Home
        RevB 818 Series Manual
         Contents
1.  satura   tion described immediately above should be used whenever pulsed power  measurements are being made  Alternatively  when the detector is used for  observing pulse shape  placing an attenuator in the beam should affect only  the pulse amplitude  not the pulse shape     The photodetector surface  window material and the attenuator all reflect  light  The amount of reflected light depends upon the angle of incidence  and the polarization of the beam  Reflected light does not get absorbed by  the detector  and therefore is not included in the detector signal  The New   port detector and attenuator calibration include the loss due to reflection  for incoherent light incident normal to the detector  For accurate power  measurements the detector should therefore be used at near normal inci   dence     When a photon is absorbed in the photodiode  an electron hole pair is  formed within the device and a voltage is developed across the diode junc   tion  If the photodiode terminals are connected a photocurrent propor   tional to the light intensity will be generated  Measuring this photocurrent  provides a measurement of the optical power incident upon the detector     Figures 3 a  and 3 b  show common methods of measuring the photocur   rent generated by the photodiode  In circuit  a  a bias voltage is used to  drive the current through a load resistor and the voltage drop is measured   This is called the photoconductive mode of operation  The bias voltage en   hances the spe
2.  trade secrets of Newport  and shall  devote its best efforts to ensure that all its personnel protect the Newport Programs as trade secrets of  Newport Corporation  Customer shall not at any time disclose Newport s trade secrets to any other person   firm  organization  or employee that does not need  consistent with Customer   s right of use hereunder  to  obtain access to the Newport Programs and Related Materials  These restrictions shall not apply to  information  1  generally known to the public or obtainable from public sources   2  readily apparent from  the keyboard operations  visual display  or output reports of the Programs  3  previously in the possession  of Customer or subsequently developed or acquired without reliance on the Newport Programs  or  4   approved by Newport for release without restriction     Service Information  This section contains information regarding factory service for the source  The user should not attempt any    maintenance or service of the system or optional equipment beyond the procedures outlined in this manual   Any problem that cannot be resolved should be referred to Newport Corporation        Technical Support Contacts    North America    Asia Europe   Newport Corporation Service Dept  Newport MICRO CONTROLE S A   1791 Deere Ave  Irvine  CA 92606 Zone Industrielle   Telephone   800  222 6440 Beaune la Rolande  FRANCE    Telephone   33  02 38 40 51 56  Asia  Lot J3 8  Wuxi Export Processing Zone  Jiangsu Province  P R  214028  Ch
3. 818 Series Photodetector Guide       User s Manual    DD    Nevvport     Experience   Solutions       818 Series  Photodetector Guide    ii       Warranty    Newport Corporation warrants that this product will be free from defects in material and workmanship and  will comply with Newport s published specifications at the time of sale for a period of one year from date of  shipment  If found to be defective during the warranty period  the product will either be repaired or replaced  at Newport s option     To exercise this warranty  write or call your local Newport office or representative  or contact Newport  headquarters in Irvine  California  You will be given prompt assistance and return instructions  Send the  product  freight prepaid  to the indicated service facility  Repairs will be made and the instrument returned  freight prepaid  Repaired products are warranted for the remainder of the original warranty period or 90 days   whichever is longer     Limitation of Warranty    The above warranties do not apply to products which have been repaired or modified without Newport s  written approval  or products subjected to unusual physical  thermal or electrical stress  improper installation   misuse  abuse  accident or negligence in use  storage  transportation or handling  This warranty also does not  apply to fuses  batteries  or damage from battery leakage     THIS WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES  EXPRESSED OR IMPLIED   INCLUDING ANY IMPLIED WARRANTY OF MERCH
