Contents

1. IEC60601 1 Standard DigiPump SR8x Syringe Pump Operator s Manual 1 6 Product Description The DigiPump SR8x syringe pump contains the following components 1 The microcomputer system the brain of the whole system that control and manage the whole system and process the detected signals Two single chip Microcontrolers are used for mutual backup copy and supervision When one Microcontroller goes wrong the other one will give an immediate warning signal and cut the power of the host computer which will then stop completely thus ensuring patient s safety 2 The Pump Device the heart of the whole system and the main driving force of the infusion Driven by stepper motor the lead screw moves the infusion piston forward 3 The Sensor System various kinds of sensors such as displacement sensor detecting liquid flow rate and flow volume pressure sensor detecting occlusion etc They supply corresponding signals which will after being amplified be sent over to the computer for signal processing 4 The alarm device after the signal given by the sensor is processed by the microcomputer an alarm control signal triggers the corresponding audio and visual alarm bringing operator s attention for right action There are mainly two kinds of alarms Visual alarm LED Alarm Message in the screen and sound alarm buzzer 5 The Keypad and Display device It is used to set infusion parameters such as infusion
2. Syringe Pump Operator s Manual 1 10 Pump Installation 1 The syringe pump is used in horizontal position 2 Make sure the screw center with thread hole in the enclosure then rotate the screw to fasten the clip on the syringe pump 3 Check the stability of the IV stand 4 Rotate the second clip screw to fasten the syringe pump on the IV stand Setting of Menu Parameters 2 Infusion Settings The DigiPump SR8x syringe pump has three Infusion Modes Rate Mode 1 Time Mode 2 and Body Weight Mode 3 The Rate Mode is the default mode and is automatically selected when the pump is turned on All settings are done using the rotary knob 1 Rotate the knob to move the cursor 2 Press the knob to select 3 Rotate the knob to change the value of the selected setting 4 Press the knob to save the setting DigiPump SR8x Syringe Pump Operator s Manual 2 1 Rate Mode After turned on the Rate mode 1s displayed on the screen and the cursor points to Rate Rate Set the Rate 1n ml h O 0 mi gt ate 1 0 midh Flow 1 0 mih SEN PreS Set the Volume to be infused in ml 0 indicates PreS function is OFF Note Not available in Time Mode ii 7 Nl Go mi Time OH 0 Min Flow 1 0 nl h 5 Drug Set the drug being used rope Cire BEDS IN oo 1 0 slo rug Penicillin Flow 1 0 nish SE Bolus The bolus rate for the 5 10 and 20ml syringes are fixed at 100 200 and 400ml h respectively The bolu
3. depends on these settings and can be changed by the user To enter in the Configuration screen under STOP condition press CLEAR SET keys simultaneously Pressing CLEAR SET keys again returns to the original screen E Cra qa ie Boe iht f ah E Ho wal if a dh Sd E a E ep inh H E Hifi E ET AA ote Set pl The parameters are preset in the factory as table 1 Table 1 The parameter selected Suggested parameter value Manutact BD galie after shut down Manufact BD value after shut down Manufact BD value after shut down Manufact BD value after shut down Manufact BD value after shut down 0 IMpa Save setting value after shut down function to close KVO The after four digit of the Machine Serial No Save setting value after shut down Save setting value after shut down Near completion alarm ON Save setting value after shut down 15 DigiPump SR8x Syringe Pump Operator s Manual 4 2 Configuration Menu Description 1 5 50 syringe This parameter is determined by the length of the syringe in milimeters This value sets the syringe rate accuracy Specific instructions to change the syringe manufacturer or type are described on 4 4 Syringe Calibration Change 2 Basic occlusion value 0 04 0 16Mpa Default value is 0 1Mpa This value sets the occlusion alarm sensitivity of different syringes 3 KVO speed the adjustable range is 0 1 5 0ml h the default value is 0 1ml h It ent
