1. No Rinse Hexane Methanol Name Company Telephone 603 893 3663 METHANOL Dispense sec Saturate sec Soak sec Drain 0 sec Soak Elute ___ Sec _ sec ___ Sec _ sec ___ Sec _ sec ___ Sec _ sec ___ Sec __ sec ___ Sec _ sec ___ Sec _ sec ___ Sec _ sec ___ Sec ___ sec sec sec Phone Telefax 603 893 4994 62 Appendix B Sample Bottle Measuring Scale Place the opening of the sample bottle over the measuring scale to determine the proper size Bottle Cap Adapter Appendix C Parts List SPE DEX Oil amp Grease Machine Fully Automated The following items are necessary to run the SPE DEX Oil amp Grease Machine 1 Water Sample Inlet Valves 3 for 1000XL and 3 for 3000XL 2 Liquid Handling Kit 3 Safety Coated Solvent and Water Recovery Assembly including a 5 gallon wastewater recovery bottle 4 10 of Solvent Exhaust Hose 5 The Choice of 2 SPE Disk Holder Assemblies 47 mm or 90 mm Part Number Description SPE Disk Holder Assemblies The Disk Holder Assembly allows the use of the SPE disk Prefilters and FilterAid 400 50 0808 Handles 47 mm disks for SPE DEX 3000XL Series Extractors 50 0810 Handles 90 mm disks for SPE DEX 3000XL Series Extractors 165 00JT Handles 50 mm JT Baker Speedisks for SPE DEX 3000XL Extractors Bottle Cap Adapters for SPE DEX 3000XL Series Extractors The Water Sample Inlet Valves for the SPE DEX 3000 Series Extractors are designed to
2. Use the checklist below to ensure that all of the Extractor components have been received Check the Controller kit for the following items C SPE DEX 3000 Controller P N 50 0412 L 12 Volt AC Power Supply and Power Cord L User s Guide Check the Extractor kit for the following items L SPE DEX 3000 XL Extractor or SPE DEX 1000XL Extractor L Water Sample Inlet Valves three 3 for 3000XL System or one 1 for 1000XL System P N 50 0471 L Disk Holder Assemblies 47 mm and or 90 mm JT Baker Speedisk Adapter LI Sensor Arm Clips various sizes if using the 50 mm JT Baker Speedisk C 2 5 Liter Safety Coated Solvent Bottles total of four 4 for 3000XL System or 2 for 1000XL System P N 27 0042 C Solvent Delivery Kit P N 50 0486 01 L Water and Solvent Waste Lines P N 50 0485 L Black Air Line P N 50 0484 L 10 of Solvent Vapor Exhaust Hose longer lengths are available P N 14 0404 _ 5 gallon Safety Coated Water Waste Recovery Bottle P N 180 0005 01 L Water Waste Recovery Assembly P N 50 0051 03 C Solvent Recovery Vacuum Assembly P N 50 0043 03 L Vacuum Supply Line Assembly P N 50 0211 02 C Collection Vessel Retaining Clip s Blue P N 22 0687 C 19 22 Adapter for 40 ml VOA Vessel P N 160 0001 L 125 ml 19 22 Taper Erlenmeyer Flask P N 27 0476 The following items are also required to complete the installation Supplied by the customer C A gas supply dry Nz capable of minimum 60 psi maximum 80 psi
3. LC A vacuum source capable of 24 26 Hg 10 Hg at 1 CFM PN 50 0583 L Cap Adapters contact Horizon Technology for the proper cap adapter to use C Additional 19 22 Collection Vessel with 40 ml vials or 125 ml size Erlenmeyer flask L Solvents to perform the proper chemistry 5L n Hexane 85 purity or greater 99 0 min sat C6 isomers residue less than 1 mg L 5L Methanol ACS grade residue less than 1 mg L L SPE extraction disks 47 50 90 mm size Prefilters and 3M Filter Aid 400 are also recommended for handling dirty samples Accessories LI Dry Trap for the Vacuum Source not supplied P N 50 0572 Rev B Pg 13 4 2 Installing the Oil amp Grease Machines The SPE DEX Oil amp Grease Machines are equipped with an internal solvent vapor exhaust fan permitting bench top operation Ten feet of solvent vapor exhaust hose is the standard length shipped with the system but lengths of 15 and 20 feet are available Therefore position the Extractor System close to a suitable vent A workspace of at least five 5 feet wide will be required to accommodate the System along with safety coated Solvent Delivery Bottles safety coated Liquid Recovery Bottles both solvent and water a compressed gas source dry nitrogen and a vacuum source Step 1 Place the Extractor unit in the area where it will be used Position the unit at an angle to allow easy access to the rear connection fittings from the left side Step
4. it may be necessary to press the CLEAR key several times Alternatively the sensor hours and number of samples processed can be reset by disconnecting the sensor lead to the circuit board connector inside extractor and then using the A key to CLEAR the memory AUXILIARY 2 The initial screen will verify the operating configuration of the Extractor i e a fully automated extractor is connected to the Controller The LCD will display Aux Summary Screen Rinse Type Auto Sample Type Auto Rev B Pg 38 At the bottom of the screen the letter function keys will display the testing functions that are available The information listed below provides a detailed explanation of each function WARNING Read this section carefully before initiating any test as several operations are irreversible and can erase vital system information making the Controller non functional DIAG A key DEFAULT B key UPLOAD E key Runs a complete diagnostics on the System Reloads the Factory Preset Methods Uploads the System software into the Controller This function will allow the latest version of software code to be uploaded into the Controller A PC with a serial port and Null Modem cable are required If this function is to be performed see Section 7 9 or contact Horizon Technology Inc Press the DIAG A key to access the Test Screen The following function keys will be displayed EEPROM A key RELAY B key This function will test
5. methods 24 and 25 have a 10 second Air Dry Time after the methanol prewets The second methanol rinse has been removed from these methods e The Factory Preset method times methods 21 through 24 and 26 through 30 for dispensing the hexane and methanol prewets have been reduced by approximately 2 seconds e To avoid inadvertently erasing the code which would cause the controller to become inoperable a lengthy sequence of keys must be pressed before the operating code can be erased The commands are displayed on the screen and only by following these steps can the code be erased e New sets of counters have been added to the Aux 1 screen The Aux 1 screen will display the total number of hours that the liquid sensors have been operated plus a counter will display the total number of samples that have been processed Warning In order for the new software to be uploaded the current operation software is erased from the EEPROM If this occurs and you do not have a copy of the original or the latest software contact Horizon Technology Inc 1 Determine the current version software running in the Controller Press the Auxiliary 2 key followed by pressing the E key UPLOAD The current version will be displayed If a newer version is available obtain the latest software version from Horizon Technology The new software will have 2 files the Upload exe file and the Superc bin file Create a new directory on the PC named Up
6. to the extractor and use the blue retaining clip to hold it in place When using a method designed for a 90 mm disk a 125 ml collection vessel is used to collect the extract A 40 ml VOA will not be sufficient to hold the greater volume of rinse solvents used Rev B Pg 30 Loading a Full Sample Bottle The SPE DEX Oil amp Grease Machines are designed for use with a wide variety of 1L sample bottles Several sizes of cap adapters are available that allow the various sample bottles to fit the Water Sample Inlet Valve The standard Water Sample Inlet Valve is designed for bottles with a 33 x 400 mm thread Refer to Appendix B for the sample bottle measuring scale and Appendix C for the parts list for the cap adapters available The following steps detail the method for loading a full sample bottle onto the Extractor Step 1 Lower the Water Sample Holder Arm Step 2 Fill a sample bottle with approximately 250 to 500 ml of water A full liter can be used but it will take longer to process Step 3 Take the Water Sample Inlet Valve and visually ensure the ball valve is closed and that the bottle gasket is in place Carefully thread the Water Sample Inlet Valve onto the sample bottle Tighten the valve only with moderate force Over tightening could create leaks if the valve body is distorted or the threads stripped If using a larger bottle firmly attach the appropriate size cap adapter to the bottle and then attach the Water Sample In
7. 3 Hexane Rinse 2 sec 2 sec 2 sec 2 sec 2 sec 5 sec 5 sec 5 sec 5 sec Soak 5 sec 5 sec 5 sec 30 sec 1 min 5 sec 5 sec 5 sec 5 sec Elute 0 sec 0 sec 0 sec 30 sec 1 min 0 sec 0 sec 0 sec 0 sec CODE Rin 4 Hexane Rinse 2 sec 2 sec 2 sec 2 sec 2 sec 5 sec 5 sec 5 sec 5 sec 3M 3M Soak 30 sec 1 min 1 min 30 sec 5 sec 5 sec 1 min 30 sec 45 sec 1 min Elute 30 sec 1 min 1 min 30 sec 0 sec 0 sec 1 min 1 min 1 min 1 min W Whatman Rin 5 Hexane Rinse 2 sec 2 sec 2 sec 2 sec 2 sec 5 sec 5 sec 5 sec 5 sec CPI CPI Soak 5 sec 5 sec 5 sec 30 sec 1 min 5 sec 5 sec 5 sec 5 sec Elute 0 sec 0 sec 0 sec 30 sec 1 min 0 sec 0 sec 0 sec 0 sec PF prefilter Rin 6 Hexane Rinse 2 sec 2 sec 2 sec 4 sec 4 sec 5 sec 5 sec 5 sec 5 sec FA Filter Aid400 Soak 30 sec 1 min 1 min 30 sec 1 min 1 min 1 min 30 sec 45 sec 1 min Elute 30 sec 1 min 1 min 30 sec 1 min 2 min 2 min 1 min 1 min 1 min Rin 7 Hexane Rinse 4 sec 4 sec 4 sec 8 sec 8 sec 8 sec 8 sec Soak 1 min 1 min 2 min 1 min 1 min 1 min 1 min Elute 1 min 2 min 4 min 5 min 3 min 4 min 5 min Ly Lt Ip A AI H V N i 9619 S TECHNOLOGY 8 Commerce Drive Atkinson NH 03811 Method Worksheet Method number LCD Screen Page 1 HEXANE Dispense sec Saturate sec Soak sec Drain sec AIR DRY min LCD Screen Page 2 Step Type Rinse sec Sec Sec Sec Sec Sec Sec Sec Sec 0 sec S NN O O1 N The choices for the Step Type of solvent rinse are as follows
8. 50 mm 20 box 50 017 HT CPI Nu Phase SPE Disks Oil amp Grease Disks 47mm 20 box 50 018 HT CPI Nu Phase SPE Disks Oil amp Grease Disks 90mm 10 box Accessories 30 0544 Solvent Rod o rings 50 0471 Water Sample Inlet Valves 14 0404 10 10 of Solvent Exhaust Duct Hose 14 0404 15 15 of Solvent Exhaust Duct Hose 14 0404 20 20 of Solvent Exhaust Duct Hose 30 0452 01 O ring for J T Baker Adapter 1000 3000XL 02 0792 01 47 mm gasket for 3M disks used with 47mm Disk Holders 02 0793 90 mm gaskets for 3M disks used with 90 mm Disk Holders SpeedVAP Solvent Evaporation System 200 1000 SpeedVAP II 9000 Solvent Evaporation System 01 0284 Locator Rack for 9 70mm aluminum pans Used with 47 mm Extracts 01 0284 01 Locator Rack for 5 105mm aluminum pans Used with 90 mm Extracts 50 002 HT 70mm Oil and Grease Aluminum Weighing Pans 100 box 50 002 02 HT 105mm Oil and Grease Aluminum Weighing Pans 100 box 50 001 HT Oil and Grease Transfer Pipettes 500 box 50 0572 Dry Trap Solvent Trap Solvent Recovery System 200 2000 Solvent Trap Solvent Recovery System 65 Standards and Supplies 40 001 HT 40 003 HT 40 007 HT 40 010 HT 50 003 HT 3M FilterAid 400 1 5 Kg bottle Whatman GMF 150 47mm Prefilters 2 micron 40 box Whatman BS2000 42 5mm Prefilters 32 box Whatman 90mm Prefilters 16 box Oil amp Grease Standard Kit 4 x 30 ml vials 8 mg ml Vacuum Requirements 50 0583 50 0572
9. A Factory Programmed Methods And Method Worksheet Method 21 Method 22 Method 23 Method 24 Method 25 Method 26 Method 27 Method 28 Method 29 Method 30 disk manufacturer Pacific 3M Pacific W Pacific W Baker BakerFA Pacific 3M Pacific W Pacific W Pacific W CPI Purge CPI PF CPI PR FA CPI CPI PF PF FA size mm 47mm 47mm 47 mm 50 mm 50 mm 90 mm 90 mm 90 mm 90 mm no disk PW Hexane Dispense 6 sec 6 sec 6 sec 4 sec 4 sec 12 sec 10 sec 10 sec 10 sec 1 sec Saturate 1 sec 1 sec 1 sec 1 sec 1 sec 2 sec 1 sec 1 sec 1 sec 0 sec Soak 1 min 1 min 1 min 1 min 1 min 1 min 1 min 1 min 1 min 0 sec Drain 1 min 1 min 1 min 1 min 1 min 1 min 1 min 1 min 1 min 5 sec PW Methanol Dispense 6 sec 6 sec 6 sec 4 sec 4 sec 12 sec 10 sec 10 sec 10 sec 1 sec Saturate 1 sec 1 sec 1 sec 1 sec 1 sec 1 sec 1 sec 1 sec 1 sec 0 sec Soak 1 min 1 min 1 min 1 min 1 min 1 min 1 min 1 min 1 min 0 sec Drain 0 sec 0 sec 0 sec 10sec 10sec 0 sec 0 sec 0 sec 0 sec 5 sec Air Dry 5 min 8 min 8 min 5 min 8 min 8 min 5 min 8 min 8 min 0 sec Rin 1 Methanol Rinse 2 sec 2 sec 2 sec 4 sec 4 sec 5 sec 5 sec 5 sec 5 sec hexane 3 sec Soak 5 sec 5 sec 5 sec 30 sec 30 sec 5 sec 5 sec 5 sec 5 sec 0 sec Elute 0 sec 0 sec 0 sec 45 sec 1 min 0 sec 0 sec 0 sec 0 sec 5 sec Rin 2 Methanol Rinse 2 sec 2 sec 2 sec 2 sec 5 sec 5 sec 5 sec 5 sec methanol 3 sec Soak 30 sec 30 sec 30 sec 5 sec 1 min 30 sec 45 sec 1 min 0 sec Elute 45 sec 1 min 1 min 30 sec 0 sec 2 min 1 min 1 min 1 min 15 sec Rin
10. A series of rinse steps are applied to the SPE disk to achieve high and consistent Oil amp Grease recoveries Rinse Solvent Dispense Time The rinse solvent is dispensed for the duration of time specified in the method program Solvent dispense times depend on the type of disk prefilter and disk size used The volume of rinse solvent dispensed is dependent on the solvent bottle pressure and solvent dispense time Rinse Solvent Soak Time After the sample bottle has been rinsed the solvents are introduced to the SPE disk and allowed to soak The program allows for a variable soaking time for each rinse solvent introduced to the SPE disk The solvent soak time can be adjusted to maximize the extraction efficiency and optimize recoveries Rinse Solvent Air Dry Time During the Air Dry of the Rinse Steps the extraction solvent is pulled into the collection vessel along with the Oil amp Grease Additional Features The SPE DEX 3000 Controller has additional features that allow for flexibility and increased ease of use Purge Key The Purge Cycle is used predominantly for three reasons 1 To remove any air in the solvent lines during the initial system installation 2 To confirm that during the initial system installation that all of the solvent solenoid valves are functioning properly 3 Verify system operation on a regular basis There is a distinct and important difference between a Purge Cycle and a normal method
11. Assembly warning label and the label identifying the bottle on the Waste Recovery Bottle Locate the blue Water Waste line attached to the Extractor from Step 8 If desired cut the line to the necessary length Using one of the barb fittings located in the original bag firmly press the barb fitting into the end of the line Once the barb fitting is firmly secured attach the barb fitting to one of the three lower elbow fittings on the safety coated Water Waste Recovery Assembly Refer to Figure 4 7 As with the Solvent line first counter rotate the tubing several times before making the connection This will assure the tubing is firmly attached to the elbow fitting Based on the location of the vacuum pump source position Figure 4 7 Water Waste both the safety coated Solvent and Water Waste Recovery Recovery Assembly Bottles between the vacuum pump source and the Extractor unit Rev B Pg 16 Step16 Step 17 Step 18 Locate the package labeled P N 50 0211 02 This package contains a short piece of smaller O D tubing with fittings attached Gently connect one end of the tubing to the very top elbow fitting on the Solvent Recovery Assembly mounted on the Solvent Recovery Bottle Figure 4 6 Using a t fitting connect the other end of the tubing to the very top elbow fitting on the Water Waste Recovery Assembly Refer to Figure 4 7 From the same bag as Step 16 P N 50 0211 02 locate the larger O D tubing Gently conne
12. B Pg 7 3 0 Oil amp Grease Machine Overview 3 1 SPE DEX 3000 Controller Figures 3 1 and 3 2 identify the major components of the Controller unit e LCD Display The LCD displays all System parameters and operating status e Keypad The user interface for method programming and system inputs The letter keys change functionality based on the programming level e Power Switch Turns the System On Off e Power Input Connects the Controller to a 120 VAC power source via the 12V supply e Comm Port Allows software uploads via flash memory e Extractor Cable Connects the Controller to the Extractor unit LCD Display Keypad Figure 3 1 Front of Controller Power Switch Extractor Cable Comm Port Power Input Figure 3 2 Back of Controller Rev B Pg 8 SPE DEX 3000 Controller Keypad STATUS LEARN METHODS AUXILIARY 1 AUXILIARY 2 A E Allows viewing of the current operational state of the three extractor stations simultaneously The Status page will indicate the method currently loaded into each station and the operating status of each station This key is also used to access control over each extractor station Allows the operator to teach each station optimal SPE conditions in real time Once a method is learned by the Controller it is stored in memory and can be used on any extractor station Displays the methods section of the Controller memory Methods are stored in a non volatile
