Home
Technical Data Sheet
Contents
1. AE EE EE AE Vd Vo V LED I V Curve Vd Forward voltage of the LED Rd LED operating resistance Vo LED operating voltage lo LED operating current Use the load s unique CR LED operating mode to test LED drivers This function allows users to configure the LED s operating resistance and forward voltage to simulate the loading behavior of typical LEDs www bkprecision com Programmable DC Electronic Load 8600 Series Remote control and programming Powerful communication interfaces The 8600 Series provides standard GPIB USB and RS232 interfaces for remote communication These interfaces offer SCPI and USBTMC standard communication protocols to control your electronic load from a PC Battery test function Typical battery discharge curve Volts 0 0 5 1 0 1 5 2 0 2 5 Hours The built in battery test function uses CC mode to calculate the battery capacity using a fixed current load discharge Users can specify cut off voltage level capacity level and time stop conditions Built in rise and fall time measurement 8600 Series display External analog programming and monitoring interface In addition to front panel and remote interface control current values can also be programmed with an analog control signal The electronic loads can be externally controlled from zero to full scale with a 0 10 V input signal A BNC output is available on the rear for monitoring the current with a 0 10 V
2. 150 kQ 300 kQ MQ AC input 110 V 220 V 10 50 60 Hz Operating temperature 32 F to 104 F 0 C to 40 C Storage temperature 14 F to 140 F 10 C to 60 C Humidity Indoor use lt 95 Safety EN61010 1 2001 EU Low Voltage Directive 2006 95 EC 8600 Series Specifications Model 8600 8601 8602 Input ratings Input voltage 0 120V 0 120V 0 500 V Input Low 0 3A 0 6A 0 3A current High 0 30A 0 60A 0 1I5A Input power 150 W 250 W 200 W Minimum Low 0 11 Vat3A 0 18 Vat6A I1Vat3A operating voltage High L I Vat 30A 1 1 Vat60A 45Vat ISA CV mode Low O 18V 0 50V Range High 0O 120V 0 500 V Low mV Resolution High 10 mV Low 0 05 0 025 0 05 0 02 FS 0 05 FS 0 025 FS Accuracy Hish 0 05 0 025 0 05 8 0 025 FS 0 05 FS 0 025 FS CC mode Low 0 3A 0 6A 0 3A Range High 0 30A 0 60A 0 ISA Low 0 1 mA Resolution High mA Low 0 05 0 05 FS Accuracy High 0 05 0 05 FS CR mode Low 0 05 Q 10 Q 0 3 Q 1002 Range High 10 Q 7 5 kQ Resolution 16 bit Low 0 01 0 08 S Accuracy High 0 01 0 0008 S CW mode Range 150 W 250 W 200 W Resolution 10 mW Accuracy 0 1 0 1 FS 0 2 0 2 FS 0 1 0 1 FS Transient mode CC mode TI amp T2 20 us 3600 s Resolution 10 us Accuracy S Us 100 ppm Low 0 001 2 5 A ms 0 001 A ms Slew Rate 2 High 0 001 2 5 A us 0 001 1 A us Electromagnetic compati
3. output signal Voltage on Von latch operation VA AA Load Sink Current t Control the input turn on state for the DC electronic load by configuring the Von latch function This can be used to start and stop discharging of a battery or other power source at a specified voltage level Oscilloscope display Delay 3 00ms MQ 5 00 vidi The 8600 Series can measure the rise or fall time from a specified start and stop voltage level of the measured input without the need for an oscilloscope This function can also be used as an internal timer to count how long the input has been enabled Voltage 0 6882 V p Seam ave ow O D e Current 1 4376 A a 1 ther pee um penne Power LIIW o Oo in Application software thew ia ae a ina inte Wik biii Pipe Lor Ma eee PC software is provided for front panel emulation generating and executing test sequences or logging measurement data without the need to write source code Additionally this application software integrates with NI Data Dashboard for LabVIEW apps which allows users to create a custom dashboard on a tablet computer or smartphone to remotely monitor 8600 Series DC loads via this PC software m Remote monitoring on iOS Android or Windows 8 compatible tablets or smartphones via NI Data Dashboard for LabVIEW apps m Log voltage current and power values with timestamp m Run transient operation and list mode
4. programs remotely m Create an unlimited number of external list files to be executed from PC memory Adjustable slew rate AA i J Slew Rate t In CC mode users can control the rate or slope of the change in current in a transient response test Set the slew rate to as slow as 0 001 A ms or as fast as 2 5 A us depending on the model and selected current range www bkprecision com Programmable DC Electronic Load Readback voltage Low O I8V O I18V 0 50V Range High 0 120V 0 120V 0 500 V Low 0 1 mV Resolution High mV Accuracy 0 025 0 025 FS Readback current Low O 3A O 6A O 3A Range High O 30A 0 60A 0 ISA Low 0 01 mA 0 1 mA 0 01 mA Resolution High 0 1 mA mA 0 1 mA Renae 0 05 0 05 0 05 y 0 05 FS 0 1 FS 0 05 FS Readback power Range 150 W 250 W 200 W Resolution 10 mW Accuracy 1 0 1 FS 0 2 0 2 FS 0 1 0 1 FS Protection range typical OPP 150 W 250 W 200 W Low 3 3 A 6 6 A 3 33 A OCP High 33 A 66 A 16 5 A OVP 120 V 120 V 500 V OTP 185 F 85 C General typical Short circuit Low 3 3 3 A 6 6 6 A 3 3 3 A Current CC High 33 30 A 66 60 A 16 5 15 A Voltage CV OV Resistance CR 35 mQ 30 mQ 300 mQ Input terminal impedance
