Home
        CombiFlash Rf User Manual with Rf150
         Contents
1.       5 4 1 Run Control Buttons    e Play     The Play button starts the run  During the run   other run control buttons are active     e Pause     The Pause button holds the  B at the current  value while the system continues to operate  sometimes  called an isocratic hold   Note that pausing the run  extends the run length  While in the Paused state  you       5 9    CombiFlash   Rf User Manual    can resume the run by clicking the Play button  or stop  the run by clicking the Stop button  If you resume the  run  the system continues with the gradient curve at the   B when the system was paused     e Stop     The Stop button suspends the entire run   Unlike the Paused state  the pump  peak detection  and  fraction collection will not operate  While stopped  you  can end the run by clicking the Rewind button  resume  the run by clicking the Play button  or skip the rest of  the run sequence by clicking the Fast Forward button         CAUTION     Rf200 and Rf200i only     When the Stop button is pressed   solvent remains in the tubing from the injection valve to the fraction  collector s drop former  Follow the steps listed below  considering  your instrument s configuration       Solid Sample Cartridge with air purge enabled   If you suspend the run with the Stop button  do not remove  the column and or solid sample cartridge without first  pressing the Fast Forward button to purge solvents from all  lines  If you stop the run because your peak of interest has  been collecte
2.       Comments                                4 15    CombiFlash   Rf User Manual       CombiFlash   Rf  User Manual       Section 5 Operation    This section provides abbreviated operating instructions for  the CombiFlash Rf systems  For complete instructions  refer to  Help menu option from PeakTrak   s main menu     5 1 Sample Preparation    Before starting a run  consider how the sample will be  introduced to the column media  This section discusses three  methods  liquid injection  solid sample cartridges  and  preloading the sample on the column     5 1 1 Liquid Sample Injection  If the sample is soluble in the starting mobile phase  it can be  prepared as a solution and injected onto the column when  prompted during the purification run     5 1 2 Solid Samples  Some compounds are not soluble in solvents that are  compatible with the chromatography  In addition  the  compounds may have very limited solubility in any solvent   resulting in sample volumes that are impractical for good  chromatography  The answer in this case is solid sample  introduction  Here the reaction mixture or dissolved sample is  mixed with the same media in the chromatography column     Some samples of this type may be simply dissolved and placed  into a prefilled cartridge  You can then use the cartridge  immediately or dry it before placing it on the system     Other samples may need more care  Typically you would  create a mixture of 20  sample load to media  w w   Then dry  the mixture under 
3.       Figure 5 3 Align and push cartridge into the cap  and then rotate the  cartridge to secure it        5 4    Section 5 Operation    5 1 3 Preloading on Column    You can load the reaction mixture or sample directly onto the  column  sometimes called    direct    or    dry    loading     M Note   If using this method  use care to ensure that a column equilibration  does not wash away the sample  Bypass the column equilibration  at run time by selecting  Preload on column   or    No Equilibration     for Rf 150 systems  as the sample Loading Type  Refer to Section  5 3 for more details     5 2 Loading a RediSep Rf Column  To load a RediSep Rf column      NWARNING    The CombiFlash Rf has redundant safety devices to limit  pressure to less than 210 psi  1448 kPa   RediSep columns  smaller than 100 g are CE certified using standard IEC61010 1  for use on the CombiF ash Rf  RediSep columns larger than  100 g meet Pressure Vessel Directive 97 23 EC  Teledyne Isco  strongly recommends against the use of columns rated less  than 210 psi  1448 kPa      For best results  always use RediSep Rf columns  The system will  not automatically detect other columns  including non Rf RediSep  columns  The system also limits the maximum operating pressure  to 50 psi when the column is not detected     1  Select a RediSep Rf column  The on line help includes a  column selection guide that can assist you in selecting a  stationary phase media and column size     2  Raise the injection valve  c
4.      RediSep     Select the column that will be used for the  run  See also  REDISEP RF COLUMN SELECTION GUIDE   ON LINEJ        Loading Type    Select the type of injection       Start Tube     Select Next Tube to begin at the next  available tube in the current rack or Next Rack to begin  with the first tube in the next rack     6 3 5 2 Gradient Plot Area   The Gradient Plot Area is the same as the one displayed on  the tabs for each column position  The display shows the  current channel  The top corner displays the current system  pressure     6 3 5 3 Run Control Buttons  The Run Control Buttons allow you to control the operation of  the system     6 3 6 Files  The Files window is modal  That is  its function and features  change according to the command used to open the window   Menu commands such as FILE gt OPEN  6 2 1 2  and SAVE  METHOD AS     6 2 1 6   or clicking Open and Save As buttons  will open this window  Use this window to browse the system s  files and folders  The following controls appear on the window       Current Path     The top left corner of the window  displays the path  current folder   As you browse  through the files  the path will update as you go  You can  click the folder names to return to upper folder levels       File and Folder Operation buttons       Copy     Click this button to copy a highlighted file to  the system s clipboard memory       Paste     Click this button to paste a file from the  clipboard memory  If the file already exist
5.     3 3 Connect Power    Ensure that the On Standby switch below the touch screen  panel is in the Standby position  Then  use the supplied IEC  power cord to connect the CombiFlash Rf to mains power       WARNING    Mains power must meet the voltage  frequency  and amperage  requirements listed on the serial number label     Z  WARNING    As long as the AC mains power cord is connected  power is  inside the unit  The mains power cord is the disconnect device   Position the Rf 150 system so that the power cord can be  unplugged  or use a power strip where the plug can quickly be  removed from the outlet in the event of an emergency        CombiFlash   Rf User Manual    3 4 Connect Solvent Lines    A  WARNING    Risk of fire ignited by electrostatic discharge  Never substitute  the black tubing on CombiFlash systems  The black tubing   P N 023 0503 06  is conductive  This tubing is required to  dissipate static electricity       A CAUTION    To prevent damage or premature wear to the pump and internal  valves  clean solvent should be used  The solvent should not  contain any dissolved solids     The Combiflash Rf 150 system has two solvent inlet ports on  the back panel  one A solvent and one B solvent  Figure 3 2         Solvent  Inlet Ports                                           Figure 3 2 CombiFlash Rf150 solvent inlet ports    The system requires at least one A and one B solvent to form a  binary gradient  It assumes that the B solvent is stronger than  A  When using th
6.    You can reset the method to Teledyne Isco recommended  defaults  While viewing the default method file for a selected  column size and type  click the    Reset file to Isco defaults          6 43    CombiFlash   Rf User Manual    button  The settings will revert to the original factory settings   Click the Save button to save the settings for future use     6 4 5 Defining a Gradient    The simplest way to change the gradient is to click and drag  the inflection points that define the shape of the gradient  Add  a point by clicking the Insert button  then click the gradient  curve to add a point  You can then drag the new point to the  desired location  Delete a point by clicking the Delete button  and then clicking the undesired inflection point     The above methods work in the MAIN WINDOW  6 3 1  and  METHOD EDITOR  6 3 2  window  Alternatively  you can modify   add  insert  and delete points using a tabular view on the  Method Editor window  Click the Edit Gradient button to open  this view  then use the controls to modify the settings  Click  the Edit Gradient button again to close the table view     The system can also optimize a gradient from your TLC  results  To use this feature  select a single solvent system and  run two TLC plates  each with a different concentration of  solvent  For example  run one plate at 1 1 hexane ethyl  acetate  the other at 1 3 hexane ethyl acetate  If the resulting  Rf values are between 0 2 and 0 8  you can enter the TLC data  in the G
7.   1 3 Operating Overview    The Combiflash Rf system is equipped with a touch screen  display for local control     The system supports TCP IP communication  This allows  direct control of the system by a Windows computer via a  cross over cable between Ethernet ports of the CombiFlash Rf  system and the computer  The Windows computer must have  Microsoft Internet Explorer 6 or newer     TCP IP communication also allows remote control of the  system via an established network  Remote controlling devices  on the network can be a Windows PC or an Apple mobile  digital device  iPod Touch  iPhone  and iPad      M Note   Teledyne Isco recommends that you obtain assistance from your  Information Technology department before attempting direct or  network connections     1 3 1 Multiple Control Possibilities    The system can be accessed from the built in touch panel and  up to ten network devices  The touch panel shares control with  all connected devices  The system performs the most recent  command from any control input     1 3 2 File Storage    To support operation from a variety of direct and network  connections  the software and all files are stored in the  CombiFlash Rf on an internal hard drive  This ensures that  your compound purification methods and run history files can  be viewed from any connection  Optionally  individual run  history files may be saved to a USB flash drive and the hard  drive or a networked computer  and all run history files can be  saved to a networ
8.   465 3799    Laboratory Equipment for Light Industrial Commercial Environments    Combiflash RF75  2008    EN 61010 1  2  Edition   Safety Requirements for Electrical Equipment for  Measurement  Control  and Laboratory Use   EN 61326 2006   EMC Requirements for Electrical Equipment for  Measurement  Control  and Laboratory Use                                              Standard Description Severity Applied Performance Criteria   EN61000 4 2 Electrostatic Discharge Level 2   4kV contact discharge A  Level 3   8kV air discharge   EN61000 4 3 Radiated RF Immunity 80 Mhz to 2 7Ghz 80  AM at 1kHz A  Level 2   3V m   EN61000 4 4 Electrical Fast Transient Level 2   1kV on AC lines   EN61000 4 5 Surge on AC Lines Level 2   2kV common mode   Level 2   1KV differential mode   EN61000 4 6 Conducted RF on AC lines 150 kHz to 80 MHz  A  Level 1   1V rms  80  modulated   EN61000 4 11 Voltage Dips 1 10 25 cycles  100 40 70  A   Voltage Interruption 250 cycles  100  drop c   CISPR11  RF Emissions Group 1  Class A Industrial  Scientific  and   EN 55011 Medical Equipment   EN61000 3 2  3 3 Harmonic and Flicker                         The undersigned  hereby declares that the design of the equipment specified above conforms to the above Directive s  and Standards    as of September 5  2008     USA Representative    odis       William Foster    Vice President of Engineering ZAN  TELEDYNE ISCO       60 5232 506             A Teledyne Technologies Company                         DECLARATION OF 
9.   7 10    Section 7 Remote Interfaces       Local Area Connection Properties    E Client for Microsoft Networks  File and Printer Sharing for Microsoft Networks  Y 3Com BCAITDI DMI TDI          Internet Protocol  TCP IP  Properties     192 168  1  200          Figure 7 7 Internet Protocol Properties Window       7 11    CombiFlash   Rf User Manual    7 2 5 Completing the Direct Connection    1  Connect the PC to the Combiflash Rf system using the  cross over cable    2 Turn the system   s On Standby switch to ON  Wait for the  PeakTrak operating screen to appear on the Combiflash Rf  system  then start the PC    3  Start Internet Explorer and type the factory default  address    http   192 168 1 51     The browser will display the  PeakTrak screen     The PeakTrak screen indicates that the direct connection is  working     The CombiFlash Rf system and the PC are now configured for  communication through a direct connection  Refer to  PeakTrak s on line help for operating instructions   Abbreviated instructions can be found in Section 5 of this    guide     7 3 Direct Printing    The CombiFlash Rf system supports direct printing to a USB  printer  The printer must support postscript  To set up the    printer    1  Connect the printer with the USB cable to the USB port on  the back panel of the system    2  Select the Tools   Configuration menu command    3  Select the Network Configuration tab    4  Select the Postscript option for the Printer Type    5  For the Connection URL
10.   8 5 8 Monthly Flow Cell Cleaning               8 10  8 6 Injection Valve Maintenance                   8 10  8 7 Electrical Troubleshooting                     8 11       vi    CombiFlash   Rf  User Manual       Section 1 Introduction    1 1 Documentation Overview  This Installation Guide provides   e Safety Information    Unpacking instructions       nstallation instructions  including connections with a  Windows computer      Certification and warranty information     Once you are operating the CombiFlash Rf  you may refer to  the Help menu for operating instructions     M Note   This Installation Guide will use the term  Combi Flash Rf  when the  context applies to all systems  This installation guide will refer to  the specific model name when the discussion is limited to a  particular model or configuration     1 2 Product Overview    The Teledyne Isco CombiFlash Rf flash chromatography  system gives you high productivity automation   programmable gradients  UV detection and peak separation   and automatic detection of columns and collection tube racks   with the exception of the Rf150   Its small size makes it a  great    personal system  and well suited for operation within  chemical hoods and other limited indoor spaces     Z WARNING    Avoid hazardous practices  If you use this instrument in any  way not specified in this manual  the protection provided by  the instrument may be impaired  this may increase your risk  of injury        CombiFlash   Rf User Manual    
11.   Flow Cell Pathlength    0 1 mm   25        UV Detection and  Monitor Wavelength    Selectable from 200 to 360 nm range  Optional 200 to  780 nm UV Vis detector              Wavelength Accuracy  5 nm  Gas Inlet Pressure 60 to 70 psig  Gas Consumption  lt 2 5 SLPM       Spray Chamber  Temperature    Setting range  10 to 60   C  limited to minimum of 5   C below ambient       Drift Tube Temperature       Setting range  30 to 90   C   Must be 5   C above spray chamber temperature  Minimum temperature is 5   C above ambient  Maximum temperature is 60   C above ambient          1 9    CombiFlash   Rf User Manual    Table 1 2 CombiFlash Rf 200i  Specifications    Continued     Split Flow Rate 0 75 mL min   10   Fraction Accuracy   2mL    flow rate   60      Electrical Safety per EN 61010 1       Pollution Degree 2       Installation Category II       Maximum Altitude 2000 meters          Note 1  All specifications are subject to change   Note 2  For columns undetected by RFID  the max pressure is limited to 50 psi           Section 1 Introduction    Table 1 3 CombiFlash Rf 4x System    Module Dimensions    1    Specifications    61 cm  24  x21 6cm  8 5  x43 2 cm  17    HxWxD        Module Weight    13 6 kg  30 Ibs        Overall Dimensions    61 cm  24  x57 6cm  22 6   x43 cm  17    HxWxD        Overall Weight    40 8 kg  90 Ibs        Power Options    Line Frequency       Ambient Temperature    Input voltage range from 100 to 240 VAC  50 60 Hz   2 5 Amps system maximum  0 5 amps 
12.   M Note   When using higher density solvents such as dichloromethane   DCM   ensure that the waste container is no more than 3 feet  below the supply solvent levels  Placing solvent containers above  the system and waste containers on the floor may cause the  internal check valves to open and allow solvent to flow through the  system when in the standby state     2 6 External Gas    2 6 1 CombiFlash Rf 200 and Rf 200i    The CombiFlash Rf 200 and Rf 200i systems use an internal  air pump to provide compressed air for solvent and waste level  sensing  Optionally  the system can deliver a compressed gas   Inert or other gases may be more suitable than air for some  applications     If ambient air is compatible with your solvents and waste   simply leave the Optional External Gas port open     If you want to use a gas instead of air  connect user supplied  lg  I D  tubing to the Optional External Gas port  Connect the  other end to the user supplied compressed gas  The external       2 8    Section 2 Preparation  Rf 200 and Rf 200i Systems    compressed gas should be dry  filtered  and regulated to 2   5  psi  0 14   0 34 atm  13 8   34 5 KPa      2 6 2 CombiFlash Rf 200i Only  ELSD operation requires a carrier gas  To connect the carrier  gas to the CombiFlash Rf 200i     1  Locate the 4s inch O D  FEP tubing  part number  023 0503 02  from the accessory package     2  Push one end of the tubing into the Nitrogen Inlet port on  the back of the system  The tubing should be full
13.   The Rf150 cannot read the RFID tag     7 Sample Injection Port     Luer type fitting to accept the  sample though either a solid load sample cartridge  shown   or a liquid injection using a syringe or similar device     8  Injection Valve     This six way valve is automatically  positioned by the Combiflash Rf according to its current  mode of operation  Modes include column equilibration   sample injection  elution  column flushing  valve cleaning   and system purging  The injection valve is not available on  the standard Rf 150 systems     9  Upper Column Mount   Secures the column inlet   10 Lower Column Mount   Secures the column outlet     11  Top Shelf     Allows storage of accessories and solvent  bottles     12  Column Expansion Ports   These ports allow the system  to add column channels with an optional module  The 4x  module option is not available for the Rf 150 systems        Figure 1 6 CombiFlash Rf Features  Side        Section 1 Introduction       Figure 1 7 CombiFlash Rf Features  Back     13 Upper Drain Tube     Liquids spilled on the top shelf are  carried away through this tube to a user supplied  container     14 External Detector  All systems except the Rf 150      The system can accept an analog peak voltage signal from  an external detector  The system will process this analog  signal to determine the presence of a peak using  PeakTrak s slope or threshold algorithms  See section 2 10  more information     15 Ethernet Port     An 8P8C jack for a netw
14.   enter  usb  dev usb lp0     6  Click OK to save the settings  The CombiFlash Rf system  will send a test page to the printer     After successfully printing a test page  the local printer will be  available for printing using the File  Print menu command or  through the Automatically Print Report at End of Run option  on the Tools   Configuration User Preferences tab        7 12    CombiFlash   Rf  User Manual       Section 8 Maintenance    8 1 System Standby and Shut Down    During extended periods of inactivity  you can place the  system in Standby to conserve power  To do so  log off the  system  File gt  Log out  and place the     On Standby switch in  Standby     When in the Standby state  normal system operation is no  longer available from the touch screen or remotely  However   some internal components are still powered       WARNING    As long as the AC mains power cord is connected  power is  inside the unit  The mains power cord is the disconnect device   Position the CombiFlash Rf system so that the power cord can  be unplugged  or use a power strip where the plug can quickly  be removed from the outlet in the event of an emergency     When you first place the system in Standby  internal  components continue to operate for almost one minute while  performing file maintenance and preparing the system for  possible power removal      N CAUTION    Removing the AC mains power cord before the file maintenance is  complete might corrupt files on the internal hard driv
15.   s internal hard  drive     6 2 1 3 Manage Files   Choose this command to open a window from which you can  archive or delete method and run files from the current  directory  You can archive files by connecting a USB Flash  drive and selecting the Copy or Move files to flash drive  option     To remove a file from the internal hard drive  highlight the file  or enter the filename and click the Delete button     To remove multiple files  click the Delete by Age button  Then  specify the age limit in months and click OK     Use the Move and Delete options with caution as the files  cannot be recovered from the internal hard drive once the  action is complete     6 2 1 4 Set Data Path      This opens the SET DATA PATH  6 3 7  window  Use this window  to specify the default storage folders for users  These folders  contain multiple METHOD FILE  6 1 2  and RUN FILE  6 1 3   resources created by system users   See also USER  MANAGEMENT  6 3 12       6 2 1 5 Save Method   Choose this menu command to save the settings of the current  method file to the system s internal storage  The file will be  saved under the current file name     Note  DEFAULT METHODS  6 1 2 1  are an exception and cannot  be overwritten by the Save Method command  If a default  method is open and you attempt to save any modifications to       6 4    Section 6 User Help Reference    that method  the FILES  6 3 6  window appears so you can  rename the method file  This preserves the default method  If  you do wi
16.   the Main window displays peak  collection data as well as the settings used for that run  The  Main window dynamically changes to display the controls  required for the current state of the system     The top region of the Main window always displays a  GRADIENT PLOT  6 3 1 1  area  The bottom region of the window  displays the RUN SETTINGS  6 3 1 2   RUN CONTROL BUTTONS   6 3 1 3   and PEAK COLLECTION BUTTONS  6 3 1 4   After a run   the left side of the window changes to show the collection rack  map or the method parameters in a PEAK COLLECTION DATA   6 3 1 5  pane     Note  The main window is expanded for CombiFlash Rf 4x  systems  The main window the displays the settings and views  in four tabs  one for each column channel  A fifth tab     BATCH   6 3 5      contains the controls to run sequential separations     6 3 1 1 Gradient Plot   The gradient plot area depicts the current gradient that will  be used for the run  The X axis of the plot displays the run  time in minutes or COLUMN VOLUMES  oN LINE   The X axis scale  can be adjusted by typing or selecting the Run Length  The  left Y axis displays absorbance units and the right Y axis  indicates the percentage of solvent B  By default  PeakTrak  automatically scales the left Y axis to best display the       6 10    Section 6 User Help Reference    absorbance trace  PeakTrak also sets the right Y axis to 100    You can override the Y axes scales by clicking the Format  Graph button and setting the desired values    
17.  13  3 10 2 General Settings                   005  3 13  3 10 3 Set Date Time                 00000005 3 13  3 10 4 Vapor Limit  Optional                    3 13  3 10 5 Set Default Tube Volumes                3 14  3 10 6 User Preferences             00000 ee eee 3 14  3 10 7 Restart the system                 005  3 17  3 11 Prime the Solvent Lines                       3 18  3 12 System Verification              0 0000 RR KK 3 19  3 13 Installation Qualification Checklist              3 20  4  Preparation   Optional 4x Module  4 1 Unpacking the Unit                     sees  4 1  4 2 Instrument Location                0 00 cee eee 4 2  4 8 Connect Power                  cece eee eee 4 2  4 4 Connect Solvent Lines               0 00000 4 3  4 5 Connect Waste Lines                  RR KK 4 4  4 6 Control Connection               0 0c RR RR KK 4 11  4 7 Position the System                    RR KK 4 11  4 8 Remove Shipping Hardware                   4 11  4 9 Turn on Power           kk KK KK eee ees 4 12  4 10 Prime the Solvent Lines                       4 13  4 11 System Verification              0 20000 eee eee 4 14  4 12 Installation Qualification Checklist             4 15  5  Operation  5 1 Sample Preparation                  000 KK 5 1  5 1 1 Liquid Sample Injection                   5 1  5 1 2 Solid Samples                     RR KK 5 1  5 1 8 Preloading on Column                    5 5       Table of Contents    5 2 Loading a RediSep Rf Column                    5 3
18.  3 4     command instead        6 14    Section 6 User Help Reference      Save As     Click this button to open the FILES  6 3 6   window  From this window you can rename the current  method and save it on the system   s internal storage       Exit     Click this button to close the Method Editor and  return to the MAIN WINDOW  6 3 1        Gradient Optimizer     Click this button to open the  GRADIENT OPTIMIZER  6 3 3  window  This feature  produces an optimized gradient optimized based on your  TLC retention factor data       Column Data      Rf200  Rf200i  and RfAX  Click this  button to open the COLUMN DATA  6 3 4  window  This  window reports information about the column installed  in the system       Time to CV   CV to Time     Click this button to toggle  the Run Length  X axis  units between minutes and  COLUMN VOLUMES  oN LINE   You can configure  Peak Trak s default Run Length units by changing the  CONFIGURATION  6 3 10  settings     6 3 2 2 Run Settings      RediSep Column      Rf200  Rf200i  and Rf4X  The  RediSep column media and size are detected by the  system and reported here  The column should match the  one that will be used for the run  You may override the  detected column size by selecting a different size  See  also  REDISEP RF COLUMN SELECTION GUIDE  oN LiNE  and  ABOUT REDISEP RF COLUMN DETECTION  ON LINE        Sample Name     The sample name is a text entry box  in which you can label the run  Type a unique descriptor   PeakTrak saves the run info
19.  4  TCP IPv4  and click the  Properties button  The PC displays the Internet Protocol  Version 4  TCP IPv4  Properties window  Figure 7 2      M Note   If the Internet Protocol Version 4  TCP IPv4  option is not listed   click the Install button to add it  If assistance is needed  press the  F1 keyboard button to display Windows Help     The following steps change the local area connection settings  If  the PC must be restored to its original network configuration   record the present Internet Protocol  TCP IP  settings now     5  Select the    Use the following IP address  option   6  Enter the IP address  192 168 1 200    7  Enter the Subnet Mask     255 255 255 0        8  Click the OK button to close the Internet Protocol Version 4   TCP IPv4  Properties window  Click the OK button to close  the Local Area Connection properties window     9  Shut down the PC and place the CombiFlash Rf system in  Standby        CombiFlash   Rf User Manual       Networking          Connect using   V VMware Accelerated AMD PCNet Adapter          This connection uses the following items   1E Client for Microsoft Networks   8  QoS Packet Scheduler    v   v   IV  dB File and Printer Sharing for Microsoft Networks  IV   4  Intemet Protocol Version 6  TCP IPv6   a n  v  v                                    Intemet               Link Layer Topology Discovery Mapper I O Driver    4  Link Layer Topology Discovery Responder                               JOa     Transmission Control Protocol Intemet 
20.  8  The post run valve cleaning fluids share a common port for  channels 2 through 4  Select the 48 inch  120 cm  length of  black tubing from the accessory package and connect it to  the Valve Waste Out bulkhead fitting on the Rf 4x module   Route the other end of this tubing to one of the waste  bottles   The channel 2 waste bottle is shown in Figure 4 8    You can alternatively route it to a container other than  those used for channels 2 through 4  however the system  will not be able to detect the level and prevent overfilling        4 9    CombiFlash   Rf User Manual        ae       To Waste  Bottle 2    Figure 4 8 Valve Waste Out routed to Channel 2    When using higher density solvents such as dichloromethane   DCM   ensure that waste containers are no more than 3 feet  below the supply solvent levels  Placing solvent containers above  the system and waste containers on the floor may cause the  internal check valves to open and allow solvent to flow through the  system when in the standby state     M Note  The Hf 4x back panel ports labeled A and B are not used        Section 4 Preparation  Optional 4x Module    4 6 Control Connection    Operation of the Rf 4x module is controlled by the CombiFlash  Rf system via a USB connection  Connect the USB cable from  the Rf 4x module to USB port on the system back panel     4 7 Position the System    After completing the fluid  air  and control connections  the  system can be moved to its operating position      N CAUTION    U
21.  8 Maintenance    If the numerical values of the Raw Lamp Energy  Figure 8 5  do  not change  or if the first number remains at zero  contact  Teledyne Isco s Technical Service department     Table 8 1 Suggested Flow Cell Cleaning Solvents           Dichloromethane   DCM     Solvent Description   Methanol Miscible with most LC solvents  Methanol will force other  solvents through the system so they will not interfere with  the next sample  It breaks down the tertiary structure of  silica and allows it to flush through    Acetone Common organic solvent with high dissolving properties     Acetone absorbs UV light at 254 nm  Therefore  the Raw  Lamp Energy gauge will always read red when acetone is  in the flow cell  After pumping acetone  pump a different  solvent through the system to remove the acetone and  read the Raw Lamp Energy gauge     Effective at solubilizing organic molecules        Dimethylsulfoxide   DMSO     Strong solvent that will clear organic sample        Dimethylformamide   DMF     Strong solvent if other solvents do not produce the  desired results                    aqueous solution       Tetrahydrofuran Strong solvent     THF    Water DMF  A 50 25 25 solvent mixture effective in clearing some  methanol obstructions    Water DMF  30 30 30ml solvent mixture effective in clearing some  Acetic Acid obstructions    Mildly acidic 0 5M acetic acid in distilled water can be used to clear  aqueous solution samples that are not removed by organic solvents   Mildly 
22.  An automatic run extension is a five minute isocratic hold  added to the end of the run   s maximum  B gradient profile   During the extension  the system continues to pump the  maximum  B solvent mixture  Should the system still detect  a peak after an extension  the system will add another  up toa  maximum of three extensions     Occasionally  compounds might come off the column once the   B returns to the minimum value at the end of a run  If the  Run Length Extension is enabled  it will automatically extend  the run one time to clear the column and plumbing of any  remaining material     Extension 1 Extension 3                   Minutes    Figure 3 9 Automatic Run Extension Examples    Figure 3 9 illustrates the possible run extensions  The original  method was programmed to rise to the maximum  B over  seven minutes  hold for two minutes and return to the  minimum  B for a final minute  At nine minutes  a peak was  eluting  The system extended the run  holding the  B at the  maximum level for another five minutes  This also occurred at  fourteen and nineteen minutes  resulting in the second and  third extensions  At 24 minutes  the  B solvent strength  returned to zero  Before the final minute elapsed  more  compound was detected  causing the system to extend the run  for a final five minutes at the minimum  B        3 16    Section 3 Preparation  Rf 150 Systems    Enable rapid equilibration     select this option to  equilibrate the column at a high flow rate     M Not
23.  B     The gradient plot area also provides a convenient method to  edit the gradient curve  You can click and drag any of the  points to change the shape of the curve  or use the buttons at  the bottom of the plot area to insert and delete points  The  following controls appear in the Gradient Plot Area       Insert     Click this button to enable the gradient point  insert mode  When this mode is active  click the gradient  curve to add a single point  You can then drag the new  point to any desired position  Click the insert button  once for each gradient point that must be added to the  plot area       Delete    Click this button to enable the gradient point  delete mode  When this mode is active  the system will  delete the point nearest the next click on the gradient  curve  Click the Delete button once for each point that  must be deleted from the plot area       Run Length     Type or select the length of the run   Note that this will change the scale of the X axis on the  gradient  Points that define the gradient will be  automatically scaled to fit the new run length  The run  length can be expressed in minutes or COLUMN VOLUMES   on LINE   CV   You can configure PeakTrak   s default Run  Length units by changing the CONFIGURATION  6 3 10   settings       Solvent A     Select the desired solvent from the list   The available solvents are those defined by the  CONFIGURATION  6 3 10  settings       Solvent B     Select the desired solvent from the list   You can progr
