1. Operating condition Dimension w x d x h 125 x 70 x 95 mm line voltage 230 V 10 50 60 Hz MASS 0 5 kg Storage temperature humidity range NS 10 50 C max 75 RH Case resistance to temperature 70 C intrinsic reference eror 24 R 14D Operating eror 15 R 45D Note R Reading D Digit 0 000 1 999 mA 0 001 mA Copying reprinting or any other way of purveying of this User s Manual is allowed with written permission of ILLKO s r o only PL 03 14 R5 Copyright O 2004 2014 ILLKO s r o 11
2. due to transport unsuitable storage conditions etc Explanation of symbols AN Warning concerning a point of danger Read User s Manual and observe all precautions Protection class double insulation i Hint notice Copyright O 2004 2014 ILLKO s r o 2 User s Manual PL 2051 N Notices The test socket 1 of the PL 2051 must not be used for continuous power supply of appliance It is intended for test purposes only max test duration is 30 min 12A For cleaning of the instrument s surface use soft cloth slightly moistened with lukewarm soap water Disconnect all cables connected from the PL 2051 and disconnect the PL 2051 from the Test socket of the PL 2051 before cleaning Do not spill cleaning liquid over the instrument Do not use cleaning liquids based on petrol hydrocarbons etc After cleaning wait until the instrument becomes totally dry before using it Unauthorized persons are not allowed to open the PL 2051 There are no replaceable components inside the instrument 1 2 Applied standards A SAKA dace A EN 61010 1 NS 7 JA ANN EE A EN 55 022 EN 61000 4 2 EN 61000 4 3 EN 61000 4 4 EN 61000 4 5 EN 61000 4 6 EN 61000 4 11 Measurement ri all EN 60601 1 EN 62353 1 3 General description The PL 2051 in conjunction with the measuring instrument ESC 2051 or REVEXplus USB measures following leakage currents According to fig 3 EN 62353 Accor
3. for testing of such DUTs which have electrically electronically controlled switching element and thus it is important to test it in its operating conditions Patient parts e Firstly test earth bond and also insulation resistance if it is possible Then you can execute leakage current measurement e Equipment under test must be isolated from ground from the earth potential during measurement Step 1 Connect mains cable of the ESC 2051 or REVEXplus to grounded mains outlet 230 V Connect mains plug of the PL 2051 to the Test socket of the ESC 205lor REVEXplus Connect mains plug of the DUT to the Test socket of the PL 2051 Connect patient parts of the DUT with Patient terminal of the PL 2051 by the P 2030 cable Step 2 Select fig 4 EN 62353 function by the PL 2051 rotary switch Select Ipe function on the ESC 2051 or REVEXplus e Do not touch exposed conductive parts of the DUT while measurement is in progress risk of electric shock e The DUT is activated during measurement Take into consideration all safety warnings referred to the DUT Press the START key and keep it pressed You can also lock the START key proceed as follows Press the A key and keep it pressed Press the START key Release both keys Both keys are on the ESC 2051 Copyright 2004 2014 ILLKO s r o 8 User s Manual PL 2051 Switch on the mains switch of the DUT The measurement of leakage current
4. is in progress The measured value is displayed on the PL 2051 LED display reading on the ESC 2051 or REVEXplus display has no meaning the measured value is not written to the ESC 2051 memory e Switch off the mains switch of the DUT Release the START key or cancel the locking by pressing of the A key The LED display is turned off e Repeat the measurement with interchanged L N mains terminals of the DUT select Ire inv function on the ESC 2051 or REVEXplus and repeat the measurement described in Step 2 Reconnect all ground connections which was disconnected before the measurement was accomplished 4 4 Leakage current according to fig 7 EN 62353 The leakage current between exposed conductive parts and patient parts of the DUT is measured The test voltage is generated by an auxiliary power supply 230 V The DUT is activated during measurement This measuring method is suitable for testing of such DUTs which have electrically electronically controlled switching element and thus it is important to test it in its operating conditions Patient parts PEO i Test voltage is 230 V short circuit current lt 3 5 mA Step 1 Connect mains cable of the ESC 2051 or REVEXplus to grounded mains outlet 230 V Connect mains plug of the PL 2051 to the Test socket of the ESC 205lor REVEXplus Connect mains plug of the DUT to the Test socket of the PL 2051 Connect patient parts of the DUT wit
