1. Polysonics SX30 Portable Doppler Flow meter User Guide P N 1 0563 007 Revision J 2000 Thermo Fisher Scientific Inc All rights reserved Sil Glyde is a registered trademark of American Grease Stick Company All other trademarks are the property of Thermo Fisher Scientific Inc and its subsidiaries Thermo Fisher Scientific Thermo Fisher makes every effort to ensure the accuracy and completeness of this manual However we cannot be responsible for errors omissions or any loss of data as the result of errors or omissions Thermo Fisher reserves the right to make changes to the manual or improvements to the product at any time without notice The material in this manual is proprietary and cannot be reproduced in any form without expressed written consent from Thermo Fisher This page intentionally left blank Chapter 1 Chapter 2 Chapter 3 Chapter 4 Chapter 5 Thermo Fisher Scientific Contents Product O Very OW ccsezes coca seeks ects cas vcn casctesbenedecas cts cecsctusx pects cseavot dEn pas nianna ia 1 1 Introducido Mine dd do reads dao 1 1 Ordering Information saisie ira on dicos er SU te 1 1 o A A ae AE te era uae 1 2 Performancen A weaned nea ane aes 1 2 Pl 1 2 Funcional anita 1 3 il Sateen atta ea ane ee a E 1 4 WA TO lo sieial 1 4 Installation amp WUritng ccsssssssssessesssssesesseesesseseesssseesseseseesseseenesesneateaeeneens 2 1 Installing the Trans da fics catesaceteteasc
2. 67 inch with 3 8 SS screw 10705 0003 Nut driver 3 8 inch 10808 0002 Coupling compound silicone RTV108 2 8 oz 10823 0005 Coupling compound Sil Glyde 4 oz 20192 0001 SS tag Polysonics SX30 User Guide 1 1 Product Overview Specifications Part Number Description 20336 0001 Quick mount clamp assembly set of 2 22366 0002 22490 0001 22686 0002 22688 0002 22689 0002 22702 0001 22703 0001 22705 0002 22262 0006 Specifications Performance Cable adapter dedicated transducer to portable meter Cable adapter set portable transducer to dedicated meter PCB assembly Doppler Transducer cables set 16 ft with BNC connectors PCB assembly motherboard Front panel assembly includes keypad and display Transducer assembly set of 2 15 ft cable length PCB assembly CPU 4 20 mA board Velocity Range 0 2 18 ft sec 0 06 5 5 m sec Accuracy 1 of velocity full scale Fluids Liquids containing particulate or entrained gas bubbles Pipe Size 0 5 200 in 12 5000 mm Physical Transmitter NEMA 6 IP67 environmentally sealed waterproof against accidental immersion splash proof with lid open Transducers Two encapsulated dual frequency sensor heads Encased in stainless steel shrouds with stainless steel straps and quick clamps Cable length 16 ft 5 m Weight 11 lb 4 9 kg with 12 hr battery Dimensions Refer to Figure 1 1 1 2 Polysonics SX30 User Guide Therm
3. Approvals Non Hazardous Area Certification Designed to meet CE and NRTL C CSA for non hazardous areas 1 4 Polysonics SX30 User Guide Thermo Fisher Scientific Chapter 2 Installation amp Wiring Select an installation site Installing the Transducers Keep the following questions in mind when choosing a proper installation site e Is the chosen section of pipe always full of liquid e Are there at least 5 pipe diameters upstream and 3 pipe diameters downstream from any directional changes pipe joints or narrowing widening of the pipe Flow Figure 2 1 Conditions at locations A B C D and E can interfere with the transmission of the ultrasonic wave and yield inaccurate or unreliable flow readings e A Pipe may not be full e B Down flow e C Too close to elbow e D Air collects at the top of the horizontal pipe e E Sediment collects at the bottom of the horizontal pipe Note Selecting the proper installation location is essential to flow measurement reliability Each application is unique and may require a variation of installation locations A pmi Thermo Fisher Scientific Polysonics SX30 User Guide 2 1 Installation amp Wiring Installing the Transducers 2 Select a transducer orientation Refer to Figure 2 2 If your application allows install the transducers at the 3 and 9 o clock locations Ensure the transducers are parallel to each other along the pipe circumference Transdu
