Home
        HOBO UX120-006M Data Logger Manual PDF
         Contents
1.   0 1  of  0 3  of  0 3  of  0 3  of  reading reading reading reading reading  Resolution 0 3 uA 40 uV 80 uV 160 uV 384 uV  Logger    Operating Range    Logging Rate  Logging Modes  Memory Modes  Start Modes  Stop Modes  Restart Mode  Time Accuracy  Power Source    Battery Life    Memory  Download Type    Full Memory Download  Time    LCD    Size  Weight    Environmental Rating    CE    Logging   20   to 70  C   4   to 158  F   O to 95  RH  non condensing    Launch Readout  0   to 50  C  32   to 122  F  per USB specification    1 second to 18 hours  12 minutes  15 seconds   Fixed interval  normal   burst  or statistics   Wrap when full or stop when full   Immediate  push button  date  amp  time  or next interval   When memory full  push button  or date  amp  time   Push button    1 minute per month at 25  C  77  F   see Plot A   Two AAA 1 5 V alkaline batteries  user replaceable  and USB cable    1 year  typical with logging rate of 1 minute and sampling interval of 15  seconds or greater    4 MB  1 9 million measurements  maximum   USB 2 0 interface    Approximately 1 5 minutes    LCD is visible from 0   to 50  C  32   to 122  F   the LCD may react slowly or  go blank in temperatures outside this range    10 8 x 5 41 x 2 54 cm  4 25 x 2 13 x 1 in    107 5 g  3 79 oz   IP50    The CE Marking identifies this product as complying with all relevant  directives in the European Union  EU      Note  The HOBO U Shuttle  U DT 1  is not compatible with this logger     HOBO 4 Channel A
2.   Some sensors  such as temperature sensors  can be connected  directly to the external input jacks  but others require  additional cables as described in the following sections     4 20mA Input Cable    The 4 20mA input cable  CABLE 4 20mA  measures current  from 0 to 20 1 mA  Do not expose to current above 20 mA or to    www onsetcomp com    negative current  Do not cut off the end of the gray cable where  it connects to the blue and yellow wires as it contains the  precision resistor required for current measurement     Voltage Input Cable   The logger   s external inputs can accept the voltage input cable   CABLE 2 5 STEREO   which allows a voltage to be recorded   The input line must not be exposed to signals below O V or  above 2 5 V  The voltage input cable connections are as follows     Wire Connection   Red Switched 2 5 V output  White Voltage input   Black Ground    Switched 2 5 V Output   The external input channels have a switched 2 5 V output  This  signal can be used to power a sensor directly or to trigger an  external circuit  External sensors should draw no more than 4  mA total when powered     The switched 2 5 V output turns on about 5 ms before the  external channels are measured as shown in the following  diagram  The shaded area shows the 100 ms period for each  enabled channel during which the logger samples the input  signals     Ch1 Ch2 Ch3 Ch4  100ms   100ms   100ms 100ms gt     ims    When using multiple voltage and or current inputs  the     from  
3.  mounting materials  see Mounting the  Logger   After logging begins  you can read out the logger at  any time  see Reading Out the Logger for details      Connecting External Sensors    The logger can accept up to four external sensors  refer to  onsetcomp com for a current list of supported sensors   Plug  each sensor into one of the four input jacks  making sure each  sensor is firmly seated in the appropriate numbered jack based  on how you configured that corresponding channel in the  Launch Logger window  For example  if you selected    TMCx   HD    for sensor 1 in the Launch Logger window  then you must  plug the TMCx HD temperature sensor into the port labeled    1     on the logger otherwise the logger will not record the correct  data  Connect each sensor before logging begins  Refer to the  sensor or cable manual for more information on connecting the  sensor and wiring  if applicable     If you disconnect a sensor or if it is not fully inserted into the  jack while the logger is logging  an erroneous sensor reading  can appear on the LCD for that channel  In addition  erroneous  readings will be logged and saved in the data file depending on  the logging interval  e g  if a sensor is disconnected for 5  minutes and the logging interval is set to 1 minute  then there  will be 5 erroneous data points while the sensor was  disconnected   If you reconnect the sensor  the correct values  will display on the LCD again and will be logged and saved in  the data file      
