Home

Handyscope HS5 instrument manual

image

Contents

1. Digital external trigger Input Auxilary 1 O connector Range 0 to 2 5 V TTL Coupling DC Jitter Depending on source and sample frequency Source channel lt 1 sample Source external or generator Sample frequency 500 MS s lt 8 samples Sample frequency lt 500 MS s lt 4 samples Chapter 9 9 4 Arbitrary Waveform Generator Output channel 1 analog BNC DAC resolution 14 bit 240 MS s Output range 12 to 12 V open circuit Amplitude Range 0 12 V 1 2 V 12 V open circuit Resolution 12 bit Accuracy 0 4 of range DC offset Range 12 V to 12 V open circuit Resolution 12 bit Accuracy 0 4 of range Noise level 0 12 V 900 uVrms 12V 1 3 mVRMS 12 V 1 5 mVRMS Coupling DC Impedance 50 Q Overload protection Output turns off when overload is applied Instrument will tolerate a short circuit to ground indefinitely System Trueform CDS Memory Standard model 256 KiSamples XM option 64 MiSamples Operating modes Continuous triggered gated Sampling rate 240 MS s 200 MS s 100 MS s or 50 MS s depend ing on model Sampling source Internal TCXO Accuracy 0 0001 Stability 1 ppm over 0 C to 55 C Time base aging 1 ppm per year Waveforms Standard Sine square triangle pulse noise DC Built in arbitrary Exponential Rise and Fall Sin x x Cardiac
2. Figure 4 3 Driver install Instrument is still connected Clicking No will bring back the previous screen The instru ment should now be disconnected Then the removal of the existing driver can be continued by clicking Next Clicking Yes will ignore the fact that the instrument is still connected and continue removal of the old driver This option is not recommended as removal may fail after which installation of the new driver may fail as well When no existing driver was found or the existing driver is removed the location for the pre installation of the new driver can be selected Driver installation 13 Setup TiePie engineering instrument driver Select Destination Location Where should TiePie engineering instrument driver be installed A Setup will install TiePie engineering instrument driver into the following folder To continue click Next If you would like to select a different folder click Browse C Program Files TiePie TiePie instrument criver Browse Atleast 1 1 MB of free disk space is required 5 en ers Figure 4 4 Driver install step 3 On Windows XP the installation may inform about the drivers not being Windows Logo Tested The driver is not causing any danger for your system and can be safely installed Please ignore this warning and continue the installation is Setup TiePie engineering instrument driver Information Please
3. Table 3 1 Maximum sampling frequencies The Handyscope HS5 supports high speed continuous streaming measurements The maximum streaming rates are Introduction resolution channels model 530 model 220 model 110 model 055 E CH1 20 MS s 10 MS s 5 MS s 2 MS s i CH1 CH2 10 MS s 5 MS s 2 MS s 1 MS s ai CH1 20 MS s 10 MS s 5 MS s 2 MS s it CH1 CH2 10 MS s 5 MS s 2 MS s 1 MS s CH1 16 bit 6 25 MS s 3 125 MS s 1 25 MS s 625 kS s CH1 CH2 Table 3 2 Maximum streaming rates The Handyscope HS5 is available with two memory configura tions these are memory model 530 model 220 model 110 model 055 standard model 128 KiS 128 KiS 128 KiS 128 KiS option XM 32 MiS 32 MiS 32 MiS 32 MiS Table 3 3 Maximum record lengths per channel Optionally available for the Handyscope HS5 are SureConnect connection test and resistance measurement SureConnect connec tion test tells you immediately whether your test probe or clip actually makes electrical contact or not No more doubt whether your probe doesn t make contact or there really is no signal This is useful when surfaces are oxidized and your probe cannot get a good electrical contact Simply activate the SureConnect and you know whether there is contact or not Also when back probing con nectors in confined places SureConnect immediately shows whether the probes make contact or not Models of the Handyscope HS5 with SureConnect c
4. Chapter 5 6 3 5 4 ao TiePie instrument xx Device driver software installed successfully Figure 5 2 Hardware install step 2 Once ready Windows will report that the driver is installed Your device is ready to use gt X Device driver software installed successfully Figure 5 3 Hardware install step 3 Now the driver is installed the measurement software can be installed and the Handyscope HS5 can be used Plug into a different USB port When the Handyscope HS5 is plugged into a different USB port some Windows versions will treat the Handyscope HS5 as different hardware and will ask to install the drivers again This is controlled by Microsoft Windows and is not caused by TiePie engineering Operating conditions The Handyscope HS5 is ready for use as soon as the software is started However to achieve rated accuracy allow the instrument to settle for 20 minutes If the instrument has been subjected to extreme temperatures allow additional time for internal tempera tures to stabilize Because of temperature compensated calibration the Handyscope HS5 will settle within specified accuracy regardless of the surrounding temperature Hardware installation Chapter 5 Combining instruments When more channels are required than one instrument can offer multiple instruments can be combined into a larger combined in strument To combine two or more instruments the instruments need to be connected
