Home
        User Manual - Arteriograph start
         Contents
1.     6  Troubleshooting    Problem    No communication    No arterial function  parameters are  displayed    No print preview    The PDF report cannot  be opened    Database import  cannot be completed    Cannot send e mail  successfully    Error messages    Error codes  as shown on the LCD  and their explanation are specified below  It    Possible causes    Communication adapter is not  connected    The device is not recognized by the  PC    No or low batteries in the device  The device was turned off    Loose cuff    Arrhythmia    Tremor    No printer have been installed to  the PC     Adobe Reader is not installed to  the PC     Not the proper database was  opened or the database was  damaged     No e mail client is installed to the  PC     What to do    Plug in the adapter and follow the  instructions of Section 2 5     Go to Control Panel Bluetooth  Devices and check whether the  device is present  If not  follow the  instructions of Section 5 2     Replace the batteries     Turn on the device by pressing its  button twice     Tighten the cuff     The method cannot reliably assess  arterial function parameters in  arrhythmic patients     If possible  try to terminate  tremor     Install a printer or use the built in  virtual printer of Windows     Download and install the latest  version of Adobe Reader     Import the database by following  the instructions of Section 5 4 8 3     Install an e mail client and set it to  be the default one     Example     should be noted th
2.    s characteristics  see Section  5 4 2 1 1           E Confirmation            2  Arm circumference  26 cm  Suggested cuff type  2  bls  26x8cm   Operators name  ARTERIOGRAM    o A med                     The measurement process starts with the SW connecting to the Arteriograph device  The scanning of  the active ports and the communication process can be tracked on the monitor  While this  communication is in progress     CONNECT     See Section 5 4 1  will appear on the LCD of the Device     After a short time measurement will start automatically by performing a blood pressure  measurement in the first phase  During measurement the recorded pulse wave curves can be tracked  on the screen    In the second phase the pulse wave recording for pulse waveform analysis is performed    e First the Device will inflate back to the diastolic pressure level and register the pulse waves  for a previously determined duration  8 sec by default     e Second  the Device inflates further to the suprasystolic pressure level  by default to the  measured systolic value   35mmHg  occluding the brachial artery completely in this way and  registers the pulse waves for a previously determined duration  8 sec by default      The time and pressure offset values can be modified through the system settings function in the  above described manner  See Section 5 4 1     During the measurement there is a real time data transfer enabling the user to see the recorded  oscillatory waves on the screen  At t
3.   You ane logged in as ATER    F  e y Togh    e   l  7 E 4   BS START pm 7 med E  Y a ap aati    Copa      bei I Lier B dec pa Y Fi u  Liane 2 Earl rurea    D  a    Time Far ra  gt  ade at Det LH T   HUM HU    23 1  55   229  x   230  ma   12   ma   i ei     Mali      Pad        Pe vy u un We q dd  yr  4 8  m s     1 85  ms     LE 7  mmHg     117 7  ma     Movement artefacts    Movement artefacts distort the waveforms and lead either to not analysable measurements or  to results with inflated variation     How to recognize  the baseline of the recording undulates or some pulse waves noticeably differ  from the others     How to avoid  emphatically ask the patient not to move  Manual evaluation  see Section 5 4 5   can help to find a regular sequence in the suprasystolic measurement phase     DB Temoltes    Lnerogrer   You are logged in e ARTESIOGRAM iu  Vie Vom Took      I 7 AR stant    a Y mod         n Fy   11   A  Le     ere min    po rem ar     E pere Eees 6 AID as 5 P 0    Emna E rh ram    Das   Tma Fed raro      sin 14 o   150    Y we    gar       5 4 5 How to analyse the data    Select a patient and an examination thereof you intend to analyse  You have two evaluation options  available  automatic and manual        The former is fast and convenient  Automatic evaluation will take    E Automatic analysis 5 39     D       CEA      place immediately after a patient s examination is selected  The    same can be achieved by clicking first on Automatic analysis S35  for supr
4.   the separation distance for the higher frequency range applies     NOTE 2 These guidelines may not apply in all situations  Electromagnetic propagation is affected by absorption and reflection from  structures  objects and people        Notes    Notes    Notes    UN    putting precision first       Arteriograph    Medexpert Arteriograph User Manual     2011 Medexpert Ltd  All rights reserved     
5.  Print the Report    40   5 4 6 3 How to send a Report in PDF format    43    5 4 7 Export the measurement data    ss 43    6    5 4 8 Backup and restore the database               00   000onnsoseennsssrerrrsssrerssssrerssssrerossereesssreresssrereesssreresseree  5 4 8 1 Backup E A eee  5 4 8 2 Restore datan id cie  5 4 8 3 Import previous database  nase een een   IFOUDIESNOOLNG ns ee ee   Technical Specilicallon     una  pen eo   7 1 Electromagnetic compatibility                                                     ss    1  Introduction   1 1 Preface  Thank you for choosing Medexpert Arteriograph     This state of the art device and method enables  recording and analysing arterial pulse waves in an exceptionally easy  reliable and accurate way  With  this entirely new implementation of oscillometric method a complete arterial function evaluation can  be achieved     1 2 Contents of the manual    This manual helps you in setting up and starting to use the Medexpert Arteriograph TL2 device as  well as installing and setting up the Medexpert Arteriograph Software up to versions 3 0 0 0     Attention  Before first use  please read and understand this document carefully      gt     1 3 General information    e When active  the Device does not cause any electromagnetic disturbance and its immunity to  environmental effects is also sufficient  The measurement data are downloaded into the PC  through wireless Bluetooth communication  The electromagnetic compatibility between the  Dev
6.  TCrep_TestElisabeth 26032007 07 43 pdf     As a next step the default e mail client     if there is any     is loaded automatically with the generated   pdf file attached to a new letter     5 4 7 Export the measurement data  Formerly saved measurement results  see Section 5 4 5  can be exported into a formatted text    E file by choosing Export data under the Tools menu or clicking the Export data    button on the  toolbar        Step 1 Select a User whose patients    measurement you would like to export by clicking the  checkbox next to it  The Select all option is a quick way to select all the listed users  All  the registered users are listed in this window  Selection is confirmed by clicking on the  arrow     Export data   1  Step   Hanse nalad uzena     Select all  Ga API TOME Mhecairent Naghain Ha  1       Step 2 Select the patients whose measurements you would like to export by clicking on its row   All the patients belonging to the selected user are listed here in alphabetical order of the  family name    In case there is a great number of patients  a particular patient can be found by using the  built in search function  For a quick search click anywhere in the patient list and type in  the initials of the family name of the given patient  The built in search is not case    Step 3    Step 4    Step 5    sensitive    Multiply selection can be done in the common Windows way  using Ctrl or Shift while  clicking   The Select all option is a quick way to select all the 
7.  The Information Menu    Med    Arteriograph   You are logged in as ARTERIOGRAM       The most important information about the manufacturer and the   mes   e Manufacturer and software version information    software can be accessed through the Information menu  The menu  gt    also includes this User   s Manual          200 ms cm   Y axis compression 2   Baseline    Family name  First name    3 2 2 The Toolbar of the Program    The toolbar can be found under the menu bar and contains icons  which allow direct access to the  major functions of the program  The Toolbar can be displayed or hidden by the Toolbar command of  the View Menu     3 2 3 The examinations window    All the previous   if there are any     examinations of the selected patient are listed here by the date  and time of recording     3 2 4 Theresults block    The calculated hemodynamic parameters are displayed here when an analysis of measurement data  is performed  automatically or manually      3 2 5 Patient details field    The name and date of birth of the patient are displayed in this field     3 2 6 The measurement data window    During the measurement the data are displayed here real time in the form of non calibrated  pressure curves  oscillations  as recorded by the device  The recorded and saved pressure waves can  be opened and displayed at any later time  The User can also select to display the amplitudes of the  oscillometric curves throughout the blood pressure measurement phase  or the estimated
8.  are computed during the measurement   Sys     Brachial systolic blood pressure  mmHg   Dia     Brachial diastolic blood pressure  mmHg   HR   Heart Rate  beat min     MAP   Mean Arterial Pressure  mmHg   As calculated from the brachial systolic and diastolic blood pressure   MAP   diastolic blood pressure    systolic     diastolic blood pressure  3    PP     Brachial Pulse Pressure  mmHg   PP is the difference between the systolic and diastolic blood pressure in mmHg     ABI   Ankle   Brachial Index  Dimensionless index   The ratio of the ankle systolic blood pressure and brachial systolic blood pressure values     Aix brachial     brachial augmentation index       Augmentation index   The difference between the amplitudes of the late  backward  systolic wave   P2  and the early  forward  systolic wave  P    over the pulse pressure  PP  and multiplied with 100  Aix    P   P    PP 100    Aix aortic     central augmentation index      Calculated on the basis of a very strong  R gt 0 9  linear relationship between the brachial and central  augmentation index     ED   Ejection Duration of the left ventricle  ms   It is the period of the mechanical systole  i e  the time span between the opening and closing of the  aortic valves     RT   Return time  ms   The RT is the time of the pulse wave travelling from the aortic root to the bifurcation and back     PWVao   aortic pulse wave velocity  m s    The velocity of the pulse wave in the aorta    This is calculated from the travelled
