Home
User Manual
Contents
1. 8 Dissect away the vein using scissors to cut the adipose and connective tissue between the artery and vein One method is to cut the vein in one position and af terwards gently to pull the vein away from the artery In this way a fine membrane of connective tissue becomes visible between the adipose tissue and the artery Carefully cut the fine membrane to remove the vein and adipose tissue while avoiding any direct contact between the scissor and artery fig 5 4 A B Figure 5 4 Removal of vein Figure 5 5 A B and C Cutting free the artery to be studied DMT Single wire myograph system Model 310A 9 Clean the artery by removing any remaining adipose or connective tissue Gently pull away adipose or connective tissue to make the connective tissue membrane become visible Cut the membrane to remove the tissue 10 Cut the distal end of the artery section to be investigated Afterwards cut the proximal end while ensuring that the vessel segment has the correct length fig 5 5A C 5 2 Mounting protocol for small arteries The procedure involves attaching the mounting wires to jaws which are in turn mount ed on the force transducer This force transducer is capable of measuring with a sensitivity of about 0 01 mN 1 mg but can be damaged if the applied force exceeds about 1 N 100 g Therefore care must be taken to avoid pressing the jaws too hard together A movement of 20 um after they have touched is sufficient to hold
2. C Figure 5 9 A Band C Mounting step 4 Mounting step five e Move the jaws apart fig 5 10A Take a second wire holding it about one third down from the far end using a forceps Align the wire parallel with the vessel segment such that the wire can be passed into the far end of the lumen Gently feed the wire through the lumen of the vessel segment in one movement using the first mounted wire as a guide fig 5 10B C Hold the wire at a point at least 10 mm from the vessel to prevent the vessel being stretched during the manoeuvre Be careful not to touch the lumen of the vessel with the end of the wire and when pushing the wire end through the near end of the lumen Once the wire has success fully passed through the lumen of the vessel seg ment place the wire in a position which ensures sufficient length for the wire to be secured both at the near and far fixing screws on the right hand jaw Figure 5 10 A B and C Mounting step 5 User manual 310A Chapter 5 33 Mounting step six e Carefully move the jaws together while ensuring that the second mounted wire lies underneath the first one secured on the left hand jaw fig 5 11A The proce dure clamps the second wire to prevent it from damag ing the vessel segment when securing the wire to the right hand jaw connected to the transducer Secure the near end of the wire in a clockwise direction under the far fixing screw on the right hand jaw fig 5 11B B
3. SA iat eae R EA ts Ne SSR E LENS REESE SVS ein Four allen screws for fine alignment of the myograph jaws Pipe for filling the myograph end eae chamber using the 40 mm ee Pine Peas Supply Microm etere funnel Suction pipe for connection TO vacuum pump Myograph jaw connected es s amp Myograph jaw connected TO micrometef vst to force transducer NS Possseerssseeereseeesee Force transducer pin SEES e Window at the bottom of the myograph chamber for imaging 12 DMT Single wire myograph system Model 310A Chapter 2 Setting up 2 1 The complete myograph 310A system PC data acquisition and analysis Power supply PowerLab data acquisition system software optional Myo Interface front panel optional _ pm D D gt eE obo Myo Interface rear panel pH meter optional Myograph unit Vacuum pump optional l O Electronic valve optional Power supply Suction bottle ere Optional Figure 2 1 The complete single wire myograph system model 310A 2 2 Setting up step by step The chapter contains a complete step by step description of how to set up a complete Myograph 310A System as illustrated in fig 2 1 1 Myograph unit Myo Interface connection Connect the myograph 310A unit to the Myo Interface using the grey 25 pin con nection cable The end of the cable with the temperature probe is used with the myograph unit 2 Myo Interface
4. This equipment generates uses and can radiate radio frequency energy and if not installed and used in accordance with the instructions may cause harmful interference to radio communications However there is no guarantee that interference will not occur in a particular installation If this equipment does cause harmful interference to radio or television reception which can be determined by monitoring the interference while turning the equipment off and on the user is encouraged to correct the interference by one or more of the following measures e Reorient or relocate the receiving antenna e Increase the separation between the equipment and receiver e Connect the equipment into an outlet on a circuit different to that which the receiver is connected to e Consult the dealer or an experienced radio TV technician for help Approvals Complies with the EMC standards EMC 89 336 EEC EN 61326 2 6 2005 EN 61000 3 2 Certified with the safety standards Directive 2006 95 EC EN 61010 1 2001 EN 61010 1 Corr 1 2003 EN 61010 1 Corr 1 2003 EN 61010 2 101 2003 DMT Single wire myograph system Model 310A Certificate of Conformity DMT A S Skejbyparken 152 8200 Aarhus N Denmark hereby declares its responsibility that the following product Single Wire Myograph System Model 310A Version 3 3 is covered by this certificate and marked with CE label conforms with the following standards EN 61010 1 2001 Safety requirements f
5. functions Numeric keys Enter data or setting values Enter Confirm entered numeric values or change between available choices Esc Delete incorrectly entered numeric values In some submenus used to return to main menu A Y Scroll up and down through the display as only four lines are capable of being displayed at a time Being in the top line of a submenu use A to change to the previous main menu The active line in the menu is indicated by a gt symbol Main menu The Main menu displays the current values from the force transducer the actual temperature probe reading and heating setting in C the heating status as well as the pH probe reading optional Temperature To change the temperature setting move the Set Temp line to the top of the display the gt symbol is displayed in that line Use the numeric keys to enter a new temperature setting and press Enter to confirm Heating To turn on the myograph heating move the Heat is line to the top of the display and use the Enter key to switch between on and off Press the Info key and the display shows the F key options associated with the Main menu Choose one of the F keys to proceed with the following options or press Info again to return to the Main menu F1 Change to the next main menu Recorder F2 Opens the Valve Activation submenu An optional electronic vacuum valve is available for the system which can be activated here F3 Opens the Zero Force submenu press F1 to z
6. is why it is necessary to first induce a contraction to be able to observe the relaxa tion to acetylcholine In this protocol the contraction is induced by noradrenaline The required concentration of noradrenaline needs to be optimized since a too low concentration makes it impossible to evaluate the relaxation On the other hand it may be difficult to relax super maximally contracted arteries which may lead to an un derestimation of the sensitivity to acetylcholine Therefore it is recommended to apply a concentration of noradrenaline inducing 60 70 of maximal contraction response In practice this concentration is found by performing a noradrenaline concentration response curve as described in the previous section The vessel segment is exposed to the noradrenaline concentration and when the response has stabilised increasing concentrations of acetylcholine are added to relax the vessel Each concentration is applied until a steady response has been reached and then the next concentration is applied When the vessel segment is either fully relaxed or does not relax more upon increasing the acetylcholine concentration the experiment is ended User manual 310A Chapter 5 39 40 Protocol Prepare the following stock solutions e Acetylcholine 10 10 107M e Noradrenaline 107M 1 Mount and normalize the vessels as described in Chapter 5 2 Perform a standard start and check the vessel segment for endothelium func tion as descri
7. 5 Endothelium function 1 sssssseseeeee essere kt Et ELSE EL ERE REELLE REESE ERE NEL SERENE EL EEN E ERE REELLE RENS TEE 37 5 6 In vitro experiment 1 Noradrenaline contractile reSPONSE cccccssssssssssssersencussnsenecserssseas 38 5 7 In vitro experiment 2 Acetylcholine relaxation CUIVE s cscccceeessseeeeceeeaaseeeeseeeeaseeeeeesseaes 39 Appendix 1 Terms Of warranty areannvnannnannnnnnnnnnnnnennnennnennnennnennnennnennnennnnnnnnnn 43 Appendix 2 Service check nrnnnnnnnnnnnnnnnnnnnnnnnnnnnnennnnnnennnennnnnnennnnnnennennnennnnnnnnn 4A Appendix 3 Shipping INStructiOns ccsccsscssscesecssenssensenssnsenssensenssensenssens 45 Appendix 4 Myograph accessories and spare parts ssscssscssscnseessensees 46 A4 1 General myograph ET EN Lae 46 A4 2 Myograph 310A system accessories ccccccssssseceecceeeeseeeeeceeanaeeeeeceeanaeeeeseesaaeeeeeseesasseeseeees 47 24 3 Myograph 310A SYSTEM spare parts siimcinsiniacccsssavannaneratesviennatensanasienieanuanacsanenteamranstasbtdveenns 47 Appendix 5 Fuse replacement nnnannnnnnannnnnnannnnnnannnnnnnnennnnnnnnnnnnnennnnnnennnennennr 48 Appendix 6 Normalization theory r sarvannnavvnnnnannnnnnannnnnnnnnnnnnnnennnnnnennnnnnennn 49 Appendix 7 Calibration of eyepiece reticule suser 51 Appendix 8 How to read a millimeter micrometer nnrrnnnvnnnennnennvennunnnnenn 53 Appendix 9 System specificatiOnS ar annnnannnnnnnnnnnnnnnnnnnnnnnnn
