Home
        System User Manual.book
         Contents
1.     Access Doors    PN 89 00002 00 109 Rev  A    The System             1  Cheminert Fitting 2  Probe Holder  3  Probe Adjustment Screw 4  Luminex XYP Instrument Sample Probe    Figure 2 2 Fluidics Components    The Luminex 200 analyzer has three access doors  Two of the access  doors are on the front  and the third is on the back  The front left  access door supplies access to the sheath filter  The front center  access door supplies access to the syringe  The rear access door  supplies access to the air intake filter  See Figure 2 3     z    Lumi  nex       1 2    1  Left door  access to service panel 2  Center door  access to syringe    Figure 2 3 Luminex 200 Analyzer Access Doors    Luminex 200 System User Manual    Air Intake Filter    Syringe    Sheath Filter    Air  Waste  and  Sheath Fluid  Connectors    Luminex Sheath  Delivery System    Waste Fluid  Container  Optical  xMAP Technology  Reagents  2 10    xMAP Technology    A replaceable air intake filter cleans the air used to pressurize sheath  fluid  This filter is enclosed behind an access door located on the  back of the Luminex 200 analyzer     The syringe delivers a sample from the 96 well microtiter plate to the  cuvette     The sheath filter removes particles greater than ten microns in  diameter from the sheath fluid     The air  waste  and sheath connectors  located on the left side of the   analyzer  connect to the SD system and waste fluid containers using   clear tubing  The air connector is green  the s
2.     PN 89 00002 00 109 Rev  A    xMAP Technology    Pressurization    Symptom    Pressurization fails or  pressure is too low     Pressure too high     PN 89 00002 00 109 Rev  A    Troubleshooting    Normal air and sheath pressure readings vary between 6 9 psi while  the compressor runs  If the system pressure is out of range  your  sample acquisition will fail or return poor results     Possible problem  The sheath and waste lines are  not fully connected   The sheath or waste bottle  fittings are cracked     There is a leak in the system     The compressor does not  engage     The Cheminert fitting is loose     Fluid leaks in the system     The sheath bottle has an air  leak     The sheath bottle is overfilled     The Sheath Delivery System is  overfilled     Regulator not adjusted properly     Solution    Make sure the lines between the  sheath and waste bottles and the  analyzer are fully connected     Inspect the fittings to be sure  they form a tight seal     Check for system leaks  A leak  will be obvious if there is fluid  on the surface on which the  system sits     Run a Prime command  If you  do not hear the compressor turn  on  contact Technical Support     Ensure that the fitting connects  tightly above the sample probe   below the blue light     See Fluid Leak on page 4 4     Disconnect the sheath and waste  bottle connections from the  analyzer  Run a Prime  command  If pressure builds   remove and retighten the sheath  fluid bottle cap  then reconnect  fluid l
3.    Warning Warning  refer to    On  refer to manual  manual   Catalog Number Consult instructions  O Off for use    SPP       In vitro Diagnostic  Medical Device    Batch Code Serial Number    o  4       y  ud  ce          Temperature  Limitation    Manufacturer    European Union  Conformity          Expiration Date Date of Manufacture    SBOE    Underwriter   s Canadian  Laboratory Underwriter   s   Laboratory  WEEE   Do not    throw in trash                   id    K       Safety Precautions    Fluidics    Read the following safety information before setting up or using the  Luminex 200 analyzer  A user should be present during operation   This system contains electrical  mechanical  and laser components  which  if handled improperly  are potentially harmful  In addition   biological hazards may be present during system operation   Therefore  we recommend that all system users become familiar with  the specific safety advisories below  in addition to adhering to  standard laboratory safety practices  The protection provided by the  equipment may be impaired or the warranty voided if the system is  used in a manner not specified by the instructions or by Luminex  Corporation     This system contains fluidics  In the event of a fluid leak  turn off all  power to the system and disconnect all power cords  The on off    PN 89 00002 00 109 Rev  A    xMAP Technology    Luminex 200 Analyzer  Laser    Barcode Reader Laser  Mechanical    PN 89 00002 00 109 Rev  A    Safety    switch
4.    place one alignment sphere into  the selected well     Select the Maintenance tab  then click Eject Retract to eject  the plate holder  Place the 96 well microtiter plate on the  Luminex XYP instrument plate holder with position A1 in the  top left corner     Verify that the correct location is selected in the Luminex IS 2 3  Software  and that you are using the appropriate number of  alignment discs  Click Eject Retract to retract the plate     PN 89 00002 00 109 Rev  A    xMAP Technology    After Running Samples    Routine Tasks    Sheath and Waste Fluids    PN 89 00002 00 109 Rev  A    Maintenance and Cleaning    5  Loosen the probe adjustment screw on the probe holder one third  to one half turn  Pull it upward until it touches the top of the  adjustment slide  Tighten the thumbscrew     6  Click Sample Probe Down to lower the sample probe     7  Gently push the probe down until it just touches the top of the  alignment discs or sphere     8  Tighten the probe adjustment screw   9  Click Sample Probe Up to raise the sample probe     10  Replace the plastic shield that covers the sample probe area         After running samples     1  Sanitize with a 10  to 20  household bleach solution   2  Run two Wash cycles with distilled water   3  Soak with distilled water  Wait until the soak completes     4  If desired  turn off the Luminex 200 analyzer     Replace the sheath fluid and empty the waste container as required   Be careful not to touch the seal in the waste bottle ca
5.   e Classification control microspheres  CONI    e Reporter control microspheres  CON2    e Luminex xMAP Sheath Fluid    PN 89 00002 00 109 Rev  A    xMAP Technology    Required  Laboratory  Reagents    Luminex IS 2 3  Software    Luminex 200  Performance  Specifications    Speed    Accuracy and Precision    PN 89 00002 00 109 Rev  A    The System    e Household bleach   e 70  isopropanol or 70  ethanol    Mild detergent   e Distilled water    Luminex IS 2 3 software provides complete control of the system  and performs data analysis  Your Luminex 200 system is preloaded  with the Luminex software  However  we supply a software CD  should you need to reinstall the software     This software requires a dedicated system  Unauthorized additional  software is prohibited and may result in improper operation of the  system     e Speed  greater than or equal to1 7 GHz Intel   Pentium   IV  processor with 256 MB RAM       USB communications link for fast data transfer    e Automatic transfer of assay templates and new reagent  information into the system via a 32  diskette or large capacity  read write CD    e Installation   lt  4 hours   e System calibration   lt  10 minutes   e System controls   lt  10 minutes   e Barcode reader entry of sample IDs   e Automatic post analysis   e Analyze one 96 well plate hour depending on manufacturer s kit  e Up to 100 xMAP microsphere sets per sample   e Sheath flowrate  90 uL sec   5 uL   e Sample injection rate into detector area  1 uL sec   0
6.   water     Make sure there is sheath fluid  in the sheath container  Prime  the system until all air is out of  the system     Add additional detergent to the  assay buffer  For example  add   02  to 0 1  Tween 20   Triton   X100  or SDS     View a list of incompatible  solvents at the Luminex  Technical Support website    http   luminexcorp custhelp com   If the solvent you are using is  listed  switch solvents     Use only Luminex sheath fluid  in the Luminex 200 analyzer   Other fluids may damage your  analyzer and may void your  warranty     Luminex 200 System User Manual  Error States    Error Message    Disconnected    Refill Sheath    Running  Sheath Empty    System Error  Messages    xMAP Technology    The following error states display on the Status Bar on the Run Batch  tab when an error occurs during system operation     Possible problem    The software has not made  initial connection     The communication cables are  not connected     Sheath fluid is low     Sheath fluid is empty     Solution    On the toolbar  click Connect   wait for the status to change     Remove and reconnect the cable  connections     Close the software application  and reopen it     Close the software application   then turn off the Luminex  system  Power on the system  and check the status     Refill the sheath container  then  click Resume     Replace or fill the sheath fluid  container with sheath fluid  Run  two Prime commands     There are two types of error messages  system error me
7.  05 uL    e System warmup  30 minutes  Systems that remain inactive for at  least four hours will require a warm up to restart the lasers  After  acquiring sample  running system calibrators  running system  controls  and warming up the instrument  the system resets the  four hour internal clock     e Sample uptake volume    5   e Classification of xMAP microspheres   gt  80     Luminex 200 System User Manual    Sensitivity    Capacity    Luminex 200    Analyzer General    xMAP Technology    Misclassification of xMAP microspheres      2   may vary by  xMAP microsphere product lines  Refer to the specific product  information sheet for further details     Temperature control  0 C to   2  C of target  Internal sample carry over   lt  0 9     Soluble background fluorescence emission at 575 nm  automatically subtracted from fluorescence intensity values    Detect 1000 fluorochromes phycoerythrin  PE  per xMAP  microsphere    Reporter channel dynamic range  3 5 decades of detection    The specifications below reflect minimum capacity values     10 GB hard drive   Store data for up to 240 000 test results   1 4 MB 3     diskette   100 MB read write CD ROM   Analyze multiple 96 well plates per batch   Analyze multiple assay templates per plate   Distinguish a minimum of 1 to a maximum of 100 unique xMAP  microsphere sets in a single sample   Detect and distinguish surface reporter fluorescence emissions at  575 nm on the surface of 1 100 unique xMAP microspheres sets  in a single sa