13. Check for crimped or pinched solvent lines that could prevent solvent flow Also verify all solvent line connections Check for liquid leaks Remove the back panel of the extractor and verify that the pressure regulator on the right when facing back of unit is set at 15 psi Rev B Pg 54 4 Prewet solvent overflows Verify the dispense time programmed for the controller method With the unit sitting idle and solvent bottles pressured does solvent leak from the Prewet Arm This would indicate a stuck open prewet valve For a stuck open prewet valve program a method with only the prewet solvent that is leaking and run the method several times to flush the valve The solvent bottle pressure can be increased incrementally by adjusting the pressure gauge inside the extractor Do not raise the pressure above 30 psi Always use the solvent filters provided with the system at the end of the lines in the solvent delivery bottles This will extend the life of the valves refer to Section 4 2 5 Prewet solvents do not drain Verify that the main vacuum source is on and the gauge set at 24 Hg to 26 Hg Verify that the manifold plugs on the Solvent Waste and Water Waste Recovery Bottles are securely sealed and not losing vacuum Make sure the fittings on the manifold plugs are tight and that all lines are connected to the Waste Bottles are secure Examine the solvent to waste line for any crimps or damage that may impede the
14. Enter a Soak time for the second Methanol rinse A short soak time of 5 seconds is recommended The maximum soak time that can be entered is 10 minutes Press the A STEP key to advance to the next line Enter the Elute time for the second rinse step This time is based on the SPE disk used and if a prefilter is used This should be sufficient to pull all of the Methanol and residual water off of the disk and into the collection vessel Select Hexane as the third rinse solvent Input a Rinse time of 2 seconds for a 47 or 50 mm SPE disk or 5 seconds if using a 90 mm SPE disk The maximum rinse time that can be entered is 10 seconds Press the A STEP key to advance to the next line A Soak time for the n Hexane follows Enter a soak time for the disk The maximum soak time that can be entered is 10 minutes Rev B Pg 37 Step 36 Press the A STEP key to advance to the next line Step 37 The Elute time is the next step Enter a time based on the SPE disk used Step 38 Continue this process of entering Rinse Soak and Elute times for all the rinses to be programmed Note The n Hexane rinse steps will remove or extract the Oil and Grease from the disk and pull it into the collection vessel It is recommended to make the final Elute time the longest to ensure that all of the solvent is pulled into the collection vessel Step 39 Save the method created by selecting the E Save key The method can now be downloaded to the Sta
15. Remove the internal components of the assembly Note the order of the components and the orientation of the seat Clean the components and verify that the spring is in good condition 9 Reassemble the Check Valve and insert it back into the Adapter Taper of the extractor Make sure to align the small vacuum opening on the Check Valve Assembly with the vacuum opening on the Adapter Taper If the vacuum openings are not correctly aligned no vacuum will be supplied to the collection vessel and therefore the solvent rinses will not elute 10 Test the Check Valve by pouring some water into the Disk Holder Cup and observe if it leaks the from the bottom luer If the leak persists replace the Check Valve with a new one Rev B Pg 53 8 0 Troubleshooting The SPE DEX Oil amp Grease Machines are engineered to perform in a demanding environment This system is a mixture of electrical mechanical and pneumatic subsystems Unfamiliarity with the operation of the system especially with new system start ups can be intimidating Before calling for assistance attempt to rectify any issues by referring to the trouble shooting tips below This will assist a Horizon Technology technical representative in pinpointing the problem quickly No matter what the problem is remember that opening the cover of either the Controller or the Extractors voids the warranty unless instructed to do so by a Horizon Technology technical representative Product quality
16. acts of nature accident misuse abuse negligence or modifications of or to any part of the Product This warranty does not cover damage due to improper operation or maintenance or attempted repair of any electrical components No other warranties are expressed or implied For service assistance or resolution of a service problem or for product information or operation call Horizon Technology Inc 1 800 997 2997 1 603 893 3663 Or write to Horizon Technology Inc 45 Northwestern Drive Salem NH 03079 EMAIL spe horizontechinc com Website www horizontechinc com 67
17. disks having a greater surface area will typically have greater flow rates than 47 mm disks Sample introduction rate is dependent upon the filtration rate As more water is drained to waste using vacuum additional sample flows into the Disk Holder Assembly The Liquid Sensor on the prewet arm detects the presence or absence of the water sample At the start of the sample processing step the Sample Inlet Valve opens to deliver the sample into the Disk Holder Assembly There is a 20 seconds delay to equilibrate the sensor before the water to waste valve opens to drain the sample to the waste When there is sample in the Disk Holder Assembly the sensor remains submerged cool and the vacuum continues pulling the sample through the SPE disk and out to the Waste Water Bottle When the water sample is completed the sensor is no longer submerged and the sensor begins to heat up If the liquid sensor is not cooled by additional sample the run will advance to the air dry step Air Dry Step This step removes water from the SPE disk The disk is Air Dried for the number of minutes specified in the method In this process the vacuum physically draws air from the laboratory through the disk The actual duration of this process depends on a number of factors e g the disk being used if a prefilter is used and how clean or dirty the sample is Step 3 Solvent Rinse of the Sample Bottle This step uses two solvents to rinse the sample bottl
18. fit a 33 x 400mm Boston Round Bottle If a larger sample bottle will be used select the bottle cap adapter that will fit your sample bottles See the Sample Bottle Measuring Scale to determine the sample bottles size presently being used For custom bottle adapters contact Horizon Technology Note the ICHEM 33 x 430 Bottle is not compatible with the Water Inlet Valve of the SPE DEX 3000XL 150 0005 01 70 x 400 adapter for 1 L jar 150 0006 01 89 x 400 adapter for 1 L jar 150 0009 01 53 x 400 adapter for 1 L wide mouth bottle 150 0010 01 58 x 400 adapter for 1 L French Square 150 001 2 01 48 x 400 adapter 150 001 4 01 63 x 400 adapter 150 0015 01 83 x 400 adapter 150 0025 33 x 430 adapter 64 Part Number Description Collection Vessels 02 0216 5Teflon Washer Vial Adapter 5 pkg 160 0003 19 22 to 24 40 Adapter for Erlenmeyer Flask 160 0001 19 22 adapter for 40 mI VOA vessel 160 0008 40 mi VOA vial 72 cs 27 0476 125 ml Erlenmeyer Flask 27 0834 Separatory Funnel 19 22 125 ml SPE Disks 1664 47 HT Pacific Brand 47 mm 1664 90 HT Pacific Brand 90 mm 40 005 HT Whatman SPE Disk 47 mm DFP for Oil amp Grease 12 box 40 006 HT Whatman SPE Disk 47 mm DFP for Oil amp Grease 48 box 40 009 HT Whatman SPE Disk 90 mm DFP for Oil amp Grease 47 005 HT 3M Empore Disks Oil amp Grease 47mm Disks 20 box 47 0008 HT 3M Empore Disks Oil amp Grease 90mm Disks 10 box 50 019 HT JT Baker Speedisk Oil amp Grease Disks
19. flow 6 Continuous drain or leak of solvent to the waste Verify the drain time programmed in the method if using a method other than Methods 21 through 30 With unit sitting idle and vacuum on fill the platform opening for the Disk Holder Assembly with water Does the water drain and to which waste bottle solvent waste or water waste Flush the stuck open drain valve with water using either the DRAIN and SOLVENT functions or the DRAIN and SAMPLE functions of the controller 7 Water Inlet Valve does not open to dispense sample Verify that the main gas source is on and set at between 60 80 psi Manually turn the aluminum shaft to open and close the Water Inlet Valve If it feels tight and is difficult to turn the o ring for the aluminum shaft may be damaged or the gaskets on the ball valve need replacement The four hex screws at the bottom of the valve housing can also be loosened Do not loosen too much as this can result in sample overflows refer to Section 7 3 Verify that the Water Sample Holder Arm onto which the sample bottle and Inlet Valve rest is not touching the metal actuator key and preventing it from turning The Holder Arm may drop over time with the weight of the bottle Adjust the hinge arm so that it makes a 90 degree angle with the sheet metal by adjusting the bottom set screws refer to Section 7 7 Rev B Pg 55 The metal rectangular actuator key onto which the shaft of the inlet valve is placed
20. for 15 seconds and pull the solvent into the collection vessel Follow the instructions below for running a Purge method Step 1 Press the STATUS key until the status of all three stations is displayed Refer to Figure 5 7 fA B C D TE Step 3 Press the E DRAIN key Refer to Figure 5 8 for th Drain Function Screen Drain Menus Step 4 Press the D PURGE key Please Select Drain Option Step 5 The LCD will display Warning Process Starting Ready to start If Station 1 is ready to start press A YES The purge method will begin on Station 1 Repeat Steps 1 5 for all stations Purge each station a minimum of 3 times to ensure that all air has been removed from the lines and that the system has been properly Figure 5 8 Drain Function Screen installed Until experience is gained with the system it is suggested to purge one station at a time to understand and confirm proper operation As more experience and confidence is gained all three stations can be purged simultaneously While the purge method is running verify that both prewet solvents Hexane and Methanol are delivered into the Disk Holder Cup Hexane will dispense from the first or upper tubing on the Liquid Sensor Prewet Arm and Methanol will dispense from the second or lower tubing The rinses both Hexane and Methanol will spray into the empty bottle Carefully observe the solvent rinse spray The spray should hit the center of the base
21. in place Center the Exhaust Hose on the Exhaust Port and thread the hose on Extend the length of the hose to the appropriate hood or vent Locate the Solvent Recovery Vacuum Assembly 50 0043 03 and a 2 5L safety coated Solvent Recovery Bottle Inspect the inside of the bottle cap to ensure the Teflon washer is securely in place Screw the assembly onto the bottle and place the bottle on the right side of the extractor Place the warning label along with the label identifying the bottle on the Solvent Recovery Bottle This bottle can also be placed on the floor below the extractor Locate the clear Solvent Waste line attached to the Extractor from Step 8 If desired cut the line to the necessary length Using one of the barb fittings located in the original bag firmly press the barb fitting into the end of the line Once the barb fitting is firmly secured attach the barb fitting to one of the three lower elbow fittings on the Solvent Recovery Vacuum Assembly Figure 4 6 Counter rotate the tubing several times before making the connection such that as the tubing returns to its relaxed state the tubing will rotate clockwise assuring a secure fit Locate the safety coated 5 gallon Water Waste Bottle P N 180 0005 01 and place the bottle next to the safety coated Solvent Recovery Waste Bottle Locate the Water Waste Recovery Assembly 50 0051 03 and Figure 4 6 Solvent Waste place on the opening of the S gallon bottle Place the Recovery
22. memory Methods 1 20 can be modified for optimal conditions Method 21 30 are factory preset and cannot be modified Displays the total number of hours the liquid sensors have been heated and the total number of water samples processed on each station Allows access to diagnostics default parameters and uploading new software revisions Special function keys with changeable functions Functions will change based on the programming level being accessed The LCD displays the functionality of each key Figure 3 3 Controller Keypad Rev B Pg 9 3 2 SPE DEX Oil amp Grease Machines Figures 3 4 and 3 7 identify the major components of the Oil amp Grease Extractor Systems e Sample Bottle e Water Sample Inlet Valve e Water Sample Holder Arm e Liquid Sensor Prewet Arm e Disk Holder Assembly e Disk Holder Platform e Collection Vessel e Controller Cable Port The original sample bottle is used on the extractor system Bottle sizes from 33 x 400 mm to 89 x 400 mm openings can be used The Assembly that opens to allow the water sample to be dispensed into the Disk Holder Assembly Holds the water sample bottle and sample inlet valve in place during the extraction process Houses the liquid sensor and the two Prewet dispensing lines Holds the SPE disks and required prefilters filter aid Available in two sizes 47 mm and 90 mm diameter Holds the SPE Disk Holder Assembly Collects the solv
23. operator to troubleshoot the System and ensure proper operation Oil amp Grease samples can range from very clean to very dirty the preset methods were programmed with this in mind Method 30 is a Purge method New methods can be created to handle unique samples and or optimize the extraction parameters 6 1 Creating a Method Using the LEARN Mode The easiest way to understand how to create new methods is to use the LEARN Mode The LEARN Mode is a simple and fast way to create an analytical extraction method by observing the volume of solvent being delivered The times calculated during the LEARN Mode are then automatically entered and stored into that method number This procedure allows optimal methods to be created based on specific disks and or unusual sample conditions that might be encountered The LEARN Mode must be used with an empty sample bottle Follow the steps below to run the LEARN Mode If desired the LEARN Mode can be canceled at any time during the LEARN sequence Step 1 Turn the controller on Step 2 Load a Disk Holder Assembly with disk onto an extractor station Follow the instructions in Section 5 5 for loading an SPE disk Step 3 Verify that the ball valve of the Water Sample Inlet Valve is closed Carefully thread the Inlet Valve onto an empty sample bottle and load onto the extractor Water Sample Holder Arm Step 4 Load acollection vessel onto the bottom taper of the extractor giving a lifting and twistin
24. run When the PURGE key is pressed all Soak times will be automatically set to 0 seconds and the Sample Processing and Air Dry time will be skipped Therefore as only the Prewet and Rinse solvents will be dispensed it is very important to remember that no water sample is to be used when the PURGE key is pressed only an empty sample bottle to contain the Rinse solvents and a collection vessel to collect the rinses Drain Keys The Controller drain keys allow the user to drain liquids from the Disk Holder Assembly independent from running a method with the press of a key There are three drain functions Sample Solvent and Elute Selecting the Drain key followed by the Sample key will drain any liquid in the Disk Holder Assembly to the safety coated Water Waste Bottle Selecting the Drain key followed by the Solvent key will drain any liquid in the Disk Holder Assembly to the safety coated Solvent Waste Bottle And selecting the Drain key followed by the Elute key will elute any liquid in the Disk Holder Assembly into the collection vessel Rev B Pg 22 Pause and Resume Key The Pause key allows the user to interrupt the method on any one station without interrupting the method on the other two stations The Resume key starts the method from where it was interrupted Counter The Controller has the capability of counting the total number of samples processed on each station and the total number of hours that the liquid sensors ar