4.  the actual  sensor values every 5 minutes  In addition  the logger will take  a sample every 30 seconds and temporarily store it in memory   The logger will then calculate the maximum  minimum   average  and standard deviation using the samples gathered  over the previous 5 minute period and log the resulting  value s   When reading out the logger  this would result in 5  data series for each channel  one sensor series  with data  logged every 5 minutes  plus four maximum  minimum   average  and standard deviation series  with values calculated  and logged every 5 minutes based on the 30 second sampling      To set up statistics     1  Select Statistics for Logging Mode in the Launch Logger  window  If Statistics has already been configured for this  logger  click the Edit button in the Launch Logger window     2  Click the Maximum  Minimum  Average  and Standard  Deviation checkboxes for each of the statistics you want to  calculate during logging  Note that Average is automatically  enabled when selecting Standard Deviation  In addition  the  more statistics you record  the shorter the logger duration  and the more memory is required     Statistics       How does Statistics differ from Filters   For each enabled sensor  log the   Madmum    Minimum    v  Standard Deviation  requires Average  Sampling every  30 seconds v  The sampling interval must be less than and a factor of the logging interval  Logging Interval  1 minute    Logging Duration  83 1 days       Help Cancel O
5.  the high or low  alarm that you configured  Press and release the  Alarm Stats button until the    alm    symbol   described later in this chart  is displayed on the  screen  This symbol at left will clear depending on  how alarms were configured in HOBOware  If the  alarm was configured to clear when the logger is  relaunched  this symbol will remain on the LCD   Otherwise  it will clear when the sensor reading is  back within the alarm limits or by pressing the  Alarm Stats button for 3 seconds     This is the channel number associated with the  sensor reading  channel 1 for this example   Up to  four channels are visible at one time     This is an example of the units entered for the  sensor  which appears to the right of the channel  number  The unit type is determined by what was  entered in the LCD Units field for that sensor in  HOBOware  See Setting up the Logger for more  details  Note that units for temperature sensors  are displayed as   F or   C only     The logger has been configured to start logging on  a particular date time  The display will count down  in days  hours  minutes  and seconds until logging  begins  In this example  5 minutes and 38 seconds  remain until logging will begin     www onsetcomp com    HOBO 4 Channel Analog Data Logger  UX120 006M  Manual    LD Symbol Description                           Setting up the Logger    LOGGING TOE DEBEL I CUN ENUY ORINE Use HOBOware to set up the logger  including setting alarms     TAS OBBE Nas VEAN CONMEU
6. HOBO   4 Channel Analog Data Logger  UX120 006M  Manual    onsete       HOBO   4 channe  analog logger    HOBO 4 Channel Analog  Data Logger    UX120 006M    Included Items     e Command    strip   e Double sided tape   e Hook  amp  loop strap   e Two AAA 1 5 V alkaline  batteries    Required Items     e HOBOware 3 6 or later  e USB cable  included with  software     Sensors and cables available at  www onsetcomp com     17384 D    The HOBO 4 Channel Analog data logger has 16 bit resolution and can record up to 1 9 million  measurements or events  The four external channels accept a variety of sensors  including  temperature and split core AC current sensors as well as 4 20 mA and voltage input cables  sold  separately   Using HOBOware    you can easily configure an alarm to trip when the sensor  reading rises above or falls below a measurement that you specify  Or  you can set up burst  logging in which the logger records data at a different interval during certain conditions  The  logger can also calculate minimum  maximum  average  and standard deviation statistics  This  easy to use data logger has a built in LCD screen to check current readings and to monitor  logging status  battery use  and memory consumption in between readouts     Specifications  Logger with CABLE  CABLE 2 5  CABLE  CABLE  CABLE   Cable Type 4 20mA STEREO ADAP5 ADAP10 ADAP24  Measurement O to 20 1 mA Oto2 5V 0to5 0V Oto 10 V Oto 24V  Range  Accuracy  0 001 mA  0 1 mV  0 2 mV  0 4 mV  1 0 mV   0 2  of
