Fuji Phosphoimager, FLA7000 (PSC 559) Multi Gauge Analysis

image

Contents

1. e Next Proceed to the next mode Display Mode Following are just the differences from Image Mode a Channel color Color icon change the display color The appearance of the icon changes depending on the type of image b Matching Adjusts the positions of the overlaid images c Crosstalk Correct the crosstalk d Range Scope Enhance the contrast e Contrast Scale Type Magnifying the image of the contrast panel being displayed f Scale Type Adjust the contrast g Next Proceed to the next mode Process Mode Following are just the differences from Image Mode 4 options a Rotate Rotate Flip image and inverts pixel values b Filter Reduces noise and displays sharp images c FFT Analyzes the frequencies of image components d Etc Clips joints or positions images Annotation Mode Following are just the differences from Image Mode a Pointer tool b Arrow tool Allows you to draw arrows c Text tool Add text to image d Information tool You can add the scan settings onto the image Window opens so you can select what you want e Scale bar tool Window opens that lets you setup the details of the scale bar f Next Proceed to the next mode Measure Mode Following are just the differences from Image Mode Four Options a Quant Allows you to set the region of interest so that it fits the size of teha sample as well as specifying the background value so it can be subtracted b Profile Allows you
2. the right You can Stop the reading any time you want The image is saved at the point when you stop the scan However the unscanned part will not be saved You can t restart the scan at this point a On the window the scan results of the different scans up to 4 are displayed b The current settings are also displayed c Youcan change the display parameters of the real time window on the right top You can change the type of tone curve Exponential Linear or Sigmoid d You can adjust the density of the image by dragging the adjuster Data of low intensity on the left will be white and data of high intensity on the right will be black e Auto Range Scope if checked will automatically correct to optimum tone f Magnification You can change the display area by selecting a magnification ratio You can also use the buttons Save as Save the data with a different name Launch Launch the Multi Gauge analysis software Return Scan multiple IP Do not open the door until the Return button is no longer grayed out Analysis Software Multi Gauge v 3 x Science Lab 2006 Following is a very simple overview You can borrow the user s manual for more details The software is setup so that you will step through each mode in sequence l 2 3 4 Multi Gauge is installed on the FLA7000 computer There are limited of copies available See Debby for details Open Multi Gauge icon or setup FLA7000 scan to open ima
3. to create the profile region in the migrating direction c Plate Set the region of the plate For the background a blank well can be designated d Distance Allows you to measure the diameter of a circle the distance between 2 points and angles Calibration Mode Following are just the differences from Image Mode 4 options a Quant Used for measuring band concentration b c d 11 An Profile Used for measuring band concentration molecular wt and isoelectric point Plate Used for measuring titerplate concentration Distance No setup alysis Mode Following are just the differences from Image Mode Quantitative results are displayed in table mode HPT Table format varies depending on the quantifying mode used To change the quantifying modes click the desired tab Table option Set items displayed on the results table Export Setting Sets the way to calculate the background Optional tables Quant Profile Plate All Displays the quantitative results in order Comparisons Arranges the quantitative results so that the corresponding bands are adjacent to each other Distance displays the results in order
4. Fuji Phosphoimager FLA7000 PSC 559 Multi Gauge Analysis Software PSC 559 Every time you use the phosphoimager you MUST sign the logbook Do not leave the Imaging Plate IP on the eraser any longer than necessary Once you have erased your IP take it back to your lab Do not store any images on the hard drive If you need to save your image to a flash drive Do not under any circumstance pull the IP out of the scanner drawer let it come out on it s own see below Contact Debby Walthall 404 413 5363 dwalthall gsu edu Sonja Young 404 413 5379 sstovall gsu edu General Information 2 kinds of holders 1 Fluor Stage For fluorescent samples and radioactive samples exposed to imaging plates that are not magnetic on the back Glass costs 2400 so be careful Keep in the plastic storage case It is stored in NSC 516 and can be checked out from Kyu a geland membrane Set sample gently on the glass with the side to be scanned touching the glass Note the location on the grid b Gel with glass Set on the glass gently Fix the glass with one glass holder in the front position and two glass holders in the rear position c IP that are not Fuji and not magnetic Place on glass white side exposed side down 2 IP Stage For radioactive samples exposed to imaging plates that are magnetic on the back All of the old BAS2500 IP are magnetic The stage has 2 sides magnetic side and a side with handles Place the IP o
5. can so that the file is smaller No matter what you size the scan doesn t take too long i Top at bottom of window Returns to the Image Reader home page j Save Condition top of window Save settings 3 4 5 x k Load Condition top of window Load saved condition settings Load IP on the IP stage Place IP stage into the instrument Start Scan The scanned area is read from the left to the right You can Stop the reading any time you want The image is saved at the point when you stop the scan However the unscanned part will not be saved You can t restart the scan at this point a You can change the display parameters of the real time window on the right top You can change the type of tone curve Exponential Linear or Sigmoid b You can adjust the density of the image by dragging the adjuster Data of low intensity on the left will be white and data of high intensity on the right will be black c Auto Range Scope if checked will automatically correct to optimum tone d Magnification You can change the display area by selecting a magnification ratio You can also use the buttons Save as Save the data with a different name Launch Launch the Multi Gauge analysis software Return Scan multiple IP Do not open the door until the Return button is no longer grayed out Reading Fluorescent Samples 1 Select Fluorescence button 2 Reading Settings Window a OS pe Image Folder Sp