25. spring 7 3 Cleaning the Water Sample Inlet Valve Assembly The Water Sample Inlet Valve P N 50 0471 Figure 7 3 is the delivery mechanism by which the sample is dispensed onto the Solid Phase Extraction SPE disk An air actuator automatically controls the opening and closing of the ball valve design The Water Sample Inlet Valve should be cleaned under warm tap water after every sample Occasionally it may be necessary to clean the Water Figure 7 3 Water Sample Inlet Valve more thoroughly for example after a sample with high Sample Inlet Valve particulates or if leaks are observed through the Water Inlet Valve when in its closed position Step 1 Remove the four 4 screws from the bottom of the valve housing using a 9 64 Allen wrench Figure 7 4 gt Step 2 Separate the two halves of the housings and remove the Ball Valve Assembly Step 3 Remove the two 2 sealing gaskets PN 02 0680 on the top and bottom of the Ball Valve core The entire assembly can now be cleaned using tap water Figure 7 5 Warning Do Not Use Acetone to clean these parts The use of Acetone will deteriorate these parts and void all warranties c que T A RAS tie water Inlet Valve Step 4 Reinstall the two gaskets making sure the notched side of the gasket faces the ball valve If the gaskets look worn or if sample leaking is a problem replace PES os sha 010 with new gaskets Step 5 Place the ball valve assembly into the top
26. the Controller EEPROM chip This test will also erase all information for Methods 1 20 Only the Factory Preset Methods will remain Press YES A key to begin the test or press NO E key to return to the main Diagnostic screen This function will test the System Relay Drivers At the completion of this test the LCD will turn off Press the E key to turn the LCD back on CAUTION This test will activate all of the solvent valves Therefore either turn off the pressurized gas and vent the solvent bottles or make sure all receptacles i e sample bottles Disk Holder Assemblies and Collection Vessels are in place before running this test A2D C key This function will display all of the A2D values for the sensors The A2D 0 line displays the output signal from the sensor for Station 1 The A2D 1 line displays the output signal from the sensor for Station 2 And the A2D 2 line displays the output signal from the sensor for Station 3 The remaining lines are for future outputs Figure 6 3 Figure 6 3 A2D Screen Rev B Pg 39 KEYPAD D key FACTORY A key This function allows testing of the keypad Once this test has been initiated the keys must be pressed in the order requested by the LCD before being able to exit from this test The keys pressed are displayed and confirmed on the LCD This function provides a convenient way of clearing all Method times and solvent parameters from Methods 1 20 All ti
27. the specific chemical for handling and safety instructions MSDS Material Safety Data Sheets are the source for chemical hazard information including basic information on the manufacturer or distributor identification of the chemical the product s hazardous ingredients physical data fire and explosion data toxicity information protection information and more The laboratory is responsible for having a MSDS for every chemical or substance being used It is also the laboratory s responsibility to make the MSDS available and accessible to all employees and to provide training in the safe handling of hazardous chemicals MSDS can be obtained from the vendor All hazardous solvents and chemicals must be disposed in accordance with appropriate Federal State and local regulations The SPE DEX Oil amp Grease Machines require the use of a pressure source to deliver solvents to the system and for the operation of the Water Inlet Valve The recommended source is nitrogen gas The operator must be aware of the potential hazards of reactivity and toxicity as well as asphyxiation of even the referred to harmless gases such as nitrogen The large amount of potential energy resulting from compression of the gas makes the cylinder a potential rocket or fragmentation bomb if dropped To prevent this secure the gas cylinder to avoid tipping If you have any questions please call Horizon Technology at 800 997 2997 Inside NH 603 893 3663 Rev
28. through the the instructions in Steps 5 1 to 5 3 below for the procedure on how to free the check valve is the Sample Collect Check Valve Figure 5 1 provided with the System P N 02 0725 gently insert the needle end straight down into the opening Gently tap the head several times to free the Check Valve Ball This will move the Check Valve Ball off of the seat assembly An internal spring will gently push the ball back into the sealing position Figure 5 1 Check Valve Opening Check Valve Assembly could permanently damage the check valve ball surface Figure 5 2 Check Valve Step 6 Step 7 Step 8 Release Tool The Disk Holder Assembly is made up of three parts the Base or Cup the Collar and the Aluminum Locking Ring see Figure 5 3 Load an empty 47 mm Disk Holder Cup no SPE disk or Collar no Aluminum ring into each Extractor station Firmly press the Disk Holder Cup down to the lowest position into the center hole on the Extractor s Disk Holder Platform Verify that Collar Locking this is a snug fit A loose fitting Disk Holder Cup will Ring result in poor vacuum which will affect the performance of the system Place the screen in the cup Figure 5 3 Disk Holder Assembly Lift the Liquid Sensor Prewet Arm up and lower it into the Disk Holder Cup Move the Sensor Arm to the backside of the Cup See Figure 5 4 Attach a collection vessel onto the bottom taper of each station giv
29. to advance to the next step Repeat Steps 9 through 13 using Methanol The times for Dispense Methanol Saturate Methanol and Soak Methanol must be entered Refer to the Factory Preset Methods chart on in Appendix A for an indication of the typical times used Rev B Pg 34 NOTE The final step Drain Methanol requires a time of 0 zero seconds This is critical to ensure the disk remains active for trapping the Oil amp Grease from the sample during the filtration process If atime is entered and the disk is allowed to dry go before the introduction of the water sample recovery values may fall out of tolerance for EPA Method 1664 QC requirements Step 15 The next step is to Air Dry the disk This is the amount of time desired for keeping the vacuum on and pulling air through the disk to remove residual water from the disk Refer to the methods chart in Appendix A and press the B key to quickly advance the timer Press the E key to advance to the next step NOTE As excessive volumes of air passing through the disk can adversely affect final recoveries the total time for this operation should be limited to a maximum of 10 minutes If longer times are required due to the dirty nature of the sample recovery studies should be conducted to ensure adequate recoveries Step 16 The next step is to Rinse the sample bottle with solvent The letter keys will change function to select n Hexane B key or to select Methanol C key Press the des
30. turns with pressure The pressure gauge for this is located inside the back panel of the extractor Verify that the left gauge when facing the back of the extractor is set at 50 psi 8 Sample overflows Disk Holder Assembly Sample overflows can only occur if there is an air leak into the bottle allowing the sample to be displaced even when the Disk Holder Assembly is filled to the opening of the down tube Examine the gasket on the Water Inlet valve This is the gasket that forms a seal between the sample bottle and Inlet Valve or if using a cap adapter between the cap adapter and Inlet Valve Verify that it is correctly in place with the white side face up and that it is not worn or damaged The two halves of the Water Inlet Valve assembly should be tight enough so that sample does not overflow If too tight the Water Inlet Valve will not open Also verify that the down tube that drains the sample into the Disk Holder Assembly is snug in place If using a cap adapter to accommodate for a wide mouth bottle use a gasket to form a seal between the cap and the lip of the bottle This is one area of the cap adapter where an air leak into the bottle can occur Do not over tighten the cap adapter as this can cause the cap to flare out or crack and the threading can be stripped The other area on the cap adapter that an air leak can occur is between the cap adapter opening connection to the Water Inlet Valve Again do not over tighten and strip
31. 2 Locate the 2 5 Liter safety coated Solvent Bottles two for the 1000XL four for the 3000XL Place these bottles on the left side of the Extractor Locate the bag with P N 50 0486 01 containing the turquoise colored Bottle Cap Assemblies Connected to the end of the solvent dip tube on the Bottle Cap Assemblies is a 10 um solvent filter Figure 4 2 and 4 3 This filter will filter the solvents and protect the solvent valves from particulate that may be present in the solvents To keep the solvent dip tube filter clean place the caps on the solvent bottles Do not tighten at this time Note Do not leave or place solvent lines in a dirty area Dust and particulate can get into the lines and end up in the solenoid valves resulting in damage Step 3 Solvent Bottle labels have been provided with the User s Manual For the 3000XL Extractor label two bottles Prewet 1 and Prewet 2 and label two bottles Rinse 1 and Rinse 2 For the 1000XL Extractor label one bottle Prewet 1 Rinse 2 and label the second bottle Prewet 2 Rinse 1 Warning labels to use the safety coated bottles provided have also been included with the manual Place a warning label on each safety coated solvent bottle Caution These solvent bottles will be pressurized in order to deliver the solvents For safety reasons only use the Safety Coated Bottles provided with the system Figure 4 1 Solvent Bottles Figure 4 2 Bottle Cap Assembly with solvent fi
32. 4 RUMMING a PUGS cus cco cus cote ss tes dis tes dis aa al a Als ar al a a a ao sues 25 5 5 Loading the SPE DISK espiren tod top tod toe tod tonto e J ie toe 27 5 6 Procedure for Preparing the Extractor to Run a Practice Water Sample 30 6 0 Programming the Controller 20055 eee 33 6 1 Creating a Method Using the Learn Mode 33 6 2 Creating a New Method 2 35 6 3 Special Function Keys 3eme nement Men 38 1 0 WANTON ANIC e iiss sedan een se le td NA NA NA NA ne 41 7 1 Start Up and Shut Down Procedures sees 41 7 2 Freeing the Elute Check Valve Assembly VV 42 7 3 Cleaning the water Inlet Valve Assembly 43 7 4 Replacement of Solvent Rod Shafts and O rings VV 44 7 5 Sample Bottle Gaskets Seals 45 7 6 Installing an Inlet Solvent Eiler eee 46 7 7 Adjusting the Water Sample Holder Arm sss sees 47 7 8 Replacing the Disk Holder Cup O ring sss sees 48 7 9 Uploading New Software Version sise 49 7 10 Assembly of the Dual Diaphragm Oil Free Vacuum Pump esse eee eee eee 50 7 11 Installing a Dry Tapis ren oh dsc tee oe eee eee ee ae ace ee nna 51 Rev B Pg 3 7 12 Cleaning and Replacing the Elute Check Valve 52 8 0 RIS nle Gennes GGG CRUE ROE ORES 54 Appendix A Factory Programmed Methods and Method Worksheet 60 Appendix B Sample Bottle Measuring Scale 63 Appendbe Ta TIE ne nn eke En nn 64 Appendic DNV AIAN e cdc
33. Dual Diaphragm Oil less Vacuum Pump 24 26 Hg 3 6 SCFM Dry Trap Please Contact the Parts amp Consumables Department at Horizon for a Parts Numbers and current pricing or visit our website at www horizontechinc com 66 Appenaix D Warranty Limited Warranty Horizon Technology Inc Horizon warrantees the SPE DEX Oil amp Grease Machines Product against defects in material or workmanship as follows 1 LABOR For a period of one 1 year from the date of purchase if this Product is determined to be defective Horizon will repair the Product and will cover all labor charges The Product must be returned to Horizon prepaid for repair Horizon will pay the return shipping charges 2 PARTS Horizon will supply at no charge new replacement parts for a period of one 1 year Horizon will send the replacement parts directly to the customer site for customer installation If the customer does not wish to install these parts the Product must be returned to Horizon To obtain warranty service the Product must be delivered prepaid in the original packaging If needed new packaging is available from Horizon To obtain warranty service the Product must be delivered prepaid in the original packaging If needed new packaging is available from Horizon This warranty does not cover customer installation or set up adjustments and or Product optimization The warranty also does not cover cosmetic damage or damage due to
34. Horizon technology SPE DEX Oil amp Grease Machine User s Manual H Oil Grease Machine 1000 3000XL Extractors Automated Solid Phase Extractor Systems Horizon technology 45 Northwestern Drive Salem NH 03079 800 997 2997 603 893 3663 Copyright 2007 This document is protected by all applicable copyright laws Duplication of this manual is prohibited without the express written permission of Horizon Technology Inc SPE DEX is a registered trademark of Horizon Technology Inc Teflon is a registered trademark of DuPont Company Empore is a registered trademark of 3M Corporation Speedisk is a registered trademark of J T Baker Rev B Pg 2 Table of Contents b ED Tse H NECN W te are data mt end la ete Dl Send le Sele Dla end le Dente dela un un Dune 5 20 Product Salely NOLICE cos tases ca a An nr SR A SORA en EN Rte 6 3 0 Oil amp Grease Machine Overview uit 8 3 1 SPE DEX 3000 Controller 8 3 2 SPE DEX Oil amp Grease Machines sise 10 ARO Installation acs sen ae ce nn mnt nn tn une a ere a 13 ASV Unpacking the e T steele cr ns 13 4 2 Installing the Oil amp Grease Machines 14 5 0 Operation eiei LR ns aes Succ alias Dora din la ice Ta ae et ent 19 5 1 Solid Phase Extraction and the Oil amp Grease Machines Theory of Operation 19 5 2 The Extractlon Cycle tatcears thea a ne Nr usar ee dat ee eee 20 5 3 Verify System Installations and Operation VV 23 5
35. act of the Oil amp Grease from the SPE disk Step 1 Prewet of the SPE Disk with Solvents Prewet Conditioning Solvent The extractor performs two prewet steps using n Hexane as the first prewet solvent and Methanol as the second prewet solvent The n Hexane prewet solvent removes manufacturing and handling impurities from the SPE disk Removing any impurities ensures that the SPE disk will not contribute any background contamination The Methanol prewet is the conditioning solvent that activates the sorbent packing within the SPE disk Once the packing is activated the SPE disk must not go dry until the completion of the water sample filtration cycle The Liquid Sensor Prewet Arm is used to deliver Prewet solvents into the Disk Holder Assembly Using timed sequences the desired volume of solvent is delivered into the Disk Holder Based on the size of the SPE disk used and the use of prefilters solvent dispense solvent soak and air dry times for each prewet solvent can be optimized Prewet Solvent Dispense Time The prewet solvent is dispensed for the duration of time specified in the method program Solvent dispense times depend on the type of disk disk size and prefilter used The volume of solvent dispensed is dependent on the solvent bottle pressure and solvent dispense time Prewet Solvent Saturate Time During the saturate time vacuum pulls the solvent into the disk packing to thoroughly permeate the disk before the soak time Pr