9.  characteristic point on the curve  see Section   5 4 5     Diastolic record  This part contains the parameters resulting from the analysis of the pulse   wave curves recorded at the diastolic pressure step  There are also two graphs showing a   section of the pulse wave curves and a single curve  the last wave if several waves are   selected for analysis   The DRA  SAI  and DAI parameters can be defined only if an ED value is   available from the previous phase  Suprasystolic record     Footer      Institutional and operator data on the left hand side  if set  Section 5 4 6 1       Manufacturer   s data complemented with the version number of the SW on the right   hand side     Suprasystolic record  The first part of this page contains the quality control of the  measurement  A section of pulse wave curves recorded on the suprasystolic pressure step  are shown in a graph to enable visual control over the quality of the waveforms  The value of  SD PWVao is also represented here    Pulse Wave Velocity measurement  This part highlights the measured aortic pulse wave  velocity and return time values    PWV   Arterial Age Assessment  The current measurement is placed in an age     PWVao  graph with smoothed reference percentile curves  see Section 5 4 5 3   The point of the  measurement is coloured according to the ranges in the graph legend  The estimated arterial  age of the patient is also given here in decades except for pooling over the vascular age of  60  A more precise es
10.  could be found in the suprasystolic   17   m s  ble  measurement sequence  The SD appears in red colour in the Results field to       ms   warn the User on the quality problem  The results of the analysis should be  ppao   59 5  mmHg     rejected and the measurement should be repeated     5 4 4 2 Quality control by visual inspection    Although the SD PWVao performs considerably well in detecting erroneous measurements  further  visual inspection can help categorizing the type of the error  Consequently it can be avoided when  repeating the measurement  By visual inspection disturbances can also be detected in measurements  that were categorized to be of good quality by the SD PWVao parameter     Improper cuff size or tightness    In this case no early and late systolic wave can be differentiated on the pulse wave form  Due to  the larger cuff size or the looser fit  the cuff system loses its sensitivity to pick up higher  frequency components of the oscillatory waves  This often leads to recording small amplitudes   which case is always indicated by the software with a warning message     How to recognize  no reflected component can be seen on the wave     How to avoid  use a smaller cuff or tighten the cuff according to the instructions  see Section  5 3        F  TercioMed    Artenograph   You are logged in as ARTERIOGRAM o a  File View Took Infeemation    e  gt  am 4 sm Em EN mede   u  x oy AD mater   Y ges comprero 2   Badr D  gt   Reset a Mina arab 5 P D kb Cavas  Disc
11.  distance  measured as the suprasternal notch     pubic bone  distance  of the pulse wave in the aorta divided by the measured transit time  RT 2      SD PWVao  m s   A parameter informing about the quality of the measurement  The PWVao is calculated from each  pulse recorded and the standard deviation is featured  For more details  see Section 5 4 4 1     SBPao   central systolic blood pressure  mmHg   Calculated on the basis of the physiological relationship between diastolic BP  mean arterial BP   peripheral and central Aix and central SBP     PPao   central pulse pressure  mmHg   The difference between aortic systolic and diastolic BP values     DRA   Diastolic Reflection Area  Dimensionless index    DRA provides information about the quality of diastolic filling in the coronary arteries  taking into  consideration the duration of the diastole and the area between the expected  theoretical  diastolic  pressure curve without reflection and the truly measured diastolic curve with reflection     SAI     Systolic Area Index      SAl is the systolic part of the area under the entire pulse curve     DAI     Diastolic Area Index      DAI is the diastolic part of the area under the entire pulse curve     5 2 System setup    After installation  connect Arteriograph to the PC by following the instructions of the enclosed  Bluetooth Operation manual  The passkey of the device is OOOO   The device should not be farther than 10 metres from the PC for optimal operation     Attention  T
12.  i till TensioMed  TensioClinic 40  C    TL2  X     S N  COSA case  4 13 12 11 10   Figure 2    15             1    2 2 The accessories of the Device    Arteriograph has the following accessories     e 4pcs  AA size alkaline batteries    e Three D ring cuffs of different sizes  see below   e Medexpert Arteriograph Software installation CD    e A tape measure for measuring the jugulum     symphysis distance  e Bluetooth adapter kit  Bluetooth adapter  Driver CD  Operation manual     e A protective pouch for the device  e A User   s Manual    The cuff sizes are  as follows     Cuff type Bladder dimensions    01  02    34 x 8 cm    26 x 8 cm    62x9cm    52x9 cm    2 3 Installing the Device    Arteriograph is a battery operated device     e Insert 4 durable alkaline AA batteries into the Device with taking care of the right polarity     see Figure 3     e Or insert 4 AA sized rechargeable Ni MH or NiCd batteries as per the above instruction    Please note that new batteries must be pre charged    o For problem free operation minimum 1 500 mAh chargeable Ni MH or NiCd batteries    Note  The clock circuits of the Device are fed by integrated NiCd HA 35 battery cells  therefore    the operation of the clock is independent from the AA battery supply         gt     are recommended     Attention  If you do not intend to use the Device for a longer period of time  remove  the batteries and store them in a dry and cool place out of little children   s reach  Do  not expose them to inten
13.  in the  same way as in a normal automatic analysis        r     m  F TercioMed    Arterograph   You are logged in as ARTERIOGRAM            File View Toot Information   gt  r Am cy u  Ar     t    Da RES   START an HU med  x maic   mes COMPAR   2   Rawr 0      mS   Davor    a e    Fay rare Teszt  it nam Beteg Dale of Det 01 01 1975           t    52 0      11 3        270  ms     j     125  ms  i N   N N          em  38 pin do E J   REN 10  fern E Mes gest NTE     8 5  m s  AT    et Le      0 00  ms       28 6  mmHg     111 6  mmHg        e Incase a pulse wave of only a single heartbeat is highlighted     of either those that were  registered at the suprasystolic or those registered at the diastolic pressure level     it will  appear in a separate window displayed on the millimetre net   Please note that in this case  SD PWVao cannot be calculated      The said separate window can have two kinds of content     o The one highlighted pulse wave at the suprasystolic pressure step is represented by a    green curve  There are four markers shown on the curve  the meaning of those can    be found below              r 7  F TersioMed    Arteriograph   You are logged in as ARTERIOGRAM     el  File Wew Toot Information    Ae Ses   sas  Ls ZM medexport Y  K oat pr ZU  mam   Y aan Comp 2   Rare 0 Reset A Atomai aah 535 D ii Capa  er Fand rare Toszt  Port name Beteg Date of Beth 01 01 1975   Aix brach 121   52 0       11 3     83   265  ms j   mel   N    5  ms A     125  me  A       53  o
14.  on         When the button is on the control tools appear  With use of these the  sec   CoC      cursor can be moved in small or big steps     Note  The Cursor feature is only available at the default X axis step of 200 ms cm     Note  The oscillatory signals are not calibrated and therefore should not be considered to       represent true pressure values        The above mentioned functions are illustrated in the picture below              If TersioMed    Arteriograph   You are logged in as ARTERIOGRAM eee  ee ABS    Bm  Les EE medoxport fl  X mat step 200 malen   Y ae compeatti  n 2    Rene 0   Reset El    san anus 535 D Ci Gas  0 1 E  ad PATIENT DETAILS Fay rare Teszt  Date   Tise Feit name Beteg Date of Bem 01 01 1975  005201  0914      gus ale  Herbie    ime 4000 fru  akude 240  es 182 mmHg                          Aix brachint   181    52 0        Aux aorls    n3 x  9    265  ms    A     _ _        _        R 9     1 PA  125  ms    jH PA fo  UG SY j ha  zu 0 PE  53  cm  30 gu HOG HUB   o Ocio van  iiti DR A set  PWVao  8 5  m  or  Loris 60         ms     PPao    28 6  mmHg   ShPao  111 6  mmHg  U  Pulsa wyra Aang    ponssuam values   Asteri age assessment       5 4 5 2 Oscillation amplitudes during blood pressure measurement    By clicking on the Amps     pressure values tab of the measurement window the maximum  amplitudes of the oscillations recorded during the blood pressure measuring phase appear  The same  function can be accessed by selecting the Amps     pressur
15.  patients  a particular patient can be found by using the search  function of the software    For a quick search click anywhere in the patient list and type in the initials of the family name of the  given patient  The quick search is not case sensitive    An advanced search can be done by name or date of birth with use of the    Search for    function   Select    Name        Date of Birth    or    ID    and type in the name  date or ID you want to find into the     Search for    field  This function is case sensitive  When searching by name  the search is made in  both the first and last names and even a part of the name can be searched for not necessarily  including the initial of the name     5 4 2 2 How to enter a new patient s data    N  If you want to enter a new patient into the database  select the New patient command in the         File menu or click on the New patient button  see on the left  on the Toolbar  For entering a  new patient only the date of birth  the height  the patient   s jugulum     symphysis  JUG SY  distance  and the upper arm circumference are required to be filled out  The jugulum     symphysis distance and  the upper arm circumference can be measured by means of the tape measure provided as an  accessory  see Section 5 3   The former data is necessary for calculating the pulse wave velocity while  the latter one is needed for calculating the proper cuff size  After filling out the arm circumference  field the suggested cuff type and its size wil
16.  sortic  75  28     i        3 Lower hmb circulation  jection dutathon       ABI ED  320 ms    200 mice  4  es tinn 19 20 M 0 M ag ee Mi ee ee    M  i 100  Apr  FE Arterial age  40 50 years       Sth percentile  Age  vari      gt 508  lt   90th percentile  Par Jeria     abih percentile      ion the basis of apprommat  ts 10000  Henne performed mm Cintra  Luroce  n pope rise          TeriiofiedT Software 5 1000 BETA  10 T10114 00115    TeniioMiedT Softwar v 3 0 0000 BETA  JON TLOLI4A 00115    Page 1     Page 2     Patient details  The data specified when entering the patient   s data are shown here including   a user defined parameter and its value provided it is set  Section 5 4 1   The space for   missing or undefined data will remain blank    Risk profile  The previously entered risk profile  see Section 5 4 2 2  appears here    Medication  The list of medications entered previously  see Section 5 4 2 2  appears here    Measurement data  This part contains the date and time of the examination  the name of   the operator  and the mandatory patient data    Suprasystolic record  Peripheral and central hemodynamic parameters are listed here    excluding pulse wave velocity  The calculated Ankle Brachial Index  ABI  is also represented   here  if this option was selected  see Section 5 4 1   A characteristic pulse wave is shown with   the points of the Return Time  RTS35  and Dicrotic notch  Ejection Duration  ED  indicated on   it  The return time does not necessarily fall to a