5. IPEinv e igo ODOR 0 JO O Fig 2 4 MEASUREMENTS 4 1 Leakage current according to fig 3 EN 62353 Patient parts Copyright O 2004 2014 ILLKO s r o 5 User s Manual PL 2051 Power supply is connected between the patient parts which are short circuited and the short circuited mains terminals of the DUT The current flowing in the circuit is measured The DUT is not activated The short circuit of the mains terminals carries out the PL 2051 This measuring method is suitable for testing of such DUTs which have no electrically electronically controlled switching element If mentioned switching element is in the DUT in fact only the circuitry placed before the switching element is tested see Fig 4 544444 4 Patient parts KI Fig 4 i Test voltage is 230 V 50 Hz short circuit current lt 3 5 mA Step 1 Connect mains cable of the ESC 2051 or REVEXplus to grounded mains outlet 230 V e Connect mains plug of the PL 2051 to the Test socket of the ESC 205lor REVEXplus Connect mains plug of the DUT to the Test socket of the PL 2051 e Connect patient parts of the DUT with Patient terminal of the PL 2051 by the P 2030 cable Step 2 e Switch on the mains switch of the DUT Select fig 3 EN 62353 function by the PL 2051 rotary switch Select Ipe function on the ESC 205lor REVEXplus Do not touch exposed conductive parts of the DUT while measurement is in progres
6. USER S MANUAL PL 2051 ILLKO User s Manual PL 2051 CONTENTS 1 INTRODUC TION wisscsssccisesessssscsssssnssasccstsvcesnsonsodsssasdossisensncaninensessad A AAPP a ii ii 2 1 2 Applied STAM ALG sesi n e 3 1 3 General Deser ip tie tssccsicesscanicesscsasnserersnensecnasatscendsossncrsanbsoonianssannssnsasessonssonsassanniansacsines 3 2 DESCRIPTION OF THE INSTRUMENT 1 ccccccscsccccccsccscccccee4 e di 4 Dl A A 4 3 START UP AAA II al 3 1 Connecting the PL 2051 to the ESC 2051 or REVEXDIUG cssccssessesssssssesseees 4 4 MEASUREMENTS oi 4 1 Leakage current according to fig 3 EN 62353 ccssccssssccscsscsssssesssssecssscsssessssscssees 4 2 Leakage current according to fig 8 EN 62353 ssccssssccssssccscsesscseecssssssccescssssceees 7 4 3 Leakage current according to fig 4 EN 62353 ooccccocuncononnnnnonnncnnananonanaconoooccconocononanos 8 4 4 Leakage current according to fig 7 EN 62353 ccssccsscsccsssscssssesssssecsssscsscesessssseees 9 5 CALIBRATION AND 11 A MAMA iii iii 10 A OLE VICE ammuli mamm sum i 10 6 TECHNICAL SPECIFICATION ccscccosscccsccrccssssssscccceee L Copyright O 2004 2014 ILLKO s r o 1 User s Manual PL 2051 1 INTRODUCTION 1 1 Safety Read this instruction manual carefully and completely and follow all N instructions contained therein Otherwise using of the instrument may be dangerous for the operator f
7. ains outlet 230 V Connect mains plug of the PL 2051 to the Test socket of the ESC 205lor REVEXplus Connect mains plug of the DUT to the Test socket of the PL 2051 Connect patient parts of the DUT with Patient terminal of the PL 2051 by the P 2030 cable Step 2 Switch on the mains switch of the DUT Select fig 8 EN 62353 function by the PL 2051 rotary switch Select Ipe function on the ESC 2051 or REVEXplus Do not touch exposed conductive parts of the DUT while measurement is in progress risk of electric shock Press the START key and keep it pressed You can also lock the START key proceed as follows Press the A key and keep it pressed Press the START key Release both keys Both keys are on the ESC 2051 The measurement of leakage current is in progress The measured value is displayed on the PL 2051 or REVEXplus display reading on the ESC 2051 LCD has no meaning the measured value is not written to the ESC 2051 memory The measurement is finished after the START key is released or locking is cancelled by pressing of the A key The LED display is turned off Copyright 2004 2014 ILLKO s r o 7 User s Manual PL 2051 Switch off the mains switch of the DUT 4 3 Leakage current according to fig 4 EN 62353 The leakage current between exposed conductive parts and patient parts of the DUT is measured The DUT is activated during measurement This measuring method is suitable
