Home
        Siemens OXYMAT 61 man
         Contents
1.                                                                                                           Figure 7 5 Reference gas channel for internal reference gas supply with ambient pressure     50                                  hPa   Table 7  4   Partno    Designation Order no  Remarks   04 1 Nozzle  complete C79451 A3480 B1    6 mm  3 000 hPa   Nozzle  complete C79451 A3480 B2 1 4   3 000 hPa  04 3 Capillary C79451 A3480 D518 3 000 hPa  04 4 Reference gas supply A5E00114838 Pump module complete  OXYMAT 61   122 Operating Instructions  12 2011  A5E00123067 02    Appendix       Technical specifications    8 1 General technical specifications    Table 8  1 Technical specifications     general technical specifications    General technical specifications    Measuring ranges 4  internally and externally switchable  manual  and automatic range switching is possible   Smallest possible measuring span 3  2 vol    or 5 vol    O2   Largest possible span 0 to 100  vol  O2   Measuring ranges with suppressed zero point Any zero point can be implemented within 0        100 vol   O2 provided that a suitable reference  gas is used  see also Reference gases   Page 126      EMC  interference immunity In accordance with standard requirements of   electromagnetic compatibility  NAMUR NE21  08 98   EN 50081 1  EN 50082 1  Degree of protection IP 20 according to EN 60529  Electrical safety According to EN 61010 1   overvoltage category III  Position of use  analyzer  Front panel vertical  D
2.                  No  Message Possible causes Remedy Remarks  w8 Temperature of   More than  3   C deviation If temperature remains  Q measuring head   from the setpoint temperature   constant  no need for  outside  see also S7  action  otherwise  contact  tolerance service department  w9 External Signal from external source Check Function 72 must be configured  maintenance accordingly  request  W10 AUTOCAL Chec   Tolerances that are exceeded   Perform AUTOCAL again This message disappears only  k error in AUTOCAL Check or false when the AUTOCAL has been  sample gas  assignment to the successfully completed   measuring range is wrong        Causes of maintenance requests    OXYMAT 61  112 Operating Instructions  12 2011  A5E00123067 02    Appendix       6 6 2 Faults    The following faults result in a fault message  output in the display  and are signaled  externally if a corresponding relay has been configured using function 71  The measurement  capability of the device is restricted if these messages appear  Immediate remedial  measures should always be carried out here by qualified maintenance personnel     Function 87 can be used to switch off  disable  each fault message individually     Each maintenance requests labeled with  Q  in the left column of the following table must be    acknowledged        No   S1 Parameter memory test failed    Message    Possible causes    EEPROM contains incorrect or  incomplete data in the working range    Remedy    1  Execute RESET or swi
3.     Attenuator          Dn Restrictor  DS Pressure switch  P Pump for reference gas supply                      Figure 7 2 Sample gas channel  with hoses    OXYMAT 61  Operating Instructions  12 2011  A5E00123067 02 119    Appendix                                                                                                                                                                                                                                                                      I E    E G D     e  ep  Ho           A                                                                                                       Figure 7 3 Gas connections   Table 7  3   Part no    Designation Order no  Remarks   03 1 Restrictor C79451 A3480 C10 Gas channel hose   03 1 Attenuator A5E00118236 Attenuator with hose   03 2 O ring  FPM  Viton  C74121 Z100 A6 1 item   03 4 Flow meter C79402 Z560 T1   03 5 Pressure switch C79302 Z1210 A2   OXYMAT 61   120    Operating Instructions  12 2011  A5E00123067 02       Appendix                            Figure 7 4 Reference gas channel for external reference gas supply with 2000 to 4000 hPa    OXYMAT 61  Operating Instructions  12 2011  A5E00123067 02 121    Appendix                                                                            10           Sie 88  me  KOs    if  4                                                                                                                                                                         
4.     The display of the drift values for the sensitivity calibration depends on the setting of the  calibration method  If total calibration is selected here  only the selected measuring range for  the sensitivity calibration is displayed  For single calibration  all measuring ranges can be  individually adjusted and can therefore have different drift values  which are shown  separately     The drift values can reset to 0 0 with the  Reset drift values  instruction  When you reboot the  device all measuring ranges have 0 0 as the drift value     OXYMAT 61  Operating Instructions  12 2011  A5E00123067 02    Operation       5 4    5 4 1    OXYMAT 61    5 4 Measuring ranges    Measuring ranges    After selecting the measuring range functions in the main menu  the following screen  appears when the third softkey   Measuring ranges   is pressed     Meas  ranges    40 Range selection    41 Define meas  ranges       Figure 5 21 Measuring ranges submenu    The measuring ranges menu contains all functions you need for selecting and setting the  measuring ranges     This menu is protected by the code of code level 1     Select measuring ranges  function 40     40 Select range    0 0  0 0  0 0    0 0    Autorange       Figure 5 22 Select measuring ranges   You can select a fixed measuring range or switch to automatic measuring range switching   All selection options are subject to mutual interlocking    Automatic measuring range switching is only possible under the following conditions  
5.    5 5 2 Limits  function 51   51 Limits MC    Limit 1   0 00   v v  on relay      Alarms at decrease 0  signal    Applies to MR agg Oo    Limit alarm on off O    Limit 2         Figure 5 31 Limits    The device monitors up to four limits  and these limits can be allocated to any measuring  range  Each limit can be allocated to any relay using function 71  Ifthis was not configured   the      indicator appears in the limit screen     You can only configure positive limits up to 100     Alarms at decrease signal        Here you select whether a limit alarm is to be switched if the entered limit is exceeded or  fallen short of      Applies to measuring range        Here  you allocate the limit to the desired measuring range s  by pressing the third softkey  several times  When you do so  the pointers move over the measuring range numbers and  indicate the ranges in which the limit monitor is to be active  The menu screen shown here is  for measuring range 3      Limit alarm      The limit monitoring of every single limit can be switched off individually using function 52     Resetting the limit alarm    If the limit relay is triggered  this state is obtained even when the measured value goes back  to the permissible range  The response of a limit relay is registered in the logbook  function  3   As soon as the cause of the limit alarm has been eliminated  the limit relay is reset  automatically     You can change to the next limit using the fifth softkey   L mit            OXYM
6.    e At least two measuring ranges must be available  A measuring range is said to be  present when the following is true  Start of scale value   full scale value    e The measuring spans must become greater or smaller    e The measuring ranges must  border on  each other or overlap    Operating Instructions  12 2011  A5E00123067 02 63    Operation       5 4 Measuring ranges    64       Measuring range types    The permissible measuring range constellations result which are shown in the figure below                             mbe  i   lt  mbe  i 1                                                            i   measuring range 1   4                                  mbe  i   gt  mbe  i 1                           mbe  end of measuring range             Figure 5 23 Measuring range types    Two measuring range types are distinguished   Type A     The full scale value must be smaller than the full scale value which follows it  The top  measuring range limit therefore becomes larger with every measuring range                 mba UU OU mbe       SSV Start of scale value  FSV Full of scale value    LS Low switchover point  select smaller measuring range  HS High switchover point  select larger measuring range    Figure 5 24 Measuring range type A    OXYMAT 61  Operating Instructions  12 2011  A5E00123067 02    Operation       OXYMAT 61       5 4 Measuring ranges    The following applies to measuring range switching      20   10  of span from full scale value    Measuring range 1    A
7.   3 3 Connections       3 3 1 2 Reference gas line  The connection stems for the reference gas has a pipe diameter of 6 mm or 1 4    The reference gas line must be as short as possible and have a small cross section     If you use N2 or O2as a reference gas  you must use a metal pipe as the supply line  If air is  used as the reference gas  it is recommended to install a dryer attachment in the intake to  avoid volume errors in comparison to the gas side that can be caused by air humidity     If you subsequently convert the device to a different reference gas supply  ensure that only  trained service personnel perform the replacement work for the connector stems and the  reference gas restrictor     3 3 1 3 Pressure sensor    The device has an internal pressure sensor for correcting the effect of pressure on the  measured value     It is firmly mounted in the analyzer unit and directly measures sample gas pressure through  the reference gas supply  The pressure sensor is therefore not involved in the installation     3 3 1 4 Gas preparation    To prevent contamination of parts from contact with the sample gas  which can thus  influence the measurement  the sample gas must be sufficiently prepared before it is  introduced in the device     The following components are usually located upstream from the sample gas inlet of the  device     e Gas sampling device  e Sample gas cooler   e Filter   e A gas suction pump    Depending on the composition of the sample gas  you may need a
8.   3 nti nee ee at net ee a cide eee ties 118  7 4 Gas  eanne yee  OO 119  OXYMAT 61    Operating Instructions  12 2011  A5E00123067 02 5    Table of contents       8 Technical specifications enseina kerian aaa EA AAE AEEA EARRA E A AEE ERKE E 123  8 1 General technical specifications      oooonnoonncccnnnnnnnncnnoconccnnnnnnnnnnnnnncnnnnnnnnnnn nn nnnn rnrn nenns nn nn nnmnnn eenn 123   8 2 Reference Jasmine 126   8 3 Z  ro p  inNtErrOr s ee e A A eet 127   8 4 Materials used in the sample gas Channel  ooccccoonocccccnnaciccnononcncnnnancnn nano nc nc nnnn nar nnnn rca rra rca 128   9 Dimension drawings 2 000000 iaa 129  9 1 DIMENSION drawings ere e ta d   da de edel 129   A APDO toda ATT A TT cad dir  s 131  A 1 EC Declaration of ConforMitY   oooooncccnnnnoccccnonocccccononcnnncnnncnc nano cnn ccnnn cnn rca nc arrancar nn rnnn anar 131   A 2 Return delivery    Rinne od iia 133   B ESD  Directives  c    s sscciasersa dabei stecevceonsa longs p aapa aaaea a tena cevnetdanataecdearedbecd Oaa TAG SOAD Ti ENT pa SEENA 135  B 1 ESD guideline S ironia tatiana 135   e List of abbreviations acronyms         uussussssnssnnennnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnsnsnnnnnsnsnnnnnsnsennnnsnnnnnsnnnnnnn 137  C 1 List of abbreviations   comico e dd 137  GIOSSAlY n A Ati ties elec ai ted a Ata eset E AN 141   o A O A NT 149  OXYMAT 61    6 Operating Instructions  12 2011  A5E00123067 02    General information 1    1 1    1 2    1 3    OXYMAT 61    Operating Instructions  12 2011  A
9.   Calibration tolerances  function 78     Setting the calibration tolerances    OXYMAT 61    With this function  you define the calibration tolerances     78 Calib  tolerance MC    Calib  tolerance at zero  in   of smallest MR   10     Calib  tolerance of sens   in   of current MR  10    Signal tolerance oO  violation       Figure 5 54 Setting the calibration tolerances    Operating Instructions  12 2011  A5E00123067 02 89    Operation       3 6 Configuration    Example    90    With softkey 3  Signal tolerance violation     you activate or deactivate the tolerance    monitoring     If you previously configured a relay output to  Maintenance request  using function 71  the  device outputs the changes in the zero point or the span with respect to the last calibration    as  Maintenance request         Note    In order to be able to use this function   Total calibration   function 23  must be activated        With  Calibration tolerance at zero point       you define the max  deviation proportional to the    smallest measuring span in       With  Calibration tolerance at span         you define the max  deviation proportional to the    current measuring span in       The calibration tolerance  which can be set from 0 to 99    refers to the measuring range  with the smallest measuring span in the case of the zero point  and to the measuring range  where the total adjustment is carried out  marked with  m  in function 22  in the case of the    span        Measuring range 1    95 
10.   External N x  maintenance  request 1 to 7        Condensate overflow     Gas cooler fault  see also section    Maintenance requests and  fault messages              Deleting the N  X After deletion  the device is restored to the initial state    logbook entries If the cause of a fault or maintenance request is not  eliminated  the corresponding message reappears in  the logbook    Function check  N X Relay must be configured for functional check using    CTRL  1 to 4 function 71 if  for example  the function is to be  checked by a second device    Start AUTOCAL N  X Note     Effective activation only possible in  Measure   operating mode     AUTOCAL must be configured  functions 23  24 and       25    Measuring range N  X For measuring range switching  1to4 on Note     Automatic measuring range switching  function 52   must be turned off                         OXYMAT 61  Operating Instructions  12 2011  A5E00123067 02 83    Operation       3 6 Configuration    84       Function    Necessary activation  voltage    Remark   effect       Zero gas on       Calibration gas  on       Sample gas on    OV    24  V    24 V pulse   min  1 s     Note   Effective activation only possible in  Measure   operating mode     Relay function under 71 must be set to zero gas   calibration gas  sample gas  and the corresponding  valves to be connected    Applies only to total calibration  since only one  calibration gas can be used  function 22        Zero calibration       Span calibration   
11.   Release of hazardous gases       Gases may be released if there are leaks in the sample gas channel  For this reason  toxic  and corrosive gases are not allowed to be be measured with this device     Failure to comply with this can result in death  personal injury as well as damage to  property and the environment           CAUTION    Incorrect mounting       The device can be damaged  destroyed or its functionality impaired through improper  mounting     e Before installing ensure there is no visible damage present on the device     e Make sure that process connectors are clean  and suitable gaskets and glands are  used    e Mount the device using suitable tools  Refer to the information in  Technical data    for  example installation torques requirements              OXYMAT 61  Operating Instructions  12 2011  A5E00123067 02 17    Mounting and connecting  3 2 Installation requirements       3 2 Installation requirements    Vibration    Select an installation location which is as vibration free as possible  If you install the device  in a cabinet or in desktop housing  you must hang it on support rails  It is not sufficient to  only secure the front side  since the intrinsic weight of the device creates too much load on  the chassis     Temperature  Ensure sufficient ventilation between the devices if you install them in control cabinets   If the device is installed outdoors  it must be protected from direct sunlight   Ensure that the permissible ambient temperature range 
12.   Sample gas  pressure  O2      oo            laptop   IN     A Maintenance center Local i    at ULTRAMAT 6E E  esse     control   2 channel unit  y  nd    RS485 ELAN        Sample gas  pressure  O2            CO2    O2  sample gas  1 pressure    I          N OXYMAT6E PT       ULTRAMAT OXYMAT 68 ak  y  Bi i    Te oo  ee 2 Container C yl  en a e    ELAN    Economical Local Area Network   Remote transmission via modem    Networking of various analyzers   Up to 12 channel addresses  interfaces  can be used for measured  ULTRAMAT 6  OXYMAT 6  OXYMAT 61  ULTRAMAT 23 variables  concentrations  pressures  temperatures  etc      Centralized maintenance by coupling to   SIPROM GA  higher level computer     Measurement data available in ASCII   PROFIBUS  format for fast processing    Remote control and download of records   Technology  RS485   9600 baud   repetition rate 0 5 s    via PC range up to 500m      1  ae  12  Channel addresses   13  and Control addresses  e g  for PC  etc      Figure 2 2 Various analyzers  networked with 485    OXYMAT 61  Operating Instructions  12 2011  A5E00123067 02 15    Description       2 4 Principle of operation    2 4    16       Principle of operation    Reference gas channel   Restrictors   Reference gas channels   Microflow sensor for measured signal  Sample gas inlet   Sample chamber   Paramagnetic measurement effect  Electromagnet with alternating field strength    OMAN DOA FWD         Sample gas and reference gas outlet    OXYMAT 61  principle o
13.   automatic  calibration  and for AUTOCAL Check  automatic check that the set calibration tolerances are  complied with  without calibration      The settings for  AUTOCAL Check   4th subfunction  only refer to the check of the  calibration tolerances without calibration        OXYMAT 61  Operating Instructions  12 2011  A5E00123067 02 55    Operation       5 3 Calibration    AUTOCAL Check operating mode  With this AUTOCAL function  you configure various AUTOCAL operating modes     The AUTOCAL Check is only for checking the calibrations  As with AUTOCAL  the device  executes the sequence configured in subfunction  AUTOCAL Check sequence     In contrast  to AUTOCAL  no new calibrations are performed  only the deviations are checked with  respect to selectable calibration tolerances     Autocal     check mode    Autocal   check on off O    Start autocal  check O  cyclically    Start autocal  check via O  binary input    Trigger autocal once    Abort autocal       Figure 5 13 AUTOCAL Check operating modes     AUTOCAL Check on off      During the state  4UTOCAL off   shown as   0    the settings on the switch  AUTOCAL start  cyclically  parameter  and  AUTOCAL start via digital input  do not have an effect on the  device   Trigger AUTOCAL once cannot be selected  A previously set cycle time continues  to run  but without triggering automatic calibration                  AUTOCAL Check start cyclically      You can activate AUTOCAL to be a regularly repeated cycle if you first set 
14.   ie Digital input    4    24 V  13 to 33 V   vo ARS SAI igital input 2   0  0V  Oto 4 5 V   DONT VA Digital input EP floating via optocoupler  me    TEYAT Digital input 1 to 4 N  ag   I      Relay 6     lt    el      A    amp         _ 10       Relay 5  KY  gt     HQ                  d    Mm    1            dD 8  _ 11       Relay 4 Contact load  SS EEE max  24 V 1 A  AC DC   es OL  a relay contacts shown   2 Say coy vhs ae Relay 3 de energized relay coil   gt  FT TAVI   A  gt  gt  TB                Relay 2     AP Note   E The wire and connector must  SS Ss Relay 1 oh be shielded and be   ee pe ys connected to the housing  Sa GND potential       Figure 3 6 Motherboard pin assignments    OXYMAT 61  Operating Instructions  12 2011  A5E00123067 02    27    Mounting and connecting       3 3 Connections    Connector SUB D 37 F  optional     Connector SUB D 9F X90  PROFIBUS DP        4  SSS SS CNTR P direction control  ie   1      RxD TxD P B     7  RxD TxD N A        4      VP   5V    Figure 3 7 Pin assignments of the add on board    28       Digital input 14 P   Digital input 13 P   Digital input 12 P Floating   Digital input 11 P via optocoupler  Digital input 10 P  0    OV  0 to 4 5 V   Digital input 9 P  4    24V  13 to 33V   Digital input 8 P   Digital input 7 P   Digital input 7 to 14 N    Contact load   max  24 V 1 A  AD DC  the relay has zero  current for the shown  relay contact position    Note    The wire and connector  must be shielded and be  connected to the housing 
15.   not trigger a maintenance request or fault message     The logbook contains a max  of eight pages  with four messages per page  lt works  according to the cyclic buffer principle  i e  when all eight pages  all 32 locations  are  occupied  the oldest message is overwritten     You can delete or block logbook entries  function 60   but you can also switch them off  individually  function 87         Note    If an error occurs whose error message is switched off with function 87  there is no reaction  at any configured interface  This applies to the ELAN interface  the analog output and the  relay output        Display measuring ranges  function 4     The measuring ranges defined using function 41 are listed here  However  you cannot carry  out any changes in this function     Operating Instructions  12 2011  A5E00123067 02 49    Operation       5 3 Calibration    5 3    50       Calibration    After selecting the calibration functions in the main menu by pressing the second softkey      Calibration      the following menu screen appears with additional choices     Calibration  20 Zero calibration    21 Span calibration    22 Setpoints total    23 Total single cal     continue    Figure 5 6 Calibration submenu    This menu is protected by the code of code level 1     You can calibrate the device using the functions available in this menu  If the function you  are looking for is not included in this overview  press the fifth softkey     Continue  to go to  other functions     T
16.  15   Pressure sensor  8   Pressure switch in sample gas channel  option   OXYMAT 61    Operating Instructions  12 2011  A5E00123067 02    23    Mounting and connecting  3 3 Connections          3 3 2 Electrical connections  3 3 2 1 Mains connection  AR WARNING       Dangerous contact voltage    Danger of electric shock in case of incorrect connection    e For the electrical connection specifications  refer to the information in Chapter   Technical specifications  Page 123      e Atthe mounting location of the device observe the applicable directives and laws for  installation of electrical power installations with rated voltages below 1000 V              e The device comes with an appliance plug which may only be connected to the mains  supply line by qualified personnel  The mains supply line must comply with the valid  regulations and conditions for the place of installation and be provided with a protective  conductor which lies at the potential of the enclosure  The cross section of each wire  must be  gt  1 mm   Connect the phase conductor to the marked position in the plug     e Install the mains line separately from the signal lines     e Provide a mains disconnection device in the direct vicinity of the device  for load rating   see rating plate   lt must be readily accessible and marked     e Check whether the existing mains voltage agrees with the mains voltage specified on the  rating plate     OXYMAT 61  24 Operating Instructions  12 2011  A5E00123067 02    Mount
17.  2 Desi vi iaa aa 13  2 3 Communication interface                         24244444444Hanennnnennnnnnnannnnnnennnnnnnnennnnnnnennennnnennnnrnnennnnnnanrnnen 15  2 4 Principle Of Operation    tie ua HE Hate bade a ES 16   3 M  unting   nd conneclinGg   2   2   A ee he ee as 17  3 1 Safety INSTUCIONS ss  A A A etch sede an a ERDE Renee 17  3 2 Installation requirements                        2 2444000042424nnnannanannnnnnnnnnnnnnnannnnnnnnnnnnnnnannnnnnn nn nannnn nn 18  3 3 A A A 19  3 3 1 Gas connections and internal gas flow diagraM     oooonnncccnnniniccnnnnncoccnnnancccnn narco noc cnn rra 20  3 3 1 1   Sample gas Nosice dei 20  3 3 1 2     Reference gas ING cities 21  33  13  SO AS 21  3 31 41   GAS PEPA O E 21  ASE AA eek eier keseehle 22  3 3 2 Electrical Connection Snas liarla area A 24  8 32  IMAINS connection 4  ac eee ee aie en aie ae none sinn 24  3 3 2 2 Connection of the signal CableS      oooocnniccinnnnncicinnnoccccncnnonccnnnnnnn cnn nono cnn nro nana rca 25  3323   PINVASSIQNIMEMS ee E 27  3 4 Example of AUTOCAL Connection                 ccc cccecceceeeeeeeeeeceececeeeeeeeaaeaeceeeeeceseneanaeceeeeeeneeesiaeeeees 29   OXYMAT 61   Operating Instructions  12 2011  A5E00123067 02 3    Table of contents       COMMISSIONING A NN 31  4 1 Safety instructions    ee te 31  4 2 Preparation for COMMISSIONING  errien iaaa eene nono na nono n cnn nano cnn anno rra rra 32  4 2 1 General informatica tddi listar dict    32  4 2 2 Selection of the reference Gas  4    ek dades
