1.   7  The sample is initially a liquid  After several minutes the sample begins to evolve into a  clot  The analyzer detects this initial clot formation and calculates the time the sample  remained a liquid  ACT   During the next several minutes of analysis  the fibrinogen  converts into a fibrin gel  The analyzer calculates the rate of change in the clot signal   Clot Rate      8  When a result has been calculated  the analyzer  will beep and display the time that the sample  remained a liquid  ACT  and the Clot Rate  CR    ACT  177 CR  19  Times for results vary from 2 to 20 minutes  depending on heparin levels and clinical conditions  The numeric result is displayed  next to the corresponding text for the appropriate channel     NR    is displayed if no  results were found for the test sample     gbACT       9  The analyzer will automatically stop data collection when results for ACT and Clot Rate  have been calculated or after 30 minutes        10  Open the head and remove the probe from the  hub with a probe extractor  Avoid moving the hub  sideways  Propery discard the probe and the cuvette   Lower the head assebly to maintain temperature  control of the instrument                    2 4 Chapter 2  Running a Whole Blood Sample    Chapter 3   Maintenance and Troubleshooting       Maintenance    Cleaning    Clean after use to reduce biohazard risks  The analyzer can be sprayed or wiped with a  disinfectant approved by your institution  Close the heads when spraying to avoi
2.   This requires QC kit 900 1303     Monthly  and prior to the use of a new shipment  validate the activated cuvettes  This  requires QC kit 900 1318     Complete instructions for running each QC test are included in the kits  The kits also include  forms for recording QC results     To order quality control kits  contact Sienco at 800 432 1624 or contact your distributor        1 6    Chapter 1  Installation and Setup    Chapter 2   Running a Whole Blood Sample       The analyzer monitors mechanical changes that occur during hemostasis  This section describes  how to prepare and run a test with native whole blood samples  Use only test kits available  from Sienco  Inc  Sienco offers different tests for different applications  Choose the test best  suited for your application     Analyzer Preparation    1  Cuvettes should be placed in the convenience wells  so they will be ready when the blood is drawn  Place  probes into the lids of the cuvettes     EA  iG  a Es    ae Pee age  PBI AS  i a     A er LG  HE      Pa  II    a ne    ad    probe         cuvette       2  Ifthe analyzer has just been recently turned on  allow it to warm  with head assembly in  the down position  until the analyzer reaches 37  C  If the analyzer is not at the desired  temperature it will display an error message and not run the test     3  When the analyzer is ready the display reads     i nn gbACT  NOTE  Results from the previous test may be  displayed on the bottom of the LCD display   Press the    S
3.   in the European community    CE marking of conformity    Biological risk    Temperature limitation    Humidity limitation    Direct current  DC     Serial number    Do not dispose of electrical  equipment as municipal  waste    HES       O     j     gt  Y   E    CONTROL    Do not reuse    Model Number    Lot Number    Manufacture date    Expiration date    Consult instructions for use    Contains sufficient for  n   tests    Caution  consult  accompanying documents    Control    Decontamination Form    This decontamination form must be filled out and returned with the Sonoclot Coagulation  Analyzer when shipping the analyzer to Sienco for any reason  or a  150 00 biohazard  decontamination charge will be assessed     Please photocopy this page  complete  and include with your analyzer     We require you to thoroughly clean and decontaminate the analyzer per guidelines followed  by your institution  Here are some tips to follow when decontaminating the analyzer     e Place a clean probe on the probe adapter to protect the sensing mechanism while  cleaning  Do not spray any cleaning solvent into the head since fluid will damage  the sensor    e Decontaminate all surfaces of the analyzer using a product certified by your  Institution    e Remove and discard all probes and cuvettes before packing the analyzer     Institution      Dpmt   Serial Number  _________ Model   Decontaminated by  ___ Date     Decontaminate Used     Probe removed NE Cuvette removed ME Power cord  amp  ad