8. ding to fig 4 EN 62353 According to fig 7 EN 62353 According to fig 8 EN 62353 The results of measurements are displayed in LED display The measurement methods can be selected by rotary function switch Copyright O 2004 2014 ILLKO s r o 3 User s Manual PL 2051 2 DESCRIPTION OF THE INSTRUMENT 2 1 Front panel 7777 PL 2051 CAT Il IP20 1 2 Fig 1 Legend 1 Test socket 230V 12A 2 Rotary switch 3 LED display 4 Patient terminal 2 2 Included in the set e PL 2051 e Calibration certificate e Instruction manual 3 START UP 3 1 Connecting the PL 2051 to the ESC 2051 or REVEXplus The PL 2051 is determined for measurement in conjunction with the measuring instrument ESC 2051 or REVEXplus only Connect the PL 2051 to mains outlet Observe all precautions related to the PL 2051 mains connecting see the ESC 2051 or REVEXplus User s Manual Connect the PL 2051 to the Test socket of the ESC 2051 or REVEXplus Copyright O 2004 2014 ILLKO s r o 4 User s Manual PL 2051 e Select Ipe function on the ESC 2051 or REVEXplus Press the START key on the ESC 2051 or REVEXplus and keep it pressed LED display on the PL 2051 lights on and value 0 000 0 003 mA is displayed as long as the START key is pressed PRISTUPNA SONDA TEPLOTA ina a PL 2051 je 0 ae 5
9. h Patient terminal of the PL 2051 by the P 2030 cable Step 2 Select fig 7 EN 62353 function by the PL 2051 rotary switch Copyright 2004 2014 ILLKO s r o 9 User s Manual PL 2051 Select Ipe function on the ESC 2051or REVEXplus e Do not touch exposed conductive parts of the DUT while measurement is in progress risk of electric shock e The DUT is activated during measurement Take into consideration all safety warnings referred to the DUT Press the START key and keep it pressed You can also lock the START key proceed as follows Press the A key and keep it pressed Press the START key Release both keys Both keys are on the ESC 2051 Switch on the mains switch of the DUT e The measurement of leakage current is in progress The measured value is displayed on the PL 2051 LED display reading on the ESC 2051 or REVEXplus display has no meaning the measured value is not written to the ESC 2051 memory Switch off the mains switch of the DUT Release the START key or cancel the locking by pressing ofthe A key The LED display is turned off e Repeat the measurement with interchanged L N mains terminals of the DUT select Ire nv function on the ESC 2051or REVEXplus and repeat the measurement described in Step 2 5 CALIBRATION AND SERVICE 5 1 Calibration All measurement instruments should be regularly calibrated We recommend interval of calibration 2 years Furthermore we reco
10. mmend to carry out calibration after each repair 5 2 Service Contact your local distributor for further information Unauthorized persons are not allowed to open the PL 2051 There are no replaceable components inside the instrument 5 3 Ecology Transportation packing It is made of corrugated board it is recyclable Instrument waste at end of life Y This symbol signifies that the product should not be thrown away to municipal Please dispose of this product according to the relevant statutory requirements Copyright 2004 2014 ILLKO s r o 10 User s Manual PL 2051 Producer ILLKO s r o Masarykova 2226 678 01 Blansko ILLKOH sro Czech Republic tel fax 420 516 417 355 e mail illkoGillko cz http www illko cz 6 TECHNICAL SPECIFICATION Protective class II double insulation ambient temperature 0 C 40 C Ovet voltage clas CAT II humidity max 75 RH 23 C Pollution Ce ere v sisscsecieestdecacivesterwevadesand 2 condensation not allowed Degree of protection IP 20 Max altitude ofusing PL 2051 2 000m Reference conditions JEANS aa kaa 07 line voltage 230 V 1 50 60 Hz Impart Enero sa Sae 2J ambient temperature 23 2 C Max current 12 A 30 min humidity 45 55 RH 23 C it depends on tested equipment A arbitrary Ouiescentcurrent circa 50 mA