4. not be changed Press BOLUS key to unlock and set the parameters Press BOLUS key once again to save and lock the Configuration Menu screen ATTENTION Syringe Length parameters are in LOCK condition when the pump is on and Length parameters can only be set in UNLOCK condition 2 Syringe Length parameters refer to syringes of 5ml 10ml 20ml 30ml and 50ml LOCK and UNLOCK function is only effective to these five parameters ee Unlock Condition Lock Condition LG i Tim eimi Y Aim Arm l dE Ath mim Cn l 3 50ml syringe setting on Configuration Menu screen rotate the knob to point the cursor to Set 50ml syringe Press BOLUS to unlock Press the knob to select the setting Rotate the knob to change the setting value Press the knob to save the new setting and lock 4 Repeat the process for each syringe to change settings Press CLEAR and SET keys simultaneously to enter the Configuration Menu screen 18 DigiPump SR8x Syringe Pump Operator s Manual 5 Flow Rate accuracy for the new syringe relies on the Length adjustment The user shall do one Test infusion with water at maximum Flow Rate using one calibrated recipient or one syringe to measure the infused volume Fine adjustment can be done in the Length calibration setting using the error measured Example New 20 ml syringe adjusted with 65 mm length Set the Flow Rate to 400 ml h and PreS to 10 ml Start the Infu
5. that required accuracy 1s obtained 8 Warning the used syringe shall not be installed on other infusion devices to avoid potential risk 9 Warning Keep cell phone or other high frequency power devices away from the syringe pump within 2 0 6m 10 Warning There is no air in line detection sensor in the pump Make sure that there is no air bubble in the syringe and the infusion set before using 11 Warning Stop using the machine when there is alarm 12 Warning Do not use the syringe pump in places where flammability and explosion can occur 1 4 Product functions and features 1 Precise control of infusion rate 2 Precise control of infusion volume 3 Smooth flow 4 Warning information about infusion completion near completion occlusion low battery syringe abnormal control abnormal 5 Advanced infusion control 1 5 Technical Specifications DigiPump SR8x Maximum infusion rate 1500ml h Different syringes have different maximum rate Range of flow rate Sml syringe 0 1 100ml1 h 10ml syringe 0 1 200m1 h 20ml syringe 0 1 400ml h 30ml syringe 0 1 600ml h 50ml syringe 0 1 1500m1 h 0 Iml h Bolus rate Sml syringe 100ml h 10ml syringe 200m1 h 20ml syringe 400m1 h 30ml syringe 400 600ml h 50ml syringe 400 1500m1 h mode on the syringe specifications Time 1 2000 minutes 33 3 hours Accumulated infusion volume 0 1 999 9ml Body Weight mode Weight 0 1 300 0 Kg Drug 0 1 999 9 mg Volume
6. volume and infusion rate It displays various parameters and the current operation progress on the LCD display 6 The rechargeable battery provides internal power for the pump when there is no AC power 1 7 Symbols and Labels 1 Product label pattern placed on the bottom of the pump enclosure Syringe Pump C ov Model DigiPump SR8x B1019 y 2 pl sn 86280 AR PAN Ae 2010 06 Input AC100 240V 50 60Hz 25VA tas Digicare Biomedical Technology Inc 107 Commerce Road Boynton Beach FL 33425 USA Tel 661 689 0408 Fax 561 689 0021 e mail salesOdigicarebiomedical com www digicarebiomedical com UA A OA TI DigiPump SR8x Syringe Pump Operator s Manual 2 Label Mark Description D Production batch No O Serial Number Caution Consult user manual Class II equipment Type BF applied part Protection against vertically falling water drops p Y NO pd It indicates that the equipment should be sent to the special agencies according to local regulation for disposal after its useful life Date of production Manufacturer CE mark Alternating current o rn z European Representative mm El 3 Standard configuration in the packaging box 1 Syringe pump one set 2 AC power cord one pc 3 Operator s manual one pc X Please contact Digicare if there are components above mentioned missing when you open the packaging O DigiPump S
7. 0 1 999 9ml Dose 0 1 9999 9 restricted to unit drug solution weight and syringe specification Unit mg kg h ug kg min DigiPump SR8x Syringe Pump Operator s Manual KVO rate Adjustable range 0 1ml h 5 0mlh KVO function starts when occlusion occurs press STOP key to cancel the KVO function when it 1s unnecessary 1n use IPX1 Infusion pressure The maximum pressure 1s 0 3Mpa Occlusion alarm pressure value is 40kPa 160kPa When pressure value is minimum infusion rate is 5ml h the longest alarm time is 10 minutes AC 100 240V 50 60Hz 1600mAh 25ml h after being fully charged Battery charge When syringe pump is connected to AC power the battery will automatically recharge about 8 14 hours to recharge fully Displayed information Flow rate accumulated infusion volume syringe specification battery capacity bed No AC power indicator operation indication etc Alarm information NEAR completion OVER OCCLUSION low battery abnormal 1 communications failure abnormal 2 pump stuck abnormal 3 wrong parameter volume limited alarm no AC power 380mmx190mmx180m 14 9 x7 5 x7 1 Operating conditions Environment temperature 5 40 air pressure 80kPa 106kPa relative humidity lt 80 Storage conditions Environment temperature 40 C 55C air pressure 50kPa 106kPa relative humidity lt 95 Applicable syringes Sml 10 ml 20 ml 30ml 50 ml syringes Electrical safety In accordance with
