1. 30 dips of Ur per 25 cycles lt 5 Ur gt 95 dips of Ur per 5 seconds lt 5 Ur gt 95 dips of Ur per 0 5 cycles 40 Ur 60 dips of Ur per 5 cycles 70 Ur 30 dips of Ur per 25 cycles lt 5 Ur gt 95 dips of Ur per 5 seconds Mains power quality should be at that of a typical commercial or hospital environment If the user of the WPB requires continued operation during power mains interruptions it is recommended that WPB be powered from an uninterruptible power supply or a battery Mains power 3 A m 3 A m Mains power quality electromagnetic field should be at that of EN 61000 4 8 a typical commercial or hospital environment IACER SRL Page 19 MNPG128 01 05 05 15 Q MEDICAL DIVISION r f immunity WPB is intended for use in the electromagnetic environment specified below The customer or the user of WPB should assure that is used in such environment Immunity test Test level EN Compliance Electromagnetic 60601 1 2 level environment guidance Conducted RF 3 Veff from 3 Veff from EN 61000 4 6 150kHz to 80MHz 150kHz to 80MHz disp ER RF Radiata 3 Veff from 3 Veff from qup should be used no closer to any part of WPB including cables than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter EN 61000 4 3 80MHz to 2 5GHz 80MHz to 2 5GHz Recommended separation di
2. gt d Model WPB200 201 202 model with temperature selection with a knob Working status is showed by a Led 1 indicator e ar Main body and accessories Model WPB100 101 102 basic model without temperature selection Working status is showed by a Led indicator Model WPB200 201 202 model equipped with backlight display and keyboard with gt temperature selection Working status is showed by a Led indicator IACER SRL Page 4 MNPG128 01 05 05 15 I TEGE MEDICAL DIVISION Kit gt Hand cover 1 Paraffin wax east LA i Foot cover Main body Single use Vinyl Pack Safety plate Specifications WPB xxx has the following specifications e Class IIa equipment Directive 93 42 EEC annexed IX rule 95 Class II applied part type BF Classif CEI EN 60601 1 CE0476 certificate Equipment not protected against liquids penetration Equipment and accessories not subjected to sterilization Use of the equipment is prohibited close to flammable substances ot in environments with high concentrations of oxygen e Continuous operating mode equipment e Equipment not suited to be used in external Technical Specifications 230 VAC 50 Hz 410x245x250mm 180 W 2 3 Kg IACER SRL Page 5 MNPG128 01 05 05 15 a MEDIGAL DIVISION Symbols and labels C Paraffin Bath 0476 Model WPB 202 i Power Mains AC220 voltage 240V ad 50 60Hz Max Full 300mmHgz Po
3. 3 2 Complies establishments other than domestic establishments and those directly Voltage fluctuations Complies connected to the public low voltage flicker emissions power supply network that supplies IEC 61000 3 3 buildings used for domestic purposes Electromagnetic immunity WPB is intended for use in the electromagnetic environment specified below The customer or the user of WPB should assure that is used in such environment Immunity test Test level EN Compliance level Electromagnetic 60601 1 2 environment guidance Electrostatic 6kV contact t 6kV contact Floors should be discharge ESD 8kV air 8kV air wood concrete or EN 61000 4 2 ceramic tile If floor are covered with synthetic material the relative humidity should be at least 30 Electrical fast 2kV for power 2kV per power Mains power quality transient burst supply lines supply lines should be at that of IEC 61000 4 4 a typical commercial or hospital environment IACER SRL Page 18 MNPG128 01 05 05 15 MEDICAL DIVISION Impulses EN 61000 4 5 1kV mode differential 1kV mode differential Mains power quality should be at that of a typical commercial or hospital environment Voltage dips short interruptions and voltage variations on power supply input lines IEC 61000 4 11 lt 5 Ur gt 95 dips of Ur per 0 5 cycles 40 Ur 60 dips of Ur per 5 cycles 70 Ur
4. as shoulders or knees using a brush to form a paraffin layer Model WPB 200 201 202 1 Place the paraffin Bath on a Manopola di Step temperatura flat surface and plug in the ze 0 Step OFF p 1 Step 52 C power cable Move the LAO N 2ssp s4t ON OFF switch placed on ino a 4 Step 59 C device back side to the ON position Place 4 6 paraffin packages in the cabin and E IN close the lid Check that L L Ddistato LED is switched on Select the temperature through the knob Wait till the paraffin is completely dissolved approx 3 6 hours IACER SRL Page 9 MNPG128 01 05 05 15 I TEGN MEDICAL DIVISION 2 5 Check if the paraffin is completely dissolved then place the safety plate in the paraffin dissolver The safety plate blocks heat transmission and as a consequence the paraffin dissolution Place the hand or foot in the warm paraffin solution for 1 2 seconds then immediately remove the hand from the solution Let it dry for 4 5 seconds This activity will form a layer of paraffin on one s hand and repeating this 4 5 times will form an even and thick layer of paraffin To strengthen the heat effect and to reduce the dispersion insert the hand in a vinyl pack and later in the hand cover or foot cover C lt After 15 minutes remove the hand cover or foot e cover and the vinyl pack The treatment can be applied also on different areas as shoulders or kne