7. K    3  Set the sampling interval  which must be less than and a  factor of the logging interval  Choose either a preset  sampling interval or select Custom and enter your own  sampling interval  Keep in mind that the more frequent the  sampling rate  the greater the impact on battery life     4  Click OK when done  This will return you to the Launch  Logger window  Click the Edit button next to Logging Mode  in the Launch Logger window to make additional changes     Once logging begins  click the Alarm Stats button on the logger  to cycle through the current maximum  minimum  average  and  standard deviation data on the LCD screen  You can plot the  Statistics series once you read out the logger     Reading Out the Logger    To read out the logger  connect it to the computer with a USB  cable  In HOBOware  select Readout from the Device menu  You  can then save the data  plot it  and export it for further analysis   Refer to the HOBOware Help for details     Battery Information    The logger requires two user replaceable AAA 1 5 V alkaline or  optional lithium batteries for operation at the extreme ends of  the logger operating range  Expected battery life varies based    1 800 LOGGERS    HOBO 4 Channel Analog Data Logger  UX120 006M  Manual    Recording Internal Logger Events    The logger records the following internal events to track logger  operation and status  You can plot these events in HOBOware  after reading out the logger and opening the data file     Internal Eve
8. TEO TOBLOPAOLRING selecting the options to start and stop logging  and choosing a  when memory fills  The memory bar indicates the logging mode    approximate space remaining in the logger to    DAMB MEM record data  When first launched  all five segments 1  Install the batteries  See Battery Information for details   in the bar will be empty  In this example  the  logger memory is almost full  only one segment in 2  Connect the logger and open the Launch Logger window   the memory bar is empty   To connect the logger to a computer  use the USB cable  The logger has been configured to never stop provided  Click the Launch icon on the HOBOware toolbar  logging  wrapping   The logger will continue or select Launch from the Device menu   recording data indefinitely  with newest data   _ overwriting the oldest data  When first launched  Important  USB 2 0 specifications do not guarantee   SSEEQMEM all five segments in the memory bar will be empty  operation outside the range of 0  C  32  F  to 50  C  122  F         In this example  the memory is full  all five  segments are filled in  and new data is now  overwriting the oldest data  This will continue until  the logger is stopped or the battery runs out     3  Configure a sensor  Under Configure Sensors to Log  click  the checkbox for sensor 1  Select the type of sensor or cable    These symbol thowthemadmum minitum  that will be connected to channel 1 on the logger  Type a             max average  and standard deviation values 
9. d record data every 10 seconds  Once the  temperature rises back to 32  F  the logger will then return to  normal mode  logging every 5 minutes  To set up burst logging     1  Select Burst for Logging Mode in the Launch Logger  window  If Burst has already been configured for this  logger  click the Edit button in the Launch Logger window     2  Inthe Burst Logging window  select a sensor from the list  In  the following example  a temperature sensor was selected     Burst Logging S    While any of these condition s  are met     Sensor   OC Voltage  V  3   Temperature    F  38  Temperature    F   f       Max  212 000   F        V  High Limit 85 000  v  Low Limit 32 000      Limt value s  are set to closest  value supported by logger     Min   40 000   F If    Log data at the burst logging interval     30 seconds v    The burst logging interval must be less than the logging interval    Logging Interval  5 minutes    Logging Duration  177 6 days minimum  4 9 years maximum    Cancel  ox             3  Select the High Limit checkbox if you want to set up a  condition in which burst logging will occur when the sensor  reading rises above the high limit value  Type in the value or  drag the red upper slider     1 800 LOGGERS    HOBO 4 Channel Analog Data Logger  UX120 006M  Manual    4  Select the Low Limit checkbox if you want to set up a    condition in which burst logging will occur when the sensor  reading falls below the low limit value  Type in the value or  drag the blue low