8. 6127800 e Chamber cover DMT Item CC 61700 40 mm funnels DMT Item F 2000 User manual 310A Appendix 4 47 Appendix 5 Fuse replacement The main fuse of the myograph system is placed inside the power inlet on the Myo In terface If the fuse blows it is easily changed using the following procedure When a fuse blows and needs to be changed it is imperative that the replacement fuse is equal to the one blown The 310A system uses T1 6A 250 V 6 3 x 32 mm DMT recommends that both fuses in the fuse block are changed at the same time as it can be difficult to determine which fuse is blown To replace the fuses Use a small screwdriver to open the voltage selector block Remove the red fuse block Remove the existing fuses Insert the new fuses Replace the fuse block back into the voltage selector block Note ensure that the correct voltage for your country is displayed al cde ak te Voltage selector block Red Fuse Block 48 DMT Single wire myograph system Model 310A Appendix 6 Normalization theory The importance of making a normalization before initiating an experiment with any tubular tissue segment is described in Chapter 5 In this appendix the mathematical rationale and calculations underlying the normalization procedure are described in detail Mathematical calculations Let X Y be the pair of values representing the micrometer reading and force read ing respectively characteris
9. A7 1 Figure A7 1 Horizontal eyepiece reticule scale The basic principle is to use the eyepiece reticule typically consisting of 50 100 divi sions to measure the length of an object in terms of reticular divisions spanned by the object Having the conversion factor specific for the eyepiece reticule and used magnification the length of the object in mm is easily calculated All reticules need to be calibrated in order to determine the specific conversion factor characteristic for the particular eyepiece reticule and the used magnification For such purpose a special devise called a stage micrometer is needed A stage microm eter is simply a microscope glass slide having a scale engraved on the surface A typi cal micrometer scale is 2 00 mm long engraved with divisions of 0 01 mm equalling 10 um per division However micrometer glass slides with less fine divisions are also useful for calibrating a stereomicroscope to be used with the wire myograph Ocular calibration procedure 1 Decide which microscope magnification is to be used for the segment length measurements Use the largest possible fixed magnification where the eyepiece reticule scale still covers the whole gap of the myograph jaws 2 Place the stage micrometer on the microscope stage and focus on it Fit one of the division lines on the stage micrometer to one of the division lines of the reticule scale very precisely While keeping the stage micrometer absolutely fixed
10. Figure 5 11 AandB Mounting step 6 Mounting step seven e Secure the far end of the wire under the near fixing screw on the right hand jaw Again the wire is passed clockwise around the screw stretching the wire as the screw is tightened fig 5 12A B Move the jaws apart to slightly stretch the vessel segment Make sure that the vessel on the far side of the jaws does not extend beyond the jaws as even a small extension will af fect the normalisation procedure In case of excess of vessel on the far side of the jaws then move the jaws together again and remove excessive tissue using a forceps as described in mounting step four A better method for the skilled operator is to move the jaws slightly apart and use scissors to make a small slit in the vessel wall where the vessel is clamped B Figure 5 12 A and B Mounting step 7 DMT Single wire myograph system Model 310A 5 3 Normalization The importance of normalizing the preparation is three fold 1 Experiments with elastic preparations like vessels can only have meaning if they are performed under conditions where the size is clearly defined 2 Clearly defined conditions are required in pharmacological experiments as the sensitivity of preparations to agonists and antagonists is dependent on the amount of stretch 3 The active response of a preparation is dependent on the extent of stretch which makes it important to set the preparation to an internal circumference giv
11. PC connection Data acquisition is possible either by connecting the Myo Interface directly to a PC or through a PowerLab data acquisition and analysis system optional Direct PC Connection Connect the Myo Interface via the RS 232 port to one of the COM ports on the PC using a serial cable Note that the 310A Myo Interface has baud rates of 9600 and 19200 available entered in the Setup menu User manual 310A Chapter 2 13 14 Il PowerLab optional Connect the Myo Interface to the PowerLab unit using BNC cables Connect Rec 1 on the Myo Interface to Input 1 on the PowerLab Rec 2 to Input 2 and so forth Connect the PowerLab unit to one of the USB ports on the PC using the USB cable delivered with the PowerLab system Follow the DMT Quick Guide instructions delivered with the PowerLab system to install the PowerLab driver and Chart soft ware on the PC 3 Gas supply Connect the small pipes on the myograph chamber cover to an adjustable gas Supply using small rubber tubes internal diameter 1 5 mm 4 Suction connection Connect the large pipes on the myograph chamber cover to a vacuum pump via a suction bottle and the vacuum valve as illustrated in fig 2 1 The internal diam eter of rubber tubes for connection to the pipes on the chamber cover is 2 mm The electronic vacuum valve is optional 5 pH electrode optional Connect the pH electrode to the pH port on the backside of the Myo Interface and inst
12. baths Driven by our global customer base our number one goal is to develop and manufacture first class research equipment within the fields of physiology and pharmacology DMT Single wire myograph system Model 310A Safety The Single Wire Myograph System has been designed for use only in teaching and research applications It is not intended for clinical or critical life care use and should never be used for these purposes nor for the prevention diagnosis curing treatment or alleviation of disease injury or handicap e Do not open the unit the internal electronics pose a risk of electric shock e Do not use this apparatus near water e To reduce the risk of fire or electric shock do not expose this apparatus to rain or moisture Objects filled with liquids should not be placed on the apparatus e Do not block any ventilation openings Install in accordance with the manufacturer s instructions e Do not install near any heat sources such as radiators heat registers stoves or other apparatus that produce heat e Only use attachments and accessories specified by the manufacturer e Unplug this apparatus during lightning storms or when unused for long periods of time e This apparatus must be grounded e Use a three wire grounding type cord similar to the one supplied with the product e Do not defeat the safety purpose of the polarized or grounding type plug A polarized plug has two flat blades one being wider than th
13. eee SING 2 T 1 F sing where r is the length of the pan arm F is the force acting on the pan arm when applying the weight F is equal acceleration of gravity times the mass of the weight r is the length of the transducer arm and F is the force acting on the force trans ducer The net torque acting at center of gravity is constant for the weight calibration setup which makes equation 1 and 2 equal making it possible to calculate the force acting on the force transducer ra F sinb ray M eaa 8 sinG FL Von ig 8 sind on sin As the length of the pan arm is 2 cm the length of the transducer arm is 4 cm the weight is 2 g both angles are 900 and the acceleration of gravity is 9 81 ms the force acting on the force transducer is F 2 cm 2 gram 9 81 ms sin 90 4 cm sin 90 ES 9 81 gram m s As 1N is equal 1 kg m s gt F is equal to F 9 81 mN Weight calibration procedure This section contains a complete step by step description of how to weight calibrate the force transducers 1 Fill the myograph chamber with double distilled water and move the jaws apart Mount a 40 um stainless steel wire on the jaw connected to the force transduc ers 2 Place the calibration bridge balance and weight on the myograph unit so that the calibration kit is warmed up together with the myograph unit Turn on the heating in the Main menu on the Myo Interface 3 A
14. kPa IC T 100 mmHg 100 mmHg T User manual 310A Appendix 6 49 50 An exponential curve is fitted to the internal circumference pressure data as illus trated in fig 5 14 in Chapter 5 Now the isobar corresponding to 100 mmHg is used to calculate the IC value from the point of interception between the function of the exponential curve and the function of the 100 mmHg isobar The normalized internal circumference IC is calculated by multiplying the internal circumference corresponding to 100 mmHg IC by a factor k The factor is for rat mesenteric arteries 0 9 Again this value should be optimized for the particular tissue preparation being used by a length tension curve IC k IC The normalized internal lumen diameter is then calculated by j J 1 TU The micrometer reading X at which the internal circumference of the normalized ves sel is set to is calculated by IC IC 2 DMT Single wire myograph system Model 310A Appendix 7 Calibration of eyepiece reticule Principles of ocular calibration The purpose of calibrating the eyepiece reticule is to determine a conversion factor 6 allowing the microscope to be used for measuring vessel seg ment lengths mounted in the wire myograph D 4 o bh 60 0 30 iw 50 Several types of eyepiece reticules are available for such a purpose The most simple and yet very useful type is a horizontal scale as illustrated in fig
15. must be replaced In this instance or in case of other problems related to the force transducer please contact DMT for further instruction and advice 4 4 Force transducer replacement If the force transducer is broken and needs to be changed please follow this step by step replacement procedure carefully 1 Disconnect the myograph unit from the Myo Interface grey cable 2 Turn the myograph unit upside down and remove the bottom plate by loosening the screws fig 4 7 3 Unscrew the two screws on the side of the myograph fig 4 8 4 Once the side block is removed you have access to the underlying screws which hold the transducer in place fig 4 8 Remove these screws and then carefully disconnect the force transducer plug and remove the old transducer Note how the plug is connected to the old force transducer to prevent incorrect connection of the new force transducer 5 Remove any remaining grease from the transducer pin left inside of the trans ducer compartment of the myograph unit Also clean the hole leading from the transducer compartment to the myograph chamber 6 Plug in the new replacement transducer and screw it into place then screw the side block back in place 7 Place some high vacuum grease supplied with the system around the transduc er pin in the myograph chamber see fig 4 9 Make sure that the hole is com pletely sealed so that absolutely no buffer solution is able to enter the transducer compart