4. ANTABILITY OR FITNESS FORA PARTICULAR  USE  NEWPORT CORPORATION SHALL NOT BE LIABLE FOR ANY INDIRECT  SPECIAL  OR  CONSEQUENTIAL DAMAGES RESULTING FROM THE PURCHASE OR USE OF ITS PRODUCTS     First printing 1999    O 2012 by Newport Corporation  Irvine  CA  All rights reserved  No part of this manual may be reproduced  or copied without the prior written approval of Newport Corporation     This manual has been provided for information only and product specifications are subject to change without  notice  Any change will be reflected in future printings     Newport Corporation  1791 Deere Avenue  Irvine  CA  92606 USA    P N 90043374 Rev  B    iii       Confidentiality  amp  Proprietary Rights    Reservation of Title     The Newport programs and all materials furnished or produced in connection with them     Related  Materials     contain trade secrets of Newport and are for use only in the manner expressly permitted   Newport claims and reserves all rights and benefits afforded under law in the Programs provided by  Newport Corporation     Newport shall retain full ownership of Intellectual Property Rights in and to all development  process   align or assembly technologies developed and other derivative work that may be developed by Newport   Customer shall not challenge  or cause any third party to challenge the rights of Newport     Preservation of Secrecy and Confidentiality and Restrictions to Access     Customer shall protect the Newport Programs and Related Materials as
5. FC OMENS ee 1  Section 1     General Information   1 1  Unpacking iO AA 2  1 2 ara an SETyViCE ee ee ee 2  1 3 Usew  itha Newport Power Meier re ee een 2  I ee ee ee ee 2  1 5 Temperature and AI RAT Ada 2  1 6  Specifications A ee See 3  Section 2     Calibration Accuracy and Limitations   U SB ER ee O RE E O ee ne 5  22 Calibration Accuracy and Service essen eareR RETTET 5  2 3  UMTON O 5  DA DE ae E E E E E ee 5  2 5 Saturation with Pulsed Power Measurements rss 6  20 UI oi 6  2 7 Photodiode Operation nt A A a 6  2 8  L  w Power Measurement Considerations is ii 7  A A A 7  2 10 Temperature and Ambient A aa 7  2 11 Using the Detector for Non CW Measurement             c cccsccesscesseecsseceeceeeeeeseecsaecesecneeeeeseecsaecneeseeeennaes 8  Section 3     Service Form   E ee een 9       Unpacking and  Inspection    fe  Calibration Services    Use with a Newport  Power Meter     aa Cleanliness    Temperature and  Humidity    Section 1  General Information    This guide contains information necessary for using Model 818 Series photo   detectors  A separate data sheet is provided with each detector  All detec   tors are also provided with individual calibration data  Please read through  the guide before attempting to make optical power measurements     CAUTION    Applied voltage exceeding the detector specification  forward  bias  or optical power exceeding the damage threshold can dam   age the detector     The items included with every order are given on the data sheet  Pleas
6. Maximum  V  5 5  NEP  pW VHz  0 018 0 015     Silicon es  Material  UV Enhanced  Silicon  Active Area  cm   1 1  Shape Wand Wand  Attenuator  OD3 Built In Built In             1  When measured with beam centered and filling 80  of active area           B Spectral Response    Calibration Accuracy  and Service    E Uniformity  2 4 Saturation    Section 2  Calibration Accuracy and Limitations    The response of the detector depends on the wavelength of the incident  light  The photodiode is transparent for photon energies less than the band  gap which determines the long wavelength infrared sensitivity limit  The  short wavelength limit is determined by the photodiode manufacturing  process and possibly  in the case of silicon photodiodes  by strong window  absorption  The photodiode response is commonly measured in amps of  photocurrent per watt of incident optical power  Typical response curves  for the photodetector is shown on the data sheet  The transmission of the  attenuator  if provided  may vary considerably across the detector   s range     Statement of Calibration    The accuracy and calibration of this photodetector are traceable to  the NIST or NPL through equipment which is calibrated at planned  intervals by comparison to certified standards maintained at New   port Corporation     Newport Corporation calibrates its detectors using secondary standards  directly traceable to the United States National Institute of Standards and  Technology or National Physical Labo
