        242 TEC C-Mount LaserMount User`s Manual
         Contents
1. 543 1302 Fax   805  543 1303    sales arroyoinstruments com    www arroyoinstruments com    Copyright    2011  Arroyo Instruments  All Rights Reserved P N 530 1004 Rev E    
2. NG       The mount comes with a cover plate and nitrogen purge nipple that is  recommended when operating the device below the dew point  which is the  temperature what which moisture will begin to condensate  This is typically 10      to 15 C  but is dependent on your laboratory conditions  To prevent  condensation  after loading the device  mount the cover plate by using the  provided 4 40 flat head screws  and screw in the nitrogen purge nipple  if it is  not already installed   then connect a nitrogen source to the input nipple on the  side of the mount  A very low nitrogen flow is required to prevent condensation   typically 1 to 2 SCFH  standard cubic feet per hour      Your fixture is now ready for use  Additional technical information can be found  below     Pli arroyo Instruments    Page 6   242        C Mount LaserMount User s Manual    Connector Pin Outs       242 TEC C Mount LaserMount Rear Connectors       DB 9 Pin Description    1 amp 2 No Connection   3 Chassis Ground  GND   4 amp 5 Laser Cathode   6 Photodiode  PD  Cathode  7 Photodiode  PD  Anode  8 amp 9 Laser Anode    Laser DB 9 Connector Pin Out    DB 15 Pin Description    1  2   amp  9            3  4   amp  10            5  amp  6 Earth Ground  7 Thermistor   8 Thermistor   11 FAN       12 FAN       13 thru 15 No Connection    TEC DB 15 Connector Pin Out    Pli              Instruments    242 TEC C Mount LaserMount User s Manual   Page 7    Connection Description  Tip  12VDC  8 to 12VDC   Sleeve Fan Grou
3. USER S MANUAL              G MOUNT           arroyo Instruments    Page 2   242        C Mount LaserMount User s Manual    Introduction       Thank you for choosing the 242        C Mount LaserMount from Arroyo  Instruments  The 242 LaserMount is designed for high performance and long  term use     The 242 LaserMount integrates a large 120W Peltier cooler for precise control  and substantial heating and cooling capacity for your powerful devices    Featuring an integrated nitrogen purge  the 242 LaserMount has an operating  range of  5  C to  85  C  covering a broad range of temperature control needs     For electrical connection  the 242 LaserMount features an integrated lead  clamp  making electrical connection to the c mount lead a snap  Simply load the  c mount device into the fixture and turn the clamping knob     The 242 LaserMount comes standard with an integrated fan for additional  cooling capacity  When used with the 5300 Series TECSource temperature  controllers  no additional power supply is needed to power the fan  or use a  standard external 12V DC power supply when connecting to other temperature  controllers     The 242 LaserMount also offers all the features you would expect from a  modern c mount laser diode fixture  including     e Stainless steel alignment plate for easy device alignment during loading      Designed to be quickly integrated with Arroyo s LaserSource and  TECSource instruments    e  Industry standard D sub connectors and pin outs allow for 
4. cts  If you  use your own cables  ensure the connections are properly made  between the instrument and the mount  and that proper grounding  techniques are used  The pin out of the connectors can be found later  in this document     WARNING    Be sure you are properly ESD protected before handling your laser  For  additional information  read the section titled  Laser Diode Protection   later in this manual        Mounting your device     Start by turning the knob on the top of the fixture counter clockwise to open the  lead clamp  Make sure the clamp is high enough to allow sufficient space for  loading the device     The 242 LaserMount has a stainless steel alignment plate which is there to help  properly align the device to the mounting hole and clamping block  Using the  alignment plate to guide the device  place the device onto the mount and use  the 2 56 x 3 16  screw provided with the fixture to fasten the device to the fixture     After the device has been mounted  tighten the clamp by turning the knob on the  top of the fixture clockwise until the lead is help tightly against the clamp block   Do not over tighten the clamp or you may damage the clamping mechanism     The drawing and two photos below illustrate the unloaded configuration with the  clamp all the way open  left   and a device loaded and clamp tightened down on  the device lead  right      Pli arroyo Instruments    242 TEC C Mount LaserMount User s Manual   Page 5    LEAD CLAMP  KNOB         062 ID  TUBI
5. nd   12V Fan Connector Pin Out    Technical Specifications  _242 TEC C Mount LaserMount           242 TEC C Mount LaserMount  LASER PACKAGE SUPPORTED    Package C mount  TEMPERATURE CONTROL  Temperature Range    C   5 to  85  Sensor Type BetaTHERM 10      1     10kQ Thermistor  TE Module Imax   7 4A  Vmax   16 4V  Qmax   78W  INPUT CONNECTORS  Laser Diode DB 9  male  Temperature Controller DB 15  male  Fan 2 1mm round  8 to 12VDC  Nitrogen 1 8  barb  GENERAL  Recommended nitrogen flow 1 to 2 SCFH  Size without base  H x W x D   inimm   4 0  101 6  x 4 0  101 6  x 3 5  76 2   Mounting holes Slotted holes for 14 20  base     74 20 x 2  for post mount   M6  for post mount   4 40 x 4 for nitrogen cover plate    Calculating Temperature Using the Steinhart Hart Formula    The thermistor resistance can easily be converted to temperature using the  Steinhart Hart equation  and is expressed as         A  B  In R    C   In  R       Where R is the thermistor resistance in Ohms  and A  B  and C are listed below        Steinhart Hart Constants for Thermistor Temperature Calculation    A 1 129241x10 3  B 2 341077x10 4  C 8 775468x10 8    Pli arroyo Instruments    Page 8   242        C Mount LaserMount User s Manual    Mechanical Specifications               161 6           zno sna     4 00  100 6      4o 101 6      2 00  50 8      Front View    assolate                49 101 4         Detailed View of Device Load Area       200  558     Side View    Pli arroyo Instruments    242 TEC C Moun
