USER MANUAL - Spectrum Analyzer
Contents
1. INSTRUMENTS USER MANUAL AT437C POWER METER USER MANUAL ATTEN Electronics Co Ltd ATTEN INSTRUMENTS USER MANUAL AT437C POWER METER Contents General Information Specifications Users Methods Logic Diagram Main panel Front View Troubleshooting Equipment maintenance oO woo NN d WW Inn Warranty 2 2 2 on nnn nnn nnn nnn nnn nnn enn nennen nnnm eren nee Warning A ATTENTION refer to manual A Danger High voltage j 7 Protective ground earth terminal Input Power max 100mW of the AT11801 Power Sensor probe ATTEN Electronics Co Ltd AT TEN INSTRUMENTS USER MANUAL AT437C POWER METE
2. Press Enter key to confirm calibration operation of the equipment then REM light on the state has entered calibration Calibration operational needs sustained seconds and correct operation after the completion of the state apparatus to measure REM lights eliminate If the calibration of the instrument in the key state of waiting no follow up operation for a few seconds after the return of measuring state apparatus By cal button the instrument automatically open 50 MHz reference source to avoid apparatus for invalid operations To allow accurate calibration operation the proposed corrective action before entering the benchmark bond by the reference source and preheat to open more than 15 min ZERO CAL during REM lights must wait for the follow up only after the eradication operation At ZERO References source should be closed to ensure that no power input sensor AT TEN INSTRUMENTS USER MANUAL AT437C POWER METER 3 2 4 STORE and RECALL Instrument retained the 10 registers 0 9 for the users up to 10 user defined set of measurements and data preservation and when necessary to facilitate the transfer and use it every time before a measurement Series set up operations Deposited by Bond displayed on the display window SAVE N N 0 9 then wait for the machines in the state By selecting the A button plus 1 or W key minus 1 to select N the press the enter button the current measurement equipment will b
3. R GENERAL INFORMATION Introduction AT437C is a cost high performance microwave power meter for microwave power measurement It can measure the frequency range from 10MHz to 18GHz AT11802 It can measure power levels 0 1uW 100mW It s a portable design keeping in mind the bench and field service application AT437C power indicator from its power sensors and components The high performance and accuracy is obtained by the Power Sensor by chopper and by the linear correction circuit components Chopper is using high sensitivity low noise MOS FET switch components Series chopper Its main role is to transform the probe by the power of DC input signal modulation signal into equivalent exchanges and through the CPU to carry out the processing Microwave power meter have preamplifier as the high gain transistor and the high impedance Low barrier comprising operational amplifier with a gain amplifier exchanges about 600 hybrid amplifiers SYSTEM SPECIFICATIONS Frequency Range 10MHz 18 GHz Power Range 0 1uW 100mW Readout accuracy DATA 8 VSWR 1 25 Power Reference 50MHz 1 00mW 1 596 Display Resolution 0 01 dB Recorder output 0 1Voltage output Impedance 1KO Display W dBm dB Line Voltage Range 220VAC 10 50Hz 60Hz 10W Size 260mm L x250mm WD x110mm H Weight about 3 7 Kg ATTEN Electronics Co Ltd AT TEN INSTRUMENTS USER MANUAL AT437C POWER METER USER METHODS 3 1 NOTICE
4. e Power Voltage The Input Power supply should not be more than 220 VAC 10 otherwise no guarantee of equipment on burnt equipment due to overload e Operating Environment The specifications stating tolerances are only valid if the instrument has warmed up for 30 minutes at ambient temperature between 15 C 59 F and 30 C 86 F Values without tolerances are typical for an average instrument e Before Zeroing The sensor should access ports SENSOR set sources off to ZERO operate then open 1 mW 50 MHz reference source for the calibration e At exactitude measure it should to process frequency settings 3 2 OPERATE METHOD 3 2 1 MODE key There are three ways in the Power Meter for the power measurement W dBm and GB When GB to be shown is the first measurement Xo absolute power as a reference on the number of forms 10Lg Xn Xo by entering key to refresh reference value 0 00 dB By MODE Key the button switch means dBm dB display By ENTER Key the button switch means dBm Watt display 3 2 2 SET key The Setting key is for microwave frequencies measured in order to correctly reflect the process of measuring the measured object situation a Frequency settings This is done by setting the display window button to show FT XX the access frequency settings Frequency setting range 0 to 18 0 said that 50 MHz while other figures GHz as a unit By pressing the A button plus 1 and V key minus 1
