Home
        ScanArray Express Microarray Analysis System User Manual
         Contents
1.        Palette   Source  Black threshold   ee o  SPES     Ful color threshokt       33 Median   Auto Adjust                      Mouse is at 16823m  21602um Pixel intensity ratio  1 91       2 3 1 Improving the Image Display    You can use the controls in the lower right corner to adjust the image display and improve  the visibility  The adjustments do not affect your data  only the image on the screen        Note  ScanArray Express auto adjusts images during scanning or opening  The image of       the Geometric Test Slide that you have just scanned using parameters in the previous  section  does not need adjustments  The image is already of good quality  However  if    you want to see at this point how the adjustments work  you can find more information in     Adjusting the Display     in Chapter 3        2 3 2 Changing the Palette for the Selected Image    You can select a different palette or change the contrast  The palette used in the composite  tab cannot be changed  it is always a blending of the palettes of the control and source   experiment  images or the rainbow palette  If you open a palette when the Composite tab  is selected  the Palette window that opens allows you only to adjust the contrast     User Manual pbtsupport perkinelmer com 2 5    Chapter 2    2 4    Getting Started  Scan and Quantitate the    Click Palette  The Set Palette Options window opens     Set Palette Options    Name Palette  Red  Orange  Yellow    Blue  Purple  Gray  Rainbow    Use the pale
2.       5  Tools Scan resolution   Sum  10 um  20 um  C 30 um  50 um    Help and Directions  Focus position   should generally be 0 pm   Scan resolution   should be about 1 10th of the spot diameter     Scanspeed  should be  Full   unless you specify a resolution of 5 um  or unless you  particularly desire low signal to noise ratios at high PMT gains     Finish Cancel       Figure 4   4 Scan Protocol   Basic Information Window       Click Next to move to the next window and Back to move to the previous window      Click Finish to save all of your changes and close the wizard      Click Cancel to close the wizard without saving any of your changes     4 4 1 Basic Information    In the Scan Protocol   Basic Information window  provide a name and description for the  protocol  and choose the scanning resolution and speed     4 6 Technical Support 800 551 2121 or 617 350 9263 ScanArray        Express    Scanning with Protocols    Chapter 4    To name a scan protocol    1  Fill in the information as listed in the following table     Item    Description       Name    Description    Focus Position    Scan Resolution    Scan Speed    Enter a name for the scan protocol so you can recall it easily  You must  enter a name in order to proceed     Enter a description for the protocol     You do not need to adjust this for each scan  Leave the default value at  0 for conventional flat slide substrates     To determine the focus position for unconventional slides  use the Scan  Protocol Tools
3.       M Use Cy3    Scan Area Co ordinates  Start position  X  mm     1 04 Area width  mm     7 52    Start position  Y  mm     1292 Area height  mm     8 23    Set Scan Area to Full Microarray    Show Zoom Window       J Automatically save image files locally  To change the scan area  drag J Automatically save images in Arraylnformatics  the rectangle edges with the    mouse  or use the the box    labelled  Area Co ordinates   at Cancel  right  a       Figure 3 1 Scan Window   Run Easy Scan    3  Use the settings from the last session if you are using the same type of microarray  or  see Section 3 2 1 to change the settings     4  Click Start     You may or may not see a message asking if you want to skip warming up the lasers   Skipping the warmup does not cause any damage  but the sensitivity of the resulting scans  will vary slightly  If you are concerned with precise repeatability or uniformity of  scanning  wait for the lasers to warm up     3 2 Technical Support 800 551 2121 or 617 350 9263 ScanArray    Express    User Manual    Using Easy Scan and Easy Quant    Changing the Settings    Chapter 3    To change the Easy Scan settings  refer to the table below for information  To make it  easier to select a scan area using the mouse  you may want to obtain an image first by  scanning the full slide at 50 microns     Item    Description       Scan Type    Scan resolution    Autoloader slot from  which to load    Fluorophores    Use    PMT Gain        Scan Area    Area Co or
4.       access to your instrument      your phone number  fax number  and e mail address    For Additional Information    Additional information about the system can be found in the following        ScanArray Express Installation Guide   e Documentation provided with the workstation and its operating software      Documentation provided by manufacturer of the fluorophores that are used   e The ArrayInformatics Microarray Laboratory Integration Guide  for  ArrayInformatics users only     Preface Preface    Getting Help for Windows 2000 XP    The Windows 2000 XP online help provides information on learning how to use the  Windows operating system     To open Windows 2000 XP help  1  Click the Start button on the task bar     2  On the Start menu  click Help  Navigate to a topic using the tabs     User Manual PBTsupport perkinelmer com Pagex    Preface Preface    Introduction Ciapies 1    Chapter Summary   Overview 1 1   The ScanArray Express Main Window 1 4  Workflow for Scanning and Quantitation 1 5  The Different Ways to Scan 1 6   The Different Ways to Quantitate 1 7  Preparing to Scan 1 8    1 1 Overview  Welcome to the ScanArray   Express from PerkinElmer Life Sciences  The ScanArray    Express is an extremely sensitive microarray laser scanner with easy to use software for  microarray scanning and quantitation     1 1 1 The ScanArray Express System  The ScanArray Express system includes the scanning instrument with one to four lasers     and a Windows 2000 or Windows XP wor
5.      Select a resolution that is appropriate for your microarrays     5 um for spots that are less than 90 um diameter  10 um for spots that are 100 to 200 um diameter  20 ums for spots that are greater than 200 um diameter    Higher settings result in a quicker image with reduced detail  The default  is 10 microns     Select a scan speed  usually Full  the default setting      You may want to use Half speed for better signal to noise ratio on dim  arrays that require PMT settings greater than 75        2  Click Next to display the Scan Protocol   Scan Area window     4 4 2 Selecting the Scan Area    In the Scan Protocol   Scan Area window  select the area of the slide to be scanned  To set  the scan area to include only that part of the slide with the image  obtain an image to help  you see the area  There are two ways you can get an image     e The first is to run an Easy Scan on your slide before you start to configure the scan    protocol        The second  if you   ve already started configuring the protocol  is to run a Quick Scan  from the Tools section of this protocol   See Running the Quick Scan on page 4 15     To select the Scan Area    1  Open the Scan Protocol   Scan Area window  shown in Figure 4 5     User Manual    pbtsupport perkinelmer com    4 7    Chapter 4 Scanning with Protocols    Scan Protocol   Scan Area    J asic Information Help  amp  Directions      The area to be scanned is shown as a darted  an Area rectangle      J 3  Fluorophores To Start a Ne
6.     Change  after selecting a protocol  to modify an existing protocol       Duplicate  makes a duplicate of the highlighted protocol to help you create a  new protocol from one that is already close to what you want     Or  click     e Delete  after selecting a protocol  to delete a protocol  you cannot delete a  protocol that is part of a batch set or protocol group    e View Usage  after selecting a protocol  to display a list of batch sets and  protocol groups which include the selected protocol     4 4 The Scan Protocol Wizard    The Scan Protocol wizard opens with the Basic Information window displayed  as shown  in Figure 4 4  From this window  using the wizard  you can display and set all of the    scanning parameters     User Manual pbtsupport perkinelmer com    4 5    Chapter 4 Scanning with Protocols    The only required steps in the wizard are 1  Basic Information and 3  Fluorophores  At  least one fluorophore must be added to a new protocol for it to be created and saved        Tip  You may want to run an Easy Scan to acquire an image before configuring your    A scan protocol  to help in setting up the protocol        You can move through the windows by clicking the numbers on the left  or by using the  buttons on the bottom of the window     Scan Protocol   Basic Information  Click the number to move f  from one window display to  another   gt    2 ScanArea    1  Basic Information Name     i    Description          3  Fluorophores      Focus position  pm      o
7.     Fulwidh   Zoom to Line       CR  cancel       Figure 4 17 Focus Line Definition Window    3  Click OK to close and return to the Automatic Focus Calibration window     4  Click Start  A progress bar displays while the focus is being calibrated  You can  cancel at any time by clicking Stop     5  Upon completion of the calibration  review the results  and click Accept Changes to  accept the results        Note  If you have used the Scan Protocol Tools  and accepted changes made by the tools   be sure to save your Scan Protocol to save the changes        User Manual pbtsupport perkinelmer com 4 23    Chapter 4 Scanning with Protocols    4 6 Creating an Image Autosave Protocol    Image Autosave Protocols specify the folder  file naming convention and file type to use  when automatically saving image files during a scan     To create an Image Autosave Protocol    1  On the Main Window  click Configure in the Configure  amp  File group  In the  Configure menu that displays  click Image Autosave Protocols  The List of Image  Autosave Protocols window opens     List of Image Autosave Protocols    Description Pattern  Default  lt Fluor  lt Date gt  lt Time gt  C  Program Files PerkinElr    Change    Duplicate    Delete  View Usage    Sort By  Name    Description        Path       mage autosave protocols beginning with   cannot be executed with current hardware        Figure 4   18 List of Image Autosave Protocols Window    2  Sort the list  optional  by Name  Description  or Path
8.     n   34s siete es Soin Soe ee dw bet ee Rene a ORS 7 12  Backing up the Database 24 27o ee ah ene ee ee kaye wate e eae 7 13  Restoring the Databases  43  4ncdcuc peodined ch ei deud paweede wp bads 7 13  S  rvice Featuics     Vs cachid wies Gea dees weed ade aaa Ai aw ene Bae 7 14  Descriptions essea ra he pha iy Dewees ee ak ed aaa a ee 7 14  Setting the Instrument Network Options               00 0002 eee eee 7 16    Chapter 8 Diagnostics and Troubleshooting    OVETVIEW aa  sone ia o E See  gre wndeee hed Ge nied Bag wines peg dali deg  gaia 8 1  Diagnostic  OOS eosar tee cee hued ea RG ie yea we ane Rao he a 8 1  Start up DigenOsics  6 004 aye ote er ek 2G te el Ma rad aan iG 8 1  Instrument Statistics Accumulation              0    ce eee eee eee eee 8 3  Troubleshooting 2 id 0c trv Porat a ulate cate win E eu le Aare ey ta se WD Ore aes oa ae 8 3  Hard Wares aia  oo aeg tk oa yas TEE GIA Cees SSy eee wtesa 8 3  Ready LED Flashes Yellow e336 e853  d VEGA PARR SYS RRE SSPE RRS OS 8 3  External Laser Will Not Turn On    0 000 tenes 8 3  Microarray is Jammed      ccs seta ais ece  ua nl den nel a ihe dee Dey doe Satins ca 8 3  Poor Image Uniformity  oii Pork Oh oeren ea ee ook Gi SOE ON a Ree 8 4  BOL Wale roc  sealer eRe eee eats Matha a BAGS Hae S SE Le eee s TAATA 8 4  Unable to Open Stored  TIFF flesh S23 woke 4 cagdon ee tala nde eae 8 4  Recording the ScanArray Express System Activity          0 0 0    eee eee ee 8 4  Viewing the System Log 44044 seceded ada eoa a 
9.     on page 1 6    e Easy Scan for ways to scan and save your images     e     Scan Protocol See Chapter 3 to get started right away with  e Protocol Group Easy Scan   e Batch Set     gt   See Chapter 4 to create a Scan Protocol or  Save your images manually or Chapter 6 for a Protocol Group or Batch Set   automatically   Quantitate  See    Different Ways to Quantitate    on page  Click Quantitate and select from  1 7 for ways to quantitate and save your  data   e Easy Quant  e Quantitation Protocol  gt  See Chapter 3 to get started right away with  Easy Quant   See Chapter 5 to create a Quantitation  Protocol   Quantitate Automatically  Save your images or quantitate        See Chapter 4 to run a Scan Protocol with a  automatically after scanning  and then save checkbox for automatic quantitation     the images and quantitation results     NOTE  To practice scanning with the Geometric Test Slide included with your  system  see Chapter 2     Getting Started  Scan and Quantitate the Geometric  Test Slide       Figure 1 3 A Workflow Roadmap    User Manual pbtsupport perkinelmer com 1 5    Chapter 1 Introduction    1 4 The Different Ways to Scan    Before starting  review the ways you can scan your microarrays  You can automatically  save the images to a local network file system or ArrayInformatics database     1 4 1 Easy Scan    The Easy Scan settings are    persistent     that is  they remain from the last scanning session   so it   s easier to scan the same type of microarra
10.    1  Click the ArrayInformatics tab     Application Settings    Connection   Scanning Quantitation    Arraylnformatics Settings  Database   Lspg dey    Image path          SLspa dev lmageStore Browse       Image path to use when Arraylnformatics is unavailable        IE     ImageT empStore Browse          2  Refer to the following table and specify your settings     Item Description       Database This field displays the database that is currently active for this  ScanArray Express system  The database cannot be changed from this  window  but must be changed using Array Informatics Database Tools     Image Path Browse the network to locate and select the directory that has been set  up for saving the scan images  Images that are saved to  Arraylnformatics are saved to this location     Image Path to use This is a local directory where scanned images are saved if the  when Arraylnformatics is temporarily unavailable  or some other error  Arraylnformatics is prevents saving in ArrayInformatics  Images that are saved to this  unavailable  directory can later be sent to the Arraylnformatics database        pbtsupport perkinelmer com 7 5    Chapter 7 System Settings    7 2 5 Other Settings  Settings on the Other tab include those that affect what is seen on the user interface   To set or change these settings    1  Click the Other tab   Application Settings    Connection Scanning Quantitation    Arr  Informatics      Image Tab Settings  Action performed by clicking mouse on image
11.    3  Click one of the following in the List of Image Autosave Protocols window     e Add to create a new protocol    e Change  after selecting a protocol  to modify an existing protocol       Duplicate  makes a duplicate of the highlighted protocol to help you create a  new protocol from one that is already close to what you want       Delete  after selecting a protocol  to delete it    e View Usage  after selecting a protocol  to display a list of protocol groups and  batch sets which include the selected protocol     4 24 Technical Support 800 551 2121 or 617 350 9263 ScanArray        Express    Scanning with Protocols Chapter 4    4  Click Add  The  mage Autosave Protocol dialog box displays  In this window  you  can check each item that you want to include in the filename  To create a unique  filename  you must include the fluorophore  the date  and the time  You can also  check Custom Text and add a version number  e g   v1  v2  etc      Image Autosave Protocol    Description     Pattern for File Names  Information to include in the file names   IV Fluorophore J    Time  hours  minutes   J    Barcode J Custom text    IV Date  month day year     File name pattern          lt Fluor gt  lt Date gt        File types   Tl    BMP  bitmap   J    JPEG Joint Photographic Experts Group   J    RAW  proprietary binary format   IV TIFF  Tagged Image File     Path to output files   C  Program Files PerkinElmer Scan4rray Express lm Browse       Car       Figure 4 19 Image Autosave Pro
12.    7 3 Configuring Fluorophores    Fluorophores specify the laser wavelength and emission filter used to scan a microarray   The ScanArray Express ships with a number of pre defined system fluorophores  These  cannot be deleted and only the display palette field assigned to the fluorophore may be  changed  User defined fluorophores may be added  duplicated  changed and deleted as  specified below  A fluorophore currently specified in a protocol may not be deleted     7 3 1 Defining a Fluorophore  To add a new fluorophore to ScanArray Express     1  On the Main Window  click Configure in the Configure  amp  File group  then click  Fluorophores   The List of Fluorophores window displays     List of Fluorophores       Pe Excitation   Emission   Composite Image  Description Peak  nm    Peak  nm  Palette Palette Standard    Alexa 438 Alexa 488   Alexa 532 Alexa 532   Alexa 546 Alexa 546   Alexa 555 Alexa 555   Alexa 568 Alexa 568   Alexa 594 Alexa 594   Alexa 647 Alexa 647   Alexa 660 Alexa 660  Allophycocyanin  APC  Allophycocyanin  APC    BODIPY 530 550 BODIPY 530 550  BODIPY 558 563 BODIPY 558 563  BODIPY 564 570 BODIPY 564 570  BODIPY 630 650 BODIPY 630 650  BODIPY TMR BODIPY TMR  Calcein Calcein   Calcium Crimson Calcium Crimson  Calcium Green 1 Calcium Green 1  Calcium Orange Calcium Orange  cy 3s cy 3s   cys cy ss   Cy2 Cy2   Cy3 Cy3    Delete    View Usage    Sort By    Name    Description    Excitation  Peak  Emission  Peak    q lt  lt  lt  lt    lt  lt  lt    lt  lt     
13.    Centers point  Centers point and zooms in  Centers point and zooms out       Program Layout      nstrument controller layout     Quantitation only layout       Item Visibility    J    Hide items  protocols  protocol groups  batch sets   and fluorophores  that are incompatible with current  hardware    IV Hide diagnostic log messages          2  Refer to the following table and specify your settings     Item Description       Image Tab Settings Select a method to set the behavior of the mouse click on all image  tabs     Action performed by Centers point  mouse click on image Centers point and zooms in  Centers point and zooms out    Program Layout Select whether the ScanArray Express system is to show both scanning  and quantitation controls  or only quantitation controls       Instrument Controller layout    Quantitation Only layout   the user see only Quantitation buttons and  windows  The scan button does not display     Item Visibility Check one or both boxes     Hide items  protocols  protocol groups  batch sets and fluorophores   that are incompatible with current hardware     If this box is not checked  the incompatible items display  marked with    an asterisk   Hide diagnostic log Keeps the diagnostic instrument messages from displaying in the Log  messages tab on the Main Window        7 6 Technical Support 800 551 2121 or 617 350 9263 ScanArray     Express    System Settings Chapter 7    3  Click OK to save changes and close the Application Settings dialog box  
14.   AN    Note  For instruments with HT capability  specify the slot number of the microarray to  be scanned  before selecting a tool to run  The default slot value is 1        2  After using the Tools to refine settings  click Finish to save the protocol     4 5 Using Scan Protocol Tools    To use the tools  run a Quick Scan first  only experienced users should run the other tools     4 5 1 Running the Quick Scan    User Manual    The Quick Scan allows you to check the setup of the protocol  Any changes made during  a Quick Scan do not persist unless you save them to the protocol  The Quick Scan  performs a scan of a selected area using selected fluorophores  allowing you to adjust the  laser power and PMT gain while the scan is being performed     pbtsupport perkinelmer com 4 15    Chapter 4    Scanning with Protocols    To perform a Quick Scan    1  Click Run in the Quick Scan area of the Scan Protocol   Tools window  The Quick  Scan dialog box displays     Quick Scan    Area to Scan     Scan resolution  jm    C5 10 C 20    30    50    Focus position  um   l    PMT Laser  Fluorophores Gain     Power        Cy3    Cy5    Select a Fluorophore l  Select a Fluorophore    Select a Fluorophore l       Help  amp  Directions  The area to scan is shown as a darted rectangle     For instructions on how to select an area  click    Show Zoom  Window         Show Zoom Window   Cancel            Figure 4   12 Quick Scan Window    2  Change settings only if needed  Refer to the following t
15.   Chapter 7    System Settings    Pre printed barcode labels are commercially available  or you can print your own that  conform to the guidelines in Table 7 1  PerkinElmer has qualified the following blank  labels     Manufacturer Part Number Size  inches  Size  mm        Substrate Intermec E17512 0 920    x 0 550    23 4 x 14mm       The label containing the barcode should fit on the first 15 mm of the slide so that it does  not tilt the slide in the slide holder  and the label must be centered on the slide     PerkinElmer has qualified a printer and labels that will withstand hybridization  washing  and other processes  The recommended printer is an Intermec 3240 with a 137071206  Super Premium ribbon  The recommended label  also from Intermec is L3203011 2 mil  glass poly  with dimensions of 0 920  x 0 550  and a corner radius of 0 04 to 0 06      Information about the printer and the labels can be obtained from Intermec        Internet Address  http   www intermec com     e mail  info intermec com    You must define the barcode parameters  which can include the Image Autosave Protocol  and Scan Protocol  as described in the next section     Technical Support 800 551 2121 or 617 350 9263 ScanArray    Express    System Settings Chapter 7    7 4 2 Defining Barcode Parameters  To define barcode parameters  1  On the Main Window  click Configure in the Configure  amp  File group and then click  Barcodes to display the Barcodes window     Barcodes    IV Barcodes contain image 
16.   Each row in the spreadsheet is the data for    one spot  If you use a     GAL file for your template  the gene name and ID are imported for                         each spot   Run Composite   Cy5  Control    Cy3  Source    Spreadsheet   Scatter Plot   Distribution Piot   Log   Diagnostics   Help    Scan   General CyB  Control   Raw Spot   CyB  Cont  it Subanay   Spot Spot Spot   Spot     gt    gt   Quantitate    ity   Lacanen fied mn   cee DNa    eons iscaten re Meden   Mem Medan   en eed ean ee   StDev   Bkgnd     Bkond           fecliow    coiro fx umy pm  z   pot   Spot    Bkgnd    Bkgnd z   7   18   28  PRESAS A   1 11 1 1 G 0346 39706 640 5 100 0 00   Configure  amp  File   f2 11 2 1 AKTI Hs 71816 1453  13837 Good 48027 4026 47365 39464 662 1292 14458 56 480853 909 SDS  me fi    Ww 3 1 AKT2 He 200816 1659  13837 Good 42628 40464 42016 39852 612 608 1445261 12974 1000 1000  E 4 XI 41 AREG Hs 1257 1859  13837 Good 4780 40510 46573 39903 607 615 1496348 16831 1000 100   Configure 5  Ww 5 1 ASPH Hs 121576 2069  13837 Good 46629 40045 46083 39499 546 583 1436290 15551 1000 100    aome 5 4 6 1 BAGI Hs41714 2269  13837 Good 48168 40223 47598 393  592 608 1441559 15582 1000 100C  7 li 7 1 BCGFI  Hs 99879 2469  13837 Good 47io7 39410 46528 aeeai 579 65 1388111 14225 1000 1006 f         8 11 8 1 BCL2 Hs 79241 2679  13837 Good 48668 45410 48071 44813 597 593 10213 38 148 09 100 0 100 0    Spot Viewer 3 WwW 3 1 BDNF Hs 56023 2879  13837 Good 50843 47266 50227 46650 616 626 1044249
17.   Spot Pile Encel Spreadsheet None                                  Figure 3 6 Quantitation Results Tab    You can scroll down t    o see all of the listed spots or scroll across to see all columns of data     click View as Excel Spreadsheet to view and save as a Microsoft Excel spreadsheet     Several spot intensity    values and ratios between multiple fluorophores are listed in the    spreadsheet  Spot intensity values are automatically normalized to correct the intensity of  each spot for variations in the overall intensity of the image  with respect to a control  image  Values include     Value    Description       Median    Mean    User Manual    pbtsupport perkinelmer com    This column normalizes the ratio of the median value of all spots in each  fluorophore to the ratio of the median value of the spots in the control  fluorophore     This column normalizes the ratio of the mean value of all spots in each  fluorophore to the ratio of the mean value of the spots in the control  fluorophore     3 15    Chapter 3 Using Easy Scan and Easy Quant       Value Description   Total Each spot   s data is normalized to the data of all spots    Background Reports the intensity of each spot in the array minus the background for  Subtracted that spot        3 4 1 1 Filtering the Spreadsheet Data    You can use the buttons at the bottom of the Spreadsheet tab to filter the data  as described  in this section     Show Hide Show Hide   Set Quality Set View  Spots by Status Columns Cri
18.   is the physical object  an  array  is an ordered collection of spots on  the microarray     A network interface card  NIC  is a computer workstation circuit board or card that is  installed in a computer so that it can be connected to a network  Workstations on local  area networks  LANs  typically contain a network interface card specifically designed for  the LAN transmission technology  such as Ethernet     A range of colors assigned to image pixels based on the intensity value for the pixel     The loss of fluorescent signal due to intense laser power  This can be minimized by  lowering laser power and by subjecting the sample to the laser power for less time  The  lower the resolution  the less time the microarray is subjected to the lasers  At a 10  micron scan  with good quality dyes  and 95  laser power  typical signal loss is less  than 2 to 3  per scan     Images are made up of a 2 dimensional array of numbers called picture elements  or  pixels  The resolution of an image is determined by the size of pixels  For example  5um  pixels is higher resolution than 10um     Periodic Maintenance and Inspection  Planned maintenance  calibration  and inspection  to correct for expected wear in the system     Photo Multiplier Tube  A light sensitive vacuum tube that controls current based on the  amount of light being received by the tube  A PMT has high gain allowing it to have a  high currentto Light ratio     The PMT absorbs low level fluorescence light and converts it t
19.   see Appendix A     The template field fills with the name of the selected  GAL file    See To define a template using a  GAL file on page 3 10    Click this button to manually define the array pattern  The Template  Specification window opens  where you enter the values    See To define a template from specifications on page 3 11   Click this button to import template information from the  Arraylnformatics database  if your ScanArray Express system is    integrated with ArrayInformatics  Refer to the Microarray Laboratory  Integration Manual that ships with your ArrayInformatics software     If your system is not integrated with Arraylnformatics  this button  appears dimmed and is unavailable to select     Select Adaptive Circle  This is the default method     For more information on quantitation methods  see    Choosing a  Quantitation Method    in Chapter 5     Select LOWESS  This is the default method     For more information on normalization methods  see Chapter 5     Click this button to open the Adjust Template window to adjust the  template  and register the images if there are two images     When you start the quantitation  ScanArray Express finds the spots  using the template  and then starts quantitation  For fastest spot finding   adjust the template within a row or column or two of the microarray  spots     See Adjusting the Template and Registering Images on page 3 12        User Manual    pbtsupport perkinelmer com 3 9    Chapter 3 Using Easy Scan and Easy Qua
20.   which is adequate for pre scanning and investigative  scanning  After five minutes  they become stable enough for all but the most sensitive  applications  ScanArray Express indicates the lasers are still warming up for the full 15  minutes        1 6 1 Turning on the ScanArray Express and Warming the Lasers    To turn on the ScanArray Express    1  Turn on the ScanArray instrument  Verify that the Power and Ready indicator lights       are green   os    Check that the two     Hold the microarray   indicators are green  face up  by the    thumb space or by      i k       the edges     2  Log in to Windows if required  using your assigned user name and password     3  On the Windows workstation  double click the ScanArray Express icon     Scandrray  Express       The User Interface opens to the ScanArray Main Window     4  Check the status of the lasers  The laser buttons on the right side of the Main Window  indicate the status of the lasers  blue indicates the lasers are turned off  yellow  indicates the lasers are warming up  green indicates the laser are ready  To turn on a  laser  click the status button for that laser     1 8 Technical Support 800 551 2121 or 617 350 9263 ScanArray TMEX press    Introduction    5     Chapter 1    Begin laser warm up by clicking the laser status button     1 6 1 1 Loading the Microarray or Cassette    To load a microarray    1     2     Hold the microarray by the thumbspace or by the edges     Holding the microarray face up  and with th
21.  18850 1000 100C  iad 10 1 101 BMP2 Hs 73853 3069  13837 Good 49223 42223 48646 41646 577 598 1553903 15170 1000 100    1 1 1 1 BMPS He 1104 3278  13837 Good 50954 491293 50353 48528 60m 66 600725 16011 1000 1006 f  12 11 121 BMPS Hs 6101 3473  13837 Good 50834 49828 50238 49230 598 616 7608 20 16448 1000 100  13 11 13 1 BMP8 He 99948 3679  13837 Good 51092 49703 50478 49089 614 622 7071 20 17348 1000  100     4 11141 BMPRIB  Hs87223 387913837 Good 50767 49340 5054 48727 613 619 680031 15367 1000 100C  15 11 181 BTC Hs 73105 4079 13837 Good 50398 49164 49811 48579 585 598 579756 15403 1000 100C  16 1 1 16 1 CALML3  Hs239800 4279  13837 Good 49887 48117 49274 47504 613 639 447159 18170 1000 100C  17 X1 171 CEBPA He 76171 4473  13837 Good 48533 47574 47933 46974 600 592 411147 15838 1000 100C  18 11 18 1 CEBPB  Hs39029 4679  13837 Good 47516 47333 46949 46766 567 600 328119 17344 1000 100C  19 1 1 191 COLISAT  Hs 78409 4879 13837 Good 47939 47614 47369 47044 570 609 322812 17077 1000 100C  20 11 201 CIGF Hs 75511 5079 13837 Good 47772 469682 47158 46348 614 2649 298 33 944410 100 106 f  a X1 211 DBH He 2301 5279  13837 Good 47801 48276 47177 47652 624 639 302069 198587 1000 100  22 11 221 DBP Hs 155402 5479  13837 Good 49260 48902 48671 48313 589 629 213724 18912 1000 100C  23 11 23 1 DTR Hs 799 5679  13837 Good 50495 50429 49894 49828 601 613 2818 35 216 26 100 0 100 0  24 V1 241 E21 Hs 96055 5879  13837 Good 50547 49600 49947 49000 600 621 351579 25S  1000 106 f  25 125 1 EF2
22.  617 350 9263    A software application program  running on a computer workstation  that provides the  GUI and allows the user tocalibrate  configure  control  and monitor the instrument  The  client software communicates with the instrument software to send configuration and  control commands to and receive status and configuration information about the  instrument     A configuration file is a text file that specifies certain system parameters     Cable used for the direct connection of two workstation Network Interface Cards  NICs    NICs have dedicated send and receive lines  A crossover cable connects from the send  line of one NIC to the receive line of the other NIC     Daisy chain is a serial connection of several devices     Ethernet is the most widely installed Local Area Network  LAN   A LAN is a network of  interconnected workstations sharing the resources of a single processor or server within  a relatively small geographic area  i e  an office building   Ethernet is a set of hardware  and signaling standards for used for LANs  The most commonly installed systems are  10 100 BaseT     A facility network is the network at the facility or institution where you work that links all  of the workstations and hubs together     Cable made of strands of thin fibers  glass or plastic wire  that transmits pulses of light   Optical fiber carries much more information than conventional copper wire and is in  general not subject to electromagnetic interference     In ScanArray
23.  A Quantitation Protocol uses pre defined settings that can be named  saved  and recalled  to use again  Using a quantitation protocol is simple once it is set up  just obtain your  images  select the protocol  and start     Use a Quantitation Protocol if        You have a variety of microarrays       You use more than two fluorophores       If the printing of the slides does not vary from microarray to microarray    e You have complex and advanced settings that you want to use  once the settings  are set up  they can be saved and easily run again     1 5 3 Automatically Quantitating after Scanning    When running a Scan Protocol  you can select to automatically quantitate after the scan   While setting up to run  you must select a quantitation protocol to run  so the quantitation  protocol must already be defined  For more information  see    Running a Scan Protocol    in  Chapter 4     You also can fully automate your scanning and quantitation using Protocol Groups or  Batch Sets  For more information  see Chapter 6     User Manual pbtsupport perkinelmer com 1 7    Chapter 1 Introduction    1 6 Preparing to Scan    To perform a scan  the instrument and the workstation must be connected  and the  instrument must have completed all start up diagnostic tests  Start the ScanArray Express  and warm up the lasers for 15 minutes  and load the microarray or cassette        Note  Lasers can vary when first turned on  After one or two minutes  they become   J approximately 60  stable
