Home

Aaron Bovie 950 User Manual - Med-E

image

Contents

1. 0 500 1000 1500 2000 2500 3000 Ohm Figure A 8 Peak voltage versus power setting for fulguration mode 6500 6000 5500 5000 4500 4000 3500 3000 2500 2000 1500 1000 500 0 Open Circuit Peak Voltage Volts 0 1 5 10 15 20 25 30 35 Output Power Setting Watts User s Guide e Aaron 950 A 9 Bipolar Curves Figure A 9 Output power versus impedance for bipolar mode BIPOLAR Mode 30W 15W 50 45 500 1000 1500 2000 2500 3000 Ohm Figure A 10 Peak voltage versus power setting for bipolar mode 2000 eak Volta DD gt gt 800 600 amp 400 200 pen Circuit P 0 1 5 10 15 20 25 30 Output Power Setting Watts Bovie Aaron Medical The accessories listed in this section are recommended for use with the Aaron 950 Electrosurgical Generator Ensure all accessories are rated for at least the maximum peak output voltage of the generator User s Guide Aaron 950 B 1 FOOTSWITCHES A1203W Aaron 950 Footswitch Each A1203W ELECTROSURGICAL PENCILS The Standard Aaron 3 button pencil features up down controls and activation button as well as being able to be activated by the footswitch A901 3 button Handpiece Each A901 ESPH ESPH Holster for ES Pencil sterile Box 40 A910 4 A910 Disposable Handpiece Sheath Box 100 gt di
2. The instrument receptacles on this generator are designed to accept only one instrument at a time Do not attempt to connect more than one instrument at a time into a given receptacle Doing so will cause simultaneous activation of the instruments Use the lowest output setting necessary to achieve the desired surgical effect Use the active electrode only for the minimum time necessary in order to lessen the possibility of unintended burn injury Pediatric applications and or procedures performed on small anatomic structures may require reduced power settings The higher the current flow and the longer the current is applied the greater the possibility of unintended thermal damage to tissue especially during use on small structures Use electrosurgery with caution in the presence of internal or external pacemakers Interference produced by the use of electrosurgical devices can cause devices such as a pacemaker to enter an asynchronous mode or can block the pacemaker effect entirely Consult the pacemaker manufacturer or hospital Cardiology Department for further information when use of electrosurgical appliances is planned for patients with cardiac pacemakers If the patient has an Implantable Cardioverter Defibrillator ICD contact the ICD manufacturer for instructions before performing an electrosurgical procedure Electrosurgery may cause multiple activation of ICDs Do not use electrosurgical equipment unless properly tra
3. EE 4 Loop Electrodes 42 34 a Deg dad A AA Ad R AS id B 4 Reusable Active El6ctrodes B 5 Bipolar Forceps Reusable EEN B 5 7 B 6 WANNE Vereinen C 1 User s Guide Aaron 950 LIST OF FIGURES vi Figure 2 1 Layout of controls indicators and receptacles on the front panel 2 2 Figure 2 2 Controls for the cut blend and coag modes A 2 4 Figure 2 3 Controls for the fulguration and bipolar modes and presets 2 5 Figure 2 4 Indicators and receptacles AEN 2 6 Figure 2 5 Layout of controls and indicators on the rear and side panels 2 7 Figure 0 MOUNTING Ki aeee e a Aa 3 2 Figure 4 1 Setup procedures EEN 4 8 Figure A 1 Output power versus impedance for cut mode A 6 Figure A 2 Peak voltage versus power setting for cut mode A 6 Figure A 3 Output power versus impedance for blend mode A 7 Figure A 4 Peak voltage versus power setting for blend mode EEN A 7 Figure A 5 Output power versus impedance for coagulation mode ssssssiieeseseees A 8 Figure A 6 Peak voltage versus power setting for coagulation mode 1 A 8 Figure A 7 Output power versus impedance for fulguration mode A 9 Figure A 8 Peak voltage versus power setting for fulguration mode AA A 9 Figure A 9 Output power versus impedance for bipolar 006 A 10 Figure A 10 Peak voltage versus power setting for bipolar 006
4. ELECTROSURGICAL GENERATOR User s Guide Aaron 950 i Bovie Aaron Medical This manual and the equipment it describes are for use only by qualified medical professionals trained in the particular technique and surgical procedure to be performed It is intended as a guide for using the Aaron 950 Electrosurgical Generator only Additional technical information is available in the Aaron 950 Service Guide Equipment Covered in this Manual Aaron 950 Electrosurgical Generator 110 VAC Model No A950 220 VAC Model No A950 220 For Information Call Aaron Medical A Bovie Company St Petersburg FL 33710 U S Phone 1 800 537 2790 Fax 1 800 323 1640 e International Phone 1 727 384 2323 Fax 1 727 347 9144 www aaronmed com salesQaaronmed com EU Representative Peter J Smith 0413 Medical Products Marketing 18 Yeates Close Thame OX9 3AR UK Made in USA Printed in USA 2006 Aaron Medical A Bovie Company All rights reserved Contents of this publication may not be reproduced without the written permission of Bovie Aaron Medical Aaron Part Number MC 55 055 001 Rev 4 CONVENTIONS USED IN THIS GUIDE Indicates a potentially hazardous situation which if not avoided could result in death or serious injury CAUTION Indicates a hazardous situation which if not avoided may result in minor or moderate injury NOTICE Indicates an operating tip a maintenance suggestion or a hazard that may result in produ
5. The warranty periods for Aaron Medical products are as follows Electrosurgical Generators Two years from date of shipment Mounting Fixtures all models Two years from date of shipment Footswitches all models Ninety days from date of shipment Patient Return Electrodes Shelf life only as stated on packaging Sterile Single Use Accessories Only as stated on packaging Handpiece Only as stated on packaging User s Guide Aaron 950 C 1 This warranty is in lieu of all other warranties express or implied including without limitation the warranties of merchantability and fitness for a particular purpose and of all other obligations or liabilities on the part of Aaron Medical Aaron Medical neither assumes nor authorizes any other person to assume for it any other liability in connection with the sale or use of any of Aaron Medical s products Notwithstanding any other provision herein or in any other document or communication Aaron Medical s liability with respect to this agreement and products sold hereunder shall be limited to the aggregate purchase price for the goods sold by Aaron Medical to the customer Aaron Medical disclaims any liability hereunder or elsewhere in connection with the sale of this product for indirect or consequential damages This warranty and the rights and obligations hereunder shall be construed under and governed by the laws of the State of Florida USA The sole forum for resolving dispu