7. e  check to be sure that all items are present and are received in good condition     The models with a suffix  CM ship with a 9 pin DIN calibration module  The  models with a suffix  DB ship with a DB15 connector that houses the PROM     Newport s calibration services calibrate  or recalibrate  photodetectors   These services provide the calibration data in a written calibration record  and in a programmable read only memory  PROM  for use with a Newport  power meter     The photodetector consists of a photodiode attached to a cable terminated  with a standard BNC connector and packaged within a protective housing   Some housings provide a place for attaching accessories such as attenuators     Care should be taken not to touch the photodiode window with bare fingers   Contaminants may cause inaccurate measurements  particularly at ultravio   let wavelengths where absorption is common  Lightly clean the detector  window with reagent grade alcohol and a soft cotton cloth     Potentially large measurement errors can be generated through scratches   digs and coating damage on detector optical surfaces  The magnitude of an  error typically varies directly with the ratio of the intercepted area of the     scratch    relative to the area of the optical beam     NOTE    Kleenex and chem wipes contain wood and fiber glass  respectively   and will scratch optical surface     The photodiode sensitivity increases with temperature for wavelengths  longer than the peak response wavelengt
8. e to use the photodiode for the display of the temporal  behavior of an optical signal with an oscilloscope     CAUTION    Forward biasing the detector can destroy the diode    Reverse bias should not be used in conjunction with the Newport  Power Meters and in no circumstances should the bias voltage  exceed the breakdown voltage of the photodiode        3 0 Service Form    GO    Newport     Experience   Solutions          Newport Corporation  U S A    Office  800 222 6440  FAX  949 253 1479                               Name Return Authorization     Please obtain RA  prior to return of item   Company  Address Date  Country Phone Number  P O  Number FAX Number  Item s  Being Returned   Model   Serial    Description                      Reason for return of goods  please list any specific problems                                                     Newport Corporation  Worldwide Headquarters    1791 Deere Avenue  Irvine  CA 92606     In U S    800 222 6440  Tel  949 863 3144  Fax  949 253 1680       Internet  www  newport com  E mail  sales newport com    QAD    Newport     Visit Newport Online at  www newport com    Newport Corporation  Irvine  California  has _    been certified compliant with ISO 9001 by the Printed in the USA  British Standards Institution     
9. ed of response and the linearity of the photocurrent genera   tion but introduces additional noise and dark current  Circuit  b  shows the  use of a Op amp to enable unbiased photocurrent measurement  Operation  with zero bias is called the photovoltaic mode because the photodiode is  actually generating the bias voltage  This is the method used in Newport  Power Meters  The feedback resistance in circuit  b  is selectable        E Noise Characteristics    Low Power  Measurement  Considerations    Temperature and  Ambient Lighting    Rreepeack    Vour    Rioap  LIGHT  2 EN    Vour    LIGHT  Vous     a  Reverse Bias Circuit  Photoconductive Mode   improved  b  Op Amp Circuit  time response and linearity Photovoltaic Mode   reduced noise    Figure 5    Photodiode Operational Circuits    Measurements of very low power optical sources are possible with the  photodetector  To use the detector properly and achieve accurate results  requires the understanding of a number of effects that limit the device per   formance  which are discussed in paragraphs 2 9 and 2 10     The lower limits of optical detection are determined by the noise character   istics of the detector and or amplifier  Theory predicts that the photodiode  noise is largely thermal  Johnson  noise associated with the effective resis   tance of the photodiode and shot noise from dark current     The dark current at 10 mV bias voltage is measured and used to define the  effective resistance of the diode  the shunt resis
10. h  Best results will be obtained by  keeping the detector near the calibration temperature of 25  C  The tempera   ture range 0    50  C should not be exceeded and the detector should not   be exposed to humidity levels greater than 70  or possible damage to the  photodiode could occur     NOTE    Refer to Newport web site for the most up to date detector speci   fications        Photodetector                      Specifications  Model 818 UV 818 SL 818 IR 818 IG  Spectral Range  nm  200 to 1100 400 to 1100 780 to 1800 800 to 1650  Power Density  Average Max w   Attenuator  W cm    Oe 2 2    Power Density  Average Maximum 0 2 2 3 3  w o Attenuator  mW cm        Pulse Energy  Maximum   w   Attenuator  J cm   nl i 0   a  Pulse Energy  Maximum   w o 01 1 035 035    Attenuator  nJ cm  2       Calibration Uncertainty   Without Attenuator     4     200 219nm   2     220 349nm   1    350 949nm   4    950   1100nm    1     400 940nm   4    941   1100nm    2     780 910nm   2    911   1700nm  4     1701 1800nm    2     800 900nm   2    901   1650nm       Calibration Uncertainty    8     200 219nm   2    220 349nm     1   Y 400 940nm     5     780 910nm   2    911     5     800 900nm                                                     1  Applies to entire spectral response  2  15 ns pulse width    3  Uniformity specification applies to detector only          4  Calibration uncertainty can be varied depending on the NIST transfer standard uncertainty variation           0 Ej 0   0     W