36. aintained This completes the installation of the SPE DEX Oil amp Grease Machine The vacuum pump source power and pressurized gas source should be turned off when the Extractor is not in use Rev B Pg 18 5 0 Operation 5 1 Solid Phase Extraction and the Oil amp Grease Machines Theory of Operation The SPE DEX 1000XL and 3000XL Series Extractor Systems automate the extraction of Oil amp Grease from aqueous samples by using Solid Phase Extraction SPE disks Pressurized gas is used to deliver the solvents to the Extractors and to actuate the Water Sample Inlet Valves Vacuum is used to remove all liquids solvents and the water sample from the Extractors When a method calls for a specific solvent the appropriate solenoid valve opens allowing the pressurized solvent to be dispensed as a Prewet or a Rinse Solvent The Prewet solvents are dispensed directly into the Disk Holder via the Prewet Arm where the Liquid Sensor is located while the Rinse solvents are sprayed directly into the sample bottle and allowed to drip into the Disk Holder via the Down Tube The Prewet steps prepare the SPE disk to receive the water sample The first Prewet solvent n hexane cleans the disk The second Prewet solvent methanol conditions the disk material to receive sample The disk is not allowed to go dry prior to receiving the water sample as this will adversely affect recoveries Vacuum is used to pull all waste solvents from the Disk Ho
37. ase Machines use a single HPLC grade Check Valve Assembly per extractor station This assembly is used to prevent solvent or the water sample from dripping into the collection vessel during the Soak Steps Process Sample Step or while the system is paused This assembly also drains the rinses into the collection vessel While this Check Valve prevents leaks the highly polished surfaces can cause the internal components to stick if not used routinely To free a sticky check valve follow the directions below Step 1 Remove the Disk Holder Assembly from the Disk Holder Platform Step 2 Locate the center opening where the solvent extract is pulled into the collection vessel See Figure 7 1 Step 3 Insert the Check Valve Release Tool Figure 7 2 provided with the System P N 02 0725 straight down into the opening Gently push downward and release This will move the Check Valve Ball off of the seat assembly An internal spring will gently push the ball back into the sealing position Repeat this procedure several times to free up the Check Valve Ball Check Valve Opening Figure 7 2 Check Valve Figure 7 1 Platform Check Fool Valve Opening Rev B Pg 42 Note Do not use a metal or hard material to push on the ball as this could permanently damage the check valve ball surface If using a tie wrap or thin piece of Teflon tubing do not push further than 6 to s of an inch as this could permanently damage the check valve
38. at connects to the bottle or cap adapter Make sure that there is no blockage of the opening and that the stainless steel tubing in this opening is not missing The dark spec is indicative that the stainless steel tubing is present If the rinse opening is blocked with debris or if the stainless steel tubing is missing the rinse spray may be week and or off center The valves can be sonicated in a bath to dislodge the debris from the rinse opening Alternatively a fine needle like tool commonly used for gas chromatography instruments can be used to clear the opening 12 Rinse leak from rinse port The solvent rod o rings are susceptible to damage with prolonged exposure to the solvents If a rinse leak is occurring during the rinse spray step the solvent rod o ring needs to be replaced The solvent rods can be pulled out of the Water Inlet Valve and flipped over to use the other end with the o ring There are two solvent rods on the Water Inlet Valve The one to the right when Water Inlet Valve shaft is pointing away from you is used to introduce the rinses into the bottle and the one on the left of the Water Inlet Valve is used for balance The two can also be interchanged Also examine the solvent rod and make sure it is not damaged Rev B Pg 57 A leak from the rinse port when the extractor is sitting idle and the bottles are pressurized is an indication of a stuck solenoid valve Care must be taken to avoid dust or particles from gettin
39. bject which will not damage the face of the gasket Figure 7 10 Inspect the gasket for any surface deformation Inspect both sides of the gasket One side of the gasket is a light tan color and the other side is white If the tan side looks deformed or is pealing off from the gasket discard the gasket and replace with a new one P N 02 0560 Insert the gasket with the tan side facing DOWN onto the Inlet Valve Gently press the gasket into position Periodically inspect the top surface of the gasket and replace if worn or if leaks occur Figure 7 9 Sample Bottle Gasket Seal Figure 7 10 Removing the Gasket Seal Rev B Pg 45 7 6 Installing an Inlet Solvent Filter The solenoid valves inside the extractor are susceptible to leaking if fine particulate matter from the solvents or solvent bottles gets into these valves If these small particles of matter are not prevented from entering the valve they could cause impressions to be made when the diaphragm inside the valves close on them One of the characteristics of PTFE which the valves are made of is its ability to retain memory When the valves open and close again the particle may well be freed but an impression of the foreign body is left behind in the PTFE This impression prevents an adequate seal to be formed when the valves close again allowing for the possibility of solvent leakage to occur There is one solenoid valve for each Prewet solvent and one for each R
40. c Air Dry step 5 Two methanol rinse steps will follow the Air Dry step During the rinse steps observe that the rinse spray hits the center of the base of the bottle The pressure of the rinse spray should be sufficient to wash the sides of the bottle Rev B Pg 31 6 The rinses will elute into the collection vessel 7 At the end of the last rinse the Water Sample Inlet Valve will close and the station will return to idle When the station has finished processing the sample remove the Water Sample Inlet Valve Assembly the Disk Holder Assembly and the collection vessel The collection vessel contains actual solvent therefore take the appropriate measures for disposing solvents Rev B Pg 32 6 0 Programming the Controller The SPE DEX 3000 Controller can store thirty 30 Oil amp Grease extraction methods Methods 1 through 20 can be created and modified by the user Methods 21 through 30 are Factory Preset Methods that cannot be modified See the chart in Appendix A for a listing of the Factory Preset Methods parameters The Factory Preset Methods serve three purposes 1 To provide a series of proven methods designed to handle a full range of Oil amp Grease samples using a variety of disks 2 To allow a faster System startup by using proven methods that can not be modified vs spending time creating new methods that might not be optimal to recoveries and 3 To provide common benchmark methods to allow the
41. ct the end with the fitting to the end of the t fitting on the top of the Water Waste Recovery Assembly The other end has no fitting and needs to be connected to the vacuum pump source Refer to Figure 4 7 It is highly recommended that a Dry Trap P N 50 0572 be installed to protect the vacuum pump source from solvent vapors and moisture Refer to Section 7 11 for instructions on installing a Dry Trap With the Dry Trap bowl facing down note the arrow with the letter N on the top Insert the two blue tubings provided into the orange quick disconnects on the brass fittings of the Dry Trap Verify that they are securely in place and will not easily slip out Align the arrow with the letter N so that it points toward the vacuum pump source in the direction of the flow Install the Dry Trap between the vacuum source and Water Waste Recovery Bottle Cut the BEV A LINE tubing that goes from the vacuum pump to the Water Waste Recovery Bottle Insert the ends of the cut line into the blue tubing on the Dry Trap Use a tie wrap on both ends to hold the tubing securely in place Attach the plug cap provided to the bottom of the Dry Trap bowl black end If liquid accumulates in the bowl drain by removing the plug cap Note Due to the many different types of fittings found on vacuum pumps the user must provide the necessary fittings to complete this connection Step 19 Step 20 Step 21 Step 22 Step 23 At the regulated main
42. d load the Water Sample Inlet Valve Assembly and bottle onto the Sample Holder Arm Help guide the Water Sample Inlet Valve shaft onto the actuator key K x Firmly press the valve in place Refer to Figures 5 5 and 5 6 Figure 5 6 Loaded Sample Boitle and Inlet 5 4 Running a Purge Method Valve Purging the Extractor Stations is necessary to remove any air from both the Prewet and Rinse lines and completely fill the lines with solvent Purging also flushes the water sample pathway for the Extractor Stations In addition this sequence is useful for verifying that the system is installed and operating properly The Purge cycle performs much like an actual extraction method by introducing the selected Prewet and Rinse solvents The shortened times of a Purge cycle however are not adequate to process a sample All soak times are automatically set to zero and the Process Sample and Air Dry steps are skipped The Controller is pre programmed with Method 30 for purging the extractor stations Alternatively the DRAIN and PURGE functions can be used These functions skip the Sample Processing step thereby decreasing the time required to purge the system Method 30 is factory preset and cannot be modified As more experience is gained with the system Methods 1 20 can be created and modified by the operator If desired any of these methods 1 20 can be created and used as a Purge Method See Appendix A for a complete listing of Method 30 paramet
43. e and wash the Extractor s water sample path Multiple rinses are used to ensure thorough washing of all surfaces The rinse solvents drain onto the SPE disk where the extraction occurs and then are eluted into the collection vessel The Extractors automatically performs the rinsing without operator intervention Water Soluble Rinse Solvent The water soluble solvent Methanol is used as the first rinse for each method This solvent removes the residual water from the sample bottle and the SPE disk The water soluble solvent is left to soak on the disk for a period of time before being pulled or eluted by vacuum into the sample collection vessel The actual duration of the elution process depends on a number of factors e g the disk being used if a prefilter is used how clean or dirty the sample is etc Removing residual water from the sample bottle walls and disk allows the subsequent n Hexane solvent rinses to properly extract the Oil amp Grease off the disk Extracting Rinse Solvent The extracting rinse solvent n Hexane extracts the Oil amp Grease from the sample bottle walls all surfaces of the water sample path and the disk The extraction solvent is left to soak on the disk for a period of time before being pulled or eluted by vacuum into the sample collection vessel Rev B Pg 21 Step 4 Extraction of the SPE Disk The n hexane rinses release the analytes from the SPE disk allowing them to pass into the collection vessel
44. e heated Pressing the Auxilliary 1 key on the Controller can access this information 5 3 Verify System Installations and Operation Step 1 Turn the vacuum pump on and adjust the main vacuum between 24 to 26 Hg Step 2 Turn on the pressurized gas source and slowly increase the pressure to 60 80 psi Verify that they are no liquid or gas leaks Step 3 Remove the back panel of the Extractor by removing the screws along the outer edges Locate the two pressure regulators at the base of the system The first regulator on the left when facing the back of the system is the extractor pressure Verify that the pressure is set at 50 psi Adjust the pressure if necessary by pulling back on the regulator knob to unlock it and then set the pressure Make sure to lock the regulator setting by pushing down on the knob The second pressure regulator located on the right when facing the back of the unit is the solvent bottles pressure for delivering both the Prewet and Rinse solvents to the Extractor Verify that this regulator is set at 15 psi Adjust the pressure if necessary and lock the knob Step 4 Turn the Controller on Step 5 The SPE DEX Oil amp Grease Machines use a single HPLC grade Check Valve Assembly This assembly is used to prevent solvents and water sample from dripping into the collection vessel during the Soak Steps and Process Sample Step or while the system is paused While this Check Valve prevents leaks the highly p
45. eh Gd puter e desc acd a ae ch Go ns tee 67 Rev B Pg 4 1 0 Introduction The SPE DEX Oil amp Grease Machines 1000XL and 3000XL Extractor Systems are specifically designed to handle the extraction of Oil amp Grease from aqueous samples Capable of using 47 50 JT Baker and 90 mm SPE disks the easy to use workstations set new standards for both speed of sample extraction and reproducibility of results The SPE DEX Oil amp Grease Machines allow testing facilities to e Reduce laboratory costs Reduce manpower requirements for testing Process more samples per hour Provide more reliable results by reducing testing variability Improve operator safety by reducing exposure to solvents Provide an increased level of productivity to the laboratory The SPE DEX Oil amp Grease Machines operate with the SPE DEX 3000 Controller The Controller is pre programmed with ten 10 optimized methods to handle a full range of Oil and Grease samples and disk configurations These factory preset methods can be used or twenty 20 new methods can be created and modified Creating custom methods allows optimizing the operating parameters to meet your own unique requirements and or to take advantage of new SPE disk technology as it becomes available The SPE DEX Oil amp Grease Machines consist of two main components the Controller and the Extractor unit The Controller The Controller provides operating power and programming log
46. ements and program to all workers Personal Protective Equipment 1 2 3 Eye protection in the form of safety glasses or goggles is mandatory when operating the Extractors If solvent liquids or vapors enter the eyes follow the appropriate first aid procedures set forth in the laboratory s safety manual Lab coats must be provided for protection and convenience They should be worn at all times when operating the Extractors Protection of the hands when working with solvents or any hazardous material is essential Wear gloves selected on the basis of the hazard Equipment Safety 1 2 3 4 5 6 The SPE DEX Oil amp Grease Machines are designed to only use n Hexane and Methanol The use of any other solvent will destroy critical valve seals and components and void all warranties The SPE DEX Oil amp Grease Machines have been designed for bench top operation or fume hood operation If installed on a bench top the extractors require venting into a local exhaust device to avoid the discharge of potentially toxic vapors and fumes into the laboratory atmosphere The equipment must be set up and operated in a well ventilated area Do not work with a volatile solvent without adequate ventilation from chemical fume hoods or other protective devices The only solvent bottles that should be used with the SPE DEX Oil amp Grease Machines are the 2 5 L Plastic coated safety bottles originally shipped wit
47. emove any air in the solvent lines and to ensure all is working before beginning with the extraction of samples Follow the instructions in Section 5 4 of the manual for running a purge method The Extractor is now ready for the extraction of samples Rev B Pg 41 Shutdown Procedure Follow the instructions below for shutting down the system at the end of the extraction day Attach a collection vessel to each station Pour water into the platform cavity and then run the Solvent Drain function by selecting the following keys Status STA 1 key A DRAIN key E SOLVENT key B This process will remove any methanol left in the solvent waste lines Repeat for the elute check valve by selecting the DRAIN and then ELUTE keys Repeat for the water waste valve by selecting DRAIN and SAMPLE 1 Disconnect the vacuum pump line from the Water Waste Recovery Bottle Let the Vacuum Pump run for 2 to 3 minutes to allow it to vent Switch off the vacuum pump and reconnect the vacuum line to the Water Waste Bottle 2 Turn off the gas supply 3 Switch the Controller off 4 Flush the Water Sample Inlet valve by actuating it under warm running water Wash the valve with a non surfactant detergent if necessary make sure to rinse it thoroughly Do not use an abrasive brush for scrubbing 5 Leave the Water Sample Inlet Valve in the half open position to allow the water to drain 7 2 Freeing the Elute Check Valve Assembly The SPE DEX Oil amp Gre