5. e installed and kept in the state N and stored for the future to use can be recalled by adopting the CALL N its transfer When displayed on the display window SAVE N in a wait state apparatus such as follow up operation without waiting for a few seconds after returning to withdraw from the storage operations measurement state By transfer button to press RECALL the display window displayed on the CALL N N 0 9 then wait for the machines in the state By selecting the A button plus 1 or V key minus 1 to select N If by then enter button send equipment will be kept in storage in the N measurement and data transfer as a set of measurements and the current settings When displayed on the display window CALL N apparatus in a wait state such as follow up operation without waiting for a few seconds after the withdrawal operation out of state and return to the measurement 3 2 5 ENTER key Enter key is a mixed function keys In the frequency or attenuation settings to set confirm a role in the calibration or zero to confirm operation in the save or recall also to confirm operation in the power measurement it played the implementation of Watt dBm display mode switch When dB measured in the way also played into the refresh button reference value 0 00 dB role NOTE zero before switching to enter the bond Watt showed that only 0 0 X 3 2 6 SYSTEM key Implementation of the remote control system function ke
6. lating power base at normal P temperature should be raised to 1 mW ultra poor 3 1 5 EQUIPMENT MAINTENANCE Equipment should be regularly checked and found that fault timely repair Equipment should not increase canopies the annex should be properly kept Equipment operation keys should not be pressed with excessive force so as to avoid damage ATTEN Electronics Co Ltd AT TEN INSTRUMENTS USER MANUAL AT437C POWER METER WARRANTY All ATTEN products are warranted against defects in materials and workmanship for two years from the date of shipment ATTEN s obligation covers repairing or replacing products which prove to be defective during the warranty period Obligation is limited to the original purchaser ATTEN is not liable for consequential damages SHENZHEN ATTEN ELECTRONICS CO LTD Building A29 Tanglang Industrial Zone Xili Nanshan Shenzhen 518055 P R China http www atten com cn http www attenelectronics com Emil atten atten com cn TEL 86 755 86021376 FAX 86 755 86021347 ATTEN Electronics Co Ltd
7. to amend and adjust to the necessary requirements If the button settings and no follow up operation the machine will wait a few seconds later returned to the measurement state when Zero and cal Tuning frequency should be set to zero ATTEN Electronics Co Ltd AT TEN INSTRUMENTS USER MANUAL AT437C POWER METER b Attenuation settings Continuous press set key by the display window display At XX enter the attenuation settings Settings range 0 to 30 dB By pressing the A button plus 1 and W key minus 1 to amend and adjust to the necessary requirements in order to achieve power compensation 3 2 3 ZERO and CAL Key NOTE ATTEN Electronics Co Ltd In the process of measuring equipment to ensure the accuracy of data often need to tune the instrument zero and calibration operation Press ZERO key ZERO displayed on the display window then wait for the machines in the state and then press enter key to confirm the transfer equipment operation REM light at this time on that has entered the zero state Need for continued operation of zero seconds after the completion of transfer and operation of the state apparatus to measure REM lights goes OFF If the transfer is the key instrument in a wait state wait for a few seconds after the follow up operation to return to measuring the state apparatus Press CAL key CAL displayed on the display window CAL At this time in a wait state apparatus
8. y role Remote Control status through this button to go back to local If no RS232 interface configuration options the system keys are vacant LOGIC DIAGRAM ATTEN Electronics Co Ltd ATTEN INSTRUMENTS USER MANUAL AT437C POWER METER RF Input Power Preamoliar Attenuation Amplifier Sensor P Network Networks Zero network Control Waveform Drive RF Output P NMHz power Reference 15VDC 15VDC H12VDC 5VDC Power Source Control Power Switch on off DC Power Eee MAIN PANEL FRONT VIEW ATTEN AT437C POWER METER POWER REF MESE O LILILI we SENSOR dBm W mone A store ZERO ENTER er scs iei on ma ne ATTEN Electronics Co Ltd AT TEN INSTRUMENTS USER MANUAL AT437C POWER METER TROUBLESHOOTING SN Fault display Fault may be part Removal Method Broken fuse Replacement of fuses 1 Not boot Check input power PCBA and Power PCBA failure replacement of defective components LED Broken LED replacement Monitor not normal 2 LED display not shine Power cable with a Good sensor inserted connecting cables sensor panel signal input or replacement probe port bad contact Reduce the power source of power to 3 Display Over Data Overflow appropriate Reference vibration stops Replacement Of defective reference No reference output pti 4 power Power errors Eef renh 6 cower Re regu
Download Pdf Manuals
Related Search
Related Contents
Graco 309643B User's Manual Leia a íntegra do voto - Tribunal de Contas do Estado de São Paulo PTE13インチ JVC GR-AXM33 Camcorder User Manual cedoC-Tech.fr English AARIA MULTI Tri-Rate Serial Digital Interface Physical Layer IP Core User's Guide 農業技術の基本指針(平成25年改定)(全文)(PDF Copyright © All rights reserved.