8. DigiPump SR8x Syringe Pump Operator s Manual DRIVEN BY INNOVATION IN MEDICINE Contents Ds IMrCO UICN aida 2 EL Product Intormatoli iii 2 1 2 Indication ot Usain a a E 2 LS Sale N ii 2 1 4 Product Functions and Features sesessecececsccecesesossecesesoesesoeoe 3 1 5 Technical Speciica OS aa rE 3 LO Froduct Descrip ind 5 EL Symbols and Label ctricos iia 5 1 8 Product Components tai ects ae E 7 19 Battery Operation is 8 1 10 Pump Installation 9 Zs IDTUSION CUINA 9 DA Rate Modena ds 10 IS wane etast 11 29 Body Webbie Modernas 11 2 4 Volume and Time Indicators ccccccscccccccccccsscccccccssssccccccssees 12 Da DPI racio s 12 3 1 Basic Operation SeDS cias 12 Ay COMMOUT AON ie ias 15 Al Conticuration Sereia e a a EES 15 4 2 Configuration Menu DeSCription cccccccccccccsscccccccccsscccccsscecs 16 4 3 Configuration Menu Operation cccccccccccccccccccccccccssccccccccsssecs 16 4 4 Syringe Calibration Change sesecesessesecesesoesecesesosoecesesesoecsesese 16 4 5 Occlusion Pressure Limit Change ssssesecesesoesesesesoesesesesoeseoesee 18 4 6 Occlusion Pressure Verification cccccccccccccccscccccccccsssccccccessees 19 5S Maintenance and SOFA np a a EE 20 S20 Pump Maintenant erroreren a a 20 S2 Transport and SOFA oa 20 5 3 Ereventive Test oi 20 54 Environmental Protectoras 21 6 Electromagnetic Compatibility and Interference sceccc
9. R8x Syringe Pump Operator s Manual 1 8 Product Components LS 14 13 124 10 Battery indicator light Flashing when there is no AC power connection AC power indicator light AC indicator light is on when using AC and the pump in turned on 3 Red alarm light Flashing when there 1s alarm signals 5 After the infusion starts the three lights flash Infusion indicator light alternately from left to right When the infusion is over all three lights are off ke In STOP condition press the SET key to enter or quit the Infusion Mode Settings screen seconds Night vision function Press this key for about 5 seconds to open or close this function after turning on the pump Start infusion after the installation and parameters 7 START key tine P Turn on the pump Press this key for about 2 seconds Turn off the pump Press this key for about 3 E Press the key to stop infusion and silence the alarm at the same time Knob is used to adjust the parameters and cursor Knob Press the knob to clear the alarm for 2 minutes and confirm the value newly set Keep pressing the button to move the piston freely 10 Clutch button When unfastening the button the clutch will gear and can only be moved by the engine to inject at maximum rate It returns to original DigiPump SR8x Syringe Pump Operator s Manual speed after releasing this key On settings menu screen press BOLUS to unlock the syringes parameters The Star
10. Step 1 Get one new brand syringe of each size to change the Syringe Calibration factor 5ml 10ml 20ml 30ml and 50m l Step 2 measure the length of syringes 1 Put the syringes at the calibration ruler and measure the distance from Zero 0 to the syringe size indicator marking In the 20ml syringe figure bellow the length is from 0 to 20ml For example the BD 20ml syringe length is 70 0mm 2 This part of the process is done with empty syringes The syringe needs to be correctly installed in the pump as follow in order to achieve correct calibration Step 3 Turn On the DigiPump SR8x syringe pump Install the 50ml syringe or smaller size to change Repeat the process for other syringes decreasing the size each step Warning The correct installation of the syringe and the piston is essential when changing the calibration settings or error may occur Push the syringe piston to compressing the syringe all the way to the end Push the Syringe Piston to the end Compress the Syringe with the Syringe Holder 17 DigiPump SR8x Syringe Pump Operator s Manual Step 4 Change Syringe Parameters ATTENTION Each syringe has two calibration parameters the length in mm used to adjust the Flow Rate accuracy and the height used by the Syringe Size indicator 1 Stop condition press CLEAR and SET keys simultaneously to enter the Configuration Menu screen 2 The Length parameters are in LOCK condition and can