24.  Column buttons  Click the Stop button to halt the air purge     6 4 10 3 Raw Lamp Energy   The right side of the Manual Control window shows a Raw  Lamp Energy gauge  This provides an indication of the UV  light measured by the optical detection system at 254 nm   High lamp energy  green  means that the flow cell easily  passes through a sufficient UV light source  Lower lamp  energy  yellow or red  means that either the light source is  weak or that the flow cell is obstructed       Green   lamp energy is sufficient to detect peaks up to  2 4 Absorbance Units       Yellow   lamp energy is partially obstructed and the  peak detector performance might be impaired  The  system can be operated although the absorbance trace  might appear noisy  Yellow might be a normal indication  for systems with flow cell windows stained by  compounds       Red   lamp energy is obstructed to a degree that the  system might not reliably detect peaks  If you attempt to       6 50    Section 6 User Help Reference    operate the system  peak collection will be forced to  collect all  This prevents diverting desired compounds to  waste     Depending on what you are doing with the Manual Control  functions  low lamp energy could be a normal indication  For  example  a UV absorbing compound could be present in the  flow cell as you are pumping solvent  Or  the selected solvent  absorbs UV light at 254 nm  Abnormal indications could be a  flow cell blocked by a compound that has precipitated or there 
25.  Files to Different Column Sizes   If you want to scale a method file to a different column size    use the FILE gt Save As menu command  From the FILES  6 3 6   window  select the Scale option  then choose the new column  size and click the Save button  The system will create a new  method file with the gradient and run parameters scaled to   the new column size     This feature is useful when developing methods using small  sample volumes  After you optimize the method to successfully  extract the compounds of interest  you can then scale up the  method to load and extract larger volumes using larger  RediSep column sizes     6 4 4 Editing a Default Method  Default Methods allow you to quickly recall standard method  settings  There are standard methods for each size and type of  RediSep column  When shipped from the factory  the Default  Methods were set to the settings recommended by Teledyne  Isco     The Default Method settings can be tailored to your  application  To modify the Default Methods     1 Select the TOOLS gt EDIT DEFAULTS  6 2 3 4  menu command    2 If prompted for a password  enter the system password    The SET PASSWORD  6 3 11  window controls the password  protection feature      3 Select the column size and media from the RediSep Column  list     4 Change the settings to meet your specific application  requirements  Refer to the EDITING A METHOD  6 4 2   instructions     5 Click the Save button  or File gt Save  to replace the default  method settings  
26.  Install Collection Tube Racks    Before beginning a run  you must load collection racks with  tubes onto the system   s fraction collector tray     Your system was shipped with two collection tube racks  The  following tube rack sets are available       60 5237 013     Two racks for 13 x 100 mm test tubes   8 mL   Total tubes  216       60 5237 061     Two racks for 16 x 100 mm test tubes   14 mL   Total tubes  150        Section 2 Preparation  Rf 200 and Rf 200i Systems    60 5237 031     Two racks for 16 x 125 mm test tubes   15 5 mL   Total tubes  150     60 5237 032     Two racks for 16 x 150 160 mm test  tubes  18 mL   Total tubes  150     60 5237 033     Two racks for 18 x 150 mm test tubes   25 mL   Total tubes  140     60 5237 034     Two racks for 18 x 180 mm test tubes   80 mL   Total tubes  140     60 5237 035     Two racks for 25 x 150 mm vials  50  mL   Total vials  60     60 5237 040     One rack for twelve 480 mL French  square bottles     e 60 5394 469     Two racks for 20 mL  28 x 61 mm   scintillation vials  Total vials  54       60 5394 468     Two racks for 40 mL  28 x 95 mm   scintillation vials  Total vials  54     To load the racks   1  Insert test tubes  vials  or bottles into the rack  Fig  2 6      Z  CAUTION    Risk of broken glass or equipment damage  Do not load test tubes  longer than the length listed on the tube size label        Figure 2 6 Loading test tubes       2 13    CombiFlash   Rf User Manual    2  While holding the rack with the tube 
27.  MANAGEMENT  6 2 3 7        6 32    Section 6 User Help Reference      HELP gt SERVICE  6 2 4 2     TOOLS gt SET PASSWORD  6 2 3 6   This protects the current  password     6 3 12 User Management    Use the TOOLS gt USER MANAGEMENT  6 2 3 7  menu command to  open a window from which you can add and remove system  users  When you add a user  the system will create a user  folder that becomes the data path for that user  User  management  in conjunction with this data path feature   conveniently lets you store your working files apart from other  system users     To add a user  click the Add New button and enter the user  name  To delete a user  click the Delete button next to the  user s name and confirm the action     See also  SET DATA PATH  6 3 7     6 4 Examples of PeakTrak actions    In this section you will find instructions for typical PeakTrak  tasks  such as     e STARTING A RUN USING DEFAULT SETTINGS  6 4 1      EDITING A METHOD  6 4 2    e ALTERNATIVE WAYS TO CREATE METHOD FILES  6 4 3   e EDITING A DEFAULT METHOD  6 4 4      DEFINING A GRADIENT  6 4 5      REAL TIME GRADIENT EDITING  6 4 6      BYPASSING THE SOLID LOAD CARTRIDGE  6 4 7      MONITORING THE WAVELENGTH RATIO  6 4 8    e VIEWING RUNS  6 4 9      MANUAL CONTROL OF THE COMBIFLASH RF  6 4 10   e SETTING THE VAPOR LIMIT  6 4 11     6 4 1 Starting a run using default settings    After the CombiFlash Rf system has been set up and  automatically primed  See AUTO PRIME VALVE WASH  6 3 8          6 33    CombiF
28.  Network Type     Select the    Static IP  option     This option will require an IP Address  Netmask  and  Gateway provided by your network administrator        P Address     Enter the static IP Address for the  system       Netmask     Enter the Netmask for the system      Gateway     Enter the Gateway for the system   Network Printing The system can print to a network printer  and supports JetDirect and line printer  LPR  queues   Consult with your network administrator to determine      The IP address of selected printer     e Ifthe printer uses an LPR print queue  you must also  find out the queue name     e Ifthe printer uses a JetDirect print queue  also ask  for the port number       The printer type  either Postscript or PCL     When this information is known  you can proceed with  configuring system for network printing        6 30    Section 6 User Help Reference    Follow the on screen instructions for entering the address   queue  and printer type information  After successfully  printing a test page  the network printer will be available for  printing using the FILE gt PRINT     6 2 1 12  menu command and  the AUTOMATICALLY PRINT REPORT AT END OF RUN feature     The Network printing controls may also be configured to print  to a local USB printer connected the system s back panel USB  port  The printer must support postscript printing  To  configure the local printer  select Postscript for the Printer  Type  Then  type the following connection URL    usb  dev us
29.  Rf system     1 6 Safety    Before installing  operating  or maintaining this equipment  it  is imperative that all hazards and preventive measures are  fully understood  While specific hazards may vary according to  location and application  take heed in the following general  warnings     A  WARNING    Avoid hazardous practices  If you use this instrument in any  way not specified in this manual  the protection provided by  the instrument may be impaired     A WARNING    Liquids associated with this instrument may be classified as  carcinogenic  biohazard  flammable  or radioactive  Should  these liquids be used  it is highly recommended that this  application be accomplished in an isolated environment  designed for these types of materials in accordance with  federal  state  and local regulatory laws  and in compliance with  your company   s chemical hygiene plan in the event of a spill     A  WARNING    If you are using flammable solvents or chemicals with this  system  vapor concentration levels may exceed the maximum  exposure levels as recommended by OSHA Guide 1910 1000   To reduce those levels to a safe exposure  Teledyne Isco  recommends that you place the system in a laboratory hood  designed for the purpose of ventilation  This hood should be  constructed and operated in accordance with federal state and  local regulations  In the event of a solvent or chemical spill   your organization should have a plan to deal with these  mishaps  In all cases  use good labora
30.  Risk of broken glass or equipment damage  Do not load test tubes  longer than the length listed on the rack label        2  While holding the rack with the rack label visible  Figure  3 7   insert the racks into the system  Slide the rack in  until you feel it drop into its seated position        Rack  Label             Figure 3 7 Inserting the collection tube rack       3 11    CombiFlash   Rf User Manual        CAUTION    An incorrectly installed rack will cause the rack to be misaligned  under the fraction collector arm  Misaligned racks might cause  fractions to miss the tube opening or deposit in the wrong tube   Always ensure the tube size label is visible  that is  facing outward   and the rack is pushed in until it is seated     3 9 Turn on Power    The system   s power switch is located just below the touch  screen panel     1  Turn the switch to the ON position  The system will begin  its startup routine which includes self diagnostics     2  Monitor the movement of the fraction collector arm  The  arm should move to the left rear corner of the fraction  collector area      NCAUTION    You should not hear any grinding noises during the homing  sequence  If grinding is heard there is a problem with the homing  of the fraction collector arm  Contact the factory for assistance     3  The system is ready for operation when the PeakTrak  screen is displayed     3 10 Configure the System    Before operating the system  it should be configured for the  desired operation  
31.  Section 2 Preparation  Rf 200 and Rf 200i Systems    of level bench space  Ensure that the module has at least 3 cm   1 25   of air space behind it for ventilation  Instructions for  installing the 4x module can be found in Section 4 of this  guide     Refer to Table 1 1 for environmental conditions and power  requirements     Z WARNING    The system is heavy  Use a two person lift to prevent injury     Before making any connections to the Combiflash Rf system   place it on the bench or in the fume hood where it will be  operated  Temporarily position the system so you can access  the back panels to complete the connections     2 3 Connect Power    Ensure that the On Standby switch below the touch screen  panel is in the Standby position  Then  use the supplied IEC  power cord to connect the system to mains power       A WARNING    Mains power must meet the voltage  frequency  and amperage  requirements listed on the serial number label     A  WARNING    As long as the AC mains power cord is connected  power is  inside the unit  The mains power cord is the disconnect device   Position the system so that the power cord can be unplugged   or use a power strip where the plug can quickly be removed  from the outlet in the event of an emergency     2 4 Connect Solvent Lines    A  WARNING    Risk of fire ignited by electrostatic discharge  Never substitute  the black tubing on CombiFlash systems  The black tubing   P N 023 0503 06  is conductive  This tubing is required to  dissip
32.  Start a Default Method                    LL   5 4 During the Run            KK KK KK RR RR RR KK eee  5 4 1 Run Control Buttons                      5 4 2 Bypassing the Solid Load Cartridge          5 4 3 About Solvent Level Detection                6  User Help Reference   6 1 PeakTrak Overview             0 KK RR KK KK  6 1 1 PeakTrak Window Elements                6 1 2 Method File             0 0 0 0    KK KK eee  6 L9 Run  Bie  i  3an KA nade bee CEES SR ME kk  6 1 4 How to get PeakTrak help                   6 2 PeakTrak Menu Options                  0005   22 T Elle Lore tact aden ous A RE d  6 2 2 Method Editor                 0 0 20  6 2 3  TOONS 5 gag pues ein co lS Ale Whee CORE epos  6 24 Help  uote alee a n   eens cues de as   6 3 PeakTrak Windows              00000 RR KK K    6 3 1 Main Window             0 0000 RR RR KK  6 3 2 Method Editor                 0 002000  6 3 8 Gradient Optimizer                  05   6 3 4 Column Data              0    KK KK  6 3 5 Batch   eee RR ve ae RA  6 3 6 Eleg  ace ced rese oe Pals Rd S res  6 3 7 Set Data Path             RR  6 3 8 Auto Prime Valve Wash                    6 3 9 Manual Control               KK  6 3 10 Configuration             llle   6 3 11 Set Password            0    eee eee eee  6 3 12 User Management                sse   6 4 Examples of PeakTrak actions                   6 4 1 Starting a run using default settings         6 4 2 Editing amethod                   005   6 4 3 Alternative Ways to Create Met
33.  The Cartridge Bypass button may be used during a run to  relieve backpressure caused by impurities precipitating in the  solid load cartridge  Use this feature with care to avoid leaving  desired compound on the cartridge  Read more about this  feature in BYPASSING THE SOLID LOAD CARTRIDGE  6 4 7     If the external detector option is enabled on Rf 200 systems   the plot area will include a secondary left Y axis to display the  analog input voltage trace  Likewise  if the evaporative light  scattering detector  ELSD  is enabled on an Rf 200i system  a  secondary left Y axis is included     The gradient plot area also provides a convenient method to  edit the gradient curve  You can click and drag any of the  points to change the shape of the curve and use the buttons at  the bottom of the plot area to insert and delete points  For  complete instructions on defining gradient curves  see  DEFINING A GRADIENT  6 4 5      To zoom in on part of the plot area  click the Zoom In button to  enable the zoom action and then drag across the area of  interest  Click the Zoom Out button to revert back to the  normal display     6 3 1 2 Run Settings      RediSep      Rf200  Rf200i  and Rf4X systems  The  system reports the detected RediSep column media and  size  The column should match the one that will be used  for the run  You may override the detected column size  by selecting a different size or media  See also  REDISEP  RF COLUMN SELECTION GUIDE  oN LINE  and ABOUT  REDISEP RF COLUM
34.  The file name will be the same as the Run file  except  with a    TXT    file name extension  When selecting this  command from a remotely connected computer via a web  browser  PeakTrak displays a file download window so you can  select a location and enter a file name      You can open  edit  and print TXT files with text editing or  word processing software         6 5    CombiFlash   Rf User Manual    6 2 1 9 Import Method   Choose this command to load a Method file from an external  source  see EXPORT METHOD  6 2 1 10   onto the system s  internal storage  From the touch screen  this command opens  the FILES  6 3 6  window so you can locate and select the file on  a USB Flash storage device  When selecting this command  from a remotely connected computer  PeakTrak displays a  window so you can select the method file  After selecting the  file  click the Open button to save the method file on the  system s internal hard drive     6 2 1 10 Export Method   Choose this command to save a Method file in a location other  than the system s internal hard drive  This can be done to  archive method files or to transfer the method to another  system using the IMPORT METHOD  6 2 1 9  menu command   From the touch screen  this command opens the FILES  6 3 6   window so you can save the file on a USB Flash storage device   When selecting this command from a remote computer   PeakTrak displays a window from which you can save the  method file     6 2 1 11 Export SMM XL Method   Choose
35.  Zone   Select your time zone from the drop down list  box     3 10 3 Set Date Time    Click this button to open a window from which you can set the  system date and time     To prevent unauthorized time changes  this feature is  password protected  The system is shipped with the password  set to combiflash  Use the Tools gt Set Password menu  command to change this password for greater security     3 10 4 Vapor Limit  Optional     The system has internal vapor The sensor detects vapors  present in the immediate atmosphere  When the limit is  exceeded  the CombiF lash Rf will shut down to avoid a  hazardous condition     Teledyne Isco recommends using the default setting of 25   This value represents an organic vapor level slightly above the  ambient vapor level of a well ventilated laboratory     A setting below 25 could cause random alarms without  significant vapors present  If PeakTrak displays a Vapor Limit       3 13    CombiFlash   Rf User Manual    alarm while the Vapor Limit is set at 25  perform the following  checks on your laboratory and the instrument     e Ensure that no open containers or spills of organic  solvent are in close proximity to the system     e Ensure that the system is located in a well ventilated  area     e Ensure that there is no visible solvent leakage from the  system     If PeakTrak continues to display the Vapor Limit alarm after  you have made these checks and corrected any problems  found  it is likely that organic vapors are present in t
36.  and is collected     An automatic run extension is a five minute isocratic hold  added to the end of the run s maximum 4B gradient profile   During the extension  the system continues to pump the  maximum  B solvent mixture  Should the system still detect  a peak after an extension  the system will add another  up to a  maximum of three extensions     Occasionally  compounds might come off the column once the   B returns to the minimum value at the end of a run  If the  Run Length Extension is enabled  it will automatically extend  the run one time to clear the column and plumbing of any  remaining material     Original method   Extension 2    Final Extension  Extension 1 Extension 3                                              15 20 25   Minutes  Figure 2 9 Automatic Run Extension Examples       Figure 2 9 illustrates the possible run extensions  The original  method was programmed to rise to the maximum  B over  seven minutes  hold for two minutes and return to the       2 19    CombiFlash   Rf User Manual    minimum  B for a final minute  At nine minutes  a peak was  eluting  The system extended the run  holding the  B at the  maximum level for another five minutes  This also occurred at  fourteen and nineteen minutes  resulting in the second and  third extensions  At 24 minutes  the  B solvent strength  returned to zero  Before the final minute elapsed  more  compound was detected  causing the system to extend the run  for a final five minutes at the minimum  B     Enabl
37.  below the level  of the Rf can contribute to decreased flow due to the high vapor  pressure of DCM  This problem becomes more pronounced as the  ambient temperature increases        CombiFlash   Rf User Manual    3 8 Install Collection Tube Racks    Before beginning a run  you must load collection racks with  tubes onto the system   s fraction collector tray     Your system was shipped with two collection tube racks  The  following tube rack sets are available     60 5237 013     Two racks for 13 x 100 mm test tubes   8 mL   Total tubes  216     60 5237 061     Two racks for 16 x 100 mm test tubes   14 mL   Total tubes  150     60 5237 031     Two racks for 16 x 125 mm test tubes   15 5 mL   Total tubes  150     60 5237 032     Two racks for 16 x 150 160 mm test  tubes  18 mL   Total tubes  150     60 5237 033     Two racks for 18 x 150 mm test tubes   25 mL   Total tubes  140     60 5237 034     Two racks for 18 x 180 mm test tubes   30 mL   Total tubes  140     60 5237 035     Two racks for 25 x 150 mm vials  50  mL   Total vials  60     60 5237 040     One rack for twelve 480 mL French  square bottles     60 5394 469     Two racks for 20 mL  28 x 61 mm   scintillation vials  Total vials  54     60 5394 468     Two racks for 40 mL  28 x 95 mm   scintillation vials  Total vials  54        3 10    Section 3 Preparation  Rf 150 Systems    To load the racks   1  Insert test tubes  vials  or bottles into the rack  Fig  3 6         Figure 3 6 Loading test tubes      A CAUTION   
38.  by Tele   dyne Isco        7 2    Section 7 Remote Interfaces    To access the system  simply open an Internet Browser and  enter the IP address as    http   __       _     where the  blanks are replaced by the selected address  The browser  window will load PeakTrak after you enter the address     7 1 2 Network Printing  The CombiFlash Rf system can print to a network printer and  supports both JetDirect and line printer  LPR  queues     Consult with your network administrator to determine the IP  address of selected printer  If the printer uses an LPR print  queue  you must also find out the queue name  If the printer  uses a JetDirect print queue  also ask for the port number   When this information is known  you can proceed with  configuring system for network printing     1  Select the Tools   Configuration menu command   2  Select the Network Configuration tab     3  Follow the on screen instructions for entering the address  and queue information     4  Click OK to save the settings  The Combiflash Rf system  will send a test page to the printer     After successfully printing a test page  the network printer  will be available for printing using the File gt Print menu  command or through the Automatically Print Report at End of  Run option on the Tools gt Configuration User Preferences tab     7 1 3 Network File Save Configuration   By default  this option is disabled and post run reports are  automatically saved internally on the CombiFlash Rf system   You can recal
39.  complete the run  the run continues  If  not sufficient  the system displays the alert message  again       Ignore     This option is available when the estimated  usage rates indicates that the solvent levels are  insufficient  Select Ignore to clear the message and  continue the run  The system then ignores the estimated  solvent usage rate for the remainder of the run  but  continues to measure the solvent level  Should the level  become insufficient  the system will display another  alert message       Disable     Select this option to turn off solvent level  measurements and estimated solvent usage rates for the  remainder of the run        5 14    Section 5 Operation      Cancel   This option appears before the run is started   Select this option to cancel the run and return to the  main window which will display the current method        5 15    CombiFlash   Rf User Manual       CombiFlash   Rf  User Manual       Section 6 User Help Reference    The CombiFlash Rf system includes User Help topics that are  available from the PeakTrak touchscreen or remote browser  window  This section contains many of these topics that  describe the PeakTrak interface and its use     On line Software Help benefits from linked text to navigate  through topics of interest   a feature not available in printed  manuals  For your convenience  linked text is displayed in  SMALL CAPITALIZED TEXT  followed by the section number in  brackets  Some topics are only available in the on line help  
40.  containers are filled  then     1  Select the    Tools gt Auto prime    menu command   2  Click the Play button to start priming the system     The system pumps 100  Solvent B through the system and  out the Waste port  and then repeats with 100   Solvent A    This B A order leaves the internal lines filled with solvent  appropriate for the beginning of a purification run      When priming the first time  inspect the solvent and waste  connections to the system  If any leaks are observed  click the  Cancel button to stop the Auto Prime  Correct the leak by  tightening the fitting an additional 1   4 turn  and then restart the Auto  Prime from step 1     To advance to the next Auto Prime step  you can click the  Fast Forward button  The Auto Prime window closes when  finished     After Auto Priming  the system is ready for operation   Section 5      Use the Auto Prime feature for quick solvent changes by  pumping a fixed amount of solvent through the waste fluid  path at a high flow rate  For more advanced priming functions  such as varying the flow rates  fluid paths  etc   use the  Tools gt Manual Control option        3 18    Section 3 Preparation  Rf 150 Systems    3 12 System Verification  It is recommended that the system operation be verified  To do  so  use a pre packaged CombiFlash Rf test kit       Part number 60 5237 050  Test kit containing five of the 5 gram size solid load  cartridges pre filled with 0 5 grams of test sample  NPHE  and five 4 gram RediSep Rf sil
41.  de las  instrucciones del manual que acompa  an a este    producto              Esta se  al alerta sobre la presencia de alto voltaje en el  interior del producto        Punto del machacamiento  Sus dedos o manos seriusly  A ser  n da  ados si usted los coloca entre las piezas    m  viles cerca de estos s  mbolos                 1 7 For Additional Information    Technical assistance for the CombiFlash Rf system can be  obtained from     Teledyne Isco  4700 Superior St   Lincoln NE 68504    Phone   800  775 2965 or  402  464 0231  Fax   402  465 3001  E mail  IscoService teledyne com       1 22    CombiFlash   Rf  User Manual       Section 2 Preparation   Rf 200 and Rf 200i Systems    This section provides instructions for unpacking and installing  the CombiFlash Rf 200 and Rf 200i systems  To prepare a  system for operation  sequentially follow all instructions in  sections 2 1 through 2 13     M Note   Section 2 15 contains an Installation Qualification checklist  If  required  sign off the checklist entries as you successfully  complete the following sections     2 1 Unpacking the Unit    The CombiFlash Rf 200 or 200i system is shipped in a single  carton  The optional 4x expansion module is shipped in a  second carton  Carefully unpack the shipment and inspect the  contents      NWARNING    The system is heavy  Use a two person lift to prevent injury      NCAUTION    Do not lift the system by the fraction collector arm  Use the lifting  handles located on the instrument s
42.  is a film built up on the flow cell  If you suspect the low lamp  energy is an abnormal indication  perform the steps in the  QUICK CLEANING WHEN RECOMMENDED  ON LINE  topic     6 4 11 Setting the Vapor Limit   Rf200  Rf200i  and Rf4X  The system has an internal vapor  sensor that detects vapors present in the immediate  atmosphere  When the limit is exceeded  the CombiFlash Rf  will shut down to avoid a hazardous condition     Teledyne Isco recommends using the default setting of 25   This value represents an organic vapor level slightly above the  ambient vapor level of a well ventilated laboratory     A setting below 25 could cause random alarms without  significant vapors present  If PeakTrak displays a Vapor Limit  alarm while the Vapor Limit is set at 25  perform the following  checks on your laboratory and the instrument       Ensure that no open containers or spills of organic  solvent are in close proximity to the system       Ensure that the system is located in a well ventilated  area       Ensure that there is no visible solvent leakage from the  system     If PeakTrak continues to display the Vapor Limit alarm after  you have made these checks and corrected any problems  found  it is likely that organic vapors are present in the  ambient environment of your laboratory  In this case  increase  the setting of the Vapor Limit  Settings greater than 25 are  appropriate for laboratory environments with a somewhat  elevated background solvent vapor concentration   rep
43.  laboratory has more than one Combiflash Rf system     Time Zone   Select your time zone from the drop down list  box     2 12 3 Set Date Time  Click this button to open a window from which you can set the  system date and time     To prevent unauthorized time changes  this feature is  password protected  The system is shipped with the password       Section 2 Preparation  Rf 200 and Rf 200i Systems    set to combiflash  Use the Tools gt Set Password menu  command to change this password for greater security     2 12 4 Vapor Limit  The system has an internal vapor sensor that detects vapors  present in the immediate atmosphere  When the limit is  exceeded  the Combiflash Rf will shut down to avoid a  hazardous condition     Teledyne Isco recommends using the default setting of 25   This value represents an organic vapor level slightly above the  ambient vapor level of a well ventilated laboratory     A setting below 25 could cause random alarms without  significant vapors present  If PeakTrak displays a Vapor Limit  alarm while the Vapor Limit is set at 25  perform the following  checks on your laboratory and the instrument     e Ensure that no open containers or spills of organic  solvent are in close proximity to the system     e Ensure that the system is located in a well ventilated  area     e Ensure that there is no visible solvent leakage from the  system     If PeakTrak continues to display the Vapor Limit alarm after  you have made these checks and corrected any probl
44.  selection valve and internal  components     M Note   When using higher density solvents such as dichloromethane   DCM   place the solvent container level with or above the  CombiFlash Rf system  Placing solvent containers below the level  of the system can contribute to decreased flow due to the high  vapor pressure of DCM  This problem becomes more pronounced  as the ambient temperature increases        3 5    CombiFlash   Rf User Manual    3 5 Connect Waste Lines     A WARNING    Risk of fire or equipment damage  Failure to connect Waste  Port tubing may allow organic solvents to pool in unsafe areas   possibly creating dangerous levels of flammable vapors     Z WARNING    Risk of fire ignited by electrostatic discharges  Never  substitute the black tubing on CombiFlash systems  The black  tubing  P N 023 0503 06  is conductive  This tubing is required  to dissipate static electricity      NCAUTION  Elevated flammable vapor levels are possible  Ensure that the  waste container is adequately ventilated  preferably by placing it in  a fume hood        CAUTION    Do not overfill the waste collection containers  Regularly monitor  and empty the bottles before they overfill     The system has one waste outlet on the back panel  Diverter  Valve waste     Accessory package  P N 60 5239 009  includes a Waste Tube  Assembly  P N 60 5234 187   To connect the Waste tube     1  Connect the cut end of this assembly to the Diverter Valve  Waste port on the system     a  locate the blac
45.  solvent connections     1 Locate the ABCD label in accessory package 69 5234 008   Affix this label to the system s left side panel  just above  the solvent ports  Figure 4 1         A B C D                      Figure 4 1 Labeled ports on CombiFlash Rf system side panel    2  Remove the jumper tubing from ports A  B  C  and D     3  Locate the solvent lines on the side of the Rf 4x module   These lines are labeled A through D  Connect the solvent  lines to their respective ports on the CombiFlash Rf  system        4 3    CombiFlash   Rf User Manual    4 5 Connect Waste Lines     A WARNING    Risk of fire or equipment damage  Failure to connect Waste  Port tubing may allow organic solvents to pool in unsafe areas   possibly creating dangerous levels of flammable vapors     A WARNING    Risk of fire ignited by electrostatic discharges  Never  substitute the black tubing on CombiFlash systems  The black  tubing  P N 023 0503 06  is conductive  This tubing is required  to dissipate static electricity      N CAUTION    Elevated flammable vapor levels are possible  Ensure that the  waste container is adequately ventilated  preferably by placing it in  a fume hood     The Rf 4x module has six liquid waste ports on the back panel   Figure 4 4        one inlet to receive diverter valve waste from the  CombiFlash Rf system labeled Column Waste  From RF       four waste outlets    one for each column channel   labeled Column Waste 1  2  3  and 4       one Valve Waste outlet shared by 
46.  system uses air or  user supplied gas to measure the hydrostatic  pressure of the waste above the level sense line end   The system will suspend operation when this level is  too high  This port is not present on the Rf 150  systems       Diverter Valve Waste     The diverter valve on the  fraction collector arm sends fluid to this waste port       Inject Valve Waste     The injection valve is cleaned  after each run  The system directs fluids from this  automatic cleaning routine to this port     25 Lower Drain Tube   This tube carries away liquids  spilled onto the rack area     26  Cooling Fan      Rf 200i only  Cools the internal ELSD  components        Section 1 Introduction    27 Exhaust Port      Rf 200i only  Vents the carrier gas and  vaporized solvents           Refer to Fig  1 7  for other back  panel features        Figure 1 8 CombiFlash Rf 200i ELSD Features  Back     28  Nitrogen Inlet      Rf 200i only  Connects to the carrier gas  supply  60 to 70 psig      29 External Detector      Rf 200i only  Reserved for future  use     30 P Trap Drain      Rf 200i only  During operation   condensate from the Thermo split    process drains from  the system  An internal P trap prevents aerosol particles  from escaping through this drain     31 Pump Drain    Rf 200i only  An internal drip tray is  located below the splitter pump for the ELSD  Should the  pump or its fittings leak  liquids will drain from this port to  protect internal components        CAUTION    Discontin