11. ina  Tel   86 510 8113 299  Fax   86 510 8526 9050    Newport Corporation Calling Procedure   If there are any defects in material or workmanship or a failure to meet specifications    promptly notify Newport   s Returns Department by calling 1 800 222 6440 or by visiting our website at  www newport com returns within the warranty period to obtain a Return Material Authorization Number   RMA    Return the product to Newport Corporation  freight prepaid  clearly marked with the RMA  and  we will either repair or replace it at our discretion  Newport is not responsible for damage occurring in  transit and is not obligated to accept products returned without   an RMA      E mail  rma service newport com  When calling Newport Corporation  please provide the customer care representative with the following  information   e Your Contact Information  e Serial number or original order number  e Description of problem  i e   hardware or software   To help our Technical Support Representatives diagnose your problem  please note the following conditions   e Is the system used for manufacturing or research and development   e What was the state of the system right before the problem   e Have you seen this problem before  If so  how often   e Can the system continue to operate with this problem   Or is the system non operational   e Can you identify anything that was different before this problem occurred     vi       Table of Contents    LO 111  Technical Support Contacts ee ee v  Table O
12. ith Attenuator    ee 1700nm  4408   7    901  4     950  1100nm 1650nm  1701 1800nm  1100nm  Uniformity        2  2  2  2  Linearity      0 5  0 5  0 5  0 5  Rise Time  us   lt 5 9  lt 2 2  lt 2  Shunt Resistance  MQ   typ   gt 10  gt 10  gt 35  kQ   gt 20  Reverse Bias  Maximum  V  5 3 0 25 2  NEP  pW VHz  0 45 0 2 0 6 0 04    Silicon UV     Indium Gallium  Material Enhanced Silicon Germanium Aisenide  Active Area  cm   1 1 0 071 0 071  Active Diameter  cm  1 13 1 13 0 3 0 3  Shape Cylinder Cylinder Cylinder Cylinder  Attenuator  OD3 Detachable Detachable Detachable Detachable  Bee Tea 5  C to 50  C  5  C to 50  C   lt 70    5  C to 50  C  5  C to 50  C   AS  lt 70  RH RH  lt 70  RH  lt 70  RH                         Model 818 ST 2  UV 818 ST 2   Spectral Range  nm  200 to 1100 400 to 1100  Power Density  Average Max 02 2   w  Attenuator  W cm       Power Density  Average   Maximum w o Attenuator 0 2 2    mW cm     Pulse Energy  Maximum   01 1   w  Attenuator  uJ cm   i   Pulse Energy  Maximum   01 1    w o Attenuator  nJ cm         Calibration Uncertainty    4    200 219 nm  2    220 349 nm  1    350 949 nm  4    950 1100 nm    1   Y 400 940 nm  4     941 1100 nm       Calibration Uncertainty    8    200 219 nm  2   Y 220 349 nm    1   Y 400 940 nm                               w Attenuator 1   Y 350 949 nm 4     941 1100 nm  4    950 1100 nm   Uniformity       2  2  Linearity      0 5  0 5  Rise Time  us   lt 3 3  Shunt Resistance  MQ   typ   gt 200  gt 200  Reverse Bias  
13. k current generally increase exponentially with  detector temperature so it is best to keep the temperature close to 25  C     Some 818 Series photodetectors  e g  Models 818 SL   UV   IR   IG  have  threads for the attachment of accessories  With these models  the effects of    7    Using the Detector  for Non CW  measurements    ambient lighting can be minimized by using the Model 818 FA Fiber Adapter  Holder with the appropriate FP3 or FP4 series fiber adapter when measure   ments of power in fiber optics are being made  If free space beam measure   ments are desired  using an attenuator will reduce stray light and often  improve the ratio of signal to background  Wavelength specific filters  such  as optical cutoff  bandpass  or spike filters can also be used if the signal  wavelength spectrum permits  Other techniques to reduce stray light in   clude using apertures to admit only the laser beam  placing the detector in a  box to shield the surface and turning off the room and other lights     When the photodetector is used with a Newport Power Meter it is operated  essentially without bias voltage  as depicted in Figure 3  b   The effective  time constant of the detector amplifier combination may be much slower  than the characteristic time of the signal  Nonetheless  if the detector am   plifier combination does not become saturated  effective integration of the  signal will occur  and accurate energy measurements of very short pulses  can be made     It is also possibl
