Home
GasAlertMicroClip Technical Reference Guide
Contents
1. MicroClip 1 2 3 and 4 Gas Detector Technical Reference Guide W o by Honeywell Table of Contents Heading Page INTOCUCIO A iii satu rckauncuaevaaoresunccoxtasaccausesteaseadeosd abeneevituccaneuaneuvsosreis 1 Zeroing the SENSOMS inicio 1 Contacting BW Technologies by Honeywell cccsccceeeceeeeeeeseeeeeeeeeeneeceeeeeeee essen seen seaaeeeaeseessaesenaeseaesenessanesnessnessnes 2 Satety Information Read PINS uc dd 2 Sensor POISONS and Contaminants iii ii id da e di aiii 5 Getting SLANG Gl asnicar acc 6 Parts OF Ine GasAlertMicroG D nani 7 DISDIAY ElemeniS rrie ai r ilecii cialis 8 P USNDULOM voissa o A deannsa bass sosndeuceunwantoutusncuauccauseutoascaorondeuvinosvecucesuctateuxceoneuse 9 ACUVAUNG the Detector aura lecci n 10 A EE T EAN ee E A AE ee ee renee 10 e a e a a a a a A g 10 AUGIDIC VISUAL POST errea aa a A O a 10 Detector V ero E a aa 10 Startup MESS eiia A A E 10 O o a O a 11 Sensor and POWER Teslui anae a a a a a a a haw do o ld a a naan ins 11 Automatic Zero and O Calibration OptiONA ooooooocccccononccccnncccnnnnnnccnnonnnccnnonnnnononnncnnnnnnnnrrnnnnnnnnrnnnnrrnrnnnnnnrcnnncen 12 Calibration Due Dale optional his ii ls 12 Last Calibration FAC ON acido cias asia 13 Overdue Os 0 oe 9 tii 13 AW A Beaton a a hhh calle ot Raia a beth faced hot a a Stabe a tees a cea tedt Seat 13 BOMO Testi aa al id 13 dst BUMP TES CEINE ni oran 14 Fo
2. Auto Zero on Startup When enabled the sensors automatically zero during the startup self tests The Auto Zero on Start up option is available for the CO HS and LEL sensors Not applicable to O 1 Click the checkbox of each sensor that will be auto zeroed during startup The detector is shipped with the Auto Zero on Start up option enabled for the CO H S and LEL sensor GasAlertMicroClip Sensor Configuration O2 Auto Calibration on Startup Automatic O Calibration When enabled the O sensor is automatically calibrated during the startup self tests The detector is shipped with the O2 Auto Calibration on Startup option enabled LEL By Vol CH4 When enabled the detector displays the LEL value as CH4 assuming a methane environment The LEL By Vol CH4 option is applicable to the LEL sensor only The detector is shipped with the LEL by Vol CH4 disabled 23 GasAlertMicroClip Technical Reference Guide User Options The user options section provides detector features that can be enabled or disabled The green checkmark indicates the option is enabled Click the checkmark to disable the option User Options _ Latching Alarms Safe Mode Stealth Mode _ IR Stealth Mode Low Alarm Acknowledge Force Calibration When Overdue _ Cal Lock Force Bump When Overdue Confidence Beep O IntelliPlash Applicable to Gas4lertMicroclip 7 only mA Figure 7 Fleet Manager II IR Link User Op
3. aH a O bt auto span m rm o S IR Lock Enabled Perform calibration using IR Lock displays the IR Link with Fleet when calibration is Manager ll software or attempted insert the detector into the MicroDock II station Refer to Cal Lock Calibration IR Lock in User Options and Calibration 56 Replacement Parts and Accessories A Warning To avoid personal injury and or damage to the detector use only the specified replacement parts To order parts or accessories listed in the following table contact BW Technologies by Honeywell Table 16 Replacement Parts and Accessories MICROpeL combustible LEL sensor Oxygen O sensor MICROceL hydrogen sulfide H S sensor MICROceL carbon monoxide CO sensor Model No SR W MP75C AAY80 390R BW SR H MC SR M MC SR DUMM1 SR TOX MC DUMM Replacement dummy O sensor Replacement dummy CO or H S sensor SR W MC DUMM SR X2V Replacement dummy LEL sensor Replacement O sensor Sensor filters MC SS Sensor filters kit of 2 MC2 SS Replacement quad sensor screens GasAlertMicroClip Replacement Parts and Accessories CG Q58 4 Quad gas cylinder CH 2 5 1 O 18 0 H2S 25 ppm CO 100 ppm bal N 58 I 1 CG T34 Dual gas cylinder 50 LEL CH4 2 5 O 20 9 bal No 34 I CG Q34 4 Quad gas cylinder CH 2 5 O 18 0 H2S 25 ppm CO 100 ppm bal N 34 I Quad calibration kit with r
4. 1 Click the Force Calibration When Overdue checkbox to enable 2 Enter a value 1 365 days in the Calibration Interval Cal Interval field A Caution If 0 zero is entered in the Cal Interval field the Force Calibration When Overdue option is automatically disabled The detector is shipped with Force Calibration When Overdue disabled For more information refer to Calibration 26 IntelliFlash When enabled the green LED flashes to provide continuous visual confirmation that the detector is operating correctly Note IntelliFlash is only applicable to GasAlertMicroClip XT and XL IntelliFlash automatically deactivates during a low battery alarm a self test fail a calibration fail a bump test fail or during an alarm event To define how often IntelliFlash occurs 1 60 seconds refer to Confidence Beep and IntelliFlash Interval Default interval is 1 second The detector is shipped with IntelliFlash enabled Latching Alarms When enabled a low alarm persists until the alarm is acknowledged and gas concentrations are below the low alarm setpoint The audible alarm can be temporarily deactivated by pressing but the LCD continues to display the peak concentration values until the alarm condition no longer exists The detector is shipped with Latching Alarms disabled Low Alarm Acknowledge When enabled the audible alarm can be deactivated during a low alarm for the CO H S and LEL sensors The LED
5. From the Devices toolbar click Configure Device via IR Link Configure Device via IR Link 9 When the Device Selection dialog box displays select GasAlertMicroClip Series and click OK Device Selection 2 GasdlertMicroClip Series O GasalertMaz ET O GasAlertQuattra CIR Link GasAlertMicroClip Installing Fleet Manager I Fleet Manager II displays the Sensors tab that includes the following sections e Detector Identification e CO O H S LEL Sensor Configuration e User Options e Language Menu 17 GasAlertMicroClip Technical Reference Guide Using Fleet Manager II to Configure the Detector When Fleet Manager II is installed refer to Figure 3 and the following procedures Figure 3 Connecting the IR Link Table 6 Connecting the IR Link em esco Detector A E IR and charger interface 18 PL 2 Se SS Activate the detector and wait for the startup tests to complete Connect the USB cable to the USB port on the computer Connect the USB cable to the IR Link Insert the IR link onto the IR interface on the back of the detector Open Fleet Manager Il and access the Sensors tab Refer to Installing Fleet Manager Il From the Sensors tab click Retrieve from Device at the bottom of the window The fields will populate with the detector s current configurations Refer to the descriptions in the following sections to define settings and enable disable options
6. GasAlertMicroClip XT and XL Datalog Interval Enter a value 1 120 seconds The default datalog interval is one reading every 15 seconds GasAlertMicroClip User Options Force Bump When Overdue When enabled if a sensor s is past due for a bump test the sensor s must be bump tested immediately otherwise the detector deactivates A bump test should be performed regularly to ensure the sensors are responding correctly to gas The following screen displays when the option is enabled and the sensor s is overdue To enable Force Bump When Overdue complete the following 1 Click the Force Bump When Overdue checkbox to enable 2 Enter a value 1 365 days in the Bump Interval field A Caution If 0 is entered in the Bump Interval field the Force Bump When Overdue option is automatically disabled The detector is shipped with Force Bump When Overdue disabled For information and procedures refer to Bump Test 25 GasAlertMicroClip Technical Reference Guide Force Calibration When Overdue When enabled if a sensor s is past due the sensor s must be calibrated immediately otherwise the detector deactivates Enable Force Calibration When Overdue to ensure calibrations are performed regularly and sensors are operating correctly The following screen displays during the startup sequence when the option is enabled and the sensor s is overdue To enable Force Calibration When Overdue complete the following
7. e Detector Identification e Sensor Configuration CO O H S and LEL e User Options e Language Menu When configuration of new settings is complete click Save to Device at the bottom of the window The detector automatically updates with the new settings Detector Identification The Detector Identification section provides information about the detector current firmware revision and hardware revision Data can also be entered 25 characters per line to display as a startup message on the detector LCD each time it is activated Detector Identification Figure 4 Detector Identification Serial Number Startup Message Top Line Serial Number Enter the serial number of the detector The serial number is located on the back of the detector The serial number is listed above the S bar code This cannot be altered KA410 1020739 SiG SIRE ES S INTINNE LAO AO 5 TENER GasAlertMicroClip Detector Identification Note GasAlertMicroClip serial numbers use the KA serial number prefix Hardware Firmware Revision Hardware Firmware Revision cannot be altered The field automatically populates when data is retrieved from the detector If new firmware is downloaded to the detector the field automatically updates when data is retrieved The firmware version displays on the detector LCD during the startup self tests Startup Message Enter text 25 characters per line to display on the detector LCD during start
8. not calibrate the detector during or immediately after charging Note Calibration can be aborted at any time To abort calibration press The following screen displays DE ES Setting Span Gas Concentration Values 1 Activate the detector and allow startup to complete 2 Connect the IR Link to the computer 38 Insert the IR Link into the IR interface on the back of the detector Refer to Figure 9 Figure 9 Connecting the Detector and IR Link Table 12 Connecting the Detector and IR Link tem escritos IR and charger interface 3 IR Link GasAlertMicroClip Calibration 4 On the PC open Fleet Manager Il 5 Login to the Administration functions From the Devices toolbar click Configure Device via IR Link and select GasAlertMicroClip Series 6 Click _ Retrieve From Device The fields populate with the detector s current settings 7 Refer to Calibration Gas Concentration for span gas values 8 Ensure the sensors to be calibrated are enabled in Fleet Manager Il l 9 Using _ J select the concentration value s in the Cal Gas field for each sensor The values entered in Fleet Manager II must match the gas concentration values on the gas cylinder 10 Click to save the settings to the detector Calibrating with the IR Link A Caution Only calibrate in a fresh air environment and in a safe area Do not calibrate the detector during or immediately after c