8.  CN M020 01    German    Luminex 100 IS User Manual  Version 2 3 CN MO030 01    Greek    Luminex 100 IS User Manual  Version 2 3 CN M022 01    Italian    Luminex 100 IS User Manual  Version 2 3 CN M021 01    Spanish    Luminex 100 IS Developer Guide to xMAP CN M029 01   Technology Version 2 3   Power Cord  USA CN P001  01   Power Cord  Australia CN P002 01   Power Cord  Brazil CN P003 01   Power Cord  Denmark CN P004 01   Power Cord  Germany  Sweden  France  CN P005 01  Belgium  and Spain   Power Cord  Switzerland CN P006 01   Power Cord  Israel CN P007 01   Power Cord  UK CN P008 01   Power Cord  Italy CN P009 01   Power Cord  Japan CN P010 01   Power Cord  China CN P012 01   Reservoir  XYP CN 0022 01   Sample Needle Height Alignment Kit CN 0015 01    Sample Probe Alignment Kit    Short Sample Needle CN 0006 01   Long Sample Needle CN 0007 01   Sample Holder  Large  1 5 mL CN 0008 01   Sample Holder  Small  1 2 mL CN 0009 01   Sheath Filter with Quick Disconnect CN 001 0 01   Sheath Bottle CN 0011 01   Syringe Cylinder with Seal CN 0013 01          PN 89 00002 00 109 Rev  A       xMAP Technology    Software    XMAP Reagents    Training    PN 89 00002 00 109 Rev  A                                                                            Product Numbers  Syringe Seal  Qty 4  CN 0014 01  Cable  USB CN 0018 01  Waste Bottle CN 0012 01  Luminex SD  Luminex Sheath CN S001  01  Delivery System   Product Description Customer  Number  Luminex 100 IS Version 2 3  Software CD  CN SW0
9.  Maintenance and Cleaning 3 1  Daily Maintenance    3 1  Before Running Samples             0 0 0 0    00 eee ee 3 1  After Running Samples             0 0 0 0  c eee eee eee 3 3  Routine Tasks  eene eres bb E REP VpS E 3 3  Sheath and Waste Fluids                 0 000000 000005 3 3  Weekly  oet   d d etia Pais wean dete e du esta 3 5  Visual Inspection    3 5  Run Self Diagnostics    0 0 0 0    cee cee eee 3 5  Clean the Sample Probe            0 0 0 0    0c ee eee ee 3 5  Flush the System     0 0    2  eee eee ee eee 3 5  Monthly sesser errit ata Teno a oe vale E REEC RET EE 3 6  Clean Exterior Surfaces         0 0 0 0    0 cc eee eee ee 3 6  Calibrate the system      0    2    cece ee eee 3 6  Every Six Months ssa 0 0 0 0    cece cece 3 6  Luminex 200 Analyzer Air Intake Filter                  3 6  Luminex XYP Instrument Air Intake Filter                3 7  Syringe Seal    sc  ce bee cesis ma eh b he rry 3 9  Luminex 200 Analyzer Ventilation Filter                 3 10  Amnually 52 decani aede s o pa eda eura ESA buried 3 11  Sheath Filtri ii coe dae eee c   eceet ECT ted 3 11   AS requited  i iciesseuc nie hi  pev pd e eb Rd cer LAGOS 3 12  Ju PR                 m 3 12  Replacing the SD System with a Sheath Bottle             3 13  Storing the System      0 0 0    eee 3 13  Taking the System out of Storage                   0   3 13  Maintenance Log      0 0 0    cece cette 6 13  Troubleshooting 4 1  Troubleshooting the Luminex 200 System                     4 1  Power
10.  Supply Problems    eese 4 2  Communication     0 0    0  eee ee n 4 2  Pressuriz  ti  n iaoee ansie ea ee eee ea ye ee 4 3  Fluid Leaks   eso ies tt eee eh See Er ee et 4 4  sample Probe    os sese esu dee   e ted e does 4 4  Calibration and Control Problems                     00005 4 6  Acquisition Problems           l sleeeeeeeeeeA 4 8  Bead Detail Irregularities     2 0 0 0    0  eee eee ee eee 4 9  Brror States    ia  cea hei ned hese eet tae eee ee EU ears 4 12  System Error Messages           2 0 0 0  c eee eee eee 4 12    ii PN 89 00002 00 109 Rev  A    xMAP Technology    Sample Error Messages          lee ees 4 14  Luminex SD Problems                lseseeeeeeeeeens 4 16  luli ET 4 16  Malin cui On sss sare oa keene we Re eee are e ute etre 4 17  Draining the Reservoir    4 17  Product Numbers 5 1  Hard Wwate   el oie eee Re ER E E a OR RN ree 5 1  SOTUWAEG  Lu i wale phates te lhe a UR Rete Nahe aa te edo ene cicer oe 5 3  XMAP Reagents    22  e600  bc eae eee e he 5 3  Tr  ting   22v epe p idee b ute dap tede a podeis cs 5 3    PN 89 00002 00 109 Rev  A    Contents    iii    Luminex 200 System User Manual    xMAP Technology    PN 89 00002 00 109 Rev  A    Safety    Please become familiar with the information in this chapter before    using the equipment  Do not perform procedures on your Luminex      200    system that are not specifically contained in this manual   unless you are directed to do so by Luminex Technical Support     Intended U se The Luminex softwar
11.  at a 90 degree angle to the plate  while dispensing them     Use a fresh bottle of calibrator  microspheres     Adjust the sample probe height   See page 3 2     Clean the sample probe  See  page 3 5     Clean the sample probe  then  flush the system  See page 3 5     Verify sample probe height  Run  three Prime commands  two  Alcohol Flush commands  then  three Washes with water     Check that the seal in the waste  container cap is dry and that the  waste container cap is vented     Verify that the waste line is not  moved during system operation     View the calibration trend  report  Check for dramatic  changes in temperature  sheath  pressure  or voltages  If any of  these are evident on the report   contact Technical Support     PN 89 00002 00 109 Rev  A    xMAP Technology    Zero events collected during    calibration     Analyzer fails Controls     PN 89 00002 00 109 Rev  A    There is a problem with fluid  levels     Laser related issue     The waste container is not  vented     The Cheminert fitting is loose     The control microspheres are  not fully suspended     Wrong control lot number or  target values are entered in  Update CON Targets box     The system controls are in the  wrong well on the plate     Not enough control  microspheres added to the well     Control lot is expired     Control microspheres have been  diluted     The sample probe height is  incorrect     The sample probe is clogged     Troubleshooting    Check the sheath and waste  fluid levels  
12.  in the UV  spectrum     1 4 PN 89 00002 00 109 Rev  A    xMAP Technology    Decontaminating  the Luminex 200  Analyzer for Return  Shipment       Note  It is the user   s  responsibility to             decontaminate the       analyzer before       shipment                                                              PN 89 00002 00 109 Rev  A    Safety    Luminex Technical Support will give you a Return Material  Authorization  RMA  number if they direct you to return the system   They will explain how to return the system according to Luminex  procedures     The accessible surfaces and the internal fluidics system must be  sanitized and decontaminated before returning the analyzer  This is  particularly important when biohazardous samples have been run   Make a copy of this page to fill out and return with the system     Complete the following checklist  signed and dated  and return it  with the Luminex 200 analyzer    1  Remove all specimens  disposables  and reagents from the system    2  Disconnect the sheath line going from the SD system to the analyzer     3  Connect a sheath bottle filled with a solution of 10  to 20  household  bleach solution to the analyzer     4  Sanitize the system using the Sanitize command on the main screen of  the system  Follow this by washing twice with distilled water     5  Disconnect the system from AC power by turning off the power switch  on the rear of the system  then unplugging the analyzer power cord  from the wall source     6  Dis
13.  is not a disconnect means  power cords must be removed  from the outlet  Contact Luminex Corporation for further  information     You must monitor waste levels manually  Do not allow the waste  container to overflow  Empty the waste container each time the  sheath fluid container is filled  Do not place the waste container on  top of the instrument  Contact Luminex Technical Support before  relocating the waste container or rerouting the waste line     Warning  If biological samples have been tested with the system   use your standard laboratory safety practices when handling    system waste        The Luminex 200 system classifies per FDA 21 CFR 1040 10 and  1040 11 as a Class II laser product consisting of a Class I laser  product  Luminex 200 analyzer  and a Class II laser product   barcode reader      Under NO circumstances should you remove the Luminex 200  analyzer cover  When performing routine maintenance  turn power  to the Luminex 200 analyzer OFF and disconnect the power cord     All laser apertures are located within the Luminex 200 analyzer and  are contained within a protective housing     Warning     Use of controls or adjustments or performance of  procedures other than those specified herein may result in    hazardous radiation exposure        Do not stare into the beam or shine it into other people s eyes     Warning  During operation  this system contains exposed   moving parts  Risk of personal injury is present  Observe all  warnings and cautions     War
14.  line  and the cap is tightened     Prime the system     As the system is priming  adjust the black regulator screw inside  the center door of the Luminex 200  Turn clockwise to increase  or counterclockwise to decrease  until the sheath pressure value  is within     0 1 psi of the normal level of your instrument   Visually monitor the pressure gauge on the Run Batch tab  Make  sure pressure is centered within the green tolerance range     When the prime function is complete  loosen the sheath bottle  cap to release pressure     Tighten sheath cap and prime the system once again     Make sure that the sheath and air pressure values increase to the  normal level once again     This procedure explains the steps you should take before placing the  system into long term storage         Before storing the system     1     Run a sanitize with 10  to 20  household bleach solution   Run a sanitize with distilled water   Run four washes with distilled water     Remove the sample probe from the instrument  flush with  distilled water from the narrow end out through the larger end   replace it in the sample arm  and wrap the end with Parafilm     Follow this procedure before you begin using the system after it has  been stored long term         Before using the system     1     Turn on the Luminex 200 and XY Platform and watch for the  following indications that the instruments are responding  correctly     xMAP Technology Maintenance and Cleaning      The light above the sample probe o
