Home

AMSY-5, AMSY-6 System Verification SysVeri

image

Contents

1. 30 35 40 45 50 55 60 65 70 75 80 85 90 95 100 102 10 104 10 106 107 Amplitude deviation dB vs A dBag max dB above Amin pV allowed eu Energy deviation vs E eu above max allowed Fig 11 Example for amplitude and energy deviation diagrams Test 5 of verification A correction of the should be values for balanced positive and negative deviations here OdB is shown above the diagram No external attenuator is needed and the full dynamic range can be tested without a change of the threshold setting Another difficulty can be shown in the listing of the amplitude varied signals below Usually PCTD is incremented by 2 between two test signals Below there are two occurrences where PCTD stayed constant 140 150 Two unexpected signals of low amplitude about 30dB were triggered by noise from the function generator However those noise signals can be identified by the verification analyzer software and do not disturb the verification result as long as the signal with expected amplitude for each PCTD value can be found in the data file SysVeri1105 page 11 37 Tel 49 8178 9674 400 Copyright 2011 Vallen Systeme GmbH Fax 49 8178 9674 444 Schaftlarner Weg 26a 82057 Icking Munich Germany email info vallen de http www vallen de Vallen SysVeri Software for automated AE System Verification
2. x Each time VeriStimulator starts and finds a 33120A or 33220A function generator you will see the window in Fig 19 This is a reminder that for the first time of using such a function generator three waveforms must be loaded into the function generator s non volatile memory If not sure you should explicitely load the waveforms Click on each of the three blue items Upload on the left side of the VeriStimulator menu see Fig 20 This is only needed once per function generator or when one has overwritten one of the waveforms by another one Fig 19 Reminder that waveforms must be loaded The Martel DC Calibrator Model M2000 can only be controlled over RS232 file SysVeri1105 page 16 37 Tel 49 8178 9674 400 Copyright 2011 Vallen Systeme GmbH email info vallen de Fax 49 8178 9674 444 Schaftlarner Weg 26a 82057 Icking Munich Germany http www vallen de Vallen SysVeri Software for automated AE System Verification systeme VeriStimulator searches for both the function generator and the DC Calibrator on all available interfaces For offline use of SysVeri no function generator and no DC Calibrator is needed 10 SysVeri Step by Step 10 1 Reference Board A reference board is needed for verification of the pulser second step of the System Verification The reference board has to be equipped with special filters ASIP 2 as reference board In case of an ASIP 2 as reference boa
3. Software for automated AE System Verification systeme 4 Basic Procedure for an AE Channel Verification Fig 1 illustrates the basic procedure for a manually performed AE channel verification There must be a predefined list of test patterns to be verified A stimulus usually an AE calibrator must be set to create the predefined test patterns step by step Each pattern is fed into the equipment under verification which converts the test signal to results The operator reads the results records them in a report and checks whether or not they meet the acceptance criteria Then the next test pattern will be processed until the list is completed Finally the channel will be accepted if all results meet the acceptance criteria otherwise it will be rejected and needs repair or at least re adjustment The automated AE channel verification as well as the verification of other parts of the system will basically follow the same procedure Start Report Deviations List of Test Should be Decide Accept Reject patterns End List of Patterns measured Display of Pattern measured Stimulus Equipment under AE Calibrator Verification Fig 1 Basic Procedure for AE Channel Verification file SysVeri1105 page 5 37 Tel 49 8178 9674 400 Copyright 2011 Vallen Systeme GmbH email info vallen de Fax 49 8178 9674 444 Sch ftlarner Weg 26a 82057 Icking Munich Germany http www vallen de Vallen SysVeri So
4. Data Example Id DSET CHAN PCTD A R D E CNTS dB us us eu Ht 3912 1 138 78 0 97 6 196 2 248E03 39 Ht 3931 1 140 30 6 0 2 15 0 882E 3 1 A 30 6dB Ht 3933 1 140 80 2 97 8 196 4 392E03 39 Ht 3969 1 142 82 1 97 8 198 6 622E03 40 Ht 3989 1 144 84 0 97 8 198 8 986E03 40 Ht 4009 1 146 86 3 99 4 200 8 156E04 40 Ht 4029 1 148 88 1 99 8 201 2 247E04 40 Ht 4048 1 150 31 3 0 2 57 4 592E 3 1 A 31 3dB Ht 4067 1 150 90 0 99 8 201 4 391E04 40 Ht 4087 1 152 91 9 100 0 201 8 620E04 40 Ht 4108 1 154 94 2 100 0 202 2 980E04 40 Ht 4146 1 156 96 0 101 4 205 4 155E05 41 Ht 4165 1 158 97 9 101 6 210 8 245E05 42 Ht 4202 1 160 99 8 104 2 212 8 387E05 42 Fig 12 Listing of Test 5 Amplitude variation in 2dB steps Extract Problem False hits A lt 32dB triggered by noise of function generator Such noise is eliminated by VeriAnalyzer because only the last hit of PCTD ID counts Chan 1 Set 167 Index 167 O 0 16 11 455 3001 Time ps Chan 1 Set 1 Index 1 OQ 0 00 00 266 8968 Time ps gt gt E E Q Q ge ge a a E E lt T lt O 5 10 15 20 25 30 35 40 45 110 120 130 140 150 Fig 13 Calibration Pulse Response Fig 14 Continuous sine wave 7 9 Calibration Pulse Response This test signal Fig 13 shows the response comes from the system internal calibration pulse generator and is only used to check whether the calibration p
5. Fig 20 Vallen VeriStimulator Function Generator The commands in the group Function Generator let you control the function generator file SysVeri1105 page 18 37 Tel 49 8178 9674 400 Copyright 2011 Vallen Systeme GmbH email info vallen de Fax 49 8178 9674 444 Schaftlarner Weg 26a 82057 Icking Munich Germany http www vallen de Vallen SysVeri Software for automated AE System Verification systeme Detect Now Starts the hardware detection of the function generator It is the same routine when you start the VeriStimulator Upload HiAmp These four commands let you upload the signals which are used during the verification steps Please upload these signals when you are using the Upload Dirac function generator for the very first time or when the internal non volatile memory of the function Upload LowSin generator has been erased or used to store other signals List arb functions Lists the functions signals that are in the function generator s memory Delete all Deletes the memory of the function generator Self Test Performs a self test Error Messages Lists all the error messages of the function generator Automatic Test This tab is explained in great detail in the following sections The buttons in System Verification and Channel Verification group guide you through a semi automated process of verifying your AE System In the Parameters group you can specify the parameters
6. e PA2 deviation mVavg vs Input mW max PAS deviation mVavg vs Input m max A a PAZ noise mV pp vs Input mY max a PA3 nose mVpp vs Input m max 1034 o 5 5 04 104 10000 5000 0 5000 19000 10000 5000 0 5000 10000 e PA4 deviation mava vs Input mV max PAS deviation mVavg vs Input mV max tA a PA4 noise mVpp vs Input mY max a FAS nose mVpp vs Input mv max 410 T m i mm me 1 oee aE p ammm as ammm aa mm ae a aam e a a a a a a er 402 wi a 10000 5000 0 5000 10000 t0000 5000 0 5000 10000 e PA6 deviation m Vavg vs Input mV max e PA deviation mVavg vs Input mW max FAG nose mVpp vs Input mY max a PA noise mV pp vs Input m max Date 30 09 2010 08 55 verified by User hph initials HPH Sign Think AE think Vallen User friendly reliable up to date Copynght 2009 Vallen Systeme GmbH ISO 9001 certified by TUV CERT ven_SY S doc page of 1 Schaftlamer Weg 26a Tel 49 8178 9674 400 email info vallen de 82057 Icking Munich Germany Fax 49 8176 9674 444 http www vallen de Fig 17 Example of a System Verification Report file SysVeri1105 page 14 37 Tel 49 8178 9674 400 Copyright 2011 Vallen Systeme GmbH email info vallen de Fax 49 8178 9674 444 Schaftlarner Weg 26a 82057 Icking Munich Germany http www vallen de Vallen SysVeri Software for automated AE System Verification After measuring t
7. 1 Set 14 Index 14 O 0 00 09 720 1197 Time ps Freq dB kHz Rectangle Frame 490 2390 OODORARNONRA Amplitude mV BeN S 50 0 50 100 150 0 200 400 600 800 1000 1200 Fig 4 ASIPP response 95 850kHz to Dirac Pulse input Left in time domain right the FFT file SysVeri1105 page 8 37 Tel 49 8178 9674 400 Copyright 2011 Vallen Systeme GmbH email info vallen de Fax 49 8178 9674 444 Schaftlarner Weg 26a 82057 Icking Munich Germany http www vallen de Vallen SysVeri Software for automated AE System Verification systeme 7 2 Rectangular shaped burst Used to vary Duration and to check the parameters Duration Energy Counts Problems Settling behavior with overshoot at the beginning and the ring down behavior at the end see Fig 5 influence the AE parameters Peak Amplitude Risetime Duration Energy Counts Needs some intelligent correction Chan 1 Set 15 Index 15 0O 0 00 10 938 7710 Time us Freq dB kHz Rectangle Frame 173 2611 Amplitude mV ih 0 50 100 150 200 0 200 400 600 800 1000 1200 Fig 5 ASIPP response 95 850kHz to rectangular shaped sine wave burst 7 3 Listing example for the duration variation test Id DSET CHAN PCTD A R D E CNTS Comment dB ps ps eu Ht 633 1 18 81 0 1 2 42 8 885E01 5 Dirac Open Ht 744 1 70 61 4 9 0 225 4 106E02 43 Testptrn
8. 3 Set front panel switch Audio at the connected channel to On at all others to Off During the test you may adjust Sensitivity and Yolume to make hits audible Click Next to continue or Done when completed all channels vallen Systeme GmbH file SysVeri1105 page 26 37 Copyright 2011 Vallen Systeme GmbH Fax 49 8178 9674 444 Schaftlarner Weg 26a 82057 Icking Munich Germany email info vallen de http www vallen de Vallen SysVeri Software for automated AE System Verification The test sequence runs automatically After completion of each channel you are prompted to connect the function generator to the next AE channel Click Done to finish this verification step Tel 49 8178 9674 400 Vallen VeriStimulator Channel erification Tests Fregq Loading function generator done Test 1 amp 2 of 6 Hits amp cascaded hits done Test 3 of 6 Duration variation done Test 4 of 6 Frequency variation done Test 5 of 6 Amplitude variation done Test 6 of 6 Continuous variation vallen Systeme GmbH Vallen VeriStimulator Channel erification Tests Fregq Channel by channel 1 Use DC Bloci for ASIP 2 or AC Input for ASIFP to connect the output of the function generator to the NEXT channel to be verified 2 ASIPP only Set front panel switch Impedance to 50 Ohm 3 Set front panel switch Audio at the connected channel to On at all others to Off