11. and wires 5 4 Environmental Protection The life time of the products is 3 years Machines over its life should be discarded please contact the manufacturer or distributor for more relevant information 1 DigiPump SR8x syringe pumps that are no longer in use could be delivered to its distributors or manufacturer for proper recycling 2 Used Lithium Polymer batteries could be delivered to its distributors or manufacturers for handling or handling it as per applicable laws and regulations 6 Electromagnetic Compatibility and Interference This pump has a function to prevent external interference including high intensity radio frequency radiation magnetic field and static Users should avoid the use of the mobile phone within 0 5 meters from the machines This pump is quite low in electromagnetic frequency which will not interfere with the surrounding electronic equipments However the pump will cause a certain amount of electromagnetic radiation which is within the requirement of IEC EN 60601 1 2 and IEC EN60601 2 24 If the pump interferes with other equipment measures must be taken to reduce this kind of interaction such as the relocation of the two instruments 21 DigiPump SR8x Syringe Pump Operator s Manual 7 Alarms 7 1 Alarms and Actions Display The syringe 1s almost empty and close to end of infusion The syringe 1s empty and infusion is finished syringe or administration set with occlusion Occlusi
12. ed to the factory with prior authorization transportation prepaid Not covered by this agreement are repairs necessitated by any of the following conditions 1 Inadequate power or power failure 2 Neglect abuse or misuse of equipment 3 Servicing of equipment by person s other than DIGICARE INC 4 Any unit opened or tampered with without prior authorization When returning an infusion pump for warranty repair you must first contact DIGICARE BIOMEDICAL TECHNOLOGY INC to receive a Returned Goods Authorization Number RGA that is to be clearly marked on top of the shipping carton Please make sure that your company name shipping address area code and telephone number and person to contact is located in and or on the box ANY UNIT THAT IS RETURNED TO THE FACTORY WITHOUT AN RGA WILL BE REFUSED Model Serial Dealer Name Date Equip Purch Expiration Date _ l DIGICARE BIOMEDICAL TECHNOLOGY INC 107 Commerce Road Boynton Beach FL 33426 USA Phone 561 689 0408 Fax 561 689 0021 Email sales digicarebiomedical com www digicarebiomedical com 25 DigiPump SR8x Syringe Pump Operator s Manual
13. ers into KVO state in the course of occlusion after running Set the value to OFF to disable KVO function 4 ID No setting The value is the last four digit of the syringe pump Serial No in order to identify each single machine The adjustable range is 0 9999 5 Bed No This value is the BED No where the syringe pump is placed The range is 0 999 When no alarm is triggered the bed No is displayed at the first line of the main screen and is overwrite by any alarm message when triggered 6 Near On Off The NEAR completion alarm can be enabled or disabled in this setting 4 3 Configuration Menu Operation Enter the Configuration Menu Turn the knob to move the cursor through all different parameters The lock icon in the upper right of LCD is locked by default In the locked state the user can not change settings Press BOLUS key to enter the unlock state Press the Knob to highlight the parameter Turn the Knob to adjust the parameter setting and press the Knob to save and lock the parameter setting After all parameter settings change press CLEAR SET to return to STOP condition Unlock Condition Lock Condition 4 4 Syringe Calibration Change The DigiPump SR8x syringe pump 1s calibrated in the factory to the following disposable syringes Sizes for Reference in mm BD 440 60 700 818 895 Length from Size indicator marking to 0ml lui andi a 16 DigiPump SR8x Syringe Pump Operator s Manual