5. read the following important information before continuing On Windows XP and newer the installation may inform about the drivers not being Windows Logo tested Please ignore the warning and continue installation Ges Ne ees Figure 4 5 Driver install step 4 Chapter 4 The driver install utility now has enough information and can install the drivers Clicking Install will remove existing drivers and install the new driver A remove entry for the new driver is added to the software applet in the Windows control panel 13 Setup TiePie engineering instrument driver Ready to Install Setup is now ready to begin installing TiePie engineering instrument driver on your computer Click Install to continue with the installation or click Back if you want to review or change any settings Destination location C Program Files TiePie TiePie instrument driver Figure 4 6 Driver install step 5 13 Setup TiePie engineering instrument driver Completing the TiePie engineering instrument driver Setup Wizard Setup has finished installing TiePie engineering instrument driver on your computer Click Finish to exit Setup Figure 4 7 Driver install Finished Driver installation Chapter 4 5 1 5 1 1 Hardware installation Drivers have to be installed before the Handyscope HS5 is connected to the computer for the first time See c
6. 0 01 Settling time lt 8 ns to 10 final value Jitter RMS lt 50 ps Chapter 9 9 5 Arbitrary Waveform Generator continued Burst Waveforms Sine square triangle noise arbitrary Count 1 to 65535 Trigger Software external Sweep Available only on models with extended memory option XM Waveforms Sine square triangle noise arbitrary Type Linear logarithmic Count Up down Trigger Software external Modulation AM Carrier waveforms Sine square triangle arbitrary Modulating waveforms Sine square triangle noise arbitrary Modulating frequency 2 mHz to 20 MHz Depth 0 0 to 100 Source Internal FM Carrier waveforms Sine square triangle arbitrary Modulating waveforms Sine square triangle noise arbitrary Modulating frequency 2 mHz to 20 MHz Peak deviation DC to 20MHz Source Internal FSK Carrier waveforms Sine square triangle arbitrary Modulating waveforms 50 duty cycle square Modulating frequency 2 mHz to 20 MHz Peak deviation 1 uHz to 20MHz Source Internal Power Power From USB or external input Consumption 5 Voc 500 mA max Power adapter External Input 110 to 240 Vac 50 to 60 Hz 0 85 A Max 50 VA to 80 VA Output 5 5 Vpc 2A Dimension Height 30 mm 1 2 Width 45 mm 1 8 Length 75 mm 3 Order number
7. 3 3 with a higher sampling frequency the original signal can be reconstructed much better from the measured samples Figure 3 3 The effect of the sampling frequency The sampling frequency must be higher than 2 times the high est frequency in the input signal This is called the Nyquist fre quency Theoretically it is possible to reconstruct the input signal with more than 2 samples per period In practice 10 to 20 sam ples per period are recommended to be able to examine the signal thoroughly Aliasing When sampling an analog signal with a certain sampling frequency signals appear in the output with frequencies equal to the sum and difference of the signal frequency and multiples of the sampling frequency For example when the sampling frequency is 1000 Hz and the signal frequency is 1250 Hz the following signal frequencies will be present in the output data Multiple of sampling frequency 1250 Hz signal 1250 Hz signal 1000 1000 1250 250 1000 1250 2250 0 0 1250 1250 0 1250 1250 1000 1000 1250 2250 1000 1250 250 2000 2000 1250 3250 2000 1250 750 Table 3 4 Aliasing As stated before when sampling a signal only frequencies lower than half the sampling frequency can be reconstructed In this case the sampling frequency is 1000 Hz so we can we only observe signals with a frequency ranging from 0 to 500 Hz This means that from the resulting frequencies in the table we c