9. 1 Ht 817 1 72 61 4 9 0 449 6 219E02 88 Testptrn 2 Ht 891 1 74 61 4 9 0 919 6 462E02 182 Testptrn 3 Ht 971 1 76 61 4 9 0 1919 2 979E02 382 Testptrn 4 Ht 1053 1 78 61 4 9 0 4024 0 207E03 802 Testptrn 5 Ht 1126 1 80 61 4 9 0 8483 2 437E03 1695 Testptrn 6 Ht 1199 1 82 61 4 9 0 17907 2 925E03 3580 Testptrn 7 Ht 1290 1 84 61 1 9 0 37836 8 195E04 7568 Testptrn 8 Ht 1362 1 86 61 4 9 0 80000 0 413E04 16000 Testptrn 9 Ht 1492 1 2 75 0 1 2 36 4 220E01 4 Dirac Close Fig 6 Listing of a test with duration variations initiated by PCTD 18 and closed by PCTD 2 see signal 3 1 above PCTD parametric counter digitally controlled Each PCTD value identifies a test signal and the corresponding should be results file SysVeri1105 page 9 37 Tel 49 8178 9674 400 Copyright 2011 Vallen Systeme GmbH email info vallen de Fax 49 8178 9674 444 Schaftlarner Weg 26a 82057 Icking Munich Germany http www vallen de Vallen SysVeri Software for automated AE System Verification 7 4 Deviation diagram example for the duration variation test Test 3 The verification report shows a number of Deviation Diagrams which contain graphical and numerical results Test 3 Duration variation Frequency 200KHz Test 3 passed 5 5 5 0 0 0 5 5 5 10 104 104 10 40 102 103 104 Dur deviation vs D us max 0 9 Eny deviation vs E eu max 0 1 Cts deviation vs Cts max 0 0 Off
10. 2 4 par inputs in box 1 2 each in box 2 4 This will cause the following mapping of parametric inputs Box PA channels built in PA mapping in data file A1 Master PAO 3 PAO 3 A2 PAO 1 PA4 5 A3 PAO 1 PA6 7 A4 PAO 1 Not available Ex a number or string that identifies the Reference Board see section 3 1 It is only used to verify the calibration pulse If multiple boxes are used it must be installed in the box that is farthest away from the master box Three of the fields E1 to E4 and three of F1 to F4 must be empty only the field belonging to the last box must contain this information If all channel slots of the last box are occupied by normal boards the board in the last slot must be exchanged against the Reference Board for the system verification part The normal board can be reinstalled before the channel verification part Fx the logical channel number assigned to the Reference Board a value between 1 and 252 The user may remap any hardware address consisting of the channel address and the RAD address to any logical channel number between 1 and 252 In Vallen Acquisition program in step hardware detection select Edit mode to assign channel numbers file SysVeri1105 page 20 37 Tel 49 8178 9674 400 Copyright 2011 Vallen Systeme GmbH email info vallen de Fax 49 8178 9674 444 Schaftlarner Weg 26a 82057 Icking Munich Germany http www vallen de Vallen SysVeri Software for automated AE S
11. During the test you may adjust Sensitivity and Yolume to make hits audible Click Next to continue or Done when completed all channels vallen Systeme GmbH file SysVeri1105 page 27 37 Copyright 2011 Vallen Systeme GmbH email info vallen de Fax 49 8178 9674 444 Schaftlarner Weg 26a 82057 Icking Munich Germany http www vallen de Vallen SysVeri Software for automated AE System Verification 10 6 2 Step 2 2 Filter In order to test the 2 filter stop the Acquisition and enter the AE Channel Settings ASIP 2 only Select filter 25 850kHz ASIPP only Select filter 2 Lo Start acquisition again and make sure that Record Control switches are on enable It is VERY IMPORTANT to use the DC Bloci Otherwise the function generator will be damaged Use the AC Input in case of an ASIPP board For ASIPP boards make sure that Impedance switch is set to 500hm No user action is required during this test Tel 49 8178 9674 400 Vallen VeriStimulator Channel ertfication 2nd Filter Step by step 1 Stop acquisition and enter Acquisition Parameter Setup AE Channel Settings 2 In all channels select filter 2 Lo ASIPP3 or 25 B50kHz2 ASIP 23 3 Resume acquisition and check front panel Record Control Switches for recording enabled Do not click Next before acquisition is properly running vallen Systeme GmbH Vallen VeriStimulator Channel ertfication 2nd Filt
12. If your serial interface is occupied by the DC calibrator and you need such an interface to control the function generator you can unplug the DC calibrator It is not needed for the channel verification procedure Before you start the channel verification create a new vac file To do so stop the acquisition and select a new vac file and store it in the same folder you previously created we recommend file name of type filename asip vac Confirm the creation of a new file For the AE channel settings browse to the folder containing the vac file created during the step Create Veri Files and select it Please also make sure that you selected an input device such as the AEP3 34dB prog or the AEP4 34dB or the VS 150RIC 34dB All these devices have an impedance of 50 Ohm and a gain of 34dB Make sure that in the VeriStimulator gt Automatic Test tab gt Parameters group the Working frequency is set to a value well in between the high and low pass filter 200kHz RMS delay is set to 0 and Counter delay is set to 0 file SysVeri1105 page 25 37 Tel 49 8178 9674 400 Copyright 2011 Vallen Systeme GmbH email info vallen de Fax 49 8178 9674 444 Schaftlarner Weg 26a 82057 Icking Munich Germany http www vallen de Vallen SysVeri Software for automated AE System Verification During the channel verification you can set the Audio switch to the channel that is verified During the data acquis
13. SysVeri Adapter LPT1 the Martel DC Calibrator RS232 and the function generator RS232 or USB or Ethernet If two RS232 devices are to be used and there is only one free RS232 port SysVeri can be used but with very few restrictions LPT1 is only used for control task through ET1 adapter during the acquisition of verification data and can be used afterwards to print the verification reports If a printer is installed on this port it must be disconnected and the driver at least temporarily configured to use another port e g LPT2 before VeriStimulator is started If there are still problems uninstall the printer driver If there is only LPT1 and no LPT2 physically available and the printer shall be used again on LPT1 VeriStimulator must be closed the cable to ET1 adapter be removed the printer cable reinstalled and the printer re configured to work on LPT1 If there is no LPT1 on your PC you might use a USB to Parallel Port adapter The interconnection of the following parts can be seen in Fig 2 Abbr Description ET1 SysVeri Adapter ET 1 CbP Printer cable Sub D 25 male to Sub D 25 male CbE Externals cable 16 pole flat female to SUB D 15 female CbA Alarm cable 10 pole flat female to SUB D 9 male CbD cable to connect the DC calibrator Martel M200x with the RS232 of the control PC CbF cable to connect the function generator with the control PC either RS232 or USB or Ethernet file SysVeri1105 page 3 37 Tel
14. System Verification Click on ASIPP Start to create one page ASIPP Report per channel When processed store the file on filename ASP doc and print it Check and sign it 5 Vallen VeriAnalyzer Configuration file VWwoec TestiSysVer iisysveri youll 04 2005 chg doc Primary Data File for System verification VwodiclTest SysVerisysveriyvo11 04 2005 sys pri Primary Data File for ASIPP verification VWwoec Test SysVer isysveriyvo411 04 2005 asip pri System Settings System Type Heap about Stop System Start ASIPP Start Fig 21 VeriAnalyzer Menu 11 Verification Statistics and Pattern Lists Vallen VeriAnalyzer gathers the results of all verifications in two files that reside in c vallen SysVeri File Sys inc txt contains results of system verifications File ASIPP inc txt contains results of ASIPP verifications These are text files that can easily be imported into an EXCEL sheet and used for e g statistical purposes During the data analysis VeriAnalyzer creates two pattern list files in txt format in the Veri Folder VeriSys sys txt contains the list of patterns checked during system verification and VeriASIPP asp txt contains the list of patterns checked during ASIPP verification file SysVeri1105 page 30 37 Tel 49 8178 9674 400 Copyright 2011 Vallen Systeme GmbH email info vallen de Fax 49 8178 9674 444 Schaftlarner Weg 26a 82057 Icking Munich Germany http www vallen de
15. Verification Step2 PUISED ec ccccccsececceeeeeeeeeeceeeesseeceseeeeeseaeeeseaeesseueeseeessaeeessgeeeesaaees 35 14 4 Channel Verification o oo cece ceceecceceeceececeeeeeeeeceeceeneseeeeeneeaeseeaeeeeeeeneeaesaeceeseeaeseeseeaeeaeseenenneeseeneneeeenees 36 144 1 Aud COMIN Ol tected eyeseciteeecsneuees e ae Senn crescent E eae eid detected a aioe edad agente Na 36 14 4 2 Channel verification Step 3 Pulse through verification ccccccccceeceeeeeeeseeeeeseeeeeseeeeees 37 1 Purpose Today s Acoustic Emission AE Systems are extremely flexible and multi functional A complete verification of all the functions and results in accordance with the manufacturer s specification overcharges the majority of system operators The requirements of today s quality assurance systems to maintain documents of evidence for periodic verification and calibration of measurement equipment causes many operators to purchase expensive AE system verification services from the AE system manufacturer Vallen SysVeri is an approach for a more efficient problem solution for users of AE systems model Vallen AMSY 5 or AMSY 6 For using SysVeri with AMSY 6 please see chapter 14 Appendix Verification of AMSY 6 Please contact sales vallen de if you need SysVeri documentation for AMSY4 Vallen SysVeri is a software package containing two programs VeriStimulator and VeriAnalyzer The software and this documentation guide the operator through an automate