24.  Express  fiberoptic cable is used to deliver excitation light from the  external blue laser to the instrument     A dye used to mark DNA  It emits light at a given wavelength when excited  The scanner  needs to know its excitation wavelength and emission wavelength to perform a scan     File Transfer Protocol  a standard Internet protocol  is the simplest way to exchange files  between computers  It is used to upload and or download files from your workstation to  the instrument     A gateway is a network point that acts as an entrance to another network  It is used to  control traffic on your facilty   s network  A gateway address is the address of a particular  gateway on a network     A test sample made of photoresist printed on glass  it is supplied with a new system   When scanned  it displays geometric images on your workstation monitor  It is used to  test image geometry  jitter  and basic function     See Graphical User Interface    The portion of an application program that provides a visual    interface between the program and the user  A GUI consists of windows  dialog boxes   icons  buttons  etc  that allow a user to control the execution of the program and view its  status     An initialization process for motion control devices where the motion control device is  driven to one particular position to calibrate the control mechanism     The output of a scan of one fluorophore     ScanArray  Express    Term    Definition       Image information    Image autosave  
25.  He 121487 6079  13837 Good 50226 49130 49832 48538 594 619 457030 19849 1000 100C  2 1 21 E23 Hs 1189 6279  13837 Good 49577 48783 48945 48151 632 664 420743 25846 1000 100C  N 2 1271 EFA Hs 108371 6489  13837 Good 47868 39817 47323 39072 545 583 1617830 18524 1000  100  28 11 28 1 E25 Hs 2331 6579  13837 Good 49598 48611 48981 47994 617 639 628814 23362 1000 100C  23 11 231 EBAF He 25195 6889  13847 Good 46025 41389 46246 40810 579 603 135229 16636 1000 100C  30 11 30 1 ECGFI  Hs739 46 7089  13847 Good 47085 41815 46495 41225 590 610 1289992 19818 1000  100  31 V1 31 EGF Hs 2230 7283  13647 Good 47505 42410 46945 41850 560 560 1244902 13842 1000 1000  32 11 321 EGLI  Hs55173 7489  13847 Good A7573 42270 47015 417086 564 577 1087787 14293 1000 106    33 11 1 2 EGFL2  Hs 57652 1268  14037 Good 459681 39698 45322 39059 639 6 amp 0 1370246 16536 1000 100C  34 VW 2 2 EGFL3 He 56186 1489  14037 Good 46887 39731 46283 39127 604 644 1999507 18250 1000  100  35 WwW 3 2 EGFL4  Hs 158200 1663  14037 Good 48571 41122 479684 40515 607 623 1389542 16022 1000 100  36 Vw 4 2 EGFLS Hs 5599 1869  14037 Good 49152 41490 48540 40878 612 636 13700 78 179 65 100 0 100 0  7 Ww 5 2 EGFR Hs 77432 2069  14037 Good 47769 41338 47116 40685 653 662 1382411 17034 1000 100C  38  1 6 2 EGRI Hs 738 2269  14037 Good 46159 40301 45542 39884 617 641 1304978 14234 100 100C  33 WW 7 2 EGR4 Hs 3052 2469  14037 Good 46645 40738 46052 40143 593 624 1248259 15994 1000 100C  40 11 8 2 ENG Hs 76753 2879  14037 Good 4
26.  Quality Measurement Formulas       PBTsupport perkinelmer com Page viii    Preface    Preface    Conventions Used in this Manual    The ScanArray Express uses the Windows   2000 or Windows   XP Operating System   We assume that the operator is acquainted with the general use of the operating system   and therefore provide only an overview of using Windows programs     The names of the buttons and their locations are bold  The name of windows and dialog  boxes are in italics  For example     e In the ScanArray Express Main Window  click Scan to start a scanning procedure     All user defined file names  for example results files  must be Windows compatible   i e they cannot contain a slash  a backslash  a dot or any other Windows reserved  character      The Cancel button in all windows exits the window you are working in without saving  any changes you may have entered     Where to Get Help    If you need help installing or operating the ScanArray system  you can contact  PerkinElmer in the following ways        Web Site   http   www perkinelmer com microarray  e Customer Service by telephone   Tel    617 350 9263 or  800  551 2121   Fax   617 482 1380   E mail  PBTsupport perkinelmer com          Please be prepared with the following when you call        serial number of your instrument       version number of the application software  from Help  About on the ScanArray  Express software  Main Window       nature of the problem   e steps you have taken to correct the problem
27.  This chapter describes how to scan  including how to adjust the display of the images  and  how to quantitate  including reviewing the results  using Easy Scan and Easy Quant        by    Tip  To become familiar with how scanning and quantitation work  first use the  Geometric Test Slide that ships with the ScanArray Express  Run an Easy Scan and  Easy Quant on the Geometric Test Slide  using recommended settings  See Chapter 2      Getting Started           3 2 Using Easy Scan       AN    Note  You can stop the scan at any time by clicking Stop on the Main Window  The  image of what was scanned to that point can be saved just as a fully scanned image can  be saved  See Saving or Printing Images on page 3 7        User Manual    When scanning with Easy Scan  you can specify the PMT settings in real time  when you  are ready to scan  If you don   t change the settings  the settings remain the same as the last  time a scan was performed     pbtsupport perkinelmer com 3 1    Chapter 3 Using Easy Scan and Easy Quant    To run an Easy Scan  1  Load your microarray     2  In the Main Window  click Scan  The Scan window opens  select Run Easy Scan     Scan types on systems with an  autoloader     Scan Area        Bun a batch set    C Run a scan protocol    C Run a protocol group    Scan resolution  um    95  Mi  O20  we  5    Autoloader slot from which to load  1 20   all    PMT Laser  Fluorophore Gain     Power         Use Cy5   59 30  Ei Fields for  Easy Scan  settings           
28.  Using the Adaptive Threshold method  the initial spot mask and background mask are  constructed in the same manner as in the Fixed Circle method  but this method will  quantitate low intensity spots better  The Adaptive Threshold is refined pixel by pixel  using a modified statistical testing process  Mann Whitney test   During the test process  an adaptive threshold is determined by comparing eight  8  sample pixels inside the spot  mask with the eight  8  median background pixels in the background mask  The process is  repeated with successively brighter pixels in the spot sample until a statistically significant  difference is found between the background pixels and the spot pixels     The main advantage of the Adaptive threshold method is that it can identify all spot pixels  adaptively and can compensate for any morphological change in spots  i e   if a spot is not  exactly round   A statistical algorithm is used to define the spot  The disadvantage is that  background quantitation may become unstable if a large maximum spot diameter is set to  compensate for spot size variation     To use the adaptive threshold method    1  Select Adaptive threshold and click Set Adaptive Threshold Options     Adaptive Threshold Options x     Diameters    Maximum spot diameter  pm      Inner background diameter  pum         Outer background diameter  um      Spot Identification    pValue     goo C oon    0005    0 00    C 0025 C 9 050 C 0 100    Calculate Default Values l          2  R
29.  a single color        User Manual    pbtsupport perkinelmer com 3 5    Chapter 3 Using Easy Scan and Easy Quant    3  Click OK     3 2 2 1 Changing the Control Image or Source  Experiment  Image    The first image scanned becomes the Control Image  the image to which quantitation  compares the other images  The Source image is the one that is combined to form the  Composite  You can change which image is used for the Control or  if you have three or  more images  which is used for the Source  experiment      To change the control image    1  Onthe Main Window  click File  then Set Control  The Set Control Image window  opens     Set Control Image    The control image is the image to which quantitation compares the other images  Please select the image to  use as the control image     Fluorophore File Name    Bbc0032 H omexchange S amples    GeoS ample  NB beDO32 H omexchange S amples GeoS ample32x32  oF tif     When you click OK  the image you select will appear on the first tab after the composite on the main screen         2  Select the image to use as the Control Image and click OK  The selected image  appears in the first tab after the Composite tab in the Main Window     To change the source image    1  On the display  click Source in the lower right corner  The Select a Source Image  window opens     Select a Source Image    The composite image is formed by combining the control image  the first image after the composite  with  another image  Please select the image to
30.  and unload substrates out of a cassette  The cassette holds several substrates  20 in  the case of the ScanArray Express   HT    A set of letters and or numbers encoded into a pattern of bars that can be read and  interpreted by a device called a barcode reader  Barcodes are most often printed on  labels  with the labels affixed to objects that need to be identified and tracked  plates  and substrates  in this case      Set of scanning protocols for use with an HT     Jitter is the deviation in or displacement of some aspect of the pulses in a high   frequency digital signal  The deviation can be expressed in terms of amplitude  phase  timing or the width of the signal pulse  When the lines and dots in a scanned image  appear jagged there is bi directional jitter  The amount of allowable jitter is less than   1pixel  To adjust jitter  contact PerkinElmer Life Sciences Support     Microsoft Windows image file format     Calibration allows normalization of all ScanArray Express instruments  A calibration file  is a text file that includes the calibration data of an instrument  This file is created at the  factory and should not be changed by the user     PBTsupport perkinelmer com    Term    Definition       Client    Configuration File    Crossover Cable    Daisy chain    Ethernet    Facility Network    Fiberoptic Cable    Fluorophore    FTP    Gateway Address    Geometric Sample Slide    GUI    Graphical User  Interface    Homing    Image    Technical Support 800 551 2121 or
31.  change the palette and  adjust the display to see the spots better  Adjusting the display does not change the data for  quantitation  or the data in the file  only the appearance of the image on the screen     You can change the control image  and if there are more than two images  you can set or  change the source  experiment  image        Scan  Quantitate  Configure  amp  File  File    Configure    Spot Viewer  ite       F              Mouse is at 6699um  17097m Pixel intensity ratio  0 91       Zoom in Zoom Out Zoom Full    Dragging with mouse   Draws Circle from Side Laser 2   634 nm    Remove Circle Not Installed       Full Slide    Laser 4     488 nm   Not Installed    Laser 5   514 nm    Not Installed    Autoloader    iin    oor   Not Installed    Cassette  Load Status   Not Installed    Palette Source        j 4000  Full color threshold   53 7K Loaded  microarray           3x3 Median   Restore  None                Palette  Remove  3x3 Median Filter Auto Adjust    Technical Support 800 551 2121 or 617 350 9263 ScanArray    Express    Using Easy Scan and Easy Quant Chapter 3    To adjust the display    1  The ScanArray Express scans the microarray using the Auto Adjust settings that  provide the best contrast between the spots and the background to make the spots  more visible  Click Unadjusted to remove the auto adjustment and display the raw  data     2  Click the 3x3 Median filter to make the image appear smoother and less    noisy       cleaning up miscellaneous pi
32.  combine with the control image     Fluorophore File Hi ame    Cy5   BbcO0032 Homexchange S amples     If you wish to set the control image instead  click    File    on the main window  and then click    Set Control        Cancel         2  Select the image to use as the Source image  the selected image appears in the first  tab after the Control image tab        Note  With the images still displayed  you can immediately quantitate  See Using Easy  Quant on page 3 8        3 6 Technical Support 800 551 2121 or 617 350 9263 ScanArray     Express    Using Easy Scan and Easy Quant Chapter 3    3 2 3 Saving or Printing Images    If you did not select to save images automatically by checking the Autosave checkbox es   before scanning  you can save the images after scanning  You can also save a portion of an  image to make a smaller file     To save images that were not automatically saved    1     On the Main Window  click File in the Configure  amp  File group  then click Save or  Save as to save the selected image   Click Save All to save all of the images     Open Image Set from  Array Informatics  Save    Save in  Save As Array Informatics    Save All Save Portion of Image  Close View Header    Close All Print    Set Control Image       In the Save Image dialog box that opens  enter a filename and pathname for the  image     To save a portion of an image    1     2     Select an image tab     On the Main Window  click File in the Configure  amp  File group  then click Save  Po
33.  commas     The major sections of a  CSV file are described in Section A 4 1  Each field in the file are  described in the Section A 4 2     Description of the File Format    Table A 5 lists the major sections of the ScanArray Express  CSV file and a brief    description of each     Table A 5  Description of Major File Sections       Section Description   Header Contains information about the product and version number  and the  Array lnformatics database name  if the data is to be saved to  Arraylnformatics    General Contains date  instrument information  user log in name  and other   Quantitation Contains values for the quantitation method selected    Parameters    Quality Measurement  Critieria    Arary Pattern  Information    Image Information    Normalization  Information    Data    Contains the values for quality measurement     Contains descriptive information about the spot size  spacing  and array  layout     Contains information about the image  for example the fluorophore  the  barcode  etc     Contains values for the normalization method     The first row contains the labels and the following rows are the  measurements  one for each spot  There may be up to four Channels   Note  Chx in the following columns represents Channels x   x   1 2 3 4        A 4 2 Description of the File Fields    User Manual    The following is a description of each field in the  CSV file     BEGIN HEADER    PerkinElmer Life Sciences       name of the company    ScanArrayCSVFileFormat 2 
34.  diagram on the left side of the window is the full slide area  showing  the selected scan area outlined in a white box     Select or change a scan area by using the mouse cursor to outline the  desired area on the diagram     Alternatively  you can enter specific X and Y coordinates  or select the  full microarray  as described under Area Co ordinates  below    You can enter precise coordinates here for the scan area  instead of  selecting an area with the mouse     Start position  X  mm   Area Width  mm   Start position  Y  mm   Area Height  mm     Click here to set or reset the scan area to the full microarray     Click here to open the Scan Area for Zoom window where you can zoom  in on the display  and get help with selecting an area     pbtsupport perkinelmer com 3 3    Chapter 3    Item    Using Easy Scan and Easy Quant    Description       Automatically save  images    Automatically save in  Array lnformatics    Check this box to automatically save the scanned images locally  When  you click Start  a dialog box opens where you specify a filename and  path for saving the image s      Check this box to automatically save the scanned images in the  Arraylnformatics database     If your scanner is not integrated with Arraylnformatics  this field is  dimmed and unavailable to select        3 2 2 Adjusting the Display    3 4    The scanned image s  display on the Main Window  a tab for each fluorophore  and a  Composite tab of both fluorophores if two were selected  You can
35.  for Easy Scan    Focus position  jm      Laser power         Laser Settings  IV Allow user to skip laser warm up period       Array  Informatics Other    ro      30       2  Refer to the following table     Item    Description       Allow user to skip  laser warm up period    Settings for Easy  Scan    Focus Position  lum     Laser Power        Check this box to allow the user to start scanning before the lasers are  warmed up  5 or 15 minutes  depending on the type of laser      This does not cause any damage  but the sensitivity of the resulting  scans will vary  If the user is concerned with repeatability or uniformity of  scanning  they should wait for the lasers to warm up     This defines the settings of the focus position and laser power for all  Easy Scans  These settings are not changeable from the Scan window     The only time the focus position needs to be changed is when the user  changes to a different type of slide substrate  such as a gel or  membrane slide  where the imaging surface is at a different height than  it is with a glass side     The default is 0  zero   Valid values range from  150 to 2000  um      The default is 90 percent  Valid values range from 0 to 100 percent        User Manual pbtsupport perkinelmer com 7 3    Chapter 7    7 2 3 Quantitation Settings    System Settings    The quantitation settings let you set some default system settings for quantitation     To change quantitation settings    1  Click the Quantitation tab     Application Se
36.  for Measurement  Control  and Laboratory Use  Europe      CRDH Title 21  CFR 1040 10 Class   Laser Product  USA      EN 60825 1 1994 Laser Equipment  Classification Requirements and  Users Guide for Laser Safety   IEC 825   Class   Laser Product    FCC Part 15 Class A  Radiated and Conducted  USA    EN55011 1991  Class A  Radiated and Conducted  Europe      ICES 003  Industry Canada  Interference   Causing Equipment  Standard  Digital Apparatus  Class A    EN50082 1 Electromagnetic Compatibility   Generic Immunity Standard   Part 1  Residential  Commercial  and Light Industry    EN61000 4 2  Electrostatic Discharge  EN61000 4 3  Radiated Electromagnetic Fields  EN61000 4 4  Electrical Fast Transient Burst  EN61000 4 5  Surge Immunity Requirements    EN61000 4 6  Conducted Disturbances Induced By Radio Frequency  Fields    EN61000 4 11  Voltage Dips  Short Interruptions and Voltage Variations  Immunity Tests    This device complies with Part 15 of the FCC Rules  Operation is  subject to the following two conditions   1  This device may not cause  harmful interference  and  2  This device must accept any interference  received  including interference that may cause undesired operation     This class A digital apparatus meets all requirements of the Canadian  Interference Causing Regulations     Cet appareil num  rique de la classe A respecte toutes les exigences du  R  glement sur le mat  riel brouiller du Canada     ScanArray  Express System    Specifications Appendix C    Sc
37.  for each Channel if there are 2 or more channels    quantitated     Chx N Median   normalized median  Chx N Mean   Chx N  Median B    Chx N  Mean B     Following Columns Repeated for each Non Control Channel    Chx N Ratio of Medians  Chx N Ratio of Means  Chx N Median of Ratios  Chx N Mean of Ratios  Chx N Rgn Ratio   Chx N Log Ratio          END DATA    User Manual pbtsupport perkinelmer com    Page A 13    Appendix Supported File Formats    A 5 _  TIF Files    The  TIFF   TIF  file format    Tag ged  Image File Format    is an industry standard  16   bit file format for storing images  The  TIF files can be transferred between different  platforms     A 14 Technical Support 800 551 2121 or 617 350 9263 ScanArray  Express    Declaration of Conformity Aopen B    This device conforms to the requirements of US 21CFR1040 10 and 1040 11 at date of  manufacture     This instrument is designed and certified to meet the following regulatory and safety  standards     B 1 Electrical and Mechanical Safety Standards     The product is UL listed  USA and Canada  to the following standard      UL 3101 1 1993 Electrical Equipment for Laboratory Use  USA  amp  Canada        The product is T  V certified to the following standard   e EN 61010 1 A2 1995 Safety Requirements for Electrical Equipment for Measurement   Control  and Laboratory Use  Europe    B 2 Laser Safety Standards     CRDH Title 21  CFR 1040 10 Class I Laser Product  USA      EN 60825 1 1994 Laser Equipment  Classification R
38.  for the calculation of the  background statistics     Enter the maximum percentile value for the calculation of the  background statistics        3  Click OK     5 4 6 Normalization Methods    User Manual    Normalization corrects the intensity of each spot for variations in the overall intensity of  the images with respect to the control image     e The LOWESS method  Locally Weighted Scatter Plot Smoothing  carries out robust  locally weighted scatter plot smoothing for both equally spaced and non equally  spaced data  For more information on the LOWESS method  refer to the following    paper     Y H  Yang  S  Dudoit  P  Luu and T P  Speed  Normalization for cDNA Microarray  Data  SPIE BiOS 2001  San Jose  California  January 2001     The paper can be found at the following web address     http   stat www berkeley edu users terry zarray Html normspie html    e The Total method uses the ratio of total of all spot pixels     pbtsupport perkinelmer com 5 17    Chapter 5 Quantitating with Protocols    5 4 7    5 5    Quantitation Protocol Summary    The Summary lists all of the parameters you have selected for quantitation and their  values     To view a summary of the protocol settings    1  Open the Summary window  where you can review the settings for this protocol     Quantitation Protocol     x     Composite   Cy3  Control    Cy5      1  Basic Information    2  Image Registration Zoom In   Zoom Gut  Zoom Full         5  Summary    These are the parameters you have entered
39.  images  1  Click the tab of the image you want to look at  to bring it to the front     2  Use the mouse to zoom in on the details  or to shift the view to a different part of the  microarray     e A mouse click on either the full slide or large image centers the field of view  on the point where you click the mouse        To shift the view to a completely different part of the microarray  click a point  on the Full Slide   To shift the view slightly  click a point in the large image display     e To zoom in or zoom out on the display  click the appropriate Zoom button   located above the full slide view  The white rectangle in the Full Slide shows  the size of the field displayed in the Main Window  and it changes size with the  zoom level     2 4 Technical Support 800 551 2121 or 617 350 9263 ScanArray    Express    Getting Started  Scan and Quantitate the Geometric Test Slide Chapter 2    3  To view the average intensity of a certain area  click the Draw Circle from the  Center Side button  This enables you to draw a circle around the desired area using  the mouse  The average intensity of spots within the circle is displayed        Run Composite   Cy3  Control    Cy5   Log   Diagnostics   Help  Scan  Wt  Zoomin  Zoom Out  Zoom Full    Quantitate          Configure  amp  File  File  Configure Laser 4   488 nm   Not treed  Laser 5   514 rnm   Not Installed    Autoloader  joor   Not Installed  Cassette  Load Status   Not Installed    Artaylniomatics  erven  Not Installed   
40.  is not integrated     ArrayInformatics is a microarray gene expression database and visualization software  available from PerkinElmer  If you are working in an integrated laboratory  refer to the  Microarray Laboratory Integration Guide that shipped with the ArrayInformatics  software     1 1 1 6 The User Interface    The User Interface  the part of the ScanArray Express software that you see displayed on  the workstation monitor screen  lets you enter information or commands  and receive  status information  The user interface opens to the ScanArray Express Main Window   described on the following page     pbtsupport perkinelmer com 1 3    Chapter 1 Introduction    1 2 The ScanArray Express Main Window    When you start the ScanArray Express  the user interface displays the Main Window on  the monitor screen  as shown in Figure 1 2 below  This is your starting point for sending  commands  entering information  or receiving status information  This is also where  scanning images and quantitation results display        Note  Some ScanArray Express systems are for quantitation only  if you are working on  a Quantitation system  you will not see the Scan button or associated windows        Run  Configure  amp  File Buttons          Click Scan to select a scanning type and begin scanning  Status buttons  Display the status of       Click Quantitate to select a quantitation type and begin quantitating  the lasers and other   instrument hardware     Turn lasers on or off     Imag
41.  lt    lt    lt    lt      lt          lt    lt    lt  lt K     Fluorophores beginning with    cannot be scanned with current hardware     Figure 7 2 List of Fluorophores Window    2  Sort the list  optional  by Name  Description  Excitation Peak or Emission Peak   3  Click one of the following in the List of Fluorophores window        Add to create a new fluorophore       Change  after selecting a fluorophore  to modify an existing one       Duplicate  makes a duplicate of the highlighted fluorophore to help you create  a new fluorophore from one that is already close to what you want    e Delete  after selecting a fluorophore  to delete it     User Manual pbtsupport perkinelmer com 7 7    Chapter 7 System Settings    View Usage  after selecting a fluorophore  to display a list of protocols which  include the selected fluorophore  You cannot delete a fluorophore that is being used  by a protocol     4  Click Add  The Fluorophore dialog box displays     Fluorophore    Name     Description     Excitation Peak  nm   Emission Peak  nm      E  s7    Advanced Options      Palette   Red  Use the palette chosen above         For all images of this fluorophore       For composite images only   use the rainbow palette  for simple images     a  core       Figure 7 3 Fluorophore Dialog Box    5  Enter the name and a brief description of the fluorophore in the appropriate fields     6  Enter the Excitation Peak and Emission Peak of the fluorophore  These values can  be obtained from the 
42.  of spots in  all subarrays    Completely Reset  Template    Modify Template  Using Changes Only    This describes the position and size of each spot in a subarray     Enter the number of rows and the number of columns of spots that are  in each subarray     Enter the horizontal and vertical spot spacing  measuring from the  center of a spot to the center of the spot in the next row or column     Enter the spot diameter in um     This forces the spots into alignment for quantitation  The straightening  does not show on the displayed image but affects the quantitation  results     This resets the template     Modifies the template using only the changes just entered        3 3 1 2 Adjusting the Template and Registering Images    After loading the template  adjust it to precisely fit the subarrays  and register the images   The images are not perfectly aligned if traces of individual color such as green and red are  visible around the outside of each spot  Use the Zoom controls as needed to make it easier  to adjust the template and view the registration     To adjust the template    1  Click Adjust Template and Register Images  The Adjust Template and Register  Image window opens  see Figure 3 4 on the next page      Technical Support 800 551 2121 or 617 350 9263    ScanArray 1M Express    Chapter 3    Using Easy Scan and Easy Quant    Adjust Template  controls    Register Image  controls    User Manual    Adjust Template and Register Images    Adjust Template t 93       Zoomi
43.  save images in Arraylnformatics  the rectangle edges with the  mouse  or use the the box    labelled  Area Co ordinates   at Cancel    right        7  Use the following settings  appropriate for the Geometric Test Slide   e Set the Scan Resolution to 10 um     e Ifyou are using an HT instrument  indicate the cassette slot into which you  placed the slide        Select Cy3 as the First Fluorophore  with PMT Gain at 55  this will be the  control image  To change the fluorophore  click the button labelled with the    fluorophore name  select the appropriate fluorophore from the list and click  OK        Select Cy5 as the Second Fluorophore  with PMT Gain at 65  this will be the  experimental  source  image     2 2 Technical Support 800 551 2121 or 617 350 9263 ScanArray    Express    Getting Started  Scan and Quantitate the Geometric Test Slide Chapter 2       Set the Scan Area by entering the following values into the Area Co   Ordinates edit boxes        Item Setting for GeoSlide  Start position  X  mm  0 75   Area width  mm  7 0   Start position  Y  mm  14 00   Area height  mm  7 5       8  Click Start     The ScanArray Express acquires a pair of images from the Geometric Test Slide using the  Cy3 and Cy5 fluorophores  and automatically ejects the microarray when finished     2 3 Viewing the Images    The scanned images display in the Main Window  each with its own tab  Cy3  Cy5  and  the Composite of the two fluorophores  You can click an image tab to bring that image to  t
44.  t Hs 6540 5079  14047 Good 40926 46310 40302 624 888 1413394 160737 1000 100C  53 11 He 248110 5278  14047 Good 4gi07 41721 47484 M098 623 16727 1000 100i      aie Loaded  Spies microarray  Gene name  Show Hide   Show Hide   SetQualiy   Set View  ACVAI Spote by Statue   Columns   Citeia   SpotStatus   Spot Pisete None                         2 5 1 Spreadsheet    Each row in the spreadsheet is the data from one spot  including the gene names and ID  numbers that were imported from the  GAL file  You can scroll vertically to see the data  for each spot  You can scroll horizontally to view the 55 columns of data for any spot  or  click Show Hide Columns to hide any columns you don   t want to see     The spreadsheet is interactive  click on a row in the spreadsheet and see the spots that  created that data in the spot viewer on the left side  and also see those spots highlighted  and centered in the full images by selecting the Cy3  Cy5  or Composite image tab  Click  on a column to sort using that column        Note  If you have Microsoft Excel on your system  you can view the spreadsheet in  Excel by clicking View as Excel Spreadsheet  in the bottom right corner of the  Spreadsheet tab        User Manual pbtsupport perkinelmer com 2 9    Chapter 2 Getting Started  Scan and Quantitate the    The following buttons on the Spreadsheet tab let you set the spot status and select which  data to view  They are described fully in    Viewing the Spreadsheet    in Chapter 3     Button D
45.  that  describe the completed protocol   Parameter Value  No  of subarrays  rows 1  No  of subarrays  cols  1  Rotation of template      0 00    Pin spacing  H  mm  450  Pin spacing  V  mm  450  Rows of spots per subartay 32  Cols  of spots per subarray 32  Spot spacing  H  um  200  Spot spacing  V  um  200  Spot diameter  um  100  Subarray  0 0   top left comer  x  am  6668  top left comer  y  urn  16986  Quantitation method Fixed Circle  Signal range  low     0  Signal range  high     35  Bkand  range  low     5    Bkgnd  range  high      Bkgnd  diam   inner  um  0  Bkgnd  diam   outer  um  0                         Figure 5 6 Quantitation Protocol Summary Window    2  Click Finish to save your protocol     Creating a Spreadsheet Autosave Protocol    Spreadsheet Autosave Protocols specify the filenaming conventions  file type  and file  location to use when automatically saving the quantitation results     Technical Support 800 551 2121 or 617 350 9263 ScanArray     Express    Quantitating with Protocols Chapter 5    To create a spreadsheet autosave protocol    1  On the Main Window  click Configure in the Configure  amp  File group  In the  Configure menu that displays  click Spreadsheet Autosave Protocol  The List of  Spreadsheet Autosave Protocols window opens     List of Spreadsheet Profiles    Description Pattern    Sort By  Name  Description      Path    D       2  Sort the list  optional  by Name  Description  or Path   3  Click one of the following in the List of Sp
46.  the database        7 12 Technical Support 800 551 2121 or 617 350 9263 ScanArray        Express    System Settings Chapter 7    7 5 1 Backing up the Database    You should have a set procedure for when to back up the database  and how to name the  backup file so that  in the event you need to restore the database  the correct backup file is  easily identified     To back up the database    1  Click Back Up Database     2  Use the Browse window to select a folder  A default name for the backup is supplied   and it is recommended to use the default name     7 5 2 Restoring the Database    Restore the database only if absolutely necessary  that is  if the file has been lost or  accidentally deleted or if you have accidentally deleted a protocol        from a previous software version  for example from v1 1 when running v2 0   the    A Note  Avoid restoring from previous versions of the software  If you do restore a backup  ScanArray Express software will have to be reinstalled        To restore the database  1  Click Restore Database     2  Use the Browse window to find the backup file that was named and saved during  backup     User Manual pbtsupport perkinelmer com 7 13    Chapter 7    7 6 Service Features    System Settings    Service features are used for calibration and configuration of the instrument and for  controlling the instrument software operation  Certain service features are intended for use  by authorized personnel only  or under the direction of service or tec
47.  the previous window  the Sensitivity Calibration Area  window     To run Automatic Sensitivity Calibration    1  Open the Scan Protocol   Automatic Sensitivity Calibration window     Scan Protocol   Automatic Sensitivity Calibration    Average spot size  pm   100    Target signal intensity        90  Sensitivity adjustment method   5  Sensitivity Calibration Keep laser power fixed  vary PMT gain      Keep PMT gain fixed  vary laser power    ec 6  Tools    Finish Cancel       Figure 4   10 Scan Protocol   Automatic Sensitivity Calibration Window    2  Use the following table for information on settings        Item Description   Automatically Check this box to have AutoSensitivity run automatically on every scan  calibrate sensitivity with this protocol    for each microarray   Average spot size Enter the average diameter of your microarray spots for the greatest   um  particle dust suppression  If the spots are poorly formed  set the    average spot size to a smaller value for better results     The minimum and maximum spot sizes are 75 and 500 um  The default  spot size is 100 um     User Manual pbtsupport perkinelmer com 4 13    Chapter 4    Scanning with Protocols    Item Description       Target signal intensity   AutoSensitivity automatically adjusts the laser or PMT so the brightest       feature in the Sensitivity Calibration area is at this percentage of full   scale  The Target signal intensity for poorly formed or mottled spots  should be lowered to avoid saturation