15.  sample probe  then  flush the system  See page 3 5     Verify that the delta cal  temperature is within   3 0  degrees  Recalibrate if outside  these limits and run the sample  again     Verify that the system is  functioning properly by running  system calibration and controls   If they pass  contact the kit  manufacturer        Cannot calculate inverse  function    This error message encompasses a range of mathematical errors  often indicating the sample result was negative or invalid  This is  according to the parameters defined by the formula used to analyze  the sample results  Standards  assay controls  and unknowns may be  flagged with Unknown Formula Failure  A standard curve is plotted  using the MFI and expected concentration value  An Unknown  Formula Failure message may occur when the system is unable to  calculate a standard concentration from the standard curve     You are using the wrong  template     You are using incorrect lot  information     There is a problem with the kit     Verify that you are using the  correct template  Consult the  online help     Verify that you have entered the  correct lot information     Run system calibration and  controls  If they pass  contact  the kit manufacturer        Luminex SD  Problems    Filter    If the empty sheath fluid container is not replaced and the system  continues to operate  the Luminex SD system eventually vents  pressure to prevent air from being introduced into the Luminex 200  analyzer  This may interrup
16.  system performs self   diagnostic tests  The results display on the Diagnostics tab  The  lasers shut off during Self Diagnostics  and must warm up again for  30 minutes after Self Diagnostics completes     Danger  Be sure that the system is not performing an operation  when you remove the sample probe     Caution  The Luminex XYP instrument sample probe should slide  up easily while removing it from the sample arm  If you feel    resistance  do not force the probe up  Contact Luminex Technical  Support            To clean the sample probe     1  Remove the sample probe as follows  Unsnap the light housing  located above the probe  Then  unscrew the Cheminert fitting on  top of the probe completely  Next  gently grasp the probe and  push up  Remove the probe out of the top of the sample arm     2  Clean the sample probe using a bath sonicator or using a 10 mL  syringe  If you are using a bath sonicator  place the bottom of the  sample probe in the bath sonicator for 2 to 5 minutes  If you are  using a syringe  force distilled water through the bottom of the  sample probe     3  Replace the sample probe and adjust the vertical height  You  should adjust the vertical height anytime the probe is removed     4  Alcohol Flush the system with 70  isopropanol or 70  ethanol     Run 3 backflushes  3 drains  2 alcohol flushes and 3 washes with  distilled water     Luminex 200 System User Manual    Monthly    Clean Exterior Surfaces    Calibrate the system    Every Six Months    Lumin
17. 109 Rev  A    The air pressure is out of range     The sample probe is not  vertically aligned     Air is present in the system     The acquisition volume is set  too high     The xMAP microspheres are not  fully suspended     You are using photobleached  microspheres     There is an insufficient number  of beads in the sample     The sample is too concentrated     Troubleshooting    See Pressurization on page 4 3     Adjust the sample probe height   See page 3 2     Verify sample probe height  Run  three Prime commands  two  Alcohol Flush commands  then  three Washes with water    Set the acquisition volume to at  least 25 uL less than the actual  volume in your wells  This  setting lets the analyzer acquire  sample more efficiently with  less chance of acquiring air     Gently vortex the plate or  resuspend beads with a  multichannel pipettor to ensure  that the microspheres are  present in the solution     Replace the microspheres with a  fresh batch     Ensure that there are 2000 5000  beads per beadset per well     Dilute concentrated biological  fluids  such as serum or plasma   at least 1 5     Use these tools to assist in diagnosing system and kit related    problems       system calibrators    system controls     assay standards     assay controls     error messages    Review calibration control trend reports routinely to detect trends     Use system xMAP control microspheres to check the success of the  system calibration and for troubleshooting purposes  If there i
18. 8 01  Product Description Customer  Number  Microspheres  LX100  CL1 CL2 Calibration L100 CAL1   xMAP Classification Calibrator   Microspheres  LX100  RP1 Calibration L100 CAL2   xMAP Reporter Calibrator   Microspheres  LX100  CL1 CL2 Control L100 CON1   xMAP Classification Control   Microspheres  LX100  RP1 Control  xMAP L100 CON2  Reporter Control   xMAP Sheath Fluid  LX100 40 50000  Product Description Customer  Number  Basic Training Course CN 0070 01             Luminex 200 System User Manual    xMAP Technology    PN 89 00002 00 109 Rev  A    
19. Luminex        Luminex 200  System User Manual        LUMINEX CORPORATION  2001 2005  All rights reserved  No part of this publication may be reproduced   transmitted  transcribed  or translated into any language or computer language  in any form or by any means  without prior express  written consent of     12212 Technology Boulevard  Austin  Texas 78727 6115  U S A     Voice   512  219 8020  Fax   512  219 5195    i LUMINEX CORPORATION    Luminex   200    System User Manual    CN M064 01    PN 89 00002 00 109 Rev  A  July 2005    Luminex Corporation  Luminex  reserves the right to modify its products and services at any time  This guide  is subject to change without notice  Although prepared to ensure accuracy  Luminex assumes no liability for  errors or omissions  or for any damages resulting from the application or use of this information     The following are trademarks of Luminex  Luminex  Luminex 100  Luminex 100 IS  Luminex 200  LabMAP   XMAP  LumAvidin  Luminex XYP  Luminex SeroMAP  Luminex FlexMAP  and Luminex SD  All other trade     marks  including Windows  Cheminert  Pentium  and Dell are trademarks of their respective companies     The Luminex 100 IS software uses the VideoSoft   VsFlexGrid Pro 7 0  VsPrinter 7 0  and VsView 3 0  ActiveX controls  which are copyrighted by VideoSoft  2001     The contents of this manual and the associated Luminex software are the property of Luminex and are  copyrighted  Except as specified in the End User License Agreement  any 
20. UL 61010A 1    Shipping and storage  The allowable shipping and storage  temperature and humidity ranges are 0  C to   50  C and 20 80   noncondensing  respectively    Input voltage range  100   120 V    10   1 4 Amp  and 200 240  V  10   0 8 Amp  47 63 Hz     AC inlet fuse  3 Amp  250 V   fast acting    Reporter laser  532 nm  nominal output 10 16 5 mW  maximum  500 mW  frequency doubled diode  mode of operation   continuous wave  CW     Classification laser  635 nm  9 1 mW   6   maximum output 25  mW  diode  mode of operation  continuous wave  CW     Reporter detector  Photomultiplier tube  detection bandwidth of  565 585 nm    Classification detector  Avalanche photo diodes with temperature  compensation    Doublet discrimination detector  Avalanche photo diodes with  temperature compensation    Sheath flow rate 90 uL   5 uL second  Cuvette  200 micron square flow channel  Sample injection rate  1 uL second  Sample uptake volume  20 200 uL    Reporter channel detection  A D resolution 14 bits  Communications interface  USB  Luminex XYP instrument  communications interface  RS 232    Ambient temperature  15  C to 30  C  59  F to 86  F   Humidity  20  to 80   noncondensing  Altitude  operation up to 2400 m  7874 ft  above mean sea level    Physical dimensions  44 cm  17 25 inches  W x 60 cm  23 5  inches  D x 8 cm  3 inches  H    Weight  15 kg  33 Ibs      UL installation category  UL Installation Category II  as defined  in Annex J of UL 61010A 1    Luminex 200 S    ystem U
21. Verify that tubing  for both bottles is tightly  connected to the instrument   Check that the waste bottle cap  is vented     Verify fluid is moving through  the system by performing a  Wash  which causes fluid to go  out to waste  If no fluid is going  to waste  clean the sample  probe  then flush the system   see page 3 5   If the issue does  not resolve  contact Technical  Support     Check that the seal in the waste  container cap is dry and that the  waste container cap is vented     Make sure that the Cheminert  fitting is tightened     Vortex the control vials to  resuspend the microspheres     Verify that the correct lot  number and target values are  used     Verify that control microspheres  are in the correct well     Add at least five drops of control  microspheres to the well  For  accurate drop volume  hold the  vial upside down at a 90 degree  angle to the plate while  dispensing them     Use a fresh bottle of control  microspheres     Do not dilute the control  microspheres     Adjust the sample probe height   See page 3 2     Clean the sample probe  See  page 3 5     Luminex 200 System User Manual    There is air in the system     The waste container is not  properly vented     The waste line was moved  during system operation   resulting in unstable flow rate     Possible problem with the  lasers     xMAP Technology    Verify sample probe height  Run  three Prime commands  two  Alcohol Flush commands  then  three Washes with water    Verify that the waste co