47.  the system operation be verified  To do  so  use a pre packaged CombiFlash Rf test kit     e Part number 60 5237 050  Test kit containing five of the 5 gram size solid load  cartridges pre filled with 0 5 grams of test sample  NPHE  and five 4 gram RediSep Rf silica gel columns     e Part number 60 5237 051  Test kit containing four of the 25 gram size solid load    cartridges pre filled with 0 5 grams of test sample  NPHE  and four 4 gram RediSep Rf silica gel columns     2 15 Installation Qualification Checklist    Table 2 1 may be completed to verify and document the  installation procedures contained in Section 2 of this guide        2 22    Section 2 Preparation  Rf 200 and Rf 200i Systems    Table 2 1 Installation Qualification Checklist    Installer Operator  Description Initials Initials    UNPACKING THE UNIT       INSTRUMENT LOCATION       CONNECT POWER       CONNECT SOLVENT LINES       CONNECT WASTE LINES    2 6 EXTERNAL GAS  Optional        2 7 CONNECT AND ROUTE DRAIN  LINES       2 8 POSITION THE SYSTEM       2 9 INSTALL COLLECTION TUBE  RACKS       2 10   EXTERNAL DETECTOR   OPTIONAL        2 11 TURN ON POWER       2 12 CONFIGURE THE SYSTEM       2 13 PRIME THE SOLVENT LINES       2 14 SYSTEM VERIFICATION             Certification of Section 2 Completion       Installer Name  print         Installer Signature   Date              Operator Name  print         Operator Signature        Date           Comments                          2 23    CombiFlash   Rf User Ma
48.  this command to export a method file for use on a  CombiFlash Companion system equipped with the Solvent  Management Module  SMM  or a CombiFlash Companion XL  system  From the touch screen  this command opens the FILES   6 3 6  window so you can save the file on a USB Flash storage  device  When selecting this command from a remote computer   PeakTrak displays a window from which you can save the  method file     6 2 1 12 Print      From the touch screen  this command prints the completed  run on the network printer  see NETWORK PRINTING on the  NETWORK CONFIGURATION  6 3 10 2  window   When selecting  this command from a remotely connected computer  PeakTrak  displays the Printer window  From this window  you can select  and configure the printer of your choice  After configuring the  settings  click the PRINT button to print the completed run        6 6    Section 6 User Help Reference    6 2 1 13 Logout   Choose this command to logout of the system  After logging  out  the system displays a login screen and waits for the next  user to log in     6 2 2 Method Editor  Selecting the Method Editor menu command will open the  METHOD EDITOR  6 3 2  window  Use this window to view and  modify the advanced method file settings     6 2 3 Tools  PeakTrak   s Tools menu has the following options     6 2 3 1 Auto Prime Valve Wash   Choose this command to automatically prime the solvent  pumps  On CombiF lash Rf 200 and Rf 200i systems  this  feature will also wash the injection valv
49.  waste collection  container     Drain tubing is not required on the Pump Drain port      NCAUTION  Discontinue use of the CombiF ash Rf 200i if liquid is present at  the Pump Drain  Contact Teledyne Isco technical service for  assistance with correcting the leak        CombiFlash   Rf User Manual    2 8 Position the System    After completing the various connections  the system can be  moved to its operating position  Turn the system so that the  operator can access all of the front view features and controls   Figure 1 5   Use care not to damage the connections  tubing   and cables while moving the system         CAUTION    Ensure that the CombiFlash Rf has at least 3 cm  1 25   of air  space behind it for ventilation     Position the solvent and waste containers as necessary     When using higher density solvents such as dichloromethane   DCM   place the solvent container level with or above the  CombiFlash Rf system  Placing solvent containers below the level  of the Rf can contribute to decreased flow due to the high vapor  pressure of DCM  This problem becomes more pronounced as the  ambient temperature increases     When using higher density solvents such as dichloromethane   DCM   ensure that the waste container is no more than 3 feet  below the supply solvent levels  Placing solvent containers above  the system and waste containers on the floor may cause the  internal check valves to open and allow solvent to flow through the  system when in the standby state     2 9
50. 1000 4 3 2006   Radiated RF Immunity 80 mHz to 2 7gHz 80  AM at 1 kHz   1A1 2007  A2 2010 Level 2   3V m A   EN61000 4 4 2004   Electrical Fast Transient  EFT  on Level 2   1kV on AC lines A   1A1 2010 Mains and I O   EN61000 4 5 2005   Surge on AC Lines Level 2   1kV Line Common Mode A  Level 2   0 5kV Differential Mode   EN61000 4 6 2008   Conducted RF Immunity on Mains and   150 kHz to 80 mHz A   vo Level 1     1V rms  80  Modulated  EN61000 4 11 2004   Voltage Dips 0  during  cycle and full cycle A    70  at 25 cycles                CISPR11  RF Emissions Radiated  below 1GHz Group 1  Class A Industrial  Scientific  and PASS  EN 55011 2009 and Conducted  AC Mains Medical Equipment    A1 2010   EN61000 3 2 2005 AC Harmonics  Flicker PASS     A1 2008  A2 2009  EN61000 3 3 2008                   We  the undersigned  hereby declare that the design of the equipment specified above conforms to the above Directive s  and    Standards as of March 13  2013     USA Representative     Lb Rte       Vikas V  Padhye Ph  D     Vice President and Product General Manager    4700 Superior Street   Lincoln  Nebraska 68504     402 464 0231  402 464 0318          60 5232 509  Rev  B             TELEDYNE ISCO  j Everywhereyoulook                         iW  TELEDYNE ISCO    A Teledyne Technologies Company       4700 SUPERIOR ST  PHONE  402 464 0231  P O  BOX 82531 FAX  402 464 0318  LINCOLN  NE 68504 2531    Effective  January 5  2012    Subject  Teledyne Isco Products    RE  Bovine Spongiform 
51. CONFORMITY       ISMEA                 Application of Council Directive     Manufacturer s Name   Manufacturer s Address     Equipment Type Environment     Trade Name Model No   Year of Issue     Standards to which Conformity is Declared     2004 108 EC The EMC Directive  2006 95 EC  The Low Voltage Directive  2002 96 EC     The WEEE Directive    Teledyne Isco  Inc    4700 Superior  Lincoln  Nebraska 68504 USA  P O  Box 82531  Lincoln  NE 68501   Phone  1   402  464 0231   Facsimile  1   402  465 3799    Laboratory Equipment for Light Industrial Commercial Environments    Combiflash RF4X  2008    EN 61010 1  2    Edition   Safety Requirements for Electrical Equipment for  Measurement  Control  and Laboratory Use   EN 61326 2006   EMC Requirements for Electrical Equipment for  Measurement  Control  and Laboratory Use                                              Standard Description Severity Applied Performance Criteria   EN61000 4 2 Electrostatic Discharge Level 2   4kV contact discharge A  Level 3   8kV air discharge   EN61000 4 3 Radiated RF Immunity 80 Mhz to 2 7Ghz 80  AM at 1kHz A  Level 2   3V m   EN61000 4 4 Electrical Fast Transient Level 2   1kV on AC lines A   EN61000 4 5 Surge on AC Lines Level 2   2kV common mode   Level 2   1KV differential mode   EN61000 4 6 Conducted RF on AC lines 150 kHz to 80 MHz  A  Level 1   1V rms  80  modulated   EN61000 4 11 Voltage Dips 1 10 25 cycles  100 40 70  A   Voltage Interruption 250 cycles  100  drop c   CISPR11  RF Emissions G
52. CombiFlash   Rf  User Manual       For indoor use only    COPYRIGHT    2013 by  Teledyne Isco   4700 Superior St    Lincoln  Nebraska  U S A  68504    Phone   402  464 0231  Toll Free   800  228 4373  FAX   402  465 3022    Revision A  August 13  2013    Part  60 5233 584       CombiFlash   Rf User Manual       CombiFlash   Rf  User Manual       Table of Contents    1  Introduction    1 1 Documentation Overview               0 000 eee 1 1  1 2 Product Overview           0 0 0 KK KK KK RR KK KK KS 1 1  1 3 Operating Overview           0 0 0    1 6  1 3 1 Multiple Control Possibilities               1 6  1 3 2 File Storage cs wg vere 1 6  1 4 Specifications         kK KK KK KK KK KK KK KK KK KI 1 7  1 5 Controls  Indicators  and Features               1 13  TG  Safety        aa le Con du e ice ds 1 19  1 6 1 Hazard Severity Levels                   1 20  1 6 2 Hazard Symbols                  00000 1 20  1 7 For Additional Information                     1 22    2  Preparation   Rf 200 and Rf 200i Systems    2 1 Unpacking the Unit                     0 00085 2 1  2 2 Instrument Location              KK KK KK KK KK eee 2 2  2 8 Connect Power          0    eee KK KK 2 3  2 4 Connect Solvent Lines              0 000 cee eee 2 3  2 5 Connect Waste Lines                eese  2 6  2 6 External Gas    0    eee eee 2 8  2 6 1 CombiFlash Rf 200 and Rf 200i              2 8  2 6 2 CombiFlash Rf 200i Only                   2 9   2 7 Connect and Route Drain Lines                  2 9  2 8 P
53. Data Path     The top left corner of the  window displays the path  current folder   As you browse  through the files  the path will update as you go       New     Click this button to add and name a subfolder  within the currently selected folder       Delete     Click this button to delete the selected folder       Folder selection box     This box lists the available  folders and selection buttons     e OK     Click this button to save your selection as the  data path and close the window       Cancel     Click the Cancel button to close the window  without changing the data path     6 3 8 Auto Prime Valve Wash  The Auto Prime feature is a convenient way to prime the  pumps with solvent  This should be done whenever new  solvents are connected to the back panel ports  or to change  between the available solvents  An Auto Prime pumps 40 mL       6 25    CombiFlash   Rf User Manual    of the selected Solvent B  drawing it in through the back panel  port  through the pump  and out the waste port  On Rf 200  Rf  200i  and Rf 4x systems  the injection valve is included in the  fluid path to wash the valve  The Auto Prime then does the  same for the selected Solvent A  By priming Solvent A last  the  system is ready for the starting mobile phase of the next run     The Auto Prime window is opened by selecting the  TOOLS gt AUTO PRIME VALVE WASH  6 2 3 1  menu command  The  window contains the following controls       Progress bar     This bar displays how much of the  solvent ha
54. Drain Tubing    Process Valves    Vinyl with FEP liner    PEEK  PTFE  perfluoroelastomer       Sample Loading Cap    316 stainless steel       Sample Loading Cap  Seal    PTFE  Elgiloy          Injection Valve    PTFE  PCTFE       Flowcell    303 SST  Type ES Quartz  SIMRIZ SZ485       Pump       316 stainless steel  PTFE  Elgiloy         a  Elgiloy is a registered trademark of Combined Metals of Chicago  L L C        Section 1 Introduction    1 5 Controls  Indicators  and Features    Figures 1 5 through 1 7 illustrate the controls on the  CombiFlash Rf        Figure 1 5 CombiFlash Rf Features  Front     1  Touch Panel LCD display   Large 10 5 inch display for  system monitoring and control     2  USB Port     Convenient  front panel port that accepts USB  Flash memory drive  A Flash drive may be inserted into  this port for transferring files  importing and exporting  methods  and performing system software updates     3  Adjustable Column Mount   The injection valve  assembly and upper column mount slide along this mount  so the system can accept a variety of column sizes     4  On Standby Switch     1   On  0   standby     5  Fraction Collector Arm and Drop Former   The arm  and drop former move to deposit liquid in the collection  tubes     6  Racks and Collection Tubes     Racks hold the fraction  collection tubes for all Rf system  Racks include an RFID       1 13    CombiFlash   Rf User Manual    tag which the system uses to read the rack type and  collection tube size
55. Encephalopathy  BSE  Transmissible Spongiform  Encephalopathy  TSE     Teledyne Isco products are not manufactured using materials that are derived from  animal sources  Therefore  we can state that Bovine Spongiform Encephalopathy   BSE  Transmissible Spongiform Encephalopathy  TSE  should not be a concern     If you have any questions or require further information  please contact us or visit our  website at www isco com     Sincerely     Ut     Vikas Padhye Ph  D    Vice President and General Manager  4700 Superior Street   Lincoln  Nebraska 68504   Phone   402 464 0231    FAX   402 464 0318     This information is considered accurate and reliable as of the date appearing above and is presented in good faith   Because use conditions and applicable laws may differ from one location to another and may change with time   recipient is responsible for determining whether the information in this document is appropriate for recipient  s use   Since Teledyne Isco has no control over how this information may be ultimately used  all liability is expressly  disclaimed and Teledyne Isco assumes no obligation or liability thereof  Therefore  no warranty  expressed or  implied  is given nor is freedom from any patent owned by Teledyne Isco or others to be inferred     FmtASASMALARNEHRAS    Name and amount of Hazardous Substances or Elements in the product                                        BBTA BIRER JR  BES BR Hazardous Substances or Elements  Component Name E   Ld TOME ZBRKA   SR 
56. KE   Pb   Hg   Cd   Cr VD   PBB   PBDE   ARER  LCD Display o o o o o o   none on RF4X   SER x O O O  e  O  Circuit boards  Beek O  e   e   e  X O  Wiring  Pas Bet O O O O x O  Internal Cables  ERR O O O O x O  Line Cord  SHON X O O O X O  Stepper Motor  TAN O o X o Oo O  Deuterium lamp  Rik O  e   e   e  x O  Valve Body                         J m   h   B   SUB SWART   FAR   BB   Name and amount of Hazardous Substances or Elements in the product  o   RAT BT EYI RT ZIN II ARMA PHS RSET MEMENRELRAT     O  Represent the concentration of the hazardous substance in this component   s any homogeneous pieces is lower than the  ST  standard limitation     X  BINAB RET EI E 212022 N AR HRM rh ES   HERIN ST HMERENHREER    EAL  IZE CARE   ARS Sc ESTE EIT X    HERRARTE     A      X  Represent the concentration of the hazardous substance in this component   s at least one homogeneous piece is higher  than the ST  standard limitation    Manufacturer may give technical reasons to the    X   marks     FRE FIR 4J0 WASE  The Environmentally Friendly Use Period  EFUP  was determined through experience      E  EL REHAB TE RNS rh  BU      PECES E77 45  207 1858200755      BEJE     E RHSR BD   A       A  B   ZA    5     The date of Manufacture is in code within the serial number  The first three numbers are the year of manufacture  207 is  year 2007  followed by a letter for the month   A  is January   B  is February and so on           CombiFlash Hazmat Table 60 5233 406 Rev  B    Teledyne Isco O
57. N DETECTION  OoN LINEI       Sample Name     The sample name is a text entry box  in which you can label the run  Type a unique descriptor   PeakTrak saves the run information under this name   Therefore  PeakTrak cannot accept any characters that  are reserved by the operating system        V  etc    If you  do not type a Sample Name  PeakTrak will generate a  date time stamp for the name when you click the Play  button to begin the run        CombiFlash   Rf User Manual    Flow Rate     Type or select the desired flow rate for the  run  When using the Isco default column methods  the  system sets the flow rate to the optimum flow rate of the  column     Detector      Rf 200  Rf 200i  and Rf 4x systems   Reports the current peak detection mode  To change the  peak detection settings  open the METHOD EDITOR  6 3 2   window     Solvent A     Select the desired solvent from the list   The solvents in the list are those defined by the  CONFIGURATION  6 3 10  window  For assistance in  selecting a solvent  see the SOLVENT SELECTION  ON LINE   table     Solvent B     Select the desired solvent from the list     6 3 1 3 Run Control Buttons  From left to right  the control buttons are     Play     The Play button starts or resumes the run  After  a run has started  this button is replaced by the Pause  button     Pause     The Pause button holds the  B at the current  value while the system continues to operate  sometimes  called an isocratic hold   Note that pausing the run  extend
58. ON LINE   SOLVENT SELECTION  ON LINE    EDITING A METHOD  6 4 2    EDITING A DEFAULT METHOD  6 4 4    BYPASSING THE SOLID LOAD CARTRIDGE  6 4 7     6 4 2 Editing a method  Editing a method allows you to tailor the operation of the  CombiFlash Rf system to best separate or purify the  compounds of interest     PeakTrak has two types of method settings     basic and  advanced     6 4 2 1 Basic method settings   The basic settings allow you to control the  B gradient mix   flow rate  solvents  and run length  These settings are part of  the MAIN WINDOW  6 3 1   These settings are the most  frequently changed settings when developing or improving  purification methods     The  B gradient mix and run length are shown on the plot  area  Refer to the DEFINING A GRADIENT  6 4 5  topic for more  details  The other settings are       Flow Rate     When using the Isco default column  methods  the flow rate is automatically set to the  optimum flow rate of the column  If using a non Isco  column  refer to the manufacturer s column  documentation to determine the best flow rate for the  column  Note that if the column is not detected by the  system the operating pressure is limited to 50 psi       Solvent A and B     Select the desired solvents from the  lists  The solvents are those defined by the  CONFIGURATION  6 3 10  window     6 4 2 2 Advanced method settings   The METHOD EDITOR  6 3 2  window gives you access to  advanced settings such as equilibration volume  initial waste   air p
59. Protocol  The default  wide area network protocol that elevi communication                   General          You can get IP settings assigned automatically if your network supports  this capability  Otherwise  you need to ask your network administrator  for the appropriate IP settings       Obtain an IP address automatically     Use the following IP address    TP address    Subnet mask     Default gateway     Obtain DNS server address automatically    Use the following DNS server addresses   Preferred DNS server     Alternate DNS server                 Figure 7 2 Internet Protocol Version 4  TCP  IPv4  Properties       7 6    Section 7 Remote Interfaces    7 2 2 Windows Vista Settings    1  From the Windows Start menu  right click on Network   then select the Properties menu option  The PC displays  the Network and Sharing Center     2 Under Tasks  select Manage Network Connections     3  Right click on the Local Area Connection icon and select  the Properties option  After you provide permission to  continue  the PC displays the Local Area Connection  Properties window  Figure 7 1      4  Select Internet Protocol Version 4  TCP IPv4   then click  the Properties button  The PC displays the Internet  Protocol Version 4  TCP IPv4  Properties window   Figure 7 2      If the Internet Protocol Version 4  TCP IPv4  option is not listed   click the Install button to add it  If assistance is needed  press the  F1 keyboard button to display Windows Vista Help     The following steps 
60. RADIENT OPTIMIZER  6 3 3  window  When you accept  the data  the system will generate a gradient curve to separate  the target compound from the nearest impurity  When using  optimized gradients  PeakTrak may automatically insert an  isocratic hold to further increase the resolution     6 4 6 Real time Gradient Editing    The gradient shape can be changed during the run  Any time  after the fraction collector has positioned the drop former over  the first tube  click and drag the points on the gradient plot  area of the Main window  You can also add or remove points by  clicking the Insert or Delete point buttons  The Insert button  allows you to add a single point when you click on the gradient  profile  Repeat this action to add more points  The Delete  button allows you to remove a single point when you click on  it     Note that only the portion of the gradient that has not yet  occurred during the run can be modified        6 44    Section 6 User Help Reference    6 4 7 Bypassing the Solid Load Cartridge   Rf 200  Rf 200i  and Rf 4x Only  Occasionally  impurities  precipitate in the solid load cartridge as compounds are  purified  This may cause high back pressures resulting in  reduced flow rates and long run times  Clicking this button  changes the injection valve position  which removes the solid  load cartridge from the solvent path and relieves the back  pressure     CAUTION  Bypassing the solid load cartridge too soon during  the purification run may leave desired 
61. RING THE  WAVELENGTH RATIO  6 4 8  topic for more details on this  feature     Rf 75 and Rf 150 systems have a fixed UV detection  wavelength of 254 nm  The peak detection settings for these  systems include slope based and peak width options  These  are described in A1 and A2 above  Rf 75 and Rf 150 systems  with the variable wavelength option include a control to type  or set the detection wavelength     6 3 3 Gradient Optimizer  Use this window to automatically generate a gradient curve  from the results of two TLC plates  Follow the on screen  instructions  then enter valid TLC data in the Gradient  Optimizer window  Click the OK button and the Method  Editor window will display the new gradient curve        6 21    CombiFlash   Rf User Manual    When using the Gradient Optimizer  the system may insert an  isocratic hold into the generated gradient  Furthermore  the  system may automatically extend the isocratic hold during the  run to increase peak resolution     To open the Gradient Optimizer window  open the METHOD  EDITOR  6 3 2  window  then click the Gradient Optimizer  button     6 3 4 Column Data     Rf200  Rf200i  and Rf4X systems  The Column Data window  reports information about the RediSep Rf column installed in  the system  Compatible RediSep Rf and RediSep Rf Gold  columns have an RFID tag that identifies the column type   The system reads this information to identify the size and  media  then automatically adjusts the default method to  match the inserted c
62. Rf system  It also allows  you to configure the system for network connections for remote  access and printing  Open the Configuration window by  selecting the TOOLS gt MANUAL CONTROL  6 2 3 2  menu  command     The configuration window displays three tabs  INSTRUMENT  CONFIGURATION  6 3 10 1  NETWORK CONFIGURATION  6 3 10 2    and USER PREFERENCES  6 3 10 3      6 3 10 1 Instrument Configuration    Solvent Controls      Solvent A1 through B2     From the list of solvents   specify the solvent connected to the each port of the  system       Add Solvent button     Click this button to enter a  solvent name  The system can store up to 16 solvent  names  For assistance in selecting appropriate solvents   see the SOLVENT SELECTION  oN LINE  table     Remove Solvent button     Click this button to open a  list of all entered solvent names  Select the names you  want to remove and then click the OK button     Enable Solvent Level Sensing      Rf 200  Rf 200i   and Rf 4x systems  If this option is selected  the system  will check the solvent levels before and during a run   The system will alert you if a solvent level becomes too  low  Note that the solvent level monitoring option does  not control the operation of Waste Solvent Level  monitoring  Waste level sensing is always enabled at the  beginning of a run     Minimum Solvent Level     This value is  density dependent  thus a range is given  To understand  the range  consider that dichloromethane is more dense  than hexane  If 
63. System  Specifications 1  Continued     UV Detection    Wavelength 200 to 360 nm  optional 200 to 780 nm UV Vis       Monitor Wavelength Optional 200 to 360 nm  optional  with Purity Ratio 200 to 780 nm UV Vis          Wavelength Accuracy  5 nm       Fraction Accuracy   2mL    flow rate   60         Electrical Safety per EN 61010 1       Pollution Degree 2       Installation Category II       Maximum Altitude 2000 meters          Note 1  All specifications are subject to change   Note 2  For Rf200 columns undetected by RFID  the max pressure is limited to 50    psi           1 8    Section 1 Introduction    Table 1 2 CombiFlash Rf 200i    Dimensions    1    Specifications    61 cm  24   x36 cm  14 1   x43 cm  17    HxWxD        Weight    30 75 kg  67 8 lbs        Power Options    Input voltage range from 100 to 240 VAC  50 60 Hz   2 0 Amps system maximum     Line cord is the disconnect device        Line Frequency    50 60 Hz       Ambient Temperature    Humidity  when  connected to power     20 to 40   C  maximum temperature must be at least  15   C below the boiling point of the solvent     0 to 95  non condensing       Flow Rate Range    5 to 200 mL min       Flow Rate Accuracy     5        Pressure Limit 2    1379 kPa  200 psi        Pressure Accuracy    5  of full scale       Gradient Formation    Peak Detection Modes    Binary gradient from A1  A2  and B1  B2 solvent inlets   Solvent selection can be automatically controlled by  the method        Slope or threshold     
64. The CombiFlash Rf system is available in different  configurations        Figure 1 1 CombiFlash Rf 150 system  10 100 mL min  150 psi   without injection valve     CombiFlash Rf 150   For general purpose Flash chromatog   raphy needs  the CombiFlash Rf can be factory configured  as a 10   100 mL min  150 psi system that pumps two sol   vents to form a binary gradient  This configuration has an  easily accessible fluid path that supports liquid or solid  sample introduction  and a fixed wavelength UV detector   254 nm   Optionally  the system can be configured for  full spectrum  200   360 nm  UV detection with secondary  monitoring  manual solvent selection valves to form a  binary gradient from two of four solvent inlets  and a 4 port  manual injection valve        1 2    Section 1 Introduction       Figure 1 2 CombiFlash Rf 200 system  5 200 mL min  200 psi  with  automatic injection valve     CombiFlash Rf 200   For more advanced needs  the Combi   Flash Rf can be factory configured as a 5 200 mL min   delivery flow rate   200 psi system that forms a binary gra   dient by automatically selecting two solvents from four  inlets  This configuration has an automatic  self cleaning  injection valve for sample introduction  and full spectrum  UV  200   360 nm  or UV vis  200   780 nm  detection with  secondary wavelength monitoring  An internal air pump is  used for post run column air purging  active solvent level  sensing  and waste full bottle detection        1 3    CombiFla
65. To configure the system  open the  Configuration window by selecting Tools    Configuration from  the PeakTrak menu     The configuration window has three tabs  Instrument  Configuration  Network Configuration  and User Preferences   Network Configuration settings are discussed later in  Section 7  REMOTE INTERFACES     Adjust the following  sections 3 10 1 through 3 10 6  on the  Instrument Configuration and User Preferences tabs for the  desired operation and then click OK to save the settings     Some configuration settings do not take effect until the system  is restarted  After completing sections 3 10 1 through 3 10 6   follow the instructions in section 3 10 7        3 12    Section 3 Preparation  Rf 150 Systems    3 10 1 Solvents  When you installed the system  up to two solvents were  connected to the back panel  This section of the Instrument  Configuration tab allows you to name these solvents  To do so     The system is shipped with common solvent names already  loaded  If your solvent names are listed in the solvent drop down  list boxes on the main window  skip these steps     1  Click the    Add Solvent  button   2  Enter the solvent name and click the OK button   3  Repeat steps 1 and 2 for both solvents     3 10 2 General Settings    CombiFlash Rf Name  optional      Use this option to name  your system  The name will appear in operational displays  and run summaries  This feature is useful when your  laboratory has more than one CombiFlash Rf system     Time
66. VENT B    ethyl acetate PRIME B    PERCENTAGE SOLVENT B  50 E  FLOW RATE  ml min    PRIME 5096 B    STOP       Ly     PUMP INTO TUBE   e      VALVE POSITION  Through column    iil  DE  a         Figure 8 6 Manual Control Settings   Quick Cleaning    5  Select Through column for the Valve Position option     6  Select Next for the Pump into Tube   option  By pumping  into a collection tube  the diverter valve is also cleaned of  any residue during this operation     1  Click the Prime B button to pump 100  Solvent B through  the bypass tube and into the collection tubes     M Note   Pumping solvent B at a moderate flow rate  25 to 45 mL min  over  time will usually solubilize obstructions  Generally  the  recommended solvent is the highest polarity solvent you have  recently used  solvent B      8  Monitor the Raw Lamp Energy gauge  As the system  pumps solvent  the raw lamp energy should gradually  improve  Pump solvent for two to five minutes or until the  indicator reaches the far right of the green range     When the indicator is in the yellow  the lamp energy and flow  cell are ready for the next run  However  continuing the  cleaning for greater raw lamp energy should reduce the  frequency of flow cell cleaning messages  Figure 8 4      If after five minutes the indicator is still in the red range   repeat the cleaning steps using an alternative solvent   Table 8 1   Or  complete the steps in the Monthly Flow Cell  Cleaning procedure  section 8 5 3         8 8    Section
67. a connection to a printer on the network  Network  settings are discussed in Section 7  REMOTE INTERFACES  Do not  select this option for these initial installation steps     Default Run Units   Run units are displayed along the  X axis of the chromatogram  Select Time  in minutes  or  Column Volumes  A column volume unit is the time it takes to  pump enough solvent to exchange the volume held by the  column  The duration in minutes will vary according to the  column media and size  and the flow rate     Gradient Method     Select the standard gradient type   Figure 3 8        Linear     The system gradually mixes solvents A and B  from the last programmed  B until it reaches the next  programmed  B  This method connects the inflection  points that define the gradient with a straight line       Step     The system holds each  B value until it reaches  the next programmed  B  Because the system holds the   B values  the changes at each new value are nearly  instantaneous  producing a stepped appearance on the  plot area       Isocratic     The system holds the initial solvent mix for  the entire run     Linear Step Isocratic    Figure 3 8 Gradient types       3 15    CombiFlash   Rf User Manual    Original method Extension 2 Final Extension     Enable Run Length Extension     When enabled  this option  automatically extends the run if a peak is eluting at the end of  the maximum  B gradient  This ensures that a late eluting  peak fully comes off the column and is collected    
68. am mid run solvent B changes by clicking  the Edit Gradient button to open the GRADIENT TABLE       Edit Gradient     Although the gradient may be edited  directly within the plot area  you can also edit the  gradient in tabular form  Click this button to open the       6 17    CombiFlash   Rf User Manual    GRADIENT TABLE  On Rf 200  Rf 200i  and Rf 4x systems   the Gradient table also lets you program solvent changes  in the gradient     For complete instructions on defining gradient curves  see  DEFINING A GRADIENT  6 4 5      Gradient Table The gradient table depicts the points that  define the gradient curve in a tabular format  From this table  you can change Solvent B  Rf 200  Rf 200i  and Rf 4x systems    the duration  and  B concentration of any point on the curve   To do so  select the new value for a gradient point  Changes  that you make to the table will be reflected in the gradient plot  area     You can also change the number of points on the curve by  inserting or deleting rows in the table  To change the number  of points  first highlight a table cell or row  Then click the  appropriate Gradient Button for the action you desire       Insert Point     This command or button will insert a  row below the selected point       Delete Point     This command will delete the selected  row  Note that you cannot delete the initial point     The system automatically updates the Run Length setting or  scales the other points when you change the number of points  and thei
69. ample and observe the flow rate from the Status  Bar in the upper right corner of the main window  As long  as the system is delivering the programmed flow rate   there is no need to bypass the cartridge     4  If the flow rate is slower than that programmed  note the  slowest displayed flow rate  Use the equation below to  determine if the cartridge can be bypassed     T   Cartridge Volume    Rf   Flow Rate        5 12    Section 5 Operation    Where T is the time in minutes to wait after the system  reaches the  B solvent used to run the TLC plate  Rf is the  retention factor measured in step 1  Cartridge Volume is  the volume in mL from step 2  and Flow Rate is the current  flow rate in mL min     5  Observe the current  B concentration in the status bar   Once the system reaches the  B percentage used for the  TLC  wait at least the time established as T  After this time  has elapsed it is likely that the desired compound has come  off the cartridge and you can safely click the Cartridge  Bypass button     For example  a TLC plate run at 30 B revealed that the last  desired compound had an Rf factor of 0 35  An empty 25 gram  size solid load cartridge was filled with sample adsorbed onto  10 grams of silica  and then placed on the system  The  programmed flow rate was 60 mL min  During the run  the  flow rate dropped to 30 mL min  The equation is then     T   45 mL   0 35   30 mL min    4 3 minutes    In this example  the Cartridge Bypass feature was used 4 3  minutes aft