9. 4 Click the Save to Device button located at the bottom of the window The detector LCD automatically updates The gas type and sensor readings no longer display on the LCD for the applicable sensor Enable the sensor as soon as possible If the sensor is damaged replace it immediately Refer to Replacing a Sensor or Sensor Filter Calibration Gas Concentration A Caution The gas concentration value entered in Fleet Manager II must match the gas concentration value on the gas cylinder 1 Refer to the following list of recommended gas mixtures e CO 100 ppm balance N e H S 25 ppm balance N e LEL 50 LEL or 2 5 to 25 by vol methane balance air e O 20 9 Select enter the gas concentration value in the Cal Gas field of the applicable sensor GasAlertMicroClip Sensor Configuration Calibration Interval A Caution BW recommends that the sensors be calibrated once every 180 days 6 months Define how often a sensor should be calibrated in the Cal Interval field A different calibration interval can be defined for each sensor 1 Enter the value 0 365 days for each sensor 2 Enter 0 to disable the calibration interval option Entering zero automatically deactivates the Force Calibration When Overdue user option The detector is shipped with the factory default set to 180 days Bump Interval Define how often a bump check should be performed for each sensor in the Bump Interval field A different b
10. 9 Gas Alam SCID ONS dde 32 10 Sample Factory Alarm SetpoiNtS serii enining a a cas 33 11 Connecting the Gas Cylinder to the Detector oocccocccconoconnconcconnoncnonanncnnnonnnonanonnnnonnnnnnnonanonannnnannonanenanenanes 37 12 Connecting the Detector and IR LINK cocococoncconccccnoncnoconononoconnonononononononnnnnonnonnnonanonannnnnnannenannnannennnenanenanes 38 13 Connecting he Charging Ad a la 45 14 Replacing a Sensor or Sensor Filter ooccoconccocncccnoccnoconcconnonnnocononanncnnnonnnonononannnnnnnnnnonnnnnnnnnonnnennninaninaninnns 47 15 TP FOUDISSHOOUING ais 52 16 Replacement Parts and Accessories ccccecccsscceececseecueeceeeceeecueeceueeseeceeseuesaueseeeseeseessueesueeecuessaeessessaaesaaes 57 GasAlertMicroClip Technical Reference Guide Table Title Page vi List of Figures Figure Title Page 1 Parts OT Me Gas AlerivIICLO CID sd A ii 7 2 Bispla EIEMEM S aae a ot 8 3 COMMECUIMIO nel oa 18 4 BETA neat scant tans cade dee anol a a ce calcite bene tea rae ace 19 O CO Sensor Eon QUAN A ene A A OAD ees 20 6 DIG aDISd SONS ant tl a detal cas lo ta le ote artes 20 T Fleet Manager IlI IR Link User ODIOS sip e de 24 8 Connecting the Gas Cylinder to the Detector ooocccoccoconoconcconnonononcnonanocanonnnnonanonnnnnnnonanonannnnnnnnannnnanenanenanes 37 9 Connecting the Detectorand IR LINK a oasis 38 10 Connecting the Charging Adapter coocccoccoconcc
11. Message option 25 characters maximum of Fleet Manager II that data will display during the startup self test To enter a startup message refer to Detector Identification or the Fleet Manager I Operator s Manual GasAlertMicroClip Activating the Detector Alarm Setpoints Sensor and Power Test 4 Next the TWA STEL low and high alarm setpoints display 5 The detector then tests the sensors Note Alarm setpoints may vary by region Refer to Factory Gas Alarm TOC TFTA Setpoints i E INE INN Twa ALARM dls J6 After testing the sensors the following screen displays to verify pom wal nal all sensors have passed E ees lt Mei mm If an error message displays refer to Startup Troubleshooting Note The sensors are tested continuously while the detector is activated 11 GasAlertMicroClip Technical Reference Guide Automatic Zero and O Calibration optional 12 6 Auto Zero on Startup If enabled the H S CO and LEL sensors are automatically zeroed during startup Each sensor is enabled individually The auto zero option is enabled for each sensor upon shipment O Auto Calibration on Startup If enabled the O sensor is automatically calibrated during startup The auto calibration option is enabled for the O sensor upon shipment The detector beeps twice to signal a successful zero and span and the following screen displays AUTO ZERO If the O Auto Calibration on Startup option is
12. Zero and Oxygen Sensor Cala n 40 AA o oe uaccucheacineacnene sas R a E ws eabecseetou ecatecsereeese ace 40 SUCCESS UI SO AM iain sai outa scot e sole ee sesen e e E 41 WNISUGCCCSSIUN S PAM armaa a Taa a a Taa a aceasta tedec Eaa e aaraa 41 Calibrator Due ID AUC o e A 41 Venice a a aa a eaaa a a a E EE a 42 DITaldOS unas a ia aaar inai 43 Event LOGS aasian n EAE EE a o aca 43 Downloading Datalogs ana Event LOGS cia A TE 43 SoftWare IREQUIFEIME INS usas bocina ds 43 NWAINTENANGCG APP eee A iaaiiai 44 Bate Call Sis loaded 44 ta lolis A ea Nee ee eee eer ee ne ere eee 45 Replacing a Sensor or Sensor Filter ars od ri tt 46 Replacing the Sensor Filters crannies a ds ii 48 Replacing the H29 CO and EL SONS ON sc apio doo 48 Replacing MS OVIEDO e O e te 49 FEaASSEMDIING Me GClOClON ns ata dl da 50 TROUDIESMOOUNG ses netas 52 Startup PFOUDIESMOOUING A cade 55 Calibration TroOUDlESNOO IN ibas 56 Replacement Parts and ACCESOS a tada 57 SDCCITICAUIONS mr 60 General Datalogger Specifications cccccccscccssccesceceeeceeeceeeceueecueeceeecseeeseesueesauesaueesaeesaueeaueesaeesaeeecaeeseeesseeseeeeneeesness 61 List of Tables Table Title Page 1 Gases Montored a is 1 2 MMET MATION Al SM Si bn eta de id 4 3 Pals Or the Gas AleriMiIcro Clic 7 4 ISP MEMES do a a ecos 8 O PUSO Ni A EA A 0 AE A AAA AAA ALAS RAE 9 6 COMECINd INSUR A ce hast el ab 18 T A a E E E E A E 29 8 Computed Gas EXPOSES eso cigs ened a o a aL 31
13. an IR device must Replacing a Sensor or Sensor be used to calibrate Filter 55 GasAlertMicroClip Technical Reference Guide Calibration Troubleshooting MO E E E LM A Auto zero Attempt calibration again No Gas Detected Ensure the sensor is enabled Unsuccessful Refer to Calibration If If the gas is not Verify gas cylinder is not H2S CO or LEL ERROR displays again detected about 30 empty or past the expiration sensor fails to auto _ ePlace the sensor Refer to seconds after the date Check replace the zero or O sensor Replacing a Sensor or APPLY GAS message regulator fails to calibrate Sensor Filter is displayed the Attempt calibration again If detector fails the the calibration fails again calibration refer to Replacing a Sensor or Sensor Filter Auto Span Ensure sensor is enabled Calibration Due Date Calibration for the sensor was Unsuccessful Verify gas cylinder is not Overdue unsuccessful The due date H S CO or LEL or empty or past the expiration O sensor fails to date Check replace the A sensor displays a will not reset Attempt negative number for a calibration of the sensor regulator Attempt next due date after again If still unsuccessful calibration again If the calibration is refer to Replacing a Sensor or sensor fails the span again performed Sensor Filter Calibrate the refer to Replacing a Sensor new sensor immediately or Sensor Filter o il 3
14. the detector aaa a proceeds to the Auto Span function H2S ppm CO ppm Auto Zero Unsuccessful If the sensors fail auto zero an error message displays showing which sensor failed Refer to Calibration Troubleshooting 40 When a sufficient amount of gas has been detected approximately 30 seconds after the gas has been applied the detector beeps once flashes and D remains lit while the detector completes the span approximately 2 minutes Mi ep JC i IZAI e A i S HS ppm Q o 3 m Fa a m o Successful Span If the sensors have spanned successfully the detector beeps and the calibration procedure continues Unsuccessful Span If any sensors fail the span the following screen displays Refer to Calibration Troubleshooting GasAlertMicroClip Calibration Due Date Note Calibration If a sensor fails calibration the next due date for that sensor will not reset Refer to Calibration Troubleshooting 6 After calibration is complete CAL DUE displays and all successfully calibrated sensors automatically reset the calibration due dates according to the calibration intervals in Fleet Manager Il Se Y E O laa ro Feat EE mae pas o N R Li bi o o EE m OT m m Lo Si 7 The number of days that displays is when the next sensor calibration should be performed 41 GasAlertMicroClip Technical Refe
15. 1 G Ex ia IIC T4 Ga Sira 13ATEX2330 EN 60079 0 EN 60079 11 EN 60079 26 IECEx Ex ia IIC T4 Ga IECEx CSA 05 0015 IEC 60079 0 IEC 60079 11 IEC 60079 26 KTL 12 KB4B0 0053 This equipment has been tested and found to comply with the limits for a Class B digital device pursuant to Part 15 of the FCC Rules and ICES 003 Canadian EMI requirements These limits are designed to provide reasonable protection against harmful interference in a residential installation This equipment generates uses and can radiate radio frequency energy and if not installed and used in accordance with the instructions may cause harmful interference to radio communications However there is no guarantee that interference will GasAlertMicroClip Specifications not occur in a particular installation If this equipment does cause harmful interference to radio or television reception which can be determined by turning the equipment off and on the user is encouraged to try to correct the interference by one of more of the following measures e Reorient or relocate the receiving antenna e Increase the separation between the equipment and receiver e Connect the equipment into an outlet on a circuit different from that to which the receiver is connected e Consult the dealer or an experienced radio TV technician for help General Datalogger Specifications Storage 16 hours at 15 second intervals Memory type Wraparound memory ensures most recent data is alwa
