Home

User Manual - Rockwell Automation

image

Contents

1. Trigger i Trigger Evaluation Delay Time gt OUT4 Ready Busy OUT 1 2 Pass Fail Ready On Ouput Valid Rockwell d 10000000877 02 Automation MultiSight User Manual Start up Case 3 Output Delay Time in ms between the Trigger and the availability of results from Output 1 2 Pass Fail Note that the Output Delay must be longer than the evaluation time Oo IN1 Trigger Output Delay Includes Evaluation Time OUT4 Ready Busy OUT 142 7777777777 lt 4 Pass Fail ie Ready On Ouput Valid Case 4 Output Duration Duration in ms of the result output on Outputs 1 2 The outputs will turn OFF once the Output Duration is complete unless the Output Active parameter is set to LOW Normally Closed in which case the outputs will turn ON once the Output Duration is complete i Trigger Evaluation Time Evaluation Time x Output Duration OUT4 Ready Busy i OUT 1 2 os i gt OS Pass Fail Da Ready ON i Pare Output NOT Valid Output OFF Output Valid When the configurable delays are used the minimum total cycle time will be at least as long as the sum of the delay times This will reduce the number of inspections that can be made ina given fi period of ti
2. 0 50 0 94 Power OUT 6 Kee M12X1 0 a av at a L 14 13 1 0 51 Connector Pin Configuration aag 71 2 va Ring Light White Light 48MS RLWH je 25 4 0 98 Connector Pin Configuration ax 7 2 7 3 9 3 9 4 5 ta 5 72 10000000877 02 2 3 1 3 1 4 5 8 8 6 7 6 7 22 65 op SY NG 30 5 1 20 Rockwell Automation MultiSight User Manual Maintenance Mounting Set up Illustrations Rod clamp 200 Rod bracket Angle bracket Dovetail bracket Mounting rod Mounting plate Mounting rod Dovetail Bracket 48MS BKTDT Angle Bracket 48MS BKTANG Rod Clamp 48MS RODLINK Rod Bracket 48MS BKTROD Mounting Rod 48MS RODxxx Mounting Plate 48MS MTPLATE Rockwell Autom ation 10000000877 02 73 MultiSight User Manual Maintenance Rockwell 10000000877 02 Automation MultiSight User Manual Maintenance Rockwell Autom ation 10000000877 02 75 www rockwellautomation com Power Control and Information Solutions Headquarters Americas Rockwell Automation 1201 South Second Street Milwaukee WI 53204 2496 USA Tel 1 414 382 2000 Fax 1 414 382 4444 Europe Middle East Africa Rockwell Automation NV Pegasus Park De Kleetlaan 12a 1831 Diegem Belgium Tel 32 2 663 0600 Fax 32 2 663 0640 Asia Pacific Rockwell Automation Lev
3. Click Connect Find in MultiSight PC Software Failure Execute Ipconfig Sensor is listed with IP Start run cmd input ipconfig Execute ARP Start run arp a d Success Reset IP address of sensor i e aaa bbb ccc identical to PC IP address area IP Address area of sensor and PC identical no ddd different i e IP aaa bbb ccc identical Execute Ping Start run cmd ping IP Adr Sensor e g 192 168 100 100 Click Connect Find in MultiSight PC software Failure eran Check software and sensor version and make connection with Connect Success Rockwell 10000000877 02 17 Automation MultiSight User Manual Start up Firmware Update Before using the Update function consult your Support group and use only the original update files from the manufacturer The execution of the function must not be interrupted as this may cause failure of the MultiSight The stored detector configurations are deleted by the update and must be redefined when the function is concluded This function allows you to update the software of the MultiSight To update the software of the MultiSight proceed as follows Open the MultiSight software Find and select the desired sensor Click on the button Update and select the desired update file Switch the MultiSight Sensor off and on again Reconfigure the MultiSight Sensor with the PC software Start up Rockwell Automation supplies defect free software fo
4. nanda ad dd te dacs Rae resp eee birdie Use Of Control InputIN2 auisraudedesentand eten a ag a a SDR GA F r Control Input Activate 0 or 1 and Activate 0t031 cece cece eee eaaa eaaa aee 60 Control Input Teach Temporary or Teach Permanent eecee cece cence ee eaaa eaaa 61 3 Minimum Operating Distance and Field of View by Cat No 63 66 ACCESSO S um eee gt SINE TOG 6518 Wa ISAM AS Mate Kedar 67 Dimensional Drawing Sa scores ee aa a NATA Ag rer BA TA TE ang DEA ADE ete estere 68 Mounting Setup Illustr tionssuvaasssdsdrida arret ANE E NABA OEE Gate BE aa 73 Rockwell Autom ation 10000000877 02 3 MultiSight User Manual Components Supplied Components Supplied Bulletin 48MS MultiSight vision sensor with integrated lighting Computer software on CD ROM Installation instructions mounting bracket 48MS BKTDT with three screws Allen wrench screwdriver and protective cap for M12 Ethernet connector Two PVC M12 QD covers must be installed around the metal coupling nuts of both connection cables for ESD protection Typical Applications The MultiSight is an optical multipixel sensor with a pass fail PNP output The MultiSight uses three four different methods of evaluation pattern matching contrast brightness and con
5. Network Connection on page 8 In addition every machine in the network must be allocated a unique IP address For the MultiSight Sensor Finder software to detect a MultiSight via Ethernet all wireless Ethernet connections from the PC must be disabled Also deselect Check Point SecuRemote in the Local Area Connection Properties and disable any firewalls The IP address consists of four decimal blocks in the format aaa bbb ccc ddd where each block may have a value from 0 255 The first three decimal blocks aaa bbb ccc form the address area In order to use the configuration software the address area of the PC and the MultiSight must be identical The fourth decimal block ddd must be different The sensor finder tool Connect Find can be used to change the IP address of the MultiSight even if the address area of the PC and MultiSight are different Example If the IP address of the PC is 192 168 2 67 any address between 192 168 2 1 and 192 168 2 254 can be selected for with selected subnet mask 255 255 255 0 the MultiSight with the exception of the IP address of the PC 192 168 2 67 and the IP addresses of any other machines in the network The IP address of the MultiSight is set by the manufacturer to 192 168 100 100 In the above example the IP address of the MultiSight therefore has to be reset to the address range of the PC see Setting the IP Address of the MultiSight on page 15 Setting the IP A
6. 000 w Control Input Enable v Trigger Mode Triggered v Shutter Adjustment Manual v CA a ms ene amp Output Active H Ok Threshold Min 0 55 7 Threshold Max 1 00 Trigger Delay ms Output Delay ms 0 Output Duration ms oOo D Output Active HIGH v Threshold Min 0 55 4 Threshold Max 1 00 lt 7 Slowly adjust the manual focus on the back of the MultiSight using the small screw driver until the image comes into focus You may need to adjust the shutter speed more to get the best image The ideal is to have a high level of contrast and a sharp focus 8 Click Stop in the Image window to exit live image mode Define Search Criteria The next step is to position the Region of Interest ROI and Pattern frames The yellow ROI frame defines WHERE the sensor will look for a pattern the red Pattern frame will define WHAT the sensor will look for 9 By default the yellow ROI frame will be selected notice the dot at each corner and along each side Double click the red frame to select it Position and resize the red pattern frame around the unique feature to be inspected the pattern that will always be there on a good part and never for a bad part 10 Next adjust the yellow Region of Interest ROI frame This frame defines WHERE the sensor will look for the pattern we defined with the red box This means it must always be bigger than and completely contain
7. 29 MultiSight User Manual Start up Parameter Description Screen Shot 30 Logic continued In order to use the grouping functionality Detector Parameter Control Input must be set to Activate 0 31 for ALL active detectors The same Logic is applied to each Group of detectors Example if the detectors in Group Aare logically linked with AND the detectors in Group B will be logically linked with AND It is not possible to link Group A with AND and Group B with OR It is not possible to logically link groups with one another The functionality of the left hand check box is different for grouping than for the logical linking previously described The evaluation of a group begins with the detector last selected via the Control Input IN2 When the sensor is connected to the computer for configuration the radio button currently selected will determine which detector the sensor begins the group evaluation with The sensor evaluates each detector that has a check in the left hand box The next detector that does NOT have a check is the last detector in the group If detectors 1 3 have a check and 4 does Not Group A is detectors 1 4 See the screen shot to the right The detectors in different groups do not have to have matching parameters however those within the same group must follow the matching parameter guidelines listed above 10000000877 02 Select Detector Sensor name Live
8. New IP 192 168 100 100 Itis possible to adjust the subnet mask and standard Mask 24 255 255 255 0 gateway using the set IP function It is recommended to nn nl tew not change these settings unless necessary Gateway It is not possible to see IMPORTANT gateway than your PC 5 Switch the power to MultiSight off and on again and click on Find again to locate the sensor with the newly assigned address Note the period between the sensor re start and successful Find is approx 6 sec If necessary repeat Find operation The MultiSight is now ready for operation at the newly assigned IP address For network operation use only IP addresses which have not already been assigned Rockwell Autom ation 10000000877 02 15 MultiSight User Manual Network Settings Creating a Functioning Ethernet Connection Between a PC and a MultiSight 1 Direct connection see Electrical Installation on page 8 All cables correctly connected 24V DC Power Supply l Connect all cables correctly 1 Power and I O cable 2 Ethernet crossover cable Electrical connection ok MultiSight Pwr LED on jletwork Start Settings Network connections No red X on Local Area Connection Contact system administrator Set IP address of PC to fixed address Check settings adjust if necessary Start settings network connections right click local area connection Internet Protocol TCP IP Properties IP address
9. OUT2 position Output state is held until the next evaluation is complete Output state is de activated after the Output Duration time elapses Output ON or OFF based on setting of parameter Output Active Image The Image section contains the current camera image as well as the Region of Interest ROI Pattern and Position frames for the detector currently selected via the radio buttons in the Select Detector window and other detectors selected with the Display Icon check box in the Select Detector window 42 10000000877 02 Rockwell Automation MultiSight User Manual Start up Screen Shot Field name Description SF Display se Single mage ROI Display via Trigger Patten Post Zoom Max Pattern Size 400x160 Live Grid None Save Image Region of Interest ROI Yellow frame with a number Frames Region of Interest ROI Yellow Pattern Red Position Blue and Match Green When a detector is selected radio buttons in Select Detector window its ROI appears as a yellow frame with a number on the image The ROI can be selected by double clicking anywhere inside it A dot will be displayed just outside each corner and along each side to indicate that it is selected When the ROI frame is selected it can be moved by clicking and dragging inside the frame and it can be resized by placing the cursor on
10. Position Control Off v External Constant gt Internal lighting is off for image exposure External lighting trigger is on constantly while the sensor isin Run Shutter 3 74 ms mode or the Live Display in the Image window is selected during Trigger Delay m D configuration i e the external light is always on it does not es strobe with each image exposure Use this setting to help reduce Output Delay ms annoyance from constantly strobing light Output Duration ms 0 Output Active HIGH v Threshold Min 0 35 4 Threshold Mas 1 00 Resolution Speed Changes the resolution and therefore also the speed Zoom Normal Faster Fine Average Rough Fastest Use this setting to improve cycle time Zoom function On Off digital zoom 2 1 32 10000000877 02 Rockwell Automation MultiSight User Manual Start up Parameter Description Control Input The settings for the Control Input are described in detail on page 58 Enable Teach temporary Activate 0 or 1 Activate 0 to 31 Teach permanent If this parameter is set to Enable the default setting the 1 0 control input IN2 must be connected to V 24V DQ This is in addition to a trigger input on INT external trigger If fi the control input IN2 is not connected to V the MultiSight is in standby mode and will not execute evaluations unless the trigger mode is continuous see page 58 for more details Trigger Mode T