5. MEDICAL DIVISION USER MANUAL MNPG128 Rev 01 Edition 05 05 15 Paraffin bath WPB 100 200 300 m MEDICAL DIVISION I A C E R Srl Via S Pertini 24 A 30030 Martellago VE ITALY Tel 041 5401356 Fax 041 5402684 e mail iacer iacer it http www itechmedicaldivision com IACER SRL Page 2 MNPG128 01 05 05 15 TEX TRO E RUE MEDICAL DIVISION Summary Summary Thermotherapy and paraffin bath Main body and accessories Kit Specifications Technical Specifications Symbols and labels Purpose Contraindications Operating instructions Troubleshooting Safety cautions Cleaning Maintenance carriage and storage Disposal Warranty Assistance Spare parts EMC tables IACER SRL Page 3 o o OO QN a a Ui A KR OQ mM e om e i i i i Lm OO NN HN CO QC Ui uU HR 4 MNPG128 01 05 05 15 I TEGN MEDIGAL DIVISION Thermotherapy and paraffin bath Thermotherapy is used successfully in clinicians and rehabilitation fields for the treatment of chronic and traumatic pathologies Paraffin bath consists in a comfortable and easy thermotherapy device thanks to the direct application of the warm paraffin wax It is vety indicated for the treatment of chronic pathologies as the rheumatoid arthritis of hands and foot Paraffin is very efficient in pain reduction due to the high amount of heat that is released directly on the treated area Paraffin promotes blood circulation and muscle relaxation
6. ansmitters rated at a maximum output power not listed above the recommended separation distance d in metres m can be estimated using the equation applicable to the frequency of the transmitter where P is the maximum output power rating of the transmitter in Watts W according to the transmitter manufacturer Note 1 At 80 MHz and 800 MHz the separation distance for the higher frequency range applies 2 These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and reflection from structures objects and people WPB Series All rights reserved Logos menieat nivisionis owned by LA C E R Srl and is registered IACER SRL MNPG128 01 05 05 15 Page 21 MEDICAL DIVISION MEDICAL DIVISION I TECH MEDICAL DIVISION I A C E R S r l Sede operativa 30030 Martellago VE Via S Pertini 24 A Tel 39 041 5401356 Fax 39 041 5402684 Sede legale S Marco 2757 30124 Venezia Cod Fisc P VA IT 00185480274 R E A VEN 120250 M VE001767 Cap Soc 110 000 00 i v www iacer it iacer iacer it
7. e when it is damaged or when the power cable is damaged Before using the Paraffin Bath check the temperature through a finger Do not apply paraffin on skin with lotion wax additives or other substances Do not overfill the Paraffin Bath with wax Please remove the vinyl pack of the paraffin wax when using and do not heat the wax at a microwave or stove Please use wax supplied by the fabricant IACER SRL Page 14 MNPG128 01 05 05 15 MEDICAL DIVISI In case of rash sore swollen skin or other side effect occur immediately stop the application and consult a doctor The fabricant is responsible for the device safety and performance only in the following events e Any changes repair actions and additions are carried out by authorized personnel e The electrical system used for the device working is in compliance with the current regulations e The device is used in compliance with the instructions of the user manual Cleaning Clean the device and its accessories using a soft cloth Resistant stains can be removed using a sponge soaked in solution of water and mild soap Wait for the complete drying before using the device and its accessories Do not use chemical substances solvents ot cleanets to avoid device and accessories damage Maintenance cattiage and storage If it is used in accordance with the instructions of the user manual the device does not need a particular regular maintenance Carriage preca
8. es using a brush to form a paraffin layer La regolazione di temperatura permette di ridurre i tempi di dissolvimento della paraffina da 56 a 59 C Modello WPB 300 301 302 1 Place the paraffin Bath on a flat surface and plug in the power cable Move the ON OFF or switch placed on device back ys side to the ON position Select the manual setup or the pre adjusted start see the following paragraphs and press START key Place 4 6 paraffin packages in the cabin and close the lid Check that LED is switched on Wait till the paraffin is completely dissolved approx 3 6 hours Tasto SET Display LCD Tasto MENU L Tasto START STOP IACER SRL Page 10 MNPG128 01 05 05 15 Q TECK 2 Check if the paraffin is completely dissolved then place the safety plate in the paraffin dissolver The safety plate blocks heat transmission and as a consequence the paraffin dissolution 3 Place the hand or foot in the warm paraffin solution for 1 2 seconds then immediately remove the hand from the solution Let it dry for 4 5 seconds This activity will form a layer of paraffin on one s hand and repeating this 4 5 times will form an even and thick layer of paraffin 4 To strengthen the heat effect and to reduce the dispersion insert the hand in a vinyl pack and later Doe in the hand cover or foot cover 5 After 15 minutes remove the hand cover or foot C cover and the vinyl pack The treatment can be applied also on dif