16. C TC 1 Calibration cap Carrying Accessories GA NS 1 Neck strap with safety release GA LY Short strap 6 in 15 2 cm GA ES 1 Extension strap 4 ft 1 2 m GA ARM 1 Arm band GA CH 2 Chest harness 1 MC2 LC 1 Black leather PVC carrying holster for XT SPAK CC1 Carrying case for Sampling Pump and 1 or detector GasAlertMicroClip Replacement Parts and Accessories MCXL FC1 Replacement front enclosure Yellow 1 compatible with XL only MCXL FC1B Replacement front enclosure Black 1 compatible with XL only MCXL BC1 Replacement back enclosure Yellow 1 compatible with XL only MCXL BC1B Replacement back enclosure Black 1 compatible with XL only MC2 FPCB1 Replacement flex PCB MCXL MPCB1 Replacement main PCB and battery 1 compatible with XL only MC LCD K1 Replacement LCD MC SCREW K1 Replacement screw kit Add one of the following applicable suffixes to the end of the order number to ensure power adapter is correct for region UK for United Kingdom EU for Europe AU for Australia China 59 GasAlertMicroClip Technical Reference Guide Specifications Instrument dimensions XT 11 25 x 6 00 x 2 89 cm 4 4 x 2 4 x 1 1 in XL 11 25 x 6 00 x 3 09 cm 4 4 x 2 4 x 1 2 in Weight XT 170 g 6 0 oz XL 190 g 6 7 oz Operating temperature 20 C to 50 C 4 F to 122 F 50 C to 58 C is certified by CSA International on the combustible sensor with 5
17. Note Wind currents may cause false readings and poor calibrations Note Only use the calibration cap during calibration and bump check Figure 8 Connecting the Gas Cylinder to the Detector GasAlertMicroClip Calibration Table 11 Connecting the Gas Cylinder to the Detector A Y Calibration cap E Calibration hose O F Gas cylinder with 0 5 ml min regulator Read the following steps 1 7 before beginning calibration 1 Verify the calibration gas being used matches the span concentration value s that are set for the detector Refer to Cal Gas in Fleet Manager Il Attach a 0 5 ml min regulator to the gas cylinder To perform an automated calibration use a demand flow regulator and refer to the MicroDock II User Manual Connect the calibration hose to the calibration cap Connect the other end of the calibration hose to the regulator on the gas cylinder Refer to Calibration Setup to apply gas When calibration is complete turn off gas and disconnect the hose from the calibration cap and regulator Ensure the gas cylinder is stored according to the manufacturer s specifications 37 GasAlertMicroClip Technical Reference Guide Calibration Setup The following calibration procedures are written as calibration performance is intended If an error or failure occurs refer to Calibration Troubleshooting A Caution Only calibrate in a fresh air environment and in a safe area Do
18. accuracy Storage temperature 40 C to 50 C 40 F to 122 F Operating humidity 0 to 95 relative humidity non condensing Alarm setpoints May vary by region and are user defined All setpoints automatically display during the startup self test Detection range HS 0 100 ppm 1 0 1 ppm increments CO 0 500 ppm 1 ppm increments O 0 30 0 vol 0 1 vol increments Combustible LEL 0 100 LEL 1 LEL increments or 0 5 0 v v methane Sensor type HS CO O Single plug in electrochemical cell Combustibles Plug in catalytic bead O measuring principle Capillary controlled concentration sensor Alarm conditions TWA alarm STEL alarm low alarm high alarm multi gas alarm over limit OL alarm low battery alarm confidence beep automatic deactivation alarm Audible alarm 95 dB at 30 cm 1 ft 100 dB typical variable pulsed beeper 60 Visual alarm Red light emitting diodes LED Display Alphanumeric liquid crystal display LCD Backlight Activates when the pushbutton is pressed and deactivates after 5 seconds also activates during an alarm condition Self test Initiated upon activation Calibration Automatic zero and automatic span Oxygen sensor Automatic span upon activation enable disable User field options Startup message Confidence Beep latching alarms enable disable safe display mode oxygen measurement combustible sensor measurement sensor disable define calibrat
19. ak exposure level ppm or e Status of the sensor Downloading Datalogs and Event Logs The datalog and event log files can only be downloaded to a PC using an IR Link or the MicroDock II base station Refer to the Fleet Manager II Operator s Manual or MicroDock II User Manual Software Requirements To create spreadsheet reports of event logs datalogs and bump and calibration results the following software applications are required e Fleet Manager ll and e Microsoft Excel 43 GasAlertMicroClip Technical Reference Guide Maintenance To maintain the detector in good operating condition perform the following basic maintenance as required e Calibrate bump check and inspect the detector at regular intervals e Maintain an operations log of all maintenance bump checks calibrations and alarm events e Clean the exterior with a soft damp cloth Do not use solvents soaps or polishes e Do not immerse the detector in liquids Battery Cautions A Warning To avoid personal injury and or property damage adhere to the following e The detector must be deactivated to charge the battery e Charge the battery immediately when the detector emits a low battery alarm Refer to Charging the Battery e Charge the battery in a safe area that is free of hazardous gas in temperatures of 32 F to 113 F 0 C to 45 C e Charge the battery using the BW Multi Unit Cradle Charger or charger adapter only Do not use any other ch
20. and visual alarm indicators remain active until the alarm condition changes or the detector deactivates Press to acknowledge the low alarm and deactivate the audible alarm Note Low Alarm Acknowledge is not applicable to O The detector is shipped with Low Alarm Acknowledge disabled Safe Mode When enabled SAFE displays continuously on the LCD unless an alarm condition occurs Safe Mode provides visual confirmation that no monitored hazardous gas is present The detector is shipped with Safe Mode disabled GasAlertMicroClip User Options Stealth Mode When enabled the backlight visual alarms and audible alarms are disabled displays continuously on the LCD During an alarm the vibrator activates and readings display on the LCD The detector is shipped with Stealth Mode disabled 2 GasAlertMicroClip Technical Reference Guide Language Menu The detector can display warnings and notifications in five different languages Refer to the following illustration Language 2 English Francais Deutsch C Espa ol Portugu s Click a language When the settings are saved to the detector the LCD displays warnings and notifications in the selected language The detector is shipped with English as the default language 28 GasAlertMicroClip Alarms Alarms Table 7 describes the detector alarms and corresponding screens During an alarm condition the detector activates the backlight a
21. are used with a demand flow regulator must meet the following maximum inlet pressure specifications Disposable cylinders 0 1000 psig 70 bar Refillable cylinders 0 3000 psig 207 bar To perform an automated bump check refer to the MicroDock II User Manual Connect the calibration hose to the calibration cap Attach the calibration cap to the detector Apply gas Verify the visual and audible alarms activate Close the regulator and remove the calibration cap from the detector N NOTE The detector will temporarily remain in alarm until the gas clears from ote the sensors Only use the calibration Disconnect the hose from the calibration cap and the regulator ord bee Lead ai 35 GasAlertMicroClip Technical Reference Guide Calibration Guidelines When calibrating the detector adhere to the following guidelines 36 Recommended gas mixture CO 100 ppm balance N HS 25 ppm balance N LEL 50 LEL or 2 5 by vol methane balance air O clean air 18 To ensure accurate calibration use a premium grade calibration gas Gases approved by the National Institute of Standards and Technology NIST improve the accuracy of the calibration Do not use a gas cylinder past its expiration date Calibrate a new sensor before use Install the sensor activate the detector and allow the sensor to stabilize before starting calibration used sensor 60 seconds new sensor 5 minutes Calibrate the sensors at lea
22. arging adapters Failure to adhere to this caution can lead to fire and or explosion e The charging adapter is voltage specific to your region Use of the charging adapter outside your region will damage the charger and the detector e Do not calibrate the detector during or immediately after charging the battery 44 e The battery can only be replaced by the manufacturer Failure to adhere to this caution can lead to fire and or explosion e Warning The GasAlertMicroClip uses a lithium battery that may present a risk of fire or chemical burn hazard if misused Do not disassemble heat above 212 100 C or incinerate e Warning Lithium polymer cells exposed to heat at 266 F 130 C for 10 minutes can cause fire and or explosion GasAlertMicroClip Maintenance Charging the Battery Table 13 Connecting the Charging Adapter To charge the battery refer to Figure 10 Table 13 and the following Hem Description A A Warning The detector must be charged in a safe area that is free of hazardous gas in temperatures of 32 F to 113 F 0 C to 45 C 1 Deactivate the detector 2 Plug the charging adapter into an AC outlet A Caution The charging adapter is voltage specific to your region Use of the charging adapter outside your region will damage the charger and the detector 3 Attach the charging adapter to the charger interface Refer to Figure 10 4 Allow the battery to charge for 2 3 hours The charging i
23. by pressing O and the gas concentration is below the low alarm setpoint The LCD displays the peak concentration and the audible visual and vibrator indicators persist until the alarm condition no longer exists Enable disable Latching Alarms in Fleet Manager ll Local regulations may require Latching Alarms be enabled 29 GasAlertMicroClip Technical Reference Guide Multi Gas Alarm COEN Over Limit OL Alarm on e Sequence alternating low and high alarm TOL ART e Fast siren and alternating flash ALAR siren and flash Lt e and gas bar flash and gas bars flash e Vibrator alarm activates e Vibrator alarm activates e OL displays y m r e S l Sensor Alarm Confidence Beep and IntelliFlash e During startup sequence Error sensor rt e One beep and flash every second name displays Note e During normal operation Err displays must be acknowledged by press _ A The detector is shipped with Confidence Beep disabled and IntelliFlash enabled Low Battery Alarm Automatic Shutdown Alarm e Sequence of 10 rapid sirens and e Sequence of 10 rapid sirens and alternating alternating flashes with 7 seconds of flashes with 1 second of silence in between LEN AH T silence in between continues for sequence reactivates seven times 15 minutes x El LOW BAT and display e Hand flash LOW BAT displays e Vibrator alarm activates and the vibrator alarm activates e OFF di