10. de     For example  let   s assume  the logger has a 10 minute logging interval and logged a  data point at 9 05  Then  the high limit was surpassed  and burst logging began at 9 06  Burst logging then  continued until 9 12 when the sensor reading fell back  below the high limit  Now back in normal mode  the next  logging interval will be 10 minutes from the last burst  logging point  or 9 22 in this case  If burst logging had not  occurred  the next data point would have been at 9 15     e A New Interval event will appear on the plot  if you select  events for plotting in the Plot Setup window  each time  the logger enters or exits burst logging mode     Statistics    Statistics is a logging mode in which the logger calculates  maximum  minimum  average  and standard deviation statistics  during logging  recording the results at each logging interval  based on samples taken at a rate you specify  This will result in  up to four additional series per sensor that record the following  information at each logging interval     e The maximum  or highest  sampled value   e The minimum  or lowest  sampled value   e An average of all sampled values  and    e The standard deviation from the average for all sampled  values     www onsetcomp com    For example  let   s say the logging interval is set to 5 minutes  and the sampling interval is set to 30 seconds  with maximum   minimum  average  and standard deviation all enabled   Once  logging begins  the logger will measure and record
11. ed on  the sensor type  This means the value that triggers the  alarm may differ slightly than the value entered     e When you read out the logger  high and low alarm levels  will be displayed on the plot along with    Chan  lt   gt  Alarm  Tripped    and    Chan  lt   gt  Alarm Cleared    events showing  when the alarm tripped and cleared  The    Chan  lt   gt   Alarm Cleared    event contains the value that was  furthest out of range for the sensor before the alarm  cleared  see the Points table for the actual value      Burst Logging    Burst logging is a logging mode that allows you to set up more  frequent logging when a specified condition is met  For  example  let   s say the logger has a temperature sensor  connected to channel 1 and is recording data at a 5 minute  logging interval  Burst logging is configured on channel 1 to log  every 10 seconds when the temperature goes above 85  F  the  high limit  or falls below 32  F  the low limit   This means the  logger will record data every 5 minutes as long as the  temperature remains between 85  F and 32  F  Once the  temperature reaches 90  F  for example  the logger will switch  to the faster logging rate and record data every 10 seconds  until the temperature falls back below the high limit  or 85  F in  this case   At that time  logging then resumes every 5 minutes  at the normal logging interval  Similarly  if the temperature falls  to 30  F  for example  then the logger would switch to burst  logging mode again an
12. er slider     Repeat steps 2 through 4 for any additional sensors     Set the burst logging interval  which must be less than the  logging interval  Select either a preset burst logging interval  or select Custom and enter your own interval  Keep in mind  that the more frequent the burst logging rate  the greater  the impact on battery life and the shorter the logging  duration     7  Click OK when done  This will return you to the Launch    Logger window  Click the Edit button next to Logging Mode  in the Launch Logger window to make additional changes     Notes     e Once the logger is launched  the high and low burst  logging limits are only checked when the logger s  LCD screen refreshes once every 15 seconds  Therefore   if you set the logging interval to less than 15 seconds and  the sensor reading falls outside the limits  the burst  logging will not begin until the next 15 second refresh  cycle     e If high and or low limits have been configured for more  than one sensor  then burst logging will begin when any  high or low condition goes out of range  Burst logging will  not end until all conditions on all sensors are back within  normal range     e The actual values for burst logging limits are set to the  closest values supported by the logger based on the  sensor type     e Once the high or low condition clears  the logging  interval time will be calculated using the last recorded  data point in burst logging mode  not the last data point  recorded in    normal mo
13. eries   3  Insert two new batteries observing polarity     4  Reinsert the battery door and snap it back into place     Mounting  magnet        a Mounting    Battery door magnets    A WARNING  If using optional lithium batteries  do not cut  open  incinerate  heat above 85  C  185  F   or recharge the  lithium batteries  The batteries may explode if the logger is  exposed to extreme heat or conditions that could damage or    destroy the batteries case  Do not dispose of the logger or  batteries in fire  Do not expose the contents of the batteries to  water  Dispose of the batteries according to local regulations  for lithium batteries        HOBOware provides the option of recording the current battery  voltage at each logging interval  which is disabled by default   Recording battery life at each logging interval takes up memory  and therefore reduces logging duration  It is recommended you  only record battery voltage for diagnostic purposes         onset    2013 2015 Onset Computer Corporation  All rights reserved  Onset  HOBO  and HOBOware are    1 800 LOGGERS  564 4377     508 759 9500 trademarks or registered trademarks of Onset Computer Corporation  All other trademarks are the  HOBO Data Loggers www onsetcomp com    loggerhelp onsetcomp com property of their respective companies   17384 D    