16. channels fill out configuration report group 1 and 2 and name report like Veri01 32 cfg doc Copy this report and name it like Veri33 38 cfg doc Enter in Veri01 32 cfg doc data of channel 1 32 and in Veri33 38 cfg doc data of channel 33 38 In the VeriAnalyzer program you have to choose once Veri01 32 cfg doc and once Veri33 38 cfg doc for generating the ASIPP verification report ASIPP Start button Consequently two ASIPP verification reports will be generated The split of the configuration report has only influence on the data analysis step with VeriAnalyzer Data acquisition for system verification VeriStimulator can be done in one step for all your channels file SysVeri1105 page 21 37 Tel 49 8178 9674 400 Copyright 2011 Vallen Systeme GmbH email info vallen de Fax 49 8178 9674 444 Schaftlarner Weg 26a 82057 Icking Munich Germany http www vallen de Vallen SysVeri Software for automated AE System Verification 10 5 System Verification The System Verification checks the PAx channels the I O flags and the Pulser 10 5 1 Step 1 PAx Flags If a M200x DC calibrator is connected via_ serial interface select option 1 and continue If you have a manually controlled DC calibrator select option 2 and continue Connect all PAx channels in parallel to the DC calibrator Make sure that the PAx ranges are set to 10V Run the Acquisition program Select a new vac file and store it in the same folder yo
17. of an Channel Verification Report Fig 16 shows an example of an automatically generated AE channel verification report It is derived from a word template The yellow fields are place holders and the results were filled in by VeriAnalyzer Vallen Systeme GmbH The Acoustic Emission Company ASIPP Verification Report VeriAnalyzer Rev 2010 0921 Data Format 00 01 ASIPP Template doc Rev 18 Mar 11 Channel hc a ASIPP No ev HW 24 FW Box IdNo 45996 Filter 1 Hi Filter 2 Lo 025 850 TP4 ASP SQ TR 4M Y Energy trie Signal Source Agilent 33220A IdNo 44618 calibrated till Dec 2071 Signal fed in at BNC connector AC Test 1 Cascaded hits 160s apart Hits 1 1 Cascaded Hits 3 3 Test 1 passed Test 2 12 hits 286ps apart Hits 72 12 CascadedHits 12 12 Test 2 passed Test 3 Duration variation Filter 095 Frequency 200 KHz Test 3 passed 107 10 104 Cts deviation vs Cts max 0 0 ca 10 Dur deviation vs D vs Bsa na Eny eviaion eh i vsE eu max above 1 Offset 4 Cts allowed 5 0 Offset 24 2 us Offset 53 2 eu allowed Test 4 Frequency Response _FIt Hi 94 2 300 1 KHz Fit Lo 25 1 836 0 KHz Test 4 passed i eererrre 107 102 Risetime deviation vs R us Amax 1dB max 4 0 Risetime deviation vs R ys Amax 1dB max above 2 0 us R shouldbe 19 f allowed 10 0 above 2 0 ys R shouldbe 19 f allowed Test 5 Amplitude variation sho
18. of the parametric inputs a DC calibrator is controlled over a RS232 serial interface Attention to protect the function generator against possible 28V DC output voltage of the AE channel a DC blocker should be used 6 Software Component Interactions for the automated System and Channel Verification The SysVeri software package consists of the programs VeriStimulator and VeriAnalyzer VeriStimulator guides the operator through the data acquisition step where certain test signals are generated and recorded by the AMSY under test using the normal Vallen acquisition program After the data acquisition step VeriAnalyzer compares the measured data with expected values and generates reports file SysVeri1105 page 6 37 Tel 49 8178 9674 400 Copyright 2011 Vallen Systeme GmbH email info vallen de Fax 49 8178 9674 444 Sch ftlarner Weg 26a 82057 Icking Munich Germany http www vallen de Vallen SysVeri Software for automated AE System Verification systeme VeriStimulator Cane ele SW Generate gt gt Acquisition Acquisition settings SW Control o E PCDT to identify test step Ch Control y Digital 1 O to enable disable aaa SAPAR recording start measurement control pulser i Connect to chan Data Function Generator lt 4 DC blocker y gt Aeau ontro Voltage J y AMSY DC calibrator Connect to
19. parametric inputs gt automatic or manual control Y y Generate PRI VAC File PRI VAC File of system verif of channel verif lt Ld data file data file Template for configuration report edit manually y A VeriAnalyzer SW Compare data with Data expected values gt Analysis System Channel Verification Report Verification Report A The figure above visualizes the software components marked in yellow used for system and channel verification together with their main interfaces They are described in more detail in next sections 6 1 VeriStimulator This program controls the DC calibrator the arbitrary function generator and the parametric counter The DC Calibrator stimulates the analog inputs for external parameters the function generator generates predefined test patterns and the parametric counter identifies each test pattern on the data file for later analysis The program guides the operator to make the needed connections 6 2 Acquisition This is the normal data acquisition program which controls the AE system receives the measured data and stores it on hard disk It doesn t know that it is used in a verification mode It performs as usual This is possible due to the identification of each test pattern by a unique parametric counter value stimulated by VeriStimulator 6 3 VeriAnalyzer This program is called after all pattern stimulation for all parametric inputs and all AE channels have been completed
20. rise time is defined as the time from first threshold crossing to peak amplitude detection and the sine square shaped test signal begins very smoothly noise may contribute to some uncertainty of the exact time of the 1 threshold crossing As Fig 10 can be seen in Fig 9 the peak time is also difficult to determine Due to these inherent uncertainties a deviation of 10 for the rise time is shown as the allowed tolerance 7 8 Data example Amplitude variation amplitude and energy response Test 4 Frequency Response Hi 94 3 m a a a 0 0 0 0 go 5 O Risetime deviation vs R us Amax 1dB max 2 9 above 2 0us R shouldbe 19 f allowed 10 0 Results of frequency variation Frequency response and rise time deviation Amplitude variations are obtained using sine square shaped signals because these do not show the settling behavior as seen with rectangle shaped signals Here one difficulty was the limited dynamic range of the function generator that is only programmable in a 46dB range but we need to check a dynamic range of 70dB 30 100dBae The increase of dynamic range for the function generator was solved by using a second sine squared shape in a lower amplitude range Test 5 Amplitude variation __ should be corrected by 0 0 dB allowed 0 5 to 0 5 dB Test 5 passed
21. wave high amplitude Freq kHz oD Ext RecordDisable Low amplitude 26d0B l offs V 0 00 Upload LoAmn VJ SPO PCTD CIk Dirac pulse signal Dirac width ns App ly Upload Dirac Burst Mode Settings Other commands List ar functions Burst Mode Lizhi ll Cycles Burst Self Test Bursts trigger Error Messages Burst Period s l Trigger Mow Status Fig 22 VeriStimulator in Tab Manual Control file SysVeri1105 page 31 37 Tel 49 8178 9674 400 Copyright 2011 Vallen Systeme GmbH email info vallen de Fax 49 8178 9674 444 Schaftlarner Weg 26a 82057 Icking Munich Germany http www vallen de Vallen SysVeri Software for automated AE System Verification systeme 13 1 Type of function The selection box next to Type of function offers eight alternatives Sine wave Square wave Triangle wave Ramp wave Noise Dirac Sin HiAmp Sin LoAmp The first 5 alternatives are hard coded in the function generator The last three refer to the special waveforms loaded into the non volatile memory of the function generator by VeriStimulator With the function Dirac a sharp and short pulse of 50ns or greater can be produced With function Sin HiAmp the sine square modulated sine wave of Fig 9 can be produced The amplitude range is 50mV to 10Vpp at 5OR Sin LoAmp generates the same function but 26dB less in amplitude than what is entered in the amplitude
22. 49 8178 9674 400 Copyright 2011 Vallen Systeme GmbH email info vallen de Fax 49 8178 9674 444 Schaftlarner Weg 26a 82057 Icking Munich Germany http www vallen de Vallen SysVeri Software for automated AE System Verification systeme 3 Overview System Verification by using Vallen SysVeriis grouped into the following steps AE system verification and m AE channel verification 3 1 System Verification This step comprises the following verifications n Front panel LEDs external in outputs and parametric flags for Alarm Warning SP0 and SP17 Front panel LED external input and parametric flag for Record Disable Measurement accuracy of external parameter inputs PAO to PAx x depends on system configuration and will be in range 1 to 7 Amplitude of calibration pulser under manual and software control 3 2 Channel Verification ASIPP or ASIP 2 This step comprises the following verifications Channel internal timing parameters Duration Discrimination Time Rearm Time The burst parameters amplitude rise time duration counts energy cascaded hits cascaded counts cascaded energy The time driven parameters RMS floating threshold file SysVeri1105 page 4 37 Tel 49 8178 9674 400 Copyright 2011 Vallen Systeme GmbH email info vallen de Fax 49 8178 9674 444 Schaftlarner Weg 26a 82057 Icking Munich Germany http www vallen de Vallen SysVeri
23. It reads the acquired data from hard disk identifies each pattern by the parametric counter data and checks whether or not the measurements meet the acceptance criteria It creates an easy readable system report and one report page per AE channel file SysVeri1105 page 7 37 Tel 49 8178 9674 400 Copyright 2011 Vallen Systeme GmbH email info vallen de Fax 49 8178 9674 444 Schaftlarner Weg 26a 82057 Icking Munich Germany http www vallen de Vallen SysVeri Software for automated AE System Verification 7 AE Test Signals and Result Examples The automated verification process uses the following CRU 500MS s Sample MERE kinds of test signals which are stimulated by the arbitrary 7 oo JCh Coupling function generator under software control T S 7 l E 7 1 Single Dirac Pulse This is a sharp pulse Fig 3 used to trigger a hit for passing administrative information in the parametric counter PCTD to the VeriAnalyzer software Due to the wideband characteristics of a pulse it will pass each frequency filter installed in the AE channel Such pulses are used To indicate the data format version id of VeriStimulator To open or close a certain test series e g the variation of a certain AE parameter See set 633 in Fig 6 PCTD 18 initiates a series of duration varied hits PCTD 2 closes the series PCTD 70 to 86 point into a should be table of durations for this verification step Chan