70. an  find the MAC Address on the Help gt About PeakTrak window     5  Click the OK button to save the settings     6  Confirm that your network administrator has completed  any necessary network changes to support the system     7  Locate the CAT5 connection cable  P N 480 6545 01  in the  accessory package  Insert one end into the Ethernet port on  the back panel  Connect the other end of the cable to your  network access port     You should be able to connect to the CombiFlash Rf from a PC  on the network  Section 7 1 1  and configure other network  features such as network printing  Section 7 1 2  and network  file saving  Section 7 1 3   Networks with Wi Fi access points  may also support remote control with mobile digital devices   Section 7 1 4      7 1 1 Network PC Access    A successfully networked Combiflash Rf system can be  accessed by a PC that meets the recommendations listed in  Table 7 1     Table 7 1 Personal Computer Recommendations     Operating Systems  Microsoft Windows 2000  XP  Vista  or Windows 7       Hardware  The computer hardware must meet the minimum  required specifications of the selected operating  System     Network Protocol  TCP IP  Network Connection  IEEE 802 3  Ethernet              Display  800 x 600 pixels  256 colors  minimum       Internet Browser  Last two versions of Microsoft Internet Explorer          a  This table shows supported PC configurations  Other configurations or  AJAX compatible browsers may be possible but are unsupported
71. ary  wavelength detection  When enabled  type or select the  peak detection wavelength in nanometers  then click the  Details button to configure additional settings described  below       Slope Based     Select this option to enable  Slope based peak detection  When enabled  peaks will  be recorded if the slope algorithm indicates a peak  within the Peak Width setting       Peak Width     Select the average peak width  setting  Peak widths are measured at the baseline   The slope detector will detect peak widths ranging  from about 0 2 to 2 times the peak width setting  For  example  if you entered a peak width of 1 minute  the  range would be 12 seconds to 2 minutes  For best  operation  the peak width should be set to just over  the average peak width being separated  For  instance  if the average peak width is 45 seconds  you  should enter a peak width of 1 minute  For most flash  chromatography  1 minute is a good starting point for  a peak width       Threshold Detection     Select this option to enable  Threshold peak detection  When enabled  peaks will  be recorded if the Absorbance Units  AU  value is  exceeded       Threshold     Type or select the Absorbance Units  value to be used for Threshold detection     Note  If both Slope and Threshold peak detection  methods are checked  the system considers a peak to  be present when any one condition is met  This logical  OR operator means that the system will cut a peak  when either the Slope condition is true  or when t
72. ased and  peak width options described in A1 and 42 above  The  system will use its internal algorithms to cut detect and  cut peaks based on the analog input signal  Refer to the  External Detector instruction sheet for cable and       6 20    Section 6 User Help Reference    plumbing requirements  The external detector option is  not available on Rf 150 and Rf 200i systems       ELS Detector     Enable this option to use the  evaporative light scattering  ELS  detector on an Rf 200i  system  When enabled  click the Details button to  configure additional settings     Signal Gain     You may modify the gain to scale the  detector signal    Slope Based and Threshold     Refer to these  features described in A1 and 42 above    Spray Chamber Temperature   The default  settings are 50   C  normal phase default methods   and 15   C  reverse phase   This setting may be  adjusted from 10 to 60   C so that detection can be  optimized for the solvent system in use  However  the  lower limit is 5  C below the ambient temperature   Drift Tube Temperature     The default settings are  55  C  normal phase default methods  and 60  C   reverse phase   This setting may be adjusted from 30  to 90   C  but is limited to a range of 5   C below and 60    C above the ambient temperature     Show Ratio     If using two of the above peak detection  options  check this option to display a ratio of the  selected wavelengths  The ratio trace is often a useful  indicator of purity  Refer to the MONITO
73. ate static electricity        CombiFlash   Rf User Manual    To Solvent Supply           r    To Solvent Level Sensing  Connection    60 0923 017 Ferrule  60 0923 015  Compression Fitting Nut    60 5234 130  Label  1 of 4                       60 5235 050 38 mm Cap  or  60 5235 051 45 mm Cap       60 5234 129   Inlet Filter Assembly   Place weighted filter at  bottom of container      Figure 2 2 Recommended solvent connection     NCAUTION    To prevent damage or premature wear to the pump and internal    valves  clean solvent should be used  The solvent should not  contain any dissolved solids        2 4    Section 2 Preparation  Rf 200 and Rf 200i Systems     NCAUTION    If solvent containers are reused  particulate matter can  accumulate in the container  Containers should be thoroughly  cleaned before each refill  Failure to clean the containers may  result in damage to the solvent selection valve and internal  components     The CombiFlash Rf 200 and Rf 200i systems have four solvent  inlet ports on the back panel  two A solvents and two B  solvents     The system requires at least one    and one B solvent to form a  binary gradient  and assumes that the B solvent is stronger  than A  The second set of inputs allow you to form alternative  binary gradients  or to change the B solvent mid run     The accessory package includes several pieces to complete the  solvent inlet connections  To connect the solvent lines     1  Solvent lines between the check valves and the solven
74. ation when the PeakTrak screen is displayed     Section 4 Preparation  Optional 4x Module    4 10 Prime the Solvent Lines    The Rf 4x module was factory tested with 1 1 water isopropyl  alcohol and subsequently purged with air  If residual fluids  would affect your chromatography  prime all channels on the  system     Channel 1 should have been primed during the installation  steps in Section 2 13  page 2 21   Channel 1 can be primed  using the Auto Prime menu command     Channels 2 through 4 must be primed using the Manual  Control menu command  To do so     1  Ensure that the solvent containers are filled     2  Insert a priming tube assembly  from the accessory  package  between the upper and lower column mounts on  channels 2 through 4     3  Select the  Tools   Manual Control    menu command   4  Select the Column 2 the channel option     5  Click the Prime B button  The system pumps 100  solvent  B through channel 2     When priming the first time  inspect the solvent and waste  connections to the system  If any leaks are observed  click the  Cancel button to stop the pump  Correct the leak by tightening the  fitting an additional 1   4 turn  and then restart the prime     6  Allow the system to pump solvent B for at least 3 minutes   Then  click the Cancel button     1  Click the Prime A button  The system pumps 100  solvent  A through channel 2     8  Allow the system to pump solvent A for at least 3 minutes   Then  click the Cancel button     M Note  This B A order 
75. b lpO      Network File Save Configuration When configured  this feature  allows the system to access a network directory from which it  can save run files as PDF  text  or run monitor  To enable this  feature  select a file type to be saved and enter the remaining  settings necessary for network access  Your network  administrator will be able to provide the required settings     NOTE  All server and domain names must be fully qualified   That is  entries must include the full name  server  name domain domain      Use forward slashes      not back  slashes  when entering the network share path     6 3 10 3 User Preferences  Language Select the language for PeakTrak     Automatically Print Report at End of Run If this option is  enabled  the system will print a completed run report using  the NETWORK PRINTING feature     Default Run Units Run units are displayed along the X axis of  the chromatogram  Select  Time  Minutes  or COLUMN  VOLUMES  oN LINE   CV  run units to be displayed on program  settings and chromatogram displays  The duration in minutes  will vary according to the column media and size  and the flow  rate     Enable Run Length Extensions When enabled  this option  automatically extends the run if a peak is eluting at the end of  the programmed run length  This ensures that a late eluting  peak is still collected        6 31    CombiFlash   Rf User Manual    Gradient Method Select the desired gradient method option       If the Linear method is selected  solven
76. basic A dish type detergent solution can wash through       particulates not picked up by the acidic solution  The  emulsification effect of detergent is an effective tool as  well to help wash through oily samples  Follow the  detergent with several minutes of clean  distilled water           a  Recent synthesis solvents should also be considered        CombiFlash   Rf User Manual    8 5 3 Monthly Flow Cell Cleaning  Perform this procedure as part of your scheduled preventive  maintenance  or when Quick Cleaning As Required  section  8 5 2  does not improve the lamp energy    1  Remove the column and insert a bypass tube between the  upper and lower column mounts     bypass tube   209 0165 46  can be found in the accessory kit   60 5239 006      2  From the menu  select Tools   Manual Control  This opens  the Manual Control window     3  Set the Flow Rate to 40 mL min  Figure 8 6    4  Select Through column for the Valve Position option   5  Select Next for the Pump into Tube   option     6  Place the B1 Solvent inlet line into a reservoir of methanol   acetone  or a strong solvent that readily dissolves residual  sample material     1  Click the Prime B button to pump 100  Solvent B through  the bypass tube and into the collection tubes     8  After three minutes  click the Stop button  Allow the  system to stand for at least six hours  Overnight is  recommended     9  Return the B1 solvent line to the original solvent container     10  Perform the Quick Cleaning as Require
77. change the local area connection settings  If  the PC must be restored to its original network configuration   record the present Internet Protocol  TCP IP  settings now     5  Select the    Use the following IP address  option   6  Enter the IP address    192 168 1 200      7  Enter the Subnet Mask     255 255 255 0        8  Click the OK button to close the Internet Protocol Version 4   TCP IPv4  Properties window  Click the OK button to close  the Local Area Connection properties window     9  Shut down the PC and place the CombiFlash Rf system in  Standby        CombiFlash   Rf User Manual    7 2 3 Windows XP Settings  1  From the Windows Start button  open the Control Panel     2  Locate the Network Connection icon and open this control  panel     3  Highlight the    LAN or High speed Internet Connection     icon  Select the    Change settings of this connection    option   or right click and select    Properties     Refer to Figure 7 3     4  Highlight the    Internet Protocol  TCP IP   item in the list   and click on the    Properties    button  Refer to Figure 7 4     5  Select the Alternate Configuration tab  Click the    User  Configured  option  See Figure 7 5     6  Enter the IP address    192 168 1 200      7  Enter the Subnet Mask     255 255 255 0        8  Click the OK button to close the Internet Protocol  properties window  Click the Close button to close the Local  Area Connection properties window     9  Shut down the PC and place the CombiFlash Rf system 
78. channels 2 through 4  for automatic injection valve cleaning     The Rf 4x module can automatically detect full waste bottles  and stop operation to prevent overfilling a waste container   This safety feature requires five air line connections       one Waste Level Sense From Rf inlet to receive air or  external gas from the CombiFlash Rf system       four Waste Level Sense outlets 1  2  3  and 4     M Note  The following instructions assume that the waste from each  channel will be isolated in separate containers        4 4    Section 4 Preparation  Optional 4x Module                                  WASTE    LEVEL  SENSE  FROM RF  WASTE LEVEL SENSE  VALVE  WASTE  OUT  COLUMN  e         LO O   From  DISCONNECT SUPPLY RE  BEFORE SERVICING  MAINS  2 10O  A B 3    Figure 4 2 Rf 4x Module back panel connections    IW Note   If the waste is to be collected in a common container  skip the  Column Waste and Waste Level Sense connections  steps 1  through 7  and complete step 8 to route the Valve Waste Out from  the Rf 4x module to the common container  The waste tubing  assembly connected to the back panel of the CombiF ash Rf  system will route all waste fluids to the common collection  container and sense the level        To complete the waste line connections     1  At the waste container for the CombiFlash Rf system   Channel 1   remove the black tubing that delivers Diverter  Valve Waste fluid and the natural tubing that delivers  Waste Level Sense air  The cap and black tubin
79. chemists Stop and then Fast Forward the run after  the last compound elutes  This action skips the  post separation flush  If any of the above conditions appear   consider allowing some of the runs to continue through the       8 5    CombiFlash   Rf User Manual    flush  or run a high percentage of  B solvent through the  system for a few minutes at the end of each day     If the separation runs always continue through the flush and  the conditions still occur frequently  edit the default column  methods to extend the flush duration      NCAUTION  When using methanol as solvent B with silica column media  do  not pump more than 5096 solvent B  Higher percentages of  methanol might break down the silica structure  possibly causing  obstructions in the flow path     8 5 2 Quick Cleaning when Recommended    When the lamp energy is lower than normal  the system will    recommend flow cell cleaning  Figure 8 4  before starting a  separation run     CONTINUE AND  CANCEP RN COLLECT ALL ea    Figure 8 4 Flow cell cleaning message       When the system displays this message you can       Cancel Run  recommended    Click the Cancel Run    button so you can perform a quick cleaning described in  the following steps       Continue Collect All     Click this button to ignore the  message  Because the peak detection operation might be  impaired  the system automatically collects all fluids to  avoid diverting compounds of interest to waste       Help   Click this button to display the flo
80. compounds on the solid  load cartridge  After removing the cartridge from the fluid  path  these compounds cannot be purified     If you need to use this feature  perform the following to ensure  that you do not leave desired compounds on the cartridge     1  Run a TLC plate prior to starting the run  Measure the Rf   retention factor  of the slowest eluting compound of  interest   The Rf should be between 0 2 and 0 8     the same  range used by the GRADIENT OPTIMIZER  6 3 3    Note the  Percent B solvent used to run the TLC plate  If a compound  displays tailing peaks as streaks on TLC  the Rf used  corresponds to the bottom of the streak     2 Load the Sample in the Solid Load Cartridge and note the  cartridge size  Refer to list below to determine the cartridge  volume  If the cartridge is partially filled with silica  use  the cartridge volume as if the cartridge were completely  filled with silica     e 5 gram cartridge   10 mL volume    25 gram cartridge   45 mL volume    65 gram cartridge   110 mL volume    3 Run the sample and observe the flow rate from the Status  Bar in the upper right corner of the main window  As long  as the system is delivering the programmed flow rate   there is no need to bypass the cartridge     4 If the flow rate is slower than that programmed  note the  slowest displayed flow rate  Use the equation below to  determine if the cartridge can be bypassed     T   Cartridge Volume    Rf   Flow Rate        6 45    CombiFlash   Rf User Manual    Wh
81. conditions that will drive off the solvent  without affecting the compounds of interest  These compounds  remain bound to the media  Once dried  pour the  media sample mixture into an empty cartridge        5 1    CombiFlash   Rf User Manual    To prepare an empty solid sample cartridge     1  Ensure the empty cartridge has a bottom frit        CAUTION    Missing frits may cause equipment damage  UV detection  problems  or increased maintenance  Frits prevent solids from  entering the fluid path     2  Prepare the media and sample     a  Dissolve your sample in a minimal amount of a  suitable  volatile solvent    b  Place the media into the solvent  If using silica gel  a  particle size of 40 60 pm  240   400 mesh  is  recommended  The amount of silica required is about  four to five times the mass of your sample    c  Agitate the solvent for a moment to allow the sample  to adsorb to the silica    d  Remove the solvent with a suitable method  such as  rotary evaporation  Alternatively  you can dry the  prepared cartridge after step 5 by drawing the  solvent out using a vacuum or using the Rf 200 or Rf  200i system   s air purge    3  Load the media and sample mixture into the cartridge  Tap  the cartridge on the benchtop to settle the mixture     4  Place a frit on the top of the cartridge  Force the frit down  against the mixture using the plastic plunger     5  Wipe any residual powder inside the neck of the cartridge     On Rf 200  Rf 200i  and Rf 4x systems  you can remo
82. d  protect the next user from being sprayed by  solvents left in the lines by pressing Fast Forward so the  instrument purges itself and is safe for its next use       Solid Sample Cartridge with air purge NOT enabled   i e   purge time set to zero in Method Editor   If you suspend the run with the Stop button  remove the  column and install the Prime Tube in place of the column   Manually purge the Sample Cartridge by pressing Tools   gt   Manual Control   gt  Air Purge Cartridge  Repeat this  procedure for Air Purge Column       Liquid sample loaded onto column with air purge  enabled   If you suspend the run with the Stop button  remove the  column and install the Prime Tube in its place  Use the Fast  Forward button to ensure no sample is left in the flow cell  and fraction collection lines  possibly affecting subsequent  separations        5 10    Section 5 Operation      Liquid sample loaded onto column with air purge NOT  enabled   If you suspend the run with the Stop button  remove the  column and install the Prime Tube in its place  Manually  purge the liquid lines by pressing Tools   gt  Manual Control      Air Purge Column       Next Tube     This button advances the fraction  collector to the next tube position  This allows you to  conveniently collect eluate of interest in a new tube     When you click the Next Tube button  the system immediately  marks the tube advance on the chromatogram  However  the  system delays the movement of the drop former over the tube
83. d  section 8 5 2  and  monitor the Raw Lamp Energy     If the lamp energy is in the green or yellow range  return  the system to operation  If the lamp energy is red  contact  Teledyne Isco s Technical Service department for assis   tance     8 6 Injection Valve Maintenance    CombiFlash Rf200  Rf 200i  and Rf 4x systems have an  automatic injection valve  Under normal operating conditions   this self cleaning valve does not require any maintenance     If maintenance is required  refer to Instruction Sheet  60 5232 475  CombiFlash Rf Injection Valve Replacement        8 10    Section 8 Maintenance    8 7 Electrical Troubleshooting    If your instrument stops working and the touch panel display  is off  check the line cord connection     If the line cord is connected properly  check the fuses in the  power input module located on the system   s rear panel  These  are time delay fuses  T type  and rated according to the mains  input power voltage     CombiF lash Rf 200 and Rf 150 systems     100 to 129 VAC mains input  2 0 A fuse  one required  e 215 to 264 VAC mains input  1 0 A fuse  two required  CombiF lash Rf 200i system     100 to 129 VAC mains input  3 0 A fuse  one required    215 to 264 VAC mains input  3 0 A fuse  two required       CAUTION    Always replace fuses with the same type and rating     The CombiFlash Rf 4x module does not have a user accessible  fuse        CombiFlash   Rf User Manual             DECLARATION OF CONFORMITY    ISM1 A                Applicati
84. d an end time  column volumes may  also be used   Note that all times are relative to start of  the run  just following the sample injection  Then choose  a peak collection mode  If you select Peaks  the module  will collect fluid only when the time window is active  and a peak is present  If you select All  the module will  collect all fluid during the time window regardless of the  peak state  Define up to three time windows     e Air Purge      Rf 200  Rf 200i  and Rf 4x systems  Type  or select the duration of a post run air purge  in minutes   An air purge clears the solvent from the column and the  fluid path  forcing any liquid out the waste port  If you  are disposing the column after each run  a sufficient  post run air purge ensures that it will be empty  To  disable the air purge  enter a value of 0  zero   A column  can also be manually air purged  See MANUAL CONTROL   6 3 9      6 3 2 3 Run Notes   Use this text entry box to enter comments or notes for the run   These comments will be saved with the run and will appear in  TXT and PDF reports        6 16    Section 6 User Help Reference    6 3 2 4 Gradient Plot Area   The gradient plot area depicts the current gradient profile  that will be used for the run  The X axis of the plot displays  the run time in minutes or COLUMN VOLUMES  oN LINE   The  X axis scale can be adjusted by typing or selecting the Run  Length  The left Y axis displays absorbance units and the  right Y axis indicates the percentage of solvent
85. d from the column before the run     Loading Type options for the Rf 150 system are       Equilibration     Select this option if you intend to  equilibrate the column before introducing the liquid  or solid sample       No Equilibration     Select this option to skip column  equilibration     5  Review or update the Start Rack and Start Tube     6  Review the Minimum Run Requirements  The system will  report the estimated solvent volumes  expected waste   collection tube usage  etc  You can use this information to  verify that there will be enough solvent to complete the  run  the waste volume does not exceed the collection  container s capacity  and whether or not more tubes will be  required during the run     The Rf 200  Rf 200i  and Rf 4x systems estimate the solvent  volumes by monitoring the solvent level in the container and the  known usage rate  The systems continue to refine this estimate  during operation  Solvent level sensing is not available on the Rf  150 system        5 7    CombiFlash   Rf User Manual    If the waste level sensing tube is inserted correctly into the waste  container  the Rf 200  Rf 200i  and Rf 4x systems will  automatically suspend operation before an overflow condition  might exist  To prevent the run from being suspended before  completing the run  ensure that the container will hold the  expected waste volume  Waste level sensing is not available on  the Rf 150 system     7  Rf 200i only  If the ELSD will be used for the run  ensure  tha
86. deficiencies before returning the instrument to  operation        CombiFlash   Rf User Manual                         Figure 8 1 Collector tray drain hole    Figure 8 2 Top shelf drain hole       8 4       Vacuum or  pressurized air  applied to the outlet  end of the drain  tube must exist at  the collection tray  drain hole     Vacuum or  pressurized air  applied to the outlet  end of the drain  tube must exist at  the top shelf drain  hole     8 5 Flow Cell Cleaning    8 5 1 Post Separation    Section 8 Maintenance    As a preventive measure  all default column methods finish  the separation run with a high percentage of solvent B   Figure 8 3   This brief time  two to three column volumes  of  strong solvent flushes residual compounds from the column     flow cell  and internal tubing        End of  Separation      Separation            Z    Gradient             Time or Column Volumes            100    C  C       l     75 Qo  i F g    HF 2    C D  i     50 se    E  77   9    Ps  lt   1 jmd  2     End of     25      w     Run L   0L        0    E        Figure 8 3 Default post separation column and flow cell flush    Skipping the post separation flush may cause residual  compounds to build up and crystallize  which might result in       cross contaminating later separation runs      higher operating pressures      reduced flow cell lamp energy      a noisy baseline on the absorbance trace      frequent messages recommending flow cell cleaning     Figure 8 4      Typically  
87. e   Due to the high pressure that is possible during rapid equilibration   this option may not be desirable for some column media or  purification methods     3 10 7 Restart the system    Some configuration settings do not take effect until the system  is restarted  After completing sections 3 10 1 through 3 10 6  restart the system     1  If the Configuration window is not closed already  click the  OK button to close it     2  Place the Power switch below the LCD display in the Off  position and wait one minute or more      N CAUTION    Always wait at least one minute after placing the Power switch in  the Off position  Moving the switch to the Off position begins a  power down sequence that takes up to one minute to complete  If  you move the switch to the ON position before the sequence is  complete  the system may not start up correctly       A CAUTION    Except during an emergency  never remove AC power  immediately after placing the Power switch in the Off position  The  power down sequence includes file maintenance that takes up to  one minute to complete  Interrupting the file maintenance may  corrupt files causing abnormal operation or a complete system  failure     3  After one minute or more  place the Power switch in the On  position     4  The system is ready for operation when the PeakTrak  screen is displayed        3 17    CombiFlash   Rf User Manual    3 11 Prime the Solvent Lines    Before the first use  the system should be primed  Ensure that  the solvent
88. e  These  corrupted files can cause abnormal operation or a complete  system failure that requires service  Unless power must be  removed due to an emergency  always wait at least one minute  after placing the system in Standby before removing the AC mains  power cord        CombiFlash   Rf User Manual    8 2 Preventive Maintenance    The system requires preventive maintenance for safe and  reliable operation  Refer to the schedule below for the  minimum periodic maintenance requirements     As Needed     Perform these tasks as conditions require     Cleaning  section 8 3        Quick flow cell cleaning when recommended by a system  alert message  section 8 5 2      Every Run   Allow the separation run to finish with a high  percentage of solvent B to flush residual compounds from the  column  internal tubing  and flow cell  Refer to section 8 5 1     Monthly   Perform these tasks at least monthly  more  frequently if conditions warrant       Tubing Inspection  section 8 4      Collection rack and tray cleaning  section 8 3 1        Monthly flow cell cleaning  section 8 5 3      8 3 Cleaning    To clean the exterior surfaces  use a cleaning cloth dampened  with a mixture of distilled water and a mild detergent  Use  isopropyl alcohol for tougher stains     On printed areas such as labels  avoid rubbing vigorously or    using aggressive solvents like acetone  Each will ruin the  printed text      NCAUTION    Do not immerse the instrument in a water bath or subject it to a  liq
89. e  When you select this  command  the system opens the AUTO PRIME VALVE WASH   6 3 8  window     6 2 3 2 Manual Control   Choose this command to open the MANUAL CONTROL  6 3 9   window  From this window you can operate the system  manually  Manual control can assist with priming or purging  the internal plumbing  and aid system troubleshooting     6 2 3 3 Auto Zero  Choose this command to zero the absorbance trace during a  run     6 2 3 4 Edit Defaults   Choose this command to open the METHOD EDITOR  6 3 2   window from which you can modify the default method files   To prevent tampering with the default column programs   access to this function can be password protected using the  SET PASSWORD  6 2 3 6  menu command     See Also  EDITING A DEFAULT METHOD  6 4 4      6 2 3 5 Configuration   Choose this command to open the CONFIGURATION  6 3 10   window  From this window you specify the solvents used with  the system  set default volumes for collection tubes  set the  system date and time  configure network settings  efc        6 7    CombiFlash   Rf User Manual    6 2 3 6 Set Password  Choose this command to open the SET PASSWORD  6 3 11   window     Passwords protect unauthorized modifications to     Default methods    User management    system date and time    system password    Before any of the above can be accessed  the user must enter a  password  Password protection for the default methods can be  disabled so all users can modify default methods     6 2 3 7 User Manag
90. e  following discussion     If a pure compound is eluting  the absorbance is linearly  related to the concentration of the compound in the solvent  If  the compound absorbs differently at different wavelengths   the absorbance at each wavelength may be different  but still  linearly related  For example  assume a compound eluting  from the system has an absorbance equal to 2 times the  concentration at 254 nm  This same compound at 220 nm has  an absorbance of 1 5 times the concentration  The ratio of  these signals will be 1 33  Since the relationship of absorbance  to concentration is not variable  the ratio remains steady while  the concentration changes from the beginning to the end of the       6 46    Section 6 User Help Reference    peak  During the duration of the peak  the ratio will be 1 33  and this constant value is displayed as a horizontal line     Now assume a case where there is a second compound eluting   only slightly shifted in time from the original compound  It is  possible that the detection absorbance trace alone would  indicate a single  valid chromatographic peak  In reality  itis a  combination of two peaks  By monitoring a second wavelength   it may be possible to reveal the second compound  Because of  the slight shift in time and the different absorbance properties  of the two compounds  the changing ratio during the detected  peak would reveal the impurity  Therefore  one can therefore  assume that if the ratio is not constant for the entire dura
91. e optional valve kit  the second set of inputs  allows you to manually switch between alternative binary   gradient solvent systems  or to change the B solvent mid run        3 4    Section 3 Preparation  Rf 150 Systems     NCAUTION    To prevent damage or premature wear to the pump and internal  valves  clean solvent should be used  The solvent should not  contain any dissolved solids     An optional  manual solvent selection valve kit may be added to  provide inlets for two A solvents and two B solvents  If using the  optional valve kit  connect the kit and solvent lines according to the  instruction sheet     Accessory package P N 60 5239 009 includes pieces to  complete the solvent inlet connections     1  Locate a Solvent Inlet Filter Assembly  P N 60 5234 193   in the accessory package  Connect the tubing end to the A  Solvent inlet  Thread the nut into the Solvent A Inlet and  finger tighten     2  Connect the second Solvent Inlet Filter assembly to the B  Solvent inlet     3  Insert the filter end of each solvent inlet filter assembly  into their respective solvent containers     4  The accessory package includes labels to easily identify the  inlet lines  Apply the A and B labels to the solvent lines  above the solvent container openings      NCAUTION    If solvent containers are reused  particulate matter can  accumulate in the container  Containers should be thoroughly  cleaned before each refill  Failure to clean the containers may  result in damage to the solvent