16. of interest Start cutting through the mesenteric membrane along both sides of the vessel about 1 2 mm from the vessel To avoid accidentally cutting the artery always cut along the length of the vessels and never perpendicular to them fig 5 2A B 7 Dissect away as much adipose tissue as needed around the vessels to distin guish between the artery and vein The artery can easily be identified by the fol lowing characteristics fig 5 3 28 DMT Single wire myograph system Model 310A Second Branch Figure 5 1 Branching of the mesenteric arteries Figure 5 2 Removal of adipose tissue around the area of interest Figure 5 3 Distinguishing between artery and vein User manual 310A Chapter 5 29 30 e The branch points of arteries are V shaped whereas those of veins are more U shaped e The arterial wall contains a thick layer of smooth muscle cells compared to the vein wall which only contains a single or a few layers of smooth muscle cells The histological difference is clearly visible in the stereomicroscope e If you still have difficulty and the vein and artery still contain some blood then try to move the blood forward by very gently squeezing the vessels with a for ceps In the artery the blood will run back quickly whereas in the vein the blood will run back very slowly if it even does so Note it is important that you per form this on vessels other than those you will use as this procedure damages the vessels
17. spare parts This section contains a complete register of equipment needed to set up a Single Wire Myograph System In addition the chapter contains a list of special myograph 310A accessories and spare parts Besides the main focus on development and manufacturing DMT has specialised in offering our customers first class laboratory equipment needed for a Single Wire Myograph setup at very competitive prices Please contact DMT Sales Department for further product information and prices A4 1 General myograph equipment This section contains a complete checklist of laboratory equipment needed when set ting up a Single Wire Myograph system e Dissection stereo microscope including ocular micrometer and stage micrometer DMT recommends the Zeiss Stemi 2000 Stereo Microscope e Mounting forceps DMT Item DF 3000 DMT recommends Dumont Medical No 5 tip 0 10 mm x 0 06 mm e Dissection scissors DMT Item DS 1000 DMT recommends Geuder G 19745 8 cm straight trabeculum e Pipettes DMT recommends CappAerou pipettes e Light source DMT recommends Schott Cold Light Source either Model KL 200 or Model KL 1500 LCD e Water bath including heater DMT recommends Julabo 5 L open bath circulator with plexiglass bath tank for internal temperature applications e Glass bottle 2 L e Vacuum pump DMT recommends a membrane vacuum pump with a volume of at least 6 L min e Dissection petri dish DMT Item PD 2000 DMT recommends a 9 cm glas
18. 00 200 300 490 500 600 700 800 900 Internal Circumference pm Time min Figure 5 13 Illustration of the stepwise Figure 5 14 Illustration of the expo normalization procedure nential curve fitting and determina tion of IC oo User manual 310A Chapter 5 35 5 4 Standard start The purpose of performing a standard start is to 1 Re activate the mechanical and functional properties of the vessel segment 2 Check that responses to different types of stimuli are normal in appearance and thereby ensuring that the functionality of the vessel segment has not been dam aged during the dissection or mounting procedures 3 Ensure that the tension development gives an effective active pressure that is above the chosen accepted value usually 13 3 kPa 100 mmHg The standard start is performed after the vessel segment has been heated equili brated and normalized The present procedure is suitable for rat mesenteric arter ies Another procedure may be needed for other animal species and tissue or vessel types Principles of the standard start procedure The standard start procedure consists of a series of five stimuli and washout periods The first two stimuli are performed using a mixture of KPSS and 10 uM noradrena line to give a maximum contractile response The third stimulus is performed using a mixture of PSS and 10 uM noradrenaline to give a maximum pure agonist mediated a adrenoceptor contraction The fourth stimulus is performed
19. Important e DMT recommends use of the high vacuum grease once a week to seal up the transducer hole by frequently use e DMT takes no responsibilities for the use of any other kinds of high vacuum grease than the one to be purchased from DMT e DMT takes no responsibilities for any kind of damage applied to the force trans ducer Linear slides Check the linear slides under the black covers for grease at least once a week In case of insufficient lubrication grease the slides with the original enclosed Grease for Linear Slides at the places marked by the arrows in fig 4 10 Figure 4 9 Transducer pin hole to be sealed up with high vacuum grease seen inside the chamber left and in the transducer house right User manual 310A Chapter 4 25 Figure 4 10 Greasing points on the linear slides Myograph cleaning DMT strongly recommends that the myograph chamber and surroundings are cleaned after each experiment At the end of the experiment use the following procedure to clean the myograph chamber and supports 1 Fill the myograph chamber to the edge with an 8 acetic acid solution and allow it to work for a few minutes to dissolve calcium deposits and other salt build up Use a swab stick to mechanically clean all chamber surfaces 2 Remove the acetic acid and wash the myograph chamber and supports several times with double distilled water 3 If any kind of hydrophobic reagent have been used which might be difficult to re
20. Single Myograph System Model 310A User Manual Version 3 3 DANISA MYO T HNunNOoLULU0G DrNnISsH MES TE atimnuNOEOL0G DMT Single wire myograph system Model 310A Single Wire Myograph System model 310A User manual Trademarks PowerLab and LabChart are registered trademarks of ADInstruments Pty Ltd The names of specific recording units such as PowerLab 4 25 are trademarks of ADInstruments Pty Ltd Pentium is a registered trademark of the Intel Corporation Windows Windows 95 Windows 98 Windows ME Windows NT Windows 2000 Windows XP and Vista are registered trademarks of Microsoft Corporation All other trademarks are the properties of their respective owners DMT reserves the right to alter specifications as required This document was as far as possible accurate at the time of printing Changes may have been made to the software and hardware it describes since then New information may be supplied separately This documentation is provided with the DMT Single wire myograph system Model 310A v 3 3 Document Number 310A UG3 3A No part of this document may be reproduced by any means without the prior written permission of DMT Copyright 2008 DMT A S M DANISH MYO TECHNOL OG DMT A S DMT Asia DMT USA Inc Skejbyparken 152 Everwin Gardens 1201 Peachtree Street DK 8200 Aarhus N Rm 502 Block B 400 Colony Square Suite 200 630 Denmark 521 Wanping Nan Lu Atlanta GA 30361 Shangh
21. The purpose of the present protocol is to determine the sensitivity of rat mesenteric small arteries to the vasoconstrictor noradrenaline norepinephrine with a cumulative concentration response curve Background Noradrenaline norepinephrine causes contraction of mesenteric small arteries through activation of adrenoceptors whereas noradrenaline activation of B adrenoceptors causes vasodilatation As the purpose is to determine the contraction sensitivity to noradrenaline the vasodilatory effect of noradrenaline is eliminated throughout the experiment by the constant presence of the B adrenoceptor antagonist propranolol Rat mesenteric arteries are densely innervated by sympathetic nerves which have a highly efficient reuptake mechanism that removes noradrenaline from the neuromuscular junction The reuptake mechanism will create a concentration gradient between the solution around the vessel segment and the receptors on the smooth muscle To correctly determine the sensitivity to noradrenaline it is necessary to eliminate this concentration gradient by performing the experiment in the presence of cocaine to block the noradrenaline reuptake To determine the sensitivity to noradrenaline the vessel segment is exposed to increasing concentrations of noradrenaline Each concentration is applied until a steady response has been reached and then the next concentration is applied When the vessel segment is fully contracted or does not response mor
22. activated Contrast The actual contrast setting in the display Baudrate Data transmission rate from the Myo Interface to a PC via the RS 232 Option The option line allows access to submenus using specific access codes Use the numeric keys to change the valve delay or display contrast settings and press Enter to confirm Press the Info key and the display shows the F key options available for the Setup menu Press Info again to return to the Setup menu or one of the F keys to proceed with one of these options F1 Press F1 to proceed to the Main menu F4 Calibrate Force or pH Having chosen in the Setup menu which signal to calibrate press F4 to initi ate the calibration process Press Esc to return to the Setup menu if you do not wish to calibrate Force calibration This procedure is described in detail in Chapter 4 pH calibration optional Having chosen to calibrate the pH settings press F4 to initiate the calibration In the Calibrate pH Offset display the first line shows the value of the offset buffer first buffer solution which is always 7 00 The second line shows the temperature of the buffer solution The temperature is an important parameter in the calibration formula and is obtained automatically by placing the myograph temperature probe in the buffer solution The third line shows the output from the pH probe as raw data from the A D converter Place the pH meter electrode and temperature probe in the offset bu