24. d First Use the detector only as specified in this guide and the operator s manual otherwise the protection provided by the detector may be impaired International symbols on the detector and in this manual are explained in Table 2 Read the Cautions on the following pages before using the detector d naX Warning This instrument contains a lithium polymer battery Dispose of lithium cells immediately Do not disassemble and do not dispose of in fire Do not mix with the solid waste stream Spent batteries must be disposed of by a qualified recycler or hazardous materials handler A Cautions e Warning Substitution of components may impair Intrinsic Safety e Before using the detector refer to Sensor Poisons and Contaminants e Warning For safety reasons this equipment must be operated and serviced by qualified personnel only Read and understand the user manual completely before operating or servicing Do not use the detector if it is damaged Inspect the detector before using Look for cracks and or missing parts e If the detector is damaged or parts are missing contact BW Technologies by Honeywell immediately e Only use sensor s that are specifically designed for the GasAlertMicroClip Refer to Replacement Parts and Accessories e Calibrate the detector before first time use and then on a regular schedule depending on use and sensor exposure to poisons and contaminants BW recommends calibrating at least once ev
25. dence Beep and IntelliFlash Backlight The backlight automatically activates e during startup e when the pushbutton is pressed then deactivates after 5 seconds and e when there is an alarm condition unless Stealth is enabled 15 GasAlertMicroClip Technical Reference Guide Deactivating the Detector To deactivate the detector press and hold O The detector e performs a sequence of two sirens with alternating flashes e vibrates e initiates the deactivation countdown and e displays OFF Note IFO is released before the countdown is complete the detector will not deactivate 16 Installing Fleet Manager Il Fleet Manager II is required to configure the detector and sensors And IR Link is also required To purchase contact BW Technologies by Honeywell To install Fleet Manager ll complete the following 1 oe w W Install Fleet Manager II using the Fleet Manager CD ROM available with MicroDock II and the IR Link or download at no cost from BW Technologies by Honeywell website www gasmonitors com Follow the installation wizard When installation is complete open Fleet Manager II Click Administration located on the left toolbar Click the Login Logout button Login Logout 6 When the Password dialog box displays enter Admin password is case sensitive Enter Password Please enter the administrator password to gain access to all areas of Fleet Manager II 7 Click OK 8
26. ector during repair attempts Non BW Technologies by Honeywell repair service attempts void this warranty GasAlertMicroClip Technical Reference Guide Table 2 International Symbols lt P Approved to both U S and Canadian Standards by CSA International Cc US Ko European Explosive Protection Conforms to European Union Directives ATEX Conforms to European ATEX Directives International Electrotechnical Commission Scheme for Certification to Standards for Electrical Equipment for Explosive IECE Atmospheres Conforms to Korea Testing Laboratory KTL Certification Sensor Poisons and Contaminants Several cleaners solvents and lubricants can contaminate and cause permanent damage to sensors Before using cleaners solvents and lubricants in close proximity to the detector sensors read the following cautions and refer to the lists below A Caution Use only the following BW Technologies by Honeywell recommended products and procedures Use water based cleaners e Use non alcohol based cleaners e Clean the exterior of the detector with a soft damp cloth e Do not use soaps polishes or solvents Below are common products to avoid using around sensors Cleaners and Lubricants Brake cleaners Lubricants Rust inhibitors Window and glass cleaners Dishsoaps Citrus based cleaners Alcohol based cleaners Hand sanitizers Anionic detergents Methanol fuels and antifreezes GasAlertMicroClip Sensor P
27. egulator 1 quad gas cylinder CG Q58 4 hose and carrying case CG2 M 200 103 G0042 H25 Single gas cylinder H2S 25 ppm bal N 58 I P Single gas cylinder CO 100 ppm 1 bal N 103 I 57 GasAlertMicroClip Technical Reference Guide CG BUMP1 Bump alarm gas aerosol CH 2 5 O 10 H2S 40 ppm CO 200 ppm Charger and Accessories MC2 CO1 MC5 GA PA 1 MC5 GA PA 3 GA PA 1 GA VPA 1 Confined space kit MC CK DL MC CK CC Multi unit 5 cradle charger Multi unit wall outlet power adapter 12 24 VDC direct wire power adapter Replacement wall outlet power adapter 12 24 VDC Vehicle power adaptor Deluxe confined space kit Carry case and foam insert compatible with XL MicroDock ll and Module DOCK2 2 1C1P 00 G MicroDock II Automatic Test and Calibration System GasAlertMicroClip XT module and charging cable DOCK2 2 1P 00 G MicroDock II Automatic Test and Calibration System and GasAlertMicroClip XT module DOCK2 0 1P 00 G GasAlertMicroClip XT docking module 1 58 Datalogging Accessories GA USB1 IR IR Connectivity Kit 1 with Fleet Manager II Sampling Testing Equipment GA SPAKO2 SamplerPak motorized sampling 1 pump kit compatible with XT GA SPAK SamplerPak motorized sampling 1 pump kit compatible with XL GA SPAK NA SamplerPak motorized sampling pump 1 rechargeable kit compatible with XL MC AS01 Manual aspirator pump kit with probe 1 1 ft 0 3 m M
28. enabled and the Auto Zero on Startup option is disabled for all sensors the following screen displays AUTO ZERO Note If oxygen is configured to measure 20 8 vol the oxygen calibration screen displays 20 8 O3 Calibration Due Date optional 7 The following screen displays the number of days remaining before calibration is due The number of days that displays is when the next sensor calibration should be performed Note If the Calibration Interval option is defined as 0 the calibration due date is bypassed during startup Last Calibration Failed optional If any sensor failed the last calibration CAL FAILURE displays on the screen Note When the CAL FAILURE displays the previous calibration has failed but the calibration is still valid until the next calibration due date Overdue Calibration If any sensor is past due for calibration the detector beeps flashes vibrates and the following screen displays If calibration is overdue and the Force Calibration When Overdue option is enabled a calibration must be performed to enter normal operation Refer to Calibration GasAlertMicroClip Activating the Detector Note If calibration is not performed or O is not pressed within 2 minutes the detector automatically deactivates If the Force Calibration When Overdue is disabled press to acknowledge the warning The detector continues with the startup self tests and then
29. enters normal operation Cal IR Lock If the Cal Lock option is enabled the following screen displays Refer to Startup Troubleshooting Bump Test Note A bump test cannot be conducted if the detector has just been calibrated If the Bump Interval option is defined as 0 in Fleet Manager II the bump test is bypassed BW recommends to bump test the sensors before each day s use to confirm their ability to respond to gas by exposing the detector to a gas concentration that exceeds the alarm setpoints 13 GasAlertMicroClip Technical Reference Guide Last Bump Test Failed If any sensor failed the last calibration CAL FAILURE displays on the screen Note When the BUMPCHK FAILURE displays the previous calibration has failed but the calibration is still valid until the next bump test due date Force Bump optional 8 Ifthe Force Bump option is enabled and the sensors are due have a bump test the following screen displays 14 A bump test must be performed to enter normal operation Apply gas to the sensors Ensure the visual audible and vibrator alarms activate When the gas is removed the detector briefly remains in alarm until the gas has cleared from the sensors When the sensors successfully pass the bump check the following screen displays showing the number of days remaining until the next bump check is due I d 1 day If Force Bump is disabled press to continue w
30. er refer to e Table 14 e Figure 11 e Figure 13 and e the following procedures GasAlertMicroClip Maintenance Table 14 Replacing a Sensor or Sensor Filter tem Description e emes NS Carbon monoxide CO sensor os Hydrogen sulfide H S sensor 1 Deactivate the detector On a clean surface place the detector Figure 11 Replacing a Sensor or Sensor Filter face down 2 Remove the six machine screws on the rear shell 3 Remove the back cover by lifting the top and the bottom upwards simultaneously to prevent damaging the charger pins 47 GasAlertMicroClip Technical Reference Guide Replacing the Sensor Filter 1 Note the placement of the PCB to ensure it is replaced correctly Remove the two screws on the PCB Remove the PCB carefully A Caution Ensure no damage occurs to the battery 2 Remove the old sensor filter It may be stuck to the sensors 3 Insert the new sensor filter Note When inserting a new sensor filter ensure the black gasket is facing the front shell 4 To complete the detector refer to Reassembling the detector 48 Replacing the H S CO and LEL sensor 1 Note the placement of the PCB to ensure it is replaced correctly Remove the two screws on the PCB Remove the PCB carefully A Caution Ensure no damage occurs to the battery If the sensor filter is stuck to the sensors remove and replace the