4. Canada Thermo Fisher Scientific 14 Gormley Industrial Avenue Gormley Ontario LOH 1G0 Phone 905 888 8808 Fax 905 888 8828 On the Web www thermofisher com If it becomes necessary to return an instrument il Contact Thermo Fisher for a Return Material Authorization RMA number You must have the RMA number to return an instrument The receiving dock will not accept shipments without the RMA number Ensure the instrument is well packed in its original shipping box if available Include a letter fully explaining the symptoms of the failure as well as detail describing the application where the unit was being operated Also include a contact name phone number and purchase order authorizing repairs Polysonics SX30 User Guide 5 7 Troubleshooting amp Maintenance Warranty Warranty 5 8 Polysonics SX30 User Guide 4 Write the RMA number on the outside of the shipping box 5 Send the unit freight paid to the address above Thermo Scientific products are warranted to be free from defects in material and workmanship at the time of shipment and for one year thereafter Any claimed defects in Thermo Scientific products must be reported within the warranty period Thermo Fisher shall have the right to inspect such products at Buyer s plant or to require Buyer to return such products to Thermo Fisher plant In the event Thermo Fisher requests return of its products Buyer shall ship with transportation char
5. EESK B 2 Polysonics SX30 User Guide Thermo Fisher Scientific Thermo Fisher Scientific 81 Wyman Street P O Box 9046 Waltham Massachusetts 02454 9046 United States www thermofisher com
6. Reset the data by changing the Max field selection from Set to Clear 4 2 Polysonics SX30 User Guide Thermo Fisher Scientific General Troubleshooting Thermo Fisher Scientific Chapter 5 Troubleshooting amp Maintenance The instrument has 3 modes of operation Setup e Learn e Operating Use the following table in conjunction with the learn mode and operating mode warning tables Table 5 1 Action code Action Description A 1 Turn meter off 2 Apply more compound 3 Tighten the pipe strap 4 Restart the meter to see if warning clears B 1 Reposition transducers 2 See if signal strength increases 3 See if warning clears 4 Restart the meter C 1 Relocate transducers 2 See if signal strength increases 3 See if warning clears 4 Restart meter D 1 Turn meter off and remove the transducers 2 Grind the pipe surface if it is rough or coated 3 Remount the transducers 4 Restart the meter See if warning clears Polysonics SX30 User Guide 5 1 Troubleshooting amp Maintenance General Troubleshooting Learn Mode Warnings The following table describes the learn mode warnings and appropriate corrective actions Complete the corrective actions in the order presented Table 5 2 Learn mode warnings amp corrective actions Warning Corrective Action Invalid Signal Can t Learn 1 Refer to Table 5 1 and complete actions A D checking the meter a
7. Scientific Operation amp Configuration HydraScan 8 At this screen a Scroll to the required totalizer unit US gallons Imperial gallons million gallons cubic feet cubic meters liters oil barrels liquid barrels and press Enter b Scroll through the available scale factors x 1 10 100 1000 10000 and press Enter to select c Determine whether to reset the totalizer and scroll to the appropriate selection Press Enter 9 Set the range at this screen a Enter the flow rate that equals the 4 mA minimum reading and press Enter b Enter the flow rate that equals the 20 mA maximum reading and press Enter c Select Next to move to the next setup screen 10 Set the time and date at this screen and press Next to continue 11 Scroll through the available wall material options carbon steel cast iron concrete liner stainless steel copper plastic and press Enter to select Press Next to complete the configuration process Hyd raScan You can also use the HydraScan software to configure the instrument Refer to the HydraScan user guide P N 1 0563 005 Thermo Fisher Scientific Polysonics SX30 User Guide 3 5 This page intentionally left blank Thermo Fisher Scientific Setup Reset Config FFT Logset mm Chapter 4 Flow Menu Items You can access the Setup menu from the main Flow screen The Setup menu displays the instrument version serial number and remaining battery power