17.  vascular  age of the patient based on the arterial function assessment     3 3 First step     Register a user    If the program is used by several users with different groups of patients  it may be necessary that  users can access their own respective patients    data only after login  To this end a new user will have  to register in the database by specifying the name and login    Arteriograph SW can be used with the default login of    ARTERIOGRAM     it is  however highly  recommended to create and use your own user login     Registration of a new user    To register a new user  open the Options    window under the Tools menu and select the  Users tab  To enter a new user click on Add New and then enter the first and family names of  the user and specify a login name  user ID   Click Add to register the new user in the database  or click Cancel to interrupt the registration  The new user can log in with the login name  specified    Note  The registered users    data cannot be edited        planen                   System    Utert    Printed spart    Uiii rime    Fast name  Faih ame     Lagni mare  eses 101     User Ent       o CO                How to delete registered users    To delete registered users  select Options    under Tools and click on the Users tab in the  pop up window  The Delete button can be used for deleting a pre selected item from the  users list  In case the list contains only one user  the Delete function will not be available     4  Performing a sin
18. Arteriograph  User Manual    Medexpert Arteriograph    TL2  Medexpert Arteriograph Software                Arteriograph    Medexpert Arteriograph     Medexpert Arteriograph Software       Copyright    2011 Medexpert Ltd   Budapest  Hungary  All rights reserved  Under the copyright laws   this manual cannot be reproduced in any form without prior written permission of Medexpert Ltd     Every effort has been made to ensure that the information in this manual is accurate  Succeeding  models and manuals are subject to change without notice     Medexpert is not responsible for printing or clerical errors     This Manual is produced on the assumption that the operator is an experienced user of the Windows  2000   XP   Vista   7 Operating Systems    If the operator is not familiar with Windows operations  please refer to the online Help of Windows  or the Windows User Manual     Arteriograph is an unregistered trademark of TensioMed Ltd   Other company and product names mentioned herein may be trademarks of their respective  companies     Head office and service     Medexpert Ltd   T  rr  kk   u  5 7   1037 Budapest  HUNGARY    Hungary   Phone   36 1 250 4490   Fax   36 1 250 3795  Web  www medexpert hu    E mail  arterialcare medexpert hu    Contents    1    o A A E AES 7  1 1 e ee ee ee o 7  1 2 Contents Of the Mana rra 7  1 3 Genera lito Mal Olor ia rer 7  1 4 GUITAR TEE dd 7  1 5 Warnings and precautions                 cccccsssseccccsssessececcaessececseuseccecauuseceessuaecees
19. In the Backup field select Restore to start the restoration process  Also select the destination  directory for the backup file        r  Backup and restore database    Backup Place of backup file        Make E d   data          Restore DA  DUO   CCleaner   nokia                      Start    Cancel O 88 edexper       A confirmation window pops up for approving the restoration  Simply press No to cancel the process   To start the restoration  click on Yes           F1 Confirmation    a Are you sure you want to restore the database           Cyes J  no ET medexpert E    An information window appears when the restoration is successfully completed        r  EN Information    eo Data restore completed           5 4 8 3 Import previous database    If you have a previous version of Arteriograph database on your computer the Convert previous  database and import data command under Tools menu helps to convert the measurements into the  new database format    Select the patient list   pat  file in the Regisztratumok folder under the previous installation directory   by default  C  TensioClinic Regisztratumok  and click Open to start the import procedure     Note  Only a database in the original installation folder can be imported     Note  For a large database the conversion can take several minutes  The status of the procedure    is shown by a progress bar        Successful conversion is confirmed by a pop up window      Information a       A Data import has been successfuly done    
20. Writer  Status  Ready Print to file   Preferences    Location   CE   Find Printer       Page Range     Al Number of copies  1  11  2 2  3 3  ne   The print preview screen appears   El Pres mr      W 5     i kar             With use of the toolbar on this screen the following functions can be accessed     Zoom tool  J m The print preview page can be zoomed to fit the screen height  Zoom to fit   to fit  the screen width  Zoom to width  and to the original size  100     Navigation tool  This tool can be used to navigate to the First page  Previous page  Next    q  gt     ee  gt  en  page or to the Last page of the medical report   Search tool    Y With use of these tools a search can be performed by page number  Go to page  or by  a keyword  Search for text      Copy tool    The current page of the report can be copied to clipboard with this tool     Print tool  A printer can be selected an set up with use of the first icon  The page can be printed      with the previously determined settings by clicking on the second icon     Load   Save tool     a The Report can also be saved in a file by clicking on Save report  The destination     a  directory and the name of the saved file can be selected in a pop up dialog window  The    report can be saved as   e qrp file  This procedure generates a QuickReport file   qrp   which can be viewed only in the    Medexpert SW   e pdf file  This creates an Adobe Acrobat file  which can be opened without the use of the    Medexpert SW  This fea
21. an 35 mmHg    e Offset of diastolic pressure level determines the value which is added to the measured  diastolic pressure in order to obtain the pressure step where the Device will record diastolic  pulse wave data  volume curve   Zero  0  is the default setting  which means that data will be    collected at the pressure step corresponding to the measured diastolic value  Diagnostic  examinations should not deviate from this value    e Data transfer time at dia and sys pressure level determines the time span  in seconds  of  pulse wave data collection at the above specified pressure levels  The default value is 8 sec   Diagnostic examinations may deviate from this value in the 7 10 sec range    e Data interface  GDT  The GDT interface enables data transfer between medical devices and  general practice IT systems according to the standard produced by the German Quality  Assurance Software  QMS   A practice software system that supports GDT is required for the  communication  For setting up the GDT interface  please read and follow the GDT interface  manual located on the Arteriograph installation CD    e Calculation of Ankle Brachial Index  if selected  Ankle Brachial Index is calculated  For this  the ankle systolic blood pressure has to be measured before the pulse wave recording  see  Section 5 4 3     e Communication port  A default communication port number can be defined here  This way  the user can avoid the more time consuming process of automatic port search  The pro
22. analysis    21  5 1 Hemodynamic parameters measured by the Arteriograph                                                        21  5 2 SYSTEN SET U EEE 22  5 3 Patient prepara O IN ne aa 22  5 4 Performing the measurement sise 24   5 4 1 Device and measurement setup                                            sise 24   S542    Selecta MVE ccena ee eg 26   5 4 2 1 How to select an existing patient                                                             26  5 4 2 1 1 How to modify a patient s data                                                            26  5 4 2 1 2 How to search for a patient                                                   ss 26   5 4 2 2 How to enter a new patient s data                                                             27   5 4 3 Medical Examination with Medexpert Arteriograph                                                    een 28   5 4 4 Quality control of the measurement ss 30   5 4 4 1 Quality control index  the SD PWVao                          sise 30   5 4 4 2 Quality control by visual inspection                                                          30   5 4 5 HOw to analyse the data ss 33   5 4 5 1 Tools for visual inspection                                                            ss 36   5 4 5 2 Oscillation amplitudes during blood pressure measurement                                             37   5 4 5 3 Vascular age assessment ss 38   5 4 6 The measurement Report u    ra 38   5 4 6 1 Edit the Report case IA 39   5 4 6 2
23. asystolic data  and then on  button D  for diastolic data  after the results have been displayed  Most of the hemodynamic  parameters described above  Section 5 1  will appear in the results window  under the list of  examinations  During the automatic analysis the algorithm will analyse pulse waves of several  heartbeats and calculate the mean of the results as well as the standard deviation of the PWVao  SD  PWVao   However if there is only one evaluable pulse wave  no SD PWVao can be calculated     For manual evaluation you have to go to the end of the whole measurement sequence by using the  horizontal scroll bar at the bottom of the graphic area   The pulse wave curves recorded at the  suprasystolic pressure step are located at the end of the measurement sequence   You can use the  mouse for selecting arbitrarily any  or several  pulse wave recorded at the diastolic and suprasystolic  pressure steps  Curves can be selected even from those recorded in the blood pressure measuring  phase  but only that were recorded above the measured systole or in a  10 mmHg range of the  measured diastole     To select a series of pulse waves  or a single one  highlight  hold down the left mouse button while  moving the mouse  those from the beginning to the end of the sequence  Only pulses  i e  heart  cycles  that are highlighted in their entirety are included in the analysis  see the picture below    e When several pulse waves are selected for manual analysis the results are displayed
24. at errors can occur due to patient movement causing artefacts  in the reading  In case the Device cannot measure blood pressure  e g  due to  motion   measuring will be disrupted     Er 32    The codes of the error flags indicated by the Device shall mean the following     1    measuring time exceeded    The Device could not measure the patient s blood pressure  within the pre set time   3    low battery    Measuring was disrupted because of low battery     31    cuff missing    The cuff is not connected to the Device     32    33    34    35    90 99    100    101    102    110    111  112  113    115       cuff clogged          air leakage          the cuff is not on the arm          measuring disrupted          Device error          Faulty measuring result          motion          no pulse detected          Faulty measuring result          Faulty measuring result          Faulty measuring result       The rubber hose of the cuff is blocked or something is choking  the hose     The cuff  or the Device  leaks     While connected to the Device  the cuff is not put on the  patient     For some reason  e g  the user presses the button  measuring is  disrupted     Blood pressure measuring failed for some Device error or the  battery is low     The calculated results cannot be considered as a real blood  pressure value or the patient is having arrhythmia     The measuring conditions  e g  the patient   s movement   interfere with the process of measuring     For some reason the D