48.  to automatically save the scan images and or   Arraylnformatics quantitation data from this range of slots in the Arraylnformatics   Database database    Remove the Check this box to easily remove the quantitation protocol and the   quantitation protocol spreadsheet autosave protocol from all of the slots in the range that you   from all slots in the are configuring    range    User Manual pbtsupport perkinelmer com 6 11    Chapter 6    Item    Automating Scans and Quantitation    Description       Set protocol group    Set scan protocol    Set image autosave  protocol    Set Quantitation  protocol    Set spreadsheet  autosave protocol    Check this box to select one or more protocol groups to run on the  range of slots  Checking this box enables the Protocol Group box   where any configured protocol groups will display     Check this box to select one or more scan protocols to run on the range  of slots  Checking this box enables the Scan protocol box  where any  configured scan protocols will display     Check this box to select an image autosave protocol to save the  scanned images automatically  Checking this box enables the Image  autosave box  where any configured protocols will display     Check this box to select one or more quantitation protocols to run on the  range of slots  Checking this box enables the Quantitation protocol box   where any configured protocols will display     Check this box to select a spreadsheet autosave protocol to save the  quantitation res
49.  um      Spot diameter  pm        100    I Straighten rows and columns of spots in all subarrays  Completely Reset Modify Template ance  Template   Using Changes Only    Figure 3 3 Define Quantitation Template Dialog Box       2  Provide the subarray and spot information as described in the following table     Item    Description       Subarrays    Number of rows and  number of columns of  subarrays    Rotation  Pin spacing     horizontal  Pin  spacing  vertical    Straighten rows and  columns of subarrays    Describe the geometry of the microarray     Enter the number of subarray rows and the number of subarray  columns     If the microarray is skewed on the substrate  you can rotate the template  to align the template to the microarray     Enter the horizontal and vertical pin spacing that was used to create the  microarray     4 5   if 384 well plates were used to print the microarray   9 0   if 96 well plates were used to print the microarray   Custom   to enable the edit boxes where you can enter values for  custom pin spacing    This forces the subarrays into straight rows and columns for  quantitation     The straightening does not display on the image  but affects the  quantitation results     User Manual pbtsupport perkinelmer com 3 11    Chapter 3    Item    Using Easy Scan and Easy Quant    Description       Spots    Rows of spots and  Columns of spots per  subarray    Horizontal spot  spacing  center to  center  um     Spot diameter    Straighten rows and  columns
50.  within the brightest spots     The minimum and maximum target intensities are 50 and 100   The  default target intensity is 90      Sensitivity adjustment For the best signal to noise ratio  go to Fluorophores and set the laser  method power to 95 for all fluorophores  then choose    Keep laser power fixed   vary PMT gain           3  Click Finish to save the Scan Protocol     4 4 6 Scan Protocol  Tools    The Scan Protocol   Tools window provides tools you can use to determine or refine  correct settings for this protocol  After you run a tool and accept the results  the settings  are automatically saved back to the protocol  When you save the protocol  all settings are  saved        A    Note  The Quick Scan can be used to test protocol parameters to see the resulting  scanned image  Only experienced users should use the other scan protocol tools        All of the tools perform specialized scans  and you must have a microarray in the scanner  before you run them  Tools include     e Quick Scan  which allows you to adjust the settings during a scan and save them     e Line Scan to help you determine the appropriate focus position  laser power  and  PMT Gain for various fluorophores        Automatic Sensitivity Calibration  for determining appropriate laser power and PMT  Gain settings  and    e Automatic Focus Calibration for determining the appropriate focus setting     When a tool has finished running  click Accept the Changes to save the information to  the protocol     
51. 0  name and version of the file format    ScanArray Express 2 0   name and version of the software    pbtsupport perkinelmer com    Page A 9    Appendix    Supported File Formats  AIDBServerName  SomeName   Added for AIDB  this row is available only if the  data was intended for AIDB    Number of Columns  62  Actual Number of columns here     number of Data columns   number of data records    END HEADER   BEGIN GENERAL INFO   DateTime  12 12 01 8 00   Date and time when the image was acquired in 24 hour  format   GalFile c   images test gal   gal file which associates names and ids   Scanner    type of scanner used  Model and serial number    User Name    Windows Logged in user name   Computer Name   Computer Name   Protocol  _   Scan Protocol used to scan the images  Quantification Method  Quantiation method    Quality Confidence Calculation    Quality criteria user    User comments    aser comments   Image Orgin 0 0   image origin relative to the scan area  Temperature    temperature of the scanner in volts  Laser Powers  10 10  Naser power   Laser On Time     Laser on time for each laser in minutes  PMT Voltages  10 12   voltages    InstrumentID   Added for AIDB this is the AIDB instrument token returned by AIDB   present only if the data is intended for AIDB    END GENERAL INFO  BEGIN QUANTIATION PARAMETERS  Values in this section depend on the quantitation method selected     For Adaptive Circle  Express  method   Minimum Percentil  value   Maximum Percentile  value   Fo
52. 0C  16 11 He 239600 4273  13837 Good ABZ 49274 47504 613 639 447159 18170 1000 10C  17 11 He 76171 4473  13837 Good A7574 47933 46974 600 592 411147 15838 1000 100C  Ratio  0 83 18 11 Hs 99029 4679  13837 Good 47333 46949 46786 567 600 328119 17344 1000 100i  5 13 4 He 78409 4873  13837 Good A7614 47369 A7044 570 609 322612 17077 1000 10C  20 11 Hs 75511 5079  13837 Good 46962 47158 46348  Gl4 2649 269833 944410 1000 100C  a 4 Hs 2301 5278  13837 Good 48276 47177 47652 624 639 302069 19587 1000    100k Autoloader  22 11 Hs 155402 5479  13837 Good 48902 48671 483 569 629 213724 18912 1000 100C Door   23 ii Hs 799 5679  13837 Good 50429 49894 49828 60 613 289835 21628 1000 100C Not Installed  24 1 1 Hs 96055 5879  13837 Good 49800 49947 49000 600 621 351579 22551 1000 100C i me  2  3 Hs 121487 8073  13837 Good 43130 49632 48536 594 619 457030 19849 1000 100C eee  25 Ww Hs 1189 6273 13837 Good 48783 48945 48151 632 664 4207 43 25846 1000 100C Not aad  Neder 47200 27 Ww Hs 108371 6489  13837 Good 39517 47323 39072 545 563 1617830 18524 1000 100C  3 28 i Hs 2331 8573 13837 Good 49511 48981 47994 617 639 628814 23362 1000 100C  29 WW Hs 25195 6889  13847 Good 41389 46246 40810 579 603 135239 16636 1000 100C  30 11 Hs 73946 7089  13847 Good 4815 46495 41225 590 610 1289992 13818 1000 100C  31 11 Hs 2230 7289  13847 Good 4240 46945 41850 560 560 1244902 13842 1000 100C  32 11 Hs 55173 7489  13847 Good 42270 4705 41706 564 577 1087787 14293 1000 10C  33 11 Hs 57652 1269  14037 Good 39
53. 2  Viewing the Quantitation Results            0    cece eens 3 14  Viewing the Spreadsheet     soncdadseawodaneemetrs o Keeani dee 3 15  Filtering the Spreadsheet Data          0 000 ee 3 16  Viewing the Scatter Plot 2 244 94 ove vans ae aadwk eee aaa s 3 20  Viewing the Distribution Plot       0 4 p08 5  0 44 aana 44 ennme aes tes 3 20  Viewing Information About the Images               0 00000 e eee 3 22  Saving the Quantitation Results        ota ees eee de ieee et xe 3 23    Chapter 4 Scanning with Protocols    OVV EW 553  aa Sie e GAA ERIS We RSS hes Wee a Re 4    Running a Scan Protocol  4 0 v2 8 Ane Se ete eee ARG h eek eed hes 4    Other Scan Settings         ght ete geal MN di i Dae wee Nee aid GN 4 3  Automatic Quantitation after Scanning         0 0 ces 4 4  Creating a Scan Protocol 25 c ccc04 4  estekasa Med dihed aye d Lhihas paomadaues HE wSO 4 4  The Scan Protocol Wizard oc  2 se ealgarda ice Hawai wala late Rasa 8s awe 4 5  Basic Infofmation rese a wea Gee eae E Baud 4 6  Selecting the Scan Area srs    citer acwqa LEN GAS AS oa OOS CARE I OE 6 vores ed 4 7    Fl  orophores 5  lt  wind seed eRe eae E ETE ead Bae ee BN RE OA 4 9    Sensitivity Calibration Areas     4 ag   4cnut gee swesG a4 4a seuohagya sn 4 10  Automatic Sensitivity Calibration           0 0 00  ce eee 4 13  Scam  Protocol  To0olsi034 0 te pagu eds whee acta ee Re we RE 4 14  Using Scan Protocol Tools tases 3 ene g wears  6 oS 2d 5 ee SS eae Ve ee 4 15  Running the Quick Sans 230 0   3  Peace 
54. 3 Distribution Plot    Chapter 2    The Distribution Plot allows you to see trends that are area sensitive on the slide  for  example  if one side of the slide is being over  or under washed  or if a pin is partially    clogged              Spot Index 785  Gene name           2       Distribution Plot of Cy Ratio of Mean Spat       fowm_J    2  Zoomin  Zoom Out Zoom Full    Sia        i    anh                          You can select a column of spots to plot from any one of the images  or any images with  the control image     Select a Column to Plot    Category    Column       General  Cy5  Control   Cy3    User Manual       Ratio of Median Spot   Ratio of Mean Spot   Median of Pixel Ratios   Mean of Pixel Ratios   Regression of Pixel Ratios   M Log2  Ratio of Median Spot   M Log2  Ratio of Mean Spot   A Mean of Log2  Median Spot   4 Mean of Log2  Mean Spot    Ratio of Median Norm Spot   Ratio of Mean Norm Spot   Median of Norm Pixel Ratios   Mean of Norm Pixel Ratios  Regression of Norm Pixel Ratios  M Log2  Ratio of Median Norm Spot   M Log2  Ratio of Mean Norm Spot   4 amp  Mean of Log2  Median Norm Spot   A Mean of Log2  Mean Norm Spot   Std Dev of Pixel Ratios   p  for Regression Ratios    pbtsupport perkinelmer com 2 11    Chapter 2 Getting Started  Scan and Quantitate the    2 6 Saving the Images    To save the images       1  Onthe Main Window  click File in the Configure  amp  File group  then click Save or  Save as to save the selected image   Click Save All to sav
55. 33 nm     two Emission Filter Wavelengths of  570 nm and 670 nm    Technical Support 800 551 2121 or 617 350 9263 ScanArray TEX press    Introduction    User Manual    Chapter 1    1 1 1 3 The ScanArray Express    The ScanArray Express can be upgraded to alternative lasers  filters  barcode reader  and  autoloader  This model supports        up to five Laser Excitation Wavelengths of  488 nm  514 nm  543 nm  594 nm  612  nm  and 633 nm        up to 11 Emission Filter Wavelengths of 508 nm  522 nm  530 nm  549 nm  570 nm   578 nm  592 nm  614 nm  660 nm  670 nm  and 694 nm     1 1 1 4 The ScanArray Express HT    The ScanArray Express HT includes a built in autoloader  and can be upgraded to  alternative laser  filters  and barcode reader  This model supports        up to five Laser Excitation Wavelengths of  488 nm  514 nm  543 nm  594 nm  612  nm  and 633 nm        up to 11 Emission Filter Wavelengths of 508 nm  522 nm  530 nm  549 nm  570 nm   578 nm  592 nm  614 nm  660 nm  670 nm  and 694 nm     1 1 1 5 The ScanArray Express in an Integrated Microarray Laboratory    If your laboratory is integrated with an ArrayInformatics database that resides on a server  or workstation on the network  you can retrieve information from and send information to  the ArrayInformatics database  Selection buttons for ArrayInformatics on the ScanArray  Express user interface are activated if your ScanArray Express is configured for  integration  or dimmed and inactive if your ScanArray Express
56. 5 4 3 1 Defining a Template Using a GAL File    A  GAL file  output from the SpotArray  includes a gene list and information about each    spot on the microarray     To define a template using a  GAL file    1  In the Quantitation window under Template  click From  GAL File  The Open GAL  File dialog box opens     Open GAL File    Look in     samples    a  Contents  gal    File name     Yeh a    Files of type    GAL    gal        5 8 Technical Support 800 551 2121 or 617 350 9263 ScanArray        Express    Quantitating with Protocols Chapter 5    2  Highlight the desired file and click Open   The dialog box closes and the name and path of the selected  GAL file displays under  the From  GAL File button in the Protocol Wizard Template window     5 4 3 2 Defining a Template From Specifications    Using specifications  you establish the array pattern by entering numerical data that  defines the subarrays and spots     To define a template from specifications    1  Inthe Protocol Wizard Template window  click From Specifications to enter values  that define the geometry of the array  The Template Specifications window displays     Template Specifications    Subarrays  Number of rows of subarrays     Number of columns of subarrays   Rotation  degrees   0 00    Horizontal pin spacing  mm     45 C90    Custom   450    Vertical pin spacing  mm     45    90 C Custom  15    I Straighten rows and columns of subarrays       Spots  Rows of spots per subarray     Columns of spots per suba
57. 8660 45059 48060 44459 600 667 926554 36123 1000 1000  a1 11 3 2 EPAS I  Hs 168082 2873  14037 Good 49158 46893 48588 46323 570 620 697723 15581 1000 106 f  2 11 102 EPSI5  Hs79095 3079  14037  Good 50054 48333 49490 47769 564 609 732800 18518 1000 106 f  e 11 112 EPS8 He 2132 3279  14037 Good 51793 49793 51193 49187 606 635 658110 18140 100 106 f  44 1112 2 ERBB2  Hs 173664 3473  14037 Good 5o58 50203 50255 49600 603 622 539871 15390 1000 100  45 1 1 13 2 ERBB3    Hs  199087 3673  14037 Good 50523 49297 49935 48709 S88 609 556735 17283 1000 100C  4 11 142 ERBB4  Hs1939 3879  14037 Good 50739 48960 50095 48316 644 633 521847 15881 1000 100C  a7 V1 18 2 EREG He 115263 4073  14037 Good 48938 48105 48299 47466 639 667 486188 22116 1000 100  4 11 16 2 FGF Hs 75297 427914037 Good 49075 47914 48430 47269 645 764 433834 48323 100 106 f  43 11 17 2 FGO  Hs 248049 4479 14037 Good 48023 47384 47460 46821 563 653 418270 19879 1000 100C  50 1 1 18 2 FGFI  Hs 249165 4673  14037 Good 48719 48264 48100 47645 619  646 477686 19097 1000 1006 f  51 13 19 2 FGFI2 Hs 124752 4879  14047 Good 47333 41085 46743 40495 590 618 1432263 170 61 100 0 100 1    52 11 20 2 FGFI3  Hs6540 5079  14047 Good 46334 40926 46310 40302 624 888 1413394 160737 1000 100  53 XI 21 2 FGF4  Hs 248110 5279  14047 Good 48107 41721 47484 41098 623 648 1420954 16727 1000 100x  4 ae Loaded  Spot Index  1 5 E  Gene name  Show Hide   Show Hide   SetQualiy   Set View View as    ACVRT Spotsby Status   Columns   Crea   SpotStatus 
58. 898 45322 39059 639 660 1370246 16536 1000 100C  34 11 Hs 56186 1469  14037 Good 39731 46283 39127 604 644 1399507 18250 1000 100C  35 il Hs 158200 1689  14037 Good 41122 47964 40515 607 623 13695 42 16022 1000 100C    3 inl He 5599 1889  14037 Good 41490 48540 40678 612 636 1370078 179365 1000 100C  Madian aAA 7 t1 Hs 77432 2069  14037 Good 41338 47116 40685 653 662 1362411 17094 1000 100C  38 11 He 738 40301 45542 99684617  641 1304978 14234 1000 10C  33 Ww He 3052 40736 46052 40143 593 624 1248259 15994 1000 100C  40 WW Hs 76753 45059 48060 44459 600 667 926554 36123 1000 100C  41 WwW Hs 166082 2879  14037 Good 46893 48588 46323 570 620 697729 15581 1000 106  42  Ww Hs  73095 3073  14037 Good 48333 49490 47769 564 609 732800 18518 1000 100C  ry  Ww Hs 2132 3273  14037 Good 49793 51193 49187 606 63 6581 10 18140 1000 100  4  Ww Hs 173658 3473  14037 Good 50203 50255 43600 603 622 599871 15390 1000 100i  45 v1 Hs 199067 3679  14037 Good 49297 49935 48709 588 609 5567 35 17283 1000 100i  46 v1 Hs 1939 3879  14037 Good 48960 50095 48316 644 633 521847 15881 1000 100i  7 v1 Hs 115263 4073  14037 Good 48105 48299 47486 639 667 486188 22116 1000 100i  48 4 Hs 75297 4273  14037 Good A7914 48430 47269 645 764 433894 48323 1000 100C  43 Ww Hs 248049 4479  14037 Good 47384 47460 46821 563 653 418270 19873 1000 100C  50 il He 249165 4679  14037 Good 48264 48100 47645 619 646 477686 19097 1000 100C  51 11 Hs 124752 4879  14047 Good 41085 46743 40495 590 618 1432263 17061 1000 100C  52 
59. Accept Changes to accept the results     Automatic Sensitivity Calibration       Laser PMT    Fluorophore Power       Gain        Scanning    Pending    Progress     Cy3 Cy5       4 5 4 Automatic Focus Calibration    4 22    Only experienced users should use this tool  The Automatic Focus Calibration tool  scans a selected area of a microarray with a single fluorophore and determines the  appropriate focus position     To perform Automatic Focus Calibration    1  Click Run in the Automatic Focus Calibration area of the Scan Protocol   Tools  window  The Automatic Focus Calibration dialog box displays     Automatic Focus Calibration    Select the fluorophore to use for the calibration     PMT   Laser    Fluorophore Gain     Power        y  Fluorophore cannot be  scanned with current hardware    Cancel       Technical Support 800 551 2121 or 617 350 9263 ScanArray     Express    Scanning with Protocols Chapter 4    2  Select the fluorophore to use for the calibration  To change the area to be used for  focus calibration  click Adjust the Focus Line to move to an area with a lot of    features  The Focus Line Definition window opens  where you can move the focus  line     Focus Line Definition  six     Help  amp  Directions Composte   G8   08          Zoomin   Zoom Out  Zoom Full     Li  wT       To Madity the Sean Line    line  and release the mouse    Focus Line       Start position     mm 00    Start pe       LY  mmm   3650  Line length  mm  2200     Line thickness  mnt      
60. In the  Configure menu  click Scan Protocols     Configure    Configure  amp  File  File      Autosave    Configure Protocols    Protocols          Quantitation    Spreadsheet  Autosave  Protocols    Quantitation  Protocols    Automation    Protocol  Groups    Batch Sets  Advanced  Application Service  Settings Features    Eluorophores Barcodes    Database  Maintenance       Figure 4   2 Configure  amp  File group and Configure menu    4 4 Technical Support 800 551 2121 or 617 350 9263 ScanArray        Express    Scanning with Protocols    Chapter 4    2  The List of Scan Protocols window opens  The ScanArray Express includes several  default protocols that display in this window  along with any protocols that have    been created     List of Scan Protocols  Description    Dr  Thomas Dr  Thomas   Dynamic Repeat    Single Cy5 scans  geo slide dots  Geo Slide 32x32    Geo Slide 32x32 Demo  Geometric Geometric    Rotated Uniformity Single Cy5 scan  geo slide dots       Static Repeatabil    4 consecutive Cy5 scans  geo sl         Protocol cannot be executed with current hardware    Resolution  um        Figure 4 3 List of Scan Protocols Window    Modify  Add  Change  Duplicate    Delete    View Usage  Sort By  Name  Description    Resolution    3  Sort the list  optional  by Name  Description  or Resolution     4  Click one of the following in the List of Scan Protocols window to open the Scan    Protocol wizard     e Add to create a new protocol using the Scan Protocol wizard   
61. OWESS  C Total    Figure 3 2 Quantitate Window   Run Easy Quant    2  Use the settings from the last session if you are using the same type of microarray  or  see Section 3 3 1 to change the settings  If you make any changes  be sure to adjust  the template and register the images  See Adjusting the Template and Registering  Images on page 3 12     3  Click Start     The ScanArray Express finds the spots  a message box displays  showing the  progress of finding spots  and then quantitates immediately     3 8 Technical Support 800 551 2121 or 617 350 9263 ScanArray    Express    Using Easy Scan and Easy Quant Chapter 3    3 3 1 Changing the Settings    To change the Easy Quant settings  refer to the table below for information     Item    Description       Quantitation Type    Template    From  GAL file    From specifications    From  Arraylnformatics    Quantitation  Method    Normalization  Method    Adjust Template and  Register Images    Select From Easy Quant     This selection lets you enter or change settings  or use the default or  previous settings     A spot template must be defined that matches the layout of the arrays  on the selected scan  Load a template from one of the methods listed  below  A template of circles appears superimposed on the microarray  image  you can then adjust and align the template     Click this button to open the Open GAL File window  where you can  browse for and select a  GAL file to use for your template  For a  description of  GAL files
62. Sarwar de Oe eee See S 4 15  Runnings Line Stanie o cov saci seed   eee seis Dost ates 4 19  Automatic Sensitivity Calibration          2 0 0    cece eee eee eee 4 20  Automatic Focus Calibration  2 gs02 cas  5 Gystindte ad Sentra erie Sala a SASS 4 22  Creating an Image Autosave Protocol                00  cece eee ene 4 24    Chapter 5 Quantitating with Protocols    OVERVIOW ces i eae ee EA HEP RAR VS ae ee 5 1  Running a Quantitation Protocol        0 0    cece eee eens 5 1  Creating a Quantitation Protocol 4 5 65 426 pee euene pas Seed eee aaa as ees 5 3  The Quantitation Protocol Wizard        2 0    ccc e ee eee 5 4  Basic Information        nnne tee eaten hg ee WAN GAD 4 Rn Oe A 5 5    mage Registration 0 4  44 stn aid Ba Pewee Edw ahd Oo 4 awn le ye owe eh Y 5 5  Template w0  eae eae he Rela ed aae Rae ee ea EAS 5 6  Defining a Template Using a GAL File    l    ce 5 8  Defining a Template From Specifications        0 000 0000 ce 5 9  Quantitation and Normalization Method                 0 000000 e ee 5 10  Quantitation Methods 4 5   oyesi44  2nages say sewada  40 as Maas 5 12  Adaptive CNL san  ooh Bit PUN e he a oh aOR Wha gM ey Rata 5 12  Fixed AECL ori 5 ine Nich hectic TAM ends 2 8 bene Ae Petes era a eee es aw 5 13  Adaptive Threshold oti ahs ore seas be lee cuce rsi etrn dre aie Sled cel are ee na ene 5 15  Histograms ss Salsa Ba ra a Gell AE E EE ries 5 16  Normalization Methods         nss Geniegad He bees COGS ee eS 5 17  Creating a Spreadsheet Autosave Proto
63. Scan Next Fluor          a  Focus pasion   048   a   z Laser power       80     PMT gain   4   aa    Mouse is at  14395m  14155m Raw pixel intensity  0    Loaded  microarray   Fluorophore  Cy3  1 of 2        Loaded                Scan Next Fluor    Focus position  0         s     Controls for adjusting Laser power       80    Laser power and PMT gain PMT gain      80 4    Figure 4   13 Quick Scan Image    6  The scanned image displays on the workstation  The name of the fluorophore is  identified on the tab of each image     You can change the focus position  laser power  and PMT gain during the scan  using  the up arrow and down arrows in the bottom right corner of the window     7  Click Scan Next Fluor to scan the next fluorophore     User Manual pbtsupport perkinelmer com 4 17    Chapter 4 Scanning with Protocols    To view the average intensity of a certain area  click the Draw Circle from the  Center Side button  This enables you to draw a circle around the desired area using  the mouse  The average intensity of spots within the circle is displayed     8     ScanArray Express       Composite   Cy3  Control    Cy5   Log   Diagnostics   Help    2  Zoomin  Zoom Gut  Zoom Full        Scan    Quantitate    Ful Side       Configure  amp  File    Laver 4   488 nm    Not Installed  Laver   514 rm   Not installed    Autoloader       Spot Viewer  Composite    oor   Not Installed  Load Status   Not Installed  Arraylnloimatios    Not Installed    Palette   Source  Black threshold  
64. Step window  select the second option  A Scan Protocol  and a Quantitation Protocol  as shown in the following figure     Protocol Group   Step 1       This step executes     A scan protocol  Select a scan protocol and an image autosave protocol     ion protocol  Select a scan protocol  animage        and a spreadsheet autosave protocol         Scan protocol  Image autosave protocol  Quantitation protocol  Spreadsheet autosave protocol        Name Name Name Name     Dr  Thomas Test1   Dynamic Repeatability   Geo Slide 32x32 Demo    Geometric   Rotated Uniformity   Static Repeatability   4 L E Bp  t Ls    Protocol description  Protocol description  Protocol description  Protocol description   li BDyesStd   Test Quantitation Protocol Test Spreadsheet Protocol      Protocols beginning with   cannot be executed with current hardware     CE j    Figure 5 2 Creating a Step for Scanning and Quantitating    2  Select a Scan protocol  Image autosave protocol  Quantitation protocol  and  Spreadsheet autosave protocol     3  Click OK to save the step to the Protocol Group     4  Repeat for each step that you want to add to the protocol group  You can add as  many steps as you want     6 3 Running and Creating Batch Sets  Batch Sets perform scans on one to twenty slides in an autoloader cassette     If you haven   t done so already  you need to create a Batch Set definition that specifies  which slots in the cassette hold microarrays  and which protocol or protocol group to run  on the m
65. Sypro Ruby  Texas Red  TO PRO   3  TO PRO 1  TOTO  3  TOTO 1  YO PRO 1  YOYO   1 and the Molecular Probes logo are trademarks of Molecular Probes  Inc     FluorX is a trademark of Amersham Biosciences Ltd     Copyright 2002  PerkinElmer Life Science  Inc  All rights reserved  No part of this manual may be  reproduced or transmitted in any form without the written permission of PerkinElmer     Printed in the U S A     Part No  80M 826 Rev  B    Information in this document is subject to change without notice and does not represent a  commitment on the part of PerkinElmer     Contents    Preface  Contents of This Manual        0 0    ccc cece eee eee eens Vili  Conventions Used in this Manual          0 00 00 00 ccc cc eee eee ees ix  Where to Get Help  4 svat kdd ah iaiia We Re aak aa EEEE A BEE t ix  For Additional Information          0  0 00000 cece eee eee ee eee ix  Getting Help for Windows 2000 XP_       0    cece ene nes x    Chapter 1 Introduction    OVET EW a ian ne ote 2 Os aA asa ee ee E Oe eo i as 1 1  The ScanArray Express System    0 0 0 0    cece cee eens 1 1  The Scanning Instrument    sede ses ihe ied ioe 8 es de ROG KO 1 2  CICA ITN LEC ohn rs areas als be aes dd eh cost ite Oe dacs ob dines lk DR ek E eto din 1 2  The ScanArray Express os538 5 3  5 hg Se ah Re Ros ad ae RG aS 1 3  The ScanArray Express HT nica ea DVile ha Pela ne Daled Opinnd dheRea eae d 1 3  The ScanArray Express in an Integrated Microarray Laboratory          1 3  The User TCV I ACE is bo
66. To access the Tools    1  Open the Scan Protocol   Tools window  shown in Figure 4 11     Technical Support 800 551 2121 or 617 350 9263 ScanArray     Express    Scanning with Protocols Chapter 4    Scan Protocol   Tools    Help and Getting Started    The tools on this page can be used to adjust the focus position  and also the laser  powers and PMT gains of the individual fluorophores  All of these tools perform  specialized types of scans when they are run  and require the presence of a microarray  in the scanner     Autoloader slot from which to load microarray  1 20    7    Quick Scan    This tool performs a scan of a selected area using selected  fluorphores  allowing the user to adjust the focus position  laser power  and PMT gain while the scan is being performed     Line Scan    This tool repeatedly scans a single scan line to help determine the  appropriate focus position  and the appropriate laser power and PMT  gains for the various fluorophores     Automatic Sensitivity Calibration    This tool scans selected areas of a microarray with selected  fluorophores to determine appropriate settings for laser power and  PMT gain     Automatic Focus Calibration    This tool scans a single scan line with a single fluorophore and  determines the appropriate focus position     Finish Cancel       Figure 4   11 Scan Protocol   Tools Window    Each of the tools is described in a section below  To use these tools  the instrument  must be connected and a microarray inserted      
67. WORE ae 8 4    Appendix A Supported File Formats    OVEIVIEW uaa aaa a Ae eas Raden pa Fs Bas oP eS as Ba A 1  Gene Array List   gal  Format ia sy etka tae BERS a eons Pas ao eee dae A 1  Description of File Format          nannan nnana a aeneae A 2  Description of the Header SeCU OI ies pa ena Ry ig doa ei a eae ages A 2  Description of the Spot Data Section            nannan cee A 3  Blocks  S  b  rrays  sk esre ceense ren oiie nan a e a a wa eb abi A 4  Example GAL file  ete edhe elie rien a ae O ie RE A 4   GPR Results File Format ee snenc cei doania Wow eines vided Bhs Oa aa A 5  GPR Header son dneni dareen a i a een Be Soe eae eee A 5  GPR Dat   ouea a tie glk a a E A E E a a a N A 6     CSV Results File Format  act cox cas ek rno gh dee Coca A ae OP he xe eG ER A 9    Description of the File Format                        Description of the File Fields     2 44s assnegeenessn oes  PIE Files n 6h Le Mah Rite a theca a T ate 5 ad erat ara au ava    Appendix B Declaration of Conformity    Electrical and Mechanical Safety Standards                  Laser Safety Standards y occ u43 tewearerecwiedce se has  Electromagnetic Emissions Standards                       Electromagnetic Immunity Standards                       FCC Label for Class A Products              0 000000 e eee  ICES 003 Label for Class A Products                  04   Other EU Conformance  lt 4 6 345 fine rove ad eae eae Ms    Appendix C Specifications    Appendix D Formulas    ONEIVIEW  att hi daa kee Hata 
68. a     B 7 Other EU Conformance    The product is CE marked in conformance with the following directives        89 336 EEC  A92 31 EEC  A93 68 EED EMC Directive  Europe       72 23 EEC  A93 68 EEC Low Voltage Directive  Europe      B 2 Technical Support 800 551 2121 or 617 350 9263 ScanArray  Express    User Manual    Specifications    Appendix    The following specifications are for ScanArray Lite  ScanArray Express and ScanArray HT  The  specifications are the same for each model except where noted     Table C 1  Specifications    ScanArray  Specifications    Description       Sample Size    Scanning Field  Pixel Resolution    Excitation Wavelengths    Emission Wavelengths    Sensitivity    Scan Speed   Dynamic Range  Repeatability and Uniformity  Image File Formats    Palette Selections    Workstation  minimum  specs     Size    Scanner    External Laser    pbtsupport perkinelmer com    Standard or metric slides    Width 0 97   1 02   24 6   25 5 mm   Length 2 95   3 0   75 0   76 2 mm   Thickness 0 037   0 051     0 93   1 29 mm     22 mm x 60 mm Norm  22 x 73 allowed  User selectable for each acquisition  5 10  20  30 and 50 microns    User selectable  depending on the model   Lite  543 and 633 nm  Express and HT  Up to 5 of 488  514  543  594  612 and 633 nm    User selectable  depending on the model    Lite  570 and 670 nm   Express and HT  508  522  530  549  570  578  592  614  660  670 and  694 nm    Lite  0 05 fluorescent molecule ym   Express and HT   lt 0 1 fluore
69. a scan protocol display  as shown in Figure 4 1 above  The settings for  performing the scan  including the scan area  are included in the scan protocol itself   Unlike when using Easy Scan  you cannot change the scan area or settings in this  window     You can set or change settings when running the scan  as described in section 4 2 1     3  Select your scan protocol and image autosave protocol  You can obtain this  information from barcodes on the microarrays  See Other Scan Settings on page 4 3     4  Click Start  The scan protocol scans the microarray and saves the images using the  Image autosave protocol  The images are displayed in the Main Window     4 2 Technical Support 800 551 2121 or 617 350 9263 ScanArray     Express    Scanning with Protocols Chapter 4    4 2 1 Other Scan Settings    You cannot change the protocol settings from this window  but you can change some of  the settings for how the Scan is to run  including getting the protocol information from    barcodes     To change the scan protocol settings  refer to the table below for information        Item Description   Scan type The scan types that display depend on your system   Without an autoloader the types are  Run Easy Scan  Run a scan  protocol  Run a protocol group   On an HT autoloader system the types are  Run Easy Scan and Run a  batch set    Scan area The scan area has been defined in the scan protocol  If you want to    Obtain scan protocol  from barcode    Scan Protocol    Obtain image  autos
70. able for information  The  default value of each item is the value specified in the protocol   s parameters     Item Description       Scan resolution  um  The minimum and maximum resolutions are 5 and 50 microns   Focus position  um  This is the distance from the lens to the calibrated nominal focus   Fluorophores Check the fluorophores you wish to use for the scan     PMT Gain This value can be changed at any time during the scan  Enter the  PMT Gain for each selected fluorophore     Laser Power This value can be changed at any time during a scan  Enter the  Laser Power for each selected fluorophore        Technical Support 800 551 2121 or 617 350 9263 ScanArray     Express    Scanning with Protocols Chapter 4    3  Set the area to be scanned by clicking Show Zoom Area and drawing a rectangle   using the mouse  on the image of the slide in the Quick Scan window  Alternatively   enter the X and Y coordinates if preferred  Click OK to return to the Quick Scan  window     The scan area will default to that specified in the protocol   s parameters   4  Click Start     5  The Quick Scan image displays     ScanArray       Run Composite   Cy3 Contiol    Cy5  Source    Log   Diagnostics   Help    aE    Scan    Stop            Zoom In  Zoom Out  Zoom Full          Dragging with mouse   Draws Circle from Side  Remove Circle  Configure  amp  File  Ele  Configure    Spot Viewer  Composite       Palette    Black threshold   4000  Full color threshold   j  583K    SiS Median  Restore    