11. high A high level V on the control input IN2 means evaluation according to detector 2 configuration a low level V GND means evaluation according to detector 1 configuration more details below Case 3 Control input Activate 0 to 31 This function allows the Control Input to determine which detector is evaluated by using a series of pulses during the Ready state In period 5 Ready high V pulses are counted C type 5 ms pause 5 ms pulse 5 ms pause etc The next evaluation takes place with the detector corresponding to the number of pulses counted The count always starts at 1 regardless of which detector was active previously The pulses must be generated bounce free preferably by a control system more details below Case 4 Control input Teach temporary or Teach permanently only possible on detector 1 This function allows the MultiSight to be taught a new pattern without being connected to the software In period 5 Ready high V a high level V can be applied at IN2 to trigger a temporary saved in RAM only or permanent Teach saved in flash This must be applied at least until the next rising edge at trigger input IN1 Then the new pattern is taught in detector 1 and saved according to specification The control input IN2 must be pulled low V GND before the next rising edge at trigger input IN1 or another Teach is triggered In this mode only detector 1 is used and overwritten by the next Teach
12. to work with a larger image This eraser can be used to eliminate portions of the taught line contour depicted as a blue line that should not be included in the inspection For example another component near the part to be inspected or even a shadow may cause a contour shape in the taught image The blue contour lines created by these conditions can be eliminated by placing the cursor eraser over those sections and clicking the mouse button If a desired blue contour line is accidentally deleted the entire contour must be re taught For a repeatable inspection results using the Contour Matching 1 detector type a high contrast image is essential There are several additional parameters that appear when the Contour Matching detector type is used These parameters affect either the teach process or the inspection process They are described in more detail below For most applications these can remain in their default states Rockwell Autom ation 10000000877 02 37 MultiSight User Manual Start up Parameter Description Screen Shot Edge contrast Difference in grey scale value Detector Parameters Minimum difference in grey scale value to define a line contour i e the min difference between bright and Detector Type Contour Match dark areas to get a blue contour line Illuminati lintemalOn fal Defined in terms of 8 bit grey scale value polo intemal On if Value range 7 128 defau
13. 192 168 100 101 255 255 255 0 Confirm via LAN properties that firewalls wireless networks and check point secure remote are disabled After second attempt i me vet Failure Execute Ipconfig Click Connect Find in MultiSight PC Software Start run cmd input ipconfig Sensor is listed with IP Execute ARP Start run arp a d Reset IP address of sensor i e aaa bbb ccc identical to PC IP address area ddd different Execute Ping Start run cmd ping IP Adr Sensor e g 192 168 100 100 IP Address area of sensor and PC identical i e IP aaa bbb ccc identical Click Connect Find in MultiSight PC software Failure A ate 0 Check software and sensor version and make connection with Connect 2 Network connection see Network Connection a for standard models on page 10 Rockwell 10000000877 02 Automation All cables correctly connected MultiSight User Manual Network Settings Connect all cables correctly 1 Connecting cable 2 Ethernet cable straight Electrical connection ok MultiSight Pwr LED on Network Start Settings Network connections No red X on Local Area Connection Obtain IP address for sensor from system administrator If necessary execute IPConfig Determine IP of PC and note Check settings adjust if necessary Confirm via LAN properties that firewalls wireless networks and check point secure remote are disabled
14. 3 m 48MS ECS 3 Lighting Cable 48MS LCS Lighting Cable Right Angle 48MS LCRT Sealing Cap M12 Male Connector 889A DCAP Included with MultiSight sensor Rockwell Automation 10000000877 02 67 MultiSight User Manual Maintenance EtherNet IP Model ett 7 Cable Connector Cover nonconducting M120 Ethernet M12 D code to RJ45 converter 1585A DD4JD Ethernet Cable RJ45 to RJ45 2m 1585J M8TBJM 2 Ethernet Patchcord Crossover M12 D code to RJ45 3 m 48MS EPC 3 M12 Patchcord for Lighting Accessory M12 to M12 2 m 889D F8ABDM 2 Included with MultiSight sensor Other lengths available replace 2 with length in meters 5 m and 10 m are standard Note Additional accessories longer cordsets and additional LED colors for external lighting available with longer lead times Contact your Rockwell sales representative for additional information Approximate Dimensions mm in MultiSight Standard Models Dimensions are not drawn to scale 45 1 77 12 7 64 2 52 305 an 05 1 2 ptical Axis I Om 30 5 9 2 1 2 6 ae 597 M12x1 ZOHO 097 p 1 225 a 0 89 Rockwell 10000000877 02 Automation MultiSight EtherNet IP Models MultiSight User Manual Maintenance Dovetail Bracket 48MS BKTDT Included with MultiSight sl 4 Angle B
15. 4 100 200 2 150 5 5 5 0 100 LL LL 50 0 T T T T T 0 T T T T T 0 100 200 300 400 500 600 0 100 200 300 400 500 600 3 93 7 87 118 15 7 19 6 23 6 3 93 7 87 11 8 15 7 19 6 23 6 Operating Distance mm in Operating Distance mm in X Width Y Height Operating Distance Height of Field of View Field of View Rockwell 4 10000000877 02 Automation MultiSight User Manual Maintenance Connection M12 Micro 8 pin Male QD both PWR and I O and ethernet Ethernet connection M12 micro 4 pin D code female quick disconnect ethernet connection EtherNet IP model T Power and I O Connection Cable Wiring Diagram Trigger Sensor Trigger Sensor V Brown 2 White 1 24V DC PNP Output V Blue 7 Green 3 i GND Pass Fail PLC Position Input i F Not Pink 6 I O i Connected Ready 4 Grey 5 Control Il 4 V or PLC p Output p ere Ls Inspection Target Rockwell 10000000877 02 65 Automation MultiSight User Manual Maintenance Connector Pin Assignments PWR and 1 0 connector Pin M12 Color Use 1 White IN1 external trigger 2 Brown 24V DC V 3 Green OUT pass fail display LED 1 4 Yellow OUT4 ready O 5 Grey IN2 control input 6 Pink OUT3 external illumination trigge
16. Detector 1 can still be logically linked to other detectors more details below Additional Details on the Special Uses of the Control Input Control Input Activate 0 or 1 and Activate 0 to 31 Cases 2 and 3 The parameter Control Input can be set up to allow the control input IN2 to select between the different detectors The parameter setting is Activate 0 or 1 OR Activate 0 to 31 When one of these options is selected the following parameters must be set accordingly for all used detectors Control Input must be the same for all either Activate 0 or 1 OR IMPORTANT Activate 0 to 31 more details below Shutter Adjustment Manual Trigger mode Triggered It is recommended to keep a list of which pattern is saved in 1 which detector number Rockwell bb 10000000877 02 Automation MultiSight User Manual Maintenance When changing detectors if a detector is selected that does not have the Control Input parameter set to Activate 0 to 31 or Activate 0 or 1 itis not possible to change to another detector using the control input IN2 The control input has a different function and therefore is no longer capable of switching to another detector For this reason it must be confirmed that the control system only selects detectors that are configured to allow the control input to change detectors If an incorrectly configured detector is selected the sensor may no longer respond and therefore would need to be reconf
17. For Help press FI 5 Adjusting image focus To adjust image definition turn the focusing screw on the back of the housing see the illustration in Electrical Installation on page 8 in clockwise direction increase the focusing distance or counterclockwise direction decrease the focusing distance Check the position of the MultiSight and adjust if necessary IMPORTANT Make fine adjustments to shutter speed and focus using GOOD parts to enable optimal configuration The basic setup of the MultiSight is complete To differentiate between GOOD and BAD parts use the software to configure the MultiSight The next section describes the configuration software Rockwell Automation 10000000877 02 21 MultiSight User Manual Start up Program Description Menu The menu for the PC program has main and subsections These are selected by clicking the desired menu item or by pressing the ALT key and the first letter of the menu item at the same time Sensor Allen Bradley 48MS MultiSight Sensor Language Result Display w 1 Allen Bradley 48M5 MultiSight Sensor File Exit Exit program Options Language Select language The new language does not take effect until the software is restarted Currently English is the only available language Options Result Display Keyboard Locked The Result window is shown enlarged in a pop up window This mode locks the keyboard and limi
18. Images Recorder Off v Logic AND Al Rockwell Automation MultiSight User Manual Start up Parameter Description Screen Shot Radio buttons Designates the active detector by clicking one of the ten radio 0 9 buttons the corresponding detector is selected and its parameters 0 31 can be configured The parameters are displayed in the Detector Parameters window in the middle of the screen Only one detector 1 can be configured at a time After restart the most recently O 2 selected detector is active The selected detector evaluates a 0 3 Region of Interest ROI which is configured in the Image window If a certain detector is to be tested select that detector and click the Run command button to the left Display icon Checking this box indicates whether the yellow Region of Interest Display ROI and red pattern frames should be shown on the Image during En Run mode even if the detector is not selected with the radio button The Region of Interest ROI and pattern frames of the DU detector selected with the radio button are automatically g displayed during Run mode other detectors with a check in the display icon box are shown but greyed out This function does not affect the performance it is just included to ease the setup process Logical link The time required for execution increases with the number of logically linked active detectors fi
19. Live Display button in the Image window After a moment the image in the window begins to be repeatedly refreshed Rockwell 10000000877 02 Automation MultiSight User Manual Start up 11 Ifno live image appears check to confirm the following a No obstruction in front of the lens b Sufficient lighting Detector Parameter Illumination should be Internal ON or sufficient external lighting provided c Correct shutter settings under Detector Parameters see screen shot Increase shutter time to brighten the image d Correct focus focus screw on back of MultiSight Sensor Allen Bradley 48MS MultiSight Sensor with EtherNet IP 3 Detector Type Pattem Match v Disconnect umination IntemalOn M mages IResokuion Speed Fine Average v om Control Input Enable v Trigger Mode Triggered a Pun loo ee Shutter Adustment Manual gt Bee Postion Control Off v BE Hee Shutter 0 81 me Os mal Trigger Delay ma 9 BE pi Output Delay ms o ig PERTE 650 or He Output Duration tms w Display vie Togon O9 Dee Output Act v Bu p R er Zoom Max Pattern Size 20080 v Load Si Bos Threshold Max 1 00 Live Gid Nene 4 Ou Ore Sie Ge ine Ree 88 ois Allen Bradley O1s fee Q2 0r Copy oa ors 65 Hi freak OCT 7 92 Ors IX Peston BB Be jis ee 028 Eee Recorder g3 abs Images Oa Dr
20. Min and Max Threshold values define the minimum and or maximum evaluation score for the inspection to pass Can be adjusted by the sliders or by clicking on the threshold number and typing a new value into the pop up window Values 0 00 1 00 in 0 05 increments 35 conformity threshold would be entered as 0 35 The threshold values are depicted by small vertical black lines on the detector results bar graph when the sensor is in run mode Both thresholds can be used at the same time and in this case a section in the middle of the results bar graph can be defined as the Good zone PASS output when Threshold Min is exceeded AND Threshold Max is not exceeded Setting the Threshold Max below the Threshold Min is not possible Threshold Min and Max In practice Good parts are not 100 identical with the pattern taught Slight differences cont in position or fluctuations in light level mean that they do not conform completely to the fi reference pattern Bad parts should differ from Good parts in as many characteristics or as large a part of the pattern area as possible If a pattern has been taught using a Good part it must first be established what degree of conformity with the pattern is achieved with various Bad parts reference the length of bar graph for results The degrees of conformity with the pattern achieved with various Good and Bad parts should be as far apart as p