9. ferent areas as shoulders or knees using a brush to form a paraffin layer Temperature selection from 56 to 59 allows paraffin dissolution time to be reduced Menu setup for WPB 300 301 302 Set time e Keeping pressed O key for 5 seconds Press and O keys to select the Day Press Ma to confirm and to move to the next field Press and O keys to select the hour Press Mm key to confirm and to move to the next field IACER SRL Page 11 MNPG128 01 05 05 15 I TECH MEDICAL DIVISION Press and keys to select the minutes Press a START STOP to return to the main menu Manual setup e Keeping pressed Mi for 5 seconds When the temperatures flashing on the display press O and key to select the temperature 48 59 C Press n START STOP to statt the program Press m START STOP to stop the treatment IACER SRL Page 12 MNPG128 01 05 05 15 I TECH MEDICAL DIVISION Pre adjusted start In WPB 300 model the treatment start can be pre adjusted With this function you can set the day and the time of treatment start and also the temperature Press Mm key Press and O keys to move the cursor to the desired day Press M key Press and O keys to move the cursor _ to the desired day and select the program For example o Mon if _ is displayed the day is memorized and the machine works properly o Mon if _ disappears the day is n
10. ity to heat temperature Stop the treatment if dermatitis occur after paraffin use Stop the treatment 1f paraffin wax is too heat or too cold it could indicate patient health problems Keep it out of children reach and of persons with cognitive problems or physical disabilities Operating instructions Model WPB 100 101 102 1 Place the paraffin Bath on a flat surface and plug in the power cable Move the ON OFF switch placed on device back side to the ON position Place 4 6 paraffin packages in the cabin and close the lid Check that LED is switched on Wait till the paraffin is completely dissolved approx 3 6 hours 2 Check if the paraffin is completely dissolved then place the safety plate in the paraffin IACER SRL Page 8 MNI I TECH MEDICAL DIVISION dissolver The safety plate blocks heat transmission and as a consequence the paraffin dissolution 3 Place the hand or foot in the warm paraffin solution for 1 2 seconds then immediately remove the hand from the solution Let it dry for 4 5 seconds This activity will form a layer of paraffin on one s hand and repeating this 4 5 times will form an even and thick layer of paraffin 4 To strengthen the heat effect and to reduce the dispersion insert the hand in a vinyl pack and later in the hand cover or foot cover 5 After 15 minutes remove the hand cover or foot cover and the vinyl pack C The treatment can be applied also on different areas e
11. ot memorized e After the day memorization press key to move to TIME setup e To select START and STOP time press M key to use Hour Minute functions and lt 9 and O set the values e After setting the STOP time pressing to select the temperature by using and keys to set the value e Repeat steps 1 7 to set Day Start time Stop time Temperature for each day separately Press START STOP to save the values IACER SRL Page 13 MNPG128 01 05 05 15 MEDICAL DIVISION e Once the memorization function is activated the screen displays Res Day for 10 seconds and the Start time and Stop time are displayed continuously Troubleshooting Problem Solution WPB doesn t switch e Check the connection to the domestic plug on Check the ON OFF button e Check the power supply cable status Paraffin wax doesn t e Ifthe device is equipped with temperature dissolve selection function increase the temperature up to 56 and later decrease it gradually Safety cautions The following cautions guarantee user safety and prevent lesions or damages Read carefully the user manual before using the device Consult the doctor to receive prescription and instruction before using the product Paraffin wax must be replaced after use and the product must be cleaned before and after each use Do not drop or throw the product Check the rated voltage before using the device Do not use the devic