14. igh Alarm checkbox or drag  the red upper slider in the Configure Alarms window     Select the Low Alarm checkbox if you want an alarm to trip  when the sensor reading falls below the low alarm value   Type the reading next to the Low Alarm checkbox or drag  the blue lower slider     Set the duration before an alarm is tripped     Select either Cumulative or Consecutive  If you select  Cumulative  then the alarm will trip when the time the  sensor is out of range over the course of the deployment is  equal to the selected duration  If you select Consecutive   then the alarm will trip when the time the sensor is  continuously out of range is equal to the selected duration   For example  the high alarm for temperature is set to 85  F  and the duration is set to 30 minutes  If Cumulative is  selected  then an alarm will trip once a sensor reading has  been at or above 85  F for a total of 30 minutes since the  logger was configured  specifically  this could be 15 minutes  above 85  F in the morning and then 15 minutes above 85  F  again in the afternoon  If Consecutive is selected  then an  alarm will trip only if all sensor readings are 85  F or above  for a continuous 30 minute period     Configure Alarms em    Sensor Alarm Settings    Sensor   DC Voltage  V   Temperature    F     Temperature    F     Max  212 000   F      ay Duration of out of  range samples      displayed before alarm is raised     v High Alarm 95 000  gt  00 Hr 00 Min    V Low Alarm 25 000       Alarm vale 
15. is set to closest    value supported by logger    Min   40 000   F        Sensor Alarm Mode    Cumulative  Raise alarm when sensor is out of range for a particular number of  samples  selected above      Consecutive  Raise alarm when sensor is out of range consecutively for a particular    2 number of samples  selected above      Additional Alarm Settings  Maintain Visual Alarm Until  Host has relaunched logger    Sensor reading within limits    Cleared with button press       Help Cancel OK          Repeat steps 2 through 6 for any additional sensors     Choose how long the logger should maintain the sensor  alarm once it has tripped  Select    Host has relaunched  logger    if you want the alarm to remain visible on the LCD  until the next time you relaunch the logger  Select    Sensor  reading within limits    if you want the alarm to clear once  the sensor reading returns to the normal range between  the high and low alarm limits  Select    Cleared with button  press    if you want the alarm to remain on until you press  the Alarm Stats button on the logger     Click OK to save alarm settings     Notes     e Once the logger is launched  alarms will trip as  determined by these settings  Logger alarms will display  on the LCD screen  Note that the alarm limits are only    www onsetcomp com    checked when the logger   s LCD screen refreshes every 15  seconds     e The actual values for the high and low alarm limits are set  to the closest values supported by the logger bas
16. most label for the sensor if desired  Be sure to connect the sensor  min recently calculated by the logger  if the logging before logging begins   mode has been set to Statistics in HOBOware   4  Set up scaling  optional   You can configure some sensors  avg Press the Alarm Stats button for 1 second to cycle to scale logged data into different values and units than the  sdo mone  Ede eet iE ANVANAPIE StALSUCS y Any aarm default  If the sensor supports scaling  click the Scaling  readings  and back to the current sensor reading  l      button and type in the appropriate values and units  consult  This is the sensor reading that tripped the alarm  the sensor user manual for recommended scaling factors    Press the Alarm Stats button to view this reading  f    alm Prese iken arms toie buton arain oele Click Save and return to the Launch Logger window   through any statistics and return to the current 5  Set the LCD units  optional   Each sensor has its own  readings  default units that will appear on the logger LCD  Type in up  The launch settings are being loaded onto the to 3 characters if you want a different unit name to appear  1 afd logger from HOBOware  Do not disconnect the on the LCD than the default   Units for temperature sensors  USB cable during this process  are F or C and cannot be changed   Note that if you have  An error occurred while loading the launch configured Scaling for the sensor  then the scaled units     _   _ configurations onto the logger from HOBOwa