22. alyzer  On systems with a  Luminex Sheath Delivery  system  ensure that the sheath  fluid pressure reading is equal to  the original reading  If it is not   contact Technical Support for  instructions to adjust the  regulator  Calibrate and verify  the system     See Pressurization on page 4 3     Warm up the system  This takes  approximately 30 minutes   Calibrate and verify the system     Open the calibration trend  report for CAL2 and record the  last three reporter voltages   Contact Luminex Technical  Support with this information     Check the system to verify that  it is warmed up and that there  are no pressurization problems   Repeat the command  If the  command continues to time out   contact Technical Support     Calibrate the instrument     Luminex 200 System User Manual    Sample Error    xMAP Technology    Sample error messages occur when there is a problem with sample       Mess ag es acquisition  These errors can display on the Diagnostics tab in the  Message Log  in Reports  or on the Errors tab in a batch analysis   Message Possible problem Solution       Insufficient Bead Count    Temperature Divergence from  Calibration Temperature    Failed Control in Batch    Failed Curve Fit    There are not enough beads seen  to meet the minimum requested     You are running the system at  an ambient temperature that is  out of range of the temperature  at which the system was initially  calibrated     One or more of the assay  controls are not within the  expected ran
23. an and maintain the  Luminex 200 system  Read and follow all instructions in this section   For your convenience  a maintenance log form is included at the end  of this chapter     It is important to use only xMAP Sheath Fluid or other Luminex   approved sheath fluid  USE OF NON LUMINEX APPROVED SHEATH  FLUID SHALL CONSTITUTE    IMPROPER USE  AND MAY VOID  WARRANTY RIGHTS PROVIDED BY LUMINEX AND OR ITS  AUTHORIZED PARTNER     Warning  When analyzing potentially infectious biological  samples on the Luminex 200 analyzer  follow standard    laboratory safety practices  These safety precautions should also  be taken when cleaning or maintaining the analyzer        Do not remove the analyzer cover under any circumstances     If the system is powered on  but idle for more than four hours  click  the Maintenance tab  Click Warmup  Wait 30 minutes for the  Luminex 200 analyzer and the optics system to warm up         Before you run samples     1  Turn the Luminex 200 analyzer on  The system begins warming  automatically     2  Verify the levels of sheath fluid and waste fluid     3  Click Prime to prime the analyzer  then click OK     Luminex 200 System User Manual    Adjust the sample  probe vertical  height    Note  Verify that the microtiter  plate is not warped  Warped  plates can lead to incorrect  probe height adjustment     Note  Alignment discs can be  placed in any well as long as the  well is designated in the software     xMAP Technology    4  Click Alcohol Flush  A confi
24. ce     While facing the Luminex 200 analyzer  place your index finger  up under the right side of the analyzer  in the space between the  Luminex 200 analyzer and the Luminex XYP instrument   When  you feel the filter  push the filter toward the left of the analyzer   See Figure 3 9     Remove the filter from the left side of the Luminex 200 analyzer        Figure 3 9 Luminex 200 Analyzer Ventilation Filter    Clean the filter with a vacuum or with distilled water  Stand the  filter upright to air dry     Re install it with the arrows facing up  The filter should click  into place     Plug in and turn on the Luminex 200 analyzer power     PN 89 00002 00 109 Rev  A    xMAP Technology    Annually  Sheath Filter    PN 89 00002 00 109 Rev  A    Maintenance and Cleaning      To change the Luminex 200 analyzer sheath filter     1     N A M    Disconnect the Luminex 200 analyzer from ac power by turning  off the power switch on the rear of the analyzer  then unplugging  the analyzer power cord from the wall source     Disconnect the sheath fluid tubing before changing the filter     Open the left door on the Luminex 200 analyzer  Disconnect the  filter by pushing down on the metal clamps on each connection   See Figure 3 10        Figure 3 10 Sheath Filter    Connect the new sheath filter  matching up the color coded  fittings  The arrow on the sheath filter should be pointing up     Reconnect the sheath fluid tubing   Plug in and power on the Luminex 200 analyzer   Close the left a
25. connect the SD system and waste and sheath containers     7  Rinse the waste container with 10  to 20  household bleach solution  and drain     8  Wash all exterior surfaces with a mild detergent  followed by a 10  to  2096 household bleach solution     9  Open the front doors of the analyzer  Clean all accessible surfaces with  mild detergent followed by a 1096 to 2096 household bleach solution     10  Pack the system within a biohazard bag  place it in the corrugated box   then insert it in its original packaging or an approved shipping  container  Attach this checklist to the top of the corrugated box prior to  packaging in the crate                                   Was there an internal leak in the system  Yes No  Print Name    Signature    Date  Instrument Serial No        xMAP Technology Safety    PN 89 00002 00 109 Rev  A 1 6    The System    Theory of Operation    PN 89 00002 00 109 Rev  A    Luminex 200 technology is based on flow cell fluorometry with  Luminex developed innovations  The fluidics  optics  robotics   temperature control  software  and xMAP   microspheres work  together to enable simultaneous analysis of up to 100 analytes in a  single test sample  Assay analysis requiring temperature control is  provided through the Luminex XYP instrument heater block     There are two fluidics paths in the Luminex 200 analyzer  The first  path involves a syringe driven mechanism that controls the sample  uptake  This mechanism permits small sample uptake volumes fro
26. d press the Prime button on the SD     Contact Technical Support for additional information     Luminex 200 System User Manual    xMAP Technology    PN 89 00002 00 109 Rev  A    Product Numbers    Hardware    Note  These part  numbers are subject to  change without notice     Note  Common  description of product is  enclosed in brackets     PN 89 00002 00 109 Rev  A                                              Product Description Customer  Number  Rear Air Filter CN 0001 01  Bottom Air Filter CN 0002 01  Air Filter  Intake CN 0027 01  Alignment guide  XYP instrument CN 0016 01  Bar Code Scanner CN PCO03 01  Serial Cable  5 feet CN 0005 01  PC  Luminex 100 IS CN PC04 01  2 Amp  250 Volts  Fast Acting Fuse  Qty 10    CN 0019 01  3 Amp  250 Volts  Fast Acting Fuse CN 0051 01  Heater Block  XYP CN 0017 01  Wrench  Hexdrive  Ball Driver CN 0025 01  3 32     Luminex 100 IS  Version 2 3  complete CN L005 01  system   Luminex 100 IS  Version 2 3  complete CN L007 01    system  with laptop                   Luminex 200 System User Manual    xMAP Technology                                           Luminex 100 IS Documentation Version 2 3   CN M032 01    Manuals on CD    Luminex 100 IS User Manual  Version 2 3 CN M01 7 01    North American    Luminex 100 IS User Manual  Version 2 3 CN M018 01    International English    Luminex 100 IS User Manual  Version 2 3 CN M031  01    Danish    Luminex 100 IS User Manual  Version 2 3 CN M019 01    French    Luminex 100 IS User Manual  Version 2 3
27. e is designed to use xMAP   technology with  assay kits available through kit manufacturers  The Luminex 200  system performs a wide range of xMAP technology based laboratory  tests  measuring biomolecular reactions on the surface of xMAP  microspheres     Wa rn ings and Notes Informational notes and warnings appear in this manual     Note  A note provides general helpful information  No safety or  performance issues are involved     Caution  This message is used in cases where the hazard is minor or  only potential hazard is present  Failure to comply with the caution  may result in potentially hazardous conditions     Warning  This message is used in cases where danger to the  operator or to the performance of the instrument is present  Failure to  comply with the warning may result in incorrect performance   instrument failure  invalid results  or hazard to the operator     Danger  This message is used in cases where significant risk of  serious injury or death is present     PN 89 00002 00 109 Rev  A 1 1    Luminex 200 System User Manual    Symbols    xMAP Technology    These symbols describe warnings  cautions  and general information  used in the operation of this instrument  Some of these symbols are  further defined under    Safety Precautions           Symbol Description Description Description  Warning Warning  MN Alternating current  refer to manual   refer to manual      ac        Protective ground    Warning   refer to manual     Warning   refer to manual        
28. e sheath container ever runs dry  prime the system at least twice  until the air is removed from the system     Empty the Waste   To empty the waste container     Container  1  Disconnect the waste container from the Luminex 200 analyzer   2  Unscrew the waste container lid  being careful not to touch the  Gore Tex    seal  If the seal becomes wet or soiled  it may impair  venting   Note  There is NO warning of 3  Discard the waste from the waste container by appropriate  a high waste volume  Empty m  ns    the waste container each time   YOL TEO sel container  4  Reconnect the waste container to the Luminex 200 analyzer and  replace the cap    Any time the sheath container is disconnected from the Luminex  200 analyzer  you must remove air from the sample lines by    priming     3 4 PN 89 00002 00 109 Rev  A    xMAP Technology    Weekly    Visual Inspection    Run Self Diagnostics    Clean the Sample Probe    Flush the System    PN 89 00002 00 109 Rev  A    Maintenance and Cleaning    Open all of the Luminex 200 analyzer doors and visually inspect for  leaks  corrosion  and other signs of improper function  Check all  visible tubing connections  Check the Luminex XYP instrument air  intake filter for buildup of dust  Check the SD system and its  connection for leaks  If you see a leak  turn off the power to the  Luminex SD system and contact Luminex Corporation     Run Self Diagnostics on a weekly basis to check system integrity  On  the Maintenance tab  click Self Diag  The