92. e peak detection and collection  settings for the run  You can return to the rack and tube  display by clicking the Display Rack button     See Also     PEAKTRAK MENU OPTIONS  6 2     e STARTING A RUN USING DEFAULT SETTINGS  6 4 1       REAL TIME GRADIENT EDITING  6 4 6     e METHOD EDITOR  6 3 2  for advanced functions    6 3 2 Method Editor    The Method Editor window has several sections  The BUTTON  BAR  6 3 2 1  at the top of the window gives you quick access to  file operations  a GRADIENT OPTIMIZER  6 3 3   and COLUMN  DATA  6 3 4   Rf200  Rf200i  and Rf4X   Below the button bar  are the RUN SETTINGS  6 3 2 2  and the RUN NOTES  6 3 2 3   The  lower part of the window contains the GRADIENT PLOT AREA   6 3 2 4  and settings for PEAK COLLECTION  6 3 2 5  and PEAK  DETECTION  6 3 2 6      6 3 2 1 Button bar  The Method Editor button bar contains the following     New      Rf200  Rf200i  and Rf4X  Click this button to  open a new method file using the default method  settings for the detected column media and size     Open     Click this button to open a method file stored  on the system s internal hard drive     Save     Click this button to save any modifications to  the current method file  Note that if you attempt to save  modifications to a default method file  the FILES  6 3 6   window will open instead so you can rename the file   This preserves the default method  If you do wish to edit  and save the system s default methods  use the     TOOLS gt EDIT DEFAULTS  6 2
93. e rapid equilibration     select this option to  equilibrate the column at a high flow rate     M Note   Due to the high pressure that is possible during rapid equilibration   this option may not be desirable for some column media or  purification methods     2 12 7 Restart the system    Some configuration settings do not take effect until the system  is restarted  After completing sections 2 12 1 through 2 12 6  restart the system     1  If the Configuration window is not closed already  click the  OK button to close it     2  Place the Power switch below the LCD display in the Off  position and wait one minute or more      NCAUTION    Always wait at least one minute after placing the Power switch in  the Off position  Moving the switch to the Off position begins a  power down sequence that takes up to one minute to complete  If  you move the switch to the ON position before the sequence is  complete  the system may not start up correctly      NCAUTION    Except during an emergency  never remove AC power  immediately after placing the Power switch in the Off position  The  power down sequence includes file maintenance that takes up to  one minute to complete  Interrupting the file maintenance may  corrupt files causing abnormal operation or a complete system  failure        2 20    Section 2 Preparation  Rf 200 and Rf 200i Systems    3  After one minute or more  place the Power switch in the On  position     4  The system is ready for operation when the PeakTrak  screen is d
94. e system  will clean the injection valve     Clicking the Fast Forward button will cause the system  to skip any remaining time in the current step and  advance to the beginning of the next step  Unlike the  Rewind button  the Fast Forward button will not modify  the current method s run length setting     6 3 1 4 Peak Collection Buttons  Three Peak Collection buttons are located at the bottom of the  Main window       All     Click this button to collect all fluids in the fraction  collection tubes       Peaks     Click this button to collect only eluted peaks in  the fraction collection tubes       None     Click this button to divert all fluids to the  waste port     6 3 1 5 Peak Collection Data   Peak Collection data is displayed in the Main window after a  run  You may also open a RUN FILE  6 1 3  for viewing data from  previous runs     The peak collection data is displayed on the left side of the  Main window       Rack and tube information     Collected peaks are  color coded in the rack diagram so that you can easily  locate the peaks of interest  The tube colors correspond  to the color bars under the peaks displayed on the  chromatogram  If during the run more than one set of       6 13    CombiFlash   Rf User Manual    tube racks was filled  use the Next and Previous Rack  buttons to view the additional racks  The table below the  rack diagram displays the peak data in tabular form       Display Method     Click the Display Method button to    view a summary of th
95. e system to be configured for network  operation and a connection to a printer on the network  Network  settings are discussed in Section 7  REMOTE INTERFACES  Do not  select this option for these initial installation steps     Default Run Units     Run units are displayed along the  X axis of the chromatogram  Select Time  in minutes  or  Column Volumes  A column volume unit is the time it takes to  pump enough solvent to exchange the volume held by the  column  The duration in minutes will vary according to the  column media and size  and the flow rate     Gradient Method     Select the standard gradient type   Figure 2 8        Linear     The system gradually mixes solvents A and B  from the last programmed  B until it reaches the next  programmed  B  This method connects the inflection  points that define the gradient with a straight line       Step     The system holds each  B value until it reaches  the next programmed  B  Because the system holds the   B values  the changes at each new value are nearly  instantaneous  producing a stepped appearance on the  plot area        Section 2 Preparation  Rf 200 and Rf 200i Systems      Isocratic     The system holds the initial solvent mix for  the entire run     Linear Step Isocratic    Figure 2 8 Gradient types    Enable Run Length Extension     When enabled  this option  automatically extends the run if a peak is eluting at the end of  the maximum  B gradient  This ensures that a late eluting  peak fully comes off the column
96. e with the run     Solid     Select this option if you have already  prepared the sample  placed it into a solid sample  load cartridge  and loaded the cartridge on the  system  You must also specify the cartridge type and  size  The system will automatically proceed until the  end of the programmed run length        6 34    Section 6 User Help Reference    e Liquid     Select this option if you have prepared a  liquid sample and plan to manually inject it into the  injection port after column equilibration       None  on column    Select this option if you have  preloaded the sample on the column  The system will  skip column equilibration so that the sample will not  be flushed from the column before the run     Loading types for the Rf 75 system are       Equilibration     Select this option if you intend to  equilibrate the column before introducing the liquid  or solid sample       No Equilibration   Select this option to skip column  equilibration     5 Review or update the Start Rack and Start Tube     6 Review the Minimum Run Requirements  The system will  report the estimated solvent volumes  expected waste   collection tube usage  etc  You can use this information to  verify that there will be enough solvent to complete the  run  the waste volume does not exceed the collection  container s capacity  and whether or not more tubes will be  required during the run     Note  The Rf 200  Rf 200i  and Rf 4x systems estimate  the solvent volumes by monitoring the solvent l
97. ecomes more pronounced  as the ambient temperature increases     2 5 Connect Waste Lines     NWARNING  Risk of fire or equipment damage  Failure to connect Waste  Porttubing may allow organic solvents to pool in unsafe areas   possibly creating dangerous levels of flammable vapors      ANWARNING    Risk of fire ignited by electrostatic discharges  Never  substitute the black tubing on CombiFlash systems  The black  tubing  P N 023 0503 06  is conductive  This tubing is required  to dissipate static electricity      NCAUTION  Elevated flammable vapor levels are possible  Ensure that the  waste container is adequately ventilated  preferably by placing it in  a fume hood     The system has two waste outlets on the back panel  Diverter  Valve waste and Inject Valve waste  You can route waste to a  common collection container  or to individual containers     The Waste Level Sense uses air pressure to detect the liquid  level in the common waste container  If using individual  containers  use the waste level sense to detect the level of the  Diverter Valve waste container  Under normal operating  conditions it will fill faster than the Inject Valve waste  container  Empty both containers at the same time to prevent  overfilling the unmonitored container        2 6    Section 2 Preparation  Rf 200 and Rf 200i Systems    To Diverter Valve Waste  Connection    To Waste Level    Sense Connection m To Inject Valve Waste  Connection    60 0923 017 Ferrule    60 0923 015  Compression Fitt
98. ection  port above the column  When you click the OK  button  the system equilibrates the column and then  waits  When instructed  perform a liquid injection as  described in steps a through d above  or attach the  solid sample cartridge to the injection port and  connect the solvent delivery tubing  Click the OK  button to continue     e Ifyou selected No Equilibration  ensure that the  solvent delivery tubing is connected  The system  starts the run immediately and assumes that the  sample has been injected on the column or the solid  load cartridge is in place     5 4 During the Run    You may allow system to proceed with the run while  monitoring the progress on the PeakTrak main window  If  desired  gradient parameters may be modified during the run    See the Real time Gradient Editing on line help topic      As the run progresses  the absorbance trace is drawn on the  Gradient Plot area  The separation or purification continues  until the end of the run defined by the run length setting     During the run  data is automatically saved every few seconds   PeakTrak automatically saves all run history on the system   s  internal hard drive in its native  RUN format  You can  manually save a Run file as Text or PDF while stopped or any  later time as long as the  RUN file has not been deleted   Selected files can be manually saved to a connected Flash  drive  To automatically save a Text or PDF file after each run   configure the Network File Save feature  see Section 7 1 3
99. ement   Choose this command to open the USER MANAGEMENT  6 3 12   window  Use this window to add or remove users from the  system  This feature allows you to create a working folder on  the system for each user  Use the FILE gt SET DATA PATH      6 2 1 4  command to select your folder     To prevent unauthorized changes to user management  this  function is password protected     6 2 3 8 Calibrate Touchscreen   Choose this command from the touch screen to calibrate the  interaction between touch screen display elements and stylus  location     The calibration routine places a series of targets on the touch  screen display  Touch the exact center of each target with a  stylus  The routine will move clockwise through 20 calibration  targets  after which the system will return to the normal  display     During calibration routine  ensure that no part of your hand or  anything other than the stylus touches the screen     Note  If you select this command from a remote connection   the system will start the calibration routine  Invoking this  from a remote connection may be useful if the system is so  mis calibrated that this menu command cannot be selected  from the touch screen        6 8    Section 6 User Help Reference    6 2 4 Help  PeakTrak s Help menu has the following options     6 2 4 1 Help  Choose this command to view the Help documentation     6 2 4 2 Service   Qualified service personnel use this command to access the  system s service functions  These service function
100. ems  found  it is likely that organic vapors are present in the ambi   ent environment of your laboratory  In this case  increase the  setting of the Vapor Limit  Settings greater than 25 are appro   priate for laboratory environments with a somewhat elevated  background solvent vapor concentration  representing a safe  setting  but one with a reduced margin for error     2 12 5 Set Default Tube Volumes  Click this button to open a window from which you can set the  default volume of the collection tubes  You can increase or  decrease these values to set the desired default volume  collected in each container  Just be sure you don   t overflow the  containers by exceeding their capacity     The system will automatically advance to the next tube when  this default volume is met in each tube  Keep in mind that  advanced method settings  discussed in Section 5 and the  on line help  may override this default setting  During a run        2 17    CombiFlash   Rf User Manual    other factors such as automatic tube advances for detected  peaks will affect collected volume     2 12 6 User Preferences    View the User Preferences tab to configure system operation  for the current user     New systems will only have one user account named    common        Language   Select the user s desired language from the  drop down list box     Automatically Print Report at End of Run     if this option  is enabled  the system will print a report at the end of each  run     This option requires th
101. er the solvent gradient reached 30  B     The calculation is conservative  but allows you to minimize the  time needed to keep the solid load cartridge in the fluid path        5 13    CombiFlash   Rf User Manual    5 4 3 About Solvent Level Detection    When enabled in the TOOLS gt Configuration window    CombiF lash Rf 200  Rf 200i  and Rf 4x systems will estimate  the solvent volume usage by measuring the solvent level in the  container and the monitoring the usage rate  The system  continues to refine this estimate of available solvent during  operation     At the beginning of a run  the system calculates the amount of  solvent required to complete the run  The system will alert you  if it detects that there is not enough solvent at the start of the  run  Likewise  the system will alert you during the run if the  estimated usage rate or the measured level indicates that  there is insufficient solvent  Note that pauses and automatic  run extensions during a run will increase the required solvent  volume     The options available on the alert message will vary  depending on when the condition occurs  before or during a  run  and which method  measured level or estimated usage  rate  indicated the condition     To clear the alert message  the user must select one of the  options below       Continue   Select this option after you have added  more solvent to the container  After selecting this option  the system measures the solvent level again  If the level  is sufficient to
102. ere T is the time in minutes to wait after the system  reaches the  B solvent used to run the TLC plate  Rf is  the retention factor measured in step 1  Cartridge  Volume is the volume in mL from step 2  and Flow Rate  is the current flow rate in mL min     5 Observe the current  B concentration in the status bar   Once the system reaches the  B percentage used for the  TLC  wait at least the time established as T  After this time  has elapsed it is likely that the desired compound has come  off the cartridge and you can safely click the Cartridge  Bypass button     For example  a TLC plate run at 30 B revealed that the last  desired compound had an Rf factor of 0 35  An empty 25 gram  size solid load cartridge was filled with sample adsorbed onto  10 grams of silica  and then placed on the system  The  programmed flow rate was 60 mL min  During the run  the  flow rate dropped to 30 mL min  The equation is then     T 2 45 mL    0 35   30 mL min    4 3 minutes  In this example  the Cartridge Bypass feature was used 4 3  minutes after the solvent gradient reached 30  B     The calculation is conservative  but allows you to minimize the  time needed to keep the solid load cartridge in the fluid path     6 4 8 Monitoring the Wavelength Ratio    When using two peak detection options  such as A1 with   2 or  All Wavelength Detection with External  a ratio of the  detection sources can also be displayed  which at times can  provide the best indication of compound purity  Refer to th
103. ese  features described in A1 and 42 above    Spray Chamber Temperature   The default  settings are 50   C  normal phase default methods   and 15   C  reverse phase   This setting may be  adjusted from 10 to 60   C so that detection can be  optimized for the solvent system in use  This setting  is limited to a minimum of 5   C below the ambient  temperature    Drift Tube Temperature     The default settings are  55   C  normal phase default methods  and 60   C   reverse phase   This setting may be adjusted from 30  to 90   C  but is limited to a range of 5   C below and 60    C above the ambient temperature       Show Ratio     If using two of the above peak detection  options  check this option to display a ratio of the  selected wavelengths  The ratio trace is often a useful       6 41    CombiFlash   Rf User Manual    indicator of purity  Refer to the MONITORING THE  WAVELENGTH RATIO  6 4 8  topic for more details on this  feature     Rf 75 and Rf 150 systems have a fixed UV detection  wavelength of 254 nm  The peak detection settings for these  systems include slope based and peak width options  These  are described in A1 and A2 above  Rf 75 and Rf 150 systems  with the variable wavelength option include a control to type  or set the detection wavelength     6 4 2 3 Saving Changes to the Method File   After you have edited the method file  you can save the  changes for future use  Click the Save As button  give the  method file a descriptive name  then click the Save butt
104. ethod File    PeakTrak controls the separation runs performed by the  system through a Method file  The system uses the method file  to direct the system operation after you click the Play button     PeakTrak categorizes the method settings as Basic or  Advanced  Basic settings are the frequently used controls  accessed through the MAIN WINDOW  6 3 1   Use the METHOD  EDITOR  6 3 2  window to access the Advanced settings     Method files are stored by the system and can be opened for  review  reuse  or modification  To open a method file  use the  FILE gt OPEN  6 2 1 2  menu command  Method files can be  transferred to the system from an external storage device  using the FILE gt IMPORT METHOD  6 2 1 9  menu command   Method files use an   mtd filename extension     6 1 2 1 Default Methods   With the exception of the Rf 150  when PeakTrak is started   and every time you insert a RediSep Rf column  the system  loads a default method  A default method contains Teledyne       6 2    Section 6 User Help Reference    Isco s recommended basic and advanced settings for the use of  RediSep Rf columns on the system  There is a default method  for each size and media type of RediSep Rf column  See  REDISEP RF COLUMN SELECTION GUIDE  ON LINE      Default methods provide a starting point for your separation  or purification  From these initial settings  you can perform a  purification run  or you can modify the settings for your next  run  Subsequent runs will use the settings in the act
105. evel in  the container and the known usage rate  The system  continues to refine this estimate during operation  See  also  ABOUT SOLVENT LEVEL DETECTION  oN LINE    Note  If the waste level sensing tube is inserted  correctly into the waste container  the system  automatically will suspend operation before an overflow  condition might exist  To prevent the run from being  suspended before completing the run  ensure that the  container will hold the expected waste volume  Waste  level sensing is not available on the Rf 75 system       Rf 200i only  If the ELSD will be used for the run  ensure  that the ELSD option is selected  Otherwise  you may  disable the ELSD  The Rf 200i system will automatically  check this option if the previous run used the ELSD        6 35    CombiFlash   Rf User Manual    8 Click OK and the run begins  The system responds  according to the sample Loading Type you selected in  step 4     Rf 200  Rf 200i  and Rf 4x systems    If you selected Solid  Pause   the system equilibrates  the column with the starting  B and then waits while  you prepare the solid sample load cartridge  Place the  cartridge on the system and click OK to continue    If you selected Solid  the system equilibrates the  column and immediately starts the separation    If you selected None  on column   the system skips the  equilibration and immediately starts the separation   If you selected Liquid  the system equilibrates the  column and then waits  When instructed  perform the  f
106. f 200 only      Vacuum or  pressurized air  applied to the outlet  end of the drain  tube must exist at  the top shelf drain  hole     Figure 2 5 Top Shelf Drain    3  Route the end of both drain tubes to a suitable waste fluid  collection container        Section 2 Preparation  Rf 200 and Rf 200i Systems    It may be necessary to extend the drain tube  If so  splice the  tubing with user supplied tubing  The user supplied tubing should  have an inside diameter no smaller than the existing drain tubing   and must be compatible with the solvents used by the system   Route this extension tubing to the waste collection vessel      CombiFlash Rf 200i only  The CombiFlash Rf 200i system  has additional drains that must be connected     4  Fill a syringe with 1 mL of liquid such as water or isopropyl  alcohol and inject it into the P Trap port     5  Locate the l4 inch I D  silicone tubing  part number  029 1312 02  and a nylon tubing clamp  part number  149 2002 08  from the accessory kit  Connect one end of the  tubing to the P Trap port and secure it with the clamp     6  Route the end of the P Trap drain tube to a suitable waste  fluid collection container     M Note  If accessible  the outlet end of the P Trap drain tube may be used  to refill the P Trap     7  Connect user supplied tubing  1 2 inch I D   to the exhaust  port and route the exhaust away from the system  This will  prevent unnecessary solvent vapor alarms     M Note  The exhaust gas is dry and does not require a
107. format     6 1 PeakTrak Overview    PeakTrak is the control software for Teledyne Isco s  CombiFlash Rf system  This overview provides a description  of       The PEAKTRAK WINDOW ELEMENTS  6 1 1     The METHOD FILE  6 1 2      The RUN FILE  6 1 3      HOW TO GET PEAKTRAK HELP  6 1 4     6 1 1 PeakTrak Window Elements  The MAIN WINDOW  6 3 1  displays the current METHOD FILE   6 1 2   From this window you can access many of PeakTrak s  features  view the system status  and view or edit the method  file settings  The elements of the main window are the       Menu     The topmost item in the window is the  PEAKTRAK MENU OPTIONS  6 2  from which you can  access all of PeakTrak s features        6 1    CombiFlash   Rf User Manual      System Status     PeakTrak displays the system status  to the right of the current method  Status messages may  include     e The system mode      current position in a run  expressed in time or column  volumes      current  B     peak status     flow rate     current time and date       Main region     This area is where you can view or  modify method file settings  Frequently used method  settings are displayed on the MAIN WINDOW  6 3 1    Advanced method settings are displayed on the METHOD  EDITOR  6 3 2  window     Many of the PEAKTRAK WINDOWS  6 3  contain these main  elements  or a subset of them  Note that the available  commands and options will change according to the active  window  the state of the system  and the current file     6 1 2 M
108. g from the  Inject Valve Waste should remain attached to the waste  bottle     2  At the CombiFlash Rf system back panel  disconnect the  black tubing from the Diverter Valve Waste port and the  natural tubing from the Waste Level Sense port        4 5    CombiFlash   Rf User Manual    The Inject Valve Waste tubing must remain connected between  the Rf system back panel and its waste container     Refer to Figures 4 3 and 4 4 for steps 3 and 4     3  Locate the 20 inch  50 cm  long Column Waste Assembly   P N 60 5234 192  in accessory package 60 5239 008   Connect this tubing between the Combiflash Rf system  and the Rf 4x module    a  Connect the cut end of the black tubing to the  CombiFlash Rf system Diverter Valve Waste port  Use a  black nut and ferrule from the accessory package   Finger tighten the nut to seat the ferrule in the  bulkhead fitting     oj        Figure 4 3 Preparing a bulkhead fitting connection    Ferrule  Nut  Tubing          b  Connect the other end of the black tubing to the Rf 4x  port labeled    From Rf   in the block labeled    Column  Waste    Finger tighten the threaded fitting     4  Locate the 20 inch  50 cm  length of natural tubing from  the accessory package  Using nuts  60 0923 015  and  ferrules  60 0923 017   connect this tubing between the  CombiFlash Rf system s Level Sense port and the Rf 4x  module s Waste Level Sense From Rf port         Level Sense  Waste Solvent  air     Figure 4 4 CombiFlash Rf system to Rf 4x waste tubing connec
109. gned  hereby declare that the design of the equipment specified above conforms to the above Directive s  and  Standards as of April 12  2007     William Foster  USA Representative    Teledyne Isco Inc     William Foster   Director of Engineering  Teledyne Isco  Inc    4700 Superior Street  Lincoln  Nebraska 68504    Phone   402  464 0231  Fax   402  464 4543       60 5232 141                               DECLARATION OF CONFORMITY    I8M1 A          Application of Council Directive     Manufacturer s Name   Manufacturer s Address     Telephone    Facsimile    Equipment Type Environment   Trade Name Model No    Year of Issue    Standards to which Conformity is    2004 108 EC     The EMC Directive  2002 96 EC     The WEEE Directive  2006 95 EC     The Low Voltage Directive  Teledyne Isco  Inc    4700 Superior   Lincoln  Nebraska 68504 1398 USA  P O  Box 8253   Lincoln  NE 68501 2531    1  402  464 0231    1  402  465 3799   Laboratory Equipment for Light Industrial Commercial Environments   Rf200i   2012    EN 61010 1  2010 Safety Requirements for Electrical Equipment for                                     Declared  Measurement  Control  and Laboratory Use  EN 61326 1 2006 EMC Requirements for Electrical Equipment for  Measurement  Control  and Laboratory Use  EMC Standard Description Severity Applied Performance Criteria  EN 61000 4 2 Electrostatic Level 2     4kV contact discharge A  Discharge Level 3     8kV air discharge  EN 61000 4 3 Radiated RF 80 MHz to 2 7 GHz 80  AM a
110. he  Threshold condition is true       Monitor     Enable this option to only use the  detection source as a monitor  When enabled  the  detection source will be displayed as a trace on the  gradient plot area  but will not be used to cut peaks        6 40    Section 6 User Help Reference    e All Wavelength Detection     Enable this option to  detect peaks within a user selected range of  wavelengths  When enabled  click the Details button to  configure additional settings  These settings include the  slope based and peak width options described in 41 and  22 above  Additionally  you can type or select the  minimum and maximum wavelengths limits in  nanometers     e External Detector      Rf200  Rf200i  and Rf4X  systems only  Enable this option to use a 0 to 1 volt  analog input signal from an external detector  When  enabled  click the Details button to configure additional  settings  These settings include the slope based and  peak width options described in A1 and 42 above  The  system will use its internal algorithms to cut detect and  cut peaks based on the analog input signal  Refer to the  External Detector instruction sheet for cable and  plumbing requirements       ELS Detector     Enable this option to use the  evaporative light scattering  ELS  detector on an Rf 200i  system  When enabled  click the Details button to  configure additional settings     Signal Gain     You may modify the gain to scale the  detector signal    Slope Based and Threshold     Refer to th
111. he ambi   ent environment of your laboratory  In this case  increase the  setting of the Vapor Limit  Settings greater than 25 are appro   priate for laboratory environments with a somewhat elevated  background solvent vapor concentration  representing a safe  setting  but one with a reduced margin for error     3 10 5 Set Default Tube Volumes    Click this button to open a window from which you can set the  default volume of the collection tubes  You can increase or  decrease these values to set the desired default volume  collected in each container  Just be sure you don   t overflow the  containers by exceeding their capacity     The system will automatically advance to the next tube when  this default volume is met in each tube  Keep in mind that  advanced method settings  discussed in Section 5 and the  on line help  may override this default setting  During a run   other factors such as automatic tube advances for detected  peaks will affect collected volume     3 10 6 User Preferences    View the User Preferences tab to configure system operation  for the current user     New systems will only have one user account named    common        Language   Select the user s desired language from the  drop down list box        3 14    Section 3 Preparation  Rf 150 Systems    Automatically Print Report at End of Run   if this option  is enabled  the system will print a report at the end of each  run     This option requires the system to be configured for network  operation and 
112. hich you can set the default volume of the collection  tubes  You can increase or decrease these values to set the  desired default volume collected in each container  Be sure you  do not overfill the containers by exceeding their capacity     The system will automatically advance to the next tube when  this default volume is met in each tube  Keep in mind that the  PEAK COLLECTION  6 3 2 5  settings on the METHOD EDITOR  6 3 2   window may override this default volume  During a run  other  factors such as automatic tube advances for detected peaks  will affect collected volume     See also  LOADING A COLLECTION RACK  oN LINE         6 29    CombiFlash   Rf User Manual    Enable Automatic Column Purchase This feature requires a  current subscription in a Teledyne Isco Column Plan  Contact  your sales representative for more information on this feature   The certificate issued with the column plan includes  instructions for activating this feature     Multiple Wavelength Detection  Rf 75 and Rf 150 Only  This  feature requires a valid activation code for the variable  wavelength option  Contact your sales representative for more  information on this feature     6 3 10 2 Network Configuration   Network configurations settings require assistance from your  network administrator  Contact your Information Technology  department before changing the settings     Instrument IP Address Configuration Use these settings to  specify the instrument s network address and its parameters    
113. hod Files      6 4 4 Editing a Default Method                  6 4 5 Defining a Gradient                  05   6 4 6 Real time Gradient Editing                 6 4 7 Bypassing the Solid Load Cartridge           6 4 8 Monitoring the Wavelength Ratio            6 4 9 Viewing runs           llle  6 4 10 Manual Control of the CombiFlash Rf        6 4 11 Setting the Vapor Limit                      7  Remote Interfaces    7 1 Network Configuration                   005   7 1 1 Network PC Access              0000005       CombiFlash   Rf User Manual    7 1 2 Network Printing                    04  7 3  7 1 3 Network File Save Configuration           7 3  7 1 4 Apple Mobile Digital Devices               7 4  7 2 Direct Connection              0 00 RR RR KK KK 7 4  7 2 1 Windows 7 Settings    KK KK 7 5  7 2 2 Windows Vista Settings                   7 7  7 2 3 Windows XP Settings                     7 8  7 2 4 Windows 2000 Settings                  7 10  7 2 5 Completing the Direct Connection          7 12  7 3 Direct Printing              0 0 0 eee 7 12  8  Maintenance  8 1 System Standby and Shut Down                 8 1  8 2 Preventive Maintenance                   0085 8 2  8 3 Cleanibg   2 3 02 dae DA oH Sen eae s 8 2  8 3 1 Collection Rack and Tray Cleaning          8 3  8 4 Tubing Inspection              KK RR RR RR KK 8 3  8 5 Flow Cell Cleaning              KK RR KK KK KK 8 5  8 5 1 Post Separation                00000 ee 8 5  8 5 2 Quick Cleaning when Recommended        8 6
114. ica gel columns     e Part number 60 5237 051  Test kit containing four of the 25 gram size solid load  cartridges pre filled with 0 5 grams of test sample  NPHE  and four 4 gram RediSep Rf silica gel columns     The test kits include instructions        3 19    CombiFlash   Rf User Manual      3 13 Installation Qualification Checklist    Table 3 1 Installation Qualification Checklist    Installer Operator  Description Initials Initials    3 1 UNPACKING THE UNIT       3 2 INSTRUMENT LOCATION       3 3 CONNECT POWER       3 4 CONNECT SOLVENT LINES       3 5 CONNECT WASTE LINES       3 6 CONNECT AND ROUTE DRAIN  LINES       3 7 POSITION THE SYSTEM       3 8 INSTALL COLLECTION TUBE  RACKS       3 9 TURN ON POWER       3 10 CONFIGURE THE SYSTEM       3 11 RESTART THE SYSTEM       3 12 SYSTEM VERIFICATION                Certification of Section 3 Completion       Installer Name  print         Installer Signature   Date           Operator Name  print         Operator Signature           Date     Comments                          3 20    CombiFlash   Rf  User Manual       Section 4 Preparation   Optional 4x Module    This section provides instructions for unpacking and installing  the Combiflash Rf 4x module after completing the  CombiFlash Rf 200 or Rf 200i system installation  Section 2    To prepare the system for operation  sequentially follow all  instructions in sections 4 1 through 4 10     Section 4 12 contains an Installation Qualification checklist  If  required  sign off 
