Air Plus User Manual
Contents
1. The Mattress can now be moved independently and will remain inflated for several hours It should be reconnected with the Power Unit and mains electrical supply as soon as possible Warning The Mattress will not alternate when disconnected from the Power Unit and or the mains electrical supply and hence will not provide optimal pressure relief 4 1 Side Formers The side formers normally remain fully inflated and providing a firm edge to the Mattress which creates a sense of security for the patient and assists when transferring from the bed to a chair However for some patients it is beneficial to be able to deflate the formers during transfer to from a chair Deflation is achieved by depressing the actuator button on the single quick connect valve located in the umbilical hose approximately 15 cms from the Power Unit Both the Formers and the Head Cell are deflated while the remaining transverse Cells remain inflated Note For security the Static mode should be selected prior to moving a patient on or off the Mattress Once the patient transfer has been completed the valve should be reconnected in order to re inflate the Formers and Head Cell The Low Pressure alarm may become activated and should be cancelled by switching off the Power Unit using the on off switch and immediately switching it back on ISSUE 2 PAGE 3 OF 6 AIR PLUS USER MANUAL 5 Alarms Alarm conditions are indicated by a fl
2. AIR PLUS USER MANUAL Air Plus User Manual OJ arie Form Alternating pressure mattress for prophylaxis and treatment of pressure sores Indications for use e The Air Plus has been developed as a state of the art pressure relieving system for high to very high risk patients Contra indications e As with all alternating mattress systems the Air Plus is contra indicated for use where patients have unstable fractures or spinal injuries CONTENTS 1 INTRODUCTION isssssiisiinsusanasssnanaunsnanannansusnannauwavaaunneuinsaunanauasuaunasanadaansanaian aaas vanauaan naana 2 Ze ANS TALE DION E A A E A A A 2 Be OPERATION sri ssistsscsactcssepasticnccawsinsayacttescnascnwsavastansd cousiusipasanehnanspautnasdsawenadapasttendsadinusavanandsuaenin 2 4 TRANSPORTATION so ssncstvestnscnaicscstviesistsanntenestseucsautsnienercveadsneueinnncaatidsenindenataaatserincianaienaenns 3 Si ALARM Le E 4 6 MAINTENANCE PROCEDURES iieitinicnssnnsesenueanitencnnnnsesveaninnsinniusasseamatanbuinsasuesuunumdnniensnncssn 4 7 TECHNICAL DATA icesinssnnnatnannssnincinenninannrsnsannnninusncsndicenasunhuanensntcnmcnniinnteiinsanudssulucinsedseanessnie 5 8 SUPPLIER ADDRESS ieasiisiestsisissracinsassainvensusinsteveiicinusnstsucenerinesasndiaudenerneeaurecinudeniaaenusetiinnueneny 6 ISSUE 2 PAGE 1 OF 6 AIR PLUS USER MANUAL 1 2 Introduction The product consists of an air filled mattress that rests on top of the bed p
3. The CPR Valve consists of a manually operated rotary action valve which when opened connects all the Cells and Formers to atmosphere and so achieving rapid Mattress deflation For convenience and accessibility the Valve is mounted in the umbilical hose between the Mattress and the Power Unit Turning the red operating knob clockwise to the CPR position activates the Valve As the Mattress deflates the Low Pressure alarm is activated and should be cancelled by switching the Power Unit off Warning Close the CPR Valve by rotating the red knob anti clockwise to the Normal position before switching the Power Unit on again 4 Transportation To change the location of the bed switch the system to Static and allow all the Cells to fully inflate Switch off the Power Unit using the on off switch and disconnect the electrical supply cable from the mains socket The bed can now be moved to a new location where it must immediately be reconnected to the mains electrical supply and the Power Unit switched back on Once the Mattress has been refilled the Dynamic mode will automatically be selected To move just the patient and mattress e g by trolley switch the system to Static and allow all the Cells to fully inflate Switch off the Power Unit using the on off switch and disconnect the umbilical hose by depressing the actuator buttons on the 3 quick connect valves on the side of the Power Unit and withdrawing the probes
4. ashing red display accompanied by an audible warning In each case the user should respond by turning the Power Unit s switch off and investigating the cause 5 1 High Pressure Alarm This condition could be caused for example by a kinked umbilical hose or visitors and others sitting suddenly on the Mattress The high pressure alarm condition is shown by 0 0 0 flashing red plus an audible alarm 5 2 Low Pressure Alarm This condition could be caused for example by opening the CPR Valve or a leak in the Mattress due to a cut or puncture The low pressure alarm condition is shown by _ __ flashing red plus an audible alarm 5 3 Valve Failure Alarm If either of the Solenoid Valves should fail electrically either open circuit or short circuit this causes a failure of the Mattress to alternate This condition is shown by 0 _ 0 flashing red plus an audible alarm 5 4 Mains Failure Alarm If mains power is lost the main display will not function This alarm condition is shown by an additional red LED close to the Mode switch flashing red and an audible alarm 6 Maintenance Procedures 6 1 Safety Only qualified technicians trained or formally approved by Rober Ltd in the operation and maintenance of Rober products may carry out maintenance modification or repair work on Rober equipment Unqualified personnel attempting to work on Rober Power Units risk serious injury to themselves and others and possibly death by electroc