21. by a single detector The evaluation result of each detector is then logically combined with the evaluation results of all the other detectors resulting in a single output OUT1 LED Q1 If the AND All logic is selected all selected detectors must pass evaluation for the pass fail output OUT1 to give a PASS output If any one detector fails its evaluation output OUT1 gives a FAIL output If the OR Search logic is selected OUT1 gives a PASS output if any one of the detectors passes its evaluation See the parameter descriptions below for an explanation of how to logically link detectors Detectors can also be grouped The grouping of detectors makes it possible to check several characteristics on different parts without reconnecting the software during part changes For example during the first production shift of the day Part A is inspected for several characteristics then during the second production shift Part B is inspected for different characteristics The sensor can be changed via the Control Input on the PWR and 1 0 connector from Part A to Part B without having to reconnect to the software via the Ethernet port See Use of Control Input IN2 section on page 58 Rockwell 0 10000000877 02 Automation MultiSight User Manual Recorder Image window of the software during Run mode Selecting this mode increases evaluation cycle time whether the sensor is connected IMPORTANT to the software or not Desel
22. control relay in the trigger circuit or by using a PLC to drive the trigger as opposed to directly wiring from a trigger sensor Rockwell Automation 10000000877 02 61 MultiSight User Manual Maintenance The save operation for permanent save Teach permanent may take up to 500 ms During this period in continuous moving applications the output locks to FAIL regardless of the part evaluation i e both good and bad parts give a FAIL output When the Teach Temporary function is used the new pattern is stored only in the RAM this requires much less time In this case the pattern is no longer available after the MultiSight is reset and must therefore be re taught Rockwell 2 10000000877 02 Automation MultiSight User Manual Maintenance Trigger Mode Parameter Triggered or Continuous When Triggered is selected the sensor initiates an exposure and evaluation when the external trigger INT transitions from low GND to high V 24V DO or the Trigger command button in the computer software is clicked When the Continuous Trigger Mode is selected the sensor initiates a new exposure and evaluation as soon as the last evaluation is complete i e the sensor cycles as fast as possible This Continuous internal triggering is initiated as soon as the sensor is put in Run mode Setting Up the Trigger Sensor During operation of the MultiSight sensor the position of the additional ext
23. hazard avoid a hazard and recognize the consequences SHOCK HAZARD Labels may be on or inside the equipment for example drive or motor to alert people that dangerous voltage may be present BURN HAZARD Labels may be on or inside the equipment for example drive or motor to alert people that surfaces may reach dangerous temperatures 1 lt lt We recommend that you save this user manual for future use The following symbols are used in this manual for clarification 8 This symbol indicates supplemental sections of the text which contain useful tips Identifies features and functions that are only available with EtherNet IP models These features and functions are not available on the standard MultiSight models MultiSight User Manual Important User Information Table of Contents Important User Information and Definition of Symbols 00 cece cee cee eaaa eee e eaaa eens 2 Components Supplied tyes EN NG uti a GS Aga 4 TypicalAppli ationsarxssarresrsssrkt ree cre ee eee ernie NG anal Pte 4 Requirements for WS usa ae Eee CG Oa eon 4 5 Description of Operation 16 Install tionInstrudtions ie suse entre vit set eee eee eteren AG a Aa EA a a akan ai Ka i Install tions sutue sata 00 dre He stale ig dale Neeb fe ers Electrical Installations na sscvs3 4 pas Yon ass oa vater GEAR Shae Resten 8 Netw
24. 02 Automation MultiSight User Manual Network Settings Setting the IP Address of the MultiSight Contact the system administrator to determine which IP addresses are allowed on your network for the MultiSight The IP address set for the MultiSight should be marked on the enclosed label After installation stick the label on the sensor in a clearly visible position Before setting the IP address make sure that the computer is connected as described in Electrical Installation and Basic Computer Settings see Electrical Installation on page 8 and Basic Settings for PC and MultiSight on page 12 To set the IP address on the MultiSight proceed as follows in the PC software MultiSight Sensor Finder 1 3 IP Type Version Mode Description 9192168100 100 48MS Object 40 Run 1 Open the MultiSight software 2 Click on the Find button After a few seconds a list of sensors in the network appears If this does not happen repeat the Find process 3 Select the desired sensor by clicking on it es 8 5 Update LocallPofPC 192 168 100 105 Subnetmask PC 255 255 255 0 4 Click on the Set IP button and enter the desired IP address for the MultiSight same as the IP address of the PC except for the last block see Basic Settings for PC and Ss MultiSight on page 12 Click Set to accept
25. Detector Parameter configurations Detector Type any Illumination identical Resolution Speed any Zoom identical Control Input identical Trigger Mode identical Shutter Adjustment manual Position Control any Shutter identical Trigger Delay identical Output Delay identical Output Duration identical Output Active identical Threshold Min any Threshold Max any The Logic field has the following options None No logical link or grouping i e only the selected detector determines the output AND All and OR Search The inspection output OUT1 pass fail is based on the evaluation of all detectors selected via the left hand check box In the screen shot detectors 1 6 are logically linked AND All OUT1 is active if ALL selected detectors pass evaluation OR Search OUT1 is active if ANY of the selected detectors passes evaluation AND and OR group The group functionality allows the MultiSight to make several different inspections on multiple parts without reconnecting to the software to change between the parts This functionality is described more thoroughly above in the general description of the Select Detector window Select Detector Sensor name L aA D 0 44 000 MN OOOO 000000000 Live Images Recorder Off v Logic AND All OOOOOOO0000 78 70 70 70 G o o ooo Rockwell Automation 10000000877 02
26. Last IP connects to the most recent IP address used Default IP connects 10 the default IP address 192 168 100 100 New IP type in IP address to connect to Find Opens sensor finder use when IP address of MultiSight is unknown Listed IP addresses recently used IP addresses Connect to an address by clicking on it Disconnect Existing connection to MultiSight is disconnected MultiSight goes into run mode Teach Teach the pattern or contour currently in the red pattern box in the image window Also used to determine the shutter speed when Shutter Adjustment parameter is set to Automatic The Teach command does not affect other configuration parameters Other configuration parameters are changed on fi the MultiSight as soon as they are changed in the software Teach from MultiSight image the MultiSight will take a new picture and learn the pattern or contour in the red pattern frame Note this image could be different than what is currently displayed if the part has moved or a setting has been changed Teach from PC image Recommended method the MultiSight will learn the pattern or contour in the red pattern frame of the image currently displayed in the image section of the software 24 Rockwell 10000000877 02 Automation MultiSight User Manual Start up Command Description Run Stop Run gt Evaluation mode MultiSight is automatically in Run mode when not connected to the
27. Live Grid None vil Save Image Field name Description Display ROI Pattern and For currently non selected detectors whose Display Position Icon check box is selected the check boxes determine whether each respective frame is displayed on the image Single Image Allows the user to capture a single image initiated by via Trigger the external or software trigger including trigger delay parameter settings This allows the user to capture an image for teaching without stopping the production line This also can be used to identify the ideal trigger delay setting Redraw Brightness When using the brightness detector type the image detector type only will color all of the dark or light pixels blue or other selected color 44 Region of Interest yellow frame with detector should always fully surround the pattern to be detected It defines the maximum search area in which the pattern will be recognized Pattern red frame Defines the pattern or contour to be taught During the inspection process the pattern frame does not have any function other than to mark the original location of the taught pattern The green match frame indicates where the pattern was detected Position frame blue The Position Control OUT2 determines whether the detected pattern s center is within this frame Rockwell 10000000877 02 Automation MultiSight User Manual Start up
28. MultiSight Vision Sensors USER MANUAL 48MS Rockwell ALLEN BRADLEY lt ROCKWELL SOFTWARE Automation Important User Information In no event will Rockwell Automation Inc be responsible or liable for indirect or consequential damages resulting from the use or application of this equipment The examples and diagrams in this manual are included solely for illustrative purposes Because of the many variables and requirements associated with any particular installation Rockwell Automation Inc cannot assume responsibility or liability for actual use based on the examples and diagrams No patent liability is assumed by Rockwell Automation Inc with respect to use of information circuits equipment or software described in this manual Reproduction of the contents of this manual in whole or in part without written permission of Rockwell Automation Inc is prohibited Throughout this manual when necessary we use notes to make you aware of safety considerations Identifies information about practices or circumstances that can cause an explosion in a hazardous environment which may lead to personal injury or death property damage or economic loss IMPORTANT Identifies information that is critical for successful application and understanding of the product Identifies information about practices or circumstances that can lead to personal injury or death property damage or economic loss Attentions help you identify a
29. Recommendations for selecting a pattern or contour For reliable evaluation The pattern contour should have a unique characteristic some aspect that does not occur on a bad part Patterns or pattern characteristics should have clear contours shapes Pattern contour characteristics should NOT vary The pattern area should be set as large as possible within the image A pattern that fills 10 of the image will be more reliably detected than a pattern that only fills 1 of the image If necessary move the sensor closer to the target The pattern area red frame should be entirely within the yellow frame region of interest For higher resolution and speed do not set the yellow frame larger than necessary relative to the size of the pattern The larger the area that requires evaluation namely the ROI relative to the size of the pattern the slower the evaluation time Sensor Image Region of Region of Interest Interest Sensor Image Patt att ern More Consistent Detection Less Consistent Detection Roc 3 ell 10000000877 02 45 Automation MultiSight User Manual Start up Region of Region of Interest Interest Sensor Image Sensor Image Faster Slower Position Deviation Check and Frame Type Position control can be activated when the pattern matching or contour matching detector types are selected If position control is selected a blue frame is displayed on image screen The X Y coordinates for the
30. The evaluation with logical linking affects both the primary inspection pass fail output OUT1 and the secondary position result OUT2 Rockwell Autom ation 10000000877 02 31 MultiSight User Manual Start up Detector Parameters To configure a detector select the appropriate radio button in the Select Detector window and then set the parameters in the Detector Parameters and Image windows Each detector is configured independently Detector parameter configurations can be copied into other detector parameters using the Copy command see page 24 Parameter Description Detector Type There are 3 4 different methods of evaluation Detector Parameters Pattern Matching page 35 Contrast page 36 Detector Type Pattem Match Brightness page 35 Illumination Internal On v Contour Matching page 37 i fm Resolution Speed These detector types are described in detail starting on page 35 pcan es Hemel Medhan ad Each of the 10 detectors can have a different detector type e g Zoom Off v Detector 1 Pattern Match Detector 2 Contrast etc This is a true even if the detectors are logically linked or grouped together Control Input Enable x Illumination Internal On gt Internal lighting is on for image exposure Trigger Trigger mode Triggered for external lighting is generated Internal Off gt Internal lighting is off for image exposure Trigger oo Le for external lighting is generated