17. nal   Click the Alarms button if you  want to configure an alarm to trip when the sensor reading  is above or below a value you specify  See Setting up Alarms  for details     e When the logger has stopped logging  the LCD screen will  remain on until the logger is offloaded to a computer or   unless launched with the    Turn LCD off    option   Once  the logger has been offloaded and disconnected from the  computer  the LCD will turn off automatically after 2  hours  The LCD will turn back on the next time the logger  is connected to the computer     7  Configure filters  optional   Click the Filters button to  create additional filtered data series  Any filtered series will  be available automatically upon reading out the logger     1 800 LOGGERS 3 www onsetcomp com    8  Configure additional sensors  Repeat steps 3 through 7 to  configure up to three more sensors     9  Select the Logging Interval  Select a logging interval from 1  second to a maximum of 18 hours  12 minutes  and 15  seconds     10  Select the Logging Mode     e Fixed Interval  In Fixed Interval mode  data will always be  recorded at the regular logging interval set in the  previous step  This is the default setting     e Burst  In Burst mode  logging will occur at a different  interval when a specified condition is met  See Burst  Logging for more information     e Statistics  In Statistics mode  maximum  minimum   average  and standard deviation statistics are calculated  for the temperature during loggi
18. nalog Data Logger  UX120 006M  Manual    Specifications  continued     LCD Screen  This logger is equipped with an LCD screen that    displays details about the current status  This example shows all             25 0  0 0     a  a  T  25 0  S  i A  m        50 0  amp  CH1  a            75 0  h    100 0  CH3   20 25 25 47 5 70  amp     Temperature    C     Plot A  Time Accuracy         LCD Symbol     START  Logger Components and Operation  Start Stop Button  STOP  Alarm Stats Button  USB Port  CLEAR  LCD Screen a maw          External Inputs    3    Start Stop Button  Press this button for 3 seconds to start or  stop logging data  or to resume logging on the next even  logging interval  This requires configuring the logger in  HOBOware with a push button start or stop  and with    Resume  logging on next button push    selected  see Setting up the  Logger   You can also press this button for 1 second to record  an internal event  see Recording Internal Logger Events  or to  turn the LCD screen on if the option to turn off the LCD has  been enabled  see Setting up the Logger      Alarm Stats Button  Use this button to clear a tripped alarm   see Setting up Alarms  or to switch between statistics  alarm  readings  and the current sensor reading     Mounting Loops  Use the two mounting loops  only one visible  in the diagram  to mount the logger with the hook and loop    strap  see Mounting the Logger   a NA  Me    External Inputs  Use these 2 5 mm jacks  not visible in the  diag
19. ng  sampling at an  interval you specify  See Statistics for more information     11  Choose when to start logging   e Now  Logging begins immediately     e At Interval  Logging will begin at the next even interval as  determined by the selected logging interval     e On Date Time  Logging will begin at a date and time you  specify     e Push Button  Logging will begin once you press the  Start Stop logging button for 3 seconds     12  Choose when to stop logging     e When Memory Fills  Logging will end once the logger  memory is full     e Never  Wrap When Full   The logger will continue  recording data indefinitely  with newest data overwriting  the oldest     e Push Button  Logging will end once you press the  Start Stop logging button for 3 seconds  Note that if you  also choose Push Button to start logging  then you will  not be able to stop logging until 30 seconds after logging  begins     If you select the Push Button setting  then you also have  the option to select    Allow button restart     This allows  you to stop and then restart logging during the  deployment by pushing the Start Stop button on the  logger for 3 seconds     Important  When    Allow button restart    is selected and  you use the Start Stop button to stop and restart logging   logging will restart on the next even logging interval  not  at the time the button was pushed  For example  a logger  started logging at 7 00 AM with a logging interval set to 1  hour  If you press the Start Stop button t