23. ai 200030 USA China Tel 45 87 411100 Tel 86 0 21 64869685 Tel 1 770 612 8014 Fax 45 87 41 11 O1 Fax 86 0 21 64280591 Fax 1 678 302 7013 www dmt dk www dmt asia com www dmt usa com sales dmt dk sales dmt asia com sales dmt usa com support amp dmt dk Support dmt asia com Ssupport dmt usa com User manual 310A Trademarks Introduction Until the mid 1970s most of the details about the mechanical morphological and pharmacological properties of vascular smooth muscle was obtained from studies on relatively large vessels At that time rat tail arteries were the smallest vessels to be investigated in detail due to limitations in the available in vitro techniques For example studies measuring the contraction force were routinely performed with only one of the mounting wires secured Furthermore relatively large wires LOO 200 um were used which precluded the use of small vessels and the vessel segment had to be directly manipulated with the dissecting equipment causing inevitable mechanical trauma Investigations of smaller vessels were therefore limited to in vivo perfusion experiments and histological examinations In 1976 Professors Mulvany and Halpern described for the first time a new technique that made it possible to investigate highly isometric responses from vessels with internal diameters as small as 100 um The mounting procedure was refined twofold both ends of each mounting wire were secured under tension wit
24. all the software flash update as described in the instruction provided Perform a pH calibration as described at the end of Chapter 3 2 3 The first weight calibration Prior to the shipment of the Single wire myograph 310A system it has gone through two days of continuous testing including a final weight calibration However in order to ensure that the myograph is working at highest performance DMT recommends that a new weight calibration is performed before starting to use the myograph system The weight calibration procedure is described in detail in Chapter 4 DMT Single wire myograph system Model 310A Chapter 3 The Myo Interface 3 1 Turning on the Myo Interface When the 310A Myo Interface is switched on the display shows that NIYOINTERF E the system is starting up and initializing after which the start up Here VERSION 3 30 message depicted to the right is shown 28 02 2008 The display automatically shows the Main menu after several sec onds Otherwise press F1 to proceed to the Main menu When the start up message is active the A and V keys can be used to adjust the display contrast setting 3 2 Menus and submenus General navigation The following controls are used to display the various menus choose varying menu options and change values F1 Change to the next menu F2 F4 Have varying functions depending on the current active menu Info Push this key in any menu to display all specific associated F key
25. ark e 1 CD Dissection and mounting of small vessels in wire myographs DMT Single wire myograph system Model 310A Chapter 1 System overview com 1 1 Myo Interface front panel Up and down arrow keys Scroll up and downe Analog recorder output Connection to exter in the menus nal data acquisition system optional i Myo Interface display e i 3 F keys Have different but specific func Numeric keys Enter data or setting values tions dependent of the active main menu Enter key Confirm entered numeric values or INFO key Push this key in any menu to change between available choices display all specific F key functions EE cosooococococcocosocooooooooooo D Esc key S Delete incorrectly entered numeric values In some submenus used to return to main menu povesstseeseeeenetnerneeeeneenernenseneenetnerneeenetnetnenseernernenseneevernennenseeenein e ON OFF switch prresesvarvevannasennanvasesvanvavasnasvavasnasvavasvaseesasvasesvasnasesvasvasasvassevasvasssvansene e Power connector Ho RKI E REGU ATOR RS 32 pH MYOGRAPH e 6 Regulator port for elec Connection port for tronic valve optional 310A Myograph unit RS 232 port for serial connection to PC Connection port for pH Meter optional USB port ER RE EE EN EE e 12V DC port User manual 310A Chapter 1 11 1 3 Single Wire Myograph Unit Port for connection to Myo Interface Gr
26. aw on the transducer pin and adjust the supports according to the instructions provided earlier in this chapter DMT Single wire myograph system Model 310A 4 6 Myograph maintenance and cleaning The Single Wire Myograph System Model 310A is a very delicate and sophisticated piece of research equipment DMT recommend that the following sections are read carefully and that the instructions are followed at all times Myograph chamber pipes To prevent the pipes from being blocked by buffer salt deposits after an experiment remove the chamber cover from the myograph and turn on the vacuum pump and vacuum valve for about 10 seconds Wait to turn off the oxygen supply until turning off the vacuum pump Wipe off any buffer remaining on the outside of the pipes using a piece of paper Force transducer The force transducer is the most delicate and fragile component of the myograph system Therefore careful handling is necessary One of the jaws in the myograph is connected to the transducer pin The transducer pin enters the myograph chamber through a pinhole in the chamber wall located below the surface level of the buffer see fig 4 9 To prevent the buffer from running into the transducer house the hole is filled with high vacuum grease As a part of daily maintenance it is very important to inspect the greasing of the transducer hole before starting any experiment Insuf ficient greasing causes damage and malfunction of the force transducer
27. bed at the start of this chapter 3 Add noradrenaline to obtain a response around 60 of maximum determined from the previous noradrenaline concentration response curve When the con tractile response is stable add increasing concentrations of acetylcholine to the chamber using Table 6 2 as a guideline Wait for a stable contractile response or a standard time such as two minutes between each application 08 1plof10M gt Po OB tof OSM 8 7Sukof108M 8 P dulof10M gt In calculating the ACh in the myograph chamber the applied volume of ACh is ignored DMT Single wire myograph system Model 310A 5 8 Buffer recipes Physiological saline solution PSS To make 1 L of PSS Solution 1 Chemical NaCl KCl MgSO 7H 0 KH PO Solution 2 Chemical CaCl 2H O Solution 3 Chemical NaHCO EDTA Glucose MW g mol 58 44 74 56 246 48 136 09 MW g mol 147 02 MW g mol 84 01 372 24 198 77 Conc mmol L 118 99 4 69 1 17 1 18 Conc mmol L 2 50 Conc mmol L 25 00 0 03 5 50 Conc g L 6 95 O35 0 29 0 16 Conc g L 0 37 Conc g L 2 10 0 01 1 09 1 Dissolve the chemicals in approximately 100 mL double distilled H O as three individual solutions as described in the table above Gently heat solution 3 to dissolve the EDTA 2 Solution 1 is added to a graduated bottle and the bottle is filled with double distilled HO to a final volume of 500 mL 3 Solut
28. bove the horizontal line represents 1 mm where each fifth line is marked by a longer line and a number which designates the length in mm Each part of division below the horizontal line is placed between each 1 mm mark scale above the horizontal line and represents 0 5 mm Thimble scale The thimble is divided into fifty equal parts and one complete rotation of the thimble is indicated by the smallest division on the sleeve which equals 0 5 mm Each divi sion on the thimble scale is 10 um If the thimble scale falls between two lines then a number between O and 10 um must be approximated Example 1 1 Note that the thimble has stopped at a point beyond 10 on the sleeve indicat ing 10000 um 10 mm 2 Note that there is no mark completely visible between the 10 mm mark and the thimble 3 Read the value on the thimble corresponding to the intersection with the hori zontal line on the sleeve A Reading on sleeve 10000 um B No additional mark visible O um eT cain em C Thimble reading 380 um tal bill Total reading 10380 um soe Example 2 Figure A7 2 Example 1 reading 10380 um 1 Note that the thimble has stopped at a point be yond 16 on the sleeve indicating 16000 um 16 mm 2 Note that this time a mark is visible between the 16 mm mark and the thimble indication 500 um 3 Read the value on the thimble corresponding to the intersection with the horizontal line on the sleeve A Reading on sl
29. carbogen 95 O 5 CO for at least 20 minutes If necessary wait further for the pH of the buffer to reach pH 7 4 High Potassium physiological saline solution KPSS To make 1 L of KPSS Use the recipe for regular PSS but replace the desired concentration of NaCl with KCI For example to make 60 mM KPSS Solution 1 Chemical MW g mol Conc mmol L Conc g L NaCl 58 44 64 86 3 79 KCI 74 56 58 82 4 39 MgSO 7H 0 246 48 1 17 0 29 KH PO 136 09 1 18 0 16 Calcium free physiological saline solution Ca free PSS To make 1 L of Ca free PSS Use the recipe for regular PSS but omit the CaCl and add EGIA to buffer the residual Ca in solution The concentration of EGTA in the PSS should be 1 5 mM to ensure sufficient buffering DMT Single wire myograph system Model 310A Appendix 1 Terms of warranty Warranty DMT A S warrants to the original user that myograph systems manufactured by DMT A S will be free from defects in materials and workmanship for a period of three years after the date of delivery DMT A S will repair or replace any defective part subject to the conditions limitations and exclusions Exclusions Force and pressure transducers separately or part of myograph systems manufac tured by DMT A S are disclaimed from any warranty Limitations This warranty shall not apply to equipment subjected to accidental damage improper use alteration or deterioration Warranty on third party products w
30. ck for service or repair please read the following shipping instructions very carefully Before you start packing the myograph system please remember that you are dealing with very delicate equipment and therefore care must be taken DMT recommends that each part of the myograph system be wrapped individually i e with bubble wrap and placed together in a large box preferable the box you once received the myo graph system in Place the wrapped items in the middle of the box and fill out the Surroundings with chips of expanded polystyrene Important Before closing the box make sure that no enclosed items can be shaken around as transport by road or air from time to time can be quite roughly Address the box to DMT A S Skejbyparken 152 DK 8200 Aarhus N Denmark Make sure that all four sides of the box are marked fragile or similar Make an indi cation on the top of the box that it contains goods returned for repair service Customers outside the EC must further enclose a pro forma invoice stating that the box contains goods being returned for repair or service If arranging transportation through a courier please keep in mind the high value of the myograph system and that a standard insurance provided by the courier in most cases is insufficient to cover damage or loss of the myograph system In most cases an additional insurance coverage is needed User manual 310A Appendix 3 45 Appendix 4 Myograph accessories and