8. EM 2 65544 edus Connect the instrument to the computer 5 2 1 Found New Hardware Wizard Plug into a different USB port Operating Conditions N aa aa au wen a 6 Combining instruments 7 Front panel real 7 2 1 3 CH1 and CH2 input connectors AWG output connector ss a Power indicators cnesta Ee 2a GE OR EE EEUE 8 Rear panel 8 1 8 2 8 3 PEER 1 Ge ome are he RE DE SO AR A 8 1 1 Power adapter o es saes nasas ewi 8 1 2 USB power cable USB pa a wann a ae N Extension Connector s aoso SS SS ss 13 13 13 13 19 19 20 20 21 21 23 25 25 25 25 Contents 84 AUR TO Sen en EE Re 30 Specifications 3l 9 1 Acquisition system 2 2 2 2 nn 31 9 2 Acquisition system continued 32 9 3 Trigger system cy aka mas ee 32 9 4 Arbitrary Waveform Generator 33 9 5 POWER ce peu pa aak a e oe i e e 35 DG PODES ses Sue ne een she 36 gr Physical ss DR aan 36 g8 TIO ie eo AO RE ER EE EF 36 99 Interface AI 36 9 10 System requirements 2 22 2 nn nenn 37 9 11 Environmental conditions 37 9 12 Certifications and Compliances 37 9 13 Package contents ooa o redoa otada Paadi 37 Safety When working with electricity no instrument can guaran tee complete safety It is the responsibility of the person who works with the instrument to operate it in a save way Maximum security is achieved
9. Handyscope HS5 User manual rl TiePie engineering engineering ATTENTION Measuring directly on the line voltage can be very dangerous The outside of the BNC connectors at the Handyscope HS5 are connected with the ground of the computer Use a good isolation transformer or a differential probe when measuring at the line volt age or at grounded power supplies A short circuit current will flow if the ground of the Handyscope HS5 is connected to a positive voltage This short circuit current can damage both the Handyscope HS5 and the computer Copyright 2013 TiePie engineering All rights reserved Revision 2 8 November 2013 Despite the care taken for the compilation of this user manual TiePie engineering can not be held responsible for any damage resulting from errors that may appear in this manual Contents 1 Safety 2 Declaration of conformity 3 Introduction 3 1 3 2 3 3 3 4 3 5 DAMPA on gba Hal be be Pa de a lw le ee Sample frequency 44 264 ch a a ei 3231 Allasne ee ge RR RR RE eS EE he Dieitizine 22 ke MR eA dd OR wd oe S Signal coupling saoe si eui eee run nen Probe compensation gt so otaa s tanab sews 4 Driver installation 4 1 4 2 4 3 Introduction ridad e ae Where to find the driver setup Executing the installation utility 5 Hardware installation 5 1 5 2 5 3 5 4 Power the instrument 5 1 1 External power R
10. Haver sine Lorentz D Lorentz Specifications Arbitrary Waveform Generator continued Signal characteristics Sine Frequency range 1 uHz to 5 10 20 or 30 MHz depending on model Amplitude flattness Relative to 1 kHz lt 100 kHz 0 1 dB lt 5 MHz 0 15 dB lt 20 MHz 0 3 dB lt 30 MHz 0 4 dB Spurious lt 100 kHz 75 dB 100 kHz to 1 MHz 70 dB 1 MHz to 10 MHz 60 dB 10 MHz to 15 MHz 55 dB 15 MHz to 20 MHz 45 dB 20 MHz to 30 MHz 35 dB Square Frequency range 1 uHz to 5 10 20 or 30 MHz depending on model Rise fall time lt 8 ns Overshoot lt 1 Variable duty cycle 0 01 to 99 99 Asymmetry lt 0 of period 5 ns 50 duty cycle Jitter RMS lt 50 ps Triangle Frequency range 1 uHz to 5 10 20 or 30 MHz depending on model Nonlinearity of peak output lt 0 01 Symmetry 0 to 100 0 1 steps Pulse Period 100 ns to 1000 s Pulse width 15 ns to 1000 s Variable edge time 20 ns to 1 s Overshoot lt 1 Jitter RMS lt 50 ps Noise Bandwidth typical 30 MHz Arbitrary Frequency range 1 uHz to 5 10 20 or 30 MHz depending on model Length 1 to 64 MiSamples Sample rate model HS5 530 240 MS s model HS5 220 200 MS s model HS5 110 100 MS s model HS5 055 50 MS s Rise fall time lt 8 ns Nonlinearity of peak output lt
11. Hz 1 A The Auxiliary I O connectors use HDMI type C sockets but are not HDMI compliant They can not be used to connect an HDMI device to the Handyscope HS5 Chapter 8 9 1 Specifications To achieve rated accuracy allow the instrument to settle for 20 min utes When subjected to extreme temperatures allow additional time for internal temperatures to stabilize Because of temperature compensated calibration the Handyscope HS5 will settle within specified accuracy regardless of the surrounding temperature Acquisition system Number of input channels 2 analog CH1 CH2 BNC Type Single ended Resolution 12 14 16 bit user selectable Accuracy 0 25 1 LSB of full scale Range 200 mV to 80 V full scale Coupling AC DC Impedance 1 MQ 25 pF Maximum voltage 200 V DC AC peak lt 10 kHz Maximum voltage 1 10 probe 600 V DC AC peak lt 10 kHz Bandwidth 3dB at 75 of full scale input Chi 250 MHz Ch2 100 MHz AC coupling cut off freq 3dB 1 5 Hz SureConnect Optionally available option S Maximum voltage on connection 200 V DC AC peak lt 10 kHz Resistance measurement Optionally available option S Ranges 100 Ohm to 2 MOhm full scale Accuracy 3 Response time to 95 lt 5 ms Specifications 9 2 Acquisition system continued Maximum sampling rate Depending on model 12 b
