Home

GM82009 Tunable Laser Source Sweep Module User Manual

image

Contents

1. to open or close the communication port e Control laser output state set the laser state to ON to enable the laser output Set it to OFF to turn laser output off e Current frequency setting select frequency value on frequency drop down box then click set value button the frequency setting will be sent to module If click the Set to Max or Set to Min button the current frequency will be set to maximum or minimum laser frequency of the module 10300003 R001 www ucinstruments com 7 iel UC INSTRUMENTS CORP GM82009 Tunable Laser Source Sweep Module V2 11 e Current wavelength setting select wavelength value on wavelength drop down box then click set value button the wavelength setting will be sent to module If click Set to Max or Set to Min button the current wavelength will be set to maximum or minimum laser wavelength of the module e Sweeping mode Frequency sweep specify the beginning and end values step and pause in seconds for frequency sweeping and select Freq on mode drop down box then click the set value button to enable this setting When ready click Sweep button to start the sweeping operation Use Abort button to stop the sweeping in progress Wavelength sweep wavelength sweep can be accomplished through steps similar to those for frequency sweep Sweep parameters description Frequency sweep parameter beginning value end value and step value in GHz the s
2. 00003 R001 www ucinstruments com 2 iel UC INSTRUMENTS CORP GM82009 Tunable Laser Source Sweep Module V2 11 Physical Dimensions Active led Optical out Power USB Trigger out 158405 100403 149 51405 133 5403 Specification 10 pm ipy lt 5 pm 0 05 dB 1hour Typ 0 1 dB 24 hours 0 3 dB typ 0 5 dB max 10300003 R001 www ucinstruments com 3 iel UC INSTRUMENTS CORP GM82009 Tunable Laser Source Sweep Module V2 11 a Software Installation 1 First install the USB driver Double click the CP210x_ VCP Win XP S2K3 Vista exe file to install USB driver according to the system prompt Silicon Laboratories CP210x VCP Drivers f InstallShield Wizard Complete The InstallShield Wizard has successfully copied the Silicon Laboratories CP210x WCP Drivers for Windows 2000 4P 2003 Server Vista to your hard drive The driver installer listed below should be executed in order to install drivers or update an existing driver Click Finish to complete the Silicon Laboratories CP210x CP Drivers for Windows 2000 P 2003 Server Vista setup 2 Next install the NET 3 5 by double clicking DotNET_Framework35 exe This will unpack the file Make sure you have internet connection when performing this step Please note this step can be skipped for Windows 7 e Double click the DotNET_Framework35 exe file to unpack it Extracting Files Extracting File WO SAS Wot tf ramenwork dot eth S Pb To D
3. GM382009 Tunable Laser Source Sweep Module User Manual UC INSTRUMENTS CORP iel UC INSTRUMENTS CORP GM82009 Tunable Laser Source Sweep Module V2 11 Description UC INSTRUMENTS GM82009 is a high power fast and continued tuning tunable laser source module The output can be tuned to any wavelength within C band or L band or C L band The GM82009 is also capable of fast sweeping for any specified wavelength frequency range The GM82009 module is compact in size powered by a 3 3v external DC supply The control of the module is made through a USB connection and can be readily integrated into various test and measurement systems The module is suitable for tests and measurements of high dense wavelength division multiplexing devices waveguide grating arrays planar lightwave circuits Erbium doped optical fiber amplifiers and other fiber optic components It can also be applied in fiber grating and other fiber optic sensors The GM82009 is the first of its kind that combines high performances compact size and cost effectiveness Applications e CWDM DWDM Fiber Components Module System testing e OPM Interleaver DPSK WSS PLC AWG components Modules testing e Fiber Grating Sensor Testing e Fiber Optic Test Equipments Inspection and Testing e Laser Sweep Optical Spectrum Analysis e R amp D and lab application e Passive components light path adjustment monitor e Easy to be integrated into customers system 103
