accuracy 1 5 of reading 0 04 x range div Input impedance 1 MQ 1 15 pF 2 pF 500 MS s Maximum real time sample rate Number of 2 digitizers Time base range Maximum record length 3000 points per input in Scope mode 27 500 points per input in ScopeRecord roll mode 5 ms div to 2 min div Accuracy 0 01 of reading 1 pixel Glitch capture 50 nsec 5 usec div to 1 min div 5 ns div to 5 s div 10 ns div to 5 s div Display and acquisition Display 144 mm full color LCD with backlight Display modes Input A input B dual average Replay Visible screen width 12 divisions Waveform mathematics A B A B A B all with user selectable scaling of resultant A versus B X Y mode Frequency spectrum using FFT analysis Acquisition modes Normal auto single shot scopeRecord roll glitch capture waveform compare waveform compare with automatic Pass Fail testing Bus Health test mode 225C and 215C only Eyepattern Display of single ended or differential bus signal Fluke 225C and 215C only Trigger and delay Source Input A input B external trigger input All input references isolated from each other and from ground Modes Automatic Connect and View free run single s
14. er 100 MHz 1 GS s with Bus Health Test SCC 190 kit Fluke 199C Color ScopeMeter 200 MHz 2 5 GS s Fluke 199C S Color ScopeMeter 200 MHz 2 5 GS s SCC190 kit Fluke 196C Color ScopeMeter 100 MHz 1 GS s Fluke 196C S Color ScopeMeter 100 MHz 1GS s SCC190 kit Fluke 192C Color ScopeMeter 60 MHz 500 MS s Fluke 192C S Color ScopeMeter 60 MHz 500 MS s SCC190 kit Fluke 125 Industrial ScopeMeter 40 MHz Fluke 125 S Industrial ScopeMeter 40 MHz SCC120 kit Fluke 124 Industrial ScopeMeter 40 MHz Fluke 124 S Industrial ScopeMeter 40 MHz SCC120 kit Fluke 123 Industrial ScopeMeter 20 MHz Fluke 123 S Industrial ScopeMeter 20 MHz SCC120 kit SCC190 FlukeViewe Software Cable Case 190 Series 5CC120 FlukeViewe Software Cable Case 120 Series PM9080 Optically Isolated RS 232 adapter cable OC4USB Optically Isolated USB interface cable BPIZ6 Differential Voltage Probe for Fluke 120 Series BHT 190 Bus Health Test break out adapter for DB 9 RJ 45 and M12 connection systems P1206 Optically Isolated External Trigger Input for Fluke 120 series SWSOW FlukeView ScopeMeter Software for Windowse C190 Hard Shell Carrying Case for Fluke 190 series C120 Hard Shell Carrying Case for Fluke 120 series SCC kit includes Hard shell carrying case optically isolated USB interface cable and FlukeView for Windows software ScopeMeter 215C 225C and 190C Series 120 Series Fluke Corporation 9 Selection guide
15. f 1 5 hours Recording mode Continuous with automatic vertical scaling and horizontal time compression Measurement speed 2 5 measurements per second maximum Horizontal scale Time from start General specifications Case Design Rugged shock proof with integrated protective holster Drip and dust proof IP51 according to IEC529 Shock and vibration Shock 30 g according to MIL PRF 28800F Class 2 par 3 8 4 2 and 4 5 5 3 1 Shock and vibration Vibration 3 g according to MIL PRF 28800F Class 2 par 3 8 5 1 and 4 5 5 4 1 Display Bright LCD with CCFL backlight 60 35 cd m2 with without power adapter Size 72 mm x 72 mm 2 8 in x 2 8 in Resolution 240 x 240 pixels Contrast and brightness User adjustable temperature compensated Memory save and recall 20 Fluke 125 124 and 10 Fluke 123 instrument screens with user set ups and user text Real time clock Time and date stamp TrendPlot recording Power Line power Country specific line voltage adapter battery charger included Battery power Rechargeable Ni MH BP120MH installed Battery operating time Up to 7 hours using BP 120MH Battery charging time 7 hours Battery power saving functions Auto power down with adjustable power down time On screen battery power indicator Mechanical data Size 50 mm x 115 mm x 232 mm 2 in x 4 5 in x 9 1 in Weight 1 2 kg 2 64 lb Safety Compliance FN61010 1 2001 Pollution degree 2 CAN CSA C22 2 No 6