31. UT4 Ready indicates 1 that processing is complete and 2 that the outputs OUT1 result of evaluation and OUT2 position are now valid A and can be processed by the control system unless output duration 0 see Configurable Delays section on page 40 4 The next rising edge at the trigger input INI starts the next evaluation 5 Period between rising edge at output OUT4 Ready and next rising edge at trigger input IN1 period which can be used for changing from detector to detector activate 0 or 1 activate 0 to 31 or Teach temporary permanently by control input IN2 OUT3 Trigger of external illumination high as long as image is taken shutter open unless Illumination parameter is set to External Constant in which case OUT3 is always ON See the signal sequence diagram on page 58 for more information about the use of trigger and output delay timers Rockwell Autom ation 10000000877 02 59 MultiSight User Manual Maintenance Case 1 Control Input Enable Control input Enable means that IN2 must be connected to V 24 V DO to allow the external trigger If control input enable default state IN2 grey wire MUST be connected to V 24V DC for the external trigger to work Case 2 Control input Activate 0 or 1 This function allows the Control Input to determine whether detector 1 or detector 2 is evaluated A change in the state of the control input B can take place in period 5 Ready
32. center of the pattern detected green cross at the center of the green frame with reference to the top left hand corner of the image zero point is displayed in the result window The increment for X Y coordinate values while using Detector Type Pattern Matching is 2 pixels so only even coordinate values will occur The center of the detected green pattern area green cross is inside the blue frame for correctly positioned parts In Run mode this will activate the position output OUT2 The center of the detected pattern indicated by the green cross in the green frame is within the blue position frame Position output OUT2 would be active Rockwell ag 10000000877 02 Automation MultiSight User Manual Start up Pattern The Pattern window shows the taught pattern or contour for the currently selected detector FF Display Zoom RO 6 Patterr Save Pattern gt Saves the taught pattern or contour as a bitmap file bmp Zoom gt Opens a window showing an enlarged version of the pattern 2 1 ratio digital zoom Eraser size contour matching only determines the size in pixels of the square eraser that can be used to edit the taught contour which is depicted as a blue line The cursor acts as the eraser when clicked on the contour in the pattern window Undesired contour lines such as those caused by other parts or shadows can be deleted Results The Results window shows the status of OUTT and OUT2 along wit
33. ckwell Autom ation 10000000877 02 55 MultiSight User Manual Maintenance Appendix Frequently Asked Questions How do I change the IP address of the MultiSight sensor Proceed as described in Network Settings and Setting the IP Address of the MultiSight Sensor sections on pages 11 and 15 respectively What is the easiest way of testing the connection to the MultiSight In the start menu of the operating system click on Run and then cmd In the window which appears enter the command Ping and then the MultiSight IP address e g ping 192 168 100 100 If all connections are correct and the sensor and the PC address section are identical and the MultiSight is ready for operation the command Ping will produce a positive signal If not re check your network configurations and consult your network administrator How do I avoid image dissipation and reflections indicated by light colored areas of the image Mount the MultiSight and set the shutter manually Adjust the illumination of the object and avoid fluctuations in brightness e g due to ambient light lam getting dissipation of image and brightness for the same object The brightness values at the edge of the illuminated image zone differ from image to image What can I do Set the shutter to Manual in the window Detector Parameters and set the slide so that there is no fluctuation in brightness Generally speaking a higher shutter
34. creen shot on page 20 This virtual detector is stored in a non volatile flash memory on the sensor and is re activated for standalone operation after switching off and on Standalone operation is therefore possible with or without PC or PLC Evaluation Method During operation the MultiSight continuously compares the pattern of the detected image with that of the reference image The pixel grey scale values of the sample are analyzed and a level of conformity to the original sample is determined If this level falls within the user adjustable threshold settings the sensor turns the PNP output ON or OFF i e the part passes or fails the inspection The EtherNet IP connection is capable of communicating additional data about the inspection Rockwell Autom ation 10000000877 02 5 MultiSight User Manual Installation Instructions Setup The PC must be connected to a MultiSight sensor via the Ethernet port to load the configuration software Also Run mode of the sensor is automatically stopped when the software connects to the sensor The software is tightly integrated with the firmware on the sensor As parameters are changed in the software they are changed in the sensor There is not a typical download or upload process but the write to flash memory button in the software must be pushed to save the settings from the sensor s RAM to the sensor s nonvolatile flash memory The Teach command is used only for
35. d pattern is detected with 100 conformity The green outline on the image indicates where the taught pattern is found The green bar graph in the Results amp Select Detector windows indicates the degree of similarity to the taught pattern The black line on the bar graph is the current Alten Bradiey threshold 1 Further objects can be set using the same procedure described above Select a new detector in Select Detector section select the desired parameters in Detector Parameter section and repeat the above steps If several objects or detectors are defined they can be logically linked or grouped see Select Detector Logic on page 26 Set the threshold value required degree of conformity of the pattern detected with the pattern taught This can be achieved by testing a variety of good parts and adjusting the Threshold Min parameter in the Detector Parameters window Rockwell eg 10000000877 02 Automation MultiSight User Manual RESET Power On In the opposite figure MultiSight sensor has evaluated the object as bad Notice that the red LED is on and the hex nut is missing from the bracket RESET Power On To reset the MultiSight pull out the plug or turn off the power supply and then restore power Maintenance Cleaning Clean the MultiSight with a clean dry cloth Do not use solvents Do not use sharp objects Do not scratch Ro
36. ddress of the PC Contact your system administrator to determine which IP IMPORTANT addresses are allowed in your network or locally in your PC and to carry out the necessary settings on your PC The IP address of the computer needs to be set when the MultiSight is directly connected with the computer see Electrical Installation on page 8 and only if the computer has no IP address In all other cases the IP address of the MultiSight has to be adjusted to that of the computer see Electrical Installation on page 8 Rockwell 19 10000000877002 Automation MultiSight User Manual Network Settings To set the IP address of the computer carry out the following procedure in the operating system 1 Click on the Start button on your computer 2 Click on Settings 3 Click on Network Connections 4 Double click on Local Area Connection Rockwell Automation Network Connections Edit View Favorites 0 9 File Tools Advanced Help 5 JO search Folders Ez Address e Network Connections Network Tasks Create a new connection Change Windows Firewall General Support 4 Local Area Connection Status LAN or High Speed Internet Local Area Connection onnected Integrated Fast PIE Connection Status Duration Speed Activity Bytes Connected 2 days 03 45 36 100 0 Mbps Sent Recei
37. ect this option for the faster evaluation time Off gt No images are recorded on the sensor Failed images gt The MultiSight records up to the last 30 images of evaluations that FAIL Allimages gt The MultiSight records up to the last 30 images both PASS and FAIL See Commands Recorder Images on page 24 for instructions on transferring the images from the MultiSight to the computer Live Images Recorder Off v Logic AND All pe O2 MO 03 MO 04 MO Os MO Os MO O7 Obes Os OOP Os Obes Ow DO Start up Parameter Description Screen Shot Sensor name Assign a name for the connected MultiSight This name appears in Select Detector the MultiSight Sensor Finder software and allows for easier Select Detector identification of the correct sensor when multiple MultiSight Sensor name sensors are on a single network Live Images When this checkbox is selected the live image is displayed in the o Rockwell Automation 10000000877 02 27 MultiSight User Manual Start up Parameter Description Screen Shot IMPORTANT Logic The Logic field allows the user to define how the detectors are linked together to determine the single OUT1 pass fail output A detector can be standalone inspection of a single characteristic logically linked multiple inspections on a single part or grouped multiple inspections on different parts The concept is more th
38. el 14 Core F Cyberport 3 100 Cyberport Road Hong Kong Tel 852 2887 4788 Fax 852 2508 1846 Publication 10000000877 02 December 2013 i h Copyright 2013 Rockwell Automation All rights reserved Printed in Germany
39. er C Program Files Rockwell Automation MultiSight a link is added to the start menu standard location Start Programs Rockwell Automation MultiSight and an icon is added to the desktop icon MultiSight Sensor Finder This User Manual can also be read directly from the CD ROM supplied Un install Procedure The software can be un installed by using the operating system fi Windows 2000 and XP under Start Settings Control Panel Add or Remove Programs Scroll down and select MultiSight Sensor then click the Remove button Configure the network settings according to the instructions in the Network Settings section on page 11 Start the PC software by double clicking the MultiSight Sensor Finder icon After the installation of the PC software only one related icon MultiSight Sensor Finder is visible on the desktop Double click on this icon to open the software which then executes one find loop and shows the list of all MultiSight sensors found on the local ethernet network The IP address of the computer is shown at the bottom of the window gt MultiSight Sensor Finder 1 3 192 168 100 100 48MS Object Local IP of PC 192 168 100 105 Subnetmask PC 255 255 255 0 Rockwell Automation 10000000877 02 19 MultiSight User Manual Start up 10 20 Click the desired sensor to select it in order to Connect SetIP or Update Click the Connect button in order to es
40. ernal trigger sensor can be optimized To do this make the following parameter settings to the PC software Trigger mode Triggered Live Images in Run mode check box selected live images update activated Click the Run command button to put the sensor in Run mode Observe the exposure that is continuously updating in the Image window Note where the pattern to be inspected falls within the Field of View Adjust the position of the trigger sensor to optimize its position in relation to the exposure Field of View Minimum Operating Distance and Field of View by Cat No Min Operating Distance mm Field of View at Min Operating Catalog Number in Distance mm in Focal Length of Lens 48MS SE1PF1 M2 48MS SN1PF1 M2 48MS SE1PF2 M2 48MS SN1PF2 M2 20 0 8 18x 14 0 7 x 0 6 6mm 20 0 8 8x6 0 3 x 0 2 12mm An operating distance of more than 100 mm may require external lighting The typical focus range is operating distance 5 The 48MS SE1PF1 M2 and 48MS SN1PF1 M2 6 mm focal length MORAN models typically require external lighting because the integrated lighting does not illuminate the entire field of view i e the edges of image are dark Rockwell Autom ation 10000000877 02 63 MultiSight User Manual Maintenance Field of View per Operating Distance 12 mm Lens 6mm Lens 200 400 350 xX X 150 300 E E 250 Y
41. ernet network Select the sensor and set the IP address if necessary The IP address and subnet mask of the PC is shown at the bottom of the window 3 Select the MultiSight by clicking on its IP address Establish the connection by clicking Connect A warning will pop up asking if you want to stop run mode Click Yes The configuration software will now open Optimize Image 4 Click the Live Display button under the Image The MultiSight illumination LEDs will begin flashing and you will get updated images live Single Image ae Display S EPan m Posion Zoom Max Pattern Size 20080 v Live Gid None v 5 Place the target terminal block or other item for inspection in the Image area 6 Notice that the Image is either too bright or dark and out of focus The first step in a new application is to perfect the image Two factors affect the image shutter speed and the manual focus The shutter parameter is in the Detector Parameters section third column about half way down the list Adjust the shutter speed by clicking on the current shutter speed and typing in a new speed or dragging the slider bar Make the image darker or brighter by adjusting the shutter speed lower or higher Rockwell Autom ation 10000000877 02 49 MultiSight User Manual Setup Quick Guide Detector Parameters Detector Type Pattern Match w Illumination Internal On NG Resolution Speed Fine Average v Zoom