8. lana 2 1 IA a Oi 2 4 DC UR A A 2 5 Transducers oia E O A ARN 2 5 4 20 mA Current L peri o eiii seis did 2 5 Operation amp Configuration 0 0 cesssssssssessesssssssssseeeeseseeseeseeesteseeseeneeneens 3 1 A E 3 1 Screen Contarini es r AKEE aeni NAKE Ease ANRE 3 1 CONO ena iea ai e areari 3 2 Auto Modest ree nenea a R a a aR 3 2 Mangal Moderna a 3 3 O OS 3 5 Flow Menu MOIS iia leasassdsininsivaiusasbanteinnaveiual 4 1 O NS 4 1 Reset ett ate eal sts A A Raa atl ruta eae aca 4 1 A del oli eta 4 1 A easy antes ae diab sec cain ann cates vised send buadeae talons aN asioteasceadese 4 E cars eae jaSintat vant sec bos od dnataonsndaract E uol awed sinies 4 1 Accessing Saving amp Loading Log Piles cui cda 4 2 Tienden a A as 4 2 I T ele oa tess eae E EA E EE A E Shade ct de Madre 4 2 Makaa A a 4 2 Troubleshooting amp Maintenance ooococccccccnconnonnmerererererreraars 5 1 General Trouble honra eas 5 1 Polysonics SX30 User Guide v Contents vi Appendix A Appendix B Polysonics SX30 User Guide Lear ModeW at petita ais 5 2 Operating Mode Warnings coriuaispi iii do 5 2 Checking the FET to edi ina ia a ih 5 3 Signal Quality amp Strength ssseseeesssesesesesesssrsisisisisisisesesiseseseseseseses 5 5 IAEA CO it e a a a aaae aa 5 6 A an a A RORA NR 5 6 o AA II II A 5 6 Contact tor OM la 5 7 Is A O 5 8 Obtaining Pipe 1D 0 as A 1 Toxic amp Hazardous Substances Tables ooococonnicacinaconnnorosesnoneren
9. 0 17 80 16 02 18 16 16 20 18 16 16 00 18 54 16 18 18 54 16 00 18 19 50 18 22 19 50 18 00 19 92 18 18 19 92 18 00 20 34 18 20 20 34 18 00 20 78 18 22 20 78 18 00 20 21 60 20 26 21 60 20 00 22 06 20 22 22 06 20 00 22 54 20 24 22 54 20 00 23 02 20 24 23 02 20 00 24 25 80 24 28 25 80 24 02 26 32 24 22 26 32 24 00 26 90 24 28 26 90 24 00 27 76 24 26 27 76 24 00 30 31 74 28 98 32 00 29 94 32 40 30 00 32 74 30 00 33 10 30 00 33 46 30 00 36 37 96 35 98 38 30 36 00 38 70 35 98 39 16 36 00 39 60 36 00 40 04 36 00 42 44 20 42 00 44 50 41 94 45 10 42 02 45 58 42 02 48 50 50 47 98 50 80 47 96 51 40 47 98 51 98 48 00 54 56 66 53 96 57 10 57 10 57 80 54 00 58 40 53 94 60 62 80 60 02 63 40 60 06 64 20 60 20 64 82 60 06 72 75 34 72 10 76 00 72 10 76 88 72 10 84 87 54 84 10 88 54 84 10 All measurements in inches A 2 Polysonics SX30 User Guide Note For pipes with cement linings reduce the pipe ID by twice the lining thickness Standard and double cement lining thicknesses are listed in the next table A Thermo Fisher Scientific Obtaining Pipe ID Table A 3 Ductile iron pipe standard classes Nominal Inside Diameter Cement Lining Pipe Size OD Class 50 Class51 Class52 Class53 Class54 Class55 Class56 Std Double T
10. ESS FOR PURPOSE Thermo Fisher Scientific Appendix A Obtaining Pipe ID If you do not know the pipe ID try to get pipe OD using one of the following methods e Read the pipe OD and schedule on the pipe e Read the OD and schedule from system drawings e Use a tape measure to measure the pipe circumference and divide the measured circumference by 3 14 e Use Thermo Fisher s Ultrasonic pipe Thickness Gauge P N 0704 0187 and perform the following calculation Pipe ID Pipe OD 2 x Pipe Thickness Table A 1 Steel stainless steel amp PVC pipe standard schedules Nom OD Sch Sch Sch Sch Sch Sch Sch Sch Sch Sch Sch Std XSTG Pipe 5 10 20 30 40 60 80 100 120 140 160 Wall Size light wall 1 1 315 1 185 1 097 1 049 0 957 1 25 1 660 1 530 1 442 1 380 1 278 1 5 1 900 1 770 1 682 1 610 1 500 2 2 375 2 245 2 157 2 067 1 939 2 5 2 8 5 2 709 2 635 2 469 2 323 3 2 900 3 5 3 364 4 3 826 5 4 813 6 5 761 8 7 625 10 9 750 12 11 750 14 13 000 16 15 15 15 15 15 15 000 18 18 00 17 670 17 500 17 376 17 124 16 876 16 500 124 15 688 15 255 14 876 14 438 17 250 17 000 20 20 00 19 634 19 500 19 250 19 000 18 812 18 376 17 938 17 438 17 000 16 500 16 062 19 250 19 000 Thermo Fisher Scientific Polysonics SX30 User Guide A 1 Obtaining Pipe ID Nom OD Sch Sc