31. der to be feed into the wire into the ves sel fig 5 7A C Be careful not to stretch the vessel segment if the end of the wire catches the vessel wall Mounting step three e Once the vessel segment is threaded onto the wire catch the free end of the wire nearest you with the forceps and move the jaws apart e While controlling the movement of the wire with the forceps use the other forceps to gently pull the ves sel segment along the wire until the area of interest is situated in the gap between the jaws The near end of the vessel segment shall lie about 0 1 mm inside the jaw gap to insure no point of contact fig 5 8A e Still controlling the free wire end with the forceps move the jaws together to clamp the wire and in one movement secure the wire under the near fixing screw on the left hand jaw Again in a clockwise di rection so that tightening the screw also tightens the wire fig 5 8B DMT Single wire myograph system Model 310A Figure 5 7 A B and C Mounting step 2 B Figure 5 8 A and B Mounting step 3 Mounting step four e Using forceps gently rub the vessel segment on the far side of the jaw to separate any excess ves sel segment from the area of interest clamped in the gap between the jaws fig 5 9A Make sure that the vessel segment is separated as close as possible to the jaws The excessive vessel segment is finally dissected free and removed from the myo graph chamber fig 5 9B C
32. e other A grounding type plug has two blades and a third round grounding pin The wide blade or the third prong is provided for your safety If the provided plug does not fit into your outlet consult an electrician for replacement of the obsolete outlet Be advised that different operating voltages require the use of different types of line cord and attachment plugs Check the voltage in your area and use the correct type See the table below Line plug according to standard 110 125 V UL817 and CSA C22 2 No 42 220 230 V CEE 7 page VII SR section 107 2 D1 lEC 83 page C4 240 V BS 1363 of 1984 Specification for 13A fused plugs and switched and unswitched socket outlets Protect the power cord from being walked on or pinched particularly at power plugs and the point where they connect to the apparatus Refer all servicing to qualified service personnel Servicing is required when the apparatus has been damaged in any way such as the power supply cord or plug is damaged liquid has spilled onto or objects have fallen into the apparatus the apparatus has been exposed to rain or moisture does not operate normally or has been dropped User manual 310A Safety EMC EMI This equipment has been tested and found to comply with the limits for a Class B Digital device pursuant to part 15 of the FCC rules These limits are designed to provide reasonable protection against harmful interference in residential installations
33. e to 9 81 mN If the force reading is different from 9 81 mN then try to calibrate the fore transducer once again starting with step 3 After calibrating the myograph carefully remove weight balance and calibration bridge The myograph system is now ready for use Figure 4 6 Illustration of how to fit the balance between the wire and the gap in the support DMT Single wire myograph system Model 310A 4 3 Checking force transducer The myograph force transducer is a strain gauge connected in a Wheatstone bridge The force transducer for each myograph unit is located in a separate compartment transducer house see Chapter 1 3 While this provides some mechanical protection the force transducers are still very vulnerable to applied forces exceeding 1 newton 100 gram or fluid running into the transducer compartment due to insufficient greasing of the transducer pinhole If the force reading continues to be unstable in spite of a recent weight calibration then repeat the weight calibration and note down the relative force reading values shown in the Calibration displays on the Myo Interface e If the value is O or above 6500 then the force transducer is broken and needs to be changed e If the value is between 1 499 or 3001 6250 then contact DMT for further instructions Important If at any time the message OFF is displayed in the Force menu on the Myo Inter face this indicates that the force transducer is broken and
34. e upon increasing the noradrenaline concentration the experiment is ended Protocol Prepare the following stock solutions e Noradrenaline 104 103 10 M e Propranolol 10 M e Cocaine 10 M 1 Mount and normalize the vessels as described in Chapter 5 2 Perform a standard start as described at start of this chapter 3 Incubate the vessel segment in 1 uM propranolol add 5 uL of 10 Mto 5 mL PSS in chamber and 3 uM cocaine add 15 UL of 10 M to 5 mL PSS in chamber for at least 10 minutes 4 Add increasing concentrations of noradrenaline into the bath use Table 6 1 as a guideline Wait for a stable contractile response or a standard time such as 2 minutes between each application DMT Single wire myograph system Model 310A 08 4 lof10M gt 05 4 lof10M gt 3 amp 10 gt In calculating the NA in the myograph chamber the fe volume of no radrenaline is ignored 5 7 In vitro experiment 2 Acetylcholine relaxation curve The purpose of the present protocol is to determine the sensitivity of the endothelium dependent vasodilator acetylcholine in noradrenaline pre contracted rat mesenteric small arteries Background Acetylcholine causes relaxation of rat mesenteric small arteries by activating of mus carinic M3 receptors at the endothelial cell layer leading to release of endothelium derived relaxing factors Rat mesenteric arteries do not show spontaneous tone in the wire myograph which
35. eeve 16000 um B One additional mark visible 500 um C Thimble reading 280 um Figure A7 3 Example 2 Total reading 16780 um reading 16780 um User manual 310A Appendix 8 53 54 Appendix 9 System specifications Vessel size Chamber Chamber material Chamber volume Chamber cover Force range Force resolution Micrometer Weight calibration Heating Temperature range Temperature resolution Temperature probe Output reading Analog output Serial output Voltage gt 60 um Single bath Acid resistant stainless steel Max 10 mL Yes 100 200 mN 0 01 mN Manual precision micrometer Manual Built into chamber independent of superfusion Ambient temp 50 C 0 1 C External Force mN Up to four outputs 1 0V F S for all acquired signals user defined Serial Interface RS232 RS485 100 to 240 VAC auto 50 60Hz DMT Single wire myograph system Model 310A DANISA M10 T aHnNO Ootu ppGt DMT A S Skejbyparken 152 DK 8200 Aarhus N Denmark Tel 45 87 41 11 00 Fax 45 87 41 11 01 www dmt dk sales dmt dk Ssupport dmt dk DMT Asia Everwin Gardens Rm 502 Block B 521 Wanping Nan Lu Shanghai 200030 China Tel 86 0 21 64869685 Fax 86 0 21 64280591 www dmt asia com sales dmt asia com Support dmt asia com DMT USA Inc 1201 Peachtree Street 400 Colony Square Suite 200 630 Atlanta GA 30361 USA Tel 1 770 612 8014 Fax 1 678 302 7013
36. ennnennennnennnennnnnn 54 DMT Single wire myograph system Model 310A About this manual This manual contains a complete list of procedures describing how to install maintain and get started using the Single wire myograph system model 310A version 3 3 Chapter 1 provides an overview of the construction and basic features of the Myo In terface and the Single wire myograph unit Chapter 2 describes step by step how to set up a complete 310A wire myograph sys tem including accessories Chapter 3 is a complete manual to the 310A Myo Interface version 3 3 The chapter contains a step by step description of how to navigate in the menus and how to use the special features of a 310A myograph Chapter 4 contains procedures describing general as well as daily maintenance of the myograph unit e g adjustment of Supports weight calibration of the force trans ducer and cleaning instructions Chapter 5 describes how to get started using the wire myograph system This in cludes a complete dissection and mounting procedure for small mesenteric arteries and a few basic pharmacological experiments Appendices contain additional information about normalisation theory ocular calibra tion myograph service shipping instructions system specifications equipment lists accessories and spare parts and fuse replacement User manual 310A About this manual Unpacking the myograph system Please take a few minutes to carefully inspect your n
37. ero the force or F4 to cancel F4 Displays a condensed Main menu Press F4 to toggle between the condensed and the normal view User manual 310A Chapter 3 16 F1 MAIN MENU F4 F1 Change menu F2 Valve ON F3 Zero force F4 Condensed menu I Activate Valve F3 Yes F1 No F4 F1 RECORDER MENU Rec4hi F4 Reset SETUP Change menu Calibrate Force CALIBRATE FORCE Force 0 15 mN Tr Rec1 FORCE Calibrate FORCE Act temp 36 8 C Rec 1lo 20 Valve dly 5 SeLtemp 3100 Rec l hi 20 Contrast 35 Heat is ON Rec 2 TEMP Baudrate 9600 Rec 2lo 0 Option 0 Rec 2hi 50 Rec 3 No use Rec 3 lo 9 Er Change menu Rees Start timer Rec 4 No use Stop Pause Rec4lo F3 Apply no force Push F3 when stable Force 1715 Zero force Yes Fl No F4 Force 0 15 mN Act temp 36 8 C Settemp 37 0 C Heat ON CALIBRATE FORCE Place 2 gram on pan Push F3 when stable Force 1880 Calibrate Force DONE Push Esc Force zero 17157 Force gain 168 9 Figure 3 1 Menu overview the green panels indicate the menu appearance options when the optional pH meter is installed Recorder menu The 310A Myo Interface has four analog output ports on the front panel for connec tion to a data acquisition system These default settings for these output ports are Rec 1 Myograph force FORCE outp