18.  2011  A5E00123067 02 117    Appendix       7 3 Electronics                                                                                           02 1 1    02 2                            Table 7  2    A                         2                      er T           Part no    Designation  02 1 Motherboard    Order no     C79451 A3480 D501  C79451 A3480 D502  C79451 A3480 D503  C79451 A3480 D504  C79451 A3480 D505    Remarks    Motherboard and firmware German   Motherboard and firmware English  Motherboard and firmware French  Motherboard and firmware Spanish  v Italian          02 1 1 Firmware  FlashPROM  C79451 A3460 S501 German  C79451 A3460 S502 English  C79451 A3460 S503 French  C79451 A3460 S504 Spanish  C79451 A3460 S505 Italian   02 2 Add on board C79451 A3480 D511 Relay  A5E00057307 PROFIBUS PA  A5E00057312 PROFIBUS DP  A5E00057164 Firmware update PROFIBUS       02 3 Plug in filter    W75041 E5602 K2       02 4 G  type fuse    W79054 L1010 T630  W79054 L1011 T100    200 V to 240 V  T 0 63 A 250 V  100 V to 120 V  T 1 A 250 V       02 5 LC display    W75025 B5001 B1       02 6 Adapter board  LCD keyboard    C79451 A3474 B605             02 7 Front panel       A5E00259978          With membrane keyboard       OXYMAT 61  Electronics    118    OXYMAT 61    Operating Instructions  12 2011  A5E00123067 02    Appendix       7 4 Gas channel                    Restrictor   inserted into the hose and fastened with a  hose clamp             1 2 Sample gas  4 Reference gas   
19.  3 l min and 1 l min     OXYMAT 61  Operating Instructions  12 2011  A5E00123067 02 35    Commissioning  4 2 Preparation for commissioning       Calibrating the zero point    The physical zero point is calibrated with function 20  It is applied to all configured measuring  ranges     Calibrating sensitivity  The sensitivity is calibrated with function 21     Compensating the effect of temperature    The compensation of the temperature effect is stored permanently in the software  firmware   for the OXYMAT 61  No further actions are required for this     Noise suppression    Noise in the measured signal can be suppressed using function 50  This function allows you  to configure a low pass filter  which is assigned a time constant of maximum 100 s     OXYMAT 61  36 Operating Instructions  12 2011  A5E00123067 02    Commissioning       4 2 4 Calibration examples    4 2 Preparation for commissioning    The following examples reflect typical applications for the OXYMAT 61     O2 monitoring in gases    Measurement task  Measurement of oxygen in N2   Measuring range  0 to 5   O2  Reference gas  N2  Calibration gas  4 7   O2                                  Procedure Function   Input Remarks  no   Selection of measuring range start of scale value 41 0 0  gt  0 2 4  mA  Selection of measuring range full scale value 41 5 5  gt  20 mA  Specification of the setpoint for the physical zero point 22 0 Setpoint for physical zero point  Specification of the setpoint for the sensitivity 22 4 7
20.  33  4 2 3 Initial  COMMISSIONING 100 A 35  424 Calibration  example Sesen ri cita 37  Opera  teeth ech e detache eee hag eet eatin ae ae Ses ee aie 39  5 1 General information   e eiea nenne nad steadied taeda nade ge ant 39  5 1 1 Measured value display and control panell                 cccccceeeeeeeececeeeeeeeeeeeeaeeeeeeesesecaaeaeeeeeeeeeeeenaees 39  5 1 2 Device operating  Modes    aida ita 41  5 1 3 Ecning US e o o 43  5 1 4 Operating  SsegUENGE  diia 43  STAL Introduction to the man menu    ee kette el 44  91 472  Entering a submen   ono oi 45  9 1 4 3    Returning to  display Mode oca td ge 46  9 1 4 4 East function Selection cti 46  5 1 5 Overview  of operating fungtions    n  n ee len E AS 47  5 2 Analyzer Status dida 48  5 2 1 Analyzer configuration  function 1     o ooooocnnnncccnnnnacicinnnancccnnnorccnnnnorr cnn narran rr 48  5 2 2 Diagnostic values  function Zj irs aias aa raae a AANE na raaa aaa AARET fia Ea Aa ia ia aA 49  5 2 3 Logbook  function IJss niera aniani ic 49  5 2 4 Display measuring ranges  function 4      oooonnncccnnnnniciconnncccnnnnonancnn nono nnnnnnnn cnn nnnn narran rca nnnn nn 49  5 3 Calre o R T E A E 50  5 3 1 Calibration of the zero point  function 20                uuressnnnnnesnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnennnnnnn nn 51  5 3 2 Span calibration  function 21       ooononcccnnnnniccnonocccccnnonnnnnnnnancnc nan ncncnnno nn nr rnnn rca r cnn rca rara 52  5 3 3 Zero span setpoints  function 22     o ooccconnnnccccnnnoccccnononccc
21.  5 7 General information    The following overview is a list of the device functions     This list corresponds to software version 4                             Main menu item Function number Name of function  Analyzer status 1 Factory data  2 Diagnostics values  3 Logbook  4 Display measuring ranges  Calibration  code 1  20 Calibrate zero  21 Calibrate span  22 Zero span setpoints  23 Calibration settings  Total  Single   24 AUTOCAL  25 Drift values  Measuring ranges 40 Select measuring ranges   code 1  41 Define measuring ranges  Parameters  code 1  50 Electric time constant  51 Limits  52 On off functions  53 Status signals  54 Graphic measured value display  55 Measured value display  56 LCD contrast  57 Magnetic field frequency  58 Date time  59 Measuring point switchover  60 Logbook settings  Configuration  code 2  70 Analog output  71 Relay assignment  72 Digital inputs  73 ELAN configuration  74 Reset  75 Save  load data  76 Suppress short noise signals  77 Measured value memory  analog output   78 Calibration tolerance  79 Change codes  80 Analyzer test  81 Select language  82 Pressure correction  83 Correction of cross interference  84 Phase adjust  85 Switch valves  86 Linear temperature compensation  87 Faults on off  90 PROFIBUS configuration          Overview of operating functions    Operating Instructions  12 2011  A5E00123067 02    47    Operation       5 2 Analyzer status    5 2 Analyzer status    After selecting the diagnostic functions in the main menu by 
22.  Activation initiates calibration  See also  Example    digital inputs        Autorange    N  X    Activate  Automatic measuring range switching        Autocal check    N  X    Note   Effective activation only possible in  Measure   operating mode     Start Autocal check  function 24           Measuring  protection          N  X          It is possible to define a digital input  Measuring  protection  which has the following effect    If the device is in the  Measure  operating mode   functional check deactivated   it remains in this state   i e     e The device can no longer be decoded     e The device can no longer be set to  Remote    The message  Measuring protection activated   appears in the status line of the measured value  display           OXYMAT 61  Operating Instructions  12 2011  A5E00123067 02    Operation       5 6 Configuration    5 6 4 ELAN configuration  function 73   With this function  you set the parameters for an ELAN network   73 ELAN config   Channel address    Measured value  telegram     ELAN tag       Figure 5 48 Configuring ELAN    Channel address     Set the channel address for the device here  Addresses from 1 to 12 can be set  In an ELAN  network  each address may only be used once  Addresses of devices that are used to  correct the pressure or the influence of interfering gas may not be entered at this location     Measured value telegrams on off     Here you can switch the cyclic  independent transmission of measured values on or off every  50
23.  O       Figure 5 58 Selecting a language    The device is always delivered in the ordered language  Usually  English is included as a  second language  if English is the first language  Spanish is set as the second language      5 6 13 Pressure correction  function 82     Variations in the sample gas pressure can be corrected within the range from 800 to 1200  hPa  absolute  with the OXYMAT 61     This function offers the possibility for pressure correction using one of the following options   e Pressure correction using an internal pressure sensor    e Pressure correction using an external pressure sensor via analog input 2     example as  shown     e Pressure correction using an external pressure sensor via ELAN  RS485      82 Pressure corr  0    ID T    With ext  pressure signal     on analog input 2    Analog inp  2  O to 20 mA       for range     0    1013 hPa       Figure 5 59 Pressure correction via analog output 2    The pressure correction can also be switched off using function 52   On off functions       OXYMAT 61  Operating Instructions  12 2011  A5E00123067 02 93    Operation       3 6 Configuration    Pressure correction using an external pressure sensor via ELAN    82 Pressure corr  0     With ext  pressure signal     from ELAN    Channel    4  Oo  NO  994 hPa CTRL       Figure 5 60 Pressure correction via ELAN       Note  Measured value  Pressure     The measured value  Pressure  is an internal value in the OXYMAT 61 gas analyzer which  can be supplied to ano
24.  Setpoint for sensitivity  Start zero point calibration 20 Introduce N2  Start sensitivity calibration 21 Introduce calibration gas          Room air monitoring    Measuring range  15 to 21   Oz  Reference gas  Air  20 95   Oz   Calibration gas 15 3                                 O2   Procedure Function   Input Remarks  no    Selection of measuring range start of scale value 41 15 15  gt  0 2 4  mA  Selection of measuring range full scale value 41 21 21  gt  20 mA  Specification of the setpoint for the physical zero point 22 20 95 Setpoint for physical zero point  Specification of the setpoint for the sensitivity 22 15 3 Setpoint for sensitivity  Start zero point calibration 20 Introduce air  Start sensitivity calibration 21 Introduce calibration gas             OXYMAT 61  Operating Instructions  12 2011  A5E00123067 02    37    Commissioning       4 2 Preparation for commissioning    O2 measurement in flue gas    Measuring range  0 to 10   Oz  Reference gas  Air  Calibration gas  Na       NOTICE       Reference gas out of range       section 5      In this example  the O2 content of the reference gas is not within the measuring range of up  to 10   O2  Since the measuring span is greater than 5    however  an exception can be  made in the selection of the reference gas     In this case  the pressure correction must be necessarily activated  see also Function 82 in                                  Procedure Function   Input Remarks  no   Selection of measuring range start of
25.  Table 5  2    Device operating modes    5 7 General information       Mode  Coded display mode    Properties    e Measured value display is  shown    e Protected submenus can  only be reached by entering  a code   e The current operating mode  of the device  except for   Measure   is displayed in  the bottom line   e Functional check not active    Remarks    The device only supplies  reliable measured values in this  mode  as far as it s in the   Measure  operating mode   From the operator control mode   you can reach this mode by  pressing the  MEAS  button  twice        Decoded display mode    e Measured value display is  shown    e The submenus protected by  the entered code are  accessible   e The current operating mode  ofthe device  except for   Measure   is displayed in  the bottom line   e Functional check active    e Measured value can be  influenced    From the operator control mode   you can reach this mode by  pressing the  MEAS  button  once and confirming or  discarding the made entries        Operator control mode          e Menu or function is  displayed    e Settings and inputs can be  made    e Functional check active    e Measured value can be  influenced       From  Coded display mode  you  can reach this mode by entering  the code of the corresponding  operation level     In this mode  you can  configure calibrate the device           OXYMAT 61  Operating Instructions  12 2011  A5E00123067 02    41    Operation  5 7 General information       Schematic di
26.  and commissioning  as well as for  safety during the operation and maintenance of the described product  Only qualified  personnel have the required professional knowledge for correctly interpreting the generally  valid safety notes and warnings in this manual in each specific case and to act accordingly     This manual is an inherent part of the scope of delivery  despite the fact that it can be  ordered separately for logistic reasons     General information       1 4 Special information and warnings    Due to the variety of technical details  it is not possible to consider every single detail for all  versions of the described product and for every conceivable case in the set up  operation   maintenance and use in systems  For further information  or in the case of problems which  are not covered in enough detail in this document  please request the required information  from your local or responsible Siemens regional office        Note    In particular  before using the device for new research and development applications  we  recommend that you first contact us to discuss the application in question        1 4 Special information and warnings    This manual provides you with information on using  installing  operating  and maintaining  the device     Pay particular attention to all special information and warnings  Information of this type is set  apart from the rest of the text and is marked with the corresponding pictograms  This  information provides you with useful 
27.  chamber  it can be compensated with this function     86 Lin  temp  comp  MC    Aftercompensation of    the zero point    Aftercompensation of      the span    Figure 5 67 Linear temperature compensation    Operating Instructions  12 2011  A5E00123067 02 99    Operation       5 6 Configuration    Temperature compensation at the zero point    Example     Assuming an average temperature Tm  two different correction values can be specified for  ranges of increased temperature and decreased temperature     If the zero point changes by  0 3    relative  in relation to the difference between 100   O2  and the start value of the smallest span due to a temperature increase in the sample  chamber changes from Tm to Tw     the value determined by the following formula should be  entered under  A  for a temperature increase       0 3     A      Tu   Tm     x 10    10   C     A factor can be determined for a temperature decrease in the same manner     If only a correction factor is determined  the first value of opposite sign is entered for the  second correction value     Temperature compensation in the measured value     Example     100    The procedure is the same as for the zero point  however  the percentage change refers to  the measured value itself     If the measured value changes from 70   to 69   with a temperature increase of 4   C  the  percentage change is    Ze x 100   1 42    4   C     and therefore  A   3 55    10  C         Note    If the zero point changes to negative
28.  control mode    Generic term for a number of independent states which the device can assume during  operation  e g  measure  standby  pause  etc       Various modes into which the device can be put by intervening from the outside  Three  modes are distinguished     e Coded display mode  e Decoded display mode  e Operator control mode    Generic term for a number of independent states which the device can assume during  operation  e g  measure  standby  pause  etc       Certain operating area  menus and functions  which is either freely accessible or protected  by a code     Operator control mode    Sample gas    Sealing    OXYMAT 61    Operating mode in which the device is partially or entirely decoded and the functional check  is activated  The device is configured in this mode     A gas which has been extracted from a process for analysis     Protection of the materials against escaping of gases     Operating Instructions  12 2011  A5E00123067 02 145    Glossary       Signaling contact    Usually a floating contact contained in electronic components  e g  relay  which signals the  occurrence of an event defined as a fault to a control unit     Span    Difference between defined start of scale value and defined full scale value     Span calibration    Adjustment of the span using a suitable calibration gas     Spike    English   Thorn  pin  prong  Undesired  briefly occurring peaked interference of a measured  signal     Standard configuration    Totality of standard featur
29.  following query appears     Return to meas  mode    Accept modifications        Figure 5 4 Query before jumping back    e Press  YES  to accept changes permanently into the working memory of the parameter  memory or  NO  to discard the changes  The device then changes to  Decoded display  mode      By pressing the  ESC  brings you back to the most recent function screen  Changes are  accepted here without a further query     Coding the device    After jumping back to  Decoded display mode  with  ESC  or  MEAS   if you press  MEAS   again you put the device back into  Coded display mode   Only now is the function check  once again deactivated and all states invoked by the decoding are released again     Fast function selection    In order to directly access the desired function when operating frequently from the  Coded  display mode   a  Power user operation  was created  This allows you to directly access the  desired function by entering the function number  This allows you to skip menu levels   However  the functions can only be directly called from the measured value display  For fast  function selection  proceed as follows     1  Enter the number of the desired function using the number keys     2  Press the softkey of the component with the arrow     If the desired function is protected  by a code  you will be asked to enter the code     OXYMAT 61  Operating Instructions  12 2011  A5E00123067 02    Operation       5 1 5    OXYMAT 61    Overview of operating functions   
30.  frequency back to the factory default of 8 095 Hz        NOTICE       Calibration of zero point and sensitivity          After each frequency change  the zero point and sensitivity must be calibrated again        Operating Instructions  12 2011  A5E00123067 02 75    Operation  5 5 Parameters       5 5 9 Date time  function 58     With this function  you can set the date and time     58 Date Time MC    New date  dd mm yy 24h da   01   03 07 y3      New time   12   53    Set clock e    Actual date  Actual time     01 03 2007 12 53       Figure 5 40 Setting the date time    The device features a system clock that is not buffered against mains failure  no a real time  clock   When starting the device  the clock starts with the date 1 1 1995     When the function is called  the cursor is placed at the first place of the date display  Enter  the new settings in the order  day  month  year  By pressing  ENTER   jump to the next input  field  With a 24 hour based system  hours  minutes   set the time in the same way        NOTICE             When the device is switched off  the clock stops and is not updated        The settings are especially important for troubleshooting  Errors which are always stored in  the logbook  function 60  can be allocated more easily with the help of the date and time     Press the third softkey  Set clock  in order to accept the set data  These then appear at the  bottom edge of the display     OXYMAT 61  76 Operating Instructions  12 2011  A5E00123067 0
31.  function 41  The start of scale or full scale values are assigned to the 0 2 4  or  20 mA of the analog output     If you define multiple measurement ranges  we recommend that you set measurement range  1 for the smallest measuring span  MS etc  In general  the following applies then   MS1  lt  MS2  lt  MS3  lt  MS4    Setting the physical Zero point    If the composition of test and reference gas are identical  their O2 difference is therefore zero  and there is no measuring signal  We refer to this as the physical zero point  Depending on  the reference gas  the physical zero point may therefore have any value between 0 and   100   Oz  The setpoint of the physical zero point is specified under function 22     Setting the setpoint of the sensitivity    The setpoints of the sensitivity must be as far from the physical zero point as possible  at  least 60   of the respective measuring span   The setpoint or setpoints are entered with  function 22  The calibration gases required for this must be made available for the  adjustment     Selecting the calibration method  Set the calibration mode to total or single calibration using either function 23 or 52     e Single calibration means that each measuring range is calibrated using its own calibration  gas     e Total calibration means that you adjust one  leading measuring range  and all other  measuring ranges are  tracked  to this using the switching ratio     When performing the calibration  ensure that the gas flows between 0
32.  in normal Measuring range limit in case of fault     output   mA operation CTRL  Start of scale Full scale value     Start of scale Full scale value    value   mA mA value   mA mA   0 20  1 21 0 21   2 20 1 21 2 21   4 20 2 21 2 21   4 20  NAMUR  13 8 20 5 3 21 5                            Note    If the electronics is defective  it is possible that the analog output may get stuck at approx   1  mA or approx   24 mA        Operating Instructions  12 2011  A5E00123067 02    79    Operation       3 6 Configuration    5 6 2 Relay outputs  function 71     80       With this function  you can assign each relay functions listed in the table  Relay outputs   possible functions    whereby you may only assign each function once  In other words  for  example   Fault  may not be assigned to two relays at the same time     71 Relay outputs    R 1 Fault   R 2 Maint  req   R 3 Funct  cont   R 4 not used    continue    Figure 5 45 Assigning the relays  There are six freely configurable relays included in the standard equipment whose  switchable output contacts  max  24 V 1 A  can be used for signaling  actuating valves  etc     If these six relays are insufficient  you can retrofit additional electronics with eight further  relays  add on board      The connection assignments for the individual relays in the zero current state can be found  in the routing assignments  Motherboard pin assignments   The relays are preset at delivery     Press the fifth softkey      continue    to access the 
33.  person  coming into contact with the materials indicated  These values correspond to IEC 801 2                specifications   Voltage in kV   kv  A  16 O Synthetic material  15  14 N    Wool  13 N  8  Antistatic material  for example wood  12 N or concrete  11 SO  10    9 N  8 N  7 N  N  8    5 N  ES  4 N  a D x eC     m    2       1   DS  HHH  gt   5110 20 a 40 50 60 70 80 90 10 Relative humidity in       Figure B 1 Electrostatic voltages on an operator    Basic protective measures against electrostatic discharge    e Ensure good equipotential bonding   When handling electrostatic sensitive devices  ensure that your body  the workplace and  packaging are grounded  This prevents electrostatic charge     e Avoid direct contact   As a general rule  only touch electrostatic sensitive devices when this is unavoidable  e g   during maintenance work   Handle the modules without touching any chip pins or PCB  traces  In this way  the discharged energy can not affect the sensitive devices     Discharge your body before you start taking any measurements on a module  Do so by  touching grounded metallic parts  Always use grounded measuring instruments     136    OXYMAT 61  Operating Instructions  12 2011  A5E00123067 02    List of abbreviations acronyms                                                                                                             C 1 List of abbreviations  Abbreviation   Meaning   gt  Greater than   lt  Smaller than   gt  Greater than or equal to   lt  S
34.  potential     Connector SUB D 9M X90  PROFIBUS PA    OXYMAT 61  Operating Instructions  12 2011  A5E00123067 02    Mounting and connecting  3 4 Example of AUTOCAL connection       3 4 Example of AUTOCAL connection    Function in the device Connection end                                                                                       VO  d  o    Connector SUB D 37 F  optional   Zero  gas  M  V1  d  c    Tham  E  ug ET  A Calibration  EPA gas 1     a      V2  d  Cu      d  c    AAA Calibration  gas 2     V3  d  c         Calibration  PE    V4  d  c    A  ce Calibratio  gas 4       AAA 24 V power Sample  M supply   max  gas  AL   In the relay contact position shown   the relays are de energized VM  d  c      d  o   de energized  open  d  c   de energized closed    Sample gas inlet    Gas preparation elements       Figure 3 8 Connector pin assignment and valve diagram of an AUTOCAL connection    OXYMAT 61  Operating Instructions  12 2011  A5E00123067 02 29    Commissioning 4    4 1 Safety instructions    Explosion hazard    The analyzer must not be operated in potentially explosive atmospheres  The supply of  gases with combustible components in concentrations above the lower explosive limit  LEL   must be clarified in consultation with the relevant hazard experts and is ultimately the  responsibility of the operator              warn ING  Dangerous contact voltage       Certain parts of this device are under dangerous voltage  In order to avoid contact with live  parts  t