11. It also contains 2 submenus Reset and Config Select this item to reset all configuration parameters to defaults Refer to Chapter 3 for details on setting configuration parameters Each parameter is discussed in the section Manual Mode Refer to Checking the FFT in Chapter 5 To set up the data logging function using the instrument 1 Select LogSet from the Flow screen 2 The screen displays the number log points remaining Select the log rate intervals of 0 5 1 5 15 or 60 minutes and press Enter 3 Turn the logger on 4 Select Flow to return to the Flow screen Note You cannot select the file in which you want to store a specific log set The meter saves the data in sequenced files 0 to 9 The meter advances to the next file each time the logger is turned on or when a file logs more than 9000 points A Polysonics SX30 User Guide 4 1 Flow Menu Items Trend Accessing Saving You can use HydraScan to download save load and erase log files Refer amp Loading Log Files the HydraScan user manual 1 0563 005 Trend Trend runs ata real time rate of action To view the flow trend select Trend from the Flow menu Rate From the Trend screen select Rate Scroll through the available rate options 0 1 0 5 1 5 15 60 minutes reading Press Enter Max The meter records the maximum flow date and time of the maximum flow since the last reset You can view this information from the Max screen
12. able 6 3 in Chapter 6 if the meter displays any of the following warnings e Warning e Alarm e Failed No Flow Pipe ID screen a Enter the pipe ID and press Enter b Use the scroll key to select the measurement units inches mm Press Enter to select the unit c Select Next to move to the next screen At the Flow Units screen a Use the scroll key to select the flow units US gallons million gallons cubic meters oil barrels Imperial gallons cubic feet liters liquid barrels and press Enter b Use the scroll key to select the time base for flow measurement seconds minutes hours days and press Enter c Select Next to continue Enter the Maximum Flow Rate Select Next At this screen a Scroll through the options available for the Velocity Range 2 4 8 16 32 ft s and press Enter to select Scroll to the next parameter b Enter a damping coefficient 0 99 1 unit 15 s Press Enter and scroll to the next parameter The damping coefficient suppresses short term fluctuations in the indicated flow rate Increasing the coefficient increases the response time to changes Keep damping at a minimum unless the flow rate fluctuates wildly In this case increase damping just enough to reduce the fluctuation to an acceptable degree Polysonics SX30 User Guide 3 3 Operation amp Configuration Configuration to smooth small fluctuations without diminishing the meter s response Note The Smar
13. cer orientations on Transducer orientations on horizontal pipes vertical pipes Good 3 and 4o clock Good 2and3o clock 9 and 8 o clock 10 and 9 o clock Better 2 and 4o clock Better 2 and 4 o clock 10 and 8 o clock 10 and 8 o clock Best 3and9o clock Best 3 and 9 o clock Another directly opposite position Figure 2 2 3 Attach the transducers to the pipe strap a Use a screwdriver to loosen the mounting screw and slide the lock back on each transducer Figure 2 3 2 2 Polysonics SX30 User Guide Thermo Fisher Scientific Installation amp Wiring Installing the Transducers b Place the pipe strap across the mounting and slide the lock back into place Figure 2 4 c Tighten the screws Figure 2 5 4 Mount the transducers ensuring the transducers face the same direction as shown in Figure 2 6 a Wrap the pipe strap around the pipe and slide the end of the pipe strap through the tension nut assembly b Pull the strap tightly and clamp the tension nut down to secure the strap to the pipe c Loosen the mounting screw of a transducer and move the transducer to the 3 o clock position on the pipe Tighten the mounting screw to secure the transducer d Loosen the mounting screw of the second transducer and move the transducer to the 9 o clock position Tighten the mounting screw to secure the transducer e Remove the strap from the pipe and apply sonic coupling compound to each t