29. ex 200 Analyzer Air  Intake Filter    Note  Hold on to the tubing   Do not allow the tubing to fall  inside the the analyzer     xMAP Technology        To clean exterior surfaces  follow these steps     1     Disconnect the system from ac power by turning off the power  switches and unplugging the Luminex 200 analyzer  the Luminex  XYP instrument  and the Luminex SD system     Wipe all exterior surfaces with mild detergent  then a 10  to  20  household bleach solution  and finally with plain distilled  water     Open both doors of the analyzer  Clean all accessible surfaces  with detergent followed by a 10  to 20  household bleach  solution and then plain distilled water     Dry the sheet metal surfaces to prevent corrosion     Plug in and power on the Luminex 200 analyzer  the Luminex  XYP instrument  and the Luminex SD system     You should calibrate the system on a monthly basis as part of    regularly scheduled maintenance  For instructions on calibrating the    system and verifying calibration  see the Luminex online help or the  Luminex IS Software Manual For Version 2 3         To replace the analyzer air intake filter     1     Disconnect the Luminex 200 analyzer from ac power by turning  off the power switch on the rear of the analyzer  then unplugging  the power cord from the wall source     On the back of the Luminex 200 analyzer  in the upper left  corner  remove the screw at the top of the panel and open the  panel door     Grasp the tubing and pull the filter 3 
30. ge     The calculation could not occur  for the assay standards for data  interpretation     Ensure there are 2000 5000  beads per set per well    Ensure proper beads are  selected    Adjust sample probe height  See  page 3 2    Resuspend beads in the wells   Check for clogs     Calibrate and verify the system   If this message still appears  after calibration  contact  Technical Support     Verify that you selected the  correct template     Verify that the correct assay lot  number and concentration  values were selected and entered  correctly     Verify that the system is  functioning properly by running  system calibration and controls   If the system controls pass   contact the kit manufacturer     Verify that you selected the  correct template     Verify that the correct control  lot number and target values  were selected and entered  correctly     Verify that the system is  functioning properly by running  system calibration and controls   If the system controls pass   contact the kit manufacturer     PN 89 00002 00 109 Rev  A    xMAP Technology    Sample High or Low    Sample Timed Out    Sample Empty Detected    PN 89 00002 00 109 Rev  A    The results do not fall within the  highest and lowest standards     The wrong control lot number  or target values are selected     The samples are concentrated     Not enough sample loaded into  the well     The sample probe is clogged   Sample probe height is    incorrect     There is a clog in the sample  line     The system c
31. ging the    plunger  injury could result if the system is not unplugged        1  Disconnect the Luminex 200 analyzer from ac power by turning  off the power switch on the rear of the analyzer  then unplugging  the power cord from the wall source     2  Open the center most door on the front of the Luminex 200    analyzer     3  Locate the syringe  a glass cylinder with a metal rod   See  Figure 3 8             1  Syringe Seal 2  Syringe  Figure 3 8 Syringe and Syringe Seal    4  Unscrew the knob on the syringe arm  at the bottom of the  syringe   and forcefully push the syringe arm down     Unscrew the syringe from the top of its housing     Pull the plunger out of the syringe     D  9v d     Remove and replace the plunger seal and the black    O    ring   8  Return the plunger to the syringe   9  Screw the syringe back into its housing     10  Return the syringe arm to its original position     Luminex 200 System User Manual    Luminex 200 Analyzer  Ventilation Filter    11   12     13     14          xMAP Technology    Hand tighten the screw on the syringe arm     Plug in the power cord and turn the Luminex 200 analyzer power  on     Prime the system twice  watching for any leaks in the syringe  area     When the prime finishes  close the center door   To clean the Luminex 200 analyzer ventilation filter     Disconnect the Luminex 200 analyzer from ac power by turning  off the power switch on the rear of the analyzer  then unplug the  analyzer power cord from the wall sour
32. he optics assembly consists of two lasers  One laser excites the dye  mixture inside the xMAP microspheres and the second laser excites  the fluorophore bound to the surface of the xMAP microspheres   Avalanche photo diode detectors measure the excitation emission  intensities of the color coding classification dye mixtures inside the  xMAP microspheres and a photomultiplier tube detects the excitation  emission intensity of the reporter molecule bound to the surface of  the xMAP microspheres  High speed digital signal processors and  advanced computer algorithms provide analysis of the xMAP  microspheres as they are processed through the Luminex 200  analyzer  Results of the analyses are processed and provided in a  report format     The Luminex 200 system includes the following hardware     e  Luminex 200 analyzer   e Computer  PC   monitor  and accessories   e Luminex XYP instrument   e  Luminex Sheath Delivery System  Luminex SD      e Power cables   e Alignment guide     Two long sample probes   e Reservoir   e Shield   e Heater block   e Sheath fluid container   e Waste container   e Sheath fluid line     Air line     Sheath fluid intake line   e Communications  1 serial communication cable  e Communications  1 USB communication cable  e Communications  1  CANBUS cable  short cable   e Barcode reader   e Sample probe height alignment kit   e 3 32 Hexdrive  Balldriver wrench    e Classification calibration microspheres  CAL 1   e Reporter calibration microspheres  CAL2  
33. heath fluid connector is  blue  and the waste fluid connector is orange     For proper operation  place the Luminex SD system at the same level  as the base of the Luminex X YP instrument  Do not put it on top of  the Luminex 200 analyzer  If you are not using the SD system  sheath  fluid levels must be monitored manually  Check the sheath fluid level  before starting a run or procedure     Warning  If biological samples have been tested with the  The waste fluid container receives waste from the system  The waste  container should not be placed on top of the instrument  Ensure that  the waste tubing is not elevated above the level of the Luminex 200  analyzer at any point  Contact Luminex Technical Support before    you relocate the waste fluid container  To maintain a stable flow rate   do not move the waste line or container during system operation     Caution  Waste levels must be manually monitored  Do not  allow the waste container to overflow     The optical system consists of the optical assembly and the excitation  lasers  The optical assemblies do not require manual adjustment by  the user     The xMAP technology reagent system consists of classification  calibration microspheres  reporter calibration microspheres   classification control microspheres  and reporter control  microspheres     PN 89 00002 00 109 Rev  A    Maintenance and Cleaning    Daily Maintenance    Before Running Samples    PN 89 00002 00 109 Rev  A    To ensure accurate test results  properly cle
34. hen click on the  Support link to open the FAQ list     Luminex highly recommends using an uninterruptible power supply   UPS  to protect your system from power outages  Choose one that  can provide 1050 Watts for at least 45 minutes  The UPS should be  UL listed  CSA certified  and CE marked when used internationally     If you do not use a UPS  use a surge protector  Choose a protector  that meets your needs  Factors to consider include electrical  environment  endurance  suppressed voltage rating  and method of  protection  It should have six outlets  rated at least 1500 Watts  and    PN 89 00002 00 109 Rev  A    xMAP Technology    Printer    Barcode Labels  Vortex    Bath Sonicator    System Overview    Electronics  Power Input  Module  Communications  Ports  SB9 PIN   Luminex 200  Analyzer    Ventilation Filter    Luminex XYP  Instrument  Ventilation Filter    PN 89 00002 00 109 Rev  A    The System    be UL listed  CSA certified  CE marked for nondomestic use when  used internationally     Printer  HP LaserJet 2300 or available equivalent    Use the Code 128 barcode label type when scanning barcode labels  into the system as patient identities     VWR product number 58816 121  Speed range 0 3200 rpm or  equivalent    Cole Parmer   product number 08849 00  Operating frequency 55  kHz or equivalent    The system consists of three subsystems  electronic  fluidic  and  optical  The following section describes the user accessible  components of each subsystem     The power i
35. ines to the analyzer  If  pressurization fails again   replace the sheath bottle     Ensure that the sheath bottle is  not filled above the fill line     Drain the SD reservoir and refill  it  See page 4 17     If using bottles  open the center  door on the Luminex analyzer   Use a screwdriver to adjust the  regulator to fit in the center of  the green region on the Run  Batch tab     Luminex 200 System User Manual    Fluid Leaks    xMAP Technology    Fluid leaks can result in poor pressurization and failed sample    acquisition        Symptom    Possible problem    Solution       Pressure too low     Large amount of fluid pooled  around instrument     Fluid dripping from the  sample probe     Fluid is leaking from the front  of the analyzer     Sample Probe    The sample probe is clogged     The syringe seal leaks     The syringe valve leaks     Fittings or fluid lines are  damaged     The sample probe is clogged   The sample three way valve is  faulty     The syringe seal leaks     The syringe valve leaks     Clean the sample probe  See  page 3 5     Replace the syringe seal  See  page 3 9     Hand tighten the syringe  connection  silver knob  on the  syringe valve  Run a Prime  command  If leaks continue   contact Technical Support     Contact Technical Support     Clean the sample probe  See  page 3 5     Contact Technical Support     Replace the syringe seal  See  page 3 9     Hand tighten the syringe  connection  silver knob  on the  syringe valve  Run a Prime  If  leak