71. ady close to what you want     Or  click     e Delete  after selecting a protocol  to delete a protocol     5 4 The Quantitation Protocol Wizard    The Quantitation Protocol wizard opens with the Basic Information window displayed  as  shown below  From this window  using the wizard  you can display and set all of the  quantitation parameters  You can move through the windows by clicking the numbers on    the left  or by using the buttons on the bottom of the window     Quantitation Protocol    Zoomed View    Click the numbers to move  from one window display to  another        1  Basic Information    2 Template    3  Quantitation and Normalization      i summay       1  Basic Information       Protocol name        Description    EZ Quant default protocol          2  Zoom In  Zoom Dut  Zoom Full                  Figure 5 2 The Quantitation Protocol Wizard    e Click Next to move to the next window and Back to move to the previous window   e Click Finish to save all of your changes and close the wizard   e Click Cancel to close the wizard without saving any of your changes     You can change the image display  making it larger or smaller  using the Zoom buttons     5 4 Technical Support 800 551 2121 or 617 350 9263    ScanArray TM Express    Quantitating with Protocols Chapter 5       Zoom In  Zoom Out  Zoom Full   Click these buttons to zoom in or zoom out the  display  or to restore the display to the full slide view  These controls remain visible  and selectable throughout 
72. all Columns   Hide all Columns Show all Columns     Hide all Columns  in all Categories in all Categories in this Category in this Category    Cancel l       2  Check the corresponding boxes for the columns you wish to show or hide   To quickly select entire groups  click on the appropriate Category     3  Click OK to implement the changes     User Manual pbtsupport perkinelmer com 3 17    Chapter 3    Using Easy Scan and Easy Quant    To set or change quality measurements    1  Click Set Quality Criteria to set or change quality measurement criteria The  following window opens     Set Quality Measurement CICH X     Quality measurement method        Signal to noise     Signal to background    Quality Measurement Parameters    Maximum footprint  um   100    Minimum signal to noise ratio  3    Lower limit     Multiplier       400    1 70    Apply Set Criteria as    New Default    Eancel    2  Set the criteria you want to use  The default method is Footprint  Refer to the  following table for information           Item Description   Quality Select one of the methods below  Footprint is the default method    measurement   menog The formulas used by the different methods use the settings in the  ScanArray Express Application Settings window  The formulas are  provided in Appendix D    Footprint This is the distance between the expected position of a spot and its    Signal to noise    Signal to background    Quality  Measurement  Parameters    actual position  Choosing this method 
73. amples GeoSample32x32 Cy5 tif    C  Program Files PerkinElmer ScanArray  Express Samples GeoSample32x32 Cy3 tif     PMTVolts 600 600    NormalizationFactor RatioOfMedians 0 819944    NormalizationFactor RatioOfMeans 0 848769    NormalizationFactor MedianOfRatios 0 843367    NormalizationFactor MeanOfRatios 0 427747    NormalizationFactor RegressionRatio 1 0055      Jpegimage       RatioFormulation W2 W11  543 633      pbtsupport perkinelmer com    data fields  columns    Type of ATF file   Date and time when the image was acquired     The name of the settings file that was used for  analysis     The Array Content List file used to associate  Names and IDs to each entry     Type of scanner used to acquire the image   User entered file comment    Resolution of each pixel in um    Name of each image buffer     File name for each image buffer  In a multi image  tiff file  the same name is repeated     The voltage of the PMTs during acquisition   The Ratio of Medians Normalization Factor  The Ratio of Means Normalization Factor   The Median of Ratios Normalization Factor   The Mean of Ratios Normalization Factor   The Regression Ratio Normalization Factor     Name of optional compressed image saved with  the results file     The ratio formulation of the ratio image  showing  which image is numerator and which is  denominator     Page A 5    Appendix    Entry    Supported File Formats    Description        Barcode 00331    ImageOrigin 0  0    JpegOrigin 390  4320      Creator ScanA
74. anArray    Specifications Descnpiion       Other EU Conformance The product is CE marked in conformance with the following directives   89 336 EEC  A92 31 EEC  A93 68 EED EMC Directive  Europe   72 23 EEC  A93 68 EEC Low Voltage Directive  Europe         Specifications as of September  2002  PerkinElmer Life Sciences reserves the right to  change these specifications without notice     User Manual pbtsupport perkinelmer com Page C 3    Appendix C Specifications    C 4 Technical Support 800 551 2121 or 617 350 9263 ScanArray  Express System    Formulas    D 1 Overview    Appendix    This chapter lists the formulas used by ScanArray Express for Quantitation     D 1 1 Quality Measurement Formulas    ScanArray Express calculates a set of raw measurements and marks spot quality based on  the method and corresponding threshold values specified in the application settings  see    Chapter 7   and reports them in the export file     Next to each quality measurement below are its raw measurement and criterian for    marking spots as    good        Table D 1  Quality Measurement Formulas       li   riterian for markin   Quality Raw Measurement Calculation   t a orma g spots as   Measurement Good   Footprint For each subarray  calculate the difference Spots with a calculated footprint less than the  between the center of the nominal spots and maximum specified in the application settings  the center of the found spot  then shift the are good   nominal spots by the difference  Let the  sh
75. anArray Express can read barcodes on the  microarrays  Using barcodes as unique identifiers on individual substrates provides a  valuable way for tracking samples and workflow  and automating analysis of microarray  experiments  Using barcodes allows a gene sample to be tracked through the entire  microarray process including printing  hybridization  scanning  quantitation  and  visualization     For systems that are integrated with ArrayInformatics  unique barcodes on microarrays are  required and allow you to retrieve information from and automatically save data to the  ArrayInformatics database     7 4 1 Supported Barcodes    The ScanArray Express barcode reader supports Code 128  Code 39  and Interleaved 2 of  5  ITF  symbologies by default  The ScanArray Express barcode reader may be  configured to support additional symbologies with help from technical support  The  maximum number of digits supported by each symbology depends on the minimum  element width  the barcode    pitch     in the printed barcode  Alphabetic and numeric digits  can be combined  but alphabetic digits take more space than numbers  thus decreasing the  maximum number of allowed digits     Table 7 1 lists the maximum number of alphanumeric characters for substrate barcodes     Table 7 1  Substrate Barcodes       Element Width Number of Characters   Inches Mm Code 128 Code 39 ITF   010     254 mm 6 2 6   0075     19 mm 12 4 10   005     127 mm 22 8 16       User Manual pbtsupport perkinelmer com 7 9  
76. ar So gota eed a hee ang WMP ORY ENE Reet Oa ER 1 3  The ScanArray Express Main Window                0 0 cece cence ences 1 4  Workflow for Scanning and Quantitation         nosaus eee eee eee 1 5  The Different Ways to Scatti  i4ae cies eee GG ae GA een oe ewe 1 6  Easy Scans ccc  keenest oma wend ee tA Rae eee hee Ga See ee 1 6  Scan Protocols aaa e 9G  4g waste eae bg deed AE 4 Simoes tw N ea ea ANT 1 6  Protocol Groups and Batch Scans             0    cece eee eee 1 6  The Different Ways to Quantitate   co 2c0 4vatiadasecenwaeaeohieiasda s 1 7  Easy CAIN ses cis ore  aided aren bg hee gdh A hdl gy aoe Faded A Aah a Bae ae 1 7  Cuianititation PIOOcOhe 0 46  Aetauig dd edonne ede ew as Re OREO RES 1 7  Automatically Quantitating after Scanning                 00 0000 0s 1 7  Preparitig  to San 43 ses ell gs ene se eee a Rates edo eae Wes eee RS 1 8  Turning on the ScanArray Express and Warming the Lasers              1 8  Loading the Microarray or Cassette      o   n unnan nnana cee 1 9  Exiting the ScanArray Express Software              0 cece cece eee teens 1 9    Chapter 2 Getting Started  Scan and Quantitate the Geometric Test Slide    OVETVIEW Ci seh gal Lita  eae hg etn ta aac charge E a e E eit nca cars Ast ath a Ghia 9X 2 1  Scanning the Geometric Test Slide 23 sia  fywidwd aetteteeueetewdd aves 2 1  Wel Wit OLGA OES ns n n a2 a  6  chau a 4 209 Wd a aed Renee sae DS 2 3  improving the Image Display v4 0 gcse a4 oe  o eGR A RAG SESS 2 5  Changing the Palette for t
77. atics    GAL file from which to read the  gene list and spot statuses       Bbc0032 HomeXchange   S amples GeoS ample3       Quantitation method  Easy Quant    Adaptive circle    Adaptive threshold settings     Fixed circle C Histogram    Adaptive Circle Options   Current options match defaults     Normalization method      LOWESS  C Total       Adjust Template and    Register Images       3  Load an array content template in the form of a  GAL file     e click From  GAL File   e In the Open GAL File dialog box that opens  browse to   C  Programs Files PerkinElmer ScanArray Express Samples    and choose the file    GeoSample32x32 gal               The path may be different if the ScanArray Express software was installed in a  folder other than the default         click Open     A template of circles appears superimposed on the microarray images        Note  When scanning the Geometric Test Slide  you don   t have to adjust the template   When scanning your own slides  you need to adjust the template as described in steps 4  and 5  For more information  see    Adjusting the Template and Registering Images    in  Chapter 3        User Manual    4  Click Adjust Template and Register Images  This opens a Zoom window  where  you can align the template precisely  and register the images if necessary     pbtsupport perkinelmer com 2 7    Chapter 2    8     9     Getting Started  Scan and Quantitate the       Adjust the template by dragging it with the mouse until it is aligned with 
78. ation method 5 16    Image autosave protocols  creating 4 24   Image file formats C 1   Image  poor uniformity 8 4   Images    changing the palette 2 5  improving image display 2 5  registering 3 14  5 5  saving a portion 3 7   saving or printing 3 7  viewing 2 3    Instrument    setting network options 7 16  shutting down 1 10    Instrument dimensions C 1  Instrument Laser  excitation filter 1 2    Laser power  setting 4 10    Index    Laser safety C 2  Line scan  running 4 19  LOWESS normalization 5 17    Mask  for quantitation 5 12  Microarray  jammed slide 8 3    Normalization method    LOWESS 5 17  Total 5 17  Palette    changing 2 5  Pixel resolution See Scan resolution  PMT gain 4 10  Protocol groups  creating 6 3  description 6 1  running 6 2  Protocols    for quantitation 5 1   for scanning 4 1   image autosave 4 24  protocol group 4 1  5 1  spreadsheet autosave 5 19    Quantitation    different ways to quantitate 1 7  saving results 4 1  5 1    Quantitation method    adaptive circle 5 12  adaptive threshold 5 15  choosing 5 11   fixed circle 5 13  histogram 5 16    Quantitation Protocol 5 1    creating 5 3   defining atemplate 5 4  running 5 1   spreadsheet profile 4 1  5 1  wizard 5 3    Quantitation Protocol Wizard    basic information 5 4  quantitation methods 5 10  summary 5 18   template 5 6    Quantitation Results     CSVfile format A 9   GPR file format A 5    Quantitation results    distribution plot 3 20  scatter plot 2 10  spreadsheet 2 9  viewing 2 9    Q
79. ave protocol  from barcode    Image autosave  protocol    Perform automatic  quantitation    Quantitation  Protocol    Spreadsheet  Autosave Protocol    Automatically Save in  Arraylnformatics    change the scan area  you must select a different protocol     Check this box to read the barcode of the microarray for the appropriate  scan protocol  If a barcode reader is not installed  or if the barcodes are  not configured  this box is dimmed and unavailable to select  See     Configuring Barcodes    in Chapter 7     To select a scan protocol  click the Select a Scan Protocol button  and  select a protocol from the List of Scan Protocols window that opens  The  button on the Scan window fills with the name of the protocol     Check this box to read the barcode of the microarray for the appropriate  image autosave protocol  If a barcode reader is not installed  or if the  barcodes are not configured  this box is dimmed and unavailable to  select  See    Configuring Barcodes    in Chapter 7     To select an image autosave protocol  click the Select an Image  Autosave Protocol button  and select a protocol from the List of Image  Autosave Protocols window that opens  The button on the Scan window  fills with the name of the protocol     Make sure this box is not checked  unless you want to automatically  quantitate after scanning  See Automatic Quantitation after Scanning  on page 4 4     Checking this box enables the following fields for selecting a  quantitation protocol and spr
80. batch set  to modify an existing set    e Duplicate  makes a duplicate of the highlighted batch set to help you create a  new batch set from one that is already close to what you want     Or  click        Delete  after selecting a batch set  to delete the set     6 8 Technical Support 800 551 2121 or 617 350 9263 ScanArray    Express    Automating Scans and Quantitation    Chapter 6    4  Click Add  The Batch Set window opens     Batch Set    Name    LEST    Description     TEST    Slot Status Protocol Group    IV Read scan protocols from barcodes  IV Read image autosave protocols from barcodes  IV Send email upon completion of scan    Email address     Save in  Arrayinformatics    Image Spreadsheet    Scan Protocol Autosave Protocol Quantitation Protocol Autosave Protocol       20       19       From Barcode       From Barcode          18  a       16  15       14                                           From Barcode       From Barcode          Configure a Range of Slots      items beginning with   cannot be executed with current hardware        Figure 5 3 Batch Set Window    5  Enter a name for the batch set  and specify parameters for the set  and configure each  slot to be used  Refer to the following table for information        Item Description   Name Enter a name for the batch set  use a name that will help you identify the  batch set    Description Enter a description for the batch set     Read scan protocols  from barcodes    Read image autosave  protocols from  barc
81. ce using it  to ensure that undesirable pixels are not  included in the results  You should first run some quantitations and look at the View Spot  Pixels in the Spreadsheet tab to understand what is being included in your quantitation    results     To use the histogram method    1  Select Histogram and click Set Histogram Options     Quantitation Options   Histogram Method    Histogram for the Average Spot Patch  Cy3       Frequency             Plotted fluorophore     Cys       Signal Level    Histogram Key  The plot line shows the histogram for the selected fluorophore   The background range is drawn as a black rectangle   The signal range is drawn as a dark green rectangle      You should select ranges that do not overlap  If the ranges do  overlap  the overlap area is drawn as a dark red rectangle           Percentile Ranges  Signal range         Background range               5 16 Technical Support 800 551 2121 or 617 350 9263 ScanArray     Express    Quantitating with Protocols    Chapter 5    2  Set the signal range and background range  Refer to the following table     Item    Description       Signal range      Low    High    Background range    Low    High    Enter the percentile for the following values  where percentile is the  percentage of pixels in the sample     Enter the minimum percentile value for the calculation of the spot  statistics     Enter the maximum percentile value for the calculation of the spot  statistics     Enter the minimum percentile value
82. code Units X Units Per Pixel  Y Units  Per PixeL X Offset  Y Offset Status   An example of the following rows    ch1 F  Program Files Packard BioChip DSOMA ExperimentSets qa   4 Images TAMRA TIF TAMRA04030270  EOO0CS Microns 20 00  20 00  0   0 Control Image    END IMAGE INFO    BEGIN NORMALIZATION INFO    Normalization Method     Normalization method  Normalization Factor     normalization Factor  END NORMALIZATION INFO  BEGIN DATA    The first row contains the labels and the following rows are the measurements  one for  each spot     Note  Chx in the following columns represents Channels x   x   1 2 3 4   The labels of the columns   Following columns present for each spot    Index   Array Row  Array Column  Spot Row  Spot Column  Name   ID   X   Y   Diameter   F Pixels   B Pixels   chx Footprint  Flags  Following columns Repeated for each channel  Chx Median  Chx Mean    A 12 Technical Support 800 551 2121 or 617 350 9263 ScanArray  Express    Supported File Formats    Chx SD   Chx B Median   Chx B Mean   Chx B SD   Chx   gt B 1SD  Chx    gt  B  2 SD  Chx F   Sat    Chx Median     B   Chx Mean     B   Chx Signal Noise Ratio       Following columns Repeated for Each Non Control channel    Chx Ratio of Medians  Chx Ratio of Means  Chx Median of Ratios  Chx Mean of Ratios  Chx Ratios SD   Chx Rgn Ratio   Chx Rgn R    Chx Log Ratio    Following 2 columns are present if there 2 or more channels quantitated    Sum of Medians  Sum of Means    Appendix    Following Columns Repeated
83. col             0    c eee eee eee eee 5 18    Chapter 6 Automating Scans and Quantitation    Overview ny hart fete eri VE ee ae al eat eda att A et amet bh ene ale Ce Ly 6 1  Running and Creating Protocol Groups             0    c eee eee eee 6 2  Creating a Protocol Group  os 64 ae ja tmcee a annann eare 6 3  Running and Creatine Batch Sets aicas yc cue ei Rae Roe 8 hal Se IVE hala 6 6  Creating  Batch Set  144 cceua iia e a gay aes Cau ae aes 6 8  Co  feunng AS lOc ycz 5 2 they iria Sead aE OS ISS ES 6 10  Configuring a Range Of Slots  a5 0r2e58 Sh Sah we oa SO am 6 11    Chapter 7 System Settings    OVERVIEW  ater iin r dad alot ea aa i Cae eae ee wae ea aaa aa 7 1  ScanArray Express Settiigs  lt  5 aru unnan AG eee He ta eeeen on tangs 7 2  Connection Settings        n    susunan are gd ahs Ss ee eda ood 7 2    SCANNING Setting Ss an tees aka oe ai oid Wee wees Boas Sn dole een BS 7 2    Quantitation Settings   44 04 44 sande teeeu4e ada aeusle eed ae ckaeadgaeas 7 4  Arraylnformatics Settings  oasis dette aie tA etree nneur 7 5  Other Settings rrgs hed a othe Bod a ued eee at tet Le ote a 7 6  Configuring Fluorophores            0 00 e een n eee 7 7  Defining a Fluorophore  4 2004 saosin de Oe eae Rea eae ae eens 7 7  Contivdring Barcodes    244 20ideaavadate poy eee ete COG deers 7 9  Supported Barcodes    dics eaieds Svea w see a aa Rae ea dO Rs eG eS 7 9  Defining Barcode Parameters ys   lt        nnn dence acorns  5  a ane 8 Sane Pagers 7 11  Database Maintenance   
84. ctions    The ability to register images is provided so  that you can test the quantitation template   The amount by which you rotate or shift the  images will NOT be saved as part of the  quantitation protocol    If the image you would like to shift is not part   posite image  click the button    Set  Source for Composite    and select the image  you wish to shift        The first image in the composite  the control  image  can be rotated  but cannot be shifted                    Figure 5 3 The Quantitation Protocol  Image Registration Window    Click the Composite tab  and adjust the Source  experiment  image to the Control       To register the next image  click Set Source for Composite Image and register it to    the Control image in the Composite tab  Do this for each image that needs to be  registered     4  Click Next or click    Template    to go to the next window     5 4 3 Template    Specify the template to be used by this protocol  A template must be defined that matches  the layout of the array on the scanned image  You can import a template from a  GAL file  or you can manually define the array pattern  If you use a  GAL file  it can be a local file or  on a network drive  or it can be accessed from ArrayInformatics  for ArrayInformatics    users   The advantage of using a  GAL file is that it also imports gene names for the spots  from the file     The defined template overlays onto the scanned image a template of circles that is closely  aligned to the microar
85. dinates    Set Scan Area to Full  Microarray    Show Zoom window    Select Run Easy Scan     The scan types that display depend on your system  You may see Run  Easy Scan  Run a scan protocol  Run a protocol group  or on HT  autoloader systems  you will see Run Easy Scan and Run a batch set     Select a scan resolution  Scan resolution represents the individual pixel  size in the scanned image  and can be set to 50  30  20  10  or 5  microns  Higher settings provide a quicker scan with reduced detail     The default resolution is 10 microns     For instruments with an autoloader  specify the slot number of the  microarray to be scanned  The default is 1  If your system does not have  an autoloader this field is dimmed and unavailable     Select one  first checkbox  or two fluorophores to be scanned and set  the PMT Gain for each     Check the box for a fluorophore  and click the button below it to open  the Select a Fluorophore window  Find and select your fluorophore in  the list  and click OK  The window closes  and the button fills with the  fluorophore name     Set a value for the PMT Gain for each selected fluorophore  The value  defaults to the value used in the last scan session     NOTE  Increasing the PMT Gain percentage improves sensitivity  but  also increases background noise and causes saturation  A PMT Gain of  approximately 50 percent lowers the background  The laser power is set  in the ScanArray Express application settings and cannot be changed  here     The
86. disk space to extend the amount of RAM available to a  program  Windows 2000 supports virtual memory and allows you to specify the amount  of hard disk space to use as virtual memory  Windows 2000 also manages the swapping  of data from RAM and the hard disk space so that it appears to your application  program s  as though the virtual memory is RAM  This reduces the amount of physical  RAM required     ScanArray Express uses RAM to save image data  The following chart provides a guide  for the amount of additional virtual memory to allocate on your workstation when running  other applications concurrently with ScanArray Express         Scans    Additional Virtual  Acquisition Scan Resolution Memory Required  1 5 microns 185 MB   1 10 microns 45 MB   2 5 microns 245 MB   2 10 microns 60 MB   3 5 microns 305 MB   3 10 microns 75MB   4 5 microns 365 MB   4 10 microns 90 MB   5 5 microns 425 MB   5 10 microns 105 MB       To change Virtual Memory   1  Login as Administrator   2  Select Start Settings System   3  Select the Performance tab   4  Click the Virtual Memory Change button     The three axes associated with the sample  X is across the narrow dimension of the  slide  Y is the length of the slide  Z is perpendicular to the slide  In ScanArray Express   lines are in the X direction  Lines are selected by moving the slide in Y and focus is lens  movement in Z        User Manual    PBTsupport perkinelmer com    Index     GAL files  format A 1   GPR files  format A 5   TIFF f
87. e Display Area Click a tab to view images or results        Run  Cyanine 3  Control   Log   Diagnostics   Help          Zoom In  Zoom Out  Zoom Full   Dragging with mouse   Draws Circle from Side    Scan             Quantitate  Remove Circle  Configure  amp  File  Eile    FullSide       Configure  Not Installed    Spot Viewer    Laser5  Cyanine 3  514 nm     Not Installed        Palette  Black threshold             2320  Full color threshold        45K Loaded  microarray             3x3 Median   Unadjusted  None                Mouse is at 14305m  12915m Raw pixel intensity  2166           Click File to open or save a file       Click Configure to create protocols  protocol groups  and batch sets  and to configure system settings        Figure 1 2 The ScanArray Main Window    1 4 Technical Support 800 551 2121 or 617 350 9263 ScanArray TMEX press    Introduction Chapter 1    1 3 Workflow for Scanning and Quantitation    The ScanArray Express performs two basic operations  scanning and quantitation   Depending on a workflow that you choose  scanning and quantitating can be done  separately and individually  with manual or automatic saving of results  or the two  operations can be partly or completely automated using protocols and protocol groups  A  Batch Set is used for scanners with autoloaders     The diagram below is an overview of the ways to scan and quantitate  and points you to  information about each     Scan    Click Scan and select from  See    Different Ways to Scan
88. e all of the images    a Op Open Image Set from  ESE T AATE  Save      Save in  Save As Array Informatics  Save All Save Portion of Image  Close View Header  Close All Print  Set Control Image  2     In the Save Image dialog box that opens  enter a filename and pathname for the  image     2 7 Saving the Results    You can save the results as a  GPR file or as a  CSV file  The format of these files is  described in Appendix A     To save the results    1  Click the Spreadsheet  Scatter Plot  or Distribution Plot tab     2  On the Main Window  click File in the Configure  amp  File box  then click Save and    provide a filename and file type     2 8 Scan and Quantitate Your Microarrays    You have completed scanning and quantitating the Geometric Test Slide and saved  the results  Use the same process for your own microarrays  specifying at each step  along the process the settings that are appropriate for your microarray  See Chapter 3  for more information on selecting settings for Easy Scan and Easy Quant     Technical Support 800 551 2121 or 617 350 9263 ScanArray    Express    Using Easy Scan and Easy Chapter 3    Quant    Chapter Summary    Overview 3 1   Using Easy Scan 3 1   Using Easy Quant 3 8   Viewing the Quantitation Results 3 14    3 1 Overview    The Easy Scan and Easy Quant settings  once set  remain set until you change them  This  makes it easy to scan microarrays with the same spot configuration and quantitate the  resulting images using the same settings    
89. e barcode or label toward you  slide the  microarray into the slot until it touches the back  The ScanArray Express will  automatically align the microarray     To load a cassette into the autoloader    If you have a ScanArray Express system with an HT autoloader  you can load up to 20  slides in the cassette        Pull the cassette straight up and out     Inside the cassette are    fingers    that hold the slide in place  To prevent breaking  these fingers  load the microarray slides from the back of the cassette    Insert a microarray  face up into each slot you want to use  slot 1 is on the bottom   slot 20 on the top  The barcode or label should be oriented so that it is toward the  front of the instrument when the cassette is replaced in the autoloader     Align the microarrays on the front of the cassette  using the metal alignment block  that ships with the autoloader  The slides must be aligned to be    recognized       Replace the cassette into the autoloader and close the door     1 7 Exiting the ScanArray Express Software    User Manual    You can exit the ScanArray Express user interface and the system saves the current  settings     pbtsupport perkinelmer com 1 9    Chapter 1    Introduction    To exit the ScanArray Express    1  Click the X in the top right hand corner of the ScanArray Express Main Window  The  following message displays     Exiting ScanArray Express    Exit the program        2  Click Yes to exit     3    You may leave the computer or instrume
90. e in the calibration   1   Allophycocyanin  APC         Fluorophore cannot be scanned with current hardware    Settings  Average spot size  um     100    Target signal intensity        90    Sensitivity adjustment method   C Keep laser power fixed  vary PMT gain    Keep PMT gain fixed  vary laser power    cost           Figure 4 16 Automatic Sensitivity Calibration Window    2  Select the fluorophores to include in the calibration  and specify the settings  Refer to  the following table        Item Description   Check the Check each fluorophore to be included  and adjust the calibration area  fluorophores to for each fluorophore  if necessary    include in the   calibration    Adjust the Calibration Select a fluorophore and click this button to adjust the area in the  Area for the Selected window that opens   Fluorophore    Average spot size Enter a value  The default is 100  Valid values are 75  um  to 500  um     um    Target signal intensity Enter a value  The default is 90   Valid values are 50  to 100           Sensitivity Select a method to use     Adjustment Method Keep laser power fixed  vary PMT gain  This is the default     Keep PMT Gain fixed  vary laser power        User Manual pbtsupport perkinelmer com 4 21    Chapter 4 Scanning with Protocols    3  Click Start to run Automatic Sensitivity Calibration  It can be cancelled at any time  by clicking Stop  then Cancel     4  A progress bar displays 100  when complete  Upon completion  review the results   and click 
91. eadsheet autosave protocol     You must select an already defined Quantitation Protocol  Click the  Select a Quantitation Protocol button to select a protocol from the List  of Quantitation Protocols window that opens     You must select an already defined Spreadsheet Autosave Protocol for  saving the quantitation results  Click the Select a Spreadsheet  Autosave Protocol button to select a protocol from the List of  Spreadsheet Autosave Protocols window that opens     Check this box to automatically save the scanned image files to the  specified location  If your system is not integrated with Arraylnformatics   this box is dimmed and unavailable to select        User Manual    pbtsupport perkinelmer com    4 3    Chapter 4 Scanning with Protocols    4 2 1 1 Automatic Quantitation after Scanning    To automatically quantitate the images after scanning    1  Inthe Scan window  select your scanning settings  then check the Perform  automatic quantitation box     2  Select a quantitation protocol and spreadsheet autosave protocol     3  Click Start     The ScanArray Express scans the microarray and saves the images  As the image is  completed  ScanArray Express immediately starts spot finding  using the template called  for in the quantitation protocol  and starts quantitating  When quantitation is finished   ScanArray Express saves the results     4 3 Creating a Scan Protocol  To create a scan protocol    1  Onthe Main Window  click Configure in the Configure  amp  File group  
92. ed data     e Total  uses the intensity of each spot in relation to all spots   See Normalization Methods on page 5 17     4  Click Test Spot Finding to ensure that the template you have set up will find the  spots     5  When you are finished  click Summary to see a summary of the quantitation  protocol     5 4 5 Quantitation Methods    After a template is imported and adjusted in the protocol wizard Template window  the  ScanArray Express determines the center for each spot  and determines the corresponding  patch  The patch is a rectangle that is constructed around the center of the spot with the  dimensions indicated in the template     Both the spot and the background must be defined with the patch  The quantitation method  selected then constructs masks for the spot and the background  A mask is a pixel by pixel  map that indicates the property of each pixel             Background inner  Spot mask           _  Background outer  Spot diameter        Background          Patch          _            5 4 5 1 Adaptive Circle    The Adaptive Circle method fits all spots in the image with circles  the circle diameter is  estimated separately for each spot on the microarray  with a minimum and maximum spot  diameter as specified in the Adaptive Circle Options dialog box  Adaptive Circle is the  best method to use when all circles are not the same size  This is the default method     Technical Support 800 551 2121 or 617 350 9263 ScanArray     Express    Quantitating with Protoco
93. efer to the table below and set the parameters        Item Description   Dimensions   Maximum spot Enter the maximum spot diameter in microns    diameter   Inner background Enter the inner background diameter in microns This must be larger  diameter than the maximum spot diameter    Outer background Enter the outer background diameter in microns  This must be larger  diameter than the inner background diameter     IMPORTANT  The inner and outer background dimensions can never  be equal     pbtsupport perkinelmer com 5 15    Chapter 5    Item    Quantitating with Protocols    Description       Spot Identification  p Value    Calculate Default  Values    The p value parameter controls the confidence for testing whether some  spot pixels differ statistically from the eight median pixels in the  background mask  The p value represents a probability that the  sampled eight spot pixels associated with an adaptive threshold are  from a different population than that of the eight median background  pixels     The smaller the p value  the more confident the derived spot threshold   All pixels above the threshold are used for the spot intensity calculation     Calculates the default values based on the array pattern selected in the  microarray  The defaults for your microarray depend on the template  that is loaded        3  Click OK     5 4 5 4 Histogram    Using the Histogram method  the histogram for each spot is calculated  Use caution with  this method if you don   t have experien