31. ery 180 days 6 months e BW recommends to bump test the sensors before each day s use to confirm their ability to respond to gas by exposing the detector to a gas concentration that exceeds the alarm setpoints Manually verify that the audible and visual alarms are activated Calibrate if the readings are not within the specified limits e BW recommends the combustible sensor be checked with a known concentration of calibration gas after any known exposure to catalyst contaminants poisons sulfur compounds silicon vapors halogenated compounds etc e The combustible sensor is factory calibrated to 50 LEL methane If monitoring a different combustible gas in the LEL range calibrate the sensor using the appropriate gas e Caution High off scale readings may indicate an explosive concentration e Only the combustible gas detection portion of this instrument has been assessed for performance by CSA International e Protect the combustible sensor from exposure to lead compounds silicones and chlorinated hydrocarbons e Sensor exposure to certain organic vapors such as leaded gasoline and halogenated hydrocarbons may temporarily inhibit sensor performance After exposure a bump test or calibration is recommended e For use only in potentially explosive atmospheres where oxygen concentrations do not exceed 20 9 v v e Any rapid up scaling reading followed by a declining or erratic reading may indicate a gas concen
32. etween continues for 15 minutes ALARM 34 e If battery power becomes critically low and LOW BAT display The detector performs a sequence of 10 rapid sirens with 1 second of silence in between sequence reactivates seven times The detector then displays OFF and the detector deactivates Charge the battery immediately Refer to Charging the Battery Note Confidence Beep automatically disables during a low battery alarm Automatic Deactivation Alarm An automatic deactivation alarm will occur if the battery voltage is too low to operate the detector calibration is due but not performed when the Force Calibration option is enabled bump test is due but not performed when the Force Bump options is enabled and e all sensors fail during the startup self test The detector performs a sequence of 10 rapid sirens with alternating flashes with 1 second of silence in between Sequence reactivates seven times OFF then displays and the detector deactivates GasAlertMicroClip Bump Test Bump Test Gas Cylinder Guidelines Bump Test e To ensure an accurate bump check use a premium grade gas Use gases approved by the National Institute of Standards and Technology e Do not use a gas cylinder that is past its expiration date Gas Cylinder Connection 1 Connect the calibration hose to the 0 5 l min regulator on the gas cylinder For use with the MicroDock II use a demand flow regulator NOTE Cylinders that
33. g average that accumulates over an 8 hour average If the worker is in the field longer the oldest accumulated values first hour are replaced by the newest values ninth hour This continues for the duration of the work shift until the detector is deactivated TWA Alarm applies to CO and H S sensors only 1 Refer to Factory Gas Alarm Setpoints for the factory alarm setpoints 2 Enter the setpoint in the TWA Alarm field ALARM wa a H2S ppm CO ppm STEL Alarm The short term exposure limit STEL is the maximum permissible gas concentration a worker can safely be exposed to for short periods of time 5 15 minutes maximum STEL Alarm applies to CO and H S sensors only Note Standard factory Alarm Setpoints vary by region Refer to Factory Gas Alarm Setpoints for OSHA factory settings 1 Refer to the applicable regulatory requirements in your area for defining STEL alarm setpoints 2 Enter the setpoint for the CO and H S sensor in the STEL Alarm field Proceed to STEL Interval STEL Interval STEL Interval provides protection for workers from over exposure to high concentrations of gas and is based on used defined 5 15 minute intervals When the maximum STEL is reached the detector alarms to notify the worker A Caution Follow all safety procedures as defined by your employer Enter the interval 5 15 minutes in the STEL Interval field The detector is shipped with the factory default setting of 15 minutes
34. harging To calibrate the detector with the IR Link complete the following procedure 1 Complete steps 1 10 under Setting Span Gas Concentration Values 2 Click on the Device Operations tab 3 Click Calibrate The detector begins calibration Refer to Auto Zero and Oxygen Sensor Calibration Calibration Procedure A Caution Only calibrate in a fresh air environment and in a safe area Do not calibrate the detector during or immediately after charging 1 Press and hold The detector performs the OFF countdown Continue holding as the detector briefly deactivates 2 The detector then reactivates and performs the CAL countdown Continue holding until the CAL countdown is complete Note IFO ts not held for the entire countdown the detector will deactivate 39 GasAlertMicroClip Technical Reference Guide Auto Zero and Oxygen Sensor Calibration Auto Span Note 4 When auto zero is complete APPLY GAS displays Do not apply calibration gas until APPLY GAS displays otherwise the auto zero function will fail 3 flashes while the detector zeroes the combustible and toxic sensors and calibrates the oxygen sensor 5 Attach the calibration cap to the detector Refer to Figure 8 Open the valve on the regulator and apply gas at a flow rate of 250 500 ml min Do flashes and displays When auto zero is complete the detector beeps twice Auto Zero Successful If the sensors successfully zero
35. iefly deactivates The LCD reactivates and then begins the CAL countdown Release _ when the CAL countdown is complete To activate the backlight in normal operation press To acknowledge latched alarms press To acknowledge a low alarm and disable the audible alarm press if the Low Alarm Acknowledge option is enabled GasAlertMicroClip Technical Reference Guide Activating the Detector A Caution Only activate the detector in a fresh air environment and ina safe area To activate the detector press Self Test The following startup tests are written as startup performance is intended If an error occurs refer to Startup Troubleshooting When the detector is activated it performs several startup tests Confirm the following tests occur Battery Test The detector performs a battery test during startup If the battery has insufficient power to operate the following screen displays Charge the battery for 2 3 hours before restarting the detector Refer to Charging the Battery 10 Audible Visual Test 1 All of the LCD elements display simultaneously as the detector beeps flashes vibrates and activates the backlight aba ALARM amp aa ae TIE SN Ee Ane SEN ANA NA IN INTA INTA IN ZN IZNIIZN AN ZN AN KN HS ppm BHBBBGE Detector Version 2 The current firmware version of the detector then displays on the LCD Startup Message 3 If data is entered in the Startup
36. ion interval force calibration calibration lock force bump define bump interval bump due lock stealth mode low alarm acknowledge language selection enable disable automatic oxygen calibration enable disable auto zero at startup define alarm setpoints span concentration values define STEL calculation period IntelliFlash and Confidence Beep and IntelliFlash Interval Battery operating time 1 rechargeable lithium polymer battery 10 hours typical Year of manufacture The detector s year of manufacture is determined from the serial number The second and third number after the letters determines the year of manufacture E g KA410 000001 2010 year of manufacture Approved battery Approved batteries for GasAlertMicroClip XT Narada NL 503759 Approved batteries for GasAlertMicroClip XL Narada NLP883759LT20 Rechargeable battery Lithium polymer Temperature code 20 C lt Tas 50 C T4 Battery charger GasAlertMicroClip charging adapter A Warning Charge only in a safe area that is free of hazardous gas and within temperatures of 0 C to 45 C 32 F to 113 F First time charge XT 2 3 hours XL 5 6 hours Normal charge XT 2 3 hours XL 5 6 hours Warranty 2 years including sensors Approvals Approved by CSA to both U S and Canadian Standards CAN CSA C22 2 No 157 and C22 2 152 ANSI UL 913 and ANSI ISA 12 13 01 Part 1 CSA Class I Division 1 Group A B C and D ATEX CE 0539 amp II
37. ith the startup self tests Note If BUMPCHK todAY displays again after performing a bump check refer to Startup Troubleshooting Self Test Pass When the detector has passed all startup self tests it enters normal operation The LCD displays the ambient gas readings pa tr o m tS S The detector automatically begins e recording the peak maximum gas exposure e calculating the short term exposure level STEL and e calculating the time weighted average TWA exposures Self Test Fail If the following error message displays after entering normal operation refer to Startup Troubleshooting GasAlertMicroClip Activating the Detector Battery Test The battery is tested when the detector is activated and continuously thereafter A newly charged battery will typically operate for 10 12 hours e Battery power is continually displayed during normal operation If battery power is low and flashes The detector performs a sequence of 10 rapid sirens and alternating flashes with 7 seconds of silence in between continues for 15 minutes If battery power becomes critically low and LOW BAT display The detector performs a sequence of 10 rapid sirens with 1 second of silence in between sequence reactivates seven times The detector then displays OFF and the detector deactivates Note If enabled Confidence Beep and IntelliFlash automatically deactivate during a low battery alarm Refer to Confi
38. lty sensor s Replace the sensor Refer to Replacing a Sensor or Sensor Filter Features and options are not operating as Changes have been made in Fleet Manager ll Verify settings in Fleet Manager Il are correct expected Charging Battery has been charging for 3 hours Battery is trickle charging Battery is fully charged and ready for operation XT model or 6 hours XL model The charging indicator on the detector LCD shows the battery is still charging Battery indicator does not display when Detector is depleted below normal levels Charge the battery for 8 hours Detector LEDs may charging light during first 5 hours This is normal If the battery indicator does not light after charging for 8 hours contact BW Technologies by Honeywell When detector is activated after charging Battery is defective Contact BW Technologies by Honeywell the battery indicator does not display 54 GasAlertMicroClip Startup Troubleshooting Startup Troubleshooting Mo LE ME UA E A E Sensor Error Calibrate the sensor s IR Lock Enabled Perform calibration using the The sensor failed Refer to Calibration If the IR Lock screen IR Link with Fleet Manager II during the self test Reactivate the detector If displays an IR device software or insert the detector error displays again replace is required to calibrate into the MicroDock II station the sensor Refer to the sensors Refer to Cal Lock Replacing a Sensor o