25. by PC  To  turn it on the button must be pressed twice again     2 5 System  Bluetooth communication  setup    In case your PC does not have a built in Bluetooth  at least v2 0   install the supplied Bluetooth  communication adapter by reading and following the instructions of the enclosed Bluetooth  operation manual  If necessary  ask for the help of your system administrator    After successful installation a Bluetooth icon appears in the system tray of the PC indicating an active  Bluetooth service    For connecting Arteriograph to the PC see Section 5 2    Data transmission can only be initialized with the Arteriograph Software     3  The software    3 1 Medexpert Arteriograph Software installation and setup  3 1 1 Preliminary information about the software    The Medexpert Arteriograph Software  Arteriograph SW  is a Windows based software  The common  Windows mouse operations are to be used in the program  left single click  double click  drag  etc       Button and command names ending in     indicate that a dialog window will open upon clicking on  those items     Arteriograph SW allows you to perform arterial function measurement  to print and share the  measurement report and to modify the measurement settings     3 1 2 Minimum system requirements    e Pentium IV PC  256MB memory  2GB available HDD capacity  CDROM  1024 768 resolution   e Windows 2000 operating system with Service Pack 4  Windows XP with Service Pack 2   Windows Vista or Windows 7 installed     e Act
26. cation  is in progress     Successful data download will be confirmed on the PC     Note  The downloaded settings cannot be cancelled  They can be overwritten through re     programming        5 4 2 Select a patient  5 4 2 1 How to select an existing patient    E After logging in to the program use the Open command in the File menu or click on the Open  5 button  see on the left  on the toolbar for displaying the list of the patients that belong to the  user logged in  Find and select the patient you are interested in  by double clicking on the row which  contains the patient   s details or by highlighting the row with a single click and then click on OK            The selected patient   s fullname and date of birth are displayed in the Patient Details block under the  Toolbar  In addition  the patient   s previous examinations will also appear in a list  if there are any        FI TenuoMed    Arteriograph   You are logged im at ARTERIOGRAM       fie View Tools information      Des BRA  5 fas  Les GY medexpert fl  K ands step AD malen vw Y sn compression 2 v Darie     Preset 5 Atomatic ana 55 D LA Canu  E narnrustiorn N Hard re Teszt   Date   Time Fest nar Beteg Date of Beth 01 01 1975    5 4 2 1 1 How to modify a patient s data    Select the patient  as described above  whose data you wish to modify  by a single click in the  Patients    List  and press Edit  Click on OK to save the modified data     5 4 2 1 2 How to search for a patient    In case there is a great number of
27. dexpert Arteriograph  using the simple blood pressure measurement function  see Section 4   In  the latter case please use the appropriate cuff size  the same rules apply as for arm cuffs  see Section  4 1      Calculation of Ankle Brachial Index  ome Su       Systolic blood pressure measured on the ankle mmHg       Com lo    After the measurement is completed the automatically calculated blood pressure and arterial  function data appear in the Results field  see Section 3 2   while the recorded pulse wave curves are  displayed in the Data window  see Section 3 2      F TersioMed    Arteriograph   You are logged in as ARTERIOGRAM a ON  File View Took Infeemation   gt    E 17 4 SBS stant Ea EN med  H    dep AD waters   Y ges Bonner 2   Padre 0 Reset a Brahma nahm SJ D ES Curra    Far nme Test  Fest rere Boteg Date of Beth 01 01 1975       ae   51 2  5     11 7  5   270  ms   125  ms  A A j    53  cm     0 5  m s  m  e a 1m i    0 21  ma       28  7  mmHg     111 7  mmttg           Every examination is saved automatically  An already saved examination can be deleted by selecting  the row  which contains its date and time in the Examinations list and then pressing the Delete  button on the keyboard     5 4 4 Quality control of the measurement    Conditions  movement  arrhythmia  respiration  etc   disturbing the measuring process may distort  the shape of the recorded pulse wave  which may result in false hemodynamic parameter values   After the measurement is completed its qual
28. e values option under View menu    By using this feature the quality of blood pressure measurement can be evaluated visually           IF  TersioMed    Arteriograph   You are logged in as ARTERIOGRAM S    Das SG   fas  Cas UE meaoxport il  Y ges Conmgretnon 2 v   Eeamunnaorn PATIENT DETAILS Fady name Teszt  Date   Tem Fest name Beteg Date of Beth 01 01 1975  ETE RE   121   83   36   30             Y 7 Y 7 T Y Y Y Y T Y Y 7 Y T Y Y Y Y T Y 7 Y 7 T    0    0 100 150 20 20 30 prets   Pulte waren Ampt   piacnse valias   Fell age Ansmann       5 4 5 3 Vascular age assessment    By clicking on the Vascular age assessment tab of the measurement window the estimated vascular  age of the measured patient can be displayed  The same function can be accessed by selecting the  Vascular age assessment command under View menu    This tab shows the Age     PWVao relationship with smoothed reference percentile curves     Note  Percentile is a value below which a certain percentage of the known measurements fall   e g  the 75  percentile is the value below which 75  of the observations can be found  Age   specific percentile curves show certain percentiles in different ages     Note  The reference curves were derived from a healthy central   European population including  approx  10 000 subjects        The current measurement is indicated with a coloured dot in the graph  with the colour depending  on which range the measured PWVao value falls  The ranges are as follows   e Normal   green colo
29. ear on the third page of the  print out        Note  The report is assigned to a specific examination  and can be modified later  For other  measurements of the patient this medical advice will not be saved or displayed        To display institutional data on the printed Report  select Options under Tools  In the dialog box  choose the Printed Report tab where you can specify the following   e Doctor s identification data    e Department details  e Operator   s name    Department details along with the doctor   s name and registration number will appear in the footer  of the printed Report  These data are user specific and therefore will appear on the report of every  patient of the user    The Operator   s name will appear in the Study data section of the printed report  It is measurement  specific  therefore can be modified for each measurement  By default the Operator   s name is the  login name of the User  The Operator   s name for the default user  ARTERIOGRAM  cannot be  modified                       Con IS E m0     5 4 6 2 Print the Report     gt  You can print out a Report by choosing Print preview and print under the File menu or clicking     the Print preview and print button on the toolbar     A Print options dialog window appears where printer selection and printer settings can be made and  other common print options can be set  Click OK to continue         Print    1                WE   General    Select Printer    Add Printer    Fax  Eh Microsoft XPS Document 
30. ection    X axis step   Horizontal axis   The scale of the horizontal  time axis of the pulse wave curve as  displayed on the millimetre net can be set from 10 ms cm to  1 600 ms cm in the pre defined steps  This function can be accessed directly from the toolbar  X axis  step  or by selecting a value under Horizontal axis in the View menu       anis step  200 meter    Y axis compression   Vertical axis compression   Here the vertical  amplitude axis of the pulse wave curve  as  displayed on the millimetre net  can be set  The set value indicates  the vertical compression rate of the curve  In case the set value is 1  the curve will be displayed in full  size  This function can be accessed directly from the toolbar  Y axis step  or by selecting a value  under Vertical axis compression in the View menu     Y asi compression     Baseline shift   The pulse wave curve on the millimetre net can be  moved upwards or downwards relative to the zero line   The original position can be restored by clicking on the Reset button     Baseline  U   Reset      View pressure curve   To display the actual cuff pressure values  press the View pressure curve button on the     toolbar  When it is on  the baseline cuff pressure can be seen during the whole measurement  on the millimetre net           Cursor   EE Lasa  Use the Cursor function for reading the co ordinates of the oscillatory signal  pulse  i   waves  and the value of the baseline cuff pressure  if the View pressure curve function  is
31. ed field strength in the  location in which the TL2 device is used exceeds the applicable RF compliance level above  the TL2 device should be observed to verify    normal operation  If abnormal performance is observed  additional measures may be necessary  such as re orienting or relocating the  TL2 device         Over the frequency range 150 kHz to 80 MHz  field strengths should be less than 3 V m     Recommended separation distances between portable and mobile RF communications equipment and the  TL2 device    The TL2 device is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled  The customer or  the user of the TL2 device can help prevent electromagnetic interference by maintaining a minimum distance between portable and  mobile RF communications equipment  transmitters  and the TL2 device as recommended below  according to the maximum output  power of the communications equipment     Separation distance according to frequency of transmitter  Rated maximum output m    power of transmitter 150 kHz   80 MHz 80 MHz     800 MHz 800 MHz   2 5 GHz  d  3 5 3 vP d  3 5 3 vP d  7 3 vP    For transmitters rated at a maximum output power not listed above  the recommended separation distance d in metres  m  can be  estimated using the equation applicable to the frequency of the transmitter  where P is the maximum output power rating of the  transmitter in watts  W  according to the transmitter manufacturer    NOTE 1 At 80 MHz and 800 MHz