94. em Settings    7 2 ScanArray Express Settings    You can dynamically change most software settings without restarting the instrument   You must have local administrator privileges to change these settings     To set or change application settings  1  Inthe Configure menu  click Application Settings  The Application Settings dialog  displays   Application Settings    Other    Array  Informatics    Connection Scanning   Quantitation    Connection to Instrument      Specify IP address  C Specify computer name       Instrument IP address    10     Computer name     3  Cancel       Figure 7 1 Application Settings Dialog Box    2  Click the tab for the settings you want to set or change  and refer to the appropriate  section below     3  Click OK when you are finished with all settings that you wish to set or change     7 2 1 Connection Settings    On the Connection tab  select either Specify IP address or Specify computer name  The  appropriate field becomes active  where you can enter the IP address or the computer  name  The application must be restarted for the new instrument IP address computer name    to take effect     7 2 2 Scanning Settings    You can allow the user to skip the laser warm up  and set default focus position and laser  power for Easy Scan     7 2 Technical Support 800 551 2121 or 617 350 9263 ScanArray        Express    System Settings    Chapter 7    To set the scanning settings    1  Click the Scanning tab     Application Settings    Connection    Settings
95. enables the Maximum footprint  field below     Uses the ratio of the spot intensity to the standard deviation of the local  background of all spots on the microarray  Choosing this method  enables the Minimum signal to noise ratio field below     Spots with a low signal to noise ratio can be quickly identified for later  review     Uses the comparison of the mean of spot intensity to the mean of local  background  Choosing this method enables the Lower Limit and  Multiplier fields below     The following fields will be enabled  or dimmed and unavailable for  selection  depending on which method you select     3 18 Technical Support 800 551 2121 or 617 350 9263 ScanArray    Express    Using Easy Scan and Easy Quant    User Manual    Item    Chapter 3    Description       Maximum footprint     um     Minimum signal to   noise ratio     Lower limit     Multiplier    Enter the maximum size of the footprint in microns  This measurement  helps to distinguish spots from artifacts on the slide     The default setting is 0  The minimum maximum settings are 100  through 72 000     This field is enabled when you choose the Signal to noise method  The  default value is 3  Valid values are 1 through 10 000     For Signal to background method  Valid values are 0 through 64 000    More accurate spot intensities can be produced with a low intensity   uniform background around all spots     For Signal to background method  The valid values are 1 through  65000        To set the status of a sp
96. equirements and Users Guide for Laser  Safety   IEC 825   Class I Laser Product   B 3 Electromagnetic Emissions Standards     FCC Part 15 Class A  Radiated and Conducted  USA       EN55011 1991 Class A  Radiated and Conducted  Europe       ICES 003  Industry Canada  Interference Causing Equipment Standard  Digital Apparatus     Class A    B 4 Electromagnetic Immunity Standards       EN50082 1  Electromagnetic Compatibility   Generic Immunity Standard  Part 1   Residential  Commercial  and Light Industry      EN61000 4 2  Electrostatic Discharge      EN61000 4 3  Radiated Electromagnetic Fields      EN61000 4 4  Electrical Fast Transient Burst    User Manual PBTsupport packardbioscience com B 1       EN61000 4 5  Surge Immunity Requirements     EN61000 4 6  Conducted Disturbances Induced By Radio Frequency Fields     EN61000 4 11  Voltage Dips  Short Interruptions and Voltage Variations Immunity Tests    B 5 FCC Label for Class A Products    This device complies with Part 15 of the FCC Rules  Operation is subject to the following two conditions      This device may not cause harmful interference       This device must accept any interference received  including interference that may cause  undesired operation     B 6 ICES 003 Label for Class A Products    This Class A digital apparatus meets all requirements of the Canadian Interference Causing Regulations     Cet appareil num  rique de la classe A respecte toutes les exigences du R  glement sur le mat  riel brouiller du Canad
97. er in regard to products or parts furnished by third parties  such being subject to the warranty of their  respective manufacturers  Service under this warranty shall be requested by contacting your nearest PerkinElmer Life  Sciences office     This warranty does not extend to any instruments or parts that have been subject to misuse  neglect or accident  or have  been modified by anyone other than PerkinElmer Life Sciences or have been used in violation of PerkinElmer Life Sciences  instructions  The warranty is null and void if the cover has been removed from the system     The foregoing obligations are in lieu of all other obligations and liabilities including negligence and all warranties  of  merchantability or fitness for a particular purpose or otherwise expressed or implied in fact or in law and state PerkinElmer  Life Sciences entire and exclusive liability and buyer s exclusive remedy for any claims or damages in connection with the  furnishing of goods or parts  their design  suitability for use  installation or operation  PerkinElmer Life Sciences will in no  event be liable for any special  incidental or consequential damages whatsoever and PerkinElmer Life Sciences liability  under no circumstances will exceed the contract price for the goods for which liability is claimed     This warranty shall be governed by  and construed and enforced with the substantive laws of the Commonwealth of  Massachusetts     This warranty may vary outside of the United States  In o
98. eriment      2  Double click a fluorophore to change it  or click one of the following        Add to add a new fluorophore       Change  after selecting a fluorophore  to modify an existing fluorophore       Duplicate  makes a duplicate of the highlighted fluorophore to help you create  a new fluorophore from one that is already close to what you want    e Remove  after selecting a fluorophore  to remove it from the protocol     pbtsupport perkinelmer com    4 9    Chapter 4 Scanning with Protocols    3  When you click Add  the Fluorophore window opens     Fluorophore    Laser power      PMT gain         Fluorophore  5 FAM  Alexa 488  Alexa 532  Alexa 546  Alexa 555  Alexa 568  Alexa 594  Alexa 647  Alexa 660  Allophycocyanin  APC   BODIPY 530 550  BODIPY 558 568  BODIPY 564 570  BODIPY 630 650    Calcein x   gt      Fluorophore cannot be scanned  with current hardware    Cancel         Figure 4 7 Fluorophore Window    4  Select a fluorophore and specify the Laser Power and PMT gain     e Increasing PMT gain improves sensitivity  but also increases background noise  and causes saturation  The optimal signal to noise is obtained using 70 80  for  the PMT Gain       Leave the laser at 90  unless scanning under special conditions     5  Click OK to accept the changes and return to the protocol wizard     6  Click Finish if the PMT Gain and laser settings are known for this scan and will be  kept constant for all scans performed with this Scan Protocol  No further  configuration 
99. escription       Show Hide Spot by Lets you select spots to hide or display  depending on their status   Status    Show Hide Columns Let you select which columns of information to show or hide     Set Quality Criteria You don   t need to set this for the Geometric Test Slide  For more  information  see    Viewing the Quantitation Results    in Chapter 3    Set Spot Status Lets you change the status for a spot  Good  Bad  Found  Not Found   Absent    View Spot Pixels Shows which pixels were treated as part of the spot during quantitation        2 5 2 Scatter Plot    The Scatter Plot allows you to see any column of data plotted against any other column        Composite   Cy3  Control    Cy5  Source    Spreadsheet    Scatter Plot   Distribution Plot   Log   Help      Run  Scan J Cy5 Mean Spot vs  Cy3 Mean Spot    Quantitate      A                Configure  amp  File                5 Os Ls a TTTTT    T A S LL rror oki  Seta 469 at 10000 20000 30000 40000 50000 50000    rioan ray     jene name   Paks Shown Spots   Properties   Cy3 Mean Spot   Zoomin   Zoom Out  Zoom Ful Unknown                               When you click a data point on the scatter plot  the selected data point displays in the spot  viewer and is highlighted in both the Spreadsheet and Image views when you switch back  to those tabs  The data is linked in all views     2 10 Technical Support 800 551 2121 or 617 350 9263 ScanArray     Express    Getting Started  Scan and Quantitate the Geometric Test Slide    2 5 
100. fferent area of the microarray may be used to adjust the  sensitivity of each fluorophore     The area to be used for a single fluorophore is shown as a  darted rectangle  To change the fluorophore  select a  different line from the list of fluorophores  To change the  outlines of the area  use the same directions as were given on  page 2 for changing the scan area     Figure 4 8 Scan Protocol   Sensitivity Calibration Areas Window    User Manual    pbtsupport perkinelmer com 4 11    Chapter 4    Scanning with Protocols    To set the area to be used for automatically calibrating the signal intensity  select a    fluorophore and click Show Zoom Window  The Zoom window allows greater  accuracy when defining an area     Setting as small an area as possible is recommended  In the Zoom window  draw a  rectangle on the slide image using the mouse  or enter the Area Co ordinates     Sensitivity Calibration Area for Fluorophore FITC   Zoom Window  5x   Help  amp  Directions Composite   Cy5  Conton    Cy3  Source      Symbols and Colors  p  The area of to be used for  sensitivity calbration is shown as     a darted rectangle           Zoomin   Zoom Out   Zoom Full    To Re Draw the Sensitivity       Move the mouse to the upper left     comer of the area you would ike  to use  Press the left most mouse  buton down  and hold it down   Then move the mouse to the  lower right comer of the area you     would ike to use  and release the  mouse button     To Modify the Sensitivity  Calibrati
101. for less conventional applications such as tissue arrays and three   dimensional substrates        Item Description   Focus position Do not change the focus position for conventional microarrays   Start Automatic If you have selected a spot on the sample that has measurable  Focus Calibration signals for the selected fluorophore in the protocol  click Start    Automatic Focus Calibration to automatically determine the  optimum lens position     A progress bar displays indicating automatic focus calibration   Upon completion click Accept Changes to accept the  adjustment or click Cancel to cancel the adjustment  If the signal  trace approaches 100   reduce the Laser Power and or PMT  Gain or re run the automatic focus        5  Click Start  review the results  and click Accept Changes to save the results to the  protocol     4 5 3 Automatic Sensitivity Calibration    Only experienced users should use this tool  The Automatic Sensitivity Calibration tool  scans selected areas of a microarray with the selected fluorophores to determine  appropriate settings for laser power and PMT gain     4 20 Technical Support 800 551 2121 or 617 350 9263 ScanArray     Express    Scanning with Protocols Chapter 4    To perform Automatic Sensitivity Calibration    1  Click Run in the Automatic Sensitivity Calibration area of the Scan Protocol    Tools window  The Automatic Sensitivity Calibration dialog box displays     Automatic Sensitivity Calibration ol x     Check the fluorophores to includ
102. g only  The available scan protocols and image autosave protocols display     The following example shows the selection for scan protocols only  the fields for  quantitation are dimmed and unavailable for selection     Protocol Group   Step 1    This step executes  A scan protocol  Select a scan protocol and an image autosave protocol        A scan protocol and a quantitation protocol  Select a scan protocol  an image  autosave protocol  a quantitation protocol  and a spreadsheet autosave protocol     Scan protocol  Image autosave protocol  Quantitation protocol  Spreadsheet autosave protocol     Name Jo Name   Name ooo   Dr  Thomas Testi  Dynamic Repeatability  Geo Slide 32x32 Demo   Geometric  Rotated Uniformity  Static Repeatability     is ot Ls   is      Protocol description  Protocol description  Protocol description  Protocol description   16DyesStd i         Protocols beginning with   cannot be executed with current hardware     C ca j       Figure 5 1 Creating a Step for Scanning    2  Under Scan Protocol  select the protocol you want to include in this step of the  protocol group     3  Under Image Autosave Protocol  select an Image Autosave Protocol to use in this  step     4  Click OK to send your selections to the Protocol Group window     You can add as many steps as you want to the Protocol Group     User Manual pbtsupport perkinelmer com 6 5    Chapter 6 Automating Scans and Quantitation    To select scan and quantitation protocols    1  Inthe Protocol Group   
103. ges  Click Close to close the List of Image Autosave  Protocols window     4 26 Technical Support 800 551 2121 or 617 350 9263 ScanArray     Express    Quantitating with Protocols    Chapter 5    Chapter Summary    Overview 5 1   Running a Quantitation Protocol 5 1   Creating a Quantitation Protocol 5 3   The Quantitation Protocol Wizard 5 4   Creating a Spreadsheet Autosave Protocol 5 18    5 1 Overview    A Quantitation Protocol lets you set up parameters for quantitating  including advanced  quantitation methods  and save them to be used again  A Spreadsheet Autosave Protocol is  used to save the quantitation results    the protocol specifies the folder  file naming  convention and file type to use when saving the results     Quantitation protocols can also be included in protocol groups or batch sets that are run  from the Scan window  See Chapter 6     Automating Scans and Quantitation    for more  information on protocol groups and batch sets  To quantitate automatically after a scan   see    Running a Scan Protocol  Automatic Quantitation    in Chapter 4     This chapter provides instructions for running a Quantitation Protocol and for creating  Quantitation Protocols and Spreadsheet Autosave Protocols        iY    Tip  If you are a new or infrequent user  you can use Easy Quant to become familiar with  the quantitation process before creating and using Quantitation Protocols  See Chapter 3  for information on using Easy Quant     For a tutorial that leads you through 
104. ght  values in millimeters for X and Y  The default scan area is the entire  slide  or 22 mm for width and 73 mm for height  The default start X and  Y positions are 0 00 mm  The minimum width and height for the scan  area are 0 1 mm and 0 05 mm respectively        3  Click Next to display the Scan Protocol   Fluorophores window     4 8 Technical Support 800 551 2121 or 617 350 9263 ScanArray        Express    Scanning with Protocols    4 4 3 Fluorophores    User Manual    Chapter 4    In the Scan Protocol   Fluorophores window  you can select or change fluorophores to be  included in the scan protocol and set a percentage for the PMT Gain and Laser Power   You can also change the order of the listed fluorophores  The first fluorophore listed is the    control fluorophore   To select Fluorophores  1  Open the Scan Protocol   Fluorophores window     Scan Protocol   Fluorophores       PMT    Fluorophore   Gain        Laser Modify    Power     Add    Change  Duplicate  Remove      Move    ToTop    Up    Down  To Bottom     Fluorophores beginning with   cannot be scanned with current hardware     NOTE  By default  the first fluorophore in the list will be used as the control for quantitation     Figure 4 6 Scan Protocol   Fluorophores Window       Finish Cancel    The window displays the current order of scanning and the name  PMT Gain  and Laser  Power for each fluorophore in the list  The first fluorophore scanned is the control image     the second one is the source image  exp
105. he Selected Image                     0 0 5 2 5  Quantitating the Geometric Test Slide       0 0      eee 2 6  Viewing the Quantitation Results  lt 4    4 0c asac Sexe sora an eo AS Gare als SGA 2 9  Spreadsheeta s timor ited eA a RN IG  i ite ig gt Rh aac  dea 2 9  Scatter LOD si Sintra a ea la tek Re get hk a ale Mole meee E Aircel tt De gee 2 10  DistibuUtion  Plot  pistera pronka ded pen chee 4 SOL Pee dees Pens 2 11  Saving the Imagis ssion ek Oar ead ae a ah wate is RAR hand Pre ile  amp  2 12  Saving the Results 2 4  eater dat aa aout Re REA ee Ee eS 2 12  Scan and Quantitate Your Microarrays            0 0 cece een ee 2 12    Chapter 3 Using Easy Scan and Easy Quant    OVETVIEW a Die  Pa urease oria a e og gre while ee held sas nian Bg  4 aae a a a a dew  4 55 Want 3 1  Using Easy Scan acusiud evades keca de Raw hye uae aes bee aoe 3 1  Changing the SetinGS tt ageewius a ie hull aG ewan ad Bod aioe E EA 3 3  Adjusting the Display   seauies  cbt sitaes Senna Wren thE hen a kod gp Wearanes SBag gS 3 4  Changing the Control Image or Source  Experiment  Image              3 6  Saving or Printing Images  ceo cav det we eee d CAV SA Ra awe 3 7  Using Easy Quant susmi wakes aces acd a poe Wh A alae aad ee DG See ahd 3 8  Changing the Sens 4345 Sei ddds peed ae db pad t ead pea eee dag eaad 3 9  Defining PO IE po Be done  5 die caste Slee tor ae g re nly Sods anit dices hey Uae cdl Woo a e aed eh G8 3 10  Adjusting the Template and Registering Images            0 00000005 3 1
106. he front of the display  An enlarged detail display of the image  as shown in Figure 2 1  is  in the center of the window  an image of the Full Slide is on the right side     Image display in Main Window Zoom buttons    ScanArray Express       Run Composite    Cy5  Control    Cy   Source    Log   Diagnostics   Help      Scan    Dragging with mouse   Draws Circle    Quantitate          Remove J          ra   Not Installed  Configure  amp  File       Full Slide    Laver 4   488 nm   N    Full slide with  outline of the  detail shown  in the large  Antal display    oor   Not Installed    Spot viewer       Cassette  Load Status   Not Installed       Palette   Source    Black threshold     4000       z Full color threshold   F  o 537K Loaded  microarray  3x3 Median Restore  Mouse is at 65991m  17037um    Pigel intensity ratio  0 31 S Note                         Figure 2   1 Image Display    User Manual pbtsupport perkinelmer com 2 3    Chapter 2 Getting Started  Scan and Quantitate the    On the left side of the window  the spot viewer shows enlarged details of the spots as the  mouse cursor moves over them     You can change the palette and adjust the display to see the spots better  Adjusting the  display does not change the data for quantitation  nor the data in the file  only the  appearance of the image on the screen  If there are two or more images  you can set the  control image  and if three or more images  you can change the source  experiment   image     To navigate the
107. he image     the Y coordinate in um of the center of the feature   indicator associated with the feature  where  0 0  is  the top left of the image     the diameter in um of the feature indicator     ScanArray  Express    Supported File Formats    Column Title    Appendix    Description       F633 Median    F633 Mean    F633 SD    B633 Median    B633 Mean    B633 SD       gt  B633   1 SD       gt  B633   2 SD    F633   Sat     F543 Median    F543 Mean    F543543 SD    B543 Median    B543 Mean    B543 SD     gt  B543   1 SD     gt  B543   2 SD    F543   Sat     User Manual    median feature pixel intensity at wavelength  1   633 nm      mean feature pixel intensity at wavelength  1  633  nm      the standard deviation of the feature pixel intensity  at wavelength  1  633 nm     the median feature background intensity at  wavelength  1  633 nm      the mean feature background intensity at  wavelength  1  633 nm     the standard deviation of the feature background  intensity at wavelength  1  633 nm      the percentage of feature pixels with intensities  more than one standard deviation above the  background pixel intensity  at wavelength  1  633  nm      the percentage of feature pixels with intensities  more than two standard deviations above the  background pixel intensity  at wavelength  1  633  nm      the percentage of feature pixels at wavelength  1  that are saturated     median feature pixel intensity at wavelength  2   543 nm      mean feature pixel intensity at wavele
108. hnical support    personnel     Service Features    Calibration    Calibrate Attenuator Density    Calibrate Flatness    Configuration Control of Instrument Software    View Instrument Restart the Instrument  Configuration Software and OS    Set Instrument Shut down the Instrument  Network Options Software             Save Calibration  Values    Restore Calibration  Values       Save Instrument Shut down the Instrument  Configuration Software and OS    Restore Instrument  Configuration    Figure 7 6 Service Features Window    7 6 1 Descriptions    Service features are described in the following table        Feature Description  Calibration  Calibrate Used by service personnel to calibrate critical instrument parameters     Calibrate Attenuator  Density    Calibrate Flatness    Save Calibration  Values    Restore Calibration  Values    this is password protected to prevent improper usage     Used exclusively by service personnel to calibrate the instrument  attentuator  Improper usage may cause incorrect operation of the  instrument     Intended for use under the direction of service or technical support  personnel  this feature allows calibration of the X Y stage flatness   Please consult with technical support before running this feature     Creates a file containing all instrument related calibration values  It is  recommended to save this file one time in the event of an instrument  hard drive failure  although the data is stored in the factory when the  instrument is ship
109. icroarray in that slot  See Creating a Batch Set on page 6 8     6 6 Technical Support 800 551 2121 or 617 350 9263 ScanArray    Express    Automating Scans and Quantitation Chapter 6    To run a batch set    1  On the Main Window  click Scan  The Scan window opens  where you can select  Run a batch set  You cannot changes settings from this window  including the scan  area  To make any changes  you need to select a different batch set  or edit the batch  set     Scan Areas  Not Drawn      Scan type   C Run Easy Scan    I Use barcodes for all microarrays  Batch set     Select a Batch Set       To change the scan areas   select a different batch set    Scan areas are not drawn for  batch sets         2  Click Use barcodes for all microarrays to get the protocol information from the  microarray barcodes      OR     Click Select a Batch set  Select a batch set from the window that opens  and click  OK  The window closes and returns you to the Scan window  the button fills with  the name of the batch set     3  Click Start     The ScanArray Express loads the microarray in the first configured slot  and runs the  specified protocol group or protocols  The microarray is scanned and the image s  are  saved to the location defined in the image autosave protocol  If quantitation was included  for a configured slot  ScanArray Express begins spot finding  then quantitates and saves  the results to the location defined in the spreadsheet autosave protocol  The images and  results files a
110. icrons with respect to the origin  which  is the top left corner of the slide        Required Line        Block   Column   Row   Name  Last line of the header  containing column titles for the spot data records      ID     A 2 1 2 Description of the Spot Data Section    User Manual    The spot data section contains lines that describe each spot in detail  It includes the block   column  and row numbers for spots  as well as descriptive name and identifier  information     There is one text line for each spot in the microarray  containing a field for each of the  descriptive columns  Each line contains tab delimited fields in the order described in  Table A 2     Table A 2  Spot Data       Field Description   Block Block number for the spot    Column Column where the spot is located within a block   Row Row where the spot is located within a block   Name Gene name  limited to 40 characters    ID Gene identifier  limited to 40 characters        pbtsupport perkinelmer com Page A 3    Supported File Formats    A 2 2 Blocks  Subarrays     A block is a set of spots printed by one pin within a 4 5 mm x 4 5 mm square or a 9 mm x  9 mm square area  Blocks are also called subarrays  Blocks are numbered from top left to  bottom right  For example     Spot 1 3 Spot 1 4    OO    Spot 2 3 Spot 2 4    QO    Spot 3 3 Spot 3 4    Spot 4 3       A 2 3 Example GAL file    The following simple array list file describes four blocks   BlockCount 4    each with 24  columns and 5 rows  For simplicit
111. ie EEE A RORE eae i  Quality Measurement Formulas                         Glossary    Index    Preface    Contents of This Manual    User Manual    Preface Summary    Contents of This Manual viii  Conventions Used in this Manual ix  Where to Get Help ix   For Additional Information ix   Getting Help for Windows 2000 XP_ x    The following table describes the contents of this User Manual        Chapter Description   Chapter 1 Provides an overview of the ScanArray instrument hardware and  software  and an overview of the ways to scan and quantitate using  ScanArray Express    Chapter 2 Helps you become familiar with ScanArray Express by running an  Easy Scan and Easy Quantitation on the Geometric Test Slide    Chapter 3 Provides instructions for scanning and quantitating using Easy Scan  and Easy Quant    Chapter 4 Provides instructions for scanning using a protocol  includes  instructions for creating a Scan Protocol    Chapter 5 Provides instructions for quantitating using a protocol  includes  instructions for creating a Quantitation Protocol    Chapter 6 Provides instructions for creating and using Protocol Groups and Batch  Sets to analyze multiple experiments and to automate your scanning  and quantitating    Chapter 7 Provides instructions to set system wide settings    Chapter 8 Provides diagnostic and troubleshooting information    Appendix A Formats of files supported by ScanArray Express    Appendix B Specifications   Appendix C Declaration of Conformity   Appendix D
112. iew 6 1   Running and Creating Protocol Groups 6 2  Running and Creating Batch Sets 6 6    6 1 Overview    Automating ties together all of the steps for scanning and analyzing your microarrays   Using Protocol Groups or Batch Sets  you can set up the ScanArray Express to  automatically scan  quantitate  and save your results  without intervention        A    Note  This is in addition to some of the automatic file saving  and automatic quantitation  you can implement when running a single Scan Protocol  as described in Chapter 4        User Manual    Protocol Groups is a powerful new ScanArray Express feature  A protocol group lets you  run a series of scan protocols on a microarray  one after the other  without intervention   and automatically save the results to a specified location  In addition  you can include  quantitation protocols in the protocol group  allowing you to scan  then quantitate  and  save all scanned images and quantitation results to a specified location   again  without  user intervention     Scanning From a batch set lets you run a scan protocol  different protocols  or a protocol  group  on up to 20 slides in a cassette  To use a batch set requires a ScanArray Express  system with the HT autoloader option     This chapter provides instructions for running a Protocol Group or creating a new one  and  for running a scan from a batch set or creating a new Batch Set     pbtsupport perkinelmer com 6 1    Chapter 6 Automating Scans and Quantitation    6 2 Run
113. ifted nominal position be  X Y   the found  position to be  x y   the footprint is    X  x      Y  yy    Signal to Noise Spot intensity standard deviation of Spots with a signal to noise ratio greater than    background intensity     Signal to background Multiplier   M  Lower Limit   L  r4   M x background mean    r2   L   background mean     the minimum specified in the application  settings are good     Spots with a mean greater than r4 anda  mean greater than rz are marked good        User Manual pbtsupport perkinelmer com    Appendix D Formulas    D 2 Technical Support 800 551 2121 or 617 350 9263 ScanArray  Express System    Warranty    PerkinElmer Life Sciences warrants to the customer that the ScanArray Express Microarray Analysis System will be free  from defects in material and workmanship and will meet all performance specifications for a period of one year from the  date of shipment  This warranty covers all parts and labor     In the event that the instrument must be returned to the factory for repair under warranty  the instrument must be packed in  the original packaging  The outside of the package must indicate the Returned Material Authorization Number  RMA   provided by PerkinElmer Life Sciences    PerkinElmer Life Sciences shall not be liable for any incidental  special or consequential damage loss or expense directly  or indirectly arising from the use of the ScanArray Express Microarray Analysis System  PerkinElmer Life Sciences makes  no warranty whatsoev
114. ignal range is drawn as a dark green rectangle       You should select ranges that do not overlap  If the ranges do   overlap  the overlap area is drawn as a dark red rectangle     Histogram Key          Percentile Ranges  Signal range         Background range            Diameters  Inner background diameter  pm     1st    Outer background diameter  um     282          Calculate Default Values Cancel       2  Refer to the following table and specify your settings     Item    Description       Plotted fluorophore    Percentile Ranges    Signal Low   Signal High  Background Low  Background High  Inner Background  Diameter  um  and    Outer Background  Diameter  um     Calculate Default  Values    The highlighted fluorophore will be plotted  Select the fluorophore   Enter the minimum and maximum percentile values for calculations     where percentile is the percentage of pixels in the sample     The default is 5  valid values are from 0 to100   The default is 95  valid values are from 0 to 100   The default is 5  valid values are from 0 to 100     The default is 95  valid values are from 0 to 100     Inner and outer background diameter depends on the array pattern     Calculates the default values based on the array pattern of the  microarray  The defaults depend on the template that has been loaded        3  Click OK     Technical Support 800 551 2121 or 617 350 9263 ScanArray        Express    Quantitating with Protocols Chapter 5    User Manual    5 4 5 3 Adaptive Threshold   
115. iles    description A 14  unable to open 8 4    Adaptive circle quantitation 5 12  Adaptive threshold quantitation 5 15  Additional information  where to find ix  ArrayInformatics   description 1 3  Automatic focus calibration 4 22  Automatic sensitivity calibration 4 20  Automating    automatic quantitation after scanning 4 4  using batch sets 6 1  using protocol groups 6 1    Backing up the database 7 13  Barcodes    configuring 7 9  labels 7 10  supported symbologies 7 9    Batch sets    creating 6 8  description 6 1  running 6 7    Calibrating    automatic focus 4 22  automatic sensitivity 4 20    Contents  of manual viii  Control image  changing 3 6    defined 3 6    Database   backing up 7 13  Declaration of conformity B 1  Diagnostic tools 8 1  Diagnostics   power on 8 1  Distribution Plot 3 20  dynamic range specification C 1    electrical safety C 2   electrical specifications C 2  electromagnetic emissions C 2  electromagnetic immunity C 2  Emission peak  See Fluorophore 7 8  emission wavelengths C 1   EU conformance C 3   Excitation peak  See Fluorophore 7 8  Experiment  See Source image  External laser  troubleshooting 8 3    File formats     CSV results A 9   GAL files A 1   GPR files A 5   GPR Results A 5    Fixed circle quantitation 5 13  Fluorophore    configuring 7 7  defining anew 7 8  emission peak 7 8    Index    excitation peak 7 8  trademarks 1 1    Geometric Test Slide    quantitating 2 6  scanning 2 1   tutorial 2 1    Glossary 1 1    Histogram quantit
116. ion  pm   Start Automatic Focus Calibration  a    Start position  Y  mm   35 00        Line length  mm   22 00 PMT Laser  Fluorophore  Gain      Power         Show Zoom window Cyanine 5 75   100     Clear Previous Plot Lines    Stop Accept Changes Cancel       Figure 4   15 Line Scan Window    2  Set the scan area and describe where the line is        Item Description  Start position  X  Y Enter the coordinates of the area to be scanned   Line length Enter the length of the scan line     Show Zoom Window Click to open the Zoom window where you can select an area  with greater precision     Using the mouse  draw a line over the area of interest on the  slide image  The line should be through the middle of the  selected dots        User Manual pbtsupport perkinelmer com 4 19    Chapter 4 Scanning with Protocols    3  Select the fluorophore  Change the PMT Gain and Laser Power while the scanning  is in progress to find the optimal settings        Item Description   Fluorophore Select the desired Fluorophore    PMT Gain  Laser Change the instrument sensitivity by adjusting the Laser Power  Power     and PMT Gain settings so that the signal is within 90  of    maximum to prevent saturation  The Line Scan changes  dynamically with each adjustment     Clear Previous Plot Click this to clear the display   Lines       4  Focus position rarely needs adjustment when using conventional genomic and  proteomic microarrays printed on flat substrates  Focus position may need to be  adjusted 