36. m  small reaction volumes  The syringe driven system transports a  specified volume of sample from a sample container to the cuvette   The sample is injected into the cuvette at a steady rate for analysis   Following analysis  the sample path is automatically purged with  Luminex xMAP Sheath Fluid by the second fluidics path  This  process removes residual sample within the tubing  valves  and  probe  The second fluidics path is driven by positive air pressure and  supplies sheath fluid to the cuvette and sample path     Luminex xMAP Sheath Fluid is the delivery medium of the sample  to the optics component  The analysis sample is acquired using a  sample probe from a 96 well microtiter plate via the Luminex XYP  instrument and injected into the base of the cuvette  The sample then  passes through with sheath fluid at a reduced rate resulting in a  narrow sample core to ensure that each microsphere is illuminated  individually  The sample injection rate is such that the xMAP  microspheres are introduced to the optics path as a series of single  events  The Luminex SD system lets you run samples continuously  without refilling sheath bottles  It automatically draws sheath from a  non pressurized bulk sheath container to constantly maintain a  reservoir of pressurized sheath fluid  A single 20 liter sheath  container provides enough fluid for 48 hours or more of normal  operation     Luminex 200 System User Manual    Hardware    xMAP Technology  Reagents    xMAP Technology    T
37. mple   Sample core  15 20 um core at 1 uL sec  sample inject rate  Maintain samples at a constant temperature from 35  C to 55  C   95 F to 131 F    Automatic sampling from a 96 well plate   Start sampling from any well position   Sheath container and waste container hold enough volume to run  up to two 96 well plates between refills   Microtiter plates with 96 wells must be compatible with the  Luminex XYP instrument plate holder  The following microtiter  plate types are compatible with the Luminex XYP instrument  plate holder  flatbottom  conical  round  filter bottom  half plates   overall height no more than 0 75   19 mm    any color   Microtiter plates with 96 wells must be compatible with  Luminex XYP instrument heater block temperature from 35 C to  55  C  95  F to 131  F  when performing heated assays and using  the heater block     Indoor use only  Operating temperature  15  C to 30  C  59  F to 86  F   Humidity  20  to 80   noncondensing    PN 89 00002 00 109 Rev  A    xMAP Technology    Optics    Fluidics    Electronics    Luminex XYP  Instrument General    PN 89 00002 00 109 Rev  A    The System    Altitude  Operation up to 2400 m  7874 ft   above mean sea level    Physical dimensions  43 cm  17 inches  W x 50 5 cm  20 inches   D x 24 5 cm  9 5 inches  H    Weight  maximum of 25 kg  60 Ibs      UL installation category  UL Installation Category II  as defined  in Annex J of UL 61010A 1    Pollution degree  UL Pollution Degree 2  as defined in Section  3 7 3 2 of 
38. n the Luminex 200 and the  light next to the door of the XY platform are lit      The compressor starts in the Luminex 200  It is a low  rumbling sound       Place your hand behind the Luminex 200 to feel air coming  out of the rear fan       Observe movement of the syringe inside the front middle door  of the Luminex 200 shortly after the instrument is powered on     Turn on the PC and start up the IS software     Complete a Warmup command  which will take 30 minutes     B5 ow om    Remove the Parafilm from the end of the sample probe     5  After Warmup is complete  run three Backflush commands  three  Drain commands  two Alcohol Flush commands  and three  Washes with distilled water  Make sure that the sheath bottle or  Luminex SD has a sufficient amount of sheath fluid and that the  waste container is empty  Verify that the pressure during each of  the maintenance commands reads between 6 and 9 psi in the  System Monitor     PN 89 00002 00 109 Rev  A 3 14    Luminex 200 Maintenance Log Month s   Year        Procedure  Use this form to record information over a four week period  Fill in the month s  and year above  Fill in the dates in the first line of the table  For each item listed at the left  enter  your initials under each date that you perform the item     Note  Follow your standard laboratory safety practices when cleaning or maintaining the system  Do not remove the instrument cover under any circumstances        Dates       Daily maintenance       Startup Initial
39. nalyzer door     Prime twice     Luminex 200 System User Manual    As required    Fuses    xMAP Technology    Danger  To avoid serious injury or death by electric shock     you must turn off the system and unplug it from the wall  outlet            To replace the fuses in either the Luminex 200 analyzer or the    Luminex XYP instrument     Turn off the power switch on the rear of the analyzer or  instrument  then unplug the instrument power cord from the wall  source  Remove the power cord from the analyzer or instrument     With a small  flathead screwdriver  open the module door on the  lower left corner of the back of the analyzer or instrument  See  Figure 3 11     eres          2     Figure 3 11 Opening the Module Door  Remove the red cartridge  use a flathead screwdriver    Check both fuses for damage     Replace damaged fuses with the type specified on the sticker to  the right of the power input module     Replace the module door     Plug in and power on the analyzer or instrument     PN 89 00002 00 109 Rev  A    xMAP Technology    Replacing the SD System  with a Sheath Bottle    Storing the System    Taking the System out of  Storage    PN 89 00002 00 109 Rev  A           Maintenance and Cleaning    Turn off the SD and disconnect the tubing from the Luminex 200  analyzer     Turn the black regulator screw inside the center door of the  Luminex 200 one full turn counter clockwise     Attach the sheath bottle  Make sure there is sheath fluid in the  bottle  to the fill
40. ng as simple as a disconnected cable   Use extreme care when you replace a fuse   Symptom Possible problem Solution       Analyzer will not turn on  or  XYP will not turn on     Fuses continue to open  blow      Communication    The power cord is disconnected  Verify that the power cord is    plugged in     No voltage is coming from the  electrical outlet     Verify that the electrical outlet is  operational     The power supply is faulty  Contact Technical Support     A fuse has burned out  See page 3 12 for instructions    on changing fuses     A component has a short circuit  Contact Technical Support     Communication problems described in this section involve the links  between the data system  PC and IS 2 3 software  and the Luminex  200 analyzer and XYP instrument  This section does not address  communication issues with other peripheral devices     The term    Communication    refers to       The transfer of data between the PC and the analyzer     The current status of the analyzer and XYP instrument      nstrument readbacks        nstrument control  sample acquisition  session uploading  and  start  stop and pause features        Symptom    Possible problem Solution       PC cannot initialize  communication with analyzer     Check the communications  cable connections     The communication cable is  unplugged  or plugged into the  wrong port     Turn off the PC and then turn on  the analyzer  XYP  and then the  PC     The XYP or the analyzer power  is not turned on 
41. ning  During operation  this system contains exposed   moving parts which could result in puncture hazard  Risk of  personal injury is present  Keep hands and fingers away from the  Luminex XYP instrument slot during operation     Warning  During operation  this system contains exposed   moving parts which could result in pinch point hazard  Risk of  personal injury is present  Keep hands and fingers away from the  Luminex XYP instrument slot during operation        Luminex 200 System User Manual xMAP Technology    Access doors must be closed while operating the Luminex 200  analyzer  the operator must be present during operation     Biological  Warning  Human and animal samples may contain biohazardous  infectious agents     Where exposure  including aerosol  to potentially biohazardous  material exists  follow appropriate biosafety procedures and use  personal protective equipment  such as gloves  gowns  laboratory  coats  face shields  or mask and eye protection  and ventilation  devices        Observe all local  state  and federal biohazard handling  regulations when disposing of biohazardous waste material     Heat    Warning  The heater plate of the Luminex XYP instrument may  be hot and could cause personal injury if touched  Do not touch  the heater plate        Blue Indicator Light The blue light above the Luminex 200 analyzer sample arm indicates  the on off status of the Luminex 200 analyzer  and is harmless  The  blue light emitting diode  LED  does not emit light
42. nput modules contain the on off switch and fuses     The communications port connects the Luminex 200 analyzer or the  Luminex XYP instrument to the computer  and the Luminex SD  system to the Luminex 200 analyzer     Located on the bottom of the Luminex 200 analyzer  the filter must  be checked and cleaned as necessary  For proper ventilation  do not  obstruct the area below and allow at least two inches  5 cm  of  clearance around the Luminex 200 analyzer     The XYP instrument ventilation filter cleans the air that cools the  internal parts of the Luminex XYP instrument  See Figure 2 1     Luminex 200 System User Manual xMAP Technology       1  Air intake filter access door 5  Communication Ports  DB9   2  Power Switch 6  XYP Communication Port  DB9   3  Power Input Module 7  Analyzer ventilation filter     on bottom of analyzer   4  XYP Ventilation Filter    Figure 2 1 Back of the Luminex 200 Analyzer and Luminex XYP  Instrument    Fluidics    Luminex XYP A stainless steel sample probe acquires the sample   Instrument Sample  Probe Warning  During operation  this system contains exposed moving  parts that can result in a puncture hazard  Risk of personal injury    is present  Keep hands and fingers away from the sample probe   The shield should be in place        Cheminert   Fitting   This fitting attaches the sample probe to the sample tubing     Disconnect this fitting when you remove the sample probe  See  Figure 2 2     2 8 PN 89 00002 00 109 Rev  A    xMAP Technology