32. evice does not detect the cardiac pulses     The signals are insufficient for blood pressure calculation    Artefacts  arrhythmia      The calculated results cannot be considered as a real blood  pressure value   The most likely reason is noisy measurement  signals      The pulse rate cannot be calculated or cannot be considered as  a real value     7  Technical Specification    Power supply   Four AA long life alkaline batteries or rechargeable batteries  Protection against electric shock   Internal supply by batteries  Grade of protection against electric shock   BF type on patient s side  Display   Liquid Crystal Display  LCD   Data storage   EEPROM  Data transmission   Radio  Bluetooth  PC Interface   Bluetooth Communication Adapter  PC system requirements   Windows XP with Service Pack 2 or newer OS installed  Ambient temperature   10     40   C  Dimensions   116 x 94 x 47mm  Weight  incl  batteries    250g  Blood pressure measuring method   Oscillometric  Sampling frequency   200Hz  Blood pressure measuring limits   30     280mmHg  Static accuracy     3mmHg or  2  of the measured value  Stability  2 years   Measuring accuracy   Systolic   91  91   out of 99 comparative checks fall within 5 mmHg   For 33 patients out of 33  2 of 3 comparative checks fall within 5 mmHg   0 of 33 patients did not have a value within 5 mmHg after 3 measurements   Diastolic   97  98   out of 99 comparative checks fall within 5 mmHg   For 33 patients 33  2 of 3 comparative checks within 5 mmH
33. g   0 of 33 patients did not have a value within 5 mmHg after 3 measurements   Average deviation from auscultatory measurements  Korotkov sounds     systolic diastolic   0 4  0 3 mmHg  Standard deviation  systolic diastolic   2 7 2 3 mmHg  Pressure sensor   Piezo resistive  Inflation   Automatic motor driven pump  Safety   Maximum cuff pressure  280mmHg  Deflation   Automatic  stepwise    7 1 Electromagnetic compatibility       Electromagnetic emissions    The TL2 device is intended for use in the electromagnetic environment specified below  The customer or the user of the TL2 device  should assure that it is used in such an environment       ss Emissionstest   Compliance Electromagnetic environment   guidance    RF emissions The TL2 device uses RF energy only for its internal   CISPR 11 function  Therefore  its RF emissions are very low and  are not likely to cause any interference in nearby  electronic equipment    The TL2 device is suitable for use in all establishments    CISPR 11 including domestic establishments and those directly   Harmonic emissions Not applicable connected to the public low voltage power supply   IEC 61000 3 2 network that supplies buildings used for domestic    Voltage fluctuations   flicker emissions Not applicable purposes   IEC 61000 3 3    Electromagnetic immunity    The TL2 device is intended for use in the electromagnetic environment specified below The customer or the user of the TL2 device should  assure that it is used in such an environmen
34. g 120mmHg lt   40  lt   120 mmHg    Fixed initial inflation pressure  mmHg  160  Offset of diastolic pressure level    Date and time of the device 120mmHg lt   0  lt   120 mmHg         Downloading into the device Data transfer time at dia and sys pressure level          1  lt   8  lt  10    Set default   sec       Data Interface  GDT    GDT Directory   Browse       Local Name TSM_ARG2 Local Short Name TARG    Remote Name Remote Short Name       Calculation of Ankle Brachial Index Communication port  COM 28         User defined parameter       Print preview Program language Date format      Set   English    dd mmyyy v    o M                      In the pop up window the following parameters can be set in the System tab     e Initial inflation pressure is the value Arteriograph inflates the cuff to when starting a blood  pressure measurement  This can be set as a fixed or a relative value  In this latter case the  Device adds 50 mmHg to the mean arterial pressure  MAP  value measured previously and  inflates the cuff to this pressure  The relative value is the default setting    e Offset of systolic pressure level determines the value which is added to the measured  systolic pressure in order to obtain the suprasystolic pressure step where the Device will  record pulse wave data  35mmH  g is the default setting  which means that data will be  collected at a pressure level of the measured systolic value   35mmHg  For diagnostic  examinations this value should not be set lower th
35. gle blood pressure measurement    The Medexpert Arteriograph device can be used as a simple automatic blood pressure meter   For  this function no computer is needed      4 1 Patient preparation    Sitting position is recommended for standard blood pressure measurement  The patient should be  comfortably seated with the legs uncrossed and the back and arm supported  The cuff on the upper  arm should be at the level of the heart   Allow the patient to sit in a relaxed state for at least 5 minutes before the measurement  Ensure  optimal room temperature and eliminate excessive background noise   Choose a cuff of appropriate size by considering the followings    e The length of the inflatable bladder of the cuff should be at least 80  of upper arm   circumference    e The width of the inflatable bladder of the cuff should be at least 40  of upper arm length     Please read and follow the latest guidelines on blood pressure measurement      4 2 Performing the measurement  To operate Arteriograph first install the device  see Section 2 3   There is a single push button     Section 2 1  available for operating the Device  The measured values and the status of the Device  are shown on the LCD     e The Device is ready for operation if the current PC time D 09 39  appears on the LCD     By pressing the button shortly manual blood pressure measuring will start     Note  In this case the Device performs blood pressure measurement without arterial function        e g  PWVao  Aix  evalua
36. he end of the examination the Device transmits the blood  pressure values to the PC  as well  The SW displays the followings  systolic and diastolic blood  pressure values  heart rate  MAP  pulse pressure  and the entire registered pulse wave curve        T  aber   If you want to interrupt the measurement use the Abort command  An         Abort button appears automatically next to the PW button when starting a  measurement  and the same command can be found under the Tools menu  Please note that this  function only disrupts the data collection procedure of the software and does not affect the  operation of the Device  To interrupt a measurement on the device the push button of the  Medexpert Arteriograph must also be pressed once     If the Calculation of Ankle     Brachial Index was selected  See Section 5 4 1  a dialog window appears  after the suprasystolic pulse wave recording phase is completed  Enter the formerly measured ankle  systolic blood pressure here and press OK  The calculated value will appear under the blood pressure  values in the Results field  see Section 3 2   If calculation of ABI is not needed or the ankle blood  pressure could not be measured  press Cancel    Measurement of the ankle blood pressure is advised to be performed preceding the pulse wave  recording  For measuring systolic blood pressure on the ankle  use of Doppler technique is suggested   However  if it is not available blood pressure measurement can also be performed on the ankle with  Me
37. her running Windows  applications and click Next     O      Welcome to the Arteriograph  Setup Wizard   The ll otal Tre sah 3 00 0 bea gi pour  Beamter    hi ee al you ieee al ee een ee  Canin       Step 4 Select the components to be installed from the drop down list  The type of  installation can either be Full or Upgrade   When the preferred component is selected  click Next     de Setup   Arteriograph    Sele Lomporenida   Wech components rodea be rise T       Sapo the comporeris pou man io etal  char re pou de nol mini lo  patel  Check Med when pou ore ready bo comino       sac    Ness     cancel      Step 5 The following screen offers you to create an icon on the desktop  This task can be  deselected  Click Next to start the installation process     y     Select Additional Tasks  Wish addboral asie a be parform  d         Desg ds addon iska ou wold ce Setup In plom de inst sn Anenegraph   then bak Hat    Addon ora    J Crate    dedo icon      smak   nees     Cancel    Step 6 The status of the automatic installation procedure can be tracked on a progress bar   If you want to interrupt it  click Cancel   This procedure also includes the driver installation of the hardware key  see Section  3 1 4        8 Setup   Arteriograph  betaling  Messe wet whe Setup rails AROSA on your DOUE       Estrie fies  Crogan Mes Teno Med A  erograch  one di    Step 7 When the automatic installation procedure is finished the following screen appears   Click Finish to exit the Wizard and start 
38. hol consumption should be avoided    o The patient should not sleep during the measuring phase    o At least a ten minute period of physical and mental rest is required before the  examination    o Any circumstance  e g  noise  phone ringing  persons    movement in the examining room   that may disturb the patient   s calm must be avoided    o Should    White Coat Effect    be detected the measurement must be repeated several  times with a few minutes    intervals or the examination should be repeated in another  time     5 4 Performing the measurement    To start the Program double click the Arteriograph icon on the desktop or use the Windows Start  Menu to navigate to the program  Log in to the software with the previously specified user name   see Section 3 3    When started for the first time  the Program logs in the user with the default     ARTERIOGRAM    login name  On every login after that  however  the user name must be specified      In case a wrong or non existent user name is given  the Program will offer the default name           E  Warning         A Login is incorrect   Do you wish to enter with the standard login name     ca A                   5 4 1 Device and measurement setup  Before starting the measurement changing or checking setup parameters may be needed     For setting up the Device select Options    under the Tools menu        F  Options                         Initial inflation pressure Offset of systolic pressure level    Previous BP value   35mmH
39. ice and the PC is guaranteed  see Section 7 1 for details     e To ensure the optimal performance and precise operation the Device should be maintained  at the manufacturer   s service every second year    e The cuff and the Device itself can be cleaned with a wet cloth by making sure that no liquid  enter into the Device    e The manufacturer   s products must be handled  stored  packaged  protected and shipped in  compliance with the general quality requirements     N Attention  Federal Law restricts this device to use by or on the order of a physician     or other licensed practitioner  US only    u    1 4 Guarantee    Medexpert Ltd  offers a two year guarantee period for the Device  Any repair within or beyond the  guarantee period will be performed by Medexpert Ltd  at the company   s site as indicated above     1 5 Warnings and precautions    Patient safety    The device has an integrated safety mechanism  which prevents the cuff pressure to exceed  300 mmHg  If however the inflation continues above this value or the pressurization lasts too  long  unplug the tube of the cuff from the device and remove the cuff from the subject     Do not use the device on an arm  which is being injected with intravenous injection     Do not use On neonates     Device use    Only use with cuffs supplied by Medexpert Ltd  Use of cuffs supplied by a third party can lead  to erroneous measurement results     Confirm blood pressure measurement with auscultation when erroneous results are  s