48. ent extract Connects the Extractor to the Controller and provides both power and program communication e Solvent Inlet Lines Fittings to connect the solvent delivery lines Fittings e Waste Line Fittings Fittings to connect solvent to waste and water to waste lines e Solvent Vapor Exhaust Holds the Vapor Exhaust Hose to the Extractor Rev B Pg 10 Sample Bottle Water Sample Inlet Valve Water Sample Holder Arm Sample Inlet Tube Liquid Sensor Prewet Arm Disk Holder Assembly Disk Holder Platform Collection Vessel Collection Vessel Adapter Taper and Clip Figure 3 4 Front Components of SPE DEX 3000XL Oil amp Grease Machine Water Sample Holder Arm Liquid Sensor Prewet Arm Disk Holder Assembly Disk Holder Platform Collection Vessel Collection Vessel Adapter Taper and Clip Figure 3 5 Front Components of SPE DEX 1000XL Oil amp Grease Machine Rev B Pg 11 Controller Cable Port Solvent Inlet Lines Solvent Bottle Gas Lines Solvent Vapor Exhaust Pressurized Gas Inlet Solvent Waste Line Water Waste Line Figure 3 6 Back Components of SPE DEX 3000XL Oil amp Grease Machine Controller Cable Port Solvent Vapor Exhaust Solvent Inlet Lines Solvent Bottle Gas Lines Pressurized Gas Inlet Solvent Waste Line Waste Water Line Figure 3 7 Back Components of SPE DEX 1000XL Oil amp Grease Machine Rev B Pg 12 4 0 Installation 4 1 Unpacking the System
49. ers Once the Purge Method is started the following sequence of events will occur These steps will be displayed on the LCD 1 Prewet 1 n Hexane will be dispensed into the Disk Holder Cup for One 1 second 2 The Solvent to Waste vacuum valve will open for five 5 seconds and pull the n Hexane into the safety coated Solvent Recovery Waste Bottle 3 Prewet 2 Methanol will be dispensed into the Disk Holder Cup for one 1 second 4 The Solvent to Waste vacuum valve will open for five 5 seconds and pull the Methanol into the safety coated Solvent Recovery Waste Bottle 5 The Water Inlet Valve will open even though no water sample is present Rev B Pg 25 Note When using Controller Programmed Method 30 before each extractor station will advance into the Rinse Step the liquid sensors will confirm that no water sample is present A fixed timer of 45 seconds is used to confirm the presence or absence of a water sample Once the sensor performs this test and confirms no water is present the station will advance to the Rinse Step This Water Processing Step is skipped if using the Purge Function D key 6 Rinse 1 n Hexane is sprayed into the sample bottle for three 3 seconds 7 The Sample Collect vacuum valve will open for five 5 seconds and pull the solvent into the collection vessel 8 Rinse 2 Methanol will be sprayed into the sample bottle three 3 seconds 9 The Sample Collect vacuum valve will open
50. evaporation and calculate the difference to determine if any background is being introduced from the solvents Values should be below 1 0 mg Rev B Pg 58 Program a short method with only prewet hexane no prewet methanol or rinses and set all other parameters to zero Set the dispense time so that approximately 30 ml of solvent will be collected No soak or drain time is needed Place a clean collection vessel under the prewet hexane delivery line on the Prewet Sensor Arm Start the short method and collect the solvent Pour the solvent into a pre weighed aluminum pan and evaporate Repeat for the prewet methanol this time changing the method so that only prewet methanol will be delivered Determine if any background is introduced from the extractor Program a short method with only the hexane rinse solvent and set all other parameters to zero Set the dispense time so that approximately 30 ml of solvent is collected Set the soak time to zero and drain time to 10 sec Use the 90 mm Disk Holder Assembly with no disk Place a clean collection vessel onto the adapter stem and run the short method Pour the rinse solvent into a pre weighed pan and evaporate Determine if any background is present Repeat with the methanol rinse solvent Run non acidified reagent water blank with the disk and method normally used at the lab Evaporate the collected hexane portion and determine if any background is coming from the disk Rev B Pg 59 Appendix
51. ewet Solvent Soak Time The solvent remains on the disk and soaks the disk for the time specified in the method to allow for the solvent to thoroughly clean the disk from impurities or to activate the sorbent packing Prewet Solvent Air Dry Time An Air Dry time can be specified after each prewet soak time The Air Dry time completely drains the solvent from the disk Different solvents permeate the disks at different rates and disks also vary in permeability The second prewet solvent Methanol which activates the disk sorbent packing has an Air Dry time of 0 sec so that the disk is left wet and ready to receive the water sample Rev B Pg 20 For proper chemistry to be obtained it is essential that during the activation conditioning of the SPE sorbent material Prewet 2 Methanol the Prewet Air Dry step does not allow the SPE disk to go dry Step 2 Introduction of the Sample to the SPE disk Process Sample Step After the conditioning of the SPE disk by the last prewet step Methanol the Water Sample Inlet Valve opens automatically introducing the sample onto the SPE disk During the process sample step the analytes of interest Oil amp Grease are trapped onto the sorbent material of the SPE disk The cleanliness amount of suspended solids and viscosity of the sample and the amount of surface area of the SPE disk used control the rate of filtration Clean samples will flow faster than dirty samples and 90 mm
52. g motion then secure the vessel with the blue clip Step 5 Turn the vacuum pump on and adjust the main vacuum to 24 to 26 Hg Step 6 Turn on the pressurized gas source and adjust to 60 80 psi Step 7 On the Controller press the LEARN key Rev B Pg 33 Step 8 NOTE If a factory preset method is currently loaded into a station that you wish to use the Controller will beep indicating that the method can not be modified only methods 1 20 can be modified or learned Press the STATUS key followed by key A B or C keys and select a method from 1 to 20 Press the LEARN key to return to the LEARN Mode Step 9 Step 10 Step 11 Step 12 Step 13 Step 14 The LCD will display the three stations and the methods currently loaded into each station Select the station to be used by pressing the A STA 1 B STA 2 or C STA 3 keys Refer to Figure 6 1 LA Le cio N Once the desired station has been selected the LCD will display the action that will be performed Figure 6 1 3000 Controller in i e Dispense Hexane and will indicate that the LEARN mode system is paused The letter keys will change function and indicate the available actions that can be performed Press the A key to Resume the operation N Hexane will begin dispensing into the Disk Holder Assembly Carefully observe the n Hexane and press the A key again to Pause the operation As the n Hexane is dispensing the elapsed time is dis
53. g into the solvent delivery lines so that they don t end up in the valves which could cause damage When disconnecting the lines from the extractor always place them on a clean surface and wrap the open ends to prevent dirt from getting in Always use the solvent filters at the end of the pick up lines in the solvent bottles refer to section 4 2 13 No Elution into collection vessel As part of the daily start up procedure Section 7 1 free the sample collect check valves using the check valve tool PN 02 0725 The check valves can stick when the unit is not in use The check valve requires a minimum vacuum level of 13 Hg to open Verify that the vacuum does not drop below this level when the system is in operation and pulling vacuum If the collection vessel does not make a good vacuum tight seal with the Adapter Taper vacuum will be lost and the rinses will not elute into the collection vessel or partial and weak elution may be observed When attaching the collection vessel to the Adapter Taper use a lifting and twisting motion to secure it in place VOA vials may have imperfections at the rim of the opening that could result in a poor seal between the rim and the cap of the glass joint adapter Use a different VOA vial if it looks suspect Always attach the glass joint adapter to the VOA vial by using the turquoise cap and verify that the base of the glass joint is sitting on the cap Do not screw it on by twisting the glass j
54. gas supply turn on the gas source and slowly increase the gas pressure to 5 psi Verify all solvent lines to ensure that there are no leaks If no leaks are found slowly increase the gas pressure to 60 psi minimum 80 psi maximum Connect the power cord to the 12V AC Power Adapter Attach the 12V cable to the backside of the Controller The connector is notched and must be keyed into position Next attach the power cord to the 120V outlet Connect the Extractor cable from the Controller to the top connector on the back of the Extractor unit The cable connector is keyed so carefully rotate the collar while gently pushing the cable into the socket Once the key is mated secure the cable by screwing on the outside collar ring Turn the Controller power switch on The switch is located on the back of the Controller The unit will perform an internal diagnostic test will give a series of beeps and display the main screen The Extractor fan will start working once the controller is switched on Turn the vacuum pump on and verify that the Water Waste Recovery Assembly is securely in place If the vacuum pump has an air bleed control regulator set the vacuum level to the maximum level A vacuum level of between 24 to 26 Hg is recommended to achieve the best flow rates Rev B Pg 17 Note It is strongly recommended to have a vacuum gauge on the main vacuum in order to monitor the vacuum level and ensure that the proper vacuum is being m
55. h the System Horizon Part Number P N 27 0042 The solvent bottles are pressurized to deliver solvent to the Extractor units The use of any other glass bottles could lead to an explosion All solvent bottles must have a label identifying the solvent it contains All solvent bottles must also have the safety label identifying them as Plastic coated safety bottles The only Waste Water Recovery Container that should be used with the SPE DEX Oil amp Grease Machines is the 5 gallon safety coated recovery bottle P N 180 0005 01 or equivalent High vacuum is pulled on this recovery bottle an inferior bottle could implode causing severe injury The 5 gallon Waste Water Recovery Container must have a label identifying it as a safety coated container During the Purge and Rinse cycle solvent is sprayed from the opening on the Water Inlet Valve Be sure a sample bottle is in position before pressing the PURGE or START key to avoid solvent being sprayed into the atmosphere and possibly causing injury to laboratory workers Rev B Pg 6 Chemical Safety 1 2 3 4 Organic solvents can pose inhalation skin and ingestion hazards with potential chronic health effects N Hexane is extremely flammable which could cause fire and or explosion hazards All solvents must be handled using appropriate personal protection equipment and in a properly operating fume hood to eliminate inhalation hazards Refer to the Material Safety Data Sheet for
56. hanol Selecting the No Rinse option will result in no rinse solvent delivery for that step Selecting the Manual option will cause the system to be interrupted at that rinse step The controller will display a message indicating that the station needs attention This is used if the operator wants to perform a manual rinse by adding the rinse solvent to the bottle manually rinsing the bottle by shaking and then pouring the solvent into the Disk Holder Assembly Select Methanol as the first rinse Input a Rinse time A rinse time of 2 seconds is recommended if using a 47 or 50 mm SPE disk and 5 seconds if using a 90 mm SPE disk The maximum Rinse time that can be entered is 10 seconds Press the A STEP key to advance to the next line The Soak time allows the disk to sit in the solvent for the specified time A short soak time of 5 seconds is recommended The maximum soak time that can be entered is 10 minutes Press the A STEP key to advance to the next line The Elute time allows the rinse solvent to elute into the collection vessel Enter Zero 0 seconds for the Elute Time of the first rinse Press the A STEP key to advance to the next line Select Methanol as the second rinse Input a Rinse time A rinse time of 2 seconds is recommended if using a 47 or 50 mm SPE disk and 5 seconds if using a 90 mm SPE disk The maximum Rinse time that can be entered is 10 seconds Press the A STEP key to advance to the next line
57. he next line The Soak Methanol command activates the sorbent material within the SPE disk preparing it for the sample Typical soak times range from 1 minute to 1 minute and 30 seconds The maximum time that can be entered is 5 minutes Press the A STEP key to advance to the next line The Drain Methanol step removes the solvent from the disk Warning It is critical that the Methanol is not removed from the disk Therefore set a Drain time of 0 seconds Press the A STEP key to advance to the next line The Air Dry step uses vacuum to dry the disk by removing excess water from within the pore spaces of the SPE disk This step follows the Sample Processing step Air Dry times longer than 10 minutes should not be used due to the volatile nature of some Rev B Pg 36 Step 21 Step 22 Step 23 Step 23 Step 24 Step 25 Step 26 Step 27 Step 28 Step 29 Step 30 Step 31 Step 32 Step 33 Step 34 Step 35 components of Oil and Grease Set the Air Dry time so that residual water is sufficiently removed from the disk allowing for the n Hexane extraction solvent to effectively extract the oil and grease from the disk The maximum time that can be entered is 90 minutes Press the A STEP key to advance to the next line The screen will advance to page 2 for the selection of the rinse solvents Use the C ADJ key to select the desired option The choices are as follows No Rinse Manual Hexane Met
58. ic for up to three extractor stations If the Controller is operating the 1000XL Station 1 is used if operating the 3000XL all three stations are used The use of the same Controller provides an easy upgrade from the 1000XL to the 3000XL Extractor Weight 5 Ibs Dimensions 7 W x 10 D x 5 1 2 H Power 120VAC 12VDC The Extractor The SPE DEX 1000XL Oil amp Grease Machine is a fully automated Extractor System which consists of a single extractor station The SPE DEX 3000XL Oil amp Grease Machine is a fully automated Extractor System which consists of three stations The operator simply loads the SPE disk a collection vessel and the water sample The System automatically introduces all solvents introduces the water sample rinses the sample bottle and extracts the oil and grease from the disk The solvent vapor exhaust fan allows bench top operation Studies confirmed the internal exhaust fan reduced the exposure of hexane vapor to levels below OSHA requirements SPE DEX 1000XL Weight 20 Ibs Dimensions 7 W x 15 D x 21 1 2 H SPE DEX 3000XL Weight 45 Ibs Dimensions 18 W x 15 D x 21 1 2 H Rev B Pg 5 2 0 Product Safety Notice This Safety Notice has been provided to inform and guide the operator in safe practices to avoid injury and is intended to cover general safety requirements for a laboratory Each laboratory is responsible for implementing and communicating its unique safety requir