12. stance d 1 2 AP 150kHz to 80MHz d 1 2 P 80 MHz to 800 MHz d 2 3 P 800 MHz to 2 5 GHz where P is the maximum output power rating of the transmitter in Watts W according to the transmitter manufacturer and d is the recommended separation distance in metres m Field strangths from fixed RF transmitters are determined by an electromagnetic site survey should be less than the complicance level in each frequency rage Interference may occur in the vicinity of equipment marked with the o following symbol A Recommended separation distances between portable and mobile communications equipment and the WPB WPB is intended for the use in an electromagnetic environment in which radiated RF disturbances are controlled The customer or the user of WPB can help prevent electromagnetic interferences by maintaining a minimum distance between portable and mobile RF communications equipment transmitters and WPB as recommended below according to the maximum output power of the communication equipment Rated maximum Separation distance according to the frequency of the power of the transmitter m transmitter W 150kHz to 80MHz 80MHz to 800MHz 800MHz to d 1 2 AP dz 1 2 AP 2GHz d 2 3 P 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 10 3 8 3 8 7 3 IACER SRL Page 20 MNPG128 01 05 05 15 l 9 MEDICAL DIVISION 100 12 12 23 For tr
13. tion on device must be performed by manufacturer For any assistance intervention contact the national distributor or I A C E R S r l Via S Pertini 24 a 30030 Martellago VE ITALY Tel 041 5401356 Fax 041 5402684 iacer iacer it www itechmedicaldivision com You can get any technical documentation on spare parts but only prior business authorization Spare parts Contact the distributor or the fabricant for original spare parts at following address I A C E R S r l Via S Pertini 24 a 30030 Martellago VE Tel 041 5401356 Fax 041 5402684 iacer Q iacer it www itechmedicaldivision com To preserve product warranty functionality and product safety we recommend to use only original spare parts IACER SRL Page 17 MNPG128 01 05 05 15 Q MEDICAL DIVISION EMC tables Electromagnetic emission Emission test Compliance Electromagnetic environment guidance Group 1 WPB uses RF energy only for its RE emissions internal function Therefore its RF Cispr 11 emissions are very low and are not likely to cause any interference in nearby electronic equipment Class B WPB is suitable for use in all RF emissions establishments including domestic Cispr 11 establishments and those directly connected to the public low voltage power supply network that supplies buildings used for domestic purposes Harmonic emissions Class A The WPB is suitable for use in all IEC 61000
14. utions It doesn t need any particular carriage precautions however we recommend to put away the device and its accessories in their own box after every treatment Switch off the device after each treatment Do not dispose the melted paraffin solution disposing the liquid paraffin in the toilet drain bath tub etc will plug the pipe as it solidifies Wait until the solution solidifies and then dispose it with regular waste IACER SRL Page 15 MNPG128 01 05 05 15 MEDIGAL DIVISION If wax happens to fall on the floor it may cause one to slip Please clean the floor after it occurs When removing the paraffin do not use sharp tools Storage precautions The device is protected till following environmental conditions Outside of the packaging Temperature from 5 to 40 C Humidity from 10 to 93 Pressure from 700 to 1060 hPa Inside of the packaging Temperature from 5 to 40 C Humidity from 10 to 93 Pressure from 700 to 1060 hPa Disposal The equipment is subjected to WEEE regulations see the symbol on the label concerning separate waste collection when disposing this product please use the designed areas for disposing electronic waste or contact the manufacturer Warranty Make reference to the national laws for any warranty conditions by contacting the national distributor or directly the manufacturer IACER IACER SRL Page 16 MNPG128 01 05 05 15 MEDICAL DIVISION Assistance Every interven
15. wer 180W pent 1096 e SN OOOO TA va 4ImIMXx l A C E R sri I TEGE DIVISION m Via Sandro Pertini 24 A i 30030 Martellago VE Italy Warning Read annex documents Product subject to WEEE regulations concerning separate waste collection of electronic equipment LX Applied part type BF Class II equipment IACER SRL Page 6 MNPG128 01 05 05 15 MEDICAL DIVISION C Curs Compliance with Directive 93 42 EEC MDD Manufacturing date Follow the instruction OL Operating instructions 5 N IP Device with protection degree IP21 i Keep out of the rain Humidity range for storage Temperature range for storage Do not pushing the device Do not sitting on the device eee Do not stepping on the device IACER SRL Page 7 MNPG128 01 05 05 15 I TEGO MEDIGAL DIVISION Purpose Clinical purpose Therapeutic Use Clinic Hospital and domestic use Chronic and acute pathologies treatment in particular for joints of hand and foot for example rheumatoid arthritis Thermotherapy for pain reduction Contraindications Do not use on open wound or cut on red skin in presence of neoplasias blood circulation problems acute inflammations or in the case of missing reduced sensitivity for example in diabetic patients Do not use in presence of bleeding or abnormal sensitiv