24. Vallen SysVeri Software for automated AE System Verification 12 Verification of Transient Recorder For the transient recorder verification the Vallen AMSY 5 AMSY 6 software package includes the TR Tester software that can be started from Vallen Control Panel Tab Veri Utilities This program is almost identical to Vallen Acquisition program uses the same routines with the following essential changes Instead of ADC data a sequence of predefined patterns is stored in the transient recorder hardware with each trigger of a channel trigger by threshold crossing as usual Instead of storing the data on file the hardware generated patterns are compared by the software and in case of any deviation an error is reported Acquisition settings are program defined and change with test progress All sampling rates and all record lengths are tested in a predefined sequence The program setup is self explaining and easy to use 13 Vallen VeriStimulator in Manual Mode The VeriStimulator and arbitrary function generator can also be used in a manual mode The operator can vary amplitude frequency and the envelope function of a burst It is also possible to create short pulses down to 50ns width and pulse trains e g N pulses in distance of t us to verify system response in such situations Vallen VeriStimulator Function Generator Ts TT A Automatic Tes Auto Detection i ee Output Function Other Sine modulated sine
25. a serial interface network or USB interface Function Generator Detection Select interfaces Choose your type of interface and W Serial interface standard COM ports AUTO continue with Start iis Network connection to IP address 192 168 224 111 Cancel will skip the hardware detection VISA interface USB Agilent driver required Vallen Systeme GmbH P Help cancel Start 10 3 1 Vallen VeriStimulator The VeriStimulator main window is shown below To the left hand side you find the group Function Generator In the center of the window you find two tabs Automatic Tests and Manual Control During the automated test routine you will only need the Automatic Tests tab Vallen VeriStimulator R2011 0303 Function Generator a ee Soe Auto Detection anual Contro Detect Now system Verification Channel Verification Parameters Arbitrary Functions Working Freg kHz 200 0 a ee a ee Step 1 PAx Flags Step 1 Tests Freq RMS Delay ms 0 wave high amplitude mae Upload HiAmp Counter Delay ms 0 Low amplitude 26dB step Pulser step 2nd Filter Upload LoAmp Dirac pulse signal Upload Dirac J curcan DGGE step 3 Pulse through Upload LowsSin Other commands List arb functions Delete all Self Test Le Create Ver fles Error Messages Status Pune Gen HEWLETT PACKARD 33120A 0 7 0 5 0 1 Cdeiscied on COMI Help About
26. amplitude settling time left 17 secondts vallen Systeme GmbH Next the minimum Amplitude of Vallen Veristimulator the pulser is verified ro System Verification Pulser test Turn the front panel knob Amp all the way to the left and continue Step by step 6 Turn front panel knob Amp to minimum deft end Click Next to continue vallen Systeme GmbH file SysVeri1105 page 24 37 Tel 49 8178 9674 400 Copyright 2011 Vallen Systeme GmbH email info vallen de Fax 49 8178 9674 444 Schaftlarner Weg 26a 82057 Icking Munich Germany http www vallen de Vallen SysVeri Software for automated AE System Verification systeme In the last step of the pulser Vallen VeriStimulator verification the amplitude settings are controlled by software System Yerification Pulser test Set the Amp switch to remote Rem and continue Step by step 7 Set switch Amp to Rem up Click Next to continue vallen Systeme GmbH Stop the pulsing in the Acquisition Vallen VeriStimulator program after completion of the pulser verification System Verification Pulser test Pulser Verification complete Stop Pulsing in Acquisition vallen Systeme GmbH fo Cancel 10 6 Channel Verification If you installed a reference board for the Pulser test remove it now If the reference board was installed in place of an ordinary board you can put the original board back in again
27. ation Report Vallen Systeme GmbH The Acoustic Emission Company AMSY 5 System Verification Report VeriAnalyzer Revision 2010 0921 Data Format 00 01 Sys Template Doc 18 Mar 11 Systemid No Model PAX Count 45996 AMSY 5M6 2 4 _ PAOD 3 Reference ASIPP used for Pulsertest Id No 71798 02 0077 Pulser Verification PulserAmp Amp RefASIPP Min Allowed Max allowed passed failed Max 400V dB dB dB Min 4V dB dB dB Remote 200V dB dB dB Verification of l O Flag Function Stimulating Bits SPO SP1 Waming Alarm over alarm connector and verifying them in ParStatus Verifying Parametric Input Function Signal Source Martel M2000 IdNo 43071 calibrated till Dee 2010 Signal fed in to all PAx in parallel Feeding 21 DC voltages from 10V to 10 in 1 0 V steps into all parametric inputs PAO to PAS passed Allowed Deviation 250 mVavg avg Average Allowed Noise 5 0 mVpp pp Peak Peak Max Min po a SSS SS 5 T T 5 oja Soe ee eee Reese eee eeeees o 44 I j 54 liE m i 144 10000 5000 0 5000 10000 t0000 5000 0 5000 10000 PAO deviation mavg vs Input mV max e PAT deviation mava vs Input mV max 1 a PAO noise mV pp vs Input mv max a FAT noise mv ppl vs Input mW max pE T ii oj oe Seep eee eee eo a _ 5 i l A 5000 g 5000 10000 10000 oj E E s nane ee Oe eee yp 10000 5000 0 sgg 10000
28. cceccceeseeeeeeeeeeeeeneeeeeeeaees 10 De Dirac PIS CS ater essen a cc tgs ai eee eee eee te een n i a E aaa 10 7 6 Sine square shaped DUPSEs ooiccocccccsncessciensececdeeccevesscnedentdcieededexsecsvhas teas odds eetedersbhasddueenteianncdbedersdactenss 10 7 f Data example Frequency and rise time response cccceccceeccseeeeee cece eeseeeeseeeeaeeesaeeesaeeesaeeeaaeetaes 11 7 8 Data example Amplitude variation amplitude and energy response ccceeseeeeeeeeeeeeeeeeaeeeeeees 11 7 9 Calibration Pulse RESPONSE cccccsscccessceceseeeceuseccanceceaeeceaseceaseessueeesausessageessageessauseessueessausessags 12 7 10 Continuous sine A Cette geese esis sete ea cats ecco ete ese iiaea aaa 12 7 11 Example of an Channel Verification Report ccccceccccsecccceeeceseeeeeeeeeeeeeeeeeeeeesseeeeseeeeseeeeseeeensaes 13 7 12 Example of a System Verification Report cccccccccecccceeececeeeeeeeeeeeeeeeeseeeceseeeeeseeeesseeesseeeesaeeesaaees 14 o UNS aara A A E AES 15 8 1 Configuration Report VeriCfg doc ccccccccccssseecceesseecceueeeceaseeccsauseecseuseecssageeessaeeessauseesssgeeessagseees 15 8 2 Sys Template doc and ASIPP Template doc cccccccccccsecccceesceceeceeeeeeseeeesseeeesseeeesseeeseeeesseeeessnees 16 9 3 Customize TEMpPIAICS nescence a a 16 9 WASTAN AU ON REE E eoumvecacieseueceseanuncespeecdeesen 16 file SysVeri1105 page 1 37 Tel 49 8178 9674 400 Copyright 2011 Vallen System
29. cy variation at a fixed amplitude to measure Frequency response no need for transient recorder Rise time response and with amplitude variation at fixed frequency to measure Peak amplitude response linearity Energy response squared or non squared Chan 1 Set 109 Index 109 O 0 00 33 751 6493 Time ys Freq dB kHz Rectangle Frame 161 2433 ae by x _ NOOO W E gt 0 50 100 150 200 0 200 400 600 800 1000 1200 Fig 9 Response to a sine square shaped burst file SysVeri1105 page 10 37 Tel 49 8178 9674 400 Copyright 2011 Vallen Systeme GmbH email info vallen de Fax 49 8178 9674 444 Schaftlarner Weg 26a 82057 Icking Munich Germany http www vallen de Vallen SysVeri Software for automated AE System Verification vallen systeme The impurities in the FFT above 220kHz right diagram are produced by the function generator but do not disturb the results 7 7 Data example Frequency and rise time response Fig 10 shows results of test 4 where test signals of different frequencies sine square shaped were used The upper diagram shows dBae values over the frequency Above the diagram the software reports the 3db bandwidth limits The lower diagram shows the rise time deviation vs rise time the measured vs the should be values of those signals that were measured with an amplitude above maximum 1dB Since the
30. d verification process A number of signal patterns are automatically created by a programmable signal generator measured by the system under test and verified by the VeriAnalyzer software The VeriAnalyzer creates a verification report for the system and for each channel The main difference between verification and calibration is given by the request that a calibration needs signal generators traceable to a standard and the laboratory executing calibration services must be certified according to ISO 17025 Vallen SysVeri in combination with the required hardware accessories makes system verification a fast and straight forward process that can be performed by most AE system operators even without the need of detailed system knowledge file SysVeri1105 page 2 37 Tel 49 8178 9674 400 Copyright 2011 Vallen Systeme GmbH email info vallen de Fax 49 8178 9674 444 Schaftlarner Weg 26a 82057 Icking Munich Germany http www vallen de Vallen SysVeri Software for automated AE System Verification systeme 2 Requirements SysVeri requires a fully operational AE system model AMSY 5 or AMSY 6 The PC must have the following software installed Microsoft Word version 2000 or later Vallen AE Suite software release 2011 0303 or later with software Key SysVeri enabled For system verification the following is also required A programmable DC calibrator with RS232 cable CbD in Fig 2 to stimula
31. e Fax 49 8178 9674 444 Sch ftlarner Weg 26a 82057 Icking Munich Germany http www vallen de Vallen SysVeri Software for automated AE System Verification systeme Details such as HW hardware revision and FW firmware revision can be taken from the Vallen Systeme Configuration Sheet that comes with each system or from the software tool HW Reporter which you find on the Vallen Control Panel 8 2 Sys Template doc and ASIPP Template doc Sys Template doc is a Word file used by VeriAnalyzer for the generation of the System Verification Report ASIPP Template doc is used for channel report generation ASIPP as well as ASIP 2 The templates are installed in c vallen SysVeri 8 3 Customize Templates Sys Template doc and ASIPP Template doc contain text marks and or diagram place holders where VeriAnalyzer inserts results in numerical or graphical presentation The Vallen header footer and company logo can be replaced by that of the company performing the verification make a backup of ASIPP Template doc from c vallen SysVeri open a new empty WinWord document insert the desired header footer and make sure page format there is at least 17cm free width and 24 5cm free height for printouts open C vallen SysVeri ASIPP Template doc and copy the body beginning with Verification Report till Sign _ into the new document finally replace ASIPP Template doc in c vallen SysVeri by the new document Then reports w