35.  scale value 41 0 0  gt  0 2 4  mA  Selection of measuring range full scale value 41 10 10 gt 20mA  Specification of the setpoint for the physical zero point 22 20 95 Setpoint for physical zero point  Specification of the setpoint for the sensitivity 22 0 Setpoint for sensitivity  Start zero point calibration 20 Introduce air  Start sensitivity calibration 21 Introduce N2             Purity monitoring of oxygen    Measuring range  95 to 100   Oz  Reference gas  Pure Oz  Calibration gas  95 6   O2                               Procedure Function   Input Remarks  no   Selection of measuring range start of scale value 41 95 95  gt  0 2 4  mA  Selection of measuring range full scale value 41 100 100   20 mA  Specification of the setpoint for the physical zero point 22 100 Setpoint for physical zero point  Specification of the setpoint for the sensitivity 22 95 6 Setpoint for sensitivity  Start zero point calibration 20 Introduce pure O2  100     Start sensitivity calibration 21 Introduce calibration gas             Detailed operating instructions for performing all the required testing functions are provided    in section 5  Operation      38    OXYMAT 61  Operating Instructions  12 2011  A5E00123067 02    Operation       Operation  5 1 General information  5 1 1 Measured value display and control panel    Q e  3     40     72 11  I  E   d mnnn      6    noes          Status display  Display of dimension    Status line  can be programmed under function 53    Display of measured c
36.  with temperature change  A has a positive sign and  vice versa  The same applies to a measured value which becomes smaller        OXYMAT 61  Operating Instructions  12 2011  A5E00123067 02    Operation       5 6 18    5 6 19    OXYMAT 61       5 6 Configuration    Error on off  function 87     With this function  you switch the message for maintenance requests and faults  see Tables   Causes of maintenance requests  and  Causes of error messages   off individually so that  there is no entry in the logbook  no message in the status line of the measured value display   and no external signal sent     87 Error On Off    S1  Parameter memory   2     3    S4  External fault    continue p    Figure 5 68 Switching errors on and off    Error messages which do not apply to this analyzer are identified by the absence of text  following the fault number     PROFIBUS configuration  function 90     This function can only be called if the device contains additional PROFIBUS electronics   add on board   This function is used to set the PROFIBUS parameters     90 PROFIBUS config     Address  126   TAG   OXYMATSIXTYONE  Ident number wo gy    Relay on PB    Software vers  2 0 0  Boot software  0 2 0       Figure 5 69 Configuring PROFIBUS    Operating Instructions  12 2011  A5E00123067 02 101    Operation       3 6 Configuration    You can set the following specific PROFIBUS parameters     e Address  PROFIBUS station address   The address can be set from 0 to 126   Under TAG  you can display t
37. 0 ms        Note  Independent sending of measured values  If you have set up your own communication control system  this option offers a simple way of    checking an ELAN telegram  In order to avoid unnecessarily loading the device and the  ELAN network  however  the function should only be switched on as needed        ELAN tag   Display of the tag assigned to the device in the ELAN network     For more details  refer to the ELAN interface description  Order no   C79000 B5200 C176  German  C79000 B5276 C176 English      OXYMAT 61  Operating Instructions  12 2011  A5E00123067 02 85    Operation  3 6 Configuration       5 6 5 Reset  function 74     With this function  you can carry out a warm restart     74 Reset    Trigger reset e       Figure 5 49 Carrying out a reset  After calling this function  the device switches to the  Warm up phase   which activates the    functional check at the same time  The device is only completely ready for operation again  after successfully running through this phase     5 6 6 Save  load data  function 75     You can configure the user data memory with this function     75 Save data MC    Save user data e    Load user data e    Load factory settings e       Figure 5 50 User data memory   Press the softkey 1  Save user data e g  after the system has been successfully  commissioned  All individual settings are then saved    Press the softkey 2  Load user data   to load the last user data saved     These functions are important when the device is 
38. 2    Operation       5 5 10    5 5 11    OXYMAT 61          5 5 Parameters    Measuring point switching  function 59     With this function  you can assign the device up to six measuring points  which are switched  automatically at the expiration of a configured period     59 Sample selection    Rel  2   45   Rel  1 20    M  p  switching on off    Figure 5 41 Measuring point switching  The condition is that you configured the measuring point relay beforehand using function 71   Relay outputs   which then actuates the corresponding solenoid valves     Every measuring point relay is also allocated a time period  which you can enter in the  respective input field  Values between 0 and 60  minutes  are possible for this input     Press the fifth softkey to activate deactivate the measuring point switching     You can allocate a signal relay to each measuring point relay  This allows measuring point  identification separate from the measuring point relay  You also use function 71 to configure  the signal relay     Logbook settings  function 60     With this function  you delete or block logbook entries   60 Setup logbook MC  Clear logbook e    Lock logbook a    Figure 5 42 Configuring the logbook    Status messages such as maintenance requests or faults cannot be suppressed by   blocking   They still appear  despite the blocked logbook     Operating Instructions  12 2011  A5E00123067 02 77    Operation       3 6 Configuration    5 6 Configuration    This menu contains all function
39. 4  Drift values  62    E    EC Declaration of Conformity  133   Edit mode  43   ELAN  15  85   Electric time constants   See Time constants     149    Index       Electrical connections  AUTOCAL connection  30  Pin assignments  28  Position  20  Signal connections  26   Electronics  Spare parts  118   Error messages  101   ESD guidelines  137    F    Factory data  48  Fault  110   Message  113  Fault messages  101  Flue gas measurement  38  Functions   See Operating functions   Fuses  109  124    G    Gas connections  Position  20  Spare parts  121  Gas flow diagram  23  Gas inlet  125  Gas preparation  22  32  Guidelines  ESD guidelines  137    Influencing variables  126  Initial commissioning  35  Input dialogs  43  Installation   Installation location  19   Preconditions  19   Reference gas connection  33  Interface   ELAN  85   PROFIBUS  101  Interferences  126  Interfering gas  94  128    L    Language  93  Language setting  93    150    Leak test  34  107   Limit monitoring  69  Limits  69   List of abbreviations  139  Logbook  49  77  110    M    Magnetic field frequency  75  Main menu  44  Mains connection  25  Maintenance  Maintenance intervals  103  Safety instructions  103  Maintenance interval  103  Maintenance request  110  Maintenance requests  Message  111  Measured value  Display  43  74  Graphical representation  72  Negative measured value  79  Phase adjustment  98  Progress display  72  Signal  98  Temperature compensation  99  Measuring characteristics  12
40. 5  Measuring head  105  106  Measuring point switching  77  Measuring principle  11  17  Measuring range  63  Calibration  52  select  63  Specify  66  Switching  63  Measuring range switching  63  Memory modules  87  Messages  110  Motherboard  108    N    Nameplate  115  Noise  88  Noise suppression  68    O    On off functions  70  Operating language  93  Operating mode  41    OXYMAT 61  Operating Instructions  12 2011  A5E00123067 02    Index       Operating sequence  42   Operation  39  Inputs  43  Menu structure  44  Power User  46  Sequence  42   Operator functions  Analyzer status  48  AUTOCAL  55  Calibration  50  Configuring the analog output  78  Correction of cross interference  94  Device test  91  Digital inputs  82  Drift values  62  ELAN  85  Electric time constants  68  Graphical measured value representation  72  Language selection  93  Limits  69  Load user data  86  Logbook  77  Measuring point switching  77  Measuring ranges  63  Noise suppression  68  Operating language  93  Overview  47  Phase adjustment  98  Pressure correction  93  PROFIBUS  101  Quick selection  46  Relay assignment  80  Save user data  86  Set contrast  74  Setting the calibration tolerances  89  Setting the code  91  Setting the date time  76  Setting the magnetic field frequency  75  Setting the measured display  74  Show hide error messages  101  Span calibration  52  Status messages  71  Suppressing noise  88  Switch on off functions  70  Switching valves  99  Temperature compen
41. 50021 x X  94 9 EC EN 50281 1 1 A1 x  94 9 EC EN 60079 14 sec 13 X  94 9 EC EN 954 X  94 9 EC EN ISO 13849 1 X  94 9 EC ZH 1 10  sec  1 3 4 2  2 3 1 X X  94 9 EC DIN EN 50104  94 9 EC DIN EN 50271    all enviroments included  Certificates   Zertifikate       Certificates f E31  a A C E   a A C E   a A C E   a A C E a A C E  Zertifikate E32  E33    b 0 1 b 0 1 b 0 1 b 6 7 b 2 3  E38 E39 c B c B  B c B c B  g 30 d A B G   d A B G   d AB G   d A B G d A B G  e E40 e E11 e E12  BVS 03 ATEX G013 X X  TUV 03 ATEX 2278 X X  T  V 01 ATEX 1686 X X  T  V 01 ATEX 1697 X X  T  V 01 ATEX 1708 X X  PTB 00 ATEX 2022 X X                      Inspection   Surveillance     Kontrolle     berwachung   Directive Notified Body Product Quality Assurance   No    Richtlinie Benannte Stelle Qualit  tssicherung Produktion  94 9 EC ATEX T  V NORD CERT GmbH  Am T  V 1 D 30516 Hannover  0044      Page 2 2                      OXYMAT 61  132 Operating Instructions  12 2011  A5E00123067 02    Appendix       A 2    A 2 Return delivery    Return delivery    The analyzer or replacement parts should be returned in their original packaging material  If   the original packing material is no longer available  wrap the analyzer in plastic foil and pack  in a sufficiently large box lined with additional padding material  wood shavings  foam rubber   etc    If you use wood shavings  the padding should be at least 15 cm thick on all sides     When shipping overseas  the analyzer must be additionally sealed air tight i
42. 5E00123067 02    Device model and validity    You have purchased the OXYMAT 61  a device that that undergoes continuous development  and therefore can vary in functionality and software release in comparison to other devices  of the same type     This manual describes the OXYMAT 61 gas analyzer with the software release 4 8 3     Information for our customers    Before beginning work with this device  please read this manual  lt contains important  information and data whose observation ensures proper device function and saves you  servicing costs  The manual will help you to operate the device more easily and efficiently   allowing you to achieve reliable results     General information    This device left the factory in a safe and proper condition and has been tested  In order to  maintain this condition and to ensure safe operation of this product  it should only be used in  the manner described by the manufacturer  Furthermore  proper transportation  storage   installation  operation and maintenance of the device are vital for ensuring correct and safe  operation     This manual contains the information required for the intended use of the described product     It is addressed to technically qualified personnel who are specially trained or who have the  relevant knowledge of automation technology  measuring and control systems      Knowledge and technically correct implementation of the safety notes and warnings  contained in this manual are required for safe installation
43. 7 xax0c bcdx Z e f xxx  7MB2021 xXXX0 XXXX Z 9 XXX 7MB2027 xxxx0 bxxx Z g xxx    7MB2041 XXXXX XXXX Z XXX  a  b  c  d  e  f  g  X  see tables page 2   siehe Tabellen Blatt 2       The product described above in the form as delivered is in conformity with the provisions of the  following European Directives    Das bezeichnete Produkt stimmt in der von uns in Verkehr gebrachten Ausf  hrung mit den  Vorschriften folgender Europ  ischer Richtlinien   berein     2004 108 EC Directive of the European Parliament and of the Council on the approximation of the laws of the Member States relating  to electromagnetic compatibility and repealing Directive 89 336 EEC    EMC Richtlinie des Europ  ischen Parlaments und des Rates zur Angleichung der Rechtsvorschriften der Mitgliedstaaten   ber  die elektromagnetische Vertr  glichkeit und zur Aufhebung der Richtlinie 89 336 EWG     2006 95 EC Directive of the European Parliament and of the Council on the harmonisation of the laws of Member States relating to  electrical equipment designed for use within certain voltage limits    LVD Richtlinie des Europ  ischen Parlaments und des Rates zur Angleichung der Rechtsvorschriften der Mitgliedstaaten  betreffend elektrische Betriebsmittel zur Verwendung innerhalb bestimmter Spannungsgrenzen     94 9 EC Directive of the European Parliament and the Council on the approximation of the laws of the Member States concerning  equipment and protective systems intended for use in potentially explosive atmosph
44. 78  89  Calibration method  54  Calibration tolerances  89  Code  91  Contrast  74  Date  76  Digital inputs  82  Display contrast  74  ELAN parameters  85  Electric time constants  68  Language  93  Limits  69  Magnetic field frequency  75  Measured value display  74  Noise suppression  68  Operating language  93  PROFIBUS parameters  101  Relay function  80  Setpoints  54  Status messages  71  Suppressing noise  88  Time  76   Software  7   Software release  7  48   Spans  35    152    Spare parts  115  Analyzer section  117  Electronics  118  Reference gas channel  123  Sample gas channel  121  Status messages  71  Submenu  45  Analyzer status  48  Calibration  50  Configuration  78  Measuring range  63  Parameters  67    T    Technical specifications  124  Associated gases  128  Climatic conditions  127  Gas inlet conditions  125  Inputs  126  Interfering gases  128  Measuring characteristics  125  Outputs  126  Power supply  124  Reference gas  127  Time response  125   Temperature compensation  99   Time  76   Time constants  68   Time response  125    U    User data memory  86    W  Warm restart  86    Z    Zero point  Calibration  51  Setpoint  54  Temperature compensation  99    OXYMAT 61  Operating Instructions  12 2011  A5E00123067 02    Siemens AG Subject to changes without prior notice  Industry Automation  IA  A5E00123067 02   Sensors and Communication O Siemens AG 2001  2011   Process Analytics   76181 KARLSRUHE    A5E00123067    www siemens com processautomati
45. A5E00123067 02    0 2 4 to 20 mA  isolated   Max  load 750 Q    6  with changeover contacts  can be freely  configured  e g  for measuring range  identification    Max  load  24 V AC DC   1 A floating   2  designed for 0 2 4 to 20 mA for external    pressure sensor and or correction of associated  gas influence  cross interference correction     6  designed for 24 V  isolated  freely configurable   e g  for measuring range switchover    RS485   Supplementary electronics with 8 additional  digital inputs and 8 additional relay outputs  e g   for triggering automatic calibration   supplementary electronics for PROFIBUS PA  and PROFIBUS DP    125    Appendix       Table 8 9 Technical specifications     climatic conditions    Climatic conditions   Permissible ambient temperature Storage and transport   40 to  70   C   operation   5 to  45   C   Permissible humidity  lt 90   RH 1 as annual mean  during storage and  transport 4    1  RH  relative humidity   2  Maximum accuracy is achieved after approximately 3 hours   3  Relative to sample gas pressure 1 000 hPa  absolute   sample gas flow 0 5   and ambient  temperature 25    4  Dew point low limit may not be violated    5  The error of the calibration gas must also be taken into account      Based on DIN EN 61207   IEC 1207    8 2 Reference gases    Table 8  10 Reference gases for OXYMAT 61                            Measuring range Recommended Reference gas connection pressure Remarks  reference gas  0 to 100 vol    O2 N2  4 6  t
46. AT 61  Operating Instructions  12 2011  A5E00123067 02 69    Operation       5 5 Parameters    5 5 3    70    On off functions  function 52     520n off config     Auto range    Stored function    continue         Figure 5 32 On off functions    With this function  you can easily switch other functions on off  for example  the ones listed in  this function screen     Thanks to this simplified operation  these functions do not have the longer paths through  various menus and submenus  For better orientation  the function numbers have also been  specified                 Activated functions are marked by      deactivated ones by       Using the fifth softkey       continue    jump to the next function screen with further functions     Function 52 is used to switch the following functions on and off     Table 5  3 Functions switched on off by function 52                                                             Description Fct  no  Remarks  Total calibration 23   AUTOCAL 24 Only with supplementary   electronics   Automatic measuring range switching 40   Limit monitoring 1 51   Limit monitoring 2 51   Limit monitoring 3 51   Limit monitoring 4 51   Suppression the display of negative measured values 55   Blocking of the logbook 60   Suppression the display of negative measured values 70   Fault   WA   FCTRL acc  to NAMUR 72   Measured value memory 77   Signal tolerance violation 78   Temperature recompensation of zero point 86   Temperature recompensation of span 86       OXYMA
47. H     Measuring range 2       Figure 5 25 Measuring range switching  type A    When the upper switchover point  OU  is exceeded  the next larger measuring range  available is selected  If the lower switchover point  UU  of the next smaller measuring range  available is fallen below  this is selected  The UU lies at 80    HystS  of the measuring  range  The OU lies at 90    HystE  of the measuring range     Type B     The full scale value must be greater than or equal to the full scale value following it  Since  the measuring spans must simultaneously become larger  the start of scale values of the  following measuring ranges become continuously smaller                    mbaOU uu mbe       SSV  Start of scale value   FSV Full of scale value   LS Low switchover point  select smaller measuring range  HS High switchover point  select larger measuring range    Figure 5 26 Measuring range type B    The following applies to measuring range switching     0  10  of span from start of scale value    Measuring range 1    Measuring range 2    Figure 5 27 Measuring range switching  type B    Operating Instructions  12 2011  A5E00123067 02 65    Operation       5 4 Measuring ranges    5 4 2    66       When the upper switchover point  OU  is fallen below  the next larger measuring range  available is selected  If the lower switchover point  UU  of the next smaller measuring range  available is exceeded  this is selected     The UU lies at 10    100     HystE  of the measuring range  The O
48. OXYMAT 61    Oxygen analyzer for standard applications    7MB2001    Operating instructions    8       OXYMAT 61       Continuous gas analysis    SIEMENS    SIEMENS    Continuous gas analysis    Oxygen analyzer for standard  applications  OXYMAT 61    Operating Instructions    7MB2001    12 2011    A5E00123067 02    General information       Description       Mounting and connecting       Commissioning       Operation       Maintenance and servicing       Spare parts list       Technical specifications       Dimension drawings       Appendix       ESD Directives       List of  abbreviations acronyms    O uw  gt   O    N O oa A  W N       Legal information    Warning notice system    This manual contains notices you have to observe in order to ensure your personal safety  as well as to prevent  damage to property  The notices referring to your personal safety are highlighted in the manual by a safety alert  symbol  notices referring only to property damage have no safety alert symbol  These notices shown below are  graded according to the degree of danger     indicates that death or severe personal injury will result if proper precautions are not taken           WARNING  indicates that death or severe personal injury may result if proper precautions are not taken                 CAUTION  with a safety alert symbol  indicates that minor personal injury can result if proper precautions are not taken              CAUTION  without a safety alert symbol  indicates that property dam
49. OXYMAT 61    The reference gas line is a restrictor  which decreases the incoming reference gas pressure  to allow a continuous flow of 5 to 10 ml min     The reference gas line is installed and removed as follows   1  Loosen the reference gas line at the nozzle and on the analyzer section   2  Remove the reference gas line     To install  carry out the above steps in reverse order        CAUTION    Clogged reference gas channel       It is important to ensure that no liquid or dust can enter the connection nozzles or the  reference gas line           The measurement function may deteriorate or the device may even fail as a result        Principle of operation  Page 16     Operating Instructions  12 2011  A5E00123067 02 105    Appendix       6 2 2    Disassembling the analyzer section    Removing the measuring head    Proceed as follows for this   1  Pull out the measuring head connector cable out of its socket     2  Loosen socket screws   Now you can remove the measuring head     To install  carry out the above steps in reverse order  It is important here to ensure that all O   rings are re inserted  In additional  the screen cup should never be removed from the  measuring head     Cleaning the sample chamber    106    The sample chamber is not particularly susceptible to faults  If condensation passes through  the analyzer as a result of a fault in the gas conditioning  although there is a temporary fault  in the measurement  highly oscillating values are displayed   the d
50. Output signal variation  lt  1 0 75   of the smallest possible measuring  range according to rating plate  with electronic  time constant of 1 s  corresponds to   0 25  at    20   Zero point drift  lt    0 5   month of respective span  Measured value drift  lt    0 5   month of respective span  Repeatability  lt   1  of respective span  Linearity deviation  lt   1  of respective span    OXYMAT 61  124 Operating Instructions  12 2011  A5E00123067 02    Appendix       Table 8 7 Technical specifications   influencing variables    Influencing variables 3   Ambient temperature    Sample gas pressure    Associated gases    Sample gas flow    Power supply     lt  0 5   10 K relative to smallest possible span  according to rating plate    With disabled pressure compensation     lt  2   of the current measuring range   1    pressure change   With enabled pressure compensation     lt  0 2   of the current measuring range   1    pressure change    Deviation from zero point corresponding to  paramagnetic or diamagnetic deviation of  accompanying gas  see table 3 2      lt  1   of smallest possible span according to  rating plate with 0 1   change in flow within the  permissible flow range     lt  0 1   of the current measuring range with rated  voltage   10     Table 8 8 Technical specifications     electric inputs and outputs    Electric inputs and outputs  Analog output    Relay outputs    Analog inputs    Digital inputs    Serial interface  Options    Operating Instructions  12 2011  
51. Suppressing noise    Spikes are created by electromagnetic interferences or sometimes by mechanical shocks   These faults can be suppressed by entering a  blanking time  of O to 5 s  The entered time  has the effect that the last measured value before the spike occurred is displayed so that the  measured result is not influenced     Times can be entered in steps of 0 1 s     Under  Level in   of smallest MR     enter the threshold value in   of the smallest measuring  range above which the noise signals are to be suppressed        Note  If a change in concentration directly follows a fault  this may be displayed after a delay     When this function is activated  the settings of function 50   Electric time constants   must be  taken into account  since these are executed first        OXYMAT 61  Operating Instructions  12 2011  A5E00123067 02    Operation       5 6 8    5 6 9       5 6 Configuration    Measured value memory  function 77     This function is used to define the response of the analog output or the digital interface when  the device assumes a specific state     77 Store MC    Analog out  to meas value M  Analog out  to 0 2 4 mA oO    Analog out  to 21 mA oO    Store on off O    Figure 5 53 Setting the analog output  In case of a fault  S  or CTRL  decoding  calibration  warm up phase   one of these values is  sent to the analog output     e The last registered measured value  or    e 0  2 4  mA  or    e 21mA       Memory on   m  activates the respective setting   