6. 1 Turn the unit off 2 Verify that the unit is connected to the correct line voltage 3 If the error code reappears record the number and contact Aaron customer service E5 Internal temperature of the unit exceeded limit 1 Turn the unit off 2 Allow the unit to cool for 20 minutes 3 Turn the unit on 4 If the error code reappears record the number and contact Aaron customer service If the unit displays any other error code it requires service 6 2 Bovie Aaron Medical REPAIR POLICY AND PROCEDURES Refer to this section for information on O The Manufacturer s Responsibility O Returning the Generator for Service User s Guide e Aaron 950 RESPONSIBILITY OF THE MANUFACTURER Aaron Medical is responsible for the safety reliability and performance of the generator only under the following circumstances The user has followed the installation and setup procedures in this user s guide Persons authorized by Aaron Medical performed assembly operation readjustments modifications or repairs The electrical installation of the relevant room complies with local codes and regulatory requirements such as IEC and BSI Equipment use is in accordance with the Aaron Medical instructions for use For warranty information refer to Appendix C Warranty RETURNING THE GENERATOR FOR SERVICE Before you return the generator call your Aaron Medical representative for assistance If i
7. Monopolar modes include Cut Blend Coagulation and Fulguration 7 Set the output power either by using the dial on the front of the unit see figure 4 1 letter J or by the up and down buttons on the handpiece see figure 4 1 letter K When power level adjustment is being made by the handpiece an audible tone will sound to indicate that the power level has been changed Depressing and holding the up or down buttons will cause the power settings to change more rapidly for quick adjustment of the output power Power out put is displayed in one watt increments for Cut Blend and Coagulation mode The maximum power for each of these modes is 60 watts Power is displayed in 1 watt increments below ten watts and in whole numbers from ten to 35 watts NOTICE The output settings cannot be adjusted when the unit is being activated 8 The unit is now ready to perform surgery Refer to Activating the Unit later in this section PREPARING FOR BIPOLAR SURGERY 1 Insert the two connectors from the bipolar cable into the bipolar cord receptacles see figure 4 1 letter D 2 Connect the desired forcep to the operating end of the bipolar cord 3 Plug the footswitch cable into the footswitch jack see figure 4 1 letter E A footswitch is required to activate the Bipolar mode NOTICE Dispersive electrodes are not utilized during bipolar procedures 4 Select the Bipolar mode by pressing the membrane switch on the front of the unit see f
8. Publication No 96 128 September 1996 Bovie Aaron Medical NOTICES Do not clean the generator with abrasive cleaning or disinfectant compounds solvents or other materials that could scratch the panels or damage the generator User s Guide Aaron 950 1 6 Bovie Aaron Medical High Frequency DESICCATOR foc AP ar SEG bees CONTROLS INDICATORS AND RECEPTACLES This section describes The Front Rear and Side Panels Controls Indicators and Receptacles User s Guide e Aaron 950 2 1 FRONT PANEL Figure 2 1 Layout of controls indicators and receptacles on the front panel Recall Set Asz D IJ gt Patient Plate Monopolar Alarm Bipolar Footswitch j 4 2 2 Bovie Aaron Medical Symbols on the Front Panel The following table lists descriptions for symbols found on the front panel of the Aaron 950 SYMBOLS DESCRIPTION Generator Controls Cut mode AE Blend mode Coagulation mode High Frequency Desiccator Controls Bipolar mode ZN Fulguration mode Presets IS Select next preset Set new preset Indicators RF ground referenced F Defibrillator proof type BF equipment RF Isolated patient connections are isolated from earth at high frequency Read instructions before use Caution high voltage gt m e Handpiece Connectors Monopolar handpiece
9. 4 1 letter A Select the desired power to be stored by utilizing the power output control knob see figure 4 1 letter J Once all the settings are selected depress and hold the Presets Set Button see figure 4 1 letter for 2 seconds To indicate the settings have been stored the letter on the Presets indicator will blink To recall a preset simply press the Presets Recall button to toggle through all the presets NOTICE A small dot blinking in the lower left corner of the Presets indicator indicates that the unit is not presently set to a user defined preset PREPARING FOR MONOPOLAR SURGERY Cut Blend and Coagulation modes require a patient return electrode Applying the Patient Return Electrode Refer to the manufacturer s instructions for application site and placement procedures When using metal plate patient return electrodes use a conductive gel specifically designed for electrosurgery Select a patient return electrode site with good blood flow While a properly applied electrode results in minimal tissue heating beneath the electrode a good blood flow helps carry heat away from the site 1 Plug the handpiece into the monopolar output on the lower left of the front of the unit see figure 4 1 letter B The plug is designed to fit in only one direction Plug the smaller round connector from the handpiece into the receptacle on the bottom of the unit The three button handpiece is designed to give the doctor complete
10. Medical REUSABLE ACTIVE ELECTRODES A834 Angled Fine Needle Each A811 Reusable Electrode Each A830 Angled Sharp Electrode Each A831 Short Angled Ball Each N A834 A832 Short Straight Ball Each ee A832M Short Straight Ball 3 32 Each A833 Short Straight Needle 3 16 Each e A835 Long Straight Ball Each A836 Long Straight Needle Each sil sil sll SI gl eil s Si SI 2 SI 21 X Reusable Bipolar Forceps are available in micro 1 0 and 2 0mm tip size and in lengths ranging from 4 5 to 8 These forceps are guaranteed for twenty cycles of autoclaving Reusable Bipolar cord is also available with quality silicone construction s A827EU Forceps Cord Each not pictured A826 5 Straight Fine Smooth Each A825 5 1 2 Straight Med Smooth Each A824 5 Curved Fine Smooth Each A823 7 Curved Fine Smooth Each A N A Q A u A822 7 Straight Fine Smooth Each 2 Q E 2 2 2 A821 7 1 2 Bayonet Med Smooth Each A820 A840 A841 A842 A843 A844 A845 All sizes and styles not pictured 7 1 2 Bayonet Fine Smooth Each 4 Tenzel with 5mm Tip Coated Each 3 1 2 McPherson Curved with 5mm Tip Coated Each 3 1 2 McPherson Straight with 5mm Tip Coated Each 4 Tenzel with 5mm Tip Uncoated Each 3 1 2 McPherson Curved with 5mm Tip Uncoated Each 3 1 2 McPherson Straight with 5mm Tip Uncoated Each User s Guide e Aaron 950 B 5 MOUNTING OPTIONS The Aaron 950 comes with a wall mo
11. A 10 Bovie Aaron Medical SEGTIGH 1 INTRODUCING THE AARON 950 ELECTROSURGICAL GENERATOR This section includes the following information Key Features Components and Accessories O Safety CAUTIONS Read all warnings cautions and instructions provided with this generator before using Read the instructions warnings and cautions provided with electrosurgical accessories before using Specific instructions are not included in this manual User s Guide Aaron 950 1 1 KEY FEATURES The Aaron 950 Electrosurgical Generator includes the latest technology This unit offers unsurpassed performance flexibility reliability and convenience It includes the following features Two levels of coagulation Pinpoint Coagulation and Fulguration Pinpoint Coagulation provides precise control of bleeding in localized areas Fulguration provides greater control of bleeding in highly vascular tissue over broad surface areas e Presets The unit incorporates six user defined presets for easy recall of frequently used settings Isolated RF output for Cut Blend and Coag modes This minimizes the potential of alternate site burns Ground Referenced RF output for Fulguration mode Self diagnostics These diagnostics continually monitor the unit to ensure proper performance COMPONENTS AND ACCESSORIES You should receive the following components with your generator e Aaron 950 Electrosurgical Generator A901 Handpie
