1. Time Electronics 1017 DC Multifunction Voltage Current Resistance Calibrator Technical Manual MUL TIFUWCTION ALIBRATOR a 10 nte 7077 Time Electronics V1 2 01 11 10 Time Electronics Ltd Botany Industrial Estate Tonbridge Kent TN9 1RH Tel 44 0 1732 355993 Fax 44 0 1732 770312 Email mail TimeElectronics co uk Web Site www TimeElectronics com 2 Contents Led ela 1 3 General DescripHon oe eI oe retten uude 1 2 Specificatl ng oen edi eroi ea eia tee ita a ded SGIRCUTE Descriptio inte nece eee 1 4 Power Supply Description 5 ecran rhe nnns baa etuer nnmnnn eua 2 Front Panel Controls spere prec pi qure naniii aaia Bi TOME TATION E UU D UU UU UU IITIIILLITLCTT S i Preparing for Use eee eap etie an e n i eee create 3 2 Using the 1017 eii eiu eb d be ende cetus 3 3 Thermal EME S A oI hec tede 3 4 Using the Deviation Control eeeeeeeeeeeeeeeeeeereee eene nennen nnns 4 SG ALTA ELON asse ena aes aes ears RE Rae Ra a a eR Ge UR ERN EY 41 Calibrating the 1017 Ee cea tee cia eie 4 2 Zero Adjustment aapa aai simu eaten adus ER te 4 3 Calibrating the Full Scales 11 cierre reiecit recen eer en rane nin rne reru n uere 4 4 Calibrating the Deviation Control eeeeeeeeeeeeeeeeeennne nnns 5 Guarantee amp Servicilig sies reda rra RA EA YEAR EREN TER AERE EA A ERE RE CAR ERR REA EA XAR SUV 10
2. 1017 Technical Manual 7 4 Calibration 4 1 Calibrating the 1017 The 1017 can be calibrated with a high accuracy D M M with D C voltage current and resistance ranges with a specification of at least 4 times greater than that of the 1017 s specifications Calibration is best carried out on fully charged batteries without the mains supply connected to ensure that no mains supply interference takes any effect on the readings VR9 DEVIATION ADJ VR7 100mA FS ADJ VR8 DEVIATION ADJ VR5 a TOV FS ADJ VR6 100V FS ADJ ET d 100mV FS ADJ qas j and 10mV 4 2 Zero Adjustment Range Allowance Adjust Pot 1V 6uV 1V ZERO ON MODULE 1017 Technical Manual 8 4 3 Calibrating the Full Scales After checking the zero settings you may then calibrate the full scales Set the digit switch to 999999 Adjust the 1V Full Scale with the CAL trimmer which is on the potted module Adjust this to read 1V to within 20uV Adjust the 10V F S trimmer VR5 to read 10V to within 200uV Adjust the 100mV amp 10mV F S trimmer VR3 to read 100mV to within 4 3u V Done on 100mV Range Check 10mV F S is within 1 3yV Adjust the 100V F S trimmer VR6 to read 100V to within 3 5mV Adjust the 100mA F S trimmer VR7 to read 100mA to within 6uA 4 4 Calibrating the Deviation Control Set the output to the 1V on the 1V range and set the deviation to 0 99996 Adjust the outp
3. All Time Electronics instruments are subject to continuous development and improvement and in consequence may incorporate minor detail changes from the information contained herein 1017 Technical Manual 2 1 Introduction MULTIFUNCTION CALIBRATOR wean 1017 NT UUUUEXLI 1 1 General Description The 1017 is a high performance portable calibrator for use in the field or laboratory With voltage ranges from 10nV to 100V current from 100nA to 100mA and resistance from 10mO to 10kQ it makes the 1017 an extremely versatile instrument capable of calibration or simulation The digital deviation control allows the output to either increase or decrease directly in a percentage from 0 001 to 0 999 This is a useful feature for checking the accuracy of difficult scales Long term stability is guaranteed by the use of high quality components and the latest in resistor technology The 1017 s basic reference source is a specially aged computer selected precision reference diode The diode provides the input to the instruments active circuitry which incorporates several high performance chopper stabilised amplifiers with a zero stability of less than 0 5uV per C and a warm up time of less than 1 minute The precision resistors used in the 1017 are high performance types with temperature ratio matching of less than 2ppm per C The output terminals are low thermal E M F types typical E M F of less than 0 2uV per C The