39. n roughly the same spot on the front shell place the new sensor Lower the PCB over the oxygen sensor 49 GasAlertMicroClip Technical Reference Guide Reassembling the detector XT Model 1 To re assemble the detector perform the following e Verify the PCB is seated correctly and inserted exactly as it was removed sensors facing the front shell e Replace the two PCB screws e Visually inspect the battery to ensure no damage has occurred e When replacing the rear shell ensure the charging pins bottom of inside rear shell are aligned with the corresponding holes on the PCB If the contact pins are bent the battery will not charge correctly Note Ensure the rib on the interior rear shell O inserts between the battery and the PCB 2 Figure 13 Replacing the Rear Shell 50 e Press the front and rear shells together firmly to ensure a proper seal Ensure the front and rear shell have a uniform tight 1 16 in 1 mm seal on all sides of the detector e When replacing the screws they must be seated properly to prevent cross threading Turn the screw counter clockwise until a click is heard and then begin tightening the screw clockwise New sensors must be calibrated Activate the detector and calibrate the sensor s Refer to Calibration GasAlertMicroClip Maintenance 51 GasAlertMicroClip Technical Reference Guide Troubleshooting If a problem occurs refer to the solutions in the Troublesho
40. ndicator flashes on the LCD while the detector is being charged h mm low mid high Figure 10 Connecting the Charging Adapter 45 GasAlertMicroClip Technical Reference Guide 46 5 When charging is complete the charging indicator stops flashing and displays m to indicate a full charge Remove the charging adapter and activate the detector If the battery indicator does not display refer to Troubleshooting 6 Charge the battery after each workday Note To reach full battery capacity allow the battery to fully charge and fully discharge three times Charging the detector in temperatures above 45 C 113 F will greatly reduce the number of charges the detector can accept The detector may be warm immediately following charging This is normal Replacing a Sensor or Sensor Filter A Warning To avoid personal injury only use sensors that are specifically designed for the detector Refer to Replacement Parts and Accessories Use proper ESD handling practices e Each sensor has a high degree of resistance to common vapors and gases To clear a sensor move the detector to a non hazardous environment and wait 10 to 30 minutes e Do not expose a sensor to vapors of inorganic solvents such as fumes from paint thinners or organic solvents such as benzoic acids and acrylic acids e Ensure hands are clean or wear gloves before handling components To replace a sensor or sensor filt
41. o aa iris 25 Force Calibration VV MEM OVETOUC o le E 26 BRIVIS il cure scabpcctecth non amuntes irecas essen avataeasmencigean a ease sssanamncteatoansean enna thes aiuainces a aR 26 Lateng Ala M Saaria esata a faced iaeiesaeta nahin a aulenle hash ues iey sansiuuk ye twaieh janamninm auld ieee annual binaetwereasau 26 LOW ALARM ACKNOW d GO cissie aa E a a al 27 A o A A are cee 27 O E a E 27 PAN GUAGE Meninas 28 A A N E E A E E A A E E E E E P lscdeusuaws E E S E 29 Computed GaS EXPOSUOS lena 31 Viewing GaS EXPOS UOS mnie nn ana iR e R a aa aa OA EE a ia iE 31 Glearing Gas EX DO SU CS rsr A lid 32 cas Aldim SeN cer a a a E I 32 Factory GaS Alarm ClDOINS sou Sassen leet sets eee Ea ee 33 Changing Alai SU ONE cepto reabre eee ta es daros 33 Stopping a Gas A ali 33 ANA A a a a a a 34 LON Bary AlI MRa di id ia 34 Automatic De acia Alarm cere a ii N 34 BUMP estaa a atan 35 Gallbra lO Micra R est acu meubensia auinewssaoinenasmeceetenessadvecens 36 o tec N enews hick alle al ach ancausslueeg aa mee nasa inant 36 DIagnostes TeS Essen dol dd E e a di tte nl Aba tt SD eso sane bd 36 Connecting the Gas Cylinder to the Detector cooccoocncccncconcconnconnoconocanonanononnonnnonanonnnonnnnnnnonnnnnnnnnnnnnnnnnnnninaninaninns 37 Salas A O a a mead teu ese 38 Setting Span Gas Concentration Values ui A as 38 GasAlertMicroClip Technical Reference Guide Title Page calibrating WISE 39 Calibr aton PFOCCOUIC sad lisis 39 Auto
42. oints Gas alarms are activated when detected gas concentrations are above or below the user defined setpoints Gas alarms are described below Table 9 Gas Alarm Setpoints Low Toxics and combustibles Ambient gas level above low alarm setpoint Oxygen Ambient gas level may be set above or below 20 9 or 20 8 High Toxics and combustibles Ambient gas level above high alarm setpoint Oxygen Ambient gas level may be set above or below 20 9 or 20 8 Toxics only Accumulated value above the TWA alarm setpoint STEL Toxics only Accumulated value above the STEL alarm setpoint Downscale Toxics only lf sensor reading is negative half of the TWA setpoint Multi gas Two or more gas alarm conditions Over Limit OL OL displays when readings are above or below the sensor detection range Refer to specifications for detection ranges Factory Gas Alarm Setpoints Note Standard factory alarm setpoints may vary by region Table 10 lists the factory alarm setpoints as defined by Occupational Safety and Health Association OSHA Table 10 Sample Factory Alarm Setpoints Gas TWA sei iow High 19 5 vol 23 5 vol Note To disable an alarm set the alarm setpoint to 0 zero in Fleet Manager II Refer to Using Fleet Manager II to Configure the Detector for complete instructions GasAlertMicroClip Alarms Changing Alarm Setpoints To change alarm setpoints use the base station or IR Link and refer to
43. oisons and Contaminants Silicones e Silicone cleaners and protectants e Silicone based adhesives sealants and gels e Hand body and medicinal creams containing silicone e Tissues containing silicone e Mold releasing agents e Polishes Aerosols e Bug repellents and sprays e Lubricants e Rust inhibitors e Window cleaners GasAlertMicroClip Technical Reference Guide Getting Started The list below provides the standard items included with the detector If the detector is damaged or parts are missing contact the place of purchase immediately e Sensors H S CO O and combustible LEL e Calibration cap and hose e Charging adapter e Printed Operator s Manual e Supplementary Booklet including a Quick Reference Card e CD ROM including translated operator s manuals Configuration Software The detector is configured with Fleet Manager Il software It can be downloaded for free from BW Technologies by Honeywell website www gasmonitors com Fleet Manager II CD ROM is shipped with the MicroDock II base station and IR Link kit The detector is shipped with the sensors and rechargeable battery installed Battery Replacement To replace the battery contact BW Technologies by Honeywell The battery can only be replaced by the manufacturer Charge Battery and Replace Sensors To charge the battery and replace the sensors and or sensor filter refer to the following e Battery Cautions e Replacing a Sensor or Sensor Filter To order re
44. onccocncncnonnnoconoconconnnonnnonnnonnnrnnnrnnnnnnnnrnnnenannnnnennnennnnnnnennnennnenanes 45 T1 Replacing a Sensor OF Sensor Fiter as a a e a eden 47 12 o a aaa aa a teceen ace 48 13 Replacing ME Rear nal aaa o ci 50 vii GasAlertMicroClip Technical Reference Guide Figure Title Page viii Introduction A Warning To ensure personal safety read Safety Information Read First and the Cautions before using the detector The GasAlertMicroClip XT and XL the detector warns of hazardous gas at levels above user defined alarm setpoints The detector is a personal safety device It is your responsibility to respond properly to the alarm Table 1 lists the gases monitored GasAlertMicroClip Zeroing the Sensors To zero the sensors refer to steps 1 3 in Calibration Procedure Table 1 Gases Monitored Combustible gases LEL 1 percent of lower explosive Field selectable for limit LEL 2 percent by volume methane 0 5 0 v v GasAlertMicroClip Technical Reference Guide Contacting BW Technologies by Honeywell To contact BW Technologies by Honeywell call USA 1 888 749 8878 Canada 1 800 663 4164 Europe 44 0 1295 700300 Other countries 1 403 248 9226 Address correspondence to BW Technologies by Honeywell 2840 2 Avenue S E Calgary AB T2A 7X9 CANADA Email info gasmonitors com BW Technologies by Honeywell s website www gasmonitors com ISO 9001 Safety Information Rea
45. oting section If the problem persists contact BW Technologies by Honeywell Table 15 Troubleshooting The detector does not activate Depleted battery Charge battery Refer to Charging the Battery Damaged or defective detector Contact BW Technologies by Honeywell when activated New sensor wait 5 minutes Hazardous environment Leave the area immediately Deactivate and reactivate in a safe area that is free of hazardous gas in an atmosphere of 20 9 oxygen The activation self test fails General fault Contact BW Technologies by Honeywell Sensor failure Replace the sensor Refer to Replacing a Sensor or Sensor Filter Detector automatically deactivates during Battery power too low to operate Charge battery Refer to Charging the Battery startup Force Calibration When Overdue option is Calibrate the sensor s immediately Refer to enabled and calibration is not attempted Calibration Force Bump When Overdue option is Bump test the sensor s immediately Refer to enabled and a bump test is not attempted Bump Test 52 GasAlertMicroClip Troubleshooting Table 15 Troubleshooting Detector Operation Detector does not display expected gas Sensor not stabilized Used sensor wait 60 seconds readings after activation self test New sensor wait 5 minutes Sensor s requires calibration Calibrate the sensor s Refer to Calibration Target gas is present Detector is operating properly Use caution in suspect areas De