12. ESSP Scratch Pads sterile Box 40 3 d ESSP mg A1204 DISPERSIVE ELECTRODES Aaron disposable dispersive electrodes with super adhesive gel are designed for single use with safety and quality built in Where reusability is needed a reusable electrode and cable are available A1204Reusable Metal Plate and Cord Each A1204P Reusable Replacement Plate for A1204 Each A1204C Reusable Replacement Cord for A1204 Each A1207 Reusable Metal Pediatric Plate Each A1202C Reusable Connecting Cord for A1202 ESRS Each A1202 Disposable Solid Adult Return Electrode w o Cable Box 5 ESRS Disposable Solid Adult Return Electrode w o Cable Box 50 ESRS B 2 Bovie Aaron Medical DISPOSABLE ACTIVE ELECTRODES AII electrodes utilize the standard 3 32 stainless steel shafts All electrodes feature safety grip insulators combining patient and user safety with easy insertion into and removal from the surgical pencil All disposable electrodes are manufactured to the highest standards They come individually packaged and are sterile Blade Electrodes ESOL Standard Blade Box 50 ES01 m ES37 Modified Blade Box 25 ES37 ES18 ES18 Angled Blade Electrode Box 25 04 ES04 ES39 ES55 ES54 NOI aaa ES39 Ext Modified Blade Box 25 _ ZZA 4 Standard Blade Box 50 not pictured 4 Modified Blade Box 25 not pictured Needle Electrodes ES02 Standard Needle Box 25 ES02 WA zakk ES38 Modified Ne
13. Manufacturer ENNEN 7 2 Returning the Generator for Genee ENNEN 7 2 Step 1 Obtain a Returned Goods Authorization Number 7 2 Step 2 Clean the Generator ENEE 7 2 Step 3 Ship the Generator 7 3 iv Bovie Aaron Medical A 1 Performance Characteristics EEN A 2 Inp t Power A 2 7 A 2 Dimensions and Weight AEN A 2 Operating Parameters nada anta a UO AA AR AA EE A 2 TransSport atid Storages een A 2 Audio VOlUME et deg Deeg eek to A3 Low Frequency 50 60 Hz Leakage Current A 3 High Frequency RF Leakage Current EEN A 3 Standards and IEC Classifications AEN A 4 Class Equipment IEC 60601 1 A 4 Type BF Equipment IEC 60601 1 Defibrillator Proof E A 4 Drip Proof IEC 60601 2 2 A 4 Electromagnetic Interference ENEE A 4 Electromagnetic Compatibility IEC 60601 1 2 and IEC 60601 2 2 A 4 Voltage Transients Emergency Generator Mains Transfer A 4 Output Characteristics nissan eilt A 5 Maximum Output for Bipolar and Monopolar Modes A 5 Output Power GL AA t A 6 000 A 6 Monopolar Goag CuIVES iin i BA A A R A A Ast A 8 Bipolar CUNES acz Eege EE ee A 10 000950168 say Ed O AE B 1 08 lieh 2 2 Dispersive Electrodes waza SOA O AE ROA WRA an B 2 Disposable Active Elechodes ENEE B 3 0 3 Needle Electrodes a a ar winnie hendmieiahiannaahiantaadadiaies B 3 Ball El6CtrodeS rennen einen B 3 REET Z Electrodes 5 5 B 4 Ball
14. Patient return electrode DA Footswitch Bipolar forceps User s Guide e Aaron 950 2 3 CUT BLEND AND COAG CONTROLS Figure 2 2 Controls for the cut blend and coag modes Power Display watts Indicates the power set for Cut Blend Active Indicator the selected mode from the ground Control Knob Coag Indicator Turn clockwise to Indicates when Cut Indicator ann Indicates when Indicates that the power HE pure cut mode has been activated in the selects the pure is selected RE cut mode t 1 Blend Mode Bov E GENERATOR I Selector i When pressed e Cut Coag Fulguration Bipolar selects the M Active Indicator blended cut mode i Indicates that the power has been activated in the NN Blend I selected mode EEV I Blend Indicator S Power Control Indicates when I blended cut I mode is selected i I I RF Isolated Indicator Coagulation i Indicates when the I RF output is isolated Power Output I I coagulation mode i increase power output is counterclockwise to Coag Mode Selector I decrease power output When pressed selects the coagulation mode 2 4 Bovie Aaron Medical FULGURATION BIPOLAR AND PRESET CONTROLS Figure 2 3 Controls for the fulguration and bipolar modes and presets Power Display watts Cut Blend Active Indicator jndicates the power set for Indicates that the power Fulguration Fulguration Indi
15. Returned Goods Authorization Number B Be sure the generator is completely dry before you pack it for shipment Package it in its original shipping container if available C Ship the generator prepaid to the address given to you by the Aaron Medical Service Center User s Guide Aaron 950 7 3 7 4 Bovie Aaron Medical TECHNICAL SPECIFICATIONS All specifications are nominal and subject to change without notice A specification referred to as typical is within 20 of a stated value at room temperature 25 C 77 F and a nominal input power voltage User s Guide Aaron 950 A 1 PERFORMANCE CHARACTERISTICS Input Power 120 VAC 10 220 VAC 10 Mains line frequency 50 60 Hz Mains line frequency 50 60 Hz Power consumption 240 VA Power consumption 240 VA Fuses two 2 0 A Slow Blow Fuses two 1 0 A Slow Blow Duty Cycle Under maximum power settings and rated load conditions Pure Cut 60 watt 500 Q load the generator is suitable for activation times of 10 seconds on 30 seconds off for one hour Dimensions and Weight Width 26 cm 10 25 in Depth 11 4 cm 4 5 in Height 17 8 cm 7 0 in Weight lt 15 4 kg lt 7 lbs Operating Parameters Ambient temperature range 10 to 40 C 50 to 104 F Relative humidity 30 to 75 non condensing Atmospheric pressure 700 hPa to 1060 hPa Warm up time If transported or
16. al equipment and physiological monitoring equipment are used simultaneously on the same patient place any monitoring electrodes as far as possible from the surgical electrodes Monitoring systems incorporating high frequency current limiting devices are recommended Do not use needles as monitoring electrodes during electrosurgical procedures Inadvertent electrosurgical burns may result To avoid the possibility of an electrosurgical burn to either the patient or the physicians do not allow the patient to come in contact with a grounded metal object during activation When activating the unit do not allow direct skin contact between the patient and the physician Examine all accessories and connections to the electrosurgical generator before use Ensure that the accessories function as intended Improper connection may result in arcs sparks accessory malfunction or unintended surgical effects When not using active accessories place them in a holster or in a clean dry nonconductive and highly visible area not in contact with the patient Inadvertent contact with the patient may result in burns Studies have shown that smoke generated during electrosurgical procedures can be potentially harmful to patients and the surgical team These studies recommend adequately ventilating the smoke by using a surgical smoke evacuator or other means 1 U S Department of Health and Human Services National Institute for Occupational Safety and He