42. h additional information co OCD X Bosttion 160 Evaluations 1 LED OUT 1 and OUT2 Y Position 150 Good Patterns 1 Direction Center Good Positions 1 Rockwell Autom ation 10000000877 02 47 MultiSight User Manual Start up Field Description Result Bar Graph X Y Position The degree of conformity of the current pattern with the pattern taught is shown as a bar graph The status of the overall inspection is shown as a green pass or red fail dot The result is determined by whether the degree of conformity is within the configured thresholds For a standalone detector the bar graph indicates the degree of conformity For logically linked detectors the bar graph represents the detector with the highest degree of conformity measurement value The threshold value set in Detector Parameters section is shown as a black line Shows the X and Y coordinates of the detected pattern center of green frame with reference to the zero point upper left hand corner of the field of view and a coordinate system of X 640 and Y 480 units Direction nu 7 With center top top right right etc shows the direction in which the current pattern deviates from the taught pattern with reference to position Center is displayed or LED green OUT2 high as long as the center of the green frame green cross is located inside the blue position frame Co
43. hen the external trigger IN1 transitions low GND to high V 24V DO while the control input IN2 is high V 24V DC The signal sequence is as follows High level V 24V DO on control input IN2 Rising edge GND to V to trigger input IN1 Note that if a second trigger on IN1 occurs while the control input IN2 is held high V a second Teach function will occur Be sure that the control input IN2 is pulled low GND after the Teach function is completed Applications where the object is continuously moving Pull control input IN2 high V for at least one external trigger image exposure The new pattern for detector 1 is saved and this new configuration applies immediately If the signal at input IN2 is held high V for several trigger cycles several image exposures the last exposure is saved as the new pattern for the first detector Applications with Timer Delays Apply a High level V 24V DC to control input IN2 when the next target object is being transported to the evaluation position therefore causing a trigger signal OR when the reference object is in the correct position and triggering has already taken place due to a Trigger Delay the trigger signal must be interrupted and re triggered the teach is initiated by the rising edge of the trigger input IN1 Only after this should the signal at control input IN2 fall to low level GND Interruption and re application of the trigger signal can be achieved by installing a
44. ield of View The height and width of the field of view is determined by the operating distance see Minimum Operating Distance and Field of View by Cat No on page 63 Field of View For most applications it is recommended to align the MultiSight at an angle of approximately 10 15 with reference to the optical axis in order to avoid interfering reflections from the inspection target Approx 10 15 Target Surface Rockwell Autom ation 10000000877 02 7 MultiSight User Manual Installation Instructions Fine Adjustment IMPORTANT Fine adjustment of the MultiSight should not be carried out until after electrical connection and start up PC software installation see Start up on page 18 Electrical Installation ATTENTION A The electrical installation of the MultiSight must be carried out by a qualified electrician When installing the MultiSight disconnect all electrical components from the power supply When the unit is used on a network make sure that the network address IP address of the MultiSight set by the manufacturer at 192 168 100 100 is still free and is not in use by any other unit connected to the system If necessary reset the IP address of the MultiSight as described in Network Settings on page 11 When the MultiSight is in use the protective caps supplied must be covering the M12 socket Ethernet if it is not in use Failure to do this may cau
45. igured by connecting to the software Recommended Procedure Configure detector 1 with the desired parameters including the above required parameters Control Input Activate 0 to 31 Shutter Adjustment Manual Trigger Mode Triggered Use the Copy command to copy the required parameters from Detector 1 to the other detectors required for use instructions on pages 25 and 29 Set up the control system to generate the appropriate number of pulses for detector selection during the OUT4 Ready signal Testing Saved Detectors When all detectors to be used have been configured as described above a detector can be selected using the radio buttons in the Select Detector window Put the sensor into Run mode Run command The sensor starts with the last detector selected via the radio buttons or the control input IN2 The detector can now be tested using various good or bad parts Parameters such as Shutter or Threshold value can be adjusted if required To test other detectors select the radio button of that detector Control Input Teach Temporary or Teach Permanent Case 4 The parameter Control Input can be set up to allow the control input IN2 to allow a pattern Teach for the first detector This gives the input on the PWR and 1 0 connector the same functionality as the Teach command button so that the software does not have to be connected for a Teach function to occur Only for the first detector The Teach function occurs w
46. ing the red frame Double click the yellow frame to select it Move and resize the frame Rockwell 50 10000000877 02 Autom ation MultiSight User Manual Setup Quick Guide 11 In the Detector Parameters section third column change the Trigger Mode setting to Continuous from Triggered While in run mode this will cause a new inspection to start as soon as the last one finishes In the Select Detector section second column click the Live Images checkbox so that images will be displayed in the software while the sensor is connected 12 Click the Teach button in the Commands section Click Teach From PC Image Once the teach process is complete the taught pattern appears in the Pattern window on the right Function Test 13 Click the Run button in the Commands section Confirm that the inspection is passing by referencing the indicators Based on the results of several good and bad parts you may need to adjust the threshold settings Rockwell Autom ation 10000000877 02 51 MultiSight User Manual Setup Quick Guide Sensor Allen Bradley 48MS MultiSight Sensor with EtherNet IP Select Detector Detector Parameters R Allen Bradley For Help press F1 14 Based on the results of several good and bad parts you may need to adjust the threshold settings Click the Stop button in the Commands section to return to configuration mode and make adjustments Example Applicat
47. ion Example Detecting whether hex nuts are welded on mounting brackets by using the detector type Pattern Matching During machine assembly inspect mounting brackets to ensure that hex nuts are welded in specified places Condition Mount the MultiSight sensor in position see Installation on page 6 and attach all cables and wires per Electrical Installation on page 8 and Connector Pin Assignments on page 66 The software connection to the MultiSight is described in the Start up section on page 18 Rockwell 52 10000000877 02 Automation MultiSight User Manual Setup Quick Guide When the MultiSight is powered up the image shown in the Image section of the computer software is the last image recorded For setting up a new inspection click on the Reset Command button To view and continuously update the live image click the Live Image button To bring the object into focus turn the focusing screw on the back of the sensor To improve the contrast use the Shutter in Detector 71 Parameters 4 Set up Select Detector Type to be Pattern Match in the Detector Desse Type Paranta 8 Illumination Intemal On v Parameters section Resolution Speed Noimal Medium Zoom Off v Set region of interest ROI by double clicking onthe yellow box coop matis and placing it on the image This is the area in which the Tinos
48. l Number Technology 640 x 480 CCD Integrated Lighting 8 LEDs 6 x white 2 x red Integrated Lens Focal Length 6 or 12 mm adjustable focus Lens adjustable to infinity 6 12 Min Operating Distance in mm 20 20 Min Field of View X x Y in mm 18x14 8x6 Field of View 200 mm 150 x 100 60x 40 Depth of focus approx 5 of operating distance Mechanical Data Length x Width x Height Standard model 77 45 45 mm including connector EtherNet IP model 81 x 45 45 mm including connector Weight Approx 160 g Vibration Shock EN 60947 5 2 Operating Temperature C F 0 50 32 122 80 humidity noncondensing Storage Temperature C F 20 60 4 140 80 humidity noncondensing Enclosure Type Rating IP 67 O For use in NFPA79 applications only Rockwell Automation 10000000877 02 57 MultiSight User Manual Maintenance Technical Data continued Connections PWR and 1 0 connector M12 8 pins Ethernet M12 8 pins Ethernet M12 4 pins D code Housing Material aluminum plastic Functions and Characteristics Number of test detectors per sensor up to 10 32 detectors with definable parameters for each Detector Types pattern match with without position detection area test brightness area test contrast contour match with without position detection position data via EtherNet IP onl
49. lt 50 Zoont I off A Edge gradient Edge steepness Control Input cae 8 I Jeg Maximum number of pixels for the edge width within ates id x j which the above edge contrast difference should take Trigger Mode Triggered vi place Value range 6 19 default 12 Shutter Adjustment Manual vi Min edge length Minimum length of a contour element Position Control Off v For removing or allowing small contour lines i Value range 4 24 default 8 Shutter 1 21 ms 4 r 1 Search mode Complete search Searches the entire ROI for the Ker Delve ma 0 contour with the best match If multiple matches are Output Del 1 found the best match is identified Ap Dole irs L 0 Until 15 good Stops search as soon as first contour Output Duration ms 0 match is found Results in faster application speed but risks matching similar line contours in the ROI Output Active HIGH vi This parameter gives the user the capability to improve Edge contrast 50 the speed of the application at the expense of accuracy Edge gradient 12 Edge width Position tolerance Min chain length 8 7 Allows for more or less tolerance in matching a contour j Sr to the taught contour Defined in number of pixels Search mode Complete search v Used for deviations in shape Value options 3 5 7 9 default 5 Contour width 5 vi Min scale 1 0 7 Max scale 1 0 7 Starting angle 170 lt 7 Total angle 360 Threshold Contour oe 0 60 R
50. me A new part cannot initiate a trigger signal until the previous evaluation timing delays and output duration are complete Rockwell Autom ation 10000000877 02 41 MultiSight User Manual Start up Special Case Control of Outputs with Use of Output Duration Parameter Output Behavior with 0 ms Output Duration default setting Behavior with gt 0 ms Output Duration 0014 Ready This output shows when the sensor is ready to make an evaluation and when the results of the previous evaluation are valid Turned ON while sensor is waiting for an external trigger Turned OFF when an evaluation is initiated either with the external trigger or Trigger Delay after the external trigger Turned ON again when the result outputs OUT1 and OUT2 are valid This output shows when the sensor is ready to make an evaluation but DOES NOT show when the results of the previous evaluation are valid Turned ON while sensor is waiting for an external trigger Turned OFF when an evaluation is initiated either with the external trigger or Trigger Delay after the external trigger Turned ON again after the evaluation is processed AND the Output Duration has elapsed OUT1 and OUT2 are reset outlined below at the same time as OUT4 is turned ON This means that there is no output to confirm OUT1 and OUT2 data validity The PLC must account for the evaluation time before reading the output results OUT1 pass fail and