32. e GmbH email info vallen de Fax 49 8178 9674 444 Schaftlarner Weg 26a 82057 Icking Munich Germany http www vallen de Vallen SysVeri Software for automated AE System Verification systeme 10 SVS VCR SIC DV SED e e EA EEE E E E E 17 10 1 Reference Board ananananonnnennnnnnnnnnnnnrnnrrnrnnnnnrnnnrn rnrn reren nran nA sane APADA DE PAPE antl heajeeuanctn taroumneneeetntentateneded eee 17 0 2 CO O OS fji AA AEN AAAA 17 10 3 Starting VeriStimulator ccccccccccssccccseseeeceesscecceaseeecseeeeceeuseeessuaeesseeeesseaseeeseageeeesegseesseaseeeessageees 18 10 3 1 Vallen VeriStimulator ecciuiiedsced chins ccncxdiawsinianiodansdenenciansinteadiienaemmcedeasaneiabsisieibewedsmian tecknsuneneveden 18 10 3 2 Make Veri Folder and create Ver i Fil S c cccececceceececeececeeeeseceeeeeeeeeseeeeeeeaeseeeeeeeeeneseneees 19 10 4 Enter Configuration Report ccccccccccsseecccesseecceeeeeceeeeeessaeeesseeeeeseeaeeeeseaeeeessaeeessegeeessaeeeeesaaeeees 19 10 4 1 Configuration Report for more than 32 Channels ccccccccseeeeeceeeeeeeeeeeeeseeeeeeeaeeseeseaees 21 10 5 System V SPIGA OD a sicisecscsenscinaneicanieabenneaince icmiateasanmneinesaiehesintabwae seticmdntnatsbcuanbadeisinatwbanatesis iSjoumduaiainswnnaaminednaianus 22 Oh oy MS gt 0 pal PFPA Flags aea e ee ee ee eee see ee 22 1032 Ep 0 nena nano nn ee ee 23 10 6 Channel Verification ooisiedies tein tetncsedine sas ennitin a icin
33. e or USB or network to PC If the only free serial interface is occupied by the DC calibrator System Verification can be done without function generator and Channel Verification can be done without DC calibrator Only one of them needs to be connected at a time Connecting several AMSY chassis e AMSY 5 Connect all AMSY 5 boxes by using the foreseen cables CBL1 1 5 PC to 1st AMSY5 box length 1 5m CBL1 0 7 from box to box length 0 7m The Pulse BNC connectors of all boxes must be connected in parallel by using BNC cables and BNC T pieces Make sure that the switch settings on rear panel of master slave units are correct master unit switch 5 and 6 set to on slave units switch 5 and 6 set to off e AMSY 6 Please see AMSY 6 Operation Manual chapter Multi chassis setup on how to connect several AMSY 6 chassis in case you want to test several chassis as one system file SysVeri1105 page 17 37 Tel 49 8178 9674 400 Copyright 2011 Vallen Systeme GmbH email info vallen de Fax 49 8178 9674 444 Schaftlarner Weg 26a 82057 Icking Munich Germany http www vallen de Vallen SysVeri Software for automated AE System Verification 10 3 Starting VeriStimulator In the VeriUtilities tab of the Vallen Control Panel start the VeriStimulator and follow the step by step instructions below After starting the VeriStimulartor it Vallen VeriStimulator R2011 0303 checks for the function generator which can be connected vi
34. er Channel by channel Use DC Bloc1 for ASIP 2 or AC Input for ASIPP to connect the output of the function generator to the FIRST channel to be verified Click Next to continue or Done when completed all channels vallen Systeme GmbH Vallen VeriStimulator Channel ertfication 2nd Filter Frequency variation test in progress vallen Systeme GmbH file SysVeri1105 page 28 37 Copyright 2011 Vallen Systeme GmbH Fax 49 8178 9674 444 Schaftlarner Weg 26a 82057 Icking Munich Germany fo Cancel email info vallen de http www vallen de Vallen SysVeri Software for automated AE System Verification 10 6 3 Step 3 Pulse through verification This step checks whether the calibration pass works as expected Stop the acquisition and change Wallen VeriStimulator the AE channel filter settings Channel Yerification Pulse through ASIP 2 only Select filter 95 300kKHz Step by step ASIPP only Sel fil 4 Hi 1 Stop acquisition and enter Acquisition Parameter Setup AE Channel Settings elect filter 1 Hi 2 In all channels select filter 1 Hi ASIFP or 95 300kHz ASIP 2 The front panel switch Amp has 3 Make sure the front panel switch Amp is in position Rem 4 Remove all cables from all channels to be in position Rem remote 5 Resume acquisition and check front panel Record Control Switches for recording enabled Do not click Next before acquisition is properly
35. field 13 2 Burst Mode Settings The selection box next to Burst Mode offers three alternatives Continuous Burst FG Burst SW Select continuous mode for a continuous sine wave square wave triangle wave ramp wave or for continuous noise Burst example A rectangle gated burst of 1ms duration and 200kHz frequency Select Sine wave and Burst FG In field cycles burst enter the duration of the burst in number of full waves cycles At a frequency of 200kHz a cycle lasts 5us For a burst of 1 ms define 200 cycles per burst When the check box Endless is NOT checked the function generator produces one burst per click on Trigger Now For a periodic repetition of bursts enter the desired burst period and check endless and click once Trigger Now For a periodic trigger of the Sine square modulated sine wave of Fig 9 or the Dirac pulse select the mode Burst SW SW means that the periodic trigger is generated by software Click on Trigger Now to trigger one burst Check Endless to produce periodic bursts in the defined burst period Burst SW supports a minimum burst period of about 1 second file SysVeri1105 page 32 37 Tel 49 8178 9674 400 Copyright 2011 Vallen Systeme GmbH email info vallen de Fax 49 8178 9674 444 Schaftlarner Weg 26a 82057 Icking Munich Germany http www vallen de Vallen SysVeri Software for automated AE System Verifica
36. for the SysVeri procedure Manual Control In chapter 13 you can find out more about running the VeriStimulator in Manual Control mode Upload LoAmp 10 3 2 Make Veri Folder and create Veri Files To create the Veri files click the button Create Veri files First create the Veri Folder in which all working files and results of the verification are stored e g D SysVeri ldentifier and specify a file name This step creates a set of files including newname VAC acquisition settings newname VAE VisualAE settings newname Cfg doc configuration file 10 4 Enter Configuration Report Start Word and open NewName cfg doc as created in VeriStimulator The file contains three groups of information to be entered see also Fig 18 Group 1 Information about the used signal sources The yellow fields indicate text to be transferred to the verification reports To change such a text double click on the yellow field This opens a menu where you can change the text in item Standard text Do not delete such a yellow field this would cause improper function of the VeriAnalyzer software file SysVeri1105 page 19 37 Tel 49 8178 9674 400 Copyright 2011 Vallen Systeme GmbH email info vallen de Fax 49 8178 9674 444 Schaftlarner Weg 26a 82057 Icking Munich Germany http www vallen de Vallen SysVeri Software for automated AE System Verification Up to four system boxes can be specif
37. ftware for automated AE System Verification systeme 5 Hardware Set up for the automated AE System Verification Personal Computer USB ASyC a Parallel i Ethernet PCl card Interface RS232 RS232 for AMSY 5 LPT1 v 5 e z i O O T O O AMSY 5 resp mies hb AE System Adapter Arbitrary Vallen Systeme ET1 DC Function i Vallen Calibrator Cenia Dig I O Alarm Warning Syst Parametric Counter CBE M2000 33120A digitally controlled PCTD Martel 33220A N Parametric Analog Inputs BNC Agilent N AE Channel ASIP 2 1 x BNC AE Channel ASIP 2 N N N Fig 2 SysVeri Hardware Setup for an automated AE System Verification Fig 2 shows the hardware setup for an automated AE system verification The PC controls the AE system as usual In case of AMSY 6 the connection to PC is done with an USB cable in case of AMSY 5 over a high speed PCl interface board ASyC For the channel verification the PC also controls an arbitrary function generator e g Agilent 33220A to generate the pre defined AE test patterns and over a parallel interface LPT1 a parallel cable CbP and the ET1 Adapter the AE system inputs for data recording control for a parametric counter and for additional I O signals For the verification
38. he calibration pulse by using a Reference Board the correct operation and routing of the 4 signals at the alarm connector Alarm Warning SPO SP1 are verified Then VeriStimulator controls a programmable DC Calibrator model Martel M2000 to generate accurate DC voltages that are fed into all parametric inputs in parallel The green circles in the diagrams above indicate the deviation of the measured from the should be values The red squares indicate the peak peak noise for each PA channel 8 Templates SysVeri interacts with a number of predefined templates that are described hereafter These templates are installed in subdirectory c vallen SysVeri by the installation program 8 1 Configuration Report VeriCfg doc VeriCfg doc resides on the application folder and is copied to a user defined folder and name by VeriStimulator The operator shall fill in the details of the configuration VeriCfg doc is used as a SOURCE for VeriStimulator It contains all configuration information relevant for the verification process such as the type ID number and current address of the AE system boxes in which the AE channels are installed for verification The number of parametric inputs to be verified The address of the Reference Board see para 3 1 to be used for pulser verification the identification and recalibration dates of the used test signal generators and for each AE channel the ID number and current address of each AE channe