14. for TET The parameter is Refer to item 4 4 Syringe installation A seine en ie Or E incorrect Solution for No 3 Something Contact Digicare Technical wrong with the Support sensor of syringe pump Something wrong with the Solution for No 1 CPU control Contact Digicare Technical the operation Support Alarm for of Electric Solution for No 2 malfunction gt Stor Machines is Contact Digicare Technical n wrong Support WRONG2 alarm check the Solution for No 3 parameter Reset all of the parameter incorrectly 7 2 Troubleshooting Press key The syringe The syringe is not properly Reinstall the syringe Check but the light for pump does not installed if the right indicator light is LED operation is work The syringe configuration on Or configure the syringe not on parameter 1s not rightly set settings Turn the knob but The speed of The syringe pump cannot be Press key to stop the the speed of syringe pump set during operation syringe pump Turn the knob syringe does not does not change to set the needed value and change press the knob to save the value Low Battery The syringe Bad battery The syringe Connect the syringe pump to pump willturn pump can not be used before A C power supply before off if not it is connected to A C operation connected to AC electric supply power 23 DigiPump SR8x Syringe Pump Operator s Manual The occlusion light is on and the syringe pum
15. l h and Start infusion To access the internal sensor output value 19 DigiPump SR8x Syringe Pump Operator s Manual screen pressing the SET STOP keys simultaneously Fotentio m L Battery Cap Verify in the manometer the pressure at witch the occlusion alarm is being triggered The default is 0 1Mpa 14 5 PSI 1 8kg in the Sensor display Press SET STOP keys to turn back to the infusion screen 5 Maintenance and Storage 5 1 Pump maintenance 1 Keep the pump clean Use a piece of cloth to wipe the syringe pump clean when any liquid drops on the pump 2 Special attention must be paid when cleaning the syringe pump Use a tampon moistened with 75 of alcohol to wipe the outer case 3 Make sure the AC power is disconnected and the pump is turned off 4 Do not clean the syringe pump by using xylene acetone or similar corrosive cleaning products to avoid outer case damage The above mentioned operation is just for guidance More proper methods should be used for sanitizing effect inspection 5 2 Transport and Storage Transport and Storage environmental conditions temperature 40C 55C air pressure 50 kPa 106kPa relative humidity lt 95 5 3 Preventive testing 1 Flow Rate accuracy check Use one volume calibrated recipient to verify Flow Rate and Volume accuracy every time new syringes are being used or every six months 2 Battery Inspection O Batteries have limited life time Replace the batteries when the
16. me press the knob to select and rotate the Knob to change the time programmed for the infusion in minutes and press the Knob to save the Time setting Rotate the Knob to Vol Volume to be infused Press the Knob to select and rotate the Knob to change the Volume setting The Flow Indicator automatically indicates the resulting programmed Flow Rate in ml hr Press the Knob to save the settings Press SET key to return to main screen 2 3 Body Weigh Mode In the Body Weight mode the user sets the following parameters and the resulting Flow Rate is automatically calculated and programmed Body Weight BW Patients body weight in Kg Note 1 0 Kg 2 2 1b Drug concentration Drug amount measured in mg Volume Vol Volume of Drug concentration per ml Dose Desired Drug dose in mg kg h or ug kg min Press the SET key and using the rotary Knob set all required parameters 1 Goin E L U my kog h Flow 1 0 mieh 11 DigiPump SR8x Syringe Pump Operator s Manual X Body Weight Mode Flow conversion mg kg h flow rate DOSE x WEIGHT x FLUID VOLUME DRUG ug kg min flow rate 60 x DOSE x WEIGHT x FLUID VOLUME DRUG 1000 2 4 Volume and Time Indicators During infusion in RUN status the accumulated infusion volume is displayed Turn the adjusting knob to display the Preset Volume Preset Time Drug and Bolus indicators 3 Operation 3 1 Basic Operation Steps Step 1 Power connection Attention Insert the powe