12. TP UE15WCP1 055200SPA Replaceable mains plugs for EU US AU UK Specifications 9 6 9 7 9 8 9 9 Probes Model HP 9250 Bandwidth 1 1 6 MHz 1 10 250 MHz Rise time 1 1 58 ns 1 10 1 4 ns Input impedance 1 1 1 MQ oscilloscope impedance 1 10 10 MQ incl 1 MQ oscilloscope impedance Input capacitance 1 1 47 pF oscilloscope capacitance 1 10 17 pF Compensation range 1 1 1 10 10 to 35 pF Working voltage 1 1 300 V CAT I 150 V CAT II DC peak AC 1 10 600 V CAT I 300 V CAT II DC peak AC Physical Height 25 mm 1 0 Length 170 mm 6 7 Width 140 mm 5 2 Weight 430 g 15 ounce USB cord length 1 8 m 70 I O connectors CH1 CH2 BNC AWG BNC USB Fixed cable with USB type A plug 1 8 m Extension connector D sub 9 pins female Power 3 5 mm power socket Auxiliary I O connectors 1 2 HDMI type C socket Interface Interface USB 2 0 High Speed 480 Mbit s USB 1 1 Full Speed 12 Mbit s compatible Chapter 9 9 10 9 11 9 12 9 13 System requirements PC 1 O connection USB 1 1 USB 2 0 or newer Operating System Windows 2000 XP Vista 7 8 32 and 64 bits Environmental conditions Operating Ambient temperature 0 C to 55 C Relative humidity 10 to 90 non condensing Storage Ambien
13. This will install TiePie engineering instrument driver 6 0 0 0 on your computer It is recommended that you close all other applications before continuing Click Next to continue or Cancel to exit Setup Figure 4 1 Driver install step 1 When drivers were already installed the install utility will re move them before installing the new driver To remove the old driver successfully it is essential that the Handyscope HS5 is disconnected from the computer prior to starting the driver install utility When the Handyscope HS5 is used with an external power supply this must be disconnected too Chapter 4 p Setup TiePie engineering instrument driver Information Please read the following important information before continuing Before installing the new drivers existing old drivers will be uninstalled from the computer Therefore make sure that all TiePie engineering instrument devices are disconnected from the computer When any instrument is used with an external power supply make sure that the extemal power supply is disconnected from the instrument as well Figure 4 2 Driver install step 2 When the instrument is still connected the driver install utility will recognize it and report this You will be asked to continue anyway Setup f Ej There still appears to be a TiePie engineering instrument connected to EP the system Make sure to disconnect it Continue anyway
14. an only see the 250 Hz signal in the sampled data This signal is called an alias of the original signal If the sampling frequency is lower than twice the frequency of the input signal aliasing will occur The following illustration shows what happens Y Tin 0 8 ms AAC Tsample 1 ms Tapparent 4 ms Figure 3 4 Aliasing In figure 3 4 the green input signal top is a triangular signal with a frequency of 1 25 kHz The signal is sampled with a fre quency of 1 kHz The corresponding sampling interval is 1 1000Hz Introduction EN lms The positions at which the signal is sampled are depicted with the blue dots The red dotted signal bottom is the result of the reconstruction The period time of this triangular signal appears to be 4 ms which corresponds to an apparent frequency alias of 250 Hz 1 25 kHz 1 kHz A To avoid aliasing always start measuring at the highest sam pling frequency and lower the sampling frequency if required 3 3 Digitizing When digitizing the samples the voltage at each sample time is converted to a number This is done by comparing the voltage with a number of levels The resulting number is the number cor responding to the level that is closest to the voltage The number of levels is determined by the resolution according to the following relation LevelCount 2Resolution The higher the resolution the more levels are available and the more accurate the input si
15. are installation HS5 Refer to paragraph 8 1 for specifications of the external power intput When the Arbitrary Waveform Generator is used the power that the Handyscope HS5 requires may strongly increase It is rec ommended to use the external power supply when the Handyscope HS5 Arbitrary Waveform Generator is used 5 2 Connect the instrument to the computer After the new driver has been pre installed see chapter 4 the Handyscope HS5 can be connected to the computer When the Handyscope HS5 is connected to a USB port of the computer Windows will report new hardware The Found New Hardware Wizard will appear Depending on the Windows version the New Hardware Wizard will show a number of screens in which it will ask for information regarding the drivers of the newly found hardware The appearance of the dialogs will differ for each Windows version and might be different on the computer where the Handyscope HS5 is installed O The driver consists of two parts which are installed sepa rately Once the first part is installed the installation of the second part will start automatically Installation of the second part is identical to the first part therefore they are not described individ ually here 5 2 1 Found New Hardware Wizard 3 Installing device driver software gt Click here for status Figure 5 1 Hardware install step 1 Windows has detected new hardware and starts installing the drivers