17. alth NIOSH Control of Smoke from Laser Electric Surgical Procedures HAZARD CONTROLS Publication No 96 128 September 1996 Bovie Aaron Medical ACTIVATING THE UNIT Monopolar Activation 1 If the unit is not already set up follow the set up procedure to prepare the unit for operation 2 Remove the handpiece from the holder Place the handpiece in the desired position 3 To activate the unit depress the activation button on the handpiece see figure 4 1 letter L or depress the pedal on the footswitch While the unit is activated the appropriate audible tone is sounded and one of the activation LEDs will illuminate see figure 4 1 letter N 4 When the procedure is completed turn the unit off 5 Return the handpiece to the holder on the right side of the unit and remove the electrode The electrode should be disposed of after each procedure If contamination has occurred to the handpiece the handpiece should be sterilized NOTICE When sterilizing the handpiece follow the manufacturer s sterilization instructions that accompany the handpiece Bipolar Activation 1 If the unit is not already set up follow the set up procedure to prepare the unit for operation 2 Place the forceps in the desired position 3 To activate the unit depress the footswitch pedal While the unit is activated an audible tone is sounded and the blue activation LED will illuminate see figure 4 1 letter N 4 When the procedure is compl
18. atient place any monitoring electrodes as far as possible from the surgical electrodes Monitoring systems incorporating high frequency current limiting devices are recommended Do not use needles as monitoring electrodes during electrosurgical procedures Inadvertent electrosurgical burns may result To avoid the possibility of an electrosurgical burn to either the patient or the physicians do not allow the patient to come in contact with a grounded metal object during activation When activating the unit do not allow direct skin contact between the patient and the physician Examine all accessories and connections to the electrosurgical generator before use Ensure that the accessories function as intended Improper connection may result in arcs sparks accessory malfunction or unintended surgical effects When not using active accessories place them in a holster or in a clean dry nonconductive and highly visible area not in contact with the patient Inadvertent contact with the patient may result in burns Studies have shown that smoke generated during electrosurgical procedures can be potentially harmful to patients and the surgical team These studies recommend adequately ventilating the smoke by using a surgical smoke evacuator or other means 1 U S Department of Health and Human Services National Institute for Occupational Safety and Health NIOSH Control of Smoke from Laser Electric Surgical Procedures HAZARD CONTROLS
19. cator saa a i SE Mode Selector Indicates when selected mode When pressed fulguration mode selects the is selected fulguration mode Bipolar Mode Hen FREQUENCY DESICCATOR Selector N When pressed Coag Fulguration Bipolar selects the Active Indicator bipolar mode Indicates that the power has been activated in the Bipolar Indicator Power Control _ SPE mode Indicates when bipolar mode is selected RF Ground Presets Referenced Indicator Indicates when the RF output is ground refer Power Output enced Only applies to Control Knob the fulguration mode Turn clockwise to Presets Recall increase power output Selector Presets Set Button counterclockwise to Presets Display When pressed When depressed for decrease power output Indicates which preset is selects the next 3 makes selected AF barid d are preset setting the current settings displayed in lowercase letters the preset letter No Presets Indicator shown in the indica Dot blinks to indicate tor window when no user defined preset is being used User s Guide e Aaron 950 2 5 INDICATORS AND RECEPTACLES Figure 2 4 Indicators and receptacles L 1 ja IJ x Patient Plate Monopolar Alarm Bipolar Footswitch j 4 5 1 6 IR 5 Monopolar Handpiece Patient Plate Bipolar Cord Receptacle Footswitch Receptacle Receptacle Grounding Receptacle Accepts the Aaron A827EU Accepts the Aaron Accepts the Aaron A901 Accepts the Aaron bipola
20. ce 50 sharp and 50 blunt non sterile dermal tips Ten A910 handpiece drapes Five disposable electrodes 3 blades 1 ball 1 needle Hospital grade power cord One reusable grounding cord Wall mount bracket Five disposable grounding pads e User s Guide SAFETY The safe and effective use of electrosurgery depends to a large degree on factors solely under the control of the operator There is no substitute for a properly trained and vigilant medical staff It is important that they read understand and follow the operating instructions supplied with this electrosurgical equipment Physicians have used electrosurgical equipment safely in numerous procedures Before starting any surgical procedure the surgeon should be familiar with the medical literature complications and hazards of using electrosurgery in that procedure To promote the safe use of the Aaron 950 Electrosurgical Generator this section presents the warnings and cautions that appear throughout this user s guide So that you can operate this equipment with maximum safety it is important that you read understand and follow the instructions in these warnings and cautions It is also important that you read understand and follow the instructions for use in this user s guide 1 2 Bovie Aaron Medical Hazardous Electrical Output This equipment is for use only by trained licensed physicians Danger Fire Explosion Hazard Do not use the Aaron 950 electrosu
21. ct damage User s Guide e Aaron 950 iii TABLE OF CONTENTS Equipment Covered in this Manual EEN iii Forlntormati n 0 ii Conventions Used in this Guide iii Introducing the Aaron 950 Electrosurgical Generator 1 1 Key Ala a on AAAA 1 2 Components and Acrceseortes ENEE 1 2 Safety AA PAZ OR AKA DOO GI 1 2 Controls Indicators and Receptacles 2 sassanen aus nea wena ne aaa zee aaa zwana zee ana 2 1 RER E 2 2 Symbols on the Front Panel wa edd 2 3 Cut Blend and Coag Controls ENEE 2 4 Fulguration Bipolar and Preset Controls E 2 5 Indicators and ReCeptacles ENEE 2 6 Rear and Side Panels aaa dA aa Aaaa eigen nn 2 7 Symbols on the Rear Panel EEN 2 7 Symbols on thie Side 00 2 7 0 3 1 Initial 60 gia AR AO A ARIA MKA RANNA A GAAS Sie te A 3 2 Installing the HO 32 2 o ee lee 3 2 Ving he AARON E 4 1 Inspecting the Generator and Accessories AEN 4 2 Setup Salet ne en essen nieder ee 4 2 Setting AAO Ea 4 3 Using and Understanding The Aaron 950 Memory Features EEN 4 4 Storing and Recalling Preset Gettmgs EEN 4 4 Preparing for Monopolar Surgery ENEE 4 4 Applying the Patient Return Electrode NENNEN 4 4 Preparing for Bipolar Surgery aa da a cnn 4 5 e EE E 4 6 Activating the Units et NOWO AAAA kak dk oda ki k AE 4 7 else EE 4 7 Bipolar eu EE 4 7 Maintaining the IG EI 5 1 E e DE 5 2 5 2 Troubleshooting E 6 1 7 1 Responsibility of the