38. ew wire myograph system for damage which may have occurred during handling and shipping If you Suspect any kind of damage please contact DMT immediately and the matter will be pursued as quickly as possible If the packing material appears damaged please retain it until a possible claim has been settled We recommend that you store the packing material for any possible future transport of the Wire Myograph System In case of transport and the original packing material is unavailable please contact the DMT Sales Department for advice and packing instructions After unpacking your new Wire Myograph System please use the following list to check that the system is complete 1 Myo Interface e Myograph connection cable with a temperature probe e Power cord the shape of the AC plug varies by country be sure that the plug has the right shape for your location 2 Single wire myograph unit e Calibration kit including bridge balance and 2 g weight e Chamber cover e Chamber separator 3 Accessories e 1 roll of stainless steel wire 40 um diameter e 1tube of high vacuum grease e 1 tube of grease for linear slides e 4 spare screws for mounting of jaws e 2 Allen keys e 1small screwdriver 4 Manuals e User manual for Single wire myograph system model 310A e Procedures for investigation of small vessels using small vessel myograph by Professor M J Mulvany Department of Pharmacology Aarhus University Denm
39. ffer solution and turn on stirring When the relative pH output in the bottom line is stable press F3 to proceed The first line now displays the value of the slope buffer second buffer solution which is always 4 00 Place the pH and temperature probes in the slope buffer solution and turn on stirring When the relative pH output in the bottom line is stable press F3 to proceed The pH calibration is now finished The parameters are stored in the internal memory of the Myo Interface Press F3 to return to the Setup menu User manual 310A Chapter 3 17 Chapter 4 The single wire myograph unit This chapter contains a complete explanation of how to adjust calibrate and maintain the 310A myograph unit to ensure the equipment performs to the highest standard 4 1 Adjustment of supports Successful mounting of any kind of tubular tissue segment in the wire myograph depends upon perfectly aligned supports The supports are factory aligned prior to shipment but with daily use of the myograph and repeated greasing of the transducer pinhole the supports will need to be adjusted with time Note The transducers are fragile and sensitive to mechanical strain Be very cau tious not to put strain on the transducer when changing or adjusting the mounting supports In addition very little force should be applied to the screws in order to avoid breaking the threads Adjustment of the supports is performed using the following step by step procedu
40. fter approximately 20 30 minutes the whole system will have reached the target temperature normally 37 C Place the warmed calibration bridge and balance on the myograph unit as illustrated in fig 4 5 4 Make sure that the tip of the transducer arm on the balance is placed in the gap between the wire and the jaw as illustrated in fig 4 6 Carefully move the calibra tion bridge until the tip of the transducer arm is placed freely in the gap which User manual 310A Chapter 4 21 Figure 4 5 Weight calibration setup means it does not touch either the wire or the jaw Note The weight should not be placed on the balance yet Go to the Setup menu on the Myo Interface and choose to calibrate the myo graph force see fig 3 1 Press F4 to start calibration Make sure that absolutely no force is applied on the force transducer by check ing that the tip of the transducer arm is not touching either the wire or jaw Also check that the relative force reading in the display is stable Press F3 to proceed with calibrating Carefully place the 2 g weight on the pan as illustrated in fig 4 6 The force ap plied on the force transducer should mimic the stretch created by the contrac tion of a mounted ring preparation Wait until the relative force reading is stable Press F3 to finish the calibration Press Esc to return to the Setup menu and then press F1 to show the Main menu The force reading on the Myo Interface should now be very clos
41. g a standard start to make sure that the vessel segment is viable The present procedure is for use with rat mesenteric arteries Another procedure may be needed for other animal species and tissue or vessel types Principles of checking endothelium function Stimulating a vessel segment with acetylcholine causes a release of nitric oxide NO also known as EDRF from the endothelium cells and subsequent relaxation of the vascular smooth muscle cells If the endothelium is undamaged by the dissection and mounting procedures then a substantial relaxation will occur With complete removal or damaged endothelium a partial relaxation or no relaxation to acetylcholine is observed It is important to note that the amount of NO or EDRF in a vessel is often dependent upon its size In certain vessels endothelium derived hyperpolarizing factor EDHF can contribute more or less than EDRF and in other vessels the same stimulation with ACh can promote release of endothelium derived contracting factor EDCF Therefore it is important to check the existing literature in order to determine the expected response in your particular vessel with the given concentration of agonist Step 1 PSS 3 uM NA Stimulate for 3 minutes Step 2 Add 10 pl 5 x 10 M ACh Wait 2 minutes Step 3 Wash out 4 x with PSS Ready for experiment User manual 310A Chapter 5 37 5 6 In vitro experiment 1 Noradrenaline contractile response
42. g vascu lature including part of the superior mesenteric artery Be careful not to damage the vasculature during this procedure The proximal end of the intestine section must be about 10 cm from pylorus Make a cut in the proximal end of the intes tine for later identification 3 Place the excised intestine section in a Petri dish about 9 cm in diameter coated with a 5 mm thick layer of Sylgard at the bottom to hold the fixing pins Immedi ately fill the Petri dish with cold PSS which has been well aerated with carbogen The dissection is performed without further oxygenation of the PSS The cold tem perature ensures the artery is relaxed during the mounting procedure 4 Pin down the proximal end of the intestine section on the left hand side of the Petri dish without stretching the vessels Pin down the remaining of the intestine section in an anti clockwise direction In this configuration proximal end at the left side distal end at the right side and running anti clockwise from proximal to distal side the feeding vasculature is on the far side of the intestine and the veins are usually uppermost 5 Select the vessel segment to be investigated fig 5 1 First time myograph users are recommended to start dissecting and mounting vessel segments from the first or second branch from the superior mesenteric artery internal diameter 200 300 um 6 Use high quality forceps and ocular dissection scissors to dissect the vessel segment
43. h a few times with double distilled water Incubate the myograph chamber and supports with an 8 acetic acid solution for 10 minutes and continue the mechanical cleaning with a swab stick 6 Wash the myograph chamber and supports several times with double distilled water Important e In exceptional cases it may be necessary to unmount the supports and clean them and the myograph chamber separately to ensure that all surfaces are cleaned User manual 310A Chapter 4 27 Chapter 5 Getting started Chapter 5 contains some practical protocols and procedures describing how to quickly get started using the Single Wire Myograph System The chapter mostly ad dresses users who are not yet familiar with the wire myograph technique However it may contain some hints that will also be useful for the experienced myograph user 5 1 Dissection protocol for small mesenteric arteries The Single Wire Myograph technique is versatile in that a large variety of physiological and pharmacological studies of ring preparations from different species are possible The wire myograph is predominantly used for investigation of small blood vessels and as an example this chapter describes the dissection of rat mesenteric arteries 1 A laboratory rat is euthanized in accordance to the local national law and regula tions A midline laparotomy is performed to expose the mesenteric bed 2 Use scissors to remove about 10 cm of intestine together with its feedin
44. h unit displaying vertical adjustment V X xX we ew Sew V x x Figure 4 3 Illustration of correctly aligned supports for small vessels left and incor rectly aligned supports middle and right User manual 310A Chapter 4 19 4 2 Force transducer calibration As a part of the general maintenance of the myograph DMT recommends that the myograph is weight calibrated at least once every month DMT also recommends that the myograph is weight calibrated every time the system has been moved or has not been used for a long period of time Principles of weight calibration Weight calibrating the force transducer is based on simple physics the net torque acting on a balance when applying a certain amount of weight The magnitude of the torque T about a point of rotation P is defined by LJ T r F sin where r is the distance from the point of rotation to the point on the object where the force F is acting with the angle of 8 Applying the physics to the weight calibration setup is illustrated in fig 4 4 an OS Kl Transducer Arm TA Figure 4 4 Theoretical principle of the weight calibration 20 DMT Single wire myograph system Model 310A Applying the weight on the pan arm creates a net torque acting at the center of grav ity resulting in a force FT acting on the force transducer The following two equations describe the forces working in the weight calibration system 1 T FE SING ho
45. hout any direct manipulation of the vessel segment Segments of small vessels could now be atraumatically mounted as ring preparations in a myograph for recording of highly isometric force measurements During the late 1970s some improvements were made to the myograph and in 1981 a new dual myograph that allowed simultaneous testing of two vessels was introduced In parallel the technique became widely acknowledged resulting in a growing interest in the myograph systems In 1986 the growing demand resulted in the foundation of the private company J P Trading with the purpose of making the myograph systems commercially available worldwide At the same time J P Trading initiated a comprehensive improvement programme for the existing myograph systems as well as a development programme of new myograph systems in close co operation with Professor M J Mulvany and the University of Aarhus During the late 1980s and through the 1990s several improvements were applied to the myograph systems such as a new mechanical design a more robust transducer and new electronic systems In addition new systems were introduced like the Automatic Dual Myograph 510A the Dual Myograph 410A the Multi Myograph 610M and the Confocal Myograph 120CW In 2000 J P Trading changed its company structure and became known as DMT Today DMT is one of the world s leading designers and manufacturers of wire myographs pressure myographs culture myographs and organ tissue