117. ion method      Adaptive circle     Adaptive threshold    Set Histogram Options    Test Spot Finding    Normalization method     LOWESS     Total    Help and Directions    All of your changes to the quantitation method  and the method options will be saved as part of  the quantitation protocol                       Figure 5 5 Quantitation Protocol Quantitation and Normalization Window    2  Select a quantitation method     Adaptive circle  the preferred method  uses a minimum and maximum spot    diameter  percentages of the nominal spot diameter  to construct a mask  See  Adaptive Circle on page 5 12     Fixed circle uses the parameters of the spot diameter and background inner and    outer dimensions to create a spot mask and background mask  See Fixed Circle  on page 5 13     Adaptive threshold uses the parameters of the spot diameter and background  inner and outer dimensions to create a spot mask and background mask  then    refines the mask on a pixel by pixel basis  See Adaptive Threshold on page 5   15     Histogram uses the same mask for both the spot and background and creates a    graph of each pixel quantity versus the pixel intensity  See Histogram on  page 5 16     3  Select a normalization method  LOWESS or Total     User Manual pbtsupport perkinelmer com 5 11    Chapter 5    Quantitating with Protocols    e  LOWESS  Locally Weighted Scatter Plot Smoothing  carries out robust  locally weighted scatter plot smoothing for both equally spaced and non   equally spac
118. is required for the protocol     Click Next if the maximum fluorescence values are significantly different from  microarray to microarray  While it will significantly increase individual scan time   ScanArray Express can automatically calibrate each microarray for maximum  fluorescence values with minimal saturation before scanning takes place     44 4 Sensitivity Calibration Areas    In the Scan Protocol   Sensitivity Calibration Areas window  you can select an area of the  microarray to be used to adjust the sensitivity of each fluorophore  In this window  you  select the area  in the next window  Automatic Sensitivity Calibration  you specify the  settings for the calibration     4 10 Technical Support 800 551 2121 or 617 350 9263 ScanArray     Express    Scanning with Protocols    Chapter 4    The larger the area that you specify as the calibration area  the more time it will take to    perform the calibration     To set the Sensitivity Calibration Area    1  Open the Scan Protocol   Sensitivity Calibration Areas window     Scan Protocol   Sensitivity Calibration Areas    Basic Information    3  Fluoropho  Calibration    y Calibration    6  Tools    Sensitivity Area for  Selected Fluorophore     Show Zoom Window          Area   Area     Width x Height   ram     Fluorophore Top Left   ram   0 00  0 00 22 00    22 00 x 73 00      Fluorophores beginning with   cannot be scanned  with current hardware     Copy Current Area to all Fluorophores    Help  amp  Directions       di
119. j    Zl   Ful color threshold  HJ      jsx Loaded          3x3 Median   Auto Adjust                Mouse is at  16823pm  21602um Pivel intensity ratio  1 91       Figure 4   14 Viewing Average Intensity of an Area    9  When the Quick Scan is done  a Quick Scan Results window appears     Quick Scan Results    As part of Quick Scan  you made the following changes   Parameter New   Focus Position  um    Cy3  Laser Power       Cy3  PMT Gain        Accept these changes as part of the scan protocol        10  Click Yes to accept the changes as part of the scan protocol     Technical Support 800 551 2121 or 617 350 9263 ScanArray       Express    Scanning with Protocols Chapter 4    11  To cancel a Quick Scan at any time during the scan  click Stop on the Main Window     4 5 2 Running a Line Scan    Line Scan should be used only by advanced users  The Line Scan repeatedly scans a  single scan line to help determine the laser power and PMT gains for the various  fluorophores  and displays the signal intensity for the specified line scan area on an  oscilloscope like screen  You can adjust the focus position  Y start position  laser power  and or PMT gain while reviewing real time changes in signal intensity     To perform a Line Scan    1  Click Run in the Line Scan area of the Scan Protocol   Tools window  The Line  Scan dialog box displays     Line Scan    Signal vs  X Position    9 10 11 12 13 14 15 16 17 108 19 0   X Position  mm         Start position  X  mm         Focus posit
120. kstation running the ScanArray Express  software     User Manual pbtsupport perkinelmer com 1 1    Chapter 1    Introduction    Figure 1 1 shows the ScanArray instrument models  from left to right   the ScanArray  Lite  the ScanArray Express  and the ScanArray Express HT with 20 slide autoloader           ep Snares       The ScanArray Lite    The ScanArray Express The ScanArray Express HT          1 2    Figure 1 1 The ScanArray Express Instrument Models    All three models run the ScanArray Express software on the workstation  The software is  used to control all aspects of scanning  quantitating and saving the scanned images and  quantitation results  using a network connection between the workstation and the  instrument     1 1 1 1 The Scanning Instrument    The instrument includes one to four lasers  depending on the system configuration   Usually lasers are mounted inside the scanner  however  large lasers or lasers with a  significant power dissipation may be externally mounted  The laser includes appropriate  beam shaping optics  an excitation filter  a shutter and beam alignment optics  Lasers turn  off automatically after 120 minutes of inactivity to maximize the service life of the laser     Optional add ons to the instrument include a barcode reader and autoloader     1 1 1 2 ScanArray Lite    The ScanArray Lite has two internal lasers  and can be upgraded with an optional barcode  reader  This model supports        two Laser Excitation Wavelengths of  543 nm and 6
121. l does not turn on  contact PerkinElmer     8 3 1 3 Microarray is Jammed    If a microarray becomes jammed the ScanArray Express attempts to correct the problem  with a PURGE sequence     To run a manual PURGE sequence    1  The HT system includes sensors that monitor the location of the microarray  If a  microarray becomes jammed  a PURGE sequence is initiated by the software that  locates the jam and places the microarray back into the cassette  An error message  displays on the client computer     2  Ifa PURGE Sequence does not fully eject the microarray into the cassette  a message  displays and you will have to remove the microarray manually by removing the front  cover  removing the cassette and extracting the microarray from the system     User Manual pbtsupport perkinelmer com 8 3    Chapter 8 Diagnostics and Troubleshooting    8 3 1 4 Poor Image Uniformity    Contact Technical Support at 800 551 2121  or 617 350 9263     8 3 2 Software    8 3 2 1 Unable to Open Stored  TIFF files  1  Make sure the stored file names do not include dots     other than the dot preceding  the file extension  e g    tif   csv   gpr   slashes      back slashes      or dashes     in    the file name     2  Make sure there is enough free space on the hard disk to open the file  If there is not   locate any unnecessary image files and delete them to free up space     3  Ifyou are still unable to open the files  contact PerkinElmer     8 4 Recording the ScanArray Express System Activity    A s
122. levator  Homing    Moves the hardware to accurately locate the home sensor and applies  the appropriate calibration     Includes all initialization and homing of Y axis hardware     Moves the hardware to accurately locate the home sensor and applies  the appropriate calibration offset     Downloads operational code to the GIM board     Includes all initialization and homing of X axis hardware     Moves the hardware to accurately locate the home sensor and applies  the appropriate calibration     Downloads operational code to the PAM board     Verifies presence of 1 5 laser modules  and determines type of laser   internal external  manufacturer      Verifies serial communication exists with serial laser  if one is installed   Serial lasers include internal and external and are located in laser  positions 4 and 5  A failure to communicate with internal serial laser 4  indicates a potential hardware problem     Includes all initialization of the optional barcode reader  This includes  establishing serial communication with the barcode reader and  downloading of all configuration parameters     Verifies serial communication exists with the barcode reader     Includes all initialization of the optional the HT autoloader  This includes  programming  downloading configuration parameters and homing of the  HT     Downloads operational code to the HT controller board     Downloads configuration calibration parameters to the controller     Moves the elevator hardware to accurately loca
123. ls Chapter 5    User Manual    To use the adaptive circle method  1  Select Adaptive circle and click Adaptive Circle Options     Adaptive Circle Options    Diameters    Minimum spot diameter    of nominal      Maximum spot diameter    of nominal        Calculate Default Values J    Cancel             2  Enter the minimum and maximum spot diameters  These are not absolute values  but  are a percentage of the nominal spot diameter as specified in the  GAL file or  template      OR     Click Calculate Default Values to calculate the default values based on the array  pattern selected in the microarray  The defaults for your microarray depend on the  template that is loaded     3  Click OK     5 4 5 2 Fixed Circle    The Fixed Circle method fits all spots in the image with circles of fixed diameter  All  spots must be of the same size and shape  Using the Fixed Circle Method  the spot mask  and the background mask are constructed using the parameters of the spot diameter  and  the background inner and outer dimensions     pbtsupport perkinelmer com 5 13    Chapter 5    Quantitating with Protocols    To use the fixed circle method    1  Select Fixed Circle and click Set Fixed Circle Options     Quantitation Options   Fixed Circle Method    Histogram for the Average Spot Patch  Cyanine 5       Frequency    Signal Level       Plotted fluorophore     The plot line shows the histogram for the selected fluorophore   Cyanine 3 The background range is drawn as a black rectangle    The s
124. manufacturer of the fluorophore  Once these values are entered   the ScanArray Express automatically selects the laser and an appropriate filter for the  fluorophore     7  To override the system selection  click Advanced Options and enter the desired  Laser and Filter  then click OK     Advanced Fluorophore Options    Help and Directions    These controls allow you to select which laser and filter will be used  with this fluorophore  overriding the choice of laser and filter which  would normally be made by the software     It is not recommended that you use this approach unless you are  certain that you understand the consequences     J Specify a filter    Filter 4  570nm J          Note  Changing the filter or laser settings is not recommended  as the ScanArray Express  has optimized these selections based on the Excitation Peak and Emission Peak entered        7 8 Technical Support 800 551 2121 or 617 350 9263 ScanArray        Express    System Settings Chapter 7    8  Click the Palette button to display the palette list  Select the desired palette and click  OK        providing a spread of intensities instead of limited shades of just one color  This allows    A Note  Using the rainbow palette for the simple images provides more visual data by  you to differentiate more spots than you can with a single color        9  Click OK to close the Fluorophore dialog box and save the new fluorophore     7 4 Configuring Barcodes    With an optional barcode reader installed  the Sc
125. med     Protocol group  Scan protocol  Image autosave protocol  Quantitation protocol  Spreadsheet autosave protocol     Name Name Name  Dr  Thomas TEST  Dynamic Repeatability i  Geo Slide 32x32 Demo   Geometric   Geometric New 919  Rotated Uniformity i  Static Repeatability    ois His  gt is       Protocol group description  Protocol description  Protocol description  Protocol description  Protocol description   16DyesStd i            tems beginning with   cannot be executed with current hardware     Cancel       Figure 5 4 Cassette Slot Window    The corresponding boxes are enabled for a protocol group or protocols and display a  list of available protocols     3  Ifyou want to save your results automatically in ArrayInformatics  check the box to  save the images  and results if quantitation is performed  in ArrayInformatics  If  your ScanArray Express is not integrated with ArrayInformatics this box will be  dimmed and unavailable to select     4  Select the protocol group  or protocols to execute  and click OK     6 10 Technical Support 800 551 2121 or 617 350 9263 ScanArray    Express    Automating Scans and Quantitation Chapter 6    6 3 3 Configuring a Range of Slots  To configure a range of slots    1  Click Configure a Range of Slots on the Batch Set window  The Range of Cassette  Slots window opens     Range of Cassette Slots    First slot  1 20     IV Set slot status I  Set use of Arraylnformatics J    Remove the quantitation   f 5 rotocol and spreadsheet   Last slo
126. n   Zoom Out   Zoom Ful      Template Properties    Subarray Properties       Spot Properties    Dragging with mouse       Moves or resizes   all subarrays     Rotates all subarrays  C Moves a subarray  C Moves a spot    A  Zoom to Subarray    Register Images  Image to shift or rotate   ior     Offset  pile  0  0  qv   J gt   br      Rotate Image 180            Set Source for  Composite Image             e          Figure 3 4 Adjust Template and Register Images Window  with Zoomed Display     2     Click on a subarray of the template   The example above shows just one subarray    The selected subarray turns blue and the Adjust Template group is activated  You can  move the highlighted subarray template by clicking Move a subarray under the  Dragging with Mouse  options  then dragging the subarray along the substrate     Or     you can click Subarray Properties to enter specific information about the location  of the subarray template and spots in the subarray     Subarray Template Properties    Subarray top left corner  x  um      Subarray top left corner  y  pm      J Straighten spot rows and columns in this subarray    Cancel       Figure 3 5 Subarray Template Properties Dialog Box    pbtsupport perkinelmer com    Chapter 3 Using Easy Scan and Easy Quant    3  To move a spot  click Move a Spot under    Dragging with Mouse     Four blue  triangles appear around the spot  you can drag the spot in any direction to align it  perfectly to the circle of the template        4  To 
127. nEimer i  Run Composite   Cy5  Contra    Cy3  Source    Spreadsheet   Scatter Plot   Distibution Plot   Log   Diagnostics   Help    Scan     General Cy5  Control   Raw Spot Cy5  Control   Mise  ga   FEE Subanay   S  Spot Spot   S AER  Quantitate Inder   Center  usesres   Goon Na    Cora Nare nese Kara Mace es peel  pee re ee ee Std Dev   Bkgnd     Bkgnd         eolrow     colon   eumy um  i  Bkgnd    Bkgnd   PKs  g 1Std   2Std Laser 2  Configure  amp  File 2 LT 18 14458 56  i 3 WW He 200816 2 1445261    4 WW Hs 1257 1869  13897 1496348  Crime W  S WW He 121876 2069  13837 Good 40045 46083 33499 546 S63 1438290 18551 1000 100C    js WW Hs 41714 2269  13837 Good 40223 47598 39631 592 608 1441559 15582 1000 10C Laser 4  7 ii Hs 99873 2469  13837 Good 3910 46528 3883 573 605 13881 11 14225 1000 10C eee  ES    8  Hs 79241 2679  13837 Good 45M0 48071 44813 597 593 1021338 14809 1000 100i ineasat  Spot Viewer 3 v1 Hs 56023 2873  13837 Good 47266 50227 468650 6GIG 625 1044243 18850 1000 100i ro  ie 10 v1 Hs 73853 3069  13837 Good 42223 48646 41646 577 598 1553303 15170 1000 100i Ei    S  Hs1104 3279  13837 Good 49123 50353 48528 60 BIS 8007 25 16011 1000 100 Installed  12 Ww Hs 6101 3473  13837 Good 49828 50235 43230 598 6I6 760820 16448 1000 100i  13 v1 Hs 39948 3679  13837 Good 49703 50478 43089 614 622 7071 20 17348 1000 100i  4 il Hs 87223 3873  13837 Good 49340 5054 48727 613 619 680031 15367 1000 100C  15 11 Hs73105 4079  13837 Good A964 49811 48579 585 598 579756 15403 1000 10
128. ngth  2  543  nm      the standard deviation of the feature intensity at  wavelength  2  543 nm      the median feature background intensity at  wavelength  2  543 nm      the mean feature background intensity at  wavelength  2  543 nm      at wavelength  2  543 nm    at wavelength  2  543 nm      the percentage of feature pixels with intensities  more than two standard deviations above the  background pixel intensity  at wavelength  2  543  nm      the percentage of feature pixels at wavelength  2  that are saturated     pbtsupport perkinelmer com Page A 7    Appendix    Column Title    Supported File Formats    Description       Ratio of Medians    Ratio of Means    Median of Ratios    Mean of Ratios  Ratios SD   Rgn Ratio   Rgn R     F Pixels  B Pixels    Sum of Medians    Sum of Means     Log Ratio  Flags  Normalize   F1 Median   B1  F2 Median   B2  F1 Mean   B1  F2 Mean   B2  SNR 1    F1 Total Intensity    Index     User Defined     the ratio of the median intensities of each feature  for each wavelength  with the median background  subtracted     the ratio of the arithmetic mean intensities of each  feature for each wavelength  with the median  background subtracted     the median of pixel by pixel ratios of pixel  intensities  with the median background subtracted     intensities  with the median background subtracted   the standard deviation of pixel intensity ratios   the regression ratio     the coefficient of determination for the current  regression value     the 
129. ning and Creating Protocol Groups    Once a protocol group is set up  it is easy to select and run  To create a Protocol Group  see  Section 6 2 1     To run a protocol group    1  Onthe Main Window  click Scan  The Scan window opens  where you can select  Run a protocol group     Scan Areals      Scan type     Run Easy Scan        Runa scan protocol    Protocol group     Select a Protocol Group    IT Automatically save in Anaylnformatics       To change the scan area s    select a different protocol  group        2  Click Select a protocol group and select a group from the List of Protocol Groups  window that opens  and click OK  The window closes  and returns you to the Scan  window  the button fills with the protocol group name     3  Click Start     The ScanArray Express loads the microarray  and runs the first protocol in the Protocol  Group  The microarray is scanned and the image s  are saved to the location defined in the  image autosave protocol  If quantitation was included in a step of the Protocol Group   ScanArray Express begins spot finding  then quantitates and saves the results to the  location defined in the spreadsheet autosave protocol  The images and results files are  closed  and ScanArray Express runs the next protocol step     6 2 Technical Support 800 551 2121 or 617 350 9263 ScanArray    Express    Automating Scans and Quantitation Chapter 6    This process continues until all protocols in the group have run  The images and results  from the last prot
130. nt    3 3 1 1 Defining a Template    Specify a template to be used for quantitation  You can import a  GAL file  which also  assigns gene names to each spot on the microarray  or you can manually define the array  pattern by entering values to describe the subarray and spots  For a detailed description of  a  GAL file  see Appendix A     To define a template using a  GAL file    1  In the Quantitation window under Template  click From  GAL File  The Open GAL  File dialog box opens     Open GAL File  Look in     samples    a  Contents  gal    Files of type  GAL    gal    Cancel       2  Highlight the desired file and click Open  The name and location of the selected   GAL file displays in the Quantitate window under the From  GAL File button     3 10 Technical Support 800 551 2121 or 617 350 9263 ScanArray    Express    Using Easy Scan and Easy Quant    Chapter 3    To define a template from specifications    1  In the Quantitation window under Template  click From Specifications  The  Template Specifications window opens     Template Specifications    Subarrays    Number of rows of subarrays     Number of columns of subarrays     Rotation  degrees      Horizontal pin spacing  mm     45    90    Custom   250    Vertical pin spacing  mm     45    390 C Custom  450    I  Straighten rows and columns of subarrays    Spots    Rows of spots per subarray     Columns of spots per subarray     Horizontal spot spacing  center to center  ym   l    Vertical spot spacing  center to center 
131. nt running  after exiting the ScanArray  Express        Note  If you need to shut down the instrument  use the Service Features  as described  below        To shut down the instrument    1  Inthe Main Window  click Configure in the Configure  amp  File group  The Configure  menu displays     2  Click Service Features  then click Shut down the instrument software and OS     3  Wait for the Ready LED on the front of the instrument to turn off  then wait 30  seconds and turn off the power switch on the right side of the instrument     To shut down the computer    1  Your computer should automatically turn off  If not  click the Start menu on the task  bar  then click Shutdown     A message displays on the monitor screen that it is okay to turn off the computer  and  you can turn off the power     Technical Support 800 551 2121 or 617 350 9263 ScanArray TMEX press    Getting Started  Scan and Chapter 2  Quantitate the Geometric Test    Slide    2 1 Overview    Chapter Summary    Overview 2 1   Scanning the Geometric Test Slide 2 1  Viewing the Images 2 3   Quantitating the Geometric Test Slide 2 6  Viewing the Quantitation Results 2 9   Scan and Quantitate Your Microarrays 2 12    This Getting Started chapter leads you through a typical scanning and quantitation session   using Easy Scan  Easy Quant and the Geometric Test Slide that ships with the ScanArray  Express system  The Geometric Test Slide is a test sample made of photo resist printed on  glass     2 2 Scanning the Geome
132. o an electrical signal  The  signal is amplified  filtered  and converted to a digital value for each pixel     The ratio of light energy to current for the PMT output  The higher the Gain the smaller  the amount of light needed for a given amount of current     PBTsupport perkinelmer com    Term    Definition       Protocol    Regedit    Repeatability    Resolution    Saturation    Scanning protocol    Spot    Subnet Mask    Substrate    TCP IP    Technical Support 800 551 2121 or 617 350 9263    A formula that defines an experiment or portion thereof  See Scanning protocol   QuantArray     In Microsoft Windows 95  Windows 98  Windows NT and Windows 2000 and Windows  XP operating systems  the Registry is a single place for keeping information such as   what hardware is attached  what system options have been selected  how computer  memory is set up  and what application programs are to be present when the operating  system is started  Regedit is an editing program used to update installed and uninstalled  application programs  It is also used to update the Registry file     The change in light level detected at a given point on the slide between two scans   Mechanical  electronic and optical changes all affect repeatability     The pixel size that is being scanned for one data point  ScanArray Express can scan as  small as 5 microns     A fluorescence signal that is strong enough to drive the PMT detector to its full scale  maximum voltage and output signal is said to cause 
133. ocol are saved as well  but remain displayed on the screen  You can close  these files  or open any of the other images results that have been saved     6 2 1 Creating a Protocol Group    To create a Protocol Group    1  From the Main Window  select Configure in the Configure  amp  File group  then click  Protocol Groups  The List of Protocol Groups window opens     List of Protocol Groups    Description    Duplicate    Delete    View Usage    Sort By  Name    Description      Protocol group cannot be executed with current hardware       2  Sort the list  optional  by Name or Description   3  Click one of the following to open the Protocol Groups window        Add to create a new protocol group       Change  after selecting a protocol group  to modify an existing group   If you change and save a protocol group that is currently in use  the ScanArray  Express uses the updated protocol group the next time that protocol group is  called  The protocol group currently being run continues with the original  settings       Duplicate  makes a duplicate of the highlighted protocol group to help you  create a new group from one that is already close to what you want     Or  click        Delete  after selecting a protocol group  to delete the group      View Usage to see where the protocol group is being used     User Manual pbtsupport perkinelmer com 6 3    Chapter 6    To create a protocol group    Automating Scans and Quantitation    1  Click Add  The Protocol Group window opens  A p
134. odes    Send email upon  completion of scan    Slot Status    Configure a Range of  Slots    Check this box to read information in the microarray barcode on which  protocols to run     Check this box to read the image autosave protocol information from the  microarray barcodes     Check this box to send an email after scanning has completed running  the batch set  enter an appropriate email address     Set or change the status of each slot by clicking the Status button next  to the slot number  The options are Full  Empty  or Ignored  The default  is Empty     See Configuring a Slot on page 6 10   Click this button to open the Range of Cassette Slots window where you    can set the same parameters for an entire range of slots  See  Configuring a Range of Slots on page 6 11        User Manual    pbtsupport perkinelmer com    6 9    Chapter 6 Automating Scans and Quantitation    6 3 2 Configuring a Slot  To configure a single slot    1  Click the status button for the slot  and change the status to Full  The Cassette Slot   Number  window opens     2  Select the button for what you want to execute for this slot     e a scan protocol   e both a scan protocol and quantitation protocol  or     a protocol group  a scan protocol     Cassette Slot 20  This slot executes     A scan protocol  Select a scan protocol and an image autosave protocol        protocol group  Select only a protocol group     I Save images in Arraylnformatics  also save the spreadsheet if quantitation is perfor
135. on  1 0     Second line of an ATF file     8  number of optional header lines plus 1    5  number of spot data columns      Type of file  the same in all GAL files        Optional Lines        BlockCount 4    BlockType 0      URL          Supplier PerkinElmer Life  Sciences      ArrayerSoftwareName   SpotArray      ArrayerSoftwareVersion 2 0      Blockn      A 2 Technical Support 800 551 2121 or 617 350 9263    Number of blocks described in the file  4 in this example   Type of block described  rectangular in this example   0   rectangular     1   orange packing  1   2   orange packing  2     The URL for the Go To Web command     The manufacturer that supplied the array or arrayer     The name of the arrayer software     The version number of the arrayer software     The position and dimensions of each block  There is one record for each block  and  each record contains 7 fields  Each field is separated by a comma followed by a  space          X position of center of top leftmost spot in the block  in um    2  Y position of center of top leftmost spot in the block  in um    3  Nominal diameter of spots within the block  in um    4    Number of columns of spots in the block     ScanArray  Express    Supported File Formats Appendix    Record    Description    5  Column spacing of spots  center to center  in the block  in um    6  Number of rows of spots in the block   7  Row spacing of spots  center to center  in the block  in um      Note  Positions on arrays are measured in m
136. on Area     Move the mouse over the edge  that you would ike to ch   The mouse pointer will change  shape when you are close  enough  Press the leftmost mouse  button down  and hold it down     release the mouse button        Area Co ordinates  Start position  X  mm   258    Stat position Ymm     13 38  Area with fom    358    Area height  mt 300    Full Microarray   Zoom to Area                        Figure 4 9 Drawing a Rectangle on the Slide Image   Zoom Window    The default scan area is the full substrate  for each fluorophore     A different area of the microarray may be used to adjust the sensitivity of each  fluorophore  or the same area may be used for all fluorophores in the protocol by  clicking Copy Current Area to all Fluorophores        Note  Be sure to set the area for all fluorophores  If you select an area for one  fluorophore  but neglect to set the area for all fluorophores  any fluorophores not set will    scan the whole slide  the default   which could take a long time  possibly hours  for dense  arrays        Technical Support 800 551 2121 or 617 350 9263    Click Next to display the Scan Protocol   Automatic Sensitivity Calibration window     ScanArray TM Express    Scanning with Protocols Chapter 4    4 4 5 Automatic Sensitivity Calibration    In the Scan Protocol   Automatic Sensitivity Calibration window  you can adjust the  scanner so that the brightest features in a defined area of an array will have a chosen signal  level  The area was defined in
137. ot    1  Highlight the    individual spot  and click Set Spot Status  The options are  Found  Not    Found  Absent  Good or Bad     Found  Not Found  Absent  Good   Bad       2  Select the status you want to assign     To view spot pixels    1  Click View Spot Pixels     xs       indicate pixels which were treated as patt of the spot during quantitation     The image shows which pixels were treated as part of the spot during quantitation     pbtsupport perkinelmer com 3 19    Chapter 3 Using Easy Scan and Easy Quant    3 4 2 Viewing the Scatter Plot    The Scatter Plot tab displays a scatter plot of spot brightnesses from both fluorophores   Any data that appears on the spreadsheet can be selected and viewed in the scatter plot   The axis measurements and scales can be changed by clicking on the axis label button                    Run Composite   Cy3  Control    Cy5  Source    Spreadsheet    Scatter Plot   Distribution Plot   Log   Help    so   Cy5 Mean Spot vo  Cy3 Mean Spot pan  Quantitate F  A f iscam  J Laser2  Configure  amp  Fil  Eie  Configure  Spot Viewer    Cy5 Mean Spot                I eryr retir rA errire  20000 30000 40000 50000 60000  Spot Inde 469 a          ie name     PAKS Shown Spots   Properties   Cy3 Mean Spot                 gt     Zoom in   Zoom Out   Zoom Full Unknown          Figure 3 7 Scatter Plot Tab    Data can be filtered and displayed based on status  by clicking the Shown Spots button at  the bottom of the plot  Set axis boundaries by clicking 
138. p gt     Perkin    life sciences     TM    ScanArray Express  Microarray Analysis System  User Manual    User Notice   Users are advised that the use of this ScanArray instrument  or the SpotArray  instrument  for certain applications in analyzing microarrays may require a license  from one or more third parties that have patents in this area  The following are some of  the companies that have patents related to microarrays     Affymetrix  Inc    Incyte Genomics   Oxford Gene Technology  Ltd  Hyseq  Inc     There may be other third parties that have patent rights in this area  PerkinElmer is not  encouraging or suggesting that the users of this equipment employ this equipment in a  way that would infringe on the patent rights of any third parties  PerkinElmer suggests  that users consult their own legal advisors for counsel on non infringing or licensed  uses of this machine in analyzing microarrays     The ScanArray Express is a registered trademark of PerkinElmer Life Sciences  Inc  Windows   Windows 2000  and Windows XP are registered trademarks of Microsoft Corporation  GenePix is a  registered trademark of Axon Instruments  Inc     Alexa 488  Alexa 532  Alexa 546  Alexa 555  Alexa 568  Alexa 594  Alexa 647  Alexa 660   BODIPY 530 550  BODIPY 558 568  BODIPY 564 570  BODIPY 630 650  BODIPY TMR   Calcium Crimson  Calcium Green 1  Calctum Orange  Magnesium Green  Magnesium Orange   Oregon Green 448  Oregon Green 500  Rhodamine Green  Rhodamine Red  Ribo Green  SYBR  Green  
139. pbtsupport perkinelmer com 7 11    Chapter 7 System Settings       Item Description  Scan protocol ID is Select First  lower barcode  or Second  upper  barcode to indicate  printed on whether the scan protocol ID information is on the lower or upper    barcode on the slide     Enter the starting position on the barcode and the length  1 through 13   of the scan protocol ID  Barcodes cannot overlap and there must be no  common characters        3  Click OK to save the barcode configuration     7 5 Database Maintenance    The ScanArray Express database is where your protocols  autosave protocols  and data are  stored  Make a copy of the ScanArray Express database regularly  to ensure that in the  event the database becomes corrupted  a recent backup copy of the database is available to  restore  Database maintenance is also used by the database administrator to backup   restore  and repair the database     Database Maintenance Eq    Restore Database    Compact and Repair  Database    Last backed up     Unknown  Last compacted       Unknown       Figure 7 5 Database Maintenance Window       Button Description  Backup Database Allows you to backup the database   Restore Database Allows you to restore the database     Compact and Repair Allows you to compact and repair the database  This feature   the Database defragments the database in a way that is similar to defragmenting your  hard drive for better space utilization  and lets you use the smallest  amount of space possible for
140. ped     Used exclusively by service personnel  Restoring invalid calibration  values can cause hardware damage to the instrument     7 14 Technical Support 800 551 2121 or 617 350 9263 ScanArray        Express    System Settings    User Manual    Feature    Chapter 7    Description       Configuration    View Instrument  Configuration    Set Instrument  Network Options    Save Instrument  Configuration    Restore Instrument  Configuration    Control of  Instrument Software    Restart the  Instrument Software  and OS    Shut down the  Instrument Software    Shut down the  Instrument Software  and OS    Information pertaining to the instrument hardware  what is installed   version numbers  and software  version numbers      See Figure 7 7     Allows the user to remotely change the network settings on the  instrument     See Setting the Instrument Network Options on page 7 16     Creates a file containing all instrument related configuration information   It is recommended to save this file one time in the event of an  instrument hard drive failure  although all data is stored in the factory  when the instrument is shipped  See Figure 7 7 below     Used exclusively by service personnel  Restoring invalid instrument  configuration information could cause improper operation of the  instrument     This is used to cause the instrument re initialize itself  This should be  used when the instrument is determined to be in an inoperable state and  the user would typically cycle the inst