5. bed platform ensuring that the removable waterproof cover is on top and that the umbilical hose is at the foot end of the bed Secure the Mattress to the bed by fastening the two head end straps to the head platform area of the bed and its mid position and foot end straps around the bed platform The straps should not be too tight and should not interfere with the operation of the bed They should be checked regularly and secured on the moving parts only Cover the Mattress with a loose fitting sheet 2 2 Power Unit Hang the Power Unit onto the footboard The mounting hooks swivel to suit the thickness of the footboard or rail Connect the Umbilical Hose to the Power Unit by inserting the 3 male probes into the female quick connect sockets The Head Cell Side Former supply probe is physically non interchangeable with the other probes Correct connection is indicated by an audible click Check that the CPR Valve is turned to the Normal position Operation Plug the mains cable into a suitable 230v mains socket and switch on the Power Unit using the on off switch The LED display is illuminated and will flash 3 green horizontal bars followed by 3 green zeros 000 as the Mattress is inflated Once the Mattress is fully inflated the Power Unit automatically switches into the Dynamic mode and the display changes to alternatively flash _0_ and 0 _ 0 in green The A cells are allowed to deflate for 3 minutes then du
6. latform and is connected via an umbilical hose to a separate power control unit The objective of the product is to relieve the pressure created under a patient when lying on a bed so that blood flow is maintained and tissue oxygenation achieved It does this by intermittently relieving the pressure from the patient by sequentially deflating the supporting air cells of the mattress The Mattress consists of a series of 17 transverse air cells located within 2 longitudenal side formers all protected by a vapour permeable waterproof cover Sixteen of the transverse cells are arranged into alternate pairs of A and B cells which are filled and emptied in sequence while the single head end cell and the formers remain inflated A foam underside is placed within the Mattress providing secondary support and appropriate height partnership The software based system within the Power Unit controls solenoid valves enabling air to flow into or out of the air cells as required according to the operating mode selected It also maintains the air pressure within the mattress at the required level and controls the action of the audible visual alarm system in the event of mains supply failure or over or under inflation pressure A CPR Valve located in the umbilical hose permits the rapid deflation of the Mattress in an emergency Installation 2 1 Mattress Remove the bedding from the conventional mattress Place the Air Plus Mattress directly on to the
7. ring the next 2 minutes they are re inflated to the working pressure and held there The process is then repeated for the B cells so that the whole cycle repeats every 10 minutes The Static mode in which all the cells remain inflated is selected by pressing the mode switch on the control panel The LED display changes to 0 0 0 and blinks every few seconds as an indication that the system is operating correctly If the mode has not already been changed by the user the Static mode is automatically limited to a maximum time of 2 hours after which for reasons of clinical safety operation reverts to the Dynamic mode 3 1 Comfort Version 2 software This facility enables the preset mattress inflation pressure to be reduced and so achieve a softer mattress ISSUE 2 PAGE 2 OF 6 AIR PLUS USER MANUAL The Soft Dynamic mode is selected by selecting the Static mode as above and then pressing the mode button again The display changes to o_o and _o_ and the dynamic cycle operates but at with reduced mattress pressure The Soft Static mode is selected by pressing the mode button again The display changes to ooo and the mattress remains inflated at the reduced pressure Repeatedly pressing the mode button enables the Dynamic Static Soft Dynamic and Soft Static modes to be selected in turn 3 2 CPR Deflation
8. ution 6 2 Cleaning Procedures Warning Before cleaning the System make sure that the Power Unit is disconnected from the mains electricity supply Do not immerse the Power Unit in water or other fluids Do not autoclave nor use phenol for cleaning Do wash hands before commencing the cleaning process Don appropriate protective clothing such as gloves apron and a mask Ensure all work surfaces are cleaned before and after contact with the Mattress 6 3 Cleaning the Mattress 1 With cover left on the Mattress disconnect the Mattress from the Power Unit 2 Clean the surface of the wash down table with Hypochlorite solution or equivalent disinfectant 3 Wash Mattress top using hot water 60 degrees C containing detergent dry with a paper towel 4 Fold Mattress in half from head to foot end clean table and the exposed bottom half of Mattress Dry with paper towel 5 Invert folded Mattress to top half of table clean table and the exposed half of Mattress Dry with paper towel ISSUE 2 PAGE 4 OF 6 AIR PLUS USER MANUAL 6 Repeat steps 3 to 5 using Hypochlorite solution 1 000 parts per million available chlorine 7 Using suitable brush hot water detergent and Hypochlorite solution clean Umbilical Hose and CPR Valve Dry with paper towel 8 If required the Mattress Cover may be removed and machine washed at a temperature of 90 degrees C for not less than 10 minutes The individual Air Cells can be
9. wiped down with established disinfectants 9 The Mattress Cover may be sterilised using ETO or Draeger processes 6 4 Cleaning the Power Unit The Power Unit can be cleaned by wiping with a cloth dampened with a detergent solution Hyperchlorite solution or by sterilising using ETO or Draeger processes 7 Technical Data 7 1 Power Unit Serial Number Electrical Supply Power Consumption On rear label 230 volt 50 Hz Max 70 watts Fuses 2 x 500 mA Protection against shock Class 2 Class of Equipment Class B Noise Level Approx 30 dB A Dimensions 295 x 220 x 90 mm Weight 2 5 kg Service Interval 12 months Mattress Serial Number Internal at foot end Number of Air Cells 17 Dimensions 900 x 1970 x 130 mm Nominal Weight 5 5 kg ISSUE 2 PAGE 5 OF 6 AIR PLUS USER MANUAL 8 Supplier Address Direct Healthcare Services Limited Unit 8 Withey Court Western Industrial Estate CF83 1BF United Kingdom Tel 0845 459 9831 Fax 0845 459 9832 Dyna Form is a registered trade mark of Direct Healthcare Ltd Products undergo continuous further development We therefore reserve the right to change the specifications of the product and the warranty conditions without prior notice Issue 2 08 04 10 Manufactured in the UK ISSUE 1 PAGE 6 OF 6