40. inaon  h Fat rne Tesrt    Date   Ime Feu nara Reteg Date of Beth 01 01 1975    Pulse ware Ampt   precia valuas   Actecial aga accacemant       Arrhythmia    The oscillatory measurement method is highly affected by arrhythmia  In such cases the beat to   beat variance of the pulse wave forms renders the analysis inaccurate     How to recognize  the heart beats do not show regularity     How to avoid  this condition cannot be affected by the examiner  Manual evaluation  see Section  5 4 5  can help to find a regular sequence in the suprasystolic measurement phase     Y rocked  Arerogriah  You se logged in m ARTIRIRAM eee     Fie Wee Took    L   m   Des PRA   imm  Las HE mea    n d  ni ar i   AA    Bose a Rami    E E    Essais i    arai FRS    Dime  Te Fact rarer Cas of Eo Ol  1 re       65 2  5    46        i      296  ma  i It     13   ma    EE  cm  LR ne J in A cpt ee   ss  ms    Im   ms    56 h  mmHg     162 6  matig     Pubs por Anpa   pe a uc A eee    Tremor or mechanical vibration    Tremor introduces an unwanted high frequency noise into the measurement  This often leads to  unrecognizable and not evaluable waveforms     How to recognize  the measurement is noisy  a sawtooth like signal is superimposed to the pulse  waves     How to avoid  tremor cannot be affected by the examiner  Make sure  however that the patient  is isolated from external mechanical vibration  e g  do not place the patient   s arm onto a  vibrating machine  like a PC      D fensotted    artenograph 
41. ity should be checked to decide whether the result of  the subsequent analysis can be accepted   5 4 4 1 Quality control index  the SD PWVao   SD PWVao  see Section 5 1  is the standard deviation of the PWVao values calculated from the pulse  waves that could be analysed by the algorithm of the software  This variable shows how much the  PWVao data of each analysed pulse curve differ  therefore reflects on the identity of the recorded    pulse wave forms  Large deviation of the measured data shows some disturbance in the  measurement     Normal SD ve     JUG SY i   51  cm  54  If the SD PWVao is greater than or equal to O and less than or equal to 1 0 m s  ue HR  the measurement is considered to be of good quality  ST ea 2     ABI     1 08   PPao   54 3  mmHg   High SD   RT 36  If the SD PWVao  SD  is greater than 1 0 m s  the measurement is considered   1   to be of unacceptable quality and the results of the current analysis should be a i  rejected  The SD appears in red colour in the Results field to warn the User on        1m s  An  the quality problem  For advanced users the use of manual evaluation  see IA  Section 5 4 5  can help to find a good quality sequence in the suprasystolic   prac N   j     40 1  mmH   measurement phase  If manual evaluation is still not acceptable  the mm  measurement should be repeated   No SD AT 107   80  ms    JUG SY dy  If the automatic analysis calculated arterial function data but no SD PWVao      lem  then only one evaluable pulse curve
42. ive Bluetooth v2 0 port     Although the Program can start in a less powerful environment  in that case we cannot take  responsibility for its fast and reliable operation     3 1 3 Software Installation    For installing the Arteriograph SW on the computer the user must have administrator privileges   After the installation is completed  please visit our website  www medexpert hu  for the latest  software upgrades  Installation instructions for those can also be found on our site     Step 1 Insert the CD into the CD tray of your PC  The installation procedure will start  automatically  In case it does not  start the installation by double clicking the  setup exe file on the disc with use of a file manager program    Should a security warning pop up  simply press Run to start the installation    Open File   Security Warning  The publisher could not be verified  Are you sure you want to  run this software   CE Name  E  settclin3000b exe      Publisher  Unknown Publisher  Type  Application  From  E  settclin3000b exe    CAN   ca     run software from publishers you trust       This file does not have a valid digital signature that verifies its   X  publisher  You should only    How can   decide what software to run     Step 2 To start the Setup Wizard  choose a preferred language for installation and click OK     Select Setup Language x      Select the language to use during the installation       A          Step 3 The wizard starts with a welcome screen  Please close all the ot
43. l automatically appear next to it     Note  To achieve good measurement quality always use the cuff suggested by the software     The cholesterol unit can be selected from a drop down list  mmol l or mg dl     The Comment field is advised to be filled out with important information on the measuring  conditions  e g  the measuring position  pregnancy     Fill out the Risk profile field with information on the known cardiovascular risk factors  such as family  history  duration and frequency of smoking  etc        Cos        Address                            5 4 3 Medical Examination with Medexpert Arteriograph    MES START   To start a single measurement select PW read under Tools or press the START button    e    on the toolbar     W     To start a scheduled sequence of measurements click the Repeated Measurements      RE button on the toolbar  In the dialog window the length of the test and the measurement  frequency can be set  To start the measurements  click Start  The measurement process can be  interrupted by pressing the Stop button     Repeated measurements    Length of test  7   KEN Start      Frequency  5 min    3 Stop     7  Save results automatically    Number of remaining measurements  3     Cancel es medexpert       When starting a measurement a confirmation window will pop up with some important  measurement information  If the listed data are correct  press Yes to start the measurement process   If these data need to be modified  press No and edit the patient
44. listed subjects  Selection is  confirmed by clicking on the arrow           2  Step     Please select patientis   Select all  Tet  Beleg Palo of Bethy  C8 OM  145 Adder          Esport al manzuremaeia     oS        By setting a time range here  only those measurements that were performed in the  specified range will be exported  By selecting Export all measurements  every  measurement of the formerly selected patients will be exported  Click Export to start the  process     2081  juut    HE fx E    l 3 3   i 7 A 16  W i gt  16 N   H M F    E    Saa  3  Step        amp  5 E   Flense smi 1 M    1106 20  J  Fram AMM Ey Ta nu    Export  Export all messurements    et NE mec Y             The status of the export process is shown by a progress bar  In case of a large database    the process can take several minutes   Export data          Export patients data       Export measurements          When the export is completed a message window will pop up confirming on the    successful save   SS Information les     eo File saved successfully     IE 8 000                     The save destination of the completed text file can be selected  The exported file is a tab delimited  text file  which can easily be opened or imported in common spreadsheet applications  e g  Microsoft  Excel  or statistical packages  e g  SPSS  SAS      Note  The exported text file contains each measurement in a different row  Several  measurements for one patient are listed under each other  If the result were no
45. m  8 UN   aes al      8 5  m s    ATH        ms      78 6  mmt ig    ER l   Cursor 4 Clove   111 6  mmHg    Pulte wee Ampt  Geecouee vales   Actor age atoetemert    o When selecting only one pulse wave at the pressure step corresponding to the    diastolic value several curves appear in the window  the original pulse wave     represented by a red curve   the systolic component  green  and the diastolic wave    component  black   The curve shows two markers indicating     from left to right     the    locations of the peak of the systolic wave and the start of the diastolic wave              r TA  F TerioMed   Arterograph   You are lagged in as ARTERIOGRAM   a  file View Tools Information    Dala ACA E  Sas   La BM medoxpor  I  K moat tep ZU  mam   Y aes Companion  2   arre Reset 8 Astonia arabe 5 35 D ii Caves  vor Fady name Teszt  Feit name Beteg Date of Beth 01 01 1975   attexstart  121   300  ms    RA   66 393   N ENI pad AR A         u   j     Al NL J Zu     33 8  X    N     l   A  gt  e     66 2         een   e  gt  l    Cu 4 Chow  Pula w Text peca o valves   Asiana age accecemert    With use of the tools in the separate window the single pulse wave can further be examined   The curve can be enlarged and reduced in size within pre set limits  see the relevant value of this in  the upper left hand corner of the window  Furthermore  you can examine the curves from point to    point by means of a cursor that can be activated     The window can be closed separately    Charac
46. mination is performed while the patient is seated   indicate this fact in the    Comment    field of the SW when the patient   s data are entered     e Place the cuff in a way that the hose is in anterior position    e To avoid the irritation of the skin a long sleeved shirt made of some thin material can be  worn under the cuff    e Connect the hose of the cuff to the air connector of the Device  For this  push the plug on the  hose into the air connector of the device and twist the plug slightly until it securely clicks   Make sure that the cuff is properly attached by checking the tightness of the connection    e Measure the suprasternal notch     pubic bone distance  Jug Sy  with use of the supplied tape  measure     Note  The suprasternal notch  jugulum      pubic bone  symphysis  distance should be measured in  a Straight line  Measuring on the body surface can lead to overestimation  e g  in case of obese  patients         e Ensure optimal measuring conditions    o Speaking  muscle movements  especially the movement of arm muscles must be avoided  as these activities significantly increase the measuring time  may result in measurement  failure or may reduce the precision rate thereof  Muscle movements  tremor  distort the  form of the measured pulse wave and render its evaluation impossible    o Within 3 hours before the examination the patient should not consume a greater  amount of food  drink coffee or smoke  Additionally within 10 hours before the  examination alco
47. nd mobile RF communications  equipment should be used no closer to any part of  the TL2 device  including cables  than the  recommended separation distance calculated from  the equation applicable to the frequency of the  transmitter     Recommended separation distance    Conducted RF 3 Ver d 1 2VP  IEC 61000 4 6 150 kHz     80 MHz    Radiated RF 3 V m d 1 2VP 80 MHz     800 MHz  IEC 61000 4 3 80 MHz   2 5 GHz d 2 3VP 800 MHz     2 5 GHz    where P is the maximum output power rating  of the transmitter in watts  W  according to  the transmitter manufacturer and d is the  recommended separation distance in metres   m     Field strengths from fixed RF transmitters  as  determined by an electromagnetic site  survey  should be less than the compliance  level in each frequency range       Interference may occur in the vicinity of  equipment marked with the following symbol     W    NOTE 1 At 80 MHz and 800 MHz  the higher frequency range applies    NOTE 2 These guidelines do not apply in all situations  Electromagnetic propagation is affected by absorption and reflection from  structures  objects and people      Field strengths from fixed transmitters  such as base stations for radio  cellular cordless  telephones and land mobile radios  amateur  radio  AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy  To assess the electromagnetic  environment due to fixed RF transmitters  an electromagnetic site survey should be considered  If the measur