39. ied one box per line Field A1 must indicate the master which is the box with the lowest RAD address Additional boxes shall be defined in the sequence of their RAD address Parametric inputs of the first box are mapped to PAO PAx where X is equal to D1 1 With AMSY 5 parametric inputs in the second box are mapped to PAy PAz Where y corresponds to D1 and z to D1 D2 1 etc up to PAZ If less than 4 boxes are used for verification the corresponding lines must be empty The following information shall be entered into the fields Ax Type of system box e g AMSY 5 M6 M6 2 M16 M16 2 M37 M37 2 or AMSY 6 MB2 MB6 MB19 An extension unit such as AMSY 5 E20 E20 2 or AMSY 6 EB21 can only be entered for A2 A3 or A4 Bx the Id number of the system box see label on rear panel Cx the rack driver address RAD Addr of the system box is configured on the rear panel For AMSY 5 this is done by a slide switch a value between 01 and 14 For AMSY 6 this is done by a rotating switch with value 1 to 9 Note The box with lowest RAD address must be configured as Master box and all others as extension box If switch setting is wrong software will point that out to the operator Cx must be entered in increasing sequence C4 gt C3 C3 gt C2 C2 gt C1 Dx the number of parametric inputs in the box usually 0 2 4 or 8 The fields D2 to D4 are considered up to a total number of 8 Example D1 4 D2 D3 D4
40. ill be created with user defined header footer None of the files should be write protected 9 Installation IMPORTANT If you have modified templates Sys Template doc ASIPP Template doc see 8 3 make a backup BEFORE starting a new installation because the existing templates are deleted during installation After installing the software please make a backup of ASIPP Template doc from c vallen SysVeri open your modified template from backup open C vallen SysVeri ASIPP Template doc and copy the body beginning with Verification Report till Sign ___ into your modified template finally replace ASIPP Template doc in c vallen SysVeri by the new document Vallen SysVeri is automatically installed with the AE Software Suite Release 2005 04 on The operation needs the SysVeri key enabled on the user s KeyFile The software installation is described in the Info Suite chapter Getting Started VeriStimulator and VeriAnalyzer can be started from the Vallen Control Panel Tab Verification Tools If the function generator Agilent 33220A shall be controlled over an USB interface the Agilent VISA drivers must be installed See Info Suite chapter AE Accessories Installing Agilent VISA drivers Model 33220 can alternatively be controlled over Network and Model 33120 can be controlled over RS232 both without a need to install the VISA drivers The IP Address of the function generator can be obtained from its menu
41. information is shown during the hardware detection phase of Acquisition F Nif no TR transient recorder module is installed Y if TR module is installed To be seen in the Hardware Detection of Acquisition G The cut off frequency 3dB of the high pass 1 in KHz See Configuration Sheet H The cut off frequency 3dB of the high pass 2 in kHz See Configuration Sheet The cut off frequency 3dB of the low pass module in kHz See Configuration Sheet Note The cut off frequency in the configuration report is that of the filter module Since the ASIPP frequency response slightly decreases towards higher frequencies it is normal that the cutoff frequency reported in the Verification Report is lower than that reported for the filter module The empty fields have no influence on the verification The configuration report is read by the VeriAnalyzer program when all verification data have been acquired It might be useful to make a hardcopy you will need to know some contents later 10 4 1 Configuration Report for more than 32 channels The ASIPP table of the configuration report defines rows for up to 32 channels We do not recommend to add more lines to this table if your system has more than 32 channels because this can lead to problems with VeriAnalyzer Program resource limitation of Microsoft Word Instead please use several configuration reports each listing up to 32 channels Example verification of AMSY 6 chassis MB19 with 38
42. ing acquisition file SysVeri1105 page 33 37 Tel 49 8178 9674 400 Copyright 2011 Vallen Systeme GmbH email info vallen de Fax 49 8178 9674 444 Schaftlarner Weg 26a 82057 Icking Munich Germany http www vallen de Vallen SysVeri Software for automated AE System Verification 14 2 System Verification Step1 PAx Flags Corresponding chapter for AMSY 5 is chapter 10 5 1 The following table shows where special actions are necessary for the AMSY 6 Acquisition Steps in VeriStimulator Info given by VeriStimulator Vallen Acquisition program with AMSY 6 connected Start System Verification Step1 PAx Flags Concerning item 2 Set all range switches at Step by step PAx to 10V see on the left on AMSY 6 no 1 Connect all parametric Inputs PA0 PAX in parallel to the output of the DC generator HW switches for the parametric channels exist for negative input Instead the input range of the parametric 2 Set all range switches at PAx to 10 3 Start acquisition and check front panel Record Control Switches for recording channels must be set by software to 10V enabled Before starting data acquisition with Go to data Click Next to continue 21 PA values acquisition set the Input range of the parametric channels to 10 V usually default value in Acquisition Parameter Setup Acquisition Parameter Setup General Settings AE Channels Parametric Fre External parametric values can be conver
43. ition the noise will vary You can adjust the volume and sensitivity with the according knobs at the SF module only AMSY 5 10 6 1 Step 1 Test Freq The working frequency has to be set well in between the interval specified by the filter settings in the AE channel e g to 200kHz for a 95 300kHz band pass filter ASIP 2 only Select filter 95 300kHz ASIPP only Select filter 1 Hi ASIP 2 only It is VERY IMPORTANT to use the DC Bloci Otherwise the function generator will be damaged ASIPP only Use the AC Input and make sure that Impedance switch is set to 50Ohm Tel 49 8178 9674 400 Vallen VeriStimulator Channel erification Tests Fregq Step by step 1 The edit field Working Freq 200 0 kHz must show a frequency well within the bandwidth of the selected filter in the channel to be verified 2 Stop acquisition and enter Acquisition Parameter Setup AE Channel Settings 3 In all channels select filter 1 Hi 4SIPP or 95 300kHz ASIP 2 4 Resume acquisition and check front panel Record Control Switches for recording enabled Do not click Next before acquisition is properly running vallen Systeme GmbH Vallen VeriStimulator Channel erification Tests Fregq Channel by channel 1 Use DC Bloci1 for ASIP 2 or AC Input for ASIFP to connect the output of the function generator to the FIRST channel to be verified 2 ASIPP only Set front panel switch Impedance to 50 Ohm
44. l plug in the channel s hardware and firmware revision the installed options squarer transient recorder and bandwidth limits of the frequency filters vallen Vallen Systeme GmbH susteme The Acoustic Emission Company AMSY 6 AMSY 5 Configuration Report VeriCfg doc 28 March 2011 Version 3 1 Signal Source used for PAx Input Stimulation IdNo calibrated till Signal Source used for ASIPP Verification IdNo calibrated till System Box Type IdNo No of PA Channels Ref ASIPP Id Ref ASIPP Logical Adr AMSY 6 MBxx C eee AMSY 6 EB21 EC S ee 82174 05 001 P O TR 4M O High Pass 1 High Pass 2 Low Rem Verified ASIP 2 Vallen ID H F SQ K TR 2 K Hi Lo Pass Ch No wI W_ YIN Y N HP4 HP4 BP TP4 ia cm amet ha a as on 1 e 410 it J w E 2 rr AMSY 5 Mxx 2 a AMSY 5 Testrack 42296 04 ASIPP Test No AS 3500 os ma 850 o 708304003 025 850 4 708304004 1 1 R2010 Y Y 095 025 850 82174 05 005 21 R20 10 Y Y 095 025 850 6 8217405106 21 R20 10 Y Y 095 025 850 Fig 18 Extract of a Configuration Report VeriCfg doc filled in by the operator with details about the devices used for test signal generation the configuration of the system and the channels under verification file SysVeri1105 page 15 37 Tel 49 8178 9674 400 Copyright 2011 Vallen Systeme GmbH email info vallen d
45. le SysVeri1105 page 35 37 Tel 49 8178 9674 400 Fax 49 8178 9674 444 Copyright 2011 Vallen Systeme GmbH Sch ftlarner Weg 26a 82057 Icking Munich Germany email info vallen de http www vallen de Vallen SysVeri Software for automated AE System Verification systeme Steps in VeriStimulator Info given by Vallen Acquisition program with AMSY 6 VeriStimulator connected Step Dy step Set pulse amplitude in DAC online control to 8 Set switch Amp to Rem up 200V You can use the move left move right keys to position exactly after selecting the ruler with the mouse Ignore Item 8 Set switch AMP to Rem up because this switch does not exist on AMSY 6 Pulse amplitude ns 7 Amplitude 200 w mm ay i A 1 Pulser Verification complete No difference to AMSY 5 Stop Pulsing in Acquisition 14 4 Channel Verification Click Next to continue 14 4 1 Audio control Corresponding chapter for AMSY 5 is chapter 10 6 As described in chapter 10 6 you may set the Audio switch to the channel that is verified During the data acquisition the noise will vary In contrast to the description in chapter 10 6 no knobs exist on the AMSY 6 to adjust volume and sensitivity of the audio output For this purpose please open instead during acquisition the Audio Pulse Control DAC online control by clicking on the Other button Record Threshold Contral Ch
46. mum left end Click Next to continue Set the frequency filter settings of your reference ASIP 2 board to 25 850 kHz Acquisition Parameter Setup Concerning item 2 Check settings on Special Function Unit SF see on the left must be ignored on AMSY 6 no HW knob is present to set manually the amplitude of the pulser the switch AMP does not exist because the pulser is always controlled remotely by software the switch Calib does not exist pulsing can be only started by software Start acquisition with Go to data acquisition Open the Audio Pulse Control DAC online control by clicking on the Other button Help Record Chrl T Audio Pulse Control Log file Show data sets instead of file size Chrl Lb and set the Pulse amplitude to maximum 450V DAC online control Z Output enable Iw Audio output Sensitivity dB volume dB Pulse amplitude Amplitude 450 v The Audio output parameters play no role for automatic verification ltem 4 has been done already before see above i e you can ignore it here For Item 6 select single channel with reference channel e g 4 in screenshot below and click Start Pulsing Same as for AMSY 5 Fulsing Function start Pulsing Instead of Item 7 set pulse amplitude in DAC online control to minimum before clicking next in the VeriStimulator Pulse amplitude Amplitude 1 e fi