16. hot edge delay video video line selectable pulsewidth dual slope N cycle Connect and View Advanced automatic triggering that recognizes signal patterns automatically sets up and continuously adjusts triggering time base and amplitude Automatically displays stable waveforms of complex and dynamic signals like motor drive and control signals Can be switched off if desired Video triggering NTSC PAL PAL SECAM Includes field 1 field 2 and line select 2 Fluke Corporation Horizontal Pulse width triggering Pulse width qualified by time Allows for triggering lt t gt t t t where t is select able in minimal steps of 0 01 div or 50 nsec Time delay One full screen of pre trigger view or up to 100 screens 1200 divisions of post trigger delay Dual slope triggering Triggers on both nsing and falling edges alike N cycle triggering Triggers on N th occurrence of a trigger event N to be set in the range 2 to 99 Automatic capture of 100 screens The instrument ALWAYS memorizes the last 100 screens no user setup required When an anomaly occurs on screen the REPLAY button can be pressed to review the full screen sequence over and over Instrument can be set up for triggering on glitches or intermittent anomalies and will operate in baby sit mode capturing 100 events Replay Manual or continuous replay Displays the captured 100 screens as a live animation or under manual control Each scree
17. itude pulse width or baud rate risetime fall time jitter signal distortion noise HF noise LF in band noise Recorder mode ScopeRecord roll mode Dual input waveform storage mode Source and display Input A input B dual Memory depth 27 500 points per input Each point consists of Min Max pair Min Max values Min Max values are measured at high sample rates ensuring capture and display of glitches Recorded timespan 6 sec to 24 hr Glitch capture 20 MS MS Recording modes Single sweep continuous roll start on Trigger through external Stop on Trigger through external Stop on Trigger through external ScopeRecord mode can be stopped by an individual trigger event or by an interruption of a repetitive trigger signal Horizontal scale Time from start time of day Zoom Up to 100x Memory Up to 2 dual input ScopeRecord waveforms can be saved for later recall and analysis TrendPlot recording single or dual input electronic paperless chart recorder Plots displays and stores meter and scope measurements Source and display Input A input B or DMM input Memory depth 138 000 points recording Per record point a minimum a maximum and an average value plus a date and timestamp are recorded Ranges 5 s div to 30 min div in normal view mode 5 min div to 48 hr div in view all mode giving overview of total record Recorded timespan Up to 22 days with a resolution of 1 minute Recording mode
18. monics dis play using FFT processing which shows the relative amplitudes of the 1st up to the 33rd harmonic Harmonics frequency range DC to 33rd harmonic fundamental lt 60 Hz DC to 24th fundamental lt 400 Hz Analyzed waveform Voltage waveform Ch A current waveform Ch B or power Ch A x Ch B automatically generated Display Bargraph showing 1st up to 33rd harmonic amplitude displayed in relative to fundamental Timebase setting 5 ms div Measurements Relative amplitude of individual harmonic THD in r or of ScopeMeter 215C 225C and 190C Series 120 Series Fluke Corporation 5 Dual input meter The specified accuracy is valid over the temperature range 18 C to 28 C 64 F to 82 F Add 10 of specified accuracy for each C below 18 C or above 28 C 64 F to 82 F Max meter bandwidth 40 MHz Fluke 125 124 20 MHz Fluke 123 Voltage measurements Measurement selection V dc V ac rms V ac dc rms Vpeak max Vpeak min Vp p Ranges 500 mV 5 V 50 V 500 V 1250 V Full scale reading 5 000 counts Accuracy V dc 0 5 5 counts V ac rms 1 Hz to 60 Hz 1 10 counts 60 Hz to 1 kHz 2 5 15 counts 20 kHz to 1 MHz 5 20 counts V ac dc true rms DC to 60 Hz 1 10 counts 60 Hz to 1 kHz 2 5 15 counts 20 kHz to 1 MHz 5 20 counts V ac pwm Measures the effective output voltage of pulse width modulated motordrives and frequency inve