43. ntainer is  properly vented and that the seal  is not wet or soiled     Verify that the waste line is not  moved during system operation     Check the system control trend  report for consistent failures  If  any are evident on the report   contact Technical Support        Acquisition  Problems       Symptom    Possible problem    Solution       Acquisition fails or slows     Slow or unsuccessful sample  acquisition     The air pressure is out of range     The sample probe height is  incorrect     The sample probe is clogged     The sheath bottle has a leaky  seal     The sheath or waste lines are  not fully connected     The calibration microspheres  have expired     The wrong wells are selected for  the calibration microspheres     The calibration lot number or  target values selected in setup is  incorrect     The sample probe is clogged     See Pressurization on page 4 3     Adjust the sample probe height   See page 3 2     Clean the sample probe  See  page 3 5     Make sure that the sheath bottle  lid is tightened  Remove and  replace the sheath bottle lid     Disconnect and reconnect the  lines  you should hear them  click     Replace old microspheres with a  fresh lot     Ensure that the correct wells are  selected in the Maintenance tab     Enter the correct calibration lot  number and target values in the  Update CAL Targets box     Clean the sample probe  See  page 3 5     PN 89 00002 00 109 Rev  A    xMAP Technology    Bead Detail  Irregularities    PN 89 00002 00 
44. ompletely acquired  the entire sample volume     There is no sample in the  designated well     The probe is clogged     Troubleshooting    Verify that you selected the  correct template     Verify that the correct control  lot number and target values  were selected and entered  correctly     Verify that the system is  functioning properly by running  system calibrators and controls   If the sample is noted as High   dilute the sample following the  kit manufacturer   s suggestions   If the sample is noted as Low   vortex the sample to ensure  homogeneity     Verify that the samples are  properly mixed     Verify that you loaded the  correct sample volume to the  well     Clean the sample probe  See  page 3 5     Adjust the sample probe height   See page 3 2     Clean the sample probe  then  flush the system  See page 3 5     Verify that the delta cal  temperature is within   3 0  If  the system is not within these  limits  recalibrate  then run the  sample again     Verify that you properly mixed  the samples before dispensing  into the wells     Verify that you loaded the  sample to the correct well on the  plate     Verify that the sample volume  added to the well is correct     Clean the sample probe  See  page 3 5     Luminex 200 System User Manual    The sample probe is not  properly aligned     There is a clog in the fluid line     Delta Cal temp out of range     There is a problem with the kit     xMAP Technology    Align the sample probe  See  page 3 2     Clean the
45. p or allow it to  become wet or dirty  This could cause the bottle to become  pressurized  causing pressure errors in the system  If the seal gets  wet  let it air dry  Touching the seal can contaminate it  If it gets  dirty  you should replace the waste bottle cap     If you are using the waste line  do not move it while the system is  running  While you can move the line along a horizontal surface  do  not permanently reroute the waste line elevation without first  contacting Technical Support  You can move the line temporarily for  cleaning and maintenance purposes     You must manually monitor waste container levels  The Run Batch  tab displays a warning when you need to refill  or replace  the sheath  fluid container  See Figure 3 4     Luminex 200 System User Manual xMAP Technology       Yh Lumines    100 IS Software angl xl  File Tools Help i  Loma  Run Batch   Maintenance nuc  mer integrated System 2 3    lt No Batch Loaded gt     112 8 4 5   6 7 8  9  10 11  12     s  AOQOQOOOOO00000     e be   BC OOOOOOOOOOOQO Fs Zy  X  eCOOCOGOOCQOOO CQOOOQO re A Se A          BO00000000000                CmdNo Command PlateLoc __ Cmd Status  1 Wash A1 Plate1 Succeeded  23 sec                 Inna Warmed Up o al  Figure 3 4 Refill Sheath Fluid Warning  Circled in Photo     Refill the Sheath     To refill the sheath fluid container     Fluid Container  1  Release system pressure by removing the lid from the sheath    fluid container     2  Refill the sheath fluid container     If th
46. reproduction in whole or in part is  strictly prohibited     Contents    Safety  Intended Use irec aite dade 4a re o DE CE SEP ded  Warnings and Notes          2 0    cee eee eee ene  Syinbols  3 exstat eaedem e acean ved eade Urea  case  Safety Precautions         llle  PIUIGICS EE  Luminex 200 Analyzer Laser    esee   Barcode Reader Laser           0 0    ee ee     Blue Indicator Light             0 0 0    2c eee ee eee  Decontaminating the Luminex 200 Analyzer for Return  Sliipment   2  a tae tat Sah ade ah ah ed ees ke ied ae    The System    Theory of Operation    0 0 0    eee eee eee  Hardwate  iu esie dere gare ip E Rabe PR eL S ina A  xMAP Technology Reagents           lesser  Required Laboratory Reagents             lllleeleeeeesss  Luminex IS 2 3 Software       Capacity AP   Luminex 200 Analyzer General                  0000000000   OPCS  soi seed ee bed Pe ek nib hed ee eee ae  EluidiGs eee toD ees iege tate owls  BlectrODiCs  iu x el Ra ede es Ex a dee  Luminex XYP Instrument General                      000   Luminex SD System General                0 000000 eee eee  PC Specifications     0    0    eee e  Recommended Additional Equipment                      System Overview         0    cece eee eee  BlectrOnICS     3 2  bI aa ee le    PN 89 00002 00 109 Rev  A    Luminex 200 System User Manual xMAP Technology    Fluidics ies l   lA ea weRREREORIPEFGG 2 8  Optical sisted ii i hd No eR ent eR e dope ama 2 10  xMAP Technology Reagents                 0 000000 2 10 
47. rmation dialog box opens  Click    Eject Retract to eject the Luminex XYP instrument tray  Place  at least 1 2 mL of 70  isopropanol or 70  ethanol in the  reservoir  Click OK     Click Wash  In the confirmation dialog box  click on the drop   down arrow located to the right of the Eject Retract button to  choose the desired location  Click Eject Retract  The Luminex  XYP instrument tray ejects  Place distilled water in the selected  well or reservoir on the plate  Click OK  Perform this step twice     Check that the Luminex XYP instrument sample probe has been  vertically aligned for the plate used in the kit     To adjust the sample probe vertical height     Remove the clear plastic shield that covers the sample probe  area     In a 96 well microtiter plate where overall height is no more than  0 75 inches  19 mm   place the appropriate alignment tool in the  plate     For a standard plate with flat bottom wells   stack two of the  larger  5 08 mm diameter  alignment discs together and place  them into the selected well     For a filter bottom plate   stack three of the larger  5 08 mm  diameter  alignment discs together and place them into the  selected well     For a half volume plate with flat bottom wells   stack two of  the smaller  3 35 mm diameter  alignment discs together and  place them into the selected well     For a round bottom  U bottom  plate   stack two of the  smaller  3 35 mm diameter  alignment discs in the selected well     For a plate with conical wells
48. s   for each item listed at the left  fill in your initials under each date that you perform the item        Laser warmup       Check sheath fluid       Check waste level       Tighten sheath cap       Alcohol Flush  70  isopropanol or 70   ethanol        Wash twice with distilled water       Shut down Initials   for each item listed at the left  fill in your initials under each date that you perform the item        Sanitize  10  to 20  household  bleach solution        Wash twice  distilled water        Soak  distilled water        Loosen sheath cap                               Tum system off  optional                                                                                                                             Weekly   Visual inspection Date I nitials  Date I nitials  Date I nitials  Date I nitials   Monthly   Clean sample probe Date I nitials   Wipe exterior surfaces Date I nitials   Calibrate and verify Date I nitials   Every six months   Replace air intake filter  Date I nitials   analyzer   Replace air intake filter  XYP Date I nitials   Replace syringe plunger seal  or Date   nitials   syringe   Check analyzer ventilation filter Date I nitials   Yearly   Replace sheath filter Date I nitials   As required   Replace fuses Date I nitials   Comments              PN 89 00002 00 109 Rev  A       Troubleshooting the  Luminex 200  System    PN 89 00002 00 109 Rev  A    Troubleshooting    Troubleshooting procedures help users isolate  identify  and remedy  problem
49. s a  problem with your kit results  xMAP controls can help determine if  the problem is analyzer related  If Calibration and Controls are  successful  contact the kit manufacturer     Luminex 200 System User Manual xMAP Technology    A normal bead detail display is shown below  It depicts a tight bead  population within a white region     Bead Details          Figure 4 1 Normal Bead Detail    The histogram for the bead detail above looks like this           Figure 4 2 Normal Histogram          Symptom Possible problem Solution  xMAP microspheres classify You may be using Replace the calibration  too high  photobleached calibration microspheres with a fresh batch   microspheres  To avoid photobleaching   protect your microspheres from  light        4 100      2    00000000 PN 89 00002 00 109 Rev  A    xMAP Technology    xMAP microspheres hit the  lower right of the region     Beads appear scattered     UT  IU    Microspheres appear as a long  diagonal line     r  Pg    PN 89 00002 00 109 Rev  A    You may be using  photobleached xMAP  microspheres     There is air in the system     The sheath fluid is empty     The xMAP microspheres have  agglutinated     The solvent is incompatible     You are using incompatible  sheath fluid     Troubleshooting    Replace microspheres with a  fresh batch  To avoid  photobleaching  protect your  microspheres from light     Verify sample probe height  Run  three Prime commands  two  Alcohol Flush commands  and  then three washes with distilled