115. ide panels     If there is any damage to the shipping carton or any  components  contact the shipping agent and Teledyne Isco  or  its authorized representative  immediately        CombiFlash   Rf User Manual     NWARNING    If there is any evidence that the system has been damaged in  shipping  do not plug it into AC power  Contact Teledyne Isco  or its authorized representative for advice     Compare the contents of the boxes with the enclosed packing  slips  If there are any shortages  contact Teledyne Isco  immediately     The fraction collector arm was stowed to prevent damage  during shipping  The arm stowing bracket should be removed    at this time   1  Loosen the bracket s two thumbscrews along the sides of  the arm  Figure 2 1                               Thumbscrew  Thumbscrew    Figure 2 1 Remove arm stowing bracket and spacer    2  Push the arm to the left or right remove the spacer     3  Store the bracket and spacer  These parts should be  reinstalled if the system must be shipped again     2 2 Instrument Location    The CombiFlash Rf 200 and Rf 200i systems have a relatively  small footprint  requiring about 1550 square centimeters  240  in   of level bench space  Ensure that the system has at least 3  cm  1 25   of air space behind it for ventilation  Additional  space may be required for solvent and waste containers     The optional 4x expansion module is installed to the left of the  system  This module requires an additional 933 cm   145 in         2 2   
116. in  Standby       lni xl       ge             Figure 7 8 LAN or High speed Internet Connection       7 8    Section 7 Remote Interfaces        1  Local Area Connection Properties          Internet Protocol  TCP IP  Properties          Figure 7 5 Alternate Configuration Settings       CombiFlash   Rf User Manual    7 2 4 Windows 2000 Settings    1  From the Windows Start button  select    Settings gt Control  Panel    to view the control panels     2  Locate the    Network and Dial up Connections    icon and  open this control panel     3  Highlight the    Local Area Connection    icon  View the  properties by right clicking and selecting    Properties     The  Local Area Connection Properties Window will open   Figure 7 6      4  Highlight the    Internet Protocol  TCP IP     item in the list   and click on the    Properties    button  The Internet Protocol  Properties Window will open  Figure 7 7      The following steps change the local area connection settings  If  the PC must be restored to its original network configuration   record the present Internet Protocol  TCP IP  settings now     5  Click the    Use the following IP address    option   6  Enter the IP address    192 168 1 200      7  Enter the default Subnet Mask    255 255 255 0        8  Click the OK button to close the Internet Protocol  properties window  Then  click the OK button to close the  Local Area Connection properties window     9  Shut down the PC and place the CombiFlash Rf system in  Standby      
117. ing Nut        60 5234 143  Waste Tubing Assembly    Ventilation Hole    Luer fitting    If necessary  connect   user supplied tubing and    See Top View  route to fume hood    Tubing Stop   ferrule  60 5235 053 38 mm Waste Cap  or   60 5235 054 45 mm Waste Cap    about 2 inches         5 cm      Top View     Diverter Valve Waste  Waste Level Sense O e Ventilation  Inject Valve Waste       Figure 2 8 Recommended waste connections  with plastic plate       2 7    CombiFlash   Rf User Manual    The accessory package includes parts to complete the waste  connections  Figure 2 3 shows the recommended connections  for user supplied waste containers with either a 38 or 45 mm  opening  GL 38 or GL 45      Refer to the steps in section 2 4 to connect the tubing to the  system using the compression fittings  Then  complete the  connections  refer to Figure 2 3      1  Push the three tubes through the cap  The tubing will fit  snugly in the cap and may require some force     2  Push the Waste Level Sense tubing until the Tubing Stop  is against the cap  This ensures that the outlet is about two  inches  5 cm  below the container s maximum level     Should the container not have a GL 38 or GL 45 opening  omit  the use of the waste cap  Ensure the solvent tubes are secured  so they drain into the waste container  Secure the Waste Level  Sense air tubing so its outlet is at least two inches  5 cm   below the container s maximum level  Seal the container  opening to avoid solvent vapors   
118. ired instrument to the customer  Teledyne  Isco will not pay air freight or customer s packing and  crating charges  This warranty does not cover loss  damage   or defects resulting from transportation between the  customer s facility and the repair facility     The warranty for any instrument is the one in effect on date  of shipment  The warranty period begins on the shipping  date  unless Teledy ne Isco agrees in writing to a different  date     Excluded from this warranty are normal wear  expendable  items such as charts  ribbon  lamps  tubing  and glassware   fittings and wetted parts of valves  and damage due to  corrosion  misuse  accident  or lack of proper maintenance   This warranty does not cover products not sold under the  Teledyne Isco trademark or for which any other warranty is  specifically stated     This warranty is expressly in lieu of all other warranties  and obligations and Teledyne Isco specifically disclaims  any warranty of merchantability or fitness for a  particular purpose     The warrantor is Teledyne Isco  4700 Superior  Lincoln   NE 68504  U S A       This warranty applies to the USA and countries where Teledyne Isco does not have an authorized dealer  Customers in countries  outside the USA  where Teledyne Isco has an authorized dealer  should contact their Teledyne Isco dealer for warranty service        Before returning any instrument for repair  please call  fax  or e mail the Teledyne Isco Service  Department for instructions  Many problem
119. isplayed     2 13 Prime the Solvent Lines    Before the first use  the system should be primed  Ensure that  the solvent containers are filled  then     1  Select the    Tools gt Auto prime    menu command     2  Select the desired A and B solvents from the drop down list  boxes     3  Click the Play button to start priming the system     The system pumps 100  Solvent B through the system and  out the Waste port  and then repeats with 100  Solvent A    This B A order leaves the internal lines filled with solvent  more appropriate for the beginning of a purification run      When priming the first time  inspect the solvent and waste  connections to the system  If any leaks are observed  click the  Cancel button to stop the Auto Prime  Correct the leak by  tightening the fitting an additional 1   4 turn  and then restart the Auto  Prime from step 1     To advance to the next Auto Prime step  you can click the  Fast Forward button  The Auto Prime window closes when  finished     After Auto Priming  the system is ready for operation   Section 5      Use Auto Prime for quick solvent changes by pumping a fixed  amount of solvent through the waste fluid path at 100 mL min  For  more advanced priming functions such as varying the flow rates   fluid paths  etc   or to purge the system with air before changing  between normal and reversed phase solvent systems  use the  Tools gt Manual Control option        CombiFlash   Rf User Manual    2 14 System Verification  It is recommended that
120. ive  window  If you have modified the settings  you can save the  method file for future use     If you find that the default methods are not a practical  starting point for your applications  the default methods can  be changed to meet your specific requirements  See EDITING A  DEFAULT METHOD  6 4 4      6 1 3 Run File    When the system has finished a run  it saves the run history  in a Run file  The history includes the METHOD FILE  6 1 2   settings  rack and tube information  and a chromatogram  You  can open and review the Run files stored by the system  To  open a Run file  use the FILE gt OPEN  6 2 1 2  menu command   Run files use an   run filename extension     6 1 4 How to get PeakTrak help  PeakTrak help can be displayed by selecting Help   Help from  PeakTrak s menu     If you need additional help  refer to Teledyne Isco s CONTACT  INFORMATION  ON LINE      6 2 PeakTrak Menu Options    PeakTrak menus include     FILE  6 2 1       METHOD EDITOR  6 2 2       TOOLS  6 2 3     HELP  6 2 4   6 2 1 File    The file menu lists the following commands        6 3    CombiFlash   Rf User Manual    6 2 1 1 New  Choose the New command to open a new method file     Rf150   The main window opens using the currently selected  method     Rf200 and Rf200i   PeakTrak will open the main window  using the default program settings for the detected column  size     6 2 1 2 Open   Choose this command to open a METHOD FILE  6 1 2    mtd  or a  RUN FILE  6 1 3    run  stored on the system 
121. k drive        1 6    Section 1 Introduction  1 4 Specifications    Table 1 1 CombiFlash Rf 150 and Rf 200 System  1    Specifications    Rf 150 Systems Rf 200 Systems   Without Inject Valve   With Inject Valve     Overall Dimensions 61 x 36 x 43 cm  24 x 14 1 x 17     HxWxD           Weight 22 3 kg  49 2 Ibs  24 5 kg  54 Ibs        Power Options Input voltage range from 100 to 240 VAC  50 60 Hz   2 0 Amps maximum     Line cord is the disconnect device        Line Frequency 50 60 Hz       Ambient Temperature 20 to 40   C  maximum temperature must be at least  15   C below the boiling point of the solvent                       Humidity  when 95  relative humidity maximum at 20 to 40   C   connected to power    Flow Rate Range 10 to 100 mL min 5 to 200 mL min   Flow Rate Accuracy  10   5     tested with water at   138 kPa or 20 psi    Pressure Limit 2 Up to 1379 kPa  200 psi   1 034 kPa  150 psi   column dependent    Pressure Accuracy 10  of full scale 5  of full scale   Gradient Formation Binary gradient from A1 Binary gradient from A1   and B1 solvent inlets  A2  and B1  B2 solvent  Optional manual valves inlets  Solvent selection  to select from A1  A2  can be automatically  and B1  B2 inlets  controlled by the method    Gradient Accuracy   2  of full scale  typical   1  of full scale   5  maximum          Peak Detection Modes   Slope or threshold       Flow Cell Pathlength 0 1 mm   25              1 7    CombiFlash   Rf User Manual    Table 1 1 CombiFlash Rf 150 and Rf 200 
122. k nut and ferrule in the accessory  package  Slide the nut and ferrule onto the tubing as  shown in Figure 3 3     Ferrule Tubing  Nut          3 6    Section 3 Preparation  Rf 150 Systems    Figure 3 3 Preparing a bulkhead fitting connection    b  Fully insert the tubing into the threaded bulkhead  fitting on the back panel of the system    c  Finger tighten the nut onto the threaded bulkhead  fitting  This will seat the ferrule in the fitting     2  Route the tubing and insert the weighted end into the  user supplied waste container     3 6 Connect and Route Drain Lines    Z  WARNING    Risk of fire or equipment damage  Failure to connect drain lines  may allow organic solvents to pool in unsafe areas  creating a  potential for dangerous levels of flammable vapors  Improper  draining may damage the instrument s internal components        3 7    CombiFlash   Rf User Manual    The CombiFlash Rf 150 system has drain tubes extending  from its top shelf and back panel  The tubes drain away any  liquid spilled on the top shelf and the tray beneath the fraction  collection racks     1  Test the fraction collector drain by connecting a vacuum or  air supply source to the outlet end of the drain tube  Then   verify the presence of such vacuum or air supply source on  the drain  Figure 3 4      Vacuum or   pressurized air   applied to the outlet   end of the drain   tube must exist at     the collection tray  drain hole                                      Figure 3 4 Fraction Collect
123. l any run and view it from the system as long as  the  RUN file has not been deleted  To save the post run report  for viewing elsewhere  you can save the run history as text or  PDF files as needed  This Disabled option is useful when there  is no need  or only an occasional need to save a run history file  externally     To save all run histories automatically as a text or PDF file   configure the Network File Save option  This allows the  CombiFlash Rf system to access the corporate network and  save the file in a selected Network Share folder  Consult with  your network administrator to configure these fields        7 3    CombiFlash   Rf User Manual    7 1 4 Apple Mobile Digital Devices    You can wirelessly monitor and control a CombiFlash Rf  system using the PeakTrak app  The app is designed for a  mobile digital device  iPod Touch  iPhone  and iPads  running  iOS 4 2 or later  The PeakTrak app is available for download  from the App Store     Each CombiFlash Rf system requires an active license before  using the PeakTrak app  A license may be ordered from  Teledyne Isco using part number 60 5237 073 which contains  the necessary software on a USB flash drive     Refer to the instructions supplied with 60 5237 073 to install  the license key and configure the system and mobile digital  devices to support remote control     M Note   Remote communication requires the CombiF ash Rf system and  mobile digital device to share a common network with Wi Fi  access in the remo
124. lash   Rf User Manual    the system is ready to perform a run using the default  settings  These default settings are preset to the most common  configurations used by chemists  and have been optimized for  the use of RediSep Rf columns  The default settings will       collect all fluid in the fraction collector rack    e differentiate between UV  254 nm  vs  non UV absorbing  fluid and place the fluids into separate tubes      cut UV absorbing peaks based on slope detection or level  threshold     To start the run from the PeakTrak Main window     1 Enter a sample name  If you choose not to  PeakTrak will  enter the date and time as the sample name when you start  the run     2 Review the PeakTrak Main window settings  If you want to  change any settings  refer to the Editing a Method on line  help topic     3 Click the Play button     4 Select the sample Loading Type from the list  The Loading  Type should support the sample injection technique you  intend to use  see SAMPLE INJECTION METHODS  oN LINEI    Possible sample loading types for the Rf 200  Rf 200i  and  Rf 4x systems are     Solid  Pause      Select this option if you intend place  the sample into a solid sample load cartridge  but  have not yet prepared the sample  You must also  specify the cartridge type and size  When you click  the run button  the system will perform a column  equilibration and then wait while you prepare the  cartridge  After you have placed the cartridge on the  system  click OK to continu
125. leaves the internal lines filled with solvent  appropriate for the beginning of a purification run     9  Repeat steps 4 through 8 for channels 3 and 4  Be sure to  change the channel option before starting the pumps        CombiFlash   Rf User Manual    4 11 System Verification    It is recommended that the system operation be verified for  each channel  To do so  use a pre packaged CombiFlash Rf test  kit   e Part number 60 5237 050  Test kit containing five of the 5 gram size solid load  cartridges pre filled with 0 5 grams of test sample  NPHE  and five 4 gram RediSep Rf silica gel columns     e Part number 60 5237 051  Test kit containing four of the 25 gram size solid load  cartridges pre filled with 0 5 grams of test sample  NPHE  and four 4 gram RediSep Rf silica gel columns     The test kits include instructions        4 14    Section 4 Preparation  Optional 4x Module    4 12 Installation Qualification Checklist    Table 4 1 Installation Qualification Checklist    Installer Operator  Description Initials Initials    UNPACKING THE UNIT       INSTRUMENT LOCATION       CONNECT POWER       CONNECT SOLVENT LINES       CONNECT WASTE LINES       CONTROL CONNECTION    POSITION THE SYSTEM       REMOVE SHIPPING HARDWARE       TURN ON POWER       PRIME THE SOLVENT LINES       SYSTEM VERIFICATION                Certification of Section 4 Completion       Installer Name  print         Installer Signature   Date           Operator Name  print      Operator Signature   Date        
126. ludes self diagnostics     2  Monitor the movement of the fraction collector arm  The  arm should move to the left rear corner of the fraction  collector area  You should not hear any grinding noises  during the homing sequence  If grinding is heard there is a  problem with the homing of the fraction collector arm     3  The system is ready for operation when the PeakTrak  screen is displayed     2 12 Configure the System    Before operating the system  it should be configured for the  desired operation  To configure the system  open the  Configuration window by selecting Tools   Configuration from  the PeakTrak menu     The configuration window has three tabs  Instrument  Configuration  Network Configuration  and User Preferences   Network Configuration settings are discussed later in  Section 7  REMOTE INTERFACES     Adjust the following  sections 2 12 1 through 2 12 6  on the  Instrument Configuration and User Preferences tabs for the  desired operation and then click OK to save the settings     Some configuration settings do not take effect until the system  is restarted  After completing sections 2 12 1 through 2 12 6  follow the instructions in section 2 12 7     2 12 1 Solvents    When you installed the system  up to four solvents were  connected to the back panel  This section of the Instrument  Configuration tab allows you to manage these solvents  To do  S0     The system is shipped with common solvent names already  loaded  If your solvent names are listed in the s
127. module  maximum     Line cord is the disconnect device   50 60 Hz    20 to 40   C  maximum temperature must be at least  15   C below the boiling point of the solvent        Humidity  when  connected to power     95  relative humidity maximum at 20 to 40   C       Flow Rate Range    5 to 200 mL min       Flow Rate Accuracy     5        Pressure Limit 2    1379 kPa  200 psi        Pressure Accuracy    5  of full scale       Gradient Formation    Binary gradient from A1  A2  and B1  B2 solvent inlets   Solvent selection can be automatically controlled by  the method        Peak Detection Modes    Slope or threshold       Flow Cell Pathlength    0 1 mm   25        UV Detection and  Monitor Wavelength    Selectable from 200 to 360 nm range  Optional 200 to  780 nm UV Vis detector        Wavelength Accuracy     5 nm       Fraction Accuracy         2mL    flow rate   60            CombiFlash   Rf User Manual    Table 1 3 CombiFlash Rf 4x System  Specifications    Continued     Electrical Safety per EN 61010 1       Pollution Degree    2       Installation Category       Maximum Altitude       2000 meters       Note 1  All specifications are subject to change   Note 2  For columns undetected by RFID  the max pressure is limited to 50 psi     Table 1 4 Component Materials List    Component    Materials       Teledyne Isco RediSep  Rf Cartridge    Virgin Polypropylene  silica based media  polyethylene   HDPE  frits       Process Tubing    Carbon impregnated fluoropolymer tubing       
128. n the  instrument s technical reference manual        The lightning flash and arrowhead within the triangle is a  warning sign alerting you of    dangerous voltage  inside  the product     The pinch point symbol warns you that your fingers or  hands will be seriously injured if you place them  between the moving parts of the mechanism near these  symbols        gt   gt        Symboles de s  curit         Ce symbole signale l existence d instructions  importantes relatives au produit dans ce manuel        Ce symbole signale la pr  sence d un danger  d   lectrocution        Risque de pincement  Ces symboles vous avertit que  les mains ou les doigts seront bless  s s  rieusement si  vous les mettez entre les   l  ments en mouvement du  m  canisme pr  s de ces symboles     gt     gt        Warnungen und Vorsichtshinweise       Das Ausrufezeichen in Dreieck ist ein Warnzeichen  das  Sie darauf aufmerksam macht  daB wichtige  Anleitungen zu diesem Handbuch geh  ren        Der gepfeilte Blitz im Dreieck ist ein Warnzeichen  das  Sei vor  gef  hrlichen Spannungen  im Inneren des  Produkts warnt        Vorsicht Quetschgefahr  Dieses Symbol warnt vor einer  unmittelbar drohenden Verletzungsgefahr f  r Finger und  H  nde  wenn diese zwischen die beweglichen Teile des  gekennzeichneten Ger  tes geraten        gt   gt                    1 21    CombiFlash   Rf User Manual    Table 1 5 Hazard Symbols  Continued     Advertencias y Precauciones    Esta se  al le advierte sobre la importancia
129. nal  computer  the summary can be printed on any  installed printer  If you are attempting to print from  the systems touch panel display  you must first set up  NETWORK PRINTING in the NETWORK CONFIGURATION   6 3 10 2  settings       Save As PDF     Click this button to save the run  summary in PDF file format  When viewing the run  from a remote personal computer  the summary can  be saved to any connected storage device  If you are  attempting to save a PDF file from the touch panel  display  insert a USB Flash drive in the USB port  below the display panel       Save As TXT     Click this button to save the run  summary in an ASCII text file format  When viewing  the run from a remote personal computer  the  summary can be saved to any connected storage  device  If you are attempting to save a TXT file from  the touch panel display  insert a USB Flash drive in  the USB port below the display panel       Zoom In     Click this button to enable the zoom  feature  Then  click and drag across the region to be  magnified       Zoom Out     Click this button to restore the  chromatogram to the normal view       Set Scale    Click this button to open a window from  which you can set the left and right Y axis scales   These scales are controlled by the Absorbance and   B upper limits       Extract Method button     Click the Extract button to  load a new method file based on the parameters for the  run you are viewing       Close button     Click the Close button when you are  do
130. ndows enter a start time  and an end time  All times are relative to start of the run  just  following the sample injection  Then choose a peak collection  mode  If you select Peaks  the module will collect fluid only  when the time window is active and a peak is present  If you  select All  the module will collect all fluid during the time  window regardless of the peak state  Define up to three time  windows        6 38    Section 6 User Help Reference    Air Purge You can program the system to purge the column  with air after a run  This feature allows you to remove  solvents from the column before it is discarded  Enter the  desired time to purge the column  in minutes  To disable the  post run air purge  enter zero minutes     Peak Collection There are three peak collection options  All   Peaks  or None  The All option will collect peak and non peak  fluid in tubes  The Peaks option will collect only peak fluid in  tubes and divert all other fluids to the waste port  The None  option will divert all fluids  peak and non peak  to the waste  port  The None option is useful for developing custom methods  that perform a column wash  system cleaning  or similar  function where solvents and elute do not need to be collected     When using the A   or Peaks options  you can specify the  maximum Peak and Non peak volumes to be collected in the  tubes  Be sure to enter volumes less than the maximum tube  capacity  Different peak and non peak tube volumes can be  used to conserve t
131. ne Year Limited Warranty     Factory Service for Teledyne Isco CombiFlash   Systems    This warranty exclusively covers Teledyne Isco  CombiFlash  instruments  providing a one year limited  warranty covering parts and labor     Any instrument that fails during the warranty period due to  faulty parts or workmanship will be repaired at the factory at  no e to the customer  Teledyne Isco   s exclusive  liability is limited to repair or replacement of defective  instruments  Teledy ne Isco is not liable for consequential  damages     CombiFlash systems are designed to use Teledyne Isco  RediSep columns  RediSep columns lead the industry in  quality and performance  Customers may use other brands  of columns on their CombiFlash instruments  These non   RediSep columns may fail to separate compounds  may leak  silica into the flow path of the CombiFlash instrument  and  may leak solvents from their fittings and seals     Damage to CombiFlash instruments resulting from the  failure of a non RediSep column may cause the need for  repairs that incur billable service activities  The customer is  responsible for these charges     Teledyne Isco will pay surface transportation charges both  ways within the 48 contiguous United States if the  instrument proves to be defective within 30 days of  shipment  Throughout the remainder of the warranty period     the customer will pay to return the instrument to Teledyne  Isco  and Teledyne Isco will pay surface transportation to  return the repa
132. ne viewing the run        6 48    Section 6 User Help Reference    6 4 9 1 General instructions for reading the data   Most elements on the window are color keyed to help you  locate the tubes containing the peaks of interest  Below each  collected peak there is a color bar that matches a tube in the  map on the left  The tube map provides a visual  representation of the tubes that contain the peaks of interest   If you wish to identify the tube by number  refer to the table  below the rack diagram     If the run used multiple racks  the Previous and Next Rack  buttons are active  Click these buttons to scroll through the  available racks  The number of the currently displayed rack is  shown at the top left of the window     6 4 9 2 Reading the Chromatogram  The plot area displays the following       The red absorbance trace produced by the system s peak  detector  Absorbance units  also shown in red  that  correspond to this trace are shown on the left Y axis       Purple absorbance and green ratio traces may be visible  if you are monitoring a second wavelength  See the  MONITORING THE WAVELENGTH RATIO  6 4 8  help topic       The blue gradient curve that was used during the run   The   Solvent B scale is shown in blue on the right  Y axis       The X axis depicts the run time  shown as minutes or  COLUMN VOLUMES  ON LINE        Vertical lines appear at intervals along the X axis  These  lines indicate collection tube changes  To prevent the  plot area from being obscured b
133. nect the threaded fitting on  the waste line assembly to the Column Waste port  labeled    1        b  At the Rf 4x back panel  connect the natural tubing of  the waste line assembly to the Waste Level Sense port  labeled    1        c  At the CombiFlash Rf system back panel  the black  tubing should still be connected to the Inject Valve  Waste port  from step 1   If not  reconnect the tubing     d  At the waste container cap for channel 1  insert the  black tubing into the Diverter Valve Waste opening     e  At the waste container cap for channel 1  insert the  natural tubing into the Waste Level Sense opening   Figure 4 5   Push the stop  ferrule  on the Waste Level  Sense tubing against the cap  This ensures that the  Waste Level Sense tubing is far enough into the  container to detect when it is nearly full     f  At the waste container cap for channel 1  insert the  black tubing from the Inject Valve Waste port  The  completed waste connections for Channel 1 are shown in  Figure 4 6     6  Using one of the three remaining waste line assemblies  and waste caps  complete the waste connections for channel  2 as shown in Figure 4 7     7  Using the three remaining waste line assemblies and waste  bottle caps  complete the waste connections for channels 2  through 4 as shown in Figure 4 7        4 8    Section 4 Preparation  Optional 4x Module               All other connections  not shown     To  Waste  Bottle 2    Figure 4 7 Channels 2 through 4 waste tubing connections   
134. ng the  FILE gt OPEN  6 2 1 2  command  If you click this option  the  system reads the RFID tag on the RediSep Rf column   loads the previous method and adds a run extension  You  can then modify the method and click Play to continue  the run  skipping the equilibration and injection  This  feature can be useful if the run was interrupted or if  additional compounds of interest might be left on the  column        6 24    Section 6 User Help Reference      Load Previous Run from Detected Rack      Rf200   Rf200i  and Rf4X  This option appears when using the  FILE   OPEN  6 2 1 2  command  If you click this option  the  system reads the RFID tag on a single rack and displays  the RUN FILE  6 1 3   This feature is useful when there is  a need to recall run results at a later time       Cancel     Click the Cancel button to close the window  without saving or opening the file     6 3 7 Set Data Path  The FILE gt SET DATA PATH     6 2 1 4  menu command opens the  Set Data Path window  Use this window to select a default  folder for the current user  After selecting a folder  file  operations such as saving or opening files will use this selected  directory     The Set Data Path primarily is used with the USER  MANAGEMENT  6 3 12  feature  User Management will  automatically create a folder for each user  When using the  system  set the data path to your folder or a subfolder within   Each user has a different data path     The window contains the following controls       Current 
135. nto AC power  Contact Teledyne Isco  or its authorized representative for advice     Compare the contents of the box with the enclosed packing  slips  If there are any shortages  contact Teledyne Isco  immediately    The fraction collector arm was stowed to prevent damage  during shipping  The arm stowing bracket should be removed  at this time     1  Loosen the bracket   s two thumbscrews along the sides of  the arm  Figure 3 1                  or        Do       Thumbscrew    T    Thumbscrew    Figure 3 1 Remove arm stowing bracket and spacer    2  Push the arm to the left or right remove the spacer     3  Store the bracket and spacer  These parts should be  reinstalled if the system needs to be shipped again        3 2    Section 3 Preparation  Rf 150 Systems    3 2 Instrument Location    The CombiFlash Rf 150 has a relatively small footprint   requiring about 1550 square centimeters  240 in   of level  bench space  Ensure that the Combiflash Rf has at least 3 cm   1 25   of air space behind it for ventilation  Additional space  may be required for solvent and waste containers     Refer to Table 1 1 of this guide for environmental conditions  and power requirements     A  WARNING    The system is heavy  Use a two person lift to prevent injury     Before making any connections to the Combiflash Rf 150   place the system on the bench or in the fume hood where it  will be operated  Temporarily position the system so you can  access the back panels to complete the connections 
136. nual       2 24    CombiFlash   Rf  User Manual       Section 3 Preparation   Rf 150 Systems    This section provides instructions for unpacking and installing  the CombiFlash Rf 150 system  To prepare the system for  operation  sequentially follow all instructions in sections 3 1  through 3 11     M Note   Section 3 13 contains an Installation Qualification checklist  If  required  sign off the checklist entries as you successfully  complete the following sections     M Note   The Rf150 is rated for operation with reagent grade solvents   Operation at pressures  gt  5 bar   75 psi  with high purity hexanes    gt 99  pure  at 100  concentration may cause the pump to stall  and produce a grating sound  Many times  normal operation can  be restored by removing the high pressure load and flushing the  pump with isopropyl alcohol     3 1 Unpacking the Unit    The Combiflash Rf 150 system is shipped in a single carton   Carefully unpack the shipment and inspect the contents     Z  WARNING    The system is heavy  Use a two person lift to prevent injury      NCAUTION    Do not lift the system by the fraction collector arm  Use the lifting  handles located on the instrument side panels        CombiFlash   Rf User Manual    If there is any damage to the shipping carton or any  components  contact the shipping agent and Teledyne Isco  or  its authorized representative  immediately     A  WARNING    If there is any evidence that the system has been damaged in  shipping  do not plug it i
137. ollowing steps    a  Place the injection syringe on the injection port    b  Slowly force the liquid into the injection port    c  Chase the first injection with a second injection  solution of at least 0 25 mL  The chase solution can be  the starting mobile phase or diluted sample mixture    d  Click OK to continue the run    Rf 75 and Rf 150 systems     If you selected Equilibration  ensure that the solvent  delivery tubing is connected to the injection port above  the column  When you click the OK button  the system  equilibrates the column and then waits  When  instructed  perform a liquid injection as described in  steps a through d above  or attach the solid sample  cartridge to the injection port and connect the solvent  delivery tubing  Click the OK button to continue   If you selected No Equilibration  ensure that the  solvent delivery tubing is connected  The system starts  the run immediately and assumes that the sample has  been injected on the column or the solid load cartridge is  in place   You may allow system to proceed with the run while  monitoring the progress on the PeakTrak Main window  If  desired  gradient parameters may be modified during the run    See REAL TIME GRADIENT EDITING  6 4 6          6 36    Section 6 User Help Reference    As the run progresses  the absorbance trace is drawn on the  Gradient Plot area  The purification continues until the end of  the run defined by the run length setting     See Also    REDISEP RF COLUMN SELECTION GUIDE  