19. n has date and time stamp Replay storage Up to 2 sets of 100 screens each can be saved for later recall and analysis FFT Frequency spectrum analysis Shows frequency content of oscilloscope waveform using Fast Fourier Transform Window Automatic Hamming Hanning or None Automatic window Digitally re samples acquired waveform to get optimum frequency resolution in FFT resultant Vertical scale Linear logarithmic in volts Frequency axis Logarithmic frequency range automatically set as function of timebase range of oscilloscope Waveform compare and pass fail testing Waveform compare Provides storage and display of a reference waveform for visual comparison with newly acquired waveforms Reference is derived from an acquired waveform and can be modified in the ScopeMeter or externally using FlukeView Software Pass Fail Testing In waveform compare mode the Color Scopemeter can be set up to store only matching Pass or only non matching Fail acquired wave forms in the replay memory bank for further analysis Automatic scope measurements V dc V ac rms V ac dc Vpeak max Vpeak min Vpeak to peak A ac Adc A ac dc frequency Hz risetime falltime power factor Watts VA VA reactive phase pulse width pos neg dutycycle pos neg temperature C temperature F dBV dBm into 50 Q and 600 Q Vpwm ac Vpwm ac dc for measurement on pulse width modulated motordrives and frequency inverters Cur
20. n x 6 6 in x 2 5 in Weight 2 kg 4 4 lbs Case Design Rugged shock proof with integrated protective holster Drip and dust proof IP51 according to IEC529 Shock and vibration Shock 30g Vibration sinusoidal 3g according to MIL PRF 28800F Class 2 Display Bright full color LCD with backlight Brightness 80 Cd m typ using power adapter Display size 115 2 mm x 86 4 mm 4 54 in x 3 4 in 144 mm 5 67 in diagonal Resolution 320 x 240 pixels Contrast and brightness User adjustable temperature compensated Brightness 80 cd m typ using power adapter Power Line power Country specific line voltage adapter battery charger included Battery power Rechargeable NiMH installed Battery operating time 4 hours Battery charging time 4 hours Battery power saving functions Auto power down with adjustable power down time On screen battery status indicator Safety Compliance FN61010 1 2001 Pollution degree 2 UL61010B with approval CAN CSA C22 2 No 61010 1 04 with approval ANSI ISA 82 02 01 4 Fluke Corporation Input voltage ratings Maximum probe voltage CAT II 1000 V CAT III 600 V Maximum voltage between 10 1 probe tip VPS200 and reference lead Floating voltage CAT II 1000 V CAT III 600 V Maximum voltage between earth ground and any terminal signal input or shielding Independently isolated inputs CAT II 1000 V CAT III 600 V Maximum voltage between any terminal of one input or probe
21. ows via optional OC4USB USB or PM9080 RS 232 interface cable Warranty Three years parts and labor on main instrument One year on accessories FlukeView ScopeMeter software helps you get more out of your ScopeMeter e Store instrument s screen copies on the PC in color with Fluke 190C Series only or in black and white e Copy screen images into your reports and documentation Capture and store waveform data from your ScopeMeter on your PC Create and archive waveform references for automatic Fluke 190C Series or visual Fluke 190B and 190C Series comparison e Includes waveform analysis e g FFT spectrum analysis e Copy waveform data into your spreadsheet for detailed analysis e Use cursors for parameter measurement e Extended recording of up to four user selected measurements help you monitor and analyze slow moving signals and related events e Logging of other readings directly into other application programs eg spreadsheet e Add user text to instrument setups and send these to the instrument for operator reference and instructions e Capture complete Replay sequence into the PC for further analysis and documentation Fluke 190C series e English French and German versions included on a single CD ROM Note Some functionality may be available with specific ScopeMeter models only System requirements e Pentium 90 or better e CD ROM drive e Microsoft Windows 2000 and beyond e One