48. per  port number can be checked under Control Panel   gt  Bluetooth devices  If this field is left  blank  the software performs an automatic port search when starting the measurement  This  process may take up to one minute    e User defined parameter  One new parameter can be added to the patient   s data set through  this function  Enter the name of the parameter into the input field  the parameter will be  displayed in the Patient details field on the print out  see 5 4 6 2  and press Set     Note  A new parameter can be added to a patient   s data only once and cannot be modified  or deleted later  The new parameter wil automatically be added to all the patients in the    database  After setting it into either a new patient   s data or an existing patient   s data  the  patient characteristics will feature this newly specified parameter  as well        Non Device specific settings     e Program language  the language of the Arteriograph SW can be selected  The change only  takes place upon the next SW start    e Date format  one of the following date formats can be selected  Year month day  day   month year  and month day year     Default settings can be restored by clicking on Default Settings  The settings must be downloaded  into the Device to take effect  For this use the Downloading into the Device button     The download of the settings to the Device can be tracked on the screen  The CONKT  text    CONNECT    will be displayed on the LCD of the Device while communi
49. program window appears with the following components   For a better understanding the  picture below shows the program window after a completed measurement      Menu bar   Toolbar   Examinations window  Results field   Patient details field  Data window    DAS e SS    x E  TersioMed    Arteriograph   You are logged in as ARTERIOGRAM    File SS Tools Information  z  en    200 mub   Y ass compresi  n 2   Eee u Rebel    6 Autom  anak 5 35    Tear  Beteg    Fary repre  Fast name    E PATIENT    Data    Tine Date of Bath    Soil  ie  83  1 25  ms  E N  53  cm       E i    85  o s  lt Le Ma     0 21  ms     ee 7  mmHg     1 1 7  mmHg   Pubs wave   dings   postie values   Aulenal age acperimeni    3 2 1 The Menu of the Program    01 01 1975    lor     Des Bea S Bee el CB medexpont fl    D    BE Cura     lt        5    The menu bar of the software can be found under the title bar and contains the menu items through    which all the major functions of the program can be accessed     The File Menu    The basic functions of Arteriograph SW can be accessed through the File  menu  You can enter new patients  open  edit or delete existing    database records  save the result of an analysis  furthermore edit  print  and send medical reports  as well as log out and exit     The View Menu    The View Menu enables you to choose to display the pulse waves  recorded or the maximum amplitudes of the oscillatory signals at each  cuff pressure step  deflated stepwise with 10 mmHg  during blood  pre
50. ransmission inhibiting objects  e g  metal objects  can cause signal  weakening and consequently malfunction in the communication  The use of the  supplied protective pouch does not affect the transmission signal strength     When an active Bluetooth device  e g  a PC  tries to add Arteriograph the sign   shown on the right appears on the LCD  No actual data transmission takes place B uel INK  yet  Adding Arteriograph is only possible by entering its passkey  see above     Data transfer can only be initiated with use of the Arteriograph SW     5 3 Patient preparation    Wrap the cuff around the dominant upper arm  When doing this procedure  please consider the  followings     e Use the cuff suggested by the software  see Section 5 4 2    e Make sure that the cuff is tightened properly  without causing harm to the patient  The cuff  should fit the patient   s arm evenly  Tightness is sufficient if your finger cannot or can only  hardly be inserted under the cuff    e Make sure that the patient   s skin is not pinched by the cuff     Note  If the cuff is too loose or big  the SW does not provide measurement results and gives an  error message     Small amplitudes          The cuff should not contact the patient   s chest during the measurement because the respiratory  movements may cause pressure alterations and artefacts in the cuff        According to the guidelines for measuring arterial function routinely the supine position is  recommended for measurements  In case the exa
51. se heat because it may cause a short circuit  Regarding both  rechargeable and alkaline batteries  the policies of use and waste collection as  governed by detailed environmental and labour safety requirements are to be    considered      amp  LR6 1 5V  AA         LR6 1 5V  AA  EL  LES  LR6 1 5v  AA      O LR61 5V  AA   amp L    Figure 3    34   43 cm  26  33 cm  18  25 cm    Once the batteries are inserted in the Device  the following checks are performed automatically     e Voltage control of the batteries  The measured value appears on the 50V  LCD for a few seconds  The supply voltage is sufficient if the  measured value is between 6 0 V and 5 4 V for alkaline batteries and  between 4 8 V and 4 6 V for rechargeable batteries     e  f the voltage drops below 4 4V  the batteries must be replaced  A LOwbot t    warning symbol of low battery appears on the LCD     e The Device is ready for operation if the current PC time appears on D  09  39  the LCD     2 4 Operation instructions    There is a single push button available for operating the Device  Section 2 1   The patient can use the  only push button for issuing a total of two commands     1  By pressing the button shortly manual blood pressure measuring will start  Note  In this case  the Device performs blood pressure measurement without any arterial function  e g  PWVao   Aix  evaluation     2  The Device can be turned off by pressing the button twice  In OFF  FF  state the device cannot be operated either manually or 
52. seuseceessuuseeeessugaeeeessagaeees 8   The Medexpert Arteriograph device                                                        9  2 1 The device at a 2 Ace  een on aa 9  2 2 The accessories of the Device sise 10  2 3 installing the Device  ae ee da eee ee eee eee 10  2 4 Operation instructions    ar NAS 11  2 5 System  Bluetooth communication  setup                                                     11   TNS OT AE ee ee euere 12  3 1 Medexpert Arteriograph Software installation and Setup                    oocccccocccnnonccncnnnnccnnnnacononacnonnnos 12   3 1 1 Preliminary information about the software                  cccccooocccncconocccnncncnnnconnnonanocnnnonononnnnonanononos 12   3 1 2 Minimum system requirements    ss 12   3 1 3 Software Installation near 13   SIE AVS TU A KV a ee ee het 15  3 2 The software at a elder innatas 15   32 1 The Menu of the Program  een 16   3 2 2 The Toolbar of the Program    ss 17   3 2 3 The examinations window                                                    ss 17   3 2 4 The results e  o  a nn  a ea ine o OOO nee  17   3 2 5 Patient details Held assim een 17   3 2 6 The measurement data window                                               ss 17  3 3 First step   Register a user seen rer ER 18   Performing a single blood pressure measurement                           nennen 19  4 1 Panent preparato De ee Ne a een 19  4 2 Performing the measurement sisi 19  4 3 Acoustic signals ann un een an 20   Performing pulse wave recording and 
53. ssure measurement  Furthermore you can set the scale both  horizontally and vertically  hide the Toolbar  and get information about  the currently studied patient            TemiohMird    Arterisgraph You are legged in a ARTERIOGRAM      Eile   Wiw Toch Information  New patient   Open  Sivi nls  Eda report  Print preview and prat    Send POF report    Exit    Ciria M  Cirle    Dile Fi    a im  a  3 Oe START x  pompei E  B    ls Fanta rare    Fit nan            F TensioMed    Artenograph   You are logged in as ARTERIOGRAM         File  View Jools Information  Y    Pulse wave    Amps   pressure values    X ai    Arterial age assessment    Exar  Honor tal aus    Date      Y Toolbar    Patient info       4    Vertical aus compression    Ctre Fl ES START    K   Ct   F2  Cents 3 we 2 B    a F arniky nam  Fasi name    7    The Tools Menu       FA TersioMed    Arteriograph   You are logged in as ARTERIOGRAM    The Tools Menu enables you to start automatic pulse wave data  collection by means of the Arteriograph Device  as well as interrupt an  ongoing data collection process  Previous versions of databases can be  converted and imported into the new database format  furthermore  exports and backups of the actual database can be created  All the  Program and Device settings can also be accessed and modified through  this Menu     Pir Nica  Tools   Imlermatinn    1   PIN read a   N        oda php x  Convert previous database and import data    Erama  Dae Tims Expert data     7         
54. t       f Electromagnetic environment    Immunity test IEC 60601 test level Compliance level    Electrostatic discharge  ESD    6 KV contact   6 kV contact Floors should be wood  concrete or ceramic   IEC 61000 4 2   8 KV air  8 kV air tile  If floors are covered with synthetic  material  the relative humidity should be at  least 30       feces A   ee    IEC 61000 4 4   1 kV for input output lines  IEC 61000 4 5   2 kV line s  to earth    Voltage dips  short interruptions    lt 5  U  Not applicable  and voltage variations on power     gt 95  dip in Ur     supply input lines for 0 5 cycle  IEC 61000 4 11    40  Ur   60  dip in Ur   for 5 cycles    70  Ur   30  dip in Uy   for 25 cycles     lt 5  Ur    gt 95  dip in Ur   for5s  Power frequency If image distortion occurs  it may be   50 60 Hz  necessary to position the TL2 device further  magnetic field from sources of power frequency magnetic  IEC 61000 4 8 fields or to install magnetic shielding  The  power frequency magnetic field should be  measured in the intended installation  location to assure that it is sufficiently low        NOTE  Uy is the AC mains voltage prior to application of the test level     Electromagnetic immunity    The TL2 device is intended for use in the electromagnetic environment specified below  The customer or the user of the TL2 device  should assure that it is used in such an environment     Immunity test IEC 60601 test level Compliance level   Electromagnetic environment     guidance    Portable a
55. t saved for a    given measurement the exported text file will only contain the patient characteristics and the  blood pressure measurement data in that row        5 4 8 Backup and restore the database    To protect your measurements against accidental data loss  the database should regularly be backed  up  For this use the Backup command under Tools menu   The Backup and restore dialog window appears  where you can set whether to make or to restore a  database  The destination directory can also be selected     5 4 8 1 Backup data    In the Backup field select Create to start the backup process  Also select the destination directory for  the backup file        FE  Backup and restore database       Backup Place of backup file  O  Make  E c   packard bell      gt   Restore  gt  C         Advancedinstallers  ar SA   bg BG      catroot   Ca catroot2   com   config   cs CZ   da DK   de DE   Dism a    emre MEE                         In case there is a previously saved backup file in the selected database  a confirmation window pops  up for overwriting the file  If you don   t want to overwrite it  press No and rename the previous  backup file or choose another destination directory  If you approve overwriting  simply click on Yes           EAN Confirmation    Q Backup file already exists  Do you want to overwrite it           An information window appears if the backup is successfully completed           EAN Information    eo Backup completed              5 4 8 2 Restore data    