51. mods Tiggered E MultiSight will look for the taught pattern Shutter Adusiment Manual i i Position Control Of vi i icki 0 Shutter 3 12 ms 7 i Set pattern by double clicking on the red frame and placing it EE PER 0 on the image where the inspection needs to be performed E al z Display Output 9 Display 5 EA Dutput Active HIGH Position 7 Save Max Pattem Size 200x80 v Image Threshold Min 0 35 Threshold Max 1 00 In most cases the majority of the parameters can be set using the IMPORTANT standard settings A different setting may be necessary to improve detection or shorten cycle times Rockwell Autom ation 10000000877 02 53 MultiSight User Manual Setup Quick Guide Select Live Images check box in the Select Detector section Click the Run button The pattern is now evaluated continuously Click on the Stop button to stop live image TT Change the trigger mode to Continuous This makes it easier to check good and bad parts Le w si Click on the Teach and select Teach from PC image button to 1 KI i teach the pattern Click on the Stop button after a few i i seconds a 7 z HF Display z z Notice the taught pattern display in the Pattern Section Diep ie Position gt Max Pattern Size 0 v commands The desire
52. n Performance Features Vision sensor Easier handling and setup than traditional vision systems 640 x 480 resolution CCD chip Integrated white light illumination Image display on computer using included software Standalone sensor function Compact sturdy industrial housing IP67 with standard 8 pin micro M12 QD connectors Detection distance adjustable from 20 mm to infinity see Minimum Operating Distance and Field of View by Cat No on page 63 for corresponding field of view Four outputs PNP type sourcing MOSFETs two inputs sinking Short evaluation time Three or four evaluation methods Pattern Matching Brightness Contrast and Contour Matching Ten virtual detectors or 32 virtual detectors Individual virtual detectors can be logically linked or grouped for evaluation of different objects with several characteristics for inspection Ethernet connection for setup EtherNet IP connection for communication of 1 0 information RSLogix 5000 Add on Profile for ease of integration into a Logix based control system Description of Operation Basics The image recorder of the MultiSight is a CCD chip with a resolution of 640 x 480 pixels Evaluation is based on 320 x 240 scanning points The sensor is designed for inspection and cannot provide measurement information To operate the MultiSight at least one virtual detector must be configured by designating a Region of Interest and teaching a sample image see s
53. ockwell 38 10000000877 02 Automation MultiSight User Manual Start up Parameter Description Screen Shot Min scaling Minimum factor for scaling the size of the contour Values 0 5 1 0 default 1 0 Detector Type Contour Match v Max scaling Maximum factor for scaling the size of the contour Illumination lInemalOn Values 1 0 1 5 default 1 0 Zoom 4 Starting angle Angle at which the search starts Search in clockwise Control Input Enable v direction Tii igger Mode T d v Values 180 179 default 10 ee 0 12 00am on a clock Spate dement Manual Position Control Off Total angle Maximum search angle Used to reduce the angular positions that are inspected so that only a contour Shutter 1 21 ms with minimal displacement from the original angular Trigger Delay ms 0 position is matched E G The application is to Output Delay ms 0 confirm that a part is in the correct orientation so reduce the Total angle If the part is not in the correct Output Duration ms 0 orientation the contour will not be found Output Active HIGH v Threshold Contour2 Threshold Contour defines the minimum evaluation Edge contrast 50 score for the inspection to pass Threshold is adjusted Edge gradient 12 by the slider Maly Values 0 00 1 00 in increments of 0 05 Macan engh 8 The threshold value is depicted bya small vertical black Search mode C
54. omplete searct v line on the detector results bar graph when the sensor est is in run mode ae 5 x Min scale 1 0 Max scale 1 0 Starting angle 170 Total angle 360 Threshold Contour 0 60 Note The default settings for all Contour Matching parameters can be used for most applications 1 Parameter for teaching 2 Parameter for evaluation The angle information displayed in the Results area and available via the EtherNet IP connection is only a rough estimation To help improve the precision of angle data the Contour Matching parameters should use the following settings Edge contrast set value as high as possible Edge gradient set value as low as possible Min edge length set value as high as possible Search mode Complete Search Edge width set value as low as possible Rockwell Autom ation 10000000877 02 39 MultiSight User Manual Start up Configurable Delays There are three timing mechanisms that can be used Trigger Delay Output Delay and Output Duration Each has a range of 0 3000 ms The following timing diagrams show the behavior of the outputs for each timing mechanism condition Case 1 No delay timers used default state IN1 Trigger Evaluation Time of Sensor OUT4 Ready Busy Cla am Gil 5 Pass Fail Ready On Ouput Valid Case 2 Trigger Delay Time in ms between the Trigger Input turning on rising edge and the image exposure o IN1
55. ork SEHINGGA SA st NEG KN iiss cede tas as pina eae ey enue 11 Basic Settings for PC and MultiSight Sensor ve 12 Setting the IP Address of the PC 2 00 cee cece eee eaaa eaaa eee een aee aa eea 12 Setting the IP Address of the MultiSight 0 ccc e cece eee c eee e eee eaaa eaaa eea 23 15 Firmwate Updates ees a atdet ang bergen 18 Start Up 6 te Na a ra A NA Aa Nah oad aed dele OEE AE ior Greene peke dekade Sende 18 Program Description 22 MENU Bits gt ngae tg prat he kr tad NANDE Ale eens OR DEE Hege cts 22 FilG EXit Madre beta a Ree dale naam ame vidd 0 ariere karse at 22 HEIP HEIP Lasser anbe alen A A ad PAD fe pene oi iailen eee RAs pee E 22 Hlelp ABOUts sasana tcc kr 5 Auda 22 ProgramiInterface sunneste ga aw Mi Ba a a ede a DEI 23 Commands Image aaa aaa a ahaaa Jaa aa sahi maan a a Laga Aga NA kana I ana a daa baka stale a ada a Ta A asp te det 5 0 teten a NG RESUMES sas nasa seere San rs te der wane eye aga Setup Quick Guide Example Application RESET POWEr OM asset gore oe et dn eee eget ee ine Maintenances 6 ajana ket aga a BAGE BA ead skh weg ermet dere tee iin aa te jod ele alee ile hee 6 NA pa a rnc Mban log Game baad BA a Na KANG Ka KEN Frequently Asked Questions vi Technical Data ss
56. oroughly described in the above general description of the Select Detector window The state of the inspection output OUT1 passes fails LED Q1 is determined by how the individual detector evaluation results are linked in the Logic field The evaluation of logically linked detectors is based on a single image This means that certain detector parameters must be the same for all detectors that are logically linked together see the chart below The recommended procedure is to set the first detector in the group to the desired settings then use the Copy command in the Commands window on the left of the software screen see page 24 to copy the configuration of that detector to the other detectors that will be logically linked The software will give the error message group parameter check failed when the sensor is put into run mode if detectors that are linked together have nonmatching parameters that should match Select Detector Sensor name SEs Live Images Off v Logic AND All Recorder 1 pe O2 oes 03 MO 04 MO Os MO Os MO O7 DO Os OOP Os Obes Ow OOP 28 10000000877 02 Rockwell Automation MultiSight User Manual Start up Parameter Description Screen Shot Logic continued Select radio buttons 0 9 0 31 in order to configure the parameters for that detector Detectors that are logically linked should follow these guidelines for the
57. ossible The threshold value should be set approximately at the center of this difference between the Good and Bad parts Rockwell 10000000877 02 Automation MultiSight User Manual Start up Detector Types Detailed Description The following detector types are available Pattern Matching Brightness and Contrast The EtherNet IP model also has a Contour Matching detector type Pattern Matching With this method the detector looks for a pattern in a defined Region of Interest ROI By defining this ROI as larger than the pattern there is some tolerance in positioning of the inspected part The ROI is the yellow frame on the image and the pattern is defined by the red frame In run mode the sensor evaluates the ROI on the current image to see if the pattern is present The evaluated pattern must match the taught pattern to a degree of conformity greater than the minimum threshold value and less than the maximum threshold value Example The sensor is taught the following pattern In the next row of three patterns each pattern is recognized with 100 conformity because the evaluated pattern is exactly the same as the taught pattern even though itis in a different position in the Region of Interest ROI It differs only in direction X and Y and is not rotated In the next row of three patterns each pattern is recognized but with less than 100 conformity Pass or Fail result is dependant on the thre
58. ource Detector Configuration to the other selected detectors Write to Flash Memory Writes all detector parameter settings to nonvolatile flash memory so that they are maintained if the sensor loses power Rockwell Automation 10000000877 02 25 MultiSight User Manual Start up Command Description Recorder Images Transfers the images saved on the MultiSight to the computer The MultiSight can save up to 30 images on its local memory To save images the Recorder parameter in the Select Detector window must be turned on see page 27 Based on the setting of that parameter the MultiSight can save either the last 30 images or the last 30 images of inspections that resulted in an output of FAIL When the images are transferred to the computer they are deleted from the MultiSight memory In order to keep the image it must be saved to the computer using the Save button on the Saved Pics pop up window that opens when the images are transferred It is possible to save either the current image or all the images by using the Save or Save All buttons The images are stored as bitmap files bmp Images are saved in dynamic memory and are therefore lost if the sensor loses power or is turned off Select Detector The MultiSight can be configured with up to ten 32 virtual detectors i e inspection of ten 32 different characteristics In order to configure one of the ten 32 detectors selec
59. r 7 Blue GND V 8 Red OUT2 position display LED Q2 Indicates sensor evaluation is valid for OUT1 and OUT2 except in special cases noted in the Configurable Delays section on page 40 Control Input IN2 if the Control Input parameter Detector Parameter window is set to Enable default setting the control input IN2 must be connected to V 24V DC for the sensor to operate If the control input IN2 is not connected to V the MultiSight will not execute evaluations if the trigger mode is set to triggered standby mode see page Use of Control Input IN2 on page 58 Ethernet M12 D Code 4 Pin Connector 66 10000000877 02 Rockwell Automation MultiSight User Manual Maintenance Accessories Dovetail Bracket O 48MS BKTDT Angle Bracket 48MS BKTANG Rod Bracket 48MS BKTROD Mounting Rod 200 mm 48MS R0D200 Mounting Rod 400 mm 48MS R0D400 Rod Clamp 48MS CLAMP Mount Plate 48MS MTPLATE RJ45 Connector 48MS RJ45CONN Ethernet Crossover Cable RJ45 to RJ45 48MS ECROSS Area Light White Light 48MS ALWH Ring Light White Light 48MS RLWH Angle Bracket for Ring Light 48MS BKTANG2 MultiSight Test Box 48MS TESTBOX Standard Model Cordset 2m 48MS CS 2 Cordset 5m 48MS CS 5 Cordset 10 m 48MS CS 10 Cordset Right Angle 2 m 48MS CSRT 2 Cordset Right Angle 5 m 48MS CSRT 5 ENET Cordset Crossover 3 m 48MS ECSCROSS 3 ENET Cordset
60. r installation in the customer s plant and integration in the customer s computing system Rockwell Automation is not liable for any interference failure or damage to the customer s system caused by the installation or integration of this software These conditions are accepted by the customer without reservation by clicking the button at the start of the software IMPORTA installation process The MultiSight functions only with software supplied Software updates are available online at www ab com go multisight The PC Software must always match the corresponding sensor firmware Installation of older versions of the software using Windows 98 2000 and XP requires administration rights Please download the latest version of the software from the above location Rockwell a 10000000877 02 Automation MultiSight User Manual Start up To start the MultiSight proceed as follows 1 Switch on the power supply 24V DC for the MultiSight If necessary connect the control input IN2 with V 24V DO see note on page 11 For the first start up of the MultiSight install the supplied software on the computer that will be used for configuring the sensor When the CD is inserted the installation program starts automatically or can be started using the setup program Software MultiSight exe Then follow the instructions on the screen During installation the file MultiSight Sensor Finder EXE is created standard fold