4. PT The PC assigns a com port number automatically SHE REW TEM Hah fS BE ERA 03 DVD CD ROM 3ESDE 2 IDE ATA ATAPI EARS MicrochipTools ASS See PERAZA Y D COM 1 LPT A CP2101 USB to UART Bridge Controller COMA TENA O LPTI MIMO COM F MEMO COM a QE 5 HUE te HE 5 En AIRE 5 ER MAMIE 5 7 MAMERTINES O BAST SHIH 5 28 PRENE AES 5 9 ERE 3 Run the demonstration program select the above comport COM in this example baud rate 1s set to 115200 press open button to establish the communication between the PC and the GM82009 10300003 R001 www ucinstruments com 6 iel UC INSTRUMENTS CORP GM82009 Tunable Laser Source Sweep Module V2 11 Operation When the module communicates with pc the interface is shown below The user may use this test program to set the current output frequency wavelength perform sweep operation and turning laser on off y GM82009 Tunable Laser Source Sele UC INSTRUMENTS CORP GUI SendFile ComPort Baudrate 115200 Update Clear FREQUENCY GHZz WAVELENGTH nm 196100 0 1550 000 5 D Min 191700 0 Min 0 ETA Max 196100 0 Max 0 Freq Scan Wave Scan Pause 0 0001 1528 00 to 1565 00 step 0 001 Pulse 0 0001 Set Value Abort oo Pause Power 0 Laser State ON OOFF e Open Close communication port Use the Open and Close button
5. etting minimum is 1 0GHz Wavelength sweep parameter beginning value end value and step value in nm the setting minimum is 0 001nm 1pm Pause there is a delay between two sweeping steps during this delay the trigger output port of GM82009 will output a negative pulse The delay time is in second Pulse pause the time width of the negative pulse from the trigger output port In the pause the trigger output is OV otherwise the trigger output is 3 3V Caution There are many accurate optical components in the module caution for static in the condition of storage removing and operation e If the environmental temperature is less than 15 centigrade then before operating the module powers the module and waits for about 10 minutes for pre heat 10300003 R001 www ucinstruments com 8 iel UC INSTRUMENTS CORP GM82009 Tunable Laser Source Sweep Module V2 11 Caution Invisible Laser Radiation Do not view directly with optical instruments Viewing the laser output with certain optical instruments e g eye loupes magnifiers and microscopes within a distance of 100 mm may pose an eye hazard Laser power up to 100Mw at 1 55 um could be accessible if optical connector is open or fiber is broken Use of controls adjustments and procedures other than those specified herein may result in hazardous laser radiation exposure UC INSTRUMENTS CORP CONTACT INFORMATION UC INSTRUMENTS CORPORATION Add 37498 Glenmoor D
6. irectory d H Odebd cb 355760106 fa 10300003 R001 www ucinstruments com 4 iel UC INSTRUMENTS CORP GM82009 Tunable Laser Source Sweep Module V2 11 e After unpacking be ready to install the program for operating environment First download the program during downloading the pc must be connected with INTERNET When the downloading operation ends begins the installation M Microsoft NEI Framework 3 5 SP1 22t FF Microsoft AUTRE NET Framework e ES A IE E HEM MICROSOFT 44h 70 Fe Fe if E MICROSOFT NET FRAMEWORK 3 5 SP1 FOR Page Down HA ES PRSTENE OBRAR ARPA W SEAS IES BES Microsoft Corporation S AF SEU RASHES E FREAK 61 MB thit FERA 2 SAT 28 4356 kbps 16 H512 kbps is e The program installation for operating environments is finished M Microsoft NET Framework 3 5 SPI Zf Microsoft RFE Net Framework Microsoft NET Framework 3 5 SPI CRUE ip ENPLE TRH RR AA Service Pack MELE RM AXES AU Hindows Update 10300003 R001 www ucinstruments com 5 iel UC INSTRUMENTS CORP GM82009 Tunable Laser Source Sweep Module V2 11 System set up 1 Connect the provided 3 3V power supply to the Power connector 2 Connect the PC to the GM82009 with the supplied USB cable the PC will prompt for installing the hardware automatically After completing the installation open the windows device manager and look for added comport under comport COM and L
7. r Fremont CA 94536 USA Tel 1 510 366 7353 Fax 1 510 353 1809 Email sales ucinstruments com Website www ucistruments com MAINTENANCE SERVICE CENTER GUILIN CHINA Address No 16 Yi Feng south road GuiLin China Tel 86 773 5850657 Fax 86 773 5814532 Website www ucigl com 10300003 R001 www ucinstruments com 9