4.  cuvette     Chapter 1  Installation and Setup 1 5    Warnings  Cautions  and Hazards    1     WARNING  As with any laboratory test result  diagnosis should not be based solely  on the analyzer test results  The attending physician is responsible for interpreting the  analyzer test results in conjunction with the patient   s condition  other test results  and  clinical observations     2  WARNING  To safeguard against electrical shock  only use a grounded electrical  outlet   3  WARNING  Use only as specified by Sienco  Inc  Otherwise  equipment protection  may be impaired   4  CAUTION  Only properly trained laboratory personnel and or other appropriate health  care professionals should use the analyzer   5  CAUTION  A biomedical engineering department should check the analyzer at least  once a year for compliance with leakage standards   6  HAZARD  Human blood is a biohazardous material  The operator should wear  appropriate protective gear when handling blood and or test cuvettes containing  BON blood samples  Biocontaminated materials should be handled and disposed of  YY E      properly in accordance with hospital  local  state  and federal regulations   7  CAUTION  Do not place the analyzer on counter tops where other vibrating instruments   such as centrifuges  are located  Mechanical vibration may cause erratic results   Quality Control    Proper performance requires regular quality control  QC  tests     1   2     Daily  run a reference viscosity test for the analyzer
5. 420 1148        3 2    Chapter 3  Maintenance and Troubleshooting    Factory Service or Repair    Contact Sienco prior to shipping your analyzer for repairs     The analyzer should be shipped in its original packing materials  if possible  If you do not have  the original packing materials please contact Sienco  We will ship you packing materials or  you may use the packaging materials from your loaner     We charge a minimum fee for testing the instruments  even if you decide not to have it repaired     Damages caused by poor packaging are your responsibility     To ship  follow these steps        Remove cuvettes  probes  power cord  and other supplies   Photocopy the decontamination form at the end of this manual     Clean and decontaminate the analyzer according to your institutional guidelines and  fill out the decontamination form  Failure to fill out form and include with the analyzer  will result in a  150 decontamination fee     4  Carefully slide the spacer tabs between the instrument case  and platen  Tape the back of the platen and the top of the  instrument case as shown  Tape the instrument head down as  shown  Place the analyzer in a large  clean plastic bag and  fold the open end under the instrument     Place the instrument in the molded foam in  the bottom of the box  with the ribbon running  underneath the analyzer as shown     If you are also returning your power adaptor f a  and cord  fold the power cords  secure each   Tamem a nn m  with a rubber band  an
6. Sonoclot   Coagulation Analyzer    Operator   s Manual  SCI Revision 2 1       SONOCLOT       C oagulation Analyzer    For in Vitro Diagnostic Use       Manufactured for  Sienco  Inc   7985 Vance Drive  Suite 104 Arvada  CO 80003 USA  303 420 1148 1 800 432 1624 Fax 303 379 4403  www sienco com e mail  sienco sienco com    Copyright    1999 2009 Sienco    Inc  All Rights Reserved     Manufactured For  Sienco    Inc    al  gt   Vance Drive  Suite 104 Arvada  CO 80003 USA  303  420 1148 303  379 4403  FAX  e sienco sienco com  e mail   http   www sienco com    Sonoclot is a registered trademark of Sienco    Inc     Sonoclot Analyzers are protected under U S  and foreign patents     Ec  Rep  QNET BV    Hommerterweg 286  C     6436 AM Amstenrade  The Netherlands    Table of Contents    Chapter 1  Installation and Setup  Principle of Operation and Intended Use  Head Assembly  Model SC1   Front View  Model SC1   Back View  Package Contents  Installation  Good Operating Procedures  applicable for all tests   Warnings  Cautions  and Hazards  Quality Control       NO N      gt   gt  e e e e ah    I ll l    OO UVUWWND             Chapter 2  Running a Whole Blood Sample  Analyzer Preparation    Cuvette and Probe Setup 2 2  Obtaining the Blood Sample 2 3  Running the Analyzer 2 3  Chapter 3  Maintenance and Troubleshooting 3 1  Maintenance 3 1  Troubleshooting 3 2  Warning and Error Messages 3 2  Factory Service or Repair 3 3  Appendix A 1  References A 1  Warranty A 2  Technical Specifi
7. aptor returned  Y NE N AE    Contact Person  __        o    o o o o o Oo Phone     Reason for Service     We will call with an estimated cost for repairs before any work is done  Please make sure to  include a contact name and phone number     Ship the Sonoclot Coagulation Analyzer to   Sienco  Inc   Attn   Service Manager  7985 Vance Drive  Suite 104  Arvada  CO 80003 USA    If you have any questions  please contact  Service Manager  800 432 1624  303 420 1148       Appendix A 5    