52. T 61  Operating Instructions  12 2011  A5E00123067 02    Appendix       Magnet circuit    A magnetic pole shoe is glued into each casing section  The cut strip wound core placed  upon it is held down by a tension spring  This method deflects the forces of the  magnetostriction away from the sample chamber     Sample chamber    The sample chamber consists of a 1 mm thick center plate in which the sample gas channel  is punched  and two 0 3 mm thick cover plates  which contain the openings for the supply of  the sample gas and the reference gas  Since the sample gas only comes into contact with  the chamber plates  the OXYMAT 61 can be used for numerous combinations of sample  gas  The other channels of the measuring system are purged by the reference gas  In the  fully assembled analyzer section  the sample chamber is installed between the two casing  sections     Measuring head    The microflow sensor of the measuring circuit is located in the measuring head  This is  installed in a thermostated aluminum block  The strong magnetic field required to produce a  sufficient measurement effect causes interference in the microflow sensor  The aluminum  block with the preamp electronics are therefore shielded to eliminate this effect     This screen is composed of    e Screen cup  and    e Restrictor plate This includes restrictors  the function of which is described in section 2 5     Both parts are made of a material with high permeability     Reference gas channel    See also    
53. T 61  Operating Instructions  12 2011  A5E00123067 02    Operation       5 5 Parameters    Besides the functions listed in the table  Functions switched on off by function 52   function  52 can also be used to address other service functions  These are restricted to service  engineers and are only visible when the service code is entered  code level 3      5 5 4 Status messages  function 53     53 Status messages    Display automat   calibration  CAL     Display stored  0     value    Displa   Timit YLIM   Displa   duc orange  AR   Display control  function  CTRL        Figure 5 33 Status messages    With this function  you can configure the status message display in the status line of the  measured value display  You can display a maximum of four different status messages  The  message CODE m o provides information on the current operating mode   Coded display                mode   Decoded display mode   and is therefore always displayed     The device activates the functional check  softkey 5 in the function display  when it    determines that the measured value was influenced    This is the case    e During the warm up phase   e After entry of a code  i e  switch to  Operator control mode      e During automatic calibration  AUTOCAL  or remote calibration    OXYMAT 61  Operating Instructions  12 2011  A5E00123067 02    71    Operation                                              5 5 Parameters   Table 5  4 Status messages   Status What appears in the display depends on fun
54. U lies at the start of the  measuring range  mba      Define measuring ranges  function 41     41 Define range    MR Start End  No     0 00  50   0 00 60     1  2  3 0 00 80   4 0 00    Figure 5 28 Define measuring ranges    You can define a maximum of four measuring ranges whose start of scale values are  allocated to the bottom value  0  2  4 mA  and whose full scale values are allocated to the  top value  20 mA  of the analog output     If the message  Measuring ranges not plausible   appears  this means that autoranging is  not possible     If the start of scale and full scale values are  0   the measuring range is deactivated        Note    If a start of scale value other than  0  is defined  you must read the section  Preparations for  start up   temperature influence         OXYMAT 61  Operating Instructions  12 2011  A5E00123067 02    Operation       5 5 Parameters    5 5 Parameters    Parameters  50 El  time constants    51 Limits    52 On off configuration    53 Status messages    continue       Figure 5 29 Parameters submenu    The parameters menu contains all functions which are required for configuring the device     After selecting the parameter functions in the main menu by pressing the fourth softkey    Parameters    the menu screen for the selection of parameter functions 50 to 53 appears   Pressing the fifth softkey      Continue  brings you to additional parameter functions     This menu is protected by the code of code level 1     OXYMAT 61  Operating Inst
55. a value for  Time  from AUTOCAL to AUTOCAL  cycle time   using subfunction  AUWTOCAL Check cycle  parameter         AUTOCAL Check start via digital input      If you previously configured a digital input using function 72  you can initiate AUTOCAL via a  digital input     The  AUTOCAL start via cycle parameter  and  AUTOCAL start via digital input    operating  modes can be activated simultaneously in order e g  to verify a weekly calibration and control  this check via a digital input     OXYMAT 61  56 Operating Instructions  12 2011  A5E00123067 02    Operation  5 3 Calibration        Trigger AUTOCAL once     In the  AUTOCAL on state  you can start an AUTOCAL sequence at any time using the  softkey  Trigger AUTOCAL once    providing the device is in the  Measure  operating mode  A  sequence initiated this way has no influence on the time cycle of an AUTOCAL  i e  the cycle  time continues to run independent of this     After the initial trigger  the dot disappears until the process is finished      Abort AUTOCAL     A running automatic calibration process can be exited at any time using the softkey  Abort  AUTOCAL   With this  all calibration data determined up to that point are discarded and the  calibration data used before AUTOCAL was started are used further  zero and span      Canceling has no influence on the time cycle  All valid calibration processes are retained     AUTOCAL sequence  The AUTOCAL sequence is configured in this submenu     Autocal  check sequence  Ze
56. age can result if proper precautions are not taken              NOTICE  indicates that an unintended result or situation can occur if the relevant information is not taken into account                 If more than one degree of danger is present  the warning notice representing the highest degree of danger will  be used  A notice warning of injury to persons with a safety alert symbol may also include a warning relating to  property damage     Qualified Personnel    The product system described in this documentation may be operated only by personnel qualified for the specific  task in accordance with the relevant documentation  in particular its warning notices and safety instructions   Qualified personnel are those who  based on their training and experience  are capable of identifying risks and  avoiding potential hazards when working with these products systems     Proper use of Siemens products    Trademarks    Note the following           WARNING    Siemens products may only be used for the applications described in the catalog and in the relevant technical  documentation  If products and components from other manufacturers are used  these must be recommended  or approved by Siemens  Proper transport  storage  installation  assembly  commissioning  operation and  maintenance are required to ensure that the products operate safely and without any problems  The permissible  ambient conditions must be complied with  The information in the relevant documentation must be obse
57. agram of the operating sequence with operating modes       Changes are  accepted             CAL McTRL Main menu Functions menu Functions       Diagnosis    THC  gt  E Calibration                            Configuration  gt    Oo  gt     papoo s   Jaz  jeuy          apo Induj  apou uonreledo                 Display mode    Query  only    following changes  CE    Accept L Measured value  modifications   lt  o  gt  memory for analog    Yes  gt  Functional check output  function 77   No  gt  relay  CTRL   function 71     only following changes       Figure 5 2 Operating sequence with operating modes    OXYMAT 61  42 Operating Instructions  12 2011  A5E00123067 02    Operation       5 1 3    5 1 4    OXYMAT 61       5 7 General information    Editing inputs    The values in the menus shown in this chapter are meant as examples     An active input field is shown with colons  e g   10   as a limiter  The cursor blinks under the  number to be entered     By pressing the  ENTER  button  you finish your input and the value is stored  If there are  several input fields on one function screen  the cursor positions itself at the next input field at  the same time        Note    Confirm every entered value  even the last of several values in a function  before exiting the  function with  ENTER         With the  CLEAR  button  you can clear a number which you have begun to enter  The  cursor then jumps back to the first position of the input field     Graphic symbols        activated  ON 
58. as calibration  15 minutes     Calibration with calibration gas 1  10 minutes     Purging with sample gas  8 minutes     Sample gas intermediate operation  30 minutes   Calibration with calibration gas 2  5 minutes   Calibration with calibration gas 3  8 minutes   Calibration with calibration gas 4  11 minutes    Purging with sample gas  8 minutes       ONOAAR ODN    Short term signaling contact in order to be able to start AUTOCAL on another device     The specified AUTOCAL sequence is shown in the following function displays   Autocal  check sequence    Zero gas 1  15 0  min      Cal  gas 1 10 0 min      SG purging 8 0 min      Int SG mode 30 0 min    continue    Figure 5 15 Example of AUTOCAL sequence    OXYMAT 61  Operating Instructions  12 2011  A5E00123067 02    Operation       OXYMAT 61                                                 5 3 Calibration   Autocal  check sequence   5  Cal  gas 2   5 0  min    6  Cal  gas 3   8 0  min    7  Cal  gas 4  11 0  min    8  Int SG mode  8 0  min    continue   Figure 5 16 Example of AUTOCAL sequence  Autocal  check sequence   9  Sig  cont  I  min    10       min    11  min    12  min    continue   Figure 5 17 Example of AUTOCAL sequence  List for the AUTOCAL sequence    Step AUTOCAL sequence  Zero gas 1 Function code 1  Zero gas 2 Function code 2  Calibration gas 1 Function code 3  Calibration gas 2 Function code 4  Calibration gas 3 Function code 5  Calibration gas 4 Function code 6  Flush sample gas Function code 7  Sample gas in
59. authorized access by codes     Commissioning    Totality of measures and actions required to make a machine or system capable of running     Containment system    Gas channel within the gas analyzer     Control panel    Panel with operating elements used to make inputs on the device     OXYMAT 61  Operating Instructions  12 2011  A5E00123067 02 141    Glossary       Correction of cross interference    Computed correction of the measured value falsification caused by the interfering gas     Cross interference deflection    The falsification of a measurement caused by an interfering gas     Cursor    Tool  insertion mark  writing mark  input mark  for identifying the current processing position  of a program     Decoded display mode    Operating mode in which the measured value is displayed  the device is partially or  completely decoded  and the functional check is active     Dialog language    Language in which communication between the user and device takes place     Display unit    Device component which outputs shows device information and visualizes the  communication via the control panel     Factory functions    Function for device maintenance  This function is protected by the highest code level and is  only accessible to maintenance personnel     Factory setting  Standard settings of the device at the time of delivery    Front panel  Front part of a device  usually with clear identification features  e g  device name   manufacturer logo  etc      Function    Numbered 
60. binet     e Nitrogen  oxygen  air  high pressure variant   Pay attention to the purity of the gas  4 6    The supply is here comes from a compressed gas cylinder with a pressure setting of  2000 to 4000 hPa over the sample gas pressure  A sintered metal frit  porous filter  is in  the coupling to prevent entry of dirt particles into the analyzer     OXYMAT 61  Operating Instructions  12 2011  A5E00123067 02 33    Commissioning       4 2 Preparation for commissioning    Introducing reference gas    Always introduce the reference gas before beginning the measurements  Reference gas  should always continue to flow even if the measurements are temporarily interrupted  The  increased consumption caused by this is negligible when compared to reference gas line is  sealed     Compressed gas cylinder    If the reference gas is taken from a compressed gas cylinder  the reference gas line must be  purged prior to commissioning  You must then check the line for leaks  because leakage  losses are often greater than the reference gas consumption  To do this  close the valve at  the compressed gas cylinder  If the pressure indicator at the reducer value of the gas  cylinder does not drop by more than 1000 hPa min  the gas connection is sufficiently sealed   The reference gas pressure must always be 2000 hPa or more above the sample gas  pressure     Checking the reference gas flow  To do this  proceed as follows   1  Close the sample gas inlet couplings     2  Lay a hose with an internal d
61. cnnoncccnnnoncnc canon cnc nano naan nano nn cc nan nn cc Edenia 54  5 3 4 Calibration setting  function 23   rrerak aA E A LRE R 54  5 3 5 AUTOCAL  function  24  02 tlf eiii ras aiii 55  5 3 6 Drift valles  unction 285            2 234  ltd ernennen 62  5 4 Measuring ranges ici iia A a a 63  5 4 1 Select measuring ranges  function 40         assssessrecssrsesrsrnssriransernnnannnnenntnnasnnnanntennnannannanennaananna 63  5 4 2 Define measuring ranges  function 41      oonnncccnnnnicinnnnacccnnnnnccnnnnnnno nan nnnnn cnn rnnnn rca rra rara rr 66  5 5 Parameters ive  cite ieee ets ciate N el oh es ati at ie ih dn A ae 67  5 5 1 Electric time constants  function 50      ooccocnnnnnccinnnoncccnnnonccnnnn nono nnnononn nn EREE arrancar nn rra 68  5 5 2 Limits  TUNCION S ae a elle kein ici 69  5 5 3 On offf  nctions  f  nelion  S2 ascii een sehen earmunlerrigennenndeenaden 70  5 5 4 Status  Messages  function incita da la rein 71  5 5 5 Graphical representation of measured values  function 54   ccccconnccccninococccononnncnanancnnnnnancnnnnnons 72  5 5 6 Measured value display  function 55                    444srnnnnnennnnnnnennnnnnnennnnnnnnnnnnnnnnnnnnnnnnnn nn mann 74  5 5 7 LCD  contrasti fUnctiON D6    curacion datada 74  5 5 8 Magnetic field frequency  function 57     ooooonnnncccnnnnanocinnnonccnnononnnnn nono nnn nano nan ran n rca rro rn rra 75  5 5 9 Date time   TUNCION 58  aci sr sia 76  5 5 10 Measuring point switching  function 9        oocconnncccnnnnaccccnnnoncccnn
62. composition       Operating hours   operating date                   Service report             RH no         Date received  Date returned  Technician           134    OXYMAT 61  Operating Instructions  12 2011  A5E00123067 02    ESD Directives B    B 1 ESD guidelines    Definition of ESD    All electronic modules are equipped with large scale integrated ICs or components  Due to  their design  these electronic elements are highly sensitive to overvoltage  and thus to any  electrostatic discharge     The electrostatic sensitive components modules are commonly referred to as ESD devices   This is also the international abbreviation for such devices     ESD modules are identified by the following symbol           CAUTION       ESD devices can be destroyed by voltages well below the threshold of human perception   These static voltages develop when you touch a component or electrical connection of a  device without having drained the static charges present on your body  The electrostatic  discharge current may lead to latent failure of a module  that is  this damage may not be  significant immediately  but in operation may cause malfunction           OXYMAT 61  Operating Instructions  12 2011  A5E00123067 02 135       ESD Directives       B 1 ESD guidelines    Electrostatic charging    Anyone who is not connected to the electrical potential of their surroundings can be  electrostatically charged     The figure below shows the maximum electrostatic voltage which may build up on a
63. ctions 52 and 53   Fet  53   Fet  52  0  Fct  52  m   ies Fct  53     Fct  53      Calibration  CAL None CAL Oo CAL m CAL  calibration running   also in Autocal    Measured value None  STO a STO   STO  analog output applied to  memory  STO memory  see function 77    Limit  LIM None LIM o LIM   LIM  limit has been violated  see   function 51    Automatic None  AR DAR   AR  measuring range switching  measuring range  actively possible   see function  switching  AR 40   Functional check  None O CTRL or m CTRL o CTRL m CTRL  device is in the  Warm   CTRL  functional check up phase  or in  Operator control   cannot be switched mode  or calibration is in progress  off using function 52    If an error occurs during operation  the message  Maintenance request    Fault  or   Measuring protection  appears in the status line  depending on the severity of the error  This  message is alternately displayed with the status messages    In addition  when communication is over the ELAN interface  the message  Remote  is  displayed   also alternating     with display of the device status    5 5 5 Graphical representation of measured values  function 54     With this function  you can follow the trend of the measured values for the last ten minutes or  24 hours on the display     54 G  signal display    Period 10 min    Period 24 h       Figure 5 34 Graphical measured value representation    OXYMAT 61  72 Operating Instructions  12 2011  A5E00123067 02    Operation       OXYMAT 61       5 5 Para
64. dance with these operating instructions              Qualified Personnel    Qualified personnel are people who are familiar with the installation  mounting   commissioning  and operation of the product  These people have the following qualifications     e They are authorized  trained or instructed in operating and maintaining devices and  systems according to the safety regulations for electrical circuits  high pressures and  aggressive as well as hazardous media     e For explosion proof devices  they are authorized  trained  or instructed in carrying out  work on electrical circuits for hazardous systems     e They are trained or instructed in maintenance and use of appropriate safety equipment  according to the safety regulations     Operating Instructions  12 2011  A5E00123067 02 9    General information       1 7 Notes on warranty    1 7    1 8    1 9    10    Notes on warranty    The contents of this manual shall not become part of or modify any prior or existing  agreement  commitment or legal relationship  The sales contract contains all obligations on  the part of Siemens as well as the complete and solely applicable warranty conditions  Any  statements regarding device versions described in the manual do not create new warranties  or modify the existing warranty     The content reflects the technical status at the time of publishing  Siemens reserves the right  to make technical changes in the course of further development     Delivery information    The respective 
65. dditional equipment such  as a wash bottle  additional filters and pressure reducers     Also provide for the removal of corrosive components or components that interfere with the  measurement process by using appropriate absorption filters     OXYMAT 61  Operating Instructions  12 2011  A5E00123067 02 21    Mounting and connecting       3 3 Connections                                           o       Gas sample probe  heated if required   Sample gas line  heated if required   Gas cooler  Filter trap  if required    Sample gas pump   Control value   Flow indicator  if device has no  sample gas monitoring     To sample gas input  Condensation drain                                  Figure 3 2 Example of gas preparation    3 3 1 5 Gas flow diagram          Figure 3 3 Gas path OXYMAT 61 with installed reference gas pump    OXYMAT 61  22 Operating Instructions  12 2011  A5E00123067 02    Mounting and connecting       3 3 Connections          E49          Figure 3 4  OXYMAT 61 gas channel  with reference gas connection 3 000 to 5 000 hPa absolute       Key for gas flow diagrams                                           1   Sample gas inlet 9   Purging gas  2   Sample gas outlet 10   Restrictor in reference gas channel  outlet   3 n c  11   Pressure switch for reference gas monitoring  4   Reference gas inlet 12   Pump  5   Restrictor in reference gas channel 13   Filter  6   Analyzer section 14   Flow indicator in sample gas channel  optional   7   Restrictor in sample gas channel
66. e  e g  for measurement range switchover  processing  of external signals from sample preparation     Can be extended by eight additional digital inputs and eight additional relay outputs for  automatic calibration with max  four calibration gases    RS485  standard    Optionally available      AK interface for the automotive industry with extended functions    Converter for RS232  USB and Ethernet   Integration in networks via PROFIBUS DP PA interface      SIPROM GA software as the service and maintenance tool    Operating Instructions  12 2011  A5E00123067 02 13    Description       2 2 Design       Status line for display of analyzer state  configurable    Two code levels according to NAMUR  maintenance and specialist levels    Display of activated measuring ranges   Five softkeys for menu control   ESC key for returning to previous input step  with application of any changes made  INFO key for help functions for the current menu   MEAS key for returning to measuring mode   CLEAR key to delete numerical inputs   ENTER key to accept numerical values    1  2  3  4  5  6  7  8    o    10 Numerical keypad for entering numerical values  11 Display of start of scale and full scale values  12 Display of concentrations as digits and bars    Figure 2 1 Membrane keyboard and graphic display    OXYMAT 61  14 Operating Instructions  12 2011  A5E00123067 02    Description  2 3 Communication interface       2 3 Communication interface         Pa Container A EN  j      Y  nr             
67. e causes Remedy  S8 Signal of selected pressure Sample gas blocked at outlet Caution  Q sensor outside tolerance   gt  2 000 hPa with internal pressure   The internal pressure sensor may be  sensor or  gt  3 000 hPa with external    destroyed if the system pressure  or system pressure too high exceeds 4 000 hPa   1  Perform a suitable action to  reduce the flow resistance at the  device inlet until  the sample gas is back  within the tolerance limits  2  Regulate the system pressure  accordingly  3 Check for leaks  see section 4 2    Preparation for commissioning    If leak is found  contact service  department  Sample gas pressure too low Set system pressure  gt  500 hPa    lt  500 hPa   S9 Signal too high Sample gas pressure  gt  3 000 hPa or   Reduce pressure or O2 concentration  Q Oz concentration in range or  2 000 to 3 000 hPa too high contact service department  S10 RAM check RAM PROM Replace motherboard or  Q contact service department   11 Reference gas supply failed Reference gas line leaks  is Check reference gas flow  see section  interrupted or blocked 4 2 Preparation for commissioning   Reference gas source is empty Connect new reference gas source   12 Mains power supply Mains voltage outside tolerance Line voltage must be within the  Q limits specified limits specified by the rating  plate   13 Hardware   line frequency Mains frequency outside tolerance Connect power supply stabilizer  Q limits  Crystal or external ADC defective Replace motherboard   14 Measur
68. e g  a screwdriver  to pry them out and removed them     109    Appendix       6 5 Cleaning the device    User Interface    The front panel is washable  We recommend using a damp sponge or cloth containing a mild  detergent for this  Yu should only apply slight pressure when cleaning  especially in the  display area  to avoid damaging the thin membrane  In addition  ensure no water penetrates  the interior of the device during cleaning     Interior  If necessary  you can gently remove dust and dirt from the interior using a compressed air  gun    6 6 Maintenance requests and fault messages    The device is able to detect functional irregularities  These appear  depending on the  severity  as either  Maintenance request  or  Fault  in the status line of the display  At the  same time  they are recorded in the logbook  function 3  and can also be called up there   Pressing the button next to the corresponding log entry acknowledges it  It will appear again   however  if you do not eliminate the cause     Each new message is written to the next free space in the logbook  A total of 32 storage  spaces are available  which means that the oldest of 32 log entries is overwritten by a new  entry  A power failure deletes all log entries     You can disable the log or delete the messages contained in it using function 60     The display of messages is especially distracting during testing  They can be disabled under  the function 87  You should not do this during normal operation  sinc
69. e you need to be able to  detect any faults that may occur     Maintenance requests    If there are indications of changes in internal device parameters   Maintenance request   appears in the status line of the display  Changes of this type do not affect the measurement  capability of the device or affect it only minimally the first time they appear  In order to  continue to guarantee the measuring capability  however  it may be necessary to carry out  remedial measures     If the relay output of the device was configured accordingly  see also section 5  Function 71    external signaling may also be performed     OXYMAT 61  110 Operating Instructions  12 2011  A5E00123067 02    Appendix                6 6 1 Maintenance requests  The following error messages result in a maintenance request  output in the display  and are  signaled externally if a corresponding relay has been configured using function 71   Function 87 can be used to switch off  disable  each maintenance request individually   Each maintenance requests labeled with  Q  in the left column of the following table must be  acknowledged   No  Message Possible causes Remedy Remarks  w1 Calibration Calibration gas was changed   Repeat calibration Calibration tolerance  see also  Q tolerance Drift characteristics Check that drift is normal function 78  l  exceeded Drift of device according to  technical specifications   Zero point  1  of FSV week   Span   gt 1  of FSV week  W2 Signal span to Reference gas contains too C