5. Data Sheet Programmable DC Electronic Load 8600 Series INPUT CCCC c AAA ee The 8600 Series programmable DC electronic loads provide the performance of modular system DC electronic loads in a compact benchtop form factor With fast transient operation speeds and high 6 bit measurement resolution these standalone DC loads can be used for testing and evaluating a variety of DC sources such as DC power supplies DC DC converters batteries battery chargers and photovoltaic arrays The DC loads can operate in constant current CC constant voltage CV constant resistance CR or constant power CW mode and be configured to provide a dynamically changing load to the DC source with fast load switching times Versatile internal external and remote triggering options allow the dynamic load behavior to be synchronized with other events Increase productivity by saving your test parameters into any one of the 100 memory areas for Quick system recall All load parameters such as voltage current slew rate and width can be set via the front panel or programmed remotely The 8600 Series provides standard USB USBTMC compliant GPIB or RS 232 serial interfaces for remote communication To ensure the reliability of your testing the 8600 Series provides a power on system self test and numerous protection features overtemperature OTP overvoltage OVP overcurrent OCP overpower OPP and local remote rever
6. bility Meets EMC Directive 2004 108 EC EN 61000 3 2 2006 EN 61000 3 3 1995 A1 2001 A2 2005 EN 61000 4 2 3 4 5 6 11 EN 61326 1 2006 1 Fast pulse trains with large transitions may not be achievable 2 The slew rate specifications are not warranted but are descriptions of typical performance The actual transition time is defined as the time for the input to change from 10 to 90 or vice versa of the programmed current values In case of very large load changes e g from no load to full load the actual transition time will be larger than the expected time The load will automatically adjust the slew rate to fit within the range high or low that is closest to the programmed value 5 v012015 Dimensions W x H x D 8 5 x 3 5 x 14 214 5 x 88 2 x 354 6 mm Weight Standard accessories I 1 Ibs 5 kg Three Year Warranty User manual power cord certificate of calibration amp test report Optional accessories Model TLPWRI high current test leads IT EI51 rack mount kit www bkprecision com
7. l data subject to change B amp K Precision Corp 2015 www bkprecision com Programmable DC Electronic Load 8600 Series Bright dual line display The 8600 Series display shows both measured input values and set parameters simultaneously Front panel BK PRECISION 8600 120V J0A 150W DC E ectronic Load Rotary control knob High current test lead accessory model TLPWRI Numeric keypad Function keys Cursor keys Load input terminal Intuitive user interface The numeric keys and rotary knob provide a convenient interface for setting the operating mode and desired current voltage and resistance levels quickly and precisely Rear panel RS 232 USB Cooling fan interface interface GPIB interface AT 2S IM SEL I i a Po O f a f RSP332 use Cc IEER 488 SK je Remote Control Monitor DA Current monitor BNC output Outputs a 0 10 V signal that follows O full range of the input current I O terminal block Line voltage selector External trigger external analog programming external input On Off control voltage fault pin and remote sense terminals 2 www bkprecision com Programmable DC Electronic Load 8600 Series Flexible operation Transient operation A level B level Slew Rate Transient operation enables the load to periodically switch between two load levels A B A power supply s regulatio
8. n and transient characteristic can be evaluated by monitoring the supply s output voltage under varying combinations of load levels t duty cycle and slew rate The 8600 Series can simulate these conditions in CC CV CW and CR mode Low voltage operation List mode Not limited to just switching between two levels list mode lets you generate more complex sequences of input changes with several different levels Up to 7 groups of list files can be saved and executed in CC mode via internal or external trigger Each list can contain up to 84 steps with varying slew rates and a minimum width time of 20 Us per step The 8600 Series can operate at low voltages for applications such as fuel cell and solar cell testing Voltage Typical minimum operating voltage at full scale current 8600 8601 8602 L I V L I V 4 5 V Automatic test mode The 8600 Series can execute multiple test sequences in automatic test mode Up to 100 different sequences can be linked to run steps of various operating modes and loading conditions Each sequence can also be programmed with upper and lower limit Pass Fail criteria When applied in production testing you can easily judge whether the test parameters of your devices are within the specification limits and adjust your process according to the Pass Fail verdict CR LED mode
9. se voltage LRV RRV protection Special Applications The 8600 Series provides a built in battery test mode to measure the ampere hour Ah characteristic of a battery and a unique CR LED mode to simulate the loading behavior of a typical LED Features m Voltage range up to 500 V m Current range up to 60 A m CC CV CR CW operating modes m 16 bit voltage and current measurement system providing mV 0 1 mA resolution m Transient mode up to 25 kHz in CC mode m List mode function m Store and recall up to 100 setups m Adjustable slew rate in CC mode m Flexible triggering options via front panel external input timer or bus m Built in battery test function with voltage level capacity level and timer stop conditions m Test modes to validate the OCP OPP protection functions of a power supply m CR LED mode to simulate the loading behavior of typical LEDs m Remote sense m Analog current control and monitoring m Thermostatically controlled fan m Standard USB USBTMC compliant RS232 Power 150 W 250 W 200 W and GPIB interfaces supporting SCPI Operating Voltage 0 120V 0 120 V 0 500 V commands for remote control Rated Current 0 30A 0 60A O I5A m OVP OCP OPP OTP including local and remote reverse voltage LRV RRV protection m Compact 19 half rack form factor allows for side by side rack mounting of two units eu a Y rie ivi ELECTRONIC TEST INSTRUMENTS BK Technica