14. ctions B D checking the meter Signal Status after performing each action to see if the problem was corrected 2 Check the FFT 3 Contact Thermo Fisher IS Invalid Signal 1 Check the FFT 2 Contact Thermo Fisher 1 From the Flow screen select FFT 2 Refer to Figures 5 1 through 5 6 to determine the FFT condition 3 Refer to the table below to determine the corrective action 4 Return to the Flow screen Table 5 5 FFT Shape Possible Cause Corrective Action Ideal Doppler N A 1 Restart the meter Broadband Poor pipe coupling or 1 Refer to Table 5 1 and complete action noise multiple noise generators codes A D checking the meter after performing each action to see if the problem was corrected 2 Relocate transducers 3 Turn off source of noise Steady noise spike Electromagnetic noise 1 Turn off source of noise 2 Relocate transducers Fluctuating Constantly changing flow 1 Increase damping reference Doppler or noise Configuration in Chapter 3 2 Relocate transducers No Doppler Ultrasound signal not 1 Refer to Table 5 1 and complete action reflection transmitting codes A D checking the meter after lor ll performing each action to see if the problem was corrected Polysonics SX30 User Guide 5 3 Troubleshooting amp Maintenance General Troubleshooting 5 4 Polysonics SX30 User Guide B Figure 5 1 Ideal Doppler shape Figure 5 2 Broadban
15. d noise Figure 5 3 Steady noise spike with Doppler Figure 5 4 Fluctuating Doppler profile Thermo Fisher Scientific Signal Quality amp Thermo Fisher Scientific Strength Troubleshooting amp Maintenance Signal Quality amp Strength H Figure 5 5 No Doppler reflection B Figure 5 6 No Doppler reflection Il Signal quality and signal strength are indicators of how well suited an application is for a Doppler flowmeter To determine the quality of your application go to the Flow screen Note the SS and SQ numbers in the bottom right corner Refer to the following figure to determine the quality of the application Also refer to Table 5 1 for actions you can take to attempt to improve signal quality and strength Excellent Flow readings consistent and accurate Excellent i l Good Flow readings fairly consistent and accurate Adequate Minimum quality flow 35 readings inconsistent or inaccurate Signal Quality SQ Adequate 25 Failure Meter not reading flow 0 25 25 10 Signal Strength SS Figure 5 7 Polysonics SX30 User Guide 5 5 Troubleshooting amp Maintenance Maintenance Maintenance General Upgrades 5 6 Polysonics SX30 User Guide The instrument is easy to maintain The transducers and flowmeter are factory service only components and maintenance free The following table describes the instrument components appropriate maintenance and a recommended schedu
16. eses B 1 Thermo Fisher Scientific Introduction Thermo Fisher Scientific Ordering Information Chapter 1 Product Overview Thermo Fisher s Polysonics SX30 portable Doppler flowmeter generates two independent ultrasonic signals at different frequencies By correlating these frequencies the instrument automatically identifies and eliminates noise errors from sources such as variable frequency drives In addition operation of the instrument is enhanced by an Expert System that allows the flowmeter to automatically learn the application parameters As a result the flowmeter can be easily commissioned in a fraction of the time required to configure competitive ultrasonic flowmeters Refer to the following list of part numbers to order spare parts for your instrument Table 1 1 Part Number Description 22767 0001 Poly CD specification sheet meter installation video CD 1 0563 007 User guide 1 0561 005 HydraScan software manual 01008 0006 RS232 cable standard serial 6 ft female to male DB9 10228 0002 Power adapter 90 264 Vac input 15 Vdc output 10312 0001 Battery 12 V 7 0 A hr 24 hour continuous operation 10312 0002 Battery 12 V 7 0 A hr 12 hour continuous operation 10435 0007 Cable coupling BNC to extend 22688 0001 2 required 10520 0005 Thermo Fisher accessory bag with shoulder strap 10605 0001 Pipe strap SS 32 inch with 3 8 SS screw 10605 0003 Pipe strap SS
17. fter performing each action to see if the problem was corrected 2 Contact Thermo Fisher No Flow Can t Learn 1 Check for flowing fluid 2 Make sure the pipe is full 3 Check connections 4 Restart meter See if warning clears 5 Contact Thermo Fisher Warning Low S Strength 1 Turn meter off 2 Refer to Table 5 1 and complete actions A D checking the meter after performing each action to see if the problem was corrected 3 Contact Thermo Fisher Warning Poor S Quality Operating Mode Warnings Table 5 3 Warning 1 Go into operating mode 2 Check the FFT Warnings displayed while the meter is in operating mode contain a level of severity Table 5 3 and a reason for the warning Table 5 4 Level of Severity Warning Flow reading marginal Alarm Flow reading unreliable Failed No Flow 5 2 Polysonics SX30 User Guide Meter will not read flow and will read zero until fault clears Thermo Fisher Scientific Checking the FFT Thermo Fisher Scientific Troubleshooting amp Maintenance General Troubleshooting Table 5 4 Operating mode warnings amp corrective actions Warning Corrective Action PO Poor 1 Check the FFT Signal Quality 2 Refer to Table 5 1 and complete actions B D checking the meter after performing each action to see if the problem was corrected 3 Contact Thermo Fisher FS Failed 1 Refer to Table 5 1 and complete a
18. ges paid by the Buyer to Thermo Fisher plant Shipment of repaired or replacement goods from Thermo Fisher plant shall be F O B Thermo Fisher plant A quotation of proposed work will be sent to the customer Thermo Fisher shall be liable only to replace or repair at its option free of charge products which are found by Thermo Fisher to be defective in material or workmanship and which are reported to Thermo Fisher within the warranty period as provided above This right to replacement shall be Buyer s exclusive remedy against Thermo Fisher Thermo Fisher shall not be liable for labor charges or other losses or damages of any kind or description including but not limited to incidental special or consequential damages caused by defective products This warranty shall be void if recommendations provided by Thermo Fisher or its Sales Representatives are not followed concerning methods of operation usage and storage or exposure to harsh conditions Materials and or products furnished to Thermo Fisher by other suppliers shall carry no warranty except such suppliers warranties as to materials and workmanship Thermo Fisher disclaims all warranties expressed or implied with respect to such products EXCEPT AS OTHERWISE AGREED TO IN WRITING BY Thermo Fisher THE WARRANTIES GIVEN ABOVE ARE IN LIEU OF ALL OTHER WARRANTIES EXPRESSED OR IMPLIED AND Thermo Fisher HEREBY DISCLAIMS ALL OTHER WARRANTIES INCLUDING THOSE OF MERCHANTABILITY AND FITN
19. h Sch Sch Sch Sch Sch Sch Sch Sch Sch Std XSTG Pipe 5 10 20 30 40 60 80 100 120 140 160 Wall Size light wall 24 23 500 23 250 22 876 22 624 22 062 21 562 20 938 20 376 19 876 19 312 23 250 23 000 30 29 376 29 00 28 750 28 500 29 250 29 000 36 35 376 35 00 34 750 34 500 31 876 31 312 32 250 36 000 42 42 000 41 000 40 750 40 500 41 250 41 000 48 48 000 47 250 47 250 47 000 1 All measurements in inches Table A 2 Cast iron pipe standard classes Nom Class A Class B Class C Class D Class E Class F Class G Class H Pipe Size 0D ID OD ID OD ID OD ID OD ID OD ID OD ID OD ID 3 3 80 3 02 396 3 12 3 96 3 06 3 96 3 00 4 480 3 96 500 410 5 00 4 04 5 00 3 96 6 6 90 6 02 7 10 6 14 7 10 6 08 7 10 6 00 7 22 606 7 22 6 00 7 38 6 08 7 38 6 00 8 905 8 13 9 05 803 930 818 930 8 10 942 810 942 810 9 60 8 10 9 60 8 00 10 11 10 10 10 11 10 9 96 11 40 10 16 11 40 10 04 11 60 10 12 11 60 10 00 11 84 10 12 11 84 10 00 12 13 20 12 12 13 20 11 96 13 50 12 14 13 50 12 00 13 78 12 14 13 78 12 00 1408 12 14 14 08 12 00 14 15 30 14 16 15 30 13 98 15 65 14 17 15 65 14 01 15 98 14 18 15 98 14 00 16 32 14 18 16 32 14 00 16 17 40 16 20 17 40 16 00 17 80 16 2
20. hickness Thickness 0 250 0 375 0 500 All measurements in inches Note For pipes with cement linings reduce the pipe ID by twice the lining thickness listed above A Thermo Fisher Scientific Polysonics SX30 User Guide A 3 This page intentionally left blank Appendix B Toxic amp Hazardous Substances Tables English and Chinese versions The Toxic and Hazardous Substances tables can be found on the following page Thermo Fisher Scientific Polysonics SX30 User Guide B 1 Toxic amp Hazardous Substances Tables Toxic amp Hazardous Substances Table SX30 For Chinese Regulation Administrative Measure on the Control of Pollution Caused by Electronic Information Products Names and Content of Toxic and Hazardous Substances or Elements Parts Name Po He os cer Pas re mos FCI FCI e IEC zomon fe e mersa fe e forsee fe fe Doppler Board o Indicates that this toxic or hazardous substance contained in all of the homogeneous materials for this part is below the limit requirement in SJ T11363 2006 X Indicates that this toxic or hazardous substance contained in at least one of the homogeneous materials used for this part is above the limit requirement in SJ T11363 2006 INIER ERKA ZIRAN Cr6 PBB PBDE EMOS MARIE RE Or KRZA EA AAA OE ES IESIT 11363 2006 NEHLE HRE ERLA F X RAIA RA AED AAA AED BAS PL SI T 11363 2006 ba VERGE WR
21. in the configuration process by selecting Auto or Manual operating mode Once you have completed the configuration process you can access these screens from the Flow screen by selecting Setup gt Config Auto Mode 1 Pipe ID screen a Enter the pipe ID and press Enter b Use the scroll key to select the measurement units inches mm Press Enter to select the unit c Select Next to move to the next screen 2 Flow Units screen a Use the scroll key to select the flow units US gallons million gallons cubic meters oil barrels Imperial gallons cubic feet liters liquid barrels and press Enter b Use the scroll key to select the time base for flow measurement seconds minutes hours days and press Enter c Select Next to enter learn mode 3 The meter displays status of the learning process Refer to Learn Mode Warnings Table 6 2 in Chapter 6 if you receive one of the following messages e Invalid Signal Can t Learn e No Flow Can t Learn e Warning Low S Strength e Warning Poor S Strength 3 2 Polysonics SX30 User Guide Thermo Fisher Scientific Manual Mode Thermo Fisher Scientific Operation amp Configuration Configuration Once the meter completes the learning process the Flow screen appears a Ifthe flow reading is not accurate and you can verify the flow from a calibrated flowmeter or other certified source perform a calibration b Refer to Operating Mode Warnings T
22. le Table 5 6 Suggested maintenance Component Maintenance How Often Transducers None Factory service item N A Coupling compound Add more compound Annually or whenever 1 Compound diminishes 2 Repositioning transducers 3 Relocating meter Flowmeter None Factory service item N A Pipe strap Retighten Annually Cable connections Ensure connections are secure Remove any buildup within the connections Complete as part of your facility s routine maintenance The most current software is installed in the instrument prior to shipment You can use the RS232 port and a remote terminal to upgrade the software as newer versions become available Contact Thermo Fisher to obtain the most recent versions of the following software e WinLoader used to upgrade meter software e Meter software e HydraScan data logging software Thermo Fisher Scientific Thermo Fisher Scientific Contact Information Troubleshooting amp Maintenance Contact Information The local representative is the first contact for support and is well equipped to answer questions and provide application assistance Your representative also has access to product information and current software revisions You can also contact Thermo Fisher directly In the United States Thermo Fisher Scientific 1410 Gillingham Lane Sugar Land TX 77478 Phone 713 272 0404 Fax 713 272 2272 In
23. o Fisher Scientific Product Overview Specifications 12 79 in 324 90 mm 10 86 in 275 87 mm 12 90 in 327 69 mm Figure 1 1 Outputs 4 20 mA into 750 ohms 12 bit 5 kV opto isolated loop powered Power Supply Built in lead acid gel battery 90 264 Vac 50 60 Hz Functional Standard 12 hours continuous operation Optional 24 hours continuous operation Charger 15 V at 1 A typical input supply voltage 2 A maximum Mains supply voltage fluctuations to charger not to exceed 10 of nominal supply voltage Auxiliary Supply 12 15 Vdc auxiliary power port 15 V at 2 A typical supply voltage Must be an SELV source Auxiliary terminals are A supplied from the SELV source as per IEC1010 1 Annex H Keypad 21 key with tactile action Display Backlit 240 x 60 dot high resolution graphics display Data Logger 90000 points programmable in intervals of 30 s 1 5 15 30 and 60 min Programming Via HydraScan software or integral keypad Thermo Fisher Scientific Polysonics SX30 User Guide 1 3 Product Overview Specifications Serial Interface RS232 with DB9 connector Temperature Range Transducers 40 F to 250 F 40 C to 121 C Electronics With CSA approval 41 F to 104 F 5 C to 40 C Without CSA approval 4 F to 140 F 20 C to 60 C Environmental Pollution degree 2 Installation category I for meter II for battery charger
24. ransducer surface f Reposition the strap on the pipe placing the transducers in the proper positions Thermo Fisher Scientific Polysonics SX30 User Guide 2 3 Installation amp Wiring Wiring g Tighten the tension nut assembly with a 3 8 inch hex nut driver to secure the strap Note Increasing the pipe strap tension improves signal transmission A Flow Figure 2 6 Wiring Wiring is accomplished using the connections on the side of the instrument enclosure Refer to Figure 2 7 and the following sections AUX POWER IN 4 20 mAIN 4 20 mA OUT TRANSDUCERS Figure 2 7 2 4 Polysonics SX30 User Guide Thermo Fisher Scientific DC Auxiliary Power Ay Transducers 4 20 mA Current Loop Thermo Fisher Scientific Installation amp Wiring Wiring Locate the auxiliary power connector on the side of the instrument and connect a minimum 10 5 V power supply that supplies 0 25 A Note The battery will not charge while the unit is operating from the auxiliary power source A Caution Avoid damaging the instrument Do not use power supplies greater than 18 V A Refer to Figure 2 7 to locate the transducer cable connections on the side of the instrument Locate the red in and black out 4 20 mA connectors on the side of the instrument Connect as shown in Figure 2 8 Current Loop Terminals Black Red Remote device or current meter Figure 2 8 Polysonics SX30 U
25. ser Guide 2 5 This page intentionally left blank The Interface Screen Contrast Thermo Fisher Scientific Chapter 3 Operation amp Configuration LCD Menu selection keys DON Numeric keys Figure 3 1 Keypad features Following is a description of the components called out in Figure 3 1 e LCD Displays menu items e Scroll Press to scroll to the next parameter displayed on the LCD e Enter Press once you have made an entry e Backspace Deletes the last value entered e Numeric keys Use to enter values when configuring the meter e Menu selection keys Press to select the corresponding menu item displayed on the LCD 1 Press the key 2 When the Adjust Contrast screen appears press the key again and the software scrolls through the levels of contrast Press the key to stop the scrolling 3 Use the scroll and backspace keys to make fine adjustments to the contrast Polysonics SX30 User Guide 3 1 Operation amp Configuration Configuration Config uration Note the following before configuring the meter e You must know the pipe internal diameter ID to ensure reliable flow measurements If you do not know the pipe ID go to Appendix A e There must be flow in the pipe when you start the meter There are two modes of operation Consider using auto mode if your application is a constant process The instrument default is the auto mode Apply power to the meter and beg
26. t Filter used by the meter allows the damping function to large flow changes even when configured with a high damping coefficient A c Select regular or high Transmit Power and press Enter In applications where the signal strength is low the meter automatically selects high transmit power You can manually select regular or high power in applications where the signal strength is marginal or where you feel increased or decreased power output may improve signal quality 5 At the next screen a Enter the Low Flow Cutoff Press Enter and scroll to the next parameter When a zero flow condition occurs internal sloshing and other fluid movement can prevent the flowmeter from reading total zero This can result in totalizer errors Minimize these errors by setting a low flow cutoff that drives the flowmeter to zero for flow rates at or below specified value b Enter the Signal Strength Cutoff and press Enter c Select Next to continue 6 At this screen a Enter the Signal Quality Cutoff and press Enter b Enter the SNR signal to noise ratio and press Enter c Select Next 7 This screen displays flow and velocity readings You can also enter a site calibration factor a Determine the flow using another source and calculate the site calibration adjustment by dividing the actual flow by the meter flow b Enter the site calibration adjustment c Select Next 3 4 Polysonics SX30 User Guide Thermo Fisher