46. ill be as determined by their respective manufac turer DMT A S shall not be liable for consequential incidental special or other direct or indirect damages resulting from economic loss or property damage sustained by you or any end user from the use of the products sold or services rendered hereunder Warranty returns A Return Material Authorisation RMA number is required for all returns This number should be clearly indicated on all returned myograph systems Products damaged due to improper or inadequate packaging when returned for RMA purposes are not granted warranty coverage User manual 310A Appendix 1 43 44 Appendix 2 Service check A myograph working at optimal performance is extremely important for success when studying small blood vessels or other small tubular tissues To make sure that our customers always are dealing with first class myographs DMT offers a Myograph Service Check at a very favourable price The Myograph Service Check includes a complete separation for inspection of all mechanical and electronic parts The myograph is then reassembled adjusted and finally all electronic and mechanical parts are tested Please note that the service does not include replacement of transducers or any other needed spare parts Please contact DMT for information about prices DMT Single wire myograph system Model 310A Appendix 3 Shipping instructions If the myograph system needs to be sending ba
47. ing maximal response The aim of the normalization procedure is to stretch the segment to a so called normalised internal circumference IC defined as a set fraction of the internal cir cumference IC that a fully relaxed segment would have at a specified transmural pressure For small rat arteries the target transmural pressure is typically LOO mmHg 13 3 kPa Principles of the normalization procedure In practice the normalization is performed by distending the segment stepwise and measuring sets of micrometer and force readings fig 5 13 step 1 4 These data are converted into values of internal circumference um and wall tension T mN mm respectively Plotting wall tension against internal circumference reveals an exponential curve and by applying the isobar curve corresponding to 100 mmHg IC o is calculated from the point of intersection using the Laplace relation fig 5 14 IC is calculated from IC oo by multiplying a factor giving an internal circumference at which the active force production as well as the sensitivity to agonists of the segment is maximal For rat mesenteric arteries the factor is 0 9 but both this factor as well as the transmural pressure has to be optimized for each particular segment The normalised internal diameter is calculated by dividing IC with T Appendix 6 contains a complete description of the mathematical rationale and calcu lations of the normalization procedure Force mN mm 1
48. ing each step in the normalization procedure Y is the force reading at the start position of the normalization procedure where the wires are just separated and the force reading is approximately zero Then given that tension on the vessel is equal to force divided by wall length the wall tension at the i th mi crometer reading is calculated by ee 25e a a where is the microscope eyepiece reticule calibration factor in mm per division and a and a are the vessel end points when measuring the length of the mounted vessel segment The internal circumference of the mounted vessel at the i th reading is calculated by IC IC FN where IC is the internal circumference of the mounted vessel when the wires are just separated and is given by IC 2 7 where d is the wire diameter For 40 um wires IC 205 6 um Using the Laplace relation the effective pressure P is calculated for each pair of readings The effective pressure is an estimate of the internal pressure which is nec essary to extend the vessel to the measured internal circumference I a The stepwise distension is continued until the calculated effective pressure exceeds the target transmural pressure The target value needs to be optimized for the indi vidual tissue preparation optimal active force as determined by the length tension relationship for that tissue For rat mesenteric arteries the target transmural pres sure is normally 100 mmHg 13 3
49. ion 3 is added to the graduated bottle which afterwards is filled with additional double distilled HO to a final volume of about 850 mL 4 Aerate the solution with carbogen 95 O 5 CO for about 20 minutes 5 Solution 2 is added and the graduated bottle is filled with additional double distilled H O to reach the final volume of 1000 mL Continue the carbogen bubbling until the pH of the buffer solution reaches 7 4 25x Concentrated PSS To make 1 L concentrated PSS Solution 1 Chemical NaCl KCl CaCl 2H O Solution 2 Chemical MgSO 7H 0 KH2PO Solution 3 Chemical EDTA MW g mol 58 44 74 56 147 02 MW g mol 246 48 136 09 MW g mol 372 24 Conc mmol L 118 99 4 69 2 50 Conc mmol L 1 17 1 18 Conc mmol L 0 03 User manual 310A Conc g L 173 85 8 75 9 20 Conc g L 7 23 4 02 Conc g L 0 25 Chapter 5 41 42 1 Dissolve the chemicals for solution 1 in about 800 mL double distilled HO in a 1000 mL graduated bottle Dissolve the chemicals for solutions 2 and 3 in 75 mL double distilled HO in individually cylinders Gently heat solution 3 to dissolve the EDTA 2 Solution 2 and 3 is added to solution 1 and the graduated bottle is filled with additional double distilled HO to reach a final volume of 1000 mL Before use 3 Dilute the 25 x PSS stock solution 1 25 with double distilled H O 4 Add 1 091 g L Glucose 2 100 g L NaHCO 5 Aerate the solution with
50. ment and damage the force transducer Important The new force transducer must be weight calibrated prior to running an experiment User manual 310A Chapter 4 23 24 Figure 4 7 Remove the bottom plate by loosening the screws r Figure 4 8 Remove the screws and then carefully disconnect the force transducer plug and remove the old transducer 4 5 Changing the myograph window glass The glass in the myograph chamber window is fixed in place and kept waterproof by a thin layer of high vacuum grease on the circular edge between the glass and the myo graph chamber base The following procedure describes how to change the myograph window glass 1 Screw the jaws as far apart as possible and carefully remove the myograph jaw from the transducer pin side it should not be necessary to remove the jaw on the micrometer side 2 Loosen the glass from the myograph chamber by gently pushing up on the glass from below the window with a blunt tool 3 Remove the old grease and clean the area thoroughly with 96 ethanol 4 Carefully apply a small continuous amount of high vacuum grease around the edge of the window Using forceps place the new window glass in place Push down gently around the edges to create a seal between the glass the grease and the chamber base 5 Check that the new window forms a tight seal by filling the chamber with distilled water If there is a leak repeat the replacement procedure 6 Finally replace the j
51. move using step 1 and 2 then try incubating the chamber and supports with 96 ethanol or a weak detergent solution 4 To remove more resistant or toxic chemicals incubate the myograph chamber and supports with 1 M HCI for up to 1 hour In exceptional cases incubate the chamber and supports with an up to 3 M HNO solution for about 15 minutes 5 Wash the myograph chamber and supports several times with double distilled water Important Notes e Be very careful using step 3 and 4 repeatedly as strong reagents may damage the myograph unit e After cleaning ALWAYS check that the greasing around the transducer pin is suf ficient to keep the buffer solution away from the transducer compartment In cases of red or brown discolorations appearing on the chamber sides or on the supports the following cleaning procedure will work in most cases 1 Incubate the myograph chamber and supports for 30 minutes with 20 uL of a 2 MM T 1210 Tetrakis 2 pyridylmethyl ethylenediamine solution dissolved in double distilled water 2 Use a cotton swab stick to mechanically clean all the affected surfaces during the last 15 minutes of the incubation period 3 Wash the myograph chamber and supports several times with double distilled water 4 Incubate the myograph chamber with 96 ethanol for 10 minutes while continu ing the mechanical cleaning with a swab stick 26 DMT Single wire myograph system Model 310A 5 Remove the ethanol solution and was
52. nnnnnnnnnnnnennnnnnennnennennnennnnnnennnn 13 2 1 The complete Myograph SLOA SYSTEM vvs aaa Eaa 13 2 2 Setting up step by step cceeseeceeescessseeeeeesccssseneeseccesssareesecoasseaseeseceasasasencecessaaesencecusssateeteceas 13 2 Se 14 Chapter 3 The Myo Interface ar nanunaunnannnannnnunnnnnnnnnnnnnnnnnnnnevnnnennnnnnnnnnennnnr 15 3 1 Turning on the Myo Interface sees ELLER LEE ELLE ERE ELLEN 15 PEN 15 Chapter 4 The Single Wire Myograph Unit anrnannnnnnnnnnnnnnnennnnnnennnennennnennen 18 4 1 Adjustment of SUPPOPTS cccccccscssessencuseusecccseueuscuseesencuseusanausereuscusussencusuueacauarenseusussenaueerescauees 18 4 2 NNVF 20 4 3 Checking force transducer vieresecescesiesairereddeascnte exssnnnsinevcncsesewevyees seusveesdtedicbaxsaveetvnentxacaveeasextserese 23 4 4 Force transducer replacement ivvsescicinsevessenecdiavsdacsccsnweuesaensuendavdnseunsinadavenenceedwanecsewesanneneateniotenss 25 4 5 Changing the myograph window PIASS ccsccccccccceeeeeessseeeececeeeeeceeeaeauseeseeeeeeseeseesaaesseeeeeeees 24 4 6 Myograph maintenance and cleaning Harran 25 Chapter 5 Getting started nenannnnennnnnnnnnnennnnnnnnnnnnnnnnnnnnnnnnnnnnnnnennnnnnennnennennnr 28 5 1 Dissection protocol for small mesenteric arterieS snrrrrnnnnnnnnnnnnnnnnnnnnnnnvnnnnnnrnennnnnsennnnnsseennn 28 5 2 M nt ng protocol for small EEE EN 31 5 61411 45 9 EE EE ne EEE OEREO EENEN 35 SSE gels 9 ks EEE TE NE EE EE rene 36 5