46. placement parts refer to Replacement Parts and Accessories To become oriented with the features and functions of the detector refer to the following figures and tables e Figure 1 and Table 3 describes the detector s components e Figure 2 and Table 4 describes the detector s display elements e Table 5 describes the detector s pushbutton GasAlertMicroClip Parts of the GasAlertMicroClip Parts of the GasAlertMicroClip Table 3 Parts of the GasAlertMicroClip em eses O Hydrogen sulfide H S sensor Figure 1 Parts of the GasAlertMicroClip GasAlertMicroClip Technical Reference Guide Display Elements AS E ALARM A AUTO SPAN gt a T N m A O PL LS AN gt S22 Cis ES np ES O O 3 3 p Pte J ME O AYOLEL9 Y Figure 2 Display Elements Table 4 Display Elements em eses entero 5 atomica sponsor GasAlertMicroClip Pushbutton Pushbutton Table 5 Pushbutton Pushbutton Description IC To activate the detector press To deactivate the detector press and hold gt until the OFF countdown is complete and the LCD deactivates To view the TWA STEL and peak maximum readings press twice To clear the TWA STEL and peak readings press when the LCD displays RESET To initiate calibration deactivate the detector Press and hold _ while the detector performs the OFF countdown Continue holding C while the LCD br
47. r Calibration IR Lock in User sensor Filter Options and Calibration Calibration Overdue Press C to continue and Bump Test Fail Perform another Bump Test Displays when calibrate the sensor s A bump test has just Ensure the cylinder is not calibration is immediately Refer to been performed The empty and that the cylinder is overdue If the Force Calibration detector is prompting not past the expiry date Calibration When If the IR Lock enabled for another bump test Ensure the regulator is fully Overdue option is screen displays the because a sensor s __ opened to apply gas If Bump enabled the MicroDock II station or the has failed Check Today displays again sensor s must be IR Link with Fleet Manager calibrate the sensors Refer to calibrated to enter Il must be used to calibrate Calibration If the calibration is normal operation unsuccessful refer to Replacing a Sensor or Sensor Filter Forced Calibration Press and hold to Sensor Fail Perform a Bump Test and If Force Calibration calibrate the sensors or A sensor has failed reactivate the detector If the When Overdue is press _ and release to during the startup sensor fails again perform enabled the sensors deactivate the detector r Lt self test Calibration Reactivate the must be calibrated to Refer to Calibration If the detector again If the sensor enter normal IR Lock enabled screen y still does not pass refer to operation displays
48. rce BUM OPONA iaa 14 AN O A ener T eee eee ae 15 GasAlertMicroClip Technical Reference Guide Title Page A ee ee eee eee 15 Baen TES unica sind e a bonosta toco caca dba 15 e o PI ae slalePaasee a eenastl 15 DCACHVAUING TNC Detector nainen ona a E A aude ceetensveneneevetaes 16 INSTALLING FICCE Manager Mlasiinarran a A Ri ceive eee 16 Using Fleet Manager ll to Configure the Detector occcoccccccncccncocnccccncocnonononnnononnnnnnnnnnnnnnnonnnnnnnonnnnnnnnnnnnnnnanenaninninons 18 Detector IGSENTITIC ATION lt a As 19 SOTA NUM DGI srera aaea aA eE a aa iid 19 Hardware FINMWatre RYU oi ola 19 Staup MESS AS E 19 Sensor Configuratio M ni dic 20 Sensor DISADIO Oso illinois 20 Calibration Gas CONCE ac 21 Gallbra ON NCCT alucino isos 21 BUMD EMS Vado ia 21 A a a coees tees 22 FONALAT dpe oe AAA dd 22 A OES A OAC Were en enT een eee emacs Tee 22 STELLA AA A A SA 22 AA A 23 AUTO Zero ON AU A ol 23 O2 Auto Calibration on Startup Automatic O CalibratiON oooocconncccnconcncnncnnoncnnnnnancnconnnccnnonnnnnnnnnnnncnnonnnnrncnnnnncnnnnnns 23 LE Te By VOCHA o R A A EA 23 USA ta 24 Cal Lock Callbratlonm IR LOCK ni ana 24 COMMCGCNGCE SP 25 Confidence Beep and IntelliFlash Interval coocccccncccnccccnononcconononccnnnononocanonnnnnnnnonononannnnnnnnnnnonanonanenannnnanenanes 25 GasAlertMicroClip Technical Reference Guide Title Page Datal0d MEVA si ii died ridad 25 Force BUMP When OVEFUC ui ii
49. rence Guide Failed Sensor Past Calibration Due Date If a sensor fails the span and it is past the calibration due date the following three screens display 42 1 Press O to acknowledge the warning The detector returns to normal operation Note A negative number indicates that sensor is overdue for a calibration Verification 1 After calibration is complete and the detector returns to normal operation verify the calibration using a gas cylinder other than the one used for calibration The gas concentration should not exceed the sensor s detection range Confirm the LCD shows the expected concentration To ensure the readings are accurate apply the verification gas for the same amount of time as was applied to the sensor when it was calibrated Example H S span time 2 minutes therefore apply verification gas for 2 minutes Datalogs The detector records various information that can be compiled to create a report The detector is capable of storing 16 hours of information when recording a datalog every 15 seconds When the memory is full the detector replaces the oldest datalogs with the most recent datalogs To change the datalog interval refer to Datalog Interval GasAlertMicroClip Datalogs Event Logs The detector records the 10 most recent gas alarm events The following information is recorded e Serial number of the detector e Start time of alarm e Type level and duration of alarm e Pe
50. sensor filter into the front shell 2 Slide the sensors out Note Detectors that are configured for 1 2 or 3 gases may contain a dummy sensor in one of the four sensor locations 3 Insert the new sensor s For sensor positioning refer to Figure 12 Figure 12 Sensor Positioning 4 To complete the detector refer to Reassembling the detector GasAlertMicroClip Maintenance Replacing the Oxygen Sensor 5 Carefully replace the circular rigidified flex PCB atop the metal sensor posts Ensure the plastic sensor post is inserted into the Note clear plastic hole Take care not to tear the flex cable Detectors that are configured for 1 2 or 3 gases may contain a 6 Press down to secure the circular rigidified flex PCB atop the dummy sensor in one of the four sensor locations metal sensor posts Take care not to press down too hard and 1 Gently remove the circular rigidified flex PCB atop the sensor accidentally activate the detector from the metal sensor posts Take care not to tear the flex cable 7 To complete the detector refer to Reassembling the detector Plastic sensor post Rigidified flex PCB Metal sensor posts Oxygen sensor Flex cable 2 Note the placement of the PCB to ensure it is replaced correctly Remove the two screws on the PCB A Caution Ensure no damage occurs to the battery 3 Lift the PCB straight up The oxygen sensor will stay in the front shell Remove the sensor 4 O
51. splays before deactivating e After 15 minutes of the Low Battery alarm the Automatic Shutdown Alarm Normal Shutdown sequence begins e Sequence of two sirens and alternating flashes OFF displays before deactivating e Vibrator alarm activates e Countdown initiates e OFF displays 30 GasAlertMicroClip Alarms Computed Gas Exposures Viewing Gas Exposures A Warning To view the TWA STEL and peak maximum readings press O twice To avoid possible personal injury do not deactivate the The LCD first displays the TWA gas exposures detector during a work shift TWA STEL and MAX readings reset once the detector is deactivated TWA Table 8 Computed Gas Exposures nn m TWA Time weighted average TWA based on H2S and CO accumulated exposure to toxic gases only averaged over a workday according to US OSHA method OSHA 8 hour moving average STEL Short term exposure limit STEL to gas H2S and CO based on a 5 15 minute user defined only period Peak Peak concentration encountered during maximum work shift For oxygen it is the highest or the lowest concentration encountered 31 GasAlertMicroClip Technical Reference Guide Clearing Gas Exposures A Caution Follow all safety procedures as defined by your employer Confirm with your supervisor before clearing TWA and STEL alarms To clear the TWA STEL and peak exposure readings press O when the LCD displays RESET 32 Gas Alarm Setp
52. st once every 180 days depending on use and sensor exposure to poisons and contaminants Calibrate the detector if the gas readings varies during startup Calibrate the sensor before defining the alarm setpoints Calibrate only in a safe area that is free of hazardous gas in an atmosphere of 20 9 oxygen Do not calibrate the detector during or immediately after charging is complete The oxygen sensor can be automatically calibrated each time upon activation if this feature is enabled Activate the detector in a normal 20 9 20 8 oxygen atmosphere Allow the detector to stabilize for 1 minute after activation before performing a calibration or bump test e If a certified calibration is required contact BW Technologies by Honeywell Diagnostics Test The detector tests the air auto zero and the span gas that is applied auto span to ensure it meets expected values Auto zero sets the zero gas level of the sensor Auto Zero If target gas is present the zero level will be incorrect and the sensor will fail If a sensor fails an error message displays Auto Span If the target gas does not meet expected values an error message displays A sensors that fails to span retains the previous span value and does not continue with the calibration process Connecting the Gas Cylinder to the Detector Refer to the following Figure 8 Table 11 and procedures to connect the gas cylinder to the detector for calibration