8. cations A 2  Environmental Conditions for Transportation and Storage A 2  Environmental Conditions for Use A 3  Electrical Classification A 3  Disposal Instructions A 3  Glossary of International Symbols A 4  Decontamination Form A 5    Chapter 1     Installation and Setup       Principle of Operation and Intended Use    The Sonoclot Coagulation Analyzer calculates onset of clot formation and clot rate for in vitro  coagulation studies by monitoring mechanical changes that occur in blood samples during  hemostasis  The mechanism is a tubular probe that moves up and down within a blood sample   As the sample progresses through various stages of clotting  the electronic circuitry detects  increasing resistance  This produces a series of electronic signals that are processed by a  microcomputer  Results are reported on the LCD display panel     Head Assembly          analyzer head    probe mount hub    convenience wells    cuvette    holder    platen    head alignment pin    1 1    Model SC1   Front View     J     analyzer head  O    Par tine era       vt  convenience b   well platen  LCD  display          Start Stop  button    LOT    Coagulation Anal y zer       For in Vitro Diagnostic Use    Select Test  button    Model SC1   Back View    serial number AS  labe   a re ili The Neiherands j   a  een electrical information  Arvada  Colorado 20003 MADE IN USA  s power cord    pol receptacle          1 2 Chapter 1  Installation and Setup    Package Contents    e Sonoclot Coagulatio
9. d contaminating  the transducers  Avoid excess wetting  Place a clean probe on the probe adaptor to protect  the sensing mechanism while cleaning  Gently wipe around the cuvette holders or inside  the heads  Do not clean the probe mount hubs  unless absolutely necessary  The transducers  attached to the probe mount hubs can easily be damaged by debris or liquids     NOTE  Do not use isopropyl alcohol or other solvents on the front panel or LCD cover     Calibration  amp  Service    The user cannot calibrate or service the analyzer  Running a reference viscosity QC test   PN 900 1302  will verify if the analyzer is operating within calibration standards  Sienco  recommends performing the reference viscosity QC test daily  prior to use of the analyzer     If the analyzer requires calibration or service  contact Sienco  Inc  or your distributor     3 1    Troubleshooting  If you have a problem with the analyzer  follow these steps     1  Review    Warnings  Cautions  and Hazards    on page 1 6    2  Review    Good Operating Procedures    on page 1 5    3  Review the operating instructions that come with the test you are running   4  If problems persist  contact the Sienco service department     Phone  303 420 1148   Toll Free  800 432 1624   Fax  303 379 4403  E mail  sienco sienco com    Mail  Sienco Service Department  7985 Vance Drive  Suite 104  Arvada  CO 80003 USA    Warning and Error Messages    The Sonoclot Analyzer will occasionally display messages that may not be fami
10. d place in the spaces  on either side of the analyzer     Sonoctor       Place the top of the molded foam over the salvar MENE sure sa     is securely  packed around the analyzer head     Make sure the decontamination form is complete  Be sure to include a brief description  of the problem along with a contact name  phone number  and purchase order     Place the completed form on top of the molded foam     Tape the box shut and ship the package to     Sienco  Inc    Service Department   7985 Vance Drive  Suite 104  Arvada  CO 80003 USA    Chapter 3  Maintenance and Troubleshooting 3 3    Appendix       References    Sonoclot Analysis is used in many clinical and research applications and consequently is  referenced in numerous studies and articles  For acomplete list of references  please visit our  web site at www sienco com     1  Nilsson CU and Engstr  m  Monitoring fondaprinux with the Sonoclot  Blood Coagulation  and Fibrinolysis  2007  18  619 622     2  Ganter MT  Monn A  Tavakoli R  et al  Monitoring activated clotting time for combined  heparin and aprotinin application  in vivo evaluation of a new aprotinin insensitive test  using Sonoclot  Eur J Cardiothorac Surg  2006  30 2   278 84     3  Shibata T  Sasaki Y  Hattori K  et al  Sonoclot analysis in cardiac surgery in dialysis  dependent patients  Ann Thorac Surg  2004  77 1   220 05     4  Liszka Hackzell JJ and Ekback G  Analysis of the information content in Sonoclot data  and reconstruction of coagulation test 
11. eep blood  dirt  and other substances away from the probe mount hub to avoid  contaminating the electromechanical transducer    e Never reuse a probe or cuvette to prevent thrombin contamination     For consistent results  the cuvettes should be warmed prior to analysis  To warm  place  the cuvette in its holder at least 30 seconds prior to running a test     Always insert and remove the probe by moving it vertically over the probe mount  hub  Never move the hub horizontally  Make sure the probe is fully seated to avoid  interference with the stir bar     Always remove the cap from the cuvette prior to placing it into the cuvette holder   Failure to do so can damage the transducer     The cuvette must be fully seated in the