70. ed value greater than Sample gas pressure exceeds the Check the sample gas pressure and  operating full scale value    5      pressure correction range of 2 000 or   possibly reduce it or switch to an  3 000 hPa external pressure sensor with an  appropriate measuring range  Incorrect calibration of the measuring   Repeat calibration and possibly check  range the sample test  S 15 Calibration aborted Fault if the device is in AUTOCAL Remedy the cause  Q mode  Fault in calibration via digital input  S16 Sample gas flow is too low Ensure sufficient flow       Causes of fault messages    114    OXYMAT 61    Operating Instructions  12 2011  A5E00123067 02    Appendix       Spare parts list F      7 1 General information    This spare parts list corresponds to the technical status of February 2011     The rating plate shows the year of construction of the gas analyzer  coded in the production    no     SIEMENS  OXYMAT 61 C     MLFB no   7MB2001 0FA00 0AAO  F  Nr  N1 JD 352  Prod  no     230 V48  63Hz 70VA    min 5   O2 A    Siemens AG  Industry Sector  76181 Karlsruhe  German        An order for spare parts must contain the following information     Quantity  Designation of the spare part  Order no  of the spare part    Device name  order number  and production no  of the gas analyzer for which the spare  part is intended    Example for ordering    2 measuring heads  C79000 A3460 B525  for OXYMAT 61  MLFB 7MB2001 0FAO00 0AAO  Prod  no  N1 JD 352    OXYMAT 61    Operating Instructio
71. ement  one distinguishes between an input signal and an output  signal     Measured value    The measured value is an output value which reflects a determined variable     Measured value display    Totality of the displayed information in  Coded display mode  and  Decoded display mode    The following are displayed  for example  measured value  status line with status messages   footer  measuring ranges  components  etc     Measuring point    Location where a measured value is recorded     Measuring point relay    Relay to which a measuring point is allocated     Measuring range    Range within which the result of a measurement moves  The measuring range has a  reference number  e g  1  It is characterized by a start of scale value and a full scale value     Measuring range switching    Automatic switching over of measuring ranges  Also referred to as  autorange  in the  software     Membrane keyboard    Keyboard whose keys are protected against atmospheric influences by a membrane     Motherboard    Board which contains the basic data and firmware for the device     OXYMAT 61  144 Operating Instructions  12 2011  A5E00123067 02    Glossary       Noise    Operating mode    Operating mode    Operation level    Totality of all phenomena which could disturb the transmission and or recording of  information     Various modes into which the device can be put by intervening from the outside  Three  modes are distinguished     e Coded display mode  e Decoded display mode  e Operator
72. er of digits and the maximum  number of decimal places     Note that a maximum of five digits can be displayed  decimal point also counts as one digit      5 5 7 LCD contrast  function 56     74       With this function  you can make the display contrast brighter or darker     56 LCD contrast MC  brighter e    darker e    Basic setting e    Test O    Figure 5 38 Setting the LCD contrast    OXYMAT 61  Operating Instructions  12 2011  A5E00123067 02    Operation       5 5 8    OXYMAT 61    5 5 Parameters    If the contrast is misadjusted  you can restore the factory settings using the third softkey   Basic settings        With the fourth softkey    Zest     you can carry out an LCD test  Various test screens are then  displayed in succession     From  Coded decoded display mode   you can restore the basic setting by entering   8   8   8   8   ENTER      Magnetic field frequency  function 57     You can use this function  for example  to minimize vibration dependent frequencies  superimposed on the analog output  beat  or even eliminate them in the ideal case     57 Magn  field freq     Frequency   8 086  Hz  Default value 8 095 Hz E       Figure 5 39 Magnetic field frequency  To do this  enter the desired frequency value in the  Frequency  input field by calling function  57  The permitted values for this are between 7 to 11 Hz     Ifthe change to a specific frequency does not have the desired effect  enter different  frequency values     Pressing the fifth softkey resets clock
73. eres   ATEX Richtlinie des Europ  ischen Parlaments und des Rates zur Angleichung der Rechtsvorschriften der Mitgliedstaaten f  r    Ger  te und Schutzsysteme zur bestimmungsgem    en Verwendung in explosionsgef  hrdeten Bereichen     Karlsruhe  22 06 2009  Siemens AG        Markus Nj  klas  Wilhelm Eckhard  Research  amp  Development   Entwicklung Quality   Qualit  t   Namg  function 1 Funktion   Name  function   Funktion     Annex A is integral part of this declaration   Anhang A ist integraler Bestandteil dieser Erkl  rung     This declaration certifies the conformity to the specified directives but contains no assurance of properties  The safety documentation  accompanying the product shall be considered in detail    Diese Erkl  rung bescheinigt die   bereinstimmung mit den genannten Richtlinien  ist jedoch keine Beschaffenheits  oder Haltbarkeitsgarantie nach  443 BGB   Die Sicherheitshinweise der mitgelieferten Produktdokumentation sind zu beachten     Siemens Aktiengesellschaft  Chairman of the Supervisory Board  Gerhard Cromme  Managing Board  Peter Loescher  Chairman  President and Chief Executive Officer  Wolfgang Dehen  Heinrich Hiesinger   Joe Kaeser  Jim Reid Anderson  Hermann Requarat  Siegfried Russwurm  Peter Y  Solmssen  Registered offices  Berlin and Munich  Commercial registries  Berlin Charlottenburg  HRB 12300  Munich  HRB 6684  WEEE Reg  Nr  DE 23691322  Page 1 2    Operating Instructions  12 2011  A5E00123067 02 131    Appendix  A T EC Declaration of Co
74. es contained in the device at the time of delivery without optional  extensions     Start up state    Settable operating mode to which the device changes after the warm up phase has been  successfully run through     Status message    Selectable output of various messages in the status line of the measured value display     Subfunction    Independent functional unit within a function     Submenu    Menu which is listed under a higher order menu item     Time constant    System parameter which determines the way a system relevant variable changes with time     OXYMAT 61  146 Operating Instructions  12 2011  A5E00123067 02    Glossary       Warm restart    Restarting the device from its  warm  state with the last set parameters     Warm up phase    Time which the analyzer needs to reach operating temperature  The warm up phase counts  as one of the operating modes     Weighting    Evaluation of single factors of a solution approach with regard to their importance  It has the  effect that more relevant factors have a greater influence on the result     Zero calibration    Calibration of the zero point with a suitable zero gas     Zero gas   Gas used to calibrate the zero point   Zero point   Smallest possible point of a measuring range  usually the start of the measuring range    OXYMAT 61    Operating Instructions  12 2011  A5E00123067 02 147    Glossary       OXYMAT 61  148 Operating Instructions  12 2011  A5E00123067 02    Index    A    Access code  91  Add on board  108  Ana
75. evice is ready for  measurements again once the sample chamber has dried out  However  one of the inlets for  the reference gas may become clogged and measurement may fail if there is extremely high  contamination  This will be apparent from extremely strong fluctuations in the display of  measured values  If such situations  you need to clean the sample chamber as follows     1  Remove the measuring head  as described in the section  Removing the measuring  head        2  Blow compressed gas into the sample chamber to clean it  The compressed air passes  through the sample gas outlet and through the reference gas channels in the upper part  of the casing     You can also flush the sample chamber with trichloroethylene or alcohol   In such  situations  you need to dry it afterwards with a stream of dry air   3  Install the measuring head again     Ifthe cleaning procedure described above does not produce the desired result  the sample  chamber must be completely removed and cleaned in an ultrasonic bath  If this measure  does not succeed  you may need to replace the sample chamber     OXYMAT 61  Operating Instructions  12 2011  A5E00123067 02    Appendix       Removing the analyzer section    Proceed as follows for this     1  Pull out the measuring head connector cable out of its socket on the magnetic field  terminal board     2  Remove the reference gas inlet hose from the analyzer section   3  Disconnect the hose at a suitable location     4  Remove the analyzer section t
76. f cross interference   By pressing the first softkey you now select the type of cross interference correction  The  following distinctions are possible    e No correction of cross interference   e Correction of cross interference with constant influence of interfering gas   e Correction of cross interference with variable influence of interfering gas via analog input    e Correction of cross interference with variable influence of interfering gas via ELAN       Note    The correction of cross interference is deactivated for the duration of a calibration  process  zero or span   It is active again when calibration has been completed and the  device has returned to the  Measure  operating mode        Correction of cross interference with constant influence of interfering gas    83 Interf  correction      With const  interference  correction             Valid for range                Interference        Figure 5 62 Constant influence of interfering gas    OXYMAT 61  Operating Instructions  12 2011  A5E00123067 02 95    Operation       3 6 Configuration    With constant residual gas composition and low concentration of O2  a residual gas effect  occurs  which varies only by fluctuations in O2 content and can therefore be regarded as  approximately constant     The device must be informed of the value of the zero point offset  interfering gas swing   see  Example 1      Example 1     The residual gas  sample gas without O2  zero gas  is 50   propane  remainder is N2  N2 is  used a
77. f operation    In contrast to almost all other gases  oxygen is paramagnetic  This property is utilized as a  measuring effect by the OXYMAT 61 channel     Due to their paramagnetism  oxygen molecules in an inhomogeneous magnetic field always  move in the direction of increased field strength  When two gases with different oxygen  contents meet in a magnetic field  a pressure difference is produced between them     In the case of OXYMAT 61  one gas  1  is a reference gas  N2  O2 or air   the other is the  sample gas  5   The reference gas is introduced into the sample chamber  6  through two  channels  3   One of these reference gas streams meets the sample gas within the area of a  magnetic field  7   Because the two channels are connected  the generated pressure  difference creates a flow  This flow is converted into an electric signal by a microflow sensor   4   The pressure difference is proportional to the difference in the oxygen content of sample  and reference gas     The microflow sensor consists of two nickel plated grids  which are heated to approximately  120   C  Together with two supplementary resistors they form a Wheatstone bridge  The  pulsating flow results in a change in the resistance of the Ni grids  This leads to an offset in  the bridge which is proportional to the oxygen content of the sample gas     Because the microflow sensor is located in the reference gas stream  the measurement is  not influenced by the thermal conductivity  the specific heat or 
78. gure 5 10 Setting the setpoints    The function display shows the setpoint input with total calibration  The third measuring  range is chosen here as the leading measuring range     With single calibration  there is no choice for the leading measuring range     5 3 4 Calibration setting  function 23     23 Total single cal     Total range calibration       Figure 5 11 Select calibration method    With this function  select between total and single calibration of the measuring ranges     Total calibration means that you adjust one  leading measuring range  and all other  measuring ranges are calibrated using the switching ratio     If the total calibration is not activated  as shown here on the function display  you must  calibrate every measuring range individually     OXYMAT 61  54 Operating Instructions  12 2011  A5E00123067 02    Operation  5 3 Calibration       5 3 5 AUTOCAL  function 24        Note    You can only make use of automatic calibration  AUTOCAL  if your device contains  additional electronics  add on board   If it does not  a corresponding message will appear on  the display when the AUTOCAL function is called     Automatic calibration can only be started if the device is in the  Measure  operating mode        24 Autocal   check    Autocal  check mode    Autocal  check sequence    Autocal  check cyclic  parameters    Autocal check       Figure 5 12 AUTOCAL function       Note    The settings for  Acal Check      subfunctions 1 to 3  are valid both for AUTOCAL
79. hPa corresponds to one millibar  mbar    HS High switchover point   HU Height unit  1 HU correspondence to 1 3 4 inch  i e  44 45 millimeter   HystS Start point of the hysteresis   HystE End point of the hysteresis   Hz Hertz   i e  In other words   IEC International Electrotechnical Commission   IP Ingress Protection Degree of protection   K Kelvin   kg Kilogram   NP No parity   I Liter   LCD Liquid Crystal Display   LED Light Emitting Diode   LEL Lower explosion limit   LS Low switchover point       OXYMAT 61    Operating Instructions  12 2011  A5E00123067 02       List of abbreviations acronyms       OXYMAT 61    C 7 List of abbreviations                                                                                                                            Abbreviation   Meaning   m Meter   mA Milliampere   Mat  no  Material number   max  Maximum   min Minute   min  Minimum   MC Measured component   MLFB Machine readable product designation   mm Millimeter   MR Measuring range   MR Maintenance request   MS Span   N2 Nitrogen   NAMUR Standards working committee for measuring and control technology in the chemical  industry   NC No connection   NO Nitrogen oxide   No  Number   O2 Oxygen   OP Odd parity   PA Process Automation   PA Polyamide  a plastic  commercial name  e g  nylon   PC Personal computer   PROFIBUS   Process fieldbus   PTFE Polytetrafluoroethylene  a plastic  commercial name  e g  Teflon   Q Acknowledgment required   RAM Random access memory  memory with ra
80. hat the sample gas pressure in the analyzer  does not accumulate  When several devices are connected in series  you must ensure that  there are no restrictions in the gas channels of the downstream devices  free exhaust gas  flow   You must remove any sample gas restrictors installed in the device  The only    restrictor that can be retained is one between the sample gas line and the first gas analyzer  unit     After removing the sample gas restrictors  the sample gas monitoring  pressure switches   of the downstream analyzers will be inoperable  To avoid error messages  you must have  disable the corresponding error messages   Gas flow rate is too low   in the configuration  function of the software  section 5 6 18  Function 87  Page 101   Error S16   You should  also note in this case that the assignment of a relay to the  Sample gas flow  error  message has no function     If the sample gas flows to an exhaust gas line  you must note the following     e The flow resistance in the exhaust gas line should be kept as low as possible with a line  as short as possible or with a large diameter junction     e The exhaust gas line must be free of rapid pressure fluctuations  If this is not the case   either a special exhaust gas line must be installed  or a damping vessel   gt  1 I  with  restrictor must be installed between the device and exhaust gas line  pneumatic low  pass            OXYMAT 61    20 Operating Instructions  12 2011  A5E00123067 02       Mounting and connecting
81. he device allows you to perform calibration manually or automatically  function 24   The  latter is only possible with an add on board  which contains eight additional digital inputs and  eight relay outputs     The setpoints for the zero and span calibrations are entered using function 22   The drift values are set with function 25  the single or total calibration with function 23     Following the selection of functions 20 and 21  the required gases must be introduced  manually     OXYMAT 61  Operating Instructions  12 2011  A5E00123067 02    Operation  5 3 Calibration       5 3 1 Calibration of the zero point  function 20     20 Zero calib     Setpoint   0 00   viv    Act  value  15 60   v v    Start calibration    CANCEL       Zero calibration    In order to calibrate the zero  the device must be in the  Measure  operating mode  If this is  not the case  calibration is not possible  The display will show the message  Analyzer is not  measuring      The zero point is calibrated simultaneously for all measuring ranges  even when the span of  the measuring ranges is calibrated individually     Only initiate calibration when the measured value  actual value  has stabilized after applying  the zero gas  Otherwise  the calibration may be imprecise     If there is a lot of noise or the measured value is unstable over time  increase the time  constant before calibrating  function 50      OXYMAT 61  Operating Instructions  12 2011  A5E00123067 02 51    Operation       5 3 Calibrat
82. he housing must be closed and grounded upon completion of all assembly and  installation work     Failure to comply with this can result in death  personal injury and or damage to property           Awarn ING  Release of hazardous gases       Gases may be released if there are leaks in the sample gas channel  For this reason  toxic  and corrosive gases are not allowed to be be measured with this device     Failure to comply with this can result in death  personal injury as well as damage to  property and the environment              OXYMAT 61  Operating Instructions  12 2011  A5E00123067 02 31    Commissioning       4 2 Preparation for commissioning    4 2 Preparation for commissioning    4 2 1 General information    Gas preparation    Make all devices in the gas channel upstream from the analyzer ready for operation  e g    gas sampling devices  gas cooling device  condensate vessels  filters and any connected  controllers  recorders or indicators     Read carefully the instructions and information in the operation manuals of the devices     Operation    Before switching on the device  familiarize yourself with its operation  see section 5 of this  manual      Interfaces    Make sure that all interfaces are properly assigned and configured     Electrical connections    Make sure that all electrical connections have been correctly made  Read also the  information in the  Electric connection  section     OXYMAT 61  32 Operating Instructions  12 2011  A5E00123067 02    Commissi
83. he tag assigned to a device in the ELAN network  or the  first 16 characters thereof      e The configuration behavior of the device is configured with the   dent number   softkey 2    Parameters 0  1 and 3 can be selected and have the following meaning              Parameter Meaning   0 Only the PROFILE ID number is positively acknowledged   1 Only the device specific ID number is positively acknowledged   3 Only the PROFILE ID number for multivariable devices  complex analyzers  is  positively acknowledged                    Note    In order to work with the provided GSD and DD  the value 1 must be entered for  Ident  number         e Relay by PB  The 8 relays of the additional electronics  add on board  can be controlled  over PROFIBUS using this  To allow activation  none of these relays must be occupied by  a device internal function        Note    The parameter  Relay by PB  is only available with software version 2 0 0 and higher of  the PROFIBUS additional electronics  add on board         OXYMAT 61  102 Operating Instructions  12 2011  A5E00123067 02    Maintenance and servicing    6 1 Safety instructions         A WARNING       Work on an open device  Disconnect the gas and power supply beforeopening the device     Only carry out adjustments using appropriate tools  for example  to avoid short circuits on  the electronic PCBs     Leaks in the gas channel caused by incorrect installation or calibration may result in the  release of gas  which in turn could result in 
84. heck reference gas Zero gas and reference gas    Q high with zero    much oxygen       point adjustment    Zero gas contains too much  oxygen    Check zero gas    should be identical       Signal voltage  Q too low with    Zero gas contains too little  oxygen    Check calibration gas       span adjustment    Reference gas flow is too low    Check reference gas flow  and correct if necessary       An incorrect measuring range  has been selected for the  calibration    Select the correct  measuring range                                        W4 Set clock The device was switched off Reenter the date and time See function 58  Q  W6 Temperature of   Ambient temperature outside   Make sure that the ambient  Q LCD is too high   tolerances specified in the temperature is in the range  or too low technical specifications  5 to from 5 to 45   C  45   C   W7 Temperature of   Ambient temperature too high   Check ambient temperature  Q analyzer section    245   C   max  45   C   especially  with built in analyzers  Measuring head temperature   Contact service department  too high  278   C   only applies  to non heated model   With a heated sample No error   chamber  selected setpoint Wait until the analyzer  temperature is too low or section has cooled down to  heating is switched off  the new setpoint  Message W7 appears until the   temperature  new setpoint temperature has  been reached  OXYMAT 61    Operating Instructions  12 2011     A5E00123067 02    111    Appendix                 
85. iameter of 4 mm from the sample gas outlet nozzle into a  beaker filled with water     3  Observe the glass   The reference gas must be slow with 1 to 2 bubbles per second coming through the  water     Checking for leaks in the sample gas channel with hosed devices  To do this  proceed as follows   1  Close the reference gas connection   2  Establish a pressure of approx  100 hPa in the sample gas channel     3  Wait about 1 minute   During this time  the inflowing sample gas will adapt to the ambient temperature     4  Note the pressure   5  Wait another 15 minutes     6  Note the pressure again   The sample gas channel is sufficiently tight when the pressure has changed by no more  than 1 hPa  1 mbar  within these 15 minutes     OXYMAT 61  34 Operating Instructions  12 2011  A5E00123067 02    Commissioning  4 2 Preparation for commissioning       4 2 3 Initial commissioning    Switch on the power supply    After you have ensured that all connections have been made and there are no leaks  switch  on the device  After a short period  the measured value display appears on the operator  display  The status display is located above this in the top row  for details on the operator  display see section 5 1      The measuring head is in the warm up phase during the first five minutes  During this time   the display signals CTRL  function check      Setting the measuring span    Specify the desired spans  measuring range full scale value   measuring range start of scale  value  with
86. imensions  analyzer  See Section  Dimensional drawing   Weight  analyzer  Approx  13 kg    Table 8  2 Technical specifications     power supply    Power supply  Auxiliary power 100 to 120 V AC  rated range of use 90 V to   see rating plate  132 V   48 to 63 Hz or  200 to 240 V AC  rated range of use 180 V to  264 V   48 to 63 Hz  Power consumption  analyzer  Approx  37 VA  Fuse ratings 100     120 V 1T 250    200     240 V 0 63T 250    OXYMAT 61  Operating Instructions  12 2011  A5E00123067 02 123    Appendix       Table 8 3 Technical specifications     gas inlet conditions    Gas inlet conditions    Permissible sample gas pressure 800 to 1 200 hPa absolute   with external reference gas supply    atm  pressure range   with built in pump     Sample gas flow 18 to 60 I h  0 3 to 1 l min   Sample gas temperature 0 to 50   C  Sample gas moisture content  lt  90   RH       Table 8 4 Technical specifications   time response    Time response    Warmup time At room temperature   lt  30 min 2   Display response time  too  3 58  Damping  electronic time constant  0 to 100 s  can be configured    Dead time  purging time of the gas channel inthe Approx  0 5 to 2 5 s  depending on model  device at 1 l min     Time for signal processing in the device  lt 1s    Table 8 5 Technical specifications     pressure correction range    Pressure correction range  Pressure sensor 500 to 2 000 hPa  abs      Table 8 6 Technical specifications   measuring characteristics    Measuring characteristics    