53. tector does not respond to pushbutton Battery is depleted Charge battery Refer to Charging the Battery Detector is performing operations that do not Pushbutton operation restores automatically when require user input the operation ends Detector does not accurately measure Sensor s requires calibration Calibrate the sensors Refer to Calibration gas Detector is colder hotter than ambient gas Allow the detector to attain ambient temperature before use Sensor filter is blocked Replace the sensor filter Refer to Replacing a Sensor or Sensor Filter Detector does not enter into alarm Alarm setpoint s defined incorrectly Reset alarm setpoints Refer to Factory Gas Alarm Setpoints and Sensor Configuration Alarm setpoint s set to zero Reset alarm setpoints Refer to Factory Gas Alarm Setpoints and Sensor Configuration Detector is in calibration mode Complete the calibration procedure 53 GasAlertMicroClip Technical Reference Guide Table 15 Troubleshooting Detector intermittently enters alarm Ambient gas levels are near alarm setpoint or Detector is operating normally Use caution in without reason the sensor is exposed to a puff of the target suspect areas Check the peak maximum gas gas exposure reading Alarm setpoints defined incorrectly Reset alarm setpoints Refer to Factory Gas Alarm Setpoints and Sensor Configuration Detector requires calibration Calibrate the sensors Refer to Calibration Missing or fau
54. the following under Sensor Configuration e Low Alarm e High Alarm e TWA Alarm e STEL Alarm Stopping a Gas Alarm The low and high alarms stop when the ambient gas concentration returns to the acceptable range Note If alarms are set to latch press to reset the alarms The detector calculates the TWA value based on OSHA standards and the STEL value based on a user defined 5 to 15 minute period Refer to STEL Interval To stop a TWA or STEL alarm perform one of the following 1 Deactivate and reactivate the detector 2 Reset the TWA STEL peak exposure readings Refer to Viewing Gas Exposures A Warning Follow all safety procedures as defined by your employer Confirm with your supervisor before clearing TWA and STEL alarms 33 GasAlertMicroClip Technical Reference Guide Sensor Alarm The detector tests for missing or defective sensors during the startup self test and continuously thereafter e During start up if a sensor fails Error and the sensor name displays e If a sensor fails the self test Err displays above the gas type of the failed sensor The detector will enter alarm Press to acknowledge the sensor alarm and stoop the alarm Refer to Troubleshooting Low Battery Alarm e Battery power is continually displayed during normal operation If battery power is low and flashes The detector performs a sequence of 10 rapid sirens and alternating flashes with 7 seconds of silence in b
55. tions 24 Cal Lock Calibration IR Lock When enabled the sensors can only be calibrated using an infrared IR device to ensure calibrations are recorded The following are IR devices e IR Link with Fleet Manager refer to Calibration Procedure e MicroDock II base station refer to the MicroDock II User Manual If Cal Lock is enabled and calibration is attempted the following screen displays Note If Cal Lock is enabled the detector will still auto zero the sensors The detector is shipped with Cal Lock disabled Confidence Beep When enabled the confidence beep provides continuous audible confirmation that the detector is operating correctly by beeping once every second Note Confidence beep automatically disables during a low battery alarm a self test fail a calibration fail a bump test fail or during an alarm event To define how often Confidence Beep occurs 1 60 seconds refer to Confidence Beep and IntelliFlash Interval Default setting is 1 second The detector is shipped with the Confidence Beep disabled Confidence Beep and IntelliFlash Interval Enter a value 1 60 seconds to define how often IntelliFlash occurs and the detector beeps Intelliflash and or Confidence Beep must be enabled in order to define Confidence Beep and IntelliFlash Interval Refer to Confidence Beep and IntelliFlash Interval Default interval is 1 second Note IntelliFlash and IntelliFlash Interval are only applicable to
56. tration beyond upper scale limit which may be hazardous GasAlertMicroClip Safety Information Read First e Only calibrate the detector in a fresh air environment and in a safe area e Use only BW approved batteries for the GasAlertMicroClip detector Refer to Specifications e Charge the detector before first time use BW recommends the detector be charged after every workday e Charge the detector using the recommended charging adapter only Do not use any other charging adapter Failure to adhere to this caution can lead to fire and or explosion e Extended exposure of the detector to certain concentrations of combustible gases and air may stress a detector element that can seriously affect its performance If an alarm occurs due to high concentration of combustible gases calibrate the detector If necessary replace the sensor e Do not test the combustible sensor s response with a butane cigarette lighter doing so will damage the sensor e Do not expose the detector to electrical shock and or severe continuous mechanical shock e Do not attempt to disassemble adjust or service the detector unless instructions for that procedure are provided in the technical reference guide and or that part is listed as a replacement part Use only BW Technologies by Honeywell replacement parts Refer to Replacement Parts and Accessories e The detector warranty will be voided if customers personnel or third parties damage the det
57. udible visual vibrator alarms and displays the current ambient readings If more than one type or level of alarm occurs simultaneously a multi gas alarm results If Stealth is enabled the audible and visual alarms are disabled and only the vibrator alarm activates To change the factory defined alarm setpoints refer to Low Alarm High Alarm TWA Alarm and STEL Alarm Table 7 Alarms Low Alarm A TWA Alarm A LowWA iN odi Siowrsiran PANEN Y rwa e Slow siren Slow alternating flash L Li e Slow alternating flash O 3H S pp and gas bar flash E and gas bar flash MEM co pan C LEL e Vibrator alarm activates PERES PE e Vibrator alarm activates High Alarm a STEL Alarm e Fast siren e Fast siren ALARM e Fast alternating flash e Fast alternating flash me Ie 2S ppm and gas bar flash i and gas bar flash pe T LEL e Vibrator alarm activates EA eae e Vibrator alarm activates iar Note If Low Alarm Acknowledge is enabled the audible alarm can be disabled during a low alarm condition The vibrator and visual alarm indicators remain active until the alarm condition changes or the detector deactivates Press to acknowledge the low alarm and deactivate the audible alarm If the alarm escalates to a high TWA or STEL alarm the audible alarm reactivates If enabled Latched Alarms causes the low and high gas alarms audible visual and vibrator to persist until the alarm is acknowledged
58. ump interval can be defined for each sensor 1 Enter the value 1 365 days for each sensor 2 Enter 0 to disable the Bump Interval option Entering 0 automatically disables the Force Bump When Overdue option The detector is shipped with the factory default set to 0 days Note BW recommends to bump test the sensors before each day s use to confirm their ability to respond to gas by exposing the detector to a gas concentration that exceeds the alarm setpoints Verify that the audible and visual alarms activate Calibrate if the readings are not within the specified limits 21 GasAlertMicroClip Technical Reference Guide Low Alarm Enter the low alarm setpoints for each sensor Refer to Factory Gas Alarm Setpoints for factory defined alarm setpoints Applicable to all sensors TN ALARM A q 3 Y shy m o S i j High Alarm Enter the high alarm setpoints for each sensor Refer to Factory Gas Alarm Setpoints for factory defined alarm setpoints Applicable to all sensors me ALARM HE Ea DE O o m wer l p Jl F x i m i o S TWA Alarm The time weighted average TWA is a safety measure used to determine accumulated average exposure to gases An average is determined using the US Occupational Safety and Health Administration OSHA method to ensure the worker is warned when the maximum average is accumulated 22 The US OSHA method is defined as a movin
59. up Enter information such as employee name plant area emergency numbers etc Depending upon the length of the message it will either a display on the LCD for 3 seconds shorter message or b scroll twice on the LCD longer message 19 GasAlertMicroClip Technical Reference Guide Sensor Configuration Settings for the sensors are configured individually Enter values or enable disable options Refer to Factory Gas Alarm Setpoints for setpoint values Note Depending upon the sensor the options may vary Carbon Monoxide CO Disabled Cal Gas ppm Cal Interval RENE Bump Interval lens Low Alarm ppm High Alarm ppm TWA Alarm ppm STEL Alarm ppr STEL Interwal minutes CO uto ero on Startup Figure 5 CO Sensor Configuration 20 Sensor Disabled A Warning Use extreme caution when disabling a sensor The disabled sensor cannot detect and alarm against the applicable gas 1 Click Retrieve from Device to populate the fields with the current detector settings 2 Click the Disabled checkbox for the required sensor Hydrogen Sulphide H25 v Disabled Cal Gas io ppm Cal Interval iso day s Bump Interval MIA days Low Alarm 10 0 ppr High Alarm 15 0 ppm TWA Alarm 10 0 ppm STEL Alarm iso ppm STEL Interval Is minutes Figure 6 Disabled Sensor The fields for the applicable sensor become inactive greyed out until the sensor is again enabled 3
60. ys saved Datalog Interval The default interval is one reading every 15 seconds The user has the option to change the intervals to rates from 1 to 120 seconds Data recorded All sensor readings all alarm conditions calibrations event flags battery status sensor status confidence beep activation and detector status with the time and date for each reading and unit serial number Operation Requires no user intervention automatic Compatible with Desktop PC computer or laptop Operating system Windows 2000 or higher Download via IR device IR Link adapter or MicroDock II base station Software required e Fleet Manager II application or e Microsoft Excel optional to create custom reports 61 GasAlertMicroClip Technical Reference Guide 62 A Wear yellow Work safe 50120681 002 EN B2 English BW Technologies 2014 All rights reserved
Download Pdf Manuals
Related Search
Related Contents
A780 Table of Contents Manuale - Pico Technology C5500 User Manual - Surveillance Cameras Viewsonic Graphic Series CDP6530T SONY Trinitron ®Color TV FilterStream AIRTAMER A300 User's Manual conexion del amplificador Line Tracking C-9881 kH + Ca 1, 2, 3 Copyright © All rights reserved.
Failed to retrieve file