17. nfusion 13 DigiPump SR8x Syringe Pump Operator s Manual 2 Press START key the syringe pump starts infusion The infusion indicator in the lower right side of the screen displays the infusion status The Infusion indicator speed is proportional to the Infusion Rate Step 7 BOLUS Infusion During the infusion process the 3 infusion indicator lights will flash alternately from left to right If Bolus infusion is required press and hold the key during infusion the machine will infuse at maximum rate for the syringe in use The Infusion Rate returns to original rate after releasing the BOLUS key 400 1500 mlM Step 8 Infusion complete 1 When the infusion is near completion the alarm indicator light flashes and LCD screen displays NEAR to remind the user that the infusion will complete soon Attention press the knob to temporarily stop the alarm It sounds again in two minutes 2 When the infusion 1s finished the pump stops infusion and the OVER message is displayed on the LCD screen Attention press the knob to temporarily stop the alarm It will sound again after two minutes Step 9 Turn off the power Press POWER for 3 5 seconds and then release it the machine power is turned off 14 DigiPump SR8x Syringe Pump Operator s Manual 4 Configuration 4 1 Configuration Screen The DigiPump SR8x syringe pump 1s pre configured with default settings The accuracy of the syringe pump for each different syringe
18. on h m Occlusion alarm alarm C need sensitivity J adjustment Syringe pump pressure sensor has problem Battery too low Bad battery aging or something wrong with the battery charge circuit LO BAT ALARM 22 DigiPump SR8x Syringe Pump Operator s Manual l press the knob to cancel the alarm the alarm will ring again after two minutes 2 Press key to stop the syringe pump and cancel the alarm 1 Press the knob to cancel the alarm the alarm will ring again after two minutes 2 Press key to stop the syringe pump and cancel the alarm Solution for No 1 Press ISTOPI key to stop the syringe pump and cancel the alarm eliminate infusion occlusion then press the START key to start infusion once again Solution for No 2 Refer to Occlusion Pressure Limit Change 4 5 Solution for No 3 Contact Digicare Technical support Solution for No 1 Connect to AC power supply to charge the battery Solution for No 2 Contact Digicare Technical Support No AC power A supply for the AC A EEE syringe pump indicator is e p 2 0 2 m1 MA Something OFF E o l wrong with the _ power supply circuit of syringe pump Solution for No 1 Check if the power cord is not plugged in or not properly plugged in Solution for No 2 Contact Digicare Technical Support The syringe fall Solution for No 1 off or is not Reinstall the syringe installed well Solution for No 2 Alarm
19. p stops The occlusion alarm signal sounds soon after the syringe pump starts operation The indicator light of the corresponding syringe is not on Wrong syringe size 7 3 Service Attention The occlusion sensitivity value of the syringe pump 1s not properly set The compressed strut is not tightly pressed on the syringe Or the syringe settings is not rightly set Reset the occlusion sensitivity value Pull the compressed strut of the syringe up reinstall the syringe and press it onto the syringe tightly Or reset the syringe settings 1 If there is any untreatable fault or question please stop using the Digipump SR8x Syringe Pump and contact Digicare for Technical Support and Service 2 Unauthorized person has no right to repair this product otherwise Digicare will take no responsibility 24 DigiPump SR8x Syringe Pump Operator s Manual WARRANTY TERMS amp CONDITIONS DIGICARE BIOMEDICAL TECHNOLOGY INC covers all of their infusion pumps with a 1 year warranty Infusion Pump DigiPump SR8x 1 year parts and labor Accessories 90 days on accessories only DIGICARE BIOMEDICAL TECHNOLOGY INC will provide the necessary parts and labor to maintain the monitor s listed on the Warranty Certificate in a usable condition during the covered period DIGICARE BIOMEDICAL TECHNOLOGY INC will at its option repair or replace any product which proves to be defective during the warranty period if return
20. r cord into the power supply 50 60 HZ AC 100 V 240V Connect the power cord as shown Step 2 Turn the Power On Turn on the power the syringe pump will display version and ID information as shown 12 DigiPump SR8x Syringe Pump Operator s Manual Step3 Install the syringe Install the syringe as shown in the figure 1 Press key 2 Press clutch button drag piston to fit the desired syringe 3 Pull the Syringe Holder and move 90 to the left or right 4 Before the installation manually remove all the air inside the syringe and then place the syringe on the mounting grooves 5 Return the Syringe Holder to original position compressing and holding the syringe in place When the Syringe Holder compresses the syringe the syringe size indicator is displayed in the screen Step 4 Select the Infusion Mode and Flow Rate settings 1 Press key once 2 Set the infusion parameters as Flow Rate Volume to be Infused and Infusion Time Step 5 Clear accumulation 1 Make sure the pump is in stop condition 2 Press CLEAR key to eliminate the accumulated volume Note accumulated volume can be cleared only in STOP condition Step 6 Start the infusion 1 Verify the syringe installation and the numerical indicators and make sure the corresponding syringe indicator light is on Attention Press CLEAR key to clear the accumulated volume before pressing START key if you don t want to add this volume to the new i