61. r and I O Connection b Connector pinouts and a detailed wiring diagram are shown in the Connector Pin Assignments section on page 66 If the Control Input parameter of the active detector is set to Enable the default setting the control input IN2 must be connected to V 24V DC in order to operate the sensor with an external trigger input IN1 see page 66 Network Settings IMPORTANT The following instructions indicate how to change the network configuration of the PC and the MultiSight If incorrect settings are used the network links in the computer may be lost Note the current settings for later use before making any changes It may be necessary to re start the system during this procedure In order to determine which IP addresses are allowed in your network or locally in your PC and to carry out the necessary settings on your PC contact your system administrator The illustrations dialogues and menus originate from the operating system Microsoft Windows xP The illustrations are similar in other operating systems Rockwell Automation 10000000877 02 11 MultiSight User Manual Network Settings Basic Settings for PC and MultiSight To configure the MultiSight Sensor with a PCit is essential that Ethernet capability is installed in the PC and the LAN connection on the PC is active with a TCP IP protocol This also applies when the PC is not connected to a network see
62. racket 48MS BKTANG 5 0 19 45 0 17 Dia 9 0 35 8 10 Las 03 o39 60 0 07 20 0 78 4 0 48 H as 0 57 Rockwell 10000000877 02 Automation 69 MultiSight User Manual Maintenance Angle Bracket for Ring Light Rod Bracket 48MS BKTANG2 48MS BKTROD io 12 5 0 49 iH p i 1 10 0 39 ole Ev re 0 PU LL ka 32 1 26 1 20 0 78 r 10 0 34 M4 10X Mounting Rod 200 48MS ROD200 200 7 87 12 0 47 Dia Mounting Rod 400 48MS ROD400 15 74 400 P 0 47 Dia Rod Link 918 H7 8 473 Rockwell s 10000000877 02 Automation MultiSight User Manual Maintenance Rod Clamp 48MS CLAMP Mounting Plate 48MS MTPLATE 150 5 90 100 3 93 50 1 96 20 0 78 il 7 oy A y 7 EN 1 1 1 INA 1 1 17 EN Q ae if 1 0 43 m 6 8 0 26 r 1H 10 0 39 5 Ls 0 24 Dia 4X Rockwell 10000000877 02 Automation MultiSight User Manual Maintenance Area Light White Light 48MS ALWH 45 1 77 et 27 pe
63. riggered gt External trigger signal IN1 or the software command button Trigger initiates image exposure and evaluation Continuous gt A new evaluation is started as soon as the previous evaluation completes Shutter Adjustment Position Control Manual gt Manual setting of the shutter speed recommended Automatic Teach gt The shutter value is set automatically by the MultiSight when the Teach command is executed Activate evaluation of position deviation X and or Y pass fail output only positional information not available as a discrete output E position data is available via EtherNet IP Off gt Position evaluation de activated Only X gt Evaluation direction X Only Y gt Evaluation direction Y X Y rectangle gt Evaluation direction X and Y with rectangular deviation area blue frame X Y ellipse gt Evaluation direction X and Y with elliptical deviation area blue frame Position Control is only used with Detector Types Pattern Match and Contour Match When the pattern or contour is successfully detected i e OUT is active the Position Control analyzes whether the pattern is within the blue position frame The Position Output OUT2 is active when the pattern is detected and the center of the pattern the green cross in the center of the green detected pattern frame is inside the blue position frame The area for the position evaluation can be marked by double clicking the blue frame and adju
64. se malfunction EtherNet IP models only The supplied black PVC M12 covers must be installed around the coupling nut of each connection cable Power I O and ethernet to protect from ESD when using standard cables There is a plastic nut on the MultiSight specific cables to provide this ESD protection so the PVC cover is not required for 48MS CS 2 and 48MS ECSCROSS 3 LED Indicators Focus Adjustment M12 Connection Socket LL 24 V DC and I O om LO Pwr Err Q1 02 Focus D M12 Connection Socket ENET AT 8 pin on standard models 4 pin D code on EtherNet IP models Rockwell 10000000877 02 Automation MultiSight User Manual Installation Instructions Sensor Display LED Description Name Color Meaning Pw gren Operatingvoltage sid Err red Error 01 yellow Result 1 pass fail OUT1 02 yellow Result 2 position OUT2 Table 1 Table Display LEDs For electrical installation connect the MultiSight as shown IMPORTANT Use only the correct connection cables as listed below see Accessories on page 67 The Ethernet connection is only required for setup During normal operation only the PWR and I O cable needs to be connected unless using the EtherNet IP functionality Rockwell Automation 10000000877 02 9 MultiSight User Manual Installation Instr
65. shold settings degree of conformity Pattern detection is also tolerant of up to 5 of rotation The next two images would be recognized Due to the slightly rotated orientation of the pattern the conformity would be less than 100 even though there is 100 pixel conformity Patterns with more than 5 of rotation are not recognized Brightness The MultiSight uses an 8 bit grey scale CCD chip to capture images The chip assigns each pixel a value of 0 black to 255 white The Brightness Detector Type evaluates all pixels that are darker than grey scale value 128 as black The detector examines the Region of Interest ROI and determines the Brightness level White 100 Black 0 If this level is between the configured threshold minimum and maximum the detector passes the evaluation The shutter speed has a significant effect on the number of pixels which are evaluated as black and thus which images pass or fail this inspection Rockwell Autom ation 10000000877 02 35 MultiSight User Manual Start up For example here is the same image with three different shutter speeds Region of Interest Shutter setting Shutter Shutter Shutter 7 The setting of the threshold determines the maximum number of pixels detected as being dark which is required to pass the evaluation The position of each dark pixel within the ROI does not affect the evaluation Example These two images produce exactl
66. software Stop gt Configuration mode Configuration parameters can only be changed in this mode Trigger Save Manual triggering of MultiSight image evaluation active only in Run mode and if Trigger parameter is set to triggered see Detector Parameters on page 32 or when triggered image button in image section is pushed Save the detector configuration for one detector or all detectors Load Load previously saved detector configuration s Reset The configuration parameters for a single detector or for all detectors are reset to the default settings Copy Copy the configuration of one or many parameters from one Source Detector Configuration to other detectors Logically linked or grouped detectors must have identical settings for certain parameters see page 28 Recommended procedure Set the parameters of one detector to the desired configuration 2 Press the Copy button 3 In the Copy command pop up window select the configured detector from the Source Detector Configuration drop down list 4 Select the detectors along the left side of the window that need to match the Source Detector Configuration e g detectors that are logically linked or grouped with the source detector 5 Select Default Par at the bottom right of the window This automatically selects all the parameters that need to match for logically linked or grouped detectors 6 Press the OK button to copy the S
67. sting the position and shape using the mouse Approach the line with the mouse pointer until it changes to a horizontal or vertical double arrow Keeping the left hand mouse button pressed the line can now be moved to the desired position Rockwell Autom ation 10000000877 02 33 MultiSight User Manual Start up Parameter Description Shutter x xx ms Exposure time slide for setting shutter value manually Increase to brighten image decrease to darken image The shutter speed can also be adjusted by clicking on the number and typing a new value into the pop up window Trigger Delay ms Time between trigger and exposure evaluation in 0 3000 ms see Configurable Delays on page 40 Output Delay ms Time between trigger and availability of result on OUT1 and OUT2 in 0 3000 ms Output Duration ms Time in 0 3000 ms that the output state is maintained Please review the notes in the Configurable Delays section on page 40 Output Active HIGH or LOW inverting of OUT1 and OUT2 HIGH gt gt Passing evaluation results in the PNP output turning ON sourcing current Failing evaluation results in the output turning OFF normally open LOW lt Passing evaluation results in the PNP output turning OFF Failing evaluation results in the output turning ON normally closed The result indicators in the software do not change with this parameter i e green dot always indicates passing evaluation Threshold
68. t the appropriate numbered radio button in the Select Detector window The current configuration of that detector appears and can be edited in the Detector Parameters section If that detector on the MultiSight has not been previously configured the default values appear Parameters that are changed in the software are immediately changed on the MultiSight in dynamic memory In order to permanently save changes in Flash memory on the MultiSight click the Write to Flash button in commands window When the sensor is in run mode each detector is independently evaluated The results of these evaluations are indicated to the right of the radio buttons for each detector The small grey dots change color to represent the results of the individual detector evaluation with green indicating PASS and red indicating FAIL The first dot is for inspection pass fail OUT1 LED Q1 and the second dot is for position OUT2 LED Q2 While in run mode a green or blue line on the bar graph indicates the degree of similarity to the pattern or contour the sensor has been taught for that detector When using contrast or brightness detector types the bar graph indicates the level of contrast or brightness The black line on the bar graph represents the threshold configured in the Detector Parameters on page 34 Detectors can be logically linked together For example the sensor can be set up to inspect multiple components on a product Each component is inspected
69. tablish a network connection with the sensor and start the configuration software If no sensor can be found with the Find command it is possible to connect manually with function Add IP by filling in the known IP address of the sensor Click Yes to stop Run mode of the MultiSight and connect to the sensor The configuration software looks like this Sensor Allen Bradley 48MS MultiSight Sensor with EtherNet IP Allen Bradley HOPNPROQ0N0000000000000000000000 Q0000000000 PEPE FETE Recorder Images Fer Hebb press FY The configuration software can only be opened when the fi computer is connected to a powered and functional MultiSight If the software was able to make a connection with the MultiSight all the buttons are activated after a few seconds and the image currently seen by the sensor is displayed in the Image window If the buttons are not activated i e greyed out the software was not able to connect with the MultiSight Check cable connections and IP address of both the MultiSight and the computer If necessary set the IP address of the PC and or the sensor as described in the Network Settings section on page 11 Use the MultiSight Sensor Finder soft ware to attempt to reconnect If unable to connect read the Network settings section on page 11 and make the nec essary settings to the PC and or sensor Click the
70. teaching the pattern contour or shutter speed in automatic shutter adjustment mode The active detector in the software is the active detector in the sensor This means that if the sensor is switched to run mode whatever detector was most recently selected in the software will determine the output of the sensor unless it is changed via EtherNet IP If an EtherNet IP connection is established the Logix add on profile output tag FirstDetector will control which detector is active and override the configuration software setting To give control of which detector is active back to the configuration software set the FirstDetector tag value to 1 Installation Instructions Installation To help provide maximum accuracy of detection the MultiSight should be protected from vibration IMPORTANT Secure the connection cables with cable binders to prevent crushing or slipping Select a position for the MultiSight in which interfering factors such as slight differences in the position of the object or variations in lighting have little or no effect Attach the MultiSight onto the dovetail mounting bracket 48MS BKTDT supplied with the unit and then onto a suitable object above the target Use only the 48MS BKTDT for mounting the unit see Approximate Dimensions mm in on page 68 Rockwell 10000000877 02 Automation MultiSight User Manual Installation Instructions Operating io Distance Ma F