cuvette holder to avoid interference between  the probe and stir bar     Do not overfill the cuvette  The proper fill level is slightly below the inner rim of the  cuvette     Native whole blood must be analyzed within 2 minutes or less of collection     Perform QC testing to verify proper operation of the analyzer and activation cuvettes   See page 1 6   Use proper biohazard handling techniques to dispose of probes and cuvettes     On rare occasions  mechanical disturbances may cause incorrect results  Always  inspect results to ensure that they are consistent     High viscosity blood samples  immersion response  gt  25 clot signal units as displayed  on the instrument  can stratify  Use an external device to mix the blood sample before  placing it in the
12. elect Test    button on the front panel until the desired test is displayed on  the LCD          2 1    Cuvette and Probe Setup    1  Open the head by tilting it backwards  The display  will read    Open           2  Sharply tap the cuvette on a hard surface to dislodge activation powder from the sides  and lid     3  With a slight twisting motion  use the cuvette to seat  the probe on the mount hub  This motion should  result in the probe sliding straight over the probe  mount hub  The hub should not move sideways  The  probe must be fully seated for proper operation     4  Remove the cuvette  leaving the probe on the hub        5  Remove the lid from the cuvette by placing it in a  convenience well and popping the lid off with your  thumb        2 2 Chapter 2  Running a Whole Blood Sample    6  Insert the cuvette into the cuvette holder  Make sure  it is fully seated  Close the head  Allow the cuvette to  warm in the cuvette holder for at least 30 seconds     DO NOT close the head on a lidded cuvette  It will  damage the instrument     Different tests have different set up requirements   Please refer to the product insert for detailed    ve  lt a ae Lbs       AR   Instructions             Obtaining the Blood Sample    Native whole blood must be analyzed within 2 minutes of collection  When drawing blood   observe the following precautions     1  Carefully decide where to draw the sample  Contamination  from a heparinized line  a  heparin impregnated catheter  or a surgery p
13. liar  Below are  some examples of these messages and procedures to follow  if necessary        NOISE    indicates that the clot signal is being disturbed by either mechanical noise  such as  bumping the analyzer  or by interference between the probe and the cuvette  Check to ensure  that the probe is tight against the mount hub and the cuvette is in contact with the bottom of  the cuvette holder  The message should disappear within a few seconds        ERROR    indicates that the self test routine has detected a problem  Normally  the message  will disappear after several seconds  You can resume testing then  If the error message remains   make a note of the error number and call Sienco   s service department        TEST NOT RUN  NOT AT TEMP    indicates that the analyzer is not to temperature and the  test was not run  The analyzer must be at 37  C to run a test        NR    indicates that no results were found for the last test analysis  This can occur due to  operator error  i e  failure to close instrument head within 60 seconds of    Close Head    message    instrument error  or an unusual clot that cannot be analyzed     NUMBERED ERROR CODES  The analyzer is programed to display various numbered error  codes  Should your analyzer display a numbered error code  record the number displayed  and contact your distributor or Sienco   s service department     If you have any questions in regards to a display message  call Sienco   s service department at  800 432 1624 or 303 
14. n Analyzer  Model SC1   e nower adaptor and cord   e probe extractor   e operator   s manual   e USB cord   e SonoView installation CD    Installation    1  Unpack the contents of the shipping container and check you have all the above  Items     NOTE  Sienco recommends that you retain the original shipping box in case you need  to return your instrument for service        2  Pull the ribbon to remove the packaging from the  top of the instrument  Then pull the ribbon to lift the  instrument out of the packaging foam in the bottom  of the box     E   3  Remove the analyzer from the plastic bag     4  Carefully remove the tape from the top and back of  the instrument     5  Remove the spacer tabs from between the analyzer  case and the head assembly platen  Keep these tabs  with the original packaging     6  Plug the power cord into the analyzer and an easily accessible grounded wall outlet   Do not use an ungrounded extension cord or adaptor  Do not obstruct access to the  outlet     7  The analyzer will automatically turn on  First  the LCD display will illuminate  Next   the analyzer will display the version number  product name  and copyright     8  The analyzer will begin warming  While  warming  the display reads     gbACT    O  Heating 36 3       9  When the analyzer has reached 37 0  C  the display indicates the default test  The  analyzer is now ready to run a test     10  Review the remainder of this chapter     Chapter 1  Installation and Setup 1 3    Installing  am