16. by selecting the proper in struments and following save working procedures Save working tips are given below Always work according local regulations Work on installations with voltages higher than 25 Vac or 60 Vpc should only be performed by qualified personnel Avoid working alone e Observe all indications on the Handyscope HS5 before con necting any wiring Check the probes test leads for damages Do not use them if they are damaged Take care when measuring at voltages higher than 25 Vac or 60 Vpc Do not operate the equipment in an explosive atmosphere or in the presence of flammable gases or fumes Do not use the equipment if it does not operate properly Have the equipment inspected by qualified service personal If necessary return the equipment to TiePie engineering for service and repair to ensure that safety features are main tained Measuring directly on the line voltage can be very danger ous The outside of the BNC connectors at the Handy scope HS5 are connected with the ground of the computer Use a good isolation transformer or a differential probe when measuring at the line voltage or at grounded power sup plies A short circuit current will flow if the ground of the Handyscope HS5 is connected to a positive voltage This short circuit current can damage both the Handyscope HS5 and the computer Safety Declaration of conformity J TiePie TiePie engineering Koperslagersstraat 37 8601 WL Sneek The Ne
17. d as external digital trigger input for the acquisition system and or the generator of the Handyscope HS5 Also each pin can be configured to output one of the following function generator outputs e Generator start e Generator stop e Generator new period The I C pins are equipped with internal 2 2 kOhm pull up re sistors connected to 3 V Rear panel 29 Pin 8 Power OUT has the same potential as the Handyscope HS5 power supply When the Handyscope HS5 is USB powered it is at USB power level When the Handyscope HS5 is externally powered it is at the same level as the external power input 8 4 AUX I O The Handyscope HS5 has two Auxiliary I O connectors at the rear of the instrument These can be used to combine multiple instru ments to a single combined instrument to perform synchronized Figure 8 6 Auxiliary I O connector measurements Pin Description Pin Description 1 GND 11 Data OK P EXT LVDS 2 EXT CLK P LVDS 12 Data OK N EXT LVDS 3 EXT CLK N LVDS 13 reserved 4 GND 14 Generator Trigger I O 5 Data OK I O 15 reserved 6 reserved 16 reserved 7 GND 17 GND 8 Ext Trigger I O 18 reserved 9 reserved 19 GND 10 GND Table 8 3 Pin description Auxiliary I O connector The I O signals pins 5 8 and 14 can be used as input and output They are digital signals switching between 0 V and 2 5 V The LVDS external clock pins 2 and 3 must be 10 M
18. electrical and electronic equipment WEEE Ee Restriction of Hazardous Substances The Handyscope HS5 has been classified as Monitoring and Con trol eguipment and is outside the scope of the 2002 95 EC RoHS Directive AN Introduction Before using the Handyscope HS5 first read chapter 1 about safety Many technicians investigate electrical signals Though the measurement may not be electrical the physical variable is of ten converted to an electrical signal with a special transducer Common transducers are accelerometers pressure probes current clamps and temperature probes The advantages of converting the physical parameters to electrical signals are large since many in struments for examining electrical signals are available The Handyscope HS5 is a portable two channel measuring in strument with Arbitrary Waveform Generator The Handyscope HS5 is available in several models with different maximum sam pling frequencies 50 MS s 100 MS s 200 MS s or 500 MS s The native resolutions are 12 bits and 14 bits and a user selectable resolution of 16 bits is available too with adjusted maximum sam pling frequencies resolution channels model 530 model 220 model 110 model 055 ee CH1 500 MS s 200 MS s 100 MS s 50 MS s it CH1 CH2 200 MS s 100 MS s 50 MS s 20 MS s CH1 14 bit 100 MS s 50 MS s 20 MS s 10 MS s CH1 CH2 CH1 16 bit 6 25 MS s 3 125 MS s 1 25 MS s 625 kS s CH1 CH2