138. olumn     If the column is reusable  the system will also write  information to the RFID tag that can help you manage its use   The    Number of times used        First used on     and    Last used  on    information can help you determine when the column  should be replaced  The    Last fluid used    will help you  determine if any solvent remaining in the column will be  miscible with the solvent currently used in the system     To view the Column Data window  open the METHOD EDITOR   6 3 2  window  then click the Column Data button     6 3 5 Batch    The Batch tab contains the controls to stage and run  sequential separations on systems with a 4x Module     6 3 5 1 Queue  The Queue section of this tab lets you enable column positions  and specify operating parameters       Queue     Check the box by the column positions that  are to be run  Note that the Rf 4x system performs the  separations sequentially  When starting a batch run  the  system starts with position 1  then continues in numeric  order through all selected column positions       Sample Name     The sample name is a text entry box  in which you can label the run  Type a unique descriptor        6 22    Section 6 User Help Reference    PeakTrak saves the run information under this name   Therefore  PeakTrak cannot accept any characters that  are reserved for the operating system        V  etc      If you do not type a Sample Name  PeakTrak will  generate a date time stamp for the name when you begin  the run  
139. olumn mount for Rf 150  systems  and insert the column into the top column mount   Note that the column fittings are keyed to ensure the  correct flow direction     3  Slowly lower the injection valve while aligning the bottom  column fitting  The spring loaded injection valve will hold  the column in place        CombiFlash   Rf User Manual    4  To seal the column fittings  give the column a slight twist   14 turn      After loading a RediSep Rf column  the system  with the exception  of the Rf 150  will use RFID technology to automatically detect the  media type and column size  PeakTrak displays the detected  column size on the Main and Method Editor windows  If the  system does not detect the column  manually select the column  media and size     5 3 Start a Default Method    After completing the system installation steps  Section 2  through 4  as necessary   preparing the sample  and inserting  the column  you are ready to perform a run with the default  method  The system s default methods are factory set with run  parameters typically used by chemists  Default methods are  optimized for the use of RediSep Rf columns  The default  settings will       collect all fluid in the fraction collector rack      differentiate between UV  254 nm  vs  non UV absorbing  fluid and place the fluids into separate tubes      cut UV absorbing peaks based on slope detection or level  threshold     To start the run from the PeakTrak Main window     1  Enter a sample name  If you choose n
140. olvent drop down    list boxes  skip steps 1 through 3        2 15    CombiFlash   Rf User Manual    1  Click the    Add Solvent    button   2  Enter the solvent name and click the OK button   3  Repeat steps 1 and 2 for all solvents     4  For the Solvent A1 drop down list box  select the solvent  name connected to the A1 Solvent Inlet Port  Repeat for  Solvents A2  B1  and B2     5  Select the  Enable solvent level sensing  check box to  enable this feature  recommended   Clear this box to  disable this feature     The solvent level sensing feature will monitor solvent  usage to minimize the risk of running out of solvent during  a purification run  The system compares the volume  required for a purification run with the estimated volume  in the solvent container and alerts you when there is not  enough     6  If using the solvent level sensing feature  enter the  Minimum Solvent Level  density dependent  as a range in  centimeters     To understand this range  consider that dichloromethane is  more dense than hexane  If the selected range is 2 4  the  system triggers an alert when the dichloromethane level  falls to about 2 cm above the weighted filter  At the same  selected range  the system would trigger an alert when the  hexane falls to about 4 cm above the weighted filter     2 12 2 General Settings  CombiFlash Rf Name  optional      Use this option to name  your system  The name will appear in operational displays  and run summaries  This feature is useful when your 
141. on   The method file will be stored by the CombiFlash Rf system  and will be available for future runs     6 4 3 Alternative Ways to Create Method Files    Other than editing a method file on the CombiFlash Rf  system  there are more ways to create method files     6 4 3 1 Importing Method Files from another System   If a method file is located on another CombiFlash Rf system   you can use the FILE gt EXPORT METHOD  6 2 1 10  menu  command to save the method file on an external storage  device  From the touch screen panel  this storage device is a  USB Flash Drive connected to the USB port below the display   If connected to the system via an Internet browser  the storage  device may be any device that can be accessed by the local  computer     To move the method file onto the system  insert the USB flash  drive into the USB port or connect to the system from the  remote computer  then use the FILE gt IMPORT METHOD  6 2 1 9   menu command to locate and import the method file     6 4 3 2 Extracting a Method File from a Previous Run   You can perform real time editing during a run to fine tune  the method  After the run  you can extract the run parameters  as a method file so it can be used on future runs  To do so  open  the Run file and click the Extract button  The system will load  a new method with identical run parameters  You can then       6 42    Section 6 User Help Reference    save the method using the FILE gt Save or Save As menu  commands     6 4 3 3 Scaling Method
142. on of Council Directive      2004 108 EC  The EMC Directive  200286 E C  The WE E EDirective  73 23 EEC     The Low Voltage Directive  Manufacturers Name  Teledyne Isco  Inc   Manufacturer s Address  4700 Superior  Lincoln  Nebraska 68504 USA  Mailing Address  P O  Box 82531  Lincoln  NE 68501  Equipment Type Environment  Laboratory Equipment for Light Industrial Commercial Environments  Trade Name Model No  CombiFlash Rf  Year of Issue  2007  Standards to which Conformity is Declared  EN 61010 1 2    edition Safety Requirements for Electrical Equipment for   Measurement  Control  and Laboratory Use   EN 61326 1 2006 EMC Requirements for Electrical Equipment for  Measurement  Control  and Laboratory Use                                  Standard Description Severity Applied Performance Criteria   EN61000 4 2 Electrostatic Discharge Level 2   4kV contact discharge A  Level 3   8kV air discharge   EN61000 4 3 Radiated RF Immunity 80 MHz to 1000MHz 80  AM at 1kHz A  Level 2   3V m   EN61000 4 4 Electrical Fast Transient Level 2   1kV on AC lines A   EN61000 4 5 Surge on AC Lines Level 2   1kV common mode  A  Level 2   0 5KV differential mode   EN61000 4 6 Conducted RF on AC lines 150 kHz to 80 MHz  A  Level 1   1V rms  80  modulated   EN6100 4 11 Voltage Dips Interruptions 100  drop  10ms A   CISPR11  RF Emissions Group 1  Class A Industrial  Scientific  and   EN 55011 Medical Equipment                EN61000 3 2  3 3   Harmonic and Flicker                         We  the undersi
143. or Drain    2  Test the top shelf drain by connecting a vacuum or air  supply source to the outlet end of the drain tube  Then   verify the presence of such vacuum or air supply source on  the drain  Figure 3 5      3  Route the end of both drain tubes to a suitable waste fluid  collection container     M Note   It may be necessary to extend the drain tube  If so  splice the  tubing with user supplied tubing  The user supplied tubing should  have an inside diameter no smaller than the existing drain tubing   and must be compatible with the solvents used by the system   Route this extension tubing to the waste collection vessel        3 8    Section 3 Preparation  Rf 150 Systems    Vacuum or  pressurized air   A applied to the outlet  end of the drain  tube must exist at  the top shelf drain  hole     Figure 3 5 Top Shelf Drain    3 7 Position the System    After completing the various connections  the system can be  moved to its operating position  Turn the system so that the  operator can access all of the front view features and controls   Figure 1 5   Use care not to damage the connections  tubing   and cables while moving the system        CAUTION    Ensure that the CombiFlash Rf has at least 3 cm  1 25   of air  space behind it for ventilation     Position the solvent and waste containers as necessary     When using higher density solvents such as dichloromethane   DCM   place the solvent container level with or above the  CombiFlash Rf system  Placing solvent containers
144. ork connection  using a standard CAT5 cable  or for a direct connection to a  computer using a cross over cable     16  RS 232 Port     Serial communication port to interface with  expansion modules     17 USB Port   For control of peripheral equipment such as  the Rf 4x module  This rear panel port is also used in  service related functions     18  Cooling Fan   Cools the internal electronic assemblies        CombiFlash   Rf User Manual    Z WARNING    Do not touch the fan     19  Check Valves     Externally mounted pump check valves to  simplify pump maintenance  The check valves are not  present on the Rf 150 systems     20 Solvent Level Sensing Outlet Ports     uses air or  user supplied gas to measure the hydrostatic pressure of  the solvent above the weighted tubing outlet  The ports are  not present on the Rf 150 systems     21 Solvent Inlet Ports     Accepts A and B solvents  The Rf  150 system does not use the 4 port block  Refer to Section  3 4 for more information on Rf 150 solvent inlet ports     22 Optional External Gas Port   An inlet for an internal  pump which supplies pressurized air or user supplied gas  for solvent level sensing  waste container full detection   and system column purges  The external gas should be  regulated to 2   5 psig before connecting the source to this  port  The port is not present on the Rf 150 systems     23 Mains Power     Connects the system to AC line voltage   24  Waste Ports     From left to right       Waste Level Sense   The
145. osition the System             0 000 cece KK KK 2 12  2 9 Install Collection Tube Racks                   2 12  2 10 External Detector  Optional                     2 14  2 11 Turn on Power                     KK KK KK KK 2 15  2 12 Configure the System                 00 0 eee 2 15  2 12 1 Solvents ie ss a   ca rw ccc cee 2 15  2 12 2 General Settings               00 000 KK 2 16  2 12 3 Set Date Time               0 0 00 000 ee 2 16  2 12 4 Vapor Limit                  eee eee 2 17  2 12 5 Set Default Tube Volumes                 2 17  2 12 6 User Preferences              0000 eee eee 2 18  2 12 7 Restart the system                 00055 2 20    2 13 Prime the Solvent Lines                    0   2 21       CombiFlash   Rf User Manual     14 System Verification                RR RR RR KK 2 22   15 Installation Qualification Checklist             2 22    NN    3  Preparation   Rf 150 Systems    3 1 Unpacking the Unit                      005  3 1  3 2 Instrument Location              0 0 00 RR KK K8 3 3  3 3 Connect Power                cece eee 3 3  3 4 Connect Solvent Lines                00000005 3 4  3 5 Connect Waste Lines                00000 K3 3 6  3 6 Connect and Route Drain Lines                   3 7  3 7 Position the System                00 RR RR KK 3 9  3 8 Install Collection Tube Racks                  3 10  3 9 Turn on Power                 dae di sl d  ye GAS 3 12  3 10 Configure the System                 000000  3 12  3 10 1 Solvents         0    ccc eee 3
146. ot to  PeakTrak will  enter the date and time as the sample name when you start  the run     2  Review the PeakTrak Main window settings  If you want to  change any settings  refer to the Editing a Method on line  help topic     3  Click the Play button     4  Select the sample Loading Type from the list  This Loading  Type should support the sample preparation you chose in  Section 5 1  Possible sample loading types for the Rf 200   Rf 200i  and Rf 4x systems are       Solid  Pause    Select this option if you intend place  the sample into a solid sample load cartridge  but       5 6    Section 5 Operation    have not yet prepared the sample  You must also  specify the cartridge type and size  When you click  the run button  the system will perform a column  equilibration and then wait while you prepare the  cartridge  After you have placed the cartridge on the  system  click OK to continue with the run       Solid     Select this option if you have prepared the  sample and placed it into a solid sample load  cartridge  You must also specify the cartridge type  and size  The system will proceed automatically until  the end of the programmed run length       Liquid   Select this option if you have prepared a  liquid sample and plan to manually inject it into the  injection port after column equilibration       None  on column    Select this option if you have  preloaded the sample on the column  The system will  Skip column equilibration so that the sample will not  be flushe
147. r duration on the curve     To close the Gradient Table  click the Edit Gradient button   See Also  DEFINING A GRADIENT  6 4 5     6 3 2 5 Peak Collection  Use the Peak Collection buttons to set the collection mode       All     Click this button to collect all fluids in the fraction  collection tubes during a run       Peaks     Click this button to collect only eluted peaks in  the fraction collection tubes during a run       None     Click this button to divert all fluids during a  run to the waste port       Tube Volume     You can also specify the Tube Volume  for collected fluids  This volume can be the default  maximum volume  Max option  as entered in the       6 18    Section 6 User Help Reference    CONFIGURATION  6 3 10  window  or a method specific  volume less than the capacity defined by the  CONFIGURATION  6 3 10  window  Note that the actual  fraction size may be less if a newly detected peak causes  a tube change  or if you click the Next Tube button       Peak     Type or select the desired volume to be  collected in each tube when the system detects a  peak       Non peak     Type or select the desired volume to be  collected in each tube when peaks are not detected   This setting allows you to conserve tubes without  diverting non peak elute to waste  This setting is  ignored when the peak collection mode is set to Peaks  or None     6 3 2 6 Peak Detection   This section of the window contains option buttons to enable  and disable various peak detection op
148. re making any connections to the CombiFlash Rf 4x  module  place it on the bench or in the fume hood where it will  be operated  Temporarily position the system so you can  access the back panels to complete the connections     4 3 Connect Power    Ensure that the On Standby switch below the CombiFlash Rf  system touch screen panel is in the Standby position  Then   use the supplied IEC power cord to connect the CombiFlash Rf  4x module to mains power      N WARNING    Mains power must meet the voltage  frequency  and amperage  requirements listed on the serial number label     N  WARNING    As long as the AC mains power cord is connected  power is  inside the unit  The mains power cord is the disconnect device   Position the CombiFlash Rf system and Rf 4x module so that  the power cord can be unplugged  or use a power strip where  the plug can quickly be removed from the outlet in the event  of an emergency        4 2    Section 4 Preparation  Optional 4x Module    4 4 Connect Solvent Lines      WARNING    Risk of fire ignited by electrostatic discharge  Never substitute  the black tubing on CombiF ash systems  The black tubing   P N 023 0503 06  is conductive  This tubing is required to  dissipate static electricity      NCAUTION    To prevent damage or premature wear to the pump and internal  valves  clean solvent should be used  The solvent should not  contain any dissolved solids     The CombiFlash Rf system has four solvent ports on the left  side panel  To complete the
149. resenting a safe setting  but one with a reduced margin for  error        6 51    CombiFlash   Rf User Manual       6 52    CombiFlash   Rf  User Manual       Section 7 Remote Interfaces    This section provides instructions for controlling the  Combiflash Rf system through a remote interface  The system  can be accessed by several personal computers through a  corporate network  or by a single personal computer  PC   through a direct connection     7 1 Network Configuration    The procedures described in this section will require assistance  from your network administrator  Contact your Information  Technology department before proceeding     The system has a factory assigned IP address   192 168 1 51   In many cases  this address must be reconfigured for use on a  corporate network  To reconfigure the IP address     1  From the touch screen panel  select the  Tools  Configuration menu command     2  Select the Network Configuration tab   3  Select Static IP as the Network Type     The CombiF ash Rf system also provides limited support for  DCHP connections  If DHCP is desired  select DCHP as the  network type and skip steps 4 through 7  The system will display  instructions required to complete a DHCP connection     4  Enter the IP Address  Netmask  and Gateway information  provided by your network administrator        CombiFlash   Rf User Manual    Should your network administrator request the MAC address  a  label on the system   s back panel lists the Address  You also c
150. ridge     Rf 200  Rf 200i  and Rf 4x Only  Occasionally  impurities  precipitate in the solid load cartridge as compounds are  purified  This may cause high back pressures resulting in  reduced flow rates and long run times  Clicking this button  changes the injection valve position  which removes the  cartridge from the solvent path and relieves the back pressure       A CAUTION    Bypassing the solid load cartridge too soon during the purification  run may leave desired compounds on the solid load cartridge   After removing the cartridge from the fluid path  these compounds  cannot be purified     If you need to use this feature  perform the following to ensure  that you do not leave desired compounds on the cartridge     1 Run a TLC plate prior to starting the run  Measure the Rf  of the slowest eluting compound of interest   The Rf should  be between 0 2 and 0 8     the same range used by the  Gradient Optimizer  6 3 3    Note the Percent B solvent  used to run the TLC plate  If a compound displays tailing  peaks as streaks on TLC  the Rf used corresponds to the  bottom of the streak     2  Load the Sample in the Solid Load Cartridge and note the  cartridge size  Refer to list below to determine the cartridge  volume  If the cartridge is partially filled with silica  use  the cartridge volume as if the cartridge were completely  filled with silica     e 5 gram cartridge   10 mL volume  e 25 gram cartridge   45 mL volume  e 65 gram cartridge   110 mL volume    3  Run the s
151. rmation under this name   Therefore  PeakTrak cannot accept any characters that  are reserved by the operating system        V  etc    If you  do not type a Sample Name  PeakTrak will generate a  date time stamp for the name when you click the Play  button to begin the run       Flow Rate     Type or select the desired flow rate for the  run  When using the Isco default column methods  the  system sets the flow rate to the optimum flow rate of the  column        6 15    CombiFlash   Rf User Manual      Equilibration Volume     Type or select the volume of  solvent that is pumped through the column and out the  waste port before the sample is introduced  Data is not  collected while this volume is being pumped  Three or  more column volumes are recommended for most  applications     Initial Waste  amp  Time Windows     Click this button to  open a window from which you can view and modify  these settings  For the Initial Waste  type or select the  volume of eluate that is allowed to pass through the  column to the waste port after the sample is injected   After this volume has been diverted  the system is ready  to collect peak or all fluids in the collection tubes  Use  this feature to clear the internal lines of any elute left  from a previous run and to conserve collection tubes by  diverting all fluids until the first peak is expected     Time Windows can limit the fraction collection to specific  time durations of the run  To use time windows  first  enter a start time an
152. roup 1  Class A Industrial  Scientific  and   EN 55011 Medical Equipment   EN61000 3 2  3 3 Harmonic and Flicker                         The undersigned  hereby declares that the design of the equipment specified above conforms to the above Directive s  and Standards    79 APO A    as of September 5  2008     USA Representative       William Foster    Vice President of Engineering ANA  TELEDYNE ISCO       60 5232 505             A Teledyne Technologies Company                            DECLARATION OF CONFORMITY     Y       Application of Council Directive     Manufacturer s Name   Manufacturer s Address     Equipment Type Environment   Trade Name Model No    Year of Issue    Standards to which Conformity is Declared     2004 108 EC  The EMC Directive  2002 96 EC   The WEEE Directive   2006 95 EC  The Low Voltage Directive   Teledyne Isco   4700 Superior  Lincoln  Nebraska 68504 USA   Mailing Address  P O  Box 82531  Lincoln  NE 68501   Phone   1  402  464 0231   FAX   1  402  465 3799   Laboratory Equipment for Light Industrial Commercial Environments  Rf 150   2013    EN 61326 2006 EMC Requirements for Electrical Equipment for Measurement  Control  and  Laboratory Use   EN61010 1 2010 Safety Requirements for Electrical Equipment for Measurement  Control   and Laboratory Use                                Standard Description Severity Applied Performance Criteria   EN61000 4 2 2008   Electrostatic Discharge Level 2   4kV contact discharge A  Level 3   8kV air discharge   EN6
153. s   This delay accounts for the volume held in the tubing between the  detector and the drop former  sometimes called the  delay  volume     It is important to remember that the chromatogram  displays activity in real time while the fraction collector activity lags  behind  The duration of this delay will vary with the flow rate       Rewind     This button is active when the system has  completed the run  or if the run was stopped by clicking  the Stop button  The Rewind button returns you to the  Main window with the current method     If you have stopped the run before its programmed run  length has elapsed  clicking the Rewind button changes the  current method to the new  shorter run length  This modi   fied method is ready for the next run  or can be saved for  future runs       Fast Forward     Click this button to jump to the next  step of a run     Once you have started a run  the system performs several  steps  The first step is to deliver solvents using the pro   grammed gradient for the entire run length  When this step  is complete  the system will purge the column with air and  clean the injection valve  Clicking the Fast Forward button  will cause the system to skip any remaining time in the  current step and advance to the beginning of the next step   Unlike the Rewind button  clicking the Fast Forward but   ton to complete the run will not modify the current  method s run length setting        CombiFlash   Rf User Manual    5 4 2 Bypassing the Solid Load Cart
154. s are  password protected to restrict their use to only trained  personnel     6 2 4 3 Software Update  Choose this command to open a window used to install patch  files to update your CombiFlash Rf system software     6 2 4 4 Export Log Files   During operation  the system writes operating data to a log  file  Teledyne Isco service personnel can interpret this data to  optimize system performance or to troubleshoot difficulties   Teledyne Isco recommends that you use the menu command  only when advised by a qualified service technician     6 2 4 5 Last Reorder Code  This command should only be used by RediSep Rf Column  Plan customers     6 2 4 6 About PeakTrak  Choose this command to view information about PeakTrak     6 2 4 7 About Isco  Choose this command to view contact information for Teledyne  Isco    6 3 PeakTrak Windows    This section contains descriptions of the windows used for  most PeakTrak tasks  In this section you will find information  on the following     e MAIN WINDOW  6 3 1   e METHOD EDITOR  6 3 2        6 9    CombiFlash   Rf User Manual      GRADIENT OPTIMIZER  6 3 3      COLUMN DATA  6 3 4      BATCH  6 3 5      FILES  6 3 6    e SET DATA PATH  6 3 7      AUTO PRIME VALVE WASH  6 3 8     MANUAL CONTROL  6 3 9      CONFIGURATION  6 3 10    e SET PASSWORD  6 3 11      USER MANAGEMENT  6 3 12     6 3 1 Main Window    The Main window gives you control of the frequently used  functions before and during the run  After a run  or when  viewing previous runs
155. s been pumped      Solvent A and B selections     Select the desired  solvents from the list  The solvent choices are those  defined by the CONFIGURATION  6 3 10  window   Rf 200   Rf 200i  and 4x only       Play button     Click this button to start the Auto  Prime      Fast Forward button     Click this button to skip to the  next Auto Prime step       Cancel button     Click this button to stop the Auto  Prime and close the window     6 3 9 Manual Control  Manual Control can assist with method development   maintenance of the system plumbing  and system  troubleshooting  The Manual Control window can be opened  by selecting TooLS    MANUAL CONTROL  6 2 3 2  from the  PeakTrak menu     The Manual Control Window has several controls     e Channel     Select the channel for manual control   Rf 4x  systems only       Solvent A and B Selection    Rf200  Rf200i  and RfAX   Select the desired solvents from the list  The solvent  choices are those defined by the CONFIGURATION  6 3 10   window       Percentage Solvent B   type or select the mixture  percentage   Rf 75 150 systems fixed at 50          6 26    Section 6 User Help Reference      Flow Rate   Type or select the pump   s flow rate in  mL min       Pump into Tube        Rf200  Rf200i  and Rf4X  As a  default setting  solvent is pumped to the waste port  during manual control  If you would like to collect the  solvent in a tube  select the option button and enter the  tube number  This function cannot be used with the  Was
156. s can often be diagnosed and corrected over the  phone  or by e mail  without returning the instrument to the factory     Instruments needing factory repair should be packed carefully  preferably in the original carton   and shipped to the attention of the service department  Small  non fragile items can be sent by  insured parcel post  PLEASE BE SURE TO ENCLOSE A NOTE EXPLAINING THE DEFECT     Shipping Address  Teledyne Isco   Attention Repair Service  4700 Superior Street    Lincoln NE 68504 USA    Teledyne Isco  PO Box 82531  Lincoln NE 68501 USA    Repair service   800 775 2965  lab instruments    800 228 4373  samplers  amp  flow meters   Sales  amp  General Information  800 228 4373  USA  amp  Canada     Fax   402  465 3001  Email     Mailing address     Phone     iscoservice teledyne com Web site  www isco com    2g TELEDYNE ISCO    Everywhereyoulook     February 28  2012 P N 60 2132 281 Rev D    
157. s in the       6 23    CombiFlash   Rf User Manual    current folder  the system will ignore the Paste  command to prevent the original file from being  overwritten       Delete     Click this button to delete a highlighted  file or a folder and its contents      New Folder     Click this button to create and name  a new folder       Up     Click this button to browse the contents of the  next higher folder level     Search     Enter a keyword and click the Search  button to find matching file names  Click the Clear  Search button to clear the results     File Management options     Click the Copy Files to  Flash Drive or Move Files to Flash Drive to create  archive copies  PDF or Text  of the files on a  connected flash drive  Click the Delete by Age to  specify an age limit  beyond which will be deleted  from the internal hard drive     File folder info scroll box     This box lists the  contents of the current folder  The contents can be sorted  by clicking the column headings     File Type     This option is shown when you can limit  the display to certain types of files  Select the desired file  type from the list     File Name     This text entry box is used to identify the  currently selected  highlighted  file or folder when  browsing and opening files  When using a Save As  command  use this box to name the file     Open Save     This button performs the listed action     Continue Last Run on Detected Column      Rf200   Rf200i  and Rf4X  This option appears when usi
158. s the run length  While in the Paused state  you  can resume the run by clicking the Play button  or stop  the run by clicking the Stop button  If you resume the  run  the system continues the gradient curve from the   B when the system was paused     Stop     This button suspends the entire run  Unlike the  Paused state  the pump  peak detection  and fraction  collection will not operate  While stopped  you can save  the run data by clicking the Save button  abort the run  by clicking the Rewind button  or resume the run by  clicking the Play button     Next Tube     This button advances the fraction  collector to the next tube position  This allows you to  conveniently collect elute of interest in new tube     Rewind     This button is active when the system has  completed the run  or if the run was Stopped by clicking       6 12    Section 6 User Help Reference    the stop button  The Rewind button returns you to the  MAIN WINDOW  6 3 1      If you have stopped the run before its programmed run  length has elapsed  the current method is modified using  the new run length  This modified method is ready for  the next run  or can be saved for future runs       Fast Forward     Click this button to jump to the next  step of a run  Once you ve started a run  the system  performs several steps  The first step is to deliver  solvents using the programmed gradient for the entire  run length  When this step is complete  the system will  purge the column with air  Following this  th
159. se care not to damage the connections  tubing  and cables while  moving the system     1  Turn the CombiFlash Rf system so that the operator can  access all of the front view features and controls  Figure  1 5      2  Position the Rf 4x module along the left side of the  CombiFlash Rf system     3  The Rf 4x module has a tab extending from the right side of  its base plate  Slightly raise the Combiflash Rf system and  slide the Rf 4x module so that the system will rest on the  module   s tab  The weight of the CombiFlash Rf system on  the tab will hold the two together     4 8 Remove Shipping Hardware    To prevent damage to the injection valve assemblies during  shipping  each is secured by a thumbscrew which must be  removed before operating the system  Firmly grasp the  cartridge guide above the injection valve and then remove the  thumbscrew  Figure 4 9   Gently lower the injection valve once  the thumbscrew is removed     Save the thumbscrews should there be a need again to stow the  valves for shipping        CombiFlash   Rf User Manual    Thumbscrew     gt  lt   one of three   E                   Figure 4 9 Remove thumbscrews    4 9 Turn on Power    Turn the CombiF ash Rf system switch to the ON position   The system will begin its startup routine which includes self  diagnostics  During the startup routine  the system will detect  the connected Rf 4x module and automatically configure the  user interface to access the additional channels  The system is  ready for oper
160. sh   Rf User Manual       Figure 1 3 CombiFlash Rf 200i system  with evaporative light  scattering and UV  or UV vis  detection    CombiFlash Rf 200i   This system has the same high perfor   mance features as the Rf 200 system  but includes an inter   nal evaporative light scattering detector  ELSD   During  operation  this detector can be combined with the UV  200     360 nm  or UV vis  200   780 nm  detection to isolate visible  and UV absorbing compounds  as well as compounds with  little or no chromophores        Section 1 Introduction       Figure 1 4 CombiFlash Rf 4x system    CombiFlash Rf 4x   To perform multiple  sequential purifi   cations in a personal or workgroup setting  a 4x module can  be added to the Rf 200 or Rf 200i system  This configuration  is known as the CombiFlash Rf 4x system  a fully auto   mated four channel Flash chromatography system  The  four channels are independent   including separate waste  containers with full bottle detection     All Combiflash Rf systems are optimized for use with  Teledyne Isco s RediSep9 Rf columns  which are pre packed  with a variety of media  For example  the CombiFlash Rf can       1 5    CombiFlash   Rf User Manual    purify samples from 4 milligrams to more than 33 grams using  RediSep Rf columns loaded with 4 to 330 grams of silica gel     Applications include purification of organic compounds for  drug discovery  as well as research in agrochemicals   petrochemicals  natural products  polymers  and catalysts   