59. ing a lifting and twisting motion to secure the vessel and ensure a vacuum tight seal If using a VOA vial and adapter make sure that they are screwed together firmly Do this by holding the cap not the glass piece when attaching the adapter to the vial If the glass adapter is allowed to rotate within the cap this could lead to the Teflon seal in the adapter to be cored which would cause the integrity of the vacuum seal to be lost Use the blue retaining clip to secure the collection vessel Figure 5 4 Disk Holder Cup amp Prewet Arm Rev B Pg 24 Preparing and loading the sample bottle properly is critical to the successful operation of the SPE DEX Oil amp Grease Machines Cap adapters of various sizes are available for different bottle sizes Contact Horizon Technology if custom cap adapters are needed The Following will explain the proper procedure for loading the sample bottle Use an empty bottle Step 9 Lower the Water Sample Holder Arm Step 10 The Water Sample Inlet Valve Assembly will fit a 33 x 400 mm Boston Round Bottle For larger sized bottles use the Figure 5 5 Sample appropriate Cap Adapter If the bottle to be used requires an Bottle and Inlet Valve adapter attach the larger cap adapter to the bottle and then attach the Water Sample Inlet Valve to the adapter Step 11 Orient the aluminum shaft of the Water Sample Inlet Valve such that it points toward the system Step 12 Invert the bottle an
60. inse solvent If one of the Prewet solenoids is leaking solvent will constantly dispense from one of the lines on the Liquid Sensor Prewet Arm Indication of a Rinse Solvent solenoid valve leak occurs when a constant leak is observed from the rinse opening on the Sample Holder Arm The Prewet or rinse leaks are observed once the solvent bottles have been pressurized To reduce the chances of damage to the solenoid valves an inlet solvent filter is placed in the solvent bottles at the end of each solvent delivery line Installing an Inlet Solvent Filter Locate one of the Inlet Solvent Filters PN 40 012 HT This is a metallic cylinder of approximately 11 2 inches in length and 1 2 inch in diameter Refer to Figure 7 11 Make sure that the small stem on the Inlet Solvent Filter has a plastic ferrule attached to it Unscrew the solvent draw up line from the solvent bottle cap and remove 1 Cut the plastic solvent draw up line to a length of eleven inches While holding the plastic solvent draw up line with a piece of 3M Scotch Brite for providing a good grip slide the ferrule cover of the Inlet Solvent Filter inside the end of the solvent draw up line and press it in as far as it will go It may be necessary to expand the LE end of the line to allow the Inlet Solvent Filter stem to ER Eee DES enter the solvent line and become firmly attached If the Figure 7 11 Inlet Solvent Filters solvent draw up line needs to be expanded then insert a
61. ired key Step 17 The letter keys will again change function and appear as they did during the initial Prewet learning sequence Follow the instructions on the LCD using the Factory Preset Methods chart in Appendix A as a guide Enter values for Rinse Soak and Elute Step 18 When all the rinses and the final Elution steps are complete press the E key DONE to exit the LEARN Mode Step 19 Press the METHOD key and use the A or B key to scroll to the method number that was LEARNED The time parameters were saved and are now displayed in the method This method can now be loaded to any of the extractor stations and used It is also possible to change any of the automatic times calculated during the Learn Mode 6 2 Creating a New Method The Controller allows the operator to create 20 new analytical methods Once an understanding of the requirements for optimal recoveries has been gained it will no longer be necessary for a learned method to be used The operator will become familiar with amounts of solvents delivered and rinse times needed and it will only be necessary to use the Create Method option When creating a method no solvents are actually delivered or sprayed only the times are entered The steps below detail how to create a new method To help create new methods use the Methods Worksheet in Appendix A Based on the SPE disk and prefilters that will be used we recommend using the Factory Preset Methods chart as a guide to get a
62. is 10 seconds Press the A STEP key to advance to the next line The Soak Hexane command allows the SPE disk to be cleaned from manufacturing impurities by soaking Typical soak times range from 1 minute to 1 minute and 30 seconds The maximum time that can be entered is 5 minutes Press the A STEP key to advance to the next line The Drain Hexane command pulls n Hexane completely through the disk and into the Waste Solvent Collection Bottle Typical times are 1 minute to 1 minute and 30 seconds The maximum time that can be entered is 5 minutes Press the A STEP key to advance to the next line The Dispense Methanol command will cover the SPE disk with Methanol Use the C ADJ and the D key ADJ to increase or decrease the desired time for dispensing the Methanol solvent Holding the keys down will quickly advance or decrease the times Visually ensure a sufficient volume of Methanol is dispensed in order to cover the disk and prefilters thoroughly Press the A STEP key to advance to the next line The Saturate Methanol command allows a small amount of the Methanol to be pulled into the pore spaces of the SPE disk to soak and activate the sorbent bed inside of the SPE disk Typical times are between 1 and 2 seconds Pulling too much Methanol through the SPE disk may allow the SPE disk to prematurely dry reducing recoveries The maximum time that can be entered is 10 seconds Press the A STEP key to advance to t
63. izon Systems The following instructions will explain installation for both applications Installing a Dry Trap for the use with the Extractor System s Vacuum Pump In order to extend the life of the Vacuum Pump and maintain the warranty it is highly recommeded that a Dry Trap PN 50 0572 be installed to capture solvent vapors or moisture that could be pulled through the vacuum pump With the Dry Trap bowl facing down note the arrow with the letter N on the top Refer to Figure 7 19 Insert the two blue tubing provided into the orange quick disconnects on the brass fittings of the Dry Trap Verify that they are securely in place Align the arrow with the letter N so that it points toward the vacuum pump source in the direction of the flow Install the Dry Trap between the vacuum source and Water Waste Recovery Bottle Cut the BEV A LINE tubing that goes from the vacuum pump to the Figure 7 18 Dry Trap Water Waste Recovery Bottle Insert the ends of the cut line into the blue tubings on the Dry Trap Use a tie wrap on both ends to hold the tubing securely in place Attach the plug cap provided to the bottom of the Dry Trap bowl black end If liquid accumulates in the bowl drain by removing the plug cap Installing a Dry Trap for use with the Speed VAP Air Source In order to extend the life of the Speed Vap and maintain its warranty a Dry Trap is recommended to capture moisture or particles that could be forced int
64. lder Assembly into the safety coated Waste Solvent Recovery Bottle This occurs automatically after the programmed disk soak times have elapsed This automated feature minimizes operator handling of solvents and or exposure to solvent vapors The collected solvent waste can then be disposed of properly Once the disk has been conditioned the Extractor will automatically actuate the Water Sample Inlet Valve and deliver the water sample to the disk The Liquid Level Sensor will then detect the presence of the water and after a short sensor equilibration period 20 seconds the water waste vacuum valve will open and filter the water through the disk and out to water waste It is during this filtration process that the Oil amp Grease in the sample gets retained onto the SPE disk All of the water waste is collected in a separate safety coated 5 gallon Water Waste Bottle When the water sample has been completely filtered the sensor will be exposed to air and undergo a change in temperature warms up The extractor station will then advance into the Air Dry step for the programmed duration Consistent vacuum times are critically important to achieve high recoveries This is due to the partial volatility of some of the lighter molecular weight hydrocarbons At the completion of the Air Dry time the vacuum will automatically shut off Once the Oil amp Grease has been retained on the SPE disk the SPE DEX Oil amp Grease Machine will automatically sp
65. let Valve to the cap adapter Step 4 Position the sample bottle Inlet Valve so that the aluminum shaft of the Inlet Valve points in the direction of the extractor With the left hand firmly grasping the bottle place the right index finger over the left opening of the Solvent Rod Rinse Port Refer to gt Figure 5 19 Step 5 Invert the sample bottle and load it onto the Sample Holder Arm Guide the Water Sample Inlet Valve stem onto the actuator key and firmly press the valve in place Refer to Figures 5 6 Step 6 Verify that the main vacuum pump is on and set between 24 Hg and 26 Hg f Figure 5 19 Water Step 7 Verify that the main gas source is on and set between 60 80 Sample Inlet Valve psi Rinse Port Step 8 Press the START key on the controller The LCD will display Warning Process starting ready to start Press the YES key The extraction sequence programmed as Method 1 will begin Observe the station for the following steps 1 Methanol Prewet solvent dispenses into the Disk Holder Assembly 2 After a 10 sec soak the Water Sample Inlet Valve will open and deliver the water sample into the Disk Holder Assembly 3 There will be a 20 sec delay for the sensor to equilibrate before the Water to Waste Valve opens to drain the water sample to the safety coated Water Waste Recovery Bottle 4 Once the water from the bottle has been completely drained to the waste the extractor will proceed to the 10 se
66. llection vessel to the extractor station with the leaking Check Valve Insert a Disk Holder Cup with the support screen into the platform opening Fill the Disk Holder Cup with warm water Go to the Status screen of the Controller and select the extraction station Use the Drain key followed by the Elute key to drain the water into the collection vessel Continue this process by flush lots of water through the Check Valve remembering to empty the collection vessel as it fills This should clear the Check Valve of dirt and debris If the Check Valve continues to leak disassemble it to clean the internal components Remove the three screws at the bottom of the Check Valve These screws hold the Check Valve Assembly into the Collection Vessel Adapter Taper Figure 7 20 Note the opening toward the back This is the vacuum opening which supplies vacuum to the collection vessel The vacuum opening is aligned with the vacuum opening on the Collection Vessel Adapter Taper Once the screws have been removed try pulling the Check Valve Assembly out from the Adapter Taper The Check Valve Assembly may be stuck and difficult to remove In this situation use a flat edged blade and gently ease the Check Valve Assembly out by placing the blade between the Check Valve base and the Adapter Taper Be careful not to score or damage the Teflon Adapter Taper Rev B Pg 52 8 Remove the top cover of the assembly Be careful not to lose any of the components
67. llon safety coated bottle until sufficient vacuum is created to securely hold it in place 5 Free the Elute Check Valves with the Check Valve Tool Refer to Section 7 2 6 Load an empty Disk Holder Cup with the Support Screen in place onto the platform for each station 7 Lift and move the Liquid Sensor Prewet Arm into the Disk Holder Cup 8 Attach a collection vessel to each sample station 9 Lower the Water Sample Holder Arm 10 Rinse the Water Inlet Valve by actuating it under warm running water Check for freedom of rotation If the valve is tight and does not freely rotate adjust the four hex head locking screws by turning each one counterclockwise one quarter turn at a time until the valve moves freely 11 Make sure that the Bottle Gasket in the Sample Inlet Valve is orientated such that the tan side is facing down and the white side is visible Replace if the gasket is worn or damaged Attach an empty sample bottle to the Water Inlet Valve Make sure the valve is closed Load the sample bottle and Water Sample Inlet Valve assembly onto the Holder Arm 12 Purge each station Gas Pressure Adjust the gas source pressure in increments checking for leaks as the pressure is increased to a minimum of 60 psi and a maximum of 80 psi Vacuum Level The main vacuum source gauge should read at least 24 Hg Check that there are no kinks or crimps in the vacuum lines Run the purge Method twice Purge all extractor stations to r
68. load Copy the 2 software files to the Upload directory S e K re Attach a Null Modem cable from the serial port on the PC to the Comm Port on the backside of the Controller 6 On the PC select Windows START followed by RUN Type in this command C UPLOAD UPLOAD EXE SUPERC BIN 1 38400 1024 for COM 1 OR C UPLOAD UPLOAD EXE SUPERC BIN 2 38400 1024 for COM 2 7 Press the A YES key on the Controller The LCD will turn off and the Controller will erase the current software version from the EEPROM The Controller is now ready to receive the new software 8 Press ENTER on the PC This will initiate the upload process The upload process will be displayed on the monitor When the upload is complete the Controller will reinitialize and reboot 9 The current software version will be displayed on the main startup screen when the Controller is rebooted Rev B Pg 49 7 10 Assembly of the Dual Diaphragm Oil Free Vacuum Pump The dual diaphragm oil free vacuum pump offers several advantages over other vacuum pumps The most noticeable advantage is seen when using 90 mm SPE Disks The dual diaphragm pump is capable of maintaining adequate vacuum levels under the increased draw from a 90 mm disk in an air dry step This translates to a minimal drop in vacuum levels when several extractors are running simultaneously The SPE disk is allowed to dry more thoroughly yielding less water in the final extract and therefore a better ext
69. lter Figure 4 3 Solvent Filter Rev B Pg 14 Step 4 Step 5 Step 6 Step 7 Step 8 Step 9 Fill the Prewet 1 and Rinse 2 safety coated solvent bottles with n Hexane 85 purity 99 0 min saturated C6 isomers and residue less than 1 mg L Securely attach the turquoise caps to the bottles Fill the Prewet 2 and Rinse 1 safety coated solvent bottles with Methanol ACS grade residue less than 1 mg L Securely attach the turquoise caps to the bottles 3000XL Extractor Connect the fitting from the Prewet 1 n Hexane bottle to the top most left bulkhead fitting labeled PW1 on the back of the Extractor Refer to Figure 4 4 Repeat this operation for the Prewet 2 bottle Methanol use the next fitting labeled PW2 Connect Rinse 1 methanol bottle and Rinse 2 Hexane bottle to the fittings labeled RIN1 and RIN2 respectively Using a wrench make sure the fittings are snug but do not over tighten 1000XL Extractor Connect the fitting from the Prewet 1 Rinse 2 n Hexane bottle to the top left bulkhead fitting Connect the fitting from the Figure 4 4 3000XL Extractor Prewet 2 Rinse 1 bottle Methanol to the top Connections right bulkhead fitting Using a wrench make sure the fittings are snug but do not over tighten Figure 4 5 Attached to each Solvent Bottle Delivery Line is a black gas line Gently connect each gas line to the plastic fitting directly below the solve
70. mes and variables will automatically be reset to zero When future software revisions are introduced this function will allow new Factory settings to be downloaded into the Controller Press the A key for YES press the E key for NO Warning Only perform this operation if new software is to be uploaded In order to prepare the Controller to receive new code the code presently running the Controller is first erased making the Controller inoperable Rev B Pg 40 7 0 Maintenance The following information is designed to assist with the general maintenance and replacement of standard components on the Oil amp Grease Extractor Systems If any questions arise contact Horizon Technology Inc for assistance 7 1 Start up and Shut down Procedures Start up Procedure Activate the Power Switch which is located on the back of the SPE DEX 3000 Controller and allow the Liquid Sensor to stabilize for a period of between 3 to 5 minutes 1 Verify that the safety coated Solvent Bottles are filled with clean contamination free solvents and that their caps are securely fastened Ensure the solvent lines are not kinked or crimped 2 Empty both safety coated waste bottles if necessary Water and Solvent 3 Turn on the pressure source nitrogen gas and slowly increase the pressure to 60 80 psi Inspect the solvent delivery lines for any liquid leaks 4 Turn on the vacuum pump source Hold the Waste Water Recovery Assembly onto the 5 ga