19. gnal can be reconstructed In figure 3 5 the same signal is digitized using two different amounts of levels 16 4 bit and 64 6 bit LA SE EES Figure 3 5 The effect of the resolution The Handyscope HS5 measures at e g 14 bit resolution 2 4 16384 levels The smallest detectable voltage step depends on the input Chapter 3 3 4 3 5 range This voltage can be calculated as VoltageStep FullInput Range LevelCount For example the 200 mV range ranges from 200 mV to 200 mV therefore the full range is 400 mV This results in a smallest detectable voltage step of 0 400V 16384 24 41 uV Signal coupling The Handyscope HS5 has two different settings for the signal cou pling AC and DC In the setting DC the signal is directly coupled to the input circuit All signal components available in the input signal will arrive at the input circuit and will be measured In the setting AC a capacitor will be placed between the input connector and the input circuit This capacitor will block all DC components of the input signal and let all AC components pass through This can be used to remove a large DC component of the input signal to be able to measure a small AC component at high resolution When measuring DC signals make sure to set the signal coupling of the input to DC Probe compensation The Handyscope HS5 is shipped with a probe for each in
20. hapter 4 for more information Power the instrument The Handyscope HS5 is powered by the USB no external power supply is required Only connect the Handyscope HS5 to a bus powered USB port otherwise it may not get enough power to op erate properly External power In certain cases the Handyscope HS5 cannot get enough power from the USB port When a Handyscope HS5 is connected to a USB port powering the hardware will result in an inrush current higher than the nominal current After the inrush current the current will stabilize at the nominal current USB ports have a maximum limit for both the inrush current peak and the nominal current When either of them is exceeded the USB port will be switched off As a result the connection to the Handyscope HS5 will be lost Most USB ports can supply enough current for the Handyscope HS5 to work without an external power supply but this is not always the case Some battery operated portable computers or bus powered USB hubs do not supply enough current The exact value at which the power is switched off varies per USB controller so it is possible that the Handyscope HS5 functions properly on one computer but does not on another The Handyscope HS5 comes with a universal power supply that can be connected to a power outlet using the appropriate adapter The 3 5 mm connector attached to the power supply must be plugged into the power connector at the rear of the Handyscope Hardw
21. instrument through USB When the driver is not installed or an old no longer compatible version of the driver is installed the software will not be able to operate the Handyscope HS5 properly or even detect it at all The installation of the USB driver is done in a few steps Firstly the driver has to be pre installed by the driver setup program This makes sure that all required files are located where Windows can find them When the instrument is plugged in Windows will detect new hardware and install the required drivers Where to find the driver setup The driver setup program and measurement software can be found in the download section on TiePie engineering s website and on the CD ROM that came with the instrument It is recommended to install the latest version of the software and USB driver from the website This will guarantee the latest features are included Executing the installation utility To start the driver installation execute the downloaded driver setup program or the one on the CD ROM that came with the instrument The driver install utility can be used for a first time Driver installation installation of a driver on a system and also to update an existing driver The screen shots in this description may differ from the ones displayed on your computer depending on the Windows version 13 Setup TiePie engineering instrument driver Welcome to the TiePie engineering instrument driver Setup Wizard
22. it measuring one channel 500 MS s 200 MS s 100 MS s or 50 MS s 12 bit measuring two channels 200 MS s 100 MS s 50 MS s or 20 MS s 14 bit 100 MS s 50 MS s 20 MS s or 10 MS s 16 bit 6 25 MS s 3 125 MS s 1 25 MS s or 625 kS s Maximum streaming rate Depending on model 12 14 bit measuring one channel 20 MS s 10 MS s 5 MS s or 2 MS s 12 14 bit measuring two channels 10 MS s 5 MS s 2 MS s or 1 MS s 16 bit 6 25 MS s 3 125 MS s 1 25 MS s or 625 kS s Sampling source Internal TCXO Accuracy 0 0001 Stability 1 ppm over 0 C to 55 C Time base aging 1 ppm per year time base aging External LVDS on auxilary connectors Input range 10 MHz Memory Standard model 128 KiSamples per channel XM option 32 MSamples per channel 64 MSamples when measuring CH1 at 500 MS s 9 3 Trigger system System Digital 2 levels Source CH1 CH2 digital external OR generator start generator new period generator stop Trigger modes Rising edge falling edge any edge inside window outside window drop inside window drop outside window pulse width Level adjustment 0 to 100 of full scale Hysteresis adjustment 0 to 100 of full scale Resolution 0 024 12 bits 0 006 14 16 bits Pre trigger 0 to 32 MSamples 1 sample resolution Post trigger 0 to 32 MSamples 1 sample resolution