87. ing and connecting       3 3 2 2    OXYMAT 61    3 3 Connections    Connection of the signal cables       AR WARNING       Inadequate signal lines    If signals  e g  4 to 20 mA analog output  are to be introduced into a hazardous area of  zone 1  they must be intrinsically safe  Additional upgrading or retrofitting of the device with  energy limiting modules is also required in this case     Failure to comply with this can result in death  serious personal injury and or damage to          property        The Ex marking of these energy limiting modules must be clearly visibly on the device        AL WARNING    Only connect the signal lines to devices which have reliable electric isolation from their  power supply                 e The connection lines to the relay outputs  the digital inputs  the analog inputs and the  analog outputs must be shielded     e The reference ground of the analog inputs is the potential of the housing     e The analog output is floating     Operating Instructions  12 2011  A5E00123067 02 25    Mounting and connecting       3 3 Connections    e Asa measure to suppress sparking across the relay contacts  e g  limit relays   RC  elements must be connected as shown in the following figure  Note that the RC element  results in a drop out delay for an inductive component  e g  solenoid valve   The RC  element should be sized according to the following rule of thumb    R Q  02xRL Q   C pF  I A   You must also use a non polarized capacitor for the RC eleme
88. injury to personnel  e g  poisoning symptoms    as well as damage to the equipment itself  e g  danger of corrosion               Maintenance    The device must be subjected to annual maintenance to verify its electrical safety and  correct operation  in particular the tightness of the sample gas channel  containment  system      The operator can extend the maintenance interval in individual cases if no negative effects  are expected with respect to chemical corrosion to the seals that come in contact with the  sample gas     To allow maintenance work  the top cover can be removed and the front panel swung  forwards  If maintenance work is interrupted for more than two hours  the analyzer must be  closed again     Service    All modules within the housing can be replaced in the event of a defect  Repairs to these  modules are not envisaged     OXYMAT 61  Operating Instructions  12 2011  A5E00123067 02 103    Appendix       6 2 Analyzer section  6 2 1 Design of the analyzer section  Analyzer section    104    The analyzer section consists of the functional parts magnetic circuit  sample chamber and  the measuring head                                         Measuring head  Restrictor plate   O ring measuring head  Pressure sensor  Electrical terminal board  Retaining plate   Cut strip wound core  Tension spring                   OLISI    Coil   Mounting screws   Bottom part of casing  Sample chamber   O ring sample chamber  Top part of casing  Reference gas connection    OXYMA
89. ion    5 3 2 Span calibration  function 21     A single or total calibration is carried out depending on the setting made with function 23   This function is specific for the component     Single calibration   Measure  operating mode required     21 Span calib     Calibrate    Calibrate    Calibrate    Calibrate       Figure 5 7 Carry out single calibration  Single calibration is only possible if the  7ofa  calibration subfunction has been deactivated  in function 23     The display will show then number of measuring ranges that were set using function 41  The  function screen above is therefore an example of the single calibration of four measuring  ranges     If you would like to calibrate measuring range 3  for example  press the corresponding  softkey     The setpoint and current actual value of the selected measuring range then appear in the  display     21 Sp  cal  MR 3  Setpoint   80 00   viv    Act  value  79 79   viv    Start calibration    CANCEL       Figure 5 8 Calibrating the measuring range  single calibration     OXYMAT 61  52 Operating Instructions  12 2011  A5E00123067 02    Operation       OXYMAT 61       5 3 Calibration    When the actual value has stabilized  you can initiate the calibration process by pressing the  fourth softkey  The actual value is now made to agree with the setpoint     If you make a mistake during the calibration  e g  wrong calibration gas   reload the original  calibration by pressing the  Cancel calibration  softkey   Total calibra
90. log output  78  89  Analyzer section  104  107  Design  104  Measuring head  105  Sample chamber  105  Spare parts  117  Analyzer status  48  Application examples  37  Area of application  11  Associated gas  94  128  AUTOCAL  Check  56  Connection  30  Cycle  60  Function  55  Operating mode  56  Sequence  57    B    Board  Replacing  108    C    Calibration  107  AUTOCAL  55  Automatic calibration  55  Calibration tolerances  89  Measuring range  52  Sensitivity  52  Setpoints  54  Single calibration  52  54  Total calibration  53  54  Zero point  51   Cleaning  Front panel  110  Interior  110  Sample chamber  106   Climatic conditions  127   Code  91    OXYMAT 61  Operating Instructions  12 2011  A5E00123067 02    Commissioning  Initial commissioning  35  Preparations  32  Safety instructions  18  31  Communication interface  15  Configuration  78  Analog output  78  Correction of cross interference  94  Device test  91  Digital inputs  82  ELAN  85  Pressure correction  93  Relay assignment  80  Temperature compensation  99  User data memory  86  Connection  Electrical connections  20  Gas connections  20  Mains connection  25  Reference gas  22  33  Sample gas  21  Contrast  74  Control keys  14  40  Control panel  39    D    Date  76  Device parameters  48  Device test  91  Dialog language  93  Digital inputs  82   See Digital inputs   Dimension drawings  130  Display  Contrast  74  Measured value  39  43  74  Display field  14  39  Display unit  14  39  72  Contrast  7
91. lp of this function   84 Phase adjust 0  144349  9  231 2   Meas  value 20 95    Phase adjust  Figure 5 65 Phase adjustment  With the maximum possible signal  sample gas  e g  air  the gain is automatically adjusted  so that V assumes a value of approximately 500 000  The phase angle   is then calculated  using this signal value and saved  whereby V is maximized and W minimized   This angle was determined at the factory and should only be readjusted when the magnetic  field frequency is changed   OXYMAT 61  98 Operating Instructions  12 2011  A5E00123067 02    Operation       5 6 16    5 6 17    OXYMAT 61       5 6 Configuration    Switch valves  function 85     With this function  you can manually switch up to six valves  This is done via the relays which  are allocated to the individual valves and are available on the motherboard and add on  board  The precondition is that the corresponding relays have been previously configured  with function 71   Relay outputs      85 Switch valves    01 Zero gas       Figure 5 66 Switching valves  The  Switch valves  function only applies to the relay functions  Zero gas    Calibration gas 1  to 4  and  Sample gas      Only one of the six valves can be switched at a time  since the valves are mutually  interlocked     Linear temperature compensation  function 86     The OXYMAT 61 is temperature compensated for sensitivity  If an additional temperature  error occurs during operation  for instance  due to light contamination of the sample 
92. maller than or equal to    Equal to    Not equal to    Congruent    Plus minus    Approximately equal  ES Inch  1 inch corresponds to 25 4 mm     Percent  1 percent corresponds to the hundredth part of the whole   vol    Volume percent  R Degrees  angular     C Degrees Celsius    F Degrees Fahrenheit  A Amperes  AC Alternating current  AC Automobile channel  Approx  Approximately  ASCII American Standard Code for Information Interchange  AUTOCAL Automatic calibration  CHa Methane  cm Centimeter  co Carbon monoxide  CO2 Carbon dioxide  D Slow blow  fuse values   DB Data bits  DC Direct current  DD Device Description  DIN Deutsche Industrie Norm  German industrial standard   DP Decentralized periphery  a PROFIBUS version  DS Pressure switch       OXYMAT 61    Operating Instructions  12 2011  A5E00123067 02 137    List of abbreviations acronyms       C 7 List of abbreviations                                                                                                             Abbreviation   Meaning   e g  For example   EEPROM Electrically erasable programmable read only memory  ESD Electrostatic sensitive device   ELAN Economical Local Area Network   EN European standard   EP Even parity   EPROM Erasable Programmable Read only Memory   ETX End of text   Fet  Function   FCTRL Function check   FKM Fluoroelastomer  name according to DIN ISO 1629   FSV Full of scale value   GA Gas analyzers   GND Ground   GSD Device master data   h Hour  from lat   hora     hPa Hectopascal  1 
93. meters    Select the desired time period with softkey 1 or 2     The device now graphs the measured value vs  time       display 10 min MC      viv Parameter      min0 1234567890    Figure 5 35 Measured value trend    The most recent measured value is on the time axis at the left att  0   Select softkey 1  Parameters    Here  assign a certain measuring range to the measured value axis   Meas  value disp  par   Optimum meas  val  dis   Range 1  Range 2    Range 3    Range 4       Figure 5 36 Parameters for measured value representation    If you select  Optimum measured value display     the software automatically scales the  measured value axis  The device adapts the scale to the measured value scatter     Operating Instructions  12 2011  A5E00123067 02 73    Operation       5 5 Parameters    5 5 6    Measured value display  function 55     With this function  you can configure the display of measured values     55 Select digits    Suppress negative  measured values    Automatic    Total digits    Digits after 3  decimal point    The decimal point counts  as a digit       Figure 5 37 Configuring the measured value display    You have the following options for this   e With softkey 1  you suppress the display of negative measured values     e With softkey 2  Automatic   you can activate the automatic display of the measured value  with 5 digits  The number of decimal places depends on the size of the measured value     e With the softkeys 3 and 4  you can choose the total numb
94. n polyethylene  foil at least 0 2 mm thick and a drying agent  e g  silica gel  should be added  In addition  the  transport container should be lined with a layer of kraft paper     Photocopy the return delivery form  fill it in and enclose it with the returned device     In case of guarantee claim  please enclose your guarantee card     Return delivery addresses    OXYMAT 61    Spare parts service  Send your orders for spare parts to the following address     SIEMENS SPA   CSC   Tel    00333 69066677  Fax   00333 69066688   1  chemin de la Sandlach  F 67506 Haguenau  France      DP order form recipient  0011E    Repairs    To enable us to quickly identify and correct faults  return the analyzers to the following  address  applicable until further notice      SIEMENS SPA   CSC   Tel    00333 69066677  Fax   00333 69066688   1  chemin de la Sandlach  F 67506 Haguenau  France      DP order form recipient  0011E    Operating Instructions  12 2011  A5E00123067 02 133    Appendix       A 2 Return delivery    Returned deliveries form        Repair        Guarantee       Customer name       Person responsible       Delivery address       Phone       Fax       E mail       Return address   if different from  address above        Customer  original   Order number       Device name       MLFB no        Production no        Description of returned  part       Fault description       Process data at  location of use       Operating temperature       Operating pressure       Sample gas  
95. nancccnnnnncc canon cnc naar n cnc naar RE T 77  5 5 11 Logbook settings  function GO iesire a a RE 77  OXYMAT 61   Operating Instructions  12 2011  A5E00123067 02    Table of contents       5 6 Configuration  een dei onen dia ida 78  5 6 1 Analog output  function TOV iaa en a E E E 78  5 6 2 Relay outp  lts  MICHON TT  raae A E E EA E O 80  5 6 3 Binaryinputs  fUNGUON  72  area EE E TAS 82  5 6 4 ELAN configuration  function Z3 jeo a E naar nennen 85  5 6 5 Reset   unction TDi A TERO ER 86  5 6 6 Save  load data  function  9     irna A a E A 86  5 6 7 Suppression of short noise signals  function 76      ooo nncccinnnniccnnnnococcnnnaccccnnnarccnnn ronca cnn 88  5 6 8 Measured value memory  function 77     oooooccccnnnnccccnonaccccnonancccnnnanccc nano ncnnnnnn cnc r nana nc cc naar nccr nana nncnnns 89  5 6 9 Calibration tolerances  function 78     oooooocnnnncccnnnnanoccnnnoncccnnnorccnnnnrrr anea E AE m i a aaki E RE 89  5 6 10 Change codes  function 79             u   2444snnnnnnnennnnnnnennnnnnnennnnnnnennnnnnnennnnnnnennnnnnnnnnnnnnnnnnnnnnnnnn nn 91  SoTL      Device test  funetion GO  iia 91  5 6 12 Language selection  function 81                    uusersnnesennnnnnennnnnnnennnnnnennnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnn nenn 93  5 6 13 Pressure correction  function 82            uuersnneennnnnnnnennnnnnnnnnnnnnnennnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnsnnnnn nn 93  5 6 14 Correction of cross interference  function 83                 24surnnnennennnnnnnennnnnnnnnnnnnnnennnnnnnn
96. ndom  direct  access    RC R   resistor  C   capacitor   RD Received data   R H   RH Relative humidity   RS232 Serial interface for data communication   RS485 Serial interface for high speed data transfer   s Second   SB Stop bit s    SG Sample gas   SO2 Sulfur dioxide   SSV Start of scale value   STX Start of text   SUB D Subminiature D  form a plug system for data connections   SW Software             Operating Instructions  12 2011  A5E00123067 02    139    List of abbreviations acronyms       C 7 List of abbreviations    140       Abbreviation  t    Meaning  Time  from lat   tempus                                          TD Transmitted data  USB Universal Serial Bus  V Volts   V  Version   VA Volt ampere   A Difference   pA Microampere   yF Microfarad    0  Phase angle   Q Ohm       OXYMAT 61  Operating Instructions  12 2011  A5E00123067 02       Glossary    Additional electronics    Generic term for all add on boards     Back wall of housing    Part for sealing the rear housing opening of the device     Calibration    Elimination of deviations between the setpoint and actual value of certain measured  variables     Calibration gas    Gas used for calibration procedures     Code    A selectable sequence of characters for enabling a protected submenu     Code level    Area of protected functions or states which are enabled after entering a certain code     Coded display mode    Operating mode in which the measured value is displayed and the device is protected  against un
97. next function screen  thus branching to  further relays        NOTICE    Changing the configuration       Every change to the configuration of the relay assignments should always be saved in the  user data memory by means of function 75     If you neglect to do this  there is a danger that a previous  unwanted  configuration will be  loaded by  Load user data   function 75            OXYMAT 61  Operating Instructions  12 2011  A5E00123067 02       Operation                                                    5 6 Configuration  Table 5  6 Relay outputs  possible functions   Function Relay is passive when   Relay conducts Remark  current  Not assigned Relay is permanently  passive  zero current   Fault Fault Also shown in the status  line of the measured value  display  Maintenance request Maintenance request  Calibration Calibration running   For identification purposes  Measuring range 1 to 4 Measuring range 1 Measuring range  to 4 on identification  Limit 1 to 4 Limit 1  to 4  triggered Limit monitoring  Functional check Functional check Decoding  Signal when    CTRL  active  e g   when warm up phase  e Device is decoded  changing to  Operator   Autocal running e Warm up phase  control mode   in the  approx  30 min    warm up phase  when e Calibration running  Autocal is running   Autocal   when calibration is e Remote  running   Sample gas Supply of sample  gas Valve actuation for Autocal  Zero gas 1 Supply of zero gas  Calibration gas 1 to 4 Supply of calibration  gas  Meas
98. nformity          IENS    Annex A to the EC Declaration of Conformity  Anhang A zur EG Konformit  tserkl  rung    No  A5E00484554   03    Product description  Gas analyzer   Gasanalysator  Produktbezeichnung OXYMAT 61  OXYMAT 6 E   F  OXYMAT 64 E  Type   Typ  7MB2001 xxA00 xxxx Z xxx  7MB2011 xax0x bcdx Z e f xxx 7MB2017 xax0c bcdx Z e f xxx  7MB2021 xxxx0 xxxx Z g xxx 7MB2027 xxxx0 bxxx Z g xxX    7MB2041 xxxxx Xxxx Z HXXX  a  b  c  d  e  f  g  x  see tables page 2   siehe Tabellen Blatt 2       Conformity to the Directives indicated on page 1 is assured through the application of the following  standards  depending on versions     Die Konformit  t mit den auf Blatt 1 angef  hrten Richtlinien wird nachgewiesen durch die Einhaltung  folgender Normen  variantenabh  ngig                                                                                                                                     Directive Standard   Reference number   Edition f E31    a A C E   a A C E   a A C E   a A C E   a A C E  Richtlinie Norm   Referenznummer Aus  x 0 9   E32 E33  b 0 1 b 0 1 b 0 1 b 6 7 b 2 3  gabe    or oder   E38 E39   c B c B c B c B c B  datum A  Z g E30   d A B G   d A B G   d A B G   d A B G   d A B G  e E40   e E11   e E12  2004 108 EC_  EN 61326 1   x X X X X x X  2006 95 EC__  EN 61010 1 2001 X X X X X X X  94 9 EC EN 50014 1997 X X  94 9 EC EN 60079 0 2006 x  94 9 EC EN 50016 1995 x x  94 9 EC EN 60079 2 2004 x  94 9 EC EN 50020 1994 X X  94 9 EC EN 60079 11 X  94 9 EC EN 
99. nnnnnnn mann 94  5 6 15 Phase adjustment  function 84  sesters a EE E A A RARR 98  5 6 16 Switch valves  function 85  ya ai E AS E 99  5 6 17 Linear temperature compensation  function 86        oooconncccinonncccccnnnonacnnononancnn nono nan n nono ncn rca nara 99  9 6 18  gt    Error On Off  FUNCIONO  as 101  5 6 19 PROFIBUS configuration  function 9       ooonnnccccnnnociccnnnocccccononcncnnnoncncnnno cnn c nano nn nr nnno rca r rana mann 101  6 Maintenance and ServiciOQ      ooooonccnnccnocononanonononoonnncnnnnnnnnnnn ccoo cnn nnnnn nono nono nnnnn nn corran nn n nn nnmnnn 103  6 1 Safely MASCAR 103  6 2 Analyzer section ci tdo td 104  6 2 1 Design of the analyzer SECtION     ooooonnncccnnnnnniccnnnoccccnononcccnn narco ca ronca rca rr rr 104  6 2 2 Disassembling the analyzer section     ooooonnncccnnnniciccnnnanccnnnoncccnn corn ccnn acc rra 106  6 2 3 Removing the sample gas restrictor    oooooonnccccnnncicinonoccccnnnorccannnorr cnn nn cnnn nr rca 108  6 3 Replacement of motherboard and add on board                        44usrsssssnennennnnennnnnnnnnnennennnnen 108  6 4 Replacing USGS coccion cir ii data 109  6 5 Cleaning the device sucre ti rbd 110  6 6 Maintenance requests and fault messages    oocoocccccnnocicicononoccnnnnnncccnn non n cana no noc nn rra 110  6 6 1 MEE 111  6 6 2 Eaults ri Aa tad 113  7 Spareparts list aoia A A E E AAE E EA AE AAE aE E 115  7 1 General information                  nenn anr denen 115  7 2 Analyzer SOCUOM il laicidad 117  7 3 Electro
100. ns  12 2011  A5E00123067 02 115    Appendix       Overview of the modules                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                Analyzer section    01    Electronics    02    Sample gas channel    03    Reference gas channel    04    OXYMAT 61 modules    Figure 7 1    OXYMAT 61    Operating Instructions  12 2011  A5E00123067 02    116    Appendix       7 2 Analyzer section                                                                                                                                                                                        Table 7  1  Part no    Designation Order no  Remarks  01 1 Analyzer section complete  material no  C79451 A3460 B31  1 4571   01 2 Measuring head C79451 A3460 B525   01 2 1 O ring C79121  Z100 A32 1 item  01 3 Sample chamber  material no  1 4571 C79451  A3277 B535   01 4 Magnet terminal board C79451 A3474 B606   01 5 O ring  FPM  Viton  C71121 Z100 A159 1 item                      OXYMAT 61  analyzer section    OXYMAT 61  Operating Instructions  12
101. nt     Connector SUB D 25F                Power supply      24 V max     vay                R  0    0 2 x R   0   C  HF    1   A     Da  a      a  O      e  O      O  6      a  25         o    pS  Figure 3 5 Example of spark suppression on a relay contact    When operated with direct current  a spark suppression diode can be installed instead of the  RC element     e Connect the signal lines to the SUB D plug connectors at the rear of the device     e Refer to the ELAN interface description  Order No  C79000 B5200 C176 German   C79000 B5276 C176 English  for details on the interface cable     OXYMAT 61  Operating Instructions  12 2011  A5E00123067 02    Mounting and connecting       3 3 2 3 Pin assignments    Connector SUB D 9F  RS 485     GND    GND R_Level N     NC  RD TD N  RD TD P   5V R_Level P   NC    NC       Connector SUB D 15F       NC  NC       Digital inputs  Analog inputs   analog outputs  z       Analog output 1 P  Analog output 1 N    Analog input 2 P  Analog input 2 N    3 3 Connections    Terminating resistors for the bus can be  connected to pins 7 and 9    Analog outputs floating   mutually isolated as well   R    lt  750 Q    Correction of     compressed or  interfering gas    Analog inputs  non floating  O to       Analog input 1 P i 20 mA 500 Q or  Analog input 1 N   Correction of 0 to 10 V  low  Digital input 6 P interfering gas impedance   Digital input 5 P  Digital input 5 to 6 N  GND  Connector SUB D 25F  A A GND  AAA Digital input 4 P Digital inputs 
102. o 100 vol    O2  suppressed The reference gas flow is set  zero point with measuring O2 2 000 to 4 000 hPa above sample gas   automatically to 5 to 10  range pressure ml min  full scale value 100 vol     max  5 000 hPa absolute   O2   Around 21   O2   suppressed zero with 21 Air Atmospheric pressure with internal  vol    O2 within the span  reference gas pump       OXYMAT 61  126 Operating Instructions  12 2011  A5E00123067 02    Appendix       8 3 Zero point error    Table 8  11 Zero point errors of different accompanying gases                                                                               Associated gas  concentration 100 vol     Zero point Associated gas Zero point  deviation in  concentration 100 vol     deviation in vol   vol    absolute   absolute   Organic gases Inert gases   Ethyne  acetylene  C2H2   0 29 Argon Ar   0 25   1 2 butadiene C4H6   0 65 Helium He   0 33   1 3 butadiene CaHe   0 49 Krypton Kr   0 55   n butane CaH10   1 26 Neon Ne   0 17   Isobutane CaH10   1 30 Xenon Xe   1 05   1 butene C4Hs   0 96 Inorganic gases   Isobutene C4H8   1 06 Ammonia NH3  0 20   Dichlorodifluoromethane CCl2F2  1 32 Hydrogen bromide HBr  0 76    R12    Ethanoic acid CH3COOH   0 64 Chlorine Cle   0 94   n heptane C7Hi6   2 40 Hydrogen chloride HCl   0 35   n hexane CeH14   2 02 Dinitrogen monoxide N20   0 23   Cyclo hexane C  H12   1 84 Hydrogen fluoride HF   0 10   Methane CHa   0 18 Hydrogen iodide HI   1 19   Methanol CH30H   0 31 Carbon dioxide CO2   0 30   n 
103. octane CeH18   2 78 Carbon monoxide co   0 07   n pentane CsH12   1 68 Oxygen O2  100   Isopentane CsH12   1 49 Nitric oxide NO  42 94   Propane C3H8   0 87 Nitrogen N2 0 00   Propylene C3He   0 64 Nitrogen dioxide NO2  20 00   Trichlorofluoromethane  R11  CClaF  1 63 Sulfur dioxide SO2   0 20   Vinyl chloride C2H3Cl   0 77 Sulfur hexafluoride SFe   1 05   Vinyl fluoride C2H3F   0 55 Hydrogen sulfide H2S   0 44   1 1 vinylidene chloride C2H2Cle   1 22 Water H20   0 03   Hydrogen H2   0 26                               Zero point error due to diamagnetism or paramagnetism of some residual gases with nitrogen as the reference gas at 60   C  and 1 000 hPa absolute  according to IEC 1207 3     Conversion to other temperatures    For this  multiply the deviations from the zero point listed in Table 3 2 by a correction factor   k     e For diamagnetic  gases  k   333 K   0    C    273 K    e For paramagnetic gases  k    333 K   8     C    273 K         All diamagnetic gases have a negative deviation from zero point    OXYMAT 61  Operating Instructions  12 2011  A5E00123067 02 127    Appendix       8 4    128    Materials used in the sample gas channel                                                             Standard   Gas channel 19  rack unit   With hoses Procedure 1 4571  Hose FPM  Viton   Sample chamber 1 4571  Nozzles for sample chamber 1 4571  Restrictor PTFE  Teflon   O rings FPM  Viton   Hose coupling Polyamide 6   Options   Flow indicator Measuring tube Duran glass  Floa