22. eas In addition place patient return electrodes according to the manufacturer s instructions Potential for alternate site burns increases if the return electrode is compromised 4 2 Bovie Aaron Medical CAUTIONS Do not stack equipment on top of the generator or place the generator on top of electrical equipment These configurations are unstable and or do not allow adequate cooling Provide as much distance as possible between the electrosurgical generator and other electronic equipment such as monitors An activated electrosurgical generator may cause interference with them Nonfunction of the generator may cause interruption of surgery A backup generator should be available for use Do not turn the activation tone down to an inaudible level The activation tone alerts the surgical team when an accessory is active When using a smoke evacuator in conjunction with the electrosurgical generator place the smoke evacuator a distance from the generator and set the generator volume control at a level that ensures that the activation tones can be heard NOTICES If required by local codes connect the generator to the hospital equalization connector with an equipotential cable Connect the power cord to a wall outlet having the correct voltage Otherwise product damage may result SETTING UP 1 If the unit is not already installed refer to Section 3 of this manual for the installation procedure 2 Turn on the generator by p
23. edle Box 25 ES03 ES03 Extended Needle Box 25 ES40 ES40 Extended Modified Needle zn t Box 25 Ball Electrodes ES20 ES20 3 16 Ball Box 25 el V ES21 5 32 Ball Box 25 ak User s Guide Aaron 950 B 3 LLETZ ELECTRODES All Aaron LLETZ Loop and Square electrodes feature tungsten wire for superior shape and integrity throughout the excision procedure All loops are packaged with a unique protective shell to prevent damage during shipping Ball electrodes are available in 3mm amp 5mm sizes for fulguration and desiccation while the needle electrodes are available for pinpoint coagulation Ball Electrodes E ES06 Extended 3mm Ball Box 5 1 507 Extended 5mm Ball Box 5 eet Loop Electrodes ES08 5mm X 5mm Loop Box 5 ESO9 10mm X 10mm Loop Box 5 A ES10 15mm X 8mm Loop Box 5 ES11 15mm X 10mm Loop Box 5 ES12 20mm X 8mm Loop Box 5 ES14 ES13 20mm X 15mm Loop Box 5 ES31 20mm X 20mm Loop Box 5 M ES 20mm X lmn Loop Boys ES43 15mm x 15mm Loop Box 5 All sizes not pictured ES44 15mm x 5mm Loop Box 5 ES45 13mm x 13mm Loop Box 5 ES46 10mm x 5mm Loop Box 5 ES47 25mm x 10mm Loop Box 5 ES51 13mm X 8mm Loop Box 5 ES52 10mm X 8mm Loop Box 5 ES53 15mm X 12mm Loop Box 5 ES41 5mm X 10mm Square Box 5 ES14 5mm X 5 mm Square Box 5 ES15 10mm X 4mm Square Box 5 ES16 10mm X 8mm Square Box 5 For additional products for LLETZ LEEP procedures contact your Aaron Distributor B 4 Bovie Aaron
24. eted turn the unit off 5 Remove the forceps from the bipolar cord and sterilize NOTICE When sterilizing the forceps follow the manufactures sterilization instructions that accompany the forceps User s Guide e Aaron 950 4 7 Figure 4 1 Setup procedures 4 8 Bovie Aaron Medical 1a ler 1 OP la MAINTAINING THE AARON 950 This section covers the following topics O Cleaning Periodic Inspection User s Guide e Aaron 950 5 1 Aaron Medical recommends that you complete periodic inspection and performance testing Perform inspections and performance testing every six months A qualified biomedical technician should conduct this testing to ensure that the unit is operating effectively and safely CLEANING After each use clean the unit Electric Shock Hazard Always turn off and unplug the generator before cleaning NOTICE Do not clean the generator with abrasive cleaning or disinfectant compounds solvents or other materials that could scratch the panels or damage the generator 1 Turn off the generator and unplug the power cord from the wall outlet 2 Thoroughly wipe all surfaces of the generator and power cord with a mild cleaning solution or disinfectant and a damp cloth Follow the procedures approved by your institution or use a validated infection control procedure Do not allow fluids to enter the chassis Do not sterilize the generator PERIODIC INSPECTION Every si
25. etting Watts A 6 Bovie Aaron Medical Figure A 3 Output power versus impedance for blend mode BLEND Mode 60W 30W 100 90 80 70 60 W 50 40 RECT 30 f Dee 24 idid bki 10 0 0 500 1000 1500 2000 2500 3000 Ohm Figure A 4 Peak voltage versus power setting for blend mode Open Circuit Peak Voltage Volts 0 1 5 10 15 20 2 30 35 40 45 50 55 60 Output Power Setting Watts User s Guide e Aaron 950 A 7 Monopolar Coag Curves These measurements were taken using short lt 0 5 meter leads Figure A 5 Output power versus impedance for coagulation mode COAG Mode 60W 30W 100 90 80 70 60 W 50 40 Pe 4 30 en ane aaa eeeee eee An R NZ 20 D 107 0 0 500 1000 1500 2000 2500 Ohm Figure A 6 Peak voltage versus power setting for coagulation mode 3000 4000 3500 3000 2500 2000 1500 1000 Open Circuit Peak Voltage Volts 500 0 Output Power Setting Watts A 8 0 4 5 10 15 20 2 30 35 40 45 50 55 60 Bovie Aaron Medical Figure A 7 Output power versus impedance for fulguration mode FULGURATION Mode 35W 18W 45 40 35 30 W 25 20 a 0a am 15 2 e 7 an 10 5 0
26. fingertip control of the power settings The Aaron A901 handpiece is unique handpieces manufactured by other manufac turers will not function with this unit Do not use the Aaron A901 handpiece on other brand units 2 Slide the desired active electrode into the handpiece until it is firmly seated see figure 4 1 letter M The handpiece will accept most standard 3 32 24 cm electrodes 3 Slide the handpiece from above into the holder on the right side of the unit 4 Plug the male end of the reusable grounding cord into the Patient Plate receptacle located to the right of the monopolar output see figure 4 1 letter C Remove the disposable dispersive electrode from its pouch and attach to the snap connector on the end of the reusable grounding cord NOTICE A return electrode is not required for the fulguration mode The patient plate alarm is not used for this mode Procedures may be performed without the use of a return electrode 5 An optional footswitch may be used with monopolar procedures If the footswitch is utilized plug the footswitch cable into the footswitch jack see figure 4 1 letter E While using a footswitch the output will be delivered via the handpiece The activation button on the handpiece will continue to function while a footswitch is connected to the unit 4 4 Bovie Aaron Medical 6 Choose the Monopolar mode of operation by pressing the desired membrane switch on the front panel see figure 4 1 letter A
27. gure 3 1 CAUTION The unit is not to be utilized in the horizontal position as liquids may easily spill into the unit If mounting on a wall surface a qualified individual should be consulted to avoid damage to the wall surface cut ag end say Wi y S W 1 a io or gongi E M 3 ik pe e ef 356 ight Ze Te 2000 PT 4 all zi l Me We 500 uty 308 ae 2 Plug the female end of the supplied power cord into the base of the unit and the male end into a grounded Figure 3 1 Mounting kit wall receptacle Connect the power cord to a properly polarized and grounded power source with the frequency and voltage characteristics that match those listed on the back of the unit 3 2 Bovie Aaron Medical ac Ja Zu ba dla hem USING THE AARON 950 This section contains the following procedures Inspecting the Generator and Accessories O Setup Safety O Setting Up O Preparing for Monopolar Surgery O Preparing for Bipolar Surgery O Activation Safety O Activating the Unit CAUTIONS Read all warnings cautions and instructions provided with this generator before use Read the instructions warnings and cautions provided with electrosurgical accessories before use Specific instructions are not included in this manual User s Guide Aaron 950 4 1 INSPECTING THE GENERATOR AND ACCESSORIES Before each use of the Aaron 950 Electrosurgical Generator ve