141. pots to plot from any one of the images  or any images with  the control image     Select a Column to Plot    Category    Column       General  Cy5  Control   Cy3       Ratio of Median Spot   Ratio of Mean Spot   Median of Pixel Ratios   Mean of Pixel Ratios   Regression of Pixel Ratios   M Log2  Ratio of Median Spot   M Log2  Ratio of Mean Spot   A Mean of Log2  Median Spot   4 Mean of Log2  Mean Spot    Ratio of Median Norm Spot   Ratio of Mean Norm Spot   Median of Norm Pixel Ratios   Mean of Norm Pixel Ratios  Regression of Norm Pixel Ratios  M Log2  Ratio of Median Norm Spot   M Log2  Ratio of Mean Norm Spot   4 amp  Mean of Log2  Median Norm Spot   A Mean of Log2  Mean Norm Spot   Std Dev of Pixel Ratios   p  for Regression Ratios    pbtsupport perkinelmer com 3 21    Chapter 3 Using Easy Scan and Easy Quant    3 4 4 Viewing Information About the Images    You can view information about one of the displayed  non composite  images  including  whether or not it has been saved  the filename if saved  and the settings that were used     To view image information  1  Click the tab for which you want to view information     2  On the Main Window  click File in the Configure  amp  File group     Open Image Set from  __Aay Informatics    Save in  Array Informatics       Save All Save Portion of Image      Close   View Header    Close All i Print    Set Control Image       3  On the File menu  click View Header  A Header Information window opens  the  example below is for the cu
142. press  A  GPR file supports  only two fluorophores and does not save all of the data columns  See the description  below      CSV   A text file of quantitation results  A  csv file saves all information that is in the  quantitation spreadsheet results      TIF   the images can be saved as TIF files  as well as bitmap   BMP    JPEG  and RAW  files     A 2 Gene Array List   gal  Format    The gene array list format is written to an ASCII text file that is viewable by applications  such as Microsoft   Notepad and Wordpad  The file contains header information and  information for each spot in the microarray     GAL files conform to the ATF format  a standard tab delimited text file format readable  by many downstream scanning and quantitation software applications  They can be  created in Microsoft Excel by saving an Excel spreadsheet as Text  Tab delimited      User Manual pbtsupport perkinelmer com A 1    Appendix    A 2 1    Supported File Formats    Description of File Format    Gene array list files contain two major sections  the header section and the spot data    section     A 2 1 1    Description of the Header Section    The header section describes basic file information and provides information about each  of the blocks  Each line is explained in Table A 1     Record    Table A 1  GAL File Format    Description       Required Lines       ATF 1 0    85     Type GenePix Array List v1 0     First line of an ATF file  the same in all GAL files   File format  ATF  and versi
143. profile IDs    Image profile ID is printed on   i First  Lov      Second  Upper  barcode    Location of the image profile ID within the first  lower  barcode     Stat at MI 2 3  4 5  6  7 8  9 10 11  12 13    CT    Length  1 2  35 6  7  8 9  10 11 12  13          IV Barcodes contain scan protocol IDs    Scan protocol ID is printed on      First  Lower  barcode    Second  Upper  barcode    Location of the scan protocol ID within the second  upper  barcode     Start at RY 15 16 17  18 19  20  21   22  23 24  25  26    Length           Figure 7 4 Barcode Dialog Box    2  Specify the settings  referring to the following table for information        Item Description   Barcodes contain Check this box to indicate that the barcode includes the identification of  image autosave the image autosave protocol to be used when saving the scanned  protocol IDs image files  Checking this box enables the barcode fields below it   Image autosave Select First  lower  barcode or Second  upper  barcode to indicate    protocol is printed on whether the image autosave protocol information in on the lower or  upper barcode on the slide     Enter the starting position on the barcode and the length  1 through 13   of the image autosave protocol information     Barcodes contain Check this box to indicate that the barcode includes the identification of  scan protocol scan protocol to be used when scanning this microarray  Checking this  information box enables the barcode fields below it     User Manual 
144. protocol    Instrument    Instrument Software    IP Address    Label    Laser    Microarray    NIC   Network Interface  Card    Palette    Photo Bleaching    Pixel    PMI    PMT    PMT  Photomultiplier  Tube  Board    PMT Gain    User Manual    All of the information associated with an image  including all system settings at the time  a scan was started  date and time the scan was performed  instrument SN and revision   instrument software revision  client software revision and user comment field     The information used to determine how images are stored  This includes format and  path     The hardware that comprises the entire Scanner  including the scanner module and the  external laser     This is the internal instrument software necessary to operate the instrument     Network address of a workstation or network  IP Addresses are always in the form of 4  number fields separated by dots  Each number field represents 1 byte Values can be in  the range of 0 255  The numbers on the left of the string define the network  the   numbers on the right define the individual workstation or Network Interface Card  NIC      A thin  flexible  adhesive backed piece of plastic or paper which a barcode or other  identifying information may be recorded     A device that amplifies light  ScanArray Express has 2 5 lasers at different wavelengths  which excite specific fluorophores     A substrate with an array of spots printed on it  the end product of a spotting instrument   The  microarray
145. ption Enter a description  optional    Pattern for File Check the box for each item you want to include in the file names  The   Names information you select will be included in the filename   For example  in the figure above the boxes are checked for Date and  Time  The fields are included in order selected  resulting in a filename  such as   080702_0819 GPR   Custom Text Check this box to enable the edit field and enter custom text  such as  descriptive information  to be included in the filename  You can enter up  to 64 characters    File name pattern Shows the fields selected  their order  and the resulting pattern     Technical Support 800 551 2121 or 617 350 9263 ScanArray     Express    Quantitating with Protocols Chapter 5    Item Description       File Types Select either one or both file types to save the data   GPR   saves the quantitation results as a Gene Results file   CSV   saves the quantitation results as an Excel spreadsheet     Path to output files Enter the desired path for output files or use the Browse button to locate  the path  which can be a local path or a path on the network        7  Click OK to save the spreadsheet autosave protocol     8  Click OK to close the List of Spreadsheet Autosave Protocols window     User Manual pbtsupport perkinelmer com 5 21    Chapter 5    5 22    Technical Support 800 551 2121 or 617 350 9263    Quantitating with Protocols    ScanArray TM Express    Automating Scans and Chapter  Quantitation    Chapter Summary  Overv
146. r Histogram Method    Low Signal    Value    Technical Support 800 551 2121 or 617 350 9263 ScanArray  Express    Supported File Formats Appendix    High Signal    Value   Low Background    value   High Background   value   For Fixed Circle    Low Signal    Value   High Signal    Value   Low Background   value   High Background   value  Nominal Spot Diameter  value  Inner Background Diameter  value  Outer Background Diameter  value  For Adaptive Threshold Method   Maximum Spot Diameter  value  Inner Background Diameter  value  Outer Background Diameter  value  p value  value    END QUANTITATION PARAMETERS  BEGIN QUALITY MEASUREMENT CRITERIA    Selected Criteria  FootPrint SignalToNoiseRatio SignalToBackground   MaximumFootprint value   or   MinSignalToNoise value   Or   LowerLimit  value   Multiplier  value    END QUALITY MEASUREMENT CRITERIA  BEGIN ARRAY PATTERN INFO    Units Microns   Array Rows      Array Columns      Spot Rows 32   Spot Columns 32   Array Row Spacing 2000 00  Array Column Spacing 2000 00  Spot Row Spacing 201 04   Spot Column Spacing 200 52  Spot Diameter  100 00  Interstitial 0 0 is off 1 is first one missing  2 is second one missing  Spots Per Array  1024   Total Spots 1024    END ARRAY PATTERN INFO    User Manual pbtsupport perkinelmer com Page A Il1    Appendix Supported File Formats    BEGIN IMAGE INFO    The first row contains the labels and the following rows are the information   one for each image    Lables    ImageID Channel Image Fluorophor Bar
147. ral scan protocols on a microarray  one after the other  without intervention  and  automatically save the results to a specified location  In addition  you can include  quantitation protocols in the protocol group  allowing you to scan  then quantitate  and  save all scanned images and quantitation results to a specified location    again  without  user intervention  For more information  see Chapter 6     1 6 Technical Support 800 551 2121 or 617 350 9263 ScanArray TM Express    Introduction Chapter 1    Using a Batch Set lets you run different scan protocols or protocol groups on up to 20  slides in a cassette  To use a Batch Set requires a ScanArray Express system with the HT  autoloader option  For more information  see Chapter 6     1 5 The Different Ways to Quantitate    Before starting  review the ways you can quantitate your microarrays     1 5 1 Easy Quant    The Easy Quant settings are       persistent     that is  they remain from the last quantitation  session  So it   s easier to quantitate the same type of microarrays  For the new or occasional  user  Easy Quant is easy to use  all user selectable settings are available from one screen   and you can change your settings as you work     Use Easy Quant if        Your lab uses only one type of two color microarray    e You want to    walk up and use    the equipment  you can change settings as you  work    e You want to test your slide  Easy Quant is a quick and easy way to do it     1 5 2 Quantitation Protocol   
148. ray spots     Technical Support 800 551 2121 or 617 350 9263 ScanArray     Express    Quantitating with Protocols Chapter 5    To define a template    1  Open the Template window        Quantitation Protocol    x     Composite   Cy3  Control    Cy5       Zoom In   Zoom Out  Zoom Full         3  Template  Define a template p   reTma    From From  GAL File   Specifications   _ Informatics    From Anay    GAL file from which to read the  gene list and spot statuses        Modify Template        Subaray Properties  Modify the ecient      Moves or resizes all subarray  Templ  ate Finaida     Moves a subanay    Zoom to Subaray    Help and Directions  All of your changes to the template willbe  saved as part of the quantitation protocol   EXCEPT for adjustments you make to the  locations of individual spots                    Figure 5 4 The Quantitation Protocol  Template Window    2  Specify the source of the template  Refer to the following table for a description of  the possible choices  Instructions for using a  GAL file or specifications are in the  sections following the table   For a description of a  GAL file  see Appendix A      After loading a template  you may need to modify it in Step 3        Item Description  Template Click a button to specify the source of a template   From  GAL file Click this button to open the Open GAL File window  where you can    browse for and select a GAL file to use     Below the button  the  GAL file from which to read the gene list and  spo
149. re closed  and ScanArray Express unloads the microarray and proceeds to the  next configured slot     User Manual pbtsupport perkinelmer com 6 7    Chapter 6 Automating Scans and Quantitation    This process continues until all microarrays in the configured slots in the batch set have  been processed  The images and results from the microarray in the last processed slot are  saved as well  but remain displayed on the screen  You can close these files  or open any  of the other images results that have been saved     After running a batch set  you can check the log to verify that all microarrays were  processed     6 3 1 Creating a Batch Set  To create a batch set    1  On the Main Window  click Configure in the Configure  amp  File group of the main  window  then click Batch Set  The List of Batch Sets window displays     List of Batch Sets  Description Modify    Dr  Thomas Dr  Thomas Add  Dynamic Repeat    Dynamic Repeatability Change  Geo Slide 32x32    Geo Slide 32x32 Demo EA   Geometric Geometric   Rotated Uniformity Rotated Uniformity  Static Repeatabil    Static Repeatability    Duplicate       Description      Batch sets beginning with    contain scan protocols that cannot be executed with  current hardware  If these batch sets are run  the incompatible protocols will be skipped        2  Sort the batch sets  optional  by Name or Description   3  Click one of the following to open the Batch Set window     e Add to create a new batch set       Change  after selecting a 
150. readsheet Autosave Protocols window        Add to create a new protocol       Change  after selecting a protocol  to modify an existing protocol       Duplicate  makes a duplicate of the highlighted protocol to help you create a  new protocol from one that is already close to what you want       Delete  after selecting a protocol  to delete it    e View Usage  after selecting a protocol  to display a list of protocol groups and  batch sets which include the selected protocol     4  Click Add     User Manual pbtsupport perkinelmer com 5 19    Chapter 5    5 20    5     Quantitating with Protocols    The Spreadsheet Autosave Protocol dialog box opens  To create a unique filename   you must include the date and time  if a barcode is available  it should be included in  the filename     Spreadsheet Autosave Protocol    Description     Pattern for File Names    Information to include in the file names   l Barcode JV Time  hours minutes     IV Date  month day year  J  Custom text    c    File name pattern          lt Date gt  lt Time gt     File types   IV GPR  Gene Results   I CSV  Excel spreadsheet     Path to output files   C  Program Files PerkinE Imer S can4rray Express Ima Browse       cancel        Figure 5 7 Spreadsheet Autosave Protocol Dialog Box       6  Fill in the information as listed in the following table    Item Description   Name Enter a name for the spreadsheet autosave protocol  Use your own  name or some other text that makes the protocol easy to find    Descri
151. register the images  use the buttons and arrows in the Register Images box to  move the selected image  in the direction of the arrow you click  until the images are  aligned     Register Images    Image to shift or rotate         Selected Image  Cy5    Offset  pixels   0   1     gt         Rotate Image 180        Set Source for  Composite Image       5  When the template is aligned and the images are registered  click OK  This returns  you to the Quantitate window  page 3 8   where you can make any additional  changes or start your quantitation     3 4 Viewing the Quantitation Results  Quantitation results display in the Main Window  They include        Spreadsheet tab  with information about each spot and subarray on the scanned  microarray  including each gene name  the accession number  subarray location  spot  location  and spot intensity values listed by fluorophore  and normalized spot  intensity values       Scatter Plot tab  lets you view spot intensity two channels at a time    e Distribution Plot  lets you see    geographic    trends for your microarrays   areas of the  slide that are being under or overwashed  for example     3 14 Technical Support 800 551 2121 or 617 350 9263 ScanArray    Express    Using Easy Scan and Easy Quant    Chapter 3    The views in all three tabs are linked    a data point selected in one tab is automatically    selected when you chi    3 4 1 Viewing the Spreadsheet    ange to another tab     An example of the spreadsheet is shown below
152. rotocol group is organized into a  series of steps  each step may include protocols for scanning  or protocols for both    scanning and quantitation     Protocol Group       Description        Scans and quantitations to perform   Each line is a Step   Scan Protocol   Image Profile  in the Protocol Group       Protocol cannot be executed with current hardware    Quantitation Protocol    Spreadsheet Profile       Modify  Add  Change    Duplicate      Remove    2  Specify a Name and Description for the Protocol Group  Use a name that will help    you identify the group later     In this window  you can Add  Change  Duplicate  or Remove a Step from the  group  A step is a selection of protocols to run  each line in this window is a step     3  Click Add to add a Step  The Protocol Group   Step 1 window opens     See To select scan protocols for a group on page 6 5     See To select scan and quantitation protocols on page 6 6     4  Once you have added all of the steps you want to include in the group  you can re   order them if you wish  using the buttons in the Move group     5  When all protocols have been selected  and they are in the order of steps in which  you want the steps to run  click OK to save the protocol group     6 4 Technical Support 800 551 2121 or 617 350 9263    ScanArray TM Express    Automating Scans and Quantitation Chapter 6    To select scan protocols for a group    1  Inthe Protocol Group   Step window  select A scan protocol to create a step for  scannin
153. rray    Horizontal spot spacing  center to center  um      Vertical spot spacing  center to center  um    Spot diameter  um      I Straighten rows and columns of spots in all subarrays    Completely Reset Modify Template CRA  Template Using Changes Only    2  Provide the information for the array pattern  as described in the following table        Item Description       Subarrays Describes the geometry of the microarray     Number of rows and Enter the number of subarray rows and the number of subarray  number of columnsof columns   subarrays    Rotation If the microarray is skewed on the substrate  you can rotate the template  to align the template to the microarray     User Manual pbtsupport perkinelmer com 5 9    Chapter 5    Quantitating with Protocols       Item Description  Pin spacing  Enter the horizontal and vertical pin spacing that was used to create the  horizontal microarray     Pin spacing  vertical    Straighten rows and  columns of subarrays    Spots    Rows of spots and  Columns of spots per  subarray    Horizontal spot  spacing  center to  center  um     Spot diameter    Straighten rows and  columns of spots in  all subarrays    4 5 if 384 well plates were used to print the microarray   9 0 if 96 well plates were used to print the microarray   Custom   to enable the edit boxes where you can enter values for  custom pin spacing    This forces the subarrays into straight rows and columns for  quantitation     The straightening does not display on the image  bu
154. rrayExpress  Microarray Analysis  System 2 0 0 0      Temperature 1 5    LaserPower 1 1    LaserOnTime 5 5    Data record column headings  Data Records    GAL FILES    The barcode symbols read from the image   The origin of the image relative to the scan area   the analysis Blocks  relative to the scan area origin     The version of the ScanArray Express software  used to create the Results file     The temperature of the scanner in volts    The power of each laser in volts    The laser on time for each laser in minutes   Column titles for each measurement  see below    Extracted data     Gene Array List files describe the size and position  of blocks  the layout of feature indicators in them   and the names and identifiers of the printed  substances associated with each feature indicator        A 3 2 GPR Data    Table A 4 below describes each column in the GPR file     Table A 4   GPR Data    Column Title    Description       Block  Column  Row    Name    Dia    Technical Support 800 551 2121 or 617 350 9263    the block number of the feature   the column number of the feature   the row number of the feature     the name of the feature derived from the Array List   up to 40 characters long  contained in quotation  marks      the unique identifier of the feature derived from the  Array List  up to 40 characters long  contained in  quotation marks      the X coordinate in um of the center of the feature   indicator associated with the feature  where  0 0  is  the top left of t
155. rrent image     Header Information for Cy5    File name  Image not saved to file  File type  Image File   Scan date  amp  time  07 06 02 17 30  Microarray barcode  2393 8464 56 S  00743 JCNPH08  Scan protocol name  Easy Scan  Fluorophore  Cys   Laser power      100   PMT gain      75   Focus position  pm   0   Scan resolution  ym   20   Scan speed  Full    Scan area  mm    Top   00 Bottom  73 00 Left   00 Right  22 00  Instrument model  Scan4rray Express  Instrument serial number  Unknown  Client software version  ScanArray Express  Microarray Analysis System 1 0 0 0    Comments        4  Add comments if you wish  and click OK     3 22 Technical Support 800 551 2121 or 617 350 9263 ScanArray     Express    Using Easy Scan and Easy Quant Chapter 3    3 4 5 Saving the Quantitation Results    You can save quantitation results as a  GPR or  CSV file or as an Excel spreadsheet  The   GPR file does not save all of the columns  if you want to save all of the data  be sure to  save it as a  CSV file  See Appendix A for a description of  GPR and  CSV files     To save the results as a GPR or CSV file    1  On the Main Window  click File in the Configure  amp  File group  then click Save or  Save As     2  In the dialog box that opens  enter a file name and select the file type  CSV or GPR        Note  To save the file in ArrayInformatics  click File in the Configure  amp  File group  and  then click Save in ArrayInformatics  If your system is not integrated with  ArrayInformatics  thi
156. rtion of Image     In the Save Portion of Image window that opens  use the mouse to select the portion  of the image that you want to save     Click OK  In the Save Portion of Image dialog box that opens  enter a filename and  click Save     To print an image    1     2     User Manual    Click the tab of the image you want to print     On the Main Window  click File in the Configure  amp  File group  then click Print     pbtsupport perkinelmer com 3 7    Chapter 3    3 3 Using Easy Quant    Using Easy Scan and Easy Quant    The ScanArray Express remembers your settings from the last quantitation session  or you  can change them as you are ready to quantitate  Scan the images to quantitate  or open  previously scanned images  by clicking File in the Configure  amp  File group on the Main  Window  then clicking Open File     To run Easy Quant    1  Inthe Main Window  click Quantitate  The Quantitate window opens  select Run    Easy Quant     Quantitate    Quantitation Template     Adjust Template and  Register Images    Quantitation type     Run a quantitation protocol    Template     From From From Array  GAL File Specifications Informatics    GAL file from which to read the  gene list and spot statuses       Bbc0032 HomeXchange S amples GeoS ample3                Quantitation method  Fields for    Adaptive circle    Adaptive threshold Easy Q ua nt     Fixed circle    Histogram setti ngs    Adaptive Circle Options   Current options match defaults     Normalization method    L
157. rument Network Options  The    following window opens     Network Instrument Settings x     Instrument name     EXPRESS           Obtain IP address from DHCP     Specify IP address  IP address   10  0    Subnet mask   255  0    Default gateway   10  0    Cancel l       Technical Support 800 551 2121 or 617 350 9263    ScanA rray    Express    System Settings    User Manual    Chapter 7    Select Obtain IP address from DHCP  which allows the instrument   s network  settings to be determined dynamically at startup  or Specify IP address  which gives  the instrument a fixed set of network settings     Click OK to send the changes to the instrument  The instrument must be restarted  after the network settings have been changed  The recommended way to restart the  instrument is to use Restart the Instrument Software and OS from the Service  Features menu     pbtsupport perkinelmer com 7 17    Chapter 7 System Settings    7 18 Technical Support 800 551 2121 or 617 350 9263 ScanArray     Express    Diagnostics and  Troubleshooting    8 1 Overview    Chapter    This chapter discusses diagnostic tools  and basic troubleshooting and error recovery  procedures that are available to help identify problems with the instrument     8 2 Diagnostic Tools    ScanArray Express diagnostic tools include Start up Diagnostics  In addition  various  LEDs on the instrument panel display the status of the instrument     8 2 1 Start up Diagnostics    This diagnostic tool is executed automatically each 
158. rument power to cause the  instrument to re initialize     Used exclusively by service personnel  Using this feature prevents the  user from being able to properly operate the instrument     This is highly recommended to be used before shutting down the  instrument  Before turning off the instrument using the power switch   click this button to observe the instrument   s ready LED turn off  wait  approximately 30 seconds and then use the instrument s power switch  to turn off the instrument        pbtsupport perkinelmer com    Chapter 7    Instrument Configuration    Instrument model    Instrument serial number    Instrument server version   Hardware device driver version  2 0 6    TCP IP address    PAM hardware version    PAM software version    GIM hardware version    GIM software version     Hardware Options    Option   Status  Barcode Reader Installed  Arraylnformatics Installed  Autoloader Not Installed  Laserl Installed  Laser2 Installed  Laser3 Installed  Laser4 Installed  Laser5 Not Installed  Filter1  Filter2  Filter3  Filter4  Filter  Filter  Filter   Filters  Filter3  Filter10  Filter11 Not Installed  Filter12 Installed             System Settings    Figure 7 7 The Instrument Configuration Window  from Save Instrument Configuration    7 6 2 Setting the Instrument Network Options    7 16    You can change the instrument name  the name seen by the network  and the network    settings     To change the network settings    1  Onthe Service Features menu  click Set Inst
159. s button is dimmed and unavailable to use        User Manual    pbtsupport perkinelmer com 3 23    Chapter 3    3 24    Technical Support 800 551 2121 or 617 350 9263    Using Easy Scan and Easy Quant    ScanArray ie Express    Scanning with Protocols Chapter 4    Chapter Summary    Overview 4 1   Running a Scan Protocol 4 1   Creating a Scan Protocol 4 4   The Scan Protocol Wizard 4 5   Using Scan Protocol Tools 4 15   Creating an Image Autosave Protocol 4 24    4 1 Overview    A Scan Protocol lets you set up parameters for scanning    including scan speed  scan  area  fluorophores  PMT Gain  laser power  and AutoSensitivity    and save the  parameters to be used again  An Image Autosave Protocol is used to save the scanned  images  it specifies the folder  file naming convention and file type to use when saving the  images     Protocols can also be defined in a Protocol Group  allowing you to use multiple protocols  to scan a single microarray  A Batch Set allows you to run multiple protocols or protocol  groups on   to 20 slides in an autoloader  See Chapter 6     Automating Scans and  Quantitation    for more information on protocol groups and batch sets     This chapter provides instructions for running a Scan Protocol and for creating Scan  Protocols and Image Autosave Protocols        i gt     Tip  If you are a new or infrequent user  you can use Easy Scan to become familiar with  the scanning process before creating and using Scan Protocols  See Chapter 3 for  informa
160. saturation  Lowering the PMT Gain  or Laser Power setting will avoid this     A formula that defines how the scanner hardware and Software execute the scanning  portion of a microarray experiment  A scanning protocol  consisting of a set of  parameters  allows the user to define the parameters one time and then repeatedly  execute the same protocol to scan multiple microarrays in an identical manner     The dried remains of sample fluid printed onto a substrate     Also called sub networks  they are used to partition network addresses for efficiency  and security  Subnet masks work by    masking    some number of the less significant  address bits on all of the workstations in the subnetwork     The piece of glass upon which the array is printed  These take the form of microscope  slides  which have chemically active coatings on the glass that allow the printed spots of  sample to bind to the substrate     Transmission Control Protocol Internet Protocol is the basic communication language or  protocol of the internet  It is also used in private networks called intranets and extranets   TCP manages the disassembling of a message or file into smaller packets that are  transmitted over the internet and received by a TCP layer that reassembles the packets  to the original message  The IP layer handles the address part of each packet so that it  gets to the right destination     ScanArray  Express    Term    Definition       Virtual Memory    XYZ axis    The concept of using hard 
161. scanning and quantitation with a sample slide  see  Chapter 2     Getting Started  Scan and Quantitate the Geometric Test Slide           5 2 Running a Quantitation Protocol    User Manual    Once a protocol is set up  it is very easy to run  After acquiring your image s  by scanning  or opening previously scanned images  using the File menu under Configure  amp  File on the  Main Window   you select a quantitation protocol to use     pbtsupport perkinelmer com 5 1    Chapter 5 Quantitating with Protocols    To quantitate an image using a protocol    1  Click Quantitate on the Main Window  The Quantitate window opens  where you  can choose a quantitation type     Quantitate    Quantitation Template     Quantitation type   C Run Easy Quant    Quantitation protocol   Easy Quant    Adjust Template and    Register Images    Stat  Cancel       Figure 5 1 Selecting to Quantitate Using a Protocol    2  Select Run a quantitation protocol  The settings for performing the quantitation are  included in the quantitation protocol  Unlike when using Easy Quant  you cannot  change the settings in this window     3  Select a protocol  click the Click to Select button and select a protocol from the  Select a Quantitation Protocol window that opens  Click OK     On the Quantitate window  the button fills with the name of the selected protocol     4  Click Adjust the Template and Register Images to adjust the template to match the  array  and register the images if more than one image is displa
162. scent molecule um     20 scan lines second  lt 2 5minutes for a 20 x 30 mm area at 10 microns  16 bit dynamic range over 4 orders of magnitude   Both less than 5  CV   TIFF  BMP  JPEG or Raw    8 user selectable  rainbow  gray  red  orange  yellow  green  blue and  purple    Windows  Pentium III 1 GHz processor  40 GB hard drive  512 MB RAM   19  color monitor  CD RW  1024 x 768 x 65K color  two 10 100 Mbs  Ethernet network interface card  NIC   Windows 2000 SP1  Internet  Explorer 5 0 or later    Lite  30  L x 13  W x 14  H  75 cm x 32 cm x 36 cm   Express  30  L x 16  W x 16  H  76 cm x 41 cm x 41 cm   HT  38  L x 16  W x 16  H  97 cm x 41 cm x 41 cm     29  L x 13  W x 12  H  74 cm x 33 cm x 30 cm     Appendix C    ScanArray  Specifications    Specifications    Description       Weight    Scanner    External Laser  Electrical    ScanArray Instrument    External Laser    additional requirements     Safety Agency  Approvals    Electrical Safety    Laser Safety    Electromagnetic Emissions    Electromagnetic Immunity    FCC Label for Class A  Products    ICES 003 Label for Class A  Products    Technical Support 800 551 2121 or 617 350 9263    Lite  70 Ibs  32 kg   Express  82 lbs  37 kg   HT  108 Ibs  50 kg     55 pounds  25 kg     Autoselecting from 100   240 V  3 5 A  50 60 Hz  Autoselecting from 100   240 V  15 A  50 60 Hz    UL  UL 3101 1 1993 Electrical Equipment for Laboratory Use  USA  amp   Canada   TUV  EN 61010 1 A2 1995 Safety Requirements for Electrical  Equipment
163. t  1 20     20 Slot status  Full C Empty    Ignored I Save in Arraylnformatics database util oes from all  slots in the range    I Set protocol group IV Set scan protocol IV Set image autosave protocol J    Set quantitation protocol J Set sprdsht autosave protocol  Protocol group  Scan protocol  Image autosave protocol  Quantitation protocol  Spreadsheet autosave protocol           Name Name Name Name  Dr  Thomas Dr   Dynamic Repeatability Sin  Geo Slide 32x32 Demo Ger   Geometric Ger  Rotated Uniformity Sin  Static Repeatability 4c  4   ois ot E   is       Protocol group description  Protocol description  Protocol description  Protocol description  Protocol description   16DyesStd      Protocols beginning with   cannot be executed with current hardware      Protocol groups beginning with    contain steps that cannot be executed with current hardware  These steps will be skipped  Cancel            Figure 5 5 Range of Cassette Slots Window    2  Refer to the following table  and set the parameters for the range of slots        Item Description   First Slot Enter the number of the first slot and the number of the last slot to   Last slot include in the range    Set slot state Check this box to set the status for all slots in the range    Set use of Check this box to enable the selection box below it for saving in   Arraylnformatics Arraylnformatics  If your system is not integrated with Arraylnformatics   this box is be dimmed and unavailable to select    Save to Check this box