104. of 5 to 45   C is maintained during  operation  see Technical Specification section    Magnetic fields    Magnetically sensitive devices should not be installed in close proximity of OXYMAT 61  since it is inherently emits stray magnetic fields  Depending on the sensitivity  distances up  to 50 cm are required  see also Function 57      OXYMAT 61  18 Operating Instructions  12 2011  A5E00123067 02    Mounting and connecting       3 3 Connections    3 3 Connections    15 pin connector  Digital 9 pin  inputs and analog inputs connector   and outputs RS485    Reference gas inlet    Sample gas outlet Sample gas inlet    Purge gas  inlet          9 pin   interface connector   optional     e g  PROFIBUS       Power supply  onnection  and miniature fuses             37 pin connector  Digital  inputs and relay outputs    tional  25 pin connector  Digital  optional     Gas connections  Nozzle 6 mm or 1 4  inputs and relay outputs    Figure 3 1 Back side with gas and electrical connections    OXYMAT 61  Operating Instructions  12 2011  A5E00123067 02 19    Mounting and connecting       3 3 Connections    3 3 1 Gas connections and internal gas flow diagram    3 3 1 1 Sample gas line    Connector stems with a pipe diameter of 6 mm or 1 4  are provided for gas connections  For    the gas supply and discharge from the device  you need to select a suitable material for the  sample gas     CAUTION    Dynamic pressure in the sample gas channel       The device must be operated in such a way t
105. ogether with the retaining plate and lift both out of the  analyzer     5  Remove the analyzer section from the retaining plate     6  Remove the sample gas lines from the analyzer section     Removing the sample chamber    Calibration    Leak test    OXYMAT 61    Proceed as follows for this     1  Remove the measuring head  as described in the section  Removing the measuring  head        2  Insert a suitable tool  e g  screwdriver  between the retaining plate and the cut strip   wound core  U core  and pry out the retaining plate until the tension spring is released     3  Remove cut strip wound core and retaining plate   4  Loosen the four mounting screws and pull the two parts of the casing apart     The sample chamber is now accessible and can be removed  To install  carry out the above  steps in reverse order  It is important to note the following in this regard     e Check all O rings to ensure they are OK  and replace them if they are damaged     e Tighten the mounting screws diagonally opposed with a torque of 6 Nm     Each time a measuring head is replaced or the analyzer section reassembled  you need to  recalibrate the device  This procedure is described in section Calibration  Page 50      A leak test must be carried out following every maintenance measure that affects the  analyzer section or the gas channel  This procedure is described in section Sealing   Page 35   If the device fails the leak test  you must replace all seals and hoses or pipes     Operating In
106. omponent   Buttons with varying function  softkeys    Display of activated measuring ranges with the current range marked  Limit marker on bargraph   Start of scale and full scale values    OMAN DOA FWD      Analog measured value display  bar chart  graph of the measured value with the start of scale and full scale value  of the current measuring range     10 Measured value    Figure 5 1 Measured value display and control panel    OXYMAT 61  Operating Instructions  12 2011  A5E00123067 02 39    Operation       5 7 General information    40                      Table 5  1 Button meanings  Button Meaning function  CLEAR Clears the commenced number input  ENTER Every number entered  except fast activation of a function  must be confirmed with   ENTER    ESC Jumps back one step in the operating structure   Changes are accepted without need for confirmation   INFO Information about current menu   current function  MEAS Jump back from every position in the operating structure to the  Decoded display  mode   you may be asked to confirm the entered data    Pressing the  MEAS  button again changes to  Coded display mode   i e  another  change in the  Operator control mode  requires the entry of the corresponding  code   Softkey Varying function  possible here are           e Submenu selection   function selection  e Selection of a subfunction  e ON OFF switch function       OXYMAT 61  Operating Instructions  12 2011  A5E00123067 02       Operation       5 1 2 Device operating modes   
107. on       
108. oning       4 2 Preparation for commissioning    4 2 2 Selection of the reference gas    When you select the reference gas  you must ensure that the various measurement spans  have at least one common point  This point is then defined as  physical zero point   This  applies to all measuring ranges  Once this point is known  the reference gas can be  selected     This is explained by the following example   There are four measuring ranges    e 17 22  02   e 15 25  O2   e 0 25  02   e 0 100   O2    These four measurement ranged have the spans from 17 to 22   O2 in common  The  physical zero point should lie within this range  Air  20 95   O2   is suitable as the  reference gas in this case     An exception to this principle is possible if the smallest measuring span 25   O2 and the  distance to the reference gas is not more than 20   Oz  Then  the physical zero point may  be located outside the measuring range  In this case  however  the pressure correction  see  Function 82 in section 5  must be activated  since there a pressure relationship due to the  large zero point offset     The purity of the reference gas has to be appropriate for the measuring task     Installing the reference gas connection  The reference gas connection is made depending on the order     e Air  low pressure variant   Air as the reference gas is sucked in with the internal pump  When the analyzer is  mounted in a cabinet  you must therefore ensure that the air sucked in is supplied from  outside the ca
109. pressed  this is displayed in the bottom line of the display   After pressing the  INFO  key  a message is displayed in plain text  The  MEAS  and  ESC   keys retain their  jump back  function    OXYMAT 61    Operating Instructions  12 2011  A5E00123067 02 91    Operation       3 6 Configuration    Relay and binary test    Analog test    92       Relay and Binary      Relay Binary    continue      Figure 5 57 Test relay outputs and digital inputs  The first function screen shows the state of the six relay and digital inputs of the standard  device  With an add on board  there are another eight relay digital inputs on a second page     With the relay test  individual relays can be activated  This is done via the input field  With   1   the relay is on  with  0   it returns to its release condition  Numbers other than  0  and  1   are not accepted by the input field     After leaving function 80  the relays are back in the state they had before calling the relay  and binary test     The current state of the digital inputs is displayed in the  Binary  column     With the analog test  the analog output is set to a constant current of 0 to 24 000 pA for  testing purposes     The analog input displays the input currents in pA     OXYMAT 61  Operating Instructions  12 2011  A5E00123067 02    Operation  5 6 Configuration       5 6 12 Language selection  function 81     With this function  you set the device to a second dialog language     81 Select language MC    English E    Deutsch
110. pressing the first  top  softkey      Analyzer status      the following menu screen appears with additional choices     Analyzer status    1 Analyzer config   2 Diagnostic values    3 Logbook    4 Display meas  ranges       Figure 5 5 Analyzer status submenu  The analyzer status functions are freely accessible  There is therefore no code query and no  change in the operating mode     The analyzer status submenu offers you various functions for displaying device parameters  and stored data     5 2 1 Analyzer configuration  function 1   When this function is selected  important device manufacturing data can be viewed   e Firmware No   Order No  of the software in the EPROM  e Order No   Information on device ordering data  e Production No   Indication of date of manufacture and device serial number  e Object version  Indication of the hardware design of the device  e Software version and date    Indication of the functional scope of the device    OXYMAT 61  48 Operating Instructions  12 2011  A5E00123067 02    Operation       5 2 2    5 2 3    5 2 4    OXYMAT 61    5 2 Analyzer status    Diagnostic values  function 2     The most important diagnostic values may be called using function 2  They may allow  conclusions to be drawn for evaluation of errors or setting work     Logbook  function 3     In the logbook  all errors which led to a maintenance request  W  or a fault  S  are listed     The limit alarm  LIM  and functional check  CTRL  are also registered  However  these do
111. ration tolerances to be checked  by AUTOCAL Check  Using functions 71 and 72  you can also select the relay output or  digital input for  AUTOCAL Check      When a calibration limit is exceeded  the maintenance request W10 is set  and also  if  configured  the relay  Acal Chk Dif      Both will be reset again after an error free AUTOCAL Check  W10 is retained in the logbook    When starting via cycle  initiate AUTOCAL Check      If this function is activated  the device executes an AUTOCAL Check when AUTOCAL is  started via a cycle  In other words  it only checks that the calibration tolerances are  observed  but does not carry out a calibration     Operating Instructions  12 2011  A5E00123067 02 61    Operation       5 3 Calibration    5 3 6    62       Sequence   1  Start the AUTOCAL Check         By means of the softkey  Trigger AUTOCAL Check once in the subfunction   Aca Check sequence       Via digital input    Via cycle    2  The device executes the sequence as configured in the subfunction  Autocal Check  sequence        Drift values  function 25     This function shows deviations occurring for the calibrations  and AUTOCAL as well    actual  value   setpoint  as a sum parameter  All rated zero point and sensitivity calibrations of each  range are calculated here for the selected measured component  Every new deviation is  added to the existing drift value     25 Drift values    Zero   0 00 vpm    Span MR4   36 00 vpm    Drift values reset o    Figure 5 20 Drift values
112. remove it   Remove the data plug from the rear of the device    Unscrew the three M3 screws located between the plugs     Pullout the plugs of the ribbon cables from the motherboard     oa kw nN  gt     Remove the motherboard from the device     OXYMAT 61  Operating Instructions  12 2011  A5E00123067 02       Appendix       Removing the add on board    Mounting    6 4    Fuses F3  F4    OXYMAT 61    Operating Instructions  12 2011  A5E00123067 02    The procedure is the same as for the motherboard  In contrast to the motherboard  the add   on board is only attached to the back of the housing with two screws     Proceed in reverse order to install the both board     Replacing fuses       AR warn ING    Dangerous contact voltage    You need to open the enclosure to do this work  This makes live parts are accessible   which may cause death or injury due to electric shock when touched        For this reason  you must always disconnect the device from the mains before replacing the  board    Improper handling   The fuses may be only be replaced by trained or instructed personnel  You may only use  fuses of the same type              The device is protected by multiple fuses  the values of which depend on the device model   e g  supply voltage  heating  etc                      Power Fuse ratings  200 to 240 V 0 63 T 250  100 to 120 V 1 T 250       Fuse ratings in Ampere    These fuses are located in a drawer of the mains connection box  To replace the fuses  use  a suitable tool  
113. ressing the associated softkey  the code of the corresponding  operation level is queried once  exception  the submenu  Analyzer status  is freely  accessible  selecting this does not change the operating mode      Decoding level 2 also decodes level 1     By entering a corresponding code  the device switches to the operator control mode   whereby the functional check is activated     The functional check  CTRL   shown in the status line of the measured value display  is  always activated by the device when an intervention endangers correct measurement  e g   when the code is entered  If you have configured a corresponding relay using function 71  with  Functional check   an extra signal is sent out via the relay contact together with the  decoding  This relay contact then signals every functional check activation  e g  even the  warm up phases and calibration states of the device     Decoding also activates the measured value memory  providing you have configured this  using function 77     The coding status of the device can be read in the status line of the measured value display  as a symbol  m CODE  for  coded  or  2 CODE  for  decoded                  Operating Instructions  12 2011  A5E00123067 02 45    Operation       5 7 General information    5 1 4 3    5 1 4 4    46    Returning to display mode    You can use the  MEAS  key to jump back to  Decoded display mode  from any location in   Operator control mode   Any input started is aborted     Before jumping back  the
114. ro gas 1  1 0 min    Zero gas 2  1 min    Cal  gas 1  2 min    Cal  gas 2 1 3 min    continue       Figure 5 14 AUTOCAL Check sequence    A AUTOCAL can be composed of up to 12 individual steps     Besides the supply of one zero gas and up to four calibration gases per component  you can  also program purging with sample gas  sample gas intermediate operation  as well as a  signaling contact  The signaling contact is only available if you have previously assigned it to  a relay output using function 71     Sample gas intermediate operation    Sample gas intermediate operation can be necessary if the system may only leave  measuring mode for a certain time  If the sum of the required purge times are greater than  the permissible downtime  you must return to the measuring mode between the individual  calibrations     OXYMAT 61  Operating Instructions  12 2011  A5E00123067 02 57    Operation       5 3 Calibration    58       Signaling contact   Use the signaling contact to initiate an automatic calibration process of a second device  for  example  or to signal the beginning or end of an AUTOCAL    Relay outputs   If you have defined relay outputs for the sample gas  zero gas  calibration gases and or  measuring calibration  function 71   these are used to activate the corresponding external  solenoid valves  The same applies to the signaling contact  AUTOCAL   This is closed for 1  s when the command is executed    Example    You want to program the following sequence      Zero g
115. ructions  12 2011  A5E00123067 02 67    Operation       5 5 Parameters    5 5 1    68    Electric time constants  function 50     50 Electr  time con MC    Effective bandwidth in   of  smallest MR 6 0      Time constant within  bandwidth ti    10    Time constant    outside  bandwidth ta   1     Actual measured  value  0 00   viv       Figure 5 30 Electric time constants    With this function  you set various time constants  which reduce the underlying noise in an  message value signal  The noise reduction is approximately equivalent to a low pass filter  with the corresponding time constant  The display delay is given as the 90   time     The time constant  ti  acts within an configurable effective interval which is defined in   of   the smallest span  It attenuates small changes in the measurement  e g  noise  on the one  hand  but becomes ineffective immediately when the measured value exceeds the effective  interval  In this case  the outer time constant  ta  attenuates the measured value     You can configure values up to 100   for the effective interval  and values up to 100 s for  the time constants  ti  and  ta    By cleverly combining these three parameters  you can  achieve a low display delay  90   time  despite high noise suppression     The effect on the configured attenuation parameters can be observed in the bottom line  The     live  measured value is displayed here     OXYMAT 61  Operating Instructions  12 2011  A5E00123067 02    Operation  5 5 Parameters    
116. rved                 All names identified by    are registered trademarks of Siemens AG  The remaining trademarks in this publication  may be trademarks whose use by third parties for their own purposes could violate the rights of the owner     Disclaimer of Liability    We have reviewed the contents of this publication to ensure consistency with the hardware and software  described  Since variance cannot be precluded entirely  we cannot guarantee full consistency  However  the  information in this publication is reviewed regularly and any necessary corrections are included in subsequent  editions        Siemens AG Order number  A5E00123067 Copyright    Siemens AG 2001   Industry Sector    02 2012 Technical data subject to change 2011   Postfach 48 48 All rights reserved    90026 N  RNBERG  GERMANY    Table of contents    1 General  informati  n ii iia 7  1 1 Device model and Validity  narnii annae aaa nano cnn aaa rca nn 7  1 2 Information for our customers a e aa aae a a conan nn eaa nn nena nn nn aE Aaa a eaaa aea 7  1 3 Gone MINO MAIN cito it A A A a aa a a 7  1 4 Special information and Warnings stisne rnea a E ncnc nano nn nc nano nn EAA nc rcnnnancnnnns 8  1 5 Pr  per USO iii oia 9  1 6 Qualified Personineliiisu u  2m reihe iii Pane Seyi add ceded 9  1 7 Netes on Walanty   2  2 22 A AA Ai ee eae 10  1 8 Deliveryinformation  AAN 10  1 9 Standards and regulations  sasani iiuair ENE rca 10   2 Description    A em AA AE AE E AE aid 11  2 1 Area 0l applieation     nee 11  2
117. s a reference gas     1  Determine the influence of interfering gas   The diamagnetic zero point offset  O2 equivalent  of the propane is   0 86   O2  With  content of 50    the interfering gas swing would be   0 43   Oz     2  Enter the interfering gas swing  her   0 43   Oz      Correction of cross interference with variable influence of interfering gas  83 Interference corr  0    With var  interference o  influence on an  input 1       v  3    Valid for range 121314  O                      Interfering gas  100     causes an  interference   42 94     Analog in 1   4  20 mA e    for range    0 00 5 00       Figure 5 63 Variable influence of interfering gas    Example of a variable cross interference correction    The concentration of the interfering component must be measured with a suitable analyzer  and the corresponding analog signal must be transmitted to the OXYMAT 61 for inclusion in  calculations     A pure residual gas should be specified as the interfering gas swing  O2 equivalent      By specifying the measuring range of the residual gas analyzer as a   as well as its  electricity output  an internal calculation of the actual O2 offset can be made     Example 2   A sample gas is composed of 4  NO and 96  Na  It should be monitored for Oz   The O2 equivalent of 100  NO amounts to 42 94  Oz     The NO analyzer has a measuring range of 5   NO and an analog output from 4 to 20 mA     OXYMAT 61  96 Operating Instructions  12 2011  A5E00123067 02    Operation  5 6 Config
118. s required for configuring the device     Configuration    70 Analog output  71 Relay outputs  72 Binary inputs  73 ELAN configuration    continue         Figure 5 43 Configuration submenu    All functions of this menu are only accessible via the code of level 2     Press the fifth softkey      continue   to branch to further configuration functions     5 6 1 Analog output  function 70     With this function  you can configure the analog output     70 Analog output MC  4   20 mA o    Output inverted a    Suppress negative a  measured values       Figure 5 44 Configuring the analog output   With softkey 1  define the start of scale value of the measuring range  The following settings   see also table  Configurations of the analog output   are possible    e 0 20mA   e 2 20mA   e 4 20mA   e NAMUR 4   20 mA  with limit at 3 8 mA      OXYMAT 61  78 Operating Instructions  12 2011  A5E00123067 02    Operation       OXYMAT 61    5 6 Configuration    You can invert the analog output with softkey 2  e g  0     20   O2   0     20 mA  gt  0     20      O2   20     0 mA    With softkey 3  you can suppress the negative measured values  If a negative value has an  unfavorable effect on further processing  the negative measured values at the analog output  can be limited to 0  or 2 4 3 8  NAMUR  mA  when this function is activated  The display  continues to show the actual measured value                    Table 5  5 Configurations of the analog output   Defined analog Measuring range limit
119. sation  99  Zero calibration  51   Order no   115   Order No   48    OXYMAT 61  Operating Instructions  12 2011  A5E00123067 02    P    Parameters  67  Analog output  78  Code  91  Date  76  Digital inputs  82  Display contrast  74  ELAN  85  Graphical measured value representation  72  Limits  69  Logbook  77  Magnetic field frequency  75  Measured value display  74  Measuring point switching  77  On off functions  70  PROFIBUS  101  Relay assignment  80  Status messages  71  Time  76  Time constants  68  Pin assignments  28  AUTOCAL  30  Power supply  124  Pressure correction  93  Principle of operation  17  Process tags  77  Production No   48  PROFIBUS  101  Proper use  9  Purity monitoring  38    Q    Qualified personnel  9    R    Reference gas  Connection  22  33  Line  105  Selection  33  Spare parts  123  Technical specifications  127   Relay assignment  80   Relay function  80   Reset  86   Return delivery  135   Room air monitoring  37   RS485  15    151    Index       S    Safety instructions  Commissioning  18  31  General  7  Mains connection  25  Maintenance  103  Servicing  103  Signal connections  26   Sample chamber  105  106  107  Cleaning  106   Sample gas  Associated gases  94  Conditioning  22  Connection  21  Correction of cross interference  94  Discharge  21  Interfering gases  94  Leak test  34  Line  21  Pressure  93  Spare parts  121   Sample gas restrictor  108   Self test  91   Sensitivity  Calibration  52  Setpoint  54   Setting  Analog output  
120. scope of delivery is listed on the shipping documents   enclosed with the  delivery   in accordance with the valid sales contract     When opening the packaging  please observe the corresponding information on the  packaging material  Check the delivery for completeness and undamaged condition  In  particular  you should compare the Order No  on the rating plates with the ordering data  if  available    If possible  please retain the packaging material  since you can use it again in case of return  deliveries     Standards and regulations    As far as possible  the harmonized European standards were the basis for the specification  and production of this device  If no harmonized European standards have been applied  the  standards and regulations for the Federal Republic of Germany are valid     When this product is used beyond the scope of these standards and regulations  the valid  standards and regulations of the country of the operating company apply     OXYMAT 61  Operating Instructions  12 2011  A5E00123067 02    Description       Description 2    2 1 Area of application    The OXYMAT 61 gas analyzers use a measuring principle based on the paramagnetic  alternating pressure method and are used to measure oxygen in gases        NOTICE             The OXYMAT 61 is not suited for measuring flammable  toxic and corrosive gases        Application examples  Measurement of oxygen  e For boiler control in incineration plants  e For quality monitoring in ultra pure gas  e For en
121. software function of a device  Functions are listed in submenus     Function display    Screen display depending on the called function     OXYMAT 61  142 Operating Instructions  12 2011  A5E00123067 02    Glossary       Functional check    Identification activated by the device if it determines that the measured value was influenced   e g  when changing to operator control mode by decoding the device      Gas analyzer    Device for quantitative analysis of gases and gas mixtures     Gas inlet   Defined point for connecting a gas to the analyzer   Gas outlet   Defined point for directing a gas out of the gas analyzer   Input field    Single or multi line area for entering data     Interfering gas    A gas which interferes with the measurement  and which may be contained in the sample  gas     Limit alarm    Signaling of violation of a high or low limit     Limit monitoring    Function which monitors adherence to adjustable limits  and signals if high or low limit is  violated     Limit relay   Relay to which a certain limit is assigned and which is switched when there is a limit alarm   Main menu   Menu of the highest hierarchical level  lt contains the submenus   OXYMAT 61    Operating Instructions  12 2011  A5E00123067 02 143    Glossary       Measured signal    Representation of measured variables in the signal path by an allocated physical quantity of  the same or different type  Depending on the position of the measured signal at the input or  output of the considered el