4. battery is fully charged This means that it is impossible to over charge the battery Approximately 30 40 hours use may be obtained between charges depending on the output setting 1017 Technical Manual 5 2 Front Panel Controls MULTI FUNCTION CALIBRATOR 1017 TIME ELECTRONICS 1 Output Terminals 2 Case Terminal 3 Reverse Off Normal Switch Selects the output polarity 4 Power On Off Switch Indication is shown by the Battery Level Indicator 5 Range Select Switch Selects the output maximum output 6 Range Output Digit Switch Selects the value of output 7 Deviation Digit Switch Selects the value of Deviation 8 Battery Level Indicator Shows the state of charge of the batteries 9 Positive Off Negative Deviation Control Applies a Positive or Negative Deviation to the output as a percentage 1017 Technical Manual 6 3 Operation 3 1 Preparing for Use Before using the 1017 you must first check to see if the battery level indicator shows that the batteries are in a good state of charge If the indicator shows that they are low then you must recharge them by plugging the 1017 into the mains supply It will take approximately 12 14 hours to fully recharge the battery pack You may of course use the 1017 when it is plugged into the mains supply irrespective of the battery state The batteries should also be checked during operation 3 2 Using the 1017 F
5. 1017 can be powered from either the mains or the internal rechargeable battery pack The batteries are automatically recharged when the mains supply is connected A front panel battery level indicator displays the batteries state of charge Output polarity can be reversed via a switch on the front panel 1017 Technical Manual 3 1 2 Specifications Outputs Voltage Range Accuracy Output Resistance Output Current All specification 3uV Current Range Accuracy Drive Voltage Resistance Range Accuracy Power Rating End Resistance Deviation Control 0 100V in 5 Ranges 0 100mA in 1 Range 0 10kQ in 1 Range 0 9 99999mvV in 10nV steps 0 99 9999mvV in 100nV steps 0 999 999mvV in 1uV steps 0 9 99999V in 10uV steps 0 99 9999V in 100uV steps 10mV range 100mV range 1V range 10V range 100V range 100mV to 10V 10mV and 100V 100mV to 10V 100V 10mV 0 99 9999mA 100mA range 10V Maximum 0 02 of setting 0 005 of range 0 01 of setting 0 004 of range 0 005 of setting 0 002 of range 0 005 of setting 0 002 of range 0 01 of setting 0 004 of range 200mQ 10 150mA 10mA Limited by 1O Output resistance in 100nA steps 0 0296 of setting and 0 00496 of range 0 9 99999kOQ in 10mQ steps 0 05 of full scale 250mW per Resistor lt 200mO Only on Voltage and Current Range 0 001 to 0 999 in 0 001 steps Accuracy 0
6. 596 Accuracy Stability Less than 10ppm C on all ranges Less than 5ppm Day at constant temperature Less than 15ppm 90 days Less than 25ppm Year Output Noise Output Noise 0 1Hz to 10Hz Typically 3ppm Power Supply The 1017 can be used continuously from a 220V 240V 50 60Hz 110V 120V to order mains supply or from the internal rechargeable Ni Cad battery pack A front panel meter indicates the state of charge at all times 1017 Technical Manual 4 Dimensions amp Weight Dimensions 290mm x 250mm x 110mm Weight 2 4Kg Operating Temperature 0 C to 50 C max For optimum performance and stability the 1017 should be stored and used in temperatures between 15 C and 25 C 1 3 Circuit Description The calibrator employs a temperature compensated zener diode as the basic reference source This provides the input to an FET chopper amplifier system which operates in a feed back stabilised mode and has a gain value determined by a set of precision wire wound resistors which are selected by a 6 decade thumbwheel switch The output voltage is variable from OV to 99 9999V in 6 ranges An output resistance of typically 150mQ is maintained on all ranges excluding the 100V range where the output resistance is 10 The maximum output current that can be drawn on the 10V and 1V ranges is automatically limited to approximately 150mA or 15mA on the 100mV range It must be noted that the current on the 10mV range is l
7. al of your old appliance please contact Rud your city office waste disposal service or return to Time Electronics 1017 Technical Manual 10
8. imited by a 10 resistance The output current on the 100V range is 10mA The 100V output is generated by driving a step up transformer with the chopped D C output from the output of the module Feedback around the complete amplifier regulates the output to the set output value When the 100mA range is selected the output current is sensed by a precision resistor The voltage generated across this resistor is used to control the output drive voltage so regulating the output current The resistance range operates by directly switching the thumbwheel switch to the output terminals The deviation control operates by injecting a current into the reference circuitry which either increases or decreases the reference voltage To ensure complete reliability of the thumbwheel switch double pole gold plated contacts are used for each position even if a contact fails the 1017 will continue to work correctly 1 4 Power Supply Description The power supply for this unit is via a Time Electronics PU2 power supply This unit provides a battery power supply for portable use which is automatically charged when mains operation is used This is mounted on the rear panel of the 1017 The power supply is set for 220V 240V AC mains unless otherwise specified When the mains is connected the charging circuitry provides the correct charge current 40 45mA for the battery and automatically reduces this to at trickle charge 3 4mA when the