164. t affects the  quantitation results     This describes the position and size of each spot in a subarray     Enter the number of rows and the number of columns of spots that are  in each subarray     Enter the horizontal and vertical spot spacing  measuring from the  center of a spot to the center of the spot in the next row or column     Enter the spot diameter in um     This forces the spots into alignment for quantitation  The straightening  does not show on the displayed image but affects the quantitation  results        3  Click Completely Reset Template to the previous settings or click Modify  Template Using Changes Only to save your changes     5 4 4 Quantitation and Normalization Method    Specify the quantitation method and normalization method to be used by this protocol   ScanArray Express uses the quantitation method to construct a pixel by pixel map that  indicates the property of each pixel in the image  Normalization corrects the intensity of  each spot for variations in the overall intensity of the images with respect to the control    image     You can use the defaults  Adaptive Circle quantitation method and LOWESS    normalization     Technical Support 800 551 2121 or 617 350 9263    ScanArray TM Express    Quantitating with Protocols Chapter 5    To choose a quantitation method    1  Open the Quantitation and Normalization window     Quantitation Protocol    Composite   Cy3  Contra   cys      1 Se a ce        4  Quantitation and Normalization     Quantitat
165. t statuses field fills with the name of the selected file     See Defining a template Using a GAL File on page 5 8     From specifications Click this button to manually define the array pattern  The Template  Specification window opens  where you enter the values     See To define a template from specifications on page 5 9     User Manual pbtsupport perkinelmer com 5 7    Chapter 5    Item    Quantitating with Protocols    Description       From  Arraylnformatics    Modify Template  Dragging with Mouse    Subarray Properties    Zoom to Subarray    Click this button to select template information from the Arraylnformatics  database  if your ScanArray Express system is integrated with  Arraylnformatics  Refer to the Microarray Laboratory Integration Guide  that ships with the ArrayInformatics software     If your system is not integrated with ArrayInformatics  this button  appears dimmed and is unavailable to select     Select one of the behaviors for the mouse     Moves or resizes all subarrays    Rotates all subarrays      Moves a subarray    Opens a dialog box where you can enter the following parameters  and  then copy them to all subarrays     Subarray top left corner  x  Subarray top left corner  y    Straighten spot rows and columns in this subarray     Click to zoom the display on a subarray        3  Modify the template  using the buttons in the Modify Template group     4  Click Next or click    Quantitation and Normalization Method    to go to the next    window     
166. te the home sensor and  applies the appropriate calibration offset        If an item fails the diagnostic test during Start up  refer to Section 8 3 for troubleshooting    tips     Technical Support 800 551 2121 or 617 350 9263    ScanArray TM Ex press    Diagnostics and Troubleshooting Chapter 8    8 2 2 Instrument Statistics Accumulation  The following statistics are recorded and available for review       Laser On time    Total slides loaded unloaded    8 3 Troubleshooting  8 3 1 Hardware    8 3 1 1 Ready LED Flashes Yellow    The instrument performs a variety of self tests of different modules upon initial power up   If any one of them fails to initialize properly the ready light will flash yellow     1  Ifan external laser is present  ensure the interlock block is installed on the rear panel  of the instrument  If it isn   t  power down the instrument  install the interlock block  and power up the instrument again     2  Ifyou do not get a steady green light after a minute or so  try removing the front  panel and putting it back on again  This will reset the two safety interlock switches     3  Power down the instrument and power it up again  If the Ready LED continues to  flash yellow  contact PerkinElmer Life Sciences     8 3 1 2 External Laser Will Not Turn On    1  Ensure the interlock on the rear panel of the external laser is plugged in  If it isn   t   power down the instrument  install the interlock block and power up the instrument  again     2  Ifthe laser stil
167. teria Spot Status   Spot Pixels    To show or hide spots by their status          1  Click Show Hide Spots by Status   Shown Spot Statuses    Spots are displayed or excluded based on spot status   Check the statuses that should be displayed     ME    V Not found    IV Absent  MV Good  IV Bad    Cancel l       The spot status definitions are as follows        Status Definition   Found The spot was found within that position on the spot template   Not found No spot was found within that position on the spot template   Absent The spot was reported as missing within the  GAL file    Good The spot met the quality criteria    Bad The spot did not meet the quality criteria        2  Check each status for which you want spots to display  uncheck each status for  which you want spots to be hidden     3 16 Technical Support 800 551 2121 or 617 350 9263 ScanArray    Express    Using Easy Scan and Easy Quant    To show or hide columns of information    Chapter 3    1  Click Show Hide Columns  The Show Hide Columns dialog box displays     Show Hide Columns    Categories of columns   Category  Cy3    Cy5  Cy5 and Control                                     Check the columns to show in the spreadsheet     Subarray Location  col row   Spot Location  col row   Gene Name   Gene ID No    Spot Location  x ymy pm     Spot Status   Sum of Median Spot  Sum of Mean Spot  Spot Size  pixels   Bkgnd  pixels   Diameter  ym   Footprint  pixels                                                   gt    Show 
168. the  spots  Leave the template in a position with the upper left hand template circle  near the upper left hand spot in the image     To register the images  click the Composite tab  The images are not perfectly aligned  if traces of individual color such as green and red are visible around the outside of  one side of each spot  Use the Zoom controls as desired to make it easier to view the  registration        Align the images by clicking the left  right  up or down arrows located at the  left side of the window  Each click on an arrow moves the image one pixel in  the direction of the arrow  Move the images until they are well registered  that  is  many spots appear yellow        Click OK   Under Quantitation Method  select Adaptive Circle   Under Normalization Method  select LOWESS     Click Start     The ScanArray Express finds the spots  a progress bar displays  showing the progress of  finding spots  and immediately begins quantitation     2 8 Technical Support 800 551 2121 or 617 350 9263 ScanArray    Express    Getting Started  Scan and Quantitate the Geometric Test Slide Chapter 2    2 5 Viewing the Quantitation Results    The quantitation results display in the Main Window as a Spreadsheet  a Scatter Plot  and  a Distribution Plot  each in a separate tab  The data is linked in all views  when you click a  data point in one view  the data point is already selected if you change to a different view                                       gt  ScanArray Express BBE  Perki
169. the Properties button     3 4 3 Viewing the Distribution Plot    The Distribution Plot is a simple yet powerful data visualization tool  This feature allows  you to select any type of quantitated data displayed in the spreadsheet and map it back to  the spatial arrangement of the microarray  using colors to indicate the value of the data     The Distribution Plot is of particular value for diagnosing or troubleshooting microarray    problems based on the idea that in a well processed array with a large number of spots   there should be no significant spatial correlation of any property     3 20 Technical Support 800 551 2121 or 617 350 9263 ScanArray     Express    Using Easy Scan and Easy Quant    User Manual    Chapter 3    For example  if ratio data were biased toward the green at the top of each subarray and  biased toward red at the bottom of each subarray  that would suggest a problem with  attachment or hybridization  Similarly  if the spots in one subarray had diameters  significantly smaller than the others  that would suggest a problem with a pin  The  Distribution Plot allows you to see all of these phenomena and more with ease        Run    Scan       Quantitate       Configure  amp  File  File          Configure           Spot Viewer  Composite     Median  4337 0  3    Median  41846 0    Spot Index  785  Gene name   ACK                Distibution Plot of Cy3 Ratio of Mean Spot    EE       Zoomin  Zoom Gut Zoom Ful                      You can select a column of s
170. the protocol wizard    5 4 1 Basic Information    In the Quantitate Protocol   Basic Information window  provide a name and description  for the protocol     To name a quantitation protocol    1  Fill in the information as listed in the following table     Item Description       Name Enter a name for the quantitation protocol so you can recall it easily    You must enter a name     Description Enter a description for the quantitation protocol        2  Click Next or click    Image Registration    to go to the next window  If only one  image is open  the next window is    Template        5 4 2 Image Registration    If more than one image is open  Image Registration automatically appears in the wizard  as Step 2  If only one image is open  this step and the associated window do not appear     Registration is done from the Composite tab  by moving the Source image so it registers to  the Control image  That is  the Control image position is fixed  and the Source image is  moved relative to the Control  Each image that needs to be registered has to be set as the  Source image for registration     User Manual pbtsupport perkinelmer com 5 5    Chapter 5 Quantitating with Protocols    To register the images    1  Open the Image Registration window     Quantitation Protocol    Composite   Cy3  Conto    Cy5      2  Zoom in  Zoom Out  Zoom Full        2  Image Registration    Image to shift or rotate  Cy5  Image offset  pixets  0  0    Rotate Image 180    set foyer eae    Help and Dire
171. ther countries  please contact your local PerkinElmer Life  Sciences office for the exact terms of the warranty     User Manual PBTsupport perkinelmer com    Page 4 ScanArray  Express    Glossary    Term    Definition       Accuracy    Administrator    Al    Application Software    Array    Arraylnformatics    HT    Barcode    Batch Set    Bi directional Jitter    BMP    Calibration File    User Manual    Is the maximum positioning error that can occur between the commanded position of a  motion control device and the actual position  Accuracy is a measure of absolute error   For example  if the motion control device is commanded to move to position X1 Y1  Z1   but actually moves to position X2  Y2  Z2  the maximumdifference between  corresponding values is the accuracy     The person in your organization that manages the facility network and or sets up your  Windows NT 2000 workstation     See Array nformatics     Software designed to perform a specific function directly for the user  ScanArray   s  application software communicates with the embedded instrument software     A collection of spots printed on a substrate  arranged in a pattern of regular rows and  columns     A microarray gene expression database and visualization software available from  PerkinElmer  Arraylnformatics can share data with ScanArray Express systems that are  configured for integration with ArrayInformatics     An accessory device in scanner that uses a motor to automatically load substrates into 
172. time the instrument is powered ON  If  the interlocks are open when the software is started  diagnostics are performed after the  interlocks are closed  A system check is performed to ensure that all major functions are  operational  The status can be reviewed by clicking the Diagnostic tab in the Main  Window  Status is indicated using Pass Fail criteria     Start up  Diagnostic Test    Purpose       Attenuator  Initialization    Attenuator Homing    Filter Wheel  Initialization    Filter Wheel Homing    Objective  Z axis   Initialization    Includes all initialization and homing of the attenuator     Moves the hardware to accurately locate the home sensor and applies  the appropriate calibration offset     Includes all initialization and homing of the filter wheel     Moves the hardware to accurately locate the home sensor and applies  the appropriate calibration offset     Includes all initialization and homing of Z axis hardware     User Manual pbtsupport perkinelmer com 8 1    Chapter 8    8 2    Start up  Diagnostic Test    Diagnostics and Troubleshooting    Purpose       Objective Homing  Servo  Y axis   Initialization    Servo Homing    Galvo Interface  Module  GIM  Boot    Galvo  X axis   Initialization    Galvo Homing    PMT Amplifier A D  Module  PAM  Boot    Laser Initialization    Laser  Communication    Barcode Reader  Initialization    Barcode Reader  Communication    Autoloader  Initialization    Autoloader Program  Download    Autoloader Setup    Autoloader E
173. tion on using Easy Scan     For a tutorial that leads you through scanning and quantitation with a sample slide  see  Chapter 2     Getting Started  Scan and Quantitate the Geometric Test Slide           4 2 Running a Scan Protocol    User Manual    A scan protocol is the most efficient and reliable method of scanning microarrays  requiring the same scanning parameters  With a user programmed protocol  each user can  scan microarray slides using the same parameters  thus eliminating user to user variability  from slide to slide     pbtsupport perkinelmer com 4 1    Chapter 4 Scanning with Protocols    To run a Scan Protocol    1  Click Sean on the Main Window  The Scan window opens  where you can choose a  scan type     Scan types on systems with  an HT autoloader         Scan Area   cc    Scan type      Run Easy Scan    Scan type        Run a batch set    Run a protocol group    I    Obtain scan protocol from barcode  Scan protocol   Select a Scan Protocol    Obtain image autosave protocol from barcode  Image autosave protocol     Select an Image Autosave Protocol          Fields for running    Perform automatic quantitation a Scan Protocol    Quantitation protocol     Select a Quantitation Protocol J    Spreadsheet autosave protocol     Select a Spreadsheet Autosave Protocol J       I Automatically save in Arraylnformatics    To change the scan area   select a different scan protocol     Cancel      Figure 4 1 Scan window    2  Select Run a scan protocol     The fields for 
174. tocol Dialog Box       Note  The items are added to the filename in the order that they were checked  as shown  in the File name pattern  If you want to change the order  uncheck all boxes  then recheck  them in the order that you want them to appear in the filename        User Manual    5  Use the following table     Item Description       Name Enter a name for the image autosave protocol  Use your own name  or  some other text that makes the protocol easy to find     Description Enter a brief description   Pattern for File Check the box for each item that you would like to include in the file  Names names  The selections include Fluorophore  Barcode  Date  Time and    Custom text  If you check Custom text  the text field becomes active and  you can enter up to 64 characters of text     To create a unique filename for each image  you must include the  fluorophore  the date  and the time     File types Check each file type that you want to use  TIFF is the preferred and  default format for storing the output image  Image information is not  supplied with  BMP or  JPEG files     pbtsupport perkinelmer com 4 25    Chapter 4 Scanning with Protocols    Item Description       Path to output files Enter the desired path for output files or use the Browse button to locate  the path  which can be a local path or a path on the network     If a directory that doesn   t exist is specified in the pathname  a folder will  be automatically created for it        6  Click OK to save the chan
175. total number of feature pixels   the total number of background pixels     the sum of the median intensities for each  wavelength  with the median background  subtracted     the sum of the arithmetic mean intensities for each  wavelength  with the median background  subtracted     log  base 2  transform of the ratio of the medians   the type of flag associated with a feature     the normalization status of the feature  included not  included      the median feature pixel intensity at wavelength  1  with the median background subtracted     the median feature pixel intensity at wavelength  2  with the median background subtracted     the mean feature pixel intensity at wavelength  1  with the median background subtracted     the mean feature pixel intensity at wavelength  2  with the median background subtracted     the signal to noise ratio at wavelength  1  defined  by  Mean Foreground 1  Mean Background 1      Standard deviation of Background 1     the sum of feature pixel intensities at wavelength   1    the number of the feature as it occurs on the array     user defined feature data read from the GAL file        Technical Support 800 551 2121 or 617 350 9263    ScanArray  Express    Supported File Formats    A 4    A 4 1    Appendix     CSV Results File Format    The  CSV Results file contains sections enclosed by BEGIN SectionName and END  SectionName  Each section contains zero  one  or multiple rows between the BEGIN and  END  Each row contains data items separated by
176. tric Test Slide    User Manual    When finished  your scanned image should match the images shown in this chapter     To run Easy Scan on the Geometric Test Slide    1     2     Turn on the instrument and client workstation   Start the ScanArray Express software   Turn on lasers   and 3  click the status button for each      Insert the Geometric Test Slide into the instrument  If you have an HT instrument  with an autoloader  insert the slide into the cassette  align the slide using the  alignment tool and insert the cassette into the instrument     pbtsupport perkinelmer com 2 1    Chapter 2 Getting Started  Scan and Quantitate the    5  Onthe Main Window  click Scan  The Scan window opens     6  Select Run Easy Scan  The Scan window displays as shown below  with the fields  for Easy Scan settings     Scan types on systems with  an autoloader         Scan Area        Scan type  Scan type   EG 3          Run a scan protocol C Runa batch set     Run a protocol group       Scan resolution  pm      65 810 Sw S0 850       Autoloader slot from which to load  1 20     1          PMT Laser    Fluorophore Gain     Power      E Use Cyanine 3 55 30  M Use Cyanine 5   65 90 Easy Scan  settings    Scan Area Co ordinates    Start position  X  mm   9 75 Area width  mm     7 00    Start position  Y  mm     14 00 Area height  mm    7 50    Set Scan Area to Full Microarray   Show Zoom Window       JE Automatically save image files locally          To change the scan area  drag I Automatically
177. tte chosen above     Forboth simple and composite images    For the composite image only   use the rainbow palette for the simple image    Contrast    0    Cancel       Click the palette you want to use  and select whether you want to use it for the  current image and composite image  or only for the composite image  using the  rainbow palette for all other images   You can adjust the contrast if needed        A    Note  Using the rainbow palette for the simple images provides more visual data by  providing a spread of intensities instead of limited shades of just one color  This allows  you to differentiate more spots than you can with a single color        Click OK     Quantitating the Geometric Test Slide    To quantitate the Geometric Test Slide    Technical Support 800 551 2121 or 617 350 9263    To generate numerical quantitation data from the images  quantitate the Geometric Test  Slide scanned in the previous section using directions in this section     With the two images generated in Section 2 2 still open  click Quantitate on the  Main Window  The Quantitate window opens     ScanArray ae Express    Getting Started  Scan and Quantitate the Geometric Test Slide Chapter 2    2  Select Run Easy Quant  The Quantitate window displays fields for easy  quantitation  Use the settings in the following steps  appropriate for the Geometric  Test Slide     Quantitate    Quantitation Template     Run a quantitation protocol    From From From Array  GAL File Specifications Inform
178. ttings    Connection   Scanning       Quantitation    Array Other  i  2 Informatics    Quality Measurement Settings    Quality measurement method        Footprint   Signalto noise     Signal to background    Quality Measurement Parameters    Maximum footprint  um   100    Minimum signal to noise ratio  3    40    Lower limit     Multiplier        2  Refer to the following table and enter your settings        Item Description   Quality These settings are used for the quantitation methods   Measurement   Settings    Quality measurement  method    Quality Measurement  Parameters    Select a method to for quality measurement  Formulas for methods are  provided in Appendix D     Footprint  Signal to noise  Lower limit and multiplier    These fields are active or inactive  depending on which Quality  measurement method you selected        7 4 Technical Support 800 551 2121 or 617 350 9263 ScanArray        Express    System Settings    Chapter 7    7 2 4 Arraylnformatics Settings    For ArrayInformatics users only  the fields on this tab let you see which ArrayInformatics  database the ScanArray Express is connected to  and lets you specify the pathname for  saving image files        AN    Note  Changing the active database must be done using ArrayInformatics Database  Tools installed on the ScanArray Express client computer  For more information  refer to  the Integration Guide that ships with the ArrayInformatics software        User Manual    To set the Arraylnformatics Settings 
179. uantitation Template    adjusting 3 12  from  GAL file 5 8  from specifications 5 9    Quick scan  running 4 16    Ready LED flashes yellow 8 3    Safety agency approvals C 2  Sample size C 1  Scan resolution    possible selections  selecting    Scan speed  specification C 1  ScanArray Express    exiting 1 9   main window 1 4  turning on 1 8  user interface 1 3    Scanning   different ways to scan 1 6  Scanning field C 1  Scatter plot 2 10  sensitivity specification C 1  Service features 7 14  Size  of instrument C 1  Size  of instruments C 1  Source image    changing 3 6  defined 3 6    Specifications  excitation wavelengths C 1  specifications C 1  Spot template  See Template  Spreadsheet 2 9  3 15  Spreadsheet autosave protocol  creating 5 19  System settings 7 1    Technical Support 1x  Template    adjusting 3 12   defining   from  GAL file 3 10  from specifications 3 11  registering images 3 14    Index    Troubleshooting    external laser 8 3  hardware 8 3   software 8 4   TIFF files 8 4   using diagnostic tools 8 1  yellow Ready LED 8 3    Tutorial    quantitating the Geometric test slide 2 6  scanning the Geometric test slide 2 2  viewing and adjusting the images 2 3  viewing quantitation results 2 9    User interface 1 3    warranty 1 3  wavelengths C 1  weight C 2  Wizards    quantitation protocol 5 4  scan protocol 4 5    workstation specifications C 1    
180. ults automatically  Checking this box enables the  Spreadsheet autosave box  where any configured protocols will display        3  When you are finished configuring the range  click OK to save your changes and  return to the Batch Set window     Technical Support 800 551 2121 or 617 350 9263 ScanArray    Express    System Settings    Chapter    Chapter Summary    Overview 7 1   ScanArray Express Settings 7 2  Configuring Fluorophores 7 7  Configuring Barcodes 7 9  Database Maintenance 7 12  Service Features 7 14    7 1 Overview    User Manual    ScanArray Express settings that are used frequently  or those that affect the operation of  the ScanArray Express system  can be specified and stored using the Configure menu   Settings include those for configuring fluorophores and barcodes  and the system wide  application settings  such as network connection  For systems that are integrated with  ArrayInformatics  the pathnames for automatically saving images and data to an  ArraylInformatics database are specified here     To access the system settings    1  Inthe Main Window  click Configure in the Configure  amp  File group  The Configure  menu displays     Configure    Image l  Autosave    Quantitation    l Spreadsheet    Autosave    Protocols      Quantitation  Protocols    Automation    Protocol    Groups Batch Sets             pareta Advanced settings covered in  n  this chapter  Eluorophores   Barcodes  Database  Maintenance  pbtsupport perkinelmer com 7 1    Chapter 7 Syst
181. w Scan Area         Move the mouse to the upper left corner of the area  y Calibration Areas you would like to scan  Press the left most mouse  button down  and hold it down  Then move the  y Calibration mouse to the lower right corner of the area you would  i like to scan  and release the mouse button     To Modify the Current Scan Area     Move the mouse over the edge that you would like to  change  The mouse pointer will change shape when  you are close enough  Press the left most mouse  button down  and hold it down  Then move the  mouse to where you would like the edge  and release  the mouse button     Show Zoom Window    Finish Cancel       Figure 4   5 Scan Protocol   Scan Area Window    2    Use the mouse to set the area to be scanned by drawing a rectangle of the scan area  on the slide image or click Show Zoom Window  A larger image allows you to  select an area with greater accuracy     Refer to the following table     Item Description       Microarray Diagram Displays the last scanned image  or the resulting image if you ran a  Quick Scan  and lets you use the mouse to select an area to scan     Full Microarray Automatically selects the entire microarray  Use this button for  50 um scans to    find    the array     Show Zoom window Opens a window with a larger diagram  for easier selection of an area  with the mouse  and edit boxes where you can enter numerical scan  area parameters instead     Enter the coordinates for the Start position  Area width and Area Hei
182. xels  The button changes to Remove Filter so you can  return to the previous appearance     3  Black threshold and full color threshold  use these two slide controls together to  adjust the image to minimize background noise and to increase the color for better  visibility     4  Repeat for each tab you want to adjust   To set palette options    You can select a different palette or change the contrast  The palette used in the composite  tab cannot be changed  it is always a blending of the palettes of the control and source   experiment  images or the rainbow palette  If you open a palette when the Composite tab  is selected  the Composite Palette window that opens allows you only to adjust the  contrast     1  Click Palette  The Set Palette Options window opens   Set Palette Options    Name Palette  Red  Orange  Yellow    Blue  Purple  Gray  Rainbow    Use the palette chosen above      For both simple and composite images   amp  For the composite image only   use the rainbow palette for the simple image    Contrast    0     ean       2  Click the color you want to use  and select whether you want to use it for the current  image  or only for the composite image  using the rainbow palette for all other  images   Adjust the contrast if needed        A    Note  Using the rainbow palette for the simple images provides more visual data by  providing a spread of intensities instead of limited shades of just one color  This allows  you to differentiate more spots than you can with
183. y  the data record information  name  ID  etc   is  included only for the first two spots     ATF 1 0   8 5    Type GenePix ArrayList V1 0    BlockCount 4     BlockType 0    URL http   www perkinelmer com microarray    Supplier PerkinElmer Life Sciences    ArrayersoftwareName SpotArray    ArraySoftwareVersion 1 0     Blockl  400  400  100  24  175  5  175    Block2  4896  400  100  24  175  5  175    Block3  400  4896  100  24  175  5  175    Block4  4896  4896  100  24  175  5  175               Block   Column   Row   Name   Tp   1 1 1 VPS8 YALOO2W  1 2 il NTG1 YALO15C       Technical Support 800 551 2121 or 617 350 9263 ScanArray   Express    User Manual    Supported File Formats     GPR Results File Format    Appendix    A  GPR file is a Results file containing general information about image acquisition and  analysis  as well as the data extracted from each individual feature     GPR Header    A sample header file and a description of each entry is included in Table A 3 below     Table A 3   GPR File Format       Entry Description  ATF 1 0 File type and version number   24 43 Number of optional header records and number of     Type PerkinElmer results 2 0    DateTime 2000 02 09 17 15 48    Settings Geo Slide 32 x 32 Demo      GalFile C Program Files PerkinElmer ScanArray  Express Samples GeoSample32x32 gal      Scanner Express Serial No   680065    Comment hyb 2673    PixelSize 10     ImageNAME 633 nm 543 nm      FileNAME C  Program  Files PerkinElmer ScanArray  Express S
184. yed   See Easy  Quantitation in Chapter 3      5  Click Start  ScanArray Express finds the spots  then immediately quantitates     After quantitation  the results display on the Main Window  See Viewing Quantitation  Results in Chapter 3     5 2 Technical Support 800 551 2121 or 617 350 9263 ScanArray     Express    Quantitating with Protocols Chapter 5    5 3 Creating a Quantitation Protocol  To create a quantitation protocol    1  On the Main Window  click Configure in the Configure  amp  File group  In the  Configure menu that opens  click Quantitation Protocols     Configure    Configure  amp  File    Eile  Configure       Spreadsheet l  Autosave  Protocols    Quantitation  Protocols       Automation    Protocol  Groups    Advanced    Application   Service  Settings Features    Eluorophores   Barcodes       Database  Maintenance       2  The List of Quantitation Protocols window opens     List of Quantitation Protocols    Description       Sort By    Name      Description         3  Sort the list  optional  by Name  Description  or Resolution     User Manual pbtsupport perkinelmer com 5 3    Chapter 5    Quantitating with Protocols    4  Click one of the following in the List of Quantitation Protocols window to open the    Quantitation Protocol wizard     e Add to create a new protocol    e Change  after selecting a protocol  to modify an existing protocol  e Duplicate  makes a duplicate of the highlighted protocol  to help you create a  new protocol from one that is alre
185. ys  For the new or occasional user  Easy  Scan is easy to use  all selectable settings are on one screen  and you can change your  settings as you work     Use Easy Scan if        Your lab uses one type of microarray and has one area of interest    e You scan only one or two colors       You want to    walk up and use    the equipment  You can change settings as you  work  including the scan resolution  the flourophores to be scanned  PMT Gain   and the area to scan     1 4 2 Scan Protocol    A Scan Protocol uses pre defined settings that can be named  saved  and recalled to use  again  a protocol is useful when scanning different types of microarrays or when you want  to use advanced features  Using a scan protocol is simple once it is set up  just load your  microarray  select the protocol  and start     Use a Scan Protocol if     e You scan more than two colors  or wish to scan any fluorophore multiple times       Many users share the scanner and you want to automate file naming and saving  for the users       Your lab uses more than one type of microarray  different image fields  labeling  protocols  etc       You wish to incorporate automation features into scanning  such as Auto  Sensitivity      You want to automatically quantitate after scanning     1 4 3 Protocol Groups and Batch Scans    Protocol Groups is a powerful new ScanArray Express feature  Using a protocol group   you can run different experiments on different areas of the same microarrays  You can run  seve
186. ystem log  or event log  keeps a record of all ScanArray Express activity in a log file  In  the file  the ScanArray Express system logs the date and time of the activity  the type and  severity of the event  the current user  as logged onto Windows 2000 XP   and a  description of the event  This log may be helpful in tracking operations or for use by  Service personnel to diagnose any problems     The system log tracks the following user requests to     s start or stop a scan   e add  update  and delete parameters      modify system configuration      start up and shut down the ScanArray Express application    and tracks the following information        changes in the status of client and instrument software communication     software and hardware failures    8 4 1 Viewing the System Log  1  On the ScanArray Express Main Window  click the Log tab     2  Use the scroll bar to move through the display as needed     8 4 Technical Support 800 551 2121 or 617 350 9263 ScanArray     Express    Supported File Formats Appa  A    Summary    Overview F 1  Gene Array List   gal  Format F 1   GPR Results File Format F 5    A 1 Overview    The ScanArray Express supports the following files  which are described in the  subsequent sections      GAL   An array content file is created by the SpotArray Microarray printer after a  printing procedure  This file can be called into ScanArray Express as a template for  quantitating      GPR   A quantitation results file created by the ScanArray Ex
    
Download Pdf Manuals
 
 
    
Related Search
    
Related Contents
初動時成功確定エンジン 『スマートディー』のトリセツ(取扱説明書)。  Mode d`emploi  Manuals  LG LMV2015SB Accessories Catalogue  Garage Door Sensor  IMUSA B120-60006 Use and Care Manual  2 Ton Low Profile Shop Crane  料理集    Razer Orochi    Copyright © All rights reserved. 
   Failed to retrieve file