15. p  Using SonoView Data Management Software  SonoView     The analyzer indicates real time results on the LCD display  It also collects and stores data much  like a digital camera  This data can be accessed by connecting the analyzer to a computer  running SonoView     To install SonoView on your computer     1  Read the installation instructions and hardware specifications on the CD sleeve   Instructions are also located in PDF format on the CD     2  Plug a USB cord into the back of the analyzer and a USB port on your computer   NOTE  The analyzer requires a 1 1 USB Port or higher     3  Open SonoView on your computer  For software instructions  please see the SonoView  operator   s manual PDF file on the installation disk        1 4 Chapter 1  Installation and Setup    Good Operating Procedures  applicable for all tests     The quality of test results depends on proper technique  Carefully follow these precautions     1     10  11    XA oe    12  13     Ner    14    Ner    Operator should read this operator   s manual in its entirely before attempting to operate  the analyzer     Only properly trained lab personnel and healthcare professionals should operate the  analyzer     Diagnosis should be based on test results in combination with clinical observations     Handle materials with care    e When taking a blood sample  avoid heparin contamination from catheters    e Never use the first sample from a new line to avoid sample contamination with  tissue thromboplastin    e K
16. ranty is limited to the replacement or  repair of defective parts and components  The warranty shall be voided if the product is     1  Used in any manner or subjected to any condition that is inconsistent with its intended  purpose or accepted industry practice  or    2  Modified in any way without prior written approval by Sienco  or    3  Repaired in any manner so as to adversely affect its operation or reliability by persons  other than Sienco or its agents     Warranty claims shall be submitted in writing to Sienco during the 2 year warranty period   The claim shall state the nature and details of the analyzer defect and the serial number  of the analyzer  The claim must accompany any defective product submitted for repair or  replacement     Purchasers must decontaminate and package the analyzer per instructions on pages 3 1 and  3 3  Such products shall  at Sienco   s election  be either repaired  replaced or returned     Technical Specifications    Width 4 25    10 8 cm  Depth 5 75 14 6 cm  Height 4 75    12 1 cm  Weight 2 5 lbs 1 13 kg  Electrical voltage requirement 100 to 240V  Electrical power requirement 9V   2A   Frequency 50 to 60 Hz    Temperature regulation of platen 37  C   0 5  C    Viscosity range for test sample  lt 80 cP    Environmental Conditions for Transportation and Storage    Transport and store within the ambient temperature range of  25  C    and 70  C     95     Transport and store to a maximum relative ambient humidity of 95         A 2 Appendi
17. rep line  will cause inaccurate results     2  Withdraw blood in a smooth  slow  and non traumatic manner  Do not use force     3  When drawing from an anesthesia or pump port  use a 2 syringe technique  Discard the  first syringe of 2 to 3 ml  Use the second syringe for the sample  You MUST use plastic  syringes to avoid uncontrolled glass activation     Running the Analyzer       1  Transfer the whole blood sample from the syringe into  the warmed cuvette  Use a blunt cannula tip for a  clean and controlled fill     Fill the warmed cuvette with the blood sample until the  fluid level is slightly below the inner rim of the cuvette   This volume is approximately 330 to 360 ul        Fill slightly below here            Dg N    Bes          4    E  x             2  Leaving the head open  immediately press the    START STOP button  QbACT  3  The magnetic stirrer will rotate and the display Mixing    will read     4  When mixing is completed  the analyzer QbACT  will beep and the display will read     Close  Close Head  Head        Chapter 2  Running a Whole Blood Sample 2 3          5  Close the analyzer head  Make sure it is fully seated  over the head alignment pin                 6  During data collection  the LCD display will  show the test being run  the clot signal reading   9 bACT 13 Ll   CS   the time in seconds  and    222    for each JACT      CR      result  When the analyzer is calculating a  result  the display will flash the number the  result is greater than      