20. nt Name Definition    Host Connected The logger was connected to the computer     Started The Start Stop button was pressed to begin  logging   Stopped The logger received a command to stop    recording data  from HOBOware or by  pushing the Start Stop button      Button Up Button The Start Stop button was pressed for  Down 1 second     Chan  lt   gt  Alarm  Tripped    An alarm has tripped on that channel     Chan  lt   gt  Alarm An alarm has cleared on that channel  This   Cleared event also contains the value that was  furthest out of range for the sensor before  the alarm cleared     The logger has entered or exited burst  logging mode     New Interval    Safe Shutdown The battery level dropped below 1 85 V  the    logger performs a safe shutdown     Mounting the Logger    There are several ways to mount the logger using the materials  included     e Use the four built in magnets on the back of the logger to  mount it to a magnetic surface     e Attach the Command strip to the back of the logger to  mount it a wall or other flat surface     e Use the double sided tape to affix the logger to a surface     e Insert the hook and loop strap through the mounting  loops on both sides of the logger to mount it to a curved  surface  such as a pipe or tubing     Protecting the Logger    The logger is designed for indoor use and can be permanently  damaged by corrosion if it gets wet  Protect it from  condensation  If the message FAIL CLK appears on the LCD  screen  there was a fail
21. o stop the  logger at 8 45 AM and then press the button again at  10 15 AM  logging will not begin immediately at 10 15   Instead  logging will begin again at 11 00 AM  which is the  next even interval time based on your 1 hour logging  interval  Therefore  depending on the logging interval   the gap between the time you press the button to  resume logging and the time actual logging begins could  be significant  The faster the logging interval  the less  time will elapse before logging resumes     1 800 LOGGERS       HOBO 4 Channel Analog Data Logger  UX120 006M  Manual    e Specific Stop Time  Logging will end at a date and time  you specify  Note that if you also configure the logger for  a Push Button stop and to    Allow button restart     then  the logger will stop logging at the date you select  regardless of how many times you stop and restart the  logger with the Start Stop button     13  Choose whether to keep the LCD on or off  By default  the  LCD will always remain on while logging  If you select the     Turn LCD off    checkbox  the LCD will not show the current  readings  status  or other information while the logger is  logging  You will  however  be able to temporarily turn the  LCD screen on by pressing the Start Stop button for 1  second if you select this option     14  Click the Start button to launch the logger  Note that the  Start button text changes based on the Start Logging  selection  Disconnect the logger from the computer and  deploy it using the
22. ram  to connect up to 4 sensors  see Connecting External  Sensors      USB Port  Use this port to connect the logger to the computer  via USB cable  see Setting up the Logger and Reading Out the  Logger      i    1 800 LOGGERS 2    A DA A TA AE Jafa fafafa re   Theberge bebe be    LOGGING    symbols illuminated on the LCD screen followed by definitions  of each symbol in the table     CLEAR    WA AVA BVA    WA AVA N  A WANG DAN CH2 yi Wid DAN       _           A   ohms MMM  ur yd WN WAN    maxminavasdoalm    NOOO ceo occ  UICC UC    Description    The logger is waiting to be launched  Press and  hold the Start Stop button for 3 seconds to launch  the logger     The logger has been launched with push button  stop enabled  press and hold the Start Stop   button for 3 seconds to stop the logger  Note  If  you also launched the logger with a push button  start  this symbol will not appear on the display for  30 seconds     An alarm is ready to be cleared  This will only  appear if    Cleared with button press    was selected  in the HOBOware alarm settings  Press the  Alarm Stats button for 3 seconds to clear the  alarm     The battery indicator shows the approximate  battery power remaining     This is an example of a temperature reading from  a temperature sensor  Temperature units are  determined by the settings in HOBOware  To  switch between Celsius and Fahrenheit  change  the Display Preferences in HOBOware before  launching the logger     A sensor reading is above or below