71. n idea of the times required Step 1 Press the METHOD key on the right hand side on the controller Method 1 is always displayed when the METHOD key is pressed Step 2 The screen will display the first page of Method 1 Prewet page Use the A NEXT and B PREV keys to scroll up or down through the method numbers Step 3 Press the E key to edit the method Step 4 The cursor will highlight the first line of the method the Dispense Hexane time for the first Prewet solvent Use the C ADJ and the D key ADJ to increase or decrease Rev B Pg 35 Step 5 Step 6 Step 7 Step 8 Step 9 Step 10 Step 11 Step 12 Step 13 Step 14 Step 15 Step 16 Step 17 Step 18 Step 19 Step 20 the desired time for dispensing the n Hexane Holding the keys down will quickly advance or decrease the times Set the time so that sufficient n Hexane solvent is delivered to thoroughly cover the disk and prefilters if used The maximum time that can be entered is 30 seconds Once a time has been entered press the A STEP key to advance to the next line The Saturate Hexane command allows a small amount of the n Hexane to be pulled into the pore spaces of the disk to optimize the soaking and washing of the disk Typical times for this step would be 1 to 2 seconds Longer times could pull too much n Hexane through the disk not leaving enough behind for proper soaking and cleaning The maximum time that can be entered
72. nt line on the back of the extractor These fittings are labeled for the corresponding solvent bottle Locate the bag containing two pieces of tubing one blue and one clear P N 50 0485 Attach the fitting on the blue tubing to the fitting marked Water Waste This is the fitting on the bottom right of the unit Attach the fitting on the clear tubing to the fitting marked Solvent Waste This is the middle fitting on the bottom of the unit Keep the other plastic fittings in this bag for later use Locate the bag containing the single piece of black tubing P N 50 0484 Attach one end of the tubing to the fitting marked Gas on the back of the unit This is the bottom left fitting Attach the other end of the gas tubing to a pressurized gas source nitrogen gas recommended capable of delivering a minimum of 60 psi Hexane Verano Figure 4 5 1000XL Extractor Connections Rev B Pg 15 Step 10 Step 11 Step 12 Step 13 Step 14 Caution A high vacuum is applied to the 5 gallon Water Waste Bottle Only the heavy walled safety coated bottle provided with the system should be used The danger of implosions can occur if other containers are used Step 15 Locate the Solvent Vapor Exhaust Hose The standard length provided with the system is 10 feet To help with the attachment of the Exhaust Hose pull the right side of the extractor unit forward The Duct Hose fitting is threaded to securely hold the Duct Hose
73. o the Speed Vap It is important to align the trap correctly for proper performance Note the N and the arrow direction on the top of the trap Orient the arrow on the trap so it faces away from the air source or in the direction of the flow Install the Dry Trap between the Air Source and Speed Vap by cutting the tubing and inserting into the fittings Figure 7 19 Top View of Dry Trap Rev B Pg 51 7 12 Cleaning and Replacing the Elute Check Valve The Elute Check Valve Assembly Figure 7 20 elutes the rinses into the collection vessel The Check Valve Assembly components include a spring a ceramic ball that sits on the spring the seat and a washer Figure 7 21 Check valve assembly mounted on extractor Vacuum opening adapter stem Checi Valve Assembly Figure 7 20 Check Valve Assembly Figure 7 21 Check Valve Components When not in use the check valve ball can stick preventing the elution of solvents into the collection vessel Prior to beginning an extraction free the Check Valve by following the instruction in Section 7 2 of the manual Dirt or debris can enter the Check Valve and get lodged between the ball and the seat When this happens an open pathway is created that allows all liquids to drip through the Check Valve and into the collection vessel during other steps in the extraction process To clean the Check Valve follow the instructions below 1 2 3 4 5 6 7 Attach a 125 ml co
74. o waste drain This usually is a result of dirt trapped in the valve and keeping it stuck open Fill the Disk Holder Assembly with warm to hot water and use the DRAIN and SAMPLE functions of the controller Continue filling and draining the Disk Holder Assembly If after the sample has completely filtered through the disk the unit remains in sample processing and does not advance to the next step rinse verify the sensor response in the A2D screen of the controller 11 No rinse spray into the bottle A gas source is required to spray solvent rinses into the sample bottle Check the main gas source and make sure the tank is not empty and that the pressure is set at between 60 80 psi Verify that there is sufficient solvent in the bottles Crimps or kinks in the solvent delivery lines will impede the flow of solvents Examine the rinse solvent delivery lines and ensure that no kinks are present Kinks usually appear at the fittings on either the solvent bottles or the connection to the extractor Straighten the kinks if possible and replace any damaged lines Examine the caps of the solvent rinse bottles for any cracks or damage that may result in loss of pressure A gas leak can usually be detected by a hissing sound Also examine the black gas lines to the solvent bottle caps for damage The Water Inlet Valves have an opening through which the rinses are sprayed The rinse opening is the dark spec located on the Inlet valve at the end th
75. of the Extractor System It is recommended to use a 47 mm Disk Holder Assembly for the practice method If not already programmed program Method 1 with the parameters given below Refer to Chapter 6 0 for programming the controller Method 1 PW Methanol dispense 5 sec saturate 1 sec soak 10 sec drain 0 sec Air Dry 10 sec Rin 1 Methanol rinse 3 sec soak 1 sec elute 5 sec Rin 2 Methanol rinse 3 sec soak 1 sec elute 10 sec Loading a method to the extractor stations Power up the controller using the switch located at the back The Controller will go through a system initialization This takes a few seconds When complete the Horizon Technology logo will appear on screen along with the software version Step 1 Press the STATUS key to display the status of all three stations See Figure 5 7 Step 2 Press one of the following keys to select the desired station to be run A STA 1 key to select Station 1 B STA 2 key to select Station 2 and C STA 3 key to select Station 3 If using the 1000XL Oil amp Grease Machine select A for Station 1 Step 3 Press the B METH key to increment the method or press the C METH key to decrease the method Holding the key down will allow you to rapidly scroll to the desired method Step 4 Load a 47 mm Disk Holder Cup with the support screen onto the extractor A disk is not necessary as this is only a practice method Attach a collection vessel 40 ml VOA vial
76. of the bottle and uniformly wash down the inside walls If the spray is not uniform contact Horizon Technology Inc Step 6 Once the Purge is complete remove the collection vessel the Water Sample Inlet Valve and sample bottle and the Disk Holder Cup Note It is recommended to run one Purge Method everyday before running actual samples to verify system operation Rev B Pg 26 Note Until all of the air has been removed from the rinse solvent lines minimal and inconsistent volumes of solvent will be collected 5 5 Loading the SPE Disk Step 1 Step 2 Step 3 Step 4 Step 5 Place the metal SPE disk support screen 1 450 into the center of the 47 mm Disk Holder Cup If using the 90 mm Disk Holder Cup use the larger 3 150 diameter support screen Place the SPE disk on the support screen Place the Disk Collar with the groove side facing up into the Disk Holder Cup Firmly screw the Aluminum Locking Ring onto the Disk Holder Cup Check to ensure the Collar tightly holds the disk in place 3M disks Due to the highly porous nature of the 3M Empore Oil and Grease disk a red gasket is required for an airtight seal Failure to use the gasket may dramatically increase sample processing time and compromise recoveries Follow the assembly instructions below when using 3M Empore Oil and Grease disk Load the Disk Holder Assembly onto the first station on the Disk Holder Platform and press down firmly to en
77. oint adapter Any cracks in the cap or damage to the PTFE seal will cause a vacuum leak If the collection vessel is not under vacuum the check valve will not open to elute the rinse solvents If the Check Valve Assembly has been replaced verify that it was correctly installed with the vacuum opening on the check valve aligned with the vacuum opening on the Adapter taper refer to Section 7 12 If no elution is observed and the disk is dry the rinses may not have been sprayed Program a method into the Controller that will have only two rinses one methanol and one hexane Run the method and observe if the rinses are sprayed into the bottle If the rinses were sprayed into the bottle one of the waste drain valves may be stuck open and the rinses are drained to waste Pour some water into the platform opening for the Disk Holder have the vacuum pump on and the unit in idle If one of the waste valves is stuck open the water in the platform opening will drain to waste Observe to which waste bottle the water is drained 14 High Background Background can be introduced from impurities in the solvents disks glassware or the Extractor itself Take approximately 30 ml of each the methanol and hexane solvents directly from the solvent delivery bottles Pour into pre weighed aluminum pans and evaporate to dryness There should be a total of 4 aluminum pans two for the prewet solvents and two for the rinse solvents Record the weights after
78. olished surfaces can cause the internal components to stick if not used routinely To confirm that the Check Valve is free securely attach a sample collection vessel onto the tapered adapter on the bottom side of the Extractor platform see Figures 3 4 and 3 5 Attach the collection vessel by giving a lifting and twisting motion to secure the vessel and ensure a vacuum tight seal Use the blue clip to hold the collection vessel in place Using the Controller keypad go to the Status screen and select STA 1 key A for Station 1 Note Facing the front of the extractor Station 1 is on the left Station 2 is in the middle and Station 3 is on the right Now select Drain key E and then Elute key C This key sequence will activate the Sample Collect vacuum valve for 30 seconds The vacuum pull can be felt by placing a finger over the Check Valve Opening Figure 5 1 or by pouring water into the platform opening the water will be pulled eluted into the collection vessel Press STOP key E to end this test Remove the collection vessel from the Extractor Verify the check valve on every station Rev B Pg 23 If the vacuum is not pulling then the Check Valve Assembly needs to be freed Follow Note Do not use a metal or hard material to push on the ball as this Step 5 1 In the center of the platform below the drain hole Step 5 2 Using the Check Valve Release Tool Figure 5 2 Step 5 3 Repeat Step 5 to confirm a vacuum pull
79. own on disk Figure 5 13 The 90 mm Disk Holder Collar Side with lip Step 5 If using a 47 mm disk place the thin edge or lip side of the Collar see Figure 5 14 face down in the Cup such that it sits on the gasket The side with the flat edge is face up in the Disk Holder Cup Thin edged side faces down on the disk Flat edged side faces up in the k Figure 5 14 47 mm Disk Holder Collar Rev B Pg 28 Step 6 Once the screen gasket disk and Collar are in place the Aluminum Locking Ring is tightly screwed onto the assembly see Figures 5 15 and 5 16 Figure 5 15 Disk Holder Assembly with Figure 5 16 Disk Holder Assembly Riser in Place ready for use The red gasket is only used with the 3M Empore Oil amp Grease disks to provide a good tight seal and optimize sample flow rate If using other disks the red gasket is not used The support screen is placed in the center recess of the Disk Holder Cup and the disk is then placed over the support screen The Collar would now be flipped over and inserted with the opposite side of that shown above in Steps 4 and 5 face down on the disk That is the grooved side of the 90 mm Collar would be face up and the side with the lip would face down holding the disk in place For the 47 mm Collar the flat side would now be placed face down on the disk Instructions if using the JT Baker Speedisk Step 1 Use the white circular adapter for the JT Baker disk Pres
80. pair of needle nose pliers into the opening of the line Gently expand the opening just enough to allow the stem to fit 4 Note Damage to the tubing may occur if over expanded If this occurs use a sharp clean razor blade to cut the end of the draw up line off and re expand the opening 2 Screw the solvent draw up line to the solvent bottle cap and insert the end of the solvent line into the solvent bottle Cap the solvent bottle tightly 3 Repeat the procedure for every solvent bottle Other steps to take e Purchase GC HPLC grade Solvents or Better See Specific Method Requirements e Filter solvents prior to use Rev B Pg 46 e Rinse the solvent delivery bottle with some solvent before using e When refilling the solvent bottles do not let the filters come into contact with a dirty surface Place them on a clean paper towel If the solvent lines are disconnected from the extractor place them on a clean surface protecting them from dirt and dust Note Horizon Technology Inc will not be responsible for damage to valves that have not been protected by the Inlet Solvent Filters even during the normal warranty period 7 7 Adjusting the Water Sample Holder Arm The Water Sample Holder Arm may become misaligned from the weight of the sample bottles Refer to Figure 7 12 When this happens the bottle with the attached Inlet Valve does not sit in a stable position on the Holder Arm This can cause problems with opening
81. played Once the Pause key is pressed the B and C keys can be pressed to increase or decrease the time of solvent dispensing which relates to the volume of n Hexane dispensed If necessary the operation can be resumed to continue this sequence When satisfied with the amount of solvent dispensed press the E key to advance to the next step The next step is to Saturate Hexane into the SPE disk This action allows a small amount of n Hexane to be pulled into the pore spaces of the SPE disk Typically only a 1 second saturation time is needed Press the A key to Resume followed by pressing the A key to Pause the operation As previously mentioned pressing the B or C keys allow you to increase or decrease the times Press the E key to advance to the next step The next step is to Soak Hexane into the disk This step is to ensure the disk is cleaned from any manufacturing impurities Soak Hexane times are typically 1 min to 1 30 min It is possible to Pause the system and press the B key to quickly advance the timer to the desired time Press the E key to advance to the next step The next step is to Drain Hexane from the disk This step will open the solvent waste vacuum valve and pull the n Hexane through the disk and out to the safety coated Solvent Waste Recovery Bottle Typical times for this operation are 1 min to 1 30 min Use the Pause key to pause the system and use the B key to quickly advance the timer to the desired time Press the E key