18. required  to correct the interference     Disposal Instructions    Sienco  Inc  expects end users to dispose of the Sonoclot Coagulation Analyzer in  an environmentally friendly way  Electrical and electronic equipment is labeled  with the following    crossed out wheeled bin    symbol indicating that the equipment  should be disposed of  by the end user  separate from other types of waste     End users should contact their dealer distributor or Sienco  Inc  for disposal  collection and  recycling options and terms and conditions in your country     In 2002 the European Union introduced the Directive on Waste Electrical and Electronic  Equipment  WEEE   The main aim of the Directive is to ensure that WEEE is collected and    Appendix A 3       treated separately  WEEE may contain hazardous substances that should not end up in the   human  environment and can have adverse effects on it if they do     WEEE is a vast source of raw materials  With the ever rising worldwide demand for new  equipment and the ever decreasing volume of natural raw materials  letting this potential  source go to waste is unacceptable  If equipment is collected separately  the equipment can  be recycled and up to 85 to 90  of the equipment can be re used as new material  saving the    use of virgin raw materials and energy of producing these     Glossary of International Symbols       se     lt e RARE       A 4 Appendix    Manufacturer    In Vitro Diagnostic medical  device    Authorized representative
19. variables  Journal of Medical Systems  2002   26 1   1 8    5  Pivalizza EG  Pivalizza PJ  Kee S  Gottschalk LI  Szmuk P  Abramson DC  Sonoclot analysis  in healthy children  Anesth Analg  2001  92 4   904 6    6  Santucci RA  Erlich J  Labriola J  et al  Measurement of tissue factor activity in whole  blood  Thromb Haemost  2000  83 3   445 54    7  Ekback G  Carlsson O  Schott U  Sonoclot coagulation analysis  a study of test variability   J Cardiothorac Vasc Anesth  1999  13 4   393 7    8  Hett DA  Walker D  Pilkington SN  Smith DC  Sonoclot analysis  Br J Anaesth  1995   75 6   771 6    9  Steer PL  Krantz HB  Thromboelastography and Sonoclot analysis in the healthy patient   J  Clin Anesth  1993  5 5   419 24    10  Shenaq SA  Saleem A  Viscoelastic measurement of clot formation  the Sonoclot  In   Ellison N  Jobes DR  eds  Effective Hemostasis in Cardiac Surgery  Philadelphia  PA   Harcourt Brace Jovanovich  Inc   1988  183 93    11  Newland MC  Chapin JW  Hurlbert BJ  Kennedy EM  Newland JR  Thrombelastograph and  Sonoclot Signature monitoring of changes in blood coagulation following cardiopulmonary  bypass  Anesthesiology  1987  67 3A   A199     12  Newlin  F   Ens  G   Leppke  L  and Hamstra  R      Heparin Control with the SONOCLOT      American Journal of Medical Technology  44  508  1978        A 1    Warranty    The Sonoclot Coagulation Analyzer has a 2 year warranty that it conforms to specifications and  is free of defects in material and workmanship  The war
20. x    Environmental Conditions for Use    Indoor use only     Use at a maximum altitude of 2000m     K Use within the ambient temperature range of 18  C and 27  C     Use at a maximum relative humidity of 80  for temperatures up to 31  C  7 decreasing linearly to 50  relative humidity at 40  C     Mains supply voltage fluctuation shall not exceed   10      The unit is intended for INSTALLATION CATEGORY II     Installation Category Il  Local level  appliances  Portable equipment etc    with smaller transient overvoltages than Installation Category II     The unit is intended for use in a POLLUTION DEGREE 2 ENVIRONMENT     Pollution Degree 2 is nonconductive pollution of the sort where occasionally  a temporary conductivity caused by condensation must be expected  This  is the usual pollution degree used for equipment being evaluated to 60950  and is suitable for equipment employed in an office environment     Electrical Classification    This equipment has been tested and found to comply with the limits for a Class A digital device   pursuant to part 15 of the US FCC Rules  These limits are designed to provide reasonable  protection against harmful interference when the equipment is operated in its installation   This equipment generates  uses and can radiate radio frequency energy and  if not installed  and used in accordance with the instruction manual  may cause harmful interference to radio  communications  If this equipment does cause harmful interference the user will be 
    
Download Pdf Manuals
 
 
    
Related Search
    
Related Contents
Recurring Contract Billing Importer  平成15年那審第12号 プレジャーボートトッチャンボーヤ2号同乗者死亡  - Magnetrol  Bedienungsanleitung-Scansonic DA88  Element 302 User Manual  Manual de usuario en español Modelo ZMRK700    HP SureStore E Disk Array 12H System Administrator`s  HomComfort HCPHSSQ Instructions / Assembly  Endpoint Protector - Mobile Device Management - User Manual    Copyright © All rights reserved. 