22. rters Fluke 125 only Vpeak Max peak or Min peak 5 of full scale Peak to peak 10 of full scale A ac dc true rms A ac A dc Current Clamp or shunt sensitivity 0 1 mV A 1 mV A 10 mV A 100 mV A 400 mV A 1 V A or 10 mV mA Ohms Ranges 500 Q 5 KQ 50 KQ 500 KQ 5 MQ 30 MQ all models 50 00 Q Fluke 125 only Max resolution 5 000 counts Accuracy 0 6 of reading 5 counts Capacitance Ranges 50 nF to 500 pF Max resolution 5 000 counts Accuracy 2 of reading 10 counts Other meter functions Frequency Up to 70 MHz Fluke 125 124 or up to 40 MHz Fluke 123 Rotational speed RPM Revolutions per minute based on 1 2 or 4 or 8 pulses per 2 revolutions Fluke 125 only Max RPM reading 50 kRPM Continuity Beeper on lt 30 Q Diode test Up to 2 8 V Duty cycle 2 to 98 up to 30 MHz Temperature C F With optional accessories Scale factors 1 mV C or 1 mV F Number of inputs 2 Input impedance 1 MQ 1 10 pF 2 pF Advanced meter functions Auto manual ranging TouchHold Relative measurements zero reference TrendPlot recording 6 Fluke Corporation Recorder mode Trendplot recording Dual input electronic paperless chart recorder Plots and displays the actual minimum maximum and average of any measurement Source and display Input A input A and B Range 15 s div to 2 days per division automatic Recorded timespan Up to 16 days with a resolution o
23. sor measurements Source Input A input B or the mathematical result trace excluding A vs B curve Dual horizontal lines Voltage at cursor 1 and 2 voltage between cursors Dual vertical lines Time between cursors 1 T between cursors in Hz voltage between markers risetime with markers falltime with markers Vrms between cursors Watts between cursors Single vertical line Min Max and average voltage at cursor position frequency and rms value of individual frequency component in FFT result ScopeMeter 215C 225C and 190C Series 120 Series Zoom Up to 16x horizontal zoom Bus health test mode Fluke 225C and 215C only Bus Health automatically analyzes the electrical signals on the industrial bus system to measure individual parameters and to give waveform information Automatically compares the measurement results to preset values and present good weak or false indicator with each parameter Bus types and reference standards used AS i EN50295 166 kb s CAN bus ISO 11898 up to 1 Mb s Modbus EIA 232 up to 115 kb s and EIA 485 up to 10 Mb s Foundation Fieldbus H1 61158 type 1 31 25 kb s Profibus DP EIA 485 up to 10 Mb s and PA 61158 type 1 31 25 kb s Ethernet 10Base2 coaxial and 10BaseT UTP 10 Mb s Ethernet 100BaseT 100 Mb s RS 232 EIA 232 up to 115 kb s RS 485 EIA 485 up to 10 Mb s Measured parameters where applicable Bias voltage level signal ampl
24. ssories like temperature probes current clamps high voltage probes cables adapters and carrying cases to further assist you in your job See the Fluke web site or contact your dis tributor for details SCC190 and SCC 120 Software Case Cable kits For user s safety the Fluke ScopeMeters are connected to a PC or printer using an optically isolated interface cable Software and cable can be ordered separately or as part of a special value kit the SCC190 or the SCC120 kit Each of these include a protective hard shell carrying case model depending on the ScopeMeter model for safe and convenient storage of instrument and accessories the FlukeView