6. ng other equipment  Ensure that any other equipment is properly  isolated to avoid any ground loop problems     Pli              Instruments    Page 10   242        C Mount LaserMount User s Manual       Arroyo Instruments warrants this product to be free from defects in material and  workmanship under normal use and service for a period of one  1  year from  date of shipment  It does not apply when the product has been misused  altered  or damaged by accident or abnormal conditions of operation  If found to be  defective during the warranty period  the product will either be repaired or  replaced at Arroyo Instruments s option     THIS WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES  EXPRESSED OR  IMPLIED  INCLUDING IMPLIED WARRANTIES OF MERCHANTABILITY OR  FITNESS FOR ANY PARTICULAR PURPOSE  ARROYO INSTRUMENTS SHALL  NOT BE LIABLE FOR ANY INDIRECT  SPECIAL  OR CONSEQUENTIAL  DAMAGES RESULTING FROM THE PURCHASE OR USE OF ITS PRODUCTS        Service and Support    For service and support  contact your local distributor or Arroyo Instruments     By mail  Arroyo Instruments  624 Clarion Court  San Luis Obispo  CA 93401    USA  By phone   1  805  543 1302  By fax   1  805  543 1303  By email  support arroyoinstruments com  On the web  http   www arroyoinstruments com    Pli              Instruments    242 TEC C Mount LaserMount User s Manual   Page 11    NOTES     Pl              Instruments        gt  arroyo instruments    624 Clarion Court  San Luis Obispo  CA 93401    Tel   805  
7. osition  However  if you to elect to connect the diode plate to the  case  it is critical that proper grounding techniques are used     Once you have properly configured the DIODE PLATE switch  if you will be  operating the mount then the only other item to setup is the power supply for the  internal cooling fan     If you are using an Arroyo Instruments 5300 Series TECSource temperature  controller  the fan supply is built directly into the TECSource  You will need to  enable the fan supply in the TECSource menu   see the TECSource manual for  additional details on how to do that     If you are using a third party temperature controller  then you will need to  provide a 12V DC power supply  The connection into the mount is a standard  2 1mm round plug with the center pin positive  and must be capable of  supplying at least 150mA     Pli              Instruments    Page 4   242 TEC C Mount LaserMount User s Manual    Connect to Laser Diode Driver and TEC Controller     Next  connect the 242 LaserMount to your laser diode driver and temperature  controller  Make sure the temperature controller s current limit is set to a  maximum value of 7 44     Where possible  we recommend the use of Arroyo Instruments laser and TEC  cables  Use p n 1262 TECSource Cable for the temperature controller  connection  and p n 1220 LaserSource Cable for the laser connection     Arroyo Instruments offers Laser and TEC cables designed to connect  directly between our LaserSource and TECSource produ
8. quick  integration into existing laser applications    e Banana plug for case ground  which can be used as a wrist strap  connection or to assure proper grounding to an optical table or test  bench        Pli              Instruments    242 TEC C Mount LaserMount User s Manual   Page 3    Installation and Use       The 242 LaserMount setup is very quick  simply follow the directions below   Configuring the mount     The only configuration of the 242 LaserMount is  determining if earth ground case is to be connected  to the diode  On the side of the 242  there is a switch  labeled DIODE PLATE  When the switch is in the  CASE position  then the diode plate is connected to  the case  and typically earth ground through the  instrument cables  of the fixture     For example  if the DIODE PLATE switch is set to  CASE  then the laser mounting plate  and therefore  the anode of the laser  will be tied to the case  If the  switch is set to N C  then the mounting plate will only  be connected to the anode connection of the laser  driver and no electrical connection to the case        In addition to the switch  there is a banana jack that provides a case connection  which can be used for a wrist strap  if the case is earth grounded  or to tie the  case to earth ground through something other than the laser driver     The purpose of the switch and banana jack is to control how your laser diode is  grounded  If no earth grounding is required  then the DIODE PLATE can be left  in the N C p
9. t LaserMount User s Manual   Page 9    Laser Diode Protection       Electrostatic discharge and current spikes can be a significant cause of damage  to laser diodes  but when proper precautions are taken  these risks can be  greatly reduced or eliminated  Arroyo Instruments  controllers offer state of art  laser diode protection  but no instrument can fully shield the laser from damage   Please take these considerations into account when operating your laser     1  Always set the current limit at or below the maximum current your laser  can handle  This prevents the device from accidentally driving the  current too high  either via the set point or from the modulation port   This also provides additional current limiting protection from ESD    2  Always work in an ESD safe operating environment  including the use  of wrist straps  ESD grounded work surfaces and floors  and ESD safe  tools    3  Where the AC power to the laser driver to temperature controller may  be noisy  use isolation transformers or uninterruptible power supplies  that provide isolation    4  Make sure all cables are securely connected and fastening screws are  screwed in tight    5  Do not route power cords or other cables in parallel with the laser or  temperature controller cables  as coupling may occur between the  cables and inject noise into the laser diode    6  While it is not possible to create a ground loop through the  LaserSource because of it s isolation of all inputs  it is possible when  usi
    