23. mains power net that supplies 100 240 Vac 50 60 Hz The external power adapter can be connected to the dedicated power connector Figure 8 3 Power adapter 8 1 2 USB power cable A special USB external power cable is supplied with the Handy scope HS5 that can be used instead of a power adapter One end of this cable can be connected to a second USB port on the computer the other end can be plugged in the dedicated power connector at the rear of the instrument The power for the instrument will then be taken from two USB ports Figure 8 4 USB power cable Chapter 8 8 2 USB The Handyscope HS5 is equipped with a USB 2 0 High speed 480 Mbit s interface with a fixed cable with type A plug It will also work on a computer with a USB 1 1 interface but will then operate at 12 Mbit s 8 3 Extension Connector EE 6 Fr o o co en Oo Figure 8 5 Extension connector A 9 pin female D sub connector is available at the back of the Handyscope HS5 containing the following signals Pin Description Pin Description 1 EXT 1 LVTTL 6 GND 2 EXT 2 LVTTL 7 Power IN 3 EXT 3 LVTTL 8 Power OUT see description 4 PC SDA 9 reserved 5 PC SCL Table 8 2 Pin description Extension connector Pins EXT 1 EXT 2 and EXT 3 are equipped with internal 1 kOhm pull up resistors to 2 5 V These pins can simultaneously be used as inputs and outputs Each pin can be configure
24. ome with re sistance measurement on all channels Resistances up to 2 MOhm can be measured directly Resistance can be shown in meter dis plays and can also be plotted versus time in a graph creating an Ohm scope With the accompanying software the Handyscope HS5 can be used as an oscilloscope a spectrum analyzer a true RMS voltmeter or a transient recorder All instruments measure by sampling the input signals digitizing the values process them save them and display them ka Chapter 3 3 1 Sampling When sampling the input signal samples are taken at fixed inter vals At these intervals the size of the input signal is converted to a number The accuracy of this number depends on the resolution of the instrument The higher the resolution the smaller the voltage steps in which the input range of the instrument is divided The acquired numbers can be used for various purposes e g to create a graph Figure 3 1 Sampling The sine wave in figure 3 1 is sampled at the dot positions By connecting the adjacent samples the original signal can be recon structed from the samples You can see the result in figure 3 2 Figure 3 2 connecting the samples Introduction 3 2 3 2 1 Sample frequency The rate at which the samples are taken is called the sampling frequency the number of samples per second A higher sampling frequency corresponds to a shorter interval between the samples As is visible in figure
25. or is the output of the internal Arbitrary Waveform Generator The outside of this BNC connector is con nected to the ground of the Handyscope HS5 Power indicator A power indicator is situated at the top cover of the instrument It is lit when the Handyscope HS5 is powered Front panel 25 Chapter 7 Rear panel OET 1 2 5 15VDC AUX VO AUX I O Figure 8 1 Rear panel 8 1 Power The Handyscope HS5 is powered through the USB If the USB cannot supply enough power it is possible to power the instru ment externally The Handyscope HS5 has two external power inputs located at the rear of the instrument the dedicated power connector and a pin of the 9 pin D sub extension connector The specifications of the dedicated power connector are Pin Dimension Description Center pin 1 3 mm positive Outside bushing 03 5 mm ground Figure 8 2 Power connector To power the instrument through the extension connector the power has to be applied to pin 7 of the extension connector Pin 6 can be used as ground The following minimum and maximum voltages apply to the power inputs Minimum 4 5 Vpc 2 A max Maximum 15 Vpc 1 A max Table 8 1 Maximum voltages Rear panel 27 Note that the externally applied voltage should be higher than the USB voltage to relieve the USB port 8 1 1 Power adapter The Handyscope HS5 comes with an external power adapter that can be connected to any