28. igure 4 1 letter A 5 Set the output power either by using the dial on the front of the unit see figure 4 1 letter J or by the up and down buttons on the handpiece see figure 4 1 letter K When power level adjustment is being made by the handpiece an audible tone will sound to indicate that the power level has been changed Depressing and holding the up or down buttons will cause the power settings to change more rapidly for quick adjustment of the output power Power is displayed in 1 watt increments below ten watts and in whole numbers from ten to 30 watts NOTICE The output settings cannot be adjusted when the unit is being activated 6 The unit is now ready to perform surgery Refer to Activating the Unit later in this section User s Guide e Aaron 950 4 5 ACTIVATION SAFETY Do not wrap the accessory cords or patient return electrode cords around metal objects This may induce currents that could lead to shocks fires or injury to the patient or surgical team Danger Fire Explosion Hazard Do not use the Aaron 950 Electrosurgical Generator in the presence of flammable anesthetics Fire Explosion Hazard The following substances will contribute to increased fire and explosion hazards in the operating room e Flammable substances such as alcohol based skin prepping agents and tinctures e Naturally occurring flammable gases that may accumulate in body cavities such as the bowel Oxygen enriched atmosphe
29. ined to use it in the specific procedure being undertaken Use by physicians without such training has resulted in serious unintended patient injury including bowel perforation and unintended irreversible tissue necrosis For surgical procedures where the high frequency current could flow through parts of the body having a relatively small cross sectional area the use of bipolar techniques may be desirable to avoid unwanted coagulation In some circumstances potential exists for alternate site burns at points of skin contact e g between the arm and the side of the body This occurs when electrosurgical current seeks a path to the patient return electrode that includes the skin to skin contact point Current passing through small skin to skin contact points is concentrated and may cause a burn This is true for grounded ground referenced and isolated output generators User s Guide e Aaron 950 1 3 1 4 To reduce the potential for alternate site burns do one or more of the following e Avoid skin to skin contact points such as fingers touching leg when positioning the patient e Place 5 to 8 cm 2 to 3 in of dry gauze between contact points to ensure that contact does not occur Position the patient return electrode to provide a direct current route between the surgical site and the return electrode which avoids skin to skin contact areas In addition place patient return electrodes according to the manufacturer s instruc
30. ining has resulted in serious unintended patient injury including bowel perforation and unintended irreversible tissue necrosis For surgical procedures where the high frequency current could flow through parts of the body having a relatively small cross sectional area the use of bipolar techniques may be desirable to avoid unwanted coagulation If the patient has an Implantable Cardioverter Defibrillator ICD contact the ICD manufacturer for instructions before performing an electrosurgical procedure Electrosurgery may cause multiple activation of ICDs In some circumstances potential exists for alternate site burns at points of skin contact e g between the arm and the side of the body This occurs when electrosurgical current seeks a path to the patient return electrode that includes the skin to skin contact point Current passing through small skin to skin contact points is concentrated and may cause a burn This is true for grounded ground referenced and isolated output generators To reduce the potential for alternate site burns do one or more of the following e Avoid skin to skin contact points such as fingers touching leg when positioning the patient e Place 5 to 8 cm 2 to 3 in of dry gauze between contact points to ensure that contact does not occur Position the patient return electrode to provide a direct current route between the surgical site and the return electrode which avoids skin to skin contact ar
31. lass I Equipment IEC 60601 1 Accessible conductive parts cannot become live in the event of a basic insulation failure because of the way in which they are connected to the protective earth conductor Type BF Equipment IEC 60601 1 Defibrillator Proof z The Aaron 950 Electrosurgical Generator provides a high degree of protection against electric shock F particularly regarding allowable leakage currents It is type BF equipment Patient connections are isolated from earth and resist the effects of defibrillator discharge Drip Proof IEC 60601 2 2 The generator enclosure is constructed so that liquid spillage in normal use does not wet electrical insulation or other components which when wet are likely to affect adversely the safety of the generator Electromagnetic Interference When other equipment is placed on or beneath an activated Aaron Medical electrosurgical generator the Aaron 950 Electrosurgical Generator operates without interference The generator minimizes electromagnetic interference to video equipment used in the operating room Electromagnetic Compatibility IEC 60601 1 2 and IEC 60601 2 2 The Aaron 950 Electrosurgical Generator complies with the appropriate IEC 60601 1 2 and IEC 60601 2 2 specifications regarding electromagnetic compatibility Voltage Transients Emergency Generator Mains Transfer The Aaron 950 Electrosurgical Generator operates in a safe manner when the transfer is made between line AC and a
32. n emergency generator voltage source A 4 Bovie Aaron Medical OUTPUT CHARACTERISTICS Maximum Output for Bipolar and Monopolar Modes Power readouts agree with actual power into rated load to within 20 or 5 watts whichever is greater Mode Output Power Output Frequency Repetition Rate Vp p max Cut 60 W 500 357 kHz 50 kHz N A 1 5 KV Blend 60 W 800 Q 357 kHz 50 kHz 30 kHz 5 kHz 2 0 KV Coagulation 60 W 1000 Q 375 kHz 50 kHz 60 kHz 5 kHz 3 8 KV Fulguration 35 W 1000 Q 575 kHz 50 kHz 30 kHz 5 kHz 6 5 KV Bipolar 30 W 200 Q 520 kHz 50 kHz 19 kHz 5 kHz 2 0 KV User s Guide e Aaron 950 A 5 OUTPUT POWER CURVES The curves that follow depict the changes for each mode at specific power settings Monopolar Cut Curves These measurements were taken using short lt 0 5 meter leads For each output power vs impedance curve the upper curve represents readings taken at full power the lower curve readings taken at half power Figure A 1 Output power versus impedance for cut mode CUT Mode 60W 30W 100 90 80 70 60 W 50 40 30 207 107 AE 0 500 1000 1500 2000 2500 3000 Ohm Figure A 2 Peak voltage versus power setting for cut mode 1400 1200 1000 800 600 400 Open Circuit Peak Voltage Volts 200 0 1 5 10 15 20 25 30 35 40 45 50 55 60 Output Power S