6. ecify where you want to save image Use Browse button to go to appropriate folder File Name Name the image Comment Comments are saved with the image Method Selection From the pull down menu select the Method that corresponds with the sample The appropriate laser and filter combination is displayed PMT You may set the voltage to be applied to the PMT The larger the value the higher the sensitivitly Click the button to increase the of scans up to 4 Click the button to reduce the of scans Pixel size The smaller the size the better the resolution but the files are huge 50 um or 100 um are good for most images Latitude This is the dynamic range L4 is the default If the band intensity is similar then L4 is fine If the band intensity is faint or over a broad range select L5 Sampling Area When you put your IP into the instrument note the location of your sample on the grid A vs 1 Use either Grid size using grid lines or Free size arbitrarily Try to size as close to exact as you can so that the file is smaller No matter what you size the scan doesn t take too long Top at bottom of window Returns to the Image Reader home page Save Condition top of window Save settings nad Nm l Load Condition top of window Load saved condition settings Load Fluorescent sample on the Fluor stage Place Fluor stage into the instrument Start Scan The scanned area is read from the left to
7. ge in Multi Gauge Different modes Image Mode has navigation panel operation panel and image window Measure Mode profile has the same panels as Image mode as well as a profile window Analysis Mode has the same panels as Image mode as well as a results table window Navigation panel buttons a Image mode Allows you to open and trim image Ifyou need to analyze multiple images open in this mode first and then proceed to the next mode b Display Mode Allows you to adjust the contrast of images opened and to make crosstalk correction c Process Mode Allows you to perform image processing such as rotaion filtering or noise reduction d Annotation Mode Allows you to put comments and arrows on image e Measure Mode Allows you to set measurement regions such as lanes bands and titer plates f Calibration Mode Allows you to draw the calibration curve of concentration and or molecular wt g Analysis Mode Allows you to display measuring results and to analyze data 10 h Save Mode Allows you to save images and results data table i Print Image Mode a Images can be opened in 3 ways Buttons for each are on side panel below image display Open image directly OPEN trim the image Trimming and then open trimmed image or combine images then open overlaid image Compose b buttons Magnify or reduce image c Hand button Scroll the image d Density Picker tool Display the density value of a pixel
8. n the magnetic side with the notch of the IP lining up with the notch on the stage the white surface should be visible Imaging Plates IP Information 1 10X more sensitive compared to X ray film 2 Reusable 3 Clean with Kimwipes Do not use H20 4 Store in the dark and flat Reading IP and Fluorescent Images 1 Turn on instrument Switch is on the left side of the instrument near the back and near the power cord Do not place IP inside instrument until the power light comes on The IP may be damaged during the self diagnostic steps 2 Let instrument warm up The power lamp on the upper left panel will be lit when warm up is complete Double Click on Image Reader software See below for software details 4 Place the IP onto the IP stage or Fluor stage 0S 7 8 Open instrument door Place IP stage or Fluor stage so that the arrow lines up with the arrow on the main body When the stage edge is placed on top of the arrow there is a tab that comes up Make sure everything fits snug Close instrument door gently As the instrument is reading the IP the busy light is lit When the read is complete the busy light goes out Open the door and remove the IP Image Reader A Home page 1 IP Mode Reading Imaging plates 2 Fluorescence Mode Reading fluorescent samples 3 Digitizing Mode Select when digitizing This is used for silver stained Coomassie stained gels 4 Method Use to register or delete lasers
9. or filters 5 Filter Module Use to register or change filters 6 Preference Set display menu file format or image data type 7 Filter The loaded filters are displayed 8 Laser Loaded lasers are displayed as well as their status ready to use OK or not ready to use No 9 Status Status of the instrument is displayed whether or not it is ready to scan Reading IP 1 Select IP button 2 Reading Settings Window Make all the following selections before reading IP a Image Folder Specify where you want to save image Use Browse button to go to appropriate folder b File Name Name the image c Comment Comments are saved with the image d Laser Filter Make sure LD650 laser and IP filter is displayed These are automatically selected when you selected the IP button e Sensitivity 3 types of sensitivity voltage applied to laser 10 000 highest sensitivity S4000 default and S1000 lowest f Pixel size The smaller the size the better the resolution but the files are huge 50 um or 100 um are good for most images g Latitude This is the dynamic range L4 is the default If the band intensity is similar then L4 is fine If the band intensity is faint or over a broad range select LS h Sampling Area When you put your IP into the instrument note the location of your sample on the grid A vs 1 Use either Grid size using grid lines or Free size arbitrarily Try to size as close to exact as you