9. irstly select the correct range on the range select switch ensuring that the output polarity switch is OFF and that nothing is dialled up on the digit switch Then dial up the required value on the digit switch The output may then be supplied to the unit under test by selecting the output polarity switch Note When the output switch is set to the OFF position there is a short circuit on the output terminals Caution must be exercised when using the 100V range against possible electrical shock 3 3 Thermal EMF s When using the 1017 to provide a precision voltage of less than 1mV care must be taken to avoid Thermal EMF s These occur where temperature differences are present at the junctions of dissimilar metals e g A normal solder to copper junction has a thermal EMF of approximately 3uV C Errors in the 1017 under stable temperature conditions are typically less than 0 2uV C 3 4 Using the Deviation Control Note The digital deviation control has no effect if used on the resistance range To apply a percentage deviation to the output select either positive or negative deviation with the control switch and dial up the percentage deviation required on the digit switch Alternatively if the output is set to the correct value and the reading on the U U T reading differs from this adjust the deviation control to give the required value The error can then be read directly as a percentage from the deviation control
10. red Time Electronics Ltd can still service your instrument As the manufacturer we have the specialised knowledge needed to keep your instrument in peak condition and we also maintain a comprehensive spare parts service Please enclose details of the service required and your full company details including a contact name when returning for servicing Returning Instruments When returning instruments please ensure that they have been adequately packed preferably in the original packing supplied Time Electronics Ltd will not accept responsibility for units returned damaged Please ensure that all units have details of the service required and all relevant paperwork Send the instrument shipping charges paid to Time Electronics Ltd Botany Industrial Estate Tonbridge Kent TN9 1RH Tel 44 0 1732 355993 Fax 44 0 1732 770312 Email mail TimeElectronics co uk Web Site www TimeElectronics com Disposal of your old equipment 1 When this crossed out wheeled bin symbol is attached to a product it means the product is covered by the European Directive 2002 96 EC 2 All electrical and electronic products should be disposed of separately from the municipal waste stream via designated collection facilities appointed by the government or the local authorities 3 The correct disposal of your old appliance will help prevent potential negative consequences for the environment and human health 4 For more detailed information about dispos
11. ut using the Deviation trimmer VRQ to read 1 01V to within 5mV Leaving the output at 1V on the 1V range set the deviation to 0 99996 Adjust the output using the Deviation trimmer VR8 to read 0 99V to within 5mV The Calibration procedure is now complete 1017 Technical Manual 9 10 5 Guarantee amp Servicing Guarantee Period This unit is guaranteed against defects in materials and workmanship for a period of one year from its delivery to the customer We maintain comprehensive after sales facilities and the unit can if necessary be returned to us for servicing During this period Time Electronics Ltd will at its discretion repair or replace the defective items For servicing under guarantee the instrument type and serial number must always be quoted together with details of any fault and the service required The purchaser of the instrument must prepay all shipping charges Time Electronics Ltd will pay return shipping charges This guarantee is void if servicing has been attempted by an unauthorised person or agent If during the guarantee period failure is due to misuse or abuse of the unit the repair will be put in hand without delay and charged unless other instructions are received Please note that if you require a new UKAS Certificate during the warranty period this will be charged at the current rate on our price list Service After Guarantee Period Even after the guarantee period has expi