14. ratory  At wavelengths where ap   propriate standards are not available  Newport calibrates its detectors by  comparison to detectors calibrated with thermal detectors  The absolute  accuracy of the photodetector calibration is indicated in the calibration  certificate     Individual detector response can change with time at different wavelengths   especially in the ultraviolet  and should be returned for recalibration at 1  year intervals to assure confidence in the accuracy of the measurement     For recalibration services  see the Accessories and Services information     Fabrication processes may cause the response of the detector to vary  slightly over the detector surface  Calibration involves illumination of ap   proximately 70  of the detector central active diameter  Optical signals  being measured should illuminate the same area  Care should be taken not  to overfill the detector if accuracy is to be maintained     For low optical power  the photocurrent is proportional to the optical signal  incident on the photodiode and the photocurrent linearly increases with  optical power  For high optical powers saturation of the detector begins   to occur and the response signal is no longer linearly proportional to the  incident power  Optical power measurements must be made in the linear  region to be valid     NOTE    The saturation is    soft     i e  the detector output does not suddenly  stop increasing  but the rate of increase slows  For Gaussian and  other signals wi
15. tance  given in the data  sheet    or   V biss  10 mV    liane    Ideally an input amplifier connected as in Figure 3  b  would have no offset  voltage and there would be no dark current  In practice though  a small bias  usually exists  and even at zero applied bias a small dark current is present   The maximum dark currents when used with Newport power meters are  given in the specifications  The drift in this dark current is generally much  smaller than the dark current so effective cancellation is often possible     For non DC measurements the light detection limit is more generally ex   pressed as the intensity of light required to produce a current equal to the  noise current  This is called the noise equivalent power  NEP  and has units  of optical power divided by the square root of the detector bandwidth     W  VHz     The NEP varies inversely with the spectral response of the photodiode and  depends on the wavelength  A  the noise frequency  f  and bandwidth  Af   NEP is therefore defined as NEP  A  f  Af       Lock In amplifier techniques can be used to approach the NEP     Drifts in temperature and ambient lighting can cause significant variations  in diode response and amplifier performance for low power measurements   particularly for wavelengths longer than the bandgap  It is therefore impor   tant to provide as constant environmental conditions as practical  including  consideration of air conditioning and roomlighting effects     In addition the noise and dar
16. th spatially varying intensities  local saturation  may occur  The onset of saturation is not always obvious and is a  common source of inaccurate measurements        Saturation with  Pulsed Power  Measurements    E Reflections    Photodiode  2 7 Operation    To determine if the detector is saturating  follow the steps below     1  Measure the photodetector current  or power   and record this  value  A      2  Place a filter or attenuator of known transmission  T  in the beam  path  Record the current again  B   A filter transmission of 0 001 is a  convenient choice     3  The power with the filter in place should be the product of the power  measured without the filter and the transmission of the filter   ie  B AxT    The transmission  T  of the filter can be determined by following the steps  below     1  Reduce the optical power to a level low enough to avoid saturation   but high enough that  when it is reduced by the filter it can still be ac   curately measured     2  Follow steps 1 and 2 in the procedure above     3  Calculate the ratio T   B A to determine the transmission of the filter  at the wavelength of light used for the measurement     The calibrated filter  or attenuator  can be used with the detector to mea   sure the power of higher power beams     Saturation effects when using pulsed lasers are a complex phenomenon   and depends upon the wavelength  peak power  pulse shape  average power   repetition rate  and on the detection circuit  However  the test for
    
Download Pdf Manuals
 
 
    
Related Search
    
Related Contents
Fiche produit - Consommables  1 Law Catalogue User Guide Index A. Introduction to the  GL750 - Comtec  2SK2719  Smeg KSEG90X2 cooker hood  iTagPlot User Manual  AEG L 72850 Washing machine User Guide Manual  FactoryCast - Guida utente - per Quantum e  Nissin MG8000 User's Manual  Trasduttore incrementale con albero cavo continuo RI 76TD    Copyright © All rights reserved. 
   Failed to retrieve file