33. nstructed to send the generator to Aaron Medical first obtain a Returned Goods Authorization Number Then clean the Generator and ship it to Aaron Medical for service Step 1 Obtain a Returned Goods Authorization Number Call the Aaron Medical Customer Service Center to obtain a Returned Goods Authorization Number Have the following information ready when you call Hospital clinic name customer number Telephone number Department address city state and zip code Model number Serial number Description of the problem Type of repair to be done Step 2 Clean the Generator Electric Shock Hazard Always turn off and unplug the generator before cleaning NOTICE Do not clean the generator with abrasive cleaning or disinfectant compounds solvents or other materials that could scratch the panels or damage the generator A Turn off the generator and unplug the power cord from the wall outlet Thoroughly wipe all surfaces of the generator and power cord with a mild cleaning solution or disinfectant and a damp cloth Follow the procedures approved by your institution or use a validated infection control procedure Do not allow fluids to enter the chassis You cannot sterilize the generator 7 2 Bovie Aaron Medical E 3 Ship the Generator Attach a tag to the generator that includes the Returned Goods Authorization Number and the information hospital phone number etc listed in Step 1 Obtain a
34. r forceps cord A1203W footswitch cord 3 button handpiece cord A1202C reusable grounding cord Patient Plate Alarm Indicator Indicates an alarm condition regarding the patient return cord or pad 2 6 Bovie Aaron Medical REAR AND SIDE PANELS Figure 2 5 Layout of controls and indicators on the rear and side panels Ce 4 5 S 3 H 8 da a a 38 38 83 gs B gt s s 85 3 3 ay 8 Symbols on the Rear Panel SYMBOLS DESCRIPTION v CCRA Volume control Fuse enclosed AN Read Instructions Before Use Symbols on the Side Panel SYMBOLS DESCRIPTION YY Non ionizing Radiation 0 Danger Explosion Risk If Used With Flammable Anesthetics User s Guide Aaron 950 2 7 2 8 Bovie Aaron Medical GETTING STARTED This section includes the following information Initial Inspection Installing the Unit User s Guide e Aaron 950 3 1 INITIAL INSPECTION When you first unpack your Aaron 950 Electrosurgical Generator inspect it visually Look for any signs of damage Verify that the shipping package contains all items listed on the packing list If the unit or any accessories are damaged notify Aaron Medical s Customer Service immediately Do not use any damaged equipment INSTALLING THE UNIT 1 Mount the Aaron 950 Electrosurgical Generator on the wall or optional stand using the mounting kit See Fi
35. res e Oxidizing agents such as nitrous oxide N20 atmospheres The sparking and heating associated with electrosurgery can provide an ignition source Observe fire precautions at all times When using electrosurgery in the same room with any of these substances or gases prevent their accumulation or pooling under surgical drapes or within the area where electrosurgery is performed Use the lowest output setting necessary to achieve the desired surgical effect Use the active electrode only for the minimum time necessary in order to lessen the possibility of unintended burn injury Pediatric applications and or procedures performed on small anatomic structures may require reduced power settings The higher the current flow and the longer the current is applied the greater the possibility of unintended thermal damage to tissue especially during use on small structures Use electrosurgery with caution in the presence of internal or external pacemakers Interference produced by the use of electrosurgical devices can cause devices such as a pacemaker to enter an asynchronous mode or can block the pacemaker effect entirely Consult the pacemaker manufacturer or hospital Cardiology Department for further information when use of electrosurgical appliances is planned for patients with cardiac pacemakers CAUTIONS The use of high frequency current can interfere with the function of other electromagnetic equipment When high frequency surgic
36. ressing the power switch ON see figure 4 1 letter G Figure 4 1 is located at the end of this Section Verify the following e All visual indicators and displays on the front panel illuminate e Activation tones sound to verify that the speaker is working properly 3 If the self test is successful a tone sounds Verify the following The unit resets to the last activated Preset setting The Preset display will display a letter from A F Preset letters b and d will display in lowercase The power display will show the power level for the last used Preset setting The mode for the last activated Preset setting is selected If the self test is not successful an alarm tone sounds An error code may appear on the power display in most cases the generator is disabled Note the error code and refer to Section 6 Troubleshooting Once the self test is successful connect the accessories and set the generator controls Refer to Preparing for Monopolar Surgery or Preparing for Bipolar Surgery later in this section User s Guide e Aaron 950 4 3 USING AND UNDERSTANDING THE AARON 950 MEMORY FEATURES The Aaron 950 incorporates six user defined Presets for easy recall of frequently used settings Storing and Recalling Preset Settings Select the desired preset location A F by pressing the Recall button see figure 4 1 letter H Select the desired mode to be stored by pressing one of the mode membrane switches see figure
37. rgical generator in the presence of flammable anesthetics Fire Explosion Hazard The following substances will contribute to increased fire and explosion hazards in the operating room e Flammable substances such as alcohol based skin prepping agents and tinctures e Naturally occurring flammable gases which may accumulate in body cavities such as the bowel Oxygen enriched atmospheres e Oxidizing agents such as nitrous oxide N50 atmospheres The sparking and heating associated with electrosurgery can provide an ignition source Observe fire precautions at all times When using electrosurgery in the same room with any of these substances or gases prevent their accumulation or pooling under surgical drapes or within the area where electrosurgery is performed Connect the power cord to a properly polarized and grounded power source with the frequency and voltage characteristics that match those listed on the back of the unit Electric Shock Hazard Connect the generator power cord to a properly grounded receptacle Do not use power plug adapters Electric Shock Hazard Always turn off and unplug the generator before cleaning Fire Hazard Do not use extension cords Patient Safety Use the generator only if the self test has been completed as described Otherwise inaccurate power outputs may result Failure of the high frequency electrosurgical equipment could result in an unintended increase of output power