71. the side or corner and then clicking and dragging The red pattern frame within the ROI is used to teach the pattern or contour to be detected When the Teach button is clicked whatever pattern is in the red frame is taught to the sensor Double click inside the red frame to select it and then resize or move it In run mode the green match frame indicates where the pattern or contour was detected The blue frame is the position frame and only appears when the Position Control parameter is active The ROI pattern and position frames of non selected detectors will appear if the Display Icon check box in the Select Detector window is checked They are greyed out to differentiate them from the currently selected detector Live Display Displays the current live image of the MultiSight and continuously refreshes the image Zoom Magnifies the image with a digital zoom 2 1 ratio The ROI and Pattern frames are displayed and can be adjusted Save image Saves the current image as a bitmap file bmp with or without the ROI pattern and position frames Max Pattern Size Select a maximum pattern size of 400 x 160 160 x 400 or 248 x 248 pixels Rockwell Automation 10000000877 02 43 MultiSight User Manual Start up Screen Shot 2 SF Display Dy sage ina Ro as V Pattem m Position Zoom Max Pattern Size 400x160 Mm f 1
72. tour matching to detect or differentiate objects by means of previously defined optical characteristics e g for separating good and bad parts The main applications are in the field of industrial automation for quality assurance purposes The MultiSight can be used for detecting presence or absence completeness position markings labeling packaging and components The MultiSight is an economical alternative to conventional vision systems Requirements for Use To configure the MultiSight the following items are required PC or notebook with minimum Pentium III 1 GHz 256 MB RAM Microsoft Windows 98 2000 XP or Vista network connection or network with TCP IP protocol Recommended screen resolution 1024 x 768 pixels Basic knowledge of computing and electronics wiring is also essential The IP address of the MultiSight is set by the manufacturer to 192 168 100 100 with sub net mask 255 255 255 0 The operation of the MultiSight is independent of a computer or PLC A PC or notebook is required only for configuring the sensor For reproducible results and avoidance of malfunctions adequate and consistent object illumination is essential Reflections or varying incident light may affect detection results IMPORTANT If necessary use an external white light source and or light screening devices to exclude incident light Rockwell 10000000877 02 Automation MultiSight User Manual Description of Operatio
73. ts the mouse pointer to the pop up window This mode can only be de activated by placing the mouse pointer over the Result indicator the grey green or red circle next to the top bar graph and pressing the key combination ALT B at the same time Help Help Display the help file a pdf of the User Manual Help About Displays software version number Rockwell 59 10000000877002 Automation MultiSight User Manual Start up Program Interface The program interface is entirely displayed on one screen It is arranged into six windows as follows Commands Select Detector Detector Parameters Image Pattern Results Sensor Allen Bradley 48MS MultiSight Sensor with EtherNet IP Disconnect Sings laos Gro folios m Paten m Position Max Pattem Size 200x80 LiveGid None Allen Bradley Recorder Images For Help press FI Rockwell Autom ation 10000000877 02 23 MultiSight User Manual Start up Commands The buttons shown along the left hand side of the screen are for the general operation of the MultiSight administration of detector configuration and displaying other information on the program 1 If a button is not active i e greyed out then that command is not available in the current mode Command Description Connect Disconnect Connect to or disconnect from MultiSight see Setup Connect
74. uctions Network Connection a for standard models Direct connection of the MultiSight to a PC with a single crossover cable recommended Ethernet Cross over Cable A8MS ECSCROSS 3 48MS M12 Direct connection of the MultiSight to a PC with an industrial ethernet cable and a separate standard RJ 45 crossover cable Ethernet Connector 48MS RJ45Conn Ethernet Cross Cable 48MS Ecross Ethernet Cable 48MS ECS 3 RJ45 PC Connection of the MultiSight to a PC via a network Network e g Switch Hub etc Ethernet Cable Standard A8MS 0 48MS ECS 3 Network Cable us PC Network Connection a for EtherNet IP Models Direct connection of the MultiSight with EtherNet IP to a PC with an industrial ethernet cable and a separate standard RJ 45 crossover cable Ethernet Connector ik vesle 1585A DD4JD Ethernet ernet Cordse Cross Cable 2 48MS a 1585D M4TBDM 2 MIT TRIS 48MS ECROSS 6 PC 48MS M12CVR Rockwell i0 10000000877 02 Automation MultiSight User Manual Network Settings Connection of the MultiSight with EtherNet IP to a PC and PLC via a network EtherNet Cable 1585D M4TBDM 2 48MS M12 48MS M12CVR Network e g Switch Hub etc Ethernet Cable RJ45 RJ45 RJ45 1585J M8TBJM 2 pA Standard Network Cable Ethernet Connector 1585A DD4JD Ethernet Cable 1585J M8TBJM 2 PLC EtherNet M12 IP Scanner Powe
75. unters After every activation of Run mode the three counters are reset to zero For the purposes of these counters an evaluation cycle of five logically linked detectors resulting in a single output is counted as one evaluation Evaluations Displays the number of evaluation cycles since the last activation of Run mode Good Patterns Number of evaluation that resulted in OUT1 pass fail output of PASS Good Positions umber of evaluation that resulted in OUT2 position output of PASS Sort result Only when logic is set to OR Displays which detector had the highest evaluation score Rotation angle Only for contour match Displays the angle of the detected contour with reference to the taught contour position This data is not highly precise typical tolerance is 10 Status Line The Status Line along the bottom right edge of the screen shows the current status of the last action performed The Status Line indicates that the software is currently executing IMPORTANT a task It is best to not attempt additional tasks while the Status Bar indicates software activity 48 Rockwell 10000000877 02 Automation MultiSight User Manual Setup Quick Guide Setup Quick Guide Preparation 1 Double click the MultiSight software icon to launch the Sensor Finder software P MultiSight Sensor 2 The Sensor Finder software will show ALL MultiSight sensors on the local Eth
76. value causes slight overexposure but also produces a more consistent result Reflections occur when the alignment or position are changed slightly e g an impression which varies in brightness How can I get a more consistent image Mount the MultiSight at an angle of 10 15 so that it is at a slight angle to the reflecting surface see the illustration on page 7 Or select another position for the pattern screen in which the reflections do not adversely affect pattern detection see Installation on page 6 Rockwell 0 10000000877 02 Automation Technical Data MultiSight User Manual Maintenance Electrical Data Certifications cULus and CE Marked for all applicable directives Supply Voltage 24V DC 10 absolute maximum 18 30 V Residual Ripple lt 5Vpp Current Consumption no 1 0 200 mA max Inputs IN1 IN2 High 10 V V 24V DC 10 low 0 3 V Input Resistance gt 20 kOhm Trigger Input rising edge 10 V V 24V DC Outputs 0011 4 PNP type sourcing MOSFET Output Rating per output 200 mA Output Current per output 1 5 A max in case of short circuit Protection Type Short circuit all outputs inverted polarity electrical shock CLASS 2 EN 61140 Set up Interfaces Ethernet LAN Network Ethernet configuration only EtherNet IP data transfer only Power On Delay Approx 6 seconds Optical Data Pixe
77. ved lt b 332 837 491 1 595 396 586 10000000877 02 13 MultiSight User Manual Network Settings 5 Click on Properties 4 Local Area Connection Properties General Authentication Advanced Connect using E9 3Com 30920 Integrated Fast Etheme This connection uses the following items Sh Client for Microsoft Networks QoS Packet Scheduler File and Printer Sharing for Microsoft Networks Description Transmission Control Protocol intemet Protocol The default wide area network protocol that provides communication across diverse interconnected networks C Show icon in notification area when connected Notify me when this connection has limited or no connectivity 6 Click on Internet Protocol TCP IP properties Internet Protocol TCP IP Properties General You can get IP settings assigned automatically if your network supports this capability Otherwise you need to ask your network administrator for the appropriate IP settings Obtain an IP address automatically Use the following IP address IP address 192 168 100 101 Subnet mask 255 255 255 0 Default gateway Obtain DNS server address autom y Use the folowing DNS server addresses Preferred DNS server Akemate DNS server 7 Set the IP address and Subnet mask per the previous page i 10000000877
78. y Typical cycle time Pattern matching 50 100 ms Contrast or brightness area test 40 50 ms Contour matching 120 500 ms Use of Control Input IN2 This section details the operation of the control input IN2 For operation using the control input set the Control Input parameter in the Detector parameters window as described below IMPORTANT The detector which is activated by Reset or Power on is the one which was activated most recently using the radio buttons in the window Select detector of the PC software or via EtherNet IP Enable Case 1 on page 60 Activate 0 or 1 Case 2 Activate 0 to 31 Case 3 Teach temporary Case 4 Teach permanent Case 4 For the signal sequence see the diagram below Pin assignments are on page 66 Rockwell 10000000877 02 Automation MultiSight User Manual Maintenance IN 1 Trigger Out 4 Ready Out 1 Pass Fail Activate 1 or 2 IN2 case 2 1or2 S______ _ _______ I IN 2 case 3 Activate 01031 ETL typ sms 5ms I IN2 case 4 Teach temp perm A EL 1 Rising edge at INI Trigger starts sensor triggers exposure and processing 2 Trigger starts falling edge at OUT4 Ready Ready stays low for duration of evaluation Duration of evaluation is indicated in Run mode with Trigger mode Continuous in sax ms at the bottom right hand of the software screen 3 Therising edge at O
79. y represent what the MultiSight looks for in run mode When the MultiSight performs an inspection in run mode it identifies these line shapes and determines if they match the taught line contour The threshold settings affect how closely the found line shapes match the taught line shape As an example a solid black circle on a white background would cause a single circular contour line A black ring on a white background would cause a large circular contour line at the outer diameter and within it a smaller circular contour line the inner diameter There is a contour line tracing border between black and white The primary advantage of Contour Matching in comparison to Pattern Matching is that this detector type is fully tolerant of shape rotation i e the object may appear in any angular position and still be detected Another advantage is that the taught contour can be edited by the user to remove parts of the contour that are not to be used for the inspection The primary disadvantage of Contour Matching in comparison to Pattern Matching is that it cannot detect as complex a pattern Contour Matching is also typically slower To edit the taught line contour place the cursor over the pattern in the Pattern section of the software on the right hand side of the screen The cursor becomes an eraser The size of the eraser is determined by the Eraser Size parameter under the pattern It is recommended to use the Zoom button in the Pattern section
80. y the same result in the brightness evaluation because in each case 9 of the 25 pixels are detected as dark HE OG E E OU Contrast i The Contrast Detector Type evaluates the difference between the brightness levels of the pixels in the ROI An image that is all one shade of one color has low contrast An image that is half white and half black has very high contrast Example Summary of Brightness and Contrast Detector Types Detector Type Brightness Contrast ROI Image bar graph bar graph 100 lt 10 ee ee al 50 gt 90 k 0 Ei For Contrast and Brightness Detector Types the teach function is only used for Automatic Shutter Adjustment It is recommended that the Shutter Adjustment parameter is set to Manual in which case the Teach command has no effect Rockwell ag 10000000877 02 Automation MultiSight User Manual Start up Contour Matching The Contour Matching detector type is essentially a simplified pattern matching algorithm Instead of the fully detailed pattern with relative grey scale levels and complex area evaluation used for comparison in the Pattern Matching detector type the Contour Matching detector type compares simple line contours and shapes In contour mode the teach process identifies the borders between dark and bright areas on the image These borders between dark and light are highlighted with a blue line and subsequentl

Download Pdf Manuals

image

Related Search

Related Contents

Operating Instructions Series    Liebert MR1 User's Manual  Tutorial do GeoOffice  PortMaster® 4 User Manual Addendum  取扱説明書  ADaM 2013 User Guide - Drake Lane Associates  JVC HR-J261EU User's Manual  SAVE VTR 150/K Installazione e assistenza  Summary  

Copyright © All rights reserved.
Failed to retrieve file