21. ring damage to the pump 3 Observe the safety of the machine avoiding drop vibration or bump the pump specially during transport 4 Do not use the pump in place which keeps chemical medicine or gives out poisonous gas Observe if the voltage frequency and power of the AC complies with pump specifications 6 Make sure there is no high frequency equipment close to the pump or it can cause interference 3 Precautions before using the machine 1 Make sure the power cord is safely and correctly connected 2 Make sure the ON OFF function and operation of the pump is ok 4 Observe if the patient s vein passage has been established 4 Precautions during operation 1 Supervise continuously if the machine and patient are normal 2 When detecting anything abnormal with the machine or patients consider the patient s safety first then stop infusion and take proper action 5 Precautions after finishing using the machine 1 Turn off the power and pull out the power cord as per requirement 2 Clean the pump before storing it for convenient use next time 6 Warning using improper syringe may cause damage to the syringe 2 DigiPump SR8x Syringe Pump Operator s Manual 7 Warning when using Non standard syringe or setting syringe parameters incorrect the accuracy may not be maintained and the discrepancy may reach 5 or above When changing Syringe types and syringe calibration settings it is the user responsibility to ensure
22. s rate for 30ml syringe is 400 ml h and can be changed from 400 to 600mlI h The bolus rate for the 50ml syringe is 400 ml and can be changed from 400 to 1500ml1 h We En E e o BEolu 400 0 mileh Flow too Flow Flow Rate programmed indicator in ml h 10 DigiPump SR8x Syringe Pump Operator s Manual When the Volume to be infused is set the Preset indicator change from N to P When the infusion is started the programmed Volume will be infused at the programmed flow rate When the programmed Volume finishes the Preset indicator returns to N When the Time to be infused is set the Preset indicator change from N to T When the infusion is started the programmed Flow Rate will infuse during the programmed Time When the programmed Time finishes the Preset indicator returns to N If both Volume and Time are preset the Preset indicator will display the first programmed and will execute the first programmed first When the first preset parameter finishes press STOP and press START to start the second parameter preset 2 2 Time Mode In Time Mode user sets the Time of the infusion in minutes the Volume to be infused in ml and the resulting Flow Rate is automatically calculated and programmed Press the SET key to change the Infusion Mode The cursor points to Rate Mode Press the Knob to select and rotate the Knob to change to Time Mode and press the Knob again to select the Time Mode Rotate the Knob moving the cursor to Ti
23. sion When the infusion finish suppose the actual volume infused in the calibrated recipient is 9 ml The error is 9 10 0 1 10 Increase the Length calibration setting 10 65 x 1 1 71 5 Repeat the Infusion in the calibrated recipient and verify if the error 1s in the specified range of 3 4 5 Occlusion Pressure Limit Change 1 The occlusion pressure default is set to 0 1 Mpa 14 5 PSI and the adjustable range is 0 04 0 16 Mpa 5 8 to 23 2PSI The actual occlusion pressure for one specific syringe will vary with the dimensions and resistance of the syringe To change the occlusion pressure alarm limit press the SET CLEAR keys Rotate the Knob to point the cursor to OCCL Gate 0 10 Mpa In the Configuration Menu Increase the Occl Gate setting in steps of 0 01 Mpa 1 45 PSI in order to decrease the Occlusion alarm sensibility Maximum is 0 16Mpa Decrease the Occl Gate setting in steps of 0 01 Mpa 1 45 PSI in order to increase the Occlusion alarm sensibility Minimum is 0 04Mpa 4 6 Occlusion Pressure Verification The Occlusion alarm pressure can be verified using one calibrated manometer The DigiPump SR8x internal pressure sensor output can be read in the screen for verification only Place one 10 ml syringe filled with water Install one administration fill with water and connect to the manometer input or just close the administration set output to create one occlusion Program the Flow Rate to 400 m