82. portion of the valve housing and press into place See Figure 7 4 Step 6 Position the bottom portion of the valve housing into SA position over the ball valve and squeeze the housing together Figure 7 5 Ball Valve and gasket sealing Step 7 Secure the assembly with the four screws Step 8 Rotate the valve with your hand and if necessary loosen the screws The ball should rotate freely and not bind up when trying to open and close the valve If not tightened sufficiently leaks can occur resulting in sample overflows Rev B Pg 43 7 4 Replacement of Solvent Rod Shaft s and O rings The Water Inlet Valve has two Solvent Rod Shafts one located on each side at the bottom of the Inlet Valve The Solvent Rod Shaft on the right side of the Inlet Valve see Figure 7 6 is used as the solvent port to deliver rinse solvents to Actuator Shaft the sample bottle and sits in the Rinse Port opening on the Sample Bottle Holder Arm The Solvent Rod Shaft on the left side of the Inlet Valve serves to balance and stabilize the valve on the Holder Arm The o rings on the Solvent Rod Shafts will wear over time and with exposure to solvents When this happens the Solvent Rod Shaft will not give a tight snug fit into the Rinse Port opening on the Sample Holder Arm During the rinse steps solvent will leak from the Holder Arm opening below the Inlet Valve olvent Rod Confirm Proper O Ring Sealing haft amp O ring Rinse Opening Step 1 Lower
83. raction Unlike the rotary vane vacuum pumps the dual diaphragm pump is not susceptible to seizing from the condensation of water and solvent vapors from the extraction Vacuum Gauge T Fitting L Fitting Bleed Valve Figure 1 Dual diaphragm Vacuum Pump with Gauge and Bleed Valve Installing the Dual Diaphragm Oil Free Vacuum Pump OO Ln E amp NN KA See Her NS Note Remove the two orange plugs located on the discharge side of the vacuum pump Thread the 2 mufflers onto these ports Remove the single orange plug located on the intake side of the pump Thread the copper portion of the assembly into the vacuum pump See Figure 1 Connect the gauge to the L fitting of the assembly Connect the bleed valve to the T fitting on the assembly The bleed valve is used to adjust the vacuum level Connect the tubing from the water recovery bottle to the pump Refer to the installation chapter of the User s Manual for installation to the water recovery bottle of the Extractor System Set the pump vacuum level at 25 Hg vacuum gauge reading using the bleed valve to increase or decrease the level It is recommended that all vacuum pumps have a Dry Trap installed HT P N 50 0572 or equivalent to prevent moisture from entering the pump See Tech Tip T35 Installing a Dry Trap Rev B Pg 50 7 11 Installing a Dry Trap A Dry Trap Figure 7 18 is used for two separate applications for the Hor
84. ray solvent into the bottle washing off any residual sample left on the bottle walls The solvent is allowed to drip into the Disk Holder and onto the SPE disk In order to ensure high and consistent recoveries the first solvent used to rinse the bottle and the disk is Methanol Since Methanol is a water soluble solvent it removes residual water trapped within the pore spaces of the disk and prefilter The Methanol rinse is then pulled into the collection vessel The next series of rinses are with n Hexane which is used to extract the Oil amp Grease from the disk These multiple rinses of n Hexane enhance recoveries by thoroughly extracting the Oil amp Grease from the SPE disk Rev B Pg 19 SPE disks offer significant benefits over other forms of extraction techniques such as Liquid Liquid Extractions Several of the benefits include reduced solvent usage no emulsions are formed faster extractions times are possible solvent exposure to workers is reduced and the analysis is operator independent Automation achieves higher and more consistent recoveries by controlling all critical extraction times and vacuum parameters 5 2 The Extraction Cycle To successfully extract Oil amp Grease from samples the SPE DEX automated Oil amp Grease Machines perform four major steps Step 1 Prewet of the SPE disk with solvents Step 2 Introduction of the water sample to the SPE disk Step 3 Solvent rinses of the sample bottle Step 4 Extr
85. reliable operation and customer satisfaction are important to us If the problems cannot be resolved please contact a Horizon Technology technical representative without delay Trouble Shooting Tips 1 The SPE DEX 3000 Controller is not responding Switch the Controller off The on off switch is located on the back of the Controller Verify that the Controller is plugged into an outlet Switch the Controller on 1 depressed If nothing is displayed on the screen verify the connection of the power source to the back of the Extractor A continuous beeping sound is an indication that a key may be stuck 2 The SPE DEX 3000 Extractor is not responding Check the connector from the Extractor to the Controller Make sure it is securely connected at both ends Note Remember to turn the Controller off before connecting or disconnecting the cables 3 Prewet solvents are not dispensed Verify that the solvent bottles are filled and that the solvent pickup tubes in the solvent bottle are long enough Verify that the gas tank is on and the gauge at the source is set between 60 80 psi Check for any air leaks Verify that the solvent bottle caps are sealed and don t have any cracks or damage Also verify that the gas lines from the solvent bottles to the extractor are securely connected and that the gas lines are not crimped or kinked Check for any punctured or damaged gas lines A hissing sound is an indication of a leak
86. rovide a proper seal then the Shaft can be replaced with the other Shaft left on the Inlet Valve Eventually the o rings will wear to a point that a proper seal cannot be achieved If Steps 2 4 confirm that all Solvent Rod Shaft o rings are not sealing properly then replace the o rings PN 30 0544 Solvent Rod Shafts will also eventually wear and require replacement Replace Solvent Rod Shafts that are worn or damaged PN 50 0667 for four PN 50 0667 01 for two Caution Failure to replace worn Solvent Rod Shafts or worn o rings could allow Rinse Solvent to be sprayed from the Assembly 7 5 Sample Bottle Gasket Seals Please follow the instructions below when replacing the Sample Bottle Gaskets Seal on the Water Sample Inlet Valves The Sample Bottle Gasket is designed to provide a leak tight seal between the Sample Bottle or the Bottle Cap Adapter and the Water Sample Inlet Valve A worn or damaged seal could result in leaks between the surfaces When leaks do occur the sample will overflow the Disk Holder Assembly during the sample processing step Inspect the Sample Bottle Gasket Seal Step 1 Step 2 Install the Sample Bottle Gasket Seal Step 3 Step 4 Remove the Sample Bottle Gasket from the Water Sample Inlet Valve This is the circular gasket that sits inside the opening of the Inlet Valve where the sample bottle threads onto Figure 7 9 Carefully remove the gasket using your fingers or with a small o
87. s the narrow end with the black o ring down firmly into the Disk Holder Platform opening Refer to Figures 5 17 amp 5 18 Step 2 Firmly press the Speedisk into the center hole of the adapter a a Figure 5 17 JT Baker Step 3 Use the Liquid Sensor Arm Clip to properly position the tip of Adapter Kit the sensor so that it does not touch the surface of the disk Simply lift the Liquid Sensor Arm and snap the clip onto the stainless steel stem Lift and move the Sensor Arm into the Speedisk Visually confirm that the sensor tip does not touch the surface of the disk Refer to Figure 5 18 Step 4 Remove the Sample Inlet Tube from the Water Sample Inlet Valve The Inlet Tube is pressed into the bottom of gt Sensor clip the Valve Grab with a twisting motion to pull the Inlet Tube out Locate the shorter Inlet Tube and install this ae on the Valve Remember to reinstall the longer Inlet k Tube if the 47 or 90 mm Disk Holder Assemblies are to be used 9 adapter Note Only the manual Speedisk can be used with the Oil amp hi Grease Machines Figure 5 18 Using the Baker Speedisk Rev B Pg 29 5 6 Procedure for Preparing the Extractor to Run a Practice Water Sample In order to gain familiarity with running the SPE DEX Oil amp Grease Machines with a water sample it is recommended to begin by using practice Method 1 This practice method will allow you to observe and understand the operation
88. s is a wearable item that will need replacement after extended use When the o ring is worn it no longer forms a tight seal and will fit loosely into the platform opening This will reduce the flow rate of the sample and can result in the insufficient drying of the disk during the Air Dry step If the disk is not sufficiently dried the extra water will decrease the extraction efficiency and thus negatively impact recoveries To remove the o ring from the Disk Holder Stem use small nose pliers Carefully grip the o ring with the pliers and pull it off Be careful not to score or damage the PTFE base To replace the o ring use the pliers to carefully stretch the o ring enough to be able to place it on the stem To avoid damage to the o ring do not overextend when using the pliers Figure 7 16 Disk Holder O Ring Rev B Pg 48 7 9 Uploading New Software Version Please follow the instructions below when uploading new software into the Controller Windows 98 is required for this upload process Call Horizon Technology if additional technical assistance is needed The most current version software for the SPE DEX 3000 Controller is 1 08 When the Controller is first powered up the initial startup screen with the Horizon Technology logo will display what version is currently in the Controller The changes from the earlier version are as follows s To optimize the flow characteristics of the J T Baker manual Speedisk the Factory Preset
89. sure a tight seal Lift the Liquid Sensor Prewet Arm up and lower it into the Disk Holder Assembly Note the position of the sensor tip It should be a few mm off the surface of the disk If the tip is touching the disk use one of the Sensor Arm clips to raise the height of the sensor If using a 47 mm disk move the Arm to the backside of the Cup Using the Disk Holder Assemblies with 3M Empore Disks The red gaskets are only used with the 3M Empore Oil amp Grease disks to provide a good tight seal and optimize sample flow rates Step 1 Place the support screen in the center of the Disk Holder Cup Figure 5 9 90 mm Disk Holder Cup Figure 5 10 47 mm Disk Holder Cup Step 2 Place the red gasket over the screen as shown in Figure 5 11 below The flat side of the gasket is face down on the screen and the side with the recessed center is face up to accommodate the disk Rev B Pg 27 Figure 5 11 Disk Holder Cup with gasket Step 3 Insert the 3M Empore Oil amp Grease disk in the recessed center of the gasket Adjust the disk around its circumference so that it sits properly in the center of the gasket 4 Figure 5 12 Disk Holder Cup with 3M Empore O amp G disk Step 4 If using a 90 mm disk place the grooved side of the Collar see Figure 5 13 face down in the Cup such that the grooved side sits on the gasket The side with the lip is face up in the Disk Holder Cup A Grooved side _ faces d
90. the threading but verify that it is tight enough to form a good seal Always use the gasket seal on the Water Inlet Valve 9 Sample is not draining to waste or draining very slowly Verify that the vacuum pump is on and working The main vacuum pump gauge should be set between 24 Hg to 36 Hg The vacuum pump requirement is 13 Hg at 1 CFM If running a dirty sample the disk may be clogged Use 90 mm disks for dirty samples and split the sample in two if necessary Consider using prefilters or Filter Aid Verify that the manifold plugs on the waste bottles are securely in place and that all waste lines water to waste and solvent to waste are free of kinks and securely connected The Disk Holder Assembly components need to be tight to form a good seal If the components are not tight air will flow through the parts when vacuum is pulling thus reducing efficiency Always use the red gaskets with the 3M Disks refer to section 5 5 Dirt may be trapped in the Water to Waste Valve Flush the valve by filling a Disk Holder Cup with warm water and using the Drain and then Sample keys for the station in question Rev B Pg 56 The Water to Waste Valve requires air pressure to open Verify that gas is being supplied to the valve Pull the black gas line off the valve fitting and using the DRAIN and SAMPLE functions on the controller activate the valve Air should come hissing out of the gas line 10 Continuous water t
91. the Water Inlet Valve to deliver sample to the Disk Holder Assembly It can also cause the solvent rinse to leak from the Solvent Rod or the rinse spray to be off to the side instead of hitting the center of the base of the bottle To adjust the Sample Holder Arm 1 Remove sample bottle from the Holder Arm 2 To view the Allen Screws located below the hinge lift the TAT A Hyr F 7 12 Misal Holder Arm to place it in its closed position The Allen Screws re rer can be adjusted to position the Holder Arm at a 90 degree angle with the front panel of the Extractor 3 Lower the Holder Arm back down to its open position Place a block or anything that can be used to measure a right angle on the Holder Arm Refer to Figure 7 13 4 Using a 5 64 Allen Wrench adjust the two screws until the side of the block is positioned such that it is completely touching and aligned with the front panel of the Extractor Figures 7 14 i e the Holder Arm is level Figure 7 13 Block on Sample Holder Arm Rev B Pg 47 5 Figure 7 15 shows the correct alignment of the Sample Holder Arm Periodically inspect the alignment of the Sample Holder Arm and adjust if necessary Figure 7 14 Block sitting gure 7 15 Correct at right angle Alignment of Holder Arm 7 8 Replacing the Disk Holder Cup O ring The stem of the Disk Holder Cup has an o ring PN 30 0452 01 to allow a snug tight fit in the platform opening of the Extractor Thi
92. the Water Sample Holder Arm Step 2 Hold the Water Sample Inlet Valve such that the Figure 7 6 Water Inlet aluminum Actuator Shaft points toward you See Figure Valve 7 7 Step 3 Slowly lower the Water Sample Inlet Valve such that the Solvent Rod Shaft can be inserted into the Solvent Rinse Port Step 4 The o ring should have a snug fit once inside the Rinse Port If the fit is snug the o ring seal should be leak tight If the fit is loose or the Water Sample Inlet Valve can be removed without lifting the Holder Arm then the o ring needs to be reversed or replaced Proceed to Step 5 Reversing Replacing the Solvent Rod Shaft Step 5 If there is not a snug fit when the Solvent Rod Shaft is in the Rinse Port then the first action is to reverse the Solvent Rod Shaft Step 6 Firmly grab the bottom of the Solvent Rod Shaft and it straight out of the Water Sample Inlet Valve body There is an upper o ring R on the Shaft so sufficient force will be needed to pull this o ring out of the Water Sample Inlet Valve body Step 7 Reverse the Solvent Rod Shaft and press the end that formerly sealed into the Holder Arm into the Water Sample Inlet Valve body Press the Shaft Figure 7 7 Positioning the Water Inlet Valve Figure 7 8 Removing the Solvent Rod Shaft Rev B Pg 44 Step 8 Step 9 completely into the Inlet Valve Follow Steps 2 4 to confirm a proper o ring seal If reversing the Shaft does not p
93. tion of choice and used for the extraction of Oil amp Grease samples 6 3 Special Function Keys AUXILIARY 1 This screen displays two functions 1 the total number of hours that the liquid sensors have been heated and 2 the total number of samples that have been processed on each station see Figure 6 2 The Controller must be powered up for the hours counter to operate The total hours that each sensor has been heated will help determine wiliary d Screen Systen Gouprter j L ours Samples longevity of the sensors and their useful Ststion 2 Hours pear ees operating lifetimes The number of ER Samples samples processed is determined only when the specific station is in the Process Sample Step and the liquid sensor has been immersed within the water sample for a minimum of two minutes If these two conditions are met the sample counter will increment by one Figure 6 2 Auxiliary 1 Screen System Counters If a sensor needs to be replaced the counters can be reset to zero or the current values can be retained Remember if the counters are not reset the operating hours for the liquid sensor will not be accurate The values for both the hours and the samples processed can be reset or cleared by switching the controller off disconnecting the Controller connector plug from the Extractor and then selecting the A key to CLEAR the values The microprocessor updates this screen on a longer time window therefore