23. re  name will appear in the Launch Logger window for the  mr Make sure the USB cable is connected to both the sensor instead of its default unit  You can still override this  ORBEK ana Ine compuler ana ty aunching again  by entering a new name in the LCD units field    gt  a The logger has been stopped with HOBOware or a  Jeu because the memory is full  reach      _  Notes  a  e You can disable the LCD screen when logging  Select E e a     Turn LCD off    when setting up the logger as described in   Sensors  the next section  When this option is enabled  you can   feo          oe a eA  still temporarily view the LCD screen by pushing the 7  2   C CABLE ADAP24  0 24 Volts DC  ed    lt Enter label here gt  LCD units  V eSa  Start Stop button for 1 second  The LCD will then remain _ EE      on for 10 minutes      m             e The LCD screen refreshes every 15 seconds while logging a  regardless of the logging interval selected in HOBOware  Lopgng beer  Aminin  If you choose a logging interval less than 15 seconds  the PEPEE erre  data will be recorded at the faster interval  but the sensor Start Logging   sa BoSS  readings will only be updated on the screen every 15 d e a  seconds    afer  e If asensor is disconnected during logging  erroneous   ees       Skip launch window neat time Cancel Start    values will display for that sensor on the LCD and return we  to normal readings once reconnected  See Connecting  External Sensors for more details        6  Set up alarms  optio
24. the current source s  and the O V line from the voltage  source s  are tied together at the logger  If these lines are at  different voltage potentials  this may cause inaccurate readings  or even damage your logger  Keep in mind that these lines may  also be tied to earth ground through the USB cable when the  logger is connected to the computer  Special precautions may  be necessary if any of the voltage or current source common  lines are not tied to earth ground  Input isolators may be  needed in industrial environments to prevent errors caused by  ground loops     A WARNING  Analog channel input cannot exceed 2 5 V DC     For sensor outputs up to 24 V DC  use the appropriate voltage  adapter cable  CABLE ADAPX         Setting up Alarms    You can set an alarm to trip when a sensor reading rises above  or falls below a specified value on any of the four sensor  channels  To set an alarm     1  Click the Alarms button from the Launch Logger window  If  the Alarms button is disabled  make sure the Logging Mode  is not set to Burst   Alarms can only be configured if the  logger is in Normal or Statistics mode      2  Inthe Configure Alarms window  select a sensor from the  list  In the example in this section  a temperature sensor  was selected     1 800 LOGGERS    9     HOBO 4 Channel Analog Data Logger  UX120 006M  Manual    Select the High Alarm checkbox if you want an alarm to trip  when the sensor reading rises above the high alarm value   Type the reading next to the H
25. ure with the internal logger clock  possibly due to condensation  Remove the batteries  immediately and dry the circuit board     Note  Static electricity may cause the logger to stop logging   The logger has been tested to 8 KV  but avoid electrostatic  discharge by grounding yourself to protect the logger  For more  information  search for    static discharge    in the FAQ section on  onsetcomp com     on the ambient temperature where the logger is deployed  the  logging or sampling interval  frequency of offloading to the  computer  number of channels that are active  if burst or  statistics logging modes are active  and battery performance   New batteries typically last 1 year with logging intervals greater    www onsetcomp com    HOBO 4 Channel Analog Data Logger  UX120 006M  Manual    than 1 minute  Deployments in extremely cold or hot  temperatures  a logging interval faster than 1 minute  ora  sampling interval faster than 15 seconds can impact battery  life  Estimates are not guaranteed due to uncertainties in initial  battery conditions and operating environment     The logger can also be powered by the USB cable when the  remaining battery voltage is too low for it to continue logging   Connect the logger to the computer  click the Readout button  on the toolbar  and save the data as prompted  Replace the  battery before launching the logger again     To install or replace the batteries     1  Open the battery door on the back of the logger   2  Remove any old batt
    
Download Pdf Manuals
 
 
    
Related Search
    
Related Contents
  STIHL RE 98  Singer 14T970C Instruction Manual  Video Projector  TeraStation User Manual  Fellowes PS60-2 User's Manual  Operating Instructions      Mark-4, Elite-4,5, 7,9 Operation Manual -DE    Copyright © All rights reserved. 
   Failed to retrieve file