38. rify that the unit and all accessories are in good working order Inspect for damage to the Electrosurgical Generator and all its connections Verify that the appropriate accessories and adapters are present Inspect all cords and connectors for signs of wear damage and abrasion Verify that no errors occur when you turn on the unit SETUP SAFETY Hazardous Electrical Output This equipment is for use only by trained licensed physicians Electric Shock Hazard Connect the generator power cord to a properly grounded receptacle Do not use power plug adapters Connect the power cord to a properly polarized and grounded power source with the frequency and voltage characteristics that match those listed on the back of the unit Fire Hazard Do not use extension cords Patient Safety Use the generator only if the self test has been completed as described Otherwise inaccurate power outputs may result The instrument receptacles on this generator are designed to accept only one instrument at a time Do not attempt to connect more than one instrument at a time into a given receptacle Doing so will cause simultaneous activation of the instruments Failure of the high frequency electrosurgical equipment could result in an unintended increase of output power Do not use electrosurgical equipment unless properly trained to use it in the specific procedure being undertaken Use by physicians without such tra
39. stored at temperatures outside the operating temperature range allow one hour for the generator to reach room temperature before use Transport and Storage Ambient temperature range 24 to 65 0 29 to 149 F Relative humidity 0 to 75 non condensing Atmospheric pressure 500 hPa to 1060 hPa A 2 Bovie Aaron Medical Audio Volume The audio levels stated below are for activation tones bipolar cut and coag and alarm tones return electrode and system alarms at a distance of one meter Alarm tones meet the requirements for IEC 60601 2 2 Activation Tone Volume adjustable 45 to 65 dB Frequency Cut 1 kHz Blend 1 kHz Coagulation 2 kHz Fulguration 2 kHz Bipolar 2 kHz Duration Continuous while the generator is activated Alarm Tone Volume not adjustable 70 dB 5 dB Frequency 2 kHz seconds 1 kHz seconds Duration 2s Low Frequency 50 60 Hz Leakage Current Enclosure source current ground open lt 500 pA Normal polarity intact ground lt 50 pA Source current patient leads all outputs Normal polarity ground open gt 50 pA Reverse polarity ground open lt 50 pA Sink current at high line all inputs gt 50 pA High Frequency RF Leakage Current Bipolar RF leakage current gt 39 MA ims Monopolar RF leakage current additional tolerance gt 150 MA uns User s Guide e Aaron 950 A 3 STANDARDS AND IEC CLASSIFICATIONS C
40. tes arising under or relating in any way to this warranty is the District Court of the County of Pinellas State of Florida USA Aaron Medical its dealers and representatives reserve the right to make changes in equipment built and or sold by them at any time without incurring any obligation to make the same or similar changes on equipment previously built and or sold by them C 2 Bovie Aaron Medical
41. tions Potential for alternate site burns increases if the return electrode is compromised Do not wrap the accessory cords or patient return electrode cords around metal objects This may induce currents that could lead to shocks fires or injury to the patient or surgical team CAUTIONS At no time should you touch the active electrode or bipolar forceps A burn could result Do not stack equipment on top of the generator or place the generator on top of electrical equipment These configurations are unstable and or do not allow adequate cooling Provide as much distance as possible between the electrosurgical generator and other electronic equipment such as monitors An activated electrosurgical generator may cause interference with them Nonfunction of the generator may cause interruption of surgery A backup generator should be available for use Do not turn the activation tone down to an inaudible level The activation tone alerts the surgical team when an accessory is active When using a smoke evacuator in conjunction with the electrosurgical generator place the smoke evacuator a distance from the generator and set the generator volume control at a level that ensures that the activation tones can be heard The use of high frequency current can interfere with the function of other electromagnetic equipment When high frequency surgical equipment and physiological monitoring equipment are used simultaneously on the same p
42. unt kit A mobile stand is available with optional instrument tray and smoke evacuator shelf A stainless steel table top stand Lee is also available A812 C Mobile Stand with Bottom Tray Top Tray Clamp and 12 Height Extender Each A812 Mobile Stand Each A812 BT Optional Bottom Tray for Smoke Evacuator Each A808 T Optional Top Tray for Instruments Each A812 A812 E Optional 12 Height Extender Each Stand Only A813 Stainless Steel Table Top Stand Each Fr A812 BT A812 C A837 Replacement Wall Mount Kit Each B 6 Bovie Aaron Medical WARRANTY Aaron Medical a Bovie Company warrants each product manufactured by it to be free from defects in material and workmanship under normal use and service for the period s set forth below Aaron Medical s obligation under this warranty is limited to the repair or replacement at its sole option of any product or part thereof which has been returned to it or its Distributor within the applicable time period shown below after delivery of the product to the original purchaser and which examination discloses to Aaron Medical s satisfaction that the product is indeed defective This warranty does not apply to any product or part thereof which has been repaired or altered outside Aaron Medical s factory in a way so as in Aaron Medical s judgment to affect its stability or reliability or which has been subjected to misuse neglect or accident
43. x months visually inspect the Aaron 950 Electrosurgical Generator for signs of wear or damage In particular look for any of the following problems Damage to the power cord Damage to the power cable receptacle Obvious damage to the unit Damage to any receptacle Accumulation of lint or debris in or around the unit Bovie Aaron Medical TROUBLESHOOTING This section includes error code descriptions and actions to take to resolve them User s Guide e Aaron 950 6 1 The Aaron 950 Electrosurgical Generator includes automatic self diagnostics If the diagnostics detect an error the system displays an error code sounds an audible tone and deactivates the unit output power Most error codes result from faults in accessories attached to the unit The following table lists the error codes describes the error and recommends actions to take to resolve the error Error Code Description Recommended Action F1 Handswitch or monopolar footswitch pedal may be stuck 1 Turn off then turn on the generator Do not press buttons or activate accessory devices during the self test 2 If the error code reappears disconnect all accessories Turn off then turn on the generator again 3 If the problem persists replace the handpiece or footswitch and repeat the restart 4 If the error code reappears record the number and call Aaron customer service E2 Line voltage error Line voltage is too high

Download Pdf Manuals

image

Related Search

Related Contents

Secteur professionnel  バルブ専用コントローラ マルティ 取扱説明書(H  NOTE RELATIVE À L`EXPÉRIMENTATION INFORMATIQUE DU  légère pour robotique rapide guide d`installation et Manuel de  The M4 Macro Processor Brian W. Kernighan Dennis M. Ritchie Bell  MANUAL DE INSTALACIÓN  USER MANUAL - Abcomrents  Document 6296218  2011 Dodge Nitro User`s Guide  Avaya CallPilot Quick Reference Guide  

Copyright © All rights reserved.
Failed to retrieve file