47. nob Amp to maximum right end Set switch Amp to local down Set switch Calib to Off down Click Next to continue vallen Systeme GmbH Cancel file SysVeri1105 page 23 37 Tel 49 8178 9674 400 Copyright 2011 Vallen Systeme GmbH email info vallen de Fax 49 8178 9674 444 Schaftlarner Weg 26a 82057 Icking Munich Germany http www vallen de Vallen SysVeri Software for automated AE System Verification In the Hardware Detection Wallen Verwmtmulator Results amp Channel Configuration dialog select an input device for System Verification Pulser test the reference channel such as the AEP3 34dB prog or the AEP4 34dB or the VS 150RIC Step by step 34dB All these devices have an impedance of 50 Ohm and a gain of 34cdB In acquisition start Single Channel pulsing and select the You can make changes in Acquisition Parameter Setup now and return to this window reference channel to send pulses when done 3 Set reference channel impedance to 50 Ohm 4 Remove all cables from all channels 5 In Acquisition click Start Pulsing using function Single channel and select the reference channel as defined in Config Report Click Next to continue vallen Systeme GmbH settings are changed there is a i delay of 20 seconds before the System Yerification Pulser test measurement is done This lets the amplitude of the pulse settle Please wait 20 seconds for pulser
48. rd make sure that it is installed in the last system box as last board Alternatively you can choose the last installed ASIP 2 in the last system box Set one of its channels filters to 25 850kHZ ASIPP as reference board In case of an ASIPP the HP high pass and TP low pass filter have to be replaced by dummy filters The reference board shall be installed in the last system box last with respect to linking of the system boxes If all slots are occupied by boards to be verified remove the last board and replace it with the reference board After the pulser step of the System Verification before the operator starts the board verification the reference board can be removed and the initial board reinstalled System must be powered down before a board is removed or installed As every board must be configured to a unique channel number make sure the channel number of the reference board is unique in the AMSY 5 respectively AMSY 6 system box Channel number 60 is recommended for the Reference Board Usually boards are configured in range 1 to 37 with AMSY 5 or 1 to 19 for AMSY 6 systems 10 2 Connections Connect LPT1 of PC printer port to ET1 Adapter using cable CbP Connect ET1 external cable CbE to AMSY 5 master unit s external connector at the rear Connect ET1 Alarm cable CbA to AMSY master unit s Alarm connector at the rear Connect DC calibrator serial interface to PC Connect Function generator serial interfac
49. rification step You can control this value by the DAC online control see previous section file SysVeri1105 page 37 37 Tel 49 8178 9674 400 Copyright 2011 Vallen Systeme GmbH email info vallen de Fax 49 8178 9674 444 Sch ftlarner Weg 26a 82057 Icking Munich Germany http www vallen de
50. rl T Audio Pulse Control Chrl 4 Ctrl L Show data sets instead of file size Then enable Output enable and use the rulers for Audio output in the DAC online control as shown in next figure DAC online control x Audio output Sensitivity dB volume dB Output enable F ulse amplitude Amplitude 200 Please note that for high sensitivity you have to move the upper ruler to the right The chassis MB2 does not have a built in loudspeaker Instead you can connect passive stereo loudspeakers to the speaker connector at the rear Please do not change the pulse amplitude value of 200V because this value is needed for Pulse through verification step file SysVeri1105 page 36 37 Tel 49 8178 9674 400 Copyright 2011 Vallen Systeme GmbH email info vallen de Fax 49 8178 9674 444 Sch ftlarner Weg 26a 82057 Icking Munich Germany http www vallen de Vallen SysVeri Software for automated AE System Verification systeme 14 4 2 Channel verification Step 3 Pulse through verification Corresponding chapter for AMSY 5 is chapter 10 6 3 Chapter 10 6 3 describes that you have to set the front panel switch Amp to position Rem This instruction you can ignore because on AMSY 6 this switch does not exist and the pulser amplitude is always controlled by software The pulser amplitude is set by the SysVeri acquisition setup file to 200V which is the value needed for this ve
51. running vallen Systeme GmbH Start the pulsing in acquisition Wallen VeriStimulator Single cycle i Channel Yerification Pulse through The test ends after the last channel completed pulsing Click Done to finish this step of Step by step the SysVeri and to return to the oe 6 In Acquisition click Start Pulsing using function Single cycle VeriStimulator q g g gle cy 7 Wait until the last channel has completed pulsing Do not click Done before the last channel has completed pulsing vallen Systeme GmbH 10 7 Data Analysis for SysVeri A printer can now be reconnected to LPT1 and the printer driver reconfigured to LPT1 Start Vallen VeriAnalyzer VeriAna exe from Vallen Control Panel Veri Utilities The menu in Fig 21 appears In frame Configuration file browse to the Veri Folder and select filename cfg doc In frame Primary Data File for System Verification browse to filename sys pri In frame Primary Data File for ASIPP Verification browse to filename asp pri Click on System Start to create the System Report When processed store the file on filename sys doc and print it Check and sign it file SysVeri1105 page 29 37 Tel 49 8178 9674 400 Copyright 2011 Vallen Systeme GmbH email info vallen de Fax 49 8178 9674 444 Schaftlarner Weg 26a 82057 Icking Munich Germany http www vallen de Vallen SysVeri Software for automated AE
52. set 20 2 us allowed 5 0 Offset 28 6 eu allowed 5 0 above1 Offset 2 Cts allowed 5 0 Fig 7 Results in Fig 6 shown in three deviation diagrams The horizontal position of each dot in a diagram represents a should be value and the vertical position represents the deviation of the measured value from a should be value in percent or dB as given in the legend The meaning of each diagram axis results and some additional numerical values are shown in different parts of the legend The Offsets indicated in the legends of Fig 7 were determined by the deviation of the first shortest measured signal from the should be values and applied to all 9 signals to correct the influence of the settling behavior and ring down effects 7 5 Dirac Pulse Series Series of short pulses are used to check duration discrimination and rearm timing internal time parameters Chan 1 Set 6 Index 6 O 0 00 06 376 1939 Time us Freq dB kHz Rectangle Frame 500 4096 ou 0 5 Amplitude mV 1 5 50 0 50 100 150 200 250 300 350 Fig 8 Response to a series of Dirac Pulses 7 6 Sine square shaped burst This is a sine wave burst in an envelope of a sine square shape Fig 9 This burst consist of 48 samples sinewave and 41 sine waves for the burst this gives a sequence of 1968 samples for the complete test signal These sine square shaped test signals are used with frequen
53. ssnatnn bbe sonal dae hot nchdcibarneiverosdeliinwiioa baivslueind saibennaire datesien lide moeiadandaeaioeteamaenis 25 106 1 Spi TESO EO nen ee ee ae e eE 26 10 6 2 Step 2 2 Filter ccccecsccecsecessecesseseecercseesceceeseceeveceevarcevauseveusevsuvaceusarceveersnsesnnvevensatereatenss 28 10 6 3 Step 3 Pulse through verification cccceecccecseeeeeeseeeeeeeeeeeeseeeeeseeeeeseeeeeeseeeeeeseeeesaeeees 29 10 7 Data Analysis for SYSVELPI cccccccceeeceeceeeeeseeeeeeeeeeeeeaeeeeeseeseeesseeeeeeseaeeeessaueeeseaeeeeaegeeeesaesseeeesaaeees 29 11 Verification Statistics and Pattern LiSts ccccscssescssescnseecneeenseecneeenseeeneseeneeeeneeeesueogesenaseanesseneusens 30 12 Verification of Transient RECOrdeS ccccceccsccseccsecceccuecneccuecceueuseuseeeeuseeseuseueueueeeeeuseuseeseueeeesseussuenes 31 13 Vallen VeriStimulator in Manual MOde cccccceccsecsecneeceecceeneeceeeuseeeseuseeeeeeeueeueeeuseeseuseeeseneeeeesenns 31 TGA VS OF TUNC IO ie E E sme dsesecaude ic tape doaduceedount ne E E tends 32 SZ EURT VIOUS Sets rectsnstalsccpsdsuetececdidensytanctpaadete cyecsesucuude E se coeteensitioneeas 32 14 Appendix Verification of FINES WH G ic scha isaac atc wide ted eee anaana aa 33 NEN SNOW coset eee erst entered eal ke ne dren amet Gentine detonate EENAA A E 33 14 2 System Verification Step1 PAX FIAQS ccccceccccssececeeeeeceeeesecesseeesseeeeseeeeeseasesseeesseeeessaeesaeeees 34 14 3 System
54. te the parametric inputs SysVeri software supports the control of model M2000 over RS232 Software also supports the use of any manually controlled DC Calibrator e g MIVC222 HPII Required accuracy 0 015 at 10V output range 0 to 10V BNC cable BNC T connectors and adapter for the calibrator output usually 4mm banana sockets to connect the DC output voltage of the DC calibrator to all parametric inputs in parallel An already verified Reference Board to measure the calibration pulse a ASIP 2 A or ASIP 2 S board with software releases higher than R2008 0101 or b ASIPP board with dummy high pass HP4 dum and dummy low pass TP4 dum filter module For AE channel verification the following is also required A programmable Arbitrary Function Generator to stimulate programmed waveforms of AE bursts SysVeri software supports control of model Agilent 33220A over Ethernet or USB alternatively model Agilent 33120A over RS232 with appropriate cable to PC CbF in Fig 2 An appropriate cable to connect function generator with PC RS232 USB or Ethernet With ASIP 2 DC Bloc1 DC Blocker for blocking the DC voltage supplied by the channel to protect the function generator A BNC cable to connect the function generator output over a DC Blocker for ASIP 2 verification to the AC input of one AE channel at a time The PC must offer a sufficient number of free LPTx RS232 USB Ethernet ports to control the ET1