11. or eguipment under test or for the instrument Observe the following safety precautions Make sure that the instrument measuring cables and all other accessories are in flawless condition e g no damaged insulation no broken cables or plugs etc The PL 2051 is determined for measurement in conjunction with the measuring instrument ESC 2051 or REVEXplus Connect the PL 2051 to the Test socket of the measuring instrument It is prohibited to connect the PL 2051 directly to mains outlet or to a Test socket of any other instrument Only a trained skilled person who is familiar with hazardous voltage operations can handle the PL 2051 Unexpected hazardous voltages can occur at appliance under test dangerously charged capacitors etc If referred to operation of the ESC 2051 or REVEXplus all relevant safety precautions and other information must be observed see the ESC 2051 or REVEXplus User s Manual The user of the PL 2051 must be familiar with the operation of the ESC 2051 or REVEXplus Device under test DUT connected to the Test Socket can be powered by mains voltage during some tests The DUT can thus be activated e It is necessary to respect all safety regulations applicable to particular measurement N The PL 2051 must not be used If it no longer functions properly If the PL 2051 the ESC 205lor REVEXplus cables connectors plugs or other accessories are damaged After extraordinary stresses
12. s risk of electric shock e Press the START key and keep it pressed You can also lock the START key proceed as follows Press the A key and keep it pressed Press the START key Release both keys Both keys are on the ESC 2051 The measurement of leakage current is in progress The measured value is displayed on the PL 2051 LED display reading on the ESC 2051 or REVEXplus display has no meaning the measured value is not written to the ESC 2051 memory The measurement is finished after the START key is released or locking is cancelled by pressing of the A key The LED display is turned off e Switch off the mains switch of the DUT Copyright 2004 2014 ILLKO s r o 6 User s Manual PL 2051 4 2 Leakage current according to fig 8 EN 62353 Power supply is connected between the patient parts which are short circuited and exposed conductive parts of the DUT The current flowing in the circuit is measured The DUT is not activated This measuring method is suitable for testing of such DUTs which have no electrically electronically controlled switching element If mentioned switching element is in the DUT in fact only the circuitry placed before the switching element is tested LO NO PE o n gt Patient parts MD Fig 5 i Test voltage is 230 V short circuit current lt 3 5 m Step 1 Connect mains cable of the ESC 2051 or REVEXplus to grounded m

Download Pdf Manuals

image

Related Search

manual manual manual forklift manualslib manual stacker manual car manuale digitale manual hoist manually meaning manual timesheet manual transmission manual wheelchair manual arts high school manually update your device drivers windows manual definition manual j load calculation manual for courts martial manual labor manual lawn mower manual muscle testing manually register devices with autopilot manual muscle testing grades manual transfer switch manualidades manual blood pressure cuff manual handling

Related Contents

W9458 - Fisher  Bosch NIN-733-V10IPS  BIO-35R-0M10 /BIO-35R-1M10 BIO-35R-0M30 /BIO-35R-1M30  Participation et développement urbain durable  Massive Wall light 55490/29/10  CUADERNO DE MANTENIMIENTO para  Instruction DREES/BES n° 2012-413 du 13  KitchenAid W10681447A Refrigerator User Manual  Alcatel IDOL 2 Mini 6016D 8GB Grey  Avaya Call Center User's Manual  