50. s continue  contact  Technical Support     Problems with the sample probe can lead to fluid leaks and  pressurization problems  as well as inhibit sample acquisition        Symptom    Possible problem    Solution       Sample probe leak     Sample arm is stuck in the up  position    The sample probe is clogged     The system isn   t properly  pressurized     Clean the sample probe  See  page 3 5     Check pressure settings  Ensure  that the sample probe is not  clogged and there are no leaks  in the syringe seal or syringe  valve     PN 89 00002 00 109 Rev  A    xMAP Technology    Sample arm is stuck in the  down position    Sample arm does not go down    smoothly     PN 89 00002 00 109 Rev  A    The sample probe height is too  low  or the path to the well is  blocked     The 96 well plate is incorrectly  seated in the XYP instrument   The 96 well plate is warped     The sample arm is misaligned     The sample probe is bent     Troubleshooting    DO NOT turn off the Luminex  XYP instrument     1  Remove the blue light  housing from the analyzer    2  Unscrew the Cheminert  fitting  The system monitor  changes from    Busy    to     Running    and sample  acquisition continues  Click  Cancel to make adjustments  before continuing with the  rest of the samples    3  Ifthe sample arm does not  raise  save the data that has  been collected  Turn off the  analyzer  but keep the XYP  instrument on    4  Exit the Luminex IS  Software    5  Turn the analyzer back on  and restart the 
51. s with the Luminex 200 analyzer and Luminex XYP  This  chapter does not troubleshoot problems with the PC  For help with  PC problems  please call Dell Technical Support at 800 624 9896     To troubleshoot a problem  select a general symptom  Next  identify  the possible problem and remedy it with one of the solutions listed     This chapter supplies information for the following topics       Power Supply Problems    Communication     Pressurization     Fluid Leaks   e Sample Probe     Calibration Problems     Acquisition Problems     Bead Detail Irregularities    Error Messages     Printing Errors     Verification    You can find answers to frequently asked questions  FAQs  on our  website  http   luminexcorp custhelp com     Luminex Technical Support is available to users in the U S  and  Canada by calling 1 877 785 BEAD   2323  between 7 00 a m  and  7 00 p m  Central Time  Monday through Friday  Users outside of the  U S   Canada and Europe can call us at  1 512 381 4397 between  7 00 a m  and 7 00 p m  Central Time  Monday through Friday   Inquiries may also be sent by email to support  luminexcorp com     Luminex 200 System User Manual    Power Supply    xMAP Technology    Users in Europe can call us at  31 162408333 between 8 30 and  5 30  Central European Time  Monday through Friday  Email  inquiries in Europe can be sent to supporteurope   luminexcorp com    Power supply problems often involve a blown fuse  faulty electronic       P roblems component  or even somethi
52. ser Manual    Luminex SD System    General    PC Specifications    Recommended Additional  Equipment    Uninterruptible  Power Supply   UPS     Surge Protector    xMAP Technology    e Pollution degree  UL Pollution Degree 2  as defined in Section  3 7 3 2 of UL 61010A 1    e Heater operating range  35  C to 55  C  95  F to 131  F  with  tolerance 0  C to  2  C   e Input voltage range  100 240 V    10   1 8 Amps  47 63 Hz   e AC inlet fuse  3 A  250 V   fast acting    e Ambient temperature  15  C to 30  C  59   to 86  F      Humidity  20  to 80   noncondensing   e Altitude  designed to operate at up to 2400m  7874 feet  above  mean sea level   e Physical dimensions  20 cm  8 inches  W x 30 cm  11 75 inches   D x 24 75 cm  9 75 inches  H   e Weight  9 kg  20 lbs    e UL installation category  UL Installation Category II  as defined  in Annex J of UL 61010A 1   e Pollution degree  UL Pollution Degree 2  as defined in Section  3 7 3 2 of UL 61010A 1   e Input voltage range  100 240 V    10   0 4 Amps  47 63 Hz   e AC inlet fuse  2 Amp  250 V   time lag    For systems using a PC  A Dell OptiPlex GX280 or Dell Optiplex  GX520  or newer PC  is shipped with the Luminex 200 system  For  systems using a laptop  a Dell D610 Notebook is shipped with the  system  Microsoft   Windows   XP is pre installed on the computers   The power requirements are 115 230 V   6 Amps  50 60 Hz    For updated information regarding the PC  notebook  or operating  system  go to http   www luminexcorp com  t
53. software    6  Runa Wash with distilled  water to remove air from  the system    7  Verify probe height     Adjust the 96 well plate     Inspect the 96 well plate   Replace it if it is warped     Readjust the sample arm  horizontal alignment     Remove the sample probe from  the Luminex 200 analyzer  Roll  it on a flat surface  If it does not  roll smoothly  replace it with a  new sample probe  Adjust the  sample probe height  see   page 3 2      Luminex 200 System User Manual    Calibration and  Control Problems    xMAP Technology       Symptom    Possible problem    Solution       Calibration is slow or fails     The calibration microspheres  are not fully suspended     Wrong lot number or target  values are entered in Update  CAL Targets dialog box     The system calibrators are in the  wrong well on the plate     Not enough calibrator  microspheres added to the well     Calibrator lot is expired     The sample probe height is  incorrect     The sample probe is clogged     There is a partial clog in the  system     There is air in the system     The waste container is not  vented     The waste line was moved  during system operation   resulting in unstable flow rate     Possible problem with the laser     Vortex the calibration vials to  resuspend the microspheres     Verify that correct lot number  and target values are used     Verify that calibrators are in the  correct well     Add at least five drops of  calibrator microspheres to the  well  Hold the vial upside down 
54. ssages  and    sample error messages     System error messages are displayed in three places       the Message Log on the Diagnostics Tab      the message log located in the Windows message log folder      the Errors tab on the Analysis window       Error Message    Possible problem    Solution       Unknown Diagnostic Error    XYP Instrument Heater  Stability Range Exceeded    Unknown varied    The heater block temperature is  not hitting the target value     Record the error code number  and message  Contact Luminex  Technical Support     Verify the heater temperature is  set correctly in the software     PN 89 00002 00 109 Rev  A    xMAP Technology    Low Voltage Detected    Runtime Sheath Pressure out  of Limits  Too High     Runtime Sheath Pressure Out  of Limits  Too Low     Low Laser Power Detected    Command Encountered Time  Out Error    Instrument Not Calibrated    PN 89 00002 00 109 Rev  A    Possible laser failure     The sheath fluid pressure is too  high     There is a pressurization  problem with the sheath fluid     The system is not sufficiently    warmed up     Possible Laser Failure    The current command failed     The instrument is out of  calibration     Troubleshooting    Turn off the analyzer  X YP  and  PC  then turn them back on   Make sure system is completely  warmed up  Calibrate and verify  the system  If the error message  occurs again  contact Luminex  Technical Support     Ensure that the sheath fluid  container is at the same level as  the an
55. t a sample and prevent further samples    from being collected     If the filter attached to the sheath intake line becomes clogged from  extended use  an alarm will sound even though the bulk sheath  container is not empty  If this happens  replace the filter  part number    CN 0037 01     PN 89 00002 00 109 Rev  A    xMAP Technology    Malfunction    Draining the Reservoir    PN 89 00002 00 109 Rev  A    Troubleshooting    If the alarm sounds even though the bulk sheath container has fluid  and the sheath filter is in good condition  the system is reporting a  malfunction  If this happens  contact Luminex Technical Support     If you need to ship the Luminex SD system back to Luminex  Corporation  drain the reservoir before you pack the system     1     In the IS 2 3 software  click Prime or Warmup to pressurize the  system     Leave the green air tubing connected between the Luminex 200  analyzer and the SD system     On the SD front panel  disconnect the blue tubing from the  intake labeled Sheath Out and disconnect the white tubing from  the intake labeled Sheath In     Insert the white tubing into the intake labeled Sheath Out and  insert the blue tubing into the intake labeled Sheath In     Turn the unit off and then on again   Press the Prime button on the front panel of the SD system     Sheath fluid will pump from the Sheath Delivery System  reservoir into the 20L Sheath Box     To refill the Sheath Delivery System  reconnect the tubing according  to color coding an
56. to 4 inches from the unit   See Figure 3 5     PN 89 00002 00 109 Rev  A    xMAP Technology Maintenance and Cleaning           ot d  Model No  g urnire x    S N  LX 100002450  Lumines Corey Blvd h    Figure 3 5 Grasping the Tubing    inex    4  Remove the filter with one hand  and hold the tubing with the  other hand     5  Connect a new filter to the tubing and position the filter inside  the panel     6  Reattach the panel door to the unit     7  Plug in and power on the Luminex 200 analyzer     Luminex XYP Instrument   To replace the XYP instrument air intake filter   Air Intake Filter    1  Disconnect the Luminex XYP instrument from ac power by  turning off the power switch on the rear of the Luminex XYP  instrument  then unplugging the Luminex XYP instrument power  cord from the wall source     2  Onthe back of the Luminex XYP instrument  to the left side   gently remove the screen from the Luminex XYP instrument  filter  Do not remove the screws  See Figure 3 6     PN 89 00002 00 109 Rev  A 3 7    Luminex 200 System User Manual xMAP Technology       Figure 3 6 Removing the Screen    3  Replace the filter and replace the screen  See Figure 3 7        Figure 3 7 Removing and Replacing the Filter    4  Plug in and turn on Luminex XYP instrument power     3 8 PN 89 00002 00 109 Rev  A    xMAP Technology    Syringe Seal    PN 89 00002 00 109 Rev  A    Maintenance and Cleaning        To replace the syringe plunger seal     Danger  The syringe arm does not deactivate when chan
    
Download Pdf Manuals
 
 
    
Related Search
    
Related Contents
Le compostage individuel dans le Haut-Rhin  FICHA DATOS TÉCNICOS GRASA GICE  Enox BFL-0622LED-DVD 22" Full HD Black LED TV  マシンコントローラ MP2300 基本モジュール  RiOs™ Essential 5, 8, 16, 24 Water Purification Systems  Basco 4500-48CLOR Installation Guide  Samsung 910MP Lietotāja rokasgrāmata  Fujitsu ESPRIMO P700 E85+  1. Appuyez sur le bouton [MENU]    Copyright © All rights reserved. 
   Failed to retrieve file