122. state  also in status message in the status line                deactivated  OFF state  also in status message in the status line         gt    access a submenu subfunction    e   trigger a function subfunction  e g  Start calibration        Operating sequence    CAL OCTRL    63 28   Vol       0 0                      Figure 5 3 Measured value display    Operating Instructions  12 2011  A5E00123067 02 43    Operation       5 7 General information    5 1 4 1 Introduction to the main menu    The device must be in measuring mode  The right side of the display field shows the  measured component  which is indicated with  MC  in the following illustrations of the  display  A right arrow   gt   appears for it  This arrow indicates a softkey  The main menu can  be opened by pressing this softkey     The main menu consists of the following commands  with the associated code level on the                   right     Function group Code level   Analyzer status Not coded  Calibration Code of code level 1  Measuring ranges Code of code level 1  Parameter Code of code level 1  Configuration Code of code level 2                The code of level 1 is factory set to  111   that of level 2 is factory set to  222    Main menu    Analyzer status    Calibration    Measuring ranges    Parameters    Configuration       OXYMAT 61  44 Operating Instructions  12 2011  A5E00123067 02    Operation       5 1 4 2    OXYMAT 61    5 7 General information    Entering a submenu    If you select a submenu by p
123. structions  12 2011  A5E00123067 02 107    Appendix       6 2 3    6 3    Removing the sample gas restrictor    Ifthe sample gas restrictor is clogged from condensate and therefore needs to be cleaned   or if it a hindrance to the measurement task for other reasons  e g  due to the use of  OXYMAT 61 with gas analyzers of the type ULTRAMAT 6 in certain measuring configuration     you must remove the restrictor     If there is no a flow indicator  optional  in the device  the sample gas restrictor is located in  the sample gas hose between the pipe inlet nozzle and the analyzer section     If there is a flow indicator  the sample gas restrictor is located between it and the analyzer  section  The restrictor is secured with a hose clamp     Remove the restrictor as follows   1  Loosen the hose section in which the restrictor is located   2  Remove the hose clamp over the restrictor     3  Use a suitable object  stick  etc   to push the restrictor out of the hose     Replacement of motherboard and add on board       AL WARNING    Dangerous contact voltage       You need to open the enclosure to do this work  This makes live parts are accessible   which may cause death or injury due to electric shock when touched     For this reason  you must always disconnect the device from the mains before replacing the  board           Replacing the motherboard    108    To do this  proceed as follows    Disconnect the device from the mains    Loosen the screws on the cover of the housing and 
124. t type meters Duran glass  Suspension boundary PTFE  Teflon   Angle pieces FPM  Viton    Sample gas Membrane FPM  Viton    pressure switch Enclosure PA 6 3T             OXYMAT 61    Operating Instructions  12 2011  A5E00123067 02          Appendix       Dimension drawings 9    9 1 Dimension drawings                                                                                 OO OULU                            OXYMAT 61       465          483          440          AAA Aa O Aa cA Ad L                            384 5                                                 Figure 9 1 Dimensions for preparing installation  in mm   front view and plan view     OXYMAT 61  Operating Instructions  12 2011  A5E00123067 02 129    Appendix                 4 3949  gt   p 351 5 a      27 5   pq 156                                                       Figure 9 2 Dimensions for preparing installation  in mm   front view and plan view     OXYMAT 61  130 Operating Instructions  12 2011  A5E00123067 02    Appendix       Appendix    EC Declaration of Conformity    OXYMAT 61    SIEMENS    Industry    EC Declaration of Conformity  EG Konformit  tserkl  rung    No  A5E00484554   03          Manufacturer  Siemens AG  Hersteller  Industry Sector  Industry Automation Division  Address  76181 Karlsruhe  Germany  Anschrift   Product description  Gas analyzer   Gasanalysator  Produktbezeichnung OXYMAT 61  OXYMAT 6 E   F  OXYMAT 64 E  Type   Typ  7MB2001 xxA00 xxxx Z Xxx  7MB2011 xax0x bcdx Z e f xxx  7MB201
125. tch  the device off and on again  If the error  message S1 appears again     2  Load user data  function 75     3  contact service department    Make sure you leave the device in  operation   in order to assist the fault diagnostics of  the service staff         2 Magnetic field supply faulty    Ribbon cable connection interrupted    Check connection       Motherboard faulty    contact service department                                         3 Microflow sensor faulty One half of the grid has been Replace measuring head or contact   Q destroyed service department   S4 External fault message Signal from external source Check  Function 72 must be configured   accordingly   S5 Temperature of analyzer section   Ambient temperature outside Make sure that the ambient   Q is too high or too low tolerances specified in the technical   temperature is in the permissible range  specifications  5 to 45   C  from 5 to 45  C  Measuring head temperature too Perform restart  RESET   if this is not  high   gt 80  C  or too low   lt 10  C  successful   Temperature sensor faulty  contact service department   gt  Temperature rises above setpoint  temperature   S7 Temperature of measuring head   More than  5   C deviation from the Replace measuring head or contact   Q outside tolerance setpoint temperature  75   C  service department   OXYMAT 61    Operating Instructions  12 2011  A5E00123067 02    113    Appendix                                                                No  Message Possibl
126. termediate operation Function code 8  Signaling contact Function code 9   59    Operating Instructions  12 2011  A5E00123067 02    Operation       5 3 Calibration    AUTOCAL Check cycle parameters    With this subfunction  you can configure various time constants for activating a cyclically  repeated AUTOCAL sequence     Autocal   check cycle    Time from autocal to auto   cal  cycle time   0   h     Time up to next autocal  15  min     Carry out span calibration  at any 1st cycle    Total range calibration  Cal  gas       Figure 5 18 AUTOCAL Check cycle parameters     Time from AUTOCAL to AUTOCAL  cycle time       Any setting between 0 and 1 000  hours  is accepted by the device   0  corresponds to no  AUTOCAL cycle      Time to first AUTOCAL   after the time of setting      If you enter  0  here and AUTOCAL is activated with  AUTOCAL on off subfunction  the  device begins immediately with the AUTOCAL sequence     If AUTOCAL is switched off  the device only starts an AUTOCAL sequence if AUTOCAL is  switched on within one minute after entering  0   If not  the total time between two AUTOCAL  cycles passes by  starting with the input of  0         Note    The clock inside the device also runs when AUTOCAL is deactivated  It starts the first time  the device is switched on with the factory set start time and must be set to the current time  using function 58         Carry out span calibration for each xth cycle      Here you set the number of cycles after which a calibration wi
127. th the calibration gas is to be  carried out     If you would like to save on calibration gas  for example  and not calibrate the span along  with every zero calibration  enter in the line  Calibrate with calibration gas at every     cycle   an integer  gt  1     OXYMAT 61  60 Operating Instructions  12 2011  A5E00123067 02    Operation       5 3 Calibration    Information in the last two lines    The information in the last  bottom  lines indicates that specified parameters relate to total  calibration with calibration gas for measuring range 3  This measuring range was selected  beforehand with function 22        Note    As long as AUTOCAL is activated  AUTOCAL 5   access to functions 20 and 21 is blocked   If you activate these functions anyway  a corresponding message appears on the display        Settings for AUTOCAL Check    OXYMAT 61    The  AUTOCAL Check  subfunction is only used for checking the calibrations     Similar to  AUTOCAL   sequence set in the  AUTOCAL Sequence  menu is performed  In  contrast to  AUTOCAL   no new calibrations are performed  only the deviations are checked  with respect to selectable calibration tolerances     Autocal check    Calib  tolerance at zero  in   of smallest MR   6     Calib  tolerance at sens   in   of current    Start only Autocal check o  cyclically    Trigger acal check once o    Abort autocal check 0       Figure 5 19 AUTOCAL Check    Functions of the AUTOCAL Check    Inthe  AUTOCAL Check subfunction  enter the desired calib
128. the internal friction of the  sample gas  This also provides a high degree of corrosion resistance because the microflow  sensor is not exposed to the direct influence of the sample gas     A magnetic field with alternating strength  8  prevents detection of the background flow in the  microflow sensor  enabling measurement to be performed independent of the sample  chamber position and thus also independent of the location where the gas analyzer is used     The sample chamber is directly in the sample channel and has a small volume  and the  microflow sensor is a low lag sensor  This results in a very fast response time for the  OXYMAT 61     OXYMAT 61  Operating Instructions  12 2011  A5E00123067 02    Mounting and connecting       Mounting and connecting    3 1 Safety instructions       DA R          Explosion hazard    The analyzer must not be operated in potentially explosive atmospheres  The supply of  gases with combustible components in concentrations above the lower explosive limit  LEL   must be clarified in consultation with the relevant hazard experts and is ultimately the  responsibility of the operator           AR WARNING  Dangerous contact voltage       Certain parts of this device are under dangerous voltage  In order to avoid contact with live  parts  the housing must be closed and grounded upon completion of all assembly and  installation work     Failure to comply with this can result in death  personal injury and or damage to property           AR WARNING
129. ther gas analyzer via ELAN  It is basically possible to use other  pressure measuring instruments or analyzers with pressure measuring function and ELAN  capability for the pressure measurement        On the device supplying the pressure data  the parameter  Measured value telegrams  must  be set to  On  using the function 73     5 6 14 Correction of cross interference  function 83     94    Associated gases contained in the sample gas  interfering gases  can affect the  measurement  This function is used to reduce such interference or eliminate it altogether in  the ideal case        Note    A correction for interfering gas is usually only useful if the corrected O2 equivalent is not  greater than the smallest span        When the reference gas and residual gas have different compositions    sample gas without  Oz content   a zero point offset  interfering gas swing  occurs triggered by the paramagnetic  or diamagnetic difference of the two gases  The value of this zero point offset can be  specified to compensate for this offset        Note    A differentiation must always be made with the correction of cross interference as to whether  the interfering gas has a constant or variable concentration        OXYMAT 61  Operating Instructions  12 2011  A5E00123067 02    Operation  5 6 Configuration       Press the first softkey  The following function screen appears on the display     83 Interf  correct  0     No interference correct  0       Figure 5 61 Carrying out correction o
130. tion   Measure  operating mode required     Total calibration is only possible if the    7ofa  calibration    subfunction has been activated  using function 23     With a total calibration  you calibrate all measuring ranges together  Define the  leading   measuring range in function 22  We recommend that you use the largest measuring range  for this     After selecting this function  the display shows the setpoint and the current value of the   leading  measuring range     21 Sp  cal  all MRs    Setpoint   100 00   v v    Act  value  97 79   viv    Start calibration 0    CANCEL e    Figure 5 9 Calibrating the measuring range  total calibration   When the actual value has stabilized  you can initiate the calibration process by pressing the  fourth softkey  The actual value is now made to agree with the setpoint     If you make a mistake during the calibration  e g  wrong calibration gas   reload the original  calibration by pressing the  CANCEL  softkey        Note  Measuring accuracy    With a switching ratio of the spans of more than 1 10  an individual calibration should be  carried out for each range to achieve a higher measuring accuracy        Operating Instructions  12 2011  A5E00123067 02 53    Operation       5 3 Calibration    5 3 3 Zero span setpoints  function 22   22 Setpoints total    setpoint for  zero  0 00    viv    setpoint for MR 1  50 00      viv  Setpoint for  MR 2  60 00   v v    Setpoint for  MR 3  80 00   viv    Setpoint for MR 4  100 00   viv       Fi
131. tips and helps avoid maloperations     OXYMAT 61  8 Operating Instructions  12 2011  A5E00123067 02    General information       1 5    1 6    OXYMAT 61    1 5 Proper use    Proper use    Proper use within the context of this manual  means that the product may be used only for  the applications described in the catalog or the technical description  and only in combination  with the equipment  components and devices of other manufacturers recommended or  permitted by Siemens     The product described in this manual has been developed manufactured  tested and  documented in compliance with relevant safety standards  When the handling rules  described for the configuration  installation  proper operation and maintenance  as well at the  safety guidelines are adhered to  therefore  there is normally no risk to the health of persons  or in respect to damage to property     This device was designed to ensure safe isolation of the primary and secondary circuits  Low  voltages that are connected must therefore also be generated with safe isolation        AR warn ING  Dangerous contact voltage       After removing the housing or protection against direct contact or after opening the system  cabinet  certain parts of of this device system will be exposed that can carry hazardous  voltage  Therefore  only appropriately qualified persons are permitted to perform work  within this device  These persons must be thoroughly familiar with all sources of danger  and service activities in accor
132. to 100   O2       Measuring range 2    90 to 100   O2       Smallest span     100   95    O2   5   O2       Measuring range in which calibration takes place    Measuring range 2       Selected calibration tolerance    6        Response threshold for zero point    5   O2 x 0 06   0 3   O2       Response threshold for span          10   O2 x 0 06  0 6   O2          If the zero   span deviates by more than the configured value compared to the last  calibration  and if  Signal tolerance violation  is activated  the corresponding relay signals a    maintenance request     OXYMAT 61  Operating Instructions  12 2011  A5E00123067 02    Operation       5 6 Configuration  5 6 10 Change codes  function 79   With this function  you replace the factory set codes   111  for level 1   222  for level 2  by  your own     79 Codes program     Code 1 2111     Code 2 222       Figure 5 55 Change codes    If you enter the value  000  for a code  code blocking is canceled and you have immediate  access to the corresponding operation level     5 6 11 Device test  function 80     You can check the functionality of the device by performing the following three tests     80 Analyzer test    Keyboard       Relay and Binary test      Analog test          Figure 5 56 Testing the device    Keyboard test  You can test all keys on the control panel by means of the keyboard test  The five softkeys  on the right edge can be used to make the associated item appear disappear  If a number  key or the sign key is 
133. undergoing repairs or maintenance work   or  for example  if new parameter settings are to be tested     OXYMAT 61  86 Operating Instructions  12 2011  A5E00123067 02    Operation       5 6 Configuration    Press softkey 3 to restore the factory settings     A confirmation query is set up in this function  In order to actually load the respective data in  the memory  you must confirm with  yes   If you select  no   this is cancelled        NOTICE       Make sure you save every change in the configuration of the relay outputs in the user data  memory using function 75     If you neglect to do this  there is a danger that a previous  unwanted  configuration will be  loaded by  Load user data   function 75               The following figure shows an overview of the interaction between the various memory  modules     Program memory Main memory Parameter memory  EPROM   Flash PROM  RAM EEPROM    Factory data    Fct  99  Program Running  only  operations service    engineers  User data    Working area       Factory basic data  change    Fct  99  only service engineers        Figure 5 51 Memory modules    OXYMAT 61  Operating Instructions  12 2011  A5E00123067 02 87    Operation       3 6 Configuration    5 6 7    88       Suppression of short noise signals  function 76     With function 76  you can eliminate undesired spikes which exceed a settable threshold     76 Suppress fault     Suppress noise signals  with a ee ar up to    0 0  s    Level in   of  smallest MR     Figure 5 52 
134. uration       Correction of cross interference via ELAN  83 Interf  correction    With var  interference  influence on ELAN       Valid for range                      Interfering gas  causes an  interference    Channel   03   NO   5        Figure 5 64 Correction of cross interference with variable influence via ELAN    If the cross interference correction is performed via the serial RS485 interface  ELAN   the  same specification must be made as for the cross interference correction with a variable  interfering gas influence     Also enter the channel number and component number of the interfering gas analyzer  The  type of gas allocated to the channel and component  the measuring range and possibly the  device status appear in the display  see also function 82  Pressure compensation       With softkey 2  you can set whether this correction of cross interference should only apply to  certain measuring ranges        Note    On the device supplying the correction data  the parameter  Measured value telegrams    function 73  must be set to    On           OXYMAT 61  Operating Instructions  12 2011  A5E00123067 02 97    Operation          3 6 Configuration  5 6 15 Phase adjustment  function 84   The physical principle of the measurement procedure and the mechanical design of the  detector result in a delayed reaction  phase offset  of the analog measured value signal  compared to the clock signal of the magnet controller  This phase shift can be adjusted if  necessary with the he
135. uring point 1 to 6 Measuring point 1 to   For taking a gas sample via  6 selected solenoid valves at various  measuring points  Measuring point signal Measuring point 1 to   For identifying the  1to6 6 selected measuring point  runs  parallel to measuring point   Signaling contact When there s a e g  in the case of  signal  the relay AUTOCAL   conducts current for   actuation of a second  a short time device  Sample gas flow Sample gas flow too   For identification purposes  low  Reference gas pressure Reference gas For identification purposes    pressure too low       AUTOCAL Check Autocal difference too  large  function 24                       OXYMAT 61  Operating Instructions  12 2011  A5E00123067 02 81    Operation       3 6 Configuration    5 6 3 Binary inputs  function 72     Digital input functions    82       Six floating digital inputs are included in the standard equipment   0    0 V  0 to 4 5 V    1     24 V  13 to 33 V       If these six inputs are insufficient  you can retrofit additional electronics with eight further  digital inputs  add on board      72 Binary inputs MC  Fault Maint r CTRL NAMUR O    Binary inputs define         Figure 5 46 Digital input function    The operation of the digital inputs are defined with softkey 1  If you activate the  NAMUR         operating mode  the digital inputs behave as in table  Digital input activation functions   marked with  N      If you deactivate the  NAMUR    0   operating mode  the digital inputs behave compatibl
136. uring ranges  depending on version 2 0   or 5 0   O2    Easy device replacement since electric connections can be simply disconnected from the  device    19  rack unit with 4 height units  U  for installation in swivel frame   19  rack unit with 4 U for installation in cabinets  with or without telescopic rails  Front plate can be swung down for servicing purposes  laptop connection   Internal gas channels  Hose made of FKM  Viton    Connections for sample gas  Pipe diameter 6 mm or 1 4     Sample chamber of stainless steel  mat  no  1 4571     OXYMAT 61  Operating Instructions  12 2011  A5E00123067 02    Description       2 2 Design    2 2 Design    Display and control panel    Inputs and outputs    Communication    OXYMAT 61    Large LCD for simultaneous display of        Measured value  digital and analog displays      Status line       Measuring ranges   Contrast of the LCD adjustable via the menu   Permanent backlighting with LED   5 digit measured value display  decimal point counts as digit    Washable membrane keyboard front panel   Menu guided operation for parameter assignment  configuration and calibration  Operating help in plain text    Graphic display of concentration trend  programmable time intervals    Two analog inputs configurable  e g  interference correction or external pressure sensor     Six relay outputs  freely configurable  e g  faults  maintenance request  calibration  limit  alarm  external solenoid valves     Six digital inputs  freely configurabl
137. vironmental protection  e For monitoring of process exhaust gas    e Forprocess optimization    OXYMAT 61  Operating Instructions  12 2011  A5E00123067 02 11    Description       2 1 Area of application    Important features    12    Four freely configurable measuring ranges  also with suppressed zero point  all  measuring ranges are linear    Electrically isolated measured value output 0 2 4 to 20 mA    Selectable automatic or manual measuring range switching  remote switching is also  possible    Measured values can be stored during calibration    Wide range of selectable time constants  static dynamic noise suppression   i e  the  response time of the device can be adapted to the respective measuring task    Simple handling through menu guided operation according to NAMUR recommendation  Fast response time  Low long term drift    Reference gas supply either externally  N2  O2 or air  approx  2000 to 4000 hPa via  sample gas pressure  or via built in reference gas pump  ambient air     Two control levels with separate authorization codes to prevent unintentional and  unauthorized operator intervention    External pressure sensor can be connected for correction of variations in the process gas  pressure in the range 800 to 1200 hPa  absolute     Automatic measuring range calibration can be configured   Fieldbus connection  optional    Monitoring of sample gas  optional    Monitoring of reference gas  only with version with internal reference gas pump   Variety of narrow meas
138. y  with the software outputs of the older version V 4 3 0  in the table  Digital input activation  functions  marked with  X                   Press softkey 2  The following function screen appears on the display   72 Binary inputs  D 1 Autocal Check    D2 Vacant    D 3 Vacant    D4 Vacant    continue    Figure 5 47 Define digital inputs    OXYMAT 61    Operating Instructions  12 2011  A5E00123067 02    Operation       5 6 Configuration    Assign one of the activation functions listed below to each input as you like  whereby every  function may only be assigned once     The connection assignments for the individual inputs can be found in the section  Electrical  connection      No digital input has been preassigned in the factory     Up to four digital inputs can be configured in a menu screen  Press the fifth softkey       continue    to access the next function screen  thus branching to further digital inputs        NOTICE    Save changes       Make sure you save every change in the configuration of the digital inputs in the user data  memory using function 75  If you neglect to do this  there is a danger that a previous   unwanted  configuration will be loaded by  Load user data   function 75               Table 5  7 Digital input activation functions             Function Necessary activation Remark   effect  voltage  OV  24  24V pulse  V  min  1 s   Not assigned No effect when activated  External fault 1 N X e g  due to   to 7       e Signal from a gas preparation system 
    
Download Pdf Manuals
 
 
    
Related Search
    
Related Contents
JEOL 7500F Quick User Guide  LOREX Technology ACC-DRW User's Manual  JobExchanger32 - MPL Automation  DeepC User Manual  Fonctionnement de l`application  Diese Anleitung ist ein Vorausgabe der gedruckten Version  Manual de usuario - Serie Vita II  Grizzly G0703 11 Dust Collector User Manual  1. Monter un projet artistique et culturel - CRDP Basse  Manuel d`utilisation et d`applications    Copyright © All rights reserved. 
   Failed to retrieve file