53. on the microscope stage find another position on both scales where the division lines also fit precisely Read the position of the two fit points on both scales and fill in the values in the Ocular Calibration Sheet Repeat the procedure twice User manual 310A Appendix 7 51 Ocular calibration sheet Date Operators O Microscope type and Ocular type Magnification Front lens Position 1 Position 2 Length 1 div div mm 2 div div mm 3 div div mm Calculations 1 Calculate the length between the two positions on the stage micrometer by sub tracting the value of position 1 from the value of position 2 Multiply the length in divisions with the length of each division to get the length in mm 2 Calculate the length between the two positions on the reticule scale by subtracting the value of position 1 from the value of position 2 3 Calculate the mean length value of both the stage micrometer and the reticular scale 4 Calculate the conversion factor 5 Mean Stage Micrometer Length mm Mean Reticule Scale div DMT Single wire myograph system Model 310A Appendix 8 How to read a millimeter micrometer Sleeve scale Thimble scale 0 I AN F s Tine PENG Figure A7 1 Overview of the micrometer parts actual reading 20000 um 20 mm Sleeve scale The micrometer sleeve scale has a total length of 25 mm divided into fifty equal parts Each part of division a
54. or electrical equipment for EN61010 1 Corr 1 2003 measurement control and laboratory use EN 61010 1 Corr 1 2003 Part 1 General requirements EN 61010 2 101 2003 Safety requirements for electrical equipment for measurement control and laboratory use Part 2 101 Particular requirements for in vitro diagnostic IVD medical equipment EN 61326 2 6 2005 Electrical equipment for measurement control and laboratory use EMC requirements Part 2 6 Particular requirements In vitro diagnostic IVD medical equipment With reference to regulations in the following direc tives 2006 95 EC 89 336 EEC User manual 310A Certificate of Conformity Contents TEC NAS soa acscine ienatsn cconn ec wenatacesttnaecesnundemsantucganssecuaseeceasayenugesssennavaseuuavecseeavecsaeases 3 OA 4 SERENE E EE E A E EE O re 5 ENG EL AA AA Ea re 6 MOI oeenn raan EAEN A Aaa Neea ENEE 6 Certificate of GonforminLarannsmrmeaaansmniivenkonbinrvtdvraannandnnannnnke 7 About this NAN 9 Unpacking the myograph system snrennvennvnnnvennvnnnvnnnnnnnnnnnnennnennnnnnnnnnnnnnnvnnnvennr 10 Chapter 1 System OvervieW annnannnnnnnnnnnnnnnnnannnnnnannnnnnnnnnnnnnnennennnennnnnnennnnnnnn 11 1 1 Myo Interface front panel isniesisvipeneivvaiveneuenienmrieanvanintumneniccndvangsewensddenbaencedenwiwanivxavensannevst ian 11 1 2 Ose Fa rear HANE EEE NE 11 1 3 Single Wire Myo grab EEE br EF ENRERE 12 Chapter 2 Setting UP nrnarnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnn
55. re illustrated in fig 4 1 1 Carefully loosen screw fig 4 1D on the top of the support connected to the force transducer Align the horizontal Support and carefully tighten the screw again 2 Loosen screw fig 4 1D on the top of the support connected to the linear slide Align the horizontal support matching the force transducer connected support as carefully as possible and gently tighten the screw again 3 Loosen screw fig 4 1C on the linear slide to roughly match the linear slide sup port to the force transducer support in the horizontal plane Tighten the screw before proceeding with step 4 4 The plate fig 4 1B on which the linear slide is mounted is balanced on top of a small stainless steel ball making it possible to finely adjust the linear slide sup port in all vertical and horizontal planes using the four Allen screws fig 4 1A Use the four Allen screws to make the final horizontal fig 4 1 and vertical fig 4 2 adjustments to match the linear slide support to the force transducer sup port The correct matching of the supports is illustrated in fig 4 3 Important Avoid continuously tightening the Allen screws during the final adjustments loosen the Allen screw placed diagonally to the allen screw being tightened otherwise there is a high risk of damaging the myograph frame DMT Single wire myograph system Model 310A Figure 4 1 Myograph unit screws for adjusting supports Figure 4 2 Myograp
56. s Petri dish coated with a 5 mm Sylgard polymer layer DMT Single wire myograph system Model 310A A4 2 Myograph 310A system accessories This section contains a list of special accessories available for the Single wire myo graph system Model 310A e ADI PowerLab data acquisition system Including Chart data acquisition and analysis software e DMT Normalisation module DMT Item MLSO65 Chart add on software et pH meter Including pH electrode and pH meter e Plastic mounting jaws DMT Item PJ 345610 With built in platinum electrodes for electrophysiological experiments e 25 um tungsten wire DMT Item TW 345610 For mounting of ring preparations with internal diameter down to 60 um e Electronic vacuum valve Item VV 9937 e Standard PC system DMT Item 80150 DMT CS 200 4 channel Current stimulators DMT Item CS 100 amp CS 200 Combined pulse and train generator A4 3 Myograph 310A system spare parts This section contains a complete list of standard spare parts available for the myo graph 310A For parts not listed in this Chapter or for special parts which may need to be custom made please contact DMT for further information e Force transducer DMT Item FT 61780 Roll of 40 um stainless steel wire DMT Item 4 SSW 123456 High vacuum grease and grease for linear slides DMT Item HVG 1000 e Stainless steel jaws DMT Item SSJ 345610 e Mounting support pins DMT Item MO 200 e Calibration kit DMT Item CK
57. the wires clamped Mounting step one e Cut lengths of 40 um wire 2 2 cm long Mount one wire on left hand jaw of myograph as follows e Holding wire at far end place centre of wire between Jaws and screw jaws together so that the wire is clamped fig 5 6A Note do not close the jaws too hard against each other See above e Bend the far end of the wire towards the left and wrap it around under fixing screw so the wire is wound clockwise tightening the screw will then tighten the wire This procedure should result in the wire being clamped between the jaws and with near end of wire pointing towards operator fig 5 6B C e Fill the myograph chamber with PSS at room tem perature UJ Figure 5 6 A B and C Mounting step 1 User manual 310A Chapter 5 31 32 Mounting step two e Using forceps to hold the handle segment transfer excised vessel from Petri dish to myograph chamber Hold the vessel as close to the proximal end as pos sible and try to mount the vessel onto the wire If the lumen is shut try one of the following possibili ties e Use the wire to gently push the lumen open blood streaming out is a good sign e Hold excised vessel about 3 mm from the cut end with one set of forceps and use the other forceps to squeeze the blood remaining in lumen out through the cut end e Pull the proximal end of the excised vessel segment along the wire such that the vessel segment acts as its own fee
58. using KPSS to give a depolarising contractile response this stimulus also includes a component from neurally released noradrenaline The final stimulus is performed using a mixture of PSS and 10 uM noradrenaline All solutions are preheated to 37 C and aerated with a mixture of 95 O and 5 CO before use Instructions for making the necessary solutions are described at the end of this chapter Repeat 1 x Stimulus 1 amp 2 Wash out KPSS 10 uM NA 4 x with PSS Stimulate for 3 minutes Wait 5 minutes Stimulus 3 Wash out PSS 10 uM NA 4 x with PSS Stimulate for 3 minutes Wait 5 minutes Stimulus 4 Wash out KPSS 4 x with PSS Stimulate for 3 minutes Wait 5 minutes Stimulus 5 Wash out KPSS 10 uM NA 4 x with PSS Stimulate for 3 minutes Ready for experiment DMT Single wire myograph system Model 310A 5 5 Endothelium function The reasons for checking endothelium function may include 1 To check whether the relaxing function of the endothelium is intact The procedure is performed to make sure that the endothelium is not damaged during the dissec tion or mounting procedure 2 If an experiment requires removal of the endothelium this procedure is useful to check whether the endothelial cells were successfully removed The procedure can be performed after the vessel segment has been heated equili brated and normalized Preferably the procedure should be done after performin
59. ut Rec 2 Temperature TEMP output Rec 3 pH reading output if installed Rec 4 No output The output order can be programmed to whatever order you desire Make the re corder line of interest active uppermost in display and use the Enter key to toggle between the available signals The full scale output from the Myo Interface is one volt In the Recorder menu it is possible to change the associated values for each analog output that correspond to O V lo and 1 V hi Timer menu The Timer menu makes it possible for the Myo Interface to act as a timer the inter face can sound an alarm after a predefined countdown To set the countdown time move the line to be programmed hour Minute or second to the top of the display Use the numeric keys to enter the time value and press Enter to confirm To control the timer function press the Info key and the display shows the F key DMT Single wire myograph system Model 310A options for the Timer menu Choose one of the F keys to proceed with the following options or press Info again to return to the Main menu F1 Proceed to the Setup menu F2 Start the countdown the time values are entered in the Timer menu F3 Stop or pause the countdown F4 Reset the entered time values or to turn off the alarm Setup menu Calibrate Displays the signal to be calibrated either force or pH Valve dly The time in seconds that the electronic valve optional to the system remains open when
60. www dmt usa com sales dmt usa com support amp dmt usa com
Download Pdf Manuals
Related Search
Related Contents
Descargar las instrucciones CAPFITOGEN Samsung 高效节能系列 双开门冰箱 540L 不锈钢银 RS542NCAESL 用户手册 KP500D - Grothe Days of Wonder Zug um Zug Europa U9809 UHF RFID Reader User`s Manual V1.2 Consejo de la Magistratura - Poder Judicial de la Nación Prometheus Feature -- Using FILES and FILES Blue Cat`s PatchWork User Manual Copyright © All rights reserved.
Failed to retrieve file