55. ted fro Conversion Specify here the range for each parametric char Oo o1oo 1 y Then Go to data acquisition and continue as instructed by the VeriStimulator window file SysVeri1105 page 34 37 Tel 49 8178 9674 400 Copyright 2011 Vallen Systeme GmbH email info vallen de Fax 49 8178 9674 444 Sch ftlarner Weg 26a 82057 Icking Munich Germany http www vallen de Vallen SysVeri Software for automated AE System Verification 14 3 Corresponding chapter for AMSY 5 is chapter 10 5 2 System Verification Step2 Pulser Steps in VeriStimulator Info given by Vallen Acquisition program with AMSY 6 VeriStimulator connected Start System Verification Step2 Pulser Step by step 1 Check front panel Record Control Switches for recording enabled 2 Check settings on Special Function Unit SF Turn front panel knob Amp to maximum right end Set switch Amp to local down Set switch Calib to Off down Click Next to continue Step by step 3 Set reference channel impedance to 50 Ohm ASIPP only 4 Set reference channel filter to 25 850kHz 5 Remove all cables from all channels 6 In Acquisition click Start Pulsing using function Single channel and select the reference channel as defined in Config Report You can make changes in Acquisition Parameter Setup now and return to this window when done Click Next to continue Step by step 7 Turn front panel knob Amp to mini
56. tion systeme 14 Appendix Verification of AMSY 6 This appendix is an addition to the previous chapters and describes points which must be observed when using the SysVeri Software R2011 0303 or later for AMSY 6 The verification of the AMSY 6 is not yet fully integrated in the SysVeri software and the previous chapters Nevertheless it is possible to verify AMSY 6 with SysVeri by taking the points described in this chapter into account Only the differences to AMSY 5 verification are described in this chapter For description of overall proceeding of system verification please see previous chapters 14 1 Overview With respect to using SysVeri software the main differences between AMSY 5 and AMSY 6 are e Inthe configuration report see chapter 10 4 please enter as system box type the chassis type of your AMSY 6 MB2 MB6 or MB19 e Input range of parametric channels is set in AMSY 5 by HW switch 1V 10V and in AMSY 6 by software Since the verification must be done with input range 10V the input range must be set by software before starting System Verification Step 1 PAx Flags e The amplitude of the pulser can be set in AMSY 5 by HW knob or by software switch Rem Local set to Rem whereas in AMSY 6 the pulser parameters are set only in software by DAC online control during acquisition This has an impact on System Verification Step2 Pulser because the amplitude of the pulser must be set by software dur
57. u created during the Create Veri files step we recommend file name of type filename sys vac Confirm the creation of a new file For the settings browse to the folder containing the vac file created during the step Create Veri Files and select it Make sure that the record control switches are in enable position In a next step the IO flags will be checked automatically Tel 49 8178 9674 400 Select This test needs a calibrated DC source Please select one option Detect an M200 calibrator attached Use other DC voltage source with manual contral Vallen VeriStimulator System Yerification Parametric inputs and flags Step by step J Cancel 1 Connect all parametric Inputs FAD PAX in parallel to the output of the DC generator 2 Set all range switches at FAx to 10 3 Start acquisition and check front panel Record Control Switches for recording enabled Click Next to continue 21 PA values vallen Systeme GmbH Vallen VeriStimulator System Yerification Parametric inputs and flags Loading function generator done Enabling recording done Enabling Led SPO done Enabling Led SP1 done Enabling Led Warning done Enabling Led Alarm vallen Systeme GmbH file SysVeri1105 page 22 37 Copyright 2011 Vallen Systeme GmbH Fax 49 8178 9674 444 Schaftlarner Weg 26a 82057 Icking Munich Germany Cancel email info
58. uldbe corected by 0 11 48 allowed 0 50 to 0 50 dB Test 5 passed pqgteoesooe 9 s e ee ee se gee 4 eee ee eee eee f 30 35 40 46 5 5 60 6 70 75 80 90 95 100 702 Amplitude deviation dB vs A dBac max Energy deviation wr vs E ka dB above 3 0 pV allowed above 2 0 eu Test 6 RMSS and Floating Threshold continuous signal corrected as with Test 5 O o a a a D o es 70 75 a0 RMSS deviation vs RMSS pV gt Floating Threshold deviation dB vs THR dBag above 3 0 pV sions dB above 1 0 pV TNR 42 Test 7 Autocalibration pulser path Pulser Voltage 200 0 V Calib Amp no cable connected filter 1 Hi 88 9 dB should be gt 55dB All tests passed Icking 30 09 2010 09 17 verified by User Rph initials HPH__Sign Think AE think Vallen User friendly reliable up to date Copynght 2009 Vallen Systeme GmbH ISO 9001 certified by TUV CERT veri_ASIP 2 doc page 1 of 10 Schaftlamer Weg 26a Tet 49 8178 9674 400 email info vallen de 82057 Icking Munich Germany Fax 49 8178 9674 444 http www vallen de Fig 16 Example of a Channel Verification Report file SysVeri1105 page 13 37 Tel 49 8178 9674 400 Copyright 2011 Vallen Systeme GmbH email info vallen de Fax 49 8178 9674 444 Schaftlarner Weg 26a 82057 Icking Munich Germany http www vallen de Vallen SysVeri Software for automated AE System Verification 7 12 Example of a System Verific
59. ulse works as expected There is no accuracy check required because the pulse signal is checked by using a Reference Board see section 3 1 during the system verification run A Reference Board is needed because the filter configuration of the ASIP board can have a big influence on the measured calibration pulse amplitude 7 10 This test signal Fig 14 is used to verify the accuracy of background noise measurement and the floating threshold behavior Due to the limited dynamic range of the function generator 300UVrRws is the lowest level that can be verified Continuous sine wave Test 6 RMSS and Floating Threshold continuous signal corrected as with Test 5 Test 6 passed 5 1 0 0 05 6 oe CC 1 5 Sy SS ee a ee es ee ee eee 100 101 102 10 104 30 35 40 45 50 55 60 65 70 75 80 85 90 95 100 RMSS deviation vs RMSS pV max 1 4 Floating Threshold deviation dB vs THR dBac max 0 3 above 2 0 uV allowed 5 0 dB above 1 0 uV TNR 4 2 allowed 1 0 Fig 15 Result of verification of background noise and floating threshold behavior Test 6 file SysVeri1105 page 12 37 Tel 49 8178 9674 400 Copyright 2011 Vallen Systeme GmbH email info vallen de Fax 49 8178 9674 444 Schaftlarner Weg 26a 82057 Icking Munich Germany http www vallen de Vallen SysVeri Software for automated AE System Verification 7 11 Example
60. vallen Vallen Systeme GmbH systeme The Acoustic Emission Company Vallen SysVeri AMSY 5 AMSY 6 System Verification Software User s Manual Contents 1 Sacer eee EE i a E e E em eae NE NE 2 2 ROC INNS sesos N E NEEE N 3 o TOU ri a a aaa e aaa a ea a aa A 4 31 System VernncalON sesi EE a 4 3 2 Channel Verification ASIPP or ASIP 2 cccccccccssseeecsesseeeseeseeeceeeeecseaseessaeeessageeeeseseeessageeessaeeeees 4 4 Basic Procedure for an AE Channel Verification cccceesseeeseeeeeeeeeeeeeeneeeeeeeeeeaneeeeeeeeeeneanseeeeeeseenes 5 5 Hardware Set up for the automated AE System Verification cccccesssesseeeeeeeeeeeessseeeeeeeeennneeeeeeeeeens 6 6 Software Component Interactions for the automated System and Channel Verification 6 oa Mia gy 1 6 110 1 06 s seee Se 7 62 PCC AON No ces sco reer eps act iew tae eerie ects a EE TEESE E aot aE a de enced eto E aE A E a AS T 6 3 VerAnalyzef sensisse enrama a aa a A R AEE Eai a EEEa a T 7 AE Test Signals and Result Examples cccccccssesseeeescenseeeeseeenseeeeeeesseeeseoeesseeeseoensseeeseoensseseeeonssseees 8 Fel ONJE Dra PS Oea E E r eE E a Ee i E EE E ES 8 TZ Rectangular shaped DUFS iieis rai ET EE a EEr EEE Ea 9 7 3 Listing example for the duration variation test ccccccceeccecceeeeeeeeeeeeeeeeeeeesseeeeeeseeeeeeseaueeesaeeesseeesees 9 7 4 Deviation diagram example for the duration variation test Test 8 ccc
61. vallen de http www vallen de Vallen SysVeri Software for automated AE System Verification The first voltage that has to Valen VeriStimulator applied is 10V 7 System Verification Parametric inputs and flags In case of the M200x one has to reverse polarity in order to get 10V The PAx verification will run automatically until OV is reached Then you are prompted to Click Next to continue provide positive voltage To do so reverse polarity again Step 1 of 21 Set the DC generator to 10 0 lf you are using a manually controlled DC calibrator change the DC settings according to the instructions given in the pop up window vallen Systeme GmbH Cancel 10 5 2 Step 2 Pulser For this test you need the aforementioned section 10 1 reference board For the instructions about the reference board see section 10 1 Please select the correct filter settings for an ASIP 2 Reference Board 25 850kHz or the dummy filters for an ASIPP board Follow the instructions of the pop Wallen VeriStimulator up window and make sure that the SF board switches and knobs System Yerification Pulser test are set correctly The Amp knob has to be turned Bien by slep all the way to the right The Amp swi tth has to beset Check front panel Record Control Switches for recording enabled to local and the Calib Switch Check settings on Special Function Unit SF has to be set to off Turn front panel k
62. ystem Verification systeme Group 3 ASIP table Information about the installed ASIPs AE channel plug in ASIPP or ASIP 2 Brac O TR 4M O High Pass 1 High Pass 2 Low Pass Rem Verified Vallen ID ak SQ K TR 2 K Hi1 Hi2 YIN YIN HP4 HP4 BP ma an 1234 ete fet a 025 044 850 ASIPP4 1234567890 11 For channel 1 the form above shows default values They have to be replaced with the actual data for the installed ASIPP 1 You can take these values from the configuration report that came with the system at time of delivery The meaning of fields B to I are described below B ASIP 2 this is a number on a white label on the board It can be obtained with the tool Hardware Reporter Vallen Control Panel tab Utilities without removing the board ASIPP single channel board only supported in AMSY 5 not in AMSY 6 a hand written number on the board itself remove the board to read that number for safety cautions see your AMSY 5 Manual section AMSY 5 Service Manual C Hardware Revision of ASIPP or digital board of ASIP2 is given on the Configuration Sheet that was shipped with the system within the System User s Manual or can be obtained with the tool Hardware Reporter Vallen Control Panel tab Utilities D Firmware Revision can be obtained with the tool Hardware Reporter Vallen Control Panel tab Utilities EN if no squarer installed Y if squarer is installed This

Download Pdf Manuals

image

Related Search

Related Contents

Remote Card  Digitus VGA splitter  2014-2015 Sysop Manual - Cabell County Schools  LG H340 8GB 4G White  施工説明書 - 手すりをつけよう!  Netgear GS105E-200PES network switch  

Copyright © All rights reserved.
Failed to retrieve file