26. put chan nel These are 1x 10x selectable passive probes This means that the input signal is passed through directly or 10 times attenuated When using an oscilloscope probe in 1 1 the setting the bandwidth of the probe is only 6 MHz The full bandwidth of the probe is only obtained in the 1 10 setting The x10 attenuation is achieved by means of an attenuation network This attenuation network has to be adjusted to the oscil loscope input circuitry to guarantee frequency independency This Introduction is called the low frequency compensation Each time a probe is used on an other channel or an other oscilloscope the probe must be adjusted Therefore the probe is equiped with a setscrew with which the parallel capacity of the attenuation network can be altered To adjust the probe switch the probe to the x10 and attach the probe to a 1 kHz square wave signal Then adjust the probe for a square front corner on the square wave displayed See also the following illustrations Figure 3 6 correct Figure 3 7 under compensated Figure 3 8 over compensated I2 Chapter 3 O 4 1 4 2 E Driver installation Before connecting the Handyscope HS5 to the computer the drivers need to be installed Introduction To operate a Handyscope HS5 a driver is required to interface between the measurement software and the instrument This driver takes care of the low level communication between the computer and the
27. t temperature 20 C to 70 C Relative humidity 5 to 95 non condensing Certifications and Compliances CE mark compliance Yes RoHS Yes EN 55011 2009 A1 2010 Yes EN 55022 2006 A1 2007 Yes IEC 61000 6 1 EN 61000 6 1 2007 Yes IEC 61000 6 1 EN 61000 6 1 2007 Yes Canada ICES 001 2004 Yes Australia New Zealand AS NZS Yes Package contents Instrument Handyscope HS5 Probes 2x 1 1 1 10 switchable HP 9250 Accessories External power adapter USB power cable Software Windows 2000 XP Vista 7 8 Drivers Windows 2000 XP Vista 7 8 Manual Instrument manual and software user s manual Specifications Chapter 9 If you have any suggestions and or remarks regarding this manual please contact TiePie engineering TiePie engineering P O Box 290 A Koperslagersstraaat 37 8600 AG SNEEK 8601 WL SNEEK The Netherlands The Netherlands Tel 31 515 415 416 Fax 31 515 418 819 E mail support tiepie nl Site www tiepie nl
28. therlands engineering EC Declaration of conformity We declare on our own responsibility that the product Handyscope HS5 530 Handyscope HS5 530XM Handyscope HS5 220 Handyscope HS5 220XM Handyscope HS5 110 Handyscope HS5 110XM Handyscope HS5 055 Handyscope HS5 055XM for which this declaration is valid is in compliance with EN 55011 2009 A1 2010 IEC 61000 6 1 EN 61000 6 1 2007 EN 55022 2006 A1 2007 IEC 61000 6 1 EN 61000 6 1 2007 according the conditions of the EMC standard 2004 108 EC and also with Canada ICES 001 2004 Australia New Zealand AS NZS Sneek 29 5 2012 ir A P W M Poelsma Declaration of conformity Environmental considerations This section provides information about the environmental impact of the Handyscope HS5 Handyscope HS5 end of life handling Production of the Handyscope HS5 required the extraction and use of natural resources The equipment may contain substances that could be harmful to the environment or human health if improperly handled at the Handyscope HS5 s end of life In order to avoid release of such substances into the environment and to reduce the use of natural resources recycle the Handyscope HS5 in an appropriate system that will ensure that most of the materials are reused or recycled appropriately The symbol shown below indicates that the Handyscope HS5 complies with the European Union s requirements according to Di rective 2002 96 EC on waste
29. to each other using special cables The Handyscope HS5 has an advanced clock distribution sys tem making it very easy to connect multiple instruments to each other to create a large multi channel instrument that uses a shared sampling clock and shared trigger signals 1 2 AUX I O AUX I O Figure 6 1 Auxilary I O connectors Connecting is done by daisy chaining the auxiliary I O con nectors of the instruments prior to starting the software using a special coupling cable order number TP C50H The software will detect how the instruments are connected to each other and will automatically terminate the connection bus The software will combine the connected instruments to one large instrument The combined instruments will now sample using the same clock with a deviation of 0 ppm Figure 6 2 3 Handyscope HS5 s combined A six channel instrument is easily made by connecting three Handyscope HS5 s to each other Combining instruments Chapter 6 Ged 7 2 7 3 Front panel Figure 7 1 Front panel CH1 and CH2 input connectors The CH1 and CH2 BNC connectors are the main inputs of the ac quisition system The outside of the BNC connectors is connected to the ground of the Handyscope HS5 Connecting the outside of the BNC connector to a potential other than ground will result in a short circuit that may damage the device under test the Handy scope HS5 and the computer AWG output connector The AWG BNC connect

Download Pdf Manuals

image

Related Search

Related Contents

FL-60 manual.cdr  SHOUT® ULTRA GEL CONCENTRE  HERMA File labels A4 192x38 mm white paper matt opaque 175 pcs.  InLine 66904    Plan de cours non officiel - Faculté de droit de l`Université Laval  3G-SDI Splitter, SDIS  

Copyright © All rights reserved.
Failed to retrieve file