56. teristic points of the pulse wave curve are marked in this window  Blank blue circles represent  the first and reflected wave peaks and the ejection duration  Dicrotic notch   respectively  Filled red  dot represents the calculated return time  The return time is calculated as the time difference  between specific characteristic points of the curve  therefore it does not necessarily fall to a  characteristic point itself    2 1      time  200  ms     amplitude  113 5          Note  Results of the evaluation are not saved automatically  For saving the results into the  database the Save results    command should be used  This allows the examiner to save the  results of only those measurements that are of good enough quality to be exported later        E The findings of the evaluation can be saved into the database by selecting Save results    in the    File menu or by clicking the Save results    button on the toolbar  Any previous result will be  overwritten           E Confirmation            2  Results already exist  Overwrite              ve Co ecto er                 This even allows saving results different from the ones calculated by the automatic evaluation    When an examination is opened it appears with the formerly saved arterial stiffness results  If no  result is saved  the software displays the results of the automatic evaluation  However  in this latter  case no values will be exported with the Export command  see Section 5 4 7      5 4 5 1 Tools for visual insp
57. the Arteriograph Software     ko  El    Completing the Arteriograph  Setup Wizard    Stop hae fined inataling A  erogreh en your camgar The  apps pay ba launched by ears Pa eed ale ona    Click Finish to ext Setup    Ji Launch Asa         Firah      3 1 4 The hardware key    For copy protection purposes  the software is supplied with a hardware key  Before starting the  software and during its run the hardware key must be plugged in a USB port of your computer  As  the software installation wizard installs the driver of the hardware key automatically  the hardware  key is ready for use     Note  Without the hardware key the software will not start  and an error message will appear     Attention  Do not remove the hardware key while the software is running  as it  causes the software to freeze out      gt     3 2 The software at a glance    To start the Program double click the Arteriograph icon on your desktop  or use the  Windows Start Menu to navigate to the program in the Program Files        A splash screen appears along with a login dialog window   When started for the first time  simply press OK and the Program logs in the user with the default     ARTERIOGRAM    login name  On every login after that  however  the user name must be specified     Arteriograph  for professional users          Y       Login Dialog    Enter pour login name       OK I Cancel Ea CA             www medexpert hu    Manufactured by TensioMed Kft   ud K    r str 2 d 1103 Budapest  Hungary    The 
58. timation of the patient s arterial age can be seen on the SD PWVao  parameter graph  which is a simplified visualization of the reference percentile values for the  patient   s age  It shows the patient s position relative to the reference population  The  measured value is indicated in the graph  also showing the estimated arterial age at the top     Page 3  optional      Medical Report  Printed only if the Edit report function was used  see Section 5 4 6 1   This  page includes only the patient details and the edited medical report     Note  The Report is always printed out in A4 size  It will contain only as much information as was  entered previously or was calculated in the course of the analyses  which also means that a  Report can be printed out with only the patient   s data appearing     Note  Cut off  critical  values for the different parameters can be found in the available scientific  literature addressing vascular age and arterial function        5 4 6 3 How to send a Report in PDF format    You can convert the above detailed Report to PDF format by Send PDF Report under the File menu  and simply e mail it  After the command is selected  the program converts the Report to PDF format  automatically  The generated  pdf file can be saved to a selected destination with the following file  name  Tcrep_fullname_date_time pdf  where fullname is the name of the patient and date and time  are the date and time of the examination in the pre set date and time format  e g  
59. tion     By doing so  the current time will disappear from the display and then     e The LCD test pattern is displayed  REX A Be Dae Be   e The supply voltage is checked 56V   e The Device calibrates itself by setting the zero pressure CAL O  The measurement starts with the inflation of the cuff  This procedure A B I    is indicated by an arrowhead pointing upwards     During deflation an arrowhead pointing downwards indicates the  procedure     When measurement is completed  the Device displays the systolic and  diastolic values     and then the heart rate     The measuring process can be disrupted any time by a single press on  the button  The text    OFF    will appear and remain visible for about 5  seconds on the display  Thereafter the time is displayed again  which  indicates the ready to operate status for another manual or  programmed measurement     4 3 Acoustic signals    v 69    LG Y  PUL 68    OF    When the Device is active  a gentle acoustic signal is emitted when its button is pressed down     5  Performing pulse wave recording and analysis    Pulse wave registration can only be initialized by Arteriograph SW that is running on a PC with  wireless Bluetooth data transmission link to the Device  in this case the button of the device should  not be used     The oscillatory signals recorded during the whole measuring process are displayed real time on the  PC screen     5 1 Hemodynamic parameters measured by the Arteriograph  The following hemodynamic parameters
60. ture is equivalent to the Send PDF Report function  see Section  5 4 6 3     The patient   s previously saved Reports can be selected and displayed by clicking on Load Report in  the Print preview dialog box     Close    Close The Print preview screen can be closed by using this button     The Report printed out by the Arteriograph Software will contain the following     TS ann F   Blood Pressure Measurement rn expe sri Pulse Wowe Analysis  expert Y expert  ON mede tel Arteriogra ph report ond Pulse Wowe Analysis Y mede i if Arteriograph report ond Arterial Age Assessment  ME Atend  Name Postal code 1035 Country code  HU Name  ID 007  ich oy City Hud  pert Date of Birth  194905409  Date of Path  1495 05 09 Addrots Age  61 years Gender Male     eel bho ana epg  Gender  Male E mail srtenograph  medenpent hu  Weight  BMA 110 ig  31 8 kg m  Quality control  Risk profile   50  0 14 mis  farnily history of CVD A    er TAE  J be       ner     nnn MN  BE  Medication BC Pere o 10  mans Pulse Wave Velocity measurement  Pury BO ms AT 147 ms  Measurement data PWV   Arterial Age Assesoment  Date  2011 04 01 10 05 Haight  lB  cm Armin  Jiem Aight Piso  Operator ARTERIOGRAK iag  Si SSom Cuff sive 2 m  4  Suprarystole record    ia    a AT EH rachael Dad Pronto    Central Hemodynamics    sl and Pulse Wave Analy  n  Sy  122 mmHg SAPO  111 2 mmHg  Dia  81 metia  PP  41 mmHg Pac  30 2 mmHg  j MAP  95 menda    ken 63   min  HA Aix brachial 555  Aix sortie 9 6     T  Abs brachial  75     B22   Aix
61. ur  The measured value is lower than or equal to the 50    percentile for  that specific age group   e Elevated   yellow colour  The measured value is higher than the 50 percentile but lower  than or equal to the 90  percentile for that specific age group   e High    red colour  The measured value is higher than the 90   percentile for that specific age    group        File View Tools Information  ec rg    Pa ES   sm  LES EA mode     mes Comet    y    r al  FI TercioMed    Arteriograph   You are lagged in as ARTERIOGRAM   a    F andy rare Teszt  Feit name Beteg Date of Bet 01 01 1975       PWVao    PA Im   a a      4 5  gt  y e  gt        o Mth percentle       50 and  lt   Sth percentile    ES r  on the hace of paana 10000 remarca paianed in Cardin ropas popisi hd o Uh percentile       5 4 6 The measurement Report    The Report is a printed summary of the measurement results  It includes patient details  all the  calculated hemodynamic parameters  characteristic pressure curves and vascular age assessment     The following functions are only enabled if an examination of a patient has been selected     5 4 6 1 Edit the Report    TA To edit the Report  press the Edit report    button on the toolbar or select the Edit report   command under the File menu  In the report window you can enter your medical advice    treatment suggestion  which can be saved by clicking on Save  The current date can be inserted to  the report by using the Insert date button  The edited Report will app
62. uspected     Do not use a microwave device  e g  mobile phone  near the unit during measurement   Do not use the device when it is exposed to mechanical vibration  e g  in vehicles      Prevent the device to be exposed to direct sunlight  to get in contact with liquid or from  excessive mechanical impact     Do not disassemble the device        2  The Medexpert Arteriograph device    This section will give an introduction on the device along with installation information and operation  instructions     2 1 The device at a glance  Figure 1 shows the front view of the device     The name of the Device   Air connector   LCD   Push button for entering two commands to control the Device  as per Section 2 4  Symbols standing for the above two commands    a ES       Arteriograph    Vascular age assessment     is  BLUELIN    SYS mmHg DIA mmHg PULSE min    BP 10  ON OFF 2x0    B           Figure 1  Figure 2 shows the device label  The meanings of the symbols are the following     6  Name of the Device   7  Name of the Manufacturer   8  Range of ambient operating temperature   9  Manufacturing serial number   10  Applicable nominal range of supply voltage with the batteries   11  Year of manufacturing   12  Safety classification of electric shock protection  Classification  Patient side  BF   13  Warning on the importance of reading this User Manual thoroughly   14  Certification mark as a guarantee of the Device   s conformity with EU requirements  15  Type ID of the Device           
    
Download Pdf Manuals
 
 
    
Related Search
    
Related Contents
  4 - ElectricalPartManuals.com  TEFAL FF102730 Instruction Manual  1. Garantie, responsabilité  VORLAGE HOCHFORMAT  取扱説明書 - MyPOSレジスターVer.5(パソコンレジソフト)  MANUEL D`INSTALLATION ET D`UTILISATION  Manual  Midway 16-20044-101 Video Game Furniture User Manual  USER MANUAL    Copyright © All rights reserved. 
   Failed to retrieve file