ScopeMeter Software for Windows and the OC4USB interface cable For those who prefer an RS 232 link an optically isolated RS 232 cable PM9080 is available as separate item 8 Fluke Corporation ScopeMeter 215C 225C and 190C Series 120 Series FLUKE o International safety standards Measurement voltage category Summary description CAT IV Three phase at utility connection any outdoor conductors under 1 000 V e Outside and service entrance e Service drop from pole to building e Run between meter and panel e Overhead line to detached building e Underground line to well pump CAT II Three phase distribution under 1 000 V including single phase commercial lighting and distribution panels e Feeders and short branch circuits e Distribution panel devices e Heavy
25. t on cursor limited part of waveform K lied eae Somes RPM Vpwm measurement THD urea test moton oe or sanden manae pres any uses A B A B AxB A ae UE A Lissajous diagrams a ra Powermessiements O O pwm PR Pe moe Someone sopone O a Capture last 100 screens Automatic with replay capability a ee Dual input TrendPlot Yes with cursors and zoom Memory for screens 10 screens and set ups set ups 5 more memories are made available upon registration of the ScopeMeter 20 10 Memory for recordings Two each can store 100 scope screens a ScopeRecord or a TrendPlot True rms multimeter 5000 counts volts amps ohms continuity diode temp Dual fully featured 5000 counts DMM Safety certified l 600 V CAT III EN61010 1 1000 V CAT II 600 V CAT III instrument and included accessories Instrument and included accessories Battery installed 4 hr Ni MH BP190 7 hr Ni MH BP120MH e A E Using optional optically insulated adapter cable OC4USB USB or PM9080 RS 232 Three years on main instrument one year on the standard accessories Scope measurements measurements Fluke Keeping your world up and running Fluke Corporation PO Box 9090 Everett WA 98206 U S A Fluke Europe B V PO Box 1186 5602 BD Eindhoven The Netherlands For more information call In the U S A 800 443 5853 or Fax 425 446 5116 In Europe M East Africa 31 0 40 2675 200 or Fax 31 0 40 2675 222 In Canada 800 36 FL
26. yzes the electrical signals on the network to give waveform data and measure individual parameters Automatic comparison of the measurement results to the standards results in good or false indicators to be displayed per parameter Bus types and reference standards used AS i EN50295 166 kb s CAN bus ISO 11898 up to 1 Mb s Interbus S EIA 485 up to 10 Mb s ControlNet 61158 type 2 2 5 Mb s Modbus EIA 232 up to 115 kb s and EIA 485 up to 10 Mb s Foundation Fieldbus H1 61158 type 1 31 25 kb s and H2 61158 type 1 up to 10 Mb s Profibus DP EIA 485 up to 10 Mb s and PA 61158 type 1 31 25 kb s Ethernet l1OBase2 coaxial and 10BaseT UTP 10 Mb s RS 232 EIA 232 up to 115 kb s RS 485 EIA 485 up to 10 Mb s or user defined system Measured parameters Baud rate risetime falltime high level low level distortion amplitude and jitter with comparison to system s standard values Power measurements Fluke 125 only Measure types Watt VA VAR Power Factor PF Power configuration Single phase or balanced 3 phase delta configuration mains supply Voltage measurement Channel A using STL120 voltage probe or direct input Current measurement Channel B using i400s current clamp included or other compatible clamp Current clamp or shunt sensitivity 0 1 1 10 100 1000 mV A 10 mV mA and 400 mV A Harmonics mode Fluke 125 only Converts waveform information into a har

Download Pdf Manuals

image

Related Search

Related Contents

モニタワーエコノミー取扱説明書  User Manual Series B Series BA  know your pressure washer  Scosche reqKASE  Poulan PLN3516F User's Manual  Viewsonic TD2240 touch screen monitor  iSequence for iPad 2.1  LevelOne 1.25G MMF SFP Transceiver, 550m, 850nm, -20 ~ 85C  User Manual - Stream Labs  HiFi digital music server OLIVE 6HD Owner`s manual OLIVE  