161. sh to edit and save the system   s default methods  use  the    TOOLS gt EDIT DEFAULTS  6 2 3 4   command instead     6 2 1 6 Save Method As      Choose this command to open the FILES  6 3 6  window  From  this window you can rename the current method and save it  on the system   s internal hard drive     The Save Method As command also can scale a method to a  different column size  To scale a method  select the scale  option and select the new column size from the drop down list   Then  click the Save button  The system saves a new method  file with the gradient and run length scaled to the selected  column size     6 2 1 7 Save Run As PDF   Choose this command to save the displayed run on a remote  computer   s hard drive or a USB Flash storage device  connected to the system  When selecting this menu command  from the touch screen  the system saves the file on the USB  storage device  The file name will be the same as the Run file   except with a    PDF    file name extension  When selecting this  command from a remotely connected computer via a web  browser  PeakTrak displays a file download window so you can  select a location and enter a file name      You can open and print PDF files with Adobe Reader      6 2 1 8 Save Run As TXT   Choose this command to save the displayed run on a remote  computer   s hard drive or a USB Flash storage device  connected to the system  When selecting this menu command  from the touch screen  the system saves the file on the USB  drive 
162. size label visible   Figure 2 7   insert the racks into the system  Slide the  rack in until you feel it drop into its seated position        Tube Size  Label          Figure 2 7 Inserting the collection tube rack        CAUTION    An incorrectly installed rack will cause the rack to be misaligned  under the fraction collector arm  Misaligned racks might cause  fractions to miss the tube opening or deposit in the wrong tube   Always ensure the tube size label is visible  that is  facing outward   and the rack is pushed in until it is seated        When you turn the power switch to on  the system will  automatically detect the type of rack and configure program  settings accordingly  Later  in section 2 12 5  you can configure  the maximum volume for each tube size     2 10 External Detector  Optional     CombiFlash Rf systems with serial numbers greater than  209C20092 can be configured for use with an analog signal  from a user supplied external detector  The analog signal  should range from    1 2 to  1 2 volts  although custom  interface cables may convert other analog signals to this  range  Contact the factory for interface and fluid path options     The external detector option is not supported on CombiF ash  Hf 200i systems        Section 2 Preparation  Rf 200 and Rf 200i Systems    2 11 Turn on Power    The system s power switch is located just below the touch  screen panel     1  Turn the switch to the ON position  The system will begin  its startup routine which inc
163. t  inlet block were stowed for shipping  Refer to the diagram  on the back of the system and connect the check valve  tubing to the respective ports  Finger tighten the fittings     2  Connect the solvent source to the system s Solvent Inlet  ports  Figure 2 2 shows the recommended connections  when the user supplied solvent container has a 38 or 45  mm opening  GL 38 or GL 45      If the container does not have a GL38 or GL45 opening  do  not use the cap  Use an alternate method to cover the open   ing  to prevent particulate matter from entering the con   tainer  Ensure that the solvent and air tubing are secured  so that the weighted filter remains at the bottom of the sol   vent source    The Solvent Level Sensing air line connects to the back    panel with a compression fitting  This fitting must be air   tight  To connect tubing using the compression fittings     3  Slide the nut and ferrule onto the tubing as shown in  Figure 2 2     4  Fully insert the tubing into the threaded bulkhead fitting  on the back panel of the system        2 5    CombiFlash   Rf User Manual    5  Finger tighten the nut onto the threaded bulkhead fitting   This will seat the ferrule in the fitting     M Note   When using higher density solvents such as dichloromethane   DCM   place the solvent container level with or above the  CombiFlash Rf system  Placing solvent containers below the level  of the system can contribute to decreased flow due to the high  vapor pressure of DCM  This problem b
164. t 1kHz A  Immunity Level 2     3V m  EN 61000 4 4 Electrical Fast Level 2     1kV on AC Lines A  Transient  EN 61000 4 5 Surge on AC Lines Level 2   1kV common mode  A  Level 2     0 5kV differential mode  EN 61000 4 6 Conducted RF on AC   150 kHz to 80 MHz  A  Lines Level 1     1V rms  80  modulated  EN 61000 4 11 Voltage 0  during half cycle and full cycle A  Dips Interruptions 70  at 25 cycles  CISPR11 EN 55011   RF Emissions Group 1  Class A Industrial  Scientific  and N A  2009  A1 2010 Medical Equipment  EN 61000 3 2  3 3 Harmonic  Flicker N A N A                               The undersigned  hereby declares that the design of the equipment specified above conforms to the above    Directive s  and Standards as of April    11  2012        USA Representative p   A O A    William Foster  Director of Engineering       Part Number  605232253    Rev A             ANA  TELEDYNE ISCO    A Teledyne Technologies Company          END 04 005    8 29 08                   DECLARATION OF CONFORMITY       ISMEA                 Application of Council Directive     Manufacturer s Name   Manufacturer s Address     Equipment Type Environment     Trade Name Model No   Year of Issue     Standards to which Conformity is Declared     2004 108 EC The EMC Directive  2006 95 EC  The Low Voltage Directive  2002 96 EC     The WEEE Directive    Teledyne Isco  Inc    4700 Superior   Lincoln  Nebraska 68504 1398 USA  P O  Box 82531  Lincoln  NE 68501 2531  Phone   1  402  464 0231  Facsimile   1  402
165. t the ELSD option is selected  Otherwise the ELSD may  be disabled  The Rf 200i system will automatically check  this option if the previous run used the ELSD     8  Click OK and the run begins  The system responds  according to the sample Loading Type selected in step 4     Rf 200  Rf 200i  and Rf 4x systems          If you selected Solid  Pause   the system  equilibrates the column with the starting  6B and  then waits while you prepare the solid sample load  cartridge  Place the cartridge on the system and click  OK to continue     e If you selected Solid  the system equilibrates the  column and immediately starts the separation  The  prepared solid load cartridge should be in place  before clicking the OK button     e  f you selected None  on Column   the system skips  the equilibration and immediately starts the  separation  The sample must be on the column before  clicking the OK button       If you selected Liquid  the system equilibrates the  column and then waits  When instructed  perform the  following steps    a  Place the injection syringe on the injection port    b  Slowly force the liquid into the injection port    c  Chase the first injection with a second injection  solution of at least 0 25 mL  The chase solution can be  the starting mobile phase or diluted sample mixture    d  Click OK to continue the run        5 8    Section 5 Operation    Rf 150 systems        e Ifyou selected Equilibration  ensure that the  solvent delivery tubing is connected to the inj
166. te Flow Path valve position  This feature can be  useful when trying to recover a compound that has  precipitated or    crashed    somewhere in the fluid path     e Valve Position   Select the valve position for the  desired solvent flow path   Rf 200  Rf 200i  and Rf 4x  systems only        Prime A   Click the    Prime A    button to pump 100  of  solvent A at the selected flow rate       Prime B   Click the    Prime B  button to pump 100  of  solvent B at the selected flow rate       Prime X  B   Click this button to pump a mix of  Solvent A and B  according to the Percentage Solvent B  value   Rf 75 and Rf 150 systems fixed at 50         Stop   Click this button to stop the pump  This button is  only active while the pump is running     The window also includes controls to air purge Rf 200  Rf 200i   and Rf 4x systems       Air Purge Cartridge   Click this button to pump  compressed air or external gas through the solid sample  cartridge       Air Purge Column   Click this button to pump  compressed air or external gas through the column       Stop     Click this button to stop the air purge  This  button is only active while purging     The right half of the window displays the raw lamp energy   Click the Close Manual Control button to close the window   See Also  MANUAL CONTROL OF THE COMBIFLASH RF  6 4 10        6 27    CombiFlash   Rf User Manual    6 3 10 Configuration    The Configuration window allows you to define the operating  characteristics of your CombiF lash 
167. te locations     7 2 Direct Connection    A direct connection supports communication between the  CombiFlash Rf system and a single PC that meets the  recommendations in Table 7 1     This guide provides instructions for Windows 7  7 2 1     Windows Vista  7 2 2   Windows XP  7 2 3   and Windows 2000   section 7 2 4   Refer to the applicable section  then follow the  instructions in section 7 2 5 to complete the direct connection     A direct connection requires a  crossover  CAT5 cable  Standard  CAT5 network cables will not work  You can purchase a crossover  cable locally or ask your Information Technology department to  assemble one  You can also order a crossover cable  part number  480 6545 02  from Teledyne Isco     M Note  These instructions assume that the system is using the factory set  IP address of 192 168 1 51        7 4    Section 7 Remote Interfaces    The following instructions may require a user account with  Administrator privileges on the PC  If you cannot modify the  settings in sections 7 2 1 through 7 2 4  contact your Information  Technology department     7 2 1 Windows 7 Settings  1  From the Windows Start menu  select Control Panel  then    select the Network and Sharing Center option  The PC  displays the Network and Sharing Center     2  Select the Change Adapter Settings option to display the  available network connections     3  Right click on the Local Area Connection icon and select  the Properties option     4  Select Internet Protocol Version
168. the checklist entries as you successfully  complete the following sections     M Note  The optional 4x module is not compatible with the CombiFlash  Rf 75 system     4 1 Unpacking the Unit    The CombiFlash Rf 4x module is shipped in a single carton   Carefully unpack the shipment and inspect the contents  If  there is any damage to the shipping carton or any components   contact the shipping agent and Teledyne Isco  or its  authorized representative  immediately     A  WARNING    If there is any evidence that the system has been damaged in  shipping  do not plug it into AC power  Contact Teledyne Isco  or its authorized representative for advice     Compare the contents of the boxes with the enclosed packing  slips  If there are any shortages  contact Teledyne Isco  immediately        4 1    CombiFlash   Rf User Manual    4 2 Instrument Location    The CombiFlash Rf system has a relatively small footprint   requiring about 1550 square centimeters  240 in   of level  bench space  Ensure that the CombiFlash Rf has at least 3 cm   1 25   of air space behind it for ventilation  Additional space  may be required for solvent and waste containers     The optional 4x expansion module is installed to the left of the  CombiFlash Rf system  This module requires an additional  933 cm   145 in   of level bench space  Ensure that the module  has at least 3 cm  1 25   of air space behind it for ventilation     Refer to Table 1 3 for environmental conditions and power  requirements     Befo
169. the selected range is 1 2  an alert would  be triggered when the dichloromethane falls to about 1  cm above the bottom of the weighted filter  At the same  selected range  an alert would be triggered when the       6 28    Section 6 User Help Reference    hexane solvent level falls to about 2 cm above the  weighted filter     CombiFlash Rf Name Enter a name for the system  This name  will appear in the title bar of the Main window  The name will  also appear in run summaries  Unique names are helpful  when operating more than one system     Time Zone Select your local time zone     Set Date Time Click this button to set the date and time  From  the touch screen  this command opens a window that allows  you to set the instrument date and time  If you select this  command from a remotely connected computer  this menu  command automatically sets the system time to match the  computer s     Note  The system will restart after you set the time     To prevent unauthorized time changes  this feature is  password protected  See also  SET PASSWORD  6 3 11      Vapor Limit Type or select the vapor limit setting  The  CombiFlash Rf system s internal vapor sensor detects vapors  present in the immediate atmosphere  When this value is  exceeded  the system will shut down to avoid a hazardous  condition  For more information  refer to SETTING THE VAPOR  LIMIT  6 4 11   This function is an optional addition to the  Rf150     Set Default Tube Volume Click this button to open a window  from w
170. tion  of the peak  the compound eluting may not be pure     To display the ratio  open the METHOD EDITOR  6 3 2  window  and select the Show Ratio option     6 4 9 Viewing runs  After completing a run  the PeakTrak Main window is used to  display all collected run data  You can also open previous runs  to view the chromatogram and the peak tube locations  To  open a previous run     1  From the main window  select File gt Open  or click the Open  button  The FILES  6 3 6  window is displayed     2 Choose a Run file and click the Open button   3 PeakTrak displays the run in the MAIN WINDOW  6 3 1    The PeakTrak Main window will include       Rack and tube information     The left pane of the  window lists the current rack  a map for that rack  and a  table that lists the peaks and their corresponding tube  numbers  If the window is currently displaying the  collection parameters  click the Rack View button to  view this information       Method parameters     Click the Display Method  button to view a summary of the peak detection and  collection settings for the run       Chromatogram     The right side of the window  displays the chromatogram  identified by the Sample  Name in the title bar of the window        6 47    CombiFlash   Rf User Manual      Option buttons     The option buttons give you quick  access to frequently used commands while viewing a  run  The buttons are     e Print     Click this button to print a run summary   When viewing the run from a remote perso
171. tions   all other connections not shown        4 6    Section 4 Preparation  Optional 4x Module    To Rf 4x Module   To Hf 4x module  Waste Level Sense Column Waste connections    connections 1  2  3  or i 1 2 3 0r4    Channel 1 Only   To Hf system  Inject Valve Waste    60 0923 017 Ferrule              60 0923 015    Compression Fitting See step 8 for details on  Nu    this connection for  channels 2 through 4     60 5234 188  Waste Line Assembly    Ventilation Hole    Luer fitting    If necessary  connect  user supplied tubing and  route to fume hood      See Top View     Tubing Stop   ferrule     60 5235 053 38 mm Waste Cap  or  60 5235 054 45 mm Waste Cap     Top View     Column Waste  Waste Level Sense with tubing stop O eC Ventilation    Inject Valve Waste  Rf 200   or Valve Waste Out  Rf 4x        Figure 4 5 Recommended Rf 4x module channel 1 waste connections       4 7    CombiFlash   Rf User Manual    5  Select one of the four waste line assemblies  PN  60 5234 188  from the accessory package  This will be used  to complete the waste connections for channel 1 as shown  in Figure 4 6     These instructions assume that the column waste from each  channel will be isolated in separate containers  If the waste is to be  collected in a common container  omit the use of the waste caps  and secure all liquid and air lines in the common container  The  ends of all air lines should extend at least 2 inches  b cm  below  the container opening     a  At the Rf 4x back panel  con
172. tions  When an option is  enabled  the window also allows you to modify the setting  details for that peak detection option  All enabled options will  be displayed on the chromatogram  The options and settings  for the Rf 200  Rf 200i  and Rf 4x systems are described below     Note  You can select up to four peak detection options on the  Rf 200  Rf 200i  and Rf 4x systems  If using more than one  option  such as A1 with    2  the system considers a peak to be  present when any option is true  a logical OR          l and 22  wavelength 1 and 2      Enable these  options to use and configure primary and secondary  wavelength detection  When enabled  type or select the  peak detection wavelength in nanometers  then click the  Details button to configure additional settings described  below       Slope Based     Select this option to enable  Slope based peak detection  When enabled  peaks will  be recorded if the slope algorithm indicates a peak  within the Peak Width setting       Peak Width     Select the average peak width  setting  Peak widths are measured at the baseline   The slope detector will detect peak widths ranging  from about 0 2 to 2 times the peak width setting  For       6 19    CombiFlash   Rf User Manual    example  if you entered a peak width of 1 minute  the  range would be 12 seconds to 2 minutes  For best  operation  the peak width should be set to just over  the average peak width being separated  For  instance  if the average peak width is 45 seconds  yo
173. tory practices and  standard safety procedures        1 19    CombiFlash   Rf User Manual      N WARNING    The CombiFlash Rf has redundant safety devices to limit  pressure to less than 210 psi  1448 kPa   RediSep columns  smaller than 100 g are CE certified using standard IEC61010 1  for use on the CombiFlash Rf  RediSep columns larger than  100 g meet Pressure Vessel Directive 97 23 EC  Teledyne Isco  strongly recommends against the use of columns rated less  than 210 psi  1448 kPa      1 6 1 Hazard Severity Levels    This manual applies Hazard Severity Levels to the safety  alerts  These three levels are described in the sample alerts  below       A CAUTION  Cautions identify a potential hazard  which if not avoided  may  result in minor or moderate injury  This category can also warn you  of unsafe practices  or conditions that may cause property  damage       A WARNING    Warnings identify a potentially hazardous condition  which if  not avoided  could result in death or serious injury      N DANGER    DANGER   limited to the most extreme situations to identify  an imminent hazard  which if not avoided  will result in  death or serious injury     1 6 2 Hazard Symbols    The equipment and this manual use symbols used to warn of  hazards  The symbols are explained in Table 1 5        1 20    Section 1 Introduction    Table 1 5 Hazard Symbols    Warnings and Cautions       The exclamation point within the triangle is a warning  sign alerting you of important instructions i
174. ts A and B will be  mixed gradually from the last programmed  B until the  next programmed  B is reached  This method connects  the inflection points that define the gradient with a  straight line       If the Step method is selected  each  B value is held  until the next programmed  B is reached  Because the   B values are held  the changes at each new value are  nearly instantaneous  producing a stepped appearance  on the plot area     e  f Isocratic is selected  the system will pump a single  B  mix throughout the run     Enable Rapid Equilibration Select this option to turn on rapid  equilibration  When enabled  this time saving feature pumps  solvent at the maximum flow rate during column  equilibration  Note that some column sizes and types might be  adversely affected by high flow rates     6 3 11 Set Password    This window  opened by selecting the TOOLS gt SET PASSWORD   6 2 3 6  menu command  is used to enter and change the system  password  and to enable password protection for the default    methods   To enter or change a password   1  Type the password   The factory default password is combiflash   2 Type the password again to verify your entry     3 To protect the Default methods select the  Require  Password For Editing Default Methods  option     4 Click the OK button to save your settings and close the  window     The following menu commands will require a password before  displaying the requested window       TOOLS gt EDIT DEFAULTS  6 2 3 4     TOOLS gt USER
175. u  should enter a peak width of 1 minute  For most flash  chromatography  1 minute is a good starting point for  a peak width       Threshold Detection     Select this option to enable  Threshold peak detection  When enabled  peaks will  be recorded if the Absorbance Units  AU  value is  exceeded       Threshold     Type or select the Absorbance Units  value to be used for Threshold detection     Note  If both Slope and Threshold peak detection  methods are checked  the system considers a peak to  be present when any one condition is met  This logical  OR operator means that the system will cut a peak  when either the Slope condition is true  or when the  Threshold condition is true       Monitor     Enable this option to only use the  detection source as a monitor  When enabled  the  detection source will be displayed as a trace on the  gradient plot area  but will not be used to cut peaks     All Wavelength Detection     Enable this option to  detect peaks within a user selected range of  wavelengths  When enabled  click the Details button to  configure additional settings  These settings include the  slope based and peak width options described in 41 and  A2 above  Additionally  you can type or select the  minimum and maximum wavelengths limits in  nanometers     External Detector     Enable this option to use a 0 to 1  volt analog signal from an external detector  When  enabled  click the Details button to configure additional  settings  These settings include the slope b
176. ubes when using the All peak collection  mode by collecting large volumes of non peak fluid while  creating more  smaller fraction volumes of fluids of interest     The fraction collector will advance to the next tube whenever a  peak is detected  Also note that the system will advance to the  next tube if a new peak is detected before completing the last  one  sometimes called a double advance   Keep in mind that  the slope and threshold detection methods use algorithms  which optimize the peak detection  For example  peak  detection includes a hysteresis to reduce the likelihood of  multiple tube advances that may occur if there is noise at the  beginning or tail end of a peak     Peak Detection This section of the window contains option  buttons to enable and disable various peak detection options   When an option is enabled  the window also allows you to  modify the setting details for that peak detection option  All  enabled options will be displayed on the chromatogram  The  options and settings for the Rf 200  Rf 200i  and Rf 4x systems  are described below     Note  You can select up to four peak detection options   excluding the External option  on the Rf 200  Rf 200i  and Rf  4x systems  If using more than one option  such as 41 with 2        6 39    CombiFlash   Rf User Manual    the system considers a peak to be present when either option  is true  a logical OR          l and 42  wavelength 1 and 2      Enable these  options to use and configure primary and second
177. ue use of the CombiF ash Rf 200i if liquid is present at  the Pump Drain  Contact Teledyne Isco technical service for  assistance with correcting the leak        CombiFlash   Rf User Manual       Figure 1 9 Optional CombiFlash Rf 4x Module Features  Back     32  Waste Level Sense   The system uses air or user supplied  gas to measure the hydrostatic pressure of the waste above  the level sense line end  The system will suspend operation  when this level is too high  Four ports allow you monitor  the level of each waste container on the Rf 4x system     33  Column Waste     Waste fluids are sent out ports one  through four  Tubing connected to these ports are paired  with the waste level sense air lines and routed to a  collection container for each channel     34 From Rf   This port on the Column Waste block receives  fluid from the diverter valve on the fraction collector arm     35  Valve Waste Out   The injection valves on channels 2  through 4 are cleaned after each run  The system directs  fluids from this automatic cleaning routine to this shared  port     36 A B   reserved for future use     37 USB Cable   This cable connects to the USB port on the  Rf 200 or Rf 200i system back panel  see item 17   The Rf       Section 1 Introduction    system controls the operation of the Rf 4x module through  this connection     38  Mains Power     Connects the module to AC line voltage     39  Waste Level Sense From Rf     Receives pressurized air or  optional external gas from the
178. uid spray  The instrument is not watertight and these actions  could damage the internal electronics        8 2    Section 8 Maintenance    8 3 1 Collection Rack and Tray Cleaning      WARNING    Risk of fire or equipment damage  Unclean collection racks and  tray might inhibit their conductive properties  The racks and  tray must be kept clean to dissipate static electricity     The collection tube racks and tray are made of conductive  plastic  Dirt  film  or coatings might prevent their ability to  dissipate static electricity  To avoid problems that possibly  result from an electrostatic discharge  clean the racks and tray  monthly  Use distilled water with a mild detergent  For  tougher stains  use isopropyl alcohol     8 4 Tubing Inspection    A  WARNING    Risk of fire or equipment damage  Faulty tubing  fittings  and  drains may allow organic solvents to pool in unsafe areas   creating a potential for dangerous levels of flammable vapors   Improper draining may damage the instrument   s internal  components     Perform a tubing inspection monthly     1  Visually inspect the solvent  waste  and drain tubing  The  tubing must be free of any damage  kinks  or deterioration   Fittings should show no signs of leaks     2  Test the collection tray drain and top shelf drain by  connecting a vacuum or air supply source to the outlet end  of the drain tubes  Then  verify the presence of such  vacuum or air supply source on the drain hole  Figures 8 1  and 8 2      Correct any 
179. urge  peak detection  and peak collection mode        6 37    CombiFlash   Rf User Manual    Equilibration Volume Column equilibration is optional  but  you may find that separations are more effective if the column  is equilibrated before injecting the sample  The recommended  equilibration volume is at least three column volumes of the  starting mobile phase     RediSep Silica gel column volumes are      RediSep Rf 4 g   4 8 mL     RediSep Rf 12 g   16 8 mL     RediSep Rf 24 g   33 mL     RediSep Rf 40 g   48 mL     RediSep Rf 80 g   125 mL     RediSep Rf 120 g   192 mL     RediSep Rf 220 g   334 mL     RediSep Rf 330 g   443 mL    Other media types will have similar column volumes  Refer to  the REDISEP RF COLUMN SELECTION GUIDE  oN LINE  for actual  volumes     As an example  consider a RediSep Rf 4 g column and a 3  column volumes setting  Before instructing you to inject the  sample  the system will pump 14 4 mL of solvent A through  the column and out the waste port     Initial Waste  amp  Time Windows The initial waste diverts a  user determined line volume to the waste port until the eluant  is about to be collected  At the start of a run  the internal  solvent lines hold some solvent from the previous run  This  volume can be diverted to waste  along with the anticipated  volume of fluid that will pass through the column before a  compound of interest will elute     Time Windows can limit the fraction collection to specific time  durations of the run  To use time wi
180. ve solvent  from a prepared cartridge by attaching the cartridge cap and  manually purging  TOOLS gt MANUAL CONTROL  the cartridge with air     Empty solid sample cartridges also allow a variety of  adsorbents  such as diatomaceous earth  boiling chips  cotton  balls  or paper wipes  Even with alternative adsorbents   always use top and bottom frits to prevent fluid path  problems        5 2    Section 5 Operation    After you have prepared the pre filled or empty cartridge   place the solid sample cartridge on the system   1  Attach the desired adjustable cartridge cap   e 60 5237 047  fits 2 5 and 5 gram solid load cartridges   e 60 5237 048  fits 12 and 20 gram solid load  cartridges   e 60 5237 044  fits 32 and 65 gram solid load  cartridges   2  Press the lever on the side of the cap and fully extend the  plunger  Figure 5 1         Figure 5 1 Press the lever and extend the plunger    3  Slide the plunger into the cartridge until it reaches the top  frit  Figure 5 2     4  Press the lever and push the cartridge into the cap  Align  the cartridge so that it fits fully into the recess in the cap   Figure 5 3     5  Load the solid sample cartridge with cap on the sample  injection port     M Note   Earlier Combi Flash Rf solid load cartridge caps required the use  of sleeves  Contact the factory for information on upgrading these  older systems to use the new cartridges without sleeves        5 8    CombiFlash   Rf User Manual       Figure 5 2 Insert the plunger to the bed 
181. w cell cleaning  on line help topic     To perform a quick cleaning        Section 8 Maintenance    1  After clicking the Cancel Run button  select the  Tools gt Manual Control menu option     2 From the Manual Control window  Figure 8 5   note the  Raw Lamp Energy level at 254 nm  The Raw Lamp Energy  gauge has three ranges  red  yellow  and green     RAW LAMP ENERGY  1000 at 81 ms    The flow cell should be cleaned for optimal  detector performance  A dirty flow cell may cause  the detector signal to appear  noisy      FLOWY CELL CLEANING HELP    Figure 8 5 Raw Lamp Energy Gauge    e Red     lamp energy is obstructed to a degree that the  system might not reliably detect peaks  If you  attempt to operate the system  peak collection will be  forced to collect all  This prevents diverting desired  compounds to waste       Yellow   lamp energy is partially obstructed and the  peak detector performance might be impaired  The  system can be operated although the absorbance  trace might appear noisy  Yellow might be a normal  indication for systems with flow cell windows stained  by compounds       Green   lamp energy is sufficient to detect peaks up  to 2 4 Absorbance Units     3  Remove the column and insert a bypass tube between the  upper and lower column mounts  A bypass tube  P N  209 0165 46  can be found in the accessory kit   P N 60 5239 006      4  Set the Flow Rate to 40 mL min  Figure 8 6         8 7    CombiFlash   Rf User Manual    SOLVENT A    hexane PRIME A    SOL
182. y seated in  the port    3  Cut the tubing to length and connect the other end to the  user supplied carrier gas  An assortment of l s inch  adapters are supplied in the accessory kit to complete the  connection to your gas source     Teledyne Isco recommends  gt 99  pure nitrogen from a source  that can deliver 2 5 SLPM at 60 to 70 psi     2 7 Connect and Route Drain Lines      WARNING    Risk of fire or equipment damage  Failure to connect drain lines  may allow organic solvents to pool in unsafe areas  creating a  potential for dangerous levels of flammable vapors  Improper  draining may damage the instrument s internal components     The CombiFlash Rf system has drain tubes extending from its  back panel  The tubes drain away any liquid spilled on the top  shelf  Rf 200 only  and the tray beneath the fraction collection  racks    1  Test the Fraction Collector drain by connecting a vacuum  or air supply source to the outlet end of the drain tube   Then  verify the presence of such vacuum or air supply  source on the drain  Figure 2 4      2   Rf 200 only  Test the top shelf drain by connecting a  vacuum or air supply source to the outlet end of the drain  tube  Then  verify the presence of such vacuum or air  supply source on the drain  Figure 2 5         CombiFlash   Rf User Manual    Vacuum or  pressurized air  applied to the outlet  end of the drain  tube must exist at  the collection tray  drain hole                             Figure 2 4 Fraction Collector Drain    R
183. y tube change marks   PeakTrak may limit the number of marks     6 4 10 Manual Control of the CombiFlash Rf  Manual Control can assist with method development   maintenance of the system plumbing  and system  troubleshooting  To manually control the CombiFlash Rf  system  first open the MANUAL CONTROL  6 3 9  window by  clicking the Tools   Manual Control menu command        6 49    CombiFlash   Rf User Manual    6 4 10 1 Pumping Solvents   To pump either solvent  select the solvent from the Solvent A   or Solvent B lists  Then  click the Prime A or Prime B buttons  to pump the selected solvent  To pump a mixture or Solvent A  and B  adjust the Percentage Solvent B setting and then click  the Prime __  B button  When finished  click the Stop button     You can control the flow rate by adjusting the Flow Rate  mL min setting     By default  the system pumps the solvent directly to the waste  port  With Rf 200  Rf 200i  and Rf 4X systems  you can change  the fluid path by selecting a different Valve Position option   through column or through cartridge and column   For valve  positions other than Direct To Waste  you can pump the  solvent into a collection tube  To do so  select the Pump Into  Tube   option and then select which tube     next    or specific  tube number      6 4 10 2 Air Purging   On CombiFlash Rf 200  Rf 200i  and 4x systems  you can purge  the solid sample cartridge and column path with air  To air  purge a path  click the Air Purge Solid Sample or Air Purge 
    
Download Pdf Manuals
 
 
    
Related Search
    
Related Contents
Philips AZ 8357/05 User's Manual    format Acrobat    manuel d`utilisation - Ordi-zen  DALI USER'S MANUAL - Circuits and Systems  建設 機械 施工 教本 (初 級 用 )  USER MANUAL / MANUAL DE USUARIO / MANUAL DO USUáRIO  Allegato B - D.U.V.R.I.  Technical Regulations    Copyright © All rights reserved. 
   Failed to retrieve file