24. sssessceeees 21 T AVATAR EAS A A a 22 LE Alarms and ACHONS nda 22 Hz TroUblesho0tMO iii 23 Usd ASE NIC os 24 Warranty Teri asirio 25 Rev 3 DigiPump SR8x Syringe Pump Operator s Manual 1 Introduction 1 1 Product Introduction The DigiPump SR8x syringe pump is a combination of microelectronic technology and innovation in nursing medicine It can control the infusion rate and total volume precisely and continuously over long hours on a large scale The DigiPump SR8x syringe pump is a hi tech equipment which can be used in all areas of modern nursing The patient s infusion situation and alarm information e g infusion completion occlusion can all be seen at the central station This greatly improves the medical nursing quality reduces the nurse s workload and ensures the patient s nursing safety For patient s safety the DigiPump SR8x syringe pump shall always be installed in a firm place like an IV pole when used during patient transport 1 2 Indication of Use The DigiPump SR8x syringe pump is indicated for use in hospitals where patients need precise and accurate intravenous microinfusion of liquids at steady speed 1 3 Safety 1 Unqualified and untrained people are not allowed to operate the machine 2 Precautions for installing the machine 1 Place the pump in dry place 2 Do not place the pump in places where atmosphere pressure temperature humidity sunshine dust salt and 10n air may b
25. t up syringe default parameters are in locked condition and can not be changed in this condition Push the piston of the syringe condition press the key to clear the accumulated volume CAUTION On Settings gt menu screen all syringes parameters will be si ney cleared when press CLEAR key in the unlock condition Then the syringes can not be detected and you have to reset the syringes parameters Syringe Holder It holds back the syringe to avoid syringe dropping Quick Help 15 SET CLEAR In condition press both keys simultaneously to enter or quit the Settings Configuration menu SET STOP In condition press both keys simultaneously to enter or quit the Sensor screen Points to the parameter to be set 1 9 Battery Operation 1 As shown in figure 1 after the pump is connected to the AC power and turned on the AC indicator light is on and the battery begins to charge automatically When the battery 1s fully charged it will stop charge automatically 2 As shown in figure 2 the AC indicator light will be off if the AC power is disconnected And the battery indicator light is flashing which means the syringe pump is using the battery capacity The syringe pump will issue a warning sound to alert the user to charge the battery when the battery power shortage X It requires 8 14 hours to charge the battery fully after it is fully depleted AC Indicator ON j Battery charging Bat on use Bat capacity DigiPump SR8x
26. y are not holding charge If replacing batteries are required please contact Digicare Model Lithium Polymer battery 7 4V 1600mAh O Use the pump running in its internal battery at least once a month till its depletion and the power of the pump turns off automatically for insuring battery performance and 20 DigiPump SR8x Syringe Pump Operator s Manual prolonging its life After battery depletion recharge 8 14 hours for next use O The following inspection need to be taken every six months 1 Connect to AC power about 8 hours for recharging 2 Turn on the syringe pump and install 50ml syringe 3 Set up infusion rate 25ml h and start syringe 4 Observe operation of the syringe pump till it turns off based on low battery If the syringe pump takes 90 minutes or even longer time from starting infusion to shutdown the battery is in good condition If the syringe pump takes 45 90 minutes from starting infusion to shutdown the battery life close to its end If the syringe pump takes less than 45 minutes from start infusion to shutdown the battery is up to its end and recharge of battery is required 5 After battery inspection recharge the battery again for usage next time 3 Routine Maintenance intervening time Routine maintenance procedures Thoroughly cleaning of syringe pump case is prea a a el required before or